Abstract---Long-term dose reproducibility and tool to tool dose neutralization rate of 42% since 1.73 * 0.58 = 1.00; a surprising
matching in the Axcelis GSD end-station is critically dependent condition for those unfamiliar with Eaton/Axcelis end stations.
on process chamber pressure measurement and Pressure [Above 3.0E-4 torr pressure, Axcelis recommends alternative
Compensation factor selection. Pressure Compensation factor hardware for maintaining dose accuracy and repeatability;
(PCOMP) determination is well established. Pressure please refer to their Threshold Activated Dose Control Option.]
measurement in the GSD end-station depends on accurate, Also, for a high dose implant, the net time that the process is
repeatable gauge capability: incorrect pressure measurements exposed to such high pressures is typically 10% or less of total
directly lead to dose errors. For example, the dose equation using process time.
PCOMP tells us that for a modest PCOMP value of 30%, a
chamber pressure measurement error of 2E-5 torr can result in a Since pressure compensation can have a profound effect on
dose error up to 6% at normal process pressures. The original dose measurement, the effects of ion gauge deviations on
HCIG used for pressure measurement was not capable of process control when ion beam neutralization is present are
meeting the requirements for good dose control since gauge to significant. Traditional hot cathode ion gauges, or HCIGs, are
gauge differences were not controlled and gauge accuracy was known to have poor gauge to gauge repeatability (+/-15%) and
only on the order of 25%. Axcelis introduced the Granville- poor gauge accuracy (+/- 25%) [2]. As with any physical
Phillips 360 Stabil-Ion gauge to improve dose reproducibility device, HCIGs also exhibit aging effects that lead to pressure
through much improved gauge to gauge matching (+/-6%) and measurement errors over time – usually as a drop in measured
more accurate gauge output. This paper discusses the details of
pressure due to surface coatings accumulated during operation,
the care and feeding of the Stabil-Ion gauge system and it’s
impact on process dose and process trends.
especially from photoresist. HCIGs are replaced often, 1-2
times per year or more, depending on usage and process
conditions. This fact complicates the business of maintaining
I. INTRODUCTION accurate dose measurements even further.
The purpose of Pressure Compensation is to ensure dose
accuracy, repeatability and uniformity during process chamber II. PRESSURE COMPENSATION EQUATION
conditions which result in charge neutralization of the ion The Pressure Compensation equation is given as follows:
beam. In the Axcelis GSD end-station, beam current is used for
both dose measurement and as a feedback for linear scan speed Idose = Idisk*exp(Kρ) (1)
control and so beam current is also a primary determinant of Where Idose = compensated ion beam current, Idisk = ion beam
dose uniformity. Ion beam neutralization is induced by high current measured by the disk faraday, ρ = pressure[torr]
pressures in the small-volume GSD process chamber and is
measured at the process chamber ion gauge (IG3) and K =
affected both by flood system gas and photoresist out-gassing.
(1E4)*ln(1 + PCOMP/100). At a pressure of 1E-4 torr, the
By definition, at a pressure of 1.0E-4 torr, the Pressure
equation reduces to, Idose = Idisk*(1 + PCOMP/100) as described
Compensation equation yields a percentage correction factor to
earlier. An alternative formulation which delineates the role of
the measured beam current that is the Pressure Compensation
PCOMP more clearly can be given as:
(PCOMP) value itself [1]. For example, at a process chamber
pressure of 1.0E-4 torr, and a PCOMP factor of 20%, the (1E4*ρ)
Idose = Idisk*(1 + PCOMP/100) (2)
measured beam current will be multiplied by a factor of 1.20
for the purposes of calculating dose. For processes that utilize
high beam currents, i.e. 10 mA or higher, with photoresist Chart 1 illustrates the dependence of the amount of correction
coated wafers, pressures of 3.0E-4 torr and higher are not to Idisk for a given PCOMP value vs. pressure. A PCOMP value
unusual. In these instances, the beam current correction factor less than zero implies a condition of charge stripping in the
grows very large. For example, for a PCOMP value of 20% and process chamber which occurs for some high energy processes.
a pressure of 3.0E-4 torr, the beam current correction factor is If there is an error in the readback from the process chamber
1.73. A correction factor of 1.73 implies an ion beam ion gauge, IG3, it will affect dose measurement during the
entire implant. Average process pressure (largely determined case of an in-use gauge. The data is then used to populate an
by electron shower gas flow), beam current, quantity of SPC chart using data from a particular MFC setpoint. The
photoresist and optimum PCOMP value all participate in dose chosen data point represents average chamber pressure for
measurement. Large PCOMP values and/or high average most proceses. Currently, control limits are set at about +/-20%
pressures coupled with IG3 readback errors represent worst- of nominal. Though, in practice, we observe mainly that
case conditions. pressure response decreases over time and so the lower control
limit provides a signal for when to change the gauge.
PCOMP Correction v. IG3 24-Jan
4 12
Figure 1. PCOMP Correction Factor vs. IG3 pressure The plots in Figure 2 illustrate the pressure response
tracking data for one GSD system prior to implementing
III. STABIL-ION GAUGE PERFORMANCE correct degas frequency. The decline from nominal
performance to OOC can be very rapid. Of note is that the
Stabil-Ion gauge systems have been installed on gauge response at all Ar flow settings is self-consistent, i.e., if
INNOViON GSD implanters at the Portland, Oregon facility to the pressure reading at 4.0 sccm has declined by 10%, then the
improve process uniformity and dose repeatability between pressure at 9.0 sccm has also declined by 10%. The plot for 12
gauge changes [3]. OEM recommendations for the gauges April is for a replacement gauge demonstrating gauge-to-gauge
include regular, extended degas of the gauge filaments and repeatability.
pressure response tracking of the gauges as a means to evaluate
when a gauge is failing and also to determine whether a 12-Apr
replacement will deliver comparable process results. The 12
IG3 pressure, x1E-5 torr
Jun-02
Jul-02
Aug-02
Jan-02
Feb-02
Mar-02
May-02
Sep-02
(weekly or more often) extended filament degas cycles. The
Stabil-Ion gauges and their successors are considerably more
expensive than standard HCIGs and therefore require measures
Measurement Date to maximize their working life.