CALIBRATION REPORT
This calibration certificate documents the traceability through an unbroken chain of comparison to national or international standards or to other
values of natural physical constants, which realize the units of measurement according to the International System of Units (SI).
Avesco Calibration Laboratory certifies that the instrument specified above was calibrated at Avesco's permanent facility, in Cubao,
Quezon City, Philippines in accordance with OEM procedure using direct measurement method.
The Laboratory's organization and practices are aligned with ISO / IEC 17025: 2017.
Measurement uncertainties at the time of test are given as expanded uncertainties with a coverage factor of k =2 approximating normal distribution
with an estimated 95% level of confidence. Uncertainty values are calculated as per with the method described in the Guide to Expression of
Uncertainty in Measurement (GUM).
This certificate pertains only to the item described above and shall not be reproduced other than in full, without the specific written
approval of Avesco Calibration Laboratory. Calibration certificates without signatures are not valid.
No statement of compliance with specifications is made and none may be implied in this certificate.
Instrument suitability is to be determined by the user for its intended use.
Certain factors may cause the calibration item to drift out of calibration before the recommended interval has expired
The Calibration Data in the tables refer to the date of calibration and should the instrument be modified or damaged in any
way, or develop inconsistent readings, the result may not be valid and the unit will require recalibration.
The data that could be found in this certificate must be interpreted as:
As Found Result - are calibration measurements taken before any adjustments and / or repair is made. No
as found result is reflected if instrument's defect or failure type caused inability to recover data.
As Left Result - reflects calibration measurement data taken after the unit was adjusted and for repaired. No as left result is shown if no adjustments and or
repair is done.
Calibrated by:
Paget of 2 A CLF-510-01-00
CALIBRATION DATA
** Consistent unit of measurement per row, unless otherwise specified.
As found As Left Measurement
UUT Range Standard Correction
Results Results Uncertainty
Insulation Test
20.0 G Ω 500 V 5.00 MΩ 5.3 -0.3 0.1 MΩ
20.0 G Ω 500 V 10.00 MΩ 10.0 0.0 0.1 MΩ
20.0 G Ω 500 V 50.00 MΩ 50.6 -0.6 0.3 MΩ
20.0 G Ω 500 V 100.0 MΩ 102.8 -2.8 0.1 MΩ
40.0 G Ω 1000 V 50.00 MΩ 50.1 -0.1 0.3 MΩ
40.0 G Ω 1000 V 100.0 MΩ 100.9 -0.9 0.8 MΩ
40.0 G Ω 1000 V 200.0 MΩ 198 2 1 MΩ
40.0 G Ω 1000 V 1.00 GΩ 0.992 0.008 0.006 GΩ
100.0 G Ω 2500 V 200.0 MΩ 200 0 1 MΩ
100.0 G Ω 2500 V 1.00 GΩ 1.001 -0.001 0.006 GΩ
100.0 G Ω 2500 V 5.00 GΩ 4.97 0.03 0.06 GΩ
1 TΩ 5000 V 5.00 GΩ 5.01 -0.01 0.06 GΩ
1 TΩ 5000 V 10.0 GΩ 9.91 0.09 0.06 GΩ
REMARKS :
Page 2 of 2 ACLF-510-01-00