Professional Documents
Culture Documents
3. In mass spectrometer, the sample that 6. No two molecules will be fragmented and
has to be analysed is bombarded with ionized in exactly the same manner.
which of the following? a) True
a) Protons b) False
b) Electrons
Answer: a
c) Neutrons
Explanation: No two molecules will be
d) Alpha particles
fragmented and ionized in exactly the same
Answer: b manner. This is how different molecules are
Explanation: In mass spectrometer, the identified in a complex mixture.
sample which is to be analysed is
7. The procedure for mass spectroscopy
bombarded with electrons. As a result, ions
starts with which of the following
are produced.
processes?
4. Mass spectrometer separates ions on the a) The sample is bombarded by electron
basis of which of the following? beam
a) Mass b) The ions are separated by passing
b) Charge them into electric and magnetic field
c) Molecular weight c) The sample is converted into gaseous
d) Mass to charge ratio state
d) The ions are detected
Answer: c 11. In mass spectrometer, the sample gas is
Explanation: The procedure for mass introduced into the highly evacuated
spectroscopy starts with converting the spectrometer tube and it is ionised by
sample into a gaseous state. This is done by electron beam.
chemical processes. a) True
b) False
8. In a mass spectrometer, the ion currents
are measured using which of the Answer: a
following? Explanation: The sample gas is introduced
a) Scintillation counter into the highly evacuated spectrometer tube
b) Ion counter and it is ionised by an electron beam. The
c) Electrometer tube sample has to be in a gaseous state.
d) Electric fields
COMPONENTS OF MASS SPECTROMETRY
Answer: c
1. Which of the following is not a
Explanation: The ion currents are measured
component of mass spectrometer?
using a sensitive electrometer tube. The ions
a) Inlet system
reaching the collecting plate are measured.
b) Sweep generator
9. Which of the following ions pass through c) Ion transducer
the slit and reach the collecting plate? d) Mass analyser
a) Negative ions of all masses
Answer: b
b) Positive ions of all masses
Explanation: Sweep generator is not a
c) Negative ions of specific mass
component of a mass spectrometer. It is a
d) Positive ions of specific mass
component of NMR spectrometer.
Answer: d
2. Which of the following can be introduced
Explanation: Positive ions of specific mass
into the ionization chamber directly?
pass through the slit and reach the collecting
a) Solid samples with low vapour
plate. These ions are measured.
pressure
10. Which of the following statements is not b) Solid samples with high vapour
true about mass spectrometry? pressure
a) Impurities of masses different from c) Liquid samples with low density
the one being analysed interferes with d) Liquid samples with high density
the result
Answer: a
b) It has great sensitivity
Explanation: Solid samples with low vapour
c) It is suitable for data storage
pressure can be introduced into the
d) It is suitable for library retrieval
ionization chamber directly.
Answer: a
3. Inlet system is also known as which of
Explanation: Impurities of masses different
the following?
from the one being analysed does not
a) Initial system
interfere with the result in mass
b) Sample reservoir
spectroscopy. This is a major advantage of
c) Sample handling system
this technique.
d) Element injection system
Answer: c 7. Mass analyser is similar to which of the
Explanation: Inlet system introduces the following in optical spectrometer?
sample into the ion source. Hence, it is called a) Source
a sample handling system. b) Monochromator
c) Detector
4. Which of the following is normally done
d) Sample
to convert the sample into the gaseous
state? Answer: b
a) Sample is pressurized Explanation: Mass analyser is similar to
b) Chemical reactions are made to occur monochromator in an optical spectrometer.
c) Sample is heated It separates ions according to their
d) Sample is cooled mass/charge ratio.
12. Given below is the block diagram of mass MAGNETIC DEFLECTION MASS
spectrometer. Identify the unmarked SPECTROMETER
component.
1. Which of the following produces the
electron beam in magnetic deflection
mass spectrometer?
a) Tungsten filament
a) Inlet system b) Quartz rod
b) Ionisation chamber c) Silica
c) Vacuum system d) Rhodium filament
d) Ion transducer
Answer: a
Answer: d Explanation: Tungsten filament produces the
Explanation: The unmarked component is electron beam in magnetic deflection mass
ion transducer. It will give current at the spectrometer. The electrons are produced
output at its output side when ions are by the heated filament.
present on the input side.
2. In magnetic deflection mass
13. Which of the following is not a spectrometer, in which of the following
characteristic of nebulizers that are ways is acceleration applied to the
commonly used? direction of motion?
a) Low cost a) In random manner
b) Low uptake rate b) Parallel to it
c) High efficiency c) Perpendicular to it
d) High uptake rate d) Along it
Answer: d Answer: c
Explanation: Commonly used nebulizers Explanation: Acceleration in the mass
spectrometer, is applied perpendicular to
the direction of motion. The velocity of the 6. Which of the following leads to the
object remains constant. limitation of resolution?
a) All ions do not have same energy
3. Direct focussing is obtained by deflecting
b) All ions do not have same charge
the ion beam along a _________ trajectory
c) All ions are not of the same size
through the magnetic field.
d) All ions do not have the same charge
a) 120o
b) 150o Answer: a
c) 190o Explanation: The resolution is limited as all
d) 180o ions do not have the same energy. Hence,
the ions would not have the same velocity.
Answer: d
Explanation: Direct focussing is obtained by 7. The electric sector field is not subject to
deflecting the ion beam along a hysteresis.
180o trajectory through a magnetic field. The a) True
gap between the poles must be large b) False
enough to contain the ion source.
Answer: a
4. Which of the following separate the ions Explanation: The electric sector field is not
according to their mass-to-charge? subject to hysteresis. Hence, the relationship
a) Ion source between mass to charge ratio and
b) Detector accelerating voltage is linear.
c) Magnetic sector
8. Which of the following components need
d) Electric sector
to be added in order to increase the
Answer: c resolution?
Explanation: Magnetic sector will separate a) Ion source
the ions according to their mass-to-charge. b) Detector
The magnet will decide the mass to charge c) Magnetic sector
ratio. d) Electric sector
Answer: b Answer: d
Explanation: A potential energy selector is Explanation: The DC ion accelerating voltage is
placed before the detector. It balances out the swept from 50 to 250 V. This is done twice per
energy of the ion beam. second.
2. Which of the following is applied in one 6. The resolution of the Bennet tube can be
or more stages where each stage is a improved by using which of the
series of three equally spaced parallel following?
grids? a) Sine wave rf signal
a) Ionisation field b) Square wave rf signal
b) Reflection c) Triangular wave rf signal
c) RF field d) Random signal
d) Acceleration
Answer: b
Answer: c Explanation: The resolution of the Bennet tube
Explanation: RF field is applied in one or more can be improved by using square wave rf signal.
stages. Each stage is a series of three equally Usually, sinusoidal wave signal is used.
spaced parallel grids.
7. In Bennet spectrometer, the rf voltage
3. The alternating voltage rf voltage is has a fixed frequency.
applied to which of the following? a) True
a) All grids b) False
b) Central grid
Answer: a
c) Alternate grids
Explanation: In Bennet spectrometer, the rf
d) None of the grids
voltage has a fixed frequency. It is modulated at
Answer: b 10 percent at 1kHz.
Explanation: The alternating voltage rf voltage is
applied to the central grid. All the other grids
are kept at ground potential.
8. The spurious lines can be reduced to a Answer: b
minimum by using which of the Explanation: The resolving power of Bennet
following? type spectrometer is primarily based on the
a) Single stage tube difference between the individual rf accelerating
b) Two stage tube stages.
c) Three stage tube
QUADRUPOLE MASS SPECTROMETER
d) Reflectron
1. Which of the following is not a feature of
Answer: c
quadrupole mass spectrometer?
Explanation: The spurious lines can be reduced
a) Low cost
to a minimum by using three stage tubes. Two
b) Light weight
stage tubes tend to produce spurious lines.
c) Low speed electronic scanning
9. Which of the following is the d) Simple in construction
disadvantage of using three stage tubes?
Answer: c
a) Reduced resolution
Explanation: In quadrupole mass spectrometer,
b) Bulky construction
electronic scanning takes place. The scanning
c) Complex
takes place at high speed.
d) Heavy
2. Which of the following is not a
Answer: a
component of quadrupole mass filter?
Explanation: The spurious lines can be reduced
a) Electrodes
to a minimum by using three stage tubes. This
b) Choke
results in reduced resolving power.
c) DC potential
10. Which of the following is not the feature d) Detector
of Bennet type spectrometer?
Answer: d
a) Lightweight
Explanation: Detector is not a part of the filter.
b) Simple
The filter uses DC potential.
c) Absence of spurious lines
d) Moderate resolution 3. The mass selection scheme uses a dc
potential and not a radio frequency
Answer: c
potential.
Explanation: Absence of spurious lines is not a
a) True
feature of Bennet type spectrometer. Spurious
b) False
lines can be reduced by adding new features.
Answer: b
11. The resolving power of Bennet type
Explanation: The mass selection scheme uses a
spectrometer is primarily based on which
dc potential. It also uses radio frequency
of the following?
potential.
a) Distance between detector and
reflectron 4. If to one pair of electrodes one potential
b) Difference between the individual rf with one sign is applied, which of the
accelerating stages following is applied to the other pair of
c) Length of the drift tube electrodes?
d) Laser beam used a) Same potential with same sign
b) Different potential with same sign not affect the resolution of the instrument.
c) Different potential with different sign Increase in resolution results in an increase in
d) Same potential with different sign the number of ions reaching the detector.
6. The potential maintained across the 9. In which of the following detectors, the
continuous dynode multiplier transducer collector is placed in an inclined manner?
is which of the following? a) Faraday cup collector
a) 1 to 5kV b) Channeltron
b) 5 to 10kV c) Micro-channel plate
c) 2.5 to 6.2kV d) Electron multiplier transducer
d) 1.8 to 2kV
Answer: a
Answer: d Explanation: The collector is placed in an
Explanation: The potential maintained across inclined manner in a faraday cup collector. This
the continuous dynode multiplier transducer is is to reflect the ions away from the entrance of
1.8 to 2kV. Electrons cascade down the tube. the cup.
7. Which of the following are not the 10. Which of the following is the least
characteristics of an electron multiplier sensitive ion detector?
transducer? a) Faraday cup collector
a) Rugged b) Channeltron
b) Reliable c) Micro-channel plate
c) Microsecond of response time d) Electron multiplier transducer
required
Answer: a
d) High current gain
Explanation: Faraday cup collector is the least
Answer: c sensitive ion detector. It is used in places where
Explanation: Electron multiplier transducer great sensitivity is not required.
requires the only nanosecond of response time.
11. Photographic plates are coated with
It provides a high current gain.
which of the following?
8. Which of the following are true about a) Lead
dynodes used in discrete dynode b) Quartz
electron multiplier? c) Silver bromide
a) The dynodes are all kept at same d) Mercuric sulphate
voltage
Answer: c
b) Successive dynodes are held at lower
Explanation: Photographic plates are coated
potential
with silver bromide. These are sensitive to
c) Successive dynodes are held at higher
energetic ions.
potential
d) All dynodes are held at very low 12. Which of the following detectors are well
potential suited for spark source instruments?
a) Faraday cup collector
Answer: c
b) Photographic plates
Explanation: In discrete dynode electron
c) Micro-channel plate
multiplier, successive dynodes are held at
d) Electron multiplier transducer
Answer: b 2. Which of the following isotope is NMR
Explanation: Photographic plates are well suited inactive?
for spark source instruments. It has great *
sensitivity.
1/1
13. Photographic plates have greater 1H
sensitivity and resolution than electron 13C
multipliers. 15N
a) True 12C
b) False
Correct answer
* Correct answer
1/1 Perpendicular and weaker field
Radioactive radiation
Excite an atom
13. Interaction between matter and Vibrate an atom
electromagnetic radiation can be observed by Vibrate a molecule
subjecting a substance to magnetic fields in Affect the nuclear spin of an atom
which of the following manner?
*
0/1
Both fields should be stationary 17. Nuclei having either the number of protons
Both fields should be varying or neutrons as odd have _______ spin.
One field should be stationary and the other should
*
be varying
It must be subjected to only one field 0/1
Integral spin
Correct answer
Half integral spin
Zero spin
One field should be stationary and the other should Positive spin
be varying
14. When energy is absorbed by the sample, the Correct answer
absorption can be observed as a change in Half integral spin
signal developed by which of the following
components? 18. The difference between the field necessary
* for resonance in the sample and in some
1/1 arbitrary chosen compound is which of the
following?
Amplifier
Photodetector *
GM counter 1/1
Radiofrequency detector
Field shift
Matrix effects
15. Which of the following are considered to be Chemical shift
the lowest form of Electromagnetic radiation?
Resonance shift
*
1/1
True
16. The amount of energy available in radio
frequency radiation is sufficient for which of the False
following?
*
1/1
21. How many signals does the aldehyde Frequency of unknown group of protons
Frequency of TMS
(CH3)3CCH2CHO have in 1H NMR and 13C
Operating frequency of the instrument
NMR spectra? All of the above
*
0/1
Correct answer
five 1H signals and six 13C signals Operating frequency of the instrument
three 1H signals and four 13C signals
five 1H signals and four 13C signals
three 1H signals and six 13C signals
Correct answer
three 1H signals and four 13C signals 1. NMR spectroscopy is used for
determining structure in which of the
following materials?
22. Which of the compounds below is used as a a) Radioactive materials
standard in proton NMR and carbon-13 NMR? b) Insoluble chemical compounds
c) Liquids
*
d) Gases
0/1
Answer: c
TMS
Explanation: NMR spectroscopy is used for
C6H6
determining structure in liquids. It is also used
CDCl3
for determining the structure in soluble
D2O
chemical compounds.
12. If the number of protons or neutrons is
Correct answer
even the spin of the nucleus will be which
TMS of the following?
a) Integral spin
23. The nuclei with a spin quantum number
b) Half integral spin
greater than _______ can exhibit the NMR c) Zero spin
phenomenon. d) Positive spin
*
Answer: c
1/1 Explanation: If the number of protons or
0 neutrons is even the spin of the nucleus will be
zero. Examples are C12 and O16.
5 13. Elementary particles such as electrons
10
and nucleus have the property of spin.
-5
a) True
b) False
24. Which of the following is inversely Answer: a
proportional to the chemical shifts positions (δ)? Explanation: Elementary particles such as
* electrons and nucleus have the property of
0/1
spin. They are known to behave as if they rotate c) Slowly varying scan voltage
about an axis. d) Rapidly varying scan voltage
Answer: a
2. In frequency sweep method, which of the
Explanation: Minimal type NMR spectrometer
following parameters are varied
uses a permanent magnet. Hence, it does not
continuously?
require a power supply and cooling system.
a) Magnetic field
b) RF signal
9. Minimal type of NMR spectrometer is c) Sample concentration
inexpensive. d) Amplification factor
a) True
b) False
Answer: b
Explanation: In frequency sweep method,
Answer: a magnetic field is held constant. RF signal is
Explanation: Minimal type of NMR spectrometer swept or varied continuously.
is inexpensive. It uses a permanent magnet and
does not require a power supply.
3. In field sweep method, which of the
following parameters are varied
10. Which of the following is used as a continuously?
coolant for magnets in continuous-wave a) Magnetic field
NMR spectrometer? b) RF signal
a) Water c) Sample concentration
b) Liquid He d) Amplification factor
c) Liquid Na
d) Heavy water
Answer: a
Explanation: In frequency sweep method, RF
Answer: a signal is held constant. Magnetic field is swept
Explanation: Water is used as a coolant for or varied continuously.
magnets in continuous-wave NMR
spectrometer. Liquid He is used for FT-NMR
4. Which of the following statements are
spectrometers.
not true about permanent magnets?
a) They are simple
CONTINUOUS WAVE NMR SPECTROSCOPY b) They are inexpensive
c) They don’t require shielding
d) They don’t require power supply
1. Which of the following components are
used to separate the nuclear spin energy
states? Answer: c
a) RF channels Explanation: Permanent magnets are simple
b) Magnet
and inexpensive. But, permanent magnets 8. The sample is contained in which of the
require extensive shielding. following components?
a) Flask
b) Capillary tube
5. Which of the following is not true of
c) Curvette
electromagnets?
d) Bore glass tube
a) They are expensive
b) They require a power supply
c) They don’t require a cooling system Answer: d
d) They don’t require extensive shielding Explanation: The sample is contained in a bore
glass tube. It should be cylindrical and thin-
walled.
Answer: c
Explanation: Electromagnets require a power
supply. They also require cooling systems. 9. The single coil probe supplies the RF
radiation to the sample and also serves
as a part of which of the following
6. For high resolution work the magnetic
circuits?
field over the entire sample volume must
a) RF channels
be maintained uniform in space and
b) Magnet
time.
c) Detector
a) True
d) Sweep generator
b) False
Answer: c
Answer: a
Explanation: The single coil probe supplies the
Explanation: For high resolution work the
RF radiation to the sample and also serves as a
magnetic field over the entire sample volume
part of the detector circuit. It has only one coil.
must be maintained uniform in space and time.
To do this all the components must be kept in
specified conditions. 10. Nuclear induction probes’ one coil is
used for signal detection. What is the
function of the other coil?
7. Which of the following must not be done
a) Sample irradiation
to maintain the magnetic field over the
b) Magnet
sample uniform in space and time?
c) Detector
a) Large pole pieces need to be used
d) Sweep generator
b) Pole faces must be polished
c) Wide pole gap must be present
d) Magnets can be permanent or Answer: a
electromagnet Explanation: Crossed coil probes are also called
nuclear induction probes. One coil is used for
irradiating the sample. The other coil is
Answer: c
mounted orthogonally for signal detection.
Explanation: Narrow pole gap must be present
to maintain the magnetic field over the sample
uniform in space and time. The pole faces must 11. Given below is the diagram of Cw NMR
be polished to optical tolerances. spectrometer. Identify the unmarked
component.
c) Generator
d) Superconducting solenoids
Answer: d
Explanation: Superconducting solenoids can be
used instead of magnets to produce the
magnetic field. The magnet must be stable and
a) Recorder homogeneous.
b) RF channel 14. How is the inhomogeneity of magnetic
c) Receiver fields compensated?
d) Sweep generator
a) With large magnetic fields
b) With small magnetic fields
Answer: d c) By using two or more magnets
Explanation: The unmarked component is d) By providing required insulation
sweep generator. Either field or frequency can
Answer: b
be swept.
Explanation: The magnet must be stable and
11. The voltage generated by the receiver homogeneous. The inhomogeneity of magnetic
coil is small and it must be amplified. fields is compensated with small magnetic
a) True fields.
b) False
FOURIER TRANSFORM NMR SPECTROSCOPY
Answer: a
Explanation: The voltage generated by the
receiver coil is small and it must be amplified. It 1. Fourier transform NMR spectrometer
is them fed to the recorder or oscilloscope. allows NMR transitions to be observed
simultaneously.
12. The amplification required for a) True
continuous-wave NMR is of the order of b) False
which of the following?
a) 101
Answer: a
b) 102 Explanation: Fourier transform NMR
c) 103 spectrometer is a type of NMR spectrometer. It
d) 105 allows samples to be observed simultaneously
instead of serially.
Answer: d
Explanation: The amplification required for
continuous-wave NMR is of the order of 105. The 2. Fourier transform NMR spectrometer has
spectrum can be recorded after amplification. which of the following characteristics?
a) Increased sensitivity, long time to
13. Which of the following can be used obtain data
instead of magnets to produce the b) Decreased sensitivity, long time to
magnetic field? obtain data
a) Inductor c) Increased sensitivity, reduced time to
b) Motor obtain data
d) Decreased sensitivity, reduced time to NMR spectrometer is that the excitation is
obtain data inefficient. Only a narrow band of frequencies
are contributing to the signal at a time.
Answer: c
Explanation: Fourier transform NMR 6. Which of the following must be done to
spectrometer has increased sensitivity. It takes improve sensitivity?
less time to obtain NMR data. a) Frequency sweep mode must be
preferred
b) Field sweep mode must be preferred
3. Which of the following cannot be done
c) Single channel excitation must be
due to the multiplex advantage?
preferred
a) Repetitive signals can be summed
d) Multichannel excitation must be
b) Repetitive signals can be averaged
preferred
c) Increases signal to noise ratio
d) Decreases signal to noise ratio
Answer: d
Explanation: Sweep techniques have less
Answer: d
sensitivity. To improve sensitivity, multichannel
Explanation: The multiplexing advantage allows
excitation must be preferred.
improvement of the signal to noise ratio. It also
allows repetitive signals to be summed and
averaged. 7. Which of the following is the
disadvantage of multichannel excitation?
a) Low sensitivity
4. Two coils are necessary for Fourier
b) Low resolution
transform NMR spectroscopy.
c) Small number of frequencies is present
a) True
d) Uneconomical
b) False
Answer: d
Answer: b
Explanation: Large number of frequencies is
Explanation: Only one coil is necessary for
present in multichannel excitation. This is
Fourier transform NMR spectroscopy. The coil
uneconomical since a receiver is required for
serves as both antenna for transmitting and
each channel.
receiving RF radiation.
Answer: a
Answer: b
Explanation: The most common spherical cells
Explanation: Sensitivity is the ratio of peak
have 4mm diameter. A 1mm capillary cell is
signal amplitude to rms noise. It is a measure of
centred along the axis of the coil.
the ability of an instrument to differentiate
signal from surrounding noise.
5. Better sensitivity with good resolution
can be achieved by concentrating the
sample inside which of the following? a) True
a) Sample cell b) False
b) Micro cell
c) Sample tube
Answer: b
d) Cylindrical tube
Explanation: The amplitude of the NMR signal
caused by the absorption of RF energy at the
Answer: b radio frequency depends on the power of the
Explanation: Better sensitivity with good RF energy applied. It also depends on the sweep
resolution can be achieved by concentrating the rate selected.
sample inside a specially designed NMR
microcell. It could be a spherical cell or a
9. Which of the following is not involved in
capillary cell.
signal averaging?
a) A system to repeatedly scan the
6. Frequency spectrum of noise occupies a spectral region of interest
band that is _________ the signal b) Some storage device to store
spectrum. Choose the most appropriate information
option. c) A system to coherently add individual
a) Narrow than spectra
b) Wider than d) Sweep generator
c) Different from
d) Same as
Answer: d
Explanation: Sweep generator is not involved in
Answer: b signal averaging. This is usually accomplished
Explanation: Frequency spectrum of noise using a small computer.
occupies a band that is wider than the signal
spectrum. Filters operate on this concept.
10. In order to have a long measuring time
for increasing sensitivity, it is better to
7. Suppression of frequencies which have several fast scans than a single slow
contain no signal does which of the scan.
following? a) True
a) Increases S/N ratio b) False
b) Decreases S/N ratio
c) S/N ratio is kept constant
Answer: a
d) S/N ratio becomes 1
Explanation: In order to have a long measuring
time for increasing sensitivity, it is better to have
Answer: a several fast scans than a single slow scan. The
Explanation: Suppression of frequencies which sensitivity increases by the square root of a
contain no signal increases S/N ratio. The noise number of scans.
decreases.
11. In Spin Decoupler, the output of the DC
8. The amplitude of the NMR signal caused amplifier is applied to which of the
by the absorption of RF energy at the following to control the power supplied
radio frequency does not depend upon to the oven heaters?
the power of the RF energy applied. a) Power diode
b) Power transistor
c) Power MOSFET
d) Power IGBT
Answer: b
Explanation: In Spin Decoupler, the output of
the DC amplifier is applied to the power
transistor to control the power supplied to the
oven heaters. The temperature is maintained at
plus or minus 1oC.
Answer: a
Explanation: In Spin Decoupler, the difference
frequency circuit provides a means of mixing
two signals and a filter to recover the low-
frequency component. It is included to provide
a read-out frequency.