Professional Documents
Culture Documents
| User Manual
|
Cat.-No. 16100/1
QduhrhnmKhrsnesgdL`mt`k
No. DATE / Rev. REVISION DESCRIPTION
1 01/2004-07 First edition
2 02/2006-08 Addition of Chapter 4.2.1 / Köhler Illumination; Update Specifications
i
ii
1 INTRODUCTION
This manual is considered as a part of the instrument; it has to be at the operator’s hand as well as at the
maintenance operator’s availability. For accurate installation, use and maintenance, please read the following
instructions carefully. In order to avoid instrument or personal damages, carefully read the ”GENERAL SAFETY
WARNINGS”, describing the suitable operating procedures. In case of breakdowns or any troubles with the
instrument, apply to the local Technical Service.
2 USER WARRANTY
HUMAN warrants that instruments sold by one of its authorised representatives shall be free of any defect in
material or workmanship, provided that this warranty shall apply only to defects which become apparent within
one year from the date of delivery of the new instrument to the purchaser.
The HUMAN representative shall replace or repair any defective item at no charge, except for transportation
expenses to the point of repair.
This warranty excludes the HUMAN representative from liability to replace any item considered as expendable in
the course of normal usage, e.g.: lamps, valves, syringes, glassware, fuses, diskettes, tubing etc.
The HUMAN representative shall be relieved of any liability under this warranty if the product is not used in
accordance with the manufacturer's instructions, altered in any way not specified by HUMAN, not regularly
maintained, used with equipment not approved by HUMAN or used for purposes for which it was not designed.
HUMAN shall be relieved of any obligation under this warranty, unless a completed installation / warranty
registration form is received by HUMAN within 15 days of installation of this product.
This warranty does not apply to damages incurred in shipment of goods. Any damage so incurred shall be reported
to the freight carrier for settlement or claim.
I
5 DISPOSAL MANAGEMENT CONCEPT
The currently valid local regulations governing disposal must be observed. It is in the responsibility of the user to
arrange proper disposal of the individual components.
All parts which may comprise potentially infectious materials have to be disinfected by suitable validated
procedures (autoclaving, chemical treatment) prior to disposal. Applicable local regulations for disposal have to be
carefully observed.
The instruments and electronic accessories (without batteries, power packs etc.) must be disposed off according to
the regulations for the disposal of electronic components.
Batteries, power packs and similar power source have to be dismounted from electric/electronic parts and disposed
off in accordance with applicable local regulations.
6 INSTRUMENT DISINFECTION
Instruments or parts which may come in contact with biological samples (patient specimens, controls etc.) should
be considered at least potentially infectious.
Before doing any servicing on the instrument it is very important to thoroughly disinfect all possibly contaminated
parts. Before the instrument is removed from the laboratory for disposal or servicing, it must be
decontaminated/disinfected. Decontamination/disinfection should be performed by authorised welltrained
personnel, observing all necessary safety precautions. Instruments to be returned have to be accompanied by a
disinfection certificate completed by the responsible laboratory manager. If a disinfection certificate is not
supplied, the returning laboratory will be responsible for charges resulting from non-acceptance of the instrument
by the servicing centre, or from authority´s interventions.
7 NOTICE
Every effort has been made to avoid errors in text and diagrams, however, HUMAN GmbH assumes no
responsibility for any errors which may appear in this publication. It is the policy of HUMAN GmbH to improve
products as new techniques and components become available. HUMAN GmbH therefore has to reserve the right
to change specifications if necessary in the course of such improvements.
II
Bnmsdmsr
1 NOTICE 3
2 APPLICATIONS 3
3 SPECIFICATIONS 3
4 STRUCTURE 4
4.1 Base 4
4.2 Illumination System 4
4.2.1 Köhler Illumination 4
4.3 Mechanical Stage (Slide Holder) 4
4.4 Ocular, Objective and Nosepiece 5
4.5 Coarse and Fine Focussing Knobs 5
4.6 Condenser 5
4.7 Immersion Oil 5
5 OPERATION 7
6 STANDARD ACCESSORIES 7
7 CARE AND MAINTENANCE: 7
8 DISPOSAL MANAGEMENT CONCEPT: 7
9 MICROSCOPE GLOSSARY 8
9.1 Darkfield Microscopy 8
9.2 Brightfield 8
9.3 Phase Contrast 8
9.4 Fluorescence Microscopy 9
9.5 Lenses 9
9.6 Focus 9
2/9
1 NOTICE
Every effort has been made to avoid errors in this text and the accompanying diagrams, however, HUMAN GmbH
assumes no responsibility for any errors which may appear in this publication. It is the policy of HUMAN GmbH to
improve products as new technologies and components become available. Human GmbH therefore reserves the
right to change specifications if necessary in the course of such improvements.
2 APPLICATIONS
The HUMASCOPE biological microscope is suited for use in medical and sanitary establishments, laboratories,
institutes, agricultural research networks, colleges and universities for clinical microscope inspections and may be
used as an instructional instrument. It can also be used in research work such as biology, bacteriology, cytology,
histology and pharmaceutical chemistry.
3 SPECIFICATIONS
Ocular
Objective
Total Magnification
Objective
Total Magnification 4X 10X 40X 100X
Ocular
10X 40X 100X 400X 1000X
16X 64X 160X 640X 1600X
,OOXPLQDWLRQ6\VWHP
In order to provide sufficient brightness to observe the specimen and make full use of the resolution of the
objectives, this instrument uses a 6V, 20W built-in variable-brightness lamp (see figure 4.0). For optimal results, the
condenser must be adjusted properly. The condenser is composed of two parts. The iris diaphragm is mounted
under the stage, the illumination source is in the base. Adjust the condenser iris (mounted under the stage) to
match the aperture of the objective. The stage can be raised and lowered using the focussing knobs. The light axis
of the condenser must coincide with the light axis of the illumination source. To correct the alignment, adjust 3
screws in the part of the condenser mounted on the stage.
If the light path is set up properly, the user has the advantage of an evenly illuminated field, a bright image
without glare and minimum heating of the specimen.
The following instructions apply to a microscope equipped for transmitted-light brightfield illumination..
1. Switch on the light source and make sure that light is passing through the field diaphragm at the base of
the microscope stand. It may help to place a piece of paper over the field stop to confirm this.
Place your specimen on the stage and turn the nosepiece (which holds the objective lenses) to the 10X or
20X lens. Open the field diaphragm as far as possible.
2. Check if the specimen is illuminated. It will help to place a piece of paper over the top of the specimen to see
if light is getting through. If you are using the brightfield condenser stop, open the iris diaphragm (or
aperture diaphragm) on the condenser turret (which contains the stops for brightfield and phase, etc) to the
maximum. The front lens should be about 1-3 mm above the specimen. Use the condenser focussing knobs
to do this.
3. Bring the specimen into focus with the coarse and fine focussing knobs. If the light is too bright, reduce the
light intensity using the light adjustment wheel.
4. When the specimen is in focus, start to close the field diaphragm and also begin to carefully move the
condenser up and down with the condenser focussing knobs. Look for a sharp image of the edge of the field
diaphragm.
5. When the edge of the field diaphragm silhouette is sharply defined, centre it with the two knurled knobs
that extend diagonally from the condenser. Close down the field diaphragm most or all the way to get it
centred properly. When it is centred, open the field diaphragm until its edge is outside the field.
6. Your specimen should now be properly illuminated and you should have a good image.
4/9
4.4 Ocular, Objective and Nosepiece
The microscope imaging system is composed of the two oculars and the objective (see figure 4.0). The objectives
are screwed into the threaded holes in the nosepiece, which can be rotated to quickly achieve the desired
magnification. The advanced, high-precision nosepiece design ensures that the field of view is always centred in
the observation area for precise focussing (parfocal). The ocular tube is inclined 450 for convenient, fatigue-free
observation.
&RDUVHDQG)LQH)RFXVVLQJ.QREV
The HUMASCOPE features coaxial coarse and fine focussing knobs on the left and right side of the instrument. The
knobs are located below the stage and are easy to operate (see figure 4.0). The coarse and fine focussing range is
30mm. The fine focussing knobs are the smaller knobs centred in the coarse focussing knobs. The lever on the left,
inside the left-side coarse focussing knob, is the coarse focussing stopper. When the specimen is in clear focus, the
stopper may be used to lock in the focus. This also allows the stage to be lowered and raised again to exactly the
same position. When the position is no longer needed, simply loosen the stopper.
&RQGHQVHU
The Abbe condenser is a specially designed lens mounted under the stage. It can be raised or lowered. It has an iris-
type aperture to control the diameter of the beam of light entering the lens system. By changing the size of the iris
and moving the lens toward or away from the stage, the diameter and focal point of the cone of light that goes
through the specimen can be controlled. Abbe condensers become most useful at magnifications above 400x. The
condenser lens system should have a numerical aperture equal to or greater than the N.A. of the objective lens
being used.
,PPHUVLRQ2LO
Immersion oil is a special oil used in microscopy only with the 100X objective lens (usually at 1000X total power). A
drop is placed on the cover slip and the objective is lowered until it just touches the drop. Once brought into focus,
the oil acts as a bridge between the glass slide and the glass in the lens. This concentrates the light path and
increasing the resolution of the image
Revolving
Nosepiece
Objective
Focussing
Knobs Mechanical
Stage
Light Source
Condenser
6/9
5 OPERATION
1. Screw the objectives into the threaded holes in the nosepiece in order of their magnification. Insert the oculars
into the inclined ocular tube.
2. Place a specimen slide on the stage. Adjust the stage so that the specimen is centred over the hole in the stage.
3. Using first the 10X objective, turn the coarse focussing knob to move the objective close to the specimen. Now
observe the image through ocular and turn the coarse focussing knob to lower the objective until the image
comes into approximate focus. Next, turn the fine focussing knob until the image is sharp. Thanks to the
instrument’s precise and reliable nosepiece, a clear focus is maintained even when switching objectives.
4. The specimen can also be observed by replacing the lower condenser with a plan-concave reflecting mirror.
6 STANDARD ACCESSORIES
ITEMS CONTENTS QUANTITY
1 Ocular WF10X, P16X 2PCS each
2 Objective 4X 10X 40X 100X 1PC each
3 Reflecting Mirror 1
4 Spare Lamp 1
5 Spare Fuse 1
6 Immersion Oil 1BOTTLE
7 User Manual 1
8 Blue Filter 1
9 Power Cable 1
10 Dust Cover HUMAN 1
The advantage of darkfield illumination is that details can be seen that are normally not resolved by the
microscope’s objective. While the actual details cannot be seen, the light reflected by them can. A nice analogy here
is that of dust in a room. Normally, in a well-lit room very small dust particles cannot be seen. However, if the lights
are turned out, a beam of light from an acute angle makes these same particles visible. Besides the optical
advantages, darkfield illumination is very beautiful and gives an almost science-fiction-like image.
%ULJKWILHOG
This is the basic microscope configuration.
In brightfield microscopy, the transparent or translucent specimen is either naturally coloured or stained and
appears dark against a bright, white background or field.
3KDVH&RQWUDVW
Phase contrast is a technique for revealing the structural features of microscopic, transparent objects that cannot
otherwise be accomplished with brightfield microscopy. Phase contrast achieves the same effect as staining a
specimen (which can kill a live specimen). In phase-contrast microscopy, the “staining” is in the optics.
In a phase-contrast microscope, the annular rings in the objective lens and the condenser split the light. The light
that passes through the central part of the light path is recombined with the light that travels around the
periphery of the specimen. The interference produced by these two paths produces images in which the dense
structures appear darker than the background.
8/9
)OXRUHVFHQFH0LFURVFRS\
The fluorescence of a substance is seen when a molecule is exposed to a specific wavelength of light (excitation
wavelength or spectrum) and the light it emits (the emission wavelength or spectrum) is of a higher wavelength.
To view this fluorescence in the microscope, several light-filtering components are needed. Specific filters are
needed to isolate the excitation and emission wavelengths of a fluorochrome. A bright light source with proper
wavelengths for excitation is also needed. For normal fluorescence applications, this is a mercury vapour arc
burner.
/HQVHV
objective lens - gathers light from the specimen
ocular - transmits and magnifies the image from the objective lens to your eye
nosepiece - rotating mount that holds 4 interchangeable objective lenses
tube - holds the ocular at the proper distance from the objective lens and blocks out stray light
)RFXV
position the objective lens at the proper distance from the specimen
coarse-focus knob - used to bring the object into the focal plane of the objective lens
fine-focus knob - used to make fine adjustments to focus the image
02/2006-08