Diasys GmbH is not responsible for any carryover interference which is caused by non-DiaSys reagents.

Carryover Pair Table (Contamination Setting “PROBE”)


Setting condition for avoiding reagent probe contamination
Influence
Substance contaminating Substance contaminated Preventive detergent
effect
1 R1 AP R1 MG Water 999
R2 AP R2 MG Water 999
2 R1 AMY R1 MG Reagent Probe Wash 2 999
R2 AMY R2 MG Reagent Probe Wash 2 999
3 R1 ASO R1 PO3 Water 999
R2 ASO R2 PO3 Water 999
4 R1 CHOL R1 LPS Reagent Probe Wash 1 999
5 R1 CKMB R1 MG Reagent Probe Wash 2 999
R2 CKMB R2 MG Reagent Probe Wash 2 999
6 R1 DBIL R1 LDH Water 999
R2 DBIL R2 LDH Water 999
7 R1 FE R1 IGM Water 999
R2 FE R2 IGM Water 999
8 R1 IGG R1 RF Water 999
R2 IGG R2 RF Water 999
9 R1 IGM R1 MG Reagent Probe Wash 2 999
R2 IGM R2 MG Reagent Probe Wash 2 999
10 R1 LDL-C R1 LPS Reagent Probe Wash 1 999
R2 LDL-C R2 LPS Reagent Probe Wash 1 999
R1 LDL-C R1 LPS Reagent Probe Wash 2 999
R2 LDL-C R2 LPS Reagent Probe Wash 2 999
11 R1 LPS R1 CA Reagent Probe Wash 2 999
R2 LPS R2 CA Reagent Probe Wash 2 999
12 R1 PAMY R1 MG Reagent Probe Wash 1 999
R2 PAMY R2 MG Reagent Probe Wash 1 999
13 R1 PL R1 CA Water 999
R2 PL R2 CA Water 999
14 R1 TBIL R1 PO3 Water 999
R2 TBIL R2 PO3 Water 999
15 R1 TP R1 CKMB Water 999
R2 TP R2 CKMB Water 999
16 R1 TRIG R1 UA TOOS Water 999
17 R1 HBUT 21 R1 MG Water 999

Order information

Reagent Probe Wash 1 Cat. No. 980 106 5x 250 mL


Reagent Probe Wash 2 Cat. No. 980 107 5x 250 mL

Carryover Pair Table DiaSys BM6010C page 1 February 2018/2