SERVICE MANUAL
Remarks :
u Before doing the calibration, please make sure that you have to adjust the
mini jumper SWA1 on the A/D board to “ADJ” location.
u After finished the calibration, you have to re-adjust the mini jumper SWA1
to LOCK location.
KEYBOARD
∧
M O DE U N I TS T →0 ← BL
P CS
1 2 3 4 5 6 7 8
-2-
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
2. LINEARITY CALIBRATION :
Press“ ”.
⇓
DISPLAY
→0 ←
0.00° g
Press“ M O DE
”(don’t release your finger)
∧
Then press“ BL
”.
⇓
DISPLAY
⇒ Program version
XXXXX 1 number
Press“ M O DE
”(don’t release) ⇓
then press“ U N I TS
” DISPLAY
⇒ Internal value
→0 ←
12532 1
01 1
-3-
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
⇓
⇒ Internal value
12532
Put the 1st calibrated mass on the platter(Please refer to the Appendix I table
list for the calibrated weight. For example : 150g x 0.005g scale, the first
calibrated weight is 30g. Then, put 30g on it).
Then, press“ ”
DISPLAY
02 1
⇓
⇒ Internal value
15358 3
⇓
Put the 2nd calibrated mass on the platter(30g. At the moment, total 60g weight
on the pan).
Then, press“ ”
DISPLAY
03 1
⇓
⇒ Internal value
26837 1
⇓
rd
Put 3 calibrated mass on the platter(30g. At the moment, total 90 weight on
the pan)
Press“ ”
-4-
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
DISPLAY
04 1
⇓
⇒ Internal value
26837 1
⇓
th
Put 4 calibrated mass on the platter(30g. At the moment, total weight 120g on
the pan)
Press“ ”
DISPLAY
05 1
⇓
⇒ Internal vale
36938 1
Put 5th calibrated mass on the platter(30g. At the moment, total weight 150g on
the pan)
Press“ ”
DISPLAY
06 1
⇓
⇒ Internal value
36938 1
-5-
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
Put 6th calibrated mass on the platter(30g. At the moment, total weight 180kg
on the pan)
Press“ ”
DISPLAY
07 1
⇓
⇒ Internal value
14151 1
Put 7th calibrated mass on the platter(30g. At the moment, total weight 210kg
on the pan)
Press“ ”
DISPLAY
08 1
⇓
⇒ Internal value
14151 1
⇓
th
Put 8 calibrated mass on the platter(30g. At the moment, total weight 240g on
the pan)
Press“ ”
DISPLAY
09 1
-6-
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
⇓
⇒ Internal value
14151 1
⇓
th
Put 9 calibrated mass on the platter(30g. At the moment, total weight 270g on
the pan).
Press“ ”
DISPLAY
10 1
⇓
⇒ Internal value
14151 1
Put 10th calibrated mass on the platter(30g. At the moment, total weight 300g
on the pan)
Press“ ”
DISPLAY
11 1
⇓
⇒ Internal value
14151 1
T
Press“ ” ⇓
Scale is reverting to 0.
While counting
backward ,remove all the
masses. Linearity
calibration is finished.
-7-
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
Press .
30g 50g 100g 300g 500g
1
Press
30g 50g 100g 300g 500g
2
Press
30g 50g 100g 300g 500g
3
Press
30g 50g 100g 300g 500g
4
Press
30g 50g 100g 300g 500g
5
Press
30g 50g 100g 300g 500g
6
Press
30g 50g 100g 300g 500g
7
Press
30g 50g 100g 500g
8
Press
30g 50g
9
Press
30g
10
Press
T
Press , the scale is returning to 0. While counting backwards, remove all the masses.
-8-
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
3. SPECIFICATION SETTING
<Key meaning >
l “
M O DE
”ð “→ ” Key “
U N I TS
”ð “← ” Key
∧
“
BL
”ð Enter key and 0 ~ 9 number keys
120000 220000
3 Division & decimal
point For one unit “ g” only, keyin “1” in the 5th digit.
-9-
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
DISPLAY
0.00° g
Press“ M O DE
”(don’t release) ⇓
Then, press“ BL
”
Then , press“ P CS
”. The digits on the display are flashing, input the specs
(please refer to the Appendix II for specification stable list)
Ex. 150g*0.005g
00000 0
Press“ BL
” keyin full capacity +9d ⇓
⇒full capacity + 9d
150045
⇓
→0←
Press“ ”
- 10 -
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
00000 0
∧
Press“ BL
”to keyin the number of weight ⇓
⇒calibrated value
15000 0
⇓
→0←
Press“ ”
00000 0
∧
BL
Press“ ”to keyin the number for division and decimal point
53000 0
Division(0.005g) Decimal(3
digits)
→0←
Press“ ”
- 11 -
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
4. TEMPERATURE CALIBRATION
∧
M O DE BL
Press (no release your finger), then, press , the display will show the
program no. xxxxxx.
M O DE T
Press (no release your finger), then, press , the display will show the
→0←
last memorized calibrated temperature values. At this moment, press
twice to delete this temperature data(make sure you already canceled this
temp. values, the display shows 0.00). Normally, BH scale has three
temperature values, high temp., low temp. and normal temp. you have to do
the same procedures for 3 times to delete 3 temp. data(be sure that you already
canceled the 3 temp. data, the display does not show any temperature internal
values).
M O DE
Press , the display showing full capacity weight, i.e. 150000(150g scale),
then, put 150g mass on the scale.
M O DE
Press , the scale will return to 0(while the display is counting backward,
remove the mass from the platter).
- 12 -
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
First, you have to check if the scale has the internal values.
If scale can display the internal value, you just need to reset the temperature
calibration.
Step 1
Step 2
Step 3
- 13 -
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
Step 4
Step 5
Step 6
Step 7
- 14 -
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
Step 8
If scale appear E1 or E2 on the display but the internal value is still between 100000
~ 600000. Please do the following procedures :
(1) Press to turn the power off. And then, press to turn the machine on
again. The display shows E1 or E2
∧
M O DE BL
(2) Press (do not release your finger), then, press , the display will show
the program version.
MODE U N I TS
(3) Press then, press , the display shows the internal value.
(4) Then, you reset the linearity, specification and temperature setting calibration
steps by steps.
- 15 -
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
In case the internal value is not at 100,000 ~ 600,000 range, load cell is
damaged, you have to replace a new one.
(I). Please refer to the electric diagram of A/D board, then, use a voltage-meter
to measure the voltage parts of the A/D board if they are at +5v or –5v normal
condition.
(II). To use a clipper to clamp the 4th and 5th pins of IC24C04 EEPROM (no
release the clipper from the pins) and press“ ”key to turn on the power.
Before the display is counting backward to 0, please release the clipper not to
touch the 4th and 5th pins of IC24C04, the display will appear the program version
(III) If the internal value is at its range, then, please refer to the calibration
procedures of E1 or E2 solution to solve the problem(to reset the linearity,
specification and temperature setting calibration).
(IV) If the display is not appearing the internal value, then, turn off the machine
and do the below procedures :
b) To remove the 4th and 5th pins of load cell cable, not connected, then, use a
voltage-meter to measure the voltage of the 4th and 5th pins if they are at 0mv ±
2% (it’d better at 0mv standard).
- 16 -
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
If the voltage of the 4th and 5th pins are normally that means the load cell is not
faulty, works ok.
C) Then, put a mass on the scale to check the load cell output value.
l Note: To check the load cell output value, you have to check the capacity
of load cell used for this scale, the weight mass you put on scale must be based
on the full capacity of load cell.
For example: BH-3kg scales, we use 3kg-load cell to make it. Then,
you put 3kg weight; the output value of load cell is 10mv.
If you put 1/3 weight (1kg) on the scale, then, the output value of load
cell is at 3mv.
e) Press “ ”key to turn on the power and reset the linearity, specification and
Then, turn off the machine and turn the power on again to check if the scale
works in normally operation (no displaying E5).
- 17 -
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
CALIBRATE
(1). Press key to turn the power off. And then, press key to turn the
machine on again., the display shows 0.00
∧
M O DE BL
(2). Hold key and press key, the display will show the program
version no.
M O DE U N I TS
(3). Then, press and keys, the display shows the internal value.
When the reading of the internal value is gradually increasing up. That is the
normal condition.
In case the reading of the internal value is shifting, unstable. The problem may
be in load cell. In this case, please open the housing, then, you change another
good load cell for doing the test.
(4) After changing the load cell, put the platter on the scale. The display would
be showing the internal value. If the internal value reading is stable. That is sure
that the former load cell is faulty, can not be used, you have to change the load
cell for the replacement.
(5) In case that you already change another load cell but the reading of the
internal value on the display is unstable, shifting, you have to check the
PCB(A/D board). There maybe the dirty or something on it. You have to clean
the board and dry the PCB with a dryer.
(6) To use a volt-meter to measure the 1st pin and 2nd pin of CN2 location(on the
A/D board) if their voltage is at +5V? If the load cell power voltage is in
- 18 -
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
normal stable condition +5V. Then, you continue to check the 9th pin of the
IC7190 at U3 if its voltage is at 2.5V.
If its voltage is under 2.5V, then, you have to change a new IC7190 at U3.
After changing the IC. You test the scale again if the scale can work stable. If it
is stable.
press key about 3 seconds, the display will show the “Calibrated weight
value”, please accord with the weight to put the same weight mass on the scale.
When the display is stable, the scale will automatically begin a self-testing
sequence and take the mass away while the display is counting backward from 9
∼ 0.
If you have tried to test the scale as the above methods but the scale reading is
still unstable. In this situation, you have to send back the defective scale to the
factory for malfunction.
1) The environment where you use the scale is very important. Be sure that you
operate the scale in a good environment, not in a windy, air conditioning
surrounding which may affect the scale unstable.
2) When turn on the machine, make sure that the scale is in stable that is the
stable sign ° indication must appear on the display, you can calibrate the scale(if
the stable sign ° indication in not appear on the display, scale can not calibrate,
that means the scale can not be stable due to the unstable environment).
- 19 -
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc
3) Hold M O DE
and press U N I TS
keys, the display shows “ CAL “ and
enter to the FUNCTION SELECTION.
4) Press P CS
to P.S. 7 The Range of Stability for filter setting. Press
∧
BL
key to select F l 105 setting. Then, press key to confirm. The
scale will return to 0.
5) Place the full capacity of weight on the scale to check if it can shows the
correct weight(if the error display weight is showing more than 50g or
less than 50g of the actual mass weight, that is allowed to continue to
do the weight calibration).
press key about 3 seconds, the display will show the “Calibrated
weight value”, please accord with the weight to put the same weight mass
on the scale. When the display is stable, the scale will automatically
begin a self-testing sequence and take the mass away while the display is
counting backward from 9 ∼ 0.
˙Remarks : In case you have tried to test the scale as the above method
but it still has the problem, can not work. In such condition, you have to
send back the scale to the factory for malfunction.
- 20 -
C:\Documents and Settings\jobecca\My Documents\Auto calibration\eqypt\98019 BH service manual.doc