You are on page 1of 66

Issue 6

05/01/11

Part Number: 63183

GMI welcomes comments on all our publications.


Your comments can be of great value in helping us to improve our customer publications.
Please send any comments that you have to our Sales Department at GMI. Contact details are
provided on the back cover of this handbook.
Instrument Service / Repair contact details are also provided on the back cover of this
handbook.

Copyright © Gas Measurement Instruments Ltd 2007


COPYRIGHT

COPYRIGHT
This User Handbook is copyright of Gas Measurement Instruments Ltd (GMI) and the information contained
within, is for use only with instruments.
Reproduction, in whole or in part, including utilisation in machines capable of reproduction or retrieval without
written permission of Gas Measurement Instruments Ltd is prohibited. Reverse engineering is not permitted.

LIABILITY
Every care has been taken in the preparation of this document, but GMI Ltd do not accept any responsibility
for errors or omissions and their consequences. Information in this document is subject to change without
notice. This document does not constitute a specification or basis for a contract. Your statutory
st rights under
law are not affected.

MODIFICATION NOTICES
GMI aim to notify customers of relevant changes in the product operation and maintain this manual up to
date. In view of the policy of continuous product improvement there may be operational differences
between the latest product and this manual.

i
OXYGEN / TOXIC - USER HANDBOOK

This Handbook is an important part of the product. Please note the following points:

• It should be kept with the instrument for the life of the product.
• Amendments should be attached.
• This Handbook should be passed on to any subsequent owner/user of the instrument.
• Although every care is taken in the preparation of this Handbook it does not constitute a
specification for the instrument.

SOFTWARE
Software supplied on EPROM or similar device for use in a particular product, may only be used in that
product and may not be copied without the written permission of Gas Measurement Instruments Ltd.
Reproduction or disassembly of such embodied programmes or algorithms is prohibited. Ownership of such
software is not transferable and GMI Ltd does not warrant that the operation of the software will be error free
or that the software will meet the customer’s requirements.

DISPOSAL ADVICE
When no longer in use, dispose of the instrument carefully and with respect for the environment.
GMI will dispose of the instrument without charge if returned to the factory.

ii
COPYRIGHT

SAFETY
• The instrument should be periodically tested, in a safe area, using a known gas concentration.
• Only GMI replacement parts should be used.
• If the instrument detects gas, follow your own organisation’s procedures and operational guidelines.
• instruments are certified as:
ATEX II 1 G Ex ia IIC T4 IECEx Ga Ex ia IIC T4 (Ta = -20oC to + 50 oC)

UL 913 Class I, Div 1 Groups A, B, C and D

MED, for oxygen instruments only. Refer to MED Declaration of Conformity for details
• This equipment is designed and manufactured to protect against other hazards as defined in
paragraph 1.2.7 of Annex II of the ATEX Directive 94/9/EC
WARNING 1: DO NOT USE INSTRUMENT IN OXYGEN ENRICHED ATMOSPHERES.
WARNING 2: SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
WARNING 3: TO PREVENT IGNITION OF FLAMMABLE OR COMBUSTIBLE ATMOSPHERES,
DISCONNECT POWER BEFORE SERVICING.
Any right of claim relating to product liability or consequential damage to any third party against GMI is
removed if the warnings are not observed.
iii
OXYGEN / TOXIC - USER HANDBOOK

AREAS OF USE
Do not use instrument in potentially hazardous atmospheres containing greater than 21% Oxygen. The
enclosure material is polycarbonate / ABS and must not be exposed to environments which are liable to
result in mechanical or thermal degradation or to damage caused by contact with aggressive substances.
Additional protection may be required in environments where the instrument enclosure is liable to damage.

STORAGE, HANDLING AND TRANSIT


The instrument is designed to handle harsh environments and is sealed to *IP54. If not subjected to misuse
or malicious damage, the instrument will provide several years of reliable service.
The instrument contains electrochemical sensors with a minimum life of two (2) years. The instrument
should only be stored for a maximum of one (1) month from the date of manufacture (as indicated on
instrument batch label). Storage in excess of this would affect the instrument warranty. The sensor contains
potentially corrosive liquid and care should be taken when handling or disposing of the sensor, particularly
when a leak is suspected.
* Note: IP rating is not verified by UL.
WARRANTY
The instrument has a warranty against faulty goods or workmanship for the lifetime of the instrument.
Consumable and Mechanical parts are not included in this. These are covered under GMI standard warranty
conditions. For details, please contact GMI Ltd (UK).

iv
REVISION RECORD

Date Issue Description of Change


23/08/2007 1 New Handbook

09/01/2008 2 ‘Switch OFF’ sequence amended,


‘Safety’ WARNING statements added,
(incl. translations updates),
Latching & Non-Latching alarm clarification,
Storage instructions amended,
Bump Test Procedure included,
Accessories list revised.

26/01/2009 3b Revised to include effect of CN 4529 and CN 4670.


Storage instructions amended and ‘Spanish’
Quick Operating Instructions included.
v
OXYGEN / TOXIC - USER HANDBOOK

Date Issue Description of Change

15/09/2009 4 Revised to include effect of CN 4686 and CN 4752.

09/03/2010 5 Revised to include effect of CN 4878.

05/01/2011 6 Revised to include effect of CN 5047 and clarify


‘Note 2’ on page 2-1.

vi
CONTENTS

CONTENTS

COPYRIGHT ...................................................................................................................................................... i
LIABILITY ........................................................................................................................................................... i
MODIFICATION NOTICES ................................................................................................................................. i
SOFTWARE ....................................................................................................................................................... ii
DISPOSAL ADVICE ........................................................................................................................................... ii
SAFETY ............................................................................................................................................................ iii
AREAS OF USE .............................................................................................................................................. iv
STORAGE, HANDLING AND TRANSIT .......................................................................................................... iv
WARRANTY .................................................................................................................................................... iv

REVISION RECORD ................................................................................................... v

1. INTRODUCTION ................................................................................................. 1-1


1.1 INSTRUMENT PART NUMBER / GAS TYPE ......................................................................................... 1-2
1.2 INSTRUMENT FEATURES ..................................................................................................................... 1-2

vii
OXYGEN / TOXIC - USER HANDBOOK

1.3 INSTRUMENT IDENTIFICATION ........................................................................................................... 1-3


1.4 INSTRUMENT DISPLAY ........................................................................................................................ 1-3
1.5 CONSTRUCTION ................................................................................................................................... 1-3

2. OPERATION ........................................................................................................ 2-1


2.1 SWITCHING THE INSTRUMENT ON .................................................................................................... 2-2
2.2 NORMAL OPERATING DISPLAY ........................................................................................................... 2-6
2.2.1. Life Remaining .................................................................................................................................... 2-6
2.2.2. Battery Low / Exhausted ..................................................................................................................... 2-6
2.3 SWITCHING DISPLAY BACKLIGHT ON ............................................................................................... 2-7
2.4 VIEWING MAXIMUM / MINIMUM VALUES ............................................................................................. 2-7
2.4.1. Clearing the Max / Min Values ............................................................................................................. 2-8
2.5 REMOTE SAMPLING .............................................................................................................................. 2-8
2.6 SWITCHING THE INSTRUMENT OFF .................................................................................................. 2-9
2.7 OPERATOR MESSAGES / FAULT FLAGS ............................................................................................ 2-10
2.8 TEST ...................................................................................................................................................... 2-11

3. ALARMS .............................................................................................................. 3-1


3.1 ALARM DISPLAY .................................................................................................................................... 3-2

viii
CONTENTS

3.2 ACKNOWLEDGE ALARM(S) .................................................................................................................. 3-4


3.3 MUTE ALARM(S) .................................................................................................................................... 3-4

4. OPERATOR MAINTENANCE .............................................................................. 4-1


4.1 CLEANING .............................................................................................................................................. 4-1
4.2 FILTER REPLACEMENT ........................................................................................................................ 4-2

5. CALIBRATION ..................................................................................................... 5-1


5.1 CALIBRATION VALIDITY ....................................................................................................................... 5-2
5.2 BUMP TEST ............................................................................................................................................ 5-3
5.2.1. Automatic Bump Test .......................................................................................................................... 5-3
5.2.2. Manual Bump Test ............................................................................................................................... 5-4

6. ACCESSORIES ................................................................................................... 6-1

7. ADDITIONAL INFORMATION ............................................................................... 7-1


Training ........................................................................................................................................................... 7-1
World Wide Web ............................................................................................................................................ 7-1

ix
OXYGEN / TOXIC - USER HANDBOOK

A. TYPICAL OPERATING PARAMETERS ................................................................ A-1

B. QUICK OPERATING INSTRUCTIONS .............................................................. B-1


English ........................................................................................................................................................... B-2
Italiano (Italian) .............................................................................................................................................. B-4
Nederlands (Dutch) ........................................................................................................................................ B-6
Português (Portuguese) ................................................................................................................................. B-8
Polski (Polish) .............................................................................................................................................. B-10
Espanol (Spanish) ........................................................................................................................................ B-12

INDEX .......................................................................................................................... i

x
INTRODUCTION

1. INTRODUCTION
The GMI instrument combines quality,
ruggedness and advanced technology in a user
friendly, portable gas detector. Small and
lightweight, it is suitably certified to recognised
International Standards.
The is a personal safety device that warns
when toxic gas or oxygen deficiency exceeds the
alarm set-points. (Refer to Chapter 3 for factory set
defaults). Its high intensity audible and bright visual
and vibrating alarms provide early warning of
dangerous gas levels. It is the responsibility of the
user to respond properly to the alarms.
The instrument is operated via a single push button
providing the user with a simple to use gas detector.
Fig. 1-1 Instrument
1-1
OXYGEN / TOXIC - USER HANDBOOK

1.1 INSTRUMENT PART NUMBER / GAS TYPE


63002: 0 to 25% Oxygen (O2)
62403 / 63324: 0 to 100 ppm Hydrogen Sulphide (H2S)
62404 / 63424: 0 to 500 ppm Carbon Monoxide (CO)

1.2 INSTRUMENT FEATURES


The main features of the range of instruments are:
• Integral impact resistant housing.
• Single button user operation.
• Clear display with screen light.
• High intensity audible and visual alarms. Also includes a vibrating alarm.
• Instrument protected to *IP54 rating, making the instrument suitable for outdoor use.
* Note: IP rating not verified by UL.
• A comprehensive range of accessories.

1-2
INTRODUCTION

1.3 INSTRUMENT IDENTIFICATION


The has a date code (mm/yy) displayed on a label on the back face of the instrument. A label on
the front face displays gas type. The text colour is coded ‘Blue’ for Oxygen, or ‘Yellow’ for Toxic gas
types. The part number and corresponding gas type can be found in section 1.1 of this chapter.

1.4 INSTRUMENT DISPLAY


The display details the current gas readings and life remaining.

1.5 CONSTRUCTION
The instrument is housed in a tough, impact resistant moulded case. The instrument is sealed
to IP54. The instrument is designed with louver to protect against rain and to withstand physical
impact testing.

The instrument has two buttons. The RH button is bright yellow (with black print of standard ON

symbol). The LH button is the same colour as the green moulding.

A robust crocodile clip is provided on the back face for attaching instrument to user’s clothing.
1-3
OXYGEN / TOXIC - USER HANDBOOK

1-4
OPERATION

2. OPERATION
Before operation of the instrument, check that the instrument is clean and in good condition
Each time you use the instrument, perform the following operation:
1) Switch instrument ON in fresh air and check the life remaining.
2) Check there are no faults.
3) The instrument is zeroed (automatically in fresh air), i.e. reading of 0 (20.9 for oxygen)
4) Attach optional accessories, as required.
5) Perform self-test by a press and hold of the green button.
6) Switch the instrument OFF, in fresh air, after use.
Note 1: You can leave the Instrument ON permanently if required. Switching OFF will reset the Max.
(Min.) value to 0 (20.9).
Note 2: A reset feature is available that allows a re-start of the instrument to overcome a stall in
operation. The instrument configuration is retained. To reset the instrument, press and hold the RH
button for 15 seconds, then press the ON button to re-start the instrument and begin the warm-up
routine as detailed in Paragraph 2.1.
2-1
OXYGEN / TOXIC - USER HANDBOOK

2.1 SWITCHING THE INSTRUMENT ON

Press and hold the RH button for one second to switch


the instrument on. The instrument begins its warm-up routine.
First, all segments are briefly shown, and the LED’s will
flash, the buzzer beeps, and the vibrator will vibrate.
An example of an initial screen display is shown in Fig 2.1
opposite.
During the warm-up cycle, a countdown timer is displayed Fig. 2-1 Initial Screen Display
at the bottom / centre of the screen

Next, the software version is displayed, as shown in example


Fig. 2-2 opposite.

Fig. 2-2 Software Version

2-2
OPERATION

The next screen displays the target gas.


The target gas in the example shown, Fig. 2-3, is Carbon
Monoxide.

Fig. 2-3 Target Gas

2-3
OXYGEN / TOXIC - USER HANDBOOK

Next, the different alarm level


values are shown. See Figures
opposite.
Note 1: If any particular alarm is
disabled, the screen will
show ‘dis.’ (disabled).
The ‘L’ and ‘NL’ symbols, at bottom
left hand and bottom right hand Fig. 2-4 Hi Alarm Level Fig. 2-5 HiHi Alarm Level
corner of display, indicate alarms
as either ‘Latching’ or ‘Non-
Latching and correspond with the
LH and RH buttons respectively.
Refer to Chapter 3 ‘ALARMS’ for
explanation of Latching and Non-
Latching alarms.

Fig. 2-6 STEL Alarm Level Fig. 2-7 LTEL Alarm Level

2-4
OPERATION

Finally, the instrument serial number is displayed with the


last four digits in large font size. In the example shown
opposite, the serial number is 100136:

Fig. 2-8 Instrument Serial Number

2-5
OXYGEN / TOXIC - USER HANDBOOK

2.2 NORMAL OPERATING DISPLAY


When warm-up is completed successfully, the normal operating display
is shown, as in example Fig. 2-9 opposite, with the gas value and the
remaining life of the instrument shown in the display.
2.2.1. Life Remaining
Initially the display will show the life remaining, at the bottom of the
screen. The time remaining (in months) reduces until
1 month remains, and is then displayed in days, as shown in Fig. 2-9 Normal Operating
Fig. 2-10 opposite. Display
Note: Remaining months / days may be reduced when
significant alarm exposures take place.
2.2.2. Battery Low / Exhausted
In shut-down mode, the instrument audible, visual and vibrating alarm
will activate once every two seconds. This will continue until either:
The user presses any button to permanently shut-down the instrument,
or,
After one hour has elapsed, the instrument will shut down and not
restart. Fig. 2-10 Life Remaining
2-6
OPERATION

2.3 SWITCHING DISPLAY BACKLIGHT ON


The display screen light can be manually switched on when surveying in poor lighting conditions.
Press the RH button once to switch the screen light on. It remains on for three seconds after the last
button press and then automatically switches off.
Note: Back-light will reduce life remaining if it is on repeatedly.

2.4 VIEWING MAXIMUM / MINIMUM VALUES


To view the maximum reading since switch ON, press the RH button
once to switch ON the back-light.
Press it again to view the maximum value since switch ON as shown
in example Fig. 2-11 opposite. (press again for minimum O2)
The reading returns to normal if no button is pressed for three
seconds.
The display screen light stays on during the viewing of the maximum
Fig. 2-11 Max / Min Values
(minimum) and switches off automatically after three seconds
following the last button press.
2-7
OXYGEN / TOXIC - USER HANDBOOK

2.4.1. Clearing the Max / Min Values


To clear the maximum (and minimum) reading, simply press and hold the RH button while viewing
the maximum (or minimum). The reading is reset to zero (20.9% for O2).

2.5 REMOTE SAMPLING

Warning (Hand Aspirator): The hand aspirator can be used for


indicative sampling, but it must be noted that when using a
hand aspirator, a reading error in the region of +/- 20% is possible.
In addition, the hand aspirator must only be used with up to 10
metres of sample line and the sample time is extended.

Remote sensing is carried out with the hand aspirator for non-
reactive gases, using the sample connector and sample tubing
supplied. Example shown in Fig. 2-12 opposite.

Fig. 2-12 Remote Sampling Set-up

2-8
OPERATION

2.6 SWITCHING THE INSTRUMENT OFF

Press and hold both the LH and RH buttons


simultaneously to switch the instrument OFF.
The instrument screen will now start a countdown from three
(3) to OFF as shown in Fig. 2-13 opposite. The buttons must
be pressed until the display shows OFF, then release the
buttons.

Fig. 2-13 OFF Sequence

2-9
OXYGEN / TOXIC - USER HANDBOOK

2.7 OPERATOR MESSAGES / FAULT FLAGS


• If the sensor or anything else fails, the audible (short beep)
and visual (red LED) alarm activate and the instrument display
will show four (4) E’s as shown in Fig. 2-14 opposite.
Switch the instrument OFF then ON again in fresh air.
If the fault persists, return the instrument for service.
• If, at the end of the warm-up cycle, the instrument audible
alarm ‘beeps’ and the visual alarm flashes, this is an indication
that the instrument has been unable to zero the sensor.
Fig. 2-14 Fault Indication
You can either mute the alarm,
or
Go to an area where there is fresh air, wait one (1) minute then
switch the instrument OFF then ON again.
If the fault persists, return the instrument for service.

2-10
OPERATION

2.8 TEST

A self-test can easily be performed, by simply pressing and holding the LH button .

Briefly, all elements of the display will be shown. The buzzer will beep, the LED’s will flash and the
instrument will vibrate. After this, the instrument will return to its normal operation screen.
Note: A self-test can only be performed from the normal operation screen with a safe gas concentration.

2-11
OXYGEN / TOXIC - USER HANDBOOK

2-12
ALARMS

3. ALARMS

When the instrument detects an alarm set-point has been reached, the audible and visual alarm
will be activated to alert the user. The instrument will also vibrate in alarm condition.
A Non-Latching alarm (default) will reset automatically once the reading returns within the pre-set
alarm limits. The alarm will not clear until the reading has returned within the limits for a period of 20
seconds. This prevents the alarm from activating / de-activating regularly when the reading is hovering
around the alarm set-point. Alternatively, the alarm can be muted.
A Latching alarm (configurable option) will stay on until the user resets by a press and hold of the
RH button once the gas reading has returned within the pre-set alarm limits.

By default, all instrument alarms are non-latching.


Note 1: Alarm levels are set in accordance with current standards.
Note 2: Warning alarms do not vibrate.

3-1
OXYGEN / TOXIC - USER HANDBOOK

3.1 ALARM DISPLAY


If an alarm occurs, the type of alarm (e.g. LoLo oxygen) is
indicated at the top of the screen with the current reading on
the display, as shown in Fig. 3-1 opposite.
To switch on the back-light during an alarm, simply press the
RH button once. It will automatically switch off after three
seconds.
Note: When a TWA alarm is active, the actual value is Fig. 3-1 Alarm Display
displayed and the alarm type (STEL or LTEL) flashes.

3-2
ALARMS

T.ex Gas
Hi HiHi Lo LoLo S TE L LTEL
Part No. Type

63002 O2 N/A 23.5% Disabled 19.5% N/A N/A

63324 10 ppm 5 ppm


H2S 10 ppm 20 ppm N/A N/A
62403 20 ppm 10 ppm

63424 30 ppm 200 ppm 30 ppm


CO N/A N/A 200 ppm
62404 35 ppm 50 ppm 35 ppm

Fig. 3.2 Alarms Table


Note: By default, all instrument alarms are non-latching.

3-3
OXYGEN / TOXIC - USER HANDBOOK

3.2 ACKNOWLEDGE ALARM(S)


To acknowledge, i.e. switch off, a latching alarm (configurable option), press and hold the RH button
when the gas reading has returned within the pre-set alarm limits.

3.3 MUTE ALARM(S)

To mute (i.e. silence) a non-latching alarm, press and hold the RH button when the alarm is
sounding. The audible / visual alarm will be silenced / extinguished. If the gas reading has returned
within the pre-set alarm limit, the alarm type will disappear from the display automatically.
If the non-latching alarm limit is exceeded again, the audible and visual alarm will be re-activated.
TWA alarms (STEL and LTEL) are non-latching and when the alarm is muted, the audible visual and
vibration will cease. The alarm type will flash until the alarm has cleared, then the alarm indication
at the top of the display (STEL or LTEL) will disappear.

3-4
OPERATOR MAINTENANCE

4. OPERATOR MAINTENANCE

4.1 CLEANING

CAUTION: Do not use polishes containing silicon or solvent to clean the instrument as these may
damage the sensor. Do not use abrasive materials or strong volatile chemical solutions
as these could damage the impact resistant casing.

The outer, impact resistant casing of the instrument may be cleaned using a non-abrasive moist
cloth. Rub the cloth over the outer casing to remove any dirt and grime.
In extreme cases, a mild soap solution may be used with a non-abrasive cloth to remove more
stubborn marks.

4-1
OXYGEN / TOXIC - USER HANDBOOK

4.2 FILTER REPLACEMENT


The instrument has a hydrophobic type filter protecting the instrument sensors. The filter should be
inspected for signs of contamination and / or damage when servicing the instrument.
This should only be carried out by trained personnel.

4-2
CALIBRATION

5. CALIBRATION

The instrument has been calibrated for a particular gas. Where any doubt exists the product should be
returned to GMI or an authorised distributor for calibration.

WARNING: The instrument must be calibrated and configured by authorised personnel only.

Two methods of calibration are possible:


• Field Calibration. See Configuration & Field Calibration Handbook
(Part No. 63187) for further details.

• The GMI Automatic Calibration System (Part No. 63122) provides controlled delivery of
gas, allowing you to calibrate in a controlled manner and maintain a record of calibration
results on a PC.

5-1
OXYGEN / TOXIC - USER HANDBOOK

Note: The detailed calibration methods, consisting of both hardware and software, are manufactured
by GMI. For more detail contact GMI or an authorised distributor.

5.1 CALIBRATION VALIDITY


Calibration validity is the responsibility of the user. Under normal operating conditions a 12 month
period can be expected. This is no guarantee, however, as the precise application of the product is
unknown to GMI. Individual codes of practice may dictate shorter periods.
Regular checking, by using bump-test procedure, establishes a pattern of reliability and enables the
calibration check period to be modified in line with operational experience. The higher the risk, the
more frequently calibration should be checked. The bump-test procedure details the application of
gas, either manually, or automatically using the Bump-Test Station.
Refer to section 5.2 for Bump Test details.

5-2
CALIBRATION

5.2 BUMP TEST


An automatic or manual Bump Test can be performed on the instrument to test the instrument’s
sensor response and alarms.

5.2.1 Automatic Bump Test


The GMI automatic Bump Test Station (GMI Part No. 63121) provides a safe and convenient method
of automatically bump testing the instrument. Up to nine (9) automatic bump test stations can be
linked in a daisy-chain fashion using accessory lead(s).
Refer to procedure detailed in Instruction Sheet ‘Bump Test Station - Installation & Use’ (GMI Part No.
63188) and by using one of the following Bump Test / Auto Calibration kits:
For H2S, use kit - GMI Part No. 63130,
For CO, use kit - GMI Part No. 63131,
For O2, use kit - GMI Part No. 63132.
Note: The Bump Test / Auto Calibration kit consists of gas cylinder, regulator, and tubing.

5-3
OXYGEN / TOXIC - USER HANDBOOK

5.2.2 Manual Bump Test


The instrument can be manually bump tested by following the procedure below and using one of the
following Field Calibration / Bump Test kits:
For H2S, use kit - GMI Part No. 63104,
For CO, use kit - GMI Part No. 63105,
For O2, use kit - GMI Part No. 63106.
Note: The Field Calibration / Bump Test kit consists of gas cylinder, regulator, and tubing complete
with connector.

1. Press and hold the LH button until the screen indicates “tESt”.
2. Apply the gas, at 250 ml/min, via the gas inlet connector on the front of the instrument, just
below the diffusion louver.
3. Apply the gas until the instrument display indicates “PASS”.
The instrument will then revert to normal operation and thus will be ready for use.
Note: If the manual bump-test is unsuccessful, the display will indicate “FAIL”.
Follow the above procedure again after allowing time to diffuse the remaining gas from the
instrument, or get the instrument re-calibrated by trained personnel.
5-4
ACCESSORIES

6. ACCESSORIES

A range of accessories is available for the following list of series instruments:


Instrument Sensor Installed
Part No.
63002 0 to 25% Oxygen (O2)
62403 / 63324 0 to 100 ppm Hydrogen Sulphide (H2S)
62404 / 63424 0 to 500 ppm Carbon Monoxide (CO)
Accessory Description
Part No.
63101 Hardhat Clip
66485 Hydrophobic Filter Assy. (in-line)
63103 Hand Aspirator c/w 3m Sample Line
66545 Ball Float Assy.

6-1
OXYGEN / TOXIC - USER HANDBOOK

Accessory Description
Part No.
66118 Sample Line (per metre)
67071 Gas Bottle Clip (for wall mounting)
63104 H2S Field Calibration Kit (gas cylinder, regulator, tubing c/w connector)
63105 CO Field Calibration Kit (gas cylinder, regulator, tubing c/w connector)
63106 O2 Field Bump Test Kit Kit (gas cylinder, regulator, tubing c/w connector)
63121 Bump Test Station c/w Power Supply
13163 5-way Lead (connects up to 4 additional bump test stations)
12988 Vehicle Power Supply (use with 63121)
63122 Auto Calibration Station
63130 H2S Bump Test Station / Auto Calibration Station Kit (gas cylinder, regulator, tubing)
63131 CO Bump Test Station / Auto Calibration Station Kit (gas cylinder, regulator, tubing)
63132 O2 Bump Test Station Kit (gas cylinder, regulator, tubing)
63120 Data Downloading Setup Pack (CD-ROM, interface adaptor, user instructions)
For a comprehensive list of accessories and calibration gases, contact your local Distributor or
alternatively, GMI Ltd.
6-2
ADDITIONAL INFORMATION

7. ADDITIONAL INFORMATION

Training
Training courses are available on all GMI products. Contact GMI Marketing Department for further
details:
Tel: +44 (0) 141 812 3211
Fax: +44 (0) 141 812 7820
e-mail: sales@gmiuk.com

World Wide Web


Visit GMI web site at www.gmiuk.com

7-1
OXYGEN / TOXIC - USER HANDBOOK

7-2
TYPICAL OPERATING PARAMETERS

A. TYPICAL OPERATING PARAMETERS

Typical operating parameters are as follows:


Instrument Gas Type Range Resolution
Part No.

63002 Oxygen (O2) 0 to 25% 0.1%


62403 / 63324 Hydrogen Sulphide (H2S) 0 to 100 ppm 1 ppm
62404 / 63424 Carbon Monoxide (CO) 0 to 500 ppm 1 ppm

Dimensions (excl. clip): 85mm (H) x 55mm (W) x 29mm (D) ; 3.3in (H) x 2.2in (W) x 1.1in (D)

Weight (incl. battery): 76g (toxic) / 86g (oxygen) ; 2.7oz (toxic) / 3.0oz (oxygen)

A-1
OXYGEN / TOXIC - USER HANDBOOK

Temperature Limits: -20oC to 50oC (-4oF to 122oF)


Humidity: 0 – 98% R.H. non-condensing
Construction / Protection Rating: Moulded polycarbonate / ABS case protected to *IP54.
* Note: IP rating not verified by UL.
Display: LCD with backlighting (Backlighting on demand).
Data Sampling System: Sampling rate - 1 minute interval. Stores (1440) logs over a period of
24 hours operation / then re-writes from longest stored log.
Previous 10 calibration results.
Previous 50 bump-tests.
ON / OFF session logs.

Power Source: Lithium Battery


Battery life - up to 24 months depending on sensor type and usage.
Audible Alarm: 95 dB @ 10 cm. (4 in.) ; 85 dB @ 30 cm. (12 in.)
Visual Alarm: 2 Red LED’s

A-2
QUICK OPERATING INSTRUCTIONS

B. QUICK OPERATING INSTRUCTIONS

The following multi-language instructions provide the user with a quick operation guide for the . .

(Single Gas) instrument.

Each language and page reference is as follows:


• English - pages B-2 to B-3
• Italiano (Italian) - pages B-4 to B-5
• Nederlands (Dutch) - pages B-6 to B-7
• Português (Portuguese) - pages B-8 to B-9
• Polski (Polish) - pages B-10 to B-11
• Español (Spanish) - pages B-12 to B-13
B-1
SAFETY AREAS OF USE
• The instrument must be regularly tested in a safe area. Do not use the instrument in potentially hazardous atmospheres containing
• Never expose instrument to extreme heat. greater than 21% Oxygen.
• Only GMI replacement parts should be used. STORAGE, HANDLING AND TRANSIT
• If the instrument detects gas, follow your own organisation’s The instrument is designed to handle harsh environments and is sealed to
procedures and operational guidelines. *IP54. If not subject to misuse or malicious damage, the instrument will provide
several years of reliable service. The instrument contains an electrochemical
• Gas can be dangerous and care should always be taken in its use. sensor with a minimum life of 2 years. The sensor contains potentially corrosive
• instruments are certified as: ATEX II 1 G Ex ia IIC T4 liquid and care should be taken when handling or disposing of the sensor,
o o
particularly when a leak is suspected. * Note: IP rating not verified by UL.
IECEx Ga Ex ia IIC T4 (Ta = -20 C to + 50 C)
1. Check that the instrument has no obvious faults.

CHECKLIST
UL 913 Class I , Div 1 Groups A, B, C and D. 2. Check accessories.
3. Read and understand handbook before use.
MED, for oxygen instruments only. Refer to MED Declaration 4. Switch ON.
of Conformity for details. 5. Check life remaining.
• This equipment is designed and manufactured to protect against 6. Check ’ZERO’ in fresh air.
other hazards as defined in paragraph 1.2.7 of Annex II of the
ATEX Directive 94/9/EC. OPERATOR MESSAGE / FAULT FLAG
If the sensor or anything else fails, the audible (short beep) and visual (red
WARNING 1: Do not use instrument in oxygen enriched atmospheres.
LED) alarm activate and the instrument will show four (4) E’s in the display.
WARNING 2: Substitution of components may impair intrinsic safety.
WARNING 3: To prevent ignition of flammable or combustible atmospheres, SELF TEST
disconnect power before servicing. All elements of the display will be shown. The buzzer will beep, the LED’s will
flash and the instrument will vibrate. After this, the instrument will return to its
Any right of claim relating to product liability or consequential damage to any normal operation screen. Note: A self-test cannot be performed while an
third party against GMI is removed if the warnings are not observed. alarm is active.

B-2
QUICK OPERATING INSTRUCTIONS
English
OPERATION
The instrument performs warm-up sequence with sensor
Press and hold to switch the instrument ON check and auto-zero. When PPM or % reading appears,
the instrument is ready for use.
Press to switch the backlight ON.

Press again to view MAX (or MIN for Oxygen instrument).

Press and hold to clear MAX (MIN).

Press and hold to acknowledge latching or fault alarm, or mute non-latching.

Press and hold to perform self-test (See explanation / opposite).

Press and hold and , for three (3) seconds, to switch the instrument OFF.

B-3
SICUREZZA AREA DI UTILIZZO
• Lo strumento deve essere regolarmente testato in una zona sicura. Non utilizzate lo strumento in atmosfere potenzialmente pericolose con
• Mai esporre lo strumento a temperature molto alte. concentrazioni di ossigeno maggiori del 21%.
• Utilizzare solo ricambi originali GMI. IMMAGAZZINARE, MANEGGIARE E TRASPORTARE
• Se lo strumento rileva la presenza di gas, seguite le procedure Lo strumento è progettato per essere maneggiato in ambienti duri, aspri. Gli
operative e le linee guida della vostra azienda. elementi sensibili (sensori) sono sigillati a norma IP54 e il resto dello strumento
• I gas possono essere pericolosi e si deve sempre fare molta risponde alle caratteristiche IP54. Se non viene utilizzato in maniera impropria
attenzione durante il loro utilizzo. o danneggiato intenzionalmente, lo strumento è in grado di funzionare per
• Gli strumenti sono certificati come: ATEX II 1 G Ex ia IIC T4 diversi anni. Lo strumento contiene un sensore elettrochimico con una vita
o o
minima di 2 anni. Il sensore contiene un liquido potenzialmente corrosivo; è
IECEx Ga Ex ia IIC T4 (Ta = -20 C to + 50 C) quindi necessario maneggiare e smaltire con attenzione il sensore, in particolar
modo se si sospetta una perdita di tale liquido.
UL 913 Classe I , Div 1 Gruppi A, B, C e D.
1. Verificate che lo strumento non abbia danni o guasti evidenti.
2. Verificate gli accessori.

CHECKLIST
MED, Solo per strumenti con ossigeno. Per dettagli fare 3. Leggete e verificate di aver capito in ogni sua parte il manuale
riferimento alla dichiarazione di conformità MED. di istruzioni.
• Questa apparecchiatura è progettata e costruita per proteggere 4. Accendete.
contro altri pericoli come definito nel paragrafo 1.2.7 di Annex II 5. Verificate la vita restante.
della direttiva ATEX 94/9/EC 6. Verificate lo ‘ZERO’ in aria pulita.
ATTENZIONE: Non usate lo strumento in atmosfere arricchite di ossigeno.
ATTENZIONE: La sostituzione di componenti potrebbe invalidare la sicurezza
MESSAGGIO OPERATORE/INDICAZIONE DI GUASTO
intrinseca. Se il sensore o qualche altra parte non funziona, gli allarmi sonoro (bip) e visivo
ATTENZIONE: Per prevenire l’accensione di atmosfere infiammabili o (LED rosso) si attivano e vengono visualizzate quattro (4) E sul display.
combustibili, scollegate l’alimentazione prima procedere alla manutenzione. AUTODIAGNOSI
Ogni diritto di rivendicazione di responsabilità della GMI per il prodotto o di Durante l’autodiagnosi vengono mostrati tutti i segmenti del display. Il cicalino
conseguenti danni a terzi, è annullato se gli avvertimenti non vengono emette un bip, i LED lampeggiano e lo strumento vibra. Dopo questa sequenza,
rispettati. lo strumento torna alla schermata operativa normale.
Nota: l’autodiagnosi non può essere effettuata quando c’è un allarme attivo.
B-4
QUICK OPERATING INSTRUCTIONS
Italiano
UTILIZZO
Lo strumento effettua la sequenza di riscaldamento con azzeramento e
Premere e mantenere il tasto per autodiagnosi del sensore. Quando la lettura di PPM o % appare, lo strumento
accendere lo strumento. è pronto per essere utilizzato.

Premere per accendere la retroilluminazione.

Premere ancora per vedere il valore MAX (o MIN per gli strumenti con Ossigeno).

Premere e mantenere per azzerare il valore MAX (MIN).

Premere e mantenere per confermare la visualizzazione degli allarmi che rientrano


automaticamente o gli allarmi di guasto, o per spegnere gli allarmi non si rientrano
automaticamente.

Premere e mantenere per effettuare l’autodiagnosi (vedi spiegazione).

Premere e mantenere e , per tre (3) secondi per spegnere lo strumento.

B-5
VEILIGHEID Alle claims aan GMI in relatie tot productaansprakelijkheid of daaruit voortvloeiende
• Het instrument moet regelmatig in een veilige omgeving schade aan een derde partij worden afgewezen indien waarschuwingen niet in acht
getest worden. genomen worden.
• Nimmer blootstellen aan extreme hitte. GEBRUIKSOMGEVING
• Gebruik alleen originele GMI-onderdelen. Gebruik het instrument niet in potentieel gevaarlijke omgevingen die meer dan
• Wanneer het instrument gas detecteert, volg de 21% zuurstof bevatten.
procedures en richttlijnen van uw organisatie. OPSLAG, GEBRUIK EN TRANSPORT
• Gas kan gevaarlijk zijn en betracht daarom altijd de nodige Het instrument is ontworpen voor ruwe omgevingen. De sensor heeft een IP54
voorzichtigheid bij het gebruik. beveiliging, de rest van het instrument is IP54. Het instrument geeft betrouwbare
• Het instrument is als volgt gecertificeerd: dienst voor een paar jaar, mits het niet aan misbruik wordt blootgesteld of beschadigd
wordt. Het instrument bevat een electrochemische sensor met een minimum levensduur
ATEX II 1 G Ex ia IIC T4 van 2 jaar. De sensor bevat potentieel corrosieve vloeistof en voorzichtigheid is van
IECEx Ga Ex ia IIC T4 (Ta = -20oC to + 50oC) belang tijdens handeling en wegwerpen, zeker als een lek is vermoed.
UL 913 Klasse I , Div 1 Groepen A, B, C en D. 1. Check dat het instrument geen fouten heeft.

CHECKLIST
2. Check de accessoires.
Alleen voor zuurstof instrumenten. Voor details, zie 3. Lees en begrijp de handleiding alvorens gebruik.
de MED declaratie van overeenstemming. 4. Schakel aan.
• Dit instrument is ontworpen en gefabriceerd om tegen de 5. Check de resterende levensduur.
gevaren gedefinieerd in paragraaf 1.2.7 van Annex II van 6. Check ‘nulstand’ in (buiten)lucht.
de ATEX richtlijn 94/9/EC te Beschermen.
GEBRUIKERS BERICHT / FOUT AANDUIDING
WAARSCHUWING: Gebruik het instrument niet inzuurstof rijke Als de sensor of iets anders faalt, dan activeren het geluidsalarms (korte piep) and
omgevingen. visuele alarm (rode LED) en het instrument toont vier (4) E’s op het scherm.
WAARSCHUWING: Vervanging van onderdelen kan de intrinsieke
veiligheid nadelig beinvloeden ZELF-TEST
Alle elementen op het scherm worden getoond. De sirene zal luiden, de LED’s
WAARSCHUWING: Sluit de voeding af voor onderhoud, om
knipperen, en het instrument zal trillen. Hienra zal het instrument terugkeren naar
ontvlamming van brandbare atmosfeer te voorkomen.
het normale scherm.
Opmerking: Een zelf-test kan niet gedaan worden als een alarm actief is.
B-6
QUICK OPERATING INSTRUCTIONS
Nederlands
GEBRUIK
Het instrument start op met een sensor check en
Houdt ingedrukt om het instrument aan te zetten automatische nulstelling. Wanneer PPM of % verschijnt
is het instrument klaar voor gebruik.
Druk de in om het licht aan te doen.

Druk nogmaals in om de MAX waarde (MIN voor zuurstof instrument) te zien.

Houdt ingedrukt om de MAX waarde te wissen

Houdt ingedrukt om een blijvend alarm uit te schakelen, of een niet-blijvend


alarm te bevestigen.

Houdt ingedrukt om een zelf-test te doen (zie uitleg ommezijde).

Houdt en , ingedrukt, voor drie (3) seconden, om het instrument uit te


schakelen
B-7
SEGURANÇASAFETY AREAS DE UTILIZAÇÃO
• O instrumento deve ser verificado e testado regularmente numa área segura. Não utilizar o instrumento em atmosferas potencialmente perigosas
• Não exponha o instrumento a fontes de calor intenso. contendo níveis superiores a 21% de Oxigénio.
• Utilizar apenas peças de substituição, marca GMI. ARMAZENAMENTO, MANIPULAÇÃO E TRANSPORTE
• Caso o instrumento detecte gás, proceda de acordo com os procedimentos O instrumento foi desenvolvido para resistir ambientes rigorosos.
e orientações da sua própria organização. Os elementos mais sensíveis estão selados de acordo com IP54 e o
• Os gases podem ser perigosos e devem ser tomadas as devidas precauções restante com a IP54. Quando utilizado de acordo com as indicações,
no seu uso. este instrumento proporciona vários anos de serviço com fiabilidade.
• Instrumentos têm certificações: ATEX II 1 G Ex ia IIC T4 O instrumento contém um sensor electroquímico com um tempo de vida
útil mínimo de 2 anos. Este sensor contém líquido potencialmente
IECEx Ga Ex ia IIC T4 (Ta = -20oC to + 50oC) corrosivo pelo que devem ser tomadas precauções especiais na sua
manipulação e destruição, principalmente caso suspeite de fuga.
UL 913 Class I , Div 1 Grupos A, B, C e D.
1. Verifique se não existem erros evidentes.
2. Verifique os acessórios.

A VERIFICAR
Só para intrumentos de oxigénio. Para referência consultar Declaração
MED (Directiva de equipamentos marítimos) 3. Leia e assegure-se de que compreendeu o guia de
utilização antes de usar o instrumento.
• Este equipamento foi desenvolvido e fabricado para protecção contra outros 4. Ligue o instrumento ON.
perigos como indicado no parágrafo 1.2.7 do Anexo II da directiva ATEX 5. Verifique a autonomia restante.
94/9/EC. 6. Verifique ‘ZERO’ em atmosfera limpa.
AVISO: Não utilizar o equipamento em atmosferas ricas em oxigénio
AVISO: A substituição de componentes pode prejudicar a segurança intrínseca do MENSAGEM DE OPERADOR / SINAL DE ERROOPERATOR
aparelho. Em caso de erro do sensor ou outro, é emitido um sinal sonoro de curta
AVISO: Para evitar a ignição de atmosferas inflamáveis ou combustíveis, desligue da duração (beep) e um visual (LED vermelho), o instrumento encontra-se
corrente antes de efectuar qualquer serviço de manutenção ou reparação. em estado de alarme mostrando quatro (4) Es no visor.
Qualquer direito de reclamação em relação à fiabilidade do produto ou dano AUTO-TESTE
consequente a terceiros cessa caso os avisos não tenham sido seguidos bem como Todos os elementos do visor são mostrados. É emitido um sinal sonoro,
as indicações. os LEDS piscam e o instrumento vibra. Após este auto-teste o instrumento
regressa ao estado operacional. Nota: O auto-teste não pode ser
B-8 realizado enquanto estiver activo um sinal de alarme.
QUICK OPERATING INSTRUCTIONS
Português
UTILIZAÇÃO
O instrumento inicia um ciclo de aquecimento com
Mantenha pressionada a tecla para ligar verificação do sensor e auto-zero. Quando valores PPM
o instrumento ou % surgirem, o instrumento está pronto a ser utilizado.

Pressione para ligar a luz do visor.

Pressione novamente para visualizar MAX (ou MIN para o de Oxigénio).

Pressione e mantenha pressionada a tecla para limpar MAX (MIN).

Pressione e mantenha pressionada a tecla para tomada de conhecimento dos


alarmes.
Pressione e mantenha pressionada a tecla para realização de auto-teste
(Ver explicação / verso).

Pressione e mantenha pressionada a tecla e , durante três (3) segundos, para desligar o instrumento
(OFF).
B-9
BEZPIECZEÑSTWO ZASTOSOAWNIE
• Detektor musi byc regularnie sprawdzany, sprawdzenie Nie nalezy stosowac detektora w atmosferze zawieracej wiêcej nie 21% tlenu.
nalezy przeprowadzac w strefie bezpiecznej.
• Nie nalezy wystawiac urzadzenia na dzialanie wysokich PRZECHOWYWANIE I STOSOWANIE
temperatur. Detektor zostal zaprojektowany do pracy w ciêzkich warunkach srodowiska. Celki
• Nalez stosowac tylko oryginalne czêsci zamienne firmy GMI. pomiarowe posiadaja stopieñ ochrony IP 54 a pozostale elementy detektora IP54.
• W przypadku gdy urzadzenie wykryje gaz nalezy postêpowac Prawidlowo uzytkowany detektor powinien pracowac kilkanascie lat. Detektor
wedlug zasad ustalonych dla danego zakladu lub instalacji. posiada celkê elektrochemiczna o minimalnym czasie pracy 2 lat. Nalezy ostroznie
• Detektor posiada nastêpujace certyfikaty: obchodzic sie z czujnikiem elektrochemicznym gdyz zawiera on korozyjny roztwór.
ATEX II 1 G Ex ia IIC T4 1. Sprawdzic czy nie ma zadnych zewnêtrznych uszkodzeñ.
IECEx Ga Ex ia IIC T4 (Ta = -20oC to + 50oC)

TESTOWA
2. Sprawdzic akcesoria.

LISTA
3. Przeczytac instrukcjê obslugi.
UL 913 Class I , Div 1 Groups A, B, C, D. 4. Wlaczyc detektor.
MED, Dotyczy tylko detektorow tlenu. Szczegoly 5. Sprawdzic stan baterii.
6. Sprawdzic “ZERO” na swiezym powietrzu.
umieszczone sa w deklaracji MED.
• Detektor ten zostal zaprojektowany i zbudowany aby chronic
przed zagrozeniami okreœlonymi w paragrafie 1.2.7 ankesu II SYGNALY AWARYJNE
dyrektywy ATEX 94/9/EC. W przypadku awarii czujnika lub innego elementu detektora uruchomiony zostanie
WARNING 1: Urzadzenie nie moze byc stosowane w atmosferze alarm dzwiêkowy(krótki sygnal) oraz swietlny(czerwona dioda LED) a na ekranie
wzbogaconej w tlen. wyswietlone zostana cztery litery “E”.
WARNING 2: Podmiana któregokolwiek z elementów moze uniewaznic SELF TEST
certyfikacjê ATEX. Na ekranie wyswietlone zostana wszystkie elementy, uruchomi siê sygnal dzwiêkowy
WARNING 3: Przed przystapieniem do prac serwisowych nalezy odlaczyc oraz zaczna migac diody LED i wlaczony zostanie alarm wibracyjny. Po
zasilanie. przeprowadzeniu testu detektor wróci do normalnego trybu pracy.
Firma GMI nie ponosi odpowiedzialnosci za szkody lub wypadki jesli nie
UWAGA: Test ten nie moze byc przeprowadzony jesli aktywny jest jakikolwiek alarm.
sa przestrzegane komunikaty i sygnaly ostrzegawcze.
B-10
QUICK OPERATING INSTRUCTIONS
Polski
FUNKCJONOWANIE/PRACA Detektor przechodzi sekwencjê uruchamiania obejmujaca
W celu wlaczenia detektora nalezy nacisnac i sprawdzenie stanu czujnika oraz auto zerowanie. W
momencie pojawienia siê oznaczenie % lub PPM detektor
przytrzymac klawisz jest gotowy do pracy.

W celu wlaczenia podœwietlenia ekranu nalezy nacisnac klawisz .

Ponowne nacisniêcie klawisza pozwoli na obejrzenie wartosci MAX


zarejestrowanego stêzenia (MIN dla tlenu).
W celu skasowania wartosci MAX(MIN) nalez nacisnac i przytrzymac klawisz

Nacisnac i przytrzymac klawisz w celu potwierdzenia alarmu zatrzaskowego,


sygnalu awarii lub alarmu niezatrzaskowego.
Nacisniêcie klawisza spowoduje uruchomieni sekwencji “self-test” .

W celu wylaczenie detektora nalezy nacisnac i przytrzymac przez 3 sek. klawisz i .


B-11
SEGURIDAD AREAS DE USO
• El equipo debe ser comprobado regularmente en una zona segura. No utilizar el equipo en ambientes potencialmente peligrosas que contengan
• No exponer el equipo a temperaturas extremas. más del 21% de oxígeno.
• Solo deben ser utilizados recambios GMI.
ALMACENAJE, MANIPULACIÓN Y TRÁNSITO
El equipo está diseñado para ser manipulado e ambientes ásperos. Los
• Si el equipo detecta gas, siga los procedimientos y normas funcionales de elementos de detección están precintados a *IP54 y el resto del equipo a
su propia empresa. *IP54. Si no está sujeto a un mal uso o a un daño malicioso, el equipo
• Un gas puede ser peligroso, así que tome precauciones a la hora de suministrará un servicio responsable durante varios años. El equipo contiene
utilizarlo. sensores electroquímicos con una vida mínima de dos años. El sensor contiene
un líquido potencialmente corrosivo y se debe tener cuidado cuando se
• El equipo está certificado como: ATEX II 1 G Ex ia IIC T4 manipule o se destruya el sensor, especialmente cuando se detecte una
IECEx Ga Ex ia IIC T4 (Ta = -20oC to + 50oC) fuga. * Nota: La clasificación IP no está verificada por UL.
UL 913 Clase I, Div 1 Grupos A, B, C y D. 1. Comprueba que el equipo no cometa fallos obvios.

CONTROL
2. Comprueba los accesorios.

LISTA DE
MED, sólo para equipos de oxígeno. Ir a Declaración de 3. Lee y entiende el manual antes de usar el equipo.
Conformidad MED para más detalles. 4. Enciéndelo.
• Este equipo ha sido diseñado y fabricado para proteger contra riesgos, 5. Comprueba el tiempo de vida restante.
tal y como se explica en el punto 1.2.7 del Anexo II de la Directriz de 6. Comprueba que en aire fresco los valores sean “CERO”.
ATEX 94/9/EC. MENSAJE DEL OPERADOR / SEÑALES DE FALLO
ADVERTENCIA 1: No usar el equipo en ambientes ricos en oxígeno. Si el sensor o cualquier otro componente del equipo falla, se activarán las
ADVERTENCIA 2: La sustitución de componentes puede perjudicar la alarmas acústica (pitido corto) y luminosa (LED rojo) y el equipo mostrará
seguridad intrínseca. cuatro letras “E” en la pantalla.
ADVERTENCIA 3: Para prevenir la ignición de ambientes inflamables o AUTO-PRUEBA
combustibles, desconecte la corriente antes de la revisión. Se mostrarán todos los elementos de la pantalla. El timbre sonará, el LED
Cualquier derecho de reclamación relacionado con la responsabilidad civil por parpadeará y el equipo vibrará. Después de esto, el equipo volverá a mostrar
productos defectuosos o como consecuencia de los daños a terceras personas la pantalla en normal funcionamiento.Nota: No se puede realizar un auto-
contra GMI será retirado si no se han tenido en cuenta las advertencias. prueba mientras haya una alarma activada.
B-12
QUICK OPERATING INSTRUCTIONS
Español
UTILIZAÇÃO
El equipo llevará a cabo una secuencia de calentamiento
Presione y mantenga para encender el equipo con comprobación del sensor y auto-cero. Cuando
aparezca una lectura PPM o %, El equipo estará listo
Presione para encender la luz de la pantalla para ser utilizado.
manualmente.

Presione otra vez para visualizar el MAX (o MIN para equipos para oxígeno).

Presione y mantenga para borrar el MAX (MIN).

Presione y mantenga para comprobar el mecanismo de bloqueo o alarma de


fallo, o mudo sin mecanismo de bloqueo.
Presione y mantenga para llevar a cabo un auto-prueba.
(Ver explicación en la página anterior).

Presione y mantenga y , durante 3 segundos, para apagar el equipo.


B-13
B-14
INDEX

INDEX

A
ACCESSORIES 6-1 AREAS OF USE iv
ACKNOWLEDGE ALARM(S) 3-4 Audible Alarm A-2
ADDITIONAL INFORMATION 7-1 Automatic Bump Test 5-3
ADVICE, DISPOSAL ii
Alarm, Audible A-2 B
ALARM DISPLAY 3-2
Alarm, Visual A-2 BACKLIGHT, DISPLAY 2-7
ALARM(S), ACKNOWLEDGE 3-4 BACKLIGHT ON 2-7
ALARM(S), MUTE 3-4 Battery Low / Exhausted 2-6
ALARMS 3-1 BUMP TEST 5-3

i
OXYGEN / TOXIC - USER HANDBOOK

DISPLAY, INSTRUMENT 1-3


DISPLAY, OPERATING 2-6
C DISPLAY, SWITCHING BACKLIGHT ON 2-7
DISPOSAL ADVICE ii
CALIBRATION 5-1
CALIBRATION VALIDITY 5-2 E
CLEANING 4-1
CONSTRUCTION 1-3 English B-1, B-2
Construction A-2 Espanol (Spanish) B-1, B-12
COPYRIGHT i
F
D
FAULT FLAGS 2-10
Data Sampling System A-2 FEATURES, INSTRUMENT 1-2
Dimensions (excl. clip) A-1 FILTER REPLACEMENT 4-2
Display A-2 FLAGS, FAULT 2-10
DISPLAY, ALARM 3-2
DISPLAY BACKLIGHT 2-7

ii
INDEX

G INTRODUCTION 1-1
Italiano (Italian) B-1, B-5
GAS TYPE 1-2
L
H
LIABILITY i
HANDLING iv Life Remaining 2-6
Humidity A-2 Limits, Temperature A-2

I M
IDENTIFICATION, INSTRUMENT 1-3 MAINTENANCE, OPERATOR 4-1
INFORMATION, ADDITIONAL 7-1 Manual Bump Test 5-4
INSTRUCTIONS, OPERATING B-1 MANUAL CALIBRATION SYSTEM 6-1
INSTRUMENT DISPLAY 1-3 MAXIMUM / MINIMUM VALUES 2-7
INSTRUMENT FEATURES 1-2 MESSAGES, OPERATOR 2-10
INSTRUMENT IDENTIFICATION 1-3 MINIMUM VALUES 2-7
INSTRUMENT PART NUMBER / GAS TYPE
1-2
iii
OXYGEN / TOXIC - USER HANDBOOK

MODIFICATION NOTICES i P
MUTE ALARM(S) 3-4
PARAMETERS, OPERATING A-1
N PART NUMBER 1-2
Polski (Polish) B-1, B-11
Nederlands (Dutch) B-1, B-7 Português (Portuguese) B-1, B-9
NORMAL OPERATING DISPLAY 2-6 Power Source A-2
NOTICES, MODIFICATION i Protection Rating A-2
NUMBER, PART 1-2

O Q
QUICK OPERATING INSTRUCTIONS B-1
OPERATING DISPLAY 2-6
OPERATING INSTRUCTIONS B-1 R
OPERATING PARAMETERS A-1
OPERATION 2-1 Rating, Protection A-2
OPERATOR MAINTENANCE 4-1 RECORD, REVISION v
OPERATOR MESSAGES 2-10 Remaining, Life 2-6

iv
INDEX

REMOTE SAMPLING 2-8 T


REPLACEMENT, FILTER 4-2
REVISION RECORD v Temperature Limits A-2
TEST 2-11
S TEST, BUMP 5-3
Training 7-1
SAFETY ii, iii, vi, viii, ii TRANSIT iv
SAMPLING, REMOTE 2-8 TYPE, GAS 1-2
Sampling System A-2 TYPICAL OPERATING PARAMETERS A-1
SOFTWARE ii
Source, Power A-2 V
STORAGE iv
SWITCHING DISPLAY BACKLIGHT ON 2-7 VALIDITY, CALIBRATION 5-2
SWITCHING THE INSTRUMENT OFF 2-9 VIEWING MAXIMUM / MINIMUM VALUES 2-
SWITCHING THE INSTRUMENT ON 2-2 7
System, Sampling A-2 Visual Alarm A-2

v
OXYGEN / TOXIC - USER HANDBOOK

W
WARRANTY iv
Weight (incl. battery) A-1
World Wide Web 7-1

vi
GMI Head Office: GMI Service & Calibration Division:
Inchinnan Business Park, 25 Cochran Close, Crownhill,
Renfrew, Milton Keynes,
PA4 9RG, MK8 OAJ,
Scotland, U.K. England, U.K.
Telephone +44 (0)141 812 3211 Telephone +44 (0)1908 568867
Fax +44 (0)141 812 7820 Fax +44 (0)1908 261056
e-mail: sales@gmiuk.com e-mail: service@gmiuk.com
http://www.gmiuk.com

You might also like