Professional Documents
Culture Documents
Disclaimer
The Good ELISA Practice (GEP) manual provides analytical results, but serves to increase the quality
a comprehensive overview for both - beginners of assessment at ELISA-analysis. The manual
and advanced analysts - in order to improve the applies in addition to the respective detailed
quality of performed ELISA analysis. It represents instructions of each test-kits. For the proper use
a guideline helping to establish reasonable of test-kits, the respective instruction manual
framework conditions and conditions of use, which is decisive and shall take priority over this GEP
have to be complied with to achieve the best manual.
possible performance when using R-Biopharm
ELISA test kits. The GEP manual is continuously up-dated. Please
make sure that this is the newest version of the manual
The GEP manual makes no claim to completeness, prior to performing the ELISA-analysis. The manual
but describes only certain requirements. can be retrieved, printed and downloaded from
The compliance with such minimum standards https://food.r-biopharm.com/technologies/elisa/
does not guarantee the achievement of correct
Introduction
The GEP manual provides a comprehensive on a well-established analytical method that
overview for both beginners and advanced meets the requirements of modern laboratories.
analysts in order to improve the quality of RIDASCREEN® tests show a high sensitivity and are
performed ELISA analysis. characterized by robustness. These test systems
are used by food manufacturers to test raw
The enzyme-linked immunosorbent assay materials or finished goods.
(ELISA) is an antibody-based test method. This
widespread technology is easy to use, sensitive, Furthermore, ELISA requires only basic lab
fast and reliable. Additionally, ELISA is robust equipment and is easy to use compared to HPLC
and mostly used for quantitative analysis. and/or LC-MS/MS.
With more than 30 years of experience in This manual is divided into three chapters, which
developing ELISA test kits, R-Biopharm relies follow the workflow of an ELISA based analysis:
Content
1 ELISA Basics 6
1.1 Antibody – Antigen Detection 6
1.2 Analyte 6
1.3 ELISA Formats 6
1.3.1 Sandwich-ELISA 7
1.3.2 Competitive ELISA (formats) 8
1.4 ELISA components 8
1.4.1 Microtiter plate (MTP) 9
1.4.2 Conjugate and Substrate 9
1.4.3 Standards (Calibrator) 9
1.4.4 Buffers 9
1.4.5 Stop-solution 10
1.4.6 Additional components 10
1.5 Lab equipment and its maintenance 10
1.5.1 Pipettes 10
1.5.2 ELISA plate washer - automated systems 11
1.5.3 Microtiter plate (MTP) reader for ELISA 11
1.5.4 Automation 12
1.5.5 Additional equipment 12
1.5.6 General words of advice 12
1.6 Good laboratory practice (GLP) 13
1.7 Test kit labeling 13
1 ELISA Basics
1.2 Analyte
The claimed (antigen) analyte of an ELISA could be: 4 A group of specific proteins e.g.
staphyloenterotoxins A, B, C, D, and E
1 A defined chemical substance e.g.
aflatoxin B1 5 A more or less defined group of proteins
from a food commodity e.g. caseins (as a
2 A group of defined chemical substances e.g. part of milk proteins)
aflatoxins B1, B2, G1 and G2
6 A food commodity e.g. peanut protein
3 A specific protein e.g. staphyloenterotoxin A
1.3.1 Sandwich-ELISA
Sandwich ELISA are often used in protein analysis washing, conjugated antibody is added which binds
such as food allergens. The analyte (antigen) to a second binding site (epitope) of the antigen
needs to be large enough to have two binding (analyte). After washing, substrate is added and
sites (epitopes). In this method a specific antigen the results can be measured via the optical density
(analyte) in the sample binds to antibodies attached (OD). The signal is proportional to the amount of
to the solid phase of a microtiter plate. After antigen (analyte) (see Figure 1, step 6, Figure 2).
Results are read in a MTP The more yellow colour the more target
6 A reader at 450 nm allergen is present in the sample
conc.
Figure 1: Example of test procedure and principle – RIDASCREEN® Allergen Sandwich ELISA
Good ELISA Practice – Manual 8
There are different formats for a competitive ELISA. method is suitable to measure samples with just
one epitope as well as small analytes such as
The competitive ELISA (Figure 3) consists of a mycotoxins or antibiotics.
microtiterplate where the antibody is bound to the
surface of the well. The analyte from the sample The difference of the direct to the indirect
and the enzyme-analyte conjugate are added to competitive ELISA (Figure 4) is that an additional
the well. The competition on antibody binding catcher antibody is bound to the microtiter plate.
sites starts. After washing, substrate is added and
the measured OD value is inversely proportional There are also competitive ELISA in which the
to the amount of analyte in the sample. The more antigen is bound to the MTP well. This is the case
analyte is present, the smaller the OD value. This for RIDASCREEN® Gliadin competitive (R7021).
Substrate
E E
E Substrate
Antigen
Substrate
The MTP (96 or 48 wells), is the basis for the of these plates. Without this activation no or only
analysis. In every well the antibody or antigen minor binding of antibodies or antigens occurs.
(depends on the format) is bound to the surface. In every well the antigen-antibody reaction and
A common plate material is polystyrene although the conjugate-substrate reaction takes place.
other materials can be used. The material is Lastly, the reaction is stopped with a specific stop-
activated by β- or γ-radiation by the manufacturer solution and the optical density is measured.
Antibodies or analytes linked to an enzyme are chromogenic mix which reacts with the conjugate
called conjugates. The linked enzyme converts its enzyme. The result is a colored solution of which
specific substrate into a colored bluish product. the optical density can be measured.
The substrate is usually a hydrogen peroxide/
RIDASCREEN Gliadin
Art. No.: R7001 Lot: 14383 Expiry: 2014-12
1.4.3 Standards (Calibrator) R-Biopharm AG, Darmstadt, Germany certifies that this batch has been approved by the Quality
Assurance Department and conforms with specifications
Standards
2.2
0.0
5.00 10.00 20.00 40.00 80.00
Please note:
1.4.4 Buffers The absorbance for the standards may decrease during the shelf life of the kit. The general shape of the
curve will remain similar, while the slope might change slightly. Furthermore refer to product leaflet 8.
Indication of instability or deterioration of reagents.
All ELISA systems contain components of biological has also an influence on test kit performance. For
Remark: This document has been created electronically and is therefore valid without a signature.
origin. For long term storage of these components the convenience of test kit users, these buffer are
and for proper function during the testing included ready-to-use or as concentrates. They
The R-Biopharm group is DIN EN ISO 9001 certified.
procedure, pH-value and ionic strength needs to are also used for sample preparation and washing www.r-biopharm.com
be constant. Often the kind of buffer component procedures of the microtiter plates.
Good ELISA Practice – Manual 10
1.4.5 Stop-solution
The stop-solution terminates the enzyme- reaction, the color will change from blue to yellow
conjugate-reaction. In most cases, sulphuric acid and will remain stable until measurement within
at low concentrations is used. By stopping the 10 min.
For most systems a positive and negative control This gives the opportunity to control the test
is recommended, for example a spiking solution. system.
1.5.1 Pipettes
A pipette is used to transfer a precisely defined • Multistepper pipettes (with the possibility to
volume of a solution to e.g. the wells of a microtiter pipette multiple times of a specific volume); used
plate. Pipettes are crucial for results with a high normally for addition of antibody or conjugate
precision. Therefore, a proper pipetting technique solutions
as well as regular calibrated pipettes are important.
Different kinds of pipettes exist: • Multi-channel pipettes (with the possibility to
pipette the solution in 8 or 12 cavities at the
• Single-channel pipettes with fixed volumes same time); used normally for washing steps or
e.g. 50 µl addition of antibody or conjugate solutions
• Single-channel pipettes (e.g. with variable • Bottle top dispenser (used normally for washing
volumes between 10 and 100 µl); used normally steps)
for samples and standards
• Fully automatic machines (all pipetting and
incubation steps are performed automatically)
11
After every incubation step (except the incubation where users load a plate and select a program. The
with the substrate) the ELISA plate has to be washers then dispense, soak and aspirate wash
washed with washing buffer. Washing is a crucial buffer from the plate in seconds. When using such
step when performing an ELISA to obtain results instruments thorough cleaning of the washer is
with high precision. important to avoid cross contamination. Refer to
the manufacturer’s instructions referring cleaning
Washing can be done manually with an 8-channel between different runs.
pipette or an 8 channel manifold (see Figure 15).
The washing procedure is an important step and
Sometimes customers use automated ELISA plate has to be validated if deviated from the described
washers. These washers are laboratory instruments product information.
1.5.4 Automation
One possibility for working with an ELISA is be validated and calibrated for your test system.
the use of an automated system which allows Examples for automation are the ThunderBolt® and
you to test your samples without any manual Bolt™. For further information please contact:
steps. Therefore, the automated system has to sales@r-biopharm.de.
Figure 7: Bolt™, Art. No. ZBOLT Figure 8: ThunderBolt®, Art. No. ZTB
Forward pipetting is a technique to dispense Before starting pipetting with an adjustable pipette:
a measured quantity of liquid by means of air 1. Adjust to the desired volume
displacement pipette. The technique is mainly 2. Check adjusted volume
recommended for aqueous solutions, such as
buffers, or diluted acids or alkalis. In case of
solutions with a high viscosity or a tendency to
foam, reverse pipetting is more suitable.
Forward Reverse
Ready position
First stop
Second stop
Figure 9: Schematic illustration of forward and reverse pipetting techniques when using a single channel pipette (piston pipette).
Blue blocks indicate the steps where the pipette needs to be placed in liquid to aspirate. Grey blocks indicate the target wells.
15
21 Press the operation button to the first stop. 21 Press the operation button to the second stop.
31 Flush pipette tips before use. Pipette tips from 31 Immerse the pipette into the liquid. Slowly
some manufacturers need to be flushed before release the operating button to ready position
aspiration and dispensing of the appropriate and wait until the desired liquid volume has
liquid. Please check the according manual. been aspirated. Ensure that not bubbles or
In case of any doubt, flush the tip before foam occurs in the pipette.
pipetting.
41 Remove excessive liquid from the outside of
41 Put the pipette tip approx. 1 cm deep into the the tip.
liquid. Slowly release the operation button to
the ready position and wait until the desired 51 Dispense the liquid into the desired well by
liquid volume has been aspirated. Ensure that pressing the operation button to the first stop.
no bubbles or foam occurs in the pipette. Ensure that no liquid remains on the outside of
the tip.
51 Remove excessive liquid from the outside of
the tip by touching the test tube with the tip. 61 For repetitive liquid pipetting, press the
operation button to the first stop and repeat
61 Dispense the liquid into the desired well by steps 3 - 5.
pressing the operation button to the second
stop. 71 Remove the tip to waste.
Organic solvents show high vapour pressures Figure 10: Pipetting of organic solvents with pipettes using the
which can affect precise pipetting. The use of air displacement technique. Particular care should be taken to
prevent evaporation into and leaking out of the tip (A).
pipettes with air displacement technique to transfer Flushing the pipette before liquid transfer helps to transfer the
organic solvents may lead to evaporation of the correct volume (B).
If single channel pipettes are used to transfer solvent at least 3 times before the desired volume
organic solvents, the pipette tip and the air inside is transferred. Use appropriate quality control
the pipette needs to be saturated with organic procedures to monitor the correctness of these
solvent vapour before pipetting the desired kinds of pipetting steps.
volume. For this aspirate and dispense the organic
17
Immersion depth
Only the top of the pipette tip is immersed into
the standard or sample solution (see Figure 11 a).
When immersing too deep, then too much liquid air
is aspirated. When the pipette tip is too close to bubble
Pre-rinsing
Pre-rinsing equalizes the air temperature and Do not touch
well wall
pressure inside the tip with the temperature of the
sample. During pre-rinsing the plunger is pressed
and released 2 to 3 times (Figure 11 c).
d e
Immersion angle
Best is to hold the pipette close to vertical. a Only the top of the tip is immersed
Anything greater than 20 ° of vertical reduces b A consistent pipetting helps to avoid jerky air aspiration
c During pre-rinsing the plunger is pressed and released
accuracy of pipetting (see Figure 11 d) 2-3 times
d The pipette is hold close to vertical during pippeting
e During releasing the top of the pipette is close to the
Dispensing
well wall
To release sample or standard solution hold the
pipette tip close to the well wall (see Figure 11 e). Figure 11: Best practice pipetting
Good ELISA Practice – Manual 18
Unwanted contamination of samples can influence immediately whenever possible. If storage cannot
test results significantly. If there are signs of be avoided, check for optimal storage conditions
unwanted contamination or spoilage do not use and analyze them as soon as possible. All samples
the sample and request a new one. Store the need to be correctly labelled and sealed to avoid
sample according to the test kit manufacturer evaporation or dry-out. Inappropriate storage
or according to the best scientific knowledge. conditions may influence later analysis and alter
Generally, samples should be prepared and tested test results.
For sample preparation please follow the maintained equipment for sample preparation. For
instructions for use provided with the test any related questions please contact R-Biopharm.
kit. Sample preparation involves typically
homogenisation and extraction of the Depending on the parameter to be tested,
analyte. Sample preparation involves typically instructions for use may contain information on
homogenisation and extraction of the analyte. how prepared samples can be stored for later
Changes or variations may lead to incorrect analysis. Please follow these instructions carefully
test results. Make sure to use only suitable and or prepare samples directly before analysis.
19
Before further use, samples must be completely should be avoided since it may denature proteins.
thawed. Thawing of frozen samples should A further sample preparation of liquid samples
be performed at 4°C or at room temperature, (e.g. in case of milk) may be necessary before
dependent on analyte stability. analysis. Please check the instructions for use for
further information.
In case of unprepared, non-liquid samples
check the instructions for use for further sample Depending on the sample, freezing and thawing
preparation and homogenization. can lead to crystallization or coagulation. Avoid
freeze-thaw cycles wherever possible, since it can
Liquid samples must be thoroughly mixed change sample integrity and alter test results. If
before they can be used for analysis. To achieve a possible, aliquot liquid samples before storage at
homogeneous sample carefully vortex or invert ‑20 °C to avoid freeze-thaw cycles.
the sample. Foam formation or intensive mixing
Certified reference materials (CRM) are naturally Figure 12: R-Biopharm offers a wide range of CRMs for
mycotoxin analysis
contaminated, homogeneous matrices whose
analyte content has been exactly and reliably
determined (Figure 12). The regular use of CRMs
is recommended for quality assurance to establish
traceability. This allows checking the trueness and
precision of experimental procedures and for the
testing of handling skills. If no reference material
is available, the use of control samples or matrices
spiked with defined analyte concentrations are
recommended.
The expiry date printed on the outer label applies Figure 13: Storage conditions and expiry date are printed
on the outer label of each ELISA test kit
to all reagents contained in the kit. To maintain the
shelf life, store the kit at conditions noted on the
outside label of the package (Figure 13).
The expiry date of the kit is labelled on the outer The exposure of the ELISA kit to cold/warm cycles
label of the kit package. At least until this date the should be kept as low as possible. We recommend
kit will perform within specifications. Additionally, that samples should be collected. Testing of higher
every kit component has its own expiry date which sample numbers at once reduces the expenditure
is identical or even exceeds the expiry date of the of time per sample. Rather test higher sample
test kit. numbers at once than test only a few samples
consecutively. Please check the instructions for use
If there is more than one kit on hand, it is for relevant limitations.
recommended to use the first in – first out
principle. This means the kit with the shortest
expiry date on the outer label should be used first.
We recommend indicating the date of the first use
on the outer label of the kit box to avoid a mix-up
of this principle.
21
2.3.3 Pre-Warming
All reagents need to be at room temperature Figure 14: Bring all test kit components to room temperature
before use and perform the test at 20 - 25 °C (68 - 77 °F)
before they can be used in the test. Take all
components out of the kit package before use
and allow them to reach room temperature
(20 - 25 °C). Larger bottles and greater 20 - 25 °C/
volumes may require more time to reach room 68 - 77 °F
temperature. Check the temperature of the
components in any case of doubt (Figure 14).
ELISA tests are sensitive to temperature should be prevented from direct exposure to
fluctuations. Therefore, try to stabilize and sunlight and ventilations. Cold lab ware and cold
control laboratory conditions. This includes the benches may also influence the temperature. It
temperature during photometric analysis. Perform is helpful to isolate the microtiter plate from the
ELISA tests between 20 and 25 °C and avoid bench surface by performing the test on a suitable
conditions that are able to drastically change the underlay. A cheap and easy solution is the use of
temperature or increase evaporation. ELISA tests paper towels.
Good ELISA Practice – Manual 22
A clean and reproducible way of working is crucial Figure 15: Carefully label all used containers and document
your labelling.
for optimal results in food and feed analysis. A very
common source of contamination is insufficiently
cleaned re-usable lab ware. To avoid this, it is
highly recommended to use solvent-resistant
single- use lab ware. If this is not possible,
re-usable lab ware should be laboratory sterile
and free of contamination. We recommend using
a laboratory dishwasher or equivalent. Use quality
control blank samples to check for contamination.
Before you start, read the instructions for use the test procedure. For the transfer of samples we
enclosed in the kit. Prepare all extraction solutions recommend single channel pipettes, or, if a pre-
and buffers according to these instructions and dilution plate is used, a multi-channel pipette. For
follow the described procedures to obtain optimal the pipetting of antibody and enzyme solutions,
results. To allow an unobstructed test procedure multistep pipettes are the best option.
it is helpful to prepare a pipetting scheme before
you start with your experiment. Between handling steps a dry-out of the wells has
to be prevented.
Depending on the ELISA, antibody and conjugate
solution may need to be diluted prior to use. Use of pre-plate: In the case of the
These dilutions should be prepared directly before RIDASCREEN®FAST Allergen test kits do not use
use and should not be stored for further use. more than three strips (24 wells) at a time. If more
Contaminated or incorrectly stored conjugate than three strips are needed, a second uncoated
solutions may have a reduced enzyme activity or plate (e.g. low binding from Greiner bio-one Cat.-
may induce a background signal. No. 655101) should be used as a pre-plate to avoid
a time shift over the microtiter plate. All standards
To obtain optimal results, test samples and control and samples are pipetted into the uncoated
samples (standards, reference samples) have to plate (at least 150 μl per well) and then quickly
be available in the same diluent. Strictly following transferred to the coated microtiter plate with an
the sample preparation protocol ensures this. The 8-channel pipette (see Figure 16). Remember: the
pipetting into the plate needs to be performed reaction starts when the first solution is added to
quickly and without interruptions at every step of the coated plate.
Figure 16: The ‘pre-plate’ contains at least 150 µl of standard standards or samples are transferred quickly to the coated
or sample per well. To the 7 strips of the pre-plate 5 standards plate. The pipette tips are flushed with the standard or sample
and 23 samples have been added in duplicate (i.e. 56 wells solution, each tip is only used once.
in total). Using an 8-channel multi pipette, exactly 100 µl of
Good ELISA Practice – Manual 24
• In the sandwich ELISA RIDASCREEN® Gliadin The activity of the chromogen may be influenced
(Art. No. R7001) the ‘antigen-antibody-reaction’ by light. Cover the plate to protect the chromogen
starts when the sample is added to the wells. So from light and store it in the brown flask in which
the ‘time sensitive’ step is the sample pipetting. it is delivered. To stop the reaction use the stop
solution delivered with the kit.
• In the competitive ELISA RIDASCREEN®
Ochratoxin A (R1311) the plate is already coated
with the specific antibody and the reaction
starts when the sample is added. So the ‘time
sensitive’ step is the sample pipetting.
Washing is a crucial step to remove all unbound Buffers, tubes, manifolds and washing needles
components that might influence reactions or need to be kept free from contamination of
lead to false results. For washing only use washing microorganisms. Keep the device clean and if
buffers recommended in the instruction for use. necessary remove contaminated parts. In case of
any doubt, use the manual washing procedure or
Many kits contain washing buffer salts or solutions contact the manufacturer of the washer.
that can be used to prepare ready to use washing
buffers. This removes the necessity to weigh Automated washing systems are not available at
reagents to prepare the own buffers. For stability all testing sites. Due to this we highly recommend
and storage information on the individual washing the use of the relatively cheap and easy to
buffers delivered with the kits, see the instructions handle bottle top dispensers with an 8 - or 12
for use. As all other components, washing buffers -fold manifold for the washing procedure. Other
need to be at room temperature before use. Follow manual washing techniques like washing bottles
the specific recommendation of the instructions may not allow to treat every well exactly the same
for use regarding the number of washing steps. and should not be applied.
At the end of the incubation steps pour the liquid Washing steps should be performed fast but
out of the wells and tap the microtiter plate holder efficiently and accurate. Make sure that the time
vigorously upside down on absorbent paper to between addition of washing buffer to the first
ensure the complete removal of the liquid from and the last well is as short as possible. This
the well. All liquid has been successfully removed ensures that wells do not dry-out and it minimizes
when no signs of liquid remains on the paper towel. differences in incubation times. Despite fast
Most ELISA tests require 250 - 300 µl of washing working speed pay attention to accuracy. Spillover
buffer per washing step and well. Add the washing of liquid from one well to another needs to be
buffer and remove the liquid by pouring out and avoided. In case of any doubt add an additional
tapping. Repeat the washing step 3 - 5 times washing step.
(see instructions for use). For washing steps it is
recommended to use a bottle-top dispenser (e.g.
Brand, 4720420) connected to a 8-channel manifold
(e.g. Brand, 704526) as shown in Figure 17.
Microtiter plates are delivered in a re-sealable bag important for components such as standard
with a pouch containing a desiccant (Figure 18). solutions that may contain organic solvents with a
In case not all wells of the plate are needed, store high vapour pressure. We recommend putting all
the rest of the wells in this bag. Put the wells and components back into the kit package for storage.
the microtiter plate together with the desiccant Store all components upright and under the
into the bag and close it. Close all flasks and make indicated conditions (Figure 19).
sure to screw the lids on firmly. This is especially
Figure 18: Unneeded plate strips should be stored together Figure 19: Until further use all components should be stored
with the desiccant in the re-sealable pouch in which they are in the kit package in an upright position
delivered
The components of each lot are thoroughly allowed. An exchange of components of kits
adjusted to deliver ELISA kits that show the optimal with the same product number is possible if the
performance. The exchange of one or more of lot number of the kit is identical. However, we
these components between different lots will recommend using only components delivered
change the performance of the tests and is not with the particular kit.
27
ELISA test kits may contain hazardous substances. laboratory security precautions. While performing
For information on hazards contained in the the test procedure use laboratory gloves, wear
substances take note of the warnings on the labels a lab coat, do not eat, drink or smoke and keep
of the components and refer to the appropriate all components away from sources of ignition.
safety data sheets (SDS). Generally, handle all Disposal of waste may differ from country to
components with care and take all usual country. Please refer to local disposal rules.
Figure 21: Parallel performance of several ELISA tests must be carefully scheduled and organized
29
0.8 A
e
10.00 0.938 c 2.00 mg/kg
0.422 stool OD b
-0.5
2.5 2.8
2.500
Std4 4
-1.0 0.6 e 3.9 21.8 2.00
Std3 0.422
8
0.8 0 0,010 s 1.4
-1.0
Std5 8 3.9 21.8
0.6
-1.5 Std5 2 Std5 4
o 2.0
2.000
40.00 1.683
10.00
B 0.635
1,5
15.00 1.399 0.6 6.00
19,5
0.204
0,091 r
b 1.2
-1.5
2.2
4.2
Std5 4
1.5
1.500
0.4 -2.0 0.9 4.7 10.5
33 0,169 a
n 1.0 Std6
-2.0 8
Std4
6.00
4.7
20.00
C 8
0.204
10.5
1.094
5,4
0.4
1.080.00
1.000
-2.5 Std6
20.00
2
1.844
56 0,370 c
e 2.1
-2.5
2.366 2.8
D 3,6
0.2 0.8
-3.0 0.5 0.2
95 0,733 0.5
0.500
-3.0
Std5
40.00
8
1.683
0.0 -3.5 0.0 160 1,188 0.6
0.0
0.000 2.2
0.00 5.00 10.00 15.00 20.00 0.222 0.666 2.00 6.00 5.00 10.00 20.00 40.00 80.00 -3.5
275 2,015 1 0.222 10
0.666 100
2.00 1000
6.00 Std6 8
Concentration (mg/L) 0.4
Concentration (ppm) Concentration (ppb) [mg Calprotectin/kg stool] 80.00 2.366
470 2,704 Concentration (ppm) 2.1
0.2
800 3,113
X-axis
Corr.Coeff.: 0.9999 Corr.Coeff.: 0.9983
slope = 0.0918 50% inhibition = 0.519 0.0 Corr.Coeff.: 0.9983
5.00
50% inhibition10.00
= 0.519 20.00 40.00 80.00
OD
Buffer1
Enzyme Solution
13084
13084
2015-01
Logit of concentration
2015-01
Conjugate
Antibody
11484
11484 OD
2015-10
2016-04
Negative control12173
Buffer1
Substrate/Chromogen
15183 OD
2015-03
2015-10
Standards
Conjugate
12494
11484
2016-05
2015-10
Spike Solution 12124 2015-02 Red Chromogen Pro 14413 2016-03 Antibody 11484 2016-04
Stop solution 15183 2018-04 Lot No. Expiry
Stop solution
Washing buffer salt
14114
051M8211
2019-02
2016-06
Negative control
Washing buffer 11243Diluent 2015-11 3 QC specification:
Red Chromogen Pro
Microwell plate
Stop solution
14413
15043
14114
2016-03
2014-12
2019-02
OD < 0,05
The relationship In Logit-log is used to Cubic spline is a numeric 4 parameter is a nonlinear Standards
Washing buffer salt
Conjugate
11353
051M8211
11353
2015-04
2016-06
2015-04
curve will remain similar, while the slope might change curve will
slightly. remain similar,
Furthermore refer to Calibrator
while the slope
product leaflet might
8. change slightly. Furthermore refer
The absorbance tostandards
for the product leaflet 8.
may decrease during the shelf life of the kit. The general shape o
sign.: Eddasimple
Rohm linear regression is Indication ofregression models.
instability or deterioration of reagents.
Date: 2014-04-29 functions
Indication of instability withofhigh
or deterioration reagents.degree Please • bottom of the curve
curve will remain similar, while the slope might change slightly. Furthermore refer to product leaflet 8
note: of instability or deterioration of reagents.
Indication
The target value of the Calibrator Calibrator has been determined
applied using the formula: sign.: N. Stork-Heininger
Quality Assurance Representative
sign.: Edda Rohm of smoothness
Quality Assurance Representative to be:
at the
Date: 2014-12-10
The absorbance
• fortop
sign.: N. Stork-Heininger
of the Date:curve
2013-09-19
the standards may decrease during the shelf life of the kit. The general
OD 1, 188shap
Date: 20
•
Quality Assurance Representative
y = mx + b
Remark: This document has been created electronically and is therefore valid without a signature.
connections of the curveQuality
will remain
Indication
EC50
similar,
Assurance
of instability or
while the slope might change slightly. Furthermore refer to product leafle
Representative
deterioration of reagents.
The rangeand(target value without± 3 SD) should becreated
within: OD 0,651 – 1
• slope at the inflection
Remark:
Remark: This document has been created electronically and is therefore valid This document
without has been created
a signature. electronically is therefore valid a signature.
polynomial pieces (knots). Remark: This document has been electronically and is therefore valid without a signature.
sign.: Edda Rohm Date: 201
Quality Assurance Representative
point of the curve =EC50
Figure 22: Overview about standard curve fittings
The R-Biopharm group is DIN EN ISO 9001 certified.
Positive controlRemark: This document has been created electronically and is therefore valid without a signature.
The R-Biopharm group is DIN EN ISO 9001 certified. The R-Biopharm group is DIN EN ISO 9001 certified.
The R-Biopharm group is DIN EN ISO 9001 certified.
www.r-biopharm.com www.r-biopharm.com www.r-biopharm.com
The target value of the Positive control Control + has been www.r-bioph
145,1 mg/kg
determined to be:
The R-Biopharm group is DIN EN ISO 9001 certified.
The range (target value ± 3 SD) should be within: 113,3 – 176,9
www.r-biopharm
Standards Standards
Standard curve Standard curve
Std. n Std. n
2.4 100
Conc.(ppb) mean Conc.(ppt) mean
CV(%) CV(%) B/B0
2.2
90
Std1 8 Std1 8
2.0
0.00 0.058 0.00 2.302
9.3 80 2.3 100.0
1.8
Std2 8 Std2 8
5.00 0.294 70
1.6 A 25.00 2.068
A 5.7 2.8 89.8
b
b s
s 1.4 Std3 8 60
o Std3 8
o 10.00 0.635
r 50.00 1.751
r 4.2
b 1.2 b 3.1 76.1
a 50
a Std4 8
n 1.0 20.00 1.094 B/B 0 c
n Std4 8
c 2.8 100.00 1.348
e 40 3.0 58.6
e
0.8 Std5 8
(%) Std5 8
40.00 1.683 30
0.6 2.2 250.00 0.719
3.4 31.2
Std6 8
0.4 20 Std6 8
80.00 2.366
2.1 750.00 0.306
0.2 4.2 13.3
10
0.0
5.00 10.00 20.00 40.00 80.00
0
25.00 50.00 100.00 250.00 750.00
Concentration (ppb)
Concentration (ppt)
Figure 23: Standard curve of an ELISA in sandwich format, Figure 24: Standard curve of an competitive ELISA, where
50% inhibition = 132.2
where absorbance is plotted over concentration
Lot No. Expiry
B/B0 is plotted over concentration
Microwell plate 15043 2014-12 Lot No. Expiry
Standards 11353 2015-04
Conjugate 11353 2015-04 Microwell plate 15313 2015-04
Buffer1 12173 2015-03 Standards 12034 2015-09
Substrate/Chromogen 15183 2015-10 Conjugate 14064 2015-07
Stop solution 15183 2018-04 Buffer1 12044 2015-12
Washing buffer 11243 2015-11 Red Chromogen Pro 11113 2015-05
Stop solution 11373 2018-08
Washing buffer salt 051M8211 2016-06
Please note:
Please note:
The absorbance for the standards may decrease during the shelf life of the kit. The general shape of the
curve will remain similar, while the slope might change slightly. Furthermore refer to product leaflet 8.
Indication of instability or deterioration of reagents. The absorbance for the standards may decrease during the shelf life of the kit. The general shape
the curve will remain similar, while the slope might change slightly. Furthermore refer to product lea
sign.: Edda Rohm Date: 8. Indication of instability or deterioration of reagents.
2013-09-19
Quality Assurance Representative
sign.: Edda Rohm Date: 2014-0
Remark: This document has been created electronically and is therefore valid without a signature. Quality Assurance Representative
Remark: This document has been created electronically and is therefore valid without a signature.
RIDASCREEN Chloramphenicol
Art. No.: R1505 Lot: 11084 Expiry: 2015-04
3.5 Determination of analyte concentration R-Biopharm AG, Darmstadt, Germany certifies that this batch has been approved by the Quality
Assurance Department and conforms with specifications
Standards
Std2 8
70
A 25.00 2.068
Std3
89.8
8
r 50.00 1.751
no values. b
a
n
50
3.1
Std4
76.1
8
c 100.00 1.348
e 40 3.0 58.6
(%) Std5 8
30
contentration 250.00
3.4
0.719
31.2
20
obtained
Std6 8
through 750.00 0.306
10 standard 4.2 13.3
curve
0
25.00 50.00 100.00 250.00 750.00
Concentration (ppt)
Please note:
The absorbance for the standards may decrease during the shelf life of the kit. The general shape of
3.6 Measuring range and dilution factor the curve will remain similar, while the slope might change slightly. Furthermore refer to product leaflet
8. Indication of instability or deterioration of reagents.
If an absorbance or a B/B0 value is obtained by the concentration of the standards, and by the
Remark: This document has been created electronically and is therefore valid without a signature.
which is below or above the standard curve, the dilution factor of the sample preparation method.
RIDASOFT® Win.NET software gives a results For example, if milk is diluted 1:4 (1+3) before
The R-Biopharm group is DIN EN ISO 9001 certified.
‘<’ (below) or ‘>’ (above) standard range. The applying to the microtiter well, the dilution factor
www.r-biopharm.com
software provides the option to extrapolate the is 4. This means, that the results read from the
concentration value. In general, these extrapolated standard curve have to be multiplied by 4 to obtain
values are only estimates and are not reliable. The the correct concentration of the analyte in the
further the sample is below the lowest or above sample. If the analyte in a sample is concentrated
the highest standard, the bigger the uncertainty by a factor of 2 during sample preparation, e.g.
of the calculated concentration is. It is advised, to by evaporation or column clean-up, the dilution
dilute samples that are above the largest standard factor is 0.5. This means, that the results read from
and to repeat the
Q U A analysis
L I T Y A S Suntil
U R A Nthe
C E Cresult
E R T I Fis
I Cwithin
ATE the standard curve Qhave
U A L I to
T Y be
A S Smultiplied
U R A N C E C Eby
R T 0.5
I F I Cto
ATE
Standards Standards
Standard curve Standard curve
Std. n Std. n
100 100
Conc.(ppt) mean Conc.(ppt) mean
CV(%) B/B0 CV(%) B/B0
90 90
Std1 8 Std1 8
0.00 2.302 0.00 2.302
80 2.3 100.0 80 2.3 100.0
Std2 8 Std2 8
70 70
A 25.00 2.068 A 25.00 2.068
b 2.8 89.8 b 2.8 89.8
s 60 s 60
o Std3 8 o Std3 8
r 50.00 1.751 r 50.00 1.751
b 3.1 76.1 b 3.1 76.1
a 50 a 50
n Std4 8 n Std4 8
c 100.00 1.348 c 100.00 1.348
e 40 3.0 58.6 e 40 3.0 58.6
(%) Std5 8 (%) Std5 8
30 30
250.00 0.719 250.00 0.719
3.4 31.2 3.4 31.2
20 Std6 8 20 Std6 8
750.00 0.306 750.00 0.306
4.2 13.3 4.2 13.3
10 10
0 0
25.00 50.00 100.00 250.00 750.00 25.00 50.00 100.00 250.00 750.00
to the microtiter well. The results have to multiplied by 4 to sample preparation 1. The result has to be multiplied with 0.5
Lot No. Expiry Lot No. Expiry
obtain the correct concentration:
Microwell plate The concentration
15313 of the
2015-04 to obtain the correct concentration:
Microwell plate The concentration
15313 of the
2015-04
Standards 12034 2015-09 Standards 12034 2015-09
analyte in the sampleConjugate
is 100
Buffer1
ng/kg x 4 = 400
14064
12044
ng/kg.
2015-07
2015-12
analyte in the sample is 100Conjugate
ng/kg
Buffer1
x 0.5 = 50 ng/kg.
14064
12044
2015-07
2015-12
Red Chromogen Pro 11113 2015-05 Red Chromogen Pro 11113 2015-05
Stop solution 11373 2018-08 Stop solution 11373 2018-08
Washing buffer salt 051M8211 2016-06 Washing buffer salt 051M8211 2016-06
The absorbance for the standards may decrease during the shelf life of the kit. The general shape of
33
50
of at least 20 blank matrix samples and then
40
calculated by the formula: Mean concentration of
LOD; 35
30
10
9-fold standard deviation of the concentrations of 0
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
blank samples (Figure 31). samples
CRM. The result is multiplied by 100 to obtain a of the results may be low especially ® at high Fast
RIDASCREEN Clenbuterol
percentage unit. The absolute difference between concentrations, Art.because
No.: R1701CsmallLot:differences
16213 Expiry: in OD or
2015-04
both values is called ‘bias’. B/B0 may cause large differences in concentration
R-Biopharm AG, Darmstadt, Germany certifies that this batch has been approved by the Quality
Assurance Department and conforms with specifications
(Figure 32).
If no CRM is available, e.g. for antibiotics in Standards
Standard curve
food of animal origin, instead of trueness, 100 Std.
Conc.(ppt)
n
mean
CV(%) B/B0
Std1 8
c Std4 8
0
or trueness is indicated in the validation report or 100.00 300.00 900.00 2700.00 8100.00
Concentration (ppt)
in the instructions for use.
Figure 32: In flat areas of the standard curve small differences
in OD produces50%large differences in concentration: 5 % OD
inhibition = 399.1
Please note:
The absorbance for the standards may decrease during the shelf life of the kit. The general shape of the
curve will remain similar, while the slope might change slightly. Furthermore refer to product leaflet 8.
Indication of instability or deterioration of reagents.
Remark: This document has been created electronically and is therefore valid without a signature.
B/B0 [%]
50
commodity (100 % level). Samples well below the 40
LOQ show no cross-reactivity. 30
20
0
the measurement of a standard curve consisting of 20 200
concentration series (Figure 33). After calculation Figure 33: The 50%-dose of the standard curve of the standard
substance (blue) is more sensitive (170 µg/kg) as the 50%-dose
of the 50 %-dose of the analyte or cross reactive of the standard curve of the cross reactive substance (210 µg/kg).
substance, the specificity or cross reactivity is The cross reactivity of the cross reactive substance is therefore:
calculated as follows: (170 µg/kg / 210 µg/kg) x 100 % = 81 %
If the binding of the antibody affects a substance or one cross-reacting substance is present in the
that is included in the scope of the method, this is sample.
called specificity of the method. If this binding is
related to unwanted substances that are not within 2nd example: The scope of a Tetracycline-ELISA is
the scope of the method it is called cross reactivity. the determination of tetracycline and its derivatives
like chlortetracycline, rolitetracycline and
1st Example: The scope of an ELISA describes that demeclocycline in different matrices to a certain
the system was developed to quantify aflatoxin degree. The calibrator material is tetracycline with
M1 in milk. Therefore, the specificity for aflatoxin a specificity of 100 %. The specificity of derivatives
M1 is 100 %. Furthermore it is stated that a cross- differs, e.g. the specificity of chlortetracycline is
reactivity of 10 % to aflatoxin M2 exists. 70 % in this system. Assuming the testing of a milk
sample, which is spiked with chlortetracycline,
Due to the principle of an ELISA system, the reveals a measured concentration of 10 μg/L.
antibody is not able to discriminate between the The real concentration is calculated to be 14
specificity to aflatoxin M1 and the cross-reactivity μg chlortetracycline per L of milk, because the
related to aflatoxin M2. If both are present in an specificity of chlortetracycline is 70 %.
unknown sample, the result is the sum of both
substances. But, due to the low cross-reactivity of The specificity of the (tetracycline) ELISA was
10 %, a 10-fold higher concentration of aflatoxin determined by analyzing the cross-reactivities to
M2 (compared to M1) is necessary to resulting corresponding substances in buffer system.
in a signal comparable to aflatoxin M1. An exact In samples, the specificity may deviate from those
quantification is only possible if only one analyte determined in the buffer system due to matrix
Good ELISA Practice – Manual 36
effects. Prior to the analysis of cross-reactive specific analyte which are not included in the
substances, the user has to determine the Limit scope of the method, the user must determine the
of Detection and the Recovery for the substance Limit of Detection and the Recovery of the specific
in the respective sample matrix. The test cannot analyte in the particular sample matrix. Please note
discriminate between analytes and cross-reactive that the specificities and cross reactivity’s were
substances. experimentally determined in the buffer system
only, as it is very time consuming and laborious
Nevertheless, ELISA tests are often used as screening to determine every specificity or cross reactivity
methods. Any positive results or concentration of any analyte or cross reactive substance in every
higher than specified threshold should be verified by matrix.
a confirmatory method e.g. LC-MS/MS.
R-Biopharm
An der neuen Bergstraße 17
64297 Darmstadt, Germany
Phone: +49 61 51 - 81 02-0
06/2020