GEAR DESIGN:

Given:

RATE VOLTAGE = 12V

No. Load Speed = 3000-3500 rpm

No. Load Current = 0.8A (starting current)

Current Rating: 2.5A

Power of motor @ max. of 12V

P= VI
Gear Computation:

tout Nin Pout


Gear Ratio = = =
Figure 10. Gear tin Nout Pin

Design 16 t
Gear Ratio = =1
16 t

P= 12V x 2.5A Gear Ratio = 1:1

Nin 3500rpm
P= 30W Gear Ratio = ; Nout = = 3500rpm
Nout 1

P=2 π TN
tin = 16
J
30 =2 π ( T ) ¿)
tout = 16t s

T = 0.08185 N-m

17
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

DESIGN PARTS OF THE VENTILATOR MACHINE (CM)

BASE PLATE (Scale 1:2)

Isometric
Front view

Side View

Side View

WALLS

Front view
Isometric

Side View Top View

18
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

LCD WALL

Isometric

Front view

Side View

Top View

FRAME 1

Front view Bottom View Isometric

Top View
Isometric

Front view

Top View Bottom View

19
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

FRAME 2

Front view Isometric


Bottom View

Top View Isometric

Top View Bottom View


Front view

CAM

Front view
Isometric

20
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

HINGE

Front view
Isometric

Isometric
Front view

Bottom View

21
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

COUPLING AND GEAR (5 mm Thickness)

Front view Front view Front view

Front view Front view Front view

Side view

Side view Side view

22
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

3D DESIGN OF THE VENTILATOR MACHINE(Solidworks)

23
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

a. Isometric View

24
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

b. Perspective View

I:E (1:2) RATIO FOR AGE CATEGORY

25
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

Figure 11. Calculation of ratio of Breathing Cycle According to the age of a patients.

A normal respiratory rate in healthy adults is roughly 12 to 20 breaths per

minute. Your respiratory rate is an important vital sign. It can potentially indicate a

more serious condition such as cardiac arrest.

If your respiratory rate is below average, it could indicate central nervous

system dysfunction. If your respiratory rate is above average, it could indicate another

underlying condition. [6]

26
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

CALIBRATION OF MOTOR SPEED BASED ON BREATH CYCLE

Figure 12. Calculation of Motor Speed.

This calculation is done manually by the researchers, by setting a timer and getting

the time of each cam revolution.

27
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

PRICING COMPONENTS:

Figure 13. Pricing of the Material used in Ventilator.

Total estimated cost is Php.7,500.00 which is cheaper than the market price

of a ventilator that costs Php.15,000.00 to Php.300,000.

28
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

CHAPTER IV

RESULT AND DISCUSSION

Breath Cycle I:E Ratio (1:2)

Age category (years) (seconds) (seconds) Motor Speed (%)

      12 V 20V

Infant (birth -1) 1.33 0.4435:0.8865 N/A 72

Toddler (1-3) 1.875 0.6250:1.2500 85 54

Pre-School (3-6) 2.143 0.715:1.42786 80 50

School Age (6-12) 2.5 0.8335:1.6665 72 44

Adult (12 and above) 4 1.3350:2.6650 50 31

Figure 14. Summary of Breathing Cycle and Percent of Motor Speed.

After the fabrication of the ventilator, the researchers calibrated the appropriate

motor speed to be used in the device in order to achieve the desired breath cycle for each age

category. The calibration is made manually and two power sources has been used, a 12V

power source and 20V as a power source. For the 12V power source, the calibrated motor for

toddlers, pre-school, school-age, and adults are 85, 80, 72, and 50 respectively; as for the

infant, the researcher did not achieve the desired breath cycle (1.33 seconds) because the full

speed of the motor can only produce a 1.6 seconds breath cycle. On the other hand, for the

20V power source, the calibrated motor speed from infant to adult is 72, 54, 50, 44, and 31

percent respectively.

29
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

FINISH PRODUCT

30
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

a. Front View b. Side View

c. Top View d. Perspective View

CONCLUSION AND RECOMMENDATION

The researchers successfully designed and fabricated a ventilator that achieves their

objectives, a device that is not only identical to those machines available in the market but

also features the same efficiency. In addition, a device that is cost-effective and easy to

handle; moreover, this will help to fill in the shortage in numbers of ventilator if it puts into

production.

31
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

Somehow, the researchers still find many deficiencies in the device because they

haven’t enough resources and time to address these shortcomings. The researchers knew that

the device is bulky and redesigning the ventilation will be possible to make it more compact

in size. Another problem they encounter is the produced noise when the cam and compressor

come into contact, to address it, adding spring as suspension in the lower part of the

compressor in order to balance the load produce by the cam in order to create smoother

interaction between the cam and follower. The researchers also consider the oxygen source,

they knew that the device did not produces pure oxygen; as a solution, an air filter will be

placed in the air intake making it safer. A casing will be use also to avoid any disturbances

that will threaten the performance of the device.

The device supposedly has a replaceable ambu bag to make it applicable for every age

category.

If these shortcomings will be resolve, the researchers believe that the device will be

more efficient and ideally, mass production of the device will help to increase the numbers

ventilators.

DESIGN OF ADJUSTABLE HINGE FOR AMBU BAG (TODDLER)

32
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

Figure 15. Design for Adjustable Hinge

Figure 16. Design for Adjustable Hinge

33
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

REFERENCE:

Abdul Mohsen Al Husseini, Heon Ju Lee, Justin Negrete, Stephen Powelson, Amelia Servi,
[1]

Alexander Slocum , Jussi Saukkonen (April 2010) DESIGN AND PROTOTYPING OF


LOW-COST PORTABLE MECHANICAL VENTILATOR.
Robinder G. Khemani, Christopher J.L. Newth ( 02 July, 2010) THE DESIGN OF
[2]

FUTURE PEDIATRIC MECHANICAL VENTILATION TRIALS FOR ACUTE LUNG


INJURY.
https://www.atsjournals.org/doi/full/10.1164/rccm.201004-0606CI

H. Seddik and A. M. Eldeib, "A wireless real-time remote control and tele-monitoring
[3]

system for mechanical ventilators," 2016 8th Cairo International Biomedical Engineering
Conference (CIBEC), 2016, pp. 64-67, doi: 10.1109/CIBEC.2016.7836121.
https://ieeexplore.ieee.org/abstract/document/7836121

M. Shahid, "Prototyping of Artificial Respiration Machine Using AMBU Bag


[4]

Compression," 2019 International Conference on Electronics, Information, and


Communication (ICEIC), 2019, pp. 1-6, doi: 10.23919/ELINFOCOM.2019.8706360.
https://ieeexplore.ieee.org/abstract/document/8706360

Aliaksei Petsiuk, Nagendra G. Tanikella, Samantha Dertinger, Adam Pringle, Shane


[5]

Oberloier, Joshua M. Pearce “Partially RepRapable automated open source bag valve mask-
based ventilator” (October 2020)
https://www.sciencedirect.com/science/article/pii/S2468067220300407

https://www.healthline.com/health/normal-respiratory-rate#normal-rate-in-kids
[6]

34
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

https://youtu.be/k5cwt6EC1kc

https://www.ncbi.nlm.nih.gov/books/NBK535395/?fbclid=IwAR2NY-r4-

t5T7cq39bxrrOKGSLs0Xq9XARgR9sMbwoCy06nMZ1d11Cmoldw

https://www.sciencedirect.com/science/article/pii/S2405844021022982?

fbclid=IwAR3PTLj3w2ZAL6VXnJicoVH3B4fIsWCvXnTIYXVFpersm0Ll0s6qKvzMp5Y

https://www.nursingtimes.net/clinical-archive/respiratory-clinical-archive/every-breath-you-

take-the-process-of-breathing-explained-08-01-2018/?

fbclid=IwAR1b89ITlSqSmR85PF9SjheInqwN7aTqfwl8F5pDno255tL7DoGN_v81yaA#:~:te

xt=When%20the%20diaphragm%20contracts%2C%20it,air%20is%20expelled

%20(exhalation)

https://www.cs.cmu.edu/~rapidproto/mechanisms/chpt6.html?fbclid=IwAR2uxUi-

1ZZsLSAU0a2z_DV2DYRwWAbhSmYvDBZ7RSo5oq3GxPTvwY-O_sI

https://ocw.metu.edu.tr/pluginfile.php/6886/mod_resource/content/1/ch8/8-3.htm?
fbclid=IwAR3ytHYR_oP9_fndOHCd1K-tYndC_T7EP5_vF3WRISActluwuvQPagZhDbs

35
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

APPENDICES

36
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

37
Republic of the Philippines
Aurora State College of Technology
Zabali, Baler, Aurora

38

You might also like