You are on page 1of 5

Rendered PDF File Page 1 of 5 DOC1138106, Rev:1

Electronic Signature Information

Name DOC1138106

Revision 1

Type Controlled Document

Title BRIVO XR115 Data Sheet

Originator 305018642_manoj_kumar_singh m b

Release Date

Obsolete Date

Name Reason For Change File Size (Bytes)


BRIVO XR115 Data Sheet v2.doc Brivo XR115 Data Sheet 1314816

Route Signer Function Status Comments Completion Date


R- 19 Jul 2012
305018642_manoj_kumar_singh m b LSD Approve OK
5066860 09:50:39 GMT
Global
R- 22 Aug 2012
212066625_christopher_g_bona MarComm Approve Approve
5066926 20:10:42 GMT
Manager
R- Reviewed and 24 Jul 2012
305008624_sadashiva_kantota_subraya Program RA Approve
5066926 approved 12:56:49 GMT
Approved based
R- 23 Aug 2012
305024142_dinesh__charak Legal Approve on Product RA
5066926 05:46:13 GMT
Approval

Periodic Review
There are no signatures or routes related to this business object.

Obsolesence Approval
There are no signatures or routes related to this business object.

* Printed versions are For Reference Only *


+ Indicates a task was reassigned from an original assignee
Approved
Rendered PDF File Page 2 of 5 DOC1138106, Rev:1

BRIVO XR115 Mobile Radiographic System

Data Sheet

Product Description

BRIVO XR115 is high-frequency analogue radiographic mobile X-ray equipment capable of generating
and controlling X-rays for diagnostic procedures when plugged into electrical outlet. The equipment can
be moved manually within the center to perform radiographic exam.

The targeted clinical users include qualified trained doctors, radiographers or radiologic technicians/
technologists, or other qualified Health Care Professionals.
Approved
Rendered PDF File Page 3 of 5 DOC1138106, Rev:1

Brief Information on the Capabilities


Brivo XR115 is indicated for use on adult and pediatric patients for general-purpose diagnostic
radiographic examinations and procedures. Its mobility enables use as needed within the emergency,
intensive care, cardiac and operating departments, and premature birth ward, for patients that may not
be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD
room. Brivo XR115 can be used in Intensive Care Units, Neonatal Intensive Care Units, Pediatric
Intensive Care Units, Emergency Room, Operating Room and Radiology department, in Hospital, Nursing
home/Retirement facility, Outpatient clinics and Diagnostic Centers.

Brivo XR115 is indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen,
pelvis and extremities, with the patient sitting, standing, or lying in the prone or supine position. The
system has medical applications ranging from but not limited to cranial, skeletal, thoracic, and lung
exposures.

Product Specifications
System
Input voltage range 230VAC +/-10%
Max current < 16 Amps
Boot time < 10 seconds
Reset time < 5 seconds
Duty cycle time 1 : 30

Generator/Tube
Output Power 4 kW
System frequency 200 kHz
kV range 40 to 100 kV
mA range 10 to 100 mA
mAs range 0.1 to 250 mAs
Exposure time range 10ms to 5sec @230 V AC
kV step Jump step of 2kV for 40-70kV, then step of 5kV till 100kV
mAs step Renard scale
Tube type Stationary anode
Tube focal spot 1.8mm
Inherent filtration <= 1.5mm Al

Collimator
Collimator rotation +90 to -90 degrees
Collimator lamp lux 160 lux @ 1m SID (As per IEC)
FOV adjustment Knob-operated collimator shutters for manual FOV adjustment
Timer for lamp Auto off <=30 seconds
Lamp watt 100W (+/-5% variation)
Approved
Rendered PDF File Page 4 of 5 DOC1138106, Rev:1

Physical specifications
Dimensions (L x W x H) <= (1280 x 600 x 1500) mm
Weight < 120 Kgs
System push/pull force < 20N (~2kgf)
Arm up/down force < 40N (~4Kgf)
Turning radius <= 1.04m
Ground clearance >= 100mm in parked condition
Positioner type Swivel arm mechanism
Brake type Foot-operated parking brakes
Tube arm bracket rotation +/- 180 degrees
Tube rotation within bracket -45 to +180 degrees
Focus to Floor distance <= 400mm to >= 2000mm

Human-Device interfaces
Control panel Control panel with membrane keypad with switches of standard size
Display The digital display has the interface to adjust & display kV and mAs
parameters. It indicates System ON, System OFF, status and fault
messages on the kV and mAs area
Console The console has provision to change the kV and mAs parameters.
Provision provided for resetting the current error.
Alarms The console informs the user once prepare is completed by audio &
visual indications. During exposure, the user is informed through
audiovisual indications. In case, error occurs during exposure, user shall
be informed by continuous alarm with fault indicator.

Other
Technique selection 2-point technique (kV, mAs)
Cassette holder Up to 6 cassettes
Mains cable length Up to 4.5 meter
SID measuring device Measuring tape of length 2 meter minimum
FOV @ 1m SID Minimum 43cm x 43cm
Handles System push handle at <95 cm from floor
Handle for Tube head rotation
Handle for Collimator rotation
Accessories holder Provided
Mains Cable winder Provided
Approved
Rendered PDF File Page 5 of 5 DOC1138106, Rev:1

Dimensional Drawing
Approved