Professional Documents
Culture Documents
Date: August 01
with solubility are not significant until you are Analysers which measure cold NOx and the NO2
measuring on the 100 ppm range. converter is after any condensate removal device has
At this level great care must be taken to ensure that probably lost 50% of the NO2 in the condensator.
all sampling components are heated to ensure that This, as previously stated, is only of real
NO condensation occurs and that the water removal consideration if the NO2 is a large constituent of NOx
device removes all free water on a continuous basis. in gas turbines.
Calibrating the analyser directly on calibration gas will HCL
set the measuring gain of the analyser but This is perhaps the most tricky of all.
subsequent connections of the same calibration gas
to the sampling point may indicate a lower reading if HCL is very soluble in water and to measure HCL
there is any water in the sampling system which will correctly the whole sampling train must replicate the
have dissolved some of the calibration gas. temperature of the stack gas. Thus heated lines,
filters, pumps, measuring cells etc. must be at least
The fact that the calibration gas comes from a
190°C.
cylinder and is very dry, makes this situation worse as
it will take many hours for this gas to equilibrate with Again, though, the calibration gas is going to be very
the water vapour present in the sampling system dry coming from a gas cylinder at 250 bar initially,
following stack gas sampling. when introducing HCL cal gas to the sampling point it
will “pull” all moisture out of the sample tubing etc.
At the 100 ppm range of measurement all of the
and will dissolve the HCL in this moisture, producing
errors of solubility start to account for a lot. Using an
hydrochloric acid vapour.
NDVV analyser allows for totally heated sampling but
Infra red requires for the water to be removed from After some time however this will equilibrate and
the sample which presents the water condensate into allow a stable reading on the analyser.
which the SO2 is dissolved. NH3
The only practical solution is to allow adequate time This largely follows the same as HCL.
for the cal gas to equilibrate until it reads through the
sampling system the same as it does when Note
calibrating the analyser directly. (1) As the introduction of soluble acid gases
NO2 into the sampling point introduces more questions
than answers, it is probably more effective to
Most NO2 solubility problems do not exist as the introduce these gases directly to the analyser for gain
majority of analysers measure NO and this is less setting and allow adequate sampling period to obtain
soluble in water. NO2 is converted into NO in the a stable reading on the analysers. The introduction of
heated part of the sampling system or analyser. a dry cal gas to the sampling point will dry out the
The only requirement is to check the NO2 converter sampling system and lose most of the cal gas and
efficiency periodically as described earlier. then switching to sample mode reverses the whole
process until this has equilibrated.