You are on page 1of 2

Document No:ES002

Date: August 01

Calibration of Gas Analysers


There are four separate aspects of calibration, as NPL offer gases traceable to National Standards and
follows:- the USA NIST and UKAS.
1. Foreword You may choose to trust the certification of your gas
2. Analyser Gain Setting/Zero Setting supplier, you may choose to have some NPL gases
to cross check from time to time the supplies from
3. System Verification
your gas supplier or you may obtain your gases from
4. Linearity of Measuring Range NPL for all your work. UKAS approved gases
5. Traceability of Calibration Gas normally means that they have been cross
6. NOx Generator referenced to NPL gases at the calibration lab of the
7. Soluble Gases gas supplier.
Foreword Cross reference means that a linearized analyser in
the lab is calibrated using an NPL calibration gas and
Gas analysers are not absolute in measurement, they then the gas suppliers own gas is introduced to the
only give a physical reading and need to be same analyser and the reading is written on the
compared with a gas of known concentration in order suppliers certificate. This is the traceability.
to provide a reading of any calibrated amount.
NOx Generator
Zeroing
Most NOx analysers only measure NO (Nitric Oxide)
The zero adjustment is normally done first using a the NO2 part of NOx is converted to NO using a NO2
gas (typically Nitrogen) which is known to have no to NO converter. NO2 + NO = NOx. From time to
gas being measured in it. You must set zero first time it is important to check if the NO2 converter is
because this will affect reading throughout all working.
measuring levels and shown as an offset.
The amount of NO2 as a percentage of NOx varies
Analyser Gain Setting with application. From normal combustion sources
Each analyser has a detector which responds to the NO2 is about 5% of the NOx but from gas turbines the
gas of interest. In order for the analyser reading to NO2 is nearer 50% due to the amount of air in the
be calibrated, a calibration gas of known combustion process. Also, as NO comes from
concentration is introduced directly to the analyser combustion and moves into the atmosphere it
and the reading adjusted by the analyser gain generally becomes NO2 in a very short period of time
controls so that it reads the correct value of the gas in the oxygen laden atmosphere.
cylinders. Therefore, the requirement for converter checking is
System Verification most critical with gas turbine combustion emissions.
Normally, a gas analyser is sent a sample via a In order to check a NO2 converter you must use NO2
sample line, filtration, cooler, pump etc. calibration gas passing through the NO2 converter to
verify that it is reading at least 95% of its value on the
In order to ensure that the sampling train integrity is analyser compared to that of NO calibration gas.
correct, it is very good practice to connect the
calibration gas to the sample point (ensuring no To avoid errors in two gases it is more accurate to
leaks) and have it passed right through to the use a noxgen instrument.
analyser or analysers. This device uses ozone, internally produced, to
Any difference between the direct analyser calibration change (oxidise) the NO calibration gas to NO2 in a
and through the sample point will indicate a problem. one for one reaction. The device finally allows you to
send NO calibration gas directly to the analyser for
The problem could be leaks, absorption, dilution, gain setting and this is done in the NO mode i.e.
solubility etc. in the sample system and with this bypassing the NO2 converter.
produce an error in the reading of the analyser.
The noxgen allows you to then start ozone generation
Normally, the analyser zero/gain settings against in increments so that the NO is converted to NO2.
calibration gas is adequate for routine calibration of The reading on the analyser in NO mode starts to
analysers. reduce as the NO is changed to NO2.
The system verification should also be carried out Once you have reached a certain level you change
prior to any test program. the analyser mode to NOx which passes the
Linearity of Measuring Range calibration gas through the NO2 converter and you
must see at least a 95% increase from the reduced
Calibrating analysers at one point is not enough. For
reading. The NO2 converter having changed the NO2
accurate readings it is important to audit the whole
back to NO.
range of the analyser using several separate points.
Soluble Gases
Checking this linearity is usually recommended on a
six monthly basis and it is not normally necessary on The most problematical of calibrations is with soluble
a more regular basis. gases - those which dissolve in water.
You can check this by using several separate Those given which are soluble in water and normally
calibration gases of different concentrations. associated with stack gas measurement are as
Normally 10 points is required. follows:-
A gas divider is a more convenient and accurate way A SO2
of splitting one calibration gas into 10 equal steps B NO2
which can then check the whole range of the
analyser. C HCL
Traceability D NH3
Obviously the calibration gas must be of the highest SO2 is normally found in relatively high
accuracy and all gas suppliers quote their accuracies. concentrations in coal/oil combustion and the errors

Source Testing Association


Unit 4, Theobald Business Centre, Knowl Piece,
Hitchin, Herts, SG4 0TY Tel +44(0)1462 457535
Email Admin@S-T-A.org Fax +44(0)1462 457157
Document No:TGN001

with solubility are not significant until you are Analysers which measure cold NOx and the NO2
measuring on the 100 ppm range. converter is after any condensate removal device has
At this level great care must be taken to ensure that probably lost 50% of the NO2 in the condensator.
all sampling components are heated to ensure that This, as previously stated, is only of real
NO condensation occurs and that the water removal consideration if the NO2 is a large constituent of NOx
device removes all free water on a continuous basis. in gas turbines.
Calibrating the analyser directly on calibration gas will HCL
set the measuring gain of the analyser but This is perhaps the most tricky of all.
subsequent connections of the same calibration gas
to the sampling point may indicate a lower reading if HCL is very soluble in water and to measure HCL
there is any water in the sampling system which will correctly the whole sampling train must replicate the
have dissolved some of the calibration gas. temperature of the stack gas. Thus heated lines,
filters, pumps, measuring cells etc. must be at least
The fact that the calibration gas comes from a
190°C.
cylinder and is very dry, makes this situation worse as
it will take many hours for this gas to equilibrate with Again, though, the calibration gas is going to be very
the water vapour present in the sampling system dry coming from a gas cylinder at 250 bar initially,
following stack gas sampling. when introducing HCL cal gas to the sampling point it
will “pull” all moisture out of the sample tubing etc.
At the 100 ppm range of measurement all of the
and will dissolve the HCL in this moisture, producing
errors of solubility start to account for a lot. Using an
hydrochloric acid vapour.
NDVV analyser allows for totally heated sampling but
Infra red requires for the water to be removed from After some time however this will equilibrate and
the sample which presents the water condensate into allow a stable reading on the analyser.
which the SO2 is dissolved. NH3
The only practical solution is to allow adequate time This largely follows the same as HCL.
for the cal gas to equilibrate until it reads through the
sampling system the same as it does when Note
calibrating the analyser directly. (1) As the introduction of soluble acid gases
NO2 into the sampling point introduces more questions
than answers, it is probably more effective to
Most NO2 solubility problems do not exist as the introduce these gases directly to the analyser for gain
majority of analysers measure NO and this is less setting and allow adequate sampling period to obtain
soluble in water. NO2 is converted into NO in the a stable reading on the analysers. The introduction of
heated part of the sampling system or analyser. a dry cal gas to the sampling point will dry out the
The only requirement is to check the NO2 converter sampling system and lose most of the cal gas and
efficiency periodically as described earlier. then switching to sample mode reverses the whole
process until this has equilibrated.

You might also like