Operation..……………………………………………………………………………………………………………………………….7
Introduction ………………………………………………………………………………………………………………….………7
Absorbance ………………..….……………………………..…………………………………………………………………….10
Transmission ……………………..………………………………………………………………………………………………..13
Folder Names……………………………………………………………………….……………………………………………..26
Lock Folder……………………………………………………………………….………………………………………………..27
Unlock Folder……………………………………………………………………….……………………………………………..27
Lock Method……………………………………………………………………….………………………………………………28
Unlock Method……………………………………………………………………….……………………………………………29
Regional ……………………………………………………………………………………………………………………………33
Preference …………………………………………………………………………………………………………………………34
Contrast ……………………………………………………………………….……………………………………………………35
About ………………………………………………………………………………………….…………………………………….35
2|P age
Print Via Computer ………………………………………………………...………………………………………………………….36
Installation ………………………………………………………………………………………………………………………….36
Accessories ……………………...…………………………………………………………………………………………………….38
3|P age
Essential Safety Notes
There are a number of warning labels and symbols on your instrument. These are there to inform you where potential
danger exists or particular caution is required. Before commencing installation, please take time to familiarize yourself
with these symbols and their meaning.
Check the contents of the package against the packing list. If any shortages are discovered, inform your supplier
immediately.
Inspect the instrument for any signs of damage caused in transit. If any damage is discovered, inform your supplier
immediately.
Ensure your proposed installation site conforms to the environmental conditions for safe operation:
Indoor use only.
Temperature range 5°C to 35°C. Note that if you use the instrument in a room subjected to extremes of
temperature change during the day, it may be necessary to recalibrate (by switching off and then on again).
Maximum relative humidity of 80% up to 31°C decreasing linearly to 50% at 40°C.
The instrument must be placed on a stable, level bench or table that can take its weight (1.55 kg) to allow air to
circulate freely around the instrument.
This equipment must be connected to the power supply with the power cord supplied. It can be used on 90 – 240 V~,
50-60 Hz supplies.
If the instrument has just been unpacked or has been stored in a cold environment, it should be allowed to come to
thermal equilibrium for 2-3 hours in the laboratory before turning the instrument on. This will prevent calibration failure
as a result of internal condensation.
Switch on the instrument via the keypad ( ) after it has been plugged in. The instrument will perform a series of
self-diagnostic checks.
Please read through this user manual prior to use.
Please contact your original supplier in the first instance if you experience technical or sample handling difficulties.
If this equipment is used in a manner not specified or in environmental conditions not appropriate for safe operation, the
protection provided by the equipment may be impaired and instrument warranty withdrawn.
4|P age
Warranty and Repair
Biochrom warrants this instrument for a period of 12 months (1 year) from the date of purchase. Where appropriate,
Biochrom will repair or replace the unit for defects of workmanship or materials. This warranty does not extend to damage
resulting from misuse, neglect, or abuse, normal wear and tear, or accidental damage. This warranty extends only to the
original consumer purchaser.
Products failing within the first 30 days of end user operation are considered dead on arrival (DOA) and where appropriate
a replacement will be given if a repair is not possible. In the instance of a DOA; Biochrom will incur shipping charges of
returning the unit back to the facility.
IN NO EVENT SHALL BIOCHROM BE LIABLE FOR INCIDENTAL OR CONSEQUENTAL DAMAGES. Some states do
not allow the exclusion or limitation of incidental or consequential damages so the above limitation to exclusion may not
apply to you. THERE ARE NO IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR USE,
OR OF ANY OTHER NATURE. Some states do not allow this limitation on an implied warranty, so the above limitation
may not apply to you.
If any defect arises within or outside the warranty period, please contact:
Goods will not be accepted for return unless a RMA (Returns Materials Authorization) number has been issued by our
Returns-Repairs Department. The customer is responsible for shipping charges. Please allow a reasonable period of time
for completion of repairs or replacement.
5|P age
Technical Specifications
Sample ID Yes
English, German, French, Spanish, Italian,
Languages
Japanese, Chinese
Dimensions 120 x 230 x 320 mm
Weight 1.55 kg
18 Vdc at max 18VA from a supplied 100 to
Power input
240V~, 50/60 Hz Mains Power Adapter
6|P age
Operation
Introduction
Your spectrophotometer is a simple-to-use visible instrument. It has no moving parts, which is the basis of the rapid
scanning operating system. It has been designed to meet the routine spectroscopy needs of customers requiring a
compact instrument that is easy to use. The product is reliable and requires low maintenance.
After it is turned on and calibrated the Novaspec III+ will show an absorbance measurement display.
To select different measurement modes, change parameters or change other settings, click on the menu key and
select the required menu item by clicking on the associated menu number using the keypad.
Measurement Modes Absorbance, Transmission, Concentration with a Factor, Concentration with a Standard,
Rate (Kinetics)
Methods Contains eight folders that can store less frequently used configured methods (eight
methods per folder)
Utilities Instrument set up (date, time, language, data output, preferences and display settings)
The instrument is supplied with a program, PVC (Print via Computer) on the accompanying CD. When used with a USB
cable to connect to a PC onto which the software has been installed, it enables the user to “print through” the PC directly
to the printer that is connected to it. The data may also be stored as an Excel spreadsheet, as an EMF graphics file, a
comma delimited (csv) data file, a tab delimited (txt) data file or in native PVC format for later access.
The instrument has a USB connector located at the front of the unit which allows data to be stored on a USB memory
stick. This data can be stored in a PVC format which can be opened using PVC software, a comma separated variable
(CSV) data file or a tab separated variable format which can be imported into Excel.
7|P a ge
Keypad and Display
The back-lit liquid crystal display is very easy to navigate around using the alphanumeric entry and navigation arrow keys
on the hard wearing, spill proof membrane keypad.
Alphanumeric Keys
Escape/Cancel
On/Standby OK Key
Key
Set Reference
Arrow Keys
Key Action
On/Standby
Turns the instrument on or back to standby.
Key
Arrow Keys Use the four arrow keys to navigate around the display and select the required
setting for the active (highlighted) option.
View Options
Used to view current options.
OK
Confirm parameters/selection.
8|P age
Software Style
The user interface is built around the options key . The instrument will open automatically to an
application and you will use the options key to select different parameters.
Allows you to toggle back and forth between how the results are
3 Graph
given, digital versus a graph.
Allows you to measure your standard when in Concentration
4 Run Standard
Standard measurement mode.
Instrument set up (date, time, language, data output,
5 Utilities
preferences and display settings).
Sub folder selection or user selected and configured method (8
sub folders each containing a maximum of 9 methods). To
6 Methods
return to the previous measurement mode, select the methods
followed by “Esc” button.
7 Sample Number Allows you to name your sample manually or can be prompted.
Use the left and right arrow key to select a folder to store in and
8 Save Method
using the alphanumeric keypad to name your method.
9|P a ge
Parameters
Absorbance
Step 1
Step 3
If not already in Absorbance mode, use left or right
arrow keys to select Absorbance mode.
Press the down arrow key.
Step 4
Use the left and right arrow keys to change the
wavelength, or use the number keypad.
Press the down arrow key.
Step 5
If required, enter a sample name. This name
followed by an incrementing sample number will be
used for naming subsequent results.
Press the down arrow key.
Step 6
Set this option to yes, if you need to measure many
different samples and assign different sample
names. The prompt will, by default, contain the
sample name last entered.
Press the down arrow key.
Step 7
10 | P a g e
Step 8
Step 9
Step 10
11 | P a g e
Options (select using key pad numbers)
12 | P a g e
Transmission
Step 1
Step 3
If not already in Transmission mode, use the left
and right arrow keys to choose Transmission
mode.
Press the down arrow key.
Step 4
Use the left and right arrow keys to change the
wavelength, or use the number keypad.
Press the down arrow key.
Step 5
If required enter a sample name. This name
followed by an incrementing sample number will
be used for naming subsequent results.
Press the down arrow key.
Step 6
Set this option to yes, if you need to measure
many different samples and assign different
sample names. The prompt will, by default,
contain the sample name last entered.
Press the down arrow key.
Step 7
13 | P a g e
Step 8
Step 9
Step 10
14 | P a g e
Options (select using key pad numbers)
15 | P a g e
Concentration with a Factor
Entering a known factor.
Step 1
Step 3
If not already in Conc. Factor mode, use left or right
arrow keys to Conc Factor mode.
Press the down arrow key.
Step 4
Use the left and right arrow keys to change the
wavelength, or use the number keypad.
Press the down arrow key.
Step 5
Using the keypad, enter the factor.
Press the down arrow key.
Step 6
The units can be manually written in using the
16 | P a g e
Press ok to save and exit the units
selection.
Step 7
If required, enter a sample name. This name
followed by an incrementing sample number will be
used for naming subsequent results.
Press the down arrow key.
Step 8
Set the sample option to yes, if you need to
measure many different samples and assign
different sample names. The prompt will, by
default, contain the sample name last entered.
Press the down arrow key.
Step 9
Step 10
Step 11
17 | P a g e
Step 12
18 | P a g e
Concentration with a Standard
Single point calibration curve with known standard and need to calculate a factor.
Step 1
Step 3
If not already in Conc. Standard mode use left or
right arrow keys to Conc Standard mode.
Press the down arrow key.
Step 4
Use the left and right arrow keys to change the
wavelength, or use the number keypad.
Press the down arrow key.
Step 5
Using the keypad, enter the concentration.
Press the down arrow key.
Step 6
The units can be manually written in using the
19 | P a g e
Press ok to save and exit the units
selection.
Step 7
If required enter a sample name. This name
followed by an incrementing sample number will
be used for naming subsequent results. Press the
down arrow key.
Step 8
Set this option to yes, if you need to measure
many different samples and assign different
sample names. The prompt will, by default,
contain the sample name last entered.
Press the down arrow key.
Step 9
Step 10
Step 11
Step 12
A Run Standard screen will appear, confirm
concentration or enter new concentration. Press
ok .
20 | P a g e
Step 13
21 | P a g e
Rate
This application has a fixed 10 second interval. I twill keep collecting data until you manually stop it by pressing the C
button.
Step 1
Step 3
Use the left and right arrow keys to choose Rate
mode.
Press the down arrow key.
Step 4
Use the left and right arrow keys to change the
wavelength, or use the number keypad.
Press the down arrow key.
Step 5
If required enter a sample name. This name
followed by an incrementing sample number will be
used for naming subsequent results. Press the
down arrow key.
Step 6
Set this option to yes, if you need to measure many
different samples and assign different sample
names. The prompt will, by default, contain the
sample name last entered .
Press the down arrow key.
Step 7
22 | P a g e
Step 8
Step 9
Step 10
Step 11
23 | P a g e
Options (select using key pad numbers)
24 | P a g e
Methods Folder
These folders are the storage locations for any user modified Applications (Methods) that are saved in the Options menu.
There are 8 total folders and can store up to 9 methods per folder. In addition, if a USB memory stick has been inserted,
it is possible to store on the memory stick.
Saved methods can be locked, unlocked and deleted using the Options menu. Select the method by pressing the relevant
Saving a Method
25 | P a g e
Use the left and right arrow keys to save the method
to a methods folder (1-8) or to a USB memory stick is
a USB stock is inserted.
1. Folder Names
2. Lock Folder
3. Unlock Folder
4. USB Memory Stick
Folder Names
26 | P a g e
Lock Folder
Press 2 to select Lock Folder.
Press the down arrow key.
Unlock Folder
27 | P a g e
Within a methods folder, press to display the
following options:
1. Delete Method
2. Lock Method
3. Unlock Method
Delete Method
Lock Method
Press 2 to select lock method.
28 | P a g e
Unlock Method
Press 3 to select unlock method.
29 | P a g e
Saving Data onto a USB Memory Stick
When a USB memory stick is inserted into the connector in front of the instrument an audible click will be heard as the
stick is recognized. Whenever a USB stick is inserted, all data will automatically be saved to the USB memory stick. The
format used is selected from the utilities menu under printer options. The options available are PVC format (the PC PVC
program is used to open the data which can then be printed or exported). TSV tab separated format or CSV comma
separator format can be imported into programs such as Excel. However, CSV format should not be selected when the
number format is set to use a comma instead of a decimal point, otherwise, the data will not load cleanly into Excel.
30 | P a g e
Data is stored on the USB memory stick under the following directory structure
Double clicking on a file opens it into the PVC application from which it can be exported or saved. Full details on the use
of PVC are covered in the PVC user manual
31 | P a g e
Utilities Folder
32 | P a g e
1. Date and Time
2. Regional
33 | P a g e
3. Export Data
4. Preference
34 | P a g e
5. Contrast
6. About
35 | P a g e
Print Via Computer (PVC)
This PC application running on a Windows computer can be used to transfer result data from an attached
instrument. The result data transferred can be printed or stored in a variety of formats including Excel
spreadsheet, graphical (EMF), comma delimited text (CSV), tab delimited text (TXT), rich text format (RTF)
and PVC.
PVC can store data either to a common data folder or be configured to save to independent directories
according to both file format and instrument connected
Data stored in PVC format can be retrieved and printed or exported into any of the supported formats
Installation
See the manual included on the PVC CDROM for installation and operating instructions.
36 | P a g e
Test Tube and Heated Cell Holder Installation
Installation
1. Lay the instrument upside down on a soft surface.
2. Unscrew the two screws and remove the cuvette cell holder by tilting the holder to allow the cell ring to clear the
baseplate.
Back of instrument
Front of instrument
3. There are two positioning areas to help indicate the orientation of the cell holders when installing. The cell holders
will not fit unless this is correctly aligned.
4. Insert the new cell holder and screw in the two screws.
Back of instrument
Front of instrument
37 | P a g e
Accessories
38 | P a g e
Icon Glossary
Status Bar Icons Meaning
Measurement in progress (lamp lit). The icons are cycled through to indicate the lamp
coming up to full brightness and then shown in reverse when the lamp is turned off.
Display and
Meaning
Menu Icons
About - used to view unit identification information.
39 | P a g e
Preferences setup.
Regional settings.
Data export setup defines the settings for the “Print” output to PC and USB memory stick.
Menu folder.
Display icons to
represent
measurement Meaning
modes
40 | P a g e
China
Biochrom
Room 1902E 19F, Building B
Zhong Shan Plaza
1065 West Zhig Shan Road
Changning District
Shanghai, China 200051
Telephone +86 21 2230 5128
Email support@hbiosci.com
Website www.biochromspectros.com
United Kingdom
Biochrom
Building 1020
Cambourne Business Park
Cambourne
Cambridge, CB23 6DW, United Kingdom
Telephone +44 (0) 1223 423 723
Fax +44 (0) 1223 4220 164
Email support@hbiosci.com
Website www.biochromspectros.com
USA
Biochrom US
84 October Hill Road
Holliston, Massachusetts 01746
Telephone (508) 893 3999
Toll Free (800) 272 2775
Fax (508) 429 5732
Email support@hbiosci.com
Website www.biochromspectros.com
41 | P a g e