Abbott i-STAT Troponin I ERROR CODES

STARTUP MESSAGES
The analyzer performs self-checks when it is turned on. If a condition that should be
corrected in the near future, but that will not affect results, is detected, a warning is
displayed. The operator presses the 1 key to access the Test Menu. The analyzer can be
customized to lock out the operator until the corrective action is taken.

MESSAGE ACTION
Electronic Simulator Test Required Insert Electronic Simulator.
PCx Glucose Strip Control Required Test controls.
Stored Memory Low Place analyzer in Downloader.
Stored Memory Full Place analyzer in Downloader.
Upload Required Place analyzer in Downloader.
Battery Low Replace batteries or recharge battery.
CLEW Expiring, Update Required Upgrade software.

GENERAL
MESSAGE CAUSE ACTION
Strip Error A wet strip inserted, strip Repeat test using proper
pulled out during testing procedure. Check Analyzer
cycle, temperature of Status under the
analyzer exceeds Administration Menu for
operation temperature for analyzer temperature.
strip testing
No display Disposable batteries dead Change disposable batteries
or rechargeable battery or recharge battery. If still no
fully discharged. Keypad display, call Support
not responding. Start Services.
switch broken.
Cartridge Locked does not Dead battery(s). Wait until analyzer turns off.
disappear after test cycle Mechanical problem. Turn analyzer on. If resets,
completed remove cartridge. If not,
change or recharge
battery(s) and turn analyzer
on.

Abbott i-STAT Troponin I Error Codes IECTnI.2.EC 5.4.12


1
QUALITY CHECK MESSAGES AND CODES

MESSAGE CAUSE ACTION


Date Invalid, Check Clock Date outside six month Select 5-Clock Set from
lifetime of software. Administration Menu.
(Password protected.)
Dead Batteries, Replace Insufficient power to Replace disposable batteries
Batteries complete a test cycle. or recharge the rechargeable
battery.
Temperature Out of Temperature outside Check analyzer temperature
Range, Check Status operating range of 16 to by pressing 1 for Analyzer
page 30 °C. Status under the
Administration Menu. Move
analyzer to warmer area if
below operating range or to
cooler area if above the
range.
Invalid or Expired CLEW Software expired or Verify that the analyzer’s
corrupt. date is correct. Change
software if expired. Update
again if not expired.
Analyzer Interrupted, Use Last cartridge run not Check the battery pack
Another Cartridge completed. inserted properly. Check for
Low Battery startup warning.
Cartridge Error Usually problem with Use another cartridge. If
sample or cartridge filling. same code repeats more
than twice, try another
analyzer.
Cartridge Preburst Calibrant pack burst Use another cartridge - do
before cartridge inserted not press on center of
into analyzer. cartridge. Check that
cartridges have not been
frozen.
Unable to Position Cartridge not sealed. Clot Use another cartridge.
Sample in sample. Aberrant
cartridge.
Sample Positioned Short Cartridge underfilled. Use another cartridge - fill to
of Fill Mark Fill Mark.
Sample Positioned Cartridge overfilled. Use another cartridge - do
Beyond Fill Mark not fill beyond Fill Mark.
Test Cancelled by User did not respond to No action required.
Operator mandatory prompt before
analyzer time out.

Abbott i-STAT Troponin I Error Codes IECTnI.2.EC 5.4.12


2
Cartridge Type Not Software does not Update software. Check to
Recognized recognize cartridge. see if cartridges are expired.
Analyzer Error, Use Analyzer detects problem Insert the Electronic
Electronic Simulator from which it is likely to Simulator. If PASS, continue
recover. to use analyzer.
Analyzer Error, See Analyzer detects problem Insert Electronic Simulator. If
Manual from which it may not PASS, insert a cartridge with
recover. sample or control. If the code
does not reappear, continue
to use analyzer.

ADDITIONAL ASSISTANCE

If the problem persists after the corrective actions have been taken, contact your local
device supplier for technical assistance.

For unresolved issues contact APPN for further advice:

Email support@appn.net.au
Website www.appn.net.au using the forum or support tab
Phone (08) 8201 7845, Monday to Friday 0900 – 1700 hours (CST)

We will endeavor to assist you further with your concern.

The following information may be required:


• Device Name
• Test
• Test cartridge Lot number
• Control History and results
• Description of the problem with reference to information codes and error messages

Abbott i-STAT Troponin I Error Codes IECTnI.2.EC 5.4.12