STARTUP MESSAGES
The analyzer performs self-checks when it is turned on. If a condition that should be
corrected in the near future, but that will not affect results, is detected, a warning is
displayed. The operator presses the 1 key to access the Test Menu. The analyzer can be
customized to lock out the operator until the corrective action is taken.
MESSAGE ACTION
Electronic Simulator Test Required Insert Electronic Simulator.
PCx Glucose Strip Control Required Test controls.
Stored Memory Low Place analyzer in Downloader.
Stored Memory Full Place analyzer in Downloader.
Upload Required Place analyzer in Downloader.
Battery Low Replace batteries or recharge battery.
CLEW Expiring, Update Required Upgrade software.
GENERAL
MESSAGE CAUSE ACTION
Strip Error A wet strip inserted, strip Repeat test using proper
pulled out during testing procedure. Check Analyzer
cycle, temperature of Status under the
analyzer exceeds Administration Menu for
operation temperature for analyzer temperature.
strip testing
No display Disposable batteries dead Change disposable batteries
or rechargeable battery or recharge battery. If still no
fully discharged. Keypad display, call Support
not responding. Start Services.
switch broken.
Cartridge Locked does not Dead battery(s). Wait until analyzer turns off.
disappear after test cycle Mechanical problem. Turn analyzer on. If resets,
completed remove cartridge. If not,
change or recharge
battery(s) and turn analyzer
on.
ADDITIONAL ASSISTANCE
If the problem persists after the corrective actions have been taken, contact your local
device supplier for technical assistance.
Email support@appn.net.au
Website www.appn.net.au using the forum or support tab
Phone (08) 8201 7845, Monday to Friday 0900 – 1700 hours (CST)