SCAN-10

Automated Sampling Instrument

User Manual

Zhuhai DL Biotech Co., Ltd

0
Content
1 Features And Usages................................................................................................2
2 Peculiarity Of Instrument........................................................................................ 2
2.1 Normal Running Conditions............................................................................. 2
2.2 Basic Parameters............................................................................................... 2
2.3 Packing List.......................................................................................................2
3 Instrument Model And Structure Illustration...........................................................3
3.1 Instrument Model..............................................................................................3
3.2 Structure Illustration......................................................................................... 3
4 Instrument Operation and Instruction......................................................................4
5 Routine Maintenance............................................................................................... 7
6 Common Problem Treatments................................................................................. 8
7 After-sales service....................................................................................................9
8 Unpacking And Inspection...................................................................................... 9
9 Delivery And Storage.............................................................................................. 9

1
1 Features And Usages
To reduce labor cost and human negligence, Scan-10 Automated sampling instrument
is used to sample the test cards by micro-computer procedure.

2 Peculiarity Of Instrument
2.1 Normal Running Conditions

Environmental temperature:5℃-40℃

Relative humidity:≤80%

Power supply:AC220V±22V,50/60HZ

Atmospheric pressure:76kpa~106kpa

2.2 Basic Parameters

Application: All kinds of test cards

Sampling time:≤3.5mins/panel

Sampling Uniformity repeatability:≥90%。

Dimension:496×365×443mm

Weight:14KG

2.3 Packing List

Automated sampling instrument 1 set

Power supply 1 piece

Tip 5 pieces

Plastic bottle 3 vials

Manual 1 piece

2
3 Instrument Model And Structure Illustration
3.1 Instrument Model: SCAN-10

3.2 Structure Illustration

Spearhead

Tip Tip loading area

Test card
area
Buttons

Screen

Front View

Power socket

Back View

3
4 Instrument Operation and Instruction

1. Turn on the instrument power switch which is on the right side. After several
seconds, the test card selection menu will be displayed on the screen.

Direction buttons

2. Use the direction buttons to choose the test card which you need. The test card
you chose will show in the middle of the screen.

3. Prepare the proper concentration of the bacterial suspension, and put it in the
bacterial suspension well.
Name Concentration

E (Enterobacteriaceae) ,
NE(Non-fermenting bacteria), 0.5 McFarland Standard
STAPH(Staphylococcus, Micrococcus)

STREP(Streptococcus) 2 McFarland Standard

FUNGUS(FUNGUS) 1 McFarland Standard

4
4. Open the other reagents, and put into the relevant holes according to the below
list.

Well Name AST Diluent Enrichment Broth Bacterial


Test card Assimilation Diluent Suspension
DL-96E Broth \ Diluent
DL-96NE Broth \ Diluent
DL-96STAPH Broth \ Diluent
DL-96STREP Broth B Broth A Diluent
DL-96FUNGUS AST Diluent Assimilation Diluent Diluent

5. Hold 1ml tip by thumb and forefinger; smoothly entangle the tip upward with a
half turn.

5
6. Put the test card (ID area on the top side, AST are on the down side) in the tray
groove, and tear the mask to the left.

ID Area

AST Area

7. Press “Calibrate” button, tip will move to A12 well automatically.

8. Adjust the tip if necessary. Tip must align the center of test card A12 well。

9. Press “run” button and start sampling.

10. After 3.5 minutes, sampling will be finished. System will go back to initial state.
Repeat step 2 to 9 can continue sampling.

Remark: When the system is running, short press of the reset button can stop the
instrument; extended press of the reset button can restart the system.

6
5 Routine Maintenance
1. This instrument should be maintained at regular intervals and keep it clean. Use
the clean soft cloth with a little anhydrous alcohol to clean it.

2. For preventing diluent rust the electronic device and effect normal operation. If
diluent is poured on the instrument surface. Please wipe it up with dry cloth at
once.

3. When stop using the instrument, please cut off the power in time. And clean the
surface.

4. If you haven’t use the instrument for a long time or just open it. The instrument
should dry run for a time and then you can use it normally.

5. When restart the instrument during running, the tip must be unloaded at first.

6. Please don’t drag the spearhead, when the instrument is power off.

7
6 Common Problem Treatments
Instrument Common Problem
No. Common Fault Cause Analysis Solution
1 Diluent is not enough in The running status is not Dry running for once before
the tip, sampling is not stable after starting. starting work and then it can be
even after turning on the used.
instrument.
2 Diluent was dropped out Step 8: Positioning Must check and adjust the straw
of the wells when doesn’t align the center positioning before
sampling. of test card well. operation.(step 8)
3 During running, there are Caused by the size 1,Check the size of tip(neither
some extra diluent drops problem or rag of tip. too big or too small is allowed).
from tip, or the tip spurts 2,Change the tip.
on the wrong place 3,If still not solve the problem,
please contact manufacturer.
4 There are bubbles in the Dosage is not enough. Add the dosage in time. Or
tip during imbibition. change the tip.
5 During running, tip stops There is foreign matter 1, Turn off the instrument first
in sudden and the lock in spearhead, and and drag the spearhead for a
instrument has rotary positioning line is distance, then restart.
noise, or its run doesn’t interfered or broken. 2, Check if the joints of vertical
accord to the program. motor and crosswire motor are
broken or not.
3, If problems not solved, please
contact the manufacturer.
6 Positioning align the Caused by tray 1, Fastening vertical motor gear
center of test card well. looseness. fixed bolt.
But diluent was dropped 2, Contact the manufacturer.
out of the wells when
sampling.
7 The font is not so clear The running status is not 1, The font will be clear after
after turning on the stable after starting. turning on for a while.
instrument. 2. Take off the bottom cap and
adjust the control of brightness.
8 The diluent is added in the Choose wrong or use the 1, Check whether it ist he wrong
wrong well. wrong test card. choice of test card. Dosage
menu is chosen wrong or not.
2, If you haven’t fault, please
contact manufacturer upgrade
the system.

8
7 After-sales service
1. Warranty
1) If any failures, caused by material or manufacturing, happen within 1 month since
the delivery date, DL is responsible to replace it with a new instrument and take
care of all the delivery expense.
2) If any failures, caused by material of manufacturing, happen within 12 months
since the delivery date, DL provides with free repair. Customers bear the cost of
freight back to repair department, while DL bear the freight charge of delivery
back to customer.
3) When the warranty is expired, DL only charges for favorable price of spare parts.
2. Warranty range
This warranty did not cover unexpected use, the use in the environment which
does not meet the requirements, and the damage from unauthorized repair or refit.

8 Unpacking And Inspection


1, User first to open the packing, please accord the sign upward to take off the
instrument. Avoid to damage the instrument, please don’t knock and pry.

2, Check the instrument and attachments accord the packing list. If there is any
omissions, please contact the supplier or our company.

9 Delivery And Storage


1, The instrument is fixed tight in the carton, anti-rain, anti-moisture and anti-shock.
During the delivery, it is not allowed to shake fiercely and collide, which may
seriously damage the product.

2, If the instrument is not use for a long period of time, it should be packed, anti-rain,
anti-moisture and anti-shock. And it should be stored at the temperature 5℃-40℃,
in the ventilated indoor with relative humidity ≤80% and no corrosive gases.

【Manufacturing Enterprise】
Company Name: Zhuhai DL Biotech Co., Ltd.
Address: 19 Jinhaian Rd, Sanzao Town, Jinwan, Zhuhai, Guangdong 519041, P.R. China
Tel: 86-756-6292028 Fax: 86-756-3391389
Website:www.medicaldl.com E-mail:dl@medicaldl.com
Shanghai International Holding Corp. GmbH (Europe)
Eiffestrasse 80, 20537 Hamburg,Germany
Tel: +49-40-2513175 Fax:+49-40-255726
Dimid No.:DE/0000040627 E-mail: shholding@hotmail.com
Version: Oct.8.2014