HumaLyzer PRIMUS

Settings for Multipurpose reagents


Software Version 3.2, Date 2015-06

Kit: y-GT liquicolor


REF: 12213, 12013, 12023, 12033

Note: Procedure 2, Sample Start, IFCC

Test Parameters GGT


Method: Kinetics
Unit: U/l
Temperature: 37°C
First Wavelength: 405nm
Second Wavelength: None
Blank: Water
Factor: 1309
Norms High: 55.0
Norms Low: 0
Delay Time: 60s
Test Time: 180s
Decimal Digits: 1
Aspirate Volume: 500 µl

Standard Test Parameters


Method:
Number of Standard:
Repeat Times:
Concentration1:

Control Test Parameters


No. AV SD Batch
1 0.00 0.0
2 0.00 0.0

Human Gesellschaft für Biochemica und Diagnostica mbH


Max-Planck-Ring 21 - D-65205 Wiesbaden – Germany
Tel.: +496122 99880 Fax: +496122 9988100 - e-mail: human@human.de