You are on page 1of 5

Review

Endogenous Antibody Interferences


in Immunoassays
Jane F. Emerson, MD, PhD,1* Keane K.Y. Lai, MD1,2

ABSTRACT detect interference, and approaches used to resolve discrepancies


between laboratory results and the clinical picture.
The potential for interfering substances to cause inaccurate laboratory
results that may cause significant adverse effects on patient care is Keywords: immunoassay, interference, heterophile antibody,
well known. Immunoassays are subject to interferences that are not antianimal antibody, autoantibody
readily detectable prior to analysis, but may cause erroneous results.

Learning Objectives
After reading this article, readers should understand potential causes
of interference that are specific to immunoassays, methods used to

Immunoassays are subject to interferences that are a misdiagnosis of cancer. In that case, fault was found with
not readily detectable prior to analysis but may cause the physicians and the hospital laboratory.2
erroneous results. There are numerous reports in the
scientific literature of patients receiving inappropriate This article emphasizes the potential impact of
medical treatment based on incorrect results of laboratory errors in immunoassays, describes the types
immunoassays. One case illustrating the impact of an of interference that are specific to immunoassays, and
immunoassay interference was reported in Lancet1 in the reviews general approaches to reducing errors and
year 2000. In this case a repeated falsely high human associated negative outcomes in patient care.
chorionic gonadotropin (hCG) result, prompted a young
woman to undergo a hysterectomy, chemotherapy,
radiotherapy, and a partial pneumonectomy before the
interference was discovered and her diagnosis corrected. Types of Interference
The case was similar to a medical malpractice case
reported in the Seattle Post-Intelligencer newspaper in All laboratory assays are subject to interferences. The
2001, in which a woman was awarded $15.5 million due to effects of hemolysis, lipemia, and bilirubinemia (ie, icterus)
on laboratory methods, for instance, are well known. Each
of these may affect the analytical measurement; good
DOI: 10.1309/LMMURCFQHKSB5YEC laboratory practice requires validating assays with regard
to potential interferences. Because hemolysis, lipemia, and
Abbreviations icterus produce measureable spectrophotometric changes
hCG, human chorionic gonadotropin; HAAAs, human antianimal in a serum or plasma specimen, automated chemistry
antibodies; HAMAs, human antimouse antibodies; TSH, thyroid-
stimulating hormone; CLSI, Clinical and Laboratory Standards analyzers routinely detect these potentially interfering
Institute; CK-MB, creatine kinase–muscle and brain; LC-MS/MS, substances. Thus, specimens for chemical analyses are
liquid chromatography–tandem mass spectrometry; FSH, follicle- prospectively evaluated for these potential interferences.3
stimulating hormone; AFP, alpha-fetoprotein; CA, cancer antigen; CEA,
carcinoembryonic antigen; PSA, prostate-specific antigen
Immunoassays are subject to other types of interference.
1
Department of Pathology, Keck School of Medicine, University of Antigen-antibody interactions are the basis of
Southern California and 2Southern California Research Center for ALPD immunoassays, and compounds or conditions that
and Cirrhosis, Los Angeles, California
alter these interactions can interfere with measurement.
*To whom correspondence should be addressed. Endogenous antibodies may bind to, bridge, or block the
E-mail: jane.emerson@usc.edu binding sites on capture and signal antibodies (Figure 1).

Downloaded www.labmedicine.com
from https://academic.oup.com/labmed/article-abstract/44/1/69/2657820 Winter 2013  |  Volume 44, Number 1  Lab Medicine 69
by guest
on 13 March 2018
Review

Capture antibody
Figure 1 Endogenous antibody
Schematic of how endogenous antibodies can interfere Detection antibody
in immunoassays to produce false-positive or false-
negative results.

Antigen

Antigen

Solid Phase
True Positive False Positive False Negative

As a result, falsely high or falsely low results may occur affect thyroglobulin immunoassays and anti-insulin
due to the presence of endogenous antibodies. In the case antibodies that interfere with immunoassays for insulin
reports referred to in the first paragraph of this article, or C-peptide.6,7,10 High titers of potentially interfering
the interference was positive and produced a falsely antibodies may occur in patients with recent infections,
high measurement hCG on which the misdiagnosis of immunizations, and transfusions.
choriocarcinoma was made, a type of cancer characterized
by high hCG levels in the absence of pregnancy.

The 3 types of endogenous antibodies known to cause incidence


interferences in immunoassays are heterophile, antianimal,
and autoantibodies. Heterophile antibodies are produced Heterophile antibodies may be present in all patients,11,12
without exposure to specific immunogens and are however the potential for immunoassay interference from
thus considered to be naturally occurring.4 Heterophile heterophile antibodies is less than for autoantibodies or
antibodies generally have low avidity but react across HAAA. The overall prevalence of interfering antibodies
multiple species with the capability of binding to multiple depends on the assumptions made about their
antigens. Heterophile antibodies may react with variable potential to cause interference. Similarly, the frequency
avidity to distinct species; also, reactivity to different of immunoassay interferences resulting from these
species may persist for different periods of time within the antibodies depends on what magnitude of bias in the
same patient.5-7 analytical method constitutes a significant interference.
In any case, the frequency of interferences is much lower
Human antianimal antibodies (HAAAs) are produced after than would be predicted by the prevalence of potentially
acute or chronic exposure to specific antigens and are interfering antibodies.
species specific.8 HAAAs generally have higher avidity
than heterophile antibodies. Human antimouse antibodies Improvements in assay formulations have decreased
(HAMAs) constitute a subset of HAAAs and are the most the occurrence of interferences from endogenous
common of this antibody type. Antianimal antibodies may be antibodies. Blocking reagents (typically mouse or rabbit
produced7,9 in response to therapeutic agents that include immunoglobulins) have been added to immunoassay
animal-derived monoclonal antibodies, or from exposure to formulations to adsorb interfering antibodies. Thus,
animals occupationally (eg, veterinarians) or to pets. although the prevalence of potentially interfering
antibodies has been reported to be as high as 40%,11 the
Autoantibodies also may interfere with immunoassays. incidence of immunoassay interference is estimated to
Examples include antithyroglobulin antibodies that be less than 2%.13 However, even with a low incidence

70 Lab Medicine  Winter 2013  |  Volume 44, Number 1


Downloaded from https://academic.oup.com/labmed/article-abstract/44/1/69/2657820 www.labmedicine.com
by guest
on 13 March 2018
Review

of interference, the number of potential errors with An algorithm has been suggested15 that would compare
serious consequences is still high because the number of immunoassay results to the prevalence of disease. In this
immunoassays performed is so large. approach results that are positive for a disease of low
prevalence, or negative for a disease of high prevalence,
Although the prevalence of interfering antibodies varies are considered suspect.
from one patient population to another, the incidence of
interference also varies by immunoassay type. Two-site Proactive Approach
assays are the most susceptible to interference.14 Table Various methods have been proposed to detect or
1 displays some assays that are relatively commonly block the effects of interfering antibodies. Suggested
affected. approaches include looking for discrepancies among
alternative methods of measurement,6,9,16 serial dilutions to
reveal nonlinearity,6,9,16 screening for antimouse antibodies
with the Tandem ICON ImmunoConcentration human
Detecting Interference chorionic gonadotropin assay9,16 (Hybritech Inc, San Diego,
CA) or pretreating specimens with blocking reagents.16-18
Both retroactive and proactive approaches exist for A study16 investigating the general usefulness of some of
detecting interfering antibodies. Retroactive refers to these approaches concluded that introducing a protocol
instances when the laboratory result is questioned to prescreen all samples for the presence of endogenous
because it is inconsistent with the clinical findings or interfering antibodies is not warranted because the
exceeds extreme limits for the analyte. A proactive approaches were associated with too low an event rate to
approach would implement a mechanism or procedure justify routine implementation or with too high a prevalence
that would detect the presence of interfering antibodies and were too nonspecific to be useful.16
prior to obtaining or reporting the result.
Another proactive approach would be to request pertinent
Retroactive Approach patient history, such as previous treatment with monoclonal
The retroactive approach has obvious shortcomings. antibody preparations or a history of erroneous laboratory
For example, quantitative results that are far beyond results. This approach would be difficult to implement and
expected values will be readily flagged for certain assays is likely to be an ineffective strategy because it is unknown
such as thyroid studies; however, this is not the case how many cases of interference could be predicted with this
for most tumor-marker assays. Thyroid-stimulating type of information.
hormone (TSH) concentrations characteristic of thyroid
disease, for example, vary to a far lesser degree than
hCG concentrations in pregnancy or in certain cancers.
Therefore, use of thresholds based on physiologically Table 1. Immunoassays Commonly Affected
unreasonable results to alert technologists of a potential by Endogenous Antibodies
interference is not practical in all immunoassays.
Endocrine Substances Tumor Markers Other

Cases in which the results are erroneously normal, or Cortisol16 AFP26 CK-MB9
abnormal but not implausible, are more difficult to detect. Estradiol9 CA 1259 Ferritin29
Retroactive detection of false positive or false negative Free thyroxine9 CA 15-327 Hepatitis B surface
FSH9 CA 19-928   antigen9
results usually requires knowledge of the clinical picture. LH CEA
9 9
Troponin I9
Laboratories do not typically receive sufficient clinical Progesterone9 hCG9

information to determine whether the laboratory result is Prolactin9 PSA16


consistent with the patient’s condition. Also, the clinical Testosterone24
Thyroglobulin25
picture is often complex, rendering interpretation of Thyroxine9
the result in clinical context beyond the scope of the Triiodothyronine9
personnel responsible for reviewing and releasing the test TSH9
results. Thus, retrospective detection of immunoassay FSH, follicle-stimulating hormone; LH, luteinizing hormone; TSH, thyroid-
interferences relies heavily on healthcare providers stimulating hormone; AFP, alpha-fetoprotein; CA, cancer antigen; CEA,
carcinoembryonic antigen; hCG, human chorionic gonadotropin; PSA, prostate-
who may have limited awareness of the potential for specific antigen; CK-MB, creatine kinase–muscle and brain.
immunoassay interference.

Downloaded www.labmedicine.com
from https://academic.oup.com/labmed/article-abstract/44/1/69/2657820 Winter 2013  |  Volume 44, Number 1  Lab Medicine 71
by guest
on 13 March 2018
Review

• Certain analytes also appear in urine. Because


endogenous antibodies are not usually present
Clinical Laboratory in urine, a discrepancy between urine and serum
Responsibility concentrations may suggest interference.
• A specimen may be preincubated with commercially
The Clinical and Laboratory Standards Institute (CLSI) available blocking reagents (some involve blocking
document, “Immunoassay Interference by Endogenous antibodies immobilized on the inside surface of a
Antibodies; Approved Guideline” 19 states that it is the test tube).
responsibility of the clinical laboratory to: • Compare values of related analytes if applicable: for
example, creatine kinase–muscle and brain (CK-MB)
• Ensure the personnel performing the assay have and troponin.
the required knowledge of endogenous antibody
interference. Many reference laboratories are capable of measuring
• Contact clinicians when there is suspicion of HAMA. If HAMA interference is suspected, quantitative
interference in a patient specimen. measurement of serum HAMA concentration may
• Follow up on complaints about clinically inconsistent be helpful; however, no immunoassay exists, to our
results. knowledge, that is capable of measuring equally well
• Notify the manufacturer/vendor of assay interference all types of HAMA found among patients8 because
problems. various assays differ in the types of HAMA detected.9,19
• Investigate mismatches between assay results Also, nonimmunometric methods such as liquid
and clinical information in conjunction with the chromatography–tandem mass spectrometry (LC-MS/MS)
manufacturer/vendor. are widely available for analytes such as testosterone, 20
• Inform clinicians regarding the nature of the interfering 25-hydroxyvitamin D, 21 and immunosuppressants. 22,23
antibody, including how it affects the immunoassay Table 2 summarizes key concepts to consider when
in question and how it may affect other immunoassay investigating potential interference in an immunoassay.
tests ordered on the patient.

Troubleshooting Interferences
Once interference is suspected, most clinical laboratories
have the capability to investigate using a number of Table 2. Immunoassay Interferences:
techniques. A multipronged approach to the investigation Key Concepts
of suspected interference is preferable to a strategy
involving only one option that may not definitively resolve Detection
  •  Retroactive
the discrepancy between clinical and laboratory findings.
   Extremes
   Clinical picture
• Obtaining patient history regarding therapy with a    Disease prevalence
monoclonal antibody preparation, animal exposure, or   •  Proactive
   Dilutions
transfusions may be helpful.
   Heterophile blocking
• The linearity of the in-house assay for the specimen    Antianimal screens
in question can be measured, taking care to use     Clinician education: alerts to patient history of endogenous
appropriate diluents. However, linearity might still     antibodies
be observed even in the presence of interfering Troubleshooting
  •  Dilutions
antibodies. 6   •  Alternate method
• The specimen may be sent to another laboratory   •  Heterophile blocking
for retesting using an alternate method, such as   •  Measure serum HAMA concentration
an immunoassay that uses a different technology   •  Test urine, if applicable
  •  Test related analytes, if applicable
(homogeneous vs. heterogeneous; competitive vs.   •  Patient history for therapeutic antibodies, animal exposure,
non-competitive) and a different reagent antibody     transfusions or autoimmune disease
source than the original immunoassay suspected to
HAMA, human antimouse antibody.
display the interference, or a nonimmunometric method.

72 Lab Medicine  Winter 2013  |  Volume 44, Number 1


Downloaded from https://academic.oup.com/labmed/article-abstract/44/1/69/2657820 www.labmedicine.com
by guest
on 13 March 2018
Review

10. Jahagirdar VR, Strouhal P, Holder G, Gama R, Singh BM.


Thyrotoxicosis factitia masquerading as recurrent Graves’ disease:
Conclusions endogenous antibody immunoassay interference, a pitfall for the
unwary. Ann Clin Biochem. 2008;45(pt 3):325-327.

Laboratory professionals must be mindful of the 11. Boscato LM, Stuart MC. Incidence and specificity of interference in
two-site immunoassays. Clin Chem. 1986;32(8):1491-1495.
potential for interference in immunoassays because of 12. Hennig C, Rink L, Kirchner H. Evidence for presence of IgG4
the possibility of erroneous results that adversely impact anti-immunoglobulin autoantibodies in all human beings. Lancet.
patient care. Knowledge of the incidence and mechanisms 2000;355(9215):1617-1618.
13. Marks V. False-positive immunoassay results: a multicenter survey
of immunoassay interferences, and the recommended of erroneous immunoassay results from assays of 74 analytes in
troubleshooting methods, is crucial to ensure the overall 10 donors from 66 laboratories in seven countries. Clin Chem.
quality of immunoassay results. Because proactively 2002;48(11):2008-2016.
14. Ismail AA, Walker PL, Cawood ML, Barth JH. Interference in
screening for interference in all specimens is not warranted immunoassay is an underestimated problem. Ann Clin Biochem.
or recommended, it is important to maintain reliable 2002;39(pt 4):366-373.
communication with other health care professionals to detect 15. Ismail AA, Ismail AA, Ismail Y. Probabilistic Bayesian reasoning can
help identifying potentially wrong immunoassays results in clinical
and minimize the impact of erroneous laboratory results. LM practice: even when they appear ‘not-unreasonable’. Ann Clin
Biochem. 2011;48(pt 1):65-71.
CE credit available for this article. For more information, 16. Emerson JF, Ngo G, Emerson SS. Screening for interference in
immunoassays. Clin Chem. 2003; 49(7):1163-1169.
visit http://labmed.ascpjournals.org/content/44/1/69.full.
17. Reinsberg J. Interferences with two-site immunoassays by human
pdf+html. anti-mouse antibodies formed by patients treated with monoclonal
antibodies: comparison of different blocking reagents. Clin Chem.
1998;44(8 pt 1):1742-1744.
18. Butler SA, Cole LA. Use of heterophilic antibody blocking
agent (HBT) in reducing false-positive hCG results. Clin Chem.
To read this article online, scan 2001;47(7):1332-1333.
the QR code, http://labmed. 19. Clinical and Laboratory Standards Institute (CLSI). Immunoassay
ascpjournals.org/content/44/1/69. Interference by Endogenous Antibodies; Approved Guideline. CLSI
full.pdf+html document I/LA30-A. Wayne, PA: Clinical and Laboratory Standards
Institute; 2008.
20. Shiraishi S, Lee PWN, Leung A, Goh VHH, Swerdloff RS, Wang
C. Simultaneous measurement of serum testosterone and
dihydrotestosterone by liquid chromatography-tandem mass
spectrometry. Clin Chem. 2008;54(11):1855-1863.
References 21. Wallace AM, Gibson S, de la Hunty A, Lamberg-Allardt C, Ashwell
M. Measurement of 25-hydroxyvitamin D in the clinical laboratory:
1. Rotmensch S, Cole LA. False diagnosis and needless therapy of
QR code generated on http://qrcode.littleidiot.be

current procedures, performance characteristics and limitations.


presumed malignant disease in women with false-positive human Steroids. 2010;75(7):477-488.
chorionic gonadotropin concentrations. Lancet. 2000;355(9205):712-
22. Keevil BG, Tierney DP, Cooper DP, Morris MR. Rapid liquid
715.
chromatography-tandem mass spectrometry method for routine
2. Skolnik S. Jury awards $15.5 million to woman misdiagnosed analysis of cyclosporin A over an extended concentration range.
with cancer. Seattle Post-Intelligencer. June 29. 2001. Accessed Clin Chem. 2002;48(1): 69-76.
December 7, 2012.
23. Bazin C, Guinedor A, Barau C, et al. Evaluation of the ARCHITECT
3. Steen G, Klerk A, van der Laan K, Eppens EF. Evaluation of the tacrolimus assay in kidney, liver, and heart transplant recipients.
interference due to haemoglobin, bilirubin and lipids on Immulite J Pharm Biomed Anal. 2010;53(4):997-1002.
2500 assays: a practical approach. Ann Clin Biochem. 2011;48(pt
24. Ramaeker D, Brannian J, Egland K, McCaul K, Hansen K. When is
2):170-175. elevated testosterone not testosterone? When it is an immunoassay
4. Levinson SS, Miller JJ. Towards a better understanding of heterophile interfering antibody. Fertil Steril. 2008;90(3):886-888.
(and the like) antibody interference with modern immunoassays. Clin 25. Preissner CM, O’Kane DJ, Singh RJ, Morris JC, Grebe SK. Phantoms
Chim Acta. 2002;325(1-2):1-15. in the assay tube: heterophile antibody interferences in serum
5. Kaplan IV, Levinson SS. When is a heterophile antibody not a thyroglobulin assays. J Clin Endocrinol Metab. 2003;88(7):3069-3074.
heterophile antibody? When it is an antibody against a specific 26. Janssen JA, Blankestijn PJ, Docter R, Blijenberg BG, Splinter TA,
immunogen. Clin Chem. 1999;45(5):616-618. van Toor H, Schalekamp MA, Lamberts SW, Krenning EP. Effects
6. Ismail AA. Interference from endogenous antibodies in automated of immunoscintigraphy with monoclonal antibodies in assays of
immunoassays: what laboratorians need to know. J Clin Pathol. hormones and tumour markers. BMJ. 1989;298(6686):1511-1513.
2009;62(8):673-678. 27. Preissner CM, Dodge LA, O’Kane DJ, Singh RJ, Grebe SK.
7. Sturgeon CM, Viljoen A. Analytical error and interference in immuno- Prevalence of heterophilic antibody interference in eight automated
assay: minimizing risk. Ann Clin Biochem. 2011;48(pt 5):418-432. tumor marker immunoassays. Clin Chem. 2005;51(1):208-210.
8. Park S, Cadeddu J, Balko JA, Tortelli MW, Wians FH Jr. Persistently 28. Berth M, Bosmans E, Everaert J, Dierick J, Schiettecatte J, Anckaert
elevated prostate-specific antigen level after successful laparoscopic E, Delanghe J. Rheumatoid factor interference in the determination
radical prostatectomy. Lab Med. 2006;37(8): 474-477. of carbohydrate antigen 19-9 (CA 19-9). Clin Chem Lab Med.
2006;44(9):1137-1139.
9. Kricka LJ. Human anti-animal antibody interferences in
immunological assays. Clin Chem. 1999;45(7): 942-956. 29. Blanco M, Varela C. Interference from heterophilic antibodies in
the Olympus ferritin method. Clin Chem. 2006;52(8):1623-1624.

Downloaded www.labmedicine.com
from https://academic.oup.com/labmed/article-abstract/44/1/69/2657820 Winter 2013  |  Volume 44, Number 1  Lab Medicine 73
by guest
on 13 March 2018

You might also like