Drawings SC210G-C-20□

Flange type Conductivity Detector


for Low Range

Unit: mm

Clearance for maintenance access


91 82
H+50

G1/2 (Note)
Approx. Cable inlet
4 -øh holes H+172
(Note) "/ANSI"Adaptor of cable inlet
/ANSI
t 45
øC
26
øD Approx.H

1/2NPT

ø15
26 28 G1/2

Flange Insertion length (Sensor length)


Model and Code Flange rating ØC ØD t Øh Model and Code H Weight (kg)
SC210G - □ - 206 JIS 10K-50-RF 120 155 16 19 SC210G - □ - 20□ - L015 - □□*A 162 Approx. 4.3
SC210G - □ - 207 ANSI CLASS150-2-RF * 120.7 152.4 19.1 19.1 SC210G - □ - 20□ - L030 - □□*A 312 Approx. 4.5
SC210G - □ - 208 JPI CLASS150-2-RF 120.6 152 19.5 20 SC210G - □ - 20□ - L050 - □□*A 512 Approx. 4.9
SC210G - □ - 20□ - L100 - □□*A 1012 Approx. 6.4
SC210G - □ - 20□ - L150 - □□*A 1512 Approx. 7.2
SC210G - □ - 20□ - L200 - □□*A 2012 Approx. 8.0
*: ANSI flange with serrations.
l SC210G Detector - converter connection cable (accessory)
Black Green
T2 ø10.5 13
Red White
T1 To detector To converter 14
White Yellow
C1 15 (A)
Brown Brown
C2 16
Black
12
Red
11
Approx. 75 L Approx. 80
(3, 5, 10, 15, 20 m) Green
13
White
14
Yellow
15 (B)
Brown
16
Black
12
Red
11

Model and Code L Weight (kg) Terminal


SC210G - □ - 20□ - L□□□ - 03 (A) M4 Ring terminal
SC210G - □ - 20□ - L□□□ - Y1 Approx. 3000 Approx. 0.9 (A) M3 Ring terminal
SC210G - □ - 20□ - L□□□ - AA (B) Pin terminal
SC210G - □ - 20□ - L□□□ - 05 (A) M4 Ring terminal
SC210G - □ - 20□ - L□□□ - Y2 Approx. 5000 Approx. 1.5 (A) M3 Ring terminal
SC210G - □ - 20□ - L□□□ - BB (B) Pin terminal
SC210G - □ - 20□ - L□□□ - 10 (A) M4 Ring terminal
SC210G - □ - 20□ - L□□□ - Y3 Approx. 10000 Approx. 3.0 (A) M3 Ring terminal
SC210G - □ - 20□ - L□□□ - CC (B) Pin terminal
SC210G - □ - 20□ - L□□□ - 15 (A) M4 Ring terminal
SC210G - □ - 20□ - L□□□ - Y4 Approx. 15000 Approx. 4.5 (A) M3 Ring terminal
SC210G - □ - 20□ - L□□□ - DD (B) Pin terminal
SC210G - □ - 20□ - L□□□ - 20 (A) M4 Ring terminal
SC210G - □ - 20□ - L□□□ - Y5 Approx. 20000 Approx. 6.0 (A) M3 Ring terminal
SC210G - □ - 20□ - L□□□ - EE (B) Pin terminal
Unless otherwise specified, differences in the dimensions are specified as: General tolerance = ±(Criteria of tolerance class IT18 in JIS B0401-1998)/2.

All Rights Reserved, Copyright © 1995, Yokogawa Electric Corporation. SD 12D08F01-02E 1/1
Subject to change without notice.
1st Edition: Jan. 1995 (YK)
9th Edition: June 2023 (YK)