You are on page 1of 13

EXAMINATION and

INTERVENTION LIGHTS
USER MANUAL

Models :
CL-4MB CL-4M
CL-2M CL-2T
CL-2D ML-10
ML-40 ML-50
ML-60 ML-70
ML-80 ML-90
1. INDEX

1. INDEX……........................................................................................................................................................................ 2
2. INTRODUCTION............................................................................................................................................................. 4
3. FOR YOUR SAFETY....................................................................................................................................................... 4
3.1. General Information.............................................................................................................................................. 4
3.2. Symbols and Explanations.................................................................................................................................... 4
3.2.1. Identification Label…........................................................................................................................................... 5
4. PACKING, CARRYING and STORAGE………........................................................................................................... 5
5. INFORMATION............................................................................................................................................................... 6
5.1. Purpose Of Use And Areas Of Usage................................................................................................................... 6
5.2. Definition of the Product and Components........................................................................................................... 7
5.2.1. Components Definition of Examination Lights.................................................................................................... 7
5.2.2. Components Definition of Intervention Lights..................................................................................................... 7
5.3. Technical Specifications........................................................................................................................................ 8
5.3.1. Intervention Lights Technical Specifications........................................................................................................ 9
5.3.2. Examination Lights Technical Specifications....................................................................................................... 9
6. INSTALLATION and USAGE of the PRODUCT…...................................................................................................... 10
6.1. Installation............................................................................................................................................................ 10
7. CLEANING and MAINTENANCE…............................................................................................................................. 11
7.1. Cleaning the Surfaces of the Operating Light....................................................................................................... 11
7.1.1. Sterilization of the Reflector Cover Glass............................................................................................................. 11
7.1.2. Lamp Head And Other Components..................................................................................................................... 11
8. FAILURE CASES and SOLUTIONS…........................................................................................................................... 11
9. EUROPEON UNION REPRESENTATIVE.................................................................................................................... 13

KK/01/ML-12.05.2008-0 www.uzumcu.com.tr 2
Copyright
All information in this document is subject to change without prior notice. No part of
this document may be reproduced or transmitted in any form by any means without the
express written permission of the manufacturer.
Declaration of Conformity
These Devices are compliant in accordance with defined requirements “Medical
electrical equipment, part 1: General requirements for safety, Collateral standard safety
requirements for medical electrical systems” in TS EN 60601-1 and TS EN 60601-2-41
and MDD/93/42/EEC Medical Devices Directives.
Patent
All brand names and product names used in this document are trademarks or registered
trademarks of ÜZÜMCÜ.
Guarantee Duration
Guarantee Duration is defined as 2 years by Ministry of Industry and Trade.

KK/01/ML-12.05.2008-0 www.uzumcu.com.tr 3
2. INTRODUCTION

Thank You for buying ÜZÜMCÜ branded Examination and Intervention Lights. Our product is designed and
produced with care in order to provide you quality and performance in a best way.
This User Manuel explains the steps of installation, operation and maintenance of ÜZÜMCÜ branded
Examination and Intervention Lights. Please read carefully all the information about operation and safety of
this product.

Important points about Carrying and Transportation


Please follow the instructions written on the boxes of the devices.
About Maintenance
In case maintenance is required, please call one of the authorized technical services.
Information about connection and installation
Please read further sections for information about connection and installation

3. FOR YOUR SAFETY


3.1. GENERAL INFORMATION
For the use of examination and intervention lights, by providing safety for both the patient and user, and its
maintenace;
• Before using examination and intervention lights, be sure that this manual is clearly read and
comprehended with all warnings and clarifications.
• The customer is responsible for the staff that will use and maintain lights to get information about the
content of this manual.
• Keep the manual in a place that the user can easily reach.
• Do not use lights for purposes except that identified in this manual.
• Keep the maintenance of examination and intervention lights in the periods that are indicated in the 7th
section of this manual named “Cleaining and Maintenance”.
• Examination and intervention lights must be defined as “Contaminate Medical Waste” after its life cycle is
ended.
ÜZÜMCÜ R&D department keeps the rights of changing the technical specifications of the device and/or
canceling the information in this manual.

3.2. SYMBOLS AND EXPLANATIONS

Symbols Explanation Symbols Explanation

Attention, please refer to the attached


Manufacturing date
documents.

Compliant in accordance with MDD


Fragile, handle with care
93/42/EEC

Do not turn upside down Keep Dry

KK/01/ML-12.05.2008-0 www.uzumcu.com.tr 4
3.2.1. IDENTIFICATION LABEL

MEDICAL DEVICES AND GAS SYSTEMS EQUIPMENTS Co. Inc.


www.uzumcu.com.tr
SURGICAL LIGHT
Model No Device Class
REF CL-2M Class I

220-240 VAC Device Power


Serial No SN 123456789
50 – 60 Hz
Lamp Type
Halojen Bulb Lamp Voltage
Manufacturing 2005 12V – 50W Lamp Power
Date

4. PACKING, CARRYING and STORAGE


Temperature of the Environment : +5oC ~ +40oC
Humidity : 30% ~ 70%
Atmospheric Pressure : 0.49atm ~ 1atm

Examination and intervention lights are dispatched after packed into a closed case and all part is separated.
After receiving the device;
 Open the package from suitable sides.
 Remove the nylon cover carefully.
 Carefully examine product and assembly parts in case it is damaged during the transportation. If so, please
contact with the related seller.
 Store product by putting one case on the top of another.

ATTENTION
 Do not use cutting tools while removing the nylon cover. Otherwise you may damage the painted
surface of the unit.

KK/01/ML-12.05.2008-0 www.uzumcu.com.tr 5
ATTENTION
 Do not carry examination and intervention lights by holding it from the parts marked with number 1.
Carry it from the part marked with number 2.

 Shape-1


5. INFORMATION

5.1. PURPOSE of USE and AREAS of USAGE


ÜZÜMCÜ branded Examination and intervention lights are designed specially to use in medical interventions
and examinations. Product design is proper for various medical interventions with powerful illumination
capacity and different joints options.
Examination and intervention lights surfaces have smooth and clear-cut design for cleaning and disinfecting.
The light is designed in accordance with hygienic needs of operating environment.

KK/01/ML-12.05.2008-0 www.uzumcu.com.tr 6
5.2. DEFINITION of the PRODUCT and COMPONENTS
5.2.1. COMPONENTS DEFINITION of EXAMINATION LIGHTS

2
5

1
3

Şekil-2
ITEM NO NAME of the ITEM ITEM NO NAME of the ITEM
1 Lamp 5 Lamp Adjustment Arm
2 On-Off Switch 6 Body Connection Screw
3 Light Intensity Switch 7 Caryying Foot

5.2.2. COMPONENTS DEFINITION of INTERVENTION LIGHTS

5
4

ITEM NO NAME of the ITEM


1 Lamp
6 2 Adjustment Arm
3 On-Off Switch
2 1 4 Holder Arm
3 5 Balanced Arm
6 Stand Holder
7 Battery
8 Casters

KK/01/ML-12.05.2008-0 www.uzumcu.com.tr 7
5.3. TECHNICAL SPECIFICATIONS
5.3.1. INTERVENTION LIGHTS TECHNICAL SPECIFICATIONS

CL-4MB CL-4M CL-2M CL-2D CL-2T

Mobile Mobile Mobile Wall Ceiling


Mount Type

24V / 75W 24V / 75W 12V / 50W 12V / 50W 12V / 50W
Bulb
Halojen Halojen Halojen Halojen Halojen

Intensity of the Light 55.000 lux/1m 55.000 lux/1m 38.000 lux/1m 38.000 lux/1m 38.000 lux/1m

Color Temperature 4.100 Kelvin 4.100 Kelvin 4.300 Kelvin 4.300 Kelvin 4.300 Kelvin

Light Head Diameter 400 mm 400 mm 245 mm 245 mm 245 mm

Length of Spring Arm 800 mm 800 mm 600 mm 600 mm 600 mm

520 up 520 up 600 up 600 up 600 up


Movement Diameter
570 down 570 down 800 down 800 down 800 down
Light Intensity
%66 - %100 %66 - %100 %66 - %100 ----- -----
Adjustment

Height of Stand 1900 mm 1900 mm 1600 mm ----- -----

Wall Mounting Arm ----- ----- ----- 600 mm -----

Ceiling Mounting Arm ----- ----- ----- ----- 1100 mm

220-240V 220-240V 220-240V 220-240V 220-240V


Input Voltage
50-60 Hz 50-60 Hz 50-60 Hz 50-60 Hz 50-60 Hz

Working Voltage 24 VDC 24 VDC 12 VDC 12 VDC 12 VDC

Battery Usage 3 hours ----- ----- ----- -----

Gross Weight 30 Kg 30 Kg 40 Kg 17 Kg 17 Kg

KK/01/ML-12.05.2008-0 www.uzumcu.com.tr 8
5.3.2. EXAMINATION LIGHTS TECHNICAL SPECIFICATIONS

ML-10 ML-40 ML-50 ML-60 ML-70 ML-80 ML-90

Mount Type Mobile Mobile Mobile Mobile Rail Type Wall Table

T-R16 22W
12V / 35W 12V / 50W 12V / 50W 12V / 50W 12V / 50W 12V / 50W
Bulb Circular
Halogen Halogen Halogen Halogen Halogen Halogen
Fluorescent
Intensity of 32.000 lux / 22.000 lux / 22.000 lux / 22.000 lux / 22.000 lux / 22.000 lux /
the Light 0,5 m 0,5 m 0,5 m 0,5 m 0,5 m 0,5 m
Light Head
120 mm 100 mm 100 mm 100 mm 100 mm 100 mm 100 mm
Diameter
Length of
----- ----- ----- 550 mm 650 mm 650 mm 650 mm
Spring Arm
Height of
mm 1720 mm 790 mm 650 mm ----- ----- -----
Stand
220-240V 220-240V 220-240V 220-240V 220-240V 220-240V 220-240V
Input Voltage
50-60 Hz 50-60 Hz 50-60 Hz 50-60 Hz 50-60 Hz 50-60 Hz 50-60 Hz
Working
12 VDC 12 VDC 12 VDC 12 VDC 12 VDC 12 VDC 12 VDC
Voltage

Gross Weight 2,8 Kg 4,7 Kg 5,5 Kg 5,5 Kg 2 Kg 2 Kg 2 Kg

KK/01/ML-12.05.2008-0 www.uzumcu.com.tr 9
6. INSTALLATION and USAGE of the PRODUCT

6.1. INSTALLATION

To prepare usage of Intervention and Examination Lights that are dispatched after mounted, it is enough to plug
the electric cable in grounded socket and position the power switch to “1” after carrying it in its place.
Follow the process below to install Intervention and Examination Lights that are dispatched with demounted
parts.

KK/01/ML-12.05.2008-0 www.uzumcu.com.tr 10
7. CLEANING AND MAINTENANCE

7.1. CLEANING the SURFACES of the INTERVENTION and EXAMINATION LIGHTS


Cleaning of Intervention and Examination Lights is performed as it explains below.
7.1.1. STERILIZATION of REFLECTOR COVER GLASS
The cover glass of the reflector must be cleaned and sterilized before the operation. The cover glass must be
wiped with a cloth which is wet with a solution containing formaldehyde or similar sterile material.
7.1.2. LAMP HEAD and OTHER COMPONENTS
Lamp head and other components must be cleaned and sterilized periodically. These components must be
wiped with a cloth which is wet with a solution containing formaldehyde or similar sterile material.
• Be careful during the cleaning process of the ceiling type operating light not to damage the ceiling
connection cover and other components or not to cause water leakage between them.
• While cleaning stand and fixed carrier of the mobile type lights, be careful not to cause water leakage
into the ÜZÜMCÜ branded Examination and Intervention Lights of the light which would damage the
device.

ATTENTION
• The main power must be turned off before cleaning or sterilizing the operating light.

RECOMMANDATION for the USAGE:


• Prevent the electric cable from any damage. Before each usage, check the electric cable durability. Do
not use the cables that are damaged.

8. FAILURE CASES and SOLUTIONS

Please contact with our technical services for other information about Intervention and Examination Lights or
their components and repairing.

• For CL-4MB models;

Failure Causes Solutions


Plug cable in socket. Check the
Electric cable may be unplugged or
electric current whether it is
electric current may be terminated.
active or not..
On-Off Switch may be on “0” Turn the On-Off Switch on the
1 Lamp does not work. position (1-0-2). device to on position (1-0-2)
Turn on the switch in steps on
Switch on lamp head is close
the lamp head.
(0-1-2).

Bulp may be out of use Change Bulb

KK/01/ML-12.05.2008-0 www.uzumcu.com.tr 11
• For CL-4M models;

Failure Causes Solutions


Plug cable in socket. Check the
Electric cable may be unplugged or
electric current whether it is
electric current may be terminated.
active or not..
1 Lamp does not work. Turn on the switch in steps on
Switch on lamp head is close
the lamp head.
(0-1-2) .

Bulp may be out of use Change Bulb

• For CL-2M / CL-2T / CL-2D models;

Failure Causes Solutions


Check the electric current
Electric current may be terminated.
whether it is active or not.
1 Lamp does not work. Turn on the switch in steps on
Switch on lamp head is close
the lamp head.
(0-1-2).

Bulp may be out of use Change Bulb

• For ML-40 / ML-50 /ML-60 / ML-70 / ML-80 / ML-90 models;

Failure Causes Solutions


Plug cable in socket. Check the
Electric cable may be unplugged or
electric current whether it is
electric current may be terminated.
active or not.
Turn on the switch in steps on
Switch on lamp head is close
1 Lamp does not work. the lamp head.
(0-1).

Bulp may be out of use Change Bulb

Transformator may be burned. Change Transformator

KK/01/ML-12.05.2008-0 www.uzumcu.com.tr 12
9. EUROPEON UNION REPRESENTATIVE

1. EXIM MED EXPORT IMPORT GmbH


Adress: HAUSANSCHRIFT LINDENSTRABE 165 40233 DUSSELDORF
GERMANY

Usage Life of the product is defined as 10 years by Ministry of Industry and Trade.