ISO 19036 Presentation Support ISO TC34 SC9 WG2!22!03-2021
ISO 19036 Presentation Support ISO TC34 SC9 WG2!22!03-2021
Measurement Uncertainty
Content
1. Introduction
2. Definitions
3. ISO 19036 – Components > ISO 19036 (2019) general approach
Combination between a global and a major component approach
4. ISO 19036 – Concepts not included in Measurement Uncertainty calculation
5. ISO 19036 – Practical approaches to estimate Measurement Uncertainty
6. ISO 19036 – Combined and Expanded Standard Uncertainty
7. ISO 19036 – Expression of Measurement Uncertainty in test reports
8. ISO 19036 – Summary
ISO 19036 –
Introduction
PURPOSE
• The international standard ISO / IEC 17025 for testing and calibration
laboratories requires the laboratories to estimate the Measurement
Uncertainty(MU).
• The laboratory’s customers use the results for taking important business
decisions globally.
• Laboratories therefore select and determine the measurement methods to
ensure the overall variability by evaluating measurement uncertainty.
• Too large or too small uncertainty may affect the reliability of the decision and
may make the situation complex and costly
• So an appropriable estimate of measurement uncertainty is an important task
performed by laboratories
Why estimate measurement uncertainty? (1)
• In particular where the result, including MU, approaches (iii) or just exceeds
to (ii) the limit of a (legal) specification, such results are questionable
Why estimate measurement uncertainty? (4)
• EURACHEM / CITAC Guide CG 4
<9.7.3> Example
A decision rule that is currently widely used is that a result implies non compliance with
an upper limit if the measured value exceeds the limit by the expanded uncertainty. With
this decision rule, then only case (i) would imply non compliance. Similarly, for a
decision rule that a result implies compliance only if it is below the limit by the expanded
uncertainty, only case (iv) would imply compliance.
<9.7.4.> In general the decision rules may be more complicated than these.
Further discussion may be found in EURACHEM/CITAC Guide: The use of
uncertainty information in compliance assessment (2007)
Why estimate measurement uncertainty? (5)
Standard Uncertainty -
uncertainty of the result of a measurement expressed as a standard deviation
Coverage Factor -
number larger than one by which a combined standard measurement
uncertainty is multiplied to obtain an expanded measurement uncertainty
ISO 19036 -
Measurement Uncertainty
Components
Uncertainty Components
Technical uncertainty – uncertainty resulting from operational
variability associated with the technical steps of the analytical
procedures
• estimated mathematically
ISO 19036 –
Concepts not included in
Measurement Uncertainty
calculation
Sampling uncertainty
Error associated with the drawing of the laboratory sample from a lot
under investigation
Caution: If intralaboratory reproducibility estimates from PT samples differ markedly from in-house estimates on “real”
samples of a similar type, the differences shall be recognized and recorded as may reflect differences in matrix and
microbial inoculum in the PT sample.
Intralaboratory reproducibility –
Non acceptable results
Criteria for excluding results from the calculation of intralaboratory reproducibility
(NOT for routine test results)
Colony-count techniques:
• results based on <30 counted colonies and counts > max countable number of colonies
MPN techniques:
• results used to be based on not less than 5 positive results across all dilutions tested for a single
test result.
Technical Uncertainty –
Intralaboratory reproducibility standard deviation
• transform cfu/g or cfu/ml results into log10 values
• calculate intralaboratory reproducibility standard deviation sIR :
• two values from each laboratory sample :
𝑛
1
𝑆𝐼𝑅 = (𝑦𝑖𝐴 − 𝑦𝑖𝐵 )2
2𝑛
𝑖=1
where:
i is the index number of the sample, i = 1 to n (n ≥10)
yiA, yiB are the log-transformed data, in log10 cfu/gr or ml, from condition A and B respectively
Technical Uncertainty –
Intralaboratory reproducibility standard deviation
• more than two values from each laboratory sample : calculate standard
deviation by one-way ANOVA (square root of within-groups mean square
value)
• Technical standard uncertainty: utech = sIR
Alternative options to estimate
Technical Uncertainty
• Reproducibility standard deviation derived from results of method
validation interlaboratory study
recommended by ISO 19036 only as a 2nd option to intralaboratory reproducibility
• For more details refer to ISO 21748 “Guidance for the use of repeatability,
reproducibility and trueness estimates in measurement uncertainty evaluation”
Reproducibility standard deviation derived
from results of method validation
interlaboratory studies
Condition
Laboratory may use the reproducibility standard deviation of an
interlaboratory validation study as an estimate of its technical
MU, if:
For a given method, not all the terms will be included in the calculation
of combined standard uncertainty
2 2 2
E.g., a colony-count method without partial confirmation: 𝑢𝑐 (y) = 𝑢𝑡𝑒𝑐ℎ + 𝑢𝑚𝑎𝑡𝑟𝑖𝑥 + 𝑢𝑃𝑜𝑖𝑠𝑠𝑜𝑛
Option 2: Combined uncertainty based on
reproducibility standard deviation alone
• log10 result with limits : <yLOQ log10 cfu/g [< yLOQ - U; yLOQ + U]
yLOQ – U = 0,959 - 0,940 = 0,019 log10cfu/g (rounded to 0,02)
yLOQ + U = 0,959 + 0,940 = 1,899 log10cfu/g (rounded to 1,90)
NOTE:
yLOQ-U can be negative, but 10yLOQ-U is always positive
Reporting Measurement Uncertainty
Include in the test report an explicit statement that
• the indicated MU is an expanded uncertainty
• statement of the confidence level
• indication that the MU has been estimated in accordance with
ISO 19036
“The reported expanded measurement uncertainty has been
estimated in accordance with ISO 19036 and is based on a standard
uncertainty multiplied by a coverage factor of k = 2, providing a level
of confidence of approximately 95 %.”
Reporting Measurement Uncertainty
If the MU is based on reproducibility standard deviation alone, this
shall be made clear in the test report:
“The reported expanded measurement uncertainty has been
estimated in accordance with ISO 19036 and is based on a standard
uncertainty multiplied by a coverage factor of k = 2, providing a level
of confidence of approximately 95 %. Combined standard
uncertainty has been taken as equal to the intralaboratory
reproducibility standard deviation.”
ISO 19036 –
Summary
Calculations
• Excel tool implementing the calculations of ISO 19036 (2019)
• Developed by Campden BRI (UK)
• Verified by WG 2 of ISO/TC 34/SC 9
• Freely available on line at
https://committee.iso.org/sites/tc34sc9/home/general-standards/content-left-area/culture-media/iso-19036-
estimation-of-measurem.html