Professional Documents
Culture Documents
2, 220e225, 2016
Ó Copyright 2016 by The International Society for Clinical Densitometry
1094-6950/19:220e225/$36.00
http://dx.doi.org/10.1016/j.jocd.2015.04.003
Technical Issues
Abstract
New densitometer installation requires cross-calibration for accurate longitudinal assessment. When replacing a
unit with the same model, the International Society for Clinical Densitometry recommends cross-calibrating by
scanning phantoms 10 times on each instrument and states that spine bone mineral density (BMD) should be within
1%, whereas total body lean, fat, and %fat mass should be within 2% of the prior instrument. However, there is
limited validation that these recommendations provide adequate total body cross-calibration. Here, we report a total
body cross-calibration experience with phantoms and humans.
Cross-calibration between an existing and new Lunar iDXA was performed using 3 encapsulated spine phantoms
(GE [GE Lunar, Madison, WI], BioClinica [BioClinica Inc, Princeton, NJ], and Hologic [Hologic Inc, Bedford,
MA]), 1 total body composition phantom (BioClinica), and 30 human volunteers. Thirty scans of each phantom
and a total body scan of human volunteers were obtained on each instrument.
All spine phantom BMD means were similar (within 1%; !0.010 g/cm2 bias) between the existing and new
dual-energy X-ray absorptiometry unit. The BioClinica body composition phantom (BBCP) BMD and bone mineral
content (BMC) values were within 2% with biases of 0.005 g/cm2 and 3.4 g. However, lean and fat mass and %fat
differed by 4.6%e7.7% with biases of þ463 g, 496 g, and 2.8%, respectively. In vivo comparison supported
BBCP data; BMD and BMC were within w2%, but lean and fat mass and %fat differed from 1.6% to 4.9%
with biases of þ833 g, 860 g, and 1.1%. As all body composition comparisons exceeded the recommended
2%, the new densitometer was recalibrated. After recalibration, in vivo bias was lower (!0.05%) for lean and
fat; 23 and 5 g, respectively. Similarly, BBCP lean and fat agreement improved.
In conclusion, the BBCP behaves similarly, but not identical, to human in vivo measurements for densitometer
cross-calibration. Spine phantoms, despite good BMD and BMC agreement, did not detect substantial lean and fat
differences observed using BBCP and in vivo assessments. Consequently, spine phantoms are inadequate for dual-
energy X-ray absorptiometry whole body composition cross-calibration.
Key Words: Cross-calibration; dual-energy X-ray absorptiometry; in vivo; spine phantom; whole body phantom.
220
DXA Cross-Calibration for Whole Body Composition 221
Fig. 1. Encapsulated spine phantoms. Three spine phantoms were evaluated for cross-calibration. A. GE Lunar (Madison,
WI), B. Hologic (Bedford, MA), and C. BioClinica Bona Fide phantom (Princeton, NJ).
Journal of Clinical Densitometry: Assessment & Management of Musculoskeletal Health Volume 19, 2016
222 Krueger et al.
vertebrae, each 1 of separate density and size. The acrylic is a between the new and existing densitometers (Table 1). The
2-phase mixture that provides a normal physiological soft tis- L1eL4 BMD mean biases between instruments were
sue of approx 24%. The BFP is the only phantom that has 0.010 g/cm2 with each phantom (Table 1).
been shown to match subject data for spine BMD by linear
regression (5,6). Total Body Phantom
A prototype total body phantom, the BBCP (Fig. 2) was The total body mean BMD and BMC were !1% different
also scanned. This phantom measures 60 cm in length, between the existing and new densitometers using the BBCP
36 cm in width, 9 cm in height, and weighs approx 16 kg (Table 1). Mean total body BMD and BMC bias was 0.005 g/
and therefore does not replicate the adult human body mass cm2 and 3.4 g, respectively (Table 1). Total body mean lean
or size. However, the phantom does replicate human body and fat differed by 4.6% and 7.8% respectively with corre-
composition proportionally in that it contains high-density sponding mean biases of þ463 and 496 g (Fig. 3A and B).
polyethylene, polyvinyl chloride, and an aperture of
aluminum to simulate different human soft tissue and bone Human In Vivo Measurement
compositions. The total body lean and fat mass initially differed by 1.6%
and 4.2% between the 2 densitometers with respective biases
Participants of þ833 and 860 g (Fig. 3C and D; Table 1). As the total
Thirty adult volunteers (15 males/15 females) were body fat value difference between instruments exceeded the
scanned once on each densitometer. Their mean (standard de- recommended 2%, the manufacturer was contacted and the
viation) age was 31.1 (10.8), range 20e60 yr; and mean (stan- new densitometer was recalibrated to improve lean and fat
dard deviation) body mass index was 24.1 (2.8), range mass agreement with the existing scanner. No adjustment
19.7e30.7 kg/m2. The entire body of all volunteers was con- was made to BMC or bone area.
tained within the scan field and positioning was per ISCD rec-
ommendations, that is, National Health and Nutrition Postrecalibration Analysis
Examination Survey style (2). Both scans in each individual After manufacturer recalibration, reanalysis of the human
were obtained on the 2 scanners within 60 min, and food in vivo scans revealed lower bias for lean and fat down to a
ingestion and voiding were prohibited between scans. mean difference of 23 and 5 g, respectively (Fig. 4A and
B; Table 2), thereby decreasing the between-scanner difference
Statistical Analysis to !0.05%. Similarly, agreement of the body composition
Agreement between the 2 densitometers was evaluated by phantom lean and fat improved, although still exceeded the
linear regression and Bland-Altman analysis using Analyse-it ISCD recommendation of !2% difference in body composi-
software, version 2.3 (Analyze-it Software Ltd, Leeds, UK). tion in that, despite recalibration, the BBCP biases for lean
and fat were reduced to 265 (2.5%) and 298 g (4.8%),
Results respectively. As spine BMD did not differ between instruments
and the intended use of this instrument was soft tissue body
Spine Phantoms composition assessment, BMD was not recalibrated, and there-
All 3 spine phantoms demonstrated a !1% difference in fore, the spine phantom reanalysis was not performed.
L1eL4 mean BMD and bone mineral content (BMC)
Discussion
In this study, the ISCD recommendations (1,2) for body
composition cross-calibration when replacing a densitometer
were implemented. To the authors’ knowledge, this study is
the first to compare data from spine phantoms and a total
body phantom, with in vivo cross-calibration. Although spine
phantoms are commonly used because of their wide accessi-
bility, these data demonstrate that using any of these 3 encap-
sulated spine phantoms does not adequately cross-calibrate
densitometers for body composition measurement. Impor-
tantly, although the L1eL4 mean BMD differed by !1%,
thus meeting the ISCD recommendations, spine phantom
data did not detect the need for manufacturer recalibration
Fig. 2. Body composition phantom. BioClinica body for total body fat and lean mass measurements. This is not
composition phantom prototype. Whole body phantom de- surprising given that spine phantoms are designed to mimic
signed to emulate bone, lean, and fat mass for the purpose spine BMD, not total body fat/lean mass and moreover are
of monitoring dual-energy X-ray absorptiometry scanners not representative of total body size. This study suggests
for body composition. The phantom measures that human in vivo measurement is necessary for optimal total
60 36 9 cm and weighs w16 kg. body DXA cross-calibration despite the same make and
Journal of Clinical Densitometry: Assessment & Management of Musculoskeletal Health Volume 19, 2016
DXA Cross-Calibration for Whole Body Composition 223
53,397
18,267
model densitometer. Utilization of these in vivo data demon-
New
1.276
unit
2420
3102
24.6
strated a need for manufacturer recalibration, which resulted
in total body fat and lean mass measurements that were
Mean
Human in vivo
virtually identical between the 2 instruments in the in vivo
Existing
54,230
17,507
unit
sample, differing by much less than the recommended 2%.
1.250
2432
3051
23.4
The ISCD also recommends cross-calibration using an
adequate total body phantom for lean and fat measurement
0.026
832.9
860.2
before and after hardware change of the same make and
11.3
50.9
1.14
Bias
model instrument (2). Although there are limited data
Total body
10,071
New
497.7
541.1
1.086
unit
6420
compared with in vivo cross-calibration (7,8). Indeed, the
37.7
ISCD Position Development Conference noted that addi-
Phantom and Human In Vivo Data and Comparisons on the Existing and New Instruments
Mean
462.9 10,534
unit
0.005 1.081
0.067 497.6
57.56 3.367 537.7
496.4 5924
2.84 34.9
The data presented here are the first to compare phantom
and in vivo evaluation for body composition cross-
calibration assessment. In this study, data from the Bio-
Bias
51.95
unit
d
d
d
Mean
d
d
d
d
d
d
44.51
48.00
unit
d
d
d
1.070
L1eL4
44.77
47.92
data point for lean and fat mass generated by this phantom
are close to the in vivo regression lines does suggests that
1.195 0.008
0.269
61.00 0.085
d
d
d
51.04
unit
1.188
d
d
d
BMC (g)
Lean (g)
Fat (%)
Journal of Clinical Densitometry: Assessment & Management of Musculoskeletal Health Volume 19, 2016
224 Krueger et al.
Fig. 3. AeD. Bland-Altman plots of lean and fat BioClinica body composition phantom data before and after recalibration.
A, B. Lean and fat differed between instruments by 4.6% and 7.8% with biases of þ463 and 496 g. C, D. After recalibration,
instrument agreement improved to 2.6% for lean and 4.8% in fat, with biases of þ265 and 298 g, respectively.
however, recalibration of the new instrument, if needed, will make and model of densitometer. As DXA technology differs
address this offset should it exist. between manufacturers, ISCD does recommend in vivo cross-
Limitations of the study include use of only a single total calibration when evaluating instruments for densitometers
body phantom, that this phantom was a prototype, and that from different manufacturers (2). Finally, as the in vivo sam-
only a single make and model of densitometer was evaluated. ple used in this trial was a relatively young and nonobese
The ability to evaluate other commercially available whole group, these data might not be replicated using other popula-
body phantoms would have been a substantial advantage. tions. However, the sample in this exercise was selected as it
However, to our knowledge, such comparisons do not exist. closely resembled the population that will be scanned on
Whether these findings would apply to other comparisons these instruments.
of like-model densitometers is not known. It is worthy of In conclusion, standard spine phantoms are inadequate for
note that this study, and the ISCD recommendations cited densitometer cross-calibration for total body fat and lean
in the introduction, apply to cross-calibration of the same measurements. Additionally, when replacing DXA scanners
Fig. 4. A, B. Lean and fat in vivo data before and after recalibration. A. It depicts the total body lean mass agreement as
assessed by linear regression and Bland-Altman plots before (open circles) and after densitometer recalibration (closed circles).
B. It depicts the same evaluations for total body fat mass before (open diamonds) and after recalibration (closed symbols).
Before recalibration, the regression equations were y 5 1.0037x þ 637.37 for lean and y 5 0.9953x 774.61 for fat. After
recalibration, the regression equations were y 5 0.9948x þ 259.91 and y 5 1.0034x 64.702, respectively. There was a change
in the slope and bias of 0.009 and 855.6 g, respectively, in lean and 0.008 and 855.6 g, respectively, in fat.
Journal of Clinical Densitometry: Assessment & Management of Musculoskeletal Health Volume 19, 2016
DXA Cross-Calibration for Whole Body Composition 225
Table 2
BBCP Phantom and Human Comparisons on the Existing and New Instruments After Recalibration
for the purpose of total body composition assessment, even 3. Miller CG. 2007 Organization of the clinical trial by the sponsor.
with the same make and model instrument, in vivo cross- In: Clinical Trials in Osteoporosis. Pearson D and Miller CG,
calibration is needed at this time to ensure comparable eds. 2nd ed. London, UK: Springer-Verlag, 75e90.
4. Anonymous 2000 Body composition procedures manual Na-
body composition results. tional Health and Nutrition Examination Survey (NHANES).
CDC, Atlanta, GA, pp 6e27 and 6e30.
5. Pearson D, Cawte SA, Green DJ. 2002 A comparison of phan-
References toms for cross-calibration of lumbar spine DXA. Osteoporos
Int 13:948e954.
1. Baim S, Binkley N, Bilezikian JP, et al. 2008 Official Positions 6. Pearson D. 2002 Standardization and pre-trial quality control.
of the International Society for Clinical Densitometry and exec- In: Clinical Trials in Osteoporosis. Pearson D and Miller CG,
utive summary of the 2007 ISCD Position Development Confer- eds. New York, NY: Springer, 43e65.
ence. J Clin Densitom 11:75e91. 7. Guo Y, Franks PW, Brookshire T, Antonio Tataranni P. 2004 The
2. Hangartner TN, Warner S, Braillon P, et al. 2013 The Official intra- and inter-instrument reliability of DXA based on ex vivo
Positions of the International Society for Clinical Densitometry: soft tissue measurements. Obes Res 12(12):1925e1929.
acquisition of dual-energy X-ray absorptiometry body composi- 8. Diessel E, Fuerst T, Njeh CF, et al. 2000 Evaluation of a new
tion and considerations regarding analysis and repeatability of body composition phantom for quality control and cross-
measures. J Clin Densitom 16(4):520e536. calibration of DXA devices. J Appl Phys 89(2):599e605.
Journal of Clinical Densitometry: Assessment & Management of Musculoskeletal Health Volume 19, 2016