Professional Documents
Culture Documents
S7 / OPMI® Sensera®
About this manual These Instructions for Use are part of the delivery package.
• Please read them carefully before using the device.
• Keep them at the site of use of the device.
• Store them for the entire service life of the device.
• Pass them on to any subsequent owner or user of the device.
Orientation aids – At the beginning of these Instructions for Use there is a summary of the
various sections that provides an overview of all the topics.
– The contents of each chapter are specified in detail at the beginning of
each chapter.
– The index facilitates searching via keywords.
Scope The following Instructions for Use apply to the OPMI® Sensera® with the
following suspension systems and identification codes:
– S7 floor stand, reference number: 000000-1087-851
– S7 ceiling mount - rigid, reference number: 000000-1103-022
– S7 ceiling mount with lifting column, reference number: 000000-1226-405
Version 13.0
Page 2 G-30-1434-en
S7 / OPMI Sensera
Section overview
Version 13.0
G-30-1434-en Page 3
S7 / OPMI Sensera
Version 13.0
Page 4 G-30-1434-en
S7 / OPMI Sensera Safety measures
Safety measures
Version 13.0
G-30-1434-en Page 5
Safety measures S7 / OPMI Sensera
Version 13.0
Page 6 G-30-1434-en
S7 / OPMI Sensera Safety measures
Key to symbols
We would like to inform you about safety requirements with which you must
comply when working with this system. This chapter contains a summary of
the most important information concerning matters relevant to instrument
safety.
Hazard symbols
WARNING Indicates a hazard which may lead to fatal injuries or severe injuries if
precautions are not followed.
CAUTION Indicates a hazard which may lead to moderately severe injuries
if precautions are not followed.
NOTE Indicates a hazard which may lead to minor injuries or property damage
if precautions are not followed.
Information symbols
The following information symbols are used in these Instructions for Use:
– Lists
Requirements for an action
• Prompt for action
Result of an action
Additional information and tips
Version 13.0
G-30-1434-en Page 7
Safety measures S7 / OPMI Sensera
Target audience
These Instructions for Use are intended for physicians, nurses and other
medical and technical staff who prepare, operate or maintain the device after
appropriate training. It is the duty of the device owner/operator to train and
brief all operating personnel.
Additional service activities are not part of these instructions for use. They will
be performed by staff specially trained for this purpose by ZEISS.
Field of use
Intended use
Normal use
Version 13.0
Page 8 G-30-1434-en
S7 / OPMI Sensera Safety measures
CAUTION Live video images and recorded video sequences, video clips (cut sequences)
and single images shall not be used for diagnosis purposes. Visualized images
may contain variations in shape, contrast and color.
Loss of warranty!
The manufacturer is not liable for damage caused by unauthorized use of
the device. Furthermore, such unauthorized modifications will render the
warranty and any claims under warranty invalid.
Version 13.0
G-30-1434-en Page 9
Safety measures S7 / OPMI Sensera
The correct use of the device is absolutely vital for safe operation. Therefore,
please thoroughly familiarize yourself with the content of these Instructions
for Use before starting up the system. Please comply with the Instructions for
Use of the other system components as well. You can obtain additional
information from ZEISS Service.
• Only operate the device within the scope of its intended use.
• This device must be used by properly instructed and trained personnel
only. It is the duty of the customer or institution operating the equipment
to train and instruct all staff using the equipment.
• Keep the Instructions for Use available at all times for the operating
personnel.
• Note the symbols and labels attached to the device! (see Page 24)
• Use only original accessories or consumables approved by Carl Zeiss.
Legal regulations • Observe the applicable legal accident prevention and occupational safety
provisions of your country.
• This device must not be modified without the manufacturer’s approval. If
the system is modified after consultation with the manufacturer, suitable
inspections and testing must be completed to ensure subsequent safe use.
The manufacturer is not liable for damage caused by unauthorized use of
the device.
Medical electrical Any additional equipment you connect to medical electrical devices must
systems demonstrably comply with the applicable standards and directives (e.g.
IEC 60950-1 for data processing equipment). In addition, all configurations
must meet the standard requirements for medical electrical systems (see
IEC 60601-1).Anyone connecting additional equipment to medical electrical
devices is a system configurer and as such is responsible for compliance of the
system with the normative requirements for medical electrical systems.
Please note that local legislation takes priority over the requirements of the
above-mentioned standards. If you have any questions, contact your local
dealer or ZEISS Service.
Version 13.0
Page 10 G-30-1434-en
S7 / OPMI Sensera Safety measures
Ambient conditions • Do not use the power-operated devices included in the delivery package
– in areas where there is a risk of explosion,
– at a distance of less than 25 cm from flammable anesthetics or volatile
solvents such as alcohol, benzine or similar substances.
• Do not use or store the device in damp rooms. Do not expose the system
to dripping water, running water, gushing water, sprayed water or
splashes.
• Ensure that fluids cannot enter the device.
• Ensure that the installation conditions and the use of the device meet
surgical requirements:
– Low vibration
– Clean environment
– Avoidance of extreme mechanical stress
Electrical safety • Always switch off the device before you disconnect connect it from the
electricity network or connect it, if you will not be using it for an extended
period of time, or want to clean its surface.
• Switch off the system at the power switch if you notice any smoke, sparks
or unusual noise. In this case, do not use the device again before it has
been repaired by our Service department.
A potential equalization connector is provided on the connection panel of the
instrument (siehe IEC 60601-1). This connector can be used to connect other
active devices to the same ground potential or for redundant connection to
the protective ground.
• Do not use multiple sockets!
• Never open the device! The device contains freely accessible live
components. If you remove the housing, you run the risk of electric shock.
• Do not touch the device if your body is electrostatically charged and the
device is not grounded.
• Please note the information on EMC (electromagnetic compatibility) in the
chapter "System data" on Page 233.
Version 13.0
G-30-1434-en Page 11
Safety measures S7 / OPMI Sensera
Transport and service • The system may be transported over long distances (e.g. during
relocation, return for repair, etc.) only in its original packaging or in special
return packaging. Please contact your dealer or ZEISS Service for this
purpose.
• This instrument is a high-grade technological product. To ensure optimum
performance and safe working order, we recommend having it checked
by our service representative as part of regular scheduled maintenance.
• In order to prevent any impairment of the device’s safety due to age,
wear, etc., the user must ensure that the device is subjected to the
necessary safety checks (see section "Care and maintenance").
• Modifications and repairs of this device or any equipment operated
together with this device may only be performed by ZEISS SERVICE or
other specialists authorized by Carl Zeiss.
Optics and light • Never look at the sun through the binocular tube, the objective lens or an
eyepiece.
• Avoid looking directly into the light source, e.g. into the microscope
objective lens or into the light guide, in order to prevent damage to the
eye!
• Start with the lowest brightness setting and gradually increase brightness
up to the necessary and still admissible level.
• With increasing age of the light source, the actual illumination intensity
delivered at the respective setting decreases (this is a normal property of
the system).
Version 13.0
Page 12 G-30-1434-en
S7 / OPMI Sensera Safety measures
Version 13.0
G-30-1434-en Page 13
Safety measures S7 / OPMI Sensera
• Before every use, make certain to compensate for any added weight; this
will enable the surgical microscope to maintain its balance in all positions
of the working range.
• Before every use, limit the stroke of the suspension arm to prevent any
contact with the patient if the surgical microscope is lowered accidentally.
• Check the surgical microscope for sufficient freedom of movement. The
device itself or accessories may be damaged if conflicting with each other.
• To ensure that no accessories can fall off the system, check the unit before
each use to makes sure that the accessories are firmly attached and that
the securing screws are properly tightened.
• When connecting accessories make sure that sufficient free space is
provided and that the surgical microscope does not touch the patient.
• Before every use, check the user settings in the software to avoid
unexpected behavior of the system.
• Always route cables so that they do not hinder the user.
• Never connect any unknown or defective devices to the electrical
interfaces (e.g. for data, video or foot control panel).
• Connect components only to properly grounded outlets.
• When configuring your system, comply with the requirements of
IEC 60601-1, chapter 16 (in some countries IEC 60601-1-1 is still valid).
• Reattach any covers or caps that were removed from the system.
Close any existing openings with the corresponding caps.
Before starting the system, remove the protective cover used for
transportation from the objective lens.
• If the xenon lamp is used beyond its maximum service life of 500 hours,
a sudden failure may occur. Change to the backup lamp (see Page 184)
or replace the lamp module in a timely manner (see Page 204). Reset the
service hour counter of the xenon illumination.
• Keep a replacement lamp (only for halogen light source) ready at hand.
• Never cover any ventilation openings. This may cause the light source of
the device to overheat and fail.
Version 13.0
Page 14 G-30-1434-en
S7 / OPMI Sensera Safety measures
• Never use force when trying to connect any electrical connectors (plugs,
sockets). If a connection cannot be made, check to make sure the plug fits
the socket. If any of the connectors are damaged, have the ZEISS service
team or authorized service representative repair them.
• Position the suspension system in such a way that you can slide or swing
it away from the patient at any time.
• Go through the checklist in the chapter "Operation".
• "Depending on the application, take appropriate precautions to ensure
that the surgical procedure or treatment can be completed without the
full functionality of this microscope (e.g. illumination) if necessary.
• If you operate the device in a sterile area, make sure that you use the
corresponding sterile accessories for the device.
The single-use sterile drapes can be used to cover the device.
• When fitting the sterile drape, ensure there is sufficient room for
movement of the microscope carrier and surgical microscope.
• Comply with the instructions for use of the sterile drape and attach the
drape in such a way that your field of view is not impaired.
More specific information about the ceiling mount:
• Move the lift arm or the lifting column down to its lower end position.
• Do not activate the lifting column or the lift arm during surgery. Do not
use the lifting column or the lift arm for focusing.
More specific information about the floor stand:
• Use the grab handle on the floor stand for moving.
• Mind the maximum vertical clearance when passing through doorways.
• Avoid any jolting.
• Be extremely careful when moving over slopes.
• Do not park the device on sloped surfaces.
• Secure the system in position by locking at least three of the tabs on the
stand base to prevent the stand from rolling away inadvertently.
• Do not exceed the instrument tray’s load capacity of 13 kg.
Version 13.0
G-30-1434-en Page 15
Safety measures S7 / OPMI Sensera
During operation
• Never look directly into the light source, e.g. a microscope objective lens
or a light guide, as this might cause phototoxic injury to the eye.
• Never leave a device unattended when its light source is switched on.
• Defective or unidentified accessories may lead to increased leakage
current on the device and injure the patient. Never connect any defective
or unidentified accessories and never touch the power outlet or video
interfaces while in contact with the patient.
• If you replace the lamp shortly after it has failed, the lamp will still be very
hot. Use heat-resistant protective gloves when replacing the lamp to
prevent burns.
• If an error, failure or malfunction occurs which you cannot correct with
the aid of the chapter "What to do in the event of malfunctions", attach
a sign to the system stating it is out of order and contact the ZEISS service
staff or authorized specialists.
• Do not pull at the light guide, power cord or any other connecting cables.
• Constantly monitor the system during use.
• Risk of crushing - mind your fingers!
Fingers may be crushed in the areas marked with the "Risk of crushing"
sign.
Do not touch these areas while the device is being moved.
• Always use the master switch to turn off the device if it is not in use.
• Insufficient, erroneous or incorrect cleaning and/or disinfection that does
not comply with these Instructions for Use can expose the patient or
medical staff to a considerable risk of infection.
• Step on at least three pedals to secure the floor stand in position.
Make sure that the floor stand is stable and cannot roll away.
Version 13.0
Page 16 G-30-1434-en
S7 / OPMI Sensera Safety measures
General
The xenon illumination is a light source with high intensity. If used improperly,
excessive illumination intensities may lead to third-degree burns. Keep the
illuminated tissue moist and make sure it is sufficiently rinsed. Carefully
monitor the effect of the illumination on the tissue, particularly under the
following circumstances:
– During longer interventions on skin and tissue using objective lenses with
short focal depth (short working distance),
– During interventions on poorly perfused tissue,
– When the brightness control for the xenon lamp is set to a high
illumination intensity.Please bear in mind that the values displayed for the
light intensity may not be used as a measuring function. They are intended
only as guide values.
• Only change the xenon lamp module after switching off the system.
The igniter generates a high voltage when the lamp is on.
• Any type of radiation has a damaging effect on biological tissue. This is
also true for the light illumination the surgical field. You should therefore
reduce brightness, the illuminated field size and the light exposure time to
the absolute minimum required.
In the event of malfunctions, if the brightness can no longer be reduced,
you must take this into account in the duration of illumination.
Version 13.0
G-30-1434-en Page 17
Safety measures S7 / OPMI Sensera
Surgery-related factors
– The selected intensity of the light source is a major factor for the risk of
injury. It should always be set to the minimum required for the surgical
procedure to be performed.
– The size of the luminous field influences the risk of injury in two respects:
– With a large illuminated-field diameter, areas of the skin are
illuminated which are not monitored as strictly by the surgeon and
which are not moistened sufficiently. These areas are particularly
exposed to the risk of injury. Such injuries can be prevented by
adjusting the luminous field size to the smallest size required for that
particular operation.
– If the luminous field is reduced in size, the intensity increases because
the light is more strongly focused. So, if possible, the intensity should
be lowered as soon as the size of the luminous field is reduced.
– A long surgical procedure increases the risk of injury, in particular if
a standard procedure takes considerably longer than usual.
– Injuries in the peripheral area can be prevented by covering this area with
wet, sterile gauze. The risk is increased if dry drapes are used to cover such
areas.
– The interaction of heat and antimicrobial substances in incision foils may
lead to an increase in the patient's reaction to these substances.
– It should also be considered that some areas of the body may be more
sensitive than others.
– Certain preparations of the surgical field, local vasoconstrictive
medications and incision drapes may also result in a higher risk of injury.
Patient-related factors
– The general condition of a patient's health may contribute to the risk of
injury.
– The skin type may also play a major role for the risk of injury.
– Certain medications also affect the sensitivity to light.
Recommendations
Due to the large number of different factors involved and the lack of scientific
publications on this topic, ZEISS cannot provide guidance on acceptable
illumination intensities and exposure durations. However, the surgical
microscope has several features that can help the user to reduce the risk of
burn injuries:
Version 13.0
Page 18 G-30-1434-en
S7 / OPMI Sensera Safety measures
– Using the buttons on the handgrip or foot control panel, the surgeon can
then set the illumination intensity to the value required for the procedure.
Please note that the intensity increases with decreasing luminous field size
if the Spot function is used. For this reason, the intensity should only be
set only after the size of the luminous field has been changed.
– Never leave a device unattended when its light source is switched on.
– - Switch off the light when the microscope is not used and make sure that
it is not aimed at unprotected bare skin.
Please note that most burns affected the skin around the incision. The most
important measures to prevent burns are reduction of the size of the
illuminated field using the spot function and the coverage of peripheral areas
with sterile, moist gauze. The area of the incision should be constantly
moistened.
Concluding remarks
ZEISS recommends that the lighting of the surgical field be reduced as far as
possible while ensuring the safety of the patient and a good microscopic
image.
Please comply with the warnings and safety instructions in this regard.
Version 13.0
G-30-1434-en Page 19
Safety measures S7 / OPMI Sensera
Version 13.0
Page 20 G-30-1434-en
S7 / OPMI Sensera Safety measures
3 2
2
Version 13.0
G-30-1434-en Page 21
Safety measures S7 / OPMI Sensera
0 No filter
1 GG475 filter: to protect the patient's eyes from unnecessary (blue)
radiation (which can cause retinal injury) during treatment.
2 KK 40 filter: to increase color temperature
3 No filter
Version 13.0
Page 22 G-30-1434-en
S7 / OPMI Sensera Safety measures
Lifting column
1 2 Selector switch
for setting the optimum viewing height of the surgical microscope or for
downward movement into the working position.
As long as you keep the selector switch in the turned position, the lifting
column in the stand base moves upwards (2) or downwards (1), depending
on the switch position. When you release the selector switch, the lifting
column stops immediately.
The switching technology causes a delay of approx. 2 seconds at the upper
and lower end positions of the lifting column. After this time elapses, you can
move the column in the opposite direction again by activating the selector
switch again.
With a maximum actuation time of one minute, the lifting column must be
turned off for at least nine minutes because its motor might overheat
otherwise.
– Before raising or lowering the stand, make sure that there is sufficient
clearance to other objects to prevent collisions.
Version 13.0
G-30-1434-en Page 23
Safety measures S7 / OPMI Sensera
Version 13.0
Page 24 G-30-1434-en
S7 / OPMI Sensera Safety measures
2
8
3
6
2
1 4
Version 13.0
G-30-1434-en Page 25
Safety measures S7 / OPMI Sensera
13
2
1
12
9
12
10 11
Version 13.0
Page 26 G-30-1434-en
S7 / OPMI Sensera Safety measures
2
7 3
4
Version 13.0
G-30-1434-en Page 27
Safety measures S7 / OPMI Sensera
Version 13.0
Page 28 G-30-1434-en
S7 / OPMI Sensera Safety measures
1 1
2 1
5 4 1 3
Version 13.0
G-30-1434-en Page 29
Safety measures S7 / OPMI Sensera
9 Rating plate
The rating label contains the following information:
– Manufacturer's symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Device Name S7 / OPMI Sensera
– Reference number
– Rated Voltage (115 V): 100 V - 125 V
(230 V): 220 V - 240 V
– Connected load (115 V): max. 850 VA
(230 V): max. 1250 VA
– Line frequency range 50 Hz - 60 Hz
– Maximum total weight (with Max. 190 kg
installed safe workload)
– CSA approval
– CE marking
Version 13.0
Page 30 G-30-1434-en
S7 / OPMI Sensera Safety measures
8 9 10
Version 13.0
G-30-1434-en Page 31
Safety measures S7 / OPMI Sensera
12 SIP-label
The SIP identification label includes the following information:
– Manufacturer (company name)
– contact details for the manufacturer,
i.e. phone number, fax number and e-mail address of the local contact
of the national ZEISS sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your system.
13 "Observe the Instructions for Use" label
Observe the Instructions for Use.
Version 13.0
Page 32 G-30-1434-en
S7 / OPMI Sensera Safety measures
11
12
13
Version 13.0
G-30-1434-en Page 33
Safety measures S7 / OPMI Sensera
Version 13.0
Page 34 G-30-1434-en
S7 / OPMI Sensera Safety measures
Version 13.0
G-30-1434-en Page 35
Safety measures S7 / OPMI Sensera
Version 13.0
Page 36 G-30-1434-en
S7 / OPMI Sensera Safety measures
1 2 1 3
5
1 6 5
9 8 7
Version 13.0
G-30-1434-en Page 37
Safety measures S7 / OPMI Sensera
10 SIP-label
The SIP identification label includes the following information:
– Manufacturer (company name)
– contact details for the manufacturer,
i.e. phone number, fax number and e-mail address of the local contact
of the national ZEISS sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your system.
11 Rating label (for S7 ceiling mount with rigid column)
The rating label contains the following information:
– Manufacturer's symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Device Name S7 / OPMI Sensera
– Reference number
– Rated Voltage (115 V): 100 V - 125 V
(230 V): 220 V - 240 V
– Connected load (115 V): max. 450 VA
(230 V): max. 450 VA
– Line frequency range 50 Hz - 60 Hz
– CSA approval
– CE marking
Version 13.0
Page 38 G-30-1434-en
S7 / OPMI Sensera Safety measures
10 11/12
Version 13.0
G-30-1434-en Page 39
Safety measures S7 / OPMI Sensera
Version 13.0
Page 40 G-30-1434-en
S7 / OPMI Sensera Safety measures
2 1
Version 13.0
G-30-1434-en Page 41
Safety measures S7 / OPMI Sensera
4 Lamp change
The plate graphically shows the three steps involved in changing the lamp.
A - Press the button
B - Pull out the lamp insert module
C - Change bulbs
Version 13.0
Page 42 G-30-1434-en
S7 / OPMI Sensera Safety measures
1 2 3 4 5 6
Version 13.0
G-30-1434-en Page 43
Safety measures S7 / OPMI Sensera
Version 13.0
Page 44 G-30-1434-en
S7 / OPMI Sensera Safety measures
2 Remote connector
Identifies a connector to which devices with a maximum breaking
capacity of 24V/0.5A may be connected.
Version 13.0
G-30-1434-en Page 45
Safety measures S7 / OPMI Sensera
Version 13.0
Page 46 G-30-1434-en
S7 / OPMI Sensera Device overview
Device overview
Version 13.0
G-30-1434-en Page 47
Device overview S7 / OPMI Sensera
Version 13.0
Page 48 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 49
Device overview S7 / OPMI Sensera
Features
Suspension systems
The S7 floor stand is a suspension system for ZEISS surgical microscopes.
It comprises a suspension arm, a carrier arm, a stand column and a stand base.
A grab handle is mounted on the stand column. It is used to move the stand
and to suspend the foot control panel. The stand column is equipped with
cable supports on the left and right for winding up cables before the unit is
relocated.The stand base is extremely easy to move on its four casters. It has
been designed in such a way that high stability is ensured even with
unfavorable loading of the stand. The locking tabs permit you to reliably
secure the floor stand in position at the site of use.
The weight may be underestimated as the suspension system can be moved
very easily. Therefore, move the stand slowly and carefully!
The S7 ceiling mount with rigid column is a suspension system for ZEISS
surgical microscopes. It comprises a suspension arm, a carrier arm, a ceiling
column and a ceiling flange.
The S7 ceiling mount with lifting column is a suspension system for ZEISS
surgical microscopes. It comprises a suspension arm, a carrier arm, a ceiling
flange and a lifting column. This motorized lifting column permits the viewing
height to be perfectly matched to the surgeon's requirements.
The carrier arm of all suspension systems contains the control unit with all
electrical supply systems required for the control of a motorized surgical
microscope. You can control the motorized functions using the handgrips, or
via a foot control panel (optional ). A light guide (fiber optic illumination)
directs the light from the lamp housing in the suspension system to the
surgical microscope.
The suspension arm of all types of suspension system permits almost effortless
Version 13.0
Page 50 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 51
Device overview S7 / OPMI Sensera
System components
Version 13.0
Page 52 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 53
Device overview S7 / OPMI Sensera
Set-up
Version 13.0
Page 54 G-30-1434-en
S7 / OPMI Sensera Device overview
1 2 3 4 5
Version 13.0
G-30-1434-en Page 55
Device overview S7 / OPMI Sensera
Version 13.0
Page 56 G-30-1434-en
S7 / OPMI Sensera Device overview
4
2
1
5
6
2
Version 13.0
G-30-1434-en Page 57
Device overview S7 / OPMI Sensera
8 Dust cover
9 Handgrips (removable)
for moving the surgical microscope and triggering the motorized OPMI
functions (see points 14 to 17).
10 Clamping screw
for optimal positioning of the handgrips.
11 Zoom wheel (emergency operation)
If the motorized functions fail, you can manually adjust the magnification
after removing the plastic cover from the zoom wheel.
12 Focusing knob
for manual adjustment of the image definition (focus, working distance).
13 Spot illumination knob
for continuous adjustment of the illuminated-field diameter.
14 Focusing knob for use of a laser micromanipulator (option)
for manual adjustment of the image definition (focus) when a laser
micromanipulator is used.
Version 13.0
Page 58 G-30-1434-en
S7 / OPMI Sensera Device overview
9 10 11 12 13
14
Version 13.0
G-30-1434-en Page 59
Device overview S7 / OPMI Sensera
Version 13.0
Page 60 G-30-1434-en
S7 / OPMI Sensera Device overview
15
16
17
F
18
19
20
21
22
Z
Version 13.0
G-30-1434-en Page 61
Device overview S7 / OPMI Sensera
Version 13.0
Page 62 G-30-1434-en
S7 / OPMI Sensera Device overview
Straight tube
1 2 3 4 5 6 7
Version 13.0
G-30-1434-en Page 63
Device overview S7 / OPMI Sensera
Version 13.0
Page 64 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 65
Device overview S7 / OPMI Sensera
Floor stand S7
Set-up
1 Suspension arm
2 Lamp housing (either xenon or halogen illumination)
3 Support arm
4 Control panel
5 Connector panel
6 Handle
7 Cable supports (2x, on the right and left of the stand column)
8 Stand column
9 Stand base
Version 13.0
Page 66 G-30-1434-en
S7 / OPMI Sensera Device overview
7
8
Version 13.0
G-30-1434-en Page 67
Device overview S7 / OPMI Sensera
Suspension arm
Version 13.0
Page 68 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 69
Device overview S7 / OPMI Sensera
Version 13.0
Page 70 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 71
Device overview S7 / OPMI Sensera
Version 13.0
Page 72 G-30-1434-en
S7 / OPMI Sensera Device overview
1 2 3 4 5 6 7 8 9 10 11
Version 13.0
G-30-1434-en Page 73
Device overview S7 / OPMI Sensera
The S7 floor stand can be equipped or retrofitted with an instrument tray (1).
In the case of retrofitting, our Service staff or an authorized person will install
the instrument tray on your suspension system.
The instrument tray (1) can carry a maximum of 13 kg. It is designed to hold
a video recorder, for example. Further accessories can be installed on the
instrument tray using the strap provided.
Version 13.0
Page 74 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 75
Device overview S7 / OPMI Sensera
S7 ceiling mount
Set-up
1 Ceiling flange
2 Stand column
3 Control panel
4 Connector panel
5 Lamp housing (either xenon or halogen illumination)
6 Suspension arm
7 Support arm
Version 13.0
Page 76 G-30-1434-en
S7 / OPMI Sensera Device overview
7
3
6
4
Version 13.0
G-30-1434-en Page 77
Device overview S7 / OPMI Sensera
Version 13.0
Page 78 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 79
Device overview S7 / OPMI Sensera
Version 13.0
Page 80 G-30-1434-en
S7 / OPMI Sensera Device overview
1 2 3 4 5 6 7 8 9
Version 13.0
G-30-1434-en Page 81
Device overview S7 / OPMI Sensera
Set-up
1 Ceiling flange
2 Lifting column
3 Control panel
4 Connector panel
5 Lamp housing (either xenon or halogen illumination)
6 Suspension arm
7 Support arm
Version 13.0
Page 82 G-30-1434-en
S7 / OPMI Sensera Device overview
7
3
6
4
Version 13.0
G-30-1434-en Page 83
Device overview S7 / OPMI Sensera
1 Lifting column
2 Selector switch
for setting the optimum viewing height of the surgical microscope or for
3 4 downward movement into the working position.
As long as you keep the selector switch in the turned position, the lifting
column moves downward (3) or upward (4), depending on the switch
position. When you release the selector switch, the lifting column stops
immediately.
The switching technology causes a delay of approx. 2 seconds at the
upper and lower end positions of the lifting column. After this time
elapses, you can move the column in the opposite direction again by
activating the selector switch again.
The lifting column is used to move the microscope into position for
surgery prior to the surgical procedure.Do not constantly move the lifting
column up and down, since a thermal cut-out will then automatically
deactivate the drive motor. If this occurs, the lifting column cannot be
moved until the motor has cooled down.
Note: With a maximum actuation time of one minutes, the lift column
must be turned off for at least nine minutes because the motor of the lift
column can overheat otherwise.
Version 13.0
Page 84 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 85
Device overview S7 / OPMI Sensera
Version 13.0
Page 86 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 87
Device overview S7 / OPMI Sensera
Version 13.0
Page 88 G-30-1434-en
S7 / OPMI Sensera Device overview
1 2 3 4 5 6 7 8 9 10
Version 13.0
G-30-1434-en Page 89
Device overview S7 / OPMI Sensera
1 Remote connection
for controlling external devices (e.g. MEDIALINK 100) with a max
switching capacity of 24 V / 0.5 A.
2 connection strip for switching element (option)
Connection option for a foot control panel
Version 13.0
Page 90 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 91
Device overview S7 / OPMI Sensera
Illumination system
Two different types of illumination are available for the suspension system.
1 Halogen illumination
The halogen light source is equipped with an illumination system for fiber
illumination. The lamp housing contains a backup lamp which
automatically swings into position when the first lamp fails.
2 XENON illumination
The xenon light source is equipped with an illumination system for fiber
illumination. The xenon lamp generates light whose spectrum resembles
that of natural daylight. Regardless of the brightness setting, the color
temperature of the light always remains the same. Normal daylight film
without any additional conversion filters can therefore be used for
photographic documentation. The lamp module contains two xenon
lamps. The second lamp is used as a backup lamp which must be manually
swung into the illumination beam path when the first lamp fails. You have
to pull out the lamp module all the way before being able to swing in the
backup lamp.
Version 13.0
Page 92 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 93
Device overview S7 / OPMI Sensera
The illumination system has been designed for fiber illumination. The lamp
housing contains a backup lamp which automatically swings into position
when the first lamp fails.
1 Lamp module
2 Ventilation grid
NOTE Overheating of the lamp module!
If ventilation grids are covered e.g. by drapes, this may lead to overheating
of the ventilation modules and deactivation of the lamp.
• Never cover the ventilation grids.
• Lamps switched off due to overheating will be automatically switched on
again when they have cooled down.
3 Flap
The flap is used as the mechanical indicator for the operating status of the
halogen lamps.
– When the flap is closed, the main lamp is operative.
– When the flap is open, the main lamp has failed. The backup lamp
is on.
4 Manual switching to the backup lamp
If the automatic activation function fails, press this button to switch on the
backup lamp.
5 Opening the lamp module
When you press this button, the slide-in lamp module is partially ejected.
Pull out the lamp module all the way to change the lamp.
6 Filter selector
The filter selector knob has four positions:
0 No filter
1 GG 475 filter: to protect the patient's eye during surgery from unnec-
essary (blue) radiation (retinal injury)
2 KK 40 filter: to increase color temperature
3 No filter
Version 13.0
Page 94 G-30-1434-en
S7 / OPMI Sensera Device overview
1
2
Version 13.0
G-30-1434-en Page 95
Device overview S7 / OPMI Sensera
Version 13.0
Page 96 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 97
Device overview S7 / OPMI Sensera
When inserting a new lamp module, make sure that the knob (2) is set to "1".
If the first lamp fails, switch to the second lamp in logical sequence.
3 Indicator: backup lamp is in use
When the segment in knob (2) lights up, the backup lamp is in use.
4 Filter selector
The filter selector knob has two positions:
0 No filter
1 Filter swung in
(no integrated filter in the standard configuration)
5 Counter
The counter records the service hours of the xenon light source.
• Change the xenon lamps after a maximum operating time of 500 hours to
avoid any rupture of the xenon lamps. Then reset the counter to "0" by
pressing the "Reset" button (6).
6 Reset button
The reset button resets the service hour counter to "0".
7 Opening the lamp module
When you press this button, the slide-in lamp module is partially ejected.
• To change the lamp, pull out the lamp module as far as it will go. Turn the
knob (2) through 180° until it snaps in place. This moves the backup lamp
into the illumination beam path.
Version 13.0
Page 98 G-30-1434-en
S7 / OPMI Sensera Device overview
3
4
Version 13.0
G-30-1434-en Page 99
Device overview S7 / OPMI Sensera
The control panel consists of a graphic display (1) with control keys (3).
All functions can be interactively controlled via menus. The display shows
the selected functions and settings. Using control buttons (3) you can move
through the menus and activate a selection field. The field activated is then
highlighted by a dark background. Press Enter button (4) to confirm your
selection or entry.
For entering, briefly press any of control buttons (3). In some settings, the
control buttons have a repeat function, i.e. constant pressing of a button
causes the command to be repeated.
The menus are clearly laid out and user-friendly. The name of the menu is
given in bar (2) at the bottom. At the right-hand end of the bar, a small
pictogram is shown for quick orientation. At the left-hand end of the bar,
the activation field (OK) is located. If the background of this field is dark,
press the "Enter" button to confirm your entry.
The following diagram depicts how the menus are linked.
Power switch (5)
For switching the system on and off. When the system has been switched on,
the green lamp in the switch is lit.
The power switch contains an automatic circuit breaker. If a short circuit
occurs, the automatic circuit breaker will respond and switch the system off.
2
4
3
Version 13.0
Page 100 G-30-1434-en
S7 / OPMI Sensera Device overview
Version ......
300 mm
USER 1
USER 2 NONE
USER 3 NONE
Off Off
Off
Standard Close
12.5 x
Version 13.0
G-30-1434-en Page 101
Device overview S7 / OPMI Sensera
User menu
The User menu comprises one standard profile and three other user profiles.
USER 1
When a user profile is selected, the system imports the settings stored under
USER 2
this profile, e.g. focusing speed, zoom link setting, zoom speed and handgrip
USER 3
settings. The stored start values ("Start at") for the magnification factor, the
focus position and the brightness are automatically set. The three user profiles
are used to adjust these settings for different users.
When the system is started up for the first time, the Standard User is
displayed; at every subsequent start-up, the user active last is displayed.
– The Standard User profile allows the adjustment of settings, but these
settings cannot be saved (fixed factory settings).
– In the Standard User profile, all "Start at" fields are deactivated. No bar is
displayed; it is not possible to adjust any settings.
Version 13.0
Page 102 G-30-1434-en
S7 / OPMI Sensera Device overview
Functions of buttons:
– When the input keyboard has been activated with the Edit button (1),
the input function is active.
2
– Select the individual characters using the arrow keys and confirm the
characters by pressing the "Enter" button.
You must activate the OK button (2) using the arrow buttons and then
press the Enter button to save the name entered under the selected user.
After selecting a user or an application, press the "Enter" button to return
to the "Surgery" menu.
Version 13.0
G-30-1434-en Page 103
Device overview S7 / OPMI Sensera
Surgery menu
The Surgery menu is constantly displayed during operation. The menu
4.9 x 300 mm 20% includes the icons (from left to right) for the following menus: Zoom, Focus,
Illumination. Below: User, Options, Video and Balance. The displays below the
Standard zoom, focus, illumination and user icons show the current settings for total
magnification, working distance, brightness level and selected User.
– The display shows the name of the user or application shortened to
10 characters.
– The displayed values are rounded. which are only provided for
information, not for measurement purposes.
Use the control buttons to activate the appropriate selection field and press
the enter button to go to the desired menu.
Zoom menu
You can adjust the settings for the following functions in the Zoom menu:
50% 1.4
Speed:
The zoom speed from minimum to maximum speed from 5% to 100% in
increments of 5%.
Start at:
the zoom start value. Enter the magnification factor at which the zoom system
should start when the system is switched on or is switched to a different user.
The magnification factor can be set and saved within a range between γ =
0.4x and γ = 2.4x in increments of 0.1.
γ is only the magnification factor of the zoom system. The total magnification
provided by the system is also determined by the eyepieces and the working
distance used.
• Activate the function desired (Speed or Start at) and move the indicator
bar using one of the arrow buttons ( , ), until the reading required is
displayed in the field at the top.
• Press the Enter button to save the value set and to return to the Surgery
menu. If you do not press any key, the display will automatically jump back
to the Surgery menu after approx. 30 seconds.
In the Standard User profile, all "Start at" fields are deactivated. No bar is
displayed; it is not possible to adjust any settings.
Version 13.0
Page 104 G-30-1434-en
S7 / OPMI Sensera Device overview
Focus menu
You can adjust the settings for the following functions in the Focus menu:
Zoom Link:
On 50% 300 mm
The Zoom Link function lets you couple the focusing speed to the current
On
magnification. This means that the focusing speed is automatically reduced
when a high magnification is used. Whether the Zoom Link function has been
activated or deactivated is stored in the respective user profile.
Speed:
The focusing speed from minimum to maximum speed from 5% to 100% in
increments of 5%.
Focus
Start at:
On
focus start value. Adjust the working distance at which the focusing system
On should start when you call your user profile. The setting can be adjusted and
saved within a range of 200 to 415 mm in increments of 1 mm.
• Activate the function desired and move the indicator bar using one of the
arrow buttons ( , ) until the reading required is displayed in the field
at the top.
• Press the Enter button to save the value set and to return to the Surgery
menu. If you do not press any button, the display will automatically jump
back to the Surgery menu after approx. 30 seconds and save the values
set.
Version 13.0
G-30-1434-en Page 105
Device overview S7 / OPMI Sensera
Illumination menu
You can adjust the settings for the following functions in the Illumination
menu:
Zoom link
On 50% 50% Intensity:
On the illumination intensity from minimum to maximum brightness from
5% to 100% in increments of 5%. This setting can be adjusted for the xenon
illumination and the halogen illumination.
Start at:
the brightness start value. Use one of the arrow buttons ( , ) to adjust the
illumination intensity in percent. The xenon or halogen lamp will then go on
at this brightness level after activation.
Illumination
The "Start at" field will be active only if "Zoom Link" is in the "Off" position.
On If Zoom Link is On, the "Start at" field will not be active. No bar is displayed;
On
it is not possible to adjust any settings.
Zoom Link:
the automatic light readjustment. When this function is active (ON), the
illumination is adjusted in such a way that the user is provided with
a subjectively constant brightness level in the eyepiece, regardless of the
zoom factor set. Whether the Zoom Link function has been activated or
deactivated is stored in the respective user profile.
• Press the Enter button to save the value set and to return to the "Surgery"
menu. If you do not press any button, the display will automatically jump
back to the Surgery menu after approx. 30 seconds and save the values
set.
Version 13.0
Page 106 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 107
Device overview S7 / OPMI Sensera
Assignment of functions:
Use the ( , ) arrow buttons to select the required function from the table.
Use the ( , ) arrow buttons to select the button required (C or D). After
you have pressed the "Enter" button, a check mark ( ) indicates that the
selected function has been assigned to the selected button.
When you press the "Enter" button again, your selection is saved and you
return to the "Surgery" menu.
Binocular Settings menu
If you select Binocular Setting in the Options menu, the following three setting
options are available:
– Selection of the eyepieces used (10x or 12.5x)
– Selection of the tube used (170 mm or 200 mm)
– Activation of the PD and Diopter Settings submenu
The selection of the tube and eyepieces is required for the correct display of
the system's total magnification in the Surgery menu. Use the arrow buttons
( , ) to select the relevant tube and eyepiece type. Press the "Enter" key to
save the eyepiece type used.
If you select the "PD and Diopter Settings" function, the program branches to
the appropriate subprogram.
Press the Enter button to save the value set and to return to the Surgery menu.
If no entry is made, the display will automatically jump back to the Surgery
menu after approx. 30 seconds.
PD and Diopter Settings"
This menu only has a memory function. It provides sliders which enable each
user to set his/her interpupillary distance and prescription using one of the
arrow buttons ( , ) and to save these values. No motorized setting is made
on the eyepieces or tube. This function is intended to allow OR staff to adjust
the surgical microscope to the surgeon's settings before the procedure and,
if necessary, to retrieve or change the stored values.
– You can set the interpupillary distance in 1 mm increments from 55 mm
to 80 mm.
– When using widefield eyepieces with a magnetic coupling, you can set
your prescription between –8 D and 8D.
If no entry is made, the display will automatically jump back to the Surgery
menu after approx. 30 seconds.
Version 13.0
Page 108 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 109
Device overview S7 / OPMI Sensera
For specific users, the ROI size can be set and saved for each user as required.
Submenu "AF - ROI Position"
This submenu permits you to define the ROI position within the microscope's
field of view.
When you start your system, the ROI is always positioned at the center of the
field of view. The ROI is visualized as a square in the graphic display. Use the
arrow buttons to shift the ROI.
Press the OK button to return to the Autofocus menu. If you do not press any
of the buttons, the display will automatically jump back to the Surgery menu
after approx. 20 seconds.
Menu item "Camera Position":
If you are using an external video camera, the orientation of the video camera
must be known to the SpeedFokus (autofocus) function.
This function is only available in systems equipped with an external video
camera.
The video image must have the same orientation as the image in the main
eyepieces of the surgical microscope.
In addition, SpeedFokus needs the information whether the video camera is
connected on the left or right, i.e. whether the beam splitter projects into the
main beam path from the left or right, to be able to perform corrections in the
right direction during the focusing process.
Submenu "Advanced Settings"
– This submenu is only available in systems equipped with an external video
camera.
– This submenu contains camera-specific settings. These settings only need
to be changed when a different video camera is installed.
– The SpeedFokus (autofocus) option does not function properly if
parameters have not been correctly set.
The "Advanced Settings" submenu contains the following camera-specific
settings:
– Video Focal Length
– Camera Chip Size (size of the CCD array)
Video Focal Length and Camera Chip Size (size of the CCD array):
These settings must conform to the technical specifications of the video
camera used.
Version 13.0
Page 110 G-30-1434-en
S7 / OPMI Sensera Device overview
Balance menu
The Balance menu has been designed as a user prompt and consists of six
images explaining the balancing procedure. You can scroll through the
Balance menu using the arrow buttons ( , ). The OK button has no
function.
Press the enter button or the arrow buttons (, , ) several times to return
to the "Surgery" menu.
Version 13.0
G-30-1434-en Page 111
Device overview S7 / OPMI Sensera
Laser menu
1
The User Selection menu permits you to access the laser selection menu via
the respective Edit button using the arrow button ( ). Here, the working
User 1 NONE distance (focus) of the Varioskop in the surgical microscope is adjusted to the
User 2 TwinER working distance of the TwinER (option) or CO2 laser, and the system is
User 3 TwinER switched to the motorized external focus module (option).
After you have activated Laser menu button (1), a pull-down menu is
automatically opened in the display.
Note: The pull-down menu will be automatically opened if you approach from
the left with the arrow button ( ). If you select the field using the
( , )arrow buttons, the pull-down menu can be opened by pressing the
Enter button.
Using the arrow buttons ( , ) , you can jump to None CO2 Laser
TwinER.
By activating one of the options None, CO2 Laser, TwinER and pressing the
enter button, you can transfer your selection to Laser menu (1).
Select "CO2 Laser" to access the "Focus Laser" menu.
Use the arrow buttons ( , ) to adjust the working distance (focus) of the
micromanipulator used. After you have pressed the OK button and confirmed
the user, the Varioskop is automatically adjusted to the selected working
distance and deactivated. At the same time, the external focus module is
activated if this option has been integrated, see also the notes on Page 153.
It is absolutely vital that the focal planes of the surgical microscope and of the
laser are identical!
Check this after mounting the micromanipulator, before every surgical
procedure (without patient) and before every use of the laser by directing
a trial laser shot at a wooden spatula located in the focal plane. If the
alignment is correct, the laser is focused precisely on the working plane,
and its therapy beam displays its maximum effect. If this is not the case,
you have to correct the working distance in the Laser menu.
Press the Enter button to save the value set and to return to the Surgery menu.
If you do not press any key, the display will automatically jump back to the
Surgery menu after approx. 30 seconds.
Version 13.0
Page 112 G-30-1434-en
S7 / OPMI Sensera Device overview
When you select "TwinER", you access the "Focus TwinER" menu (option).
The "Focus TwinER" menu only contains the following displays:
– Internal Focus is fixed
– external focus active
The Varioskop automatically adjusts to the focal length of the TwinER laser
and switches off. At the same time, the external focus module becomes
operative. When you press the laser micromanipulator focus switch, the
external focus module is automatically activated. Any manual change of
the focus triggers an error message.
This display automatically jumps back to the User menu after about
30 seconds. If you do not press any key, the display will automatically
jump back to the Surgery menu after approx. 30 seconds.
Version 13.0
G-30-1434-en Page 113
Device overview S7 / OPMI Sensera
You can delete the error message by pressing the "Enter" button. Internally,
the system keeps a record of the malfunctions occurred.
If you cannot eliminate the malfunction, you may perhaps be able to continue
working by switching off the suspension system or separating the plug-in
connection to the surgical microscope (see "Failure of main functions",
Page 178).
Acoustic signals
Version 13.0
Page 114 G-30-1434-en
S7 / OPMI Sensera Device overview
Version 13.0
G-30-1434-en Page 115
Device overview S7 / OPMI Sensera
14-function foot control panel, wired (FCP) and wireless (FCP WL)
A retrofit kit for the 14 function foot control panel is available as an option.
The prerequisite for this is that your device is equipped with the foot control
panel connector socket.
Please observe the G-30-1707 Instructions for Use which are contained in the
delivery scope for the 14 function foot control panel.
Version 13.0
Page 116 G-30-1434-en
S7 / OPMI Sensera Device overview
1 2 3
Version 13.0
G-30-1434-en Page 117
Device overview S7 / OPMI Sensera
2 1
Version 13.0
Page 118 G-30-1434-en
S7 / OPMI Sensera Preparations for use
Adjusting the system - S7 ceiling mount with rigid column .......... 140
Setting the suspension arm weight compensation ..................................140
Limiting the suspension arm's downward movement .............................140
Configuring the system - S7 ceiling mount with lifting column .... 142
Set ergonomic working height ................................................................142
Setting the suspension arm weight compensation ..................................144
Limiting the suspension arm's downward movement .............................144
Version 13.0
G-30-1434-en Page 119
Preparations for use S7 / OPMI Sensera
Version 13.0
Page 120 G-30-1434-en
S7 / OPMI Sensera Preparations for use
Medical electrical Any additional equipment you connect to medical electrical devices must
systems demonstrably comply with the applicable standards and directives (e.g.
IEC 60950-1 for data processing equipment). In addition, all configurations
must meet the standard requirements for medical electrical systems (see
IEC 60601-1).Anyone connecting additional equipment to medical electrical
devices is a system configurer and as such is responsible for compliance of
the system with the normative requirements for medical electrical systems.
Please note that local legislation takes priority over the requirements of the
above-mentioned standards. If you have any questions, contact your local
dealer or ZEISS Service.
Version 13.0
G-30-1434-en Page 121
Preparations for use S7 / OPMI Sensera
Version 13.0
Page 122 G-30-1434-en
S7 / OPMI Sensera Preparations for use
1 2 3 4 5 6 7
Version 13.0
G-30-1434-en Page 123
Preparations for use S7 / OPMI Sensera
Establishing connections
You can secure the power plug and the multipoint connector of the switching
component against inadvertent loosening by installing the two cables in the
strain relief device (1).
After you have mounted the strain relief device (1), the cable must have the
following length:
– 320 mm from the strain relief device up to and including the power
connector (9).
– 320 mm from the strain relief device up to the connector (10) of the foot
control panel.
• Form a loop with the cable (3).
• Open the flap (4).
• Feed the cable through the opening (5).
• Close the flap (6).
• Tighten the cable until it encloses the flap (7).
• Check the cable length.
Version 13.0
Page 124 G-30-1434-en
S7 / OPMI Sensera Preparations for use
2 3 4
5 6 7 8
320 mm
10
Version 13.0
G-30-1434-en Page 125
Preparations for use S7 / OPMI Sensera
Version 13.0
Page 126 G-30-1434-en
S7 / OPMI Sensera Preparations for use
1 2 3 4 5
Version 13.0
G-30-1434-en Page 127
Preparations for use S7 / OPMI Sensera
Version 13.0
Page 128 G-30-1434-en
S7 / OPMI Sensera Preparations for use
1 2
Version 13.0
G-30-1434-en Page 129
Preparations for use S7 / OPMI Sensera
Version 13.0
Page 130 G-30-1434-en
S7 / OPMI Sensera Preparations for use
1 2
Version 13.0
G-30-1434-en Page 131
Preparations for use S7 / OPMI Sensera
The image quality obtained with Y/C connections and cables shorter than
10 m is always better than that obtained with BNC connections of the same
length. If cable lengths exceed 10 m, the image quality provided by a Y/C
connection will decrease markedly; in these cases, only BNC connections
should be used.
If possible, you should position the individual units in such a way that you can
use Y/C connections. The Y/C connection provides the best image quality,
in particular for images recorded with MEDIALINK 100.
If this is not possible, position the individual units in such a way that you can
use the Y/C cable at least for the video input port of MEDIALINK 100.
Set-up
1 Standard video monitor
2 Standard video recorder, video printer, video projector
3 MEDIALINK 100
4 CCU connector panel (on back of the system) for the 1Chip HD camera
5 Connector panel of the suspension system
Version 13.0
Page 132 G-30-1434-en
S7 / OPMI Sensera Preparations for use
Devices with the SpeedFokus S7 / OPMI Sensera with 1Chip HD camera option,
(autofocus) option for an external MEDIALINK100 option and video monitor /
camera and the optional video recorder
1Chip HD camera
Video monitor 1
VIDEO IN
Video recorder 2
MEDIALINK 100 3
Y/C
SD - Y/C Remote
Version 13.0
G-30-1434-en Page 133
Preparations for use S7 / OPMI Sensera
Version 13.0
Page 134 G-30-1434-en
S7 / OPMI Sensera Preparations for use
Normally the 1Chip HD camera is located on the right side as viewed from the
main eyepieces. The iris diaphragm for video lenses on the 1Chip HD camera
allows you to re-equip your surgical microscope quickly. The 1Chip HD camera
can be attached to and removed from the beam splitter, and therefore the
output ports of the beam splitter (left/right sides) can also be changed.
After changing sides, you will have to adjust the position of the image.
For further details please refer to the Instructions for Use of the "1Chip HD
camera" (G-30-1946), in the chapters "Using the CCU to configure the 1Chip
HD camera / Configuring the picture settings". These Instructions for Use are
included automatically in systems with the "1Chip HD camera" option.
1 2 3
4
5
Version 13.0
G-30-1434-en Page 135
Preparations for use S7 / OPMI Sensera
Camera cable
DC 12 V - 15 V Permanently
from the suspension installed in
system the suspension
system
CCU
Version 13.0
Page 136 G-30-1434-en
S7 / OPMI Sensera Preparations for use
Version 13.0
G-30-1434-en Page 137
Preparations for use S7 / OPMI Sensera
Version 13.0
Page 138 G-30-1434-en
S7 / OPMI Sensera Preparations for use
• Lower the surgical microscope until it can be focused on the surgical field
(depending on the focal length of the objective lens). At the same time,
ensure that there is sufficient safety clearance to the surgical field.
• Firmly retighten locking lever (2) for the upward/downward movement.
• Lower the surgical microscope to its lowest position again, and check the
distance between the microscope and the surgical field.
– The limit of downward travel is only effective in the lower part of the
suspension arm's vertical movement range (starting from the arm's
horizontal position).
– If you wish to use the suspension arm's full range of vertical movement,
lower the suspension arm as far as it will go and then firmly retighten
locking lever (2).
1 2 3
Version 13.0
G-30-1434-en Page 139
Preparations for use S7 / OPMI Sensera
Version 13.0
Page 140 G-30-1434-en
S7 / OPMI Sensera Preparations for use
• Lower the surgical microscope until it can be focused on the surgical field
(depending on the focal length of the objective lens). At the same time,
ensure that there is sufficient safety clearance to the surgical field.
• Firmly re-tighten the locking lever (2) for the upward/downward
movement.
• Lower the surgical microscope to its lowest position again, and check the
distance between the microscope and the surgical field.
– The limit of downward travel is effective pnly in the lower part of the
suspension arm's vertical movement range (starting from the arm's
horizontal position).
– If you wish to use the suspension arm's full range of vertical movement,
lower the suspension arm as far as it will go and then firmly re-tighten
locking lever (2).
1 2 3
Version 13.0
G-30-1434-en Page 141
Preparations for use S7 / OPMI Sensera
Bring the surgical microscope into a position convenient for you before each
surgical procedure, and set the optimal ergonomic working height of the
suspension system via the lifting column (without the patient!).
As long as you keep the selector switch (2) in the turned position, the lifting
column (1) moves upward or downward, depending on the switch position.
When you release the selector switch, the lifting column stops immediately.
With a maximum actuation time of one minute, the lifting column must be
turned off for at least nine minutes because its motor might overheat
otherwise.
• Before raising or lowering the suspension system, make sure that there is
sufficient clearance from other objects so that any collision is avoided.
The lifting column is used to move the microscope into position for surgery
prior to the surgical procedure.
Do not constantly move the lifting column up and down, since a thermal cut-
out will then automatically deactivate the drive motor. If this occurs, the lifting
column cannot be moved until the motor has cooled down.
Version 13.0
Page 142 G-30-1434-en
S7 / OPMI Sensera Preparations for use
Version 13.0
G-30-1434-en Page 143
Preparations for use S7 / OPMI Sensera
Version 13.0
Page 144 G-30-1434-en
S7 / OPMI Sensera Preparations for use
• Lower the surgical microscope to its lowest position again, and check the
distance between the microscope and the surgical field.
– The limit of downward travel is effective only in the lower part of the
suspension arm's vertical movement range (starting from the arm's
horizontal position).
– If you wish to use the suspension arm's full range of vertical movement,
lower the suspension arm as far as it will go and then firmly re-tighten
locking lever (2).
1 2 3
Version 13.0
G-30-1434-en Page 145
Preparations for use S7 / OPMI Sensera
Version 13.0
Page 146 G-30-1434-en
S7 / OPMI Sensera Preparations for use
Version 13.0
G-30-1434-en Page 147
Preparations for use S7 / OPMI Sensera
Zero position • Press the button (1) until the motor of the balancing system stops.
This decouples the microscope body from its mount.The microscope has
been decoupled from its suspension and has adopted its center-of-gravity
position (zero position) if the white ring with marking (3) does not move
when you turn the microscope.
• Check whether the white ring is really in its zero position. If this is not the
case, turn it manually to the zero position.
• Press key (2).
Allow the microscope to hang in its center-of-gravity position while you
press the button (2).
After approx. 15 seconds, the microscope body has again been coupled with
its suspension. You can check this: when you move the microscope body, the
white ring (3) also moves.
• Continue pressing key (2) and check whether the microscope body
remains balanced in every tilt position. Depending on the accessories
attached, this can take some time (approx. 1 minute or longer). To check
this, lift the microscope body slightly and tilt it into different positions to
see whether the microscope body has been correctly balanced.
• If you have pressed button (2) for too long, you can correct the balance
by pressing button (1).
Version 13.0
Page 148 G-30-1434-en
S7 / OPMI Sensera Preparations for use
2
1
1
1
3
4
Version 13.0
G-30-1434-en Page 149
Preparations for use S7 / OPMI Sensera
Notes on adjustment
– To permit fast setup of the microscope, it is advisable to store the pupillary
distances and prescriptions of the different users so that they can be easily
preset during system preparation.
– Do not wear any bifocal or progressive lenses for the adjustment
procedure and subsequent work with the microscope.
Such lenses make it impossible to set the diopters correctly and result in
unsatisfactory imaging.
– Optimal image clarity in the eyepieces and at the optical port for
documentation equipment (photo, video) can be achieved only by
carefully focusing and precisely adjusting the binocular tube.
Adjusting the • Position the microscope vertically above a flat object, e.g. a piece of paper
interpupillary distance with writing on it.
• Bring the eyepieces on the binocular tube to eye distance, so that the two
eyepiece images (the object and the periphery of the field of view) merge
to produce a single image.
Adjusting the eyepieces The following procedure must be performed separately for each eyepiece in
the order described.
• Set the microscope to the smallest magnification on the microscope.
• Set the diopter adjustment ring on the eyepiece to the zero position.
• Look through the eyepiece and move the microscope to focus the image.
• Set the microscope to the highest magnification and correct the fine
focusing until the image is defined sharply.
• Now set to the smallest magnification once again without changing the
working distance.
• Set the diopter adjustment ring on the eyepiece to its maximum value
(plus direction).
• Look through the eyepiece and slowly turn the diopter adjustment ring to
the minus direction until the image is sharply defined.
Version 13.0
Page 150 G-30-1434-en
S7 / OPMI Sensera Preparations for use
Adjusting the eyecups • Adjust the eyecups so that you can survey the entire field of view.
– Viewing with eyeglasses: screw in the eyecups.
– Viewing without eyeglasses: remove the eyecups.
Version 13.0
G-30-1434-en Page 151
Preparations for use S7 / OPMI Sensera
The device can be equipped with resterilizable products for sterile use.
The asepsis sets available from ZEISS contain caps and handgrips which can
be sterilized in autoclaves. For more detailed information about sterilization,
please refer to the enclosed "Preparation of resterilizable products" manual
for each Asepsis Set.
You can also use disposable sterile drapes to cover the system. When draping
the system, make sure there is enough slack in the drapes to allow for
movement of the microscope carrier and surgical microscope.
Version 13.0
Page 152 G-30-1434-en
S7 / OPMI Sensera Preparations for use
• Please comply with the separate Instructions for Use for your
micromanipulator.
Before using the surgical microscope with a laser micromanipulator, you must
adjust the focal plane of the microscope (Varioskop) to the focal plane of the
micromanipulator.
The focal plane roughly corresponds to the working distance.
In the "User selection" menu, you can access the "Laser selection" menu via
the respective "Edit" button using the arrow button ( ) (see chapter:
Description/Laser menu/CO2laser).
• Use the arrow buttons ( , ) to adjust the working distance (focus) of
the Varioskop in the surgical microscope to the working distance of your
micromanipulator.
If the motorized fine focus (available as an option) has been integrated, the
motorized focusing function of the Varioskop automatically switches to the
fine focus mode. When the focusing buttons on the handgrip are pressed,
the complete surgical microscope including the micromanipulator is moved.
This ensures that the surgical microscope and the micromanipulator remain
focused on the same focal plane. Any manual change of the focus triggers an
error message.
Version 13.0
G-30-1434-en Page 153
Preparations for use S7 / OPMI Sensera
The therapy beam of the laser displays its maximum effect in the focal
plane of the micromanipulator, where it burns a spot on the wooden
spatula. The wooden spatula must be sharply focused in this position.
Only then will the two focal planes coincide.
If this is not the case, refocus and correct the working distance in the Laser
selection menu.
• Repeat the test until full coincidence of the focal planes has been
obtained.
The link-up of a micromanipulator with the surgical microscope via the
adapter plate results in a medical system for which the system manufacturer
must meet the necessary requirements (approval, qualification, laser
protection, etc.). Please comply with the Instructions for Use provided by
the micromanipulator manufacturer and laser manufacturer. Supplementary
information is available from our Service department or from authorized
distributors.
Version 13.0
Page 154 G-30-1434-en
S7 / OPMI Sensera Preparations for use
NOTE Mechanical forces can have undesirable impacts when the device is
being transported!
If you do not comply with the following instructions/precautions, the device
may topple over, injure persons and incur damage:
• Use the grab handle for transporting the device.
• Mind the maximum vertical clearance when passing through doorways.
• Avoid any jolting.
• Be extremely careful when moving over inclined/sloping surfaces.
• Do not park the device on inclined/sloping surfaces.
Please comply with the following instructions when relocating the stand:
• Turn off the device at the power switch.
1 • Remove the power plug from the power outlet.
2 • Move the suspension arm into its transport position (see illustration on the
right).
3
• Wind up the foot control panel cable on one of the cable holders (2),
and hang up the foot control panel on the holder (3).
• Wind up the power cord on the other cable holder (2).
• Release all the locking tabs (4) of the casters.
4 • Hold the device by the grab handle (1) and carefully push it into its new
position. Ensure that the device is standing on level ground.
• Press at least three of the locking tabs (4) and make sure that the stand is
no longer able to move away by itself.
The weight may be underestimated as the suspension system can be moved
very easily. Therefore, move the stand slowly and carefully!
Version 13.0
G-30-1434-en Page 155
Preparations for use S7 / OPMI Sensera
Version 13.0
Page 156 G-30-1434-en
S7 / OPMI Sensera Operation
Operation
Version 13.0
G-30-1434-en Page 157
Operation S7 / OPMI Sensera
Version 13.0
Page 158 G-30-1434-en
S7 / OPMI Sensera Operation
Check that the xenon illumination is on and the green indicator lamp is lit.
If the first lamp has failed and the backup lamp is in use (segment in the
switching knob lights up), make sure to have a second backup lamp
module ready at hand as a precaution.
Version 13.0
G-30-1434-en Page 159
Operation S7 / OPMI Sensera
Version 13.0
Page 160 G-30-1434-en
S7 / OPMI Sensera Operation
Balancing
The suspension arm is properly balanced. If the suspension arm is properly
balanced, the surgical microscope should remain immobile in all positions
of the working range and should not swivel up or down.
Stroke limit
Check that the minimum working distance (height) from the surgical field
has been adjusted.
Stand base
Ensure that at least three of the brake tabs have been pressed and that
these prevent the stand from starting to move by itself.
Accessories
Proper functioning of all other equipment (surgical microscope,
coobservation equipment, video system, etc.) has been checked using the
relevant user's manuals.
1 2 3 4
Version 13.0
G-30-1434-en Page 161
Operation S7 / OPMI Sensera
Version 13.0
Page 162 G-30-1434-en
S7 / OPMI Sensera Operation
Version 13.0
G-30-1434-en Page 163
Operation S7 / OPMI Sensera
Check that the xenon illumination is on and the green indicator lamp is lit.
If the first lamp has failed and the backup lamp is in use (segment in the
switching knob lights up), make sure to have a second backup lamp
module ready at hand as a precaution.
Version 13.0
Page 164 G-30-1434-en
S7 / OPMI Sensera Operation
Version 13.0
G-30-1434-en Page 165
Operation S7 / OPMI Sensera
Balancing
The suspension arm is properly balanced. When the system is properly
balanced, the surgical microscope should remain immobile in all positions
of the working range and should not swivel up or down.
Stroke limit
The minimum working distance (height) from the surgical field has
been set.
1 2 3 4
Version 13.0
Page 166 G-30-1434-en
S7 / OPMI Sensera Operation
Check that the video system and the surgical microscope have been
matched to each other, i.e. when the image is in focus in the microscope
eyepieces, the video image on the monitor must also be in focus (also
applies to the integrated video camera).
The image on the monitor has the same orientation as the image in the
main eyepieces.
Focusing process
Focusing can be activated in accordance with the programming of the
buttons:
using buttons A, B, C and/or D of the handgrips.
using button C and/or D of the foot control panel, if a foot control panel
(option) is provided.
Version 13.0
G-30-1434-en Page 167
Operation S7 / OPMI Sensera
Version 13.0
Page 168 G-30-1434-en
S7 / OPMI Sensera Operation
Check that the xenon illumination is on and the green indicator lamp is lit.
If the first lamp has failed and the backup lamp is in use (segment in the
switching knob lights up), make sure to have a second backup lamp
module ready at hand as a precaution.
Version 13.0
G-30-1434-en Page 169
Operation S7 / OPMI Sensera
Version 13.0
Page 170 G-30-1434-en
S7 / OPMI Sensera Operation
1 2 3 4
Version 13.0
G-30-1434-en Page 171
Operation S7 / OPMI Sensera
Version 13.0
Page 172 G-30-1434-en
S7 / OPMI Sensera Operation
Operating procedure
Version 13.0
G-30-1434-en Page 173
Operation S7 / OPMI Sensera
CAUTION Live video images and recorded video sequences, video clips (cut sequences)
and single images shall not be used for diagnosis purposes. Visualized images
may contain variations in shape, contrast and color.
Version 13.0
Page 174 G-30-1434-en
S7 / OPMI Sensera Operation
Version 13.0
G-30-1434-en Page 175
Operation S7 / OPMI Sensera
Version 13.0
Page 176 G-30-1434-en
S7 / OPMI Sensera What to do in the event of malfunctions
Version 13.0
G-30-1434-en Page 177
What to do in the event of malfunctions S7 / OPMI Sensera
Version 13.0
Page 178 G-30-1434-en
S7 / OPMI Sensera What to do in the event of malfunctions
1
2
3
4
Version 13.0
G-30-1434-en Page 179
What to do in the event of malfunctions S7 / OPMI Sensera
Version 13.0
Page 180 G-30-1434-en
S7 / OPMI Sensera What to do in the event of malfunctions
1
2
3
4
Version 13.0
G-30-1434-en Page 181
What to do in the event of malfunctions S7 / OPMI Sensera
Version 13.0
Page 182 G-30-1434-en
S7 / OPMI Sensera What to do in the event of malfunctions
Version 13.0
G-30-1434-en Page 183
What to do in the event of malfunctions S7 / OPMI Sensera
Version 13.0
Page 184 G-30-1434-en
S7 / OPMI Sensera What to do in the event of malfunctions
• Turn knob (1) through 180° until it snaps in. This swings the second xenon
lamp (backup lamp) into the beam path.
• Push the lamp module all the way back into the lamp housing.
Version 13.0
G-30-1434-en Page 185
What to do in the event of malfunctions S7 / OPMI Sensera
• Reset the service hour counter to "0" after replacing the lamp. Use
a pointed object and press it into the recess of reset button (1).
• Switch the suspension system back on again at the power switch.
If the first lamp has failed and the backup lamp is in use (segment in the
knob is lit), make sure to have a backup lamp module ready at hand as
a precaution.
Version 13.0
Page 186 G-30-1434-en
S7 / OPMI Sensera What to do in the event of malfunctions
Version 13.0
G-30-1434-en Page 187
What to do in the event of malfunctions S7 / OPMI Sensera
OPMI Sensera
Version 13.0
Page 188 G-30-1434-en
S7 / OPMI Sensera What to do in the event of malfunctions
S7 suspension system
Version 13.0
G-30-1434-en Page 189
What to do in the event of malfunctions S7 / OPMI Sensera
Version 13.0
Page 190 G-30-1434-en
S7 / OPMI Sensera What to do in the event of malfunctions
Version 13.0
G-30-1434-en Page 191
What to do in the event of malfunctions S7 / OPMI Sensera
Version 13.0
Page 192 G-30-1434-en
S7 / OPMI Sensera What to do in the event of malfunctions
Version 13.0
G-30-1434-en Page 193
What to do in the event of malfunctions S7 / OPMI Sensera
Version 13.0
Page 194 G-30-1434-en
S7 / OPMI Sensera Care and Maintenance
Version 13.0
G-30-1434-en Page 195
Care and Maintenance S7 / OPMI Sensera
Cleaning
Version 13.0
Page 196 G-30-1434-en
S7 / OPMI Sensera Care and Maintenance
Sterilization
Asepsis sets
The asepsis sets available from ZEISS contain rubber caps and hand grips
that can be sterilized in autoclaves. For more detailed information about
sterilization, please refer to the enclosed "Preparation of resterilizable
products" manual for each Asepsis Set.
• Loosely attach the sterile drapes so that there is sufficient space for the
microscope supporting stand and the surgical microscope to move.
Version 13.0
G-30-1434-en Page 197
Care and Maintenance S7 / OPMI Sensera
Disinfection
Version 13.0
Page 198 G-30-1434-en
S7 / OPMI Sensera Care and Maintenance
System maintenance
Maintenance intervals
Version 13.0
G-30-1434-en Page 199
Care and Maintenance S7 / OPMI Sensera
Safety check
Version 13.0
Page 200 G-30-1434-en
S7 / OPMI Sensera Care and Maintenance
Version 13.0
G-30-1434-en Page 201
Care and Maintenance S7 / OPMI Sensera
Version 13.0
Page 202 G-30-1434-en
S7 / OPMI Sensera Care and Maintenance
5 6 7
Version 13.0
G-30-1434-en Page 203
Care and Maintenance S7 / OPMI Sensera
Version 13.0
Page 204 G-30-1434-en
S7 / OPMI Sensera Care and Maintenance
– Check the function of the xenon lamp and backup xenon lamp.
– Pack the old lamp module (5) in the transport packaging of the new
lamp module. Fill in the enclosed return card and send the old lamp
module to the nearest ZEISS service representative.
– Use only the original transport packaging (2), as it also provides
explosion protection should xenon lamps be defective.
Version 13.0
G-30-1434-en Page 205
Care and Maintenance S7 / OPMI Sensera
Spare parts
Asepsis sets
Version 13.0
Page 206 G-30-1434-en
S7 / OPMI Sensera Care and Maintenance
Disposal
Version 13.0
G-30-1434-en Page 207
Care and Maintenance S7 / OPMI Sensera
Version 13.0
Page 208 G-30-1434-en
S7 / OPMI Sensera Device data
Device data
Version 13.0
G-30-1434-en Page 209
Device data S7 / OPMI Sensera
Specifications
Surgical microscope
Component Properties
Magnification Motorized zoom system, zoom ratio 1:6,
magnification factor γ = 0.4x - 2.4x.
Total magnification 2.5x - 14.9x at a working distance of 200 mm
(10x eyepiece) 1.5x - 8.9x at a working distance of 415 mm
Total magnification 3.1x - 18.6x at a working distance of 200 mm
(12.5x eyepiece) 1.8x - 11.1x at a working distance of 415 mm
Fields of view 14 mm - 83 mm at a working distance of 200 mm
(with 10x eyepiece) 23 mm - 140 mm at a working distance of 415 mm
Version 13.0
Page 210 G-30-1434-en
S7 / OPMI Sensera Device data
Component Properties
Fields of view 12 mm - 71 mm at a working distance of 200 mm
(with 12.5x eyepiece) 20 mm - 119 mm at a working distance of 415 mm
Focus Continuous motorized focusing via the integrated
Varioskop focusing range: 200 - 415 mm
Note: the values shown on the display differ from
the actual values. They are rounded values and serve
for information only.
Focal lengths f At a working distance of 200 mm: f = 278 mm,
At a working distance of 300 mm: f = 366 mm,
At a working distance of 415 mm: f = 466 mm
Light field diameter At a working distance of 200 mm
min. 11 mm - max. 85 mm
At a working distance of 415 mm
min. 19 mm - max. 142 mm
Tubes / Eyepieces Binocular 180° tiltable tube, f = 170 mm.
10x or 12.5x widefield eyepieces with magnetic
coupling.
Weight Approx. 9.2 kg (without tube and eyepieces)
Available as options
Version 13.0
G-30-1434-en Page 211
Device data S7 / OPMI Sensera
Stands/mounts
Component Properties
Power supply Only connect the floor stand to wall outlets which
are provided with a properly connected protective
ground conductor.
Rated Voltage (115 V): 100 V - 125 V
(230 V): 220 V - 240 V
Power consumption Floor stand:
(115 V): max. 850 VA
(230 V): max.1250 VA
Remote connection socket
Ceiling mount with rigid or lifiting column:
View of connector side
(115 V): max. 450 VA
2 (230 V): max. 450 VA
Rated Frequency 50 Hz - 60 Hz
3 1 Fuses Automatic circuit breaker
Electrical – Power outlet (floor stand only)
outlets (115 V): max. 400 VA
1 (230 V): max. 800 VA
2 – Surgical microscope
– Remote socket for an external signal with max.
3
24 V / 0.5 A (floor stand only)
The device is designed for continuous operation.
Fiber-optic
illumination system 2 halogen reflector lamps 12 V 100 W in
quick-change module for light guide, with
GG 475 and KK 40 filters.
Fully automatic lamp replacement.
Version 13.0
Page 212 G-30-1434-en
S7 / OPMI Sensera Device data
Version 13.0
G-30-1434-en Page 213
Device data S7 / OPMI Sensera
Floor stand S7
Mechanics
Component Properties
Support arm Length: 450 mm
Swivel angle: 360°
Suspension arm Length: 865 mm
Swivel angle: 305°
Lifting stroke: -320 mm / +400 mm
coupling Swivel angle: 350°
Stand height 1730 mm
Base dimensions 640 x 640 mm
Maximum load for the 14 kg (complete microscope equipment,
suspension arm including accessories)
Weight Approx. 177 kg (including microscope)
Maximum total weight 190 kg (with installed safe workload)
305°
Version 13.0
Page 214 G-30-1434-en
S7 / OPMI Sensera Device data
S7 ceiling mount
Mechanics
Component Properties
Support arm Length: 450 mm
Swivel angle: 296°
Suspension arm Length: 865 mm
Swivel angle: 296°
Lifting stroke: -320 mm / +400 mm
Coupling Swivel angle: 350°
Admissible max. load 14 kg (complete microscope equipment, including
on suspension arm accessories)
Weight (including Approx. 80 kg
microscope)
Version 13.0
G-30-1434-en Page 215
Device data S7 / OPMI Sensera
Version 13.0
Page 216 G-30-1434-en
S7 / OPMI Sensera Device data
Version 13.0
G-30-1434-en Page 217
Device data S7 / OPMI Sensera
Ordering data
Operate the device using only accessories that were included in the scope of
supply and have been recommended by ZEISS. Your country-specific contact
persons for orders are listed on the internet at the following website:
www.meditec.zeiss.com
Version 13.0
Page 218 G-30-1434-en
S7 / OPMI Sensera Device data
Version 13.0
G-30-1434-en Page 219
Device data S7 / OPMI Sensera
Version 13.0
Page 220 G-30-1434-en
S7 / OPMI Sensera Device data
Version 13.0
G-30-1434-en Page 221
Device data S7 / OPMI Sensera
Version 13.0
Page 222 G-30-1434-en
S7 / OPMI Sensera Device data
Version 13.0
G-30-1434-en Page 223
Device data S7 / OPMI Sensera
Ordering data for 1Chip HD camera option for the OPMI Sensera
Version 13.0
Page 224 G-30-1434-en
S7 / OPMI Sensera Device data
Version 13.0
G-30-1434-en Page 225
Device data S7 / OPMI Sensera
Version 13.0
Page 226 G-30-1434-en
S7 / OPMI Sensera Device data
Version 13.0
G-30-1434-en Page 227
Device data S7 / OPMI Sensera
Video accessories
For external video accessories for this surgical microscope, please refer to the
separate product overview G-30-1888.
Version 13.0
Page 228 G-30-1434-en
S7 / OPMI Sensera Device data
Compliance
Version 13.0
G-30-1434-en Page 229
Device data S7 / OPMI Sensera
Environmental conditions
For operation
Version 13.0
Page 230 G-30-1434-en
S7 / OPMI Sensera Device data
Version 13.0
G-30-1434-en Page 231
Device data S7 / OPMI Sensera
Version 13.0
Page 232 G-30-1434-en
S7 / OPMI Sensera Device data
Version 13.0
G-30-1434-en Page 233
Device data S7 / OPMI Sensera
Version 13.0
Page 234 G-30-1434-en
S7 / OPMI Sensera Device data
Version 13.0
G-30-1434-en Page 235
Device data S7 / OPMI Sensera
System combinations
The device S7 / OPMI Sensera can be extended with the accessories offered
by ZEISS (see ordering data/accessories) to form an already tested system.
This device combination has been approved by ZEISS.
As per IEC60601-1, section 16.2, this system's power outlet is a multiple
electrical outlet intended for use with an ME system.
Connecting electrical devices not approved by ZEISS can lead to a reduced
level of safety on the ME system.
• When configuring your ME system (using the power outlet), you must
comply with the requirements of IEC60601-1, section 16, as well as the
following measures:
– Do not place multiple electrical outlets on the floor.
– Do not connect any additional multiple electrical outlets.
– Only connect compatible components to the system.
– Do not exceed the max permissible electrical load of multiple electrical
outlets.
– Use multiple electrical outlets only for components of the system.
Medical electrical Any additional equipment you connect to medical electrical devices must
systems demonstrably comply with the applicable standards and directives (e.g.
IEC 60950-1 for data processing equipment). In addition, all configurations
must meet the standard requirements for medical electrical systems (see
IEC 60601-1).Anyone connecting additional equipment to medical electrical
devices is a system configurer and as such is responsible for compliance of
the system with the normative requirements for medical electrical systems.
Please note that local legislation takes priority over the requirements of the
above-mentioned standards. If you have any questions, please contact your
local dealer or ZEISS Service.
Version 13.0
Page 236 G-30-1434-en
S7 / OPMI Sensera Keyword index
Keyword index
Numerics
Version 13.0
G-30-1434-en Page 237
Keyword index S7 / OPMI Sensera
Version 13.0
Page 238 G-30-1434-en
S7 / OPMI Sensera Keyword index
Version 13.0
G-30-1434-en Page 239
Keyword index S7 / OPMI Sensera
Version 13.0
Page 240 G-30-1434-en
S7 / OPMI Sensera Keyword index
Photo .....................................................................................................107
Potential equalization ...............................................................................11
Potential equalization pin .............................................................72, 80, 88
Power output socket ................................................................................72
Power switch ..........................................................................................100
Preparations for use ...............................................................................119
Prior to the very first use ...........................................................................13
Protective ground connector ....................................................................13
Version 13.0
G-30-1434-en Page 241
Keyword index S7 / OPMI Sensera
Version 13.0
Page 242 G-30-1434-en
S7 / OPMI Sensera Keyword index
Weight ...................................................................................................211
What to do in the event of malfunctions ................................................177
Widefield eyepieces with magnetic coupling .............................................64
Version 13.0
G-30-1434-en Page 243
Keyword index S7 / OPMI Sensera
Version 13.0
Page 244 G-30-1434-en
S7 / OPMI Sensera
Version 13.0
G-30-1434-en Page 245
Carl Zeiss Meditec AG
Goeschwitzer Strasse 51-52 Fax: + 49 (0)7364 - 20 4823
07745 Jena E-mail: surgical@meditec.zeiss.com
Germany Internet: www.meditec.zeiss.com