You are on page 1of 278

ZEISS

S100 / OPMI® pico

Instructions for Use


G-30-1781-en
Version 7.0
2015-09-09
S100 / OPMI pico

About this manual These Instructions for Use are part of the delivery package.
• Carefully read them before using the device.
• Keep them at the site of use of the device.
• Store them for the entire service life of the device.
• Pass them on to every subsequent owner or user of the device.

Orientation aids – The chapter overview at the beginning of these Instructions for Use pro-
vides a summary of all subjects.
– The contents of each chapter are specified in detail at the beginning of
each chapter.
– A list of abbreviations, key words and technical terms in the annex facili-
tates the search for specific terms.

Area of application These Instructions for Use apply to the S100 / OPMI® pico with the following
identification:

Configuration REF. no.


®
S100 / OPMI pico (floor stand) 000000-1403-542
S100 / OPMI® pico (wall mount) 000000-1403-544
S100 / OPMI® pico (ceiling mount) 000000-1403-572
S100 / OPMI® pico (Centro suspension system) 000000-1403-573
S100 / OPMI® pico (support unit) 000000-1420-191
S100 / OPMI® pico 000000-1095-094
(support unit for non-Zeiss illumination)
S100 / OPMI® pico (Atmos support unit) 000000-1080-349

Trademarks OPMI® is a registered trademark of Carl Zeiss.

Version 7.0
Page 2 G-30-1781-en
S100 / OPMI pico

Manufacturer Carl Zeiss Meditec AG


Goeschwitzer Strasse 51-52 Fax: + 49 (0) 7364 - 20 4823
07745 Jena E-Mail: surgical@meditec.zeiss.com
Germany Internet: www.zeiss.com/med

Subject to changes in design and scope of delivery and as a result of ongoing


technical development. Printed in Germany.

Copyright © Carl Zeiss Meditec AG 2015


All rights reserved. Reprints and reproductions, including excerpts, are not
permissible. However, these Instructions for Use may be digitized and
archived for internal use. Access to this data by third parties is not permissible
and must be prevented.

Version 7.0
G-30-1781-en Page 3
S100 / OPMI pico

Version 7.0
Page 4 G-30-1781-en
S100 / OPMI pico

Chapter overview

Chapter: Safety Measures 7

Chapter: System Overview 51

Chapter: Preparations for use 89

Chapter: Operation 127

Chapter: What to do in the event of malfunctions 209

Chapter: Care and Maintenance 221

Chapter: System Data 233

Chapter: Indexes 261

Version 7.0
G-30-1781-en Page 5
S100 / OPMI pico

Version 7.0
Page 6 G-30-1781-en
S100 / OPMI pico Safety Measures

Safety Measures

Hazard symbols ...........................................................................................8


Information symbols....................................................................................8

Target group .....................................................................................9

Field of application ............................................................................ 9


Intended use ...............................................................................................9
Normal use..................................................................................................9

Notes for the operator..................................................................... 10

Requirements for operation............................................................. 13


Prior to the very first use............................................................................13
Before every use ........................................................................................14
During use.................................................................................................15
After every use ..........................................................................................16

Liability and warranty ................................................................................16

Possible burn injuries due to a high illumination intensity............... 17

Safety devices..................................................................................19

Symbols and labels on the device ....................................................22


Symbols and labels on the microscope.......................................................22
Symbols and labels on the S100 ................................................................24
Symbols and labels on the light sources.....................................................40
Symbols and video control units (optional).................................................48
Key to symbols
We would like to inform you about safety aspects which must be observed
when handling this device. This chapter contains a summary of the most im-
portant information concerning matters relevant to instrument safety.

Version 7.0
G-30-1781-en Page 7
Safety Measures S100 / OPMI pico

Hazard symbols

The following safety information has been incorporated into the Instructions
for Use. Please note this information and be particularly careful in these cases.

WARNING Warning label, which may refer to fatal injuries or severe injuries if pre-
cautions are not followed.
CAUTION Indicating a hazard, which may lead to moderately severe injuries
if risks are not avoided.
NOTE Warning label, which may refer to minor injuries and property damages
if precautions are not followed.

Information symbols

The following information symbols are used in these Instructions for Use:
– Listing
 Requirements for an action
• Prompt for action
 Result of an action
Additional information and hints

Version 7.0
Page 8 G-30-1781-en
S100 / OPMI pico Safety Measures

Target group

These Instructions for Use are intended for physicians, nurses and other med-
ical and technical staff who prepare, operate or maintain the device after ap-
propriate training. It is the duty of the equipment owner/operator to train and
brief all the operating staff.
Extended service tasks are not part of these Instructions for Use. They will be
performed by staff specially trained for this purpose by ZEISS.

Field of application

Intended use

The S100 / OPMI pico is a surgical microscope designed to illuminate and


magnify the surgical area and to support the visualization of surgical proce-
dures.

Normal use

The S100 / OPMI pico is a microscope that is operated manually on a sitting


or lying patient.

Provision: The S100 / OPMI offers in conjunction with different suspension systems (e.g.
floor stand, wall mount, ceiling mount) and flexible carrier arms, an ergo-
nomic working position for the treatment environment or surgery. The system
can be securely moved into the working position with the hand grips without
any problems.

Version 7.0
G-30-1781-en Page 9
Safety Measures S100 / OPMI pico

Treatment/surgery: The microscope is positioned and roughly focused by the physician or the sur-
geon. The most frequently used functions as the setting of focus, magnifica-
tion and light intensity are manually controlled by the physician or the sur-
geon during treatment or surgery.
A video player is connected to the system for educational purposes or to vis-
ualize the progress of the treatment/surgery.

After the treatment / surgery The physician or the surgeon swings the surgical microscope from the treat-
ment or surgical area into its standby position.
The S100 / OPMI pico can be cleaned and disinfected.

Disposal: For the disposal of electrical and electronic devices, please observe the rele-
vant national legislation and other regulations applicable in your country.

CAUTION Injury to the patient's eye!


• The S100 / OPMI pico must not be used for ophthalmic procedures!
• Make sure that no light enters the patient's eyes.

CAUTION Live video images and recorded video sequences, video clips (cut sequences)
and single images must not be used for diagnosis purposes. Visualized im-
ages may contain variations in shape, contrast and color.

Notes for the operator

The correct use of the device is absolutely vital for safe operation. Therefore,
please thoroughly familiarize yourself with the content of these instructions
for use before starting up the system. Please also observe the instructions for
use of any additional equipment. You can obtain additional information from
ZEISS Service.
• Only operate the device within the scope of its intended use.
• Only personnel who have undergone appropriate training and instruction
are allowed to use this device. It is the responsibility of the customer or
the company operating the device to train and instruct all operating per-
sonnel.
• Please keep these Instructions for Use where they are easily accessible at
all times for the persons operating the system.

Version 7.0
Page 10 G-30-1781-en
S100 / OPMI pico Safety Measures

• Please take note of the planning guidelines (brochure M-30-1385) for


planning the S100 ceiling mount or the S100 wall mount.
• Note the symbols and labels attached to the device (see Page 22)!
• Use only original accessories approved by ZEISS.

Legal regulations • Observe the legal regulations for accident prevention and occupational
health and safety applicable in the country concerned.
• This system must not be modified without the manufacturer's approval. If
the system is modified, suitable inspections and testing must be complet-
ed to ensure that it can still be used safely.
The manufacturer is not liable for damage caused by unauthorized per-
sons tampering with the system. Furthermore, this will forfeit any rights
to claim under warranty.
• Modifications and repairs on this device or any equipment operated to-
gether with this device may only be performed by ZEISS SERVICE.
• If required by the regulations and directives applicable in the country of
use, connect the system to a special emergency backup power supply.

Ambient requirements • Do not operate the system contained in the delivery package
– in explosion-risk areas,
– if inflammable anesthetics or volatile solvents such as alcohol, benzine
or similar chemicals are present at a distance of less than 25 cm.
• Do not station or use the instrument in damp rooms. Do not expose the
instrument to water splashes, dripping water or sprayed water.
• Do not place any fluid-filled containers on top of the instrument. Make
sure that no fluids can seep into the instrument.
• Make sure that the installation conditions and the use of the device meet
surgical requirements: See Ambient conditions on Page 257

Electrical safety • Always switch off the system before disconnecting it from / connecting it
to line power, for cleaning its surface or if it will not be used for a pro-
longed period of time.
• Switch off the unit at the power switch if you notice any smoke, sparks or
unusual noise. Do not use the unit until it has been repaired by our service
team.
• Do not use multiple sockets!
• Never open the device! The device contains freely accessible live compo-
nents. If you remove the housing, you run the risk of electric shock.

Version 7.0
G-30-1781-en Page 11
Safety Measures S100 / OPMI pico

• Do not touch the system if your body is electrostatically charged and the
system is not grounded.
• A potential equalization connector (see IEC 60601-1:2005) is provided on
the connector panel. This connector is optional for halogen and Xenon il-
lumination. This connector can be used for placing other active devices at
the same ground potential or for redundant grounding to protective
earth.
• Please note the information on EMC (electromagnetic compatibility) in the
chapter "System Data" on Page 245.

Transport & Service • The device may only be moved over long distances (e.g. relocation, return
for repair etc.) in its original packaging or in special return packaging.
Please contact your dealer or ZEISS Service.
• This system is a high-grade technological product. To ensure optimum
performance and safe working order, we recommend having it checked
by our service representative as part of regular scheduled maintenance.
• In order to prevent any impairment of the device's safety due to age,
wear, etc., the user must ensure that the device is subjected to the neces-
sary safety checks (see section "Care and Maintenance").

Optics and light • Never look into a light source or the sun with the binocular tube, the ob-
jective lens or an eyepiece.
• Never directly look into the light source, e.g. a microscope object lens or
a light guide to avoid phototoxic damages to the eyes.
• Adjust the brightness at the stand. Start with the lowest brightness and
slowly increase it to the required but not more than the allowed bright-
ness.
• With increasing age of the light source, the actual illumination intensity
delivered at the respective setting decreases (this is a normal property of
the system).

Version 7.0
Page 12 G-30-1781-en
S100 / OPMI pico Safety Measures

Requirements for operation

WARNING The device must not be modified without any approval!


This device must not be modified without the manufacturer's approval.
If the device is modified, appropriate examinations and tests must be per-
formed to guarantee the further secure use of the device.
The manufacturer is not liable for damages caused by an unauthorized use
of the device. Furthermore, this will invalidate any rights to make a claim
under the warranty.

Prior to the very first use

A ZEISS Service representative or an expert authorized by ZEISS will install the


system. Please make sure that the following requirements continue to be met
for further operation:
 If you determine that one of the symbols or the lettering (effective
Page 22) is missing or illegible on your device, please contact ZEISS Service
or the ZEISS authorized partner.
 The connecting components have been properly connected. The screw
connections have been firmly tightened.
 All cables and plugs are in perfect condition.
 Determine the supply voltage at the place of installation and set the volt-
age of the device accordingly.
 Check whether the device fuse used is the correct one as specified on the
fuse label.
More specific information about the floor stand:
 The main plug must only be connected to a socket with
a faultless protective ground conductor.
 The device is connected with the power cord supplied.
Additional information about wall and ceiling mount:
 The device may only be connected to a power supply network which has
a faultless protective ground conductor.

Version 7.0
G-30-1781-en Page 13
Safety Measures S100 / OPMI pico

Before every use

• When adding accessories and/or components, please ensure the permis-


sible total weight of the device is not exceeded (see label "Maximum load"
or section "Technical Data").
• When connecting accessories make sure that sufficient free space is pro-
vided and that the surgical microscope does not touch the patient.
• Before every use, perform the balancing procedure to enable the surgical
microscope to maintain its balance in all positions of the working range.
• Check the surgical microscope for sufficient freedom of movement. The
device itself or accessories may be damaged if conflicting with each other.
• To prevent accessories from falling down, check before every use that ac-
cessories are securely locked in position and securing screws are firmly
tightened.
• Always run cables in a manner that will not impede the user's movements.
• If you operate the device in a sterile area, make sure that you use the cor-
responding sterile accessories for the device.
• Make sure that the operating staff is briefed about sterile use of the sys-
tem.
• Never cover any ventilation openings, as the light source of the system
switches off in the event of overheating.
• Do not exert any force when plugging in electrical plug-in connections
(connectors, jacks). If connection is not readily possible, check whether
the plug fits the socket. If you detect any damage to the plug-in connec-
tion, arrange for the damage to be repaired by ZEISS Service.
• Please refer to the functional test in chapter "Operation" (Page 129).
More specific information about the floor stand:
• Using the brake tabs on the base, secure the stand in position. Make sure
that the stand is stable and cannot roll away.

Version 7.0
Page 14 G-30-1781-en
S100 / OPMI pico Safety Measures

During use

• Reduce the brightness and duration of light exposure of the surgical area
to the absolute minimum level required to avoid burn injuries.
• Never directly look into the light source, e.g. a microscope object lens or
a light guide to avoid phototoxic damages to the eyes.
• Never leave a device unattended with the light source still switched on.
• Not approved or defective accessories for the device may lead to elevated
leakage currents. Never connect the non approved or defective accesso-
ries to the device.
• Never touch video interfaces while in contact with the patient.
• If you need to change the lamp shortly after it has failed, the lamp will still
be very hot. Use heat-resistant protective gloves when replacing the lamp
to prevent burns.
• If the xenon lamp is used beyond its maximum service life of 500 hours
(see chapter "Maintenance intervals" on Page 222), sudden failure may
occur and interrupt operation. Change to the backup lamp or replace the
lamp module in a timely manner.
• If the halogen lamp is used beyond its maximum service life of 50 hours,
sudden failure may occur and interrupt operation. Replace the halogen
lamp in a timely manner (see chapter "Care and Maintenance" on Chang-
ing the halogen lamp on page 1).
• Do not pull at the power cord or any other connecting cables.
• If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions attach a sign to the system stat-
ing it is out of order and inform ZEISS Service.
• Never operate the system unattended.
• Always switch off the system
- before disconnecting it from / connecting it to line power
- if it will not be used for a prolonged period of time
- or for cleaning purposes.

Version 7.0
G-30-1781-en Page 15
Safety Measures S100 / OPMI pico

After every use

• Always use the master switch to turn off the device if it is not in use.
• Insufficient, incorrect or wrong cleaning or disinfection contrary to the di-
rections in these Instructions for Use may expose the patient or medical
staff to a considerable risk of infection. For more information, please see
"Maintenance of the device" in the chapter "Care and Maintenance".
• If the light guide is not connected at both ends, there may be a risk of fire
or burning when the light source is activated.
• Risk of crushing - mind your fingers!
Fingers may be crushed in the areas marked with the "Risk of crushing"
label.
Do not touch these areas while the system is being moved.

Liability and warranty

Warranty and liability depend on the applicable contractual stipulations.

Loss of warranty!
The manufacturer is not liable for damages caused by an unauthorized use
of the device. Furthermore, this will this will invalidate any rights to make a
claim under the warranty.

Version 7.0
Page 16 G-30-1781-en
S100 / OPMI pico Safety Measures

Possible burn injuries due to a high illumination intensity

CAUTION Danger of burnings caused by high light intensity!


In case of a misuse of the light source, a too high illumination intensity may
cause tertiary burns.
• Please read and note the following points.

General
This device is optionally equipped with a powerful xenon light source.
Various factors contribute to the risk of burn injuries:
Device-related factors:
– With increasing service life of the lamp, the actual light intensity delivered
at the respective setting decreases. In case of a bulb exchange, illumina-
tion increases to the originally high value.
Surgery-related factors:
– The selected intensity of the light source is a major factor for the risk of
injury. It should always be set to the minimum required for the surgical
procedure to be performed.
– A long surgical procedure increases the risk of injury, in particular if a
standard procedure takes considerably longer than usual.
– Injuries in the peripheral area may be avoided by covering the respective
parts with wet, sterile gauze.
The gauze must be moistened at regular intervals to avoid that the area
dries out or heats up.
– Certain preparations of the surgical field, local vasoconstrictive medica-
tions and incision drapes may also result in a higher risk of injury (drapes
may heat up to varying degrees depending on their color and moisture
content).
– You should also take into account that some areas of the body may be
more sensitive than others.
Patient-related factors:
– The general condition of a patient's health may contribute to the risk of
injury.
– The skin type may also play a major role for the risk of injury.

Version 7.0
G-30-1781-en Page 17
Safety Measures S100 / OPMI pico

– Certain medications also affect the sensitivity to light.


– The interaction of heat and antimicrobial substances in incision foils may
lead to an increased reaction of the patient to these substances.

Recommendations
Due to the large number of different factors involved and the lack of scientific
publications on this topic, Carl Zeiss cannot provide guidance on acceptable
illumination intensities and exposure durations. However, this system has sev-
eral features that can help the user to reduce the risk of burns.
– The start value of the illumination intensity can be set to a low value.
– The risk of burns can be reduced by constantly by irrigating the illuminated
surgical field and by keeping it moist.
– If used, drapes should also be re-moistened at regular intervals in order to
prevent heat from accumulating underneath the drape.
– Switch off the light when the microscope is not used, and make sure that
it is not pointed at unprotected bare skin.
Please keep in mind that the most published cases with burns refer to the skin
around the incision. The most important measure to avoid burns is covering
the peripheral area with wet, sterile gauze. The area of the incision should be
constantly moistened.

Final remark
ZEISS recommends that the lighting of the surgical field be reduced as far as
is possible for the safety of the patient and for a good microscopic image.
Please refer to the warnings and precautions in this case.

Version 7.0
Page 18 G-30-1781-en
S100 / OPMI pico Safety Measures

Safety devices

Halogen light source


1 Flap
The flap is used as mechanic indicator of the operating status of the hal-
ogen light source.
– When the flap is closed, the main lamp is operative.
– When the flap is open, the main lamp has failed. The backup lamp is
on.
2 Switching to the backup lamp
The lamp housing contains a backup lamp. If the first lamp fails, you can
switch on the backup lamp by pressing this button.
• Replace the defective lamp at the first opportunity (see Page 224).

Fig. 1: Safety system halogen


light source

Version 7.0
G-30-1781-en Page 19
Safety Measures S100 / OPMI pico

Xenon light source

CAUTION Limited service life of the xenon lamp!


If used longer than the indicated service life of 500 hours the xenon lamp
may fail or burst.
Dropping fragments may contaminate the sterile area. Risk of infection.
• Change the lamp module on time.

– UV/IR heat protection filter


The filter protects from UV and IR radiation.
– Temperature monitoring
A sensor constantly monitors the lamp temperature.
1 Switching to the backup lamp
The xenon light source has a lamp module containing two xenon lamps.
The second lamp is used as backup lamp. Manually swivel in the backup
lamp (Page 218) if it is used beyond its maximum service life of 500h (see
chapter "Maintenance intervals" on Page 222).
If the red segment (2) in the button (1) is lit, the backup lamp is operated.
Keep the lamp module ready if you operate the backup lamp.
If the backup lamp is operated, replace the lamp module in a timely
manner (see Page 226) before it is used beyond its maximum service life
of 500h (see chapter "Maintenance intervals on Page 222).

Fig. 2: Safety system xenon


light source

Version 7.0
Page 20 G-30-1781-en
S100 / OPMI pico Safety Measures

LED light source

CAUTION LED equipment, laser class 2


The LED light source is a class 2 LED device according to IEC 60825-1:2001.
According to EN 62471:2009 this light source is to be classified as belonging
to the free group.

NOTE Varying color impression


The color temperature of the lamps differs. This can result in a changing color
impression of the treatment / surgical field.
– Temperature monitoring
A sensor constantly monitors the LED temperature.
1 LED error indicator
If the amber LED is illuminated, the LED light source is in a limited oper-
ating mode.
In this case, the light intensity drops to 50%.
• Please contact ZEISS Service.

Fig. 3: Safety equipment:LED


light source

Version 7.0
G-30-1781-en Page 21
Safety Measures S100 / OPMI pico

Symbols and labels on the device

CAUTION Observe the warning labels and signs!


• If you notice that any of these labels/signs are missing on your device or
have become illegible, please contact the ZEISS service department.

Symbols and labels on the microscope

1 Video signal output

2 "Instructions for Use" label


Observe the Instructions for Use or accompanying documents.

OPMI pico 3 Designation of the microscope

4 Friction symbol
Explanation of the braking effect achieved by turning the friction adjust-
ment knob.
- Left turn: decrease braking effect
- Right turn: increase braking effect
5 Marking the magnification changer
Marks the magnification setting.

1
1.6 6 Labeling magnification changer
Displays the magnification value
2.5
0.4

0.6
1

Version 7.0
Page 22 G-30-1781-en
S100 / OPMI pico Safety Measures

Fig. 4: Symbols and labels on


Surgical microscope

5 6

5
1 6
2
3

Fig. 5: Symbols and labels on


surgical microscope with
integrated HD video camera

11

5 6

1
5
6
2
3

Version 7.0
G-30-1781-en Page 23
Safety Measures S100 / OPMI pico

Symbols and labels on the S100

Symbols and labels on support and suspension arm


1 Friction symbol
Explanation of the braking effect that is achieved by turning the friction
control.
- Turn left: Reducing the braking effect
- Turn right: Increasing the braking effect
2 "Risk of crushing" warning label
Fingers may be crushed. Do not touch this area while moving the surgical
microscope or parts of it.

3 Brightness setting
If the light source is turned on, the brightness can be continuously set by
turning the brightness control of the fiber illumination:
- Turn left: Minimum brightness
- Turn right: Maximum brightness

4 "Maximum load" warning label


The maximum load on the suspension arm must not exceed 7 kg.
max. 7 kg

5 Balance of suspension arm


After all accessories have been attached to the surgical microscope, this
adjustment screw can be used to set the balance of the suspension arm.
- Turn left: Reduce weight
- Turn right: Increase weight

Version 7.0
Page 24 G-30-1781-en
S100 / OPMI pico Safety Measures

Fig. 6: Symbols and labels on 1


the support and suspension
arm S100 Stands
1

2 1 2 3

4
2 5 1
2

1
1

2 1 2 3

5
4
1
2

Version 7.0
G-30-1781-en Page 25
Safety Measures S100 / OPMI pico

Symbols and labels on the S100 floor stand


1 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Your options to contact the device manufacturer,
i.e., phone number, fax number and email address of the local contact
of the national ZEISS sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your device.
2 System identification label
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CSA label
– CE label

3 Plate "Date of manufacture"


This plate indicates the date of manufacture for the device (year-month).

4 Label "Transport position"


Always hold the stand at the indicated point for transportation.
Observe the user manual or accompanying documentation.
Indicates the transport position for the device. Before transporting the
system, always ensure that it is in this position to avoid any damage.

Version 7.0
Page 26 G-30-1781-en
S100 / OPMI pico Safety Measures

Fig. 7: Symbols and labels on


the S100 floor stand

1
2
3

Version 7.0
G-30-1781-en Page 27
Safety Measures S100 / OPMI pico

5 Transport position for the carrier arm flat screen monitor


Observe the user manual or accompanying documentation.
Indicates the transport position of the carrier arm flat screen monitor. Be-
fore transporting the system, make sure that the carrier arm flat screen
monitor is in this position in order to avoid any damage.

6 "Maximum load" warning label on flat screen monitor bracket


The maximum load (accessory equipment) on the optional flat screen
monitor bracket must not exceed 9 kg!

7 "Maximum load" warning label on instrument tray


The maximum load (accessory equipment) on the optional instrument tray
must not exceed 13 kg!

Version 7.0
Page 28 G-30-1781-en
S100 / OPMI pico Safety Measures

Fig. 8: Symbols and labels on


the device carrier and monitor
arm (optional)

5
6

Version 7.0
G-30-1781-en Page 29
Safety Measures S100 / OPMI pico

Symbols and labels on the S100 wall stand


1 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Your options to contact the device manufacturer,
i.e., phone number, fax number and email address of the local contact
of the national ZEISS sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your device.
2 Device identification label (S100 wall stand with long arm)
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CSA label
– CE symbol

3 Plate "Date of manufacture"


This plate indicates the date of manufacture for the device (year-month).

Version 7.0
Page 30 G-30-1781-en
S100 / OPMI pico Safety Measures

Fig. 9: Symbols and labels on


the S100 wall stand

1
2
3

Version 7.0
G-30-1781-en Page 31
Safety Measures S100 / OPMI pico

Symbols and labels on S100 ceiling stand


1 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Your options to contact the device manufacturer,
i.e., phone number, fax number and email address of the local contact
of the national ZEISS sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your device.
2 System identification label
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CSA label
– CE label

3 Plate "Date of manufacture"


This plate indicates the date of manufacture for the device (year-month).

Version 7.0
Page 32 G-30-1781-en
S100 / OPMI pico Safety Measures

Fig. 10: Symbols and labels on


S100 ceiling stand

5
6
7

Version 7.0
G-30-1781-en Page 33
Safety Measures S100 / OPMI pico

Symbols and labels on S100 Centro stand


1 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Your options to contact the device manufacturer,
i.e., phone number, fax number and email address of the local contact
of the national ZEISS sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your device.
2 System identification label (Centro suspension system)
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CSA label
– CE label

3 Plate "Date of manufacture"


This plate indicates the date of manufacture for the device (year-month).

Version 7.0
Page 34 G-30-1781-en
S100 / OPMI pico Safety Measures

Fig. 11: Symbols and labels on


the S100 Centro stand

1
2
3

Version 7.0
G-30-1781-en Page 35
Safety Measures S100 / OPMI pico

Symbols and labels on the S100 carrier units


1 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Your options to contact the device manufacturer,
i.e., phone number, fax number and email address of the local contact
of the national ZEISS sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your device.
2 System identification label (S100 support unit)
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CSA label
– CE label
3 System identification label
(S100 support unit for secondary light)
The system identification label provides the following information:
– Manufacturer symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CE label

Version 7.0
Page 36 G-30-1781-en
S100 / OPMI pico Safety Measures

4 Plate "Date of manufacture"


This plate indicates the date of manufacture for the device (year-month).

Fig. 12: Symbols and labels


on the S100 carrier units

1
2
4

Fig. 13: Symbols and labels on


the S100 carrier units
with secondary light source

1
3
4

Version 7.0
G-30-1781-en Page 37
Safety Measures S100 / OPMI pico

Symbols and labels on the Atmos support unit


1 System identification label support unit Atmos
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CE label

2 Plate "Date of manufacture"


This plate indicates the date of manufacture for the device (year-month).

Version 7.0
Page 38 G-30-1781-en
S100 / OPMI pico Safety Measures

Fig. 14: Symbols and labels


on the S100 Atmos
support unit

1
2

Version 7.0
G-30-1781-en Page 39
Safety Measures S100 / OPMI pico

Symbols and labels on the light sources

Halogen light source


1 Indicating label "Observe the user manual"
Observe the user manual or accompanying documents.

2 S100 Halogen rating label


The rating label provides the following information:
– Manufacturer symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Illumination technique Halogen
– Rated voltage (115V): 100 V - 125 V
(230V): 220 V - 240 V
– Maximum current consumption (115 V): max. 250 VA
(230 V): max. 250 VA
– Power frequency range 50Hz - 60Hz

3 Fuse label
This plate contains the following information:
– Maximum voltage 115V/ 230V
– Rated current 6.3 A-H / 3.15 A-H
– (Tripping) characteristics T

4 "Observe disposal regulations" label


Electrical or electronic devices must not be disposed of as normal domestic
waste. For more information on the disposal of electrical and electronic
devices, please see the chapter on "Care and Maintenance".

5 "Observe the user manual" label


For information on how to replace the lamp, see Page 224.

Version 7.0
Page 40 G-30-1781-en
S100 / OPMI pico Safety Measures

Fig. 15: Symbols and labels on


the HALOGEN illumination

Version 7.0
G-30-1781-en Page 41
Safety Measures S100 / OPMI pico

6 "Hot surfaces" warning label


Hot surfaces may cause burns.

12V/100W 7 Lamp identification


When replacing lamps, only use halogen lamps with the rated line voltage.
380079-9040 Order these halogen lamps from ZEISS by stating the specified catalog
number.
8 Order number "Halogen lamp module"
When the lamp module is defective, you can order a new lamp module
000000-1086-764 from ZEISS. Please specify the imprinted order number.
9 CAUTION label
Warning of a potential hazard

Fig. 16: Symbols and labels


on lamp insert module

6 7

9 8

Version 7.0
Page 42 G-30-1781-en
S100 / OPMI pico Safety Measures

Xenon light source


1 "Observe the Instructions for Use" label
Observe the Instructions for Use or accompanying documentation.

2 S100 Xenon rating label


The rating label provides the following information:
– Manufacturer symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Illumination technique Xenon
– Rated voltage (115 V): 100 V - 125 V
(230 V): 220 V - 240 V
– Maximum current consumption (115 V): max. 350 VA
(230 V): max. 350 VA
– Power frequency range 50Hz - 60Hz

3 "Observe disposal regulations" information label


Do not dispose of electrical and electronic devices along with normal do-
mestic waste. For more information on the disposal of electrical and elec-
tronic devices, please see "Maintenance and care".

Fig. 17: Symbols and labels on


the Xenon Illumination

Version 7.0
G-30-1781-en Page 43
Safety Measures S100 / OPMI pico

4 "Observe the user manual" label


For information on replacing the lamp module, please see Page 226.

5 Warning label
The xenon light source shall not be used for ophthalmological surgeries.
Make sure that no xenon light enters the patient's eyes.
6 "CAUTION" warning plate
The lamp housing contains an integrated xenon light source.

7 Replace Lamp
The plate graphically shows the three steps involved in changing the lamp.
A - Press the button
B - Pull out the lamp insert module
C - Change bulbs

8 Switch setting label


1 2 Indicates which lamp is in operation.
1. Main lamp
2. Backup lamp

Fig. 18: Symbols and labels


on the
Xenon illumination
5
6

8 1 2

Version 7.0
Page 44 G-30-1781-en
S100 / OPMI pico Safety Measures

9 "CAUTION" label
Warning of a potential hazard

10 "Hot surfaces" warning label


Hot surfaces may cause burns.

11 Label "Xenon lamp insert module"


If the lamp insert module is defective, you can order a new lamp insert
304977-9037-000 module from ZEISS (see ordering information Page 251).
12 "Xenon" power rating

180W Xenon lamp

Fig. 19: Symbols and labels


on lamp insert module

8 9 10

11

Version 7.0
G-30-1781-en Page 45
Safety Measures S100 / OPMI pico

LED light source


1 Indicating label "Observe the user manual"
Observe the user manual or accompanying documents.

2 S100 S100 power rating plate


The rating label provides the following information:
– Manufacturer symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Illumination technique LED
– Material number 305929-9010-000
– Rated voltage 100V - 240V
– Maximum power consumption 250 VA
– Power frequency range 50Hz - 60Hz

3 "Observe disposal regulations" label


Do not dispose of electrical and electronic devices along with normal do-
mestic waste. For more information on the disposal of electrical and elec-
tronic devices, please see chapter "Maintenance and care".

4 LED warning sign


The LED light source in question is a
class 2 LED device according to IEC 60825-1:2001. According to EN
62471:2009, this light source must be classified as belonging to the free
group.
5 Laser warning sign
The LED light source is a class 2 LED device according to IEC 60825-
1:2001.

6 "Instructions for Use" label


Observe the Instructions for Use or accompanying documents.

Version 7.0
Page 46 G-30-1781-en
S100 / OPMI pico Safety Measures

Fig. 20: Symbols and labels on


the LED light source

5 4 3 2

Version 7.0
G-30-1781-en Page 47
Safety Measures S100 / OPMI pico

Symbols and video control units (optional)

1 Operating elements of the video control units


Description of the buttons on Page 76.
2 Connections to the video control units
Description of the connections on Page 86.
3 HD version label
The HD version can be identified by the 1080p resolution.
(Software upgrade)

4 Label for integrated HD video camera live & streaming with


SW recording license
Identifies the HD video camera live & streaming (full HD 1080p) with
SW recording license (optional software license)

Version 7.0
Page 48 G-30-1781-en
S100 / OPMI pico Safety Measures

Fig. 21: Video units

1 2

1
2

Version 7.0
G-30-1781-en Page 49
Safety Measures S100 / OPMI pico

Version 7.0
Page 50 G-30-1781-en
S100 / OPMI pico System Overview

System Overview

System overview..............................................................................52

Controls and connections ................................................................54


OPMI pico .................................................................................................54
Hand grips on OPMI pico...........................................................................58
Controls for the widefield eyepieces ..........................................................59
Controls for the tubes................................................................................60
Controls of OPMI pico with MORA interface..............................................62
Controls of OPMI pico with Endoport ........................................................65
S100 Suspension Systems ..........................................................................66
Attachment of carrier arm flat screen monitor to the S100 floor stand
(option) .....................................................................................................69
Halogen light source..................................................................................70
Xenon light source ....................................................................................72
LED light source.........................................................................................75
Button functions of the HD video control units ..........................................76
Remote control for the video camera live & streaming control unit (option) ...
82
Overview of operating elements ................................................................84

Version 7.0
G-30-1781-en Page 51
System Overview S100 / OPMI pico

System overview

Surgical microscope The surgical microscope (4) enables the illumination and visual magnification
of the field of view. For visual magnification, the surgical microscope can be
equipped with different tubes, eyepieces and objective lenses.

S100 suspension system The S100 comes as a floor stand, ceiling mount or wall mount. It is also avail-
able as a support unit and for adaptation to a Centro stand.
The S100 is used to control and supply the functions of the surgical micro-
scope. It consists of a carrier arm (2), suspension arm (3) and supply unit (1)
with light source. The suspension arm allows rough positioning of the surgical
microscope and moving into standby position.
The S100 floor stand can be positioned in the OR by four controllable
casters (5). The locking tabs of the casters prevent the system from being
moved or rolling away unintentionally.

Light source Depending on the order, the device may be equipped with a halogen, xenon
or LED light source. The halogen, xenon or LED light source is used to illumi-
nate the surgical area. It has a fiber illumination in the surgical microscope,
i. e. the light is directed from the light source via a light guide to the surgical
microscope. The halogen, xenon or LED light source is situated in the lamp
housing of the supply unit (1).
The halogen light source contains two halogen lamps. The second lamp is
used as backup lamp.
The xenon light source produces light whose spectrum is similar to daylight.
Regardless of the brightness setting, the color temperature of the light always
remains the same. Normal daylight film without any additional conversion fil-
ters can therefore be used for photographic documentation.
The xenon light source contains a lamp module with two xenon lamps. The
second lamp is used as backup lamp.
The LED light source has an integrated failure concept. In the event of a defect
in the LED light source, the light intensity drops to 50% and the control lamp
on the lamp housing lights up.

Version 7.0
Page 52 G-30-1781-en
S100 / OPMI pico System Overview

Fig. 22: System overview


OPMI pico on S100 stand

Ceiling-mounted stand
2

Wall-mounted stand
3
1
2 4

Carrier unit
1 3

Centro stand 2

3
1

4
2

1 3 4

Floor stand

Version 7.0
G-30-1781-en Page 53
System Overview S100 / OPMI pico

Controls and connections

OPMI pico

1 Friction adjustment button


For the fixation of the surgical microscope on the carrier arm in the desired
position.
2 Dust cover
3 Magnification changer
Use this button to manually set/adjust the magnification to one of the five
positions (γ = 0.4x, γ = 0.6x, γ = 1.0x, γ = 1.6x, γ = 2.5x).
4 Video connection socket (optional)
For the system cable to the HD video camera (optional).
5 Light guide socket
This socket is used to direct the light from the light source to the surgical
microscope.
6 Filter selection
Turn this selector to swing filters into the illumination beam path. The se-
lector has three positions: no filter, green filter and orange filter.
– The green filter enhances contrast.
– The orange filter prevents premature curing of composite fillings.

Version 7.0
Page 54 G-30-1781-en
S100 / OPMI pico System Overview

Fig. 23: Operating elements


and connectors on the sur-
gical microscope
6

5
4
1

2
3

3 2

Version 7.0
G-30-1781-en Page 55
System Overview S100 / OPMI pico

1 Friction adjustment button


For clamping the the surgical microscope at the support arm and for
mounting the surgical microscope in the desired position.
3 Magnification changer
Use this button to manually set/adjust the magnification to one of the five
positions (γ = 0.4x, γ = 0.6x, γ = 1.0x, γ = 1.6x, γ = 2.5x).
7 Cover cap
You can remove this cover cap and mount a hand grip for microscope
guidance instead.
8 Objective lens
a Objective lens with fixed focal length
Objective lenses with different focal lengths are available. The focal
length equals approximately the working distance.
Variably adjustable working distance, focus range 100 mm.
b Focusing objective lens
Objective lenses with different focal lengths are available. The focal
length equals approximately the working distance. Focusing range
13 mm.
c Varioscope 100
Variably adjustable working distance, focus range 100 mm.
9 Drilled holes for fastening accessories
The bottom of the surgical microscope contains two holes for mounting
accessories (e.g. dovetail mount, handgrip, please see Page 58).
10 Securing screw
Screw in the securing screw as far as it will go after you have inserted the
binocular tube or another module in the mount on the surgical micro-
scope.

Version 7.0
Page 56 G-30-1781-en
S100 / OPMI pico System Overview

Fig. 24: Operating elements


3
and connections on the sur-
10
gical microscope
1

7
8a
3

3
10

1
7

3 8b

3
10

3 8c

Version 7.0
G-30-1781-en Page 57
System Overview S100 / OPMI pico

Hand grips on OPMI pico

Five application-oriented handgrips are provided:


1 Handgrip, complete, rear
2 Comfort grips
On the comfort grips, the grip components can be repositioned to meet
the user's personal requirements.
3 Comfort T handgrip
4 T-shaped grip, complete
5 Handgrip, complete, on the front

Fig. 25: Handgrips on


OPMI pico 1

Version 7.0
Page 58 G-30-1781-en
S100 / OPMI pico System Overview

Controls for the widefield eyepieces

NOTE Widefield eyepieces with magnetic coupling


Please keep in mind the usual rules for the handling of magnets in regard of
eyepieces that have been removed from the tube.
• Do not place the eyepiece near instruments which may be magnetizable.
• Do not place the eyepiece on sensible devices such as infusion pumps,
pacemakers, measuring devices or magnetic data media such as floppy
disks, audio and video tapes or credit cards.
• Always store the eyepiece in its original packaging, when not in use.

CAUTION Injury to the eye due to light radiation!


Looking into a light source through a tube, an objective lens or an eyepiece
may cause eye injury.
• Never look into a light source or the sun with a tube, an objective lens or
an eyepiece.
1 Eyecup
for adjusting the distance between the eyepiece and the eye.
2 Diopter setting ring
adjustable from +5 D to -8 D
3 Diopter scale
for reading off the prescription set

Fig. 26: Operating elements


of the eyepieces

2
3

Version 7.0
G-30-1781-en Page 59
System Overview S100 / OPMI pico

Controls for the tubes

CAUTION Injury to the eye due to light radiation!


Looking into a light source through a tube, an objective lens or an eyepiece
may cause eye injury.
• Never look into a light source or the sun with a tube, an objective lens or
an eyepiece.
The tubes have the following controls:
1 Eyepiece mount
2 Pupillary distance (PD) adjustment knob
The correct position has been reached when the two eyepiece images
merge into one. You can read off the PD setting on the adjustment knob.

Fig. 27: Controls for the tubes 180° tiltable tube 45° inclined tube

1 2 1 2

Straight tube
1

Version 7.0
Page 60 G-30-1781-en
S100 / OPMI pico System Overview

Folding tube f170/f260 mm (optional)


3 Eyepiece mount
4 Pupillary distance (PD) adjustment knob
5 Scale for pupillary distance
6 Symbol "normal/larger" for magnification changer
7 Dial for integrated magnification changer
8 Symbol (released/locked) for manual tube rotation
9 Button for manual tube rotation

Fig. 28: Operating elements - Foldable Tube f170/f260


foldable tube
3 4 5 6 7 8 9

Version 7.0
G-30-1781-en Page 61
System Overview S100 / OPMI pico

Controls of OPMI pico with MORA interface.

Use The MORA interface is intended exclusively for use on the OPMI pico surgical
microscope in dentistry.

Description The MORA interface increases the microscope's mobility about its lateral tilt
axis, without changing the user's upright, ergonomic seated posture.
This offers advantages for the treatment of relatively large fields of therapy,
e.g. when several teeth are being treated at once.
A manually operated annular diaphragm permits the depth of field to be
varied.
With maximum lateral tilt of the microscope body and high magnification,
slight vignetting of the field of view may occur.

Operating elements and 1 Marking of zero position


Connectors
2 Friction of lateral tilt movement
The microscope body can be tilted by approx. 25° to the left or right, and
can be locked in any position using the friction screw.
3 Securing screw
Secures the MORA interface on the OPMI pico
4 Diaphragm setting knob
A swing-in annular diaphragm permits you to increase the depth of field
so that a larger area of the field of therapy is sharply displayed. This in-
volves a reduction in light intensity. Set the annular diaphragm to meet
your specific requirements.
5 Securing screw
Secures the binocular tube to the MORA interface.

Version 7.0
Page 62 G-30-1781-en
S100 / OPMI pico System Overview

Fig. 29: OPMI pico with


MORA interface
1

approx. ±25°

Version 7.0
G-30-1781-en Page 63
System Overview S100 / OPMI pico

Controls of MORA interface with lateral output port for documenta-


tion

CAUTION Video images not for diagnostic purposes!


The visualized images may include deviations in scale, shape and color.
• Do not use video images for diagnostic purposes as the video camera is
not calibrated for these purposes.
The MORA interface is optionally available with documentation equipment.
This version is shown in the illustration on the opposite page.
The method described below can also be used for other accessories. The op-
erating principle of the accessories is described in the relevant user manuals.

Controls and 1 Annular diaphragm


connectors We recommend that you swing the annular diaphragm out of the beam
path for image capture to ensure that sufficient light is available. With the
annular diaphragm swung into the beam path, the exposure time in-
creases accordingly.
2 Knurled ring
Attachment of dust cover or documentation accessories
3 Dust cover
4 Accessory module

Fig. 30: MORA interface with


lateral output port for docu-
mentation accessories

1 2 3 4

Version 7.0
Page 64 G-30-1781-en
S100 / OPMI pico System Overview

Controls of OPMI pico with Endoport

Use The Endoport endoscope port is intended for use on OPMI pico microscopes
in ENT surgery.

Description The endoscope port of the OPMI pico Endoport is a standardized interface for
endoscope cameras. This port permits an external endoscope camera with
special adapter to be connected using the quick-action locking mechanism.
If you want to equip the surgical microscope with the Endoport accessory,
please observe the requirements for system expansions, see Page 251.
Only attach external endoscope cameras including adapter with a maximum
weight of 1.0 kg to the endoscope port.
The endoscope port always produces an image at infinity; for this reason, the
external endoscope camera's focus must also be set at infinity. The procedure
of focusing the endoscope camera at infinity is a camera-specific procedure
and differs from camera type to camera type.
Microscope settings such as the focusing of the eyepieces do not affect the
endoscope port.

Controls and 1 Endoscope port


connectors
2 Endoscope camera

Fig. 31: OPMI pico with


Endoport

Version 7.0
G-30-1781-en Page 65
System Overview S100 / OPMI pico

S100 Suspension Systems

1 Carrier arm
2 Suspension arm
The suspension arm is equipped with a switch. This switch activates the
light source when the suspension arm is moved into its working position.
The light source is switched off in the upper rest position.
3 Mobility adjustment screw
You are able to set the lifting mobility of the suspension arm with the ad-
justment screw after the successful balance setting of the microscope
equipment.
4 Mobility adjustment screw
Use this screw to set the mobility of the coupling's rotation.
5 Mobility adjustment screw
Use this screw to set the mobility of the surgical microscope's swing
movement.
6 120° coupling
The 120° coupling improves ergonomic handling.
7 K 0/120 coupling
We recommend using the K 0/120 coupling for gynecological procedures.
This makes it possible to shift the working area downwards by 115 mm.
8 Securing screws
9 Adjustment screw for weight compensation
10 Mobility adjustment screw
You can use this screw to adjust the rotational mobility of the suspension
arm.
11 Mobility adjustment screw
You can use this screw to adjust the rotational mobility of the carrier arm.

Version 7.0
Page 66 G-30-1781-en
S100 / OPMI pico System Overview

Fig. 32: Operating elements 1


and connections on the
S100 stand

2 3 4 5

11

8
10 9 6

2 3 4

8
8
7
8

Version 7.0
G-30-1781-en Page 67
System Overview S100 / OPMI pico

12 Supply unit with light source


The connectors for the power cord, light guide, camera and video as well
as the power switch are located on the supply unit. The supply unit in-
cludes the lamp housing with the halogen, xenon or LED light source
13 Steerable casters with arrestors (on floor stand only)
Press the locking tabs down once to lock the stand into position.
To unlock lift the locking tab upward with the tip of your foot.
14 Wall flange (wall mount only)
The wall mount differs from the ceiling mount by the type of mounting to
the building.
15 Ceiling flange (ceiling mount only)
The ceiling mount differs from the wall mount by the type of mounting to
the building.

Fig. 33: Operating elements


and connections on the
S100 stand
15

12

12

14

13

12

Version 7.0
Page 68 G-30-1781-en
S100 / OPMI pico System Overview

Attachment of carrier arm flat screen monitor to the S100 floor stand
(option)

It is possible to upgrade the S100 floor stand to a documentation workstation


comprising documentation and display devices. Optionally, you can mount a
carrier arm flat screen monitor and/or an instrument tray on the S100 floor
stand. The carrier arm flat screen monitor and the instrument tray can be var-
iably positioned up to a specific height (please refer to Page 116). It is also
possible to additionally set the variable tilt of the carrier arm flat screen mon-
itor. Thanks to these adjustment possibilities, this option provides you with a
clearly arranged, ergonomic combination of the surgical microscope on the
S100 floor stand as well as a documentation and display device.

Carrier arm flat screen monitor (1)


A flat screen or a touchscreen can be mounted on the carrier arm flat screen
monitor. When mounted on the S100 floor stand, the carrier arm flat screen
monitor has a maximum load capacity of 9 kg.

Instrument tray for a documentation device(2)


The instrument tray is used to carry a documentation device with a maximum
weight of 13 kg, e.g. a video recorder.

Fig. 34: Adapting the instru-


ment tray and carrier arm flat
screen monitor to the
S100 floor stand
1

Version 7.0
G-30-1781-en Page 69
System Overview S100 / OPMI pico

Halogen light source

1 Ventilation grid
Do not cover the ventilation grid!
For example, drapes could be covering the screen. This could cause the in-
tegrated thermal switch to turn off the light source.
2 Flap
The flap is used as the mechanical indicator for the operating status of the
halogen lamps.
– When the flap is closed, the main lamp is operative.
– The open flap indicates that the backup lamp is being used.
3 Manual switching to the backup lamp
Pressing this button swivels in the backup lamp (Page 217)
4 Slide-in lamp module
5 Opening the slide-in lamp module
When you press this button, the slide-in lamp module is partially ejected.
Pull out the lamp module all the way for exchanging the lamp.
6 Brightness setting
If the light source is switched on, the brightness can be infinitely adjusted
by turning the brightness control button for the fiber illumination.
- Left position: Minimum brightness
- Right position: Maximum brightness

Fig. 35: Operating elements


and connections on the
halogen light source
1

3 6
4
5

Version 7.0
Page 70 G-30-1781-en
S100 / OPMI pico System Overview

Connector panel of the halogen light source


1 Power switch
2 Light guide socket
The light guide socket is used to direct the light from the light source to
the surgical microscope.
3 Strain relief device
The strain relief device prevents inadvertent unplugging of the power
cable.
4 Power inlet
The power cord connects the device to the power socket. The power inlet
socket contains the device fuse.
5 Display window for rated voltage and voltage switchover
The rated voltage indicated at the sliding switch must correspond to the
rated voltage of the net available on the site of installation. If the rated
voltage has not been correctly set, you can adjust the sliding switch using
a suitable tool.
6 Potential equalization connector (can be ordered separately, optional)
This connector can be used for placing other active devices at the same
ground potential or for redundant grounding to protective earth.

Fig. 36: Connector panel for


the halogen light source

4
1
5

Version 7.0
G-30-1781-en Page 71
System Overview S100 / OPMI pico

Xenon light source

CAUTION Limited service life of the xenon lamp!


If used longer than the indicated service life of 500 hours the xenon lamp
may fail or burst.
Dropping fragments may contaminate the sterile area. Risk of infection.
• Manually swivel in the backup lamp (Page 218) or replace the lamp mod-
ule in a timely manner (see Page 226).
1 Ventilation grid
2 Lamp module
3 Manual switching to the backup lamp
The xenon light source has a lamp module containing two xenon lamps.
The second lamp is used as backup lamp. Manually swivel in the backup
lamp (Page 218) if it is used beyond its maximum service life of 500h (see
also chapter "Maintenance intervals" on Page 222) or if the the first lamp
fails.
If the backup lamp is operated, keep a new lamp module ready for ex-
change (Page 226).
4 Indicator: Backup lamp is in use
If the red segment (4) in the button (3) is lit, the backup lamp is operated.
5 Opening the lamp module
When you press this button, the lamp module is partially ejected.
6 Brightness control
If the light source is switched on, the brightness can be continuously set
by turning the brightness control of the fiber illumination:

Version 7.0
Page 72 G-30-1781-en
S100 / OPMI pico System Overview

Fig. 37: Operating elements


on the xenon illumination

2
3 4

Version 7.0
G-30-1781-en Page 73
System Overview S100 / OPMI pico

Connector panel of the xenon light source


1 Light guide socket
The light guide socket is used to direct the light from the light source to
the surgical microscope.
2 Strain relief device
The strain relief device prevents inadvertent unplugging of the power
cable.
3 Power switch
When the device is switched on, the green indicator light in the power
switch is lit.
4 Power supply socket
The power cord connects the device to the power socket. Only connect
the system to a wall outlet which is properly grounded.
5 Selector switch for rated voltage and voltage switchover
The rated voltage indicated at the sliding switch must correspond to the
rated voltage of the net available on the site of installation. If the rated
voltage has not been correctly set, you can adjust the sliding switch using
a suitable tool.
6 Potential equalization connector (can be ordered separately, optional)
This connector can be used for placing other active devices at the same
ground potential or for redundant grounding to protective earth.

Fig. 38: Connector panel for


the xenon light source

4
5

Version 7.0
Page 74 G-30-1781-en
S100 / OPMI pico System Overview

LED light source

Connector panel for the LED light source


1 Light guide socket
The light guide socket is used to direct the light from the light source to
the surgical microscope.
2 Strain relief device
The strain relief device prevents inadvertent unplugging of the power
cable.
3 Power switch
When the device is switched on, the green indicator light in the power
switch is lit.
4 Power supply socket
The power cord connects the device to the power socket. Only connect
the system to a wall outlet which is properly grounded.
5 Potential equalization
This connector can be used for placing other active devices at the same
ground potential or for redundant grounding to protective earth.

Fig. 39: Connector panel for


the LED light source

Version 7.0
G-30-1781-en Page 75
System Overview S100 / OPMI pico

Button functions of the HD video control units

Depending on the integrated video camera selected, the supply unit contains
different video control units:
For the integrated SD video camera
– Integrated HD 720p control unit
– Integrated video camera live & streaming control unit
with SW recording license (optional SW license)
The area to the right of the dashed line is displayed with a 90° rotation. It is
located at the level of the connector panel.
1 Config
Use the <config> key to go to the menu.
The configuration menu permits you to set various function sequences to
meet your specific needs. The different possible settings in the configura-
tion of the application program are explained in the chapter "Operation"
(Page 127). The camera is delivered with a default configuration (factory
settings).
2 Indicator light (integrated HD 720p control unit)
The indicator light (LED) is lit when the power switch has been turned on.
3 Indicator lights (integrated video camera live & stream control unit)
The indicator light "power" is lit when the power switch has been turned
on.
The "network" indicator light flashes during network data transmission,
The indicator lamp "record" is lit when recording a video.
4 White bal
White balance for integrated HD 720p control unit (Page 165)
and for the integrated video camera live & streaming control unit
(Page 185)
5 Brightness (integrated HD 720p control unit)
You can use the and keys to adjust the brightness of the video
image.
The configuration menu () gives you the choice between two different
modes: Automatic - Manual.
6 Brightness (integrated video camera live & streaming control unit)
Use the <brightness> key to go to the "Brightness" menu (Page 182)
where you can configure settings.

Version 7.0
Page 76 G-30-1781-en
S100 / OPMI pico System Overview

Fig. 40: Button functions 1 2 4 5


HD video control unit

3 4 6 1

Version 7.0
G-30-1781-en Page 77
System Overview S100 / OPMI pico

7 Image
The <freeze> button allows you to freeze the current live video image and
to display it as a full-screen freeze frame on the video monitor. In the
meantime the text "freeze" appears in the upper left corner.
When you press the <live> key, the current live video image is displayed
again on the monitor. The text "freeze" disappears.
8 Colors (integrated video camera live & streaming control unit)
Use the <colors> key to go to the "Colors" menu (Page 186) where you
can configure the color settings.
9 Live (integrated video camera live & streaming control unit)
Closes all open dialogs, discontinues all actions except <record>. When
you press the <live> key, the current live video image is displayed again
on the monitor.
10 "Select" control keys (cursor keys)
Use can use the control keys to navigate the menu. In addition, the cursor
keys have special functions while an image is being displayed or a video is
being played (Page 80).
11 "Enter" key ( )
Press the <enter> key to confirm a selection or trigger a function. The dif-
ferent functions are described in the respective menu item of the menu.
In addition, textual information for the selected mode is displayed.

Version 7.0
Page 78 G-30-1781-en
S100 / OPMI pico System Overview

Fig. 41: Button functions 7 10 11


HD video control unit

8 9 10 11

Version 7.0
G-30-1781-en Page 79
System Overview S100 / OPMI pico

12 Record button
Press the <record> button to start a video recording. All open menus will
be closed and an image display or video playback will be terminated. The
video recording starts.
In the footer bar the text "Rec" appears on the right before USB/LAN and
the recording time appears on the left. The "record" indicator light (3) will
light up.
Press the <record> button to end a video recording. The text "Rec" and
the recording time appears. The "record" indicator light (3) goes out.
13 Capture button
Images can be created in live mode and during video recording or
video playback. Press the <capture> button to trigger the image capture.
The text "Capture" appears on the left in the header bar of the display.
The <Capture> button can be used to create individual images during
video recording or playback.
14 Play button
Press the <play> button to display the last image captured or to play the
last video recorded.
If the <play> button is pressed during video playback, the current display
appears as a still image. The text "Pause" appears on the left in the header
bar of the display.
Press the <play> button again to continue playing the video.
15 Index button
Press the <Index> button to open the index view of the files of the acti-
vated patient. Use the control keys to select a video or image. A blue
frame marks the respective selected file.
Press the <index>, <config> or <live> button, to terminate the index view.
16 Function key (F1)
The <function> key can be programmed individually (Page 190).
17 Function key (F2)
The <function> key can be programmed individually (Page 190).
18 Right cursor key (next image/10% skip forward)
With the right cursor key you can switch to the next image (in the list
below, older) while an image is being displayed, and while a video is being
played you can move forward 10% and continue playing the video from
there.

Version 7.0
Page 80 G-30-1781-en
S100 / OPMI pico System Overview

19 Left cursor key (previous image/10 % skip back)


With the left cursor key you can switch to the previous image (in the list
above, newer) while an image is being displayed, and while a video is
being played you can move back 10 % and continue playing the video
from there.
20 Down cursor key (fast forward)
Press the down cursor key to scan through the selected video sequence in
fast forward. The text "Forward" appears in the header bar of the display.
Press the <play> button to play the video at normal speed again.
21 Up cursor key (fast rewind)
Press the up cursor key to scan through the selected video sequence in fast
rewind. The text "Rewind" appears on the left in the header bar of the dis-
play.
Press the <play> button to play the video at normal speed again.

Fig. 42: Button functions


HD video control unit 16 19 21 20 18 12 13 14 15 17
live & streaming

Version 7.0
G-30-1781-en Page 81
System Overview S100 / OPMI pico

Remote control for the video camera live & streaming control unit (option)

Remote control is delivered as a standard feature with the integrated video


camera live & streaming control unit with SW recording license. As an option,
it is also available without a SW recording license with limited functions.

CAUTION Danger of infection caused by non-sterile remote control unit (op-


tion)!
The remote control and USB keyboard (option) are not intended for a sterile
operation!
If the non-sterile remote control unit (option) is used in the sterile OR, germs
may be transmitted from the user to the patient.
• The remote control unit (option) may only be used in the non-sterile area.

The button functions of the remote control are identical to those of the oper-
ator panel on the device.
The following functions can only be carried out on the HD control unit:
– White bal
– Brightness
– Colors
– Function F1
– Function F2

Remote control unit key function


1 ON
Not used
2 Input
Not used
) 1)
3 Control keys / "Enter" key (
Navigation in the menu / Confirming the selection
4 Live 1) 2)
Closes all open dialogs, discontinues all actions except <record>. If the
<Live> button is pressed, the current live video image will appear again on
the monitor.

Version 7.0
Page 82 G-30-1781-en
S100 / OPMI pico System Overview

5 Index 2)
Index view of the files of the activated patient (Page 207)
6 Play
Play the video (Page 207):
Press the <Play> button: Toggles between play video / pause video
Press the <Live> button to stop the playback
7 Record 1) 2)
Start / stop recording (Page 205):
Press the <Record> button: Toggles between start recording /
stop recording
Without SW recording license:
Press the <Record> button: Toggles between streaming ON/OFF
8 Capture 1) 2)
Show (Page 207) / create (Page 205) individual images (photo)
Without SW recording license:
Press the <Capture> button: Toggles between Freeze/Live
9 Menu (Page 176)

1)
Functions without SW recording license
2) Functions in the web interface (see Page 201)

Fig. 43: Remote control unit


3 9
9 1

8
3
7
4 6
8 4 5
5
7 2
6

Version 7.0
G-30-1781-en Page 83
System Overview S100 / OPMI pico

Overview of operating elements

Here you will find an overview of the different operating elements of the in-
tegrated HD video camera live & streaming (full HD 1080p) with SW recording
license and its functions.

Control unit Remote Web Foot


control Interface switch
Control keys / Navigation in the menu / - -
Enter key Confirm
Config Opens Main menu - -
White bal Opens White Balance - - -
submenu
Brightness Opens Brightness sub- - - -
menu
Colors Opens Colors submenu - - -
Live Closes all dialogs, discontinues all actions, live -
image appears on the monitor
Function F1 Zoom ON/OFF** - - -
Smart recording ON/
OFF*
Function F2 Switch camera profile - - -
Switch storage mode* /
Streaming ON/OFF**
Capture USB/LAN storage mode: Image capture*
Streaming storage mode: Freeze/Live*
Freeze/Live**
Record Video recording ON/OFF*
Streaming ON/OFF**
Play Play/pause video -
Index USB/LAN storage mode: Opens index* -
Streaming storage mode: Error message*
No function**

* With SW recording license


** Without SW recording license

Version 7.0
Page 84 G-30-1781-en
S100 / OPMI pico System Overview

Press the <config> key of the remote control to go to the menu. From there
you can navigate the menu using the control keys and the "enter" key and this
way configure the integrated video camera live & streaming control unit.

Version 7.0
G-30-1781-en Page 85
System Overview S100 / OPMI pico

HD video control unit connections only


1 Remote socket (HD)
A foot switch (option) can be connected here.
You can configure the foot switch for the integrated HD video camera
(HD ready 720p) as described in Page 152 or for the integrated video
camera live & streaming control unit (full HD 1080p) as described in
Page 191.
2 Connection for the camera head
The system cable connects the camera installed in the surgical microscope
to the video unit.
Switch the video unit off before plugging or unplugging the system cable
into/from the video unit. The camera may be damaged if the cable is
plugged/unplugged while voltage is still being applied.
3 "HDMI" video output
Digital connection for monitors or recorders from the consumer area.
HDMI (High Definition Multimedia Interface) is used to transfer digital
video data.
4 "YPbPr" video output
Analog connection for high resolution monitors or TV sets
(For the YPbPr cable, see Order Information, Page 254).
5 SD video
"Composite" video output (FBAS). Analog connection for video compo-
nents.
6 USB port
Use only USB media that meet at least USB 2.0 specifications.
7 Service interface
Reserved for ZEISS Service personnel.
8 LAN input/output
Network 100 Mbit
9 Video signal output Y/C
Use the Y/C output if only one analog SD monitor is available and no HD
monitor.
No menus and status information (header bar and footer bar) will be dis-
played in the Y/C video signal output. Use an HD video signal output (DVI
oder YPbPr) for operation and configuration.
We recommend the use of high resolution monitors.

10 DVI output
Analog connection for video components.

Version 7.0
Page 86 G-30-1781-en
S100 / OPMI pico System Overview

Fig. 44: Connectors 1 2


HD video control unit

6 7

1
8
9
10
4

Version 7.0
G-30-1781-en Page 87
System Overview S100 / OPMI pico

Version 7.0
Page 88 G-30-1781-en
S100 / OPMI pico Preparations for use

Preparations for use

Configuring the surgical microscope................................................91


Mounting the surgical microscope.............................................................92
Dismantling the surgical microscope..........................................................94
Mounting the tube, eyepieces and objective lens.......................................96
Mounting the hand grips...........................................................................97
Connecting the light guide ........................................................................98
Connecting an endoscope camera.............................................................99
Connect the stereo co-observer ...............................................................100
Mounting the MORA interface ................................................................101
Mounting documentation accessories on the MORA interface with lateral out-
put port...................................................................................................102

Setting the suspension arm weight compensation ........................ 103


Adjusting the surgical microscope ...........................................................104

Relocating the S100 floor stand ....................................................106

Connections................................................................................... 108
Connecting the S100 floor stand .............................................................108
Connecting the integrated video camera .................................................110
Connecting the computer to integrated video camera live & streaming Control
unit with LAN connection ........................................................................114

Attachment of instrument tray and carrier arm flat screen monitor to


the S100 floor stand ......................................................................116
Ergonomic positioning the carrier arm flat screen monitor and instrument tray
116
Mounting the flat screen monitor on the carrier arm flat screen monitor .118
Routing the cables and making the electrical connections .......................120
Adjusting the carrier arm flat screen monitor...........................................122

Version 7.0
G-30-1781-en Page 89
Preparations for use S100 / OPMI pico

Preparing the system for sterile use ..............................................125


Attach asepsis caps and drapes............................................................... 125

Version 7.0
Page 90 G-30-1781-en
S100 / OPMI pico Preparations for use

Configuring the surgical microscope

CAUTION Risk of injury due to sinking surgical microscope or due to dropping


parts!
Mounting and removing modules and accessories may lead to unintended
tilting of the surgical microscope. Parts or modules may drop.
The suspension arm may be dropped or lifted.
This may lead to personal injuries or to material damage.
• Please only use accessories approved by ZEISS.
• Never change modules and accessories during a surgical procedure or
above the patient!
• Mount documentation accessories only in accordance with the appropri-
ate instructions for use.
• After every change in configuration and before every use, make sure that
the modules and accessories are securely locked in position.
• Make sure that all securing screws and locking screws are tightened firm-
ly!
• Make sure not to exceed the maximum load capacity of the device during
the installation of accessories and components.
• Readjust the balance of the suspension arm after every change of acces-
sories (see Page 103).
• Readjust the adjustment screw for mobility. The adjustment screw has
only a braking, no locking function.

WARNING Danger of tipping due to overloading!


The mounting and removal of accessories may lead to unintentional tipping
of the surgical microscope.
• Make sure that the maximum permissible load on the suspension arm is
not exceeded.

Version 7.0
G-30-1781-en Page 91
Preparations for use S100 / OPMI pico

Mounting the surgical microscope

After you have mounted the OPMI pico, the suspension arm may move
downwards. Prevent the OPMI pico from knocking against anything.
The adjustment screw (1) on the suspension arm of the S100 suspension
system only acts as a brake but has no locking function.
Without coupling
• Tighten adjustment screw (1) (braking function).
• Insert the shaft (5) into the the suspension arm's holding fixture.
• Screw in and tighten the safety screws (2) and (4).
• The safety screw (2) contains a hex socket head screw (3), which is an ad-
ditional safety screw. Tighten it.
Including 120° coupling
• Tighten adjustment screw (1)(braking function).
• Insert the shaft (6) into the mount of the suspension arm.
• Screw in and tighten the safety screws (2) and (4).
• The safety screw (2) contains a hex socket head screw (3). This is an addi-
tional safety screw. Tighten it.
• Insert shaft (5) into the holding fixture of the coupling. Position and tight-
en the safety screws (7) and (8).
• Securing screw (8) contains a hex socket head screw (9) which is an addi-
tional securing screw. Tighten it.
Including K 0/120 coupling
We recommend using the K 0/120 coupling for gynecological procedures.
This makes it possible to shift the working area downwards by 115 mm.
• Tighten adjustment screw (1)(braking function).
• Insert the shaft (10) into the the suspension arm's holding fixture.
• Screw in securing screws (2) and (4). Tighten it.
• Securing screw (2) contains a hex socket head screw (3) which is an addi-
tional securing screw. Tighten it.
• Insert shaft (5) into the holding fixture of the coupling. Position and tight-
en the safety screws (11).

Version 7.0
Page 92 G-30-1781-en
S100 / OPMI pico Preparations for use

Fig. 45: Assembling and dis-


mantling the OPMI pico 1 2 3

120° coupling

K 0/120 coupling 9 8 7 6

10 5

11

Version 7.0
G-30-1781-en Page 93
Preparations for use S100 / OPMI pico

Dismantling the surgical microscope

Move the suspension arm into its upper position before dismantling the
surgical microscope. While the surgical microscope is dismounted, the
suspension arm may shoot up by its own spring force despite tightened
locking screw.
Without coupling
• Tighten adjustment screw (1)(braking function).
• Hold the surgical microscope securely while loosening the safety screws.
• Loosen safety screw (3).
• Loosen and remove the safety screws (2) and (4).
• Remove the surgical microscope.
Including 120° coupling
• Tighten adjustment screw (1)(braking function).
• Hold onto the surgical microscope firmly.
• Loosen safety screw (9).
• Loosen and remove the safety screws (7) and (8).
• Remove the surgical microscope.
• Hold onto the surgical microscope firmly.
• Loosen safety screw (3).
• Loosen and remove the safety screws (2) and (4).
• Remove the coupling.
Including K 0/120 coupling
• Tighten adjustment screw (1)(braking function).
• Hold onto the surgical microscope firmly.
• Loosen safety screw (11).
• Remove the surgical microscope.
• Hold onto the surgical microscope firmly.
• Loosen safety screw (3).
• Loosen and remove the safety screws (2) and (4).
• Remove the coupling.

Version 7.0
Page 94 G-30-1781-en
S100 / OPMI pico Preparations for use

Fig. 46: Assembling and dis-


mantling the OPMI pico 1 2 3

120° coupling

K 0/120 coupling 9 8 7 6

10 5

11

Version 7.0
G-30-1781-en Page 95
Preparations for use S100 / OPMI pico

Mounting the tube, eyepieces and objective lens

• Give securing screw (2) a few turns to loosen it.


• Remove dust cover (3) and store it in a safe place.
• Position binocular tube (4) on the surgical microscope and tighten secur-
ing screw firmly (2). The straight binocular tube is shown here as an ex-
ample.
• You can install further accessories between the binocular tube and the
microscope body. Fasten these assemblies in the same way using the
securing screw (2).
• Insert eyepieces (5) in the mounts provided on the binocular tube as far as
they will go. The magnetic coupling reliably secures them in position.
If required, use the sleeves supplied.
• Screw the objective lens (1) into the microscope body and tighten it firmly.

Fig. 47: Mounting the tube,


eyepieces and objective lens

1 2 3 4 5

Version 7.0
Page 96 G-30-1781-en
S100 / OPMI pico Preparations for use

Mounting the hand grips

Five handgrips are available for the OPMI pico surgical microscope (see
Page 58).
The handgrips can be positioned as required by reinserting (2) or pressing
them together (1).
• Select the handgrip suitable for your application.
• Mount the handgrip supplied with the instrument. Firmly tighten the
screw(s).

Fig. 48: Handgrips on the


OPMI pico 1

Version 7.0
G-30-1781-en Page 97
Preparations for use S100 / OPMI pico

Connecting the light guide

• Switch off the light source.


• Put the light guide into the light guide socket on the light source and on
the surgical microscope until it locks.
• Make sure that the light guide has been routed with sufficient slack. Sup-
port system and surgical microscope must be movable within the com-
plete area without the light guide being stretched, excessively bent or
twisted.

Fig. 49: Connecting the light


guide

Make sure that the light guide is not stretched or bent when the micro-
scope is turned or tilted.

Version 7.0
Page 98 G-30-1781-en
S100 / OPMI pico Preparations for use

Connecting an endoscope camera

If you want to equip the surgical microscope with an endoscope camera,


please observe the requirements for system expansions, see Page 251.

Prerequisites  Your OPMI pico has been equipped with the "Endoport" accessory com-
ponent:
 Your endoscope camera, including adapter, does not weigh more than
1.0 kg.
• Focus the endoscope camera at infinity.
The procedure of focusing the endoscope camera at infinity is a camera-
specific procedure and differs from camera type to camera type.
Microscope settings such as the focusing of the eyepieces do not affect
the endoscope port.
• Remove the cover from the endoscope port.
• Attach your endoscope camera to the endoscope port and allow the
quick-action lock to snap in.
• Connect the video cable of the endoscope camera to the endoscope sys-
tem. Please also observe the information contained in the user's manual
on the endoscope system.
• Adjust the orientation of the image on the video monitor. To achieve this,
turn the endoscope camera in the endoscope port as required.

Fig. 50: Connecting an endo-


scope camera

Version 7.0
G-30-1781-en Page 99
Preparations for use S100 / OPMI pico

Connect the stereo co-observer

Mounting the beam splitter


• Unscrew securing screw (3) by a few turns.
• Remove the inclined tube (1).
• Position the beam splitter (2) on the surgical microscope.
• Firmly tighten securing screw (5).
• Position the inclined tube (1) on the beam splitter (2) and tightenthe se-
curing screw (3) firmly.
Mounting the co-observation module
• Loosen knurled ring (4).
• Remove the dust cover (6) from the holding fixture of the beam splitter (2)
and store it in a safe place.
• Slide the stereo co-observation module (7) into the holding fixture. Make
sure that the locating block is fed into the counterpart of knurled ring (4).
• Firmly tighten knurled ring (4).
• Mount a tube (8) on the stereo co-observation module (7).
After having mounted the accessories, you must perform a weight compen-
sation procedure (see Page 103).

Fig. 51: Stereo co-observation


module
6
1

2
7
3
8
4

Version 7.0
Page 100 G-30-1781-en
S100 / OPMI pico Preparations for use

Mounting the MORA interface

• Give securing screw (3) a few turns to loosen it.


• Remove the dust cover and store it in a safe place.
• Place MORA interface (7) on the surgical microscope and firmly tighten se-
curing screw (3).
• Insert binocular tube (5) into the mount of MORA interface (7) and tighten
securing screw (6).
• Tilt the microscope body by approx. 25° to the left or right and lock it in
the required position using friction screw (2).
• Set ring diaphragm (4) to meet your specific requirements.

Fig. 52: Mounting the MORA 1 2


interface

5 6 7

approx. ± 25°

Version 7.0
G-30-1781-en Page 101
Preparations for use S100 / OPMI pico

Mounting documentation accessories on the MORA interface with lateral out-


put port

Requirement  The MORA interface with lateral port for documentation accessories has
been mounted on the surgical microscope.
The method described below can also be used for other accessories. The op-
erating principle of the accessories is described in the relevant Instructions for
Use.

Attaching documentation • Loosen knurled ring (2). An arrow with the labeling "open" is located on
accessories the shaft next to knurled ring (2).
• Remove dust cover (3) and store it in a safe place.
• Slide the accessories (4) in the image exit port holding fixture .
The mount of the image exit port is fitted with guide projections. Acces-
sory (4) has the corresponding grooves.
• Carefully turn the accessory module until the guide projections fit into the
grooves,
• and slide the accessory in the mount as far as it will go.
• Screw knurled ring (2) onto accessory module (3).
• Firmly tighten knurled ring (2).
If the accessory is mounted with the connector pointing vertically up-
wards, the image of the connected camera will be displayed with a 90°
rotation.
• To avoid this, you can mount the accessory rotated by 90° so that the con-
nector (incl. camera) points horizontally to the back.
• Note the effect of annular diaphragm (1) swung into the beam path.

Fig. 53: Mounting documenta-


tion accessories on the
MORA interface with lateral
output port

1 2 3 4

Version 7.0
Page 102 G-30-1781-en
S100 / OPMI pico Preparations for use

Setting the suspension arm weight compensation

NOTE Perform the weight compensation


• Perform a weight compensation (balancing) before using the device (not
above the patient and with a secure distance to other persons and devic-
es) so that the surgical microscope remains in its position within the
working area.

CAUTION Risk of crushing - mind your fingers!


Fingers may be crushed in the areas marked with the "Risk of crushing" label.
• Do not touch these areas while the system is being moved or brought in
its working / transport position.
• Do not perform balance setting until the surgical microscope and all ac-
cessories have been mounted!
• The maximum load on the suspension arm must not exceed 7 kg.
• Loosen the friction adjustment screw (1) of the suspension arm.
• Insert an 8 mm hex key into opening (3) of the articulated joint cover (2).
• Use the hex key to turn balance adjustment screw (4) until balance has
been achieved.
• Check the balance setting in several positions.
• Using the braking screw (1) adjust the required mobility of the lifting
movement of the suspension arm.

Fig. 54: Adjusting the weight


compensation (balance) on
1
the suspension arm

2
3
4

Version 7.0
G-30-1781-en Page 103
Preparations for use S100 / OPMI pico

Adjusting the surgical microscope

Adjustment instructions
– In order to be able to quickly set/adjust the surgical microscope, it is ad-
visable to save the interocular distance as well as the refraction values for
the different users so they can be preset within the scope of the prepara-
tions.
– Do not wear multifocal or varifocal glasses during the adjustment process
or when working with the microscope later on.
Such glasses make a proper diopter adjustment ring setting impossible
and result in unsatisfactory imaging.
– Optimum image clarity in the eyepieces and at the optical port for docu-
mentation equipment (photo, video) can only be achieved through careful
focusing and precisely adjusting the binocular tube.

Adjusting the binocular tube

Adjusting the interocular • Move the microscope to a vertical position above a flat object, e.g. a sheet
distance of paper with writing.
• Adjust the distance of the eyepieces at the binocular tube to your interoc-
ular distance so that the two eyepiece images (object and edge of visual
field) merge into one.

Adjusting the eyepieces The following procedure must be performed separately for each eyepiece in
the order described.
• Set the microscope to minimum magnification.
• Set the diopter adjustment ring at the eyepiece to zero position.
• Look through the eyepiece and move the microscope body to focus the
image.
• Set the microscope to maximum magnification and adjust the fine focus-
ing system until the image is sharply defined.
• Then reset the microscope to minimum magnification without changing
the working distance.
• Set the diopter adjustment ring at the eyepiece to its maximum value (plus
direction).
• Look through the eyepiece and slowly turn the diopter adjustment ring to
the minus direction until the image is sharply defined.
• Repeat the entire procedure for the second eyepiece.

Version 7.0
Page 104 G-30-1781-en
S100 / OPMI pico Preparations for use

 The microscope is now adjusted in such a way that the image is sharply
defined over the complete magnification range without any renewed fo-
cusing during the magnification setting.
If this is not the case, repeat the procedure.

Adjusting the eyecups • Adjust the eyecups of the eyepieces in such a way that the entire visual
field can be seen.
– Viewing with eyeglasses: Rotate eyecups inwards.
– Viewing without eyeglasses: Rotate eyecups outwards.

Adjusting the working distance and magnification


• Position the microscope above the surgical field so as to ensure conven-
ient work.
• Roughly focus on the object by changing the distance between
microscope and object.
• Set the microscope to maximum magnification and adjust the fine focus-
ing system until the image is sharply defined.
• Set the microscope to the required magnification. The focal plane is re-
tained in this process.

Fig. 55: Adjusting the surgical


microscope 1
1. Pupillary distance (PD)
adjustment knob
2. Eyepieces 3
3. Eyecup
4. Diopter setting ring 4

Version 7.0
G-30-1781-en Page 105
Preparations for use S100 / OPMI pico

Relocating the S100 floor stand

CAUTION Risk of crushing and toppling during transportation of the system!


If the following precautions are not observed, the floor stand may topple
over and injure persons!
• Move the floor stand into its transport position.
• To safely move the system, always hold the stand column at the position
marked with a label.

CAUTION Risk of crushing - mind your fingers!


Fingers may be crushed in the areas marked with the "Risk of crushing" label.
• Do not touch these areas while the system is being moved or brought
into its working / transport position.

Please observe the following points when relocating the suspension system:
• Turn off the device at the power switch.
• Remove the power plug from the wall socket.
• Disconnect the video cable between the video modules (e.g. video moni-
tor, video recorder) and the camera control unit (CCU).
• Adjust the flat screen (optional) to a secure transport position.
• Do not transport the device by pulling or pushing on the optional flat
screen!
• Avoid any uncontrolled swiveling of the optional flat screen.
• Swing the suspension arm into the transport position.
• Unlock all arrestors of the casters.
• Hold the system by the marked position on the stand column and carefully
push it along. Make sure that the system is positioned on level ground.
• Press at least two of the arrestors and make sure that the suspension sys-
tem is no longer able to roll away by itself.
The mass may be underestimated as the suspension system can be smoothly
moved. Therefore, move the suspension system slowly and carefully!

Version 7.0
Page 106 G-30-1781-en
S100 / OPMI pico Preparations for use

CAUTION Damaging of the suspension system!


It is possible that the suspension system is damaged due to a collision with
other devices, parts may drop down and cause personal injuries.
• Check the device for damage before each operation.
• Move the suspension system carefully and avoid collisions.
If the suspension system nevertheless collides with other devices, you
must have it checked by ZEISS Service.
Transport over long distances
• Only transport the device over long distances (e.g. relocation, return for
repair) in its original packaging or special return packaging. Please contact
your dealer or ZEISS Service.

Version 7.0
G-30-1781-en Page 107
Preparations for use S100 / OPMI pico

Connections

CAUTION Risk of electric shock caused by inadmissible or defective accessories!


Inadmissible or defective accessories may lead to increased leakage current!
• Never touch any power outlets or video ports during patient contact.
• When setting up your system, make sure to comply with the require-
ments of IEC 60601-1:2005, chapter 16; in some countries IEC 60601-1-
1:2000 may still be valid.
• Also comply with the Medical Device Operator Ordinance or other na-
tional codes.

Connecting the S100 floor stand

NOTE Risk of tripping!


Inappropriate routing of cables increases the risk of tripping.
• Always run cables in a manner that will not impede the user's move-
ments.

• Ensure that the rated voltage indicated in the display window (3) (only for
halogen/xenon) corresponds with the rated voltage for the network at the
site of the installation.
• Ensure that the unit has been switched off at the power switch before
plugging in or unplugging connectors at the power input socket (4).
• Make sure that power switch (1) is OFF.
• Plug the supplied power cord on the supply unit into the power inlet sock-
et (4).

CAUTION Risk of electric shock


Risk of electric shock for operator and patient if cables and plug connections
loosen and if housing parts are under voltage.
• Only connect the power cord using a socket with proper protective earth
connection.
• Connect the power cord via a socket to the power supply system.

Version 7.0
Page 108 G-30-1781-en
S100 / OPMI pico Preparations for use

• Secure the power cord using the strain relief device (2).
• If necessary, connect the potential equalization connector (5) of the sys-
tem with the potential equalization contact at the site of use.

Fig. 56: S100 Connecting the Halogen


support unit

Xenon
1 3

4
5

2
LED
3

2 5

Version 7.0
G-30-1781-en Page 109
Preparations for use S100 / OPMI pico

Connecting the integrated video camera

CAUTION Video images not for diagnostic purposes!


The visualized images may include deviations in scale, shape and color.
• Do not use video images for diagnostic purposes as the video camera is
not calibrated for these purposes.

Connecting the integrated HD video camera


• Switch off the device before plugging or unplugging the system cable into
the video unit. The camera may be damaged if the cable is plugged/un-
plugged while voltage is still being applied.

If you have an HD compatible S100 / OPMI pico, use an HD compatible mon-


itor that can show the resolution of the HD video camera 720P and 1080P
with 50 Hz and 60 Hz.
If you use another monitor, no video image will be shown.

• Slide the dust cover (3) on the system cable (camera side). Make sure that
the thread faces the connector plug on the camera side.
• Connect the camera head (1) of the surgical microscope to the camera
control unit (CCU) (5) using the included system cable (4).
• Screw the dust cover (3) into the thread (2).
Make sure not to screw on the dust cover (3) crookedly. This may damage
the thread!
• Firmly tighten screws on the camera head connector.

Version 7.0
Page 110 G-30-1781-en
S100 / OPMI pico Preparations for use

Fig. 57: Connecting the HD


video camera 1 2 3 4 5

Version 7.0
G-30-1781-en Page 111
Preparations for use S100 / OPMI pico

Connecting the video connections


• Connect the required video components using the video cable.
• Check the camera and monitor function. After switching the device on, a
color bar pattern is briefly displayed for checking the color setting, bright-
ness and geometry.
If the color bar pattern is permanently shown, the camera head of the vid-
eo unit is disconnected, unplugged or defective.
Examples of possible connections
1 External video monitor
2 Integrated video control unit
3 External video recorder

Fig. 58: Examples - possible


BNC (FBAS) connections 1
(for integrated HD 1
75 Ohm 75 Ohm
video camera) ON FBAS FBAS ON
OFF VIDEO IN
VIDEO IN OFF
3
75 Ohm
2 FBAS FBAS ON FBAS
VIDEO OUT
VIDEO OUT OFF VIDEO IN

2 FBAS
VIDEO OUT

In the HD video control unit the video signal is scaled down in the SD video
output (FBAS VIDEO OUT) for SD devices.

Version 7.0
Page 112 G-30-1781-en
S100 / OPMI pico Preparations for use

Fig. 59: Examples - possible


connections 1 1 1
HDMI (only with integrated
HD video camera) HDMI HDMI DVI
VIDEO IN VIDEO IN VIDEO IN

2 HDMI HDMI HDMI DVI HDMI


3 3
VIDEO OUT VIDEO OUT VIDEO IN VIDEO OUT VIDEO IN

2 HDMI 2 HDMI
VIDEO OUT VIDEO OUT

Fig. 60: Examples - possible


connections 1 1 1
DVI (only for integrated HD
video camera live & DVI HDMI DVI
streaming) VIDEO IN VIDEO IN VIDEO IN

2 DVI HDMI DVI DVI DVI


3 3
VIDEO OUT VIDEO OUT VIDEO IN VIDEO OUT VIDEO IN

2 DVI 2 DVI
VIDEO OUT VIDEO OUT

Fig. 61: Examples - possible


connections
1
YPbPr (only with integrated 1
HD video camera)
YPbPr YPbPr
VIDEO IN
VIDEO IN

2 YPbPr YPbPr YPbPr


VIDEO OUT 3
VIDEO OUT VIDEO IN

2 YPbPr
VIDEO OUT

If the video signal is to be processed by a DVI frame grabber* on the PC and


this frame grabber has no 1080p capability, the resolution of the video
camera can be reduced to 1080i.
* A frame grabber is an electronic module which can be used to capture
a video signal (in this case a digital DVI signal).

Version 7.0
G-30-1781-en Page 113
Preparations for use S100 / OPMI pico

Connecting the computer to integrated video camera live & streaming Control
unit with LAN connection

You can connect the integrated video camera live & streaming control unit
with LAN connection to a computer . Observe the conditions for system ex-
tension (see Page 251).
The network cable is available as an accessory (see Accessories).
Optional software is not included in our delivery package.
System requirements
 Computer with LAN connection: Network at least 100 Mbit
 Requirements for connecting the network computer:
– You will need administrator rights for the network computer.
– You will need the Ethernet data of the archiving PC to configure the
network share.
– You can obtain the data required for this purpose from the appropri-
ate IT administrator.
 Video software

• Connect the computer with the corresponding network cable with the
LAN connection on the video unit (on the stand).

Fig. 62: Connecting the PC


(LAN)

Version 7.0
Page 114 G-30-1781-en
S100 / OPMI pico Preparations for use

Version 7.0
G-30-1781-en Page 115
Preparations for use S100 / OPMI pico

Attachment of instrument tray and carrier arm flat screen


monitor to the S100 floor stand

Ergonomic positioning the carrier arm flat screen monitor and instrument tray

NOTE Overload of the instrument tray!


When overloaded, the instrument tray may break off or bend, the suspension
system may topple over and accessories may fall down.
• Make sure that the accessory equipment is positioned as securely as pos-
sible on the instrument tray.
• The instrument tray may be loaded with the maximum weight of 13 kg.

In order to ergonomically position carrier arm flat screen monitor (2) and in-
strument tray (3), loosen the adapter (1), adjust side and height and screw it
tight again:

Instrument tray • Hold the instrument tray with one hand and unscrew the four locking
screws at the adapter (1) with a few turns using the hex key until the
adapter (1) can easily be moved.
• Position instrument tray (3) in the required height (max. 800 mm above
the floor) and uniformly tighten (2.1 Nm) all four locking screws of
adapter (1) using the hex key.
• When attaching adapter (1), make sure the two components of the
adapter (1) are separated by the same distance (4) from top to bottom.

Carrier arm flat screen monitor • Hold the carrier arm flat screen monitor with one hand and unscrew the
four locking screws at the adapter (1) with a few turns using the hex key
until the adapter (1) can easily be moved.
• Position the carrier arm flat screen monitor (2) in the required height (max.
1230 mm above the floor) and uniformly tighten (2.1 Nm) all four locking
screws of adapter (1) using the hex key.
• When attaching adapter (1), make sure the two components of the
adapter (1) are separated by the same distance (4) from top to bottom.

Version 7.0
Page 116 G-30-1781-en
S100 / OPMI pico Preparations for use

Fig. 63: Ergonomic arrange-


ment of carrier arm flat screen
monitor and instrument tray 2
1

1
3 max.
1230 mm

max.
800 mm

1 4

4x

Version 7.0
G-30-1781-en Page 117
Preparations for use S100 / OPMI pico

Mounting the flat screen monitor on the carrier arm flat screen monitor

NOTE Overload of the carrier arm flat screen monitor!


When overloaded, the carrier arm flat screen monitor may break off or bend
or the suspension system may topple over.
• The monitor bracket may be loaded with the maximum weight of 9 kg.

Mounting the flat screen


Before mounting a flat screen on the carrier arm flat screen monitor, make
sure it meets the following specifications:
– Compliance with VESA standard (Video Electronics Standards Associa-
tion):
– 75 mm x 75 mm hole spacing or
– 100 mm x 100 mm hole spacing
– Maximum total weight of the flat screen: 9 kg
• Press flat screen (3) against the carrier arm so that the holes in the carrier
arm (2) coincide with those provided in the flat screen.
• Firmly tighten screws (1).

Cable routing in the carrier arm flat screen monitor


• Remove the clamping plate (5) from the cover plate (6) on the underside
of the carrier arm flat screen monitor.
• Place all connecting cables required for the monitor in the cable guide (4).
Leave sufficient cable loops in order to freely position the monitor without
damaging cables or plugs.
• Attach the clamping plate (5) to the cover plate (6) again.

Version 7.0
Page 118 G-30-1781-en
S100 / OPMI pico Preparations for use

Fig. 64: Mounting the flat


screen monitor on the monitor
bracket

11 2 3

1 1

4 5 6

Version 7.0
G-30-1781-en Page 119
Preparations for use S100 / OPMI pico

Routing the cables and making the electrical connections

CAUTION Risk of patient injury caused by electrical voltage!


• Do not touch the power outlet socket or other signal interfaces while in
contact with the patient.

• Route the power and data cables parallel to each other.


• Run the cables through the existing cable holders down to the stand base.
• Make sure that the cable loops (1) are large enough to allow the devices
to be positioned freely without cables or connectors being damaged. This
applies in particular to the area of the joints of the flat screen monitor
bracket.
• Route the cables in a bundle along the column down to the stand base.
• Fix the cables to the column of the floor stand using cable ties.
– To ensure optimum performance, we recommend having the electrical in-
stallation performed by the ZEISS Service or a professional electrician (in-
house electrician).
• Place cable cover (2) over the cables on the column.
• Only connect the power cord to wall outlets which are equipped with a
properly connected protective earth conductor.

Version 7.0
Page 120 G-30-1781-en
S100 / OPMI pico Preparations for use

Fig. 65: Routing the cables


and making the electrical con-
nections

Version 7.0
G-30-1781-en Page 121
Preparations for use S100 / OPMI pico

Adjusting the carrier arm flat screen monitor

After mounting the carrier arm flat screen monitor, you can adjust the friction
of its joints as required. The factory setting of the joints is suitable for a me-
dium load capacity which meets the requirements of most users. If one joint
is set too stiffly or too freely, it is possible to adjust the friction.

Adjusting the tilt function of the carrier arm flat screen monitor
• Check the tilt function of the monitor joint (upward/downward move-
ment).
• Use the tool supplied to increase or reduce the friction of the joint as re-
quired.

Fig. 66: Adjusting the tilt


function of the carrier arm flat
screen monitor

Version 7.0
Page 122 G-30-1781-en
S100 / OPMI pico Preparations for use

Adjusting the swivel movement of the carrier arm flat screen monitor
If the carrier arm flat screen monitor does not remain in place in the position
required, the following readjustments can be performed.
• Remove plastic cover (1) at the joint.
• Tighten the adjustment screw (2) clockwise until the required friction is
achieved.
• Reattach plastic cover (1).
• Check if the securing screw (3) is firmly tightened.

Fig. 67: Adjusting the swivel


movement of the carrier arm
flat screen monitor

Version 7.0
G-30-1781-en Page 123
Preparations for use S100 / OPMI pico

Balance setting of the carrier arm flat screen monitor


Only perform this adjustment if the carrier arm flat screen monitor does not
remain in the position required and moves downward.
Only perform the balancing procedure with the flat screen monitor attached.
• Move the carrier arm flat screen monitor into its horizontal position and
hold it with one hand.
• Check the carrier arm’s friction adjustment (upward/downward move-
ment).
• Increase or reduce the friction of the carrier arm flat screen monitor as re-
quired using the tools supplied.

Fig. 68: Setting the carrier


arm flat screen monitor
weight compensation

Version 7.0
Page 124 G-30-1781-en
S100 / OPMI pico Preparations for use

Preparing the system for sterile use

CAUTION Risk of infection caused by contamination!


• Only use the system with suitable, sterile accessories.

Attach asepsis caps and drapes

Attaching asepsis caps

CAUTION Risk of infection caused by contaminated asepsis accessories!


• Only use sterile asepsis accessories suitable for the device!
• Clean, disinfect and sterilize asepsis accessories before each use. This also
applies to the first use after delivery.
• Make sure that the operating staff is briefed about the sterile operation
of the device.
• Before each use of the surgical microscope, check whether the asepsis
caps are firmly attached to the surgical microscope.
For sterile use, the system can be equipped with resterilizable products. The
asepsis sets available from ZEISS contain caps and handgrips which can be
sterilized in autoclaves. For detailed information on sterilization, please refer
to the enclosed instructions "Preparation of resterilizable products" for the re-
spective asepsis sets.
Attaching the drapes

NOTE Overheating as a result of covering ventilation openings!


• Never cover any ventilation openings, e.g. with drapes! This may cause
the light source to overheat and lead to lamp failure.
To ensure sterile covering of the system, it is also possible to use the single-
use drapes.
When attaching the sterile drapes, make sure there is enough slack in the
drapes to allow for movement of the microscope carrier and surgical mi-
croscope. It is especially important that the drapes are completely loose
around the handgrips. The surgeon must be able to reliably operate the
controls through the drape.

Version 7.0
G-30-1781-en Page 125
Preparations for use S100 / OPMI pico

Version 7.0
Page 126 G-30-1781-en
S100 / OPMI pico Operation

Operation

Powering the system up and down ............................................... 128

Function test before use ................................................................129

Typical procedure ..........................................................................132

Integrated HD video camera (operating the video unit) ................. 134


Menu structure........................................................................................134
Initial start-up with integrated HD video camera......................................137
Opening the configuration menu (config) ................................................137
Opening the installation menu (config) ....................................................155
Performing the white balance (white bal) ................................................165
Setting the brightness..............................................................................167
Showing a live video image (image) .........................................................168

Integrated HD video camera live & streaming (Operating the video


unit)............................................................................................... 170
Menu structure........................................................................................170
Main menu* ............................................................................................176
Patient* ...................................................................................................178
Settings ...................................................................................................181
Camera....................................................................................................182
Recording* / Streaming............................................................................188
System ....................................................................................................190
Web interface..........................................................................................201
Installing the SW recording license ..........................................................204
Recording individual images and HD videos* ...........................................205
Viewing individual images (photos) and HD videos * in the index.............207
Deleting* individual images (photos) and HD videos in the Index .............208

Version 7.0
G-30-1781-en Page 127
Operation S100 / OPMI pico

Powering the system up and down

Prerequisites  The correct rated voltage has been set on the suspension system (please
refer to Page 108).
 The included power cord (floor stand) is connected (please refer to
Page 108).
 The light guide is connected (please refer to Page 98).
Proceed as follows to power the system up and down:
• Press the power switch (1) on the suspension system.
Before starting to work with the system, you have to perform a function test
(see next page). This permits you to check the system configuration and user-
specific settings and to reconfigure them, if necessary.
1

Version 7.0
Page 128 G-30-1781-en
S100 / OPMI pico Operation

Function test before use

NOTE Failure recognized too late!


• To avoid the occurrence of any defects during surgery, perform a system
test before every surgical procedure (without patient)!

NOTE Impairment of proper functioning due to malfunction!


If a function fails, do not use the system for safety reasons.
• Repair the cause of the failure (see "What to do in the event of malfunc-
tions"“) or contact ZEISS Service or the authorized partner.

Function test of the surgical microscope


• Before every use, make sure that all components and installed accessories
are securely locked in position. Pay attention to the maximum load capac-
ity.
• Check the unobstructed movement of the surgical microscope:
 All cables and accessories have been mounted such a way that
the surgical microscope can be positioned safely and without causing
damage.
• Check the eyepieces / binocular tube / object lens:
 The surgical microscope and the tube are in an ergonomic position.
 The interocular distance (pupillary distance) has been correctly set
(please refer to Page 104).
 The eyecups have been adjusted in such a way that you can see the
full visual field (please refer to Page 104).
 Verify that the correct prescription has been set on the diopter scale
(please refer to Page 104).
 Image quality is correct throughout the entire magnification range.
 Object lens and eyepieces are clean.
• Check the function of the magnification changer.
• Check the function of the filter selector knob.

Version 7.0
G-30-1781-en Page 129
Operation S100 / OPMI pico

Function test of the accessories


• Check the accessories:
 The securing screws of the video system cable have been firmly tight-
ened.
 The correct function of the other equipment has been checked with
the relevant user manual.

Function test of S100 suspension systems


• Check correct balancing:
 When fully balanced, the surgical microscope should remain stationary
in all positions within the work area and move neither upward nor
downward (see Page 103).
 The friction of the suspension arm, the carrier arm and the 120° cou-
pling (if used) has been set as required.
• Make sure that the securing screws have been firmly tightened (see
Page 66).
• Check the lamp brightness:
 The lamp brightness can be varied (see Page 70). Brightness variation
has an effect on the surgical field illumination.
• Check the unobstructed movement of the system (floor stand only):
 The system has been positioned in such a way that it can be moved
away from the patient at any time.
• Check the stand for secure positioning (floor stand only):
 Press at least two of the locking tabs to prevent the stand from start-
ing to move by itself.
• Check light source halogen / xenon / LED.

Version 7.0
Page 130 G-30-1781-en
S100 / OPMI pico Operation

WARNING Function deterioration of the light source


With increasing service life of the light source, the actual illumination inten-
sity decreases (this is a normal property of the system).
• Before every use, ensure that the available light quantity is sufficient.
• Have a backup light source (e.g. surgical lamp) available.
In case of a halogen light source:
• If necessary, exchange the lamp (see Page 224).
Check the set brightness after the lamp exchange.
In case of a xenon light source:
• Switch to the backup lamp.
If you already work with the backup lamp, be sure to replace the slide-in
lamp module (see Page 226). Check the set brightness after every lamp
replacement.

 The main and backup lamp are intact in the halogen light source.
 Main and backup lamp are intact in the xenon light source.
In case of exceeding the maximum service life of 500h of the main
lamp and the backup lamp is in operation, have a backup lamp module
available for replacement purposes as precaution (see Page 226). For
the order number, please see the chapter "Device data".
 With the LED light source, the amber LED is not illuminated.
 It is possible to switch on or off the halogen, xenon or LED light source
by moving up and down the suspension arm.
The light source is turned on within the preset working area.
The light source is turned off in the upper resting position.

WARNING Precautions in case of device failure


• We recommend you to take adequate precautions, depending on the ap-
plication, to enable the surgical procedure or treatment to be finished
without using this microscope, if required.

Version 7.0
G-30-1781-en Page 131
Operation S100 / OPMI pico

Typical procedure

Starting work

CAUTION Damage to the eye caused by high light intensity!


During surgical procedures in the facial area, no light must enter the patient's
eye.
• Make sure to protect the patient's eye from light radiation.
• During surgical procedures in the facial area, the patients must wear tint-
ed UV protective eyewear (ordering data see Page 251).

CAUTION Danger of burnings


Patient may be burnt if the microscope is unattended targeted to the patient
with the light source switched on.
• Never leave the device unattended with the light source switched on.

 All preparations for use have been completed.


• Check the system for proper function. The process is described in chapter
"Operation - Function test before use".
• Turn on the device with the power switch of the suspension system.
• Select the filter suitable for your application in the illumination beam path.
• Adjust the brightness on the suspension system. Start with the lowest
brightness and slowly increase it to the required but not more than the al-
lowed brightness.
• When focusing, make sure to carefully lower the surgical microscope and
observe the working distance from the patient. The working distance ap-
proximately corresponds to the focal length of the object lens used.
• For focusing, move the surgical microscope over the surgical field into an
ergonomic position within the working distance.
• Select the magnification required using the changer button.
• Look through the eyepieces and lower the surgical microscope until the
surgical field comes into focus. This is to achieve focusing. Ensure a suffi-
cient safety distance to the patient.
• Look through the eyepieces of the binocular tube. Adjust the eyepieces in
such a way that you can see both the edge of the visual field and the mi-
croscope image sharply. Also refer to "Adjusting the surgical microscope".

Version 7.0
Page 132 G-30-1781-en
S100 / OPMI pico Operation

Ending work

NOTE Risk of injury caused by suspended systems!


Parts of the suspension system are positioned at head level and pose a risk
of injury.
• Move the system into the standby position after use and make sure that
sufficient headroom is available.
• Move the suspension arm into its upper rest position.
• Switch off the system when not in use.

Version 7.0
G-30-1781-en Page 133
Operation S100 / OPMI pico

Integrated HD video camera (operating the video unit)

Test pattern
A test pattern (color bars) is displayed if the camera head is not connected
when you switch on the system.
• Switch off the system again, connect the camera head and switch the sys-
tem back on.

Menu structure

Menu display
– The menu text is displayed on the monitor in light green.
A light green arrow is displayed in front of a selected line.

Navigation using the control keys


Control keys <select> (cursor key)
– The control keys allow you to switch between the lines. The selected
line is displayed in green.
– The control keys allow you to change the settings.
<Enter> key
Use this key to access the corresponding submenu. The current setting, which
is displayed in green, is selected in this submenu. Text information is displayed
in the various submenus.

Version 7.0
Page 134 G-30-1781-en
S100 / OPMI pico Operation

Fig. 69: Menu structure


Configuration menu

* In case the "OPHTHALMOLOGY" application is used, the menu item


"FUNDUS SYSTEM" is also displayed here.

CAUTION Not for ophthalmic examinations!


• The S100 / OPMI pico must not be used for ophthalmological procedures!

Version 7.0
G-30-1781-en Page 135
Operation S100 / OPMI pico

In the installation menu you can install device settings that are seldom
changed and that require the camera to be restarted.
• Press the <config> key for approximately 5 seconds.
 This will take you to the following menu structure:

Fig. 70: Menu structure


Installation menu

Version 7.0
Page 136 G-30-1781-en
S100 / OPMI pico Operation

Initial start-up with integrated HD video camera

During the initial start of the surgical microscope with integrated HD video
camera, the sub menu "APPLICATION VARIANT" will be displayed for the first
time (see Page 163).
The application is set in this menu.
ZEISS Service adjusts this setting for you during initial start-up. You can read-
just this setting in the "INSTALLATION MENU". see “Opening the installation
menu (config)” on Page 155.

Opening the configuration menu (config)

The configuration menu allows you to make user-specific settings, such as se-
lecting a user or setting the brightness / color saturation.
• Press the <config> key to go to the configuration menu.
 The following menu appears on the monitor:

USER MENU STANDARD

 USER User 1 - 4
BRIGHTNESS AUTOMATIC - LOW AUTO -
MANUAL
LIGHTMETER SMALL SPOT - LARGE SPOT -
INTEGRAL
ORIENTATION STANDARD - MIRROR - FLIP -
ROTATE
ADVANCED ...
SERVICE MENU ...

EXIT

• Use the control keys to select a function.


 The selected function is displayed with an arrow.
• Use the control keys to change the settings.
If you need more information on a menu item, you can open the corre-
sponding sub menu by pressing the <enter> key. This sub menu provides
additional information on the setting options. The current setting is high-
lighted as the initial setting in this sub menu.

Version 7.0
G-30-1781-en Page 137
Operation S100 / OPMI pico

• You can exit the configuration menu by selecting menu item "Exit" or by
pressing the <config> or <live> key. If changes have been made, a prompt
is displayed for each user on exit from the configuration menu as to
whether the settings should be saved. Answer this prompt with "YES" or
"NO".
An existing "freeze" state ends when the configuration menu is activated.

Version 7.0
Page 138 G-30-1781-en
S100 / OPMI pico Operation

User sub menu


You can select among four different user profiles in the "USER SELECTION"
submenu: USER 1, USER 2, USER 3, USER 4
The user names and their configuration depend on the selected application.
The latter can be selected as required when switching on the device for the
first time and can be subsequently changed at any time.

CAUTION Not for ophthalmic examinations!


• The S100 / OPMI pico must not be used for ophthalmological procedures!

– The "OPHTHALMOLOGY" application has the "POSTERIOR" userinstead


of USER 1 and the "ANTERIOR" user instead of USER 2.
– Instead of USER 1, the "DENTRISTRY" application has user
"ENDODONTY" and user "TREATMENT" instead of USER 2.
– The Anterior, Posterior, Endodonty and Treatment user have optimized
settings.
– The video image is displayed in its unaltered condition (standard) for the
Anterior user.
– The video image is rotated 180°(ROTATE) for the Posterior user.
– The video image is displayed in its unaltered condition (standard) for the
Endodonty user.
– The video image is displayed in its unaltered condition (standard) for the
Treatment user.
The white balance is user-specific. This allows User 1, for example, to save a
different white balance than User 2. When changing users, the white balance
value is loaded for the other user. This requires that the light source is also
changed because, otherwise, the colors would be distorted. Therefore, white
balancing should always be carried out with the light source to be used for
the particular user.

Version 7.0
G-30-1781-en Page 139
Operation S100 / OPMI pico

Proceed as follows to select a user and to load and activate the respective con-
figuration:
• Press the <config> key to access the configuration menu.
• In the Configuration menu, select the menu item "User" with the con-
trol keys and press the <enter> key.
 The following menu appears:

USER SELECTION

 USER 1
USER 2
USER 3
USER 4

CANCEL

LOAD USER 1 SETTINGS

• Select the required user with the control keys.


 The selected user is displayed with an arrow.
• Press the <enter> key to temporarily save the user settings.
 The device switches back to the configuration menu. The user settings
are finally saved after you exit the configuration menu. If changes have
been made, a prompt is displayed for each user as to whether the set-
tings should be saved. Answer this prompt with "YES" or "NO".
• If you do not want to save any changes to the user, press:
– the <cancel> key to return to the configuration menu,
– the <config> or <live> key to return to live image mode.

Version 7.0
Page 140 G-30-1781-en
S100 / OPMI pico Operation

Brightness submenu
In the "BRIGHTNESS" sub menu, you can select between three different expo-
sure controls: AUTOMATIC - LOW LIGHT AUTO - MANUAL

AUTOMATIC In this mode the camera adjusts the brightness of the video image automati-
cally to the value specified by the <brightness> keys and . Exposure con-
trol affects the exposure time of the camera.
The nominal brightness is set in the AUTOMATIC mode. The camera controls
the exposure time in the range between 1/50(60)s and 1/10000s. The expo-
sure time currently selected is not displayed on the monitor. The nominal
brightness can be adjusted between 0 and 100%. Changes are made in 5%
increments.
A required value in the range between 50% and 70% usually provides op-
timum exposure results.

LOW AUTO This mode has the same function as AUTOMATIC mode, however, exposure
times longer than 1/50 (60)s are allowed, which can result in blurred move-
ments.

MANUAL In this mode the exposure time selected is displayed when the control
keys are pressed. The setting range extends from 1/6 to 1/10000.
– Increments for PAL: 1/10000, 1/4000, 1/2000, 1/1000, 1/500, 1/250, 1/
100, 1/50, 1/25, 1/12 and 1/6
– Increments for NTSC: 1/10000, 1/4000, 1/2000, 1/1000, 1/500, 1/250, 1/
120, 1/60, 1/30, 1/15 and 1/8
Greater exposure times can result in blurred movements.
The set time disappears from the screen again if the <brightness> key or
was not pressed longer than 2 seconds.

Version 7.0
G-30-1781-en Page 141
Operation S100 / OPMI pico

Proceed as follows to select an exposure control and to load the respective


configuration:
• Press the <config> key to access the configuration menu.
• Use the control keys to select the "BRIGHTNESS" menu item in the con-
figuration menu and press the <enter> key.
 The following menu appears:

BRIGHTNESS MODE

 AUTOMATIC 80%*
LOW LIGHT AUTO 80%*
MANUAL 1/16*

CANCEL

SET POINT FOR AUTOMATIC BRIGHTNESS REGULATION

The values marked with * are only examples.


• Select the desired mode with the control keys.
 The selected choice is displayed with an arrow.
 The selected mode is explained at the lower end of the display.
• You can use the control keys to change the settings.
• Press the <enter> key to save the changes.
 The selection is saved and you return to the configuration menu.
• If you do not want to save your changes, return to the configuration menu
via the "CANCEL" menu item. The mode is set as it was when the sub
menu was accessed.

Version 7.0
Page 142 G-30-1781-en
S100 / OPMI pico Operation

Lightmeter sub menu


In the "LIGHTMETER" sub menu, you can select between three different types
of exposure measurements: INTEGRAL, SMALL SPOT or LARGE SPOT.
The Lightmeter function is only enabled if Brightness is set to the AUTO-
MATIC" or "LOW LIGHT AUTO" mode.

INTEGRAL: The exposure is measured and averaged across the full video image. This
measurement is recommended for surgical fields that are fully and evenly lit.

LARGE SPOT: The exposure is measured in an area in the image center. This measurement
is ideal for working with an eclipsed surgical field edge (resulting from a re-
duced illuminated field diameter).

SMALL SPOT: The exposure is measured in a very small area in the image center. This meas-
urement is suitable for work in narrow channels when an extremely small illu-
minated field diameter is being used.
However, if the object of interest is not located in the image center, the
SMALL SPOT setting usually does not provide the desired exposure result. In
this case, select either a larger area (INTEGRAL or LARGE SPOT) or use the
manual exposure mode.

Fig. 71: Types of


exposure measurement

INTEGRAL LARGE SPOT SMALL SPOT

Version 7.0
G-30-1781-en Page 143
Operation S100 / OPMI pico

Proceed as follows to select a exposure measurement type and to load the re-
spective configuration:
• Press the <config> key to access the configuration menu.
• Select the menu item "LIGHTMETER" in the configuration menu using the
control keys and press the <enter> key.
 The following menu appears:

LIGHTMETER

 SMALL SPOT
LARGE SPOT
INTEGRAL

CANCEL

USE ENTIRE SCENE AVERAGE FOR BRIGHTNESS METERING

• Select the desired mode with the control keys.


 The selected choice is displayed with an arrow.
 The selected mode is explained at the lower end of the display.
• Press the <enter> key to save the changes.
 The selection is saved and you return to the configuration menu.
• If you do not want to save your changes, return to the configuration menu
via the "CANCEL" menu item. The exposure type is set as it was when the
sub menu was accessed.

Version 7.0
Page 144 G-30-1781-en
S100 / OPMI pico Operation

Orientation submenu
The "ORIENTATION" submenu allows setting the screen orientation. To do
this, proceed as follows:
• Press the "config" key to access the configuration menu.
• In the configuration menu, use the control keys to select menu item
"ORIENTATION" and press the "enter" key.
 The following menu appears:

ORIENTATION

 STANDARD Standard
MIRROR Mirror the image vertically*
FLIP Mirror the image horizontally*
ROTATE Turn the image through 180°*

CANCEL

STANDARD VIEW - NO ROTATION

The comments marked with * are only provided for information purposes
and do not appear on the monitor.
• Use the control keys to select the required mode.
 The selection is displayed with an arrow.
• Press the "enter" key to save the changes.
 The selection is saved and you return to the configuration menu.
• If you do not want to save any changes, select menu item "CANCEL" to
return to the configuration menu. As a result, the image orientation on the
screen will be reset to the setting present when the submenu was opened.

Version 7.0
G-30-1781-en Page 145
Operation S100 / OPMI pico

Advanced submenu
In the "ADVANCED" submenu, you can set the settings that are seldom used.
Proceed as follows to make changes:
• Press the <config> key to access the configuration menu.
• Select the menu item "ADVANCED" in the Configuration menu using the
control keys and press the <enter> key.
 The following menu appears:

USER MENU ADVANCED

 RED ±0
BLUE ±0
CHROMA ±0
COLOR ANGLE ±0
BLACK LEVEL ±0
ZOOM LEVEL 1.0 x
PEAK/AVERAGE 0

BACK

Use the control keys to select a function.


 The selected function is displayed with an arrow.
• Pressing the <enter> key displays the corresponding submenu, which
sometimes contains informational text. The current setting is selected as
the initial setting in the respective submenu.
The "ADVANCED" submenu is exited using the menu item <BACK>.

Version 7.0
Page 146 G-30-1781-en
S100 / OPMI pico Operation

Recommended color setting


The depiction of the color on the monitor is dependent on the individual
viewer's preferences.
Therefore, 2 standard color settings are outlined here:
– More subtle color impression
– Stronger color impression (preferred, e.g. in Asian part of the world)

Color perception Stronger color im-


subtle pression
 RED ±0 +1
BLUE +1 +3
CHROMA +3 +7
COLOR ANGLE ±0 ±0
BLACK LEVEL ±0 ±0
ZOOM LEVEL 1.0 x 1.0 x
PEAK/AVERAGE 6 6
The shown color settings were ascertained using a brightness setting of 55%,
(see ).
The color settings were ascertained on an 24" NDS ENDOVUE monitor with
factory settings.

Description of the setting options

CHROMA The camera's color saturation is intensified or reduced using the "Chroma"
setting.

COLOR ANGLE With the "Color Angle" setting (frequently also: HUE), you can change the
color angle as well as the hues of the camera.

BLACK LEVEL You can set the blackness value of the camera with the "Black Level" setting.
Reducing the black level generates a higher degree of contrast. Increasing a
value means that dark details are easier to recognize.

ZOOM LEVEL The camera image can be magnified electronically using the "Zoom Level" set-
ting.

Version 7.0
G-30-1781-en Page 147
Operation S100 / OPMI pico

PEAK / AVERAGE The setting "Peak / Average" is used to modify the lighting control character-
istics. A lower value results in peak value-based lighting control, which re-
sponds precisely to changes in the project brightness. A higher value sets
mean value-based lighting control, which responds less sensitively to changes
in brightness.

Executing setting options

RED submenu In this submenu, you can set the red value of the camera in a range from -20
to +20. To do this, proceed as follows:
• Select menu item "RED" in the "ADVANCED" menu using the control
keys and press the "enter" key.
• Select the "SETTINGS" menu item with the control keys.
 The selected choice is displayed with an arrow.
• Change the red value using the control keys.
• Press the "enter" key to save the changes.
 The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
• If you do not want to save your changes, return to the "ADVANCED USER
SETTINGS" menu via "CANCEL". The red value is set as it was when the
submenu was accessed.
The red value setting also influences the camera's white balance. see “Per-
forming the white balance (white bal)” on Page 165.

BLUE submenu In this submenu, you can set the blue value of the camera in a range from
-20 to +20. To do this, proceed as follows:
• Select the "BLUE" menu item in the "USER MENU ADVANCED" menu us-
ing the control keys and press the "enter" key.
• Select the "SETTINGS" menu item with the control keys.
 The selected choice is displayed with an arrow.
• Change the blue value using the control keys.
• Press the "enter" key to save the changes.
 The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
• If you do not want to save your changes, return to the "ADVANCED USER
SETTINGS" menu via "CANCEL". The blue value is set as it was when the
submenu was accessed.

Version 7.0
Page 148 G-30-1781-en
S100 / OPMI pico Operation

The blue value setting also influences the camera's white balance. see “Per-
forming the white balance (white bal)” on Page 165.

CHROMA submenu In this submenu, you can set the color saturation of the camera in a range
from -10 to +10. To do this, proceed as follows:
• Select the "CHROMA" menu item in the "USER MENU ADVANCED" menu
using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
 The selection is displayed with an arrow.
• Change the color saturation using the control keys.
A chroma setting in the range of ± 0 usually provides the best color im-
pression. Chroma setting is most effective with strongly saturated (bright)
colors. The effect is low with weakly saturated colors.
• Press the "enter" key to save the changes.
 The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". As a result, color saturation will be reset to the
setting present when the submenu was opened.

COLOR ANGLE submenu In this submenu, you can set the color angle (hue) of the camera in a range
from -10 to +10. To do this, proceed as follows:
• Select the "COLOR ANGLE" menu item in the "USER MENU ADVANCED"
menu using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
 The selection is displayed with an arrow.
• Change the color angle using the control keys.
• Press the "enter" key to save the changes.
 The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". As a result, the color angle will be reset to the
setting present when the submenu was opened.

BLACK LEVEL submenu This submenu allows setting the black level of the camera in a range from -10
to +10. To do this, proceed as follows:

Version 7.0
G-30-1781-en Page 149
Operation S100 / OPMI pico

• Select the "BLACK LEVEL" menu item in the "USER MENU ADVANCED"
menu using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
 The selection is displayed with an arrow.
• Change the black level value using the control keys.
• Press the "enter" key to save the changes.
 The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". The black level is set as it was when the submenu
was accessed.

ZOOM LEVEL submenu In this submenu, you can set the zoom level of the camera in a range from
1.0x to 2.0x. To do this, proceed as follows:
• Select the "ZOOM LEVEL" menu item in the "USER MENU ADVANCED"
menu using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
 The selection is displayed with an arrow.
• Change the zoom level using the control keys.
• Press the "enter" key to save the changes.
 The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". The zoom level is set as it was when the sub-
menu was accessed.

PEAK / AVERAGE submenu In this submenu, you can set the exposure control characteristics of the
camera in a range from 0 to +8. To do this, proceed as follows:
• Select menu item "PEAK / AVERAGE" from the "ADVANCED USER SET-
TINGS" menu using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
 The selection is displayed with an arrow.
• Change the value using the control keys.

Version 7.0
Page 150 G-30-1781-en
S100 / OPMI pico Operation

• Press the "enter" key to save the changes.


 The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". The zoom level is set as it was when the sub-
menu was accessed.

Version 7.0
G-30-1781-en Page 151
Operation S100 / OPMI pico

SERVICE submenu
You can set the settings that are seldom used in the "SERVICE MENU" sub-
menu. Proceed as follows to make changes:
• Press the <config> key to access the configuration menu.
• In the Configuration menu, use the control keys to select the menu
item "SERVICE MENU" and press <enter>.
 The following menu appears:

SERVICE MENU

 REMOTE SETUP ...


FACTORY SETTINGS ...
VERSION INFO ...

BACK TO CONFIG

• Use the control keys to select a function.


 The selected function is displayed with an arrow.
• Pressing the <enter> key displays the corresponding submenu, which
sometimes contains informational text. The current setting is selected as
the initial setting in the respective submenu.
• You can exit submenus with menu item <CANCEL>. The changes made
are not saved in this case.

REMOTE SETUP You can configure the device in this sub menu so that a selected function
(please refer to REMOTE FUNCTION) is controllable via a foot switch or a
fundus imaging system. The sub menu contains the following areas:
– REMOTE FUNCTION
– SWITCHING MODE
– AUDIO CONFIRMATION
You can assign one of the functions described below to the integrated HD
video camera under "REMOTE FUNCTION". This function is then triggered
when the foot switch is activated or by swiveling in and out of a fundus im-
aging system.
The following options are available:

Version 7.0
Page 152 G-30-1781-en
S100 / OPMI pico Operation

– FREEZE / LIVE
If the current live video image is displayed and the foot switch is pressed,
the live video image is frozen and displayed on the monitor as a freeze
frame. With renewed pressing of the foot switch the live video image is
displayed again.
– USER1 / USER2
(or Endodonty/Treatment or Anterior/Posterior)
You can switch between user 1 and user 2 by pressing the foot switch. If
the current user is neither user 1 nor user 2, user 1 is switched on with the
first activation of the foot switch.
– USER3 / USER4
You can switch between user 3 and user 4 by pressing the foot switch. If
the current user is neither user 3 nor user 4, user 1 is switched on with the
first activation of the foot switch.
– INVERTED / STANDARD (is only required with a special tube)
If the current live video image is displayed and the foot switch is pressed,
the live video image is frozen and displayed on the monitor as a freeze
frame. With renewed pressing of the foot switch the live video image is
displayed again.
You can select under "SWITCHING MODE" if a remote function shall be con-
trolled via a foot switch or a fundus imaging system. The following control
types are available:
– FOOTSWITCH / AUX
Function control is performed via a foot switch.
– FUNDUS SYSTEM
The function is controlled via a fundus imaging system. This setting is only
available under the "OPHTHALMOLOGY" application.

CAUTION Not for ophthalmic examinations!


• The S100 / OPMI pico must not be used for ophthalmological procedures!

Version 7.0
G-30-1781-en Page 153
Operation S100 / OPMI pico

You can select under "AUDIO CONFIRMATION" if there shall be an acoustic


feedback (signal tone) when pressing the foot switch.
– Enabled
Acoustic return signal is activated.
– Disabled
Acoustic return signal is deactivated.

FACTORY SETTINGS This submenu allows resetting the user and camera settings to the factory set-
tings. To do this, proceed as follows:
• In the "SERVICE ", use the control keys to select menu item "FACTORY
SETTINGS" and press the "enter" key.
• Use the control keys to select one of the following items:
– RESET USER 1
– RESET ALL USERS
– RESET CAMERA
 After selection of one of the menu items, a message informs you that
the stored data will be overwritten.
• Press the "enter" key to overwrite the settings.
 The system returns to the SERVICE MENU after approx. 2 seconds.
• If you do not want to save any changes, select menu item "CANCEL" to
return to the SERVICE MENU.

VERSION INFORMATION This submenu shows the current versions of the hardware and software com-
ponents as well as the serial numbers of the camera and CCU. Proceed as fol-
lows:
• Select menu item "VERSION INFORMATION" and press the "enter" key.
The following information is displayed:
– SOFTWARE (software version of the CCU)
– S/N CCU (serial number of the CCU)
• Press the "enter" key to return to the SERVICE MENU.

Version 7.0
Page 154 G-30-1781-en
S100 / OPMI pico Operation

Opening the installation menu (config)

You can use the installation menu to make device settings that are seldom
changed and that require a restart of the camera.
• Press the <config> key for approx. 5 seconds to access the installation
menu.
 The following menu appears on the monitor:

INSTALLATION MENU

 HDTV MODE 720P - 1080P - 1080I


DVI/HDMI OUT PC MODE - VIDEO MODE
SDTV FORMAT LETTERBOX - SIDECUT
PAL / NTSC PAL - NTSC
LIGHT SOURCE XENON/LED HALOGEN
APPLICATION OPHTHALMO - DENTISTRY -
NEURO/SPINE - ENT
EXIT
CHANGES WILL TAKE PLACE AFTER RESTARTING THE CAMERA

• Use the control keys to select a function.


 The selected function is displayed with an arrow.
• Pressing the <enter> key displays the corresponding submenu, which
sometimes contains informational text. The current setting is selected as
the initial setting in the respective submenu.
You can use the "CANCEL" menu item to exit a submenu. The changes made
are not saved in this case.

CAUTION Not for ophthalmic examinations!


• The S100 / OPMI pico must not be used for ophthalmological procedures!

Version 7.0
G-30-1781-en Page 155
Operation S100 / OPMI pico

HDTV sub menu


The "HDTV MODE" sub menu allows you to select the video camera's input
signal and adjust it to your display device.

720P Select this setting format if your display device is 720p compatible. This mode
is similar to the 1080p progressive standard, but its maximum resolution is
only 1280 x 720 pixels.

1080P* Select this setting if your display device is 1080p compatible. The image is set
up using real full frames. This prevents any interline flicker, and the image is
sharply defined and stable. The maximum resolution is 1920 x 1080 pixels.

1080I* Use the 1080i (interlace) mode if your display device supports the 1080i
standard. The image is set up by two consecutive half images. The delayed
image setup can cause slight interline flicker. The maximum resolution is 1920
x 1080 pixels.
As a requirement, you should have connected the HD compatible OPMI pico
with a 1080p version and a 1080p or 1080i compatible display device. We
recommend the use of a 1080p compatible display device.

Version 7.0
Page 156 G-30-1781-en
S100 / OPMI pico Operation

Proceed as follows to select a mode and to load the respective configuration:


• Press the <config> key for approx. 5 seconds to access the installation
menu.
• Select the menu item "HDTV MODE" in the installation menu using the
control keys and press the <enter> key.
 The following menu appears:

HDTV MODE

 720P
1080P*
1080I*

CANCEL

DVI/HDMI -> 720P - 1080P* - 1080I*


YPbPr -> 720P - 1080I*

*only in case of an optional 1080p version


In order to use the optional 1080p resolution, a 1080p compatible display de-
vice must be connected.
• Select the desired mode with the control keys.
 The selected choice is displayed with an arrow.
 At the lower end of the screen, it is explained which video signal is at
which video output.
• Press the <enter> key to save the changes.
 If no 1080p compatible display device is connected, the display re-
mains black and the HDTV sub menu is displayed again after approx.
30 seconds. No changes have been made.
 If a 1080p compatible display device is connected, the prompt "APPLY
CHANGES" appears.
• Select YES or NO with the control keys and confirm the selection with
the <enter> key.
 When you chose "YES", the camera resolution is changed and you go
back to the installation menu.

Version 7.0
G-30-1781-en Page 157
Operation S100 / OPMI pico

 When you chose "NO", the camera resolution is not changed and you
go back to the installation menu.
• If you do not want to save your changes, return to the installation menu
via "CANCEL". The video signal is set as it was when the sub menu was
accessed.

Version 7.0
Page 158 G-30-1781-en
S100 / OPMI pico Operation

DVI/HDMI OUT submenu


The "DVI/HDMI OUTPUT" allows switching between video formats "PC
MODE" and "VIDEO MODE" when a high-resolution display device (monitor,
projector, etc.) is connected. To do this, proceed as follows:
• Press the "config" key for approx. 5 seconds to access the installation
menu.
• Select menu item "DVI/HDMI OUT" in the installation menu using the
control keys and press the "enter" key.
 The following menu appears

DVI/HDMI OUTPUT RANGE

 PC MODE (255 DEFAULT)


VIDEO MODE (235)

CANCEL

ADJUST DVI/HDMI OUTPUT FOR SPECIAL HDMI MONITOR SIGNAL

• Use the control keys to select the required mode.


 The selection is displayed with an arrow.
 The properties of each format are explained at the bottom of the
screen.
• Press the "enter" key to save the changes.
 The selection is saved and you return to the installation menu.
• If you do not want to save any changes, select menu item "CANCEL" to
return to the installation menu. The video format is set as it was when the
submenu was accessed.
You can find the correct setting for your connected monitor or TV in the user
manual pertaining to the device.
If you cannot find any information on the right setting, we recommend the PC
MODE (default) setting for display devices with DVI port. We recommend the
VIDEO MODE setting for display devices with HDMI port.

Version 7.0
G-30-1781-en Page 159
Operation S100 / OPMI pico

SDTV FORMAT submenu


If a conventional display device (monitor, projector, etc.) with 4:3 format is
connected, the "SDTV FORMAT" submenu allows you to choose between
"LETTERBOX" and "SIDE-CUT" visualization.

LETTERBOX Use this setting for display devices that are not compatible with the 16:9 wide
screen format. A black bar will appear at the top and bottom of the screen
and the image will be displayed in the 16:9 format.

SIDE-CUT In this mode, the sides of the image will be "cut", leaving the center of the
image to be displayed with a 4:3 aspect ratio.
Proceed as follows to select a format and to load the respective configuration:
• Press the "config" key for approx. 5 seconds to access the installation
menu.
• Select menu item "SDTV FORMAT MODE" in the installation menu using
the control keys and press the "enter" key.
 The following menu appears:

SDTV FORMAT

 LETTERBOX
SIDE-CUT

CANCEL

DISPLAY VIDEO PICTURE WITH BLACK STRIPES AT TOP AND BOTTOM

• Use the control keys to select the required mode.


 The selection is displayed with an arrow.
 The properties of each format are explained at the bottom of the
screen.
• Press the "enter" key to save the changes.
 The selection is saved. The camera restarts after you have confirmed
that the setting is to be applied.
• If you do not want to save any changes, select menu item "CANCEL" to
return to the installation menu. As a result, the image display on the
screen will be reset to the setting present when the submenu was opened.

Version 7.0
Page 160 G-30-1781-en
S100 / OPMI pico Operation

PAL / NTSC Selection submenu


In the "PAL/NTSC" submenu, you can switch between PAL and NTSC format.
Proceed as follows to switch on the PAL / NTSC SELECTION:
• Press the "config" key for approx. 5 seconds to access the installation
menu.
• Select menu item "PAL / NTSC SELECTION" in the installation menu using
the control keys and press the "enter" key.
 The following menu appears:

PAL / NTSC SELECTION

 PAL
NTSC

CANCEL

• Use the control keys to select the required mode.


 The selection is displayed with an arrow.
 The properties of each format are explained at the bottom of the
screen.
• Press the "enter" key to save the changes.
 The selection is saved and you return to the installation menu.
• If you do not want to save any changes, select menu item "CANCEL" to
return to the installation menu. As a result, the image display on the
screen will be reset to the setting present when the submenu was opened.
After you have saved the changes, the settings are applied and the camera is
restarted.

Version 7.0
G-30-1781-en Page 161
Operation S100 / OPMI pico

LIGHT SOURCE
In the "LIGHT SOURCE" submenu, you can select between XENON / LED and
HALOGEN.
• Press the "config" key for approx. 5 seconds to access the installation
menu.
• Select menu item "LIGHT SOURCE" from the installation menu using the
control keys and press the "enter" key.
 The following menu appears:

LIGHT SOURCE

 XENON / LED
HALOGEN

CANCEL

SETUP CAMERA FOR XENON OR LED LIGHT SOURCES

• Use the control keys to select the required mode.


 The selection is displayed with an arrow.
 The properties of each format are explained at the bottom of the
screen.
• Press the "enter" key to save the changes.
 The selection is saved and you return to the installation menu.
• If you do not want to save any changes, select menu item "CANCEL" to
return to the installation menu. As a result, the image display on the
screen will be reset to the setting present when the submenu was opened.

Version 7.0
Page 162 G-30-1781-en
S100 / OPMI pico Operation

Application submenu
In the "APPLICATION" submenu, you can select between five different appli-
cations. Each application affects the user profile and the associated camera
settings.
The application selection is displayed when the device is switched on for the
first time and can be changed at any time. The selected application is dis-
played briefly during start-up. The following applications can be selected:

DENTISTRY This application is designed for displaying the best image in the lower or upper
tooth segment. The "ENDODONTY" and "TREATMENT" users have optimized
and defined settings for the brightness control. This is why the "BRIGHTNESS"
and "LIGHTMETER" menu items are not available to these users.

NEURO/SPINE This application is designed for displaying the image in the neurological area.
The difference refers to the camera settings as well as to the not selectable
image orientation.

ENT This application id for the imaging of the neck, nose, ear area. The differences
can be found in regard of the camera settings and the adjusted image section.

OPHTHALMOLOGY This application is designed for displaying the image in the ophthalmological
area. The "ANTERIOR" and "POSTERIOR" users are available for displaying the
best image in the anterior or posterior eye segment. Both users have fixed set-
tings for the brightness control. This is why the "BRIGHTNESS" and "LIGHT-
METER" menu items are not available to these users.

CAUTION Not for ophthalmic examinations!


• The S100 / OPMI pico must not be used for ophthalmological procedures!

Version 7.0
G-30-1781-en Page 163
Operation S100 / OPMI pico

Proceed as follows to select an application variant and to load the respective


configuration:
• Press the <config> key for approx. 5 seconds to go to the installation
menu.
• Select the menu item "APPLICATION" in the installation menu using the
control keys and press the <enter> key.
 The following menu appears:

APPLICATION VARIANT

OPHTHALMOLOGY
 DENTISTRY
NEURO/SPINE
ENT

CANCEL

USER SETTINGS WILL BE RESET TO DEFAULTS !

• Select the desired mode with the control keys.


 The selected choice is displayed with an arrow.
• Press the <enter> key to save the changes.
 The selection is saved and activated the next time the device is
switched on.
 You then return to the installation menu.
• If you do not want to save any changes, return to the installation menu
via the "CANCEL" menu item. The application variant is then set back to it
was when the submenu was accessed.

Version 7.0
Page 164 G-30-1781-en
S100 / OPMI pico Operation

Performing the white balance (white bal)

In the white balancing procedure, the system adjusts the video signal in such
a way that white areas in the surgical field are also white on the video image.
• Press the <start> key to activate the white balance procedure.
 The following message is displayed:
"PLEASE PUT A PIECE OF WHITE PAPER UNDER THE MICROSCOPE"
• Place a suitable white object (e. g., a piece of white paper with a matte
surface) under the surgical microscope.
• Carry out the white balance procedure using a medium brightness level.
• Press the <start> key to start the white balance procedure.
During the white balance procedure (display: "WHITE BALANCE IN
PROGRESS") the key functions are locked.
If the white balance procedure is successfully completed, the following will be
displayed: "WHITE BALANCE SUCCESSFUL". Otherwise, the following mes-
sage is displayed on the monitor: "WHITE BALANCE FAILED". Check the set-
tings described above and repeat the white balance procedure.
The white balance is saved for the user currently selected. When the user is
changed, the white balance values pertaining to that user are loaded.
After the white balance procedure is completed, all keys are enabled again.
The determined values are saved and are available to all users each time the
device is restarted.

White balance is adjusted corresponding to the set light type. The adjustment
depends on the "LIGHT SOURCE" setting in the "INSTALLATION MENU" and
the BLUE and RED settings in the "USER MENU ADVANCED".

Version 7.0
G-30-1781-en Page 165
Operation S100 / OPMI pico

The RED and BLUE values have an influence on the camera colors even after
the white balance procedure.
Recommended settings for a halogen light source
"LIGHT SOURCE" - "HALOGEN"

RED BLUE
Adjustment for a weak to medium illumi- +4 -8
nation (default)
Adjustment for a bright halogen illumina- +2 -4
tion
Without halogen light adjustment 0 0

Recommended settings for a xenon or LED light source


"LIGHT SOURCE" - "XENON/LED"

RED BLUE
No adjustment needed (every illumina- 0 0
tion level)

Version 7.0
Page 166 G-30-1781-en
S100 / OPMI pico Operation

Setting the brightness

Under "brightness" you can adjust the brightness of the video image in 5%
increments between 0% and 100% using the and keys. The value set
appears in the upper left corner of the monitor.
In the configuration menu, you can select between three different modes in
the brightness menu: AUTO - LOW AUTO - MANUAL.

AUTO
In AUTO mode the camera adjusts the brightness of the video image automat-
ically to the value specified by the and control keys. Exposure control
affects the exposure time of the camera.
The nominal brightness is set in the AUTO mode. The camera controls the ex-
posure time in the range between 1/50(60)s and 1/10000s accordingly. The
nominal brightness can be adjusted between 0 and 100%. Changes are made
in 5% increments.
A required value in the range between 50% and 70% usually provides op-
timum exposure results.

LOW AUTO
This has the same function as AUTO Mode, however, exposure times longer
than 1/50(60)s are allowed, which can result in blurred movements.

MANUAL
In MANUAL Mode the selected exposure time is displayed when the con-
trol keys are pressed. The setting range goes from 1/6 to
1/10000.
– Increments for PAL: 1/10000, 1/4000, 1/2000, 1/1000, 1/500, 1/250, 1/
100, 1/50, 1/25, 1/12 and 1/6
– Increments for NTSC: 1/10000, 1/4000, 1/2000, 1/1000, 1/500, 1/250, 1/
120, 1/60, 1/30, 1/15 and 1/8
Greater exposure times can result in blurred movements.
The set time disappears from the screen again if the or control key was
not pressed longer than 2 seconds.

Version 7.0
G-30-1781-en Page 167
Operation S100 / OPMI pico

Showing a live video image (image)

Use the <freeze> key to freeze the current live video image and to display it
on the video monitor as a full-screen still image. In the meantime the text
"freeze" appears in the upper left corner.
When you press the <live> key, the current live video image is displayed again
on the monitor. The text "freeze" disappears.

If the current live video image is being displayed and the switch connected to
the remote socket (foot switch) is pressed, and if FREEZE/LIVE is set under
SERVICE MENU - REMOTE SETUP - REMOTE FUNCTION in the CONFIG MENU,
the live video image is frozen and displayed as a still image on the monitor.
This corresponds to the "freeze" function. If you press the connected switch
(foot control panel) while the freeze frame is being displayed, the live video
image appears on the monitor again. This corresponds to the "live"function.

Version 7.0
Page 168 G-30-1781-en
S100 / OPMI pico Operation

Version 7.0
G-30-1781-en Page 169
Operation S100 / OPMI pico

Integrated HD video camera live & streaming (Operating the


video unit)

Test pattern
A test pattern (color bars) is displayed if the camera head is not connected
when you switch on the system.
Switch off the system again, connect the camera head and switch the system
back on.

Menu structure

Menu display
 The selected menu, submenu or the function is displayed in light blue.
 Selected tabs appear in white letters.

Navigation using the control keys


Control keys <select> (cursor key)
– Using the control keys, you can switch between the menus, submen-
us or the function. The selected function is displayed in light blue.
– You can use the control key to switch to the desired menu, select the
desired function, or start the desired function (e.g. white balance).

Variants
The operation and the menu differ depending on whether you have a SWQ
recoding license or not. Functions that are available only with a SW re-
cording license are identified with a "*".

Version 7.0
Page 170 G-30-1781-en
S100 / OPMI pico Operation

Fig. 72: Menu structure


Integrated HD video camera
live & streaming

Main menu*

Menu
Patient*

Menu
Settings

Submenu
Camera
- Brightness

- Image

Version 7.0
G-30-1781-en Page 171
Operation S100 / OPMI pico

Fig. 73: Menu structure


Integrated HD video camera
live & streaming

- Color

- Profiles

Submenu
Recording*/Streaming

Submenu
System
- General

- Software

Version 7.0
Page 172 G-30-1781-en
S100 / OPMI pico Operation

Fig. 74: Menu structure


Integrated HD video camera
live & streaming

- Microscope

- Network

- Video

Version 7.0
G-30-1781-en Page 173
Operation S100 / OPMI pico

Fig. 75: Schematic


Menu structure
Integrated HD video camera
live & streaming

*Available only with SW recording license

Version 7.0
Page 174 G-30-1781-en
S100 / OPMI pico Operation

Version 7.0
G-30-1781-en Page 175
Operation S100 / OPMI pico

Main menu*

The Main menu is available only with a SW recording license.


• Press the <Menu> or <config> key.
 The Main menu appears.
If you do not have a SW recording license, the "Settings" menu appears im-
mediately.

Fig. 76: Main menu 1

*5

1 Header bar
Mode: No video, Live, Image, Capture, Freeze, Pause, Play,
Forward, Rewind
Patient*: Name of the active patient, short term memory mode or
media name of the image display or video playback
Time: Time setting
2 Menu bar
Menu name is displayed
3 Patient* (only with SW recording license)
See Page 178
4 Settings (see Page 181)
Camera (see Page 182)
Recording* (with SW recording license) /
Streaming (without SW recording license) (see Page 188)
System (see Page 190)

Version 7.0
Page 176 G-30-1781-en
S100 / OPMI pico Operation

5 Footer bar
Recording medium: USB*, LAN*, Streaming, (not available).
During video recording, the information "Rec" or "Smart Rec"*
(when smart recording is active) appears next to the recording me-
dium and after it the duration of the current recording.*
Remaining recording time of the recording medium (e.g. 346:53:06)*

*Available only with SW recording license

Version 7.0
G-30-1781-en Page 177
Operation S100 / OPMI pico

Patient*

Patients can be managed in the menu "Patient". The "Patient" menu is avail-
Patient able only with a SW recording license.
• Select the "Patient" menu in the Main menu.
 The list of patients appears

Fig. 77: Patient

2 3
1 4

Creating or renaming patient records can be done faster and conveniently


using the:
– Web interface: direct keyboard entry (see Page 202)
– By creating a folder using the file manager of the computer operating sys-
tem
1 Select
Use "Select" to activate a patient.
• Use the "Select" control keys to make the selection.
• Select a patient using the control keys .
• Confirm the selection by pressing the <Enter> key or .
 The selected patient is now active. The active patient has a white bor-
der, his name appears in the header bar.
2 New
You can select "New" to create a new patient.
• Use the control keys to select "New".
• Confirm the selection by pressing the<Enter> key or .

Version 7.0
Page 178 G-30-1781-en
S100 / OPMI pico Operation

Variant 1: Select "Last name + first name" in "Patient data mode" (see
Page 189).
 A virtual keyboard appears prompting you to enter the last name.
• Please enter the patient's last name.
The maximum text length is 15 characters.
• To save the entry, select the "OK" key.
 A virtual keyboard appears prompting you to enter a first name (for
the requirement, see “Patient data mode*” on Page 189).
• Please enter the patient's first name.
The maximum text length is 15 characters.
• To save the entry, select the "OK" key.

Variant 2: Select "ID" in "Patient data mode" (see Page 189).


 A virtual keyboard appears prompting you to enter an ID.
• Enter the ID for the patient.
The maximum text length is 15 characters.
• To save the entry, select the "OK" key.
 The Patient directory is displayed.
 The new patient has been created and is active immediately.
3 Rename
You can use "Rename" to rename a patient.
• Use the control keys to select "Rename".
• Select a patient using the control keys .
• Confirm the selection by pressing the<Enter> key or .

Variant 1: Select "Last name + first name" in "Patient data mode" (see
Page 189).
 A virtual keypad showing the previous user name is displayed.
• Please change the patient's last name.
The maximum text length is 15 characters.
• To save the entry, select the "OK" button.
 The virtual keyboard appears with the patient's first name .
• Change the patient's first name.
The maximum text length is 15 characters.

Version 7.0
G-30-1781-en Page 179
Operation S100 / OPMI pico

• To save the entry, select the "OK" button.


 The Patient directory is displayed.
 The patient's name has been changed.

Variant 2: Select "ID" in "Patient data mode" (see Page 189).


 A virtual keyboard appears with the previous ID of the patient.
• Change the patient ID.
The maximum text length is 15 characters.
• To save the entry, select the "OK" button.
 The Patient directory is displayed.
 The patient ID has been changed.
4 Delete
You can use "Delete" to delete patients.
The active patient cannot be deleted.
Activate another patient and then delete the
desired patient.
• Select a patient using the control keys .
• Use the control keys to select "Delete".
• Confirm the selection by pressing the <Enter> key or .
• Confirm the warning message with "OK" to delete the patient.
 The patient and the associated data will be deleted.

Version 7.0
Page 180 G-30-1781-en
S100 / OPMI pico Operation

Settings

In the "Settings" menu you can access the "Camera", "Recording"/


Settings "Streaming" and "System" submenus.

Fig. 78: Settings

1 Camera
See Page 182
2 Recording (with SW recording license) /
Streaming (without SW recording license)
See Page 188
3 System
See Page 190

Version 7.0
G-30-1781-en Page 181
Operation S100 / OPMI pico

Camera

In the "Camera" submenu you can adjust the brightness, image, color and
profiles.
The settings in the "Brightness", "Image" and "Color" tabs directly affect the
profile set in "Profiles". Different values can be saved in the respective tabs for
each profile. The settings that are configured are saved automatically. When
the profile is changed (see "Active Profile" Page 187) the settings saved for
this profile are loaded.

Brightness
In the "Brightness" tab you can configure different exposure and brightness
settings.
Use the <Brightness> key to go directly to the "Brightness" tab.

Fig. 79: Brightness

2
3
1

4
5
6

1 Mode
In "Mode" you can choose between three different exposure controls: Au-
tomatic - Low light - Manual

Automatic In this mode the camera adjusts the brightness of the video image auto-
matically to the value specified in the "Auto level". Exposure control af-
fects the exposure time of the camera.
In "Automatic" mode the target brightness is specified and the camera ad-
justs the shutter speed in a range between 1/50 (60) s and 1/10000 s.
There is no screen output regarding the shutter speed actually set. The
target value can be set between 0 and 100%. The change is made in 5%
increments.

Version 7.0
Page 182 G-30-1781-en
S100 / OPMI pico Operation

A target value in a range from 50% to 70% usually provides the best ex-
posure results.

Low light This mode has the same function as the "Automatic" mode, but exposure
times greater than 1/50 (60) s are allowed, which can result in blurred
movements.

Manual In this mode you can set the exposure time under "Exposure time". The
setting range extends from 1/6 to 1/10000.
– Increments for PAL: 1/10000, 1/4000, 1/2000, 1/1000, 1/500,
1/250, 1/100, 1/50, 1/25, 1/12 and 1/6
– Increments for NTSC: 1/10000, 1/4000, 1/2000, 1/1000, 1/500,
1/250, 1/120, 1/60, 1/30, 1/15 and 1/8
2 Lightmeter
In "Lightmeter" you can choose between three types of exposure meas-
urement: Small spot, large spot or integral.
The lightmeter function is active only when the brightness is to set "Auto-
matic" or "Little light".

Small spot The exposure is measured in a very small area in the image center. This
measurement is suitable for work in narrow channels when an extremely
small illuminated-field diameter is being used.

Large spot The exposure is measured in an area in the image center. This measure-
ment is ideal for working with an eclipsed surgical field edge (resulting
from a reduced illuminated-field diameter).

Integral The exposure is measured and averaged across the full video image. This
measurement is recommended for surgical fields that are fully and evenly
lit.
If the object of interest to you is not in the image center, in most cases the
"Small spot" setting will not deliver the desired exposure result. Select ei-
ther a larger area ("Integral" or "Large spot") or manual exposure.

Fig. 80: Types of


exposure measurement

Small spot Large spot Integral

Version 7.0
G-30-1781-en Page 183
Operation S100 / OPMI pico

3 Black level
In "Black level" you can set the black level of the camera in a range from
-10 to +10.
4 Peak / Average
In "Peak / Average", you can set the exposure control characteristics of the
camera between 0 and +8.
5 Exposure time
In "Exposure time" you can set the exposure time of the camera. See also
"Mode" (1).
6 Auto level
In Auto level you can set the target brightness of the camera. See also
"Mode" (1).

Version 7.0
Page 184 G-30-1781-en
S100 / OPMI pico Operation

Image
In the "Image" tab you can set the display configurations.

Fig. 81: Image

1 2

1 Zoom
In "Zoom" you can set the zoom value of the camera in a range from 1.0 x
to 2.0 x in 0.1 increments.
2 Orientation
In "Orientation" you can set the image orientation.

Standard Standard
Mirror Mirror the image vertically
Flip Flip the image horizontally
Rotate Rotate the image 180°

3 White balance
With "Start white balance" and the <white bal> key you can start white
balancing.
In the white balancing procedure, the system adjusts the video signal in
such a way that white areas in the surgical field also appear in white on
the video image.
• Switch off the color filter and point the surgical microscope to a suit-
able white object (e.g. a sheet of white paper with a matte surface).
• Carry out the white balance procedure using a medium brightness lev-
el.
• Press the "Start white balance" key or the <white bal> button.
• If the white balancing was concluded successfully, a message appears.
• Confirm the notification with "OK".

Version 7.0
G-30-1781-en Page 185
Operation S100 / OPMI pico

Color
You can make the color settings in the "Color" tab.
Use the <colors> key to go to the "Color" tab.

Fig. 82: Color

2
1

3
4

1 Red
In "Red" you can set the red value of the camera in a range from -20 to
+20.
The setting of the red value also affects the white balance of the camera
(to perform white balancing, see Page 185).

2 Blue
In "Blue" you can set the blue value of the camera in a range from
-20 to +20.
The setting of the blue value also affects the white balancing of the
camera (to perform white balancing, see Page 185).

3 Color saturation
In "Color saturation" you can set the color saturation of the camera in a
range from -10 to +10.
A color saturation setting in a range of ±0 usually provides the best color
impression. The effect of the color saturation setting is the strongest for
heavily saturated (vivid) colors. The effect is small with low-saturation
colors.
4 Hue
In "Hue" you can set the hue of the camera in a range from
-10 to +10.

Version 7.0
Page 186 G-30-1781-en
S100 / OPMI pico Operation

Profiles
You can select the current profile in the "Profile" tab.

Fig. 83: Profiles

1 Active profile
A profile can be selected in "Active profile". Every profile can be renamed
and reset.
Rename Profile (Rename):
• Select the <Rename> button.
 A virtual keyboard is displayed. This virtual keyboard can be used to
rename the profile (max. 15 characters).
• You can select the characters using the control keys . Use the <En-
ter> or to adopt the selected character.
• You can delete a character by pressing "Delete".
• Press "OK" to accept a change. You can cancel the action by pressing
"Cancel".
Reset profile:
• Select the <Reset> button.
• Confirm the warning message with "OK" to reset the profile.
 The profile is reset to factory settings.

Version 7.0
G-30-1781-en Page 187
Operation S100 / OPMI pico

Recording* / Streaming

In the "Recording"* submenu (with SW recording license) or "Streaming"


Recording (without SW recording license) you can configure the recording of photos and
videos.

Streaming

Fig. 84: Recording

1 4*

*2 5*

*3 6*

1 Video quality
Here you can set the recording quality of the video to be saved. Saving a
higher video quality requires more memory space.
2 Image format*
The memory format of the still images is set here. Saving an image in
TIFF requires more memory than JPEG format. With the TIFF format, the
file is saved without data loss and is not compressed. With the JPEG
format, the file is compressed with a little data loss.
3 Recording mode*
If "Smart Recording" is active, the video image of the preceding 30 sec-
onds is recorded when the <record> button is pressed. This recording
from the past is recorded in parallel at the start of the video recording. The
recording time should be at least 30 seconds in order for the preceding 30
seconds also be to written completely in the video file.
If the recording is ended too early, not all the data can be recorded.

Version 7.0
Page 188 G-30-1781-en
S100 / OPMI pico Operation

4 Storage mode*
Here you set the output medium for the images or videos.
Without a SW recording license only "Streaming" mode is available.
In the "LAN" storage mode, a "Standard Windows Network Share" pro-
tocol is used.
In the "Streaming" storage mode, the "Real Time Protocol" (RTP) is used
and the connection is established using the "Real-Time Streaming Pro-
tocol" (RTSP).
5 Patient data mode*
Here you determine the mode of the file name.
You can enter the patient name in "Last name + first name". In ID you can
assign an ID yourself for the patient using numerical characters.
6 File mode*
When "Short Term Memory Mode" (default) is active, the data is not as-
signed to a patient, but is filed in the folder according to the date and
time.
Files that are older than 3 days are deleted automatically. In "Short Term
Memory Mode" you cannot switch the patient. Deactivate this mode in
order to be able to switch patients.
If "Short Term Memory Mode" is deactivated, patient mode is active. Your
files are then assigned to the active patient.

*Available only with SW recording license

Version 7.0
G-30-1781-en Page 189
Operation S100 / OPMI pico

System

General configurations and settings for the software, microscope, network


System and video can be made in the "System" submenu.

The "System" submenu cannot be opened when Streaming or Smart Re-


cording is activated. Therefore, deactivate Streaming or Smart Recording to
open the "System" submenu.

General
In the "general" tab you can set general configurations.

Fig. 85: General

4
1
5

2
3

1 Language
In "Language" you can select a language.
2 Function key F1 and Function key F2
In "Function key F1" and "Function key F2" you can assign to the keys
<function F1> and <function F2> the following functions:
Key <function F1>:
– "Zoom ON/OFF"
– "Smart Recording ON/OFF"*
Key <function F2>:
– "Switch camera profile"
– "Switch storage mode"* /
"Streaming ON/OFF" without SW recording license

Version 7.0
Page 190 G-30-1781-en
S100 / OPMI pico Operation

The "Zoom ON/OFF" setting of the <function F1> key will show visible
changes only if the zoom value is set greater than 1.0 (see Page 185).
3 Remote function (foot switch)
In "Remote Function" you can assign the following functions to a foot
switch:
– Video recording, which corresponds to the triggering of the button
<Record>* / "Streaming ON/OFF" without SW recording license
– Image capture, which corresponds to the triggering of the button
<Capture>* / "Freeze"/"Live" without SW recording license
The assigned function is triggered when the foot switch is pressed.
4 Sounds
In "Sounds" you can activate or deactivate the notification or acknowledg-
ment sounds.
Message: Notification in case of information and error messages
Confirmation: Confirmation when the key function is performed
5 Date / Time
In "Date / Time" you can adjust the date and the time.
Setting the date:
• Select "Date".
 A virtual keyboard appears.
• Delete the appropriate characters (backwards) by pressing the "De-
lete" button.
• Enter the date using the virtual keyboard.
Year: Month: Day
Setting the time:
• Select "Time".
 A virtual keyboard appears.
• Delete the appropriate characters (backwards) by pressing the "De-
lete" button.
• Enter the time using the virtual keyboard.
Hours: Minutes

*Available only with SW recording license

Version 7.0
G-30-1781-en Page 191
Operation S100 / OPMI pico

Software
In the "Software" tab you can obtain information on the software.

Fig. 86: Software

1
3
2

1 Version
"Version" shows you the installed software version.
2 Special
Software Update
The software can be updated via a USB medium. Old system configura-
tions are retained in this process. After this, a restart is required.
Do not switch the device off during a software update!
Factory settings
Resetting the factory settings
Service
Only for service staff.
3 Export
Log Files
Automatically generated protocols are saved as files on the USB medium.
They can be sent to ZEISS Service or an authorized partner to evaluate mal-
functions.
Open-Source Licenses
Creates a license file on the USB medium. A zip file containing notes and
the licenses is saved in the USB medium.

Version 7.0
Page 192 G-30-1781-en
S100 / OPMI pico Operation

Microscope (surgical microscope)


In the "Microscope" (surgical microscope) tab you can choose between four
different OP types. Each OP type affects the profiles and thus the camera set-
tings associated with them

Fig. 87: Microscope

1 2

1 Treatment
In "Treatment" you can choose between four different treatment settings.
Switching the type of treatment causes all the profiles to be reset to the
new default values.
Each type of treatment affects the profiles and thus the camera settings
associated with them The profiles associated with the type of treatment
selected have optimized and specified settings for the brightness control.
The following types of treatment can be selected:
– Dental: This type of treatment is designed for the optimum image set-
ting in the upper and lower tooth section area.
– ENT: This type of treatment is designed to display images in the ear,
nose and throat area.
– Gynecology: This type of treatment is designed to display images in
the gynecological area.
– Neuro: This type of treatment is designed to display images in the neu-
rological area.
2 Export / Import
In "Export / Import" you can export or import the camera settings or all
settings to or from a USB medium.

Version 7.0
G-30-1781-en Page 193
Operation S100 / OPMI pico

We recommend that you back up your configuration on a UBS data car-


rier. In the case of a malfunction of the integrated HD video camera live &
streaming, this configuration information can be used to restore the
system and your settings (including the network configuration) after the
device has been repaired or replaced.
3 Recording License
In "Recording License" you can see whether the SW recording License is
installed and activated for the video recording function (Recording). For
installation, see Page 204.

Version 7.0
Page 194 G-30-1781-en
S100 / OPMI pico Operation

Network
In the "Network" tab you can set your network configurations.

Fig. 88: Network

*2
1 3

1 IP Settings
MAC Address
The MAC Address (Media Access Control Address) is the hardware address
of each individual network adapter that serves as a unique identification
of the device in a computer network if required. The unique MAC address
is required to create a SW recording license.
DHCP activated:
Network link via a dynamic IP address "DHCP"
A dynamic IP address is an IP address which is automatically assigned to
the requesting device by an address service (DCHP server) available in the
network.
After applying the activated DHCP, you can read out the values trans-
mitted from the DHCP server using the buttons IP Address, Subnet Mask
and Standard Gateway.
DHCP deactivated:
Network connection via a static IP address.
Static IP addresses are fixed IP addresses, permanently assigned to a
system. You can obtain the data required for this purpose (IP address,
subnet mask and standard gateway) from your IT Administrator.
• Enter the data and use "Apply" to activate the changes.
For more detailed information on the network configuration, see "Config-
uring the Network Connection (default settings)" onPage 197 .
2 Network share*
The recorded video and image data can be saved in a "network folder"“
in a computer on the network.

Version 7.0
G-30-1781-en Page 195
Operation S100 / OPMI pico

Requirements for connecting the network computer:


– You may need administrator rights to the network computer to
create a network share.
– You will need the Ethernet data of the archiving PC to configure
the network share.
– You can obtain the data required for the network share from your
IT Administrator.
• Enter the data and use "Apply" to activate the changes.
3 Streaming
You can stream live camera data to a network computer with little delay.
Depending on the receiver software, delays of less than 1 second are pos-
sible.
Prerequisites for Streaming:
– You need RTSP/RTP compatible receiver software in your compu-
ter of media device.
• Enter the data and use "Apply" to activate the changes.
4 Web control / Web interface*
The web interface with activated web control is a substitute remote con-
trol that uses a web browser.
When web control is activated, you can control the following functions
over the network at the computer:
– Manage patients and input or rename patients directly using the
keyboard (see Page 202)
– Delete files or patients
– Trigger the <Record>, <Capture>, <Live> and <Index> keys (see
Page 80)
– The image is displayed or the video is played via the DVI or YPbPr
monitor output, and not in the web browser.
If web control is deactivated, control is not possible. No header bar and
footer bar or notifications are displayed in the web browser (security func-
tion e.g. for universities).

Security recommendations for networking:


– You can deactivate the use of the network. Switch over to the UBS or dis-
connect the network cable.
– You can deactivate web control over the network.

Version 7.0
Page 196 G-30-1781-en
S100 / OPMI pico Operation

– You can use access rights (user and password) to the network server or
network computer to control access to the network share.
– Share only the minimum data fields required (e.g. only for the daily re-
quirement or not all patient data).
– Use a separate share with a separate patient area for each S100 / OPMI
pico with SW recording license to prevent problems among the different
devices and users.
– You can change and adjust the network share any time.
– After using the network share, you can disable it again in the network
server.
– Install the current security updates in the network server.

The integrated HD video camera live & streaming is not an archiving system.
Therefore to protect your videos and images, you should secure a proper ar-
chiving system.

Configuring the network connection (default settings)

Fig. 89: Network

*2
1 3

In the "Network" tab (Main menu > Menu settings > System submenu > Net-
work tab) you can configure the surgical microscope for a network.
The following settings are preinstalled settings. These settings can be
changed by the IT Administrator any time that the network environment re-
quires it.
1 IP settings (static)
IP address: 192.168.0.11
Subnet mask: 255.255.255.0
Standard gateway: 192.168.0.1

Version 7.0
G-30-1781-en Page 197
Operation S100 / OPMI pico

2 Network share
Host IP: 192.168.0.10 (external computer)
Share name: recording
Work group: (or domain name)
User name: zeiss
Password: recording
3 Streaming
Client IP: 192.168.0.255 (broadcast or single client IP address)
Port: 5004 (normally does not have to be changed)
You can stream to a single receiver (client IP) by indicating a single IP ad-
dress. But the stream can also be transmitted to several receivers at the
same time if you use the broadcast address of your network area as the
client IP. In addition, each receiver must set up the connection via RTSP.
Use broadcast only in a cable-connected network and not to WLAN re-
ceivers.
If streaming is active in the S100 / OPMI pico, the following URL can be
used in the video software (additional software may be required):
URL: rtsp://192.168.0.11:8554/video
You only have to adjust the IP address or replace the name configured in
the Domain Name System (DNS).
You can read out the IP address of the surgical microscope using the "IP
Address" button even if DHCP is active (see Page 195).
4 Web control / Web interface
URL: http://192.168.0.11
(Port 80 must not be entered)
You can read out the IP address of the surgical microscope using the "IP
Address" button even if DHCP is active (see Page 195).
If a name has been saved in your Domain Name System (DNS), you can
also use this name instead of the IP address.
If you have any questions, contact your IT Administrator.

Version 7.0
Page 198 G-30-1781-en
S100 / OPMI pico Operation

Video
In the "Video" tab you can determine the display and recording options.

Fig. 90: Video

1 4

2 5

1 HDTV format
You can select the video camera's output signal and adjust it to your dis-
play device here.
720p
Select this setting if your display device is 720p compatible. The image is
set up using real full frames. This prevents any interline flicker, and the
image is sharply defined and stable. The resolution is 1280 x 720 pixels.
1080p
Select this setting if your display device is 1080p compatible. The image is
set up using real full frames. This prevents any interline flicker, and the
image is sharply defined and stable. The resolution is 1920 x 1080 pixels.
Without a SW recording license, streaming is transmitted with 720p even
if your camera and monitor are set to 1080p.
2 SDTV format
Here you can choose between a "Letterbox" and a "Sidecut" display when
connected to a display unit with Y/C analog input (monitor, projector, etc.)
mostly in 4:3 format.
Letterbox
Black bars will appear at the top and bottom of the screen and the image
will be displayed in the 16:9 format.
Sidecut
The image is cut off on the sides and only the middle of the image is dis-
played in an aspect ratio of 4:3.

Version 7.0
G-30-1781-en Page 199
Operation S100 / OPMI pico

3 Display the ZEISS logo


Here you can choose whether the ZEISS logo is to be displayed in the live
image and in the recording.
4 Video standard
You can switch from 50 Hz (PAL) and 60 Hz (NTSC) here.
After you have started the switchover, the settings will be applied and the
camera will restart. This process can take up to 30 seconds. In the mean-
time the ZEISS logo will be displayed filling the entire screen.
5 DVI / HDMI output
Here you can toggle between the video formats "PC mode" and "Video
mode" when a high-resolution display unit (monitor, projector, etc.) is
connected.
You can find the correct setting for your connected monitor or TV in the
Instructions for Use belonging to your output device.
If you do not find any information for this setting, we recommend PC
mode (default) for display units with a DVI connection. For display units
with an HDMI connection we recommend the video mode setting. PC
mode offers greater image quality than video mode.

Version 7.0
Page 200 G-30-1781-en
S100 / OPMI pico Operation

Web interface

With the help of the web interface you can use a web browser to trigger the
control buttons <Record>, <Capture> and <Live> from any computer.
• Start the web interface by entering the URL in the address line of your
browser (see Page 198).
 The start screen appears.

The link displayed on the start screen and the URL facilitate the configuration
and starting of the streaming application of your computer.

Fig. 91: Web interface 1


Start screen
2
3

5
1 Header bar
See "Main menu" on Page 176
2 Upper selection bar
Patient or Index*
Record
Capture
Live
3 Menu bar
The menu name is displayed.
4 Lower selection bar
For patient: Select, New, Rename, Delete
For index: Show/Play, Delete
5 Footer bar
See "Main menu" on Page 176

*Available only with SW recording license

Version 7.0
G-30-1781-en Page 201
Operation S100 / OPMI pico

Creating or renaming a patient using the web interface


You can rename the patient quickly and conveniently using the web interface.
• Press the "Patient" button.
 The list of patients appears.

Fig. 92: Patient web interface

You can select "New" to create a new patient.


• Press the "New" button.
 A field appears prompting you to enter the last name and first name of
the patient.
• Use the keyboard to enter the name of the patient.
• Press the "OK" button to save the entry.
 The list of patients is displayed.
 The new patient has been created and is active immediately.

Index in the web interface


You can use the web interface to open or delete the media saved in the index
(for the Index menu, see Page 207).
• Press the "Index" button in the upper selection bar or alternatively the <In-
dex> button of the remote control of the video unit.
– The "Index" is opened and the existing data is displayed.

Version 7.0
Page 202 G-30-1781-en
S100 / OPMI pico Operation

Fig. 93: Index web interface

When web control is deactivated, the Patient/Index, Record, Capture and Live
buttons are not displayed.
Without a SW recording license, the Patient/Index button is not displayed
during streaming.
A change cannot be made to the Patient menu in Short Term Memory Mode.
Only the index is available.

Version 7.0
G-30-1781-en Page 203
Operation S100 / OPMI pico

Installing the SW recording license

Fig. 94: Microscope

1 2

You need the SW recording license to be able to use all the functions such as
recording and patient management.
The SW recording license is available from your dealer. Your MAC address
(Page 195) is required for this purpose. You will receive a USB medium with
the SW recording license.
• Insert the USB medium with the license file into the USB port.
• Open the Microscope tab (Main menu > Settings menu > System sub-
menu > Microscope tab).
• Press the "Import license" button (3).
• Enter "OK" to confirm the notification.
 "Installed" appears after the button. The SW recording license is active im-
mediately. All functions can be used without restarting.

Version 7.0
Page 204 G-30-1781-en
S100 / OPMI pico Operation

Recording individual images and HD videos*

• Always check the following points (without the patient) before using the
device:
– The monitor is switched on.
– The network cable connection is established or a USB medium is con-
nected, is detected and has sufficient memory capacity.
– The correct profile was selected (see Page 187).
– The correct patient has been selected (see Page 178) or "Short Term
Memory Mode" is activated (see Page 188).
The name of the patient or "Short Term Memory Mode" (1) appears in
the header bar.

Creating individual images


• To creat individual images (photos), press the <Capture> button on the re-
mote control, the video unit, the foot switch (if configured) or the web in-
terface.

Recording HD videos
• To start video recording, press the <Record> button on the remote con-
trol, the video unit, the foot switch (if configured) or the web interface.
 During video recording, "Rec" (2), the duration of the recording (3)
e.g. 00:00:08 and the remaining recording time (4) e.g. (348:26:22)
appear in the footer bar. The "record" indicator light on the control
unit lights up with an amber color.
While recording a video the photo function (Capture) can be used to
create additional images.

Version 7.0
G-30-1781-en Page 205
Operation S100 / OPMI pico

Fig. 95: Video recording


1

3
2 4

• Press the <Record> button again to end the video recording.


 "Rec" (2) and video recording time (3) cease to be illuminated in the
footer bar.

*Available only with SW recording license

Version 7.0
Page 206 G-30-1781-en
S100 / OPMI pico Operation

Viewing individual images (photos) and HD videos * in the index

 The patient whose data you wish to view is activated (see Page 178) or
"Short Term Memory Mode" is activated (see Page 188).
The name of the patient or "Short Term Memory Mode" (1) appears in the
header bar.
• Press the <Index> button of the remote control, the video unit or the web
interface.
 The "Index" is opened and the existing data is displayed.

Fig. 96: Viewing images and


1
videos

• Button (2) shows various texts, depending on the data type that was se-
lected:
- "Show" for images,
- "Play" for videos.
• Use the keys to select the file that you want to view.
• Use the keys to select the "Show "or "Play" button and press the
<Enter> or key.
 The selected image or video is displayed or played.
• Press the <Index> or <Live> button of the remote control, video unit or
web interface to end the image display or video playback.
 The image display or video playback is ended and the "Index" is displayed.

*Available only with SW recording license

Version 7.0
G-30-1781-en Page 207
Operation S100 / OPMI pico

Deleting* individual images (photos) and HD videos in the Index

 The patient whose data you wish to delete is activated (see Page 178) or
"Short Term Memory Mode" is activated (see Page 188).
The name of the patient or "Short Term Memory Mode" (1) appears in the
header bar.
• Press the <Index> button of the remote control, the video unit or the web
interface.
 The "Index" is opened and the existing data is displayed.

Fig. 97: Viewing images and


1
videos

• Use the keys to select the file that you want to delete.
• Use the keys to select the "Delete" button (2) and press the <Enter>
or key.
• Confirm the notification with "OK" to delete the selected filed.
 The selected file is deleted and the "Index" is displayed.

Version 7.0
Page 208 G-30-1781-en
S100 / OPMI pico What to do in the event of malfunctions

What to do in the event of malfunctions

Troubleshooting ............................................................................210
For your safety ........................................................................................210
Malfunctions of the system......................................................................210
Malfunctions on S100 stands with video (optional)..................................210
Malfunctions of S100 suspension systems with video and image storage op-
tion .........................................................................................................216

Lamp failure of halogen light source .............................................217

Lamp failure of xenon light source ................................................ 218

Fault in the LED lilght source .........................................................220

Version 7.0
G-30-1781-en Page 209
What to do in the event of malfunctions S100 / OPMI pico

Troubleshooting

For your safety

• If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions“, attach a sign to the system
stating it is out of order and contact our service representative.

Malfunctions of the system

Malfunctions on S100 stands with video (optional)

Problem Possible cause Remedy Cross-


reference
No function at all Suspension system turned off. Switch on suspension system Page 128
or plug in power cord.
Power plug of stand not in- Plug in the power cord. Page 128
serted.
Defective power cord. Replace the power cord. –
Power failure Contact in-house electrician. –

Version 7.0
Page 210 G-30-1781-en
S100 / OPMI pico What to do in the event of malfunctions

Problem Possible cause Remedy Cross-


reference
Light source (surgical field illu- Power cord not plugged in. Plug in the power cord. Page 108
mination) disabled
Power switch not pressed. Press power switch. Page 108
Defective power cord. Replace the power cord. –
Power failure. Contact in-house technician. –
Defective instrument fuse. Change instrument fuse. Page 228
(Halogen)
Failure of suspension elec- Illuminate surgical area with –
tronics. an OR luminaire. Contact
service.
Light guide not properly in- Insert light guide as far as it Page 98
serted in lamp or microscope. will go.
Switched off via inclination Move suspension arm of sus- –
switch on suspension system. pension system into working
position.
Light source (surgical field illu- Brightness level set too low. Adjust brightness using –
mination) defective brightness control on suspen-
sion arm.
Light guide not properly in- Insert light guide as far as it Page 98
serted in lamp and/or micro- will go.
scope.
Defective light guide (illumi- Contact service. –
nation not uniform).
For xenon light sources: Disable light source on stand. Page 226
Old xenon lamp. Change xenon lamp.
For LED light source: Contact service. –
LED light source with limited
operating mode.

Version 7.0
G-30-1781-en Page 211
What to do in the event of malfunctions S100 / OPMI pico

Problem Possible cause Remedy Cross-


reference
Lamp in light source remains Knob for brightness control Turn knob clockwise. Page 70
dark and fan is running set to left-hand stop.
Inclination switch deacti- Move suspension arm of sur- –
vated. gical microscope down from
its standby position.
Inclination switch out of ad- Contact service. –
justment or defective
Lamp module has no contact. Insert lamp module to the –
stop.
For halogen light source: Switch to backup lamp. Page 224
Defective halogen lamp.
For halogen light source: Plug ceramic base firmly onto Page 224
Ceramic base does not have contacts of halogen lamp.
proper contact with halogen
lamp.
For xenon light sources: Switch to backup lamp. Page 218
Defective xenon lamp.
For LED light source: Contact service. –
Defective LED light source.
Defective stand electronics. Illuminate surgical area with –
an OR luminaire.
Contact service.
Lamp goes constantly off and Ventilation openings are cov- Ventilation openings must be –
on again during operation. ered or dirty. clear; clean them if necessary.
Defective fan. Illuminate surgical area with –
an OR luminaire. Contact
service.
Failure of suspension elec- Illuminate surgical area with –
tronics. an OR luminaire. Contact
service.
Defective magnification - Contact service. –
changer.
Filters in illumination beam - Contact service. –
path are defective or cannot
be exchanged.

Version 7.0
Page 212 G-30-1781-en
S100 / OPMI pico What to do in the event of malfunctions

Problem Possible cause Remedy Cross-


reference
Video image interference on Endoscope camera or video Contact manufacturer of en- –
monitor or from video re- equipment (monitor and doscope system.
corder. video recorder) on endo-
scope system defective.
Motion of surgical micro- Friction adjustment screw on Loosen friction adjustment Page 66
scope too stiff. stand tightened too firmly. screw on stand as required.

Problem Possible cause Remedy Cross


reference
No video image - only Camera head not connected. Connect the camera head. Page 86
color bars
No image, Plug of video power supply is Contact service. –
video indicator lamp is off, not properly plugged onto
Halogen lamp in light the plate of the light source.
source is lit
Defective light source. Contact service. –
Image with color tinge No white balance Perform white balance Page 165
Image without color Y/C cable between CCU and Plug connector in properly. Page 110
monitor not properly plugged
in.
Defective Y/C cable between Use different Y/C cable or BNC
CCU and monitor. cable.
No image, video indicator Monitor not switched on or Switch on monitor or set it to Please refer
lamp is lit not correctly set. correct channel. to monitor's
Instructions
Set monitor to default status, or
for Use.
increase brightness and contrast.
Monitor cannot display the Use an HD compatible monitor –
set resolution of the HD video that can display the set resolu-
camera. tion.
Connecting cable between Plug connector in properly. Page 110
CCU and monitor or between
CCU and surgical microscope
not or not properly plugged
in.
Defective connecting cable Replace the cable. –
between video unit and mon-
itor.

Version 7.0
G-30-1781-en Page 213
What to do in the event of malfunctions S100 / OPMI pico

Problem Possible cause Remedy Cross


reference
No image, video indicator If a video recorder is con- Connect CCU and monitor di- Page 110
lamp is lit nected between CCU and rectly.
monitor: Defective video re-
corder.
Defective CCU or camera Contact service. –
head in surgical microscope,
or defective connecting cable
between CCU and surgical
microscope.
Image too dark, Monitor not correctly set. Set monitor to default status, or Please refer
snow effect on monitor. increase brightness and contrast. to monitor's
Instructions
for Use.
Brightness of video unit set Increase brightness of video unit. Page 167
too low.
Surgical field illumination too Increase brightness of illumina- Page 70
dark. tion.
Image too bright, Monitor not correctly set. Set monitor to default status, or Please refer
Milky image on monitor reduce brightness and/or con- to monitor's
trast. Instructions
for Use.
Brightness of video unit is set Reduce brightness of video unit. Page 167
too high.
Surgical field illumination too Reduce brightness of surgical Page 70
bright. field illumination.
ENT application: Image de- Video unit: Lightmeter function - set "Small Spot" Page 82
tail in black speculum too
Brightness of CCU set too Reduce brightness of video unit. Page 167
bright, too shiny
high.
Brightness control is based on Center the speculum and object –
black color of speculum. in center of screen.
Use a metal speculum (gray, –
stainless steel).

Version 7.0
Page 214 G-30-1781-en
S100 / OPMI pico What to do in the event of malfunctions

Problem Possible cause Remedy Cross-


reference
Remote control does not Battery in remote control dis- Replace battery. Page 231
work. charged.
Remote control pointed in Point remote control at front –
wrong direction or distance of device; distance should be
of remote control from device more than 0.5 m.
too short.
Streaming does not work. 1920 x 1080 resolution too Switch HDTV format to Page 199
high for video. "720p" (Settings menu >
System submenu > Video).
Video quality too high. Set video quality to "low" Page 188
(Settings menu > System sub-
menu > Video).
Network settings in receiver Check network settings (Set- Page 195
computer wrong. tings menu > System sub-
menu > Network).
Firewall prevents data flow. Check firewall settings. –
Receiver software does not Check network settings (Set- Page 195
support RTSP/RTP or is config- tings menu > System sub-
ured incorrectly. menu > Network).
Check receiver software set- –
tings.

Version 7.0
G-30-1781-en Page 215
What to do in the event of malfunctions S100 / OPMI pico

Malfunctions of S100 suspension systems with video and image storage op-
tion

Fault Possible cause Remedy Cross-


(Message/acoustic signal) reference
No stick available No USB data carrier inserted. Insert USB data carrier. –
Not enough memory avail- USB data carrier is inserted Press the Index key; menu ap- –
able but there is insufficient pears on the screen, permit-
memory space. ting images to be deleted.
Write-protection activated USB data carrier is inserted Deactivate write protection. –
but write-protection is active.
Defective USB memory USB data carrier does not Replace the USB data carrier. –
please remove the stick contain any file system or a
damaged
file system.
Error Short circuit on Remote Contact service. –
Remote Short Circuit - socket - all keys have been
Disconnect Input disabled.

The messages are automatically deleted.

Version 7.0
Page 216 G-30-1781-en
S100 / OPMI pico What to do in the event of malfunctions

Lamp failure of halogen light source

The lamp housing contains a backup lamp which can be manually swung into
the illumination beam path when the first lamp fails. Open flap (1) indicates
that the backup lamp is operative.

CAUTION Risk of burns!


In the event of overheating, the lamp switches off automatically. When the
lamp has cooled down, it automatically switches on again.
If the microscope is positioned unattended above the patient, the patient can
be burnt due to the produced heat.
• Immediately activate the backup lamp after a lamp failure or switch off
the light source completely.
• Check if the light source is switched off before leaving the device unat-
tended.
• Press button (2) to manually swing the backup lamp into the beam path.
 Open flap (1) indicates that the backup lamp is operative.
• Replace the defective lamp at the first opportunity (see Page 224).
• If the backup lamp is defective and replacement is not possible, continue
surgery using an external OR illuminator.

Fig. 98: Swing in the


backup lamp

Version 7.0
G-30-1781-en Page 217
What to do in the event of malfunctions S100 / OPMI pico

Lamp failure of xenon light source

The xenon light source has a lamp module containing two xenon lamps. The
second lamp is used as backup lamp. Manually swivel in the backup lamp after
a failure of the first lamp or if the maximum service life of the first lamp of
500h (see chapter "Maintenance intervals" on Page 222) is exceeded. The
backup lamp is in operation if the red segment (3) is lit in button (2).

CAUTION Risk of burns!


In the event of overheating, the lamp switches off automatically. When the
lamp has cooled down, it automatically switches on again.
If the microscope is positioned unattended above the patient, the patient can
be burnt due to the produced heat.
• Immediately activate the backup lamp after a lamp failure or switch off
the light source completely.
• Check if the light source is switched off before leaving the device unat-
tended.

Manual switching to the backup lamp


• Turn off the suspension system at the power switch before switching to
the backup lamp.
• Press the button (4) to open the lamp module.
 The lamp module is slightly ejected.
• Pull out the lamp module (1) as far as it will go. Turn knob (2) through
180° until it snaps in place.
 This moves the backup lamp into the illumination beam path.
• Push the lamp module (1) all the way back into the lamp housing.
• Turn on the suspension system at its power switch.
• In case the backup lamp is defective and an exchange of the lamp module
is not possible, please proceed with an external surgical lamp.

Failure of backup lamp


• If you work with the backup lamp and this lamp fails during a surgery, the
lamp module of the xenon light source must be exchanged, see Page 226.

Version 7.0
Page 218 G-30-1781-en
S100 / OPMI pico What to do in the event of malfunctions

An additional xenon lamp module ready for use is recommended in regard


of the preparation of a procedure.
If the use of a backup lamp is initiated, it is absolutely required.

Fig. 99: Lamp failure of xenon


light source

1
2

Version 7.0
G-30-1781-en Page 219
What to do in the event of malfunctions S100 / OPMI pico

Fault in the LED lilght source

The surgical microscope is equipped with an LED light source.


1 LED error message
If the amber LED is illuminated, the LED light source is in a limited oper-
ating mode.
In this case, the light intensity drops down to 50 %.
• Please contact ZEISS Service.

Fig. 100: Malfunction in the


LED light source

Failure concept for the LED light source


Automatic treatment continuation due to integrated failure concept:
– The LED light source uses an integrated failure concept for automatic
treatment continuation without an interruption.
– Individual LEDS are electronically arranged in two independent strands,
where one strand automatically continues to be use in the event of a mal-
function.
 50% light intensity

Version 7.0
Page 220 G-30-1781-en
S100 / OPMI pico Care and Maintenance

Care and Maintenance

System maintenance......................................................................222
Maintenance intervals..............................................................................222
Safety check ............................................................................................223
Replacing the halogen lamp ....................................................................224
Exchanging the lamp module of the xenon illumination system ..............226
Changing the fuse - halogen illumination system .....................................228

Care of the device..........................................................................229


Cleaning ..................................................................................................229
Sterilization .............................................................................................230
Disinfection .............................................................................................230

Inserting or changing the battery ..................................................231

Environmental protection measures ..............................................232


Note on disposal .....................................................................................232

Version 7.0
G-30-1781-en Page 221
Care and Maintenance S100 / OPMI pico

System maintenance

Maintenance intervals

Regular preventive maintenance is required to ensure safe and correct opera-


tion and a long service life of the system on an ongoing basis.
The following table lists the relevant maintenance activities and shows at
which maximum intervals they must be performed.
The corresponding procedures in order to perform maintenance activities on
your own can be found in the following descriptions.
Any other maintenance activities that are not described here in detail require
device-specific knowledge.
Please contact your local ZEISS Service or an authorized partner for per-
forming maintenance activities in due time.

Interval Maintenance tasks


Max. 12 weeks Exchange xenon lamp if it is operated for 8 hours/day on 5 days/week. This
corresponds to the service life of a xenon lamp of 500 h.
Max. 25 weeks Exchange xenon lamp if it is operated for 4 hours/day on 5 days/week. This
corresponds to the service life of a xenon lamp of 500 h.
Every 2 years Visual inspection and function test of the following modules:
– S100 suspension system
– Illumination and power supply unit
– Surgical microscope with tube
– Light guide
– Power cord
– Connected accessories
– Secure seating of OPMI suspension
– Securing screws are available and tightened
According to country-specific Safety check
requirements

Version 7.0
Page 222 G-30-1781-en
S100 / OPMI pico Care and Maintenance

Safety check

CAUTION Risk of injury!


• Make sure that the regular technical safety checks required for this sys-
tem in accordance with the applicable national regulations are performed
on schedule and to the stipulated extent.

To prevent any impairment of the system's safety as a result of aging, wear,


etc., the organization operating the system must ensure, in accordance with
the applicable national regulations, that the regular technical safety checks
defined for this system are performed on schedule and to the stipulated ex-
tent.
The safety checks shall only be performed by the manufacturer or by qualified
staff.
At a minimum, the scope of the safety checks of the system should comprise
the following points:
– Availability of the user manual
– Visual inspection of the system and accessories for damage and legibility
of the labels
– Leakage current test
– Test of the protective ground conductor
– Function and wear test of casters and arrestors on the floor stand
– Function test of all switches, buttons, sockets and indicator lamps of the
system
– Securing screws are available and tightened

Version 7.0
G-30-1781-en Page 223
Care and Maintenance S100 / OPMI pico

Replacing the halogen lamp

NOTE Risk of burns caused by hot lamps!


If you change the lamp shortly after it has failed, the lamp will still be very
hot.
• Allow the lamp to cool down before changing it.
• Only use the halogen light indicated in chapter "Device data => order da-
ta".
• Turn off the suspension system at the power switch.
• Press button (3), the lamp module is slightly ejected. Pull out lamp module
(2).
• Press the button (1). The lamp holder in the lamp module springs open.
• Remove the defective halogen lamp from the spring-loaded mount.
• Pull the ceramic base (9) from the contact pins of the halogen lamp.
Make sure not to touch the bulb (8) or the inner side of the lamp reflector (7)
of the new halogen lamp.
• Plug ceramic base (9) onto the contact pins of a new halogen lamp.
• Insert the new halogen lamp.
• Press the halogen lamp into the spring-loaded mount.
• Press the lamp holder (5) into the lamp module until it snaps in.
• Push the lamp module including the new halogen lamp back into the sys-
tem.
• Turn on the system at its power switch.
• Use the brightness control to adjust the illumination when the suspension
arm is in its working position.

Version 7.0
Page 224 G-30-1781-en
S100 / OPMI pico Care and Maintenance

To use the service life of your halogen lamps as economically as possible, we


recommend that you do not set the brightness to a higher level than required
for your work.
If main lamp (4) has failed, remove it and replace it by backup lamp (6). Install
the new halogen lamp instead of the backup lamp.

Fig. 101: Changing the


halogen lamp

1
2
3

7 8 9

Version 7.0
G-30-1781-en Page 225
Care and Maintenance S100 / OPMI pico

Exchanging the lamp module of the xenon illumination system

NOTE Risk of burns caused by hot lamps!


If you change the lamp shortly after it has failed, the lamp will still be very
hot.
• Allow the lamp to cool down before changing it.

NOTE Risk of injury!


The xenon lamp may fast burst due to the high inner pressure even if it is
cold. A hot lamp is therefore more dangerous due to the higher pressure and
the used materials.
In case of burst lamps, persons may be injured by flying glass splinters.
The lamp module shall only be changed by instructed staff.
• Only change the lamp module when it is cold.
• Avoid shocks at the lamp module.
• Turn off the suspension system at the power switch.
• Press button (2). The lamp module (1) is slightly ejected.
• Pull out the lamp module (1) as far as it will go.
• Slide the original transport case (3) over the lamp module. Bolt (4) must
engage in hole(5). This unlocks the stop.
• Remove the old lamp module and insert the new module.
• Turn on the suspension system at its power switch.
• Check the function of the xenon lamp and backup xenon lamp.
• Pack the old lamp module (6) in the transport packaging of the new lamp
module. Fill in the enclosed return card and send the old lamp module to
the nearest Zeiss service representative.
Please only use the original transport packaging (6), it is an explosion protec-
tion for eventually defective xenon lamps.

Version 7.0
Page 226 G-30-1781-en
S100 / OPMI pico Care and Maintenance

Fig. 102: Changing the lamp


module of the
xenon illumination system

1
3

Version 7.0
G-30-1781-en Page 227
Care and Maintenance S100 / OPMI pico

Changing the fuse - halogen illumination system

CAUTION Use suitable device fuse


• Only use the fuse indicated for the device (see order data).
The fuses are integrated in the power inlet socket for the power cable. To
change a fuse, proceed as follows:
• Turn off the unit at the power switch.
• Unplug the power cable from the line power socket.
• Unplug the power cable from the power inlet socket.
• Insert a suitable object under two flaps (1) and pull out fuse holder (2).
• Remove the defective fuse.
• Insert the new fuse and push the fuse holder back into its compartment.
Note the voltage rating of the line fuse. This rating can be found on the type
label.
• Connect the device to the power supply.
• Turn on the device at its power switch.

Fuse rating
For 100 V...125 V~ T 6,3 A/H 250 V
For 220 V...240 V~ T 3,15 A/H 250 V

Fig. 103: Changing the fuses


1

Version 7.0
Page 228 G-30-1781-en
S100 / OPMI pico Care and Maintenance

Care of the device

Cleaning

Cleaning optical surfaces


The multi-layer T* coating of the optical components (e.g. eyepieces, objective
lenses) ensures optimum image quality.
Image quality is impaired by even slight contamination of the optics or by a
fingerprint. To protect the internal optics from dust, the system should never
be left without the objective lens, binocular tube and eyepieces. After use,
cover the system to protect it from dust. Always store objective lenses, eye-
pieces and accessories in dust-free cases when they are not being used.
Clean the exterior surfaces of the optical components (eyepieces, objective
lenses) only when necessary:
• Do not use any aggressive or abrasive cleaning agents.
• Remove dust from the optical surfaces using a squeeze blower or a clean,
grease-free brush.
For the regular cleaning of objective lenses and eyepieces of the surgical mi-
croscope, we recommend the optics cleaning set available from ZEISS. For the
catalog number, please see the section "System data - Ordering data".

Prevention of fogging
To protect the eyepiece optics from fogging, we recommend using an anti-
fogging agent. Anti-fogging agents provided by eyecare professionals for use
with eyeglass lenses are also suitable for Zeiss eyepieces.
• Please observe the Instructions for Use supplied with each anti-fogging
agent.
Anti-fogging agents do not only ensure fog-free eyepiece optics. They also
clean the eyepiece optics and protect them from dirt, grease, dust, fluff and
fingerprints.

Cleaning mechanical surfaces


All mechanical surfaces of the system can be cleaned by wiping them with a
damp cloth. Do not use any aggressive or abrasive cleaning agents.
Clean off any residue using a mixture of 50% ethyl alcohol and 50% distilled
water plus a dash of household dish-washing liquid.

Version 7.0
G-30-1781-en Page 229
Care and Maintenance S100 / OPMI pico

Sterilization

The asepsis sets available from ZEISS contain rubber caps and hand grips that
can be sterilized in autoclaves. For detailed information on sterilization, please
see the enclosed instructions "Preparation of resterilizable products" for the
respective asepsis set.
You can also use disposable sterile drapes to cover the system.
• When fitting the sterile drape, ensure there is sufficient slack to allow for
movement of the microscope carrier and surgical microscope.
• Do not cover the ventilation opens - this ensures the lamps are cooled ad-
equately and no lamp fails.

Disinfection

NOTE Risk of damaging the paint on the device!


• Use a disinfectant based on aldehyde or alcohol. The addition of quater-
nary compounds is acceptable.
To avoid damaging the surfaces, do not use any disinfecting components
other than those specified below.
The maximum concentrations are:
– For alcohol (tested with 2 propanol): 60%
– For aldehyde (tested with glutaraldehyde): 2%
– For quaternary compounds (tested with DDAC): 0.2%

Version 7.0
Page 230 G-30-1781-en
S100 / OPMI pico Care and Maintenance

Inserting or changing the battery

A battery compartment is provided on the back of the remote control.


• Slide the cover of the battery compartment open.
• Insert two LR03 alkaline batteries (size AAA) as shown below.
– Do not try to recharge the empty dry cell batteries.
– Do not bring the dry batteries in contact with metal objects. This can lead
to heat formation.
– Do not combine different battery types.
– Exchange both batteries when performing a battery change.
Remove the batteries to avoid damage by corrosion whenever the remote
control is not used for a longer period.

Fig. 104: Battery change


Remote control

Disposal of used batteries


Under no circumstances may batteries be disposed of as normal domestic
waste.
• When disposing of batteries, adhere to the appropriate domestic laws and
other regulations in your country governing the disposal of batteries.

Version 7.0
G-30-1781-en Page 231
Care and Maintenance S100 / OPMI pico

Environmental protection measures

Note on disposal

User information
on the disposal of electrical and electronic devices

This symbol means that the product must not be disposed of as normal do-
mestic waste.
The correct disposal of electrical or electronic devices helps to protect the en-
vironment and to prevent potential hazards to the environment and/or human
health which may occur as a result of improper handling of the devices con-
cerned.
For detailed information on the disposal of the product, please contact your
local dealer or the device manufacturer or its legal successor. Please also note
the manufacturer's current information on the Internet. In the event of resale
of the product or its components, the seller is required to inform the buyer
that the product must be disposed of in accordance with the applicable na-
tional regulations currently in force.

For customers in the European Union


Please contact your dealer or supplier if you wish to dispose of electrical or
electronic devices.

Information on disposal in countries outside the European Union


This symbol is only applicable in the European Union. For the disposal of elec-
trical and electronic devices, please observe the relevant national legislation
and other regulations applicable in your country.
Disposal of outer packaging material
• In order to dispose of outer packaging materials please refer to the local
guidelines and bylaws.

Version 7.0
Page 232 G-30-1781-en
S100 / OPMI pico System Data

System Data

Technical data ............................................................................... 234


Essential performance of the S100 / OPMI pico .......................................234
OPMI pico surgical microscope ................................................................235
S100 Suspension Systems ........................................................................236
Connectors and illumination ....................................................................243

EMC (electromagnetic compatibility) .............................................245


Electromagnetic interference ...................................................................246
Electromagnetic immunity for ME equipment and ME systems ................247
Electromagnetic immunity for non-life-supporting ME equipment
and ME systems.......................................................................................248
Recommended safety distances between portable and mobile
RF communication equipment and the S100 / OPMI pico ........................250

System upgrades ...........................................................................251

Ordering data ................................................................................251

Environmental conditions ..............................................................257

Regulatory information ................................................................. 259

Version 7.0
G-30-1781-en Page 233
System Data S100 / OPMI pico

Technical data

Essential performance of the S100 / OPMI pico

The S100 / OPMI pico has no essential performance within the meaning of
IEC 60601-1: 2005 + A1:2012.
The fundamental functions within the meaning of the intended use were
tested by the manufacturer for their stability.

Version 7.0
Page 234 G-30-1781-en
S100 / OPMI pico System Data

OPMI pico surgical microscope

Component Properties
Magnification Manual magnification changer,
Five settings where
γ = 0.4x, γ = 0.6x, γ = 1.0x, γ = 1.6x, γ = 2.5x
Focusing Objective lens with fixed focal distance: Focusing oc-
curs by employing the working distance to the ob-
ject
Focusing objective lens: Variably adjustable working
distance, focus range 13 mm.
Varioscope 100: Variably adjustable working dis-
tance, focus range 100 mm.
Objective lens Objective lens with fixed focal distance:
Optionally f = 200 mm, f = 250 mm, f = 300 mm,
f = 350 mm or f = 400 mm
Focusing objective lens:
Optionally f =200 mm, f = 250 mm or f = 300 mm
Varioscope 100:
AA = 200 - 300 mm
Tubes / Eyepieces Optional binocular sloped, straight or 180° swiveling
tube.
10x widefield eyepieces, alternatively: 12.5 x
Weight approx. 1.7 kg (without CCU, Endoport, tube, object
lens, eyepieces and MORA interface)

Fig. 105: OPMI pico

Version 7.0
G-30-1781-en Page 235
System Data S100 / OPMI pico

S100 Suspension Systems

S100 floor stand

Component Features
Suspension arm Length: 600 mm
Swivel angle about carrier arm: ± 150°
Lifting range: ± 300 mm
Carrier arm Length: 500 mm
Stand height 1730 mm
Base 625 x 625 mm
(dimensions)
Weight approx. 90 kg
(including
microscope)

Fig. 106: Dimensional


drawing S100 floor stand 500 mm 600 mm
Illustration with illumination

± 300 mm
1730 mm

360° ± 150°
1470 mm

Version 7.0
Page 236 G-30-1781-en
S100 / OPMI pico System Data

S100 Wall stand

Component Properties
Suspension arm Length: 950 mm
Vertical lift: ± 400 mm
Turning angle around carrier arm: 150°
Carrier arm Length: 500 mm
Turning angle around stand column: ±105°
Weight Approx. 45 kg
(including
microscope)

Fig. 107: Dimensional drawing


320 mm 500 mm 950 mm
of S100 wall mount

520 mm
± 400 mm

±105° ±150°

Version 7.0
G-30-1781-en Page 237
System Data S100 / OPMI pico

S100 Ceiling stand

Components Properties
Suspension arm Length: 950 mm
Turning angle around carrier arm: ±150°
Vertical lift: ± 400 mm
Carrier arm Length: 500 mm
Turning angle around stand column: ±163°
Weight Approx. 55 kg
(including
microscope)

Fig. 108: Dimensional drawing


of S100 ceiling mount

500 mm 950 mm
425 mm

± 400 mm

± 163° ± 150°

Version 7.0
Page 238 G-30-1781-en
S100 / OPMI pico System Data

S100 Centro stand

Component Properties
Suspension arm Length: 950 mm
Turning angle around carrier arm: 150°
Vertical lift: ± 400 mm
Carrier arm Length: 500 mm
Weight Approx. 30 kg
(including microscope
and halogen lighting)

Fig. 109: Dimensional


drawing S100 Centro stand

500 mm 950 mm

425 mm

± 400 mm
±150°

Version 7.0
G-30-1781-en Page 239
System Data S100 / OPMI pico

S100 Carrier unit

Component Properties
Suspension arm Length: 950 mm
Turning angle around carrier arm: ±150°
Vertical lift: ± 400 mm
Carrier arm Length: 500 mm
Weight Approx. 35 kg
(including microscope and
lighting)

Fig. 110: Dimensions of


S100 support unit 500 mm 950 mm

275 mm

± 400 mm
± 150°

Version 7.0
Page 240 G-30-1781-en
S100 / OPMI pico System Data

S100 support unit for non-Zeiss illumination

Component Properties
Suspension arm Length: 950 mm
Turning angle around carrier arm: 150°
Vertical lift: ± 400 mm
Carrier arm Length: 500 mm
Weight (including Approx. 25 kg
microscope)

Fig. 111: Dimensional


drawing of S100 support unit 500 mm 950 mm
for non-Zeiss illumination

275 mm

± 400 mm
± 150°

Version 7.0
G-30-1781-en Page 241
System Data S100 / OPMI pico

S100 Atmos carrier unit

Component Properties
Suspension arm Length: 950 mm
Turning angle around carrier arm: 150°
Vertical lift: ± 400 mm
Carrier arm Length: 500 mm
Weight (including Approx. 25 kg
microscope)

Fig. 112: Dimensional


drawing S100 support unit 500 mm 950 mm
Atmos

275 mm

± 400 mm
± 150°

Version 7.0
Page 242 G-30-1781-en
S100 / OPMI pico System Data

Connectors and illumination

Component Properties
Power connection Only connect the system to wall outlets which
are equipped with a properly connected pro-
tective ground conductor.
Electrical outputs HD Video output HDMI
video control units
Video output YPbPr
Video output BNC
Remote Socket
External switching contact input
Electrical outputs video Video output YPbPr
camera live &
LAN input/output
streaming control unit
Video output port Y/C
Video output port DVI
Remote Socket
External switching contact input
Halogen lighting system
Rated voltage (115 V): 100 V - 125 V
(230 V): 220 V - 240 V
Rated frequency 50 - 60 Hz
Current consumption (115 V): max. 250 VA
(230 V): max. 250 VA
Fiber optic lighting Halogen lamp, 12 V, 100 W,
backup lamp can be manually swiveled into beam
path
Fuses Fuse link
115 V ~ T 6.3 A/H 250 V
230 V ~ T 3.15 A/H 250 V
Xenon lighting system
Rated voltage (115 V): 100 V - 125 V
(230 V): 220 V - 240 V
Rated frequency 50 - 60 Hz

Version 7.0
G-30-1781-en Page 243
System Data S100 / OPMI pico

Component Properties
Current consumption (115 V): max. 350 VA
(230 V): max. 350 VA
Fiber optic illumination Xenon short-arc reflector lamp,
Color temperature: approx. 5000 K
Rated power: approx. 180 W
Backup lamp can be manually swung into beam
path
Fuses Automatic circuit breaker
LED lighting system
Rated voltage 100 V - 240 V
Maximum input 250 VA
Power frequency range 50 Hz - 60 Hz
Fuses Automatic circuit breaker

Version 7.0
Page 244 G-30-1781-en
S100 / OPMI pico System Data

EMC (electromagnetic compatibility)

The device is subject to specific precautions with regard to electromagnetic


compatibility (EMC). In order to avoid the occurrence of EMC interference, the
device may only be installed, operated and maintained in the manner indi-
cated in these Instructions for Use and only with components supplied by
ZEISS.

CAUTION Danger from electromagnetic radiation


Electrical devices can influence each other as a result of their electromagnetic
radiation. The use of non-approved components (accessories, transformers
of all types, cables) can cause increased emissions or reduce the device's im-
munity.
• With the exception of the combination of the devices described in these
Instructions for Use, do not operate the device in direct proximity to other
devices.
• Only use accessories, transformers, cables and spare parts which are
specified in these Instructions for Use or which are approved by ZEISS for
this device.
• Do not use any portable or mobile RF communication equipment near
the device as it is not possible to exclude the possibility that the function
of the device will be affected.
• Please follow the EMC guidelines in the following pages.

Version 7.0
G-30-1781-en Page 245
System Data S100 / OPMI pico

Electromagnetic interference

The S100 / OPMI pico is intended for operation in an electromagnetic environ-


ment as specified below. The customer or the user of S100 / OPMI pico is re-
sponsible for ensuring that the device is operated in such an environment.

Interference measure- Compliance Electromagnetic environment - guidelines


ments
RF emissions as per CISPR11 Group 1 The S100 / OPMI pico uses RF energy only for its in-
ternal functions. As a result, RF emissions are very low
and unlikely to cause any interference in nearby elec-
tronic devices.
RF emissions as per CISPR11 Class B The S100 / OPMI pico is suitable for use in all facilities
including locations in residential environments and
Harmonic emissions as per EN Class A
those directly connected to the public power supply
61000-3-2
network which also supplies buildings used for resi-
Emission of voltage fluctua- Compliant dential purposes.
tions/flicker as per EN 61000-
3-3

Version 7.0
Page 246 G-30-1781-en
S100 / OPMI pico System Data

Electromagnetic immunity for ME equipment and ME systems

The S100 / OPMI pico is intended for operation in an electromagnetic environ-


ment as specified below. The customer or the user of the S100 / OPMI pico is
responsible for ensuring that the device is operated in such an environment.

Immunity IEC 60601 Compliance level Electromagnetic environment -


tests test level guidelines
Electrostatic dis- ±6 kV ±6 kV Floors should be made of wood or
charge (ESD) as per contact discharge contact discharge concrete or be covered with ceramic
IEC 61000-4-2 tiles. If the flooring contains synthetic
±8 kV air discharge ±8 kV
materials, the relative humidity must
air discharge
be at least 30%.
Fast transient/burst ±2 kV for ±2 kV for The quality of the supply voltage
immunity as per power supply lines power supply lines should be that of a typical business or
IEC 61000-4-4 ±1 kV for ±1 kV for hospital environment.
input/output lines input/output lines
Surges as per ± 1 kV phase-to-neutral ± 1 kV phase-to-neutral The quality of the supply voltage
IEC 61000-4-5 voltage voltage should be that of a typical business or
± 2 kV phase/neutral to ± 2 kV phase/neutral to hospital environment.
ground voltage ground voltage
Voltage dips, short < 5% UT < 5% UT The quality of the supply voltage
interruptions and (> 95% dip of UT) for (> 95% dip of UT) for should be that of a typical business or
voltage variations 0.5 cycle 0.5 cycle hospital environment. If the user of
as per S100 / OPMI pico requires continued
IEC 61000-4-11 function even in the event of inter-
40% UT 40% UT
ruptions in the power supply, we rec-
(60% dip of UT) for 5 cy- (60% dip of UT) for 5
ommend to power the device S100 /
cles cycles
OPMI pico from an uninterruptible
power supply or a battery.
70% UT 70% UT
(30% dip of UT) for 25 (30% dip of UT) for 25
cycles cycles
< 5% UT < 5% UT
(95% dip of UT) for 5s (95% dip of UT) for 5s
Power frequency 3 A/m 3 A/m Power frequency magnetic fields
(50/60Hz) mag- should be at levels typical of business
netic field as per and hospital environments.
IEC 61000-4-8
NOTE: UT is the AC supply voltage prior to application of the test level.

Version 7.0
G-30-1781-en Page 247
System Data S100 / OPMI pico

Electromagnetic immunity for non-life-supporting ME equipment


and ME systems

The device S100 / OPMI picois intended to be operated in one of the electro-
magnetic environments specified below. The customer or the user of the de-
vice S100 / OPMI pico is responsible for ensuring that it is operated in such an
environment.

Interference re- IEC 60601 Compliance level Electromagnetic environment -


sistance test level Guidelines
Tests
Portable and mobile radio communication
equipment should not be used closer to the de-
vice S100 / OPMI pico, including cables, than the
recommended safety distance that is calculated
using the equation applicable to the transmis-
sion frequency involved.
Recommended safety distance:
Conducted RF dis- 3 Vactual value 3V
d = 1, 2 P
turbances as de-
150 kHz to 80 MHz
scribed in
EN 61000-4:-6
Radiated RF distur- 3 Vactual value 3 V/m d = 1, 2 P for 80 MHz to 800 MHz
bances as de-
80 MHz to 2.5 GHz d = 2,3 P for 800 MHz to 2.5 GHz
scribed in
EN 61000-4:-3
where P is the output power rating of the trans-
mitter in watts (W) according to the transmitter
manufacturer's specifications and d is the rec-
ommended safety distance in meters (m). The
field strength of stationary transmitters as deter-
mined by a site survey should bea less than the
compliance level in all frequency ranges.b
Interference may occur in the vicinity of equip-
ment marked with the following symbol.

Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is influ-
enced by absorption and reflection by structures, objects and persons.

Version 7.0
Page 248 G-30-1781-en
S100 / OPMI pico System Data

Interference re- IEC 60601 Compliance level Electromagnetic environment -


sistance test level Guidelines
Tests
a In theory, the field strengths of stationary transmitters such as base stations for cell phones and mobile land
radio equipment, amateur radio stations, AM and FM radio and TV transmitters cannot be predicted accu-
rately. To assess the electromagnetic environment with respect to stationary RF transmitters, a site study of
the electromagnetic phenomena should be considered. If the field strength measured at the location where
the device is used exceeds the above compliance levels, the device must be monitored to verify proper func-
tion. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or re-
locating the ME equipment or ME system.
bField strengths should be less than 3 V/m for the 150 kHz to 80 MHz frequency range.

Version 7.0
G-30-1781-en Page 249
System Data S100 / OPMI pico

Recommended safety distances between portable and mobile


RF communication equipment and the S100 / OPMI pico

The device S100 / OPMI pico is intended for use in an electromagnetic envi-
ronment in which RF disturbances are controlled. The customer or the user of
the device S100 / OPMI pico can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communi-
cation equipment (transmitters) and the S100 / OPMI pico - depending on the
output power of the communication equipment as specified below.

Rated power of the Separation distance depending on transmission frequency [m]


transmitter [W]
150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz

d = 1, 2 P d = 1, 2 P d = 2, 3 P

0.01 0.12 0.12 0.23


0.1 0.38 0.38 0.73
1st 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed in the above table, the recommended separa-
tion distance d in meters (m) can be determined using the equation indicated for each column, with P being
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufac-
turer's specifications.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
influenced by absorption and reflection by structures, objects and persons.

Version 7.0
Page 250 G-30-1781-en
S100 / OPMI pico System Data

System upgrades

The device S100 / OPMI pico can be extended with the accessories offered by
ZEISS (see ordering data/accessories) to form an already tested system. This
system combination has been approved by ZEISS.
Additional devices that are connected to electronic medical devices must
comply with the relevant standards and guidelines
(for example IEC 60950 for data processing devices).
Furthermore, any configurations must comply with normative requirements
for electronic medical devices (see IEC 60601-1:2005, chapter 16; in some
countries IEC 60601-1-1:2000 may still be valid).
The one who is connecting additional devices to electronic medical devices
acts as system configurer and is responsible for the system complying with the
normative requirements for electronic medical devices.
Please note that local legislation takes priority over the above-mentioned nor-
mative requirements. If you have any questions, please contact your local
dealer or ZEISS Service.

Ordering data

Operate the device only with accessories included in the scope of delivery and
recommended by ZEISS. The contact partners in charge of your country are
listed on the Internet on the following website:
www.zeiss.com/med

For authorized ZEISS partners, the contractual product configurations, acces-


sories and spare parts apply.
Please contact your authorized ZEISS partner for more information.

Illumination
Description Order number
Halogen lamp, 12 V, 100 W, 380079-9040-000

Version 7.0
G-30-1781-en Page 251
System Data S100 / OPMI pico

Illumination
Description Order number
Tinted UV protective eyewear 000000-0480-028
N-xenon lamp module incl. 2 xenon lamps in trans- 304977-9036-700
port case for Superlux 180

Couplings
Description Order number
120° coupling for S100 suspension system 000000-1080-351
K 0/120 coupling for use during gynecological ex- 305256-9903-000
aminations

Tubes
Description Order number
Sleeve for inclined and straight tubes 305542-0107-000
Straight binocular tube, f = 170 mm 303765-0000-000
45° inclined binocular tube 303784-0000-000
180° swiveling binocular tube 303791-0000-000
180° swiveling binocular tube, f = 200 mm 303792-0000-000
Foldable tube f170/f260 incl. rotators and 303771-9020-000
PROMAG
(not for use with a mouth switch)

Eyepieces
Description Order number
10x push-in widefield eyepiece 305542-0000-000
12.5x push-in widefield eyepiece, asph. 305543-9901-000

Objective lens
Description Order number
Focusing objective lens, f = 200 mm 000000-1159-867
Focusing objective lens, f = 250 mm 000000-1160-302
Focusing objective lens, f = 300 mm 000000-1176-319
Objective lens with fixed focal distance, 305132-0000-000
f = 200 mm

Version 7.0
Page 252 G-30-1781-en
S100 / OPMI pico System Data

Objective lens
Description Order number
Objective lens with fixed focal distance, 305705-0000-000
f = 250 mm
Objective lens with fixed focal distance, 305137-0000-000
f = 300 mm
Objective lens with fixed focal distance, 305707-0000-000
f = 350 mm
Objective lens with fixed focal distance, 305146-0000-000
f = 400 mm

Hand grips
Description Order number
Handgrip, complete, rear 000000-1058-010
Handgrip, complete 000000-1057-543
Comfort handgrip for OPMI pico 303335-9006-000
Comfort T handgrip 305929-8300-000
Splash protection
Description Order number
Drape adapter ring for VisionGuard 000000-1366-769
VisionGuard replacement lenses (sterile, 20/Box) 306001-0000-000
Vision Guard lens protectors (non-sterile, 10/box) 306002-0000-000

Asepsis
Description Quanti- Order number
ty
Sterile handgrip sleeves for OPMI pico 2 pairs 305810-9004-000
Sterile handgrips (rotatable through 360°) pack of 305810-9016-000
2
Asepsis caps 12 mm pack of 305810-9002-000
(resterilizable caps) 6
Asepsis caps 22 mm pack of 305810-9001-000
(resterilizable caps) 6
Asepsis caps for PD 180° tiltable tube pack of 305810-9003-000
(resterilizable caps) 6

Version 7.0
G-30-1781-en Page 253
System Data S100 / OPMI pico

Asepsis
Description Quanti- Order number
ty
Asepsis caps 27 mm pack of 305810-9008-000
(resterilizable caps) 6
Asepsis caps 49 mm pack of 305810-9007-000
(resterilizable caps) 2

Video accessories
Video accessories can be found in the separate product overview G-30-1888.

Standard video accessories for integrated cameras


Description Order number
USB memory stick 000000-0458-366
S VHS VIDEO connecting cable, 2 m 301687-9102-000
S VHS cable, 2x 4 pins, 5 m 000000-0300-281
10 m Component HD YPbPr connecting cable 308203-3080-000
S VHS connecting cable, 2x 4 pins, 5 m Mini DIN, 308203-3070-000
sturdy and gold-plated contacts, 10 m
DVI monitor cable, 2 m 308203-3020-000
HD cable HDMI to DVI, 5 m 308203-1070-000
Monitor cable DVI, 5 m 302681-8767-000

Foot switch

Description Cat. No.


Foot switch, 10 m 000000-1179-860

Remote control
Description Order number
Remote control 308203/-2210-000
2 x LR03 alkaline batteries AAA size ---

Version 7.0
Page 254 G-30-1781-en
S100 / OPMI pico System Data

Device shelf and flat screen holder on S100


Description Order number
Device shelf for S 100 floor stand 308202-9160-000
Carrier arm for flat screen monitor 308202-9150-000
Cable cover 000000-1300-879

Accessories
Description Order number
Beam splitter 50 301513-9902-000
Angled lens without lateral output for documenta- 000000-1099-230
tion accessories
Angled lens with lateral output for 000000-1079-239
documentation accessories
Rotating adapter 301007-0000-000
(Rotating adapter for binocular tubes)
Adapter with double iris diaphragm 303354-0000-000
Stereo coobserver tube with 2 pivot joints 000000-1063-869
Long coobservation tube with two 302706-0000-000
pivot joints

Spare parts and cleaning


Description Order number
Dust cover, blue, with ZEISS logo 000000/-1055-278
Optics cleaning kit 000000/-1216-071
3.15 A fuse + drawer 000000/-1102-199
Fuse 6,3A + KEA drawer 000000/-1102-202
Fuse T1, 6 A L, 250 V, 5 x 20 mm, IEC 127-2/3 000000/-0127-023

Version 7.0
G-30-1781-en Page 255
System Data S100 / OPMI pico

Ordering information for country-specific power cables for


S100 floor stand

Description Length of power cable Order number


Europe 6m 000000-0135-487
USA 4.5 m 397902-9066-000
UK 6m 309850-9003-000
Switzerland 6m 000000-0426-193
Argentina 5m 000000-0434-527
China 6m 000000-0475-509
Brazil 6m 000000-0527-729

To upgrade your system, we recommend the following:

Description Order number


UC: MORA interface 303950-9001-500
UC MORA interface with lateral output port 303950-9002-500
for documentation accessories
UC Xenon Superlux lighting for S100 / OPMI pico 000000-1326-941
UC HD 1080P video camera and control unit for in- 308203-1060-500
tegration in S100 / OPMI pico
UC Illumination LED 305929-9010-510
UC Varioscope 100 305929-9001-510
UC integrated video camera live & streaming (con- 305929-9021-513
trol unit)
UC integrated video camera live & streaming (con- 305929-9021-511
trol unit / camera)
SW license for recording 305929-9021-000

These upgrade kits require a service call which must be coordinated with your
local ZEISS service partner.

Version 7.0
Page 256 G-30-1781-en
S100 / OPMI pico System Data

Environmental conditions

During operation

Feature Permissible range


Temperature
without HD camera + 10°C ... + 40°C
with HD camera + 10°C ... + 35°C
Rel. humidity 30% ... 75%
Air pressure 700 hPa ... 1060 hPa

For transportation and storage

Feature Permissible range


Temperature - 40°C ... + 70°C
Rel. humidity 10% ... 90%
(without condensa-
tion)
Air pressure 500 hPa ... 1060 hPa

Version 7.0
G-30-1781-en Page 257
System Data S100 / OPMI pico

Symbols for transportation and storage

Symbol Meaning
This way up
Indicates the correct upright position of the
package.
Fragile, handle with care

Keep dry

Do not stack

Permissible temperature range


Temperature limits for storage and operation
Number of packaging units

1
Permissible range of relative humidity
Humidity limits for storage and operation
Permissible range of air pressure
Air pressure limits for storage and operation

Version 7.0
Page 258 G-30-1781-en
S100 / OPMI pico System Data

Regulatory information

Devices: S100 / OPMI® pico (Floor stand) (000000-1403-542)


S100 / OPMI® pico (Floor stand) (000000-1403-544)
S100 / OPMI® pico (Ceiling mount) (000000-1403-572)
S100 / OPMI® pico (Centro suspension system) (000000-1403-573)
S100 / OPMI® pico (Support unit) (000000-1420-191)

Electrical safety The medical device complies with the requirements


of
– IEC 60601-1:2005 + CORR. 1 (2006) +
CORR. 2 (2007),
– IEC 601-1:1988+A1:1991+A2:1995,
– CAN/CSA-C22.2 No. 601.1
Classification according to the level of protection
against electrical shock: Protection class I
Protection type (according to IEC 60529): IP X0
Electromagnetic com- According to IEC 60601-1-2:2007 the medical de-
patibility vice is classed as Class B (according to CISPR 11).
Classification The medical device is classed under Appendix IX of
the Medical Devices Directive 93/42/EEC as Class 1.
CE marking The medical device complies with the essential re-
quirements stipulated in Appendix I of the Medical
Device Directive 93/42/EEC.
In line with Appendix XII of the Medical Device Di-
rective 93/42/EEC the medical device is labeled
with

.
RoHS compliance The product is RoHS compliant in accordance with
Directive 2011/65/EU.

Version 7.0
G-30-1781-en Page 259
System Data S100 / OPMI pico

Devices S100 / OPMI® pico (Carrier unit for external lighting) (000000-1095-094)
S100 / OPMI® pico (Atmos support unit) (000000-1080-349)

Electrical safety The medical device complies with the requirements


according to
– IEC 60601-1:2005 + CORR. 1 (2006) +
CORR. 2 (2007),
– IEC 601-1:1988+A1:1991+A2:1995
Classification according to the level of protection
against electrical shock: Protection class I
Protection type (according to IEC 60529): IP X0
Electromagnetic com- According to IEC 60601-1-2:2007 the medical de-
patibility vice is classed as Class B (according to CISPR 11).
Classification The medical device is classed under Appendix IX of
the Medical Devices Directive 93/42/EEC as Class 1.
CE marking The medical device complies with the essential re-
quirements stipulated in Appendix I of the Medical
Device Directive 93/42/EEC.
In line with appendix XII of the Medical Device Di-
rective 93/42/EEC the medical device is labeled
with

Version 7.0
Page 260 G-30-1781-en
S100 / OPMI pico Indexes

Indexes

List of technical terms

Term Explanation
Apochromatic optics Optical system with maximum correction of chro-
matic aberration
Drapes Sterile covers for surgical microscopes
FlexioStill Adapter for the attachment of digital compact
cameras to ZEISS surgical microscopes
FlexioMotion Adapter for the attachment of camcorders to ZEISS
surgical microscopes
Interlace (i) In interlace mode, the even and uneven-numbered
lines are alternatively set up. Todays video stand-
ards are based on this procedure.
Progressive (p) In Progressive mode, images are displayed in full
frames. This way, the line structure cannot be rec-
ognized and interline flicker is prevented.
Radiation intensity Radiation incident or emitted at right angles per
unit time and unit area.
VisionGuard Lens guard for ZEISS surgical microscopes

Version 7.0
G-30-1781-en Page 261
Indexes S100 / OPMI pico

List of abbreviations

Term Explanation
AG Aktiengesellschaft (German acronym for "stock
company")
BEV German acronym for "special emergency backup
power supply"
BNC Bayonet Neill Concelmann
(coaxial bayonet-locking connector for high fre-
quencies, named after its inventor)
CCU Camera Control Unit
(Video camera control unit)
CE Communauté Européenne (European Community)
- The manufacturer declares that the device com-
plies with the directives of the European Union.
DIN Deutsches Institut für Normung (German stand-
ards association)
D Diopter (dioptric power of optical systems)
DVI Digital Video Interface
(digital interface for video data)
EMC Electromagnetic Compatibility - defines non-inter-
ference of electrical and electronic devices with
their environment
EN European Norm (European Standard)
H Breaking capacity of the instrument fuse
HD High Definition
(providing high resolution)
HDMI High Definition Multimedia Interface
(digital interface for video and audio data)
RF Radio frequency
i interlace (see technical term index)
IEC International Electrotechnical Commission
IOL Intraocularlens - artificial lens for the eye

Version 7.0
Page 262 G-30-1781-en
S100 / OPMI pico Indexes

Term Explanation
JPEG JPEG File Interchange Format
(file format for the storage of images)
LED Light Emitting Diode
(LED)
NTSC National Television Systems Committee
(US institution defining the first color television
standard)
OR Surgical operation
OPMI Surgical microscope
PAL Phase Alternating Line
(color encoding system in analog TV systems)
p progressive (see technical term index)
PD Pupillary distance
SD Standard Definition
SDTV Standard Definition TeleVision
(TV standard providing a lower resolution than
HDTV, e.g. PAL or NTSC)
T* T* multi-coating - minimizes reflection
TIFF Tagged Image File Format
(file format for the storage of images)
UV UltraViolet
VESA Video Electronics Standards Association
(Device to mount flat screen monitors on the wall
or ceiling in accordance with the VESA standard.
Y/C Analog interface that transmits separate brightness
signals (luminance Y) and color signals (chromi-
nance C).
YPbPr Color model that transmits brightness information
Y and color difference information Pb and Pr sep-
arately.

Version 7.0
G-30-1781-en Page 263
Indexes S100 / OPMI pico

Version 7.0
Page 264 G-30-1781-en
S100 / OPMI pico

Keyword index

Symbols

..............................................................................................................188
"Enter" key .........................................................................................78, 82

Numerics

1080 Mode
Interlace ......................................................................................156
Progressive ..................................................................................156
1080p ....................................................................................................199
720 Mode ..............................................................................................156
720p ......................................................................................................199

Activating the recording license ..............................................................194


active profile ...........................................................................................187
Adjusting the color saturation (chroma) ..................................................149
Adjusting the eyepieces ..........................................................................104
Adjusting the flat screen monitor bracket ...............................................122
Adjusting the interocular distance ...........................................................104
Adjustment screw .....................................................................................66
Adjustment screw for mobility ..................................................................66
Adjustment screw for weight compensation .............................................66
Advanced ...............................................................................................146
Advanced submenu HD Video ................................................................146
Ambient requirements ..............................................................................11
Anti-fogging agent ...................................................................................98
Application variant
Dentistry .....................................................................................163
ENT .............................................................................................163
Neuro/Spine ................................................................................163
Ophthalmology ...........................................................................163
Asepsis sets ............................................................................................125
Automatic (automatic brightness adjustment) .........................................167

Backup lamp is in use ...............................................................................72


Balance setting .......................................................................................124
Battery
change ........................................................................................231

Version 7.0
G-30-1781-en Page 265
S100 / OPMI pico

Disposal ......................................................................................231
insertion ......................................................................................231
Black level submenu HD Video live & streaming ......................................184
Black Level submenu, HD video ...............................................................149
Blue submenu, HD video .........................................................................148
Brightness .................................................................................................76
brightness
HD video .....................................................................................167
HD Video submenu .....................................................................141
submenu HD Video live & streaming ............................................182
Brightness control .....................................................................................72
Brightness setting .....................................................................................70
Burn injuries caused by illumination intensity ............................................17

Cable cover .............................................................................................120


Camera head ............................................................................................86
Camera head, connector ...........................................................................86
Camera profiles submenu HD Video live & streaming ..............................187
Capture button .........................................................................................80
Carrier arm ...............................................................................................66
Cat. No. ..................................................................................................251
Changing the battery ..............................................................................231
CHROMA ................................................................................................149
Chroma submenu, HD video ...................................................................149
Color Angle submenu, HD video .............................................................149
Color saturation submenu HD Video live & streaming .............................186
Colors .......................................................................................................78
Comfort handgrips ....................................................................................97
Config HD video .......................................................................................76
Config key ................................................................................................83
Configuring the foot-switch ....................................................................152
Configuring the network connection ......................................................197
Connecting the computer (LAN option) ...................................................114
Connection for the camera head ...............................................................86
Connections ............................................................................................108
Connector panel
Halogen light source .....................................................................71
Xenon light source ........................................................................74
Control keys ........................................................................................81, 82
control keys ........................................................................................78, 80
Control unit button functions (CCU) ..........................................................76
Controls
Tubes ............................................................................................60
Widefield eyepieces .......................................................................59
Controls and connections .........................................................................54

Version 7.0
Page 266 G-30-1781-en
S100 / OPMI pico

Correcting the black level .......................................................................149


Correcting the zoom level .......................................................................150
Coupling
120° coupling ...............................................................................66
K 0/120 .........................................................................................66
Cover cap .................................................................................................56
Creating individual images ......................................................................205

Date and time


HD Video live & streaming submenu ...........................................191
Deleting individual images ......................................................................208
Deleting photos ......................................................................................208
Deleting photos and videos ....................................................................208
Deleting videos .......................................................................................208
DHCP (dynamic IP address) .....................................................................195
Diameter of luminous field .....................................................................235
Diopters
Scale .............................................................................................59
Setting ring ...................................................................................59
Display window for rated voltage .............................................................71
Display ZEISS logo ..................................................................................200
Displaying photos ...................................................................................207
Disposal symbol ............................................................................40, 43, 46
Documentation and display device ...........................................................69
Documentation unit on S100 floor stand (option) .....................................69
Documentation workstation .....................................................................69
Drapes ....................................................................................................125
Drilled holes for fastening accessories .......................................................56
Dust cover ................................................................................................54
DVI output ...............................................................................................86
DVI/HDMI OUTPUT submenu, HD video ..................................................159

Electrical connections, making ................................................................120


Electrical safety .........................................................................................11
Electrical standard ..................................................................................235
Electrician ...............................................................................................120
EMC (electromagnetic compatibility) .......................................................245
Emergency backup power supply .............................................................11
Ergonomic position .................................................................................116
Essential performance .............................................................................234
Examples of possible connections ...........................................................112
Exchanging the lamp module .................................................................226
Exchanging the xenon lamp module .......................................................226

Version 7.0
G-30-1781-en Page 267
S100 / OPMI pico

Exposure measurement ...........................................................................143


Eyecup ......................................................................................................59
Eyepiece optics, fog-free ...........................................................................98
Eyepieces ................................................................................................252
Eyepieces, Technical Data .......................................................................235

Factory Settings ......................................................................................154


Factory Settings submenu, HD video .......................................................154
Field of application .....................................................................................9
Filter selection ...........................................................................................54
First use ....................................................................................................13
Flap ..........................................................................................................70
Flat screen holder as individual component ...............................................69
Flat screen holder on S100 floor stand ......................................................69
Focusing range .......................................................................................235
Fogging of optical surfaces .......................................................................98
Friction adjustment button ..................................................................54, 56
Full frames ..............................................................................................156
Function key (F1) ......................................................................................80
Function key (F2) ......................................................................................80
Function key F1 HD Video live & streaming menu ...................................190
Function key F2 HD Video live & streaming .............................................190
Fuse ........................................................................................................228
Fuse change ...........................................................................................228
Fuse rating ..............................................................................................228

General settings submenu HD Video live & streaming .............................190

Halogen illumination system .....................................................................70


Hazard symbols ..........................................................................................8
HD Video live & streaming
Function key F2 ...........................................................................190
HD Video live & streaming menu
Active profile ...............................................................................187
Auto level ....................................................................................184
Black level ...................................................................................184
Brightness ...................................................................................182
Camera .......................................................................................182
Color saturation ..........................................................................186
Date / Time .................................................................................191
DVI / HDMI output ......................................................................200

Version 7.0
Page 268 G-30-1781-en
S100 / OPMI pico

Export .........................................................................................192
Export / Import ............................................................................193
exposure time .............................................................................184
Function key F1 ...........................................................................190
General .......................................................................................190
HDTV format ...............................................................................199
hue .............................................................................................186
Image .........................................................................................185
Image format ..............................................................................188
Index ...................................................................................207, 208
IP settings ...................................................................................195
Language ....................................................................................190
Lightmeter ..................................................................................183
Main menu .................................................................................176
mode ..........................................................................................182
Network share ............................................................................195
Orientation .................................................................................185
Peak / Average ............................................................................184
Profiles ........................................................................................187
Recording ...................................................................................188
Recording license ........................................................................194
Recording mode ..........................................................................188
SDTV format ...............................................................................199
Settings .......................................................................................181
Software .....................................................................................192
Special ........................................................................................192
Streaming ...........................................................................188, 196
Version .......................................................................................192
Video ..........................................................................................199
Video quality ...............................................................................188
Video standard ............................................................................200
Web interface .............................................................................196
White balance .............................................................................185
Zoom ..........................................................................................185
HD Video live & streaming settings .........................................................181
HD Video menu
brightness ...................................................................................141
configuration menu ....................................................................155
HDTV format ..........................................................................................199
HDTV sub menu SD Video ......................................................................156
Hex key ..................................................................................................116
Hue submenu HD Video live & streaming ................................................186

Illumination off .........................................................................................66


Image .......................................................................................................78

Version 7.0
G-30-1781-en Page 269
S100 / OPMI pico

freeze (freeze frame) ...................................................................153


HD video .....................................................................................168
rotate 180° .................................................................................153
Image format submenu HD Video live & streaming .................................188
Image orientation ...................................................................................145
Image submenu HD Video live & streaming .............................................185
Index button .......................................................................................80, 83
Index in the web interface ......................................................................202
Indicator light ...........................................................................................76
Information symbols ...................................................................................8
Inserting the battery ...............................................................................231
Installing handgrips ...................................................................................97
Installing the recording license ................................................................204
Installing the SW recording license ..........................................................204
Instrument tray as individual component ..................................................69
Instrument tray on S100 floor stand .........................................................69
Integrated HD video camera ...................................................................134
Integrated HD video camera live & streaming ..........................................170
Intended use ...............................................................................................9
IP settings (static) ....................................................................................197
IP settings submenu HD Video live & streaming ......................................195

Key to symbols ...........................................................................................7

Lamp failure
Halogen ......................................................................................217
xenon ..........................................................................................218
LAN input/output ......................................................................................86
LED lighting system ...................................................................................21
Legal regulations ......................................................................................11
Letterbox ........................................................................................160, 199
Light guide socket ........................................................................71, 74, 75
Light guide, connection socket ...........................................................54, 75
Light guide, socket .............................................................................71, 74
Light Source submenu, HD video ............................................................162
Lightmeter sub menu SD Video ...............................................................143
Line fuses ................................................................................................228
Live ...........................................................................................................82
live ............................................................................................................78
Live video image of HD video ..................................................................168
Locking screws .......................................................................................116
Locking tab ...............................................................................................66
Low Light Auto (semiautomatic brightness adjustment) ..........................167

Version 7.0
Page 270 G-30-1781-en
S100 / OPMI pico

MAC Address .........................................................................................195


Magnification .............................................................................54, 56, 235
Magnification changer ........................................................................54, 56
Magnification factor ...............................................................................235
Manual (manual brightness adjustment) .................................................167
Manual switching to the backup lamp ................................................70, 72
Manufacturer .............................................................................................3
Maximum load
flat screen monitor bracket ...........................................................28
instrument tray .............................................................................28
Menu display ..................................................................................134, 170
Menu HD Video
User ....................................................................................146, 152
Menu HD video
Black Level ..................................................................................149
Blue ............................................................................................148
Chroma .......................................................................................149
Color Angle .................................................................................149
DVI/HDMI OUTPUT .....................................................................159
Factory Settings ..........................................................................154
Light Source ................................................................................162
Orientation .................................................................................145
PAL / NTSC Selection ...................................................................161
Peak / Average ............................................................................150
Red .............................................................................................148
SDTV Format ...............................................................................160
Version Information ....................................................................154
Zoom Level .................................................................................150
Menu key .................................................................................................83
Menu SD Video
HDTV ..........................................................................................156
Lightmeter ..................................................................................143
User ............................................................................................139
Menu SD video
Remote Setup .............................................................................152
Menu structure of the HD video camera live & streaming .......................170
Mounting the hand grips ..........................................................................97

Navigation ......................................................................................134, 170


Network share ................................................................................195, 198
Normal use .................................................................................................9

Version 7.0
G-30-1781-en Page 271
S100 / OPMI pico

Object focal distance ..............................................................................235


Objective lens ...........................................................................................56
Opening the configuration menu ............................................................176
HD video .............................................................................137, 155
Opening the lamp module ........................................................................72
Opening the Main menu .........................................................................176
Opening the slide-in lamp module ............................................................70
Operating elements (overview of HD video camera live & streaming) ........84
Operation ...............................................................................................127
OPMI pico with MORA coupling ...............................................................62
Optical surfaces ........................................................................................98
Optical surfaces, cleaning .......................................................................229
Ordering data .........................................................................................251
Orientation submenu HD Video live & streaming .....................................185
Orientation submenu, HD video ..............................................................145
Overview of HD video camera live & streaming operating elements ..........84

PAL / NTSC Selection submenu HD video ................................................161


Peak / Average submenu HD Video live & streaming ...............................184
Peak / Average submenu, HD video ........................................................150
Play button .........................................................................................80, 83
Potential equalization ...................................................................71, 74, 75
Power inlet ...............................................................................................71
Power switch ................................................................................71, 74, 75
Powering the system up/down ................................................................128
Procedure ...............................................................................................132
Protective ground conductor ....................................................................13
Protective ground conductor, properly connected ...................................120

Rated voltage, indicator window ..............................................................71


Rating label ...............................................26, 30, 32, 34, 36, 38, 40, 43, 46
Rating plate ..............................................................................................40
Record button ..........................................................................................83
Recording HD videos ...............................................................................205
Recording license ....................................................................................194
Recording mode submenu HD Video live & streaming
...................................................................................................188
Recording submenu HD Video live & streaming .......................................188
Red submenu, HD video .........................................................................148
Remote (AUX)socket (HD) .........................................................................86
Remote control (option), HD video ............................................................82

Version 7.0
Page 272 G-30-1781-en
S100 / OPMI pico

Remote control battery ...........................................................................231


Remote control key functions ...................................................................82
Remote Setup .........................................................................................152
Remote Setup sub menu HD video .........................................................152
Replacing the halogen lamp ...................................................................224
Reset
User ............................................................................................154
Reset Camera .........................................................................................154
Resolution
1280 x 720 .................................................................................156
1920 x 1080 ...............................................................................156
Retrofit kits .............................................................................................253
Ring diaphragm ........................................................................................62
Routing the cables ..................................................................................120

S100 floor stand with documentation and display device .........................69


Safety
checks .........................................................................................223
Devices .........................................................................................19
Safety equipment
LED light source ............................................................................21
Safety system
Halogen light source .....................................................................19
Xenon light source ........................................................................20
SDTV format ...........................................................................................199
SDTV Format submenu, HD video ...........................................................160
Securing screw .........................................................................................56
Security recommendations for networking .............................................196
Select control keys ....................................................................................78
Service ............................................................................................120, 152
Service submenu HD Video .....................................................................152
Setting the brightness of the HD Video ...................................................167
Setting the suspension arm weight compensation ..................................103
Side-Cut .................................................................................................160
Sidecut ...................................................................................................199
SIP label ............................................................................26, 30, 32, 34, 36
Slide-in lamp module ................................................................................70
Software Update ....................................................................................192
Standby position (illumination off) ............................................................66
Sterilization ............................................................................................230
Strain relief device ........................................................................71, 74, 75
Streaming .......................................................................................196, 198
Supply unit ...............................................................................................66
Surgical microscope
Configuration ................................................................................91

Version 7.0
G-30-1781-en Page 273
S100 / OPMI pico

Surgical microscope, adjustment .............................................................104


Suspension arm ........................................................................................66
Symbols and labels on the device ..............................................................22
System overview .................................................................................51, 52

Target group ...............................................................................................9


Test pattern ....................................................................................134, 170
Touchscreen .............................................................................................69
Trademarks .................................................................................................2
Transport ..................................................................................................12
Transport position .....................................................................................28
Tube and eyepieces for main microscope ..................................................96
Tubes, Technical Data .............................................................................235

Upgrade component ...............................................................................253


USB port ...................................................................................................86
User
Anterior .......................................................................................163
Endodonty ..................................................................................163
Posterior ......................................................................................163
Treatment ...................................................................................163
User submenu SD Video ..........................................................................139

Ventilation grid ...................................................................................70, 72


Version Information submenu, HD video .................................................154
Video ......................................................................................................199
fast forward ..................................................................................81
rewind ...........................................................................................81
skip back .......................................................................................81
skip forward ..................................................................................80
Video camera, connecting ......................................................................110
Video connection socket ...........................................................................54
Video connection, socket ..........................................................................54
Video output
Composite (FBAS) ..........................................................................86
HDMI ............................................................................................86
YPbPr ............................................................................................86
Video playback .......................................................................................207
Video quality submenu HD Video live & streaming ..................................188
Video recording ......................................................................................205
Video signal output ports Y/C ...................................................................86

Version 7.0
Page 274 G-30-1781-en
S100 / OPMI pico

Viewing individual images ......................................................................207


Viewing photos ......................................................................................207
Viewing videos .......................................................................................207

Wall flange ...............................................................................................66


Wall stand ................................................................................................66
Web control ...................................................................................196, 198
Web interface .........................................................................196, 198, 201
Create patient .............................................................................202
Index ...........................................................................................202
Rename patient ...........................................................................202
start ............................................................................................201
Weight ...................................................................................................235
Weight compensation, adjustment screw .................................................66
White balance ..........................................................................................76
White balance (white bal) HD video ........................................................165
White balance submenu HD Video live & streaming ................................185

Xenon illumination system ........................................................................72

Zoom Level .............................................................................................150


Zoom Level submenu, HD video .............................................................150
Zoom submenu HD Video live & streaming .............................................185

Version 7.0
G-30-1781-en Page 275
S100 / OPMI pico

Version 7.0
Page 276 G-30-1781-en
S100 / OPMI pico

(Blank page, for your notes . . .)

Version 7.0
G-30-1781-en Page 277
Carl Zeiss Meditec AG
Goeschwitzer Strasse 51-52 Fax: + 49 (0)7364 - 20 4823
07745 Jena E-Mail: surgical@meditec.zeiss.com
Germany Internet: www.zeiss.com/med