Professional Documents
Culture Documents
EMG/EP
Measuring System
MEB-9400
MEB-9400A
MEB-9400K
Trademarks
Microsoft and Windows are registered trademarks of Microsoft Corporation.
Other models and trademarks are the property of their respective owners.
Contents 1
2
GENERAL HANDLING PRECAUTIONS.............................................................................. i
3
WARRANTY POLICY.......................................................................................................... ii
EMC RELATED CAUTION...................................................................................................iii 3-1
Conventions Used in this Manual and Instrument............................................................... v
Warnings, Cautions and Notes.................................................................................. v 3-2
Explanations of the Symbols in this Manual and Instrument.................................... vi
3-3
Introduction....................................................................................................................... 1.2
4
Installing Measurement Programs.................................................................................... 1.3
Important Safety Information............................................................................................ 1.4
4-1
4-2
Section 2 Electrode/Accessory Handling............................................. 2.1
EMG Needle Electrode..................................................................................................... 2.3 4-3
Type of Electrode.................................................................................................... 2.4
Electrode Attachment and Connection Procedure................................................. 2.5 4-4
3-2
Section 3-4 Calculation........................................................................................... 3.4.1
General.......................................................................................................................... 3.4.2 3-3
Marking.......................................................................................................................... 3.4.3
Manual Marking................................................................................................... 3.4.3 3-4
Setting a Mark at a Time........................................................................... 3.4.3
Moving a Mark..................................................................................................... 3.4.5 4
Deleting Marks..................................................................................................... 3.4.5
Deleting an Individual Mark....................................................................... 3.4.5 4-1
Deleting All Marks..................................................................................... 3.4.5
Measurement Table Window.......................................................................................... 3.4.6 4-2
To Open the Measurement Table Window........................................................... 3.4.7
To Close the Measurement Table Window........................................................... 3.4.7 4-3
6-3
Section 4 Auditory Evoked Potential.................................................. 4.0.1
Section 4-1 Measurement – Basic.......................................................................... 4.1.1
Operation Flow............................................................................................................... 4.1.2
Entering the Patient Information.................................................................................... 4.1.4
Selecting the Examination Item..................................................................................... 4.1.5
Entering the Examination Information............................................................................ 4.1.6
Checking and Changing Measurement Settings............................................................ 4.1.7
Diagnosis................................................................................................. 4.5.19
I-L Curve Window........................................................................................................ 4.5.20
Displaying the I-L Curve.......................................................................... 4.5.20
Creating a Normative Data File............................................................... 4.5.21
Saving Data on Disk.................................................................................................... 4.5.23
Condition Table Window............................................................................................... 4.5.23
Annotating a Waveform................................................................................................ 4.5.24
7-8
Section 7 Nerve Conduction Study.................................................... 7.0.1
Section 7-1 Nerve Conduction Study – Basic....................................................... 7.1.1 7-9
General.......................................................................................................................... 7.5.2
7-2
Nerve Conduction Velocity Calculation.......................................................................... 7.5.3
Calculation with Cursors...................................................................................... 7.5.3
7-3
Calculation with Marks......................................................................................... 7.5.5
Marking.......................................................................................................................... 7.5.6 7-4
Auto Marking During Acquisition......................................................................... 7.5.6
Marking After Acquisition..................................................................................... 7.5.6 7-5
Setting One Mark at a Time....................................................................... 7.5.6
Automatically Setting Multiple Marks According to the Settings on 7-6
the Labeling Page...................................................................................... 7.5.8
Moving a Mark..................................................................................................... 7.5.8 7-7
Deleting Marks..................................................................................................... 7.5.9
Deleting an Individual Mark....................................................................... 7.5.9 7-8
Deleting All Marks..................................................................................... 7.5.9
7-9
Measurement Table Windows...................................................................................... 7.5.10
To Open the Measurement Table Window......................................................... 7.5.10
7-10
To Close the Measurement Table Window......................................................... 7.5.11
Measurement Table Window Type I Description.......................................................... 7.5.12
7-11
SCS Measurement Table Window..................................................................... 7.5.12
Annotating a Waveform........................................................................... 7.5.12
7-12
NCS Measurement Table Window..................................................................... 7.5.13
Entering Segment Waveforms................................................................. 7.5.13 7-13
Entering Normative Data......................................................................... 7.5.14
Entering Skin Temperature...................................................................... 7.5.14 7-14
Measurement Table Window Type II Description......................................................... 7.5.15
SCS Measurement Table Window..................................................................... 7.5.15 8
Entering the Site...................................................................................... 7.5.16
Changing the Waveform Order................................................................ 7.5.16
Displaying Normative Range................................................................... 7.5.16
Creating the CSV File.............................................................................. 7.5.17
Saving Data on Disk.................................................................................................... 7.5.18
Condition Table Window............................................................................................... 7.5.18
Annotating a Waveform................................................................................................ 7.5.19
General.......................................................................................................................... 7.9.2
7-3
F-wave Conduction Velocity Measurement.................................................................... 7.9.3
Calculation with Cursors...................................................................................... 7.9.3
7-4
Calculation with Marks......................................................................................... 7.9.5
Marking.......................................................................................................................... 7.9.6
7-5
Auto Marking During Acquisition......................................................................... 7.9.6
Marking After Acquisition..................................................................................... 7.9.6 7-6
Setting One Mark at a Time....................................................................... 7.9.6
Setting Multiple Marks Automatically According to the Settings on 7-7
the Labeling Page...................................................................................... 7.9.8
Moving a Mark..................................................................................................... 7.9.8 7-8
Deleting Marks..................................................................................................... 7.9.9
Deleting an Individual Mark....................................................................... 7.9.9 7-9
Deleting All Marks..................................................................................... 7.9.9
F-wave Measurement Table Window........................................................................... 7.9.10 7-10
To Open the Measurement Table Window......................................................... 7.9.10
To Close the Measurement Table Window......................................................... 7.9.10 7-11
Please read these precautions thoroughly before attempting to operate the instrument.
1. To safely and effectively use the instrument, its operation must be fully understood.
3. Before Operation
(1) Check that the instrument is in perfect operating order.
(2) Check that the instrument is grounded properly.
(3) Check that all cords are connected properly.
(4) Pay extra attention when the instrument is in combination with other instruments to avoid misdiagnosis or other
problems.
(5) All circuitry used for direct patient connection must be doubly checked.
(6) Check that battery level is acceptable and battery condition is good when using battery-operated models.
4. During Operation
(1) Both the instrument and the patient must receive continual, careful attention.
(2) Turn power off or remove electrodes and/or transducers when necessary to assure the patient’s safety.
(3) Avoid direct contact between the instrument housing and the patient.
6. The instrument must receive expert, professional attention for maintenance and repairs. When the instrument
is not functioning properly, it should be clearly marked to avoid operation while it is out of order.
9. When the instrument is used with an electrosurgical instrument, pay careful attention to the application and/or
location of electrodes and/or transducers to avoid possible burn to the patient.
10. When the instrument is used with a defibrillator, make sure that the instrument is protected against defibrillator
discharge. If not, remove patient cables and/or transducers from the instrument to avoid possible damage.
WARRANTY POLICY
Nihon Kohden Corporation (NKC) shall warrant its products against all defects in materials and workmanship for one year
from the date of delivery. However, consumable materials such as recording paper, ink, stylus and battery are excluded
from the warranty.
NKC or its authorized agents will repair or replace any products which prove to be defective during the warranty period,
provided these products are used as prescribed by the operating instructions given in the operator’s and service manuals.
No other party is authorized to make any warranty or assume liability for NKC’s products. NKC will not recognize any
other warranty, either implied or in writing. In addition, service, technical modification or any other product change
performed by someone other than NKC or its authorized agents without prior consent of NKC may be cause for voiding
this warranty.
Defective products or parts must be returned to NKC or its authorized agents, along with an explanation of the failure.
Shipping costs must be pre-paid.
This warranty does not apply to products that have been modified, disassembled, reinstalled or repaired without Nihon
Kohden approval or which have been subjected to neglect or accident, damage due to accident, fire, lightning, vandalism,
water or other casualty, improper installation or application, or on which the original identification marks have been
removed.
CAUTION
United States law restricts this product to sale by or on the order of a physician.
The following describes some common interference sources and remedial actions:
1. Strong electromagnetic interference from a nearby emitter source such as an authorized radio station
or cellular phone:
Install the equipment and/or system at another location. Keep the emitter source such as cellular
phone away from the equipment and/or system, or turn off the cellular phone.
2. Radio-frequency interference from other equipment through the AC power supply of the equipment
and/or system:
Identify the cause of this interference and if possible remove this interference source. If this is not
possible, use a different power supply.
4. Electromagnetic interference with any radio wave receiver such as radio or television:
If the equipment and/or system interferes with any radio wave receiver, locate the equipment and/or
system as far as possible from the radio wave receiver.
5. Interference of lightning:
When lightning occurs near the location where the equipment and/or system is installed, it may induce
an excessive voltage in the equipment and/or system. In such a case, disconnect the AC power cord
from the equipment and/or system and operate the equipment and/or system by battery power, or use
an uninterruptible power supply.
If the above suggested remedial actions do not solve the problem, consult your Nihon Kohden
representative for additional suggestions.
For EMC compliance, refer to “Specification - Electromagnetic Compatibility” in the Reference section of the operator’s
manual.
The CE mark is a protected conformity mark of the European Community. Products with the CE mark comply
with the requirements of the Medical Device Directive 93/42/EEC.
WARNING
A warning alerts the user to the possible injury or death associated with the use or misuse of the
instrument.
CAUTION
A caution alerts the user to possible injury or problems with the instrument associated with its use or
misuse such as instrument malfunction, instrument failure, damage to the instrument, or damage to other
property.
NOTE
A note provides specific information, in the form of recommendations, prerequirements, alternative methods or
supplemental information.
Main unit
On screen
Window maximize/resize
Check box Maximize Resize Minimize minimize button
Introduction........................................................................................................................................................... 1.2
Installing Measurement Programs........................................................................................................................ 1.3
Important Safety Information................................................................................................................................ 1.4
Introduction
1
Installing Measurement Programs
General
WARNING WARNING
Never use the system in the presence of any Never use the system in a hyperbaric oxygen
flammable anesthetic gas or high concentration chamber. Failure to follow this warning may cause
oxygen atmosphere. Failure to follow this warning explosion or fire.
may cause explosion or fire.
WARNING WARNING
Do not diagnose patients based on only the data When using an electrically controlled
from this instrument. Overall judgement must be neurovascular intervention device such as an
performed by a physician who understands the endovascular coil, follow the instructions of the
features, limitations and characteristics of the device’s manual.
instrument by reading this operator’s manual
thoroughly and by reading the biomedical signals
acquired by other instruments.
MRI examination
WARNING WARNING
When performing MRI test, remove all electrodes Never use this system in an MRI examination
and transducers from the patient which are room. The system may not operate properly due
connected to this system. Failure to follow this to high frequency magnetic noise from the MRI.
warning may cause skin burn on the patient. For
details, refer to the MRI manual.
WARNING WARNING
Do not use the system near an ESU. Noise from When the system is used with an electrosurgical
the ESU may cause the system to malfunction. unit (ESU), firmly attach the entire area of the
ESU return plate. Otherwise, the current from the
ESU flows into the electrodes of the system,
causing electrical burn where the electrodes are
attached. For details, refer to the ESU manual.
1
WARNING WARNING
While using an ESU, needle electrodes, silver When using an ESU, do the following. Otherwise,
ball electrodes and intracranial electrodes must current from the ESU flows into the electrodes
not be used for monitoring unless no other and causes skin burn.
alternatives are available. If needle electrodes, • Ensure that there is enough distance between
silver ball electrodes and intracranial electrodes the electrodes and the ESU tip or return plate.
are used, whenever possible, disconnect • Ensure that no electrodes are attached near
electrodes from the electrode junction box while the path for high frequency current between the
using the ESU. Otherwise, current from the ESU ESU tip and return plate.
might cause skin burn where the electrodes are • If there is not enough distance between the
attached. electrodes and the ESU tip, return plate or
path for high frequency current, disconnect the
electrodes and breakout box connector from
the electrode junction box while using the ESU.
• During long term monitoring, periodically check
that the electrodes are attached properly.
WARNING
Operation in close proximity (for example 1 m) to
shortwave or microwave therapy equipment may
cause instability in the ELECTRICAL
STIMULATOR output.
WARNING WARNING
Before defibrillation, all persons must keep clear Before defibrillation, remove everything including
of the bed and must not touch the patient or any electrodes and patches from the patient s chest.
equipment or cable connected to the patient. If the defibrillator paddle contacts an object on
Failure to follow this warning may cause electrical the patient’s chest, the discharged energy may
shock or injury. be insufficient and cause skin burn.
WARNING
Before defibrillation, remove from the patient all
electrodes, probes and transducers from
connectors that do not have a or mark.
Otherwise, the operator may receive electrical
shock and the connected instrument may be
damaged.
Installation
WARNING WARNING
Only use a 3-prong power cord. Using other Equipotential grounding is required for electrical
power cords may result in electrical shock or safety. Consult with a qualified biomedical
injury to the patient and operator. engineer.
WARNING WARNING
Before connecting or disconnecting instruments, When connecting external equipment to this
make sure that each instrument is turned off and system, ensure that the external equipment
the power cord is disconnected from the AC complies with the IEC 60601-1 safety standard
socket. Otherwise, the patient or operator may for medical equipment and CISPR11 Second
receive electrical shock or injury. Edition 1990-1999 Group 1 and Class B
standard, or the equivalent. When the external
equipment does not comply with the IEC 60601-1
safety standard, use a locally available medical
use isolation transformer unit between the
external equipment and the AC outlet.
WARNING WARNING
When connecting an external instrument to the Install all network devices, including printer and
connectors marked with of the system, the hubs, outside the patient environment (IEC
external instrument and system must be 60601-1-1). If they are installed inside the patient
connected according to the IEC 60601-1-1 environment, the patient or operator may receive
“Medical electrical equipment - Part 1-1: General electrical shock or injury. For installation, contact
requirements for safety - Collateral standard: your Nihon Kohden representative.
Safety requirements for medical electrical
systems”. Failure to follow this warning may
cause electrical shock to the patient and operator.
WARNING WARNING
Do not use a damaged network cable. The patient Connect the system to network as specified.
or operator may receive electrical shock when the Otherwise the patient and operator may receive
damaged part is touched. electrical shock or injury. To connect the network,
contact your Nihon Kohden representative.
CAUTION CAUTION
Use the YS-004A9 tool set for cart assembly. Select an installation location where patient
information cannot be seen or accessed by
unauthorized personnel.
1
CAUTION CAUTION
Do not install the system near equipment with a Do not install the system where it will be exposed
high power consumption, such as large X-ray to water or chemical solutions. Avoid direct
equipment. sprinkling, spray or moist air from the nebulizer or
humidifier. These cause malfunction and shorten
the life of the system.
CAUTION CAUTION
Do not install the system near a power line, Select a room with no excessive noise, vibration,
dynamo or motor which has electromagnetic sunlight, high humidity or water splashes.
induction.
CAUTION CAUTION
Make sure that there is no influence from a Select a room with a 3-prong outlet with a ground
cellular phone. third contact.
CAUTION CAUTION
Avoid locations where the system may receive When the external noise is too large, install the
strong electromagnetic interference and high- system in a shielded room.
frequency electric interference such as radio or
TV stations, cellular phones or mobile two-way
radios. Otherwise, the display may flicker or the
displayed waveform trace may be thicker.
CAUTION CAUTION
A sudden loss of power or extreme power surge Only install the specified software in the system.
can damage data and the computer. To assure an Unspecified software may cause the system to
uninterrupted power supply, use an malfunction. Nihon Kohden has no responsibility
uninterruptable power supply (UPS). for any problem caused by unspecified software.
CAUTION CAUTION
Do not delete any files in the hard disk. The Be careful not to pinch your fingers between the
system may not start. PC unit, main unit, isolation unit, printer and cart.
CAUTION CAUTION
Use only the KD-019A cart or KD-026A cart for When moving the system, select a flat path and
this system. If the specified cart is not available move the system carefully to prevent it from
and you must use another cart, use a cart that tipping over or components falling off.
complies with IEC 60601-1. Otherwise, the cart
may tip over and the cart and system
components may break. Also secure the cables
of the instrument to ensure electrical safety and
to prevent components from falling off or cable
disconnection.
CAUTION CAUTION
• Do not move the cart with the casters locked. Do not carelessly touch connectors on the
The cart may tip over. system components. The system may
• When moving the cart, hold the handle of malfunction due to electrostatic energy. Everyone
the cart. Never hold the arm for the electrode who uses this system must be trained in the
junction box. prevention of electrostatic discharge (ESD).
CAUTION
Do not connect unspecified devices such as USB
memory to the USB connectors of this instrument
because the device may have a computer virus.
Operation
Unused connectors on the electrode junction box
WARNING
During measurement, do not touch any connector
that an electrode is not connected to. If an
instrument that is connected to the patient is
faulty, the patient and operator may get electrical
shock.
WARNING
Do not put a water cup where it can spill on the
system. It may cause electrical shock to the
patient and operator and cause system
malfunction.
WARNING WARNING
Do not use a needle electrode for electric When the needle electrode is used on a
stimulation. Failure to follow this warning injures potentially contagious patient, dispose of the
the patient because the needle electrode is needle electrode by following your facility’s
damaged by electrolyzation inside the body. regulations or local laws for disposing of the
infectious medical waste.
WARNING
To prevent infection, autoclave the needle
electrodes before and after use.
WARNING WARNING
For patients with the following conditions, all For visual stimulation always prepare mouth
stimulation (activations) must be applied under gags, tongue depressors and gauze sponges to
the supervision of the physician in charge. prevent the patient from biting his tongue or
• Patients with a history of epilepsy, cortical injuring himself during testing. Visual stimulation
lesions, convexity skull defects, raised may induce seizure activity.
intracranial pressure, neurological or psychiatric
disorders or cardiac disease
WARNING
• Patients with a history of cerebral aneurysm
Do not apply photic stimulation continuously for
treatment using surgical devices such as
more than 5 minutes in any mode. After applying
vascular clips or shunts, or a neurovascular
5 minutes of photic stimulation to the patient, wait
intervention device such as an endovascular
at least 20 minutes before another photic
coil
stimulation.
• Patients under proconvulsant medications or
anesthetics
• Patients with any type of implanted biomedical WARNING
device (e.g. intracranial electrodes, cochlear When electrical stimulation is performed for a
implant or cardiac pacemaker) or any metallic long period, the EEG paste gets dry and
fragments or plate in the head electrical stimulation may cause skin burn
For those patients, there is the possibility of the because the dried paste increases the skin-
following potential hazards: electrode contact impedance. Therefore,
• Tongue or lip laceration periodically check that the paste is moist.
• Mandibular fracture
• Seizure
• Cardiac arrhythmia
• Scalp burn
• Intraoperative awareness
WARNING
Do not touch the flash lamp assembly because it
gets very hot and can burn the skin.
WARNING WARNING
Apply the contact lens electrode to the patient’s Remove the contact lens electrode from the
eye under dark red lighting supervised by a patient’s eye after approximately every 45
physician. When attaching a contact lens minutes and give the patient a 10 minute rest. If
electrode, follow the contact lens electrode the electrode is applied for more than 45 minutes,
operator’s manual. it causes a lack of oxygen to the cornea.
WARNING WARNING
When the monitored waveform is abnormal (for • Never check the skin-electrode contact
example, the waveform is flat or hum is impedance while using a contact lens
overlapping), remove the contact lens electrode electrode. Failure to follow this warning injures
from the patient immediately. the patient’s eyes because impedance check
electrical current will be applied.
• When attaching a contact lens electrode, follow
the contact lens electrode operator’s manual.
WARNING WARNING
• If redness, swelling, itch, or rash appears on Keep collodion/celloidin and acetone away from
the patient’s skin, immediately remove the fire because they are flammable.
collodion from the skin and perform medical
treatment.
• Never allow collodion/celloidin or acetone to
get in the patient’s eyes. If collodion/celloidin
or acetone accidentally gets in the eyes,
immediately and thoroughly wash eyes with
clean water and perform medical treatment
immediately.
• If collodion is swallowed, have the person drink
water and vomit collodion. Perform medical
treatment immediately.
• Collodion is a volatile solvent. Both patients
and medical staff must take extreme care not to
inhale collodion fumes. When using collodion,
make sure there is adequate ventilation. If
excess collodion fumes are inhaled, have the
person lie quietly and keep warm in fresh air.
Perform medical treatment immediately.
WARNING
Attach the electrodes under the supervision of a
physician in charge.
WARNING
When electrical stimulation is performed for a
long period, the EEG paste gets dry and
electrical stimulation may cause skin burn
because the dried paste increases the skin-
electrode contact impedance. Therefore,
periodically check that the paste is moist.
Other
CAUTION CAUTION
When using the instrument for brain death During measurement, do not change the date
diagnosis, before examination, check and adjust and time. This makes the order of the saved
the date and time of the system. The date and event data and the time of the saved waveforms
time on the screen and recording result are part incorrect.
of important information for the medical record.
CAUTION CAUTION
Do not use the sub-dermal straight needle Never check the skin-electrode contact
electrode for more than one hour as a impedance while using a needle electrode or
measurement electrode for the EEG or evoked while any electrodes are attached at the
potential measurement. When measuring the operation site on the patient’s body. Failure to
EEG or evoked potential for over one hour, use follow this caution causes electrical burn where
the EEG disk electrode. the electrodes are inserted or attached.
CAUTION CAUTION
Before disposing of needle electrodes, check with Do not turn the power off when any program is
your local solid waste officials about proper running. When turning the system off, follow the
disposal. procedure in “Power Off Procedure” in Section 2
of the Operator’s Manual.
CAUTION CAUTION
Do not use the Windows remote desktop function. Fuses cut the power off when an abnormality
This function allows unauthorized people to occurs in the systems. Solve the cause of the
remotely access this system. malfunction before replacing the fuse. Use the
correct fuse only. The fuse rating is shown on the
holder.
CAUTION CAUTION
Turn off the stimulator and the MEB-9400 system When using the somato control box, start
power before connecting or disconnecting the stimulation at 0 mA intensity and increase
stimulator for the visual evoked potential gradually. Starting at a higher intensity may
examination. deliver a 100 mA current to the patient.
CAUTION CAUTION
Do not connect or disconnect the somato control Do not use water to clean the somato control box
box cable when the system power is on. because the box contains electrical parts.
CAUTION CAUTION
If the somato control box gets wet, stop use and The sterilization temperature must not exceed
contact your Nihon Kohden representative. 70˚C (158˚F). Electrodes and accessories may
deform or melt above this temperature.
CAUTION CAUTION
Do not wipe an electrode lead or cable with Do not use alcohol or benzine as the contact lens
alcohol. Alcohol causes the lead and cable vinyl electrode cleaner. These solutions deform the
cover to harden. electrode.
CAUTION CAUTION
Do not leave the strap type ground electrode in The current density from the electric stimulator
water or physiological salt solution. It will rust and might be over 2 mA rms/cm2 depending on the
correct measurement cannot be performed. dimensions of the electrode, width of pulse,
intensity and frequency of stimulation.
NOTE
• When using the MEB-9400 application program,
close all other programs. Otherwise, the system
program may not function properly.
• Turn off any screen saver before opening the MEB-
9400 application program.
• Periodically back up data in the system. Otherwise
you may lose data if the computer crashes.
WARNING
The QP-945BK System Program is protected by
copyright law and international treaties.
Unauthorized reproduction or distribution of this
software, or any portion of it, may result in severe
civil and criminal penalties, and will be
prosecuted to the maximum extent possible
under law.
CAUTION CAUTION
Keep the disk away from strong magnetic objects Do not insert or remove a disk while the lamp on
such as a magnet, TV set or speaker. Otherwise, the disk unit is lit.
data in the disk may be lost.
CAUTION CAUTION
Do not touch the disk surface of the recorded Keep the disk away from direct sunlight and high
side (CD-ROM: opposite side of the label side). If temperature. Otherwise, the disk may become
the surface of the disk becomes contaminated deformed.
with any foreign substances such as fingerprints,
reading data may be impossible.
CAUTION CAUTION
Do not handle the disk while smoking or eating. Do not get the disk wet.
CAUTION CAUTION
Do not put a label on top of another label. Do not write on the label after the label is
Remove the old label before applying a new label. attached on the disk. Otherwise, the disk may be
damaged and reading may be impossible.
CAUTION CAUTION
Do not bend the disk, put heavy material on the Clean the disk with a disk cleaner. Do not use
disk, or give a strong impact to the disk. organic solvents such as acetone.
CAUTION
This CD-ROM is not an audio CD and cannot be
played with an audio CD player.
Maintenance
CAUTION
Dispose of Nihon Kohden products according to
your local laws and your facility’s guidelines for
waste disposal. Otherwise, it may affect the
environment.
If there is a possibility that the product may have
been contaminated with infection, dispose of it as
medical waste according to your local laws and
your facility’s guidelines for medical waste.
Otherwise, it may cause infection.
This section describes how to attach and disinfect each electrode or accessory.
Select the suitable electrode according to the examination.
WARNING
Do not use a needle electrode for electric stimulation. Failure to
follow this warning injures the patient because the needle electrode
is damaged by electrolyzation inside the body.
WARNING
While using an ESU, needle electrodes, silver ball electrodes and
intracranial electrodes must not be used for monitoring unless no
other alternatives are available. If needle electrodes, silver ball
electrodes and intracranial electrodes are used, whenever possible,
disconnect electrodes from the electrode junction box while using
the ESU. Otherwise, current from the ESU might cause skin burn
where the electrodes are attached.
WARNING
To prevent infection, discard needle electrodes after using on a
contagious patient (e.g. hepatitis). Discard electrodes according to
your local regulations.
WARNING
To prevent infection, autoclave the needle electrodes before and
after use.
CAUTION
Do not use the sub-dermal straight needle electrode for more than
one hour as a measurement electrode for the EEG or evoked
potential measurement. When measuring the EEG or evoked
potential for over one hour, use the EEG disk electrode.
CAUTION
Never check the skin-electrode contact impedance while using a
needle electrode or while any electrodes are attached at the
operation site on the patient’s body. Failure to follow this caution
causes electrical burn where the electrodes are inserted or attached.
CAUTION
The disposable needle electrodes are single use only. Do not reuse
even after sterilization.
NOTE
Refer to the needle electrode operator’s manual when using needle
electrodes.
Examination Guide MEB-9400 2.3
2. ELECTRODE/ACCESSORY HANDLING
Type of Electrode
• Concentric Needle Electrode: NM-121T/131T/151T, NM-320T/330T
Needle Length: 20 mm (NM-121T/320T)
30 mm (NM-131T/330T)
50 mm (NM-151T)
Needle Diameter: 0.45 mm dia (NM-121T/131T/151T)
0.32 mm dia (NM-320T/330T)
Extension Cable: BM-121S (1 m)
BM-115S (1.8 m)
3. Connect the EMG needle electrode end to the extension cable, aligning the
dot mark at the end of the electrode with the red dot mark at the cable side.
5. Hold the needle electrode vertically to the skin and insert the needle as
quickly as possible into the muscle to be examined.
WARNING
To prevent infection, autoclave the needle electrodes before and
after use.
Before use, autoclave the needle electrode with the tip of the needle covered with
the protective tube.
After use, disconnect the needle electrode from the extension cable. Wipe blood
off the electrode with cotton moistened with alcohol. Autoclave the electrodes
with the tip of the needle covered with the protective tube. Clean and gas-
sterilize the extension cable.
Type of Electrode
The disk electrode is mainly used as the recording electrode for the surface EMG
or evoked potential testing.
FP1 FP2
1 2
F7 F8
F3 FZ F4
13 14
3 19 4 pre-auricular
point
T3 C3 CZ C4 T4
15 5 24 6 16
11 12
A1 A2
P3 P4
T5 T6
7 20 8
17 PZ 18
O1 O2
9 10
inion
Electrode Positions
CZ
1. Divide the longitudinal line of the head into halves and divide the transverse
PZ
FZ line of the head into halves. The point where two lines cross is the vertex. 2
Attach an electrode at the vertex (Cz point) (Fig.1).
FPZ OZ 2. Divide the distance between Cz and the nasion in proportions as shown
and attach an electrode at the point Fz. Do the same for Pz. (Do not attach
electrodes at the points Fpz and Oz.) (Fig.1)
Fig.1
CZ 3. Divide the transverse line of the head into proportions of 10%, 20%, 20%,
20%, 20% and 10% as shown (Fig.2) and attach electrodes at T4, T3, C3 and
C4 C3
C4.
T4 T3
Fig.2
T4 4. Divide the peripheral line passing over Fpz, T4, Oz and T3 into proportions as
T6
F8 shown (Fig.3) and attach electrodes at Fp2, F8, T6, O2, Fp1, F7, T5 and O1.
FP2 O2
FPZ OZ
FP1 O1
F7 T5
T3
Fig.3
5. Take the line passing over F7, Fz and F8 and attach an electrode at the mid-
point (F3) between F7 and Fz as shown (Fig.4).
FP1 FP2
1 2
6. Do the same for F4. Take the line passing over T5, Pz and T6 and do the same
F7 F8
F3 FZ F4 for P3 and P4.
13 14
3 19 4
T3 C3 CZ C4 T4
15 5 24 6 16
11 12
A1 A2
P3 P4
T5 T6
7 20 8
17 PZ 18
O1 O2
9 10
Fig.4
2. Rub the site with an optimal amount of Skinpure skin preparation gel. For
impedance reduction, wipe the site with dry gauze to remove the corneum
layer of skin.
3. Apply an optimal amount of Elefix EEG paste to the cleaned site and rub the
paste in a 10 mm diameter circle around the site.
silver plate 4. Apply an optimal amount of paste to the disk (cup) of the electrode, and
lead attach it on the pasted site as shown at left.
paste
5. Fasten the electrode on the site with surgical tape if necessary.
skin 6. Connect the active and reference electrode leads to the corresponding
electrode jacks.
NOTE
To short-circuit a channel electrode jack to another channel electrode
jack, use the jumper lead (standard accessory) shown below.
CAUTION
Do not wipe an electrode lead or cord with alcohol. If the electrode is
wiped with alcohol, the lead/cord vinyl cover will harden.
1. Remove the disk electrode from the site holding the electrode part. To
prevent the internal wire from breaking, do not pull or bend the lead/cable.
2. Soak the electrodes in running water for a while, and clean them gently with
the running water.
NOTE
If any paste remains on the electrode, it will harden and be difficult to
remove.
3. Wipe them with a wet cloth which has been cleaned with invert soap and
water.
2
4. Wipe the electrode lead or cord with water or lukewarm water.
NOTE
If there is a possibility of discontinuity in an electrode lead, check the
lead continuity with a multimeter. Replace the electrode with a new one if
discontinuity is found.
CAUTION
The sterilizer temperature must not exceed 70°C (158°F). The
electrode(s) may deform or melt above this temperature.
Use ethylene oxide gas according to the sterilization procedure if sterilization (or
disinfection) is necessary.
Type of Electrode
WARNING
When electrical stimulation is continuously performed for a long
period, the felt pads and the EEG paste get dry and the electrical
stimulation may cause skin burn because the dried felt pad or paste
increases the skin-electrode contact impedance. Therefore, always
check that the felt pads are sufficiently moistened with water and that
the paste is moist.
The following three types are provided to electrically stimulate the skin surface
over an examined pathway.
NM-430S
Variable distance type, lead length 2 m
4. Push the two soaked felt pads into each pad holder on the electrode.
Black mark 5. Attach the electrode to the skin surface over the examined nerve.
Felt pad
The black marked side is the negative pole.
CAUTION
The sterilizer temperature must not exceed 70°C (158°F). The
electrode(s) may deform or melt above this temperature.
Use ethylene oxide gas according to the sterilization procedure if sterilization (or
disinfection) is necessary.
Finger Electrode
Electrode Attachment and Connection Procedure
1. Apply an optimal amount of Elefix EEG paste to the ring part of each
electrode.
2. Expand the two ring parts and attach them to a digit innervated by the
examined nerve.
NOTE
Wipe the part between the rings with a dry cloth or paper. If too much
paste is applied to the first mounted ring, it may cause a short circuit so
that the digit cannot be stimulated.
NOTE
When using the NM-450S finger electrode as a recording electrode,
connect the finger electrode to the electrode lead jack on the electrode
junction box.
CAUTION
The sterilizer temperature must not exceed 70°C (158°F). The
electrode(s) may deform or melt above this temperature.
Use ethylene oxide gas according to the sterilization procedure if sterilization (or
disinfection) is necessary.
2. Connect the somato control box cable to the SOMATO connector on the
main unit.
3. Touch the electrode probes to the skin surface over the examined nerve, and
apply stimulation.
SOMATO connector
CAUTION
When using the stimulation probe, start stimulation at 0 mA intensity
and increase gradually. Starting at a higher intensity may deliver a
100 mA current to the patient.
The somato control box RY-441B controls the stimulation on/off and intensity.
• Single Stimulation
To apply single electric stimulation, press the stimulation ON/OFF button on
either side.
• Continuous Stimulation
To apply continuous electric stimulation, press the stimulation ON/OFF
buttons on both side.
The stimulation intensity dial on the operation panel can also control the
stimulation intensity.
CAUTION
Do not use water to clean the somato control box because it
contains electrical parts.
CAUTION
Do not connect/disconnect the somato control box cord when the
power of the instrument is on.
CAUTION
If the somato control box gets wet, stop use and contact your Nihon
Kohden representative.
1. Wipe the paste off the probes with a dry cloth or paper.
Grounding Electrode
Type of Electrode
Strap type ground electrodes 1, 2 and 3 consist of an electrode lead and a strap
with Velcro tape to wind it around a limb.
2. Squeeze the strap softly, and wind it around the limb between the stimulating
and recording electrodes.
NOTE
Keep the strap electrode wet enough and wind it around the limb firmly.
Otherwise, artifact will appear.
3. Clip the electrode lead clip onto the post on the strap, and connect the plug
at the other end of the lead to the patient ground 2-pin jack on the electrode
junction box.
4. Connect the electrode lead to the patient ground 1-pin jack on the electrode
junction box.
CAUTION
Do not leave the strap type ground electrode in water or
physiological salt solution. It will rust and correct measurement
cannot be performed.
CAUTION
The sterilizer temperature must not exceed 70°C (158°F). The
electrode(s) may deform or melt above this temperature.
Use ethylene oxide gas according to the sterilization procedure if sterilization (or
disinfection) is necessary.
Collodion Electrode
WARNING
• If redness, swelling, itch, or rash appears on the patient’s skin,
immediately remove the collodion from the skin and perform
medical treatment.
• Never allow collodion/celloidin or acetone to get in the patient’s
eyes. If collodion/celloidin or acetone accidentally gets in the
eyes, immediately and thoroughly wash eyes with clean water and
perform medical treatment immediately.
• If collodion is swallowed, have the person drink water and vomit
collodion. Perform medical treatment immediately.
• Collodion is a volatile solvent. Both patients and medical staff
must take extreme care not to inhale collodion fumes. When
using collodion, make sure there is adequate ventilation. If excess
collodion fumes are inhaled, have the person lie quietly and keep
warm in fresh air. Perform medical treatment immediately.
WARNING
Keep collodion/celloidin and acetone away from fire because they
are flammable.
NOTE
Refer to the collodion and acetone operator’s manual when using them.
Type of Electrode
The collodion electrode (disk/cup electrode with brim and one or two holes)
records a potential on the scalp for a long time using a collodion or celloidin.
Necessary Accessories
The following locally available accessories are needed in addition to the optional
collodion electrode set.
2
Electrode Placement
An electrode on the patient’s head should be placed on the position according
to the international 10-20 system. Refer to “Disk Type Recording Electrode
– Electrode Placement on Head” in this section.
2. Push aside the hair covering the scalp to expose the scalp, and continue
holding the collodion electrode on the scalp until step 4.
3. Thinly spread collodion on the electrode brim, scalp and hair roots with a
cotton applicator containing the collodion as shown at left.
4. Dry the collodion with cool air from a drier while holding the electrode. The
collodion dries in about 2 minutes.
NOTE
Do not dry the collodion with warm air from the drier because warm air
causes sweat on the scalp.
5. Repeat steps 3 and 4 a few times to fasten the electrode more securely.
6. Put CardioCream ECG paste in a syringe, and inject the paste into the
fastened electrode through the hole with the syringe as shown at the left until
the paste oozes out.
NOTE
To prevent the injected paste from drying out, spread the collodion over
the hole(s) of the electrode.
When using an electrode with two holes, artifact such as sweat caused
during measurement can be reduced by injecting new paste and forcing
out the old paste and sweat.
8. Connect the grounding electrode to the E jack on the electrode junction box.
Removing Collodion
1. Spread acetone on the coagulated collodion around the electrode with a
cotton applicator containing acetone.
2. Wait a while until the electrode is loosened, then remove it from the head.
4. Remove the remaining collodion from the electrode with the acetone and
rinse the paste out of the electrode with running water. Wipe it with a dry
cloth and keep it clean for the next use.
CAUTION
The sterilizer temperature must not exceed 70°C (158°F). The
electrode(s) may deform or melt above this temperature.
Use ethylene oxide gas according to the sterilization procedure if sterilization (or
disinfection) is necessary.
Headphones/Earphones
2
NOTE
Confirm that the L and R receivers are attached to the left and right ears
respectively, before use.
WARNING WARNING
Apply the contact lens electrode to the patient’s Remove the contact lens electrode from the
eye under dark red lighting supervised by a patient’s eye after approximately every 45
physician. When attaching a contact lens minutes and give the patient a 10 minute rest. If
electrode, follow the contact lens electrode the electrode is applied for more than 45 minutes,
operator’s manual. it causes a lack of oxygen to the cornea.
WARNING WARNING
When the monitored waveform is abnormal (for • Never check the skin-electrode contact
example, the waveform is flat or hum is impedance while using a contact lens
overlapping), remove the contact lens electrode electrode. Failure to follow this warning injures
from the patient immediately. the patient’s eyes because impedance check
electrical current will be applied.
• When attaching a contact lens electrode, follow
the contact lens electrode operator’s manual.
Type of Electrode
• ERG Contact Lens Electrode KE-L*
Lens diameter: 12.0 mm
Electrode material: 24 k gold foil
Cord length: 50 cm
WARNING 2
When attaching an ERG contact lens electrode, follow the ERG
contact lens electrode operator’s manual.
1. Put a few drops of a surface anesthetic such as Benoxil in the patient’s eyes.
2. Put a drop of cornea protecting agent such as Scopisol on the inside curved
portion of the contact lens.
3. Insert the lower edge of the contact lens to the patient’s lower eyelid pulling
the lower eyelid down.
4. Insert the upper edge of the contact lens to the upper eyelid pulling the upper
eyelid up.
Disinfection
CAUTION
Do not use alcohol or benzine as the electrode cleaner. These
solutions will deform the electrode.
1. Clean the contact lens electrode with a disinfection solution such as Osvan
liquid, Hibitane liquid or invert soap.
3. Keep the electrode dry and clean for the next use.
Sterilization
Refer to the manual of the purchased electrode for details.
LED Goggles
The following goggles are used as the flash visual stimulating accessory for the
VEP and other testing.
NOTE
Do not bind the goggles strap too tightly to the head. If the strap is too
tight, artifact will appear during measurement.
2. Connect the LS-102J cord to the GOGGLE connector on the right side panel
of the main unit.
GOGGLE connector
NOTE
• A locally available CRT monitor with an analog RGB input jack can be
used as the pattern reversal/flash stimulus CRT display instead of the
Nihon Kohden CRT Monitor.
• Use an isolation transformer with a CRT monitor.
CRT Monitor
Connection
1. Connect the RGB cable provided with the CRT monitor to the D-SUB
connector on the rear of the CRT monitor.
2. Connect the other side of the cable to the PATTERN connector on the right
side panel of the main unit.
D-SUB connector
CRT Monitor
PATTERN connector
Thermistor Probe
The 409J thermistor probe is used for monitoring temperature and automatically
displaying temperature on the screen.
NOTE
Only use the 409J thermistor probe.
Connection
Connect the thermistor probe to the TEMP connector on the right side panel of
the main unit.
TEMP connector
3-1
Section 3-1 Measurement – Basic.................................................................................................................... 3.1.1
Section 3-2 Measurement – SEP/SSEP/ECSP................................................................................................ 3.2.1 3-2
Section 3-3 Measurement – ECG SSEP.......................................................................................................... 3.3.1
Section 3-4 Calculation..................................................................................................................................... 3.4.1 3-3
3-4
3-1
Operation Flow.................................................................................................................................................. 3.1.2
Entering the Patient Information........................................................................................................................ 3.1.4
Selecting the Examination Item......................................................................................................................... 3.1.5
Entering the Examination Information............................................................................................................... 3.1.6
Checking and Changing Measurement Settings............................................................................................... 3.1.7
Measurement Window Description.................................................................................................................. 3.1.11
Monitoring the Raw Waveforms (Monitor Window)................................................................................ 3.1.11
Starting Stimulation and Acquisition (Sweep Window).......................................................................... 3.1.12
Starting Averaging (Average Window)................................................................................................... 3.1.13
Stopping Monitoring, Acquisition or Averaging to Display, Measure or Save a Waveform
(Stop Window)....................................................................................................................................... 3.1.14
Operation Flow
This is a typical example. You may not need to do all these. Operating
procedures differ according to the examination program.
3
Stop averaging waveforms.
3-1
When the system power is turned on, enter the patient information at first.
1. Turn the system power on. The Main Menu window automatically opens.
If the Main Menu window is minimized:
• Press the MENU key on the operation panel.
• Or, click the Main Menu on the Task bar.
2. Click the New Patient button on the Main Menu window. The Patient
Information dialog box opens.
3. Enter the patient information such as, ID No. Name, etc. For details, refer to
the MEB-9400 operator’s manual.
On the Main Menu window, select the examination program. Each program has 3
its own measurement and stimulation settings.
3-1
To open an examination program,
Or,
Or,
When the currently displayed waveform is not saved, a dialog box opens to
prompt you to save the waveform.
You can enter comments in the Main Menu window and edit categories and
programs. Refer to the MEB-9400 operator’s manual for details.
2. Enter the examination information with the keyboard. You can enter the
preset examination information with the list box.
• To clear the entered examination information, click the Clear All button.
• To cancel entering, click the Cancel button.
Before starting measurement, you need to check the measurement settings and 3
change them if necessary on the Set Conditions window.
3-1
1. Click the button on the Tool bar to open the Set Conditions window.
2. Select the condition tab that you want to check. The Condition tabs are at the
top of the window.
Condition tab
For the procedure to change the settings, refer to the MEB-9400 operator’s
manual.
Amp. Page
Item Description
Input Select Selects the input source from either JB (Electrode junction box) or OFF (No
input). You can change the input source to JB or OFF by pressing the Ctrl key +
F1 to F4 key on the keyboard. The F1 to F4 keys correspond to channels 1 to 4.
Electrode (−) Selects a pair of electrode names for each channel to acquire the waveforms
Electrode (+) from A1, A2, A+, Fz, Pz, C3, C4, Cz, Oz, LO, RO, LT, RT, C5S, EP1, EP2, X1
to X8 and DIN.
Sens. (/div) Selects the amplifier sensitivity. When “additional amplifier” is set to “Velocity”,
this selects the amplitude of the velocity waveform.
Hi-cut Selects the high-cut filter frequency of the amplifier.
Lo-cut Selects the low-cut filter frequency of the amplifier.
Acquisition Page
Item Description
Analysis (/div) Selects the analysis time for horizontal length (scale/division) for trigger sweep
and average.
Preset Count Selects the number of waveforms to acquire for average.
Reject Level Selects the amplitude threshold (div) to select the acquired waveforms for
averaging. When the amplitude of the waveform exceeds the upper or lower
threshold, it is not included in averaging. Select “OFF” to not use this function.
Common Reject When a large noise interfered with a Sweep window waveform during averaging,
the affected waveform can be removed from averaging. Common Reject selects
the method to remove the waveform. The limit is set at Reject Level.
ON: When noise interferes with a sweep waveform acquired from one channel
and the waveform amplitude exceeds the Reject Level, the waveform and
all other sweep waveforms which were simultaneously acquired from other
channels are also removed.
OFF: When noise interferes with a sweep waveform acquired from one channel
and the waveform amplitude exceeds the Reject Level, only the sweep
waveform is removed from averaging.
Alternate Average Check this box to alternately acquire the waveforms in two stages (first stage
(Odd/Even) then second stage).
Electric Page
Item Description
Intensity Selects the intensity of the electrical stimulation signal. You can type in
the intensity with the keyboard. You can also change the intensity with the
stimulation intensity dial on the operation panel.
Duration Selects the duration of the electrical stimulation signal. You can type in the width
with the keyboard. You can also change the duration with the STIM DURATION
key.
Trigger Page
Item Description
Stim Rate Selects the stimulation frequency. You can type in the frequency with the
keyboard.
Labeling Page
Item Description
Label Selects the label for marking the waveform.
Measurement Table Selects the Measurement Table window Type I or Type II.
3
Type
3-1
Misc. Page
Item Description
Show Intensity Window Check this box to display the window which displays the stimulation intensity.
Show Temperature Check this box to display the skin temperature window.
Window*
Show Measurement Check this box to display the Measurement window for each examination
Window program when the examination program opens.
Show Nerve/Muscle Check this box to display the table to enter nerve name or muscle name when the
Table examination program opens.
Foot Switches Left, Selects the function for the left and right foot switches.
Right
Auto Store Check this box to automatically save the acquired waveform in temporary
memory when the STOP key on the operation panel is pressed.
Item Description
Trigger Mode This item selects the trigger source from the following modes to display a
time-locked waveform.
Recurrent Mode: Generates a trigger at the Stim Rate which is set in the
Trigger page of the Set Conditions window.
Random Mode: Generates a trigger at random rate within ± 50% of the Stim
Rate.
Signal Mode: Generates a trigger each time the ECG waveform exceeds the
preset threshold.
Signal Trigger Channel This item selects the channel of the waveform to be used as a trigger source.
When the Signal is selected at Trigger Mode, this setting is needed.
Trigger Slope This item selects which part of the trigger source waveform (ECG waveform)
to trigger. When the Signal is selected at Trigger Mode, this setting is needed.
Rising: Upward slope of trigger source waveform
Falling: Downward slope of trigger source waveform
LEVEL 1 (1)
LEVEL 2 (2)
ECG
Four operation modes are used for measurement in this instrument. Select a 3
mode with the MONITOR, STIM/SWEEP, ANALYSIS or STOP key on the
operation panel. 3-1
Amplifier sensitivity
Monitoring time
To display, measure or save a waveform in the hard disk, press the STOP key.
Stimulation and acquisition stops and the Sweep window changes to the Stop
window.
Sensitivity
Number of rejected
Preset number of waveforms to average waveforms
Sensitivity
General.............................................................................................................................................................. 3.2.2
Examination Programs............................................................................................................................ 3.2.2 3-2
Features.................................................................................................................................................. 3.2.2
Preparation........................................................................................................................................................ 3.2.3
Electrode Attachment and Connection.................................................................................................... 3.2.3
Checking Skin-Electrode Contact Impedance................................................................................................... 3.2.7
Measurement..................................................................................................................................................... 3.2.8
Selecting the Program and Changing Measurement Settings................................................................ 3.2.8
Measurement Procedure......................................................................................................................... 3.2.8
Deleting Waveforms................................................................................................................................ 3.2.9
Saving Data on Disk................................................................................................................................ 3.2.9
Printing Data........................................................................................................................................... 3.2.9
General
Examination Programs
• SEP: Somatosensory evoked potential
Middle or long latency (100 ms or more)
Factory default setting is analysis time at 10 ms/div, 1 channel
measurement.
• SSEP: Short-latency somatosensory evoked potential
Short latency (30 ms or less)
Factory default setting is analysis time at 3 ms/div, 4 channel
measurement.
• ESCP: Evoked spinal cord potential
Factory default setting is analysis time at 3 ms/div, 1 channel
measurement.
Features
A 2 channel or 4 channel electrode junction box can be used. With the 4 channel
electrode junction box, up to 104 response waveforms can be displayed on the
window. (4 channels × 26 stages)
SEP Stop window
Preparation
3-2
2. Connect the grounding electrode lead to the E jack.
4. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where electrodes will be attached. Rub the skin with an optimal amount of
Skinpure skin preparation gel to reduce artifact. Wipe the skin thoroughly
with dry gauze to remove any moisture and gel from the skin.
5. Attach the recording electrodes with paste to the optimal sites as shown in
the next page.
6. Attach the grounding electrode between the active electrode and the
stimulating electrode.
NOTE
The skin-electrode contact impedance should be 5 kΩ or less.
8. Soak the felt pads of the surface stimulation electrode with water.
NOTE
• Do not short-circuit between the felt pads with water.
• Do not touch the stimulating electrode with the grounding electrode.
This causes stimulus artifact.
• To prevent noise and artifact, place the patient and the electrode
junction box away from the system and bind the recording electrode
leads together with tape.
• Have the patient relax on a reclining chair or a bed with eyes closed.
• To prevent EMG of the neck, ask the patient to not move his/her head.
• To prevent EMG of the chin, ask the patient to open his/her mouth
slightly.
<SEP at the contralateral Shagass point on the brain evoked by the stimulating median nerve at the right
wrist>
Recording electrode placement Stimulation/Ground electrode placement
Grounding electrode
7cm
Stimulation electrode
CZ
2cm
A1 C3'
(A Derivation in which Fz is
reference electrode is also used.)
A1
C3' Grounding electrode
Grounding electrode
3
Stimulation electrode
3-2
Grounding electrode
(−) (+)
jumper lead C5S EPi
CH 1 CPc Fz
CH 2 CPc EPc
CH 3 C5S Fz
CH 4 EPi EPc
Fz EPc
Grounding electrode
Fpz
C5S
Medial malleolus
CPz Posterior tibial nerve
Stimulation electrode
T12S
CPi
ICc
(−) (+)
CH 1 CPi FPz
CH 2 CPz FPz
CH 3 FPz C5S
CH 4 T12S ICc
jumper lead
jumper lead
Fpz
3
WARNING
• Never check the skin-electrode contact impedance while using a
contact lens electrode. Failure to follow this warning injures the
patient’s eyes because impedance check electrical current will be 3-2
applied.
• When attaching a contact lens electrode, follow the contact lens
electrode operator’s manual.
CAUTION
Never check the skin-electrode contact impedance while using a
needle electrode or while any electrodes are attached at the
operation site on the patient’s body. Failure to follow this caution
causes electrical burn where the electrodes are inserted or attached.
3. Click the Close button on the menu bar to complete the impedance check.
Measurement
1. Select the SOMATO tab in the category tab on the Main Menu window. The
examination program buttons on the Main Menu window change.
2. Select the SEP, SSEP or ESCP examination program on the Main Menu
window. The selected examination program window opens.
3. Click the button on the Tool bar to display the Set Conditions window
to check or change the measurement settings. Refer to “Checking and
Changing Measurement Settings” in this section.
Measurement Procedure
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Check that there is no artifact.
2. Check that the stimulus intensity is set to 0 mA. If not, press the RESET key
on the operation panel.
3. Press the STIM/SWEEP key on the operation panel to display the waveforms
and stimulation output.
6. Press the ANALYSIS key on the operation panel to start averaging. The
artifact is removed from averaging.
When the averaging reaches the preset value (set at Preset Count on
the Acquisition page of the Set Conditions window), the averaging and
stimulation stop automatically.
• To stop during averaging, press the STOP key on the operation panel.
• To reset/restart averaging, press the ERASE key on the operation panel.
9. From the Stop window, you can display the measurement results in the
Measurement Table window when the waveforms are marked.
• To mark the waveforms, refer to “Marking”.
• To display the Measurement Table window, refer to “Measurement Table
Window”.
Deleting Waveforms
You can delete unnecessary waveforms by the following procedure.
1. Select the waveforms to delete on the Stop window with the ALL/INDV
key and SELECT ↓ ↑ key on the operation panel or mouse. The selected
waveforms have a different color. You can also select the waveform(s) by
clicking the All, Channel or Individual button on the Tool bar.
Individual button All button ALL/INDV key: Selects one or more waveforms. The All, Channel
or Individual button on the Tool bar indicates the
selected waveform(s).
The All button is down: All waveforms in one stage or site are
selected.
The Channel button is down: All waveforms in one channel are selected.
Channel button
The Individual button is down: One individual waveform is selected.
2. Press the ERASE key on the operation panel. The message confirming
deletion appears.
3. Check that the waveform selection is correct, and press the ERASE key. The
selected waveforms are deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
Examination Guide MEB-9400 3.2.9
Section 3-3 Measurement – ECG-SSEP
3
General.............................................................................................................................................................. 3.3.2
Features.................................................................................................................................................. 3.3.2
Preparation........................................................................................................................................................ 3.3.3
Electrode Attachment and Connection.................................................................................................... 3.3.3 3-3
Checking Skin-Electrode Contact Impedance................................................................................................... 3.3.5
Measurement..................................................................................................................................................... 3.3.6
Selecting the Program and Changing Measurement Settings................................................................ 3.3.6
Measurement Procedure......................................................................................................................... 3.3.6
Deleting Waveforms................................................................................................................................ 3.3.8
Saving Data on Disk................................................................................................................................ 3.3.8
Printing Data........................................................................................................................................... 3.3.8
General
Features
• A 2 channel or 4 channel electrode junction box can be used. With the
4 channel electrode junction box, up to 104 response waveforms can be
displayed on the window (4 channels × 26 stages). However, one of these
channels is used for ECG input (default setting: channel 1).
• Trigger level of ECG waveform can be set by using the CURSOR dial.
Preparation
3-3
3. Connect the surface stimulation electrode lead to the SOMATO connector
on the right side panel of the main unit with the somato A → B connection
cable. When using the SOMATO control box, connect the SOMATO control
box cord to the SOMATO connector on the right side panel of the main unit.
5. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where electrodes will be attached. Rub the skin with an optimal amount of
Skinpure skin preparation gel to reduce artifact. Wipe the skin thoroughly
with dry gauze to remove any moisture and the gel from the skin.
6. Attach the recording electrodes with paste to the optimal site as shown in the
next page.
7. Attach the grounding electrode between the active electrode and the
stimulating electrode.
NOTE
The skin-electrode contact impedance should be 5 kΩ or less.
9. Soak the felt pads of the surface stimulation electrode with water.
10. Attach the ECG electrodes on the optimal site as shown in the next page.
NOTE
• Do not short-circuit between the felt pads with water.
• Do not touch the stimulating electrode with the grounding electrode.
This causes stimulus artifact.
• To prevent noise and artifact, place the patient and the electrode
junction box away from the system and bind the recording electrode
leads together with tape.
• Have the patient relax on a reclining chair or a bed with eyes closed.
• To prevent EMG of the neck, ask the patient to not move his/her head.
• To prevent EMG of the chin, ask the patient to open his/her mouth
slightly.
Example:
<SSEP at the contralateral Shagass point on the brain evoked by the stimulated median nerve at the left
wrist>
7cm
2cm
C4'
C4'
ECG (
)
Reference electrode (+)
(−) (+)
CH1 ECG ECG
CH2 C4’ Reference
CH3 C5S Reference
CH4 EP1 Reference
jumper lead
jumper lead
Reference electrode
3
WARNING
• Never check the skin-electrode contact impedance while using a
contact lens electrode. Failure to follow this warning injures the
patient’s eyes because impedance check electrical current will be
applied.
• When attaching a contact lens electrode, follow the contact lens 3-3
electrode operator’s manual.
CAUTION
Never check the skin-electrode contact impedance while using a
needle electrode or while any electrodes are attached at the
operation site on the patient’s body. Failure to follow this caution
causes electrical burn where the electrodes are inserted or attached.
3. Click the Close button on the menu bar to complete the impedance check.
Measurement
1. Select the SOMATO tab in the Category tab on the Main Menu window. The
examination program buttons on the Main Menu window change.
2. Select the ECG-SSEP examination program on the Main Menu window. The
ECG-SSEP examination program window opens.
3. Click the button on the Tool bar to display the Set Conditions window
to confirm or change the measurement settings. Refer to “Checking and
Changing Measurement Settings” in this section.
Measurement Procedure
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Check that there is no artifact.
2. Check that the stimulus intensity is set to 0 mA. If not, press the RESET key
on the operation panel.
3. Press the STIM/SWEEP key on the operation panel to display the Signal
Level Trigger Monitor window for ECG trigger control and the response
waveforms evoked by each electric stimulation from the stimulation
electrode.
4. Turn the cursor dial on the operation panel to set the trigger level cursor on
the peak of R wave of the ECG waveform.
7. Press the ANALYSIS key on the operation panel to start averaging. The
artifact is removed from averaging.
When the averaging reaches the preset value (set at Preset Count on
the Acquisition page of the Set Conditions window), the averaging and
stimulation stop automatically. 3
• To stop during averaging, press the STOP key on the operation panel.
• To reset/restart averaging, press the ERASE key on the operation panel.
3-3
8. Save the waveforms temporarily in memory.
• When Auto Store is set to ON (default setting: ON) on the Misc. page of
the Set Condition window, the averaged waveforms are temporarily saved
in memory when averaging finishes.
• When Auto Store is OFF, click the Store button on the Tool bar or press
the STORE key on the operation panel to save the waveform.
[For Reference]
Stimulation output initial setting:
set with “CURSOR”
TRIGGER LEVEL
ECG
MAIN TRIGGER
WAVEMODE = TRAIN STIM INTERVAL = 100 ms
STIMULATION OUTPUT
TRAIN NUMBER = 3
STIMULATION TRIGGER
ANALYSIS TIME = 3 ms/DIV
AVERAGE
10. From the Stop window, the measurement results can be displayed in the
Measurement Table window when the waveforms are marked.
• To mark the waveforms, refer to “Marking”.
• To display the Measurement Table window, refer to “Measurement Table
Window”.
Deleting Waveforms
You can delete unnecessary waveforms by the following procedure.
1. Select the waveforms to delete on the Stop window with the ALL/INDV
key and SELECT ↓ ↑ key on the operation panel or mouse. The selected
waveforms have a different color. You can also select the waveform(s) by
clicking the All, Channel or Individual button on the Tool bar.
All button ALL/INDV key: Selects one or more waveforms. The All, Channel
Individual button or Individual button on the Tool bar indicates the
selected waveform(s).
The All button is down: All waveforms in one stage or site are
selected.
The Channel button is down: All waveforms in one channel are selected.
Channel button
The Individual button is down: One individual waveform is selected.
2. Press the ERASE key on the operation panel. The message confirming
deletion appears.
3. Confirm that the waveform selection is correct, and press the ERASE key.
The selected waveforms are deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
General.............................................................................................................................................................. 3.4.2
Marking.............................................................................................................................................................. 3.4.3
Manual Marking....................................................................................................................................... 3.4.3
Setting a Mark at a Time............................................................................................................... 3.4.3
Moving a Mark......................................................................................................................................... 3.4.5
Deleting Marks........................................................................................................................................ 3.4.5 3-4
Deleting an Individual Mark........................................................................................................... 3.4.5
Deleting All Marks......................................................................................................................... 3.4.5
Measurement Table Window.............................................................................................................................. 3.4.6
To Open the Measurement Table Window............................................................................................... 3.4.7
To Close the Measurement Table Window.............................................................................................. 3.4.7
Measurement Table Window Type I Description................................................................................................ 3.4.8
Evoked Measurement Table - Latency Window....................................................................................... 3.4.8
Entering Normative Data............................................................................................................... 3.4.8
Annotating a Waveform................................................................................................................. 3.4.9
Evoked Measurement Table - Mark Pair Window.................................................................................... 3.4.9
Selecting Measurement Items....................................................................................................... 3.4.9
Setting Pairs................................................................................................................................ 3.4.10
Entering the Waveform Number.................................................................................................. 3.4.10
Evoked Measurement Table Window Type II Description................................................................................ 3.4.11
Evoked Measurement Table Window..................................................................................................... 3.4.11
Annotating a Waveform............................................................................................................... 3.4.12
Selecting Measurement Items..................................................................................................... 3.4.12
Setting Pairs................................................................................................................................ 3.4.12
Entering the Waveform Number.................................................................................................. 3.4.13
SSEP Measurement Table Window....................................................................................................... 3.4.13
Displaying Normative Range....................................................................................................... 3.4.14
Creating the CSV File................................................................................................................. 3.4.14
Saving Data on Disk........................................................................................................................................ 3.4.16
Condition Table Window.................................................................................................................................. 3.4.16
Annotating a Waveform................................................................................................................................... 3.4.17
General
For SEP, SSEP, ESCP and ECG-SSEP, the following calculation operations are
common.
• Marking
• Displaying the Measurement Table window
• Displaying the Condition Table window
• Annotating a waveform
Marking
Manual Marking
Marks can be set manually.
There are two ways to manually set marks on the waveform.
• Setting one mark at a time
• Setting marks according to the settings on the Labeling page on the Set 3-4
Conditions window
The Latency window and Mark Label window appear, and a vertical cursor
appears on the wave area.
Latency window Mark Label window
Vertical cursor
2) On the wave area, click a waveform at the point you want to add the mark.
The selected mark appears on the clicked point. On the Mark Label window,
the lower mark label is highlighted. Up to 12 marks can be set on one
waveform.
2) Turn the CURSOR dial to move the vertical cursor to the position to be
marked.
3) Click the mark label function button you want to set. The mark is set on the
cursor point. The latency (ms) at the point is displayed under the selected
mark label function button.
When the mark label you want to set is not on the function buttons
a) Click the function button to display the next function buttons.
b) Click the desired mark label on the Mark Label window or select the
mark label with the F2 (Select Prev. Mark) or F3 (Select Next Mark)
key on the keyboard. To select marks listed in the upper Mark Label
window, press the F2 key. To select marks listed in the lower Mark
Label window, press the F3 key. The selected mark label is highlighted.
c) Press the F1 (Add Mark) key on the keyboard. The mark is set at the
cursor point.
3. Click the End Marking function button or press the F8 (End Marking) key
on the operation panel.
Moving a Mark
Do one of the following.
• Drag the mark to the desired position. You can move a mark with this
3
procedure on the Stop window without opening the Mark Label window.
• On the Mark Label window, select the mark label you want to move. Then,
do “Setting a Mark at a Time” procedure. The selected mark move to the new
position.
• When the mark label you want to delete is on the function button, you can
delete the mark by clicking the function button twice.
• Or,
1. On the Mark Label window, click the mark label to delete. The cursor moves
to the mark position.
2. Click the Add Mark function button or press the F1 key on the operation
panel.
Click the Delete Mark function button or the F6 key on the keyboard.
All marks on the selected waveform are deleted.
There are two types of Measurement Table windows, Type I and Type II. Select
either Type I or Type II on the Labeling page on the Set Conditions window.
Type I: Displays the Evoked Measurement Table - Latency window and the
Evoked Measurement Table - Mark Pair window.
Type II: For SEP, SSEP or ESCP, the Evoked Measurement Table - Latency
window and the Evoked Measurement Table - Mark Pair window are
combined into one table window.
For SSEP and ECG-SSEP, the SSEP Measurement Table is displayed.
Type I Type II
Marking
Measurement Condition Second Examination Nerve
Input
Patient
Superimpose 3
Table Table Sens. Information Information
F1 F2 F3 F4 F5 F6 F7 F8
When the Type I is selected on the Labeling page on the Set Conditions window,
the following windows open.
1. To display the Normative Data column, click the Show ND check box on the
Evoked Measurement Table - Latency window. A check mark appears in the
box, and the Normative Data column appears.
2. Click the column which you want to enter the normative data in. The cursor
appears in the clicked column.
3. Enter the normative data with the keyboard. The data remains at the next
measurement.
Annotating a Waveform
Whether or not to display the Annotation can be set on the Labeling page of the
Set Conditions window.
To annotate a waveform,
1. Click the No. you want to annotate.
3
Click the button of the Type box. The selectable items are displayed in the
drop down list. Items are Interval, Amplitude, Area, Compare and (none).
Click an item in the list. The clicked item is displayed in the Type box.
Setting Pairs
To set a pair, enter two mark labels in the Mark Pair From box and the Mark Pair
To box. Up to 3 pairs can be set. The “-” is displayed between the boxes which
make a pair.
To enter a mark label in the box, click the button of the box. Twelve mark
labels are displayed in the drop down list. Click the desired mark label. The mark
label is displayed in the box.
Example
Mark Pair From box Mark Pair To box
Type box
N11
A1
Ep
A3
Calculation value
The mark in the Mark Pair From box
Use the keyboard to enter the waveform number. When you set two waveform
numbers (only for Interval), enter waveform numbers in the box in the format
“A1-B1”.
No. box
SEP/ESCP/ECG-SSEP 3
When the Type II is selected on the Labeling page on the Set Condition window,
the Evoked Measurement Table window is displayed. The Evoked Measurement
Table - Latency window and the Evoked Measurement Table - Mark Pair
window are combined into the Evoked Measurement Table window.
SSEP
When the Type II is selected on the Labeling page on the Set Condition window,
3-4
the SSEP Measurement Table window is displayed. The Evoked Measurement
Table- Latency window and the Evoked Measurement Table - Mark Pair window
are combined into the SSEP Measurement Table window.
Preset number of
Stimulation
waveforms to average
rate
Stimulation
intensity
Stimulation
duration
Waveform number
(The waveforms which are
displayed in the wave area
are automatically displayed
in waveform number order.)
Measurement items
No. box
Annotating a Waveform
Whether or not to display the Annotation can be set on the Labeling page of the
Set Conditions window.
To annotate a waveform,
1. Click the No. you want to annotate.
2. Enter an annotation from the keyboard or the annotation list. You can edit the
list in the User Define dialog box.
To open the annotation list, click the down arrow on the right of the
Annotation column.
Click the button of the Type box. The selectable items are displayed in the
drop down list. Items are Interval, Amplitude, Area, Compare and (none).
Click an item in the list. The clicked item is displayed in the Type box.
Setting Pairs
To set a pair, enter two mark labels in the Mark Pair From box and the Mark Pair
To box. Up to 3 pairs can be set. The “-” is displayed between the boxes which
make a pair.
To enter a mark label in the box, click the button of the box. Twelve mark
labels are displayed in the drop down list. Click the desired mark label. The mark
label is displayed in the box.
Example Mark Pair From box The mark in the Mark
Pair From box
Type box Mark Pair To box
Use the keyboard to enter the waveform number. When you set two waveform
numbers (only for Interval), enter waveform numbers in the box in the format
No. box
“A1-B1”.
3
Preset number of
waveforms to average
Calls up the
CSV file
Type box
Selects measurement items
No. box
Waveform number. When a
measurement item other than
“Interval” is selected, the waveform
number is automatically entered.
Mark Pair From box The marks used for CCT calculation.
Automatically set to the 4th and 5th marks in the latency table
Mark Pair To box
CCT is displayed on the bar.
is displayed when the mark is out
of the normative range.
When a CSV file is selected in the Normative data file box, the following data is
displayed in the Normative Range (+/−2SD) row.
To display the calculation value and normative value, set the mouse at the bar,
or .
You can create a normative data file. Using a Microsoft Excel spreadsheet is
recommended. Bars in the NR(+/−2SD) column are displayed based on the
normative data file.
2. Enter text and values. Follow the formats below. Otherwise, the normative
range is not displayed.
- Type the title and comments in the first and second lines.
- Type the normative data in the third or later line.
- The data and sd are 3 digit significant figures, 2 digit decimals. Both the
data and sd must be entered.
Second line:
Comment
Third line:
Data and sd for setting Normative value and
the normative range standard deviation of CCT
Bar length in the Normative
Normative value and Normative value and standard
Range (+/−2SD) column
standard deviation of mark 1 deviation of mark 2 on the
on the Labeling page on the Labeling page on the Set
Set Conditions window Conditions window
- The contents are the same as that of the Excel spreadsheet above.
- Words or data must be separated by commas only. Never put a space or tab
before or after a comma.
3-4
3. Save the file in C:\Program files\Npk
Use any file name.
When creating the file in an Excel spreadsheet, select CSV for the file type.
When creating the file with any other software, save the file with .CSV as
the extension in the file name.
To save the measurement results and the waveforms on the hard disk, refer to the
MEB-9400 operator’s manual.
The Condition Table window displays the list of settings in which waveforms
were acquired.
To open the Condition Table window, click the Condition Table function button
on the Stop window.
F1 F2 F3 F4 F5 F6 F7 F8
Confirm settings for each waveform. Not all settings can be displayed on the
screen at the same time. To display other settings, use the scroll bar or the jump
buttons at the bottom of the screen.
To close the Condition Table window, click the button on the window.
Annotating a Waveform
You can annotate a measured waveform. The annotation display position and 3
annotation display ON/OFF on the Measurement Table window are set on the
Labeling page of the Set Conditions window. Refer to the MEB-9400 operator’s
manual for details.
1. Click the Marking function button or press the F1 key on the keyboard on
the Stop window.
3. Click the Wave Annotation function button or press the F5 key on the
keyboard to open the Wave Annotation window.
4. Click the or button of the Wave No. box on the Wave Annotation
window to display the waveform to be annotated in the box.
6. Click the OK button. The added annotation appears on both the selected
waveform on the Wave area and the Annotation column on the Measurement
Table window.
4-1
4-2
4-3
4-4
4-5
Operation Flow
This is a typical example. You may not need to do all these. Operating
procedures differ according to the examination program.
4-1
When the system power is turned on, enter the patient information at first.
1. Turn the system power on. The Main Menu window automatically opens.
If the Main Menu window is minimized:
• Press the MENU key on the operation panel.
• Or, click the Main Menu on the Task bar.
2. Click the New Patient button on the Main Menu window. The Patient
Information dialog box opens.
3. Enter the patient information such as, ID No. Name, etc. For details, refer to
the MEB-9400 operator’s manual.
On the Main Menu window, select the examination program. Each program has
its own measurement and stimulation settings.
Or,
4
1. Press the Tab key on the keyboard.
4-1
Or,
When the currently displayed waveform is not saved, a dialog box opens to
prompt you to save the waveform.
You can enter comments in the Main Menu window and edit categories and
programs. Refer to the MEB-9400 operator’s manual for details.
2. Enter the examination information with the keyboard. You can enter the
preset examination information with the list box.
• To clear the entered examination information, click the Clear All button.
• To cancel entering, click the Cancel button.
Before starting measurement, you need to check the measurement settings and
change them if necessary on the Set Conditions window.
1. Click the button on the Tool bar to open the Set Conditions window.
2. Select the condition tab that you want to check. The condition tabs are at the
top of the window.
Condition tab
4-1
For the procedure to change the settings, refer to the MEB-9400 operator’s
manual.
Item Description
Input Select Selects the input source from either JB (Electrode junction box) or OFF (No
input). You can change the input source to JB or OFF by pressing the Ctrl key
+ F1 to F4 key on the keyboard. The F1 to F4 keys correspond to channels 1 to
4.
Electrode (−) Selects a pair of electrode names for each channel to acquire the waveforms
Electrode (+) from A1, A2, A+, Fz, Pz, C3, C4, Cz, Oz, LO, RO, LT, RT, C5S, EP1, EP2, X1
to X8 and DIN.
Sens. (/div) Selects the amplifier sensitivity. When “additional amplifier” is set to
“Velocity”, this selects the amplitude of the velocity waveform.
Hi-cut Selects the high-cut filter frequency of the amplifier.
Lo-cut Selects the low-cut filter frequency of the amplifier.
Acquisition Page
Item Description
Analysis (/div) Selects the analysis time for horizontal length (scale/division) for trigger sweep
and average.
Preset Count Selects the number of waveforms to acquire for average.
Reject Level Selects the amplitude threshold (div) to select the acquired waveforms for
averaging. When the amplitude of the waveform exceeds the upper or lower
threshold, it is not included in averaging. Select “OFF” to not use this function.
Common Reject When a large noise interfered with a Sweep window waveform during
averaging, the affected waveform can be removed from averaging. Common
Reject selects the method to remove the waveform. The limit is set at Reject
Level.
ON: When noise interferes with a sweep waveform acquired from one
channel and the waveform amplitude exceeds the Reject Level, the
waveform and all other sweep waveforms which were simultaneously
acquired from other channels are also removed.
OFF: When noise interferes with a sweep waveform acquired from one
channel and the waveform amplitude exceeds the Reject Level, only
the sweep waveform is removed from averaging.
Alternate Average Check this box to alternately acquire the waveforms in two stages (first stage
(Odd/Even) then second stage).
Trigger Page
Item Description
Stim Rate Selects the stimulation frequency. You can type in the frequency with the
keyboard.
Item Description
Waveform Selects the shape of the auditory stimulation signal.
Click: Rectangular shape wave. The duration is set in “Click Duration”.
Burst: Tone burst (modified sine wave). The duration is set in “Plateau
Time”, the rising and falling period is set in “Rise/Fall Time” and the
frequency is set “Tone Freq.”.
Intensity Left Selects the intensity of the left auditory stimulation signal. 4
Intensity Right Selects the intensity of the right auditory stimulation signal.
Stim. Site Left Selects the auditory stimulation site. 4-1
OFF: No stimulation
ON: Left ear
Mask: The white noise for masking is applied to the left ear. The white noise
intensity is the intensity set in “Intensity Right” + the intensity set in
“Mask Level”.
Stim. Site Right Selects the auditory stimulation site.
OFF: No stimulation
ON: Right ear
Mask: The white noise for masking is applied to the right ear. The white
noise intensity is the intensity set in “Intensity Left” + the intensity set
in “Mask Level”.
Mask Level Selects the intensity of the signal output as white noise. The white noise is
applied to the ear which is not examined to mask it from the stimulation.
Labeling Page
Item Description
Label Selects the label for marking the waveform.
Auto Mark Check this box to automatically set the mark on the waveform. The marked
position is set in “Mark Position”. Only for the ABR examination program.
Measurement Table Selects the Measurement Table window Type I or Type II.
Type
Misc. Page
Item Description
Show Intensity Check this box to display the window which displays the stimulation intensity.
Window
Show Measurement Check this box to display the Measurement Table window for each
Window examination program when the examination program opens.
Show Nerve/Muscle Check this box to display the table to enter nerve name or muscle name when
Table the examination program opens.
Foot Switches Left, Selects the function for the left and right foot switches.
Right
Auto Store Check this box to automatically save the acquired waveform in temporary
memory when the STOP key on the operation panel is pressed.
ABR Page
Item Description
Number of Step Number of steps to perform.
For example, when this item is 6, the system automatically performs steps 1 to
6.
Ch. Select the channel to acquire the waveform from.
Stim. Side Select the side to give stimulation to.
Single ch. Ipsi When this item is checked, the ch. is automatically set as following.
Left side stimulation: Acquire waveform from odd channel
Right side stimulation: Acquire waveform from even channel
Measurement Table Selects either “Routine” or “V wave I-L curve” type of the ABR Measurement
Type II Table Type II.
Ch.: Click the button on the box to display selectable channels. The channels
4
which are set to OFF at Input Select on the Amp. page are not displayed.
When ALL is selected, waveforms from all channels are acquired. 4-1
Single ch. Ipsi: When this item is checked, the Ch. boxes become gray.
Individual setting for each step is not available. The channels
which are set to OFF at Input Select on the Amp. page are not
used.
Four operation modes are used for measurement in this instrument. Select a
mode with the MONITOR, STIM/SWEEP, ANALYSIS or STOP key on the
operation panel.
Amplifier
sensitivity
Monitoring time
To display, measure or save a waveform in the hard disk, press the STOP key.
Stimulation and acquisition stop and the Sweep window changes to the Stop
window.
4-1
Sensitivity
To display, measure or save a waveform in the hard disk, press the STOP key.
Averaging stops and the Average window changes to the Stop window.
Number of rejected
waveforms
Preset number of waveforms
to average
Sensitivity
Number of acquired or
averaged waveforms
Preset number of waveforms Number of rejected
to acquire or average waveforms
4
4-1
General.............................................................................................................................................................. 4.2.2
Features.................................................................................................................................................. 4.2.2
Preparation........................................................................................................................................................ 4.2.3
Electrode Attachment and Connection.................................................................................................... 4.2.3
Checking Skin-Electrode Contact Impedance................................................................................................... 4.2.5
Measurement..................................................................................................................................................... 4.2.6
Selecting the Program and Changing Measurement Settings................................................................ 4.2.6
Measurement Procedure......................................................................................................................... 4.2.6 4
Automatic Measurement Procedure........................................................................................................ 4.2.7
Operations on the Stop Window.............................................................................................................. 4.2.8
Deleting Waveforms................................................................................................................................ 4.2.8
4-2
Saving Data on Disk................................................................................................................................ 4.2.8
Printing Data........................................................................................................................................... 4.2.8
General
Features
• The factory default setting is: analysis time at 1 ms/div, 2 channel
measurement
• A 2 channel or 4 channel electrode junction box can be used. With the
4 channel electrode junction box, up to 104 response waveforms can be
displayed on the window (4 channels × 26 stages).
• The response waveform can be marked either automatically or manually.
• Auto marking is available.
• Latency, latency interval, amplitude and acoustic pressure are displayed in the
Measurement Table window.
• The Normative data can be put in the Measurement Table window Type II and
the I-L curve window.
• Auditory threshold can be measured automatically.
• The I-L curve can be displayed in the I-L Curve window.
ABR window
Preparation
3. Connect the headphone cord to the AUDITORY connector on the right side
panel of the main unit.
4. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where the recording electrodes will be attached. Rub the skin with an 4
optimal amount of Skinpure skin preparation gel to reduce artifact. Wipe
the skin thoroughly with dry gauze to remove any moisture and gel from the
skin.
4-2
5. Attach the recording electrodes (active and reference electrodes) and
grounding electrodes to the optimal sites on the patient’s head with paste.
NOTE
The skin-electrode contact impedance should be 5 kΩ or less.
NOTE
• To prevent noise and artifact, place the patient and the electrode
junction box away from the system and bind the recording electrode
leads together with tape.
• Have the patient relax on a reclining chair or a bed with eyes closed.
• To prevent EMG of the neck, ask the patient to not move his/her head.
• To prevent EMG of the chin, ask the patient to open his/her mouth
slightly.
Grounding electrode
Active electrode
Fpz A1
A1 A2 Cz
Fpz
A2
Cz
WARNING
• Never check the skin-electrode contact impedance while using a
contact lens electrode. Failure to follow this warning injures the
patient’s eyes because impedance check electrical current will be
applied.
• When attaching a contact lens electrode, follow the contact lens
electrode operator’s manual.
CAUTION 4
Never check the skin-electrode contact impedance while using a
needle electrode or while any electrodes are attached at the
operation site on the patient’s body. Failure to follow this caution
4-2
causes electrical burn where the electrodes are inserted or attached.
3. Click the Close button on the menu bar to complete the impedance check.
Measurement
2. Select ABR in the Main Menu window. The ABR examination program
window opens.
3. Click the button on the Tool bar to display the Set Conditions window
to check or change the measurement settings. Refer to “Checking and
Changing Measurement Settings” in this section.
Measurement Procedure
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Confirm that there is no artifact.
2. Press the L/R key on the operation panel to select the stimulating output
side. When the L/R key on the operation panel is pressed, the output changes
as follows. → LEFT → RIGHT → LEFT and RIGHT
3. Turn the stimulation intensity dial on the operation panel to increase the
stimulation intensity. The default setting is at 90 dBnHL.
4. Press the STIM/SWEEP key on the operation panel to display the waveforms
and stimulation output.
5. Press the ANALYSIS key on the operation panel to start averaging. The
artifact is removed from averaging.
When the averaging reaches the preset value (set at Preset Count on
the Acquisition page of the Set Conditions window), the averaging and
stimulation stop automatically.
• To stop during averaging, press the STOP key on the operation panel.
• To reset/restart averaging, press the ERASE key on the operation panel.
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Confirm that no artifact is
overlapping.
2. Click the Start Auto Seq. function button or press the F8 (Start Auto Seq.)
key on the keyboard.
The Start Auto Seq. function button changes to the Stop Auto Seq. function
button.
The system starts acquiring waveforms according to the steps set on the ABR
page on the Set Conditions window. Acquired waveforms are automatically
averaged and temporarily saved in memory.
3. When all steps finish, the window changes to the Stop window.
To quit automatic measurement, click the Stop Auto Seq. function button or
press the F8 (Stop Auto Seq.) key on the keyboard. To continue automatic
measurement from the step where measurement was stopped, click the
Start Auto Seq. function button or press the F8 (Start Auto Seq.) key on the
keyboard.
Deleting Waveforms
You can delete unnecessary waveforms by the following procedure.
1. Select the waveforms to delete on the Stop window with the ALL/INDV
key and SELECT ↓ ↑ key on the operation panel or mouse. The selected
waveforms have a different color. You can also select the waveform(s) by
clicking the All, Channel or Individual button on the Tool bar.
Individual button All button ALL/INDV key: Selects one or more waveforms. The All, Channel
or Individual button on the Tool bar indicates the
selected waveform(s).
The All button is down: All waveforms in one stage or site are
selected.
Channel button The Channel button is down: All waveforms in one channel are selected.
The Individual button is down: One individual waveform is selected.
SELECT ↓ ↑ key: Selects the stage if the ALL button is selected.
Selects the channel if the Channel button is
selected.
2. Press the ERASE key on the operation panel. The message confirming
deletion appears.
3. Confirm that the waveform selection is correct, and press the ERASE key.
The selected waveforms are deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
General.............................................................................................................................................................. 4.3.2
Features.................................................................................................................................................. 4.3.2
Preparation........................................................................................................................................................ 4.3.3
Electrode Attachment and Connection.................................................................................................... 4.3.3
Checking Skin-Electrode Contact Impedance................................................................................................... 4.3.5
Measurement..................................................................................................................................................... 4.3.6
Selecting the Program and Changing Measurement Settings................................................................ 4.3.6
Measurement Procedure......................................................................................................................... 4.3.6 4
Operations on the Stop Window.............................................................................................................. 4.3.7
Deleting Waveforms................................................................................................................................ 4.3.7
Saving Data on Disk................................................................................................................................ 4.3.8
Printing Data........................................................................................................................................... 4.3.8
4-3
General
MLR: Middle latency response which has middle latency component (10 to 50
ms) of the auditory evoked potential
The factory default setting is analysis time 5 ms/div, 1 channel
measurement.
SVR: Slow vertex response which has long latency component (50 ms or more)
of the auditory evoked potential
The factory default setting is analysis time 50 ms/div, 1 channel
measurement.
Features
A 2 channel or 4 channel electrode junction box can be used. With the 4 channel
electrode junction box, up to 104 response waveforms can be displayed on the
window (4 channels × 26 stages).
Preparation
3. Connect the headphone cord to the AUDITORY connector on the right side
panel of the main unit.
4. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where the recording electrodes will be attached. Rub the skin with an 4
optimal amount of Skinpure skin preparation gel to reduce artifact. Wipe
the skin thoroughly with dry gauze to remove any moisture and gel from the
skin.
NOTE
The skin-electrode contact impedance should be 5 kΩ or less.
NOTE
• To prevent noise and artifact, place the patient and the electrode
junction box away from the system and bind the recording electrode
leads together with tape.
• Have the patient relax on a reclining chair or a bed with eyes closed.
• To prevent EMG of the neck, ask the patient to not move his/her head.
• To prevent EMG of the chin, ask the patient to open his/her mouth
slightly.
Grounding electrode
Active electrode
Fpz A1
A1 A2 Cz
Fpz
A2
Cz
WARNING
• Never check the skin-electrode contact impedance while using a
contact lens electrode. Failure to follow this warning injures the
patient’s eyes because impedance check electrical current will be
applied.
• When attaching a contact lens electrode, follow the contact lens
electrode operator’s manual.
CAUTION 4
Never check the skin-electrode contact impedance while using a
needle electrode or while any electrodes are attached at the
operation site on the patient’s body. Failure to follow this caution
causes electrical burn where the electrodes are inserted or attached.
4-3
3. Click the Close button on the menu bar to complete the impedance check.
Measurement
1. Click the AUDITORY tab in the Category tab on the Main Menu window.
The examination program buttons on the Main Menu window change.
2. Select the MLR or SVR on the Main Menu window. The selected
examination program window opens.
3. Click the button on the Tool bar to display the Set Conditions window
to confirm or change the measurement settings. Refer to “Checking and
Changing Measurement Settings” in this section.
Measurement Procedure
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Confirm that there is no artifact.
2. Press the L/R key on the operation panel to select the stimulating output
side. When the L/R key on the operation panel is pressed, the output changes
as follows. → LEFT → RIGHT → LEFT and RIGHT
3. Turn the stimulation intensity dial on the operation panel to increase the
stimulation intensity. The default setting is at 90 dBnHL.
4. Press the STIM/SWEEP key on the operation panel to display the waveforms
and stimulation output.
5. Press the ANALYSIS key on the operation panel to start averaging. The
artifact is removed from averaging.
When the averaging reaches the preset value (set at Preset Count on
the Acquisition page of the Set Conditions window), the averaging and
stimulation stop automatically.
• To stop during averaging, press the STOP key on the operation panel.
• To reset/restart averaging, press the ERASE key on the operation panel.
4-3
Deleting Waveforms
You can delete unnecessary waveforms by the following procedure.
1. Select the waveforms to delete on the Stop window with the ALL/INDV
key and SELECT ↓ ↑ key on the operation panel or mouse. The selected
waveforms have a different color. You can also select the waveform(s) by
clicking the All, Channel or Individual button on the Tool bar.
Individual button All button ALL/INDV key: Selects one or more waveforms. The All, Channel
or Individual button on the Tool bar indicates the
selected waveform(s).
The All button is down: All waveforms in one stage or site are
selected.
Channel button The Channel button is down: All waveforms in one channel are selected.
The Individual button is down: One individual waveform is selected.
SELECT ↓ ↑ key: Selects the stage if the ALL button is selected.
Selects the channel if the Channel button is
selected.
2. Press the ERASE key on the operation panel. The message confirming
deletion appears.
3. Confirm that the waveform selection is correct, and press the ERASE key.
The selected waveforms are deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
General.............................................................................................................................................................. 4.4.2
Features.................................................................................................................................................. 4.4.2
Preparation........................................................................................................................................................ 4.4.3
Electrode Attachment and Connection.................................................................................................... 4.4.3
Derivation................................................................................................................................................ 4.4.4
Derivation from External Auditory Meatus.................................................................................... 4.4.4
Derivation from Tympanic Membrane............................................................................................ 4.4.4
Derivation from Tympanic Promontory (Intratympanic Derivation)................................................ 4.4.5 4
Checking Skin-Electrode Contact Impedance................................................................................................... 4.4.6
Measurement ................................................................................................................................................... 4.4.7
Selecting the Program and Changing Measurement Settings................................................................ 4.4.7
Measurement Procedure......................................................................................................................... 4.4.7
Operations on the Stop Window.............................................................................................................. 4.4.8
Deleting Waveforms................................................................................................................................ 4.4.8
Saving Data on Disk................................................................................................................................ 4.4.9 4-4
Printing Data........................................................................................................................................... 4.4.9
AP and CM Classification........................................................................................................................ 4.4.9
General
EcochG measures auditory evoked responses from the cochlear and auditory
nerves by giving sound stimulation to the ear.
Features
• AP (whole nerve Action Potential) and CM (Cochlear Microphonics) can be
automatically classified during averaging (auto processing function).
• The factory default setting is analysis time at 1 ms/div, 2 channel
measurement.
• A 2 channel or 4 channel electrode junction box can be used. With the
4 channel electrode junction box, up to 104 response waveforms can be
displayed on the window (4 channels × 26 stages).
Preparation
3. Connect the headphone cord to the AUDITORY connector on the right side
panel of the main unit. When using the electrocochleogram needle electrode
for the recording electrode, use a speaker box instead of the headphone.
WARNING
Attach the electrodes under the supervision of a physician in charge.
4-4
CAUTION
Never check the skin-electrode contact impedance while using a
needle electrode or while any electrodes are attached at the
operation site on the patient’s body. Failure to follow this caution
causes electrical burn where the electrodes are inserted or attached.
5. Put the headphones on the patient or place a specified speaker box near the
patient’s ear.
NOTE
• To prevent noise and artifact, place the patient and the electrode
junction box away from the system and bind the recording electrode
leads together with tape.
• Have the patient relax on a reclining chair or a bed with eyes closed.
• To prevent EMG of the neck, ask the patient to not move his/her head.
• To prevent EMG of the chin, ask the patient to open his/her mouth
slightly.
• When using the YZ-0073 silver ball electrode, refer to the YZ-0073 silver
ball electrode manual.
Example:
<When using the YZ-0073 silver ball electrode>
CH1 (−)
Shield wire
CH1 (+) Silver ball
electrode
Disk electrode
Grounding electrode
(Fpz or nasion)
Active electrode
(External auditory meatus)
Derivation
Derivation from External Auditory Meatus
3. Put the silver ball electrode (used as active electrode (−) ) with the ECG/
EEG paste on the meatus.
2. Put the silver ball electrode (used as active electrode (−) ) with the ECG/
EEG paste on the tympanic membrane.
CAUTION
Never check the skin-electrode contact impedance while using a
needle electrode or while any electrodes are attached at the
operation site on the patient’s body. Failure to follow this caution
causes electrical burn where the electrodes are inserted or attached.
NOTE
When using the electrocochleogram needle electrode (purchased locally),
use the specified speaker for the auditory stimulation.
1. Disinfect the external auditory meatus and tympanic membrane and apply
local anesthesia. 4
4-4
CAUTION
Never check the skin-electrode contact impedance while using a
needle electrode or while any electrodes are attached at the
operation site on the patient’s body. Failure to follow this caution
causes electrical burn where the electrodes are inserted or attached.
3. Click the Close button on the menu bar to complete the impedance check.
Measurement
1. Click the AUDITORY tab in the Category tab on the Main Menu window.
The examination program buttons on the Main Menu window change.
Measurement Procedure
1. Press the MONITOR key on the operation panel to monitor the raw 4-4
waveform from the recording electrode. Confirm that there is no artifact.
2. Press the L/R key on the operation panel to select the stimulating output
side. When the L/R key on the operation panel is pressed, the output changes
as follows.
→ LEFT → RIGHT → LEFT and RIGHT
3. Turn the stimulation intensity dial on the operation panel to increase the
stimulation intensity. The default setting is at 90 dBnHL.
4. Press the STIM/SWEEP key on the operation panel to display the waveforms
and stimulation output.
5. Press the ANALYSIS key on the operation panel to start averaging. The
artifact is removed from averaging.
When the averaging reaches the preset value (set at Preset Count on
the Acquisition page of the Set Conditions window), the averaging and
stimulation stop automatically.
• To stop during averaging, press the STOP key on the operation panel.
• To reset/restart averaging, press the ERASE key on the operation panel.
Deleting Waveforms
You can delete unnecessary waveforms by the following procedure.
1. Select the waveforms to delete on the Stop window with the ALL/INDV
key and SELECT ↓ ↑ key on the operation panel or mouse. The selected
waveforms have a different color. You can also select the waveform(s) by
clicking the All, Channel or Individual button on the Tool bar.
Individual button All button ALL/INDV key: Selects one or more waveforms. The All,
Channel or Individual button on the Tool
bar indicates the selected waveform(s).
The All button is down: All waveforms in one stage or site are
selected.
The Channel button is down: All waveforms in one channel are selected.
Channel button
The Individual button is down: One individual waveform is selected.
SELECT ↓ ↑ key: Selects the stage if the ALL button is
selected. Selects the channel if the Channel
button is selected.
2. Press the ERASE key on the operation panel. The message confirming
deletion appears.
3. Confirm that the waveform selection is correct, and press the ERASE key.
The selected waveforms are deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
AP and CM Classification 4
stimulus
response
<A> <B>
CM
AP
<A+B> <A−B>
General.............................................................................................................................................................. 4.5.3
Marking.............................................................................................................................................................. 4.5.4
Auto Marking During Acquisition............................................................................................................. 4.5.4
Marking After Acquisition........................................................................................................................ 4.5.4
Setting One Mark at a Time.......................................................................................................... 4.5.4
Setting Multiple Marks Automatically According to the Settings on the Labeling Page................. 4.5.6
Moving a Mark......................................................................................................................................... 4.5.6
Deleting Marks........................................................................................................................................ 4.5.7 4
Deleting an Individual Mark........................................................................................................... 4.5.7
Deleting All Marks......................................................................................................................... 4.5.7
Measurement Table Windows............................................................................................................................ 4.5.8
To Open the Measurement Table Window............................................................................................... 4.5.9
To Close the Measurement Table Window.............................................................................................. 4.5.9
Measurement Table Window Type I Description.............................................................................................. 4.5.10
For the MLR, SVR, EcochG Examination Program............................................................................... 4.5.10
Evoked Measurement Table - Latency Window..................................................................................... 4.5.10
Entering Normative Data............................................................................................................. 4.5.10 4-5
Annotating a Waveform............................................................................................................... 4.5.11
Evoked Measurement Table - Mark Pair Window.................................................................................. 4.5.11
Selecting Measurement Items..................................................................................................... 4.5.11
Setting Pairs................................................................................................................................ 4.5.12
Entering the Waveform Number.................................................................................................. 4.5.12
For the ABR Examination Program....................................................................................................... 4.5.13
ABR Measure Window.......................................................................................................................... 4.5.13
Setting Pairs................................................................................................................................ 4.5.14
Entering Normative Data............................................................................................................. 4.5.14
Annotating a Waveform............................................................................................................... 4.5.14
Measurement Table Window Type II Description............................................................................................. 4.5.15
For the MLR, SVR, EcochG Examination Program............................................................................... 4.5.15
Evoked Measurement Table Window..................................................................................................... 4.5.15
Annotating a Waveform............................................................................................................... 4.5.16
Selecting Measurement Items..................................................................................................... 4.5.16
Setting Pairs................................................................................................................................ 4.5.16
For the ABR Examination Program....................................................................................................... 4.5.17
ABR Measurement Table Window......................................................................................................... 4.5.17
Routine........................................................................................................................................ 4.5.17
Laency-Interval-Amp. Table......................................................................................................... 4.5.17
V wave I-L Curve......................................................................................................................... 4.5.18
Latency-Interval-Amp. Table........................................................................................................ 4.5.18
V wave I-L Curve Graph.............................................................................................................. 4.5.19
Diagnosis.................................................................................................................................... 4.5.19
I-L Curve Window............................................................................................................................................ 4.5.20
Displaying the I-L Curve.............................................................................................................. 4.5.20
General
In the ABR, MLR, SVR and EcochG examination programs, the following
calculation operations are available.
• Marking
• Displaying the Measurement Table window
• Displaying the Condition Table window
• Annotating waveforms
4-5
Marking
When marks are set on the waveform, the latency and amplitude are calculated.
The calculation values are displayed on the Evoked Measurement Table window.
1. Click the Marking function button on the Stop window or press the F1
(Marking) key on the keyboard.
The Latency window and Mark Label window appear, and a vertical cursor
appears on the wave area.
2) On the wave area, click a waveform at the point you want to add the mark.
The selected mark appears on the clicked point. On the Mark Label window,
the mark label one line below is highlighted. Up to 12 marks can be set on
one waveform.
2) Turn the CURSOR dial to move the vertical cursor to the position to be
marked.
3) Click the function button of the mark label you want to set, or press the
function key on the keyboard corresponding to the mark label. The mark is
set on the cursor point. The latency (ms) at the point is displayed under the
selected mark label on the function button.
When the mark label you want to set is not on the function buttons
a) Click the function button or press the F9 key on the keyboard to
display the next function buttons.
b) Click the desired mark label on the Mark Label window or select the
mark label with the F2 (Select Prev. Mark) or F3 (Select Next Mark) key
on the keyboard. To select marks listed in the upper Mark Label window, 4
press the F2 key. To select marks listed in the lower Mark Label window,
press the F3 key. The selected mark label is highlighted.
c) Click the Add Mark function button or press the F1 (Add Mark) key on
the keyboard. The mark is set at the cursor point.
3. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
2. On the Stop window, click the Marking function button, or press the F1 key
on the keyboard.
3. Click the function button on the right of the function buttons or press the
F9 function key on the key board.
4. Click the Auto Marking function button or press the F4 key on the keyboard.
5. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
Moving a Mark
Do one of the following.
• Drag the mark to the desired position. You can move a mark with this
procedure on the Stop window without opening the Mark Label window.
• On the Mark Label window, select the mark label you want to move. Then,
do “Setting a Mark at a Time” procedure. The selected mark move to the new
position.
Deleting Marks
Deleting an Individual Mark
When the mark label you want to delete is on the function button, you can also
delete the mark by clicking the function button twice.
Or,
1. On the Mark Label window, click the mark label to delete. The cursor moves
to the mark position.
4
2. Click the Add Mark function button or press the F1 key on the operation
panel. The mark is deleted.
There are two types of Evoked Measurement Table windows, Type I and Type
II. Select either Type I or Type II on the Labeling page on the Set Conditions
window.
<Type I>
<For the MLR, SVR, EcochG <For the ABR examination program>
examination programs>
<Type II>
4-5
Or click the tool button or the Measurement Table function button again.
2. Click the column in which you want to enter data. The cursor moves to the
clicked column. Enter normative data where the cursor is blinking. The
entered data remains in the next measurement.
Annotating a Waveform
To annotate a waveform,
1. Click the No. you want to annotate.
You can also annotate a waveform on the Wave Annotation window. Refer to the
“Annotating a Waveform” in this section.
Type box
No. box
4-5
Click the arrow button of the Type box. The selectable items are displayed in the
drop down list. Items are Interval, Amplitude, Compare and (none).
Click an item in the list. The clicked item is displayed in the Type box.
Setting Pairs
To set a pair, enter two mark labels in the Mark Pair From box and the Mark Pair
To box. Up to 3 pairs can be set. The “-” is displayed between the boxes which
make a pair.
To enter a mark label in the box, click the button of the box. Twelve mark
labels are displayed in the drop down list. Click the desired mark label. The mark
label is displayed in the box.
N11
A1
Ep
A3
Calculation value
The mark in the Mark Pair From box
To enter the waveform number in the No. box, use the keyboard. When you
set two waveform numbers (only for Interval), enter waveform numbers in the
format “A1-B1”.
4-5
Setting Pairs
To set a pair, enter two mark labels in the Mark Pair From box and the Mark Pair
To box. Up to 3 pairs can be set. The “-” is displayed between the boxes which
make a pair.
To enter a mark label in the box, click the button of the box. Twelve mark
labels are displayed in the drop down list. Click the desired mark label. The mark
label is displayed in the box.
A1
Calculation value
1. To display the Normative Data column, click the Show ND check box on the
Evoked Measurement Table - Latency window. A check mark appears in the
box, and the Normative Data column appears.
2. Click the column which you want to enter the normative data in. The cursor
appears in the clicked column.
3. Enter the normative data with the keyboard. The entered data remains at the
next measurement.
Annotating a Waveform
To annotate a waveform,
1. Click the No. you want to annotate.
You can also annotate a waveform on the Wave Annotation window. Refer to the
“Annotating a Waveform” in this section.
When the waveforms are marked, waveform measurement starts and the results
can be displayed in the Evoked Measurement Table window.
Stimulation rate
Waveform number
The waveforms
Stimulation
which are displayed 4-5
intensity left
in the wave area are
automatically displayed
in wave number order.
Stimulation
intensity right
(M means mask
noise)
Type box:
Selects measurement
items
No. box
Annotating a Waveform
To annotate a waveform,
1. Click the No. you want to annotate.
Click the arrow button of the Type box. The selectable items are displayed in the
drop down list. Items are Interval, Amplitude, Compare and (none).
Click an item in the list. The clicked item is displayed in the Type box.
Setting Pairs
To set a pair, enter two mark labels in the Mark Pair From box and the Mark Pair
To box. Up to 4 pairs can be set. The “-” is displayed between the boxes which
make a pair.
To enter a mark label in the box, click the button of the box. Twelve mark
labels are displayed in the drop down list. Click the desired mark label. The mark
label is displayed in the box.
The mark in the Mark
Pair From box
Stimulation rate 4
Stimulation intensity left
Stimulation intensity right
(M means mask noise)
Lat.-Interval-Amp. Table
Scroll bar
Sorts the waveforms and data by measurement order,
stimulation intensity or stimulation site.
Latency-Interval-Amp. Table
Latency, latency between marks, and amplitude for each waveform for the
stimulation are calculated. Up to 16 lines are displayed.
You can see all twelve marks in the Latency column by scrolling. When two or
more waveforms are acquired under the same stimulation intensity and side, and
the same mark is set on these waveforms, the latency of the waveform with the
earlier stage is displayed.
For the boxes in the Amp column, you can enter two marks to calculate the
amplitude between the marks. For other columns, items are fixed.
Stimulation rate
Stimulation intensity left
Stimulation intensity right
(M means mask noise)
Lat.-Interval-Amp. Table
Diagnosis
Latency-Interval-Amp. Table
Latency, latency between marks, and amplitude for each waveform for the
stimulation are calculated. Up to 16 lines are displayed.
You can see all twelve marks in the Latency column by scrolling. When two or
more waveforms are acquired under the same stimulation intensity and side, and
the same mark is set on these waveforms, the latency of the waveform with the
earlier stage is displayed.
For the boxes in the Amp column, you can enter two marks to calculate the
amplitude between the marks. For other columns, items are fixed.
I-L Curve window ABR Measurement Table window (V wave I-L curve)
Diagnosis 4-5
• Click the Diag List button to open the Diag. List box. Select an item and click
the Copy & Paste button.
The Diag. List can be edited. Refer to the MEB-9400 operator’s manual
section 9 “System Settings”.
The I-L Curve window displays the Intensity-Latency curve (I-L curve) and the
Interpeak Latency curve (IPL curve) for the marked waveforms.
To open the I-L Curve window, click the I-L Curve button on the Evoked
Measurement Table - Latency window or on the ABR Measurement Table
window.
The graphs are automatically displayed. To close the I-L Curve window, click the
Close button on the I-L Curve window.
When a normative data file is selected in the Normative data File box, normative
range is displayed on the graph as .
Latency
I-L curve
Stimulation intensity
IPL curve
You can create a normative data file. The normative data is displayed on the I-L
curve graph and IPL curve graph. You can compare the graphs and normative
data.
You can enter the data as text in a text file or create the file in a Microsoft Excel
work sheet and save it as a text file in CSV format. Using a Microsoft Excel
work sheet is recommended.
2. Enter text and values. Follow the formats below. Otherwise, the normative
range is not displayed.
- Type the title and comments in the first and second lines. Type the 4
normative data and sd in the third and later lines.
- Type the data in the order shown in the example below.
- Intensity: maximum 8 lines, 0 to 90 dB in 1 dB step.
- data, sd: 3 digit significant figure, 2 digit decimals
Both the data and sd need to be entered.
- Save the normative data file in C:\Program Files\Npk.
Stimulation intensity (dB) Normative value and Normative value and standard
standard deviation of mark 1 deviation of mark 2 on the
on the Labeling page on the Labeling page on the Set
Set Conditions window Conditions window
- The contents are the same as that of the Excel spreadsheet above.
- Words or data must be separated by commas only. Never put a space or tab
before or after a comma.
Examination Guide MEB-9400 4.5.21
4. AUDITORY EVOKED POTENTIAL
To save the measurement results and waveforms on the hard disk, refer to the
MEB-9400 operator’s manual.
The Condition Table window displays the list of settings in which waveforms
were acquired. 4
To open the Condition Table window, click the Condition Table function button
on the Stop window or press the F3 key on the keyboard.
Confirm settings for each waveform. Not all settings can be displayed on the
screen at the same time. To display other settings, use the scroll bar or the jump 4-5
buttons at the bottom of the screen.
To close the Condition Table window, click the button on the window.
Annotating a Waveform
You can annotate a measured waveform. The annotation display position and
annotation display ON/OFF on the Measurement Table window are set on the
Labeling page of the Set Conditions window. Refer to the MEB-9400 operator’s
manual for details.
1. Click the Marking function button or press the F1 key on the keyboard on
the Stop window.
F1 F2 F3 F4 F5 F6 F7 F8
4. Click the or button of the Wave No. box on the Wave Annotation
window to display the waveform to be annotated in the box.
6. Click the OK button. The added annotation appears on both the selected
waveform on the Wave area and the Annotation column on the Measurement
Table window.
5-1
5-2
5-3
5-4
5-5
5-1
Operation Flow
This is a typical example. You may not need to do all these. Operating
procedures differ according to the examination program.
5-1
When the system power is turned on, enter the patient information at first.
1. Turn the system power on. The Main Menu window automatically opens.
If the Main Menu window is minimized:
• Press the MENU key on the operation panel.
• Or, click the Main Menu on the Task bar.
2. Click the New Patient button on the Main Menu window. The Patient
Information dialog box opens.
3. Enter the patient information such as, ID No. Name, etc. For details, refer to
the MEB-9400 operator’s manual.
On the Main Menu window, select the examination program. Each program has
its own measurement and stimulation settings.
Or,
Or,
When the currently displayed waveform is not saved, a dialog box opens to
prompt you to save the waveform.
You can enter comments in the Main Menu window and edit categories and
programs. Refer to the MEB-9400 operator’s manual for details.
2. Enter the examination information with the keyboard. You can enter the
preset examination information with the list box.
• To clear the entered examination information, click the Clear All button.
• To cancel entering, click the Cancel button.
Before starting measurement, you need to check the measurement settings and
change them if necessary on the Set Conditions window.
1. Click the button on the Tool bar to open the Set Conditions window.
2. Select the condition tab that you want to check. The Condition tabs are at the
top of the window.
Condition tab
5-1
For the procedure to change the settings, refer to the MEB-9400 operator’s
manual.
Amp. Page
Item Description
Input Select Selects the input source from either JB (Electrode junction box) or OFF (No
input). You can change the input source to JB or OFF by pressing the Ctrl key +
F1 to F4 key on the keyboard. The F1 to F4 keys correspond to channels 1 to 4.
Electrode (−) Selects a pair of electrode names for each channel to acquire the waveforms
Electrode (+) from A1, A2, A+, Fz, Pz, C3, C4, Cz, Oz, LO, RO, LT, RT, C5S, EP1, EP2, X1
to X8 and DIN.
Sens. (/div) Selects the amplifier sensitivity. When “additional amplifier” is set to “Velocity”,
this selects the amplitude of the velocity waveform.
Hi-cut Selects the high-cut filter frequency of the amplifier.
Lo-cut Selects the low-cut filter frequency of the amplifier.
Additional Amplifier Selects the processing when acquiring the waveforms.
OFF: No processing is performed.
Integral: Displays the integral of the acquired waveforms using the reset
threshold which is selected in “Reset”.
Mean: Displays the envelope of the acquired waveforms using the time
constant which is selected in “Time Constant”. When a lower time
constant is selected, a higher resolution waveform envelope appears
(with more slices per time).
Velocity: Displays velocity waveform. When measuring waveform velocity in
the EOG examination program, select Velocity. Sens. shows waveform
velocity.
Acquisition Page
Item Description
Analysis (/div) Selects the analysis time for horizontal length (scale/division) for trigger sweep
and average.
Preset Count Selects the number of waveforms to acquire for average.
Reject Level Selects the amplitude threshold (div) to select the acquired waveforms for
averaging. When the amplitude of the waveform exceeds the upper or lower
threshold, it is not included in averaging. Select “OFF” to not use this function.
Common Reject When a large noise interfered with a Sweep window waveform during averaging,
the affected waveform can be removed from averaging. Common Reject selects
the method to remove the waveform. The limit is set at Reject Level.
ON: When noise interferes with a sweep waveform acquired from one channel
and the waveform amplitude exceeds the Reject Level, the waveform and
all other sweep waveforms which were simultaneously acquired from
other channels are also removed.
OFF: When noise interferes with a sweep waveform acquired from one channel
and the waveform amplitude exceeds the Reject Level, only the sweep
waveform is removed from averaging.
Alternate Average Check this box to alternately acquire the waveforms in two stages (first stage
(Odd/Even) then second stage).
Trigger Page
Item Description
Trigger Mode Selects the trigger mode.
The default setting for the ERG examination program is EXT1.
EXT 1: Generates a trigger signal synchronized with an external trigger signal
which is input to the TRIGGER IN/OUT connector on the rear panel of
the main unit. 5
Stim Rate Selects the stimulation frequency. You can type in the frequency with the
keyboard. 5-1
Stimulator Selects the stimulation source.
Set this item as following:
Pattern-VEP examination program Pattern
Goggle-VEP examination program Goggle
Flash-VEP, ERG examination program EXT
Visual Page
Item Description
Pattern Location Selects the location where the pattern is displayed on the CRT monitor for
pattern reversal stimulation.
Pattern Size Selects the size of the pattern displayed on the CRT monitor for pattern reversal
stimulation.
Display Pattern Selects the pattern for pattern reversal stimulation.
Check: Displays the checked pattern on the CRT monitor.
H-Bar: Displays the horizontal bars (stripes) on the CRT monitor.
V-Bar: Displays the vertical bars (stripes) on the CRT monitor.
Display Mode Selects the pattern for the visual stimulation on the CRT monitor.
Reverse: Displays the reversing pattern on the CRT monitor.
(Pattern reversal visual stimulation).
Flash: Displays the flashing pattern on the CRT monitor.
(Flash pattern visual stimulation).
The duration of each flashing pattern is set in “Pattern Display Time”.
Goggle Side Left Check the box to select the visual stimulation site from left, right or both. You
/Goggle Side Right can also change the side with the L/R key on the operation panel.
(area)
The stimulation differs according to the program. Other condition settings are common for these three
programs. Following are the stimulation conditions for each program and the default settings for the items
which need to be set. These settings can be changed on the Visual page of the Set Conditions window. Refer
to the MEB-9400 operator’s manual for details.
Pattern-VEP
Stimulation: Pattern Reversal Stimulation
Stimulation Output: FULL
Pattern Size: 16
The pattern (checker or stripe) reversal or flash stimulation is generated with the connected CRT
display. These items can be set on the Visual page of the Set Conditions window.
• Pattern Location
• Pattern Size
• Display Pattern
• Display Mode
Goggle-VEP
Stimulation: LED Goggle Stimulation
Stimulation Output: BOTH
Flash-VEP
Stimulation: Use external stimulator. Stimulation output synchronizes with the trigger of the
EMG/EP measuring system.
Labeling Page
Item Description
Label Selects the label for marking the waveform.
Measurement Table Selects the Measurement Table window Type I or Type II.
Type
Misc. Page
Item Description
Show Intensity Window Check this box to display the window which displays the stimulation intensity.
Show Measurement Check this box to display the Measurement window for each examination
Window program when the examination program opens.
Foot Switches Left, Selects the function for the left and right foot switches.
Right
Auto Store Check this box to automatically save the acquired waveform in temporary
memory when the STOP key on the operation panel is pressed.
5-1
Four operation modes are used for measurement in this instrument. Select a
mode with the MONITOR, STIM/SWEEP, ANALYSIS or STOP key on the
operation panel.
Amplifier sensitivity
Monitoring time
To display, measure or save a waveform in the hard disk, press the STOP key.
Stimulation and acquisition stops and the Sweep window changes to the Stop
window.
Sensitivity
5-1
To display, measure or save a waveform in the hard disk, press the STOP key.
Averaging stops and the Average window changes to the Stop window.
Number of averaged
waveforms
Number of rejected
waveforms
Preset number of waveforms
to average
Sensitivity
5-1
5-2
General
WARNING
All stimulation (activation) must be applied under the supervision of
the physician in charge. When the patient has seizure or history of
seizures, visual stimulation may induce seizure activity.
WARNING
Do not apply photic stimulation continuously for more than 5 minutes
in any mode. After applying 5 minutes of photic stimulation to the
patient, wait at least 20 minutes before another photic stimulation.
WARNING
Do not touch the flash lamp assembly because it gets very hot and
can burn skin.
Programs
• Pattern-VEP Visual evoked potential using the pattern/stripe reversal
CRT display*
* Locally available CRT monitor with an analog RGB input
jack.
NOTE
Use an isolation transformer with a CRT monitor.
Features
• The factory default setting is analysis time 30 ms/div, 3 channel measurement.
• A 2 channel or 4 channel electrode junction box can be used. With the
4 channel electrode junction box, up to 104 response waveforms can be
displayed on the window (4 channels × 26 stages).
Pattern-VEP window
5-2
Preparation
Stimulator Connection
CAUTION
Turn off the stimulator power before connecting or disconnecting the
stimulator.
1. Connect the RGB cable provided with the CRT monitor to the D-SUB
connector on the rear of the CRT monitor.
2. Connect the other side of the cable to the PATTERN connector on the right
side panel of the main unit.
D-SUB connector
CRT Monitor
PATTERN connector
Connect the LS-102J cord to the GOGGLE connector on the right side panel of
the main unit.
Stimulator SLS-3100
Connect the TRIGGER IN/OUT connector on the rear panel of the main unit to
the OUTPUT connector on the rear panel of the stimulator panel SLS-3100 with
the cable (SLS-3100 standard accessory).
US
F
E
GND
US
F
E
50 60Hz
100V
Visual Angle
When using the CRT display for pattern reversal stimulation, the visual angle is 5-2
calculated as follows:
360 2 d
Accordingly,
360 r r
57.3
2 d d
For example,
When the screen size is 17 inch (275 × 350 mm), the relation between the
distance and visual angles (θ and δ) is shown in the following table.
2. Connect the grounding electrode lead to the E jack as shown on the next
page.
3. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where the recording and grounding electrodes will be attached. Rub the skin
with an optimal amount of Skinpure skin preparation gel to reduce artifact.
Wipe the skin thoroughly with dry gauze to remove any moisture and gel
from the skin.
NOTE
The skin-electrode contact impedance should be 5 kΩ or less.
NOTE
• To prevent noise and artifact, place the patient and the electrode
junction box away from the system and bind the recording electrode
leads. Hook the electrode leads on the cable tie and fasten them with
tape.
• Have the patient relax on a reclining chair.
• To prevent EMG of the neck, fix the patient’s head with a headrest or a
pillow.
• When using the pattern reversal CRT display in the Pattern-VEP
program, instruct the patient to watch the view target at the center of
the screen without blinking.
• Use an eyepatch to cover the other eye when stimulating only one eye.
<Pattern-VEP measurement>
The following derivations are typically used in the Pattern-VEP measurement.
NOTE
When performing the Pattern-VEP measurement, connect the two
electrode leads used for A1 and A2 with the jumper lead because
A1 + A2 is usually used as the ground electrode in the Pattern-VEP
measurement.
MF
CH1 to 3 Reference electrode (+)
Cz
A1 A2
LO MO RO
Grounding electrode
Grounding electrode
(−) (+) A1 A2 Cz
CH 1 LO MF
or
CH 2 MO MF jumper lead
CH 3 RO MF LO MO RO
E Cz or A1+A2
jumper lead
jumper lead
MF
<Flash-VEP/Goggle-VEP measurement>
The following derivations are typically used in the Flash-VEP/Goggle-VEP
measurement.
NOTE
When performing the Flash-VEP/Goggle-VEP measurement, select A+
for the Electrode (+) on the Amp page of the Set Conditions window.
Fz Grounding electrode
Cz
A2
CH 1 to 4 Reference electrode (+)
A1
LO RO
MO
CH 4 Active electrode (–)
CH 3 Active electrode (–)
CH 2 Active electrode (–)
CH 1 Active electrode (–)
CH 1 LO A1+A2
5
CH 2 MO A1+A2
CH 3 RO A1+A2
5-2
CH 4 Cz A1+A2
E Fz
jumper lead jumper lead
jumper lead
A1 A2
WARNING
• Never check the skin-electrode contact impedance while using a
contact lens electrode. Failure to follow this warning injures the
patient’s eyes because impedance check electrical current will be
applied.
• When attaching a contact lens electrode, follow the contact lens
electrode operator’s manual.
CAUTION
Never check the skin-electrode contact impedance while using a
needle electrode or while any electrodes are attached at the
operation site on the patient’s body. Failure to follow this caution
causes electrical burn where the electrodes are inserted or attached.
3. Click the Close button on the menu bar to complete the impedance check.
Measurement
3. Click the button on the Tool bar to display the Set Conditions window
to check or change the measurement settings. Refer to “Checking and
Changing Measurement Settings” in this section.
Measurement Procedure
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Check that there is no artifact.
5
2. When performing the Pattern-VEP or Goggle-VEP test, press the L/R key on
the operation panel to select the stimulating output side. When the L/R key
on the operation panel is pressed, the output side changes as below.
When performing the Flash-VEP test, go to step 4. 5-2
For pattern reversal stimulation, you can change the size of the pattern using
the Intensity dial on the operation panel.
3. Turn the stimulation intensity dial on the operation panel to change the size
of the pattern.
NOTE
There are various patterns for pattern reversal CRT display. Select a
pattern at Pattern Location on the Visual page of the Set Conditions
window. Refer to the MEB-9400 operator’s manual for details.
4. When performing the Goggle-VEP test, attach the LED goggle on the
patient.
When performing the Flash-VEP test, hold the xenon lamp approximately 30
cm from the patient’s eyes.
6. Press the ANALYSIS key on the operation panel to start averaging. Artifact
is removed from averaging.
When the averaging reaches the preset value (at Preset Count on the
Acquisition page of the Set Conditions window), the averaging and
stimulation stop automatically.
• To stop during averaging, press the STOP key on the operation panel.
• To reset/restart averaging, press the ERASE key on the operation panel.
Deleting Waveforms
You can delete unnecessary waveforms by the following procedure.
1. Select the waveforms to delete on the Stop window with the ALL/INDV
key and SELECT ↓ ↑ key on the operation panel or mouse. The selected
waveforms have a different color. You can also select the waveform(s) by
clicking the All, Channel or Individual button on the Tool bar.
2. Press the ERASE key on the operation panel. The message confirming
deletion appears.
3. Check that the waveform selection is correct, and press the ERASE key. The
selected waveforms are deleted.
5
Saving Data on Disk
To save the measurement results and the waveforms on the hard disk, refer to the
MEB-9400 operator’s manual.
5-2
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
General.............................................................................................................................................................. 5.3.2
Features.................................................................................................................................................. 5.3.3
Preparation........................................................................................................................................................ 5.3.4
Stimulator Connection............................................................................................................................. 5.3.4
Stimulator SLS-3100..................................................................................................................... 5.3.4
Necessary Accessories/Unit................................................................................................................... 5.3.4
Electrode Attachment and Connection.................................................................................................... 5.3.5
Measurement..................................................................................................................................................... 5.3.7
Selecting the Program and Changing Measurement Settings................................................................ 5.3.7
Measurement Procedure......................................................................................................................... 5.3.7
Operations on the Stop Window.............................................................................................................. 5.3.8
Deleting Waveforms................................................................................................................................ 5.3.8
Saving Data on Disk................................................................................................................................ 5.3.8
Printing Data........................................................................................................................................... 5.3.8
5-3
General
WARNING
Apply the contact lens electrode to the patient’s eye under dark red
lighting supervised by a physician. When attaching a contact lens
electrode, follow the contact lens electrode operator’s manual.
WARNING
Remove the contact lens electrode from the patient’s eye after
approximately every 45 minutes and give the patient a 10 minute
rest. If the electrode is applied for more than 45 minutes, it causes a
lack of oxygen to the cornea.
WARNING
• Never check the skin-electrode contact impedance while using a
contact lens electrode. Failure to follow this warning injures the
patient’s eyes because impedance check electrical current will be
applied.
• When attaching a contact lens electrode, follow the contact lens
electrode operator’s manual.
Features
A 2 channel or 4 channel electrode junction box can be used. With the 4 channel
electrode junction box, up to 104 response waveforms can be displayed on the
window (4 channels × 26 stages).
ERG window
5-3
Preparation
Stimulator Connection
CAUTION
Turn off the stimulator power before connecting or disconnecting the
stimulator.
Stimulator SLS-3100
Connect the TRIGGER IN/OUT connector on the rear panel of the main unit to
the OUTPUT connector of the rear panel of the SLS-3100 with the cable (SLS-
3100 standard accessory).
OUTPUT connector
OUTPUT
FUSE
1A TIME LAG
US
F
E
GND
US
F
E
50 60Hz
100V
Necessary Accessories/Unit
• Flash stimulator SLS-3100 (option)
• ERG Contact Lens Electrode: KE-L (adult)*
KE-S (child)*
* Not available in the U.S.
WARNING
• Never check the skin-electrode contact impedance while using a
contact lens electrode. Failure to follow this warning injures the
patient’s eyes because impedance check electrical current will be
applied.
• When attaching a contact lens electrode, follow the contact lens
electrode operator’s manual.
1. Connect each contact lens electrode lead with the DIN conversion connector
to the electrode jack as shown on the next page.
2. Connect the reference electrode lead to the electrode jack as shown on the
next page.
3. Connect the grounding electrode lead to the E jack as shown on the next
page.
4. Keep the patient in a dark room and cover the patient’s eyes with an
eyepatch for about 10 minutes for sufficient dark adaptation.
5. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where the reference and grounding electrodes will be attached. Rub the skin
with an optimal amount of Skinpure skin preparation gel to reduce artifact.
5
Wipe the skin thoroughly with dry gauze to remove any moisture and the gel
from the skin.
6. Attach the reference electrode and the grounding electrode to the sites as
shown on the next page. To avoid artifact, cover the reference electrode with
a black tape. 5-3
7. Attach the contact lens electrode to the patient following the procedure
below.
1) Put a few drops of surface anesthetic such as Benoxil in the patient’s
eyes.
2) Put a drop of cornea protecting agent such as Scopisol on the inside
curved portion of the contact lens.
3) Insert the lower edge of the contact lens to the patient’s lower eyelid
pulling the eyelid down.
4) Insert the upper edge of the contact lens to the upper eyelid pulling the
upper eyelid up.
NOTE
• To prevent noise and artifact, place the patient and the electrode
junction box away from the system and bind the recording electrode
leads. Hook the electrode leads on the cable tie and fasten them with
tape.
• Have the patient relax on a reclining chair or bed with his/her eyes
relaxed and not focused.
• Ask the patient to look at the little LED of the flash stimulator without
blinking.
Example:
30 cm
A2
jumper lead
Measurement
1. Select the VISUAL tab in the Category tab on the Main Menu window. The
examination program buttons on the Main Menu window change.
2. Select the ERG examination program in the Main Menu window. The ERG
examination program window opens.
3. Click the button on the Tool bar to display the Set Conditions window
to check or change the measurement settings. Refer to “Checking and
Changing Measurement Settings” in this section.
Measurement Procedure
WARNING
When the monitored waveform is abnormal (for example, the
waveform is flat or hum is overlapping), remove the contact lens
electrode from the patient immediately.
5
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Check that there is no artifact.
2. Press the STIM/SWEEP key on the operation panel to open the Sweep
window. 5-3
3. Press the MONITOR key again. The Check Monitor window appears.
4. After confirming that the waveform on the Check Monitor window is stable,
press the FLASH switch on the SLS-3100 to give a single-flash stimulation.
The waveform synchronized with the flash stimulation appears on the Sweep
window.
5. Press the STOP key on the operation panel. The evoked waveform appears
on the Stop window.
Deleting Waveforms
You can delete unnecessary waveforms by the following procedure.
1. Select the waveforms to delete on the Stop window with the ALL/INDV
key and SELECT ↓ ↑ key on the operation panel or mouse. The selected
waveforms have a different color. You can also select the waveform(s) by
clicking the All, Channel or Individual button on the Tool bar.
Individual button All button ALL/INDV key: Selects one or more waveforms. The All, Channel
or Individual button on the Tool bar indicates the
selected waveform(s).
The All button is down: All waveforms in one stage or site are
selected.
The Channel button is down: All waveforms in one channel are selected.
The Individual button is down: One individual waveform is selected.
Channel button
SELECT ↓ ↑ key: Selects the stage if the ALL button is selected.
Selects the channel if the Channel button is selected.
2. Press the ERASE key on the operation panel. The message confirming
deletion appears.
3. Check that the waveform selection is correct, and press the ERASE key. The
selected waveforms are deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
General.............................................................................................................................................................. 5.4.2
Features.................................................................................................................................................. 5.4.2
Preparation........................................................................................................................................................ 5.4.3
Necessary Unit........................................................................................................................................ 5.4.3
Electrode Attachment and Connection.................................................................................................... 5.4.3
Checking Skin-Electrode Contact Impedance................................................................................................... 5.4.4
Measurement..................................................................................................................................................... 5.4.5
Selecting the Program and Changing Measurement Settings................................................................ 5.4.5
Measurement Procedure......................................................................................................................... 5.4.5
Measuring Velocity Waveforms................................................................................................................ 5.4.6
Operations on the Stop Window.............................................................................................................. 5.4.7
Deleting Sites.......................................................................................................................................... 5.4.7
Saving Data on Disk................................................................................................................................ 5.4.7
Printing Data........................................................................................................................................... 5.4.7
5-4
General
The electrooculogram is the variation of the eyeball static potential derived from
both the lateral and medial canthi by tracing a view target of the view target
reciprocating motion unit such as the SLE-5100 with the head immobilized.
Features
• The velocity waveforms of the nystagmogram on the odd number channels can
be displayed on the even number channels.
• Response waveforms of up to 99 measurements (sites) can be temporarily
saved in memory and reviewed.
EOG window
Preparation
Necessary Unit
Nystagmo stimulator SLE-5100: Used for inducing eyeball movement
3. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where the electrodes will be attached. Rub the skin with an optimal amount
of Skinpure skin preparation gel to reduce artifact. Wipe the skin thoroughly
with dry gauze to remove any moisture and gel from the skin.
4. Attach the recording and grounding electrodes to the optimal sites on the
patient’s head with paste as shown below.
NOTE 5
• The skin-electrode contact impedance should be 5 kΩ or less.
• To prevent noise and artifact, place the patient and the electrode
junction box away from the system and bind the recording electrode
leads. Hook the electrode leads on the cable tie and fasten them with
tape.
• Have the patient relax on a chair and place his/her head on a head
fastening stand.
5-4
• Ask the patient to relax his/her eye muscles.
• Perform the examination in a dark room.
Fpz
Grounding electrode right eye left
left eye left
CH1 Electrode (−)
CH2 Electrode (+)
Fpz
WARNING
• Never check the skin-electrode contact impedance while using a
contact lens electrode. Failure to follow this warning injures the
patient’s eyes because impedance check electrical current will be
applied.
• When attaching a contact lens electrode, follow the contact lens
electrode operator’s manual.
CAUTION
Never check the skin-electrode contact impedance while using a
needle electrode or while any electrodes are attached at the
operation site on the patient’s body. Failure to follow this caution
causes electrical burn where the electrodes are inserted or attached.
3. Click the Close button on the menu bar to complete the impedance check.
Measurement
2. Select the EOG program on the Main Menu window. The EOG examination
program window opens.
3. Click the button on the Tool bar to display the Set Conditions window
to check or change the measurement settings. Refer to the MEB-9400
operator’s manual for details.
Measurement Procedure
1. Press the MONITOR key on the operation panel to monitor the raw 5
waveform from the recording electrode.
2. After checking that there is no artifact, press the STOP key on the operation
panel.
6. If you want to save the waveforms during measurement, click the Store
button on the Tool bar or press the STORE key on the operation panel. The
preset period of the acquired waveforms before clicking the Store button or
pressing the STORE key are temporarily saved in memory. (The period is set
on the Store Time on the Acquisition page of the Set Conditions window.)
After saving, measurement continues.
7. Press the STOP key on the operation panel to stop measurement and save the
waveforms.
• When Auto Store is set to ON (default setting: ON) on the Misc. page of
the Set Conditions window, the preset period of the acquired waveforms
before pressing the STOP key are temporarily saved in memory. (The
period is set on the Store Time on the Acquisition page of the Set
Conditions window.)
• When Auto Store is OFF, click the Store button or press the STORE key to
save the waveform.
Site No. box
The term “site” means the saved measurement result and one saved
measurement result is one site. Up to 99 sites can be temporarily saved in
memory. The Site No. box on the examination window shows the current
site.
2) Enter the site information in the Site: box by the procedure below.
• From the keyboard, type the site information.
• Click the button and select a site information from the drop down
list. You can define the drop down list. Refer to the MEB-9400
operator’s manual section 9 “System Settings”.
Deleting Sites
An unnecessary site can be deleted.
1. On the Stop window, select the site you want to delete by clicking the or
button at the right of the Site No. box. The temporarily saved waveforms
for the selected site are displayed.
2. Press the ERASE key on the operation panel. A message confirming deletion
appears.
3. Check that the site selection is correct, and press the ERASE key. The
selected site is deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
5-4
General.............................................................................................................................................................. 5.5.2
Copying Waveforms to the Saved Waves Window............................................................................................. 5.5.3
Deleting All Waveforms and Measurement Results................................................................................ 5.5.4
Marking.............................................................................................................................................................. 5.5.5
Manual Marking....................................................................................................................................... 5.5.5
Moving a Mark......................................................................................................................................... 5.5.7
Deleting Marks........................................................................................................................................ 5.5.7
Deleting an Individual Mark........................................................................................................... 5.5.7
Deleting All Marks......................................................................................................................... 5.5.7
Measurement Table Window.............................................................................................................................. 5.5.8
To Open the Evoked Measurement Table Window.................................................................................. 5.5.8
To Close the Evoked Measurement Table Window.................................................................................. 5.5.8
Measurement Table Window Type I Description................................................................................................ 5.5.9
Evoked Measurement Table - Latency Window....................................................................................... 5.5.9
Entering Normative Data............................................................................................................... 5.5.9
Annotating a Waveform............................................................................................................... 5.5.10
Evoked Measurement Table-Mark Pair Window.................................................................................... 5.5.10
Selecting Measurement Items..................................................................................................... 5.5.10
Setting Pairs................................................................................................................................ 5.5.11 5
Entering the Waveform Number.................................................................................................. 5.5.11
Evoked Measurement Table Window Type II Description................................................................................ 5.5.12
Evoked Measurement Table Window..................................................................................................... 5.5.12
Annotating a Waveform............................................................................................................... 5.5.13
Selecting Measurement Items..................................................................................................... 5.5.13
Setting Pairs................................................................................................................................ 5.5.13
Entering the Waveform Number.................................................................................................. 5.5.13
Saving Data on Disk........................................................................................................................................ 5.5.14
Condition Table Window.................................................................................................................................. 5.5.14
5-5
Annotating a Waveform................................................................................................................................... 5.5.15
General
Only for EOG examination program, before the above operations, you need to
copy temporarily saved waveform to the Saved Waves window.
1. On the Stop window, select a site by clicking the or button at the right
of the Site No. box. The temporarily saved waveforms for the selected site
are displayed.
2. Select the waveforms on the Stop window with the ALL/INDV key and
SELECT ↓ ↑ key on the operation panel or mouse. The selected waveforms
have a different color. You can also select the waveform(s) by clicking the
All, Channel or Individual button on the Tool bar.
Individual button All button ALL/INDV key: Selects one or more waveforms. The All, Channel
or Individual button on the Tool bar indicates the
selected waveform(s).
The All button is down: All waveforms in one stage or site are
selected.
Channel button The Channel button is down: All waveforms in one channel are selected.
The Individual button is down: One individual waveform is selected.
SELECT ↓ ↑ key: Selects the stage if the ALL button is selected. 5
Selects the channel if the Channel button is
selected.
3. Click the Append to Saved Screen function button or press the F1 key on
the keyboard. The number in Free count at the lower right of the window
decreases.
F1 F2 F3 F4 F5 F6 F7 F8 F9
5-5
4. Click the Show Saved Screen function button or press the F2 key on the
keyboard. The Saved Waves window appears and the waveform which was
saved in step 3 is displayed.
F1 F2 F3 F4 F5 F6 F7 F8 F9
1. Click the function button or press the F9 key on the keyboard on the Stop
window. The next function buttons are displayed.
F1 F2 F3 F4 F5 F6 F7 F8 F9
2. Click the Erase All function button or press the F8 key on the keyboard. The
confirmation message window appears.
Erase
All
F8 F9
3. Click the Yes button. All waveforms and measurement results are deleted.
To cancel deletion, click the Cancel button on the message window. The
message window closes.
Marking
Manual Marking
1. Click the Marking function button on the Stop window.
The Latency window and Mark Label window appear, and a vertical cursor
appears on the wave area.
5-5
2) On the wave area, click a waveform at the point you want to add the mark.
The selected mark appears on the clicked point. On the Mark Label window,
the lower mark label is highlighted. Up to 12 marks can be set on one
waveform.
2) Turn the CURSOR dial to move the vertical cursor to the position to be
marked.
3) Click the mark label function button you want to set. The mark is set on the
cursor point. The latency (ms) at the point is displayed under the selected
mark label function button.
When the mark label you want to set is not on the function buttons
a) Click the function button to display the next function buttons.
b) Click the desired mark label on the Mark Label window or select the
mark label with the F2 (Select Prev. Mark) or F3 (Select Next Mark) key
on the keyboard. To select marks listed in the upper Mark Label window,
press the F2 key. To select marks listed in the lower Mark Label window,
press the F3 key. The selected mark label is highlighted.
c) Press the F1 (Add Mark) key on the keyboard. The mark is set at the
cursor point.
3. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
Moving a Mark
Do one of the following.
• Drag the mark to the desired position. You can move a mark with this
procedure on the Stop window without opening the Mark Label window.
• On the Mark Label window, select the mark label you want to move. Then,
do the “Manual Marking” procedure. The selected mark moves to the new
position.
Deleting Marks
Deleting an Individual Mark
• When the mark label you want to delete is on the function button, you can
delete the mark by clicking the function button twice.
• Or,
1. On the Mark Label window, click the mark label to delete. The cursor moves
to the mark position.
5
2. Click the Add Mark function button or press the F1 key on the keyboard.
The mark is deleted.
There are two types of Measurement Table windows, Type I and Type II. Select
either Type I or Type II on the Labeling page on the Set Conditions window.
Type I: Displays the Evoked Measurement Table - Latency window and the
Evoked Measurement Table - Mark Pair window.
Type II: The Evoked Measurement Table - Latency window and the Evoked
Measurement Table - Mark Pair window are combined into one table
window.
Type I Type II
Or click the tool button or the Measurement Table function button again.
When the Type I is selected on the Labeling page on the Set Conditions window,
the following windows open.
When the waveforms are marked, waveform calculation starts automatically and
the results can be displayed in the Evoked Measurement Table window.
1. To display the Normative Data area, click the Show ND check box on the
Evoked Measurement Table - Latency window. A check mark appears in the
box, and the Normative Data area appears.
2. Click the box to enter the normative data. The cursor appears in the selected
box.
3. Enter the normative data with the keyboard. The data is saved for the next
measurement.
Annotating a Waveform
Whether or not to display the Annotation can be set on the Labeling page of the
Set Conditions window.
To annotate a waveform,
1. Click the No. you want to annotate.
Click the button of the Type box. The selectable items are displayed in the
drop down list. Items are Interval, Amplitude, Area, Compare and (none).
Click an item in the list. The clicked item is displayed in the Type box.
Setting Pairs
To set a pair, enter two mark labels in the Mark Pair From box and the Mark Pair
To box. Up to 3 pairs can be set. The “-” is displayed between the boxes which
make a pair.
To enter a mark label in the box, click the button of the box. Twelve mark
labels are displayed in the drop down list. Click the desired mark label. The mark
label is displayed in the box.
Example
Type box Mark Pair From box Mark Pair To box
P100
5
Calculation value
Use the keyboard to enter the waveform number. When you set two waveform
numbers (only for Interval), enter waveform numbers in the box in the format
“A1-B1”.
5-5
No. box
When the Type II is selected on the Labeling page on the Set Conditions window,
the Evoked Measurement Table window is displayed. The Evoked Measurement
Table - Latency window and the Evoked Measurement Table - Mark Pair
window are combined into the Evoked Measurement Table window.
Waveform number
The waveforms
which are displayed
in the wave area are
displayed in waveform
number order.
Selects
Measurement Items
No. box
Enter the waveform
number
Annotating a Waveform
Whether or not to display the Annotation can be set on the Labeling page of the
Set Conditions window.
To annotate a waveform,
1. Click the No. you want to annotate.
2. Enter an annotation from the keyboard or the annotation list.
To open the annotation list, click the down arrow on the right of the
Annotation column.
You can also annotate a waveform on the Wave Annotation window.
Setting Pairs
To set a pair, enter two mark labels in the Mark Pair From box and the Mark Pair
To box. Up to 4 pairs can be set. The “-” is displayed between the boxes which
make a pair.
To enter a mark label in the box, click the button of the box. Twelve mark
labels are displayed in the drop down list. Click the desired mark label. The mark 5-5
P100
Type box
Calculation value
To save the measurement results and the waveforms on the hard disk, refer to the
MEB-9400 operator’s manual.
The Condition Table window displays the list of settings in which waveforms
were acquired.
To open the Condition Table window, click the Condition Table function button
or press the F3 key on the keyboard on the Stop window.
Check settings for each waveform. Not all settings can be displayed on the
screen at the same time. To display other settings, use the scroll bar or the jump
buttons at the bottom of the screen.
To close the Condition Table window, click the button on the window.
Annotating a Waveform
You can annotate a measured waveform. The annotation display position and
annotation display ON/OFF on the Measurement Table window are set on the
Labeling page of the Set Conditions window. Refer to the MEB-9400 operator’s
manual for details.
1. Click the Marking function button or press the F1 key on the keyboard on
the Stop window.
3. Click the Wave Annotation function button or press the F5 key on the
keyboard to open the Wave Annotation window.
4. Click the or button of the Wave No. box on the Wave Annotation
window to display the waveform to be annotated in the box.
5-5
5. Enter an annotation in the Annotation box, or click the button of the
Annotation: box and select an annotation from the drop down list. You can
customize the drop down list. Refer to the MEB-9400 operator’s manual
section 9 “System Settings”.
6. Click the OK button. The added annotation appears on both the selected
waveform on the Wave area and the Annotation column on the Measurement
Table window.
6-1
6-2
6-3
6-1
Operation Flow
This is a typical example. You may not need to do all these. Operating
procedures differ according to the examination program.
Mark the waveforms in the Stop window. Measure and edit the
acquired waveforms and data. The result is displayed in the
Measurement Table window.
When the system power is turned on, enter the patient information at first.
1. Turn the system power on. The Main Menu window automatically opens.
If the Main Menu window is minimized:
• Press the MENU key on the operation panel.
• Or, click the Main Menu on the Task bar.
2. Click the New Patient button on the Main Menu window. The Patient
Information dialog box opens.
3. Enter the patient information such as ID No. Name, etc. For details, refer to
the MEB-9400 operator’s manual.
6-1
On the Main Menu window, select the examination program. Each program has
its own measurement and stimulation settings.
Or,
Or,
Category tab
When the currently displayed waveform is not saved, a dialog box opens to
prompt you to save the waveform.
You can enter comments in the Main Menu window and edit categories and
programs. Refer to the MEB-9400 operator’s manual for details.
2. Enter the examination information with the keyboard. You can enter the
preset examination information with the list box.
Before starting measurement, you need to check the measurement settings and
change them if necessary on the Set Conditions window.
1. Click the button on the Tool bar to open the Set Conditions window.
2. Select the condition tab that you want to check. The Condition tabs are at the
top of the window.
Condition tab
For the procedure to change the settings, refer to the MEB-9400 operator’s
manual.
EMG Page
Item Description
EMG filter Sets the low and high cut values of the amplifier when the EMG filter function
button is pressed.
Lo-cut: low cut filter
High-cut: high cut filter
EMG Measurement Sets whether time (ms) or ratio (mV/ms) is displayed for “Rise Time” in the
(Rise) MUP Analysis window.
Turn Level Sets the amplitude threshold for counting turns.
Phase Level Sets the + and − threshold from the baseline for counting phases.
Findings Window Selects the types of the EMG Findings window (vertical or horizontal).
Monitor Function Key Different measurement settings can be assigned to the function buttons 1 to 3.
F1 to F3: Type the label according to the measurement setting.
F2, F3 Settings: Selects the sweep time and sensitivity. When the F1 function
button is pressed, the settings are returned to the settings set in
the Set Conditions window - Amp page and Acquisition page.
Auto Store: Check this box to automatically save the displayed waveform.
Muscle List Auto Advance: Automatically selects the next muscle name when the
waveforms are saved (STORE key is pressed).
Edit List: Opens the Muscle List dialog box to edit the muscle names for
each EMG measurement program.
EMG filter
When satellite potential is in the waveform and main MUP
components cannot be recognized clearly, EMG filter
temporarily eliminates satellite potential.
Rise Time
90%
Rise time is the time interval between 10 and 90% (peak to
Process amplitude peak) of the fastest up-going process which is also greater
MUP amplitude than 50% of the MUP amplitude
10%
Rise time
2 Turn Level
4
1 Turns is the number of peaks in which the amplitudes of
the rising and falling components both exceed the turn
amplitude threshold.
6
3
5
6-1
Phase Level
PHASE LEVEL
+ Phases is one plus the total number of MUP components
that exceed both thresholds.
1 2 3
0
3 + 1 = 4 phases
Amp. Page
Item Description
Input Select Selects the input source from either JB (Electrode junction box) or OFF (No
input). You can change the input source to JB or OFF by pressing the Ctrl key +
F1 to F4 key on the keyboard. The F1 to F4 keys correspond to channels 1 to 4.
Sens. (/div) Selects the amplifier sensitivity. When “additional amplifier” is set to “Velocity”,
this selects the amplitude of the velocity waveform.
Hi-cut Selects the high-cut filter frequency of the amplifier.
Lo-cut Selects the low-cut filter frequency of the amplifier.
Additional Amplifier Selects the processing when acquiring the waveforms.
OFF: No processing is performed.
Integral: Displays the integral of the acquired waveforms using the reset
threshold which is selected in “Reset”.
Mean: Displays the envelope of the acquired waveforms using the time
constant which is selected in “Time Constant”. When a lower time
constant is selected, a higher resolution waveform envelope appears
(with more slices per time).
Velocity: Displays velocity waveform.
Rectify Selects the rectification mode. (see the figure below)
OFF: No rectification.
HALF: Half-wave rectification
FULL: Full-wave rectification
HALF
Raw waveform
FULL
Acquisition Page
Item Description
Sampling time Selects the sampling time.
Monitor (/div) Selects the sweep time for the Monitor window for horizontal 1 div
Analysis (/div) Selects the analysis time for horizontal length (scale/division) for trigger sweep
and average.
Trigger Page
Item Description
Measured Signal Selects the trigger level and trigger point.
Trigger OFF: Acquires the preset period of the raw waveform as a sweep wave
when the raw waveform exceeds a trigger level cursor.
Window: Acquires the preset period of the raw waveform when the raw
waveform exceeds the LEVEL 1 cursor but does not exceed the
LEVEL 2 cursor.
Rise Time: Acquires the raw waveform when the raw waveform exceeds a
trigger level and the Rise Time (the time interval between 10 and
90% (peak to peak) of the raw waveform) is less than “Rise Time
Threshold”.
Rise Time Threshold Selects the time interval when “Measured Signal Trigger” is set to “Rise Time”.
OFF
Waveforms 1 and 2 are acquired.
LEVEL
TRIGGER OUTPUTsignal
Window
Only waveform 1 is acquired.
LEVEL 2
LEVEL 1
Rise Time
When “Rise Time 1” < “Rise Time Threshold” < “Rise Time 2”, only waveform
1 is acquired.
LEVEL
Display Page
Item Description
Cascade Display Check this box to display cascaded waveforms on the Monitor, Sweep and Stop
windows.
Traces per Field Selects the number of waveforms displayed on one screen.
Labeling Page
Item Description
Area Calculation Mode Selects the area calculation mode. (see the figure below)
Baseline: Calculate the area bounded by the waveform, the baseline and the
vertical lines from the two marks (L1 and L2) to the baseline.
Mark: Calculates the area bounded by the waveform and the line between the
two marks (L1 and L2).
Measurement Table Selects the Measurement Table window Type I or Type II.
Type
Baseline: Calculates the area bounded by the waveform, the baseline and the
vertical lines from the two marks (L1 and L2) to the baseline.
Mark: Calculates the area bounded by the waveform and the line between
the two marks (L1 and L2).
Misc. Page
Item Description
Show Intensity Window Check this box to display the window which displays the stimulation intensity.
Show Measurement Check this box to display the Measurement window for each examination
Window program when the examination program opens.
Show Nerve/Muscle Check this box to display the table to enter nerve name or muscle name when the
Table examination program opens.
Foot Switches Left, Selects the function for the left and right foot switches.
Right
6-1
Four operation modes are used for measurement in this instrument. Select a
mode with the MONITOR, STIM/SWEEP, ANALYSIS or STOP key on the
operation panel.
Operation status
High-cut/Low-cut
filter settings
Selected side
Monitoring time/
Amplifier sensitivity
Starting Acquisition
Press the STIM/SWEEP key on the operation panel. The Monitor window
changes to the Sweep window, and the Trigger Level Monitoring window opens.
In the Trigger Level Monitoring window, you can set the trigger level to acquire
waveforms. When the raw waveform exceeds the trigger level, a preset period of
the raw waveform before and/or after the trigger point is displayed on the Sweep
window and temporarily saved in memory. You can set the number of waveforms
to acquire.
To display, measure or save a waveform in the hard disk, press the STOP key.
Acquisition stops and the Sweep window changes to the Stop window.
6-1
To display, measure or save a waveform in the hard disk, press the STOP key.
Averaging stops and the Average window changes to the Stop window.
6-1
General.............................................................................................................................................................. 6.2.2
Preparation........................................................................................................................................................ 6.2.3
Electrode Attachment and Connection.................................................................................................... 6.2.3
Measurement..................................................................................................................................................... 6.2.4
Selecting the Program and Changing Measurement Settings................................................................ 6.2.4
Waveform Acquisition for EMG Measurement......................................................................................... 6.2.4
Operations on the Stop Window.................................................................................................... 6.2.5
Waveform Acquisition for MUP Measurement......................................................................................... 6.2.6
Operations on the Stop Window.................................................................................................... 6.2.8
Calling up Temporarily Saved Waveforms............................................................................................... 6.2.9
Selecting the Site No. to Display Waveforms on the Stop Window............................................... 6.2.9
Deleting Sites.......................................................................................................................................... 6.2.9
Saving Data on Disk................................................................................................................................ 6.2.9
Printing Data........................................................................................................................................... 6.2.9
6-2
General
Program
EMG: Measurement of voluntary/spontaneous EMG
Features
• The factory default setting is analysis time 10 ms/div, 1 channel measurement.
• A single MUP can be recorded by triggered sweeping of the waveform.
• Waveforms can be rectified and integrated in real time.
• Up to 99 waveforms of 600 seconds can be temporarily saved in memory.
Preparation
WARNING
To prevent infection, autoclave the needle electrodes before and
after use.
4. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where the grounding electrode will be attached. Rub the skin with an optimal
amount of Skinpure skin preparation gel to reduce artifact. Wipe the skin
thoroughly with dry gauze to remove any moisture and gel from the skin.
5. Attach the grounding electrode to the optimal sites with paste as shown
below.
NOTE
• In the EMG examination program, the skin-electrode contact impedance
check is not available.
• To prevent noise and artifact, place the patient and the electrode
junction box more than 1 m away from the system.
• Have the patient relax and not contract the muscle.
Electrode Connection
6
Example: JB-944BK (4 channel, without montage function)
Grounding electrode
6-2
Needle electrode
Measurement
1. Select the EMG tab in the Category tab on the Main Menu window. The
examination program buttons on the Main Menu window change.
2. Select the EMG examination program on the Main Menu window. The EMG
examination program window opens.
3. Click the button on the Tool bar to display the Set Conditions window
to check or change the measurement settings. Refer to the MEB-9400
operator’s manual for details.
Side button 4. Enter the muscle name and the examination side.
• To enter the muscle name, click the muscle name in the muscle name list,
or press the MUSCLE/TEST ↑ ↓ key on the operation panel or select the
Muscle name list muscle name by clicking the F5 Muscle List DOWN or F6 Muscle List UP
function button. The selected muscle name is surrounded by a rectangle.
You can enter the desired muscle name in the Other Muscle text box.
• To select the examination side, click the L, R or Both side button or click
the F7 Left, Right or Both function button, or press the SIDE/SET key on
the operation panel.
Other Muscle
text box When the waveform acquisition is started, the muscle name and examination
side are displayed on the title bar and lower right corner of the screen.
2. Adjust the volume of the speaker with the Sound volume dial on the
operation panel.
3. Disinfect the patient’s skin at the examined site with a piece of cotton
moistened with alcohol.
5. With the Monitor window and monitor speaker sound, monitor the EMG
of the fully relaxed potential and the voluntary contraction action potential
waveform.
• To adjust the sensitivity to display the waveform of the optimal size, press
the VERTICAL RANGE ↑ ↓ key.
• To adjust the time per division in continuous waveform, press the TIME
SCALE ← → key.
• To change the sweep time and sensitivity combination that is selected on
the Set Conditions window, use the F1 to F3 function keys.
• To display the waveform cascaded, click the Cascade Display function
button.
• To scroll the waveform vertically, turn the WAVE dial on the operation
panel.
• Turn the EMG filter ON or OFF with the EMG Filter function button.
6. Temporarily save the waveform. After you save the waveform, you can
either continue or stop measurement.
2. Adjust the volume of the speaker with the Sound volume dial on the
operation panel.
3. Disinfect the patient’s skin at the examination site with a piece of cotton
moistened with alcohol.
4. Insert the EMG needle electrode into the examined muscle. With the
patient’s cooperation, monitor the EMG of the moderate contraction action
potential waveform.
• To adjust the sensitivity to display the waveform of the optimal size, press
the VERTICAL RANGE ↑ ↓ key on the operation panel.
• To adjust the time per division in continuous waveform, press the TIME
SCALE ← → key.
• To display the waveform cascaded, click the Cascade Display function
button.
• To scroll the waveform vertically, turn the WAVE dial on the operation
panel.
• To select the muscle name, click the muscle name in the muscle list or use
the F5 Muscle List UP or F6 Muscle List DOWN function button.
• To select the examination side, click the L, R or Both side button or use
the F7 Left, Right or Both function button.
5. Press the STIM/SWEEP key on the operation panel to display the Sweep
window and the Trigger Level Monitoring window.
6. Turn the Cursor dial on the operation panel to set the trigger level cursor on
the Trigger Level Monitoring window near the peak of the waveform.
When the raw waveform exceeds the trigger level, a preset period of the raw
waveform before and/or after the trigger point is displayed on the Sweep
window.
• To adjust the sensitivity to display the waveform of the optimal size, press
the VERTICAL RANGE ↑ ↓ key.
• To adjust the time per division in continuous waveform, press the TIME
SCALE ← → key.
• To display the waveform cascaded, click the Cascade Display function
button.
• To scroll the waveform vertically, turn the WAVE dial on the operation
panel.
6
• Turn the EMG filter ON or OFF with the EMG Filter function button.
7. If necessary, press the ANALYSIS key on the operation panel to average the
waveform. When the averaging reaches the preset value (at Preset Count 6-2
on the Acquisition page of the Set Conditions window), the averaging stops
automatically.
8. Temporarily save the waveform. After you save the waveform, you can
either continue or stop measurement.
The number of saved waveforms is displayed in the Site No.: list box on the
upper right of the window.
On the Stop window, click the or button at the right of the [Site No.] to
select and display the desired waveform. To display the next site data, click the
button. To display the previous site data, click the button. The site No. of
the displayed waveform appears on the upper right of the window.
Deleting Sites
An unnecessary site can be deleted.
1. On the Stop window, select the site you want to delete by clicking the or
button at the right of the Site No. box. The temporarily saved waveforms
for the selected site are displayed.
2. Press the ERASE key on the operation panel. The message confirming
deletion appears.
3. Check that the site selection is correct, and press the ERASE key. The
selected site is deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
6-2
General.............................................................................................................................................................. 6.3.2
Copying Waveforms to the Saved Waves Window............................................................................................. 6.3.3
Deleting Waveforms................................................................................................................................ 6.3.4
Marking.............................................................................................................................................................. 6.3.5
Manual Marking....................................................................................................................................... 6.3.5
Moving a Mark......................................................................................................................................... 6.3.7
Deleting Marks........................................................................................................................................ 6.3.7
Deleting an Individual Mark........................................................................................................... 6.3.7
Deleting All Marks......................................................................................................................... 6.3.7
Measurement Table Window.............................................................................................................................. 6.3.8
To Open the EMG Measurement Table Window...................................................................................... 6.3.8
To Close the EMG Measurement Table Window..................................................................................... 6.3.8
Measurement Table Window Type I Description................................................................................................ 6.3.9
Measurement Table Window Type II Description............................................................................................. 6.3.11
Making an EMG Findings Window................................................................................................................... 6.3.13
Making an EMG Findings Summary Table Window............................................................................... 6.3.14
Calculation In the Manual MUP Window......................................................................................................... 6.3.15
Registering the Waveform........................................................................................................... 6.3.15
Deleting Registered Waveforms.................................................................................................. 6.3.15
Saving Data on Disk........................................................................................................................................ 6.3.16
Condition Table Window.................................................................................................................................. 6.3.16
Deleting All Waveforms and Calculation Results............................................................................................. 6.3.17
6-3
General
In the EMG examination program, the following operations are available for
calculation.
• Copying waveforms on the Stop window to the Saved Waves window.
• Marking
• Displaying the Measurement Table window
• Making an EMG Findings window
• Displaying the Condition Table window
1. On the Stop window, select a site by clicking the or button at the right
of the Site No. box. The temporarily saved waveforms for the selected site
are displayed.
2. Select the waveforms on the Stop window with the ALL/INDV key and
SELECT ↑ ↓ key on the operation panel or mouse. The selected waveforms
have a different color. You can also select the waveform(s) by clicking the
All, Channel or Individual button on the Tool bar.
3. Select the function button or press the F9 function key on the keyboard.
4. Click the Append to Stored Wave function button or press the F4 function
key on the keyboard.
5. Click the Show Stored Wave function button or press the F5 key on the
keyboard. The Saved Waves window appears and the waveform which was
selected in step 2 is displayed.
6-3
Append to Show Erase
Superimpose
Stored Wave Stored Wave All
F1 F2 F3 F4 F5 F6 F7 F8 F9
Deleting Waveforms
You can delete unnecessary waveforms by the following procedure.
1. Select the waveforms to delete with the ALL/INDV key and SELECT ↑
↓ key on the operation panel or mouse. The selected waveforms have a
different color. You can also select the waveform(s) by clicking the All,
Channel or Individual button on the Tool bar.
Individual button All button ALL/INDV key: Selects one or more waveforms. The All, Channel
or Individual button on the Tool bar indicates the
selected waveform(s).
The All button is down: All waveforms in one stage or site are
selected.
Channel button The Channel button is down: All waveforms in one channel are selected.
The Individual button is down: One individual waveform is selected.
SELECT ↑ ↓ key: Selects the stage if the ALL button is selected.
Selects the channel if the Channel button is selected.
2. Press the ERASE key on the operation panel. A message confirming deletion
appears.
3. Check that the waveform selection is correct, and press the ERASE key. The
selected waveforms are deleted.
Marking
Manual Marking
1. Click the Marking function button on the Stop window.
The Latency window and Mark Label window appear, and a vertical cursor
appears on the wave area.
Mark Label
window
6-3
2) On the wave area, click a waveform at the point you want to add the mark.
The selected mark appears on the clicked point. On the Mark Label window,
the lower mark label is highlighted. Up to 12 marks can be set on one
waveform.
2) Turn the CURSOR dial to move the vertical cursor to the position to be
marked.
3) Click the mark label function button you want to set. The mark is set on the
cursor point. The latency (ms) at the point is displayed under the selected
mark label function button.
When the mark label you want to set is not on the function buttons
a) Click the function button to display the next function buttons.
b) Click the desired mark label on the Mark Label window or select the
mark label with the F2 (Select Prev. Mark) or F3 (Select Next Mark) key
on the keyboard. To select marks listed in the upper Mark Label window,
press the F2 key. To select marks listed in the lower Mark Label window,
press the F3 key. The selected mark label is highlighted.
c) Press the F1 (Add Mark) key on the keyboard. The mark is set at the
cursor point.
3. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
Moving a Mark
Do one of the following.
• Drag the mark to the desired position. You can move a mark with this
procedure on the Stop window without opening the Mark Label window.
• On the Mark Label window, select the mark label you want to move. Then, do
“Manual Marking” procedure. The selected mark moves to the new position.
Deleting Marks
Deleting an Individual Mark
• When the mark label you want to delete is on the function button, you can
delete the mark by clicking the function button twice.
• Or,
1. On the Mark Label window, click the mark label to delete. The cursor moves
to the mark position.
2. Click the Add Mark function button or press the F1 key on the keyboard.
The mark is deleted.
Click the Delete Mark function button or press the F6 key on the keyboard.
All marks on the selected waveform are deleted.
6-3
There are two types of EMG Measurement Table windows: Type I and Type II.
Select Type I or Type II on the Labeling page on the Set Conditions window.
Or click the tool button or the Measurement Table function button again.
When Type I is selected on the Labeling page on the Set Conditions window, the
following window opens.
Displays the latency, amplitude and area between the marks. The latency,
amplitude and area are automatically calculated and displayed.
b
6-3
a
Area: Area bounded by the waveform and baseline. The negative area is
calculated in positive value.
Rise: This item calculates either time interval or slope according to the setting
at the Rise on the EMG page of the Set Conditions window. When Time
is selected, the time interval between 10 and 90% of the fastest up-going
process which is also greater than 50% of the amplitude is displayed.
When Ratio is selected, the slope between 10 and 90% of the fastest
up-going process which is also greater than 50% of the amplitude is
displayed.
90%
amplitude
10%
rise time
When Type II is selected on the Labeling page on the Set Conditions window, the
following window opens.
V-range: Sensitivity
No.: Waveform number
Muscle: Muscle name
Side: Examination side
Dur.: Time between the marks (duration time of the action potential)
Amp.: Difference between the maximum and minimum potential between 6
the marks
Phases: The number of phases, determined by counting negative and
positive peaks to and from the baseline in which each consecutive
peak exceeds ± 15 µV (This value is set at Phase Level in the EMG
Measurement box on the EMG page of the Set Conditions window),
6-3
equals the number of baseline crossings plus one.
Turns: When two adjacent potentials (a and b) exceed the Turn Level
b amplitude (set at Turn Level in the EMG Measurement box on the
a EMG page of the Set Conditions window), the peak is counted
as one turn. The total count appears in the Turns column in the
window.
Area: Area bounded by waveform and the baseline. The negative area
is calculated in positive value.
Rise: This item calculates either time interval or slope according to
the setting at the Rise on the EMG page of the Set Conditions
window. When Time is selected, the time interval between 10
and 90% of the fastest up-going process which is also greater
than 50% of the amplitude is displayed.
When Ratio is selected, the slope between 10 and 90% of the
fastest up-going process which is also greater than 50% of the
amplitude is displayed.
fastest up-going process
90%
amplitude
10%
rise time
STIM settings: Stimulation rate, stimulation intensity and duration time. Only
when using electric stimulation.
An EMG Findings window can be created for each site data. The operator’s
desired diagnosis items and the values can be included. Refer to the MEB-9400
operator’s manual for details.
When Site No. 1 and 2 are examinations on the same muscle and examined
site, the contents entered in the EMG Findings Site:1 window are automatically
entered in the EMG Findings Site:2 window.
1. Click the Findings function button or press the F6 key on the keyboard to
open the EMG Findings window.
F1 F2 F3 F4 F5 F6 F7 F8 F9
You can make the EMG Findings window vertical or horizontal. Select the
type on the EMG page of the Set Conditions window.
2. Do one of the following to enter the muscle name in the Muscle: box.
• Type in the muscle name in the Muscle: box from the keyboard.
• Click the button to display the drop down list box. Click the muscle
name in the list box.
• Click the Muscle List... button to display the list of defined muscles. Click
the muscle name in the list. To make a custom muscle list, refer to the
MEB-9400 operator’s manual.
The diagnosis items and drop down list can be set as you want. To customize
the diagnosis items and drop down list, refer to the MEB-9400 operator’s
manual.
6. To search for a data file quickly, you can set a key for a file in the Bookmark
check box. Files can be classified as A or B. Click the A or B check box. To
search for a file, refer to the MEB-9400 operator’s manual.
To display the EMG Findings Summary Table window, click the Findings
Summary function button or press the F4 key on the keyboard.
1. On the Stop window, select Manual MUP function button with the mouse to
display the Manual MUP window. When the Manual MUP Window check
box is selected in the Set Conditions window - Misc. page, the Manual MUP
window automatically opens when the EMG examination program is started.
2. Drag the waveform from the Stop window. Duration, Amplitude, Phases,
Turns, Area and Rise are automatically calculated and the result is displayed
on the Manual MUP window. The EMG analysis time (/div) can be set in the
Set Conditions window - EMG page.
To save the measurement results and waveforms on the hard disk, refer to the
MEB-9400 operator’s manual.
The Condition Table window displays the list of settings in which waveforms
were acquired.
To open the Condition Table window, click the Condition Table function button
or press the F3 key on the keyboard on the Stop window.
Check settings for each waveform. Not all settings can be displayed on the
screen at the same time. To display other settings, use the scroll bar or the jump
buttons at the bottom of the screen.
To close the Condition Table window, click the button on the window.
In the EMG examination program, you can delete all waveforms and calculation
results which are temporarily saved in memory.
F1 F2 F3 F4 F5 F6 F7 F8 F9
2. Click the Erase All function button or press the F8 key on the keyboard.
Erase
All
F8 F9
3. The message window confirming the deletion appears. Click the Yes button.
All waveforms and calculation results are deleted.
To cancel deletion, click the Cancel button on the message window. The
message window closes.
6-3
7-1
7-2
7-3
Section 7-1 Nerve Conduction Study – Basic................................................................................................ 7.1.1
Section 7-2 Measurement – MCS.................................................................................................................. 7.2.1 7-4
Section 7-3 Calculation – MCS...................................................................................................................... 7.3.1
Section 7-4 Measurement – SCS................................................................................................................... 7.4.1 7-5
Section 7-5 Calculation – SCS....................................................................................................................... 7.5.1
Section 7-6 Measurement – Rep.Stim........................................................................................................... 7.6.1 7-6
Section 7-7 Calculation – Rep.Stim................................................................................................................ 7.7.1
Section 7-8 Measurement – F-wave............................................................................................................... 7.8.1 7-7
7-12
7-13
7-14
– Basic
7-1
Operation Flow
This is a typical example. You may not need to do all these. Operating
procedures differ according to the examination program.
7
Stop averaging waveforms.
7-1
When the system power is turned on, enter the patient information at first.
1. Turn the system power on. The Main Menu window automatically opens.
If the Main Menu window is minimized:
• Press the MENU key on the operation panel.
• Or, click the Main Menu on the Task bar.
2. Click the New Patient button on the Main Menu window. The Patient
Information dialog box opens.
3. Enter the patient information such as ID No. Name, etc. For details, refer to
the MEB-9400 operator’s manual.
7
Selecting the Examination Item
7-1
On the Main Menu window, select the examination program. Each program has
its own measurement and stimulation settings.
Or,
Or,
Category tab
When the currently displayed waveform is not saved, a dialog box opens to
prompt you to save the waveform.
You can enter comments in the Main Menu window and edit categories and
programs. Refer to the MEB-9400 operator’s manual for details.
2. Enter the examination information with the keyboard. You can enter the
preset examination information with the list box.
• To clear the entered examination information, click the Clear All button.
• To cancel entering, click the Cancel button.
7
Checking and Changing Measurement Settings
7-1
Before starting measurement, you need to check the measurement settings and
change them if necessary on the Set Conditions window.
1. Click the button on the Tool bar to open the Set Conditions window.
2. Select the condition tab that you want to check. The condition tabs are at the
top of the window.
Condition tab
For the procedure to change the settings, refer to the MEB-9400 operator’s
manual.
Item Description
Input Select Selects the input source from either JB (Electrode junction box) or OFF (No
input). You can change the input source to JB or OFF by pressing the Ctrl key +
F1 to F4 key on the keyboard. The F1 to F4 keys correspond to channels 1 to 4.
Electrode (−) Selects a pair of electrode names for each channel to acquire the waveforms
Electrode (+) from A1, A2, A+, Fz, Pz, C3, C4, Cz, Oz, LO, RO, LT, RT, C5S, EP1, EP2, X1
to X8 and DIN
Sens. (/div) Selects the amplifier sensitivity. When “additional amplifier” is set to “Velocity”,
this selects the amplitude of the velocity waveform.
Hi-cut Selects the high-cut filter frequency of the amplifier.
Lo-cut Selects the low-cut filter frequency of the amplifier.
Acquisition Page
Item Description
Analysis (/div) Selects the analysis time for horizontal length (scale/division) for trigger sweep
and average.
Preset Count Selects the number of waveforms to acquire for averaging.
Trigger Page
Item Description
Stim Rate Selects the stimulation frequency. You can type in the frequency with the
keyboard.
Electric Page
Item Description
Intensity Selects the intensity of the electrical stimulation signal. You can type in
the intensity with the keyboard. You can also change the intensity with the
stimulation intensity dial on the operation panel.
Duration Selects the duration of the electrical stimulation signal. You can type in the width
with the keyboard. You can also change the duration with the STIM DURATION
key.
Display Page 7
Item Description
7-1
Cascade Display Check this box to display cascaded waveforms on the Monitor, Sweep and Stop
windows.
Traces per Field Selects the number of waveforms displayed on one screen.
Labeling Page
Item Description
Label Selects the label for marking the waveform.
Auto Mark Check this box to automatically set the mark on the waveform. The marked
position is set in “Mark Position”.
Mark Position Selects the amplitude mark position on the response waveform when the Auto
Mark check box is checked.
Area Calculation Mode Selects the area calculation mode.
Measure Table Type Selects the Measurement Table window.
Baseline: Calculates the area bounded by the waveform, the baseline and the
vertical lines from the two marks (L1 and L2) to the baseline.
Mark: Calculates the area bounded by the waveform and the line between
the two marks (L1 and L2).
Misc. Page
Item Description
Show Intensity Window Check this box to display the window which displays the stimulation intensity.
Show Temperature Check this box to display the skin temperature window.
Window*
Show Measurement Check this box to display the Measurement window for each examination
Window program when the examination program opens.
Show Nerve/Muscle Check this box to display the table to enter nerve name or muscle name when the
Table examination program opens.
Foot Switches Left, Selects the function for the left and right foot switches.
Right
NCS Page
Item Description
Calculate CCV Set whether or not to display the CCV on the NCS Measurement Table window.
To display the CCV, check the Calculate CCV check box.
Ratio Enter a variation of the conduction velocity per 0.1°C. Use the keyboard to enter
the ratio from 0.0 to 10.0, in steps of 0.1, in the Ratio box.
Reference Temperature Enter the reference skin temperature for the compensation (30.0 to 40.0°C, in
0.1°C steps).
CCV = CV + Ratio × (Reference Temperature − Temperature)
Measure Table Type II Select the type of the Measurement Table window Type II.
Rep. Stim 7
Amp. Page
7-1
Item Description
Input Select Selects the input source from either JB (Electrode junction box) or OFF (No
input). You can change the input source to JB or OFF by pressing the Ctrl key +
F1 to F4 key on the keyboard. The F1 to F4 keys correspond to channels 1 to 4.
Electrode (−) Selects a pair of electrode names for each channel to acquire the waveforms
Electrode (+) from A1, A2, A+, Fz, Pz, C3, C4, Cz, Oz, LO, RO, LT, RT, C5S, EP1, EP2, X1
to X8 and DIN
Sens. (/div) Selects the amplifier sensitivity. When “additional amplifier” is set to “Velocity”,
this selects the amplitude of the velocity waveform.
Hi-cut Selects the high-cut filter frequency of the amplifier.
Lo-cut Selects the low-cut filter frequency of the amplifier.
Acquisition Page
Item Description
Analysis (/div) Selects the analysis time for horizontal length (scale/division) for trigger sweep.
Preset Count Selects the number of waveforms to acquire.
Trigger Page
Item Description
Stim Rate Selects the stimulation frequency. You can type in the frequency with the
keyboard.
Electric Page
Item Description
Intensity Selects the intensity of the electrical stimulation signal. You can type in
the intensity with the keyboard. You can also change the intensity with the
stimulation intensity dial on the operation panel.
Duration Selects the duration of the electrical stimulation signal. You can type in the width
with the keyboard. You can also change the duration with the STIM DURATION
key.
Labeling Page
Item Description
Label Selects the label for marking the waveform.
Auto Mark Check this box to automatically set the mark on the waveform. The marked
position is set in “Mark Position”.
Area Calculation Mode Selects the area calculation mode.
Measure Table Type Selects the Measurement Table window type.
Marking Position
The mark position for automatic marking can be selected at the Measure Mode in the Rep. Stim
Measurement (on the Rep. stim page of the Set Conditions window). The default setting is Half.
<Half>
Area: The area bounded by the baseline, EXCLUSION PERIOD (mark1), and the first baseline crossing of
the waveform after the peak of upward deflection (mark2).
Amp.: The amplitude between the maximum peaks.
EXCLUSION PERIOD amplitude
mark2
Stimulation mark1
Mark position
mark1: appears at the position set on Exclusion Period (on the Rep. stim page of the Set Conditions
window)
mark2: appears at the first baseline crossing of the waveform after the peak of upward deflection
<Full>
Area: The area bounded by the baseline, EXCLUSION PERIOD (mark1), and the 9 div (mark2).
Amp.: The amplitude between the maximum peaks.
amplitude
EXCLUSION PERIOD area
mark2
9 div
Stimulation mark1
Mark position
mark1: appears at the position set on Exclusion Period (on the Rep. stim page of the Set Conditions
window)
mark2: appears on the 9 div from the stimulation point.
• Exclusion Period: Set the Exclusion Period in the Rep. Stim Measurement box on the Rep. stim page of
the Set Conditions window.
This item selects a time interval after the trigger point. Amplitude and area are not
calculated for this interval because stimulus artifact may be included in this interval.
After this exclusion period interval elapses, the calculation starts.
7-1
Item Description
Analysis Wave No. Select up to 10 wave Nos. to select the response waveform to be used for
decrement ratio calculation. The waveform for the selected Wave No. is used for
decrement ratio calculation.
Exclusion Period This item selects a time interval after the trigger point. Amplitude and area are
not calculated for this interval because stimulus artifact may be included in this
interval. After this exclusion period interval elapses, the calculation starts.
Measure Mode Selects either Half or Full.
Number of Bar Number of waveforms for one record on the summary graph.
Number of Sequence Number of sequences to perform.
Setting range is 1 to 12, OFF.
For example, when this item is 6, the system performs sequences 1 to 6. When
OFF is selected, the system performs no sequence.
Rec. Stim. Check the box to acquire waveforms of that sequence.
Stim Rate Stimulation frequency
Preset Count Number of waveforms to be acquired.
Tetanic/Contract Check this box to output the stimulation of the selected Stim. Rate. When the
examiner gives the patient muscle movement instruction, uncheck the box.
Train Time Duration period of stimulation.
Interval Interval between sequences. (in 1 second steps)
Misc. Page
Item Description
Show Intensity Window Check this box to display the window which displays the stimulation intensity.
Show Measurement Check this box to display the Measurement window for each examination
Window program when the examination program opens.
Show Nerve/Muscle Check this box to display the table to enter nerve name or muscle name when the
Table examination program opens.
Foot Switches Left, Selects the function for the left and right foot switches.
Right
F-wave 7
Amp. Page
7-1
Item Description
Input Select Selects the input source from either JB (Electrode junction box) or OFF (No
input). You can change the input source to JB or OFF by pressing the Ctrl key +
F1 to F4 key on the keyboard. The F1 to F4 keys correspond to channels 1 to 4.
Electrode (−) Selects a pair of electrode names for each channel to acquire the waveforms
Electrode (+) from A1, A2, A+, Fz, Pz, C3, C4, Cz, Oz, LO, RO, LT, RT, C5S, EP1, EP2, X1
to X8 and DIN
Sens. (/div) Selects the amplifier sensitivity. When “additional amplifier” is set to “Velocity”,
this selects the amplitude of the velocity waveform.
Hi-cut Selects the high-cut filter frequency of the amplifier.
Lo-cut Selects the low-cut filter frequency of the amplifier.
Right Sens. When “Dual Sens” is checked, the sensitivity of the second period is a multiple
(×2, ×5, ×10, ×20, ×50, or ×100) of the first period.
Split Pos Selects the sensitivity separating position when “Dual Sens” is checked.
Smoothing Check this box to smooth the waveform in the second period by the moving
average method when “Dual Sens.” is checked.
Smoothing Point Selects the number of sampling points when “Smoothing” is checked.
Acquisition Page
Item Description
Analysis (/div) Selects the analysis time for horizontal length (scale/division) for trigger sweep
and average.
Preset Count Selects the number of waveforms to acquire for averaging.
Trigger Page
Item Description
Stim Rate Selects the stimulation frequency. You can type in the frequency with the
keyboard.
Electric Page
Item Description
Intensity Selects the intensity of the electrical stimulation signal. You can type in
the intensity with the keyboard. You can also change the intensity with the
stimulation intensity dial on the operation panel.
Duration Selects the duration of the electrical stimulation signal. You can type in the width
with the keyboard. You can also change the duration with the STIM DURATION
key.
Display Page
Item Description
Cascade Display Check this box to display cascaded waveforms on the Monitor, Sweep and Stop
windows.
Traces per Field Selects the number of waveforms displayed on one screen.
Labeling Page
Item Description
Label Selects the label for marking the waveform.
Auto Mark Check this box to automatically set the mark on the waveform. The marked
position is set in “Mark Position”.
Measure Table Type Selects the Measurement Table window.
Misc. Page
Item Description
Show Intensity Window Check this box to display the window which displays the stimulation intensity.
Show Temperature Check this box to display the skin temperature window.
Window*
Show Measurement Check this box to display the Measurement window for each examination
Window program when the examination program opens.
Show Nerve/Muscle Check this box to display the table to enter nerve name or muscle name when the
Table examination program opens.
Foot Switches Left, Selects the function for the left and right foot switches.
Right
H-reflex 7
Amp. Page
7-1
Item Description
Input Select Selects the input source from either JB (Electrode junction box) or OFF (No
input). You can change the input source to JB or OFF by pressing the Ctrl key +
F1 to F4 key on the keyboard. The F1 to F4 keys correspond to channels 1 to 4.
Electrode (−) Selects a pair of electrode names for each channel to acquire the waveforms
Electrode (+) from A1, A2, A+, Fz, Pz, C3, C4, Cz, Oz, LO, RO, LT, RT, C5S, EP1, EP2, X1
to X8 and DIN
Sens. (/div) Selects the amplifier sensitivity. When “additional amplifier” is set to “Velocity”,
this selects the amplitude of the velocity waveform.
Hi-cut Selects the high-cut filter frequency of the amplifier.
Lo-cut Selects the low-cut filter frequency of the amplifier.
Acquisition Page
Item Description
Analysis (/div) Selects the analysis time for horizontal length (scale/division) for trigger sweep
and average.
Preset Count Selects the number of waveforms to acquire for averaging.
Trigger Page
Item Description
Stim Rate Selects the stimulation frequency. You can type in the frequency with the
keyboard.
Electric Page
Item Description
Intensity Selects the intensity of the electrical stimulation signal. You can type in
the intensity with the keyboard. You can also change the intensity with the
stimulation intensity dial on the operation panel.
Duration Selects the duration of the electrical stimulation signal. You can type in the width
with the keyboard. You can also change the duration with the STIM DURATION
key.
Display Page
Item Description
Cascade Display Check this box to display cascaded waveforms on the Monitor, Sweep and Stop
windows.
Traces per Field Selects the number of waveforms displayed on one screen.
Labeling Page
Item Description
Label Selects the label for marking the waveform.
Auto Mark Check this box to automatically set the mark on the waveform. The marked
position is set in “Mark Position”.
Measure Table Type Selects the Measurement Table window.
Misc. Page
Item Description
Show Intensity Window Check this box to display the window which displays the stimulation intensity.
Show Temperature Check this box to display the skin temperature window.
Window*
Show Measurement Check this box to display the Measurement window for each examination
Window program when the examination program opens.
Show Nerve/Muscle Check this box to display the table to enter nerve name or muscle name when the
Table examination program opens.
Foot Switches Left, Selects the function for the left and right foot switches.
Right
Blink 7
Amp. Page
7-1
Item Description
Input Select Selects the input source from either JB (Electrode junction box) or OFF (No
input). You can change the input source to JB or OFF by pressing the Ctrl key +
F1 to F4 key on the keyboard. The F1 to F4 keys correspond to channels 1 to 4.
Electrode (−) Selects a pair of electrode names for each channel to acquire the waveforms
Electrode (+) from A1, A2, A+, Fz, Pz, C3, C4, Cz, Oz, LO, RO, LT, RT, C5S, EP1, EP2, X1,
X2, X3, X4, X5, X6, X7, X8 and DIN
Sens. (/div) Selects the amplifier sensitivity. When “additional amplifier” is set to “Velocity”,
this selects the amplitude of the velocity waveform.
Hi-cut Selects the high-cut filter frequency of the amplifier.
Lo-cut Selects the low-cut filter frequency of the amplifier.
Fast Recovery Check this box to immediately recover the baseline which is affected by
stimulation artifact. The time constant is set to 15.9 seconds for 1 ms after
stimulation.
Acquisition Page
Item Description
Analysis (/div) Selects the analysis time for horizontal length (scale/division) for trigger sweep
and average.
Trigger Page
Item Description
Stim Rate Selects the stimulation frequency. You can type in the frequency with the
keyboard.
Electric Page
Item Description
Intensity Selects the intensity of the electrical stimulation signal. You can type in
the intensity with the keyboard. You can also change the intensity with the
stimulation intensity dial on the operation panel.
Duration Selects the duration of the electrical stimulation signal. You can type in the width
with the keyboard. You can also change the duration with the STIM DURATION
key.
Labeling Page
Item Description
Label Selects the label for marking the waveform.
Measure Table Type Selects the Measurement Table window.
Misc. Page
Item Description
Show Intensity Window Check this box to display the window which displays the stimulation intensity.
Show Temperature Check this box to display the skin temperature window.
Window*
Show Measurement Check this box to display the Measurement window for each examination
Window program when the examination program opens.
Show Nerve/Muscle Check this box to display the table to enter nerve name or muscle name when the
Table examination program opens.
Foot Switches Left, Selects the function for the left and right foot switches.
Right Default setting: single output from the R side.
7
Measurement Window Description
7-1
Four operation modes are used for measurement in this system. Select a mode
with the MONITOR, STIM/SWEEP, ANALYSIS or STOP key on the operation
panel.
Amplifier
sensitivity
Monitoring time
To display, measure or save a waveform in the hard disk, press the STOP key.
Stimulation and acquisition stops and the Sweep window changes to the Stop
window.
Sensitivity
Number of rejected
waveforms
Sensitivity
Number of averaged
waveforms
Sensitivity
7-2
General.............................................................................................................................................................. 7.2.2
Features.................................................................................................................................................. 7.2.2
Preparation........................................................................................................................................................ 7.2.3
Electrode Attachment and Connection.................................................................................................... 7.2.3
Measurement..................................................................................................................................................... 7.2.5
Selecting the Program and Changing Measurement Settings................................................................ 7.2.5
Entering the Nerve Name........................................................................................................................ 7.2.5
Measurement Procedure......................................................................................................................... 7.2.6
Operations on the Stop Window.............................................................................................................. 7.2.7
Deleting Waveforms................................................................................................................................ 7.2.7
Saving Data on Disk................................................................................................................................ 7.2.7
Printing Data........................................................................................................................................... 7.2.7
General
This program measures the velocity of an impulse through a motor nerve (MCV,
motor conduction velocity). MCV is calculated from the waveform latency and
the distance between two electrical stimulation points on the nerve using evoked
potential derived from electrodes.
Features
• A 2 channel or 4 channel electrode junction box can be used. With the
4 channel electrode junction box, up to 104 response waveforms can be
displayed on the window (4 channels × 26 stages).
• Marks can be set either automatically or manually. The default setting is Auto.
• The nerve conduction velocity is automatically calculated when the distance
between the stimulation sites is entered after the latency cursors are set or the
response waveforms are marked.
• Latency, duration, amplitude, area, amplitude ratio, area ratio, NCV and
CCV (NCV compensated with skin temperature) can be displayed in the
Measurement Table window.
• The settings in which the waveform was acquired can be displayed in the
Condition Table window.
7
Preparation
1. Connect the recording electrode (disk electrode) leads to the electrode jacks
as shown on the following pages.
4. Attach the recording electrodes with paste to the optimal sites as shown on
the following pages.
5. Attach the grounding electrode with Elefix between the active electrode and
the stimulating electrode.
6. Soak the felt pads of the surface stimulation electrode with water.
NOTE
• Do not short-circuit between the felt pads with water.
• Do not touch the stimulating electrode with the grounding electrode.
This causes stimulus artifact.
Grounding electrode
<Median nerve>
Recording site: Abductor pollicis brevis
Active electrode (−): Abductor pollicis brevis
Reference electrode (+): Tendon of the abductor pollicis brevis
8 cm
<Ulnar nerve>
Recording site: Abductor digiti minimi
Active electrode (−): Midportion of the abductor digiti
minimi
Reference electrode (+): Tendon of the abductor digiti minimi
(Proximal phalanx of the fifth digit)
8 cm
Stimulation site (1)
(Distal)
Radial to the flexor carpi
ulnaris tendon at the wrist
8 cm
<Peroneal nerve>
Stimulation site (1)
pillow (Distal)
Between the tendons of the tibialis
Recording site anterior and extensor hallucis
Active electrode (−): Abductor hallucis
Reference electrode (+): Hallux base (Near the big toe)
Recording site
Active electrode (−): Musculus extensor digitorum brevis
Reference electrode (+): Near the little toe
7
Measurement
1. Select the NCS tab in the category tab on the Main Menu window. The
examination program buttons on the Main Menu window change.
2. Select the MCS program on the Main Menu window. The MCS examination
program window opens. The wave area, MCS Measurement Table window,
NCS Measurement Table window, Temperature window, and Nerve Table
window are displayed in the MCS examination program window.
3. Click the button on the Tool bar to display the Set Conditions window
to check or change the measurement settings. Refer to “Checking and
Changing Measurement Settings” in this section.
1. Click the Nerve Input function button or press the F7 key on the keyboard to
open the Nerve Table.
2. From the keyboard, type in the nerve name in the text box. Or, click the
Nerve List... button and select a nerve name from the list.
3. Click either the Left or Right check box. You can also select the side by
pressing the SIDE/SET key on the operation panel.
Measurement Procedure
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Check that there is no artifact.
2. Check that the stimulus intensity is set to 0 mA. If not, press the RESET key
on the operation panel.
3. Press the STIM/SWEEP key on the operation panel to display the waveforms
and stimulation output.
4. Place the stimulating electrode on the stimulation site. The black marked
side of the NM-420S is minus (−). Fix the surface stimulation electrode to
the site with the strap (included in NM-420S).
5. While checking the waveform, turn the stimulation intensity dial on the
operation panel to gradually increase the stimulus intensity. The response
waveform elicited by each electric stimulation appears on the window.
7. Press the STOP key on the operation panel to stop the stimulation and
display the last response waveform obtained before the STOP key was
pressed.
9. Press the Reset key on the operation panel to set the stimulation intensity to
0 mA.
Deleting Waveforms
You can delete unnecessary waveforms by the following procedure.
1. Select the waveforms to delete on the Stop window with the ALL/INDV
key and SELECT ↓ ↑ key on the operation panel or mouse. The selected
waveforms have a different color. You can also select the waveform(s) by
clicking the All, Channel or Individual button on the Tool bar.
Individual button All button ALL/INDV key: Selects one or more waveforms. The All, Channel
or Individual button on the Tool bar indicates the
selected waveform(s).
The All button is down: All waveforms in one stage or site are
selected.
Channel button The Channel button is down: All waveforms in one channel are selected.
The Individual button is down: One individual waveform is selected.
SELECT ↓ ↑ key: Selects the stage if the ALL button is selected.
Selects the channel if the Channel button is
selected.
2. Press the ERASE key on the operation panel. The message confirming
deletion appears.
3. Check that the waveform selection is correct, and press the ERASE key. The
selected waveforms are deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
7-3
General.............................................................................................................................................................. 7.3.2
Nerve Conduction Velocity Calculation.............................................................................................................. 7.3.3
Calculation with Cursors.......................................................................................................................... 7.3.3
Calculation with Marks............................................................................................................................ 7.3.5
Marking.............................................................................................................................................................. 7.3.6
Auto Marking During Acquisition............................................................................................................. 7.3.6
Marking After Acquisition........................................................................................................................ 7.3.6
Setting One Mark at a Time.......................................................................................................... 7.3.6
Setting Multiple Marks Automatically According to the Settings on the Labeling Page................. 7.3.8
Moving a Mark......................................................................................................................................... 7.3.8
Deleting Marks........................................................................................................................................ 7.3.9
Deleting an Individual Mark........................................................................................................... 7.3.9
Deleting All Marks......................................................................................................................... 7.3.9
Measurement Table Windows.......................................................................................................................... 7.3.10
To Open the Measurement Table Window............................................................................................. 7.3.10
To Close the Measurement Table Window............................................................................................ 7.3.11
Measurement Table Window Type I Description.............................................................................................. 7.3.12
MCS Measurement Table Window........................................................................................................ 7.3.12
Annotating a Waveform............................................................................................................... 7.3.12
NCS Measurement Table Window......................................................................................................... 7.3.13
Entering Segment Waveforms..................................................................................................... 7.3.13
Entering Normative Data............................................................................................................. 7.3.14
Entering Skin Temperature.......................................................................................................... 7.3.14
Measurement Table Window Type II Description............................................................................................. 7.3.15
MCS Measurement Table Window........................................................................................................ 7.3.15
Entering the Site......................................................................................................................... 7.3.16
Changing the Waveform Order.................................................................................................... 7.3.16
Displaying Normative Range....................................................................................................... 7.3.16
Creating the CSV File................................................................................................................. 7.3.17
Saving Data on Disk........................................................................................................................................ 7.3.18
Condition Table Window.................................................................................................................................. 7.3.18
Annotating a Waveform................................................................................................................................... 7.3.19
General
7
Nerve Conduction Velocity Calculation
NOTE
The nerve conduction velocity measured with cursors cannot be saved.
Only the nerve conduction velocity measured with marks can be saved
in memory.
cursor A
cursor B
3. Set the vertical line cursor on the optimal onset point of the distal waveform
with the CURSOR dial on the operation panel. You can also drag and move
the vertical line cursor.
5. Set the vertical line cursor on the optimal onset point of the proximal
waveform with the CURSOR dial on the operation panel. You can also drag
and move the vertical line cursor.
7. Enter the distance in the Distance (mm) box with the keyboard or by using
the and buttons next to the Distance (mm) box.
The nerve conduction velocity appears in the NCV row in the Latency
window.
1. If automatic marking function is not on, set the marks “Lat. 1” manually on 7-3
the waveform.
2. Click the Measurement Table function button on the Stop window, or press
the F2 key on the keyboard. The MCS Measurement Table window appears.
The nerve conduction velocity appears in the NCV column in the MCS
Measurement Table window.
Nerve conduction velocity
Marking
1. Click the Marking function button on the Stop window or press the F1
(Marking) key on the keyboard.
The Latency window and Mark Label window appear, and a vertical cursor
appears on the wave area.
Vertical cursor
2) On the wave area, click a waveform at the point you want to add the mark.
The selected mark appears on the clicked point. On the Mark Label window, 7-3
the mark label one line below is highlighted. Up to 12 marks can be set on
one waveform.
2) Turn the CURSOR dial to move the vertical cursor to the position to be
marked.
3) Click the mark label function button you want to set. The mark is set on the
cursor point. The latency (ms) at the point is displayed under the selected
mark label function button.
When the mark label you want to set is not on the function buttons
a) Click the function button to display the next function buttons.
b) Click the desired mark label on the Mark Label window or select the
mark label with the F2 (Select Prev. Mark) or F3 (Select Next Mark) key
on the keyboard. To select marks listed in the upper Mark Label window,
press the F2 key. To select marks listed in the lower Mark Label window,
press the F3 key. The selected mark label is highlighted.
c) Press the F1 (Add Mark) key on the keyboard. The mark is set at the
cursor point.
3. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
With the following procedure, marks are set according to the settings on the
Labeling page on the Set Conditions window after waveform acquisition.
2. On the Stop window, click the Marking function button, or press the F1 key
on the keyboard.
4. Click the Auto Marking function button or press the F4 key on the keyboard.
5. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
Moving a Mark
Do one of the following.
• Drag the mark to the desired position. You can move a mark with this
procedure on the Stop window without opening the Mark Label window.
• On the Mark Label window, select the mark label you want to move. Then, do
the “Setting One Mark at a Time” procedure. The selected mark moves to the
new position.
Deleting Marks 7
Deleting an Individual Mark
• When the mark label you want to delete is on the function button, you can
delete the mark by clicking the function button twice. 7-3
• Or,
1. On the Mark Label window, click the mark label to delete. The cursor
moves to the mark position.
2. Click the Add Mark function button or press the F1 key on the keyboard.
The mark is deleted.
Click the Delete Mark function button or the F6 key on the keyboard.
All marks on the selected waveform are deleted.
There are two types of Measurement Table windows, Type I and Type II. Select either Type I or Type II on the Labeling
page on the Set Conditions window.
Type I: Displays the MCS Measurement Table window and the NCS Measurement Table window.
Type II: The MCS Measurement Table window and the NCS Measurement Table window are combined in one table
window. Displays waveform. The Routine window or S-impose window can be selected on the NCS page on the
Set Conditions window.
<Type I> <Type II>
7-3
When Type I is selected on the Labeling page on the Set Conditions window, the
following windows open.
When the waveforms are marked, the system automatically starts calculation and
displays the results in the Measurement Table window.
Annotating a Waveform
Whether or not to display the Annotation can be set on the Labeling page of the
Set Conditions window.
To annotate a waveform,
1. Click the No. you want to annotate.
You can also annotate a waveform on the Wave Annotation window. Refer to
“Annotating a Waveform” in this section.
7-3
1. On the NCS Measurement Table window, click the Segment column. The
cursor moves to the clicked column.
2. Enter the waveform number in the Segment column. When entering two
waveform numbers (only for Interval), insert “-” between the two waveform
numbers. For example, A1-B1.
When “*” is inserted before the waveform number, the NCV column for the
line is blank.
1. To display the Normative Data column, click the Show ND check box on the
NCS Measurement Table window. A check mark appears in the box, and the
Normative Data column appears on the NCS Measurement Table window.
2. Click the column which you want to enter the normative data in. The cursor
appears in the clicked column.
1. On the NCS Measurement Table window, click the Temp. column. The
cursor moves to the clicked box.
2. From the keyboard, type in the examined skin temperature in 0.1°C steps,
and press the Enter key.
The CCV value is displayed in the CCV column in the NCS Measurement
Table window.
7
Measurement Table Window Type II Description
When Type II is selected on the Labeling page on the Set Conditions window, the
following window opens.
7-3
When the waveforms are marked, the system automatically starts calculation and
displays the results in the Measurement Table window.
Comment (stimulation
site)
Paired waveforms
(The number is waveform order
on the wave window.)
Click the column you want to enter the site in. Then, enter the site from the
keyboard.
When the waveform order is changed on the window, the order in the table also
changes automatically.
When a CSV file is selected in the Normative data file box, following data is
displayed in the NR(+/−2SD) column.
Normative range (normative value ± 2SD)
Mark position is in normative range
Mark position is out of normative range
To display the calculation value and normative value as a pop-up, place the
mouse cursor on the bar, or .
2. Enter text and values. Follow the formats below. Otherwise, the normative
range is not displayed.
- Type the title and comments in the first and second lines.
- Type the normative data in the third or later line.
- The data and sd are 3 digit significant figures, 2 digit decimals. Both the
data and sd must be entered.
Second
line:
Comment
Third line: Second bar in the NR(+/− 2SD) column Maximum value of the
Data and sd for setting Interval column
the normative range
First bar in the NR(+/− 2SD) column
Second
line:
Comment
Third line:
Data and sd for setting
the normative range
- The contents are the same as that of the Excel spreadsheet above.
- Words or data must be separated by commas only. Never put a space or tab
before or after a comma.
To save the measurement results and the waveforms on the hard disk, refer to the
MEB-9400 operator’s manual.
The Condition Table window displays the list of settings in which waveforms
were acquired.
To open the Condition Table window, click the Condition Table function button
on the Stop window.
F1 F2 F3 F4 F5 F6 F7 F8
Check settings for each waveform. Not all settings can be displayed on the
screen at the same time. To display other settings, use the scroll bar or the jump
buttons at the bottom of the screen.
To close the Condition Table window, click the button on the window.
7
Annotating a Waveform
You can annotate a measured waveform. The annotation display position and
annotation display ON/OFF on the Measurement Table window are set on the
Labeling page of the Set Conditions window. Refer to the MEB-9400 operator’s 7-3
manual for details.
1. Click the Marking function button or press the F1 key on the keyboard on
the Stop window.
F1 F2 F3 F4 F5 F6 F7 F8
3. Click the Wave Annotation function button or press the F5 key on the
keyboard to open the Wave Annotation window.
4. Click the or button of the Wave No. box on the Wave Annotation
window to display the waveform to be annotated in the box.
6. Click the OK button. The added annotation appears on both the selected
waveform on the Wave area and the Annotation column on the Measurement
Table window.
General.............................................................................................................................................................. 7.4.2
Features.................................................................................................................................................. 7.4.2 7-4
Preparation........................................................................................................................................................ 7.4.3
Electrode Attachment and Connection.................................................................................................... 7.4.3
SCS Orthodromic.......................................................................................................................... 7.4.3
SCS Antidromic............................................................................................................................. 7.4.3
Measurement..................................................................................................................................................... 7.4.7
Selecting the Program and Changing Measurement Settings................................................................ 7.4.7
Entering the Nerve Name........................................................................................................................ 7.4.7
Measurement Procedure......................................................................................................................... 7.4.8
Operations on the Stop Window.............................................................................................................. 7.4.9
Deleting Waveforms................................................................................................................................ 7.4.9
Saving Data on Disk.............................................................................................................................. 7.4.10
Printing Data......................................................................................................................................... 7.4.10
General
This program measures the sensory nerve conduction velocity. Sensory nerve
conduction velocity is the speed of an impulse through a sensory nerve. SCV is
calculated from the waveform latency and the distance between the electrical
stimulation point and the recording point.
Features
• A 2 channel or 4 channel electrode junction box can be used. With the
4 channel electrode junction box, up to 104 response waveforms can be
displayed on the window (4 channels × 26 stages).
• The marks can be set automatically.
• The nerve conduction velocity is automatically calculated when the distance
between the stimulation sites is entered after the latency cursors are set or
when the response waveforms are marked.
• Latency, amplitude, CV (NCV) and CCV (NCV compensated for skin
temperature) can be displayed in the Measurement Table window.
7
Preparation
1. Connect the recording electrode leads to the electrode jacks as shown on the
following page. 7-4
4. Soak the felt pads of the surface stimulation electrode with water.
5. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where the recording electrode will be attached. Rub the skin with an optimal
amount of Skinpure skin preparation gel to reduce artifact. Wipe the skin
thoroughly with dry gauze to remove any moisture and gel from the skin.
6. Attach the recording electrodes (disk electrode) with paste to the optimal
sites as shown on the following pages.
7. Attach the grounding electrode with Elefix between the active electrode and
the stimulating electrode as shown on the following pages.
NOTE
• Do not short-circuit between the felt pads with water.
• Do not touch the stimulating electrode with the grounding electrode.
This causes stimulus artifact.
SCS Antidromic
5. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where the recording electrode will be attached. Rub the skin with an optimal
amount of the Skinpure skin preparation gel to reduce artifact. Wipe the skin
thoroughly with dry gauze to remove any moisture and gel from the skin.
6. Attach the recording electrodes with paste to the site as shown on the
following pages.
7. Attach the grounding electrode with Elefix between the active electrode and
the stimulating electrode as shown on the following pages.
NOTE
• Do not short-circuit between the felt pads with water.
• Do not touch the stimulating electrode with the grounding electrode.
This causes stimulus artifact.
Grounding electrode
Stimulation:
Stimulating the forefinger with a finger electrode
Stimulation site
Cathode ring: Near the second proximal interphalangeal joint
Anode ring: Second distal interphalangeal joint
14 cm 7-4
A piece of gauze or cotton
should be held between the
forefinger and middle finger.
Recording site:
Recording at the forefinger with a
finger electrode
Recording site
Active electrode (−): Near the second proximal interphalangeal joint
Reference electrode (+): Second distal interphalangeal joint
A piece of gauze or 14 cm
cotton should be held
between the forefinger
and middle finger.
Stimulation:
Stimulating the little finger with a finger electrode
Stimulation site
Cathode ring: Near the fifth proximal
interphalangeal joint
Anode ring: Fifth distal
interphalangeal joint 14 cm 90°
Recording site:
Recording at the little finger with a finger electrode
Stimulation site
Recording site Cathode electrode: Midcalf (the junction of gastrocnemius
Active electrode (−): Behind the lateral muscle and Achilles)
14 cm
malleolus
Reference electrode (+): Distal to the active
electrode
Pillow
7
Measurement
7-4
1. Select the NCS tab in the category tab on the Main Menu window. The
examination program buttons on the Main Menu window change.
2. Select the SCS program on the Main Menu window. The SCS examination
program window opens. The wave area, SCS Measurement Table window,
NCS Measurement Table window, Temperature window, and Nerve Table
window are displayed in the SCS examination program window.
3. Click the button on the Tool bar to display the Set Conditions window
to check or change the measurement settings. Refer to “Checking and
Changing Measurement Settings” in this section.
1. Click the Nerve Input function button or press the F7 key on the keyboard to
open the Nerve Table.
2. From the keyboard, type in the nerve name in the text box. Or, click the
Nerve List... button and select a nerve name from the list.
3. Click either the Left or Right check box. You can also select the side by
pressing the SIDE/SET key on the operation panel.
Measurement Procedure
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Check that there is no artifact.
2. Check that the stimulus intensity is set to 0 mA. If not, press the RESET key
on the operation panel.
5. While checking the waveform, turn the stimulation intensity dial on the
operation panel to gradually increase the stimulation intensity.
<Orthodromic>
First, increase the stimulation intensity from 0 mA until the patient feels
the stimulation. (This intensity is the Sensing Threshold.) Then, gradually
increase the intensity to three times as high as the sensing threshold.
<Antidromic>
Increase the stimulation intensity from 0 mA until the amplitude of the
response waveform becomes maximum. Too strong stimulation causes
artifact.
CAUTION
Take care not to increase the intensity excessively. Excessive
stimulation to the periphery causes pain to the patient.
When the averaging reaches the preset value (set at Preset Count on
the Acquisition page of the Set Conditions window), the averaging and
stimulation stop automatically. When the averaging is completed, the
window changes to the Stop window.
• To stop during averaging, press the STOP key on the operation panel.
• To reset/restart averaging, press the ERASE key on the operation panel.
11. Calculate the sensory nerve conduction velocity. Refer to “Nerve Conductive
Velocity Measurement”.
Deleting Waveforms
You can delete unnecessary waveforms by the following procedure.
1. Select the waveforms to delete on the Stop window with the ALL/INDV
key and SELECT ↓ ↑ key on the operation panel or mouse. The selected
waveforms have a different color. You can also select the waveform(s) by
clicking the All, Channel or Individual button on the Tool bar.
Individual button All button ALL/INDV key: Selects one or more waveforms. The All, Channel
or Individual button on the Tool bar indicates the
selected waveform(s).
The All button is down: All waveforms in one stage or site are
selected.
Channel button The Channel button is down: All waveforms in one channel are selected.
The Individual button is down: One individual waveform is selected.
SELECT ↓ ↑ key: Selects the stage if the ALL button is selected.
Selects the channel if the Channel button is
selected.
2. Press the ERASE key on the operation panel. The message confirming
deletion appears.
3. Check that the waveform selection is correct, and press the ERASE key. The
selected waveforms are deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
General.............................................................................................................................................................. 7.5.2
Nerve Conduction Velocity Calculation.............................................................................................................. 7.5.3
Calculation with Cursors.......................................................................................................................... 7.5.3
Calculation with Marks............................................................................................................................ 7.5.5 7-5
Marking.............................................................................................................................................................. 7.5.6
Auto Marking During Acquisition............................................................................................................. 7.5.6
Marking After Acquisition........................................................................................................................ 7.5.6
Setting One Mark at a Time.......................................................................................................... 7.5.6
Automatically Setting Multiple Marks According to the Settings on the Labeling Page................. 7.5.8
Moving a Mark......................................................................................................................................... 7.5.8
Deleting Marks........................................................................................................................................ 7.5.9
Deleting an Individual Mark........................................................................................................... 7.5.9
Deleting All Marks......................................................................................................................... 7.5.9
Measurement Table Windows.......................................................................................................................... 7.5.10
To Open the Measurement Table Window............................................................................................. 7.5.10
To Close the Measurement Table Window............................................................................................ 7.5.11
Measurement Table Window Type I Description.............................................................................................. 7.5.12
SCS Measurement Table Window......................................................................................................... 7.5.12
Annotating a Waveform............................................................................................................... 7.5.12
NCS Measurement Table Window......................................................................................................... 7.5.13
Entering Segment Waveforms..................................................................................................... 7.5.13
Entering Normative Data............................................................................................................. 7.5.14
Entering Skin Temperature.......................................................................................................... 7.5.14
Measurement Table Window Type II Description............................................................................................. 7.5.15
SCS Measurement Table Window......................................................................................................... 7.5.15
Entering the Site......................................................................................................................... 7.5.16
Changing the Waveform Order.................................................................................................... 7.5.16
Displaying Normative Range....................................................................................................... 7.5.16
Creating the CSV File................................................................................................................. 7.5.17
Saving Data on Disk........................................................................................................................................ 7.5.18
Condition Table Window.................................................................................................................................. 7.5.18
Annotating a Waveform................................................................................................................................... 7.5.19
General
7
Nerve Conduction Velocity Calculation
NOTE
The nerve conduction velocity measured with cursors cannot be saved.
Only the nerve conduction velocity measured with marks can be saved
in memory.
Latency cursor B
Latency cursor A
2. Press the A/B key on the operation panel to select latency cursor A.
3. Set the cursor on the optimal onset point of the distal waveform with the
CURSOR dial on the operation panel. You can also drag and move the
cursor.
5. Set the cursor on the optimal onset point of the proximal waveform with
the CURSOR dial on the operation panel. You can also drag and move the
cursor.
7. Enter the distance in the Distance (mm) box with the keyboard or by using
the and buttons next to the Distance (mm) box.
The nerve conduction velocity appears in the NCV row in the Latency
window.
1. If automatic marking function is not on, set the marks “Lat. 1” manually on
the waveform.
2. Click the Measurement Table function button on the Stop window, or press
the F2 key on the keyboard. The SCS Measurement Table window (type II) 7-5
appears.
4. Enter the distance in the Dist (mm) column in the SCS Measurement Table
window with the keyboard.
Marking
The Latency window and Mark Label window appear, and a vertical cursor
appears on the wave area.
Latency window Mark Label window
Vertical cursor
2) On the wave area, click a waveform at the point you want to add the mark.
The selected mark appears on the clicked point. On the Mark Label window,
the mark label one line below is highlighted. Up to 12 marks can be set on
one waveform.
7-5
Setting a mark with the function buttons
1) Select the waveform to be marked with the SELECT ↓ ↑ key on the
operation panel.
2) Turn the CURSOR dial to move the vertical cursor to the position to be
marked.
3) Click the mark label function button you want to set. The mark is set on the
cursor point. The latency (ms) at the point is displayed under the selected
mark label function button.
When the mark label you want to set is not on the function buttons
a) Click the function button to display the next function buttons.
b) Click the desired mark label on the Mark Label window or select the
mark label with the F2 (Select Prev. Mark) or F3 (Select Next Mark) key
on the keyboard. To select marks listed in the upper Mark Label window,
press the F2 key. To select marks listed in the lower Mark Label window,
press the F3 key. The selected mark label is highlighted.
c) Press the F1 (Add Mark) key on the keyboard. The mark is set at the
cursor point.
3. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
With the following procedure, marks are set according to the settings on the
Labeling page on the Set Conditions window after waveform acquisition.
2. On the Stop window, click the Marking function button, or press the F1 key
on the keyboard.
4. Click the Auto Marking function button or press the F4 key on the keyboard.
5. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
Moving a Mark
On the Mark Label window, select the mark label you want to move. Then, do
the “Setting One Mark at a Time” procedure. The selected mark moves to the
new position.
Deleting Marks 7
Deleting an Individual Mark
• When the mark label you want to delete is on the function button, you can
delete the mark by clicking the function button twice.
7-5
• Or,
1. On the Mark Label window, click the mark label to delete. The cursor
moves to the mark position.
2. Click the Add Mark function button or press the F1 key on the keyboard.
The mark is deleted.
Click the Delete Mark function button or the F6 key on the keyboard.
All marks on the selected waveform are deleted.
There are two types of Measurement Table windows, Type I and Type II. Select either Type I or Type II on the Labeling
page on the Set Conditions window.
Type I: Displays the SCS Measurement Table window and the NCS Measurement Table window.
Type II: The SCS Measurement Table window and the NCS Measurement Table window are combined in one table
window. Waveforms are displayed. The Routine winodow or S-impose window can be selected on the NCS page
on the Set Conditions window.
<Type I> <Type II>
7-5
When Type I is selected on the Labeling page on the Set Conditions window, the
following windows open.
When the waveforms are marked, the system automatically starts calculation and
displays the results in the Measurement Table window.
Annotating a Waveform
Whether or not to display the Annotation can be set on the Labeling page of the
Set Conditions window.
To annotate a waveform,
1. Click the No. you want to annotate.
You can also annotate a waveform on the Wave Annotation window. Refer to
“Annotating a Waveform” in this section.
7-5
1. On the NCS Measurement Table window, click the Segment column. The
cursor moves to the clicked column.
2. Enter the waveform number in the Segment column. When entering two
waveform numbers (only for Interval), insert “-” between the two waveform
numbers. For example, A1-B1.
When “*” is inserted before the waveform number, the NCS column for the
line is blank.
1. To display the Normative Data column, click the Show ND check box on the
NCS Measurement Table window. A check mark appears in the box, and the
Normative Data column appears on the NCS Measurement Table window.
2. Click the column which you want to enter the normative data in. The cursor
appears in the clicked column.
1. On the NCS Measurement Table window, click the Temp. column. The
cursor moves to the clicked column.
2. From the keyboard, type in the examined skin temperature in 0.1°C steps,
and press the Enter key.
The CCV value is displayed in the CCV column in the NCS Measurement
Table window.
7
Measurement Table Window Type II Description
When Type II is selected on the Labeling page on the Set Conditions window, the
following window opens.
7-5
Comment (stimulation
site)
Paired waveforms
(The number is
waveform order on the
wave window.)
Distance between
the waveforms in the
Segment column
(unit: mm)
Click the column you want to enter the site name and enter the site from the
keyboard.
When waveforms are moved, added or deleted in the window, the table is
automatically updated to reflect this.
When a CSV file is selected in the Normative data file box, the following data is
displayed in the NR(+/−2SD) column.
To display the calculation value and normative value, set the mouse at the bar,
or .
2. Enter text and values. Follow the formats below. Otherwise, the normative
range is not displayed.
7-5
- Type the title and comments in the first and second lines.
- Type the normative data in the third or later line.
- The data and SD have 3 significant digits and 2 decimal places. Both the
data and SD must be entered together.
Second line:
Comment
Third line:
Data and SD for setting Second bar in the NR(+/− 2SD) column
the normative range
First bar in the NR(+/− 2SD) column
Second line:
Comment
Third line:
Data and sd for setting
the normative range
- The contents are the same as that of the Excel spreadsheet above.
- Words or data must be separated by commas only. Never put a space or tab
before or after a comma.
To save the measurement results and the waveforms on the hard disk, refer to the
MEB-9400 operator’s manual.
The Condition Table window displays the list of settings in which waveforms
were acquired.
To open the Condition Table window, click the Condition Table function button
on the Stop window.
F1 F2 F3 F4 F5 F6 F7 F8
Check settings for each waveform. Not all settings can be displayed on the
screen at the same time. To display other settings, use the scroll bar or the jump
buttons at the bottom of the screen.
To close the Condition Table window, click the button on the window.
7
Annotating a Waveform
You can annotate a measured waveform. The annotation display position and
annotation display ON/OFF on the Measurement Table window are set on the
Labeling page of the Set Conditions window. Refer to the MEB-9400 operator’s
manual for details.
1. Click the Marking function button or press the F1 key on the keyboard on
the Stop window. 7-5
F1 F2 F3 F4 F5 F6 F7 F8
3. Click the Wave Annotation function button or press the F5 key on the
keyboard to open the Wave Annotation window.
4. Click the or button of the Wave No. box on the Wave Annotation
window to display the waveform to be annotated in the box.
6. Click the OK button. The added annotation appears on both the selected
waveform on the Wave area and the Annotation column on the Measurement
Table window.
General.............................................................................................................................................................. 7.6.2
Features.................................................................................................................................................. 7.6.2
Preparation........................................................................................................................................................ 7.6.3
Electrode Attachment and Connection.................................................................................................... 7.6.3
Measurement..................................................................................................................................................... 7.6.5
Selecting the Program and Changing Measurement Settings................................................................ 7.6.5 7-6
Entering the Nerve Name........................................................................................................................ 7.6.5
Measurement Procedure......................................................................................................................... 7.6.6
Automatic Measurement Procedure........................................................................................................ 7.6.7
Operations on the Stop Window.............................................................................................................. 7.6.7
Calling Up Waveforms and Data from Memory....................................................................................... 7.6.7
Deleting Waveforms................................................................................................................................ 7.6.8
Saving Data on Disk................................................................................................................................ 7.6.8
Printing Data........................................................................................................................................... 7.6.8
General
Features
• All the response waveforms appear on the window with the time scale
compressed to monitor each response waveform. (Compressed waveform
display)
• 10 response waveforms selected from all the response waveforms appear on
the window superimposed. The onset time of each waveform is shifted. (This
is called the Delayed-superimposed waveform display.)
• Up to 12 steps can be set for one automatic measurement.
• 10 waveforms can be marked, and each amplitude and area decrement ratio are
automatically calculated and displayed on the right side of the window. The
waveforms can be marked either automatically or manually.
• The response waveforms of up to 99 Rec. No.s can be saved temporarily in
memory.
• Measurement sites, amplitude, area and the decrement ratio of the waveform
are displayed in the Measurement Table window. The Rep. stim Measurement
Table window type II has the summary graph.
7
Preparation
5. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where the recording electrode will be attached. Rub the skin with an optimal
amount of the Skinpure skin preparation gel to reduce artifact. Wipe the skin
thoroughly with dry gauze to remove any moisture and gel from the skin.
6. Attach the recording electrodes (disk electrode) with paste to the optimal
sites as shown on the following pages.
7. Attach the grounding electrode with Elefix between the active electrode and
the stimulating electrode as shown on the following pages.
NOTE
• Make sure that the electrodes are properly attached and fixed.
• Do not short-circuit between the felt pads of the surface stimulation
electrode with water.
• To prevent artifact caused by muscle expansion/contraction, immobilize
the examined extremity.
Grounding electrode
Stimulation site
Radial to the flexor carpi ulnaris tendon at the wrist
Grounding electrode
7
Measurement
1. Select the NCS tab in the category tab on the Main Menu window. The
examination program buttons on the Main Menu window change.
2. Select the Rep. stim program on the Main Menu window. The Rep. stim 7-6
examination program window opens. The wave area and Nerve Table
window are displayed in the Rep. stim examination program window.
3. Click the button on the Tool bar to display the Set Conditions window
to check or change the measurement settings. Refer to “Checking and
Changing Measurement Settings” in this section.
1. Click the Nerve Input function button or press the F7 key on the keyboard to
open the Nerve Table.
2. From the keyboard, type in the nerve name in the text box. Or, click the
Nerve List... button and select a nerve name from the list.
3. Click either the Left or Right check box. You can also select the side by
pressing the SIDE/SET key on the operation panel.
Measurement Procedure
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Check that there is no artifact.
2. Check that the stimulus intensity is set to 0 mA. If not, press the RESET key
on the operation panel.
4. Place the surface stimulation electrode or the SOMATO control box probes
on the stimulation site.
The black marked side of the NM-420S surface stimulation electrode is
minus (−).
5. While checking the waveform, turn the stimulation intensity dial on the
operation panel to gradually increase the stimulation intensity. The response
waveform elicited by each electric stimulation appears on the window.
8. Press the ANALYSIS key on the operation panel to start the repetitive
stimulation at the preset stimulation rate. (The rate is set in the Trigger page
of the Set Conditions window.)
When the number of stimulations reaches the preset value (set at Preset
Count on the Acquisition page of the Set Conditions window), the
stimulation stops automatically.
To start the automatic measurement, press the ANALYSIS key on the operation
panel. While stimulation is given or during intervals, the remaining period is
displayed on the Rep.stim Measurement Table window (type II).
When the step which is set at Number of Sequence finishes, the automatic
measurement stops. You can also stop the automatic measurement by pressing 7-6
the STOP key.
Click the Rec. No. button or Rec. No. button to select the measurement
results to be displayed. Or, type in the recording number in the Rec. No. box.
The selected measurement result is displayed.
Deleting Waveforms
Waveforms and data for an unnecessary Rec. No. can be deleted.
1. On the Stop window, select the Rec. No. you want to delete by clicking
the or button at the right of the Rec. No. box. The temporarily saved
waveforms for the selected Rec. No. are displayed.
2. Press the ERASE key on the operation panel. A message confirming deletion
appears.
3. Check that the Rec. No. selection is correct, and press the ERASE key. The
selected Rec. No. is deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
General.............................................................................................................................................................. 7.7.2
Marking.............................................................................................................................................................. 7.7.3
Auto Marking During Acquisition............................................................................................................. 7.7.3
Marking After Acquisition........................................................................................................................ 7.7.3
Setting One Mark at a Time.......................................................................................................... 7.7.3
Setting Multiple Marks Automatically According to the Settings on the Labeling Page................. 7.7.5
Moving a Mark......................................................................................................................................... 7.7.5
Deleting Marks........................................................................................................................................ 7.7.6 7-7
General
In the Rep. stim examination program, the following calculation operations are
available.
• Marking
• Displaying the Measurement Table Window
• Displaying the Summary window
• Displaying the Condition Table Window
• Annotating a Waveform
7
Marking
The Latency window and Mark Label window appear, and a vertical cursor
appears on the wave area.
Latency window Mark Label window
Vertical cursor
M_lat., M_amp. 1, M_amp. 2: 2) On the wave area, click a waveform at the point you want to add the mark.
One mark per record. The selected mark appears on the clicked point. On the Mark Label window,
F_lat., F_amp. 1, F_amp. 2:
the mark label one line below is highlighted. Up to 12 marks can be set on
One mark per sweep waveform
one waveform.
2) Turn the CURSOR dial to move the vertical cursor to the position to be
marked.
3) Click the mark label function button you want to set. The mark is set on the
cursor point. The latency (ms) at the point is displayed under the selected
mark label function button.
When the mark label you want to set is not on the function buttons
a) Click the function button to display the next function buttons.
b) Click the desired mark label on the Mark Label window or select the
mark label with the F2 (Select Prev. Mark) or F3 (Select Next Mark) key
on the keyboard. To select marks listed in the upper Mark Label window,
press the F2 key. To select marks listed in the lower Mark Label window,
press the F3 key. The selected mark label is highlighted.
c) Press the F1 (Add Mark) key on the keyboard. The mark is set at the
cursor point.
3. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
With the following procedure, marks are set according to the settings on the
Labeling page on the Set Conditions window after waveform acquisition.
2. On the Stop window, click the Marking function button, or press the F1 key
on the keyboard.
7-7
4. Click the Auto Marking function button or press the F4 key on the keyboard.
5. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
Moving a Mark
Do one of the following.
• Drag the mark to the desired position. You can move a mark with this
procedure on the Stop window without opening the Mark Label window.
• On the Mark Label window, select the mark label you want to move. Then, do
the “Setting One Mark at a Time” procedure. The selected mark moves to the
new position.
Deleting Marks
Deleting an Individual Mark
• When the mark label you want to delete is on the function button, you can
delete the mark by clicking the function button twice.
• Or,
1. On the Mark Label window, click the mark label to delete. The cursor
moves to the mark position.
2. Click the Add Mark function button or press the F1 key on the keyboard.
The mark is deleted.
Click the Delete Mark function button or the F6 key on the keyboard.
All marks on the selected waveform are deleted.
7
Measurement Table Windows
There are two types of Measurement Table windows, Type I and Type II. Select
either Type I or Type II on the Labeling page on the Set Conditions window.
Type I: For 10 response waveforms, the amplitude, area and decrement ratio
are displayed.
Type II: For 10 response waveforms, the amplitude, area and decrement ratio
are displayed. A summary graph is also displayed.
7-7
Or click the tool button or the Measurement Table function button again.
When Type I is selected on the Labeling page on the Set Conditions window, the
following windows open.
For 10 response waveforms, the amplitude, area and decrement ratio are
displayed.
Decr. (%)
Amplitude decrement ratio (1 − AMPSTIM / AMPREF) × 100%
Decr. (%)
Area decrement ratio (1 − AREASTIM / AREAREF) × 100%
Annotating a Waveform 7
Whether or not to display the Annotation can be set on the Labeling page of the
Set Conditions window.
To annotate a waveform,
1. Click the No. you want to annotate.
You can also annotate a waveform on the Wave Annotation window. Refer to
“Annotating a Waveform” in this section.
7-7
Entering the Stim. Site and Recording Site
1. Select the Stim. Site column or Recording Site column. The cursor moves to
the selected column.
1. Click the button on the right of the Reference Stim. No. box. The drop
down list opens.
2. Select the desired Wave. No. in the list. Only the Wave No. which are not set
to OFF at “Analysis Wave No.” on the Rep. stim page of the Set Conditions
window appears.
Click the button on the Channel No. box. The drop down list appears. Select
the desired channel.
When Type II is selected on the Labeling page on the Set Conditions window, the
following window opens.
Preset number of
waveforms to be acquired
Decr. (%):
Amplitude decrement ratio
(1 − AMPSTIM / AMPREF) × 100%
Summary graph:
Amplitude of the waveforms in Stim.
No. are displayed. One stick stands
for a response waveform.
Select the Stim. No. column and type in the stimulation number from the
keyboard. Pressing the Default button returns the contents in the Stim. No.
column to the settings at Analysis wave No. on the Rep.Stim page of the Set
Conditions window.
7-7
Summary Graph
The amplitudes of response waveforms are shown. Graphs for up to 12 Rec. Nos.
can be displayed.
One bar stands for one response waveform. The number of the bars displayed in
the graph is set on Number of Bar on the Rep. stim page on the Set Conditions
window. When the number of response waveforms exceeds the number set at
Number of Bar, waveforms with the smaller Stim Nos. are displayed.
Bars for the waveforms of the selected Rec. No. are shown in green, and bars for
the waveforms of the other Rec. Nos. are shown in light blue.
You can jump to an other Rec. No. by clicking the corresponding graph. The
wave window and the Rep. stim Measurement Table displays data for the Rec.
No.
Below the graph, the stimulation rate and the amplitude decrement ratio of the
rightmost bar are displayed.
Stimulation rate
The results of all Rep. Stim measurements are displayed on the Rep.stim
Summary window. Data of amplitude and area decrement ratio for up to 10
response waveforms are displayed for one Rec. No. measurement.
The Rep.stim Summary Window is correlated with the Rep. stim Measurement
Table window.
To close the Rep.stim Summary window, click the button on the window or
click the Summary function button or press the F4 key on the keyboard again.
Recording number
7
Saving Data on Disk
To save the measurement results and the waveforms on the hard disk, refer to the
MEB-9400 operator’s manual.
The Condition Table window displays the list of settings in which waveforms
were acquired.
To open the Condition Table window, click the Condition Table function button
on the Stop window.
F1 F2 F3 F4 F5 F6 F7 F8
Check settings for each waveform. Not all settings can be displayed on the
screen at the same time. To display other settings, use the scroll bar or the jump
buttons at the bottom of the screen.
To close the Condition Table window, click the button on the window.
Annotating a Waveform
You can annotate a measured waveform. The annotation display position and
annotation display ON/OFF on the Measurement Table window are set on the
Labeling page of the Set Conditions window. Refer to the MEB-9400 operator’s
manual for details.
1. Click the Marking function button or press the F1 key on the keyboard on
the Stop window.
F1 F2 F3 F4 F5 F6 F7 F8
3. Click the Wave Annotation function button or press the F5 key on the
keyboard to open the Wave Annotation window.
4. Click the or button of the Wave No. box on the Wave Annotation
window to display the waveform to be annotated in the box.
6. Click the OK button. The added annotation appears on both the selected
waveform on the Wave area and the Annotation column on the Measurement
Table window.
General.............................................................................................................................................................. 7.8.2
Features.................................................................................................................................................. 7.8.2
Preparation........................................................................................................................................................ 7.8.3
Electrode Attachment and Connection.................................................................................................... 7.8.3
Measurement..................................................................................................................................................... 7.8.5
Selecting the Program and Changing Measurement Settings................................................................ 7.8.5
Entering the Nerve Name........................................................................................................................ 7.8.5
Measurement Procedure......................................................................................................................... 7.8.6
Operations on the Stop Window.............................................................................................................. 7.8.7
7-8
Calling Up Waveforms and Data from Memory....................................................................................... 7.8.7
Deleting Waveforms................................................................................................................................ 7.8.7
Saving Data on Disk................................................................................................................................ 7.8.8
Printing Data........................................................................................................................................... 7.8.8
General
Features
• Latency and amplitude of each F-wave are automatically measured and
displayed on the window in real time during measurement.
• With the Dual Sensitivity function, the M-wave is displayed on the left of the
window and the F-wave on the right.
• The F-wave conduction velocity is automatically calculated when the response
waveforms are marked and the distance between the stimulation site and the
spinal cord is entered. The marks can be set either automatically or manually.
• The following data are displayed in the Measurement table window.
- M-wave latency and amplitude
- F-wave latency and amplitude
- The latency between M-wave and F-wave
- F-wave conduction velocity
- F-wave occurrence ratio (Number of F-waves/Number of total waveform
traces)
• The following data are displayed in the F-wave histogram window.
- F-wave latency histogram
- Minimum, maximum, and mean value for each measurement
- Standard deviation
• Up to 10 measurement results can be temporarily saved in memory.
• Up to 9999 waveforms for one measurement can be temporarily saved in
memory.
7
Preparation
5. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where the recording electrode will be attached. Rub the skin with an optimal
amount of the skin preparation gel, Skinpure, to reduce artifact. Wipe the
skin thoroughly with dry gauze to remove any moisture and gel from the
skin.
6. Attach the active and reference electrodes with paste to the optimal sites.
Use the disk electrode for the active and reference electrodes.
7. Attach the grounding electrode with Elefix between the active electrode and
the stimulating electrode.
NOTE
• Do not short-circuit between the felt pads with water.
• Do not touch the stimulating electrode with the grounding electrode.
This causes stimulus artifact.
Grounding electrode
CH1(−) Active electrode (−): Midportion of the abductor digiti minimi muscle
Recording
CH1(+) Reference electrode (+): Tendon of the abductor digiti minimi muscle
electrode
(Proximal phalanx of the little finger)
Grounding electrode
7
Measurement
1. Select the NCS tab in the category tab on the Main Menu window. The
examination program buttons on the Main Menu window change.
2. Select the F-wave program on the Main Menu window. The F-wave
examination program window opens. The wave area and Nerve Table
window are displayed in the F-wave examination program window.
3. Click the button on the Tool bar to display the Set Conditions window 7-8
to check or change the measurement settings. Refer to “Checking and
Changing Measurement Settings” in this section.
1. Click the Nerve Input function button or press the F7 key on the keyboard to
open the Nerve Table window.
2. From the keyboard, type in the nerve name in the text box. Or, click the
Nerve List... button and select a nerve name from the list.
3. Click either the Left or Right check box. You can also select the side by
pressing the SIDE/SET key on the operation panel.
Measurement Procedure
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Check that there is no artifact.
2. Check that the stimulus intensity is set to 0 mA. If not, press the RESET key
on the operation panel.
4. Place the stimulating electrode on the stimulation site. Put the (−) side
proximal to the (+) side. The black marked side of the NM-420S surface
stimulation electrode is minus (−).
5. While checking the waveform, turn the stimulation intensity dial on the
operation panel to gradually increase the stimulation intensity. The response
waveform elicited by each electric stimulation appears on the window.
When the number of acquired waveforms reaches the preset value (set at
Preset Count on the Acquisition page of the Set Conditions window), the
stimulation stops automatically.
7-8
Click the Rec. No. button or Rec. No. button to select the measurement
results to be displayed. Or, type in the recording number in the Rec. No. box.
The selected measurement result is displayed on the wave area.
Deleting Waveforms
Waveforms and data for an unnecessary Rec. No. can be deleted.
1. On the Stop window, select the Rec. No. you want to delete by clicking
the or button at the right of the Rec. No. box. The temporarily saved
waveforms for the selected Rec. No. are displayed.
2. Press the ERASE key on the operation panel. A message confirming deletion
appears.
3. Check that the Rec. No. selection is correct, and press the ERASE key. The
waveforms and data for the selected Rec. No. is deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
General.............................................................................................................................................................. 7.9.2
F-wave Conduction Velocity Measurement........................................................................................................ 7.9.3
Calculation with Cursors.......................................................................................................................... 7.9.3
Calculation with Marks............................................................................................................................ 7.9.5
Marking.............................................................................................................................................................. 7.9.6
Auto Marking During Acquisition............................................................................................................. 7.9.6
Marking After Acquisition........................................................................................................................ 7.9.6
Setting One Mark at a Time.......................................................................................................... 7.9.6
Setting Multiple Marks Automatically According to the Settings on the Labeling Page................. 7.9.8
Moving a Mark......................................................................................................................................... 7.9.8
Deleting Marks........................................................................................................................................ 7.9.9
7-9
Deleting an Individual Mark........................................................................................................... 7.9.9
Deleting All Marks......................................................................................................................... 7.9.9
F-wave Measurement Table Window............................................................................................................... 7.9.10
To Open the Measurement Table Window............................................................................................. 7.9.10
To Close the Measurement Table Window............................................................................................ 7.9.10
Measurement Table Window Type I Description.............................................................................................. 7.9.11
F-wave Measurement Table Window..................................................................................................... 7.9.11
Annotating a Waveform............................................................................................................... 7.9.12
Entering the Stim. Site and Recording Site................................................................................. 7.9.12
Entering the Distance (mm)........................................................................................................ 7.9.12
Entering the Amplitude Threshold............................................................................................... 7.9.12
F-wave Histogram Window Description................................................................................................. 7.9.13
Measurement Table Window Type II Description............................................................................................. 7.9.14
F-wave Measurement Table Window..................................................................................................... 7.9.14
Selecting the Rec. No.................................................................................................................. 7.9.15
Entering the Stim. Site and Recording Site................................................................................. 7.9.15
F-wave Table Window............................................................................................................................ 7.9.15
Selecting the Rec. No.................................................................................................................. 7.9.16
Entering the Amplitude Threshold............................................................................................... 7.9.16
F-wave Sensitivity Change.............................................................................................................................. 7.9.17
Saving Data on Disk........................................................................................................................................ 7.9.18
Condition Table Window.................................................................................................................................. 7.9.18
Annotating a Waveform................................................................................................................................... 7.9.19
General
7
F-wave Conduction Velocity Measurement
and the spinal cord is entered, the F-wave conduction velocity is automatically
calculated.
Latency 1, Latency 2, Interval and FWCV are displayed in the Latency window.
NOTE
The F-wave conduction velocity measured with cursors cannot be saved.
Only the F-wave conduction velocity measured with marks can be saved
in memory.
1. Press the LAT/AMP/TRIG key on the operation panel, or click the tool
button. The Latency cursor and Latency window appear on the wave area.
When the STOP key is pressed during measurement, the LAT/AMP/TRIG
key on the operation panel or the button on the Tool bar becomes
automatically pressed.
Latency window
Latency cursor
Entered
distance
3. Set the cursor on the optimal onset point of the M-wave with the CURSOR
dial on the operation panel. You can also drag and move the cursor with the
mouse.
5. Set the cursor on the optimal onset point of the F-wave with the CURSOR
dial on the operation panel. You can also drag and move the cursor with the
mouse.
6. Measure the distance between the stimulation site and the spinal cord.
7. Enter the distance in the Distance (mm) box with keyboard or by using either
the button next to the Distance (mm) box.
The F-wave conduction velocity appears in the FWCV row in the Latency
window.
1. If automatic marking function is not on, set the marks (M-Lat., F-Lat.)
manually on the waveform. Refer to “Marking After Acquisition” in this
section.
2. Click the Measurement Table function button on the Stop window, or press
the F2 key on the keyboard. The F-wave Measurement Table window
appears.
3. Measure the distance between the stimulation site and the spinal cord.
4. Enter the distance in the Distance (mm) box with keyboard or by using either
the button next to the Distance (mm) box. The F-wave conduction
velocity appears in the FWCV column in the F-wave Measurement Table
window.
Marking
1. Click the Marking function button on the Stop window or press the F1
(Marking) key on the keyboard.
The Latency window and Mark Label window appear, and a vertical cursor
appears on the wave area.
Latency window Mark Label window
2) On the wave area, click a waveform at the point you want to add the mark.
The selected mark appears on the clicked point. On the Mark Label window,
the mark label one line below is highlighted. Up to 12 marks can be set on
one waveform.
2) Turn the CURSOR dial to move the vertical cursor to the position to be
marked.
7-9
3) Click the mark label function button you want to set. The mark is set on the
cursor point. The latency (ms) at the point is displayed under the selected
mark label function button.
When the mark label you want to set is not on the function buttons
a) Click the function button to display the next function buttons.
b) Click the desired mark label on the Mark Label window or select the
mark label with the F2 (Select Prev. Mark) or F3 (Select Next Mark) key
on the keyboard. To select marks listed in the upper Mark Label window,
press the F2 key. To select marks listed in the lower Mark Label window,
press the F3 key. The selected mark label is highlighted.
c) Press the F1 (Add Mark) key on the keyboard. The mark is set at the
cursor point.
3. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
With the following procedure, marks are set according to the settings on the
Labeling page on the Set Conditions window after waveform acquisition.
2. On the Stop window, click the Marking function button, or press the F1 key
on the keyboard.
4. Click the Auto Marking function button or press the F4 key on the keyboard.
5. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
Moving a Mark
Do one of the following.
• Drag the mark to the desired position. You can move a mark with this
procedure on the Stop window without opening the Mark Label window.
• On the Mark Label window, select the mark label you want to move. Then, do
the “Setting One Mark at a Time” procedure. The selected mark moves to the
new position.
Deleting Marks 7
Deleting an Individual Mark
• When the mark label you want to delete is on the function button, you can
delete the mark by clicking the function button twice.
• Or,
1. On the Mark Label window, click the mark label to delete. The cursor
moves to the mark position.
2. Click the Add Mark function button or press the F1 key on the keyboard.
The mark is deleted.
Click the Delete Mark function button or the F6 key on the keyboard.
All marks on the selected waveform are deleted.
There are two types of F-wave Measurement Table windows: Type I and Type
II. Select either Type I or Type II on the Labeling page on the Set Conditions
window.
Type I: Calculation values are displayed. You can open the F-wave Histogram
window from this window.
Type II: Calculation values, F-wave histogram and superimposed waveforms are
displayed. You can open the F-wave Table window from this window.
<Type I> <Type II>
Or click the tool button or the Measurement Table function button again.
7
Measurement Table Window Type I Description
When Type I is selected on the Labeling page on the Set Conditions window, the
following windows open.
7-9
Annotating a Waveform
Whether or not to display the Annotation can be set on the Labeling page of the
Set Conditions window.
To annotate a waveform,
1. Click the No. you want to annotate.
You can also annotate a waveform on the Wave Annotation window. Refer to
“Annotating a Waveform” in this section.
1. Click the Stim. Site text box or Recording Site text box. The cursor moves to
the clicked text box.
2. Type in the stimulating site and recording site in each text box.
1. Click the Distance (mm) box. The cursor moves to the text box.
2. Type in the distance between the stimulating point and the spinal cord.
1. Click the button on the right of the Amplitude Threshold box. The drop
down list opens.
To display the F-wave Histogram window, click the Histogram... button on the
F-wave Measurement Table window.
7-9
When Type II is selected on the Labeling page on the Set Conditions window, the
following window opens.
Stimulation settings
Histogram
Amplitude of a waveform is
shown as a bar.
Superimposed waveforms
(Baseline for waveforms are adjusted into the same level.)
F-wave Occurrence: Ratio of number of F-waves over the threshold to all the 7
response waveforms. (F-wave/all response waveforms)
A waveform is recognized as an F-wave when its
amplitude is over the value set at Amplitude Threshold.
Count box: Set the maximum number of F-wave occurrences for the
Y-axis of the histogram.
F-latency (ms) box: Set the maximum F-wave latency for the X-axis of the
histogram.
Latency Histogram Vertical axis: Number of F-wave occurrences. (The
maximum value is set at the Count box.)
Horizontal axis: F-wave latency (The maximum value is
set at the F-latency box.)
Click the and buttons on the Rec. No. box to select the desired recording
number.
7-9
1. Click the Stim. Site text box or Recording Site text box.
2. Type in the stimulation site and recording site in each text box.
Click the and buttons on the Rec. No. box to select the desired recording
number.
1. Click the button on the right of the Amplitude Threshold box. The drop
down list opens.
7
F-wave Sensitivity Change
The display gain (sensitivity) of the F-wave (right half of the wave area) can be
temporarily changed on the Stop window.
1. Click the Second Sens. function button or press the F4 key on the keyboard.
(The Second Sens. function button is pressed.)
When adjusting the sensitivity of the M-wave, select the Second Sens.
7-9
function button (The Second Sens. function button is unpressed), then adjust
the sensitivity with the VERTICAL RANGE ↓ / ↑ key on the operation
panel.
To save the measurement results and the waveforms on the hard disk, refer to the
MEB-9400 operator’s manual.
The Condition Table window displays the list of settings in which waveforms
were acquired.
To open the Condition Table window, click the Condition Table function button
on the Stop window.
F1 F2 F3 F4 F5 F6 F7 F8
Check settings for each waveform. Not all settings can be displayed on the
screen at the same time. To display other settings, use the scroll bar or the jump
buttons at the bottom of the screen.
To close the Condition Table window, click the button on the window.
7
Annotating a Waveform
You can annotate a measured waveform. The annotation display position and
annotation display ON/OFF on the Measurement Table window are set on the
Labeling page of the Set Conditions window. Refer to the MEB-9400 operator’s
manual for details.
1. Click the Marking function button or press the F1 key on the keyboard on
the Stop window.
F1 F2 F3 F4 F5 F6 F7 F8
4. Click the or button of the Wave No. box on the Wave Annotation
window to display the waveform to be annotated in the box.
6. Click the OK button. The added annotation appears on both the selected
waveform on the Wave area and the Annotation column on the Measurement
Table window.
General............................................................................................................................................................ 7.10.2
Features................................................................................................................................................ 7.10.2
Preparation...................................................................................................................................................... 7.10.3
Electrode Attachment and Connection.................................................................................................. 7.10.3
Measurement................................................................................................................................................... 7.10.5
Selecting the Program and Changing Measurement Settings.............................................................. 7.10.5
Entering the Nerve Name...................................................................................................................... 7.10.5
Measurement Procedure....................................................................................................................... 7.10.6
Copying the Acquired Waveforms to the Stop Window......................................................................... 7.10.7
Copying Procedure..................................................................................................................... 7.10.7
Operations on the Stop Window............................................................................................................ 7.10.8
Deleting Waveforms.............................................................................................................................. 7.10.8
Saving Data on Disk.............................................................................................................................. 7.10.8
7-10
Printing Data......................................................................................................................................... 7.10.8
General
Features
• Up to 9999 sweep waveforms can be acquired for each channel. The
measurement results and the waveforms can be saved on the hard disk.
• The data for up to 26 stages can be displayed in the Measurement Table
window.
• The latency, amplitude, and the maximum amplitude with latency of each
M-wave and H-wave can be automatically calculated and displayed in the
Measurement Table window.
• Superimposed waveforms and the Intensity-Amplitude graph are displayed on
the H-reflex Measurement Table window (type II).
7
Preparation
4. Soak the felt pads of the surface stimulation electrode with water.
5. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where the recording electrode will be attached. Rub the skin with an optimal
amount of Skinpure skin preparation gel to reduce artifact. Wipe the skin 7-10
thoroughly with dry gauze to remove any moisture and gel from the skin.
6. Attach the active and reference electrodes with paste to the optimal sites
as shown on the following page. Use the disk electrode for the active and
reference electrodes.
7. Attach the grounding electrode with Elefix between the active electrode and
the stimulating electrode.
NOTE
• Do not short-circuit between the felt pads with water.
• Do not touch the stimulating electrode with the grounding electrode.
This causes stimulus artifact.
Stimulation site
Tibial nerve at the popliteal fossa
Grounding electrode
Grounding electrode
7
Measurement
1. Select the NCS tab in the category tab on the Main Menu window. The
examination program buttons on the Main Menu window change.
2. Select the H-reflex program on the Main Menu window. The H-reflex
examination program window opens. The wave area and Nerve Table
window are displayed in the H-reflex examination program window.
3. Click the button on the Tool bar to display the Set Conditions window
to check or change the measurement settings. Refer to “Checking and
Changing Measurement Settings” in this section.
7-10
Entering the Nerve Name
Enter the nerve name to be examined and its side on the Nerve Table window.
1. Click the Nerve Input function button or press the F7 key on the keyboard to
open the Nerve Table.
2. From the keyboard, type in the nerve name in the text box. Or, click the
Nerve List... button and select a nerve name from the list.
3. Click either the Left or Right check box. You can also select the side by
pressing the SIDE/SET key on the operation panel.
Measurement Procedure
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Check that there is no artifact.
2. Check that the stimulus intensity is set to 0 mA. If not, press the RESET key
on the operation panel.
4. Place the stimulating electrode on the stimulation site. Put the (−) side
proximal to the (+) side. The black marked side of the NM-420S surface
stimulation electrode is minus (−).
5. While checking the waveforms, turn the stimulation intensity dial on the
operation panel to gradually increase the stimulation intensity until the H-
wave disappears.
The response waveform elicited by electric stimulation changes as below.
low
(1) H-wave appears
Stimulus (2) The H-wave amplitude increases and M-wave appears.
Intensity (3) The H-wave amplitude decreases and the M-wave
amplitude increases.
(4) H-wave disappears and the M-wave amplitude reaches the maximum level.
high
6. Press the STOP key on the operation panel. The wave window changes to the
Gather window and the preset number of acquired waveforms are displayed
on the Gather window. (The number of waveforms is set at Preset Count on
the Acquisition page of the Set Conditions window.)
NOTE
Acquired waveforms are displayed on the Gather window only when
Store Raw Wave is set to Gather on the Acquisition page of the Set
Conditions window.
7. Copy the necessary waveforms to the Stop window and save them.
For the procedure, refer to the “Copying the Acquired Waveforms to the
Stop Window” section.
If the Gather window is not displayed, first display the Gather window with the
following procedure.
F1 F2 F3 F4 F5 F6 F7 F8 F9
2. Click the Gather Wave function button or press the F2 key on the keyboard.
Second Gather
Sens Wave
F1 F2 F9
The Gather window opens. The latest waveform is displayed at the top. 7-10
Copying Procedure
2. Click the Copy to Stop Screen function button or press the F2 key on the
keyboard. The selected waveform is copied to the Stop window.
The Gather window closes and the Stop window is displayed. The copied
waveforms are given names A1, B1, C1,... respectively. (This is the stage
name.)
• When Auto Store is set to ON (default setting: ON) on the Misc. page
of the Set Conditions window, the waveforms on the Stop window are
automatically temporarily saved in memory as one stage.
• When Auto Store is OFF, click the Store button on the Tool bar or press
the STORE key on the operation panel to save the waveform.
NOTE
If you delete the A1 waveform on the Stop window, all acquired
waveforms in the Gather window are deleted.
To close the Gather window, click the End Gather function button or press
the F8 key on the keyboard.
Deleting Waveforms
You can delete unnecessary waveforms by the following procedure.
1. Select the waveforms to delete on the Stop window with the ALL/INDV
key and SELECT ↓ ↑ key on the operation panel or mouse. The selected
waveforms have a different color. You can also select the waveform(s) by
clicking the All, Channel or Individual button on the Tool bar.
Individual button All button ALL/INDV key: Selects one or more waveforms. The All, Channel
or Individual button on the Tool bar indicates the
selected waveform(s).
The All button is down: All waveforms in one stage or site are
selected.
Channel button The Channel button is down: All waveforms in one channel are selected.
The Individual button is down: One individual waveform is selected.
SELECT ↓ ↑ key: Selects the stage if the ALL button is selected.
Selects the channel if the Channel button is
selected.
2. Press the ERASE key on the operation panel. The message confirming
deletion appears.
3. Check that the waveform selection is correct, and press the ERASE key. The
selected waveforms are deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
General............................................................................................................................................................ 7.11.2
Marking............................................................................................................................................................ 7.11.3
Auto Marking During Acquisition........................................................................................................... 7.11.3
Marking After Acquisition...................................................................................................................... 7.11.3
Setting One Mark at a Time........................................................................................................ 7.11.3
Setting Multiple Marks Automatically According to the Settings on the Labeling Page............... 7.11.5
Moving a Mark....................................................................................................................................... 7.11.5
Deleting Marks...................................................................................................................................... 7.11.6
Deleting an Individual Mark......................................................................................................... 7.11.6
Deleting All Marks....................................................................................................................... 7.11.6
H-reflex Measurement Table Window.............................................................................................................. 7.11.7
To Open the Measurement Table Window............................................................................................. 7.11.7
To Close the Measurement Table Window............................................................................................ 7.11.8
Measurement Table Window Type I Description.............................................................................................. 7.11.9
H-reflex Measurement Table Window.................................................................................................... 7.11.9 7-11
Annotating a Waveform............................................................................................................. 7.11.10
Entering the Stim. Site and Recording Site............................................................................... 7.11.10
Measurement Table Window Type II Description........................................................................................... 7.11.11
H-reflex Measurement Table Window.................................................................................................. 7.11.11
Annotating a Waveform............................................................................................................. 7.11.12
Intensity Amplitude Graph......................................................................................................... 7.11.12
Saving Data on Disk...................................................................................................................................... 7.11.13
Condition Table Window................................................................................................................................ 7.11.13
Annotating a Waveform................................................................................................................................. 7.11.14
General
• Marking
• Displaying the Measurement Table window
• Displaying the Condition Table window
• Annotating a waveform
7
Marking
Marks can be set automatically or manually. When the waveforms are marked,
waveform measurement starts automatically and the results can be displayed in
the H-reflex Measurement Table window.
The Latency window and Mark Label window appear, and a vertical cursor
appears on the wave area.
Latency window Mark Label window
Vertical cursor
2) On the wave area, click a waveform at the point you want to add the mark.
The selected mark appears on the clicked point. On the Mark Label window,
the mark label one line below is highlighted. Up to 12 marks can be set on
one waveform.
2) Turn the CURSOR dial to move the vertical cursor to the position to be
marked.
3) Click the mark label function button you want to set. The mark is set on the
cursor point. The latency (ms) at the point is displayed under the selected
mark label function button.
When the mark label you want to set is not on the function buttons
a) Click the function button to display the next function buttons.
b) Click the desired mark label on the Mark Label window or select the
mark label with the F2 (Select Prev. Mark) or F3 (Select Next Mark) key
on the keyboard. To select marks listed in the upper Mark Label window,
press the F2 key. To select marks listed in the lower Mark Label window,
press the F3 key. The selected mark label is highlighted.
c) Press the F1 (Add Mark) key on the keyboard. The mark is set at the
cursor point.
3. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
With the following procedure, marks are set according to the settings on the
Labeling page on the Set Conditions window after waveform acquisition.
2. On the Stop window, click the Marking function button, or press the F1 key
on the keyboard.
4. Click the Auto Marking function button or press the F4 key on the keyboard.
5. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
Moving a Mark
Do one of the following.
• Drag the mark to the desired position. You can move a mark with this
procedure on the Stop window without opening the Mark Label window.
• On the Mark Label window, select the mark label you want to move. Then, do
the “Setting One Mark at a Time” procedure. The selected mark moves to the
new position.
Deleting Marks
Deleting an Individual Mark
• When the mark label you want to delete is on the function button, you can
delete the mark by clicking the function button twice.
• Or,
1. On the Mark Label window, click the mark label to delete. The cursor
moves to the mark position.
2. Click the Add Mark function button or press the F1 key on the keyboard.
The mark is deleted.
Click the Delete Mark function button or the F6 key on the keyboard.
All marks on the selected waveform are deleted.
7
H-reflex Measurement Table Window
There are two types of H-reflex Measurement Table windows: Type I and Type II. Select either Type I or Type II on the
Labeling page on the Set Conditions window.
7-11
7
Measurement Table Window Type I Description
When Type I is selected on the Labeling page on the Set Conditions window, the
following window opens.
7-11
Annotating a Waveform
Whether or not to display the Annotation can be set on the Labeling page of the
Set Conditions window.
To annotate a waveform,
1. Click the No. you want to annotate.
You can also annotate a waveform on the Wave Annotation window. Refer to
“Annotating a Waveform” in this section.
1. Click the Stim. Site text box or Recording Site text box. The cursor moves to
the clicked box.
7
Measurement Table Window Type II Description
When Type II is selected on the Labeling page on the Set Conditions window, the
following windows open.
Sensitivity
7-11
Superimposed waveforms
(Baseline for waveforms
are adjusted into the same
level.) Intensity Amplitude graph
Annotating a Waveform
Whether or not to display the Annotation can be set on the Labeling page of the
Set Conditions window.
To annotate a waveform,
1. Click the No. you want to annotate.
You can also annotate a waveform on the Wave Annotation window. Refer to
“Annotating a Waveform” in this section.
The relation between the stimulation intensity and the amplitude is shown in the
graph. According to the marks on the waveform, H-wave and M-wave are plotted
with the following symbols.
M-wave
H-wave
The horizontal axis is the stimulation current, and the vertical axis is the
amplitude of H-wave and M-wave. The scales for both axis are adjusted
automatically and cannot be changed by users.
7
Saving Data on Disk
To save the measurement results and the waveforms on the hard disk, refer to the
MEB-9400 operator’s manual.
The Condition Table window displays the list of settings in which waveforms
were acquired.
To open the Condition Table window, click the Condition Table function button
on the Stop window.
7-11
F1 F2 F3 F4 F5 F6 F7 F8
Check settings for each waveform. Not all settings can be displayed on the
screen at the same time. To display other settings, use the scroll bar or the jump
buttons at the bottom of the screen.
To close the Condition Table window, click the button on the window.
Annotating a Waveform
You can annotate a measured waveform. The annotation display position and
annotation display ON/OFF on the Measurement Table window are set on the
Labeling page of the Set Conditions window. Refer to the MEB-9400 operator’s
manual for details.
1. Click the Marking function button or press the F1 key on the keyboard on
the Stop window.
F1 F2 F3 F4 F5 F6 F7 F8
3. Click the Wave Annotation function button or press the F5 key on the
keyboard to open the Wave Annotation window.
4. Click the or button of the Wave No. box on the Wave Annotation
window to display the waveform to be annotated in the box.
6. Click the OK button. The added annotation appears on both the selected
waveform on the Wave area and the Annotation column on the Measurement
Table window.
General............................................................................................................................................................ 7.12.2
Features................................................................................................................................................ 7.12.2
Preparation...................................................................................................................................................... 7.12.3
Electrode Attachment and Connection.................................................................................................. 7.12.3
Measurement................................................................................................................................................... 7.12.5
Selecting the Program and Changing Measurement Settings.............................................................. 7.12.5
Entering the Nerve Name...................................................................................................................... 7.12.5
Measurement Procedure....................................................................................................................... 7.12.6
Operations on the Stop Window............................................................................................................ 7.12.6
Deleting Waveforms.............................................................................................................................. 7.12.7
Saving Data on Disk.............................................................................................................................. 7.12.7
Printing Data......................................................................................................................................... 7.12.7
7-12
General
Features
This examination measures polysynaptic reflex (orbicularis oculi muscle reflex)
consisting of two components, the afferent of the trigeminal sensory branches
and the efferent of facial nerve motor fibers.
7
Preparation
2. Connect the NM-550B Ground Electrode (disk type) lead to the E jack.
4. Soak the felt pads of the surface stimulation electrode with water.
5. Use a piece of cotton moistened with alcohol to clean the patient’s skin
where the recording electrode will be attached. Rub the skin with an optimal
amount of Skinpure skin preparation gel to reduce artifact. Wipe the skin
thoroughly with dry gauze to remove any moisture and gel from the skin.
7-12
6. Attach the active and reference electrodes with paste to the optimal sites as
shown on the following page.
NOTE
• Do not short-circuit between the felt pads with water.
• Do not touch the stimulating electrode with the grounding electrode.
This causes stimulus artifact.
• When the stimulating electrode position is optimum, there is less
stimulating artifact. To find the optimum position of the stimulating
electrode, turn the stimulating electrode (+ anode) while the stimulating
electrode (− cathode) is fixed.
The active electrode is placed two-thirds of the distance from the outer edge of
the lower orbicular muscle of eye, and the reference electrode is placed on the
roots of nose on the left and right sides. Simultaneous recording is performed at
the part of bilateral symmetry (2 channel measurement). Stimulation is given to
each side.
SOMATO control box
Ground Electrode
Grounding electrode
7
Measurement
1. Select the NCS tab in the category tab on the Main Menu window. The
examination program buttons on the Main Menu window change.
2. Select the Blink program on the Main Menu window. The Blink examination
program window opens. The wave area and Nerve Table window are
displayed in the Blink examination program window.
3. Click the button on the Tool bar to display the Set Conditions window
to check or change the measurement settings. Refer to “Checking and
Changing Measurement Settings” in this section.
When using the Foot switch for stimulation, check the setting of Foot Switch
Right on the Misc. page of the Set Conditions window. The factory default
setting is Single. When the “R” foot switch is pressed, the single stimulation
is performed.
7-12
1. Click the Nerve Input function button or press the F7 key on the keyboard to
open the Nerve Table.
2. From the keyboard, type in the nerve name in the text box. Or, click the
Nerve List... button and select a nerve name from the list.
3. Click either the Left or Right check box. You can also select the side by
pressing the SIDE/SET key on the operation panel.
Measurement Procedure
NOTE
Ask the patient to not blink during recording.
1. Press the MONITOR key on the operation panel to monitor the raw
waveform from the recording electrode. Check that there is no artifact.
2. Check that the stimulus intensity is set to 0 mA. If not, press the Reset key
on the operation panel.
4. Place the stimulating electrode on the stimulation site and prepare for
the stimulation output. The black marked side of the NM-420S surface
stimulation electrode is minus (−).
5. Output a single stimulation.
When using the foot switch:
The setting of Foot Switch Right is set to Single in the factory default
setting. Press the “R” foot switch. The single stimulation is output and the
two channel response waveforms appear on the Stop window.
7. Repeat the measurement several times under the same settings. Take 3 to 5
second interval between measurements.
The acquired waveforms are superimposed on the temporarily saved
waveforms.
Deleting Waveforms 7
You can delete unnecessary waveforms by the following procedure.
1. Select the waveforms to delete on the Stop window with the ALL/INDV
key and SELECT ↓ ↑ key on the operation panel or mouse. The selected
waveforms have a different color. You can also select the waveform(s) by
clicking the All, Channel or Individual button on the Tool bar.
Individual button All button ALL/INDV key: Selects one or more waveforms. The All, Channel
or Individual button on the Tool bar indicates the
selected waveform(s).
The All button is down: All waveforms in one stage or site are
selected.
Channel button The Channel button is down: All waveforms in one channel are selected.
The Individual button is down: One individual waveform is selected.
SELECT ↓ ↑ key: Selects the stage if the ALL button is selected.
Selects the channel if the Channel button is
selected.
2. Press the ERASE key on the operation panel. The message confirming
deletion appears.
3. Check that the waveform selection is correct, and press the ERASE key. The 7-12
selected waveforms are deleted.
Printing Data
For printing, refer to the MEB-9400 operator’s manual.
General............................................................................................................................................................ 7.13.2
Marking............................................................................................................................................................ 7.13.3
Marking After Acquisition...................................................................................................................... 7.13.3
Moving a Mark....................................................................................................................................... 7.13.5
Deleting Marks...................................................................................................................................... 7.13.5
Deleting an Individual Mark......................................................................................................... 7.13.5
Deleting All Marks....................................................................................................................... 7.13.5
Measurement Table Windows.......................................................................................................................... 7.13.6
To Open the Measurement Table Window............................................................................................. 7.13.6
To Close the Measurement Table Window............................................................................................ 7.13.7
Measurement Table Window Type I Description.............................................................................................. 7.13.8
Evoked Measurement Table - Latency Window..................................................................................... 7.13.8
Entering Normative Data............................................................................................................. 7.13.8
Annotating a Waveform............................................................................................................... 7.13.9
Evoked Measurement Table - Mark Pair Window.................................................................................. 7.13.9
Selecting Measurement Items..................................................................................................... 7.13.9
Setting Pairs.............................................................................................................................. 7.13.10
Entering the Waveform Number................................................................................................ 7.13.10
Measurement Table Window Type II Description........................................................................................... 7.13.11 7-13
Blink Measurement Table Window....................................................................................................... 7.13.11
Marking for Calculating with the Blink Measurement Table Window ........................................ 7.13.12
Entering the Stage.................................................................................................................... 7.13.12
Rectified Average Wave............................................................................................................ 7.13.12
Displaying the Normative Range............................................................................................... 7.13.13
Creating the CSV File............................................................................................................... 7.13.13
Saving Data on Disk...................................................................................................................................... 7.13.15
Condition Table Window................................................................................................................................ 7.13.15
Annotating a Waveform................................................................................................................................. 7.13.16
General
• Marking
• Displaying the Measurement Table window
• Displaying the Condition Table window
• Annotating waveforms
7
Marking
When marks are set on the waveform, the latency and amplitude are calculated.
The calculation values are displayed on the Measurement Table window.
1. Click the Marking function button on the Stop window or press the F1
(Marking) key on the keyboard.
The Latency window and Mark Label window appear, and a vertical cursor
appears on the wave area.
Latency window Mark Label window
7-13
Vertical cursor
2) On the wave area, click a waveform at the point you want to add the mark.
The selected mark appears on the clicked point. On the Mark Label window,
the mark label one line below is highlighted. Up to 12 marks can be set on
one waveform.
Examination Guide MEB-9400 7.13.3
7. NERVE CONDUCTION STUDY
2) Turn the CURSOR dial to move the vertical cursor to the position to be
marked.
3) Click the function button of the mark label you want to set, or press the
function key on the keyboard corresponding to the mark label. The mark is
set on the cursor point. The latency (ms) at the point is displayed under the
selected mark label on the function button.
When the mark label you want to set is not on the function buttons
a) Click the function button or press the F9 key on the keyboard to
display the next function buttons.
b) Click the desired mark label on the Mark Label window or select the
mark label with the F2 (Select Prev. Mark) or F3 (Select Next Mark) key
on the keyboard. To select marks listed in the upper Mark Label window,
press the F2 key. To select marks listed in the lower Mark Label window,
press the F3 key. The selected mark label is highlighted.
c) Click the Add Mark function button or press the F1 (Add Mark) key on
the keyboard. The mark is set at the cursor point.
3. Click the End Marking function button or press the F8 (End Marking) key
on the keyboard.
Moving a Mark 7
Do one of the following.
• Drag the mark to the desired position. You can move a mark with this
procedure on the Stop window without opening the Mark Label window.
• On the Mark Label window, select the mark label you want to move. Then, do
the “Marking After Acquisition” procedure. The selected mark moves to the
new position.
Deleting Marks
Deleting an Individual Mark
• When the mark label you want to delete is on the function button, you can also
delete the mark by clicking the function button twice.
• Or,
1. On the Mark Label window, click the mark label to delete. The cursor moves
to the mark position.
7-13
2. Click the Add Mark function button or press the F1 key on the keyboard.
The mark is deleted.
Click the Delete Mark function button or press the F6 key on the keyboard.
All marks on the selected waveform are deleted.
There are two types of Evoked Measurement Table windows, Type I and Type
II. Select either Type I or Type II on the Labeling page on the Set Conditions
window.
When the waveforms are marked, waveform measurement starts and the results
can be displayed in the Measurement Table window.
Type I: Displays the Evoked Measurement Table - Latency window and the
Evoked Measurement Table - Mark Pair window.
Type II: The Blink Measurement Table window is displayed. On the Blink
Measurement Table window, waveforms which are averaged after
rectification are displayed.
<Type I> <Type II>
F1 F2 F3 F4 F5 F6 F7 F8
Or click the tool button or the Measurement Table function button again.
7-13
When Type I is selected on the Labeling page on the Set Conditions window, the
Evoked Measurement Table - Latency window and the Evoked Measurement
Table - Mark Pair window open.
2. Click the box to enter normative data and enter normative data with the
keyboard. The entered normative data is automatically saved in the memory
and applied to the next measurement.
Annotating a Waveform 7
To annotate a waveform,
1. Click the Annotation box to enter an annotation.
You can also annotate a waveform on the Wave Annotation window. Refer to the
“Annotating a Waveform” in this section.
Type box
No. box
7-13
Click the arrow button of the Type box. The selectable items are displayed in the
drop down list. Items are Interval, Amplitude, Area, Compare and (none).
Click an item in the list. The clicked item is displayed in the Type box.
Setting Pairs
To set a pair, enter two mark labels in the Mark Pair From box and the Mark Pair
To box. Up to 3 pairs can be set. The “-” is displayed between the boxes which
make a pair.
To enter a mark label in the box, click the button of the box. Twelve mark
labels are displayed in the drop down list. Click the desired mark label. The mark
label is displayed in the box.
R1 iR2a
C1
Calculation value
To enter the waveform number in the No. box, use the keyboard. When you
set two waveform numbers (only for Interval), enter waveform numbers in the
format “A1-B1”.
7
Measurement Table Window Type II Description
When Type II is selected on the Labeling page on the Set Conditions window, the
Blink Measurement Table window opens.
When the waveforms are marked, waveform measurement starts and the results
can be displayed in the Blink Measurement Table window.
Latency at R1
(latency at R)
(L-D-Lat. latency)
(latency at R)
(R-D-Lat. latency)
Stimulation side of
each stage
Calls up a CSV
file
4 waveforms which
are averaged after
rectification
To calculate waveforms with the Measurement table window type II, set marks
and enter stages in the Stim. Site column as shown in the example below.
L-D-Lat R-D-Lat
F1 L2
To enter more than one stage name in the column, put “-” between the first stage
and the last stage.
The figure on the left means that waveforms of stages A to E were acquired by
stimulating the left side, and waveforms of stages F to L by stimulating the right
side.
When the mouse is set to the bar or mark, the calculation value and the
normative value are displayed.
You can create a normative data file. Using a Microsoft Excel spreadsheet is
recommended. Bars in the Normative Range (+/−2SD) column are displayed
based on the normative data file.
7-13
1. Create a new file.
2. Enter text and values. Follow the formats below. Otherwise, the normative
range is not displayed.
- Type the title and comments in the first and second lines.
- Type the normative data in the third or later line.
- The data and sd are 3 digit significant figures, 2 digit decimals. Both the
data and sd must be entered.
Second line:
Comment
Third line: - The contents are the same as that of the Excel
Data and sd for setting
spreadsheet.
the normative range
- Words or data must be separated by commas only.
Never put a space or tab before or after a comma.
7
Saving Data on Disk
To save the measurement results and waveforms on the hard disk, refer to the
MEB-9400 operator’s manual.
The Condition Table window displays the list of settings in which waveforms
were acquired.
To open the Condition Table window, click the Condition Table function button
on the Stop window or press the F3 key on the keyboard.
7-13
Check settings for each waveform. Not all settings can be displayed on the
screen at the same time. To display other settings, use the scroll bar or the jump
buttons at the bottom of the screen.
To close the Condition Table window, click the button on the window.
Annotating a Waveform
You can annotate a measured waveform. The annotation display position and
annotation display ON/OFF on the Measurement Table window are set on the
Labeling page of the Set Conditions window. Refer to the MEB-9400 operator’s
manual for details.
1. Click the Marking function button or press the F1 key on the keyboard on
the Stop window.
4. Click the or button of the Wave No. box on the Wave Annotation
window to display the waveform to be annotated in the box.
6. Click the OK button. The added annotation appears on both the selected
waveform on the Wave area and the Annotation column on the Measurement
Table window.
Studies)
General............................................................................................................................................................ 7.14.2
Setup..................................................................................................................................................... 7.14.2
Operation Procedure....................................................................................................................................... 7.14.3
Settings for Each Examination Program................................................................................................ 7.14.3
Operation Procedure............................................................................................................................. 7.14.8
When the Examination Order is Set............................................................................................ 7.14.8
When the Examination Order is Not Set .................................................................................... 7.14.9
Starting Measurement ............................................................................................................. 7.14.10
Marking..................................................................................................................................... 7.14.11
Changing the Examination Type................................................................................................ 7.14.11
Changing the Examination Side................................................................................................ 7.14.11
Selecting Other Nerve............................................................................................................... 7.14.11
MCS/SCS Measurement Table Window.............................................................................................. 7.14.12
F-wave Measurement Table Window................................................................................................... 7.14.15
Saving Data on Disk............................................................................................................................ 7.14.17
Printing a Report................................................................................................................................. 7.14.17
7-14
General
Motor Nerve Conduction Study (MCS), Sensory Nerve Conduction Study (SCS)
and F-wave examination can be performed on the NCS examination program.
• A nerve name and its measurement settings can be registered. You can set the
measurement settings by specifying the registered nerve name.
• Up to 42 routine examinations can be set by selecting a nerve, side and
examination program as a routine protocol.
• The measurement result can be displayed on the measurement table window.
• The measurement result can be printed as an NCS report with the Excel.
• The normative data can be used for the MCS and SCS examination.
Setup
Refer to the MEB-9400 operator’s manual as necessary to prepare the equipment
and settings.
1. Select the NCS tab in the category tab on the Main Menu window. The
examination program buttons on the Main Menu window change.
2. Select the NCS program on the Main Menu window. The NCS examination
program window opens. The wave area and Nerve Table window are
displayed in the NCS examination program window.
3. Click the button on the Tool bar to display the Set Conditions window
to check or change the measurement settings. Refer to the MEB-9400
operator’s manual and “Settings for Each Examination Program” in this
section.
7
Operation Procedure
7-14
Item Description
Nerve Select the nerve name for the measurement. By specifying the nerve name,
the measurement settings are automatically selected.
MCS, SCS, F-wave Select the measurement program for the selected nerve.
Channels selection list box Select the channel to acquire the waveforms.
Electrode (−) Select a pair of electrode names to acquire the waveforms.
Electrode (+)
Sens Select the amplifier sensitivity.
Hi-cut Select the high-cut filter frequency of the amplifier.
Lo-cut Select the low-cut filter frequency of the amplifier.
Analysis (/div) Select the analysis time for horizontal length (scale/division) for trigger
sweep and average.
Count (SCS, F-wave only) SCS: Select the number of waveforms to average.
F-wave: Select the number of waveforms to acquire.
Reject (SCS only) Select the amplitude threshold (div) to select the acquired waveforms for
averaging. When the amplitude of the waveform exceeds the upper or lower
threshold, it is not included in averaging.
Table (MCS, SCS only) For the MCS and SCS examination, the measurement table can be edited.
Edit Table button: Opens the Edit Table to enter the site, segment, distance
and normative value (NR/SD) for each nerve.
Stim. Site (F-wave only) Enter the stimulation site for the nerve.
Recording site (F-wave only) Enter the recording site for the nerve.
ND:F-Lat,(F-wave only) Enter the normative F-wave latency for the nerve.
ND:FWCV (F-wave only) Enter the normative F-wave conduction velocity for the nerve.
Cascade Display in Check this box to display the cascaded waveforms on the Monitor/Sweep
(F-wave only) window and/or Stop window. The number of waveforms displayed on one
screen can be set in the No. of Disp list box.
Histogram (F-wave only) Select the maximum horizontal and vertical scale to display the F-wave
histogram.
Dual Sens. (F-wave only) 2nd Sens.: Select the waveforms second period amplifier sensitivity.
Smoothing points: Select the number of sampling points.
Labeling Page 7
Select the marks settings set on the waveform.
Item Description
Auto Marking Select whether or not to automatically set marks on the
waveform.
Auto Magnify Select the waveform display sensitivity while marking.
(MCS, SCS only)
Auto Advance Check this box to automatically select the next
(MCS, SCS only) waveform in Fast Marking mode. Refer to “Operation
Procedure - Marking”.
Measure Table Type Select the type of the Measurement Table window.
(MCS, SCS only)
F-wave Measure Mode Select the type of the Measurement Table window. 7-14
(F-wave only)
For other settings, refer to the “Labeling Page” in Section 5 of the MEB-9400
operator’s manual.
NCS Page
Select the method to set the mark on the waveform.
Item Description
Auto Mark Detect Select the algorithm to detect a mark point.
Peak-Bottom: Detects the positive peak and
negative peak as a mark point after
detecting the rising point.
Window: Conventional mark detection method
Table Sequence Opens the Table Sequence dialog box to select the
examination order for the routine examination.
Single Stim. Check this box to apply single stimulation when
the STIM/SWEEP button is pressed. After the
stimulation, the screen changes to the STOP window.
Temp or Normative Data Select to display the normative data or temperature
for each segment in the NR (+/−2SD) column.
Stim reset timing Select the timing to set the stimulation intensity to
zero.
Reset button: when the RESET key is pressed.
Store Timing: when the waveform is saved (The
STORE key is pressed).
Trace Fixed Mode Check this box when the examination order is not
determined. You can set the examination site in the
NCS examination window before measurement.
7-14
Operation Procedure
The operation procedure is different depending on whether the examination order
is set in the Test Sequence dialog box or not. The examination order is preset in
the factory.
When the NCS examination program starts, the NCS examination program
window opens. The wave area and Nerve Table window are displayed in the
NCS examination program window.
• To do the next examination, select the Next Test function button or press the
MUSCLE/TEST ↓ ↑ key on the operation panel or select the next examination
in the list with the mouse.
• To measure the nerve or examination side that is not set in the Test Sequence
dialog box,
1) Select the Exit Sequence function button with the mouse. The Nerve Table
window opens.
2) Select the nerve name and examination side, and click the OK button.
• To return to the test sequence, click the Return Sequence function button with
the mouse.
7.14.8 Examination Guide MEB-9400
7-14. NCS (NERVE CONDUCTION STUDIES)
1. Select the nerve name with the function button. When the nerve name is
not displayed on the function button, select the Other Nerve button with the
mouse to display the nerve name.
2. Select the F3 function button to select the examination side. The currently
selected side is displayed on the F3 function button.
7-14
Starting Measurement
For each operation procedure, refer to the “Operation Procedure” section for
each examination.
Marking 7
Marks can be set on the waveform automatically or manually. For marking, refer
to “Common Operation” and “Settings for Each Examination Program - Mark
Page” in this section.
MCS/SCS: Up to 4 marks for each examination
F-wave: Up to 6 marks for each examination
1. Press the LAT/AMP/TRIG key on the operation panel to enter the Fast
Marking mode.
3. Turn the CURSOR dial to move the vertical cursor to the rising point of the
waveform.
4. Press the A/B key on the operation panel. The Latency 1, Latency 2,
Amplitude 1 and/or Amplitude 2 marks are automatically set on the
waveform.
5. Repeat steps 2 to 4 for the waveforms to be marked. When the Auto Mark
Advance check box (MCS and SCS measurement only) is checked in the
Set Conditions window - Labeling page, the next waveform is automatically
selected. 7-14
Paired waveforms
The number is waveform order
on the wave window. The site
can be selected from the list
box. To open the list box, click
the segment.
7-14
Use Trace Fixed Mode when the examination order is not determined. When the Trace Fix Mode check
box is checked, stages A1 to F1 are displayed on the Measurement Table window. Before measurement,
you can select the stage where the waveform is saved with the ALL/INDV key and SELECT ↓ ↑ key on the
operation panel or mouse operation. The selected stage is surrounded by a rectangle. In the above example,
the waveform is acquired in stage B1. After waveform acquisition, when another stage is selected, the
acquired waveform is automatically saved. To acquire another waveform to in a stage where a waveform
has already been saved, delete the saved waveform with the Erase button on the tool bar.
ND:
Normative data entered
in the Set Conditions
window - Nerve page.
7-14
5. Set the cursor on the rising point of the M- 6. Set the cursor on the rising point of the F-
wave. The M-wave latency from the trigger wave. The F-wave latency from the trigger
point to the cursor point is calculated. point to the cursor point is calculated.
ND:
Normative data entered in Histogram:
the Set Conditions window Latency of a waveform is
- Nerve page. shown as a bar
Printing a Report
When Microsoft Excel is installed in the system, the NCS measurement results
and waveforms can be printed as a report. For the report settings, refer to
“Printing a Report” in Section 8 of the MEB-9400 operator’s manual.
Report Example
7-14
When an error message is displayed on the screen or trouble occurs, take quick
action referring to the troubleshooting table. If there is any damage or the system
is suspected to be faulty, turn the power off, attach an “Unusable” or “Repair
request” label to the system and contact your Nihon Kohden (NK) representative.
Waveform Acquisition
AC Interference
50 Hz AC interference
NOTE
• When the interference is small, click the AC Interference button on the
Tool bar to check whether it is caused by AC interference.
• When using electric stimulation, set “AC Filter” to “Off” because the AC
filter distorts acquired waveforms.
• The default filter frequency is 60 Hz. When using the system in a region
with 50 Hz AC power, set it to 50 Hz.
Spike Noise
Noise from a fluorescent lamp
Other Noise
EMG interference
When the cause of the artifact is not found, change the stimulation rate in 0.1 Hz steps and
average the waveforms. This is useful to average out the artifact which appears periodically and
leave only evoked waveforms.
8