8 Disassembly

8.1 Overview
This chapter describes how to disassemble the TM80 and the BP10.

Before disassembling the TM80 and the BP10, make the following preparations:

 Ensure that the TM80or the BP10is not used for monitoring a patient and
has been disconnected from the patient.

 Disassemble the TM80 or the BP10 on the workbench of the equipment


maintenance room.

 The personnel who will disassemble the TM80 or the BP10 must be
qualified for maintenance of medical equipment or have received
trainingsgiven by Mindray or have been authorized by Mindray.

 In the disassembly process, the personnel disassembling the TM80or the


BP10 must take protective measures related toElectro-Static discharge
(ESD ).

NOTE

 During re-installation after disassembling the device, the sealing strip


must be installed correctly. Improper installation of the sealing strip
may degrade the waterproof performance of the device.

TM80 Telemetry Monitor Service Manual 8-1


8.2 Disassembling the TM80
1. Remove the label on the rear housing of the TM80 and remove the battery.
Unscrew the screws on the front and rear housing by using a Phillips
screwdriver.

One M1.6×13.3 screw

Two M1.6×3.5 screws

Five M1.6×2.4 screws

Two M1.6×3.5 screws

2. Holdthe display and let it face your left palm, open the front and rear housings,
and remove the connector of the Wi-Fi cable.

CAUTION

 Do not pull apart the Wi-Fi cable of the TM80.

8-2 TM80 Telemetry Monitor Service Manual