ATRT-03 PC Software

Rev 11.11 (09/05/03)


1. Correct problem of saving extra tests after running test using test plan. (Rev
11.9)
2. Correct print file name printout on test report (after save shot). (Rev 11.10)
3. Correct multiple file delete problem under file utilities. (Rev 11.10).
4. Correct PC print out due to change font size. (Rev 11.11)

Rev 11.12 (07/07/04)


1. Correct vector diagram printout when export file to EXCEL.
2. Change remote ITF message.

Rev 11.13 (09/05/04)


1. Add 24 more vector diagrams to program.

Rev 11.14 (10/10/04)


1. Add cut and paste commands to create test plan.
To cut and paste H or X voltage column for transformer test plan:
- Select cell and enter voltage
- Use TAB key to advance to next cell
- Click on “File”
- Click on “Utilities”
- Click on “Duplicate H & X column”

Rev 11.15 (10/10/04)


1. Allow file names to have “period”.
2. Allow starting the program by double click on the test report or test plan.
3. Add 7 new transformer type: YNzn1, YNzn7, YNzn11, Dy3, Dyn3, Dy9, Dyn9

Rev 11.16 (10/11/05)


1. Correct LTC and Voltage Regulator table generation problem.

Rev 11.17 (05/20/06)

1. Add note pad to test report


2. Allow the user bring up program by double click on test plan or test report.

Rev 11.18 (08/08/06)


1. Found bug in test report. Could not scroll through test report!
2. Found bug in retrieving transformer test report from ATRT-03. Transformer
test report tap header was incorrect.

Rev 11.19 (11/20/06)


1. Found bug in voltage regulator test plan
2. Expand COM port to COM 7
3. Allow user to start program from test plan.
Rev 12.00 (10/17/07)
1. Redesign ATRT-03 PC software. Initial release 12.0
2. Add vector diagram to PC print out

Rev 12.1 (12/02/07)


1. Add export test report into EXCEL directly.
2. Update ratio calculation after change transformer type in test plan.

Rev 12.2 (06/10/08)


1. Correct bug in additional test under LTC.

Rev 12.3 (06/16/08)


1. Correct bug in voltage regulator test plan.

Rev 12.4 (09/16/08)


1. Correct bug in additional test plan for 3-phase LTC
2. Lock out voltage column in LTC and VLT test plans
3. Increase shot path length display.

Rev 12.5 (11/29/08)


1. Allow user to change LTC test voltages after creating LTC table.
2. Found bug in downloading 33 position test report. The program reads only 20
tests in the report.
3. Revise “Export” command.
4. Allow users to start the program by clicking on the test file or test plan.
5. High-light failed test on report.
6. Verify setup file on Windows Vista

Rev 12.6 (12/06/08)


1. When open a test record from the test file. All LTC taps are shown. Extra test
is not.
2. Create C:/vanguard/ttra directory.
3. Create Test Plans and Test Records sub-directories.
4. Executable file is ttra.exe
5. Export Test Records to text file or EXCEL files
6. Export Test Plan to text file.
7. High light “Fail” test.

Rev 12.7 (08/15/09)


1. Found bug in test plan for YNd5 Phase A display (H1-H1/X3-X1). Correct to
H1-H0/X3-X1.
Rev 12.8 (02/18/10)
1. Correct “Deviation Graph”
2. Allow users to insert company logo. The Logo file is a 1000 x 100 bit map file.
The default Vanguard file is located in the TTRA folder.
Users can select not to print the company logo by clearing the logo file path.

Rev 12.9 (10/14/10)


1. Correct test number on print out
2. Allow 10 tests per page
3. Allow note pad to be printed on last page with logo

Rev 13.0 (9/14/11)


1. Add Auto-Export to pdf, XCEL, HML

Rev 13.1 (11/22/11)


1. Allow the user to start the program by double clicking on the test file

Rev 13.2 (12/30/11)


1. Found bug in “Application Setting”. Program does not save settings (COM
port, Shot paths)

Rev 13.3 (2/06/12)


1. Found bug in “Create Test Plan”. Program tried to generate the pdf report.
2. Program crashed when tried to run and old test plan.

Rev 13.4 (03/01/12)


1. Allow user to create LTC/VLTG Reg test plan with any odd or even taps
2. Fix LTC test plan with extra test bug

Rev 13.5 (03/19/12)


1. Allow users to add more than one neutral position on LTC and Voltage
Regulator test plans

Rev 13.6 (06/04/12)


1. Remember COM port setting

Rev 13.7 (10/18/12)


1. Found bug in displaying 33 positions turns ratio.

Rev 13.8 (09/05/13)


1. Test plan print out show only 17 positions on a 38 position plan.