You are on page 1of 9

Available online at www.sciencedirect.

com

The overshadowed role of electron ionization–mass


spectrometry in analytical biotechnology
Deyber Arley Vargas Medina 1,2, ]]
]]]]]]
]]

Edvaldo Vasconcelos Soares Maciel 1,2,


Natalia Gabrielly Pereira dos Santos 1,2 and
Fernando Mauro Lancas 1

Target and untargeted analysis of several compounds are on their mass-to-charge ratios. In EI–MS, gas-phase
crucial methods in important areas such as omics sciences. analyte molecules are bombarded with high-energy
Gas chromatography coupled to mass spectrometry (GC–MS) electrons within a vacuum chamber [1•].
is widely used for volatile and thermally stable compounds. In
this case, the electron ionization technique (EI) is preferable as it The EI source comprises a tungsten filament, an ion
produces highly fragmented and reproducible spectra generation compartment, and a series of lenses for ion
comparable to spectral libraries. However, only a fraction of manipulation. The filament generates an electron beam
target compounds is analyzable by GC without chemical that interacts with the sample, following a spiral path
derivatization. Therefore, liquid chromatography (LC) coupled guided by a cathode and small permanent magnets. Upon
with MS is the most used technique. Contrary to EI, collision with the electron beam, the gas-phase molecules
electrospray ionization does not produce reproducible spectra. absorb energy and undergo ionization by losing an elec­
That is why researchers have been working on interfaces tron, resulting in the formation of odd-electron ca­
between LC and EI–MS to bridge the gap between those tion radicals (M+•). These species are inherently unstable
techniques. This short review will discuss advancements, and readily undergo consecutive fragmentation through
applications, and perspectives on biotechnological analysis. intramolecular reactions, generating more stable species.
The positively charged species generated are subse­
Addresses quently ejected from the ionization source by applying a
1
Laboratory of Chromatography, Institute of Chemistry at Sao Carlos, positive potential to a repeller. Within the mass analyzer,
University of Sao Paulo, P.O Box 780, 13566590 Sao Carlos, Brazil
2 ions are separated based on their mass-to-charge ratio (m/
Clemens Schöpf Institute, Department of Chemistry, Technical
University of Darmstadt, 64287 Darmstadt, Germany z) and recorded in the EI mass spectra [2].

Corresponding author: Lancas, Fernando Mauro (flancas@iqsc.usp.br) EI–MS operates under vacuum conditions, with only ap­
proximately 1 in 10 000 molecules in the ionization source
being ionized. Fragmentation occurs exclusively through
Current Opinion in Biotechnology 2023, 82:102965 intramolecular reactions, leaving other sample constituents
This review comes from a themed issue on Analytical unaffected. Each EI–MS spectrum serves as a unique
Biotechnology fingerprint for a specific molecule, and the multitude of
Edited by Alexander Gruenberger, Janina Stephanie Bahnemann recorded fragment ions provides valuable structural in­
and Christian Dusny formation for molecular identification and elucidation.
Besides, EI spectra are highly reproducible, so numerous
EI spectra libraries have been established, with some in­
tegrated into MS software packages for automated spectra
Available online xxxx comparison and annotation of unknown analytes. The
https://doi.org/10.1016/j.copbio.2023.102965 capability for analyte identification remains the primary
0958–1669/© 2023 Elsevier Ltd. All rights reserved. advantage of EI–MS over other MS ionization sources.

The unparalleled role of electron ionization in


Electron ionization–mass spectrometry: a volatilomics
brief description During the last decades, tremendous progress in bio­
Electron ionization–mass spectrometry (EI–MS) is a technology and related fields has been made. A suc­
widely employed technique for the analysis of molecular cessful example is human genome sequencing and the
structure and composition in organic compounds. It in­ consequent appearance of important areas such as
volves the vaporization and ionization of sample mole­ genomics, proteomics, lipidomics, and so on [3••].
cules, followed by their separation and detection based Omics techniques aim to systematically analyze large

www.sciencedirect.com Current Opinion in Biotechnology 2023, 82:102965


2 Analytical Biotechnology

Abbreviations GC Gas chromatography


LC Liquid chromatography
API Atmospheric pressure ionization LEI Liquid electron ionization
DIP Direct inlet probe MS Mass spectrometry
EI Electron impact or ionization m/z Mass-to-charge ratio
E-LEI-MS Extractive-Liquid Sampling Electron MOI Microfluidic open interface
Ionization-Mass Spectrometry SPME Solid-phase microextraction
ESI Electrospray ionization SIDS Sudden infant death syndrome
FIA Flow injection analysis VOCs Volatile organic compounds

cohorts of biological-related molecules to get informa­ specific ionic fragments from each target compound sepa­
tion about an organism’s state, function, and dynamics rately [10]. GC–MS/MS, utilizing high-resolution detectors
[4••]. Volatilomics studies the emission of volatile or­ such as time-of-flight and Orbitrap, is an ideal technique
ganic compounds (VOCs) produced by metabolic path­ for accurate mass determination across the entire m/z
ways, easily found in noninvasive matrices (e.g. saliva, (mass-to-charge ratio) range [11]. Therefore, GC–MS and
urine, exhaled breath, and flowers) [5]. VOCs are en­ GC–MS/MS with low-resolution (e.g. quadrupoles or ion
dogenous or exogenous compounds (< 1500 Da). No­ traps) or high-resolution detectors have been successfully
tably, the analysis of VOCs can aid in the detection of used in bioanalytical-related applications, including medi­
diseases in humans, animals, and plants, as well as the cine [8], natural products [12], drug discovery [13], foo­
presence of microorganisms and other pathogenic agents domics [14], and biotechnology [15–17].
[5,6]. Approximately 2746 VOCs have been identified in
humans so far [7]. Regardless of the technique, un­ Summarizing, GC–EI–MS plays a crucial role in various
targeted or nontargeted analyses are challenging, often biotechnological applications, currently contributing i)
demanding several steps, including experimental design, the diagnosis cancers, respiratory disorders, and in­
data acquisition, and processing [6]. Because of the un­ fectious diseases, ii) drug development based on volatile
ique characteristics of the compounds, their variable compounds released by microorganisms, plants, or other
concentration, and the complexity of the matrices where biological sources, iii) monitoring of plant health, asses­
they are found, powerful analytical techniques are re­ sing the ripeness or quality of fruits, vegetables, and
quired for their determination [6]. other agricultural products, iv) bioprocess optimization
by monitoring VOCs as markers of the efficiency, per­
Gas chromatography coupled to mass spectrometry formance, and productivity of fermentation, microbial
(GC–MS) is widely used, especially in the case of growth, and other biotechnological processes, and v) in
VOC analysis. This technique has several advantages, microbiome studies, among many other areas [3].
such as high reproducibility, efficiency, and sensitivity
below ppt levels [6]. Concerning MS, EI is the most Given recent advancements in GC, including compre­
reliable, efficient, and used ionization technique in hensive GCxGC, and the increasing affordability of high-
combination with GC [8]. The robust ionization process resolution mass spectrometers, it is widely acknowledged
of EI (70 eV) produces highly fragmented and re­ that GC–MS(MS) satisfactorily meets the requirements
producible spectra, which, combined with its well-un­ for analyzing VOCs. On the other hand, it is widely re­
derstood fragmentation rules, allows comparison against cognized that GC is suitable for the straightforward ana­
available spectral libraries [8]. EI provides each analyte lysis of volatile and thermally stable compounds.
with a spectrum that resembles a fingerprint, which is However, for compounds that are not inherently volatile
reproducible across different MS instruments. Ad­ or thermally stable, chemical derivatization is often ne­
ditionally, matrix effects are rarely observed, as the cessary to enhance their thermal stability and volatility,
highly energetic process extinguishes most molecular thus enabling their analysis by GC [18]. It should be
interactions [9]. The reproducible fragmentation pattern noted that derivatization procedures are typically labor­
combined with the GC retention time helps identify ious, time-consuming, and susceptible to sample losses.
known and unknown compounds. Although the re­ Moreover, it is estimated that only a narrow fraction of
cognized efficiency of EI, omics studies often generate a compounds of current interest are analyzable by GC–MS
large set of data containing hundreds of compounds, without any previous procedure (Figure 1) [19••]. Of
making the MS spectra interpretation difficult [10]. course, this is a problem for many researchers.

To overcome this downside, gas chromatography–tandem For this reason, the main technique currently chosen for
mass spectrometry (GC–MS/MS) can be used to analyze nonvolatile compound analysis is liquid chromatography

Current Opinion in Biotechnology 2023, 82:102965 www.sciencedirect.com


Electron impact mass spectrometry in biotechnology Vargas Medina et al. 3

Figure 1

Current Opinion in Biotechnology

Recommended chromatography-based methods according to the classes of compounds and polarity.


Reprinted with permission from [51]. Copyright (2020) Elsevier.

coupled to MS using electrospray ionization (LC–ESI–MS). ambient of EI–MS in an efficient way. Two alternatives
LC has become a more versatile technique than GC, with could be explored for that purpose: i) the coupling of
similar or even better sensitivity and efficiency [19••]. EI–MS to LC and ii) the development of probes for
Moreover, LC can analyze target compounds in a broader directly introducing samples into the source.
range of molecular weights, with different polarities and
without thermal stability or volatility issues (Figure 1) [17]. The coupling of electron ionization–mass spectrometry
However, one drawback is the incapability of comparing with nano-liquid chromatography
ESI-generated spectra against spectral libraries due to the Although in the past, the possibility to obtain EI–MS
soft-ionization process and nonreproducible data [20]. spectra by coupling with LC was strongly limited by the
Therefore, the optimal approach involves coupling EI with considerable amount of used mobile phase (mL/min), this
GC and LC, enabling the analysis of GC- and LC-amenable is a problem almost completely superseded by the modern
compounds using EI. During the last 22 years, several works nano-LC [19,22••]. At nL/min flow rates, the impact of the
focused on the topic have been published, reinforcing its mobile phase over the EI–MS process drastically di­
importance. Here, we bring to the journal audience an minishes, and some standard EI–MS spectra can be ob­
overview of how much progress has been made, high­ tained from the liquid effluent. At the moment, three nano-
lighting interesting applications and what to expect soon. LC–EI–MS-interfacing strategies have been introduced: i)
Can EI finally reach the same popularity as its counter­ Direct EI, ii) liquid electron ionization (LEI), and iii) Cold
part ESI? EI. Although still with limitations, those interfaces are ex­
tending the range of the applicability of nano-LC–EI–MS
The potential of electron ionization–mass toward heavier and more polar compounds.
spectrometry for the analysis of nonvolatile
compounds and analytical biotechnology Direct EI relies on introducing the nano-LC effluent
Before the emergence of modern ESI sources, EI–MS into the ionization source through a fine capillary tube
was widely used for peptide sequencing, proteomic, and (20–25-µm i.d.). This strategy can provide reproducible
genomic studies [21]. Through derivatization, peptides mass spectra for several types of molecules and provides
were converted into volatile polyaminoalcohols analyz­ the first demonstration of the absence of matrix effects
able by GC–MS. Although it was a confident strategy in the analysis of biological samples via nano-
until before the emergence of modern atmospheric LC–EI–MS [22••]. Direct EI has facilitated important
pressure ionization sources, the process was tedious, la­ achievements, such as developing portable nano-
borious, and time-consuming. However, is it possible to LC–EI–MS instruments for forensic analysis [23] and
exploit the structural elucidation advantages of EI–MS determining free fatty acid profiles without needing es­
in analyzing nonvolatile or large molecules without de­ terification stages [24]. This previous study reveals par­
rivatization stages? That achievement depends on the ticular importance since it sets the bases for future
possibility of transferring the heavier analytes from a lipidomic studies based on nano-LC–EI–MS. Although
condensate phase (liquid or solid) to the vacuum Direct EI is still quite limited, this interfacing strategy

www.sciencedirect.com Current Opinion in Biotechnology 2023, 82:102965


4 Analytical Biotechnology

could bring important insights exploiting narrow-bore The potential of modern nano-liquid
(< 25) open tubular columns (OT) [25•]. OTs can be chromatography–electron ionization–mass
directly coupled to the EI source without additional spectrometry in analytical biotechnology
interfacing devices, avoiding extra column void volume Although significant advances have been made in nano-
and improving separation/detection efficiency [26•]. LC–EI–MS in recent years, the literature on analytical
Recently, a series of OT columns with 2.0-µm i.d has biotechnology primarily relies on LC–EI–MS with lim­
demonstrated outstanding efficiency and peak capacity ited reports. This scarcity can be attributed to the un­
for digested protein analysis [27,28]. The future cou­ derdeveloped nature of the technique and its restricted
pling of those columns in nano-LC–EI–MS could bring accessibility, with only a few research groups worldwide
essential contributions to the development of the tech­ actively engaged in this field.
nique and enlarge its applicability range by reduction of
the mobile- phase flow rate until the pL/min order. However, some interesting biotechnological-related
applications of nano-LC–EI–MS have been explored.
LEI is a more advanced nano-LC–EI–MS-interfacing For instance, nano-LC–EI–MS has been tested for the
technology than Direct EI and provides a more ex­ determination of pharmaceutical drugs used in treating
tended applicability range, less noised spectra, and im­ psychiatric disorders and pain management [36]. No­
proved detection limits [29]. Those improvements result table publications include the determination of ben­
from incorporating a tailored vaporization chamber that zodiazepines in alcoholic beverages [37], and the study
efficiently separates the stages of mobile-phase eva­ of trans-cinnamaldehyde absorption in dermal tissues
poration and analytes’ ionization. The merits of LEI using a nano-LC–EI–MS with a Direct-EI interface
have been mainly demonstrated in the targeted analysis [38]. Additionally, the analysis of potential genotoxic
of small molecules of diverse polarity, such as pesticides, impurities in active ingredients for drug formulation
polycyclic aromatic hydrocarbons, hormones, and phe­ has demonstrated the time-saving benefits of nano-
nols [30]. However, this interface also could allow the LC–EI–MS by avoiding extensive sample preparation
performance of untargeted metabolic studies via nano- and derivatization processes [39]. In lipidomics-related
LC–EI–MS. For example, during the study of sudden studies, nano-LC–EI–MS has shown promise with
infant death syndrome (SIDS), a newborn victim brain studies profiling free fatty acids in mussels [24] and
sample was analyzed using LEI in full-scan mode. With determining nonesterified free fatty acids (FFAs) in
the aid of deconvolution software, the technique allowed human plasma as a health indicator in diabetic patients
the easy and unequivocal identification of markers un­ [40]. Another area of interest where LC–EI–MS has
derneath the matrix peak [31]. been applied is in the analysis of dietary supplements.
Studies have quantified soy-content isoflavones in
Furthermore, Cold EI is an interface developed for the dietary supplements and analyzed urine samples from
obtention of EI spectra with enhanced molecular ions. patients treated with these products [41]. LC–EI–MS
This interface relies on the ionization of cold molecules in has also been employed for the quantification of
supersonic beams and works with both GC and micro-/ naturally occurring ephedrine alkaloids in standard
nano-LC [32]. Cold EI is also practically free of matrix reference materials [42] and the analysis of poly­
effects and provides superior identification capabilities phenols and caffeine in Camellia sinensis extracts
than standard EI. The assignment of molecular formulas certified as NIST reference materials for dietary sup­
is more confident from the enhanced molecular ion and plement quality control [43]. Overall, these applica­
its isotopic pattern [33]. LC–Cold EI–MS has demon­ tions demonstrate the potential of LC–EI–MS in
strated applicability to a more extended range of polarity analytical biotechnology, highlighting its advantages
than LC–ESI–MS and superior separation/detection ef­ and providing insights for further development in
ficiency — providing narrower chromatographic peaks — the field.
than GC–EI–MS [34]. Currently, Cold EI has made
possible the coexistence of GC–MS and LC–MS in a Direct input techniques in electron
single instrument [33]. A recent instrument introduced by ionization–mass spectrometry
the developers of the Cold EI can easily switch between Bio-solid-phase microextractio–electron
GC–EI–MS and LC–EI–MS via software and without ionization–mass spectrometry
instrumental modification. That unified instrument even Advances in the area of detection, mainly in MS, have
can record GC and LC effluents in the same chromato­ allowed the direct coupling of sample preparation tech­
gram. Although still enduring important limitations, the niques such as solid-phase microextraction (SPME) with
future improvement of unified chromatography with Cold MS, making it a tool for rapid screening and quantifi­
EI–MS could allow the complementary GC–MS/LC–MS cation of analytes. Considering that SPME requires low
information for more broad exploration of unknown volumes of solvents to desorb the fiber, it ends up being
samples. Figure 2 summarizes some examples of im­ a great option for direct coupling with EI–MS. Cappiello
portant applications of nano-LC–EI–MS. and collaborators coupled the SPME fibers to the

Current Opinion in Biotechnology 2023, 82:102965 www.sciencedirect.com


Electron impact mass spectrometry in biotechnology Vargas Medina et al. 5

Figure 2

Current Opinion in Biotechnology

Some examples of recent applications of nano-LC–EI–MS. (a) Comparative scheme of the mass spectra of C18 fatty acid obtained by nano-LC–EI–MS
with Direct-EI interface, and the spectra in the NIST database [22•]. (b) Cromatogram of the extract obtained from a SIDS victim’s brain sample
analyzed via nano-LC–EI–MS with an LEI interface [28•]. (c) Cromatogram of unified GC–EI–MS and LC–EI–MS obtained in a unified instrument
equipped with a Cold-EI interface [35].

www.sciencedirect.com Current Opinion in Biotechnology 2023, 82:102965


6 Analytical Biotechnology

Figure 3

Current Opinion in Biotechnology

Direct introduction techniques to EI–MS. (a) SPME device used for matrix analyte extraction. Soon after, use of the MOI to desorb the analytes and
introduce them into the LEI–MS. (b) FIA device for direct introduction of the sample into the Direct EI–MS. (c) DIP probe used for direct introduction of
solid samples into the EI ionization chamber for subsequent MS detection.
(a) Reprinted with permission from [44]. Copyright (2020) American Chemical Society. (b) Adapted from [45].

Current Opinion in Biotechnology 2023, 82:102965 www.sciencedirect.com


Electron impact mass spectrometry in biotechnology Vargas Medina et al. 7

LEI–MS through a microfluidic open interface (MOI) ionization chamber, where the sample is vaporized and
(Figure 3a) [44]. MOI is a desorption chamber into ionized by electron impacts, and then, the generated ions
which the analytes extracted on the SPME fiber are are detected in the MS [49]. This technique is very fast
desorbed by a solvent at atmospheric pressure and and can be used for the direct obtention of EI–MS spectra
subsequently pumped to an LEI–MS interface. of nonvolatile solid and liquid samples. For instance,
DIP–MS has been successfully employed as a rapid and
Flow injection analysis direct electron ionization–mass straightforward tool for identifying heavy and ultraheavy
spectrometry crude oils, asphaltenes, and tar sands in field applications,
Samples can be directly introduced into an EI–MS source eliminating the requirement for prior separation or treat­
via flow injection analysis (FIA). Analysis time becomes ment. By programming the probe temperature ramp,
even faster, bringing highly reliable information about the DIP–MS enabled the detection of individual components
nature of the analyte, considering the accuracy of the EI within the mixture up to masses of 950 m/z [50]. Despite
spectra. Injection via FIA consists of boosting the sample its promising potential, DIP has not been extensively
through a valve, and a carrier liquid constantly transports explored in the context of biotechnological analysis, and
the sample. With the help of a peristaltic pump, the there are few or no reports describing its applicability in
sample can be driven to Direct EI–MS (Figure 3b) [45]. this field found in the literature. Conducting additional
Analytes together with the mobile phase are converted studies to explore the use of DIP for obtaining EI–MS
into an aerosol under high temperature and vacuum. In­ spectra of proteins, other biomolecules and even viruses
side the ionization chamber, there is a hot vaporization or bacteria could yield valuable insights into the field of
surface, which allows the vaporization of the sample and biotechnological analysis.
subsequent ionization with electrons.
Techniques for direct sample introduction into the
Extractive liquid sampling–electron ionization–mass EI–MS source are excellent alternatives to conventional
spectrometry methods, such as GC–MS, for quick and easy identifi­
Those abovementioned Direct EI–MS sample in­ cation of compounds and subsequent decision-making.
troduction systems rely on the use of FIA or HPLC Rapid detection techniques are promising in the area of
(High Performance Liquid Chromatography) pumps. biotechnology, or other fields such as chemistry, biology,
However, bear in mind that those systems dispense the and clinical and forensic analysis, as they can be the
use of the chromatographic column, and work very re­ initial step within analytical studies. Fields such as the
duced back pressures, in a previous review, we asked discovery of biomarkers (whether metabolites or pro­
and raise the possibility to develop straightforward teins) for disease prognosis or diagnosis; development of
EI–MS ambient sampling systems exploiting the MS new drug candidates; analysis of metabolites and mar­
vacuum pump to propel the sample into ionization EI kers in plants; control of toxins and xenobiotics in
source [1•]. A first attempt to materialize that idea was plants; and proteins in tissues, are some possibilities that
recently introduced by the Cappiello research group. the authors believe that these techniques have potential
They developed a novel analytical strategy called ex­ to be explored.
tractive liquid sampling–electron ionization–mass spec­
trometry (E-LEI–MS), which enables direct analysis of
samples without the need for separation or pretreatment Conclusions and future perspectives
steps [46••]. E-LEI–MS combines a device for solvent EI is an ionization technique with an unparalleled ability
release and ambient sampling with an EI source in a to provide highly reproducible and comparable mass
single-quadrupole mass spectrometer (Figure 3c). This spectra. Along with GC, EI–MS is the chief technique in
system has demonstrated successful applications in the targeted and untargeted determination of volatile
analyzing active ingredients in pharmaceutical tablets, compounds in several research fields. However, only less
detecting pesticides on fruit peel, determining drugs of of the 20% of molecules of current interest are analyz­
abuse (such as cocaine) on banknotes, and analyzing able by GC–EI–MS. Developing technology that allows
unknown components on painting surfaces. exploiting the advantages of EI–MS in the analysis of
nonvolatile compounds could bring new critical insights
Direct inlet probe into biotechnological research. Hence, the coupling with
The direct inlet probe (DIP) probe is one way to in­ nano-LC, through Direct EI, LEI, and Cold-EI inter­
troduce solid or liquid samples directly into the EI–MS faces, is expanding the range of applicability of the
source, currently available in many commercial GC–MS EI–MS and opening the door for the development of
instruments. This probe consists of a stainless-steel rod unified GC/LC instruments. Although important ad­
designed, so that a quartz cup (i.d. 1 mm, length 7 mm) vances have been achieved in the last 20 years, nano-
can be fitted at the tip into which a solid sample of LC–EI–MS still endures significant limitations.
∼0.2 mg is introduced [47,48]. In this approach, the Therefore, direct sample introduction strategies are a
sample contained in the DIP is introduced into the promising alternative for obtaining EI–MS spectra of

www.sciencedirect.com Current Opinion in Biotechnology 2023, 82:102965


8 Analytical Biotechnology

heavy compounds. While probes for direct sample in­ This paper Focus on the potential of endogenous volatile organic me­
troduction into EI–MS have been commercially avail­ tabolites (VOMs) as an innovative and efficient approach to establishing
a volatilomic biosignature and potentially define a panel of prostate
able for several years, those have been little explored. cancer biomarkers as a complementary tool to be used in clinical
New practical demonstrations of their potential could practice for its diagnostic and management.
provide a bunch of new possibilities for biotechnology 5. Majchrzak T, Wojnowski W, Piotrowicz G, Gębicki J, Namieśnik J:
research. Sample preparation and recent trends in volatolomics for
diagnosing gastrointestinal diseases. TrAC Trends Anal Chem
2018, 108:38-49.
CRediT authorship contribution statement 6. Tholl D, Hossain O, Weinhold A, Röse USR, Wei Q: Trends and
Deyber Arley Vargas Medina: Conceptualization, applications in plant volatile sampling and analysis. Plant J
2021, 106:314-325.
Visualization, Writing – original draft. Edvaldo Vasconcelos
Soares Maciel: Conceptualization, Visualization, Writing – 7. Llambrich M, Brezmes J, Cumeras R: The untargeted urine
volatilome for biomedical applications: methodology and
original draft. Natalia Gabrielly Pereira dos Santos: volatilome database. Biol Proced Online 2022, 24:2-20.
Conceptualization, Visualization, Writing – original draft. 8. Mazur DM, Detenchuk EA, Sosnova AA, Artaev VB, Lebedev AT:
Fernando Mauro Lancas: Conceptualization, Visualization, GC-HRMS with complementary ionization techniques for target
Funding acquisition, Project administration, Supervision, and non-target screening for chemical exposure: expanding
the insights of the air pollution markers in Moscow Snow. Sci
Writing – review & editing. Total Environ 2021, 761:144506.
9. Vargas Medina DA, Maciel EVS, de Toffoli AL, Lanças FM:
Data Availability Miniaturization of liquid chromatography coupled to mass
spectrometry. TrAC Trends Anal Chem 2020, 128:115910.
10. Khodadadi M, Pourfarzam M: A review of strategies for
Data will be made available on request. untargeted urinary metabolomic analysis using gas
chromatography–mass spectrometry. Metabolomics 2020,
16:1-14.
Declaration of Competing Interest
11. Lubes G, Goodarzi M: GC–MS based metabolomics used for the
identification of cancer volatile organic compounds as
The authors declare that they have no competing in­ biomarkers. J Pharm Biomed Anal 2018, 147:313-322.
terests or personal relationships that could have influ­ 12. Rashid A, Ali V, Khajuria M, Faiz S, Gairola S, Vyas D: GC–MS
enced this work. based metabolomic approach to understand nutraceutical
potential of Cannabis seeds from two different environments.
Food Chem 2021, 339:128076.
Acknowledgements 13. Zhang A, Sun H, Wang X: Mass spectrometry-driven drug
This work was supported by the São Paulo Research Foundation, FAPESP discovery for development of herbal medicine. Mass Spectrom
[Grants 2017/02147-0, 2019/22724-7]; the National Council for Scientific Rev 2018, 37:307-320.
and Technological Development, CNPq [Grant 307293/2014-9]; the
Coordination for the Improvement of Higher Education Personnel 14. Valdés A, Álvarez-Rivera G, Socas-Rodríguez B, Herrero M, Ibáñez
— Brazil, CAPES [Finance Code 001]; and the Alexander von Humboldt E, Cifuentes A: Foodomics: analytical opportunities and
challenges. Anal Chem 2022, 94:366-381.
Foundation.
15. Christ B, Pluskal T, Aubry S, Weng J-K: Contribution of
untargeted metabolomics for future assessment of biotech
References and recommended reading crops. Trends Plant Sci 2018, 23:1047-1056.
Papers of particular interest, published within the period of review, have 16. Yang Q, Zhang A, Miao J, Sun H, Han Y, Yan G, Wu F, Wang X:
been highlighted as: Metabolomics biotechnology, applications, and future trends: a
systematic review. RSC Adv 2019, 9:37245-37257.
•• of special interest
•• of outstanding interest 17. Mauerhofer L-M, Pappenreiter P, Paulik C, Seifert AH, Bernacchi S,
Rittmann SK-MR: Methods for quantification of growth and
productivity in anaerobic microbiology and biotechnology. Folia
1. Maciel EVS, Pereira dos Santos NG, Vargas Medina DA, Lanças
Microbiol 2019, 64:321-360.
• FM: Electron ionization mass spectrometry: quo vadis?
Electrophoresis 2022, 43:1587-1600. 18. Moros G, Chatziioannou AC, Gika HG, Raikos N, Theodoridis G:
The authors provided a comprehensive overview of the fundamentals of Investigation of the derivatization conditions for GC–MS
EI–MS, emphasizing its development over the years, its coupling with metabolomics of biological samples. Bioanalysis 2017, 9:53-65.
gas and LC, and future trends.
19. Rigano F, Tranchida PQ, Dugo P, Mondello L: High-performance
2. Vargas Medina DA, Maciel EVS, Lanças FM: Mass spectrometric •• liquid chromatography combined with electron ionization mass
detection, instrumentation, and ionization methods. Liquid spectrometry: a review. TrAC - Trends Anal Chem 2019,
Chromatography. Elsevier; 2023:679-706. 118:112-122.
The paper provides a review of the state of the art in the field of LC
3. Carazzone C, P.G. Rodríguez J, Gonzalez M, López G-D:
hyphenated to EI MS, describing the diverse interfaces developed for
•• Volatilomics of natural products: whispers from nature.
that purpose over the last twenty years.
Metabolomics - Methodology and Applications in Medical Sciences
and Life Sciences. IntechOpen; 2021:13. 20. Famiglini G, Palma P, Termopoli V, Cappiello A, Tsizin S, Seemann
The paper discusses the study of VOCs emitted by living organisms like B, Alon T, Fialkov AB, Amirav A: Electron ionization LC-MS: what
plants, flowers, animals, fruits, and microorganisms using metabolomics is it and why use it? Compr Anal Chem 2018, 79:1-28.
tools. The paper covers the different theoretical and practical aspects of
the study of the volatilome, including sample preparation, extraction, 21. Biemann K: Laying the groundwork for proteomics: mass
instrumental analysis, and data processing. spectrometry from 1958 to 1988. J Proteom 2014, 107:62-70.
4. Berenguer CV, Pereira F, Pereira JAM, Câmara JS: Volatilomics: 22. Famiglini G, Palma P, Termopoli V, Cappiello A: The history of
•• an emerging and promising avenue for the detection of •• electron ionization in LC-MS, from the early days to modern
potential prostate cancer biomarkers. Cancers 2022, 14:3982. technologies: a review. Anal Chim Acta 2021, 1167:338350.

Current Opinion in Biotechnology 2023, 82:102965 www.sciencedirect.com


Electron impact mass spectrometry in biotechnology Vargas Medina et al. 9

The paper traces the history of LC-MS using EI and discusses the 37. Famiglini G, Termopoli V, Palma P, Cappiello A: Liquid
challenges and successes of coupling these techniques, mainly those chromatography-electron ionization tandem mass
related to high-vacuum and high-temperature conditions required for spectrometry with the direct-EI interface in the fast
the operation of EI and the need to remove the eluent carrying the determination of diazepam and flunitrazepam in alcoholic
analyte before entering the ion source. Besides, the paper provides in­ beverages. Electrophoresis 2016, 37:1048-1054.
sights into the development and application of LC-MS techniques for
biological, pharmaceutical, and environmental samples. 38. Cappiello A, Famiglini G, Termopoli V, Trufelli H, Zazzeroni R,
Jacquoilleot S, Radici L, Saib O: Application of liquid
23. Abonamah JV, Eckenrode BA, Moini M: On-site detection of chromatography-direct-electron ionization-MS in an in vitro
fentanyl and its derivatives by field portable nano-liquid dermal absorption study: quantitative determination of trans
chromatography-electron lonization-mass spectrometry (nLC- -cinnamaldehyde. Anal Chem 2011, 83:8537-8542.
EI-MS). Forensic Chem 2019, 16:100180.
39. Cappiello A, Famiglini G, Palma P, Termopoli V, Trufelli H: A new
24. Rigano F, Albergamo A, Sciarrone D, Beccaria M, Purcaro G, liquid chromatography-mass spectrometry approach for
Mondello L: Nano liquid chromatography directly coupled to generic screening and quantitation of potential genotoxic
electron ionization mass spectrometry for free fatty acid alkylation compounds without derivatization. J Chromatogr A
elucidation in Mussel. Anal Chem 2016, 88:4021-4028. 2012, 1255:286-290.
25. Vargas Medina DA, Pereira dos Santos NG, da Silva Burato JS, 40. Trufelli H, Famiglini G, Termopoli V, Cappiello A: Profiling of non-
• Borsatto JVB, Lanças FM: An overview of open tubular liquid esterified fatty acids in human plasma using liquid
chromatography with a focus on the coupling with mass chromatography-electron ionization mass spectrometry. Anal
spectrometry for the analysis of small molecules. J Chromatogr Bioanal Chem 2011, 400:2933-2941.
A 2021, 1641:461989.
The paper highlights the potential of OT-LC-MS in providing enhanced 41. Zhang LX, Burdette CQ, Phillips MM, Rimmer CA, Marcus RK:
sensitivity when coupled with ESI- sources and the unprecedented Determination of isoflavone content in SRM 3238 using liquid
detection capabilities if interfaced with a highly structural informative EI chromatography-particle beam/electron ionization mass
source. spectrometry. J AOAC Int 2015, 98:1483-1490.
26. Vargas Medina DA, Burato JS, da S, Borsatto JVB, Lanças FM: 42. Castro J, Krishna MVB, Marcus RK: Liquid chromatography-
Porous layer open tubular nano liquid chromatography directly particle beam electron ionization mass spectrometry method
coupled to electron ionization mass spectrometry. J for analysis of botanical extracts: evaluation of ephedrine
Chromatogr A 2022, 1674:463143. alkaloids in standard reference materials. J AOAC Int 2010,
93:1788-1797.
27. Xiang P, Yang Y, Zhao Z, Chen M, Liu S: Ultrafast gradient
separation with narrow open tubular liquid chromatography. 43. Castro J, Pregibon T, Chumanov K, Marcus RK: Determination of
Anal Chem 2019, 91:10738-10743. catechins and caffeine in proposed green tea standard
reference materials by liquid chromatography-particle beam/
28. Chen A, Liu S, Yang Y, Xiang P: Liquid chromatographic electron ionization mass spectrometry (LC-PB/EIMS). Talanta
separation using a 2 μm i.d. open tubular column at elevated 2010, 82:1687-1695.
temperature. Anal Chem 2021, 93:4361-4364.
44. Rocío-Bautista P, Famiglini G, Termopoli V, Palma P, Nazdrajić E,
29. Termopoli V, Famiglini G, Palma P, Piergiovanni M, Rocio-Bautista
Pawliszyn J, Cappiello A: Direct coupling of bio-SPME to liquid
P, Ottaviani MF, Cappiello A, Saeed M, Perry S: Evaluation of a
electron ionization-MS/MS via a modified microfluidic open
liquid electron ionization liquid chromatography–mass
interface. J Am Soc Mass Spectrom 2021, 32:262-269.
spectrometry interface. J Chromatogr A 2019, 1591:120-130.
30. Cappiello A: The benefits of a liquid-electron ionization liquid 45. Cerdà V, Ferrer L, Avivar J, Cerdà A: Evolution and description of
chromatography – mass spectrometry interface. LCGC 2019, the principal flow techniques. Flow Analysis. Elsevier; 2014:1-42.
32:330-331. 46. Arigò A, Famiglini G, Marittimo N, Agostini M, Renzoni C, Palma P,
31. Termopoli V, Famiglini G, Palma P, Piergiovanni M, Cappiello A: •• Cappiello A: Extractive-liquid sampling electron ionization-
Atmospheric pressure vaporization mechanism for coupling a mass spectrometry (E-LEI-MS): a new powerful combination
liquid phase with electron ionization mass spectrometry. Anal for direct analysis. Sci Rep 2023, 13:6429.
Chem 2017, 89:2049-2056. The paper Introduces a new analytical strategy called E-LEI-MS for the
direct analysis of samples without the need for separation or pre-
32. Margolin Eren KJ, Elkabets O, Amirav A: A comparison of electron treatment steps, showcasing its applications to the analysis of active
ionization mass spectra obtained at 70 eV, low electron ingredients in pharmaceutical tablets, pesticides on fruit peel, a drug of
energies, and with cold EI and their NIST library identification abuse (cocaine) determination in banknotes, and analysis of unknown
probabilities. J Mass Spectrom 2020, 55:1-15. components on painting surfaces.
33. Alon T, Amirav A: Isotope abundance analysis methods and 47. Flego C, Zannoni C: Direct insertion probe−mass spectrometry:
software for improved sample identification with supersonic a useful tool for characterization of asphaltenes. Energy Fuels
gas chromatography/mass spectrometry. Rapid Commun Mass 2010, 24:6041-6053.
Spectrom 2006, 20:2579-2588.
48. Käfer U, Gröger T, Rüger CP, Czech H, Saraji-Bozorgzad M,
34. Tsizin S, Bokka R, Keshet U, Alon T, Fialkov AB, Tal N, Amirav A: Wilharm T, Zimmermann R: Direct inlet probe – high-resolution
Comparison of electrospray LC–MS, LC–MS with Cold EI and time-of-flight mass spectrometry as fast technique for the
GC–MS with Cold EI for sample identification. Int J Mass chemical description of complex high-boiling samples. Talanta
Spectrom 2017, 422:119-125. 2019, 202:308-316.
35. Tsizin S, Fialkov AB, Amirav A: Electron ionization mass 49. Gross JH: Mass Spectrometry. Springer; 2011.
spectrometry for both liquid and gas chromatography in one
system without the need for hardware adjustments. J Am Soc 50. Zannoni C, Flego C: Direct insertion probe - mass spectrometry
Mass Spectrom 2020, 31:1713-1721. in the characterization of opportunity crudes. LC-GC Eur 2012,
10:28-35.
36. Rigano F, Russo M, Arigò A, Dugo P, Mondello L: Combining
linear retention index and electron ionization mass 51. Zeki ÖC, Eylem CC, Reçber T, Kır S, Nemutlu E: Integration of
spectrometry for a reliable identification in nano liquid GC–MS and LC–MS for untargeted metabolomics profiling. J
chromatography. J Chromatogr A 2020, 1610:460581. Pharm Biomed Anal 2020, 190:113509.

www.sciencedirect.com Current Opinion in Biotechnology 2023, 82:102965

You might also like