Series 5000 High-Range Phosphate Analyzer Quick Reference Guide* 60001-44

Key Option Description


SETUP ALARMS RESET ALARMS Reset present sample alarms and system alarms
ALARMS ENABLED† Enables/disables sample alarms
L† ALM 1 0.0 mg/l† Sample alarm 1 (High, Low, Rate; 0.0 to 50.0)
† † Sample alarm 2 (High, Low, Rate; 0.0 to 50.0)
H ALM 2 50.0 mg/l
ALARM CONFIG Configuration of sample alarms: ALARM 1 = LOW† (High, Low, Rate); ALARM 2 = HIGH†
(High, Low, Rate); POWER FAIL: OFF† (Off, On)
RELAY CONFIG Configuration of alarm relays: Relay 1 and 2 high voltage rating; relay 3 and 4 low
voltage rating)
REAGENTS RBLANK 0.0 mg/l† Concentration of reagent blank to be automatically subtracted from results
REAG LEVEL 100%† Present level of reagents — set to 100% when replaced
STD LEVEL 100%† Present level of standard — set to 100% when replaced
PRIME REAGENTS Fills reagent tubing with reagent
RECORDER †
RECMAX 50.0 mg/l Upper limit of recorder in terms of concentration
† Lower limit of recorder in terms of concentration
RECMIN 0.0 mg/l
ON ALARM HOLD — Recorder output continues holding at the output last detected before the
system alarm occurred
GO MAX — Recorder output to full scale
GO MIN — Recorder output to zero
AUTO-SET CURRENT Calibrates the range of the recorder to 4-20mA when current output is selected
AUTO-SET VOLTAGE Calibrates the range of the recorder when a voltage output is selected
MANUAL SET REC INCR FULL SCALE — Use with TEST menu, RECORDER TEST, OUTPUT REC FS
option. Use INCR FULL SCALE to increase to desired level if recorder full scale is
too small
DECR FULL SCALE — Use with TEST menu, RECORDER TEST, OUTPUT REC FS
option. Use DECR FULL SCALE to decrease to desired level if recorder full scale is too
large
INCR ZERO — Use with TEST menu, RECORDER TEST, OUTPUT REC ZERO option.
If recorder zero is too small, use INCR ZERO to increase to desired level
DECR ZERO — Use with TEST menu, RECORDER TEST, OUTPUT REC ZERO option.
If recorder zero is too large, use DECR ZERO to decrease to desired level
PRINT SETUP Prints analyzer settings
INITIAL SETUP † Current military time of day
HR:MIN 00:00
M/D/Y 01/01/94† Current date
DAY SUNDAY† Current day of week
DISPLAYED CONC Selects display to read as Phosphate† (PO4) or Phosphorus (P)
*
Press one of the following number keys on power up to select an alternative menu language: 1 for English, 2 for German, 3 for French, or 4 for Spanish

Default condition at cold start
Series 5000 High-Range Phosphate Analyzer Quick Reference Guide (continued)
SETUP INITIAL SETUP SMP CND STATUS Enables/disables† sample conditioning relay to operate external electronic control
(cont.) (cont.) module (When enabled: BACKFLUSH: OFF† — sets air backflush cycle to occur every 0
to 99 cycles; DISINFECT: OFF† — sets disinfection cycle to occur every 1,8,12, 24, 72, or
168 hours; TM TIL DF:XXX:XX — sets the initial time for first disinfection cycle to occur
COLD START Resets analyzer to default conditions
BAUD RATE 1200† Baud rate of serial I/O (300, 600, 1200, 2400, 4800, 9600)
STOP BITS 2† Number of stop bits of serial I/O (1, 2)

PARITY NONE Parity of serial I/O (Odd, Even, None)

CHAR LENGTH 8 Character length of serial I/O (7, 8)
ADJUST CONTRAST Modifies display contrast (0, 1, 2, 3†, 4, 5, 6, 7)
KYBD UNLOCKED† Prevents unauthorized keyboard access (KYBD UNLOCKED, KYBD LOCKED, KYBD
PARTIAL LOCK)
CALIB AUTO-CALIBRATION Initiates automatic calibration at user request
RECALL CAL VALUE Recalls the last calibration value
Initiates automatic calibration on a specified day and time of the week and may
AUTO-CALIB TIME be disabled
STANDARD VALUE Concentration of standard used in automatic calibration
Changes currently displayed concentration to agree with independently established
USER CALIBRATION value — (not recommended; specifications are based on automatic calibration)
DEFAULT CALIB Resets analyzer to default calibration
TEST RELAY TEST Activates alarms at user request
RECORDER TEST Confirms recorder setup for full scale, zero and 1/2 scale
PRINTER TEST Verifies normal printer operation
GRAB SAMPLE TEST Allows user to measure grab sample
RECALL G SMP VAL Recalls the last Grab Sample Value
TIME REMAINING Displays time remaining in present measurement cycle
DISPLAY TEST Verifies normal display operation
REAGENT PRESSURE Measures and displays reagent pressure
SAMPLE PRESSURE Measures and displays sample pressure to analyzer
SYSTEM RESET Resets system alarms, warnings and sample alarms and starts new cycle
CLEAR Clears data entry or returns to normal display
NEXT Use to move through menus and select fixed options
ENTER Use to enter submenus and to select values


Default condition at cold start

HACH COMPANY, P.O. Box 389, Loveland, CO 80539, (970) 669-3050 or (800) 227-4224, FAX # (970) 669-2932 4/94-1ED
Rev. 2, 12/96