You are on page 1of 11

Microscopy and Microanalysis (2020), 1–11

doi:10.1017/S1431927620024368

Original Article

Ultra-Microtome for the Preparation of TEM Specimens from Battery


Cathodes
Hanlei Zhang1,2,3, Chongmin Wang4 and Guangwen Zhou2,3*
1
Engineering Research Center of Nano-Geomaterials of Ministry of Education, Faculty of Materials Science and Chemistry, China University of Geosciences, No. 68
Jincheng Street, East Lake High-Tech Development Zone, Wuhan 430078, Hubei, P. R. China; 2Materials Science and Engineering Program & Department of
Mechanical Engineering, State University of New York, Binghamton, NY 13902, USA; 3NorthEast Center for Chemical Energy Storage, State University of New York,
Binghamton, NY 13902, USA and 4Environmental Molecular Sciences Laboratory, Pacific Northwest National Laboratory, Richland, WA 99352, USA

Abstract
With the wide application of ultra-microtome sectioning in the preparation of transmission electron microscopy (TEM) specimens with
bio- and organic materials, here, we report an ultra-microtome-based method for the preparation of TEM specimens from cathodes of
Li-ion batteries. The ultra-microtome sectioning reduces the sample thickness to tens of nanometers and yields atomic resolution from
the core region of particles of hundreds of nanometers. Analysis indicates that the mechanical cross-sectioning introduces no observable
microstructural artifacts or structural damage, such as microcracking and nanoporosity. These results demonstrate the high efficiency of
the ultra-microtome approach in preparing well-thinned specimens of particulate materials that allow for atomic-scale TEM imaging of
a large number of sectioned particles in one single TEM specimen, thereby providing statistically significant results of the TEM analysis.
Key words: transmission electron microscopy, sample preparation, artifact, atomic resolution, ultra-microtome
(Received 2 April 2020; revised 28 June 2020; accepted 7 July 2020)

Introduction most materials with enough fragility (Mori et al., 1999), but it
does not work with soft materials. Also, it can significantly change
Specimen preparation is a critical step in the transmission elec-
the macroscale morphology of materials, as well as introducing
tron microscopy (TEM) analysis of materials (Klepeis et al.,
undesirable artifacts such as microcracking and porosity (Ahn
1987; Ishitani & Yaguchi, 1996; Sun et al., 2005; Mayer et al.,
& Peacor, 1986). FIB is relatively expensive and time-consuming
2007). Well-prepared specimens reveal the authentic microstruc-
in preparing TEM specimens. FIB works well with a large number
tural features of materials by minimizing misleading artifacts gen-
of materials and usually does not change the materials’ macro-
erated by the specimen preparation process (Dai et al., 2001). To
scale morphology, but it can possibly change the atomic-scale
perform TEM imaging with clear atomic resolution, a well-shaped
morphology via ion implantation and knock-on effects (Lezec
thin-film specimen with a thickness of <100 nm is usually desired
et al., 1991; Volkert & Minor, 2007). These two effects can be
(Burghardt & Droleskey, 2006; Zhao et al., 2017). Unfortunately,
minimized via the state-of-the-art cryo-FIB technique which
most of the raw materials for observation are randomly shaped
involves freezing the specimens with liquid nitrogen while sec-
with a much larger thickness. To overcome this universal issue,
tioning (Wang et al., 2012; Zhang et al., 2016), which has demon-
multiple thinning methods have been developed for obtaining
strated to minimize the ion-beam damage to sensitive materials
TEM specimens with a desirable thickness, including mechanical
such as biomaterials (Schertel et al., 2013; Schaffer et al., 2019)
cutting and ion-beam milling (Barna et al., 1999; McCaffrey et al.,
and battery cathodes (Zachman et al., 2018, 2020). FIB can
2001; Liu et al., 2007), crushing and grinding (Zhang & Howes,
easily prepare TEM specimens as large as 10 μm (Li et al.,
1994; Zhang et al., 2011), and focused ion beam (FIB) cutting
2006; Bashir et al., 2019), meeting most requirements in specimen
(Giannuzzi & Stevie, 1999; Schaffer et al., 2012; Lu et al., 2013).
preparation. However, when analyzing particulate materials of
All these methods have different advantages and disadvantages,
hundreds of nanometers (such as the cathode particles we study
meaning that they need to be carefully chosen based on the spe-
in this work), even the largest FIB sample can only contain a
cific characterization requirements. Mechanical cutting and mill-
dozen of particles, considering their dispersion. Such a sample
ing are suitable for large, bulk materials with enough ductility
volume is not sufficient for obtaining statistically significant
(Puttick et al., 1994), while they do not work with powder or
results.
very fragile materials. Crushing and grinding can be useful for
Particulate materials ranging from nanometers to hundreds of
micrometers are widely used in many fields of the industry such
as energy storage (Yoshida et al., 2013; Siu et al., 2018; Zhang
*Author for correspondence: Guangwen Zhou, E-mail: gzhou@binghamton.edu
et al., 2018), catalysis (He et al., 2017; Stamenkovic et al., 2017),
Cite this article: Zhang H, Wang C, Zhou G (2020) Ultra-Microtome for the
Preparation of TEM Specimens from Battery Cathodes. Microsc Microanal. and structural composites (Flom & Arsenault, 1989; Kumlutaş
doi:10.1017/S1431927620024368 et al., 2003; Yang et al., 2003). Accordingly, there is an increasing
© Microscopy Society of America 2020

Downloaded from https://www.cambridge.org/core. Cornell University Library, on 01 Sep 2020 at 13:33:26, subject to the Cambridge Core terms of use, available at
https://www.cambridge.org/core/terms. https://doi.org/10.1017/S1431927620024368
2 H. Zhang et al.

requirement to characterize their microstructures and properties material (e.g., epoxy) and then sliced into thin films, as schemati-
with TEM techniques. Among the TEM specimen preparation cally shown in Figure 1c. During this process, the diamond knife
methods mentioned above, only FIB is suitable for the prepara- can push the NPs out of the matrix (Fig. 1d) instead cutting
tion of high-quality TEM specimens from particles. The other through them (Fig. 1c), failing in thinning the NPs. This phenom-
methods either fail in working with particles or generate undesir- enon is observed with multiple popular nanomaterials such as
able artifacts. Although FIB has the ability to locate and section zeolite (Haswell et al., 2003), silicates (Messenger et al., 2003),
target particles with a high precision, it is also very time- and metal oxides (Tucker et al., 1985). In these materials, the
consuming and only analyzes a relatively small area at one time, internal chemical bonding within the NP is stronger than its con-
as discussed above. Therefore, FIB is not suitable for analyzing nection with the epoxy matrix, so the interfacial connection pref-
a large number of particles of hundreds of nanometers to obtain erably disassembles upon the microtome cutting.
statistically significant information. In this work, we developed an ultra-microtome-based method
To overcome these drawbacks, we have used a TEM specimen for the preparation of TEM specimens from lithium transition
preparation method with the ultra-microtome technique, a metal oxides (battery cathode particles). Cathode materials are
mechanical sectioning method that produces millimeter-sized widely synthesized by a precipitation process involving primary
slices in lateral with a thickness of only tens of nanometers, allow- particles and secondary particles (May et al., 2018; Zhang et al.,
ing for the analysis of tens of thousands of particles in one TEM 2018; Boesenberg et al., 2019). Primary particles normally have
specimen. The ultra-microtome method was originally developed a size in the hundred nanometer range, and secondary particles
for sectioning biomaterials and organic materials (Reymond & usually have a particle size in the micrometer scale. We show
Pickett-Heaps, 1983; Harris et al., 2006; Graf et al., 2009; the feasibility of preparing cathode particles into well-thinned
Schrand et al., 2010; Brown et al., 2014), which is feasible based specimens that allow for atomic-scale TEM imaging of hundreds
on the fact that these materials are soft enough and can be cut of thousands of sectioned particles in one single TEM specimen,
smoothly with the diamond knife in microtomy. Following the thereby providing a statistically significant result of the TEM anal-
pioneering work by Malis & Steele (1990) and Mcmahon & ysis. The results also show that the ultra-microtome sectioning
Malis (1995), ultra-microtomy has been seen increasingly as an process does not introduce observable mechanical damage or
attractive alternative means of producing TEM specimens of microstructural artifacts into the specimens. This ultra-
hard and inorganic materials (Becker & Bange, 1993; microtome method may find broader applicability in sectioning
Antonovsky, 1995; Glanvill, 1995; Quintana, 1997). A main chal- other particulate materials for TEM observations.
lenge for ultra-microtomy is that the diamond knife sectioning
may generate mechanical defects in the specimen such as cracking
and porosity, as schematically shown in Figures 1a and 1b. Experimental
Another challenge of the ultra-microtomy method resides in the
Materials
cutting of robust inorganic nanoparticles (NP) such as silicates
and metal oxides (Tucker et al., 1985; Mori et al., 1992; In this work, we use the industrial-level LiNi0.80Co0.15Al0.05O2
Hochella et al., 1999). The NPs are embedded within a matrix (NCA) cathode material for Li-ion batteries as the example to

Fig. 1. Challenges in the ultra-microtome sectioning of inorganic materials. (a,b) Schematics showing the cracks and porosity generated by ultra-microtome cut-
ting. (c) Schematic showing the ultra-microtome sectioning of NPs imbedded in the epoxy. (d) Schematic showing the poor connection between the NPs and the
matrix material upon the ultra-microtome cutting.

Downloaded from https://www.cambridge.org/core. Cornell University Library, on 01 Sep 2020 at 13:33:26, subject to the Cambridge Core terms of use, available at
https://www.cambridge.org/core/terms. https://doi.org/10.1017/S1431927620024368
Microscopy and Microanalysis 3

Fig. 2. Preparation of TEM specimens with ultra-microtome cutting. (a) Bar-shaped capsule models for the preparation of epoxy bars containing cathode particles.
(b) Solidified epoxy bar containing cathode materials in its head, as marked with an orange arrow. (c) Schematic showing the slicing of the epoxy bar with ultra-
microtome. (d) Scanning electron microscopy (SEM) image of the pristine NCA secondary particles. (e) Schematic showing the NCA material used for ultra-
microtome cutting, which is secondary particles composed of primary particles of hundreds of nanometers. (f) TEM analysis of the specimen prepared with
ultra-microtome.

demonstrate the capability of the ultra-microtome method in pre- with a pointed head (Fig. 2a) and settled at 60°C for 24 h for sol-
paring TEM specimens. The large size of the primary particles in idification. Cathode particles, especially cycled samples, tend to
the NCA cathode (hundreds of nanometers) prohibits direct undergo degradation in air. Therefore, the influence of air expo-
microscopic observation of the core area. Therefore, there is a sig- sure and heating on cycled samples could be a concern for the
nificant lack of information regarding the microstructural feature ultra-microtome method and preventive measures need to be
in the bulk for such large potato-like cathode particles, mainly taken to minimize such an effect. For instance, air exposure can
due to the challenge in preparing appropriate specimens for be minimized by rapidly sealing the cathode particles with
TEM analysis (Zhang et al., 2017b, 2017c). Pristine NCA cathode epoxy when preparing the epoxy bars for ultra-microtome. Our
material was obtained from TODA America Inc. X-ray diffraction results indicate that brief air exposure (a few seconds) and the
analysis in our previous work (Zhang et al., 2017b) indicated that subsequent curing of the solidified epoxy resin at 60°C for 24 h
the pristine material has an ideal R3m crystal configuration. The do not induce noticeable changes to the morphology and atomic
pristine NCA material was prepared into battery cathodes by mix- structure of pristine and cycled NCA particles. This is confirmed
ing the active material, carbon black, and polyvinylidene fluoride by comparing the cycled samples sectioned by ultra-microtome
(PVDF) with a weight ratio of 8:1:1, using N-methyl-2-pyrroli- and FIB. As demonstrated in our previous work (Zhang et al.,
done solvent. Cathodes with 3–5 mg of active NCA were assem- 2017c), primary particles sectioned by ultra-microtome and FIB
bled in 2325-type coin cells in a glovebox filled with helium. yield the similar feature of (003) intragranular cracking and the
The electrolyte was 1 M LiPF6 dissolved in a mixture of ethylene presence of a rock-salt layer between the crack. This negligible
carbonate (EC) and dimethyl carbonate (DMC), with a volume effect is further confirmed by examining the thickness of the sur-
ratio of 1:1. The coin cells were cycled with an MPG2 multichan- face rock-salt layer formed on the 30-cycle NCA sample. Our
nel potentiostat (Biologic). Galvanostatic cycling was performed TEM imaging shows that the thickness of the rock-salt layer on
at a rate of C/10 with a current density of 15 mA/g, between the microtome-sectioned particles is similar to that on uncut par-
3.0 and 4.3 V. ticles (Zhang et al., 2017b), indicating that the sample preparation
with ultra-microtome does not induce noticeable growth of the
rock-salt layer. Solidified resin bars containing the objective mate-
Ultra-Microtome Cutting for the Preparation of TEM Specimens
rials were extracted from the capsule model and trimmed with a
The TEM specimen preparation was performed using a razor blade (Fig. 2b) for ultra-microtome sectioning. Since only
Reichert-Jung Ultracut E ultra-microtome machine, equipped the head of the resin bar is used in the microtome cutting, the
with a 45° diamond knife manufactured by DiATOME U.S. material of interest is purposely concentrated in the head region
Pristine materials and cycled cathodes scraped from the current to increase the particle concentration in the TEM specimen.
collector were mixed with white resin (Electron Microscopy Slices with a nominated thickness of ∼50 nm were sectioned
Sciences, hard grade), pulled into a bar-shaped capsule model from the tip of the resin bars using the ultra-microtome machine

Downloaded from https://www.cambridge.org/core. Cornell University Library, on 01 Sep 2020 at 13:33:26, subject to the Cambridge Core terms of use, available at
https://www.cambridge.org/core/terms. https://doi.org/10.1017/S1431927620024368
4 H. Zhang et al.

at a cutting rate of ∼0.8 mm/s, as schematically shown in sample drifting or vibration under the electron beam irradiation.
Figure 2c. More details regarding the sample-cutting procedures Zoom-in view from inside the epoxy slice (Figs. 3b, 3c) clearly
with ultra-microtome can be found in the literature (Hughes shows the presence of cathode particles embedded in the epoxy
et al., 2014; Sun et al., 2017). for TEM imaging. The epoxy film in Figures 3b and 3c is almost
Figure 2d presents an SEM image of pristine NCA particles, invisible at the high magnification, demonstrating its ultra-thin
indicating that the pristine material is composed secondary parti- thickness.
cles of 5–10 μm, which consist of primary particles of ∼300–
1000 nm. Since the atomic HRTEM imaging usually requires a
Atomic-Resolution Imaging Enabled with Ultra-Microtome
specimen thickness of less than ∼100 nm (Buseck et al., 1989;
Williams & Carter, 2009), only the surface region along the Figure 4a presents a zoom-out TEM image of a pristine NCA par-
thin edges of the primary particles is suitable for HRTEM imag- ticle sectioned with ultra-microtome. The particle has a relative
ing, while no atomic resolution can be obtained from the core uniform contrast without significantly enhanced darkness in the
area (Zhang et al., 2017a). The thinning process of the NCA par- core region, indicating that the thickness of the entire particle
ticles via ultra-microtome is schematically summarized in Figures has been uniformly reduced. Meanwhile, TEM imaging of the
2e and 2f. Before cutting (Fig. 2e), the material is secondary par- sliced particle shows an intact morphology without any observ-
ticles of micrometers, composed of primary particles of hundreds able cracking or fractured feature, indicating the absence of
of nanometers. Besides the large thickness, the overlapping of the mechanical damage from the ultra-microtome process. Figures
primary particles also prevents atomic resolution. After cutting 4b to 4e present HRTEM views obtained progressively from the
(Fig. 2f), the secondary particle is sectioned into a slice with a particle surface toward the core region (0–280 nm), which show
thickness of ∼50 nm. The internal structures of the particles are that atomic imaging is eligible across the particle, indicating
well maintained, as will be demonstrated later. The boundaries that the thickness of whole particle is reduced under ∼100 nm.
between neighboring primary particles are not cracked, and no It is worth mentioning that the most comment zone axis for
microstructural defects are introduced into the crystal lattice. observing the layered phase is along the [010] direction to reveal
This overcomes the overly high thickness and overlapping of pri- the layered feature, as shown in Figures 5a and 5b and previous
mary particles, making the slice a well-defined subject for TEM studies (Lin et al., 2014a, 2014b; Yan et al., 2017). Since the inter-
observations, as schematically illustrated in Figure 2f. layer d-spacing is very large (∼4.7 Å), the interlayer feature can be
resolved for relatively lower-resolution images, as shown in
Figure 5c. Here, we purposely avoided the commonly used
TEM Observation [010] zone axis and selected the [221] zone axis because the latter
TEM and high-resolution TEM (HRTEM) observations were per- can be used to better demonstrate the good quality of the
formed using an FEI Titan 80-300 microscope with a field emis- microtome-sectioned sample to clearly resolve the narrower lat-
sion gun (FEG) and an image aberration corrector, operated at an tice spacings. Since the Li and transition metal slabs alternatively
acceleration voltage of 300 kV. Scanning TEM (STEM) high-angle stack along the [001] direction, they are overlapped and no longer
angular dark-field (HAADF) observation was performed using an show the layered feature when viewed along the [221] zone axis,
FEI Titan 80-300 microscope with an FEG and a probe aberration as shown in Figure 5d (extracted from Fig. 4d). Coincidentally,
corrector, operated at an acceleration voltage of 300 kV. HRTEM images of the layered phase, viewing down the [221]
zone axis, have the similar lattice appearance as the rock-salt
phase along the [110] zone axis. To further confirm this feature,
SEM Observation we performed HRTEM image simulations of the layered phase
SEM observation was performed in the second electron mode along the [221] zone axis. As can be seen from Figures 5d and 5e,
using a Hitachi S-4800 FE-SEM, operated at an acceleration volt- the experimental and simulated HRTEM images match very well.
age of 2 kV. The well-shaped low-thickness particles have been repeatedly
observed in pristine NCA specimens sectioned with ultra-
microtome, as demonstrated in Figures 5f and 5g and our previ-
TEM Simulation ous work (Zhang et al., 2017b, 2017c). Figure 5f is obtained from
HRTEM simulations in this work were performed using the JEMS the bulk area of a primary particle that is ∼100 nm from the out-
TEM simulation software. The microscopic parameters used for ermost surface, revealing a clear atomic resolution of over
the simulation were from the Titian 80-300 microscope, the ∼2.19 Å. Figure 5g is obtained ∼70 nm from the outermost sur-
same as the microscope used for the experiment in this work. face, showing an atomic resolution of ∼2.95 Å. As illustrated in
Figure 5e is obtained with the established crystal structure of Figure 5, lattice points with the separation of 2.19 Å can be easily
LiNi0.80Co0.15Al0.05O2 (Mori et al., 2009), a sample thickness of resolved in the ultra-microtome sectioned cathode. In view of the
23 nm and a defocus of 41 nm. clear image contrast, a point resolution better than 2.19 Å can be
expected with ultra-microtome sectioned samples.
Here, we used the industrial-level NCA cathode to demon-
Results and Discussion strate the capability of ultra-microtome in reducing the particle
thickness. These potato-like cathode particles have the size rang-
Morphology of the TEM Specimen Prepared with
ing from ∼300 nm to 1 μm, for which HRTEM imaging is limited
Ultra-Microtome
to the surface region of 5–20 nm without access to the core area of
The epoxy slices prepared with ultra-microtome were placed on primary particles (Hwang et al., 2014; Karki et al., 2016; Zhang
copper grids coated with a lacey carbon film for TEM imaging, et al., 2017a), as shown in Figure 5h.
as shown in the SEM image in Figure 3a (pristine NCA). The Figure 5i shows a STEM-HAADF image of the pristine NCA
lacey provides a steady support for the epoxy slice, avoiding material cut with ultra-microtome, which shows a clear atomic

Downloaded from https://www.cambridge.org/core. Cornell University Library, on 01 Sep 2020 at 13:33:26, subject to the Cambridge Core terms of use, available at
https://www.cambridge.org/core/terms. https://doi.org/10.1017/S1431927620024368
Microscopy and Microanalysis 5

Fig. 3. An overview of the TEM sample prepared with ultra-microtome. (a) SEM image showing epoxy slices placed on a TEM grid. (b) Magnified view of cathode
materials imbedded in the epoxy slice. (c) Further zoom-in view showing the primary particles in the epoxy slice.

Fig. 4. Atomic-resolution imaging throughout the cathode particle. (a) Zoom-out TEM image of a cathode particle sectioned with ultra-microtome, showing a uni-
form reduction in thickness. (b–e) Zoom-in HRTEM views obtained progressively from surface toward the core, showing atomic resolution across the particle.

resolution of the R3m layered phase (same as the one in Figs. 5f, Mechanical Effects of the Ultra-Microtome Cutting
5g). Figure 5i is obtained from a region that is ∼30 nm away from
Due to the mechanical nature of the ultra-microtome cutting,
the outermost surface, indicating that specimens prepared with
there are concerns on the possibility of generating artifacts such
the ultra-microtome method are also suitable for atomic STEM
as cracking and nanoporosity. To clarify this aspect, Figure 7 stud-
imaging. On the other hand, due to the epoxy used for particle
ies the influence of ultra-microtome on the porosity of cathode
imbedding during the sample preparation, a large amount of
particles. Figures 7a to 7c present zoom-in SEM views showing
hydrocarbon compounds is present around the NCA particles.
the primary particles in (1) the pristine material, (2) a 1-cycle
These compounds act as a source of carbon, which generates con-
electrode without ultra-microtome cutting, and (3) a 1-cycle elec-
tamination when illuminated under the electron beam irradiation
trode after ultra-microtome cutting (obtained from a microtome
and prevents clear atomic-resolution STEM imaging. To eliminate
slice). The SEM images are converted into binary images to per-
this contamination, the specimen needs to be beam-showered or
form porosity analysis, as presented in Figures 7d to 7f, respec-
plasma-cleaned.
tively. The images in Figures 7a to 7c cover the area of several
Besides the NCA cathode, the ultra-microtome is also shown
micrometers and contain thousands of primary particles for
to be capable of yielding well-sectioned TEM specimens with
obtaining statistically significant results of the porosity. The
NMC811 cathode particles. Figure 6a presents a zoom-out TEM
white areas in the binary images correspond to the cathode mate-
view of NMC811 particles sectioned with the ultra-microtome.
rial, while the black areas represent voids and electrolyte.
The relatively uniform image contrast across the particle indicates
The analysis of the binary images (Figs. 7d–7f) indicates that
the evenly reduced thickness of the particle. Figures 6b and 6c are
the concentrations of cathode particles (active material) in the
obtained from regions that are ∼80 and ∼200 nm away from the
three samples are 88.75±0.06, 69.33±0.02, and 65.11±0.04%, as
outermost surface of the particle. The lattice-resolved HRTEM
shown in Figure 7g. As can be seen, the particle concentrations
images confirm that the ultra-microtome is capable of providing
in the assembled electrodes are largely reduced compared with
TEM specimens suitable for atomic-scale imaging from the differ-
the as-received pristine material. On the other hand, the particle
ent depths of the sectioned NMC811 particles without inducing
concentrations in the uncut and cut electrodes are almost the
any noticeable mechanical damage to the crystal lattice.

Downloaded from https://www.cambridge.org/core. Cornell University Library, on 01 Sep 2020 at 13:33:26, subject to the Cambridge Core terms of use, available at
https://www.cambridge.org/core/terms. https://doi.org/10.1017/S1431927620024368
6 H. Zhang et al.

Fig. 5. HRTEM observations of the R3m layered phase enabled with ultra-microtome. (a,b) HRTEM view and schematic along the [010] zone axis of the layered
phase. (c) Lower-resolution image of the layered phase, where the atomic resolution is not obtained but fringes showing the layered feature are still visible.
(d,e) HRTEM view and the corresponding HRTEM simulation along the [221] zone axis of the layered phase. (f,g) HRTEM image from the core region of primary
NCA particles, enabled with ultra-microtome cutting. (h) HRTEM image of an uncut particle, showing no atomic resolution from the core area. (i) Atomic
STEM-HAADF image of a particle cut with ultra-microtome.

Fig. 6. Sectioning NMC811 with ultra-microtome. (a) Zoom-out TEM view of NMC811 NPs. (b,c) HRTEM images obtained from different depths of the particle shown
in (a), confirming that well-defined atomic resolution can be achieved from the core of the particle.

same. The reduction in the particle concentration during the The influence of the ultra-microtome cutting on the size of the
preparation of electrodes can be attributed to the addition of elec- primary particles is also analyzed, as shown in Figure 8. Figures 8a
trolyte and binder (Nanda et al., 2011; Xiao et al., 2012), as sche- and 8c present representative SEM images of uncut and cut
matically shown in Figure 7h. The almost unchanged particle 1-cycle electrodes, showing primary particles embedded in the
concentration between the uncut and cut electrodes indicates electrolyte and binder. Figure 8a is obtained from the raw elec-
that ultra-microtome sectioning does not generate any observable trode, and Figure 8c is obtained from an epoxy slice sectioned
mechanical defects, such as microcracks and cavities. Otherwise, from the electrode. The morphology and size distribution of the
the particle concentration in the electrode after cutting (Figs. 7c, primary particles in the microtome-sectioned electrode is almost
7f) should be significantly reduced. the same as the uncut electrode. This is supported by comparing

Downloaded from https://www.cambridge.org/core. Cornell University Library, on 01 Sep 2020 at 13:33:26, subject to the Cambridge Core terms of use, available at
https://www.cambridge.org/core/terms. https://doi.org/10.1017/S1431927620024368
Microscopy and Microanalysis 7

Fig. 7. The influence of the ultra-microtome cutting on the porosity of cathode materials. (a–c) SEM images showing the primary particles of the pristine material, a
1-cycle electrode without ultra-microtome cutting, and a 1-cycle electrode after ultra-microtome cutting. (d–f) Binary images converted from the SEM images
shown in (a–c), respectively. The white areas correspond to the cathode material, and the black areas correspond to voids and electrolyte. (g) Histogram showing
the particle concentrations in the three samples shown in (a–c). (h) Schematic showing the cathode particles mixed with electrolyte and binder in the battery
electrode.

the particle size distribution of the two samples. The particle sizes “anti-core-shell” phase transformation (Zhang et al., 2017b) in
were manually measured by excluding aggregated ones in Figures the bulk of cycled NCA particles. In summary, no structural
8a and 8c, and the histograms of the size distributions are shown changes caused by the ultra-microtome sectioning has been
in Figures 8b and 8d. Gauss fitting of the histograms yields the observed at either the atomic or micrometer levels.
mean sizes of ∼58.53 and ∼63.71 nm for the uncut and
microtome-sectioned primary particles, respectively. Their close
Challenges of the Ultra-Microtome Sectioning
match indicates that the ultra-microtome sectioning does not
crush the cathode particles. Otherwise, fragments of the primary Despite the high-quality ultra-microtome specimens with ignorable
particles are supposed to show up, and a large number of “smaller mechanical artifacts, there are also several challenges that should be
particles” can be observed in Figures 8b and 8d. taken into consideration, as presented in Figure 9. The ultra-thin
According to the morphology analysis in Figures 7 and 8, no nature of the epoxy membrane makes it vulnerable to the electron
observable mechanical damage is caused by the ultra-microtome beam irradiation in TEM. The high-energy e-beam along with its
sectioning of the relatively fragile NCA cathode. Additionally, the charging effect can cause vibration of the epoxy membrane, reduc-
observed atomic configuration of the layered phase in the ultra- ing the TEM resolution. To resolve this issue, a carbon coating is
microtome sample shown in Figure 5 is identical to its established mounted on the TEM copper grid to provide a stable support for
structure, meaning that ultra-microtome has not caused any the epoxy slices. The carbon coating is usually ultra-thin (a few
phase transformation. This argument is also supported by our nanometers) or highly porous not to jeopardize the TEM resolution.
previous work, in which TEM observations of uncut and cut sam- Sometimes the high-energy e-beam even partially decomposes
ples reveal the same (003) cracking (Zhang et al., 2017c) and and evaporates the epoxy membrane to generate holes (Fig. 9a,

Downloaded from https://www.cambridge.org/core. Cornell University Library, on 01 Sep 2020 at 13:33:26, subject to the Cambridge Core terms of use, available at
https://www.cambridge.org/core/terms. https://doi.org/10.1017/S1431927620024368
8 H. Zhang et al.

Fig. 8. Influence of the ultra-microtome cutting on the size of the primary particles in the electrode. (a,c) Representative SEM images of the uncut and cut 1-cycle
electrodes, exhibiting similar morphologies of primary particles. (b,d) Histograms showing the size distributions of the primary particles in (a,c), indicating that the
two cathodes have primary particles with almost the same size distribution.

Fig. 9. Challenges of the ultra-microtome method for TEM specimen preparation. (a) Holes generated in the epoxy membrane by the electron beam irradiation. (b)
Carbon contamination on the surface of a 1-cycle NCA particle under STEM-HAADF imaging.

marked with orange arrows). The tension within the membrane organic nature of the epoxy membrane, it can be decomposed
can cause a quick expansion of these holes, which finally destroys by the e-beam irradiation into carbon and hydrocarbon species
the whole membrane. The generation of holes can be attenuated (Luft, 1961; Keller et al., 2004; Monticelli et al., 2007; Griffiths
by reducing the dose-rate of the e-beam and avoiding epoxy & Walther, 2010). When an intense e-beam, such as the STEM
films with the imperfections since the imperfections can be a electron probe, is applied on the cathode particle embedded in
source for the membrane piercing. It is worth mentioning that epoxy, its strong charging effect drives the concentration of the
defective epoxy membrane areas tend to break up under the carbon and hydrocarbon species toward the e-beam, as marked
e-beam irradiation due to stress concentrations at defect sites by the orange arrows in Figure 9b, showing up as bright spots.
such as holes and tears. Defect-free membrane areas are more The carbon and hydrocarbon contamination largely reduces the
robust and preferred for TEM observations. resolution of the STEM/TEM imaging, which is largely deter-
Another challenge of the ultra-microtome method is the car- mined by the stability of the epoxy film. In this work, the
bon and hydrocarbon contamination of specimens. Due to the epoxy bars were treated at 60°C for 24 h for solidification,

Downloaded from https://www.cambridge.org/core. Cornell University Library, on 01 Sep 2020 at 13:33:26, subject to the Cambridge Core terms of use, available at
https://www.cambridge.org/core/terms. https://doi.org/10.1017/S1431927620024368
Microscopy and Microanalysis 9

according to its technical sheet. After treatment, the epoxy bars ultra-microtome cutting (mechanical effect and introduction of
were fully solidified without remaining liquids. hydrocarbon) can be minimized without introducing artifacts to
Carbon contamination during STEM-HAADF imaging can be the microstructural feature of the cathode material while still
minimized by choosing an area with the intact epoxy film and achieving the atomic resolution of TEM imaging. Based on
showering the area with the e-beam. As shown in Figure 5d, these observations, it can be concluded that ultra-microtome is
atomically resolved STEM-HAADF images can be obtained a suitable method to prepare TEM specimens and reveal the
with the e-beam shower for ∼20 min under the STEM-HAADF atomic feature in the NCA cathode.
mode with a current intensity of ∼20 pA/(nm2 s) and an
on-screen magnification of 27,000 times, followed by
STEM-HAADF imaging using the same current intensity. Conclusion
An ultra-microtome-based method for the preparation of TEM
Comparison with Other Sample Preparation Methods specimens is introduced for sectioning particles of the cathode
material of Li-ion batteries. The specimens prepared by the
As discussed in the introduction section, there are multiple sam-
microtome method contains a large number of the sectioned par-
ple preparation methods for reducing the thickness of NPs.
ticles that are sufficiently thin to yield atomic-scale HRTEM and
Compared with the ultra-microtome cutting, these methods
STEM imaging of the microstructural feature. No observable
exhibit both advantages and disadvantages, as discussed in the
structural damages or artifacts, including lattice amorphization,
following.
microcracking, or fragmentation, are introduced from the micro-
Crushing and grinding with a pestle and a mortar is the sim-
tome sectioning. This ultra-microtome-based method can find
plest and cheapest way for reducing the particle size (Kisch, 1991;
broader applicability in the preparation of the TEM specimens
Mori et al., 1999). The particles can be crushed into pieces smaller
from particulate matter of varying size ranges.
than 50 nm, which easily meet the requirement of atomic
HRTEM imaging. However, the micrometer-scale features of the Acknowledgments. This work was supported as part of the NorthEast
particles are damaged during this process. Also, the grinding usu- Center for Chemical Energy Storage (NECCES), an Energy Frontier
ally introduces a high concentration of mechanical artifacts, such Research Center funded by the U.S. Department of Energy, Office of
as vacancies, microcracks, dislocations, and twinning. Therefore, Science, Basic Energy Sciences under Award # DE-SC0012583. The TEM
it can only yield very limited information regarding the target work was conducted in the William R. Wiley Environmental Molecular
Sciences Laboratory (EMSL), a national scientific user facility sponsored by
material.
DOE’s Office of Biological and Environmental Research and located at
Compared with ultra-microtome, FIB can precisely pick a spe-
PNNL. PNNL is operated by Battelle for the Department of Energy under
cific location for TEM sampling, which is very useful for high- Contract DE-AC05-76RLO1830. C.M.W. is supported by the Assistant
accuracy microstructure characterization. The specimen prepared Secretary for Energy Efficiency and Renewable Energy, Office of Vehicle
with FIB is very clean and in the desirable shape, which benefits Technologies of the U.S. Department of Energy under the advanced cathode
the TEM imaging. However, it also has multiple drawbacks com- materials program.
pared with the ultra-microtome method. The largest FIB lift-out
TEM specimen is ∼10 nm in lateral, which only contains a
dozen of NPs of hundreds of nanometers. The ultra-microtome References
specimen, however, can be as large as 1–2 mm in lateral dimen-
Ahn JH & Peacor DR (1986). Transmission electron microscope data for rec-
sions, which thus contains hundreds of thousands of the same torite: Implications for the origin and structure of “fundamental particles”.
NPs. Considering a 5 μm × 5 μm FIB specimen and a 1 mm × Clays Clay Miner 34, 180–186.
0.5 mm microtome specimen, the microtome specimen is Antonovsky A (1995). Ultramicrotomy of industrial alumina. Microsc Res
20,000 times larger than the FIB sample. The large number of Tech 31, 300–307.
cathode particles in the ultra-microtome specimen allows for Barna A, Pécz B & Menyhard M (1999). TEM sample preparation by ion
the statistical analysis of a large collection of particles. milling/amorphization. Micron 30, 267–276.
Ultra-microtome also exhibits other advantages over FIB, such Bashir A, Millar RW, Gallacher K, Paul DJ, Darbal AD, Stroud R, Ballabio
as its improved time-efficiency compared with FIB. The prepara- A, Frigerio J, Isella G & MacLaren I (2019). Strain analysis of a Ge micro
disk using precession electron diffraction. J Appl Phys 126, 235701.
tion of one TEM sample with ultra-microtome usually takes less
Becker O & Bange K (1993). Ultramicrotomy: An alternative cross section
than 20 min, while it usually takes more than 5 h for FIB. Also,
preparation for oxidic thin films on glass. Ultramicroscopy 52, 73–
the ultra-microtome method can potentially cut relatively large 84. http://www.sciencedirect.com/science/article/pii/030439919390023Q.
samples up to 1 mm, while FIB is not suitable for this purpose. Boesenberg U, Sokaras D, Nordlund D, Weng T-C, Gorelov E, Richardson
On the other hand, the ultra-microtome method also has TJ, Kostecki R & Cabana J (2019). Electronic structure changes upon lith-
drawbacks such as the potential mechanical damage on the sam- ium intercalation into graphite–Insights from ex situ and operando x-ray
ple, as well as the hydrocarbon/carbon contamination due to the Raman spectroscopy. Carbon 143, 371–377.
use of epoxy. The high-energy ion beam of FIB largely removes Brown AP, Brydson RMD & Hondow NS (2014). Measuring in vitro cellular
any potential hydrocarbon/carbon contamination, so the FIB uptake of nanoparticles by transmission electron microscopy. J Phys Conf
sample is usually very clean. The hydrocarbon contamination of Ser 522, 12058.
the ultra-microtome sample cannot be removed, meaning that it Burghardt RC & Droleskey R (2006). Transmission electron microscopy.
Curr Protoc Microbiol 3, 2B.1.1.
may not be suitable for in situ TEM experiments.
Buseck P, Cowley J & Eyring L (1989). High-Resolution Transmission Electron
Considering the advantages and drawbacks of different TEM Microscopy and Associated Techniques. Oxford University Press.
specimen preparation methods, it is important to select the Dai JY, Tee SF, Tay CL, Song ZG, Ansari S, Er E & Redkar S (2001).
appropriate method depending on the specific conditions of the Development of a rapid and automated TEM sample preparation method
sample and the observation requirements. In Figures 4 to 7, we in semiconductor failure analysis and the study of the relevant TEM artifact.
have demonstrated that the major potential drawbacks of the Microelectronics J 32, 221–226.

Downloaded from https://www.cambridge.org/core. Cornell University Library, on 01 Sep 2020 at 13:33:26, subject to the Cambridge Core terms of use, available at
https://www.cambridge.org/core/terms. https://doi.org/10.1017/S1431927620024368
10 H. Zhang et al.

Flom Y & Arsenault RJ (1989). Effect of particle size on fracture toughness of Luft JH (1961). Improvements in epoxy resin embedding methods. J Cell Biol
SiC/Al composite material. Acta Metall 37, 2413–2423. 9, 409–414.
Giannuzzi LA & Stevie FA (1999). A review of focused ion beam milling tech- Malis TF & Steele D (1990). Ultramicrotomy for materials science. MRS
niques for TEM specimen preparation. Micron 30, 197–204. Proceedings 199, 3. https://www.cambridge.org/core/article/ultramicrot-
Glanvill SR (1995). Ultramicrotomy of semiconductors and related materials. omy-for-materials-science/1B9BFDF4FD1AFCBE26B6460374053D48.
Microsc Res Tech 31, 275–284. May BM, Serrano-Sevillano J, Dauphin AL, Nazib A, Lima N,
Graf CM, Meinke MC, Gao Q, Hadam S, Raabe J, Sterry W, Blume-Peytavi Casas-Cabanas M & Cabana J (2018). Effect of synthetic parameters on
U, Lademann JM, Ruhl E & Vogt A (2009). Qualitative detection of single defects, structure, and electrochemical properties of layered oxide LiNi0.
submicron and nanoparticles in human skin by scanning transmission 80Co0. 15Al0. 05O2. J Electrochem Soc 165, A3537–A3543.
X-ray microscopy. J Biomed Opt 14, 21015. Mayer J, Giannuzzi LA, Kamino T & Michael J (2007). TEM sample prepa-
Griffiths AJV & Walther T (2010). Quantification of carbon contamination ration and FIB-induced damage. MRS Bull 32, 400–407.
under electron beam irradiation in a scanning transmission electron micro- McCaffrey JP, Phaneuf MW & Madsen LD (2001). Surface damage formation
scope and its suppression by plasma cleaning. J Phys Conf Ser 241, 12017. during ion-beam thinning of samples for transmission electron microscopy.
Harris KM, Perry E, Bourne J, Feinberg M, Ostroff L & Hurlburt J (2006). Ultramicroscopy 87, 97–104.
Uniform serial sectioning for transmission electron microscopy. J Neurosci Mcmahon G & Malis T (1995). Ultramicrotomy of nanocrystalline materials.
26, 12101–12103. Microsc Res Tech 31, 267–274.
Haswell R, McComb DW & Smith W (2003). Preparation of site-specific Messenger S, Keller LP, Stadermann FJ, Walker RM & Zinner E (2003).
cross-sections of heterogeneous catalysts prepared by focused ion beam Samples of stars beyond the solar system: Silicate grains in interplanetary
milling. J Microsc 211, 161–166. dust. Science 300, 105–108.
He J, Zhao Y, Wang Y, Wang J, Zheng J, Zhang H, Zhou G, Wang C, Wang Monticelli O, Musina Z, Russo S & Bals S (2007). On the use of TEM in the
S & Ma X (2017). A Fe5C2 nanocatalyst for the preferential synthesis of eth- characterization of nanocomposites. Mater Lett 61, 3446–3450.
anol via dimethyl oxalate hydrogenation. Chem Commun 53, 5376–5379. Mori D, Kobayashi H, Shikano M, Nitani H, Kageyama H, Koike S, Sakaebe
Hochella MF Jr, Moore JN, Golla U & Putnis A (1999). A TEM study of sam- H & Tatsumi K (2009). Bulk and surface structure investigation for the
ples from acid mine drainage systems: Metal-mineral association with positive electrodes of degraded lithium-ion cell after storage test using
implications for transport. Geochim Cosmochim Acta 63, 3395–3406. X-ray absorption near-edge structure measurement. J Power Sources 189,
Hughes AE, Trinchi A, Chen FF, Yang YS, Cole IS, Sellaiyan S, Carr J, Lee 676–680.
PD, Thompson GE & Xiao TQ (2014). The application of multiscale quasi Mori H, Miyoshi H, Takeda K, Yoneyama H, Fujita H, Iwata Y, Otsuka Y &
4D CT to the study of SrCrO4 distributions and the development of porous Murata Y (1992). TEM observation of iron oxide pillars in montmorillon-
networks in epoxy-based primer coatings. Prog Org Coat 77, 1946–1956. ite. J Mater Sci 27, 3197–3199.
Hwang S, Chang W, Kim SM, Su D, Kim DH, Lee JY, Chung KY & Stach Mori H, Sawada M, Higuchi T, Hasebe K, Otsuka N & Terano M (1999).
EA (2014). Investigation of changes in the surface structure of LixNi0. 8Co0. Direct observation of MgCl2-supported Ziegler catalysts by high resolution
15Al0. 05O2 cathode materials induced by the initial charge. Chem Mater 26, transmission electron microscopy. Macromol Rapid Commun 20, 245–250.
1084–1092. Nanda J, Remillard J, O’Neill A, Bernardi D, Ro T, Nietering KE, Go J &
Ishitani T & Yaguchi T (1996). Cross-sectional sample preparation by focused Miller TJ (2011). Local state-of-charge mapping of lithium-ion battery elec-
ion beam: A review of ion-sample interaction. Microsc Res Tech 35, 320–333. trodes. Adv Funct Mater 21, 3282–3290.
Karki K, Huang Y, Hwang S, Gamalski AD, Whittingham MS, Zhou G & Puttick KE, Whitmore LC, Chao CL & Gee AE (1994). Transmission electron
Stach EA (2016). Tuning the activity of oxygen in LiNi0. 8Co0. 15Al0. 05O2 microscopy of nanomachined silicon crystals. Philos Mag A 69, 91–103.
battery electrodes. ACS Appl Mater Interfaces 8, 27762–27771. Quintana C (1997). Ultramicrotomy for cross-sections of nanostructure.
Keller LP, Messenger S, Flynn GJ, Clemett S, Wirick S & Jacobsen C (2004). Micron 28, 217–219. http://www.sciencedirect.com/science/article/pii/
The nature of molecular cloud material in interplanetary dust. Geochim S0968432897000231.
Cosmochim Acta 68, 2577–2589. Reymond OL & Pickett-Heaps JD (1983). A routine flat embedding method
Kisch HJ (1991). Illite crystallinity: Recommendations on sample preparation, for electron microscopy of microorganisms allowing selection and precisely
X-ray diffraction settings, and interlaboratory samples. J Metamorph Geol 9, orientated sectioning of single cells by light microscopy. J Microsc 130, 79–84.
665–670. Schaffer M, Pfeffer S, Mahamid J, Kleindiek S, Laugks T, Albert S, Engel
Klepeis SJ, Benedict JP & Anderson RM (1987). A grinding/polishing tool for BD, Rummel A, Smith AJ, Baumeister W & Plitzko JM (2019). A
TEM sample preparation. MRS Online Proceedings Library Archive 115, 179. cryo-FIB lift-out technique enables molecular-resolution cryo-ET within
Kumlutaş D, Tavman IH & Çoban MT (2003). Thermal conductivity of native Caenorhabditis elegans tissue. Nat Methods 16, 757–762.
particle filled polyethylene composite materials. Compos Sci Technol 63, Schaffer M, Schaffer B & Ramasse Q (2012). Sample preparation for atomic-
113–117. resolution STEM at low voltages by FIB. Ultramicroscopy 114, 62–71.
Lezec HJ, Musil CR, Melngailis J, Mahoney LJ & Woodhouse JD (1991). Schertel A, Snaidero N, Han H-M, Ruhwedel T, Laue M, Grabenbauer M &
Dose-rate effects in focused-ion-beam implantation of Si into GaAs. J Möbius W (2013). Cryo FIB-SEM: Volume imaging of cellular ultrastruc-
Vac Sci Technol B Microelectron Nanometer Struct Process Meas Phenom ture in native frozen specimens. J Struct Biol 184, 355–360. http://www.sci-
9, 2709–2713. encedirect.com/science/article/pii/S1047847713002657.
Li J, Malis T & Dionne S (2006). Recent advances in FIB–TEM specimen Schrand AM, Schlager JJ, Dai L & Hussain SM (2010). Preparation of cells
preparation techniques. Mater Charact 57, 64–70. http://www.sciencedir- for assessing ultrastructural localization of nanoparticles with transmission
ect.com/science/article/pii/S1044580305002780. electron microscopy. Nat Protoc 5, 744.
Lin F, Markus IM, Nordlund D, Weng T-C, Asta MD, Xin HL & Doeff MM Siu C, Seymour I, Britto S, Zhang H, Rana J, Feng J, Omenya FO, Zhou H,
(2014a). Surface reconstruction and chemical evolution of stoichiometric Chernova N & Zhou G (2018). Enabling multi-electron reaction of
layered cathode materials for lithium-ion batteries. Nat Commun 5, 3529. ε-VOPO4 to reach theoretical capacity for lithium-ion batteries. Chem
Lin F, Nordlund D, Markus IM, Weng T-C, Xin HL & Doeff MM (2014b). Commun 54, 7802–7805.
Profiling the nanoscale gradient in stoichiometric layered cathode particles Stamenkovic VR, Strmcnik D, Lopes PP & Markovic NM (2017). Energy and
for lithium-ion batteries. Energy Environ Sci 7, 3077–3085. fuels from electrochemical interfaces. Nat Mater 16, 57–69.
Liu Y, Wang R, Guo X & Dai J (2007). A cross-sectional TEM sample prep- Sun BB, Wang YB, Wen J, Yang H, Sui ML, Wang JQ & Ma E (2005).
aration method for films deposited on metallic substrates. Mater Charact Artifacts induced in metallic glasses during TEM sample preparation. Scr
58, 666–669. Mater 53, 805–809.
Lu L, Han X, Li J, Hua J & Ouyang M (2013). A review on the key issues for Sun F, Li H, Leifer K & Gamstedt EK (2017). Rate effects on localized shear
lithium-ion battery management in electric vehicles. J Power Sources 226, deformation during nanosectioning of an amorphous thermoplastic poly-
272–288. mer. Int J Solids Struct 129, 40–48.

Downloaded from https://www.cambridge.org/core. Cornell University Library, on 01 Sep 2020 at 13:33:26, subject to the Cambridge Core terms of use, available at
https://www.cambridge.org/core/terms. https://doi.org/10.1017/S1431927620024368
Microscopy and Microanalysis 11

Tucker DS, Jenkins EJ & Hren JJ (1985). Sectioning spherical aluminum Zachman MJ, Tu Z, Choudhury S, Archer LA & Kourkoutis LF (2018).
oxide particles for transmission electron microscopy. J Electron Microsc Cryo-STEM mapping of solid–liquid interfaces and dendrites in lithium-
Tech 2, 29–33. metal batteries. Nature 560, 345–349.
Volkert CA & Minor AM (2007). Focused ion beam microscopy and micro- Zhang B & Howes TD (1994). Material-removal mechanisms in grinding
machining. MRS Bull 32, 389–399. ceramics. CIRP Ann 43, 305–308.
Wang K, Strunk K, Zhao G, Gray JL & Zhang P (2012). 3D structure deter- Zhang H, Karki K, Huang Y, Whittingham MS, Stach EA & Zhou G
mination of native mammalian cells using cryo-FIB and cryo-electron (2017a). Atomic insight into the layered/spinel phase transformation in
tomography. J Struct Biol 180, 318–326. http://www.sciencedirect.com/sci- charged LiNi0. 80Co0. 15Al0. 05O2 cathode particles. J Phys Chem C 121,
ence/article/pii/S1047847712001955. 1421–1430.
Williams DB & Carter CB (2009). Transmission Electron Microscopy: A Zhang H, May BM, Serrano-Sevillano J, Casas-Cabanas M, Cabana J, Wang
Textbook for Materials Science. New York: Springer. C & Zhou G (2018). Facet-dependent rock-salt reconstruction on the sur-
Xiao J, Chen X, Sushko PV, Sushko ML, Kovarik L, Feng J, Deng Z, Zheng face of layered oxide cathodes. Chem Mater 30, 692–699.
J, Graff GL & Nie Z (2012). High-performance LiNi0. 5Mn1. 5O4 spinel Zhang H, Omenya F, Whittingham MS, Wang C & Zhou G (2017b).
controlled by Mn3+ concentration and site disorder. Adv Mater 24, 2109– Formation of an anti-core–shell structure in layered oxide cathodes for
2116. Li-ion batteries. ACS Energy Lett 2, 2598–2606.
Yan P, Zheng J, Gu M, Xiao J, Zhang J-G & Wang C-M (2017). Intragranular Zhang H, Omenya F, Yan P, Luo L, Whittingham MS, Wang C & Zhou G
cracking as a critical barrier for high-voltage usage of layer-structured cath- (2017c). Rock-salt growth-induced (003) cracking in a layered positive elec-
ode for lithium-ion batteries. Nat Commun 8, 14101. trode for Li-ion batteries. ACS Energy Lett 2, 2607–2615.
Yang H-S, Kim D-J & Kim H-J (2003). Rice straw–wood particle composite Zhang J, Ji G, Huang X, Xu W & Sun F (2016). An improved cryo-FIB
for sound absorbing wooden construction materials. Bioresour Technol 86, method for fabrication of frozen hydrated lamella. J Struct Biol 194, 218–
117–121. 223. http://www.sciencedirect.com/science/article/pii/S1047847716300284.
Yoshida J, Nakanishi S, Iba H, Abe H & Naito M (2013). Thermal behavior Zhang Z, Wu Y, Guo D & Huang H (2011). Phase transformation of single
of delithiated Li1−xMnPO4 (0≤ x<1) structure for lithium-ion batteries. Int J crystal silicon induced by grinding with ultrafine diamond grits. Scr Mater
Appl Ceram Technol 10, 764–772. 64, 177–180.
Zachman MJ, Tu Z, Archer LA & Kourkoutis LF (2020). Nanoscale elemen- Zhao H, Qiu B, Guo H, Jia K, Liu Z & Xia Y (2017). Characterization of
tal mapping of intact solid–liquid interfaces and reactive materials in energy Li-rich layered oxides by using transmission electron microscope. Green
devices enabled by cryo-FIB/SEM. ACS Energy Lett 5, 1224–1232. Energy Environ 2, 174–185.

Downloaded from https://www.cambridge.org/core. Cornell University Library, on 01 Sep 2020 at 13:33:26, subject to the Cambridge Core terms of use, available at
https://www.cambridge.org/core/terms. https://doi.org/10.1017/S1431927620024368

You might also like