////////////// Technical Information (General Radiography / XTV) ///////////////

SVHI-E15D007, SVHI-E15E007 (1/4)


TO: Engineers June 29, 2015
Issued by Service Management Dept.
Mandatory Signature

Masakado, Matsuo
Change reliability Safety
Manager, Service Management Dept.
Information
Many thanks for your kind support.
Please take note of the following information.
Affected Models All models Some models
UD150B-40 Series (except for G-40), SONIALVISION G4

Subject
Replacement of UD CONT-2002 with UD CONT-2002X

1. Outline
The UD CONT-2002 control PCB (P/N: 502-22500, 562-22500) used on the UD150B-40 Series
(except for G-40) and SONIALVISION G4 units will be discontinued and will be replaced with the
new PCB, UD CONT-2002X. This bulletin provides information related to this change.
The new PCB will be used on products shipped from mid May 2015 and after.
Note that Production of UD CONT-2002 (P/N: 502-22500-02, 562-22500-02) for the UD150G-40 will
be continued. This Technical Information does not apply to them.

2. Description
 Applicable Boards
Old 502-22500 / 562-22500 UD CONT-2002 ASSY

New 565-11900 UD CONT-2002X ASSY

UD CONT-2002 UD CONT-2002X

/////////////////////////// Service Management Department ///////////////////////////


////////////// Technical Information (General Radiography / XTV) ///////////////
SVHI-E15D007, SVHI-E15E007 (2/4)

 Technical Information on the UD CONT-2002X board


There is no change in the board form, position of connectors/LEDs/check pins, reference number,
or name. Although adjustment procedures are also very similar, the following points are different.
(a) Settings for SW 4-2 and 4-7 dip switches have been partially changed and JP4 jumper
switch has been added.
Dip switches
Normal
SW No. Remarks
State
ON: For factory adjustment (writing FLASH in the CPU)
SW 4-1 OFF
OFF: Normal state
ON: For factory adjustment (writing FLASH in the CPU)
SW 4-2 OFF
OFF: Normal state
ON: Rated: Type L-40
SW 4-3 OFF
OFF: Rated: Type B-40/V-40
ON: During checking operation
SW 4-4 OFF
OFF: Normal state
ON: For L-40E/F (when GSC-2002S is connected)
SW 4-5 OFF
OFF: For B-40/V-40/L-40 (when GSC-2002L is connected)
ON: General radiography system (with one mA Power-2002
SW 4-6 OFF board)
OFF: Fluoroscopy system (with two mA Power-2002 boards)
ON: During forced F/W writing
SW 4-7 OFF
OFF: Normal state
ON: Complete initialization of the memory inside the board
SW 4-8 OFF
OFF: Normal state

Jumper pin
Normal
JP No. Remarks
State
+5 V power is supplied from J15 (D-sub 9-pin
JP4 1-2 connector).
[ISOLATION PW] [NC] 1-2: J15 pin 9 is open.
2-3: +5 V is output from J15 pin 9.

(b) The logic signal level has been changed to 3.3 V. Example: CP27 [EXRR]
(c) Flash waveform has been changed to rectangular waveform according to the change in
the method to control filament heating.
(Flash calibration for the XTV unit is unnecessary.)
(d) Serial port settings for terminal communication have been changed for higher-speed
communication.
Setting UD CONT-2002 UD CONT-2002X
Baud rate 9600 38400
Data 8 bit 8 bit
Parity odd none
Stop 2 bit 2 bit
Flow control none none

Notes:
 Serial port settings for the remote maintenance PC have also been changed.
 There is no change in communication with Fuji Film and Canon CR/FPD units.

/////////////////////////// Service Management Department ///////////////////////////


////////////// Technical Information (General Radiography / XTV) ///////////////
SVHI-E15D007, SVHI-E15E007 (3/4)
(e) Some terminal commands have been added/deleted.
(f) Procedure to update software has been changed to use a special application instead of
TeraTerm.
For the new board, different types of files will be written. (Extension: *.mot)
For updating software, both "*.bin" files for the UD CONT-2002 board and "*.mot" files for
the UD CONT-2002X board will be supplied. Use the file according to the board used.

For details, refer to Section 1.1 Control Board UD CONT-2002X in the corresponding
installation manual.

3. Supported Software
The table below shows the revision numbers of software and documents that support the new UD
CONT-2002X board.
Software and Documents are available on SUMITS.

Main unit software


Supported Revision No.
Instrument
UD CONT-2002X GSC-2002L/S to be used with
UD150B/V/L-40
2.662 and later 2.661 and later
(other than SONIALVISION safire II/17)
2.496 and later
UD150B/V/L-40
(DAR-8000 System Rev.04.30.13A 2.496 and later
(SONIALVISION safire II/17)
and later)
(X)UD150L-40E/F 2.094 and later 2.093 and later
7.00 and later
SONIALVISION G4 (G4 System Rev.01.04.04 and 7.00 and later
later)

Related tools
Tool Supported Revision No.
APR Import/Export Tool 3.1.0.0 and later
UD Maintenance Tool
3.0.0.1 and later
(for SONIALVISION G4)
UD Maintenance Tool
3.0.0.1 and later
(for RADspeed Pro)

/////////////////////////// Service Management Department ///////////////////////////


////////////// Technical Information (General Radiography / XTV) ///////////////
SVHI-E15D007, SVHI-E15E007 (4/4)
4. Related Documents
Supported
Document No. Title
Revision
M501-E356 UD150B-40 UD150V-40 UD150L-40 AD and later
D150BC-40/BC-45/VC-40/VC-45/LC-40/LC-45, GSC-2002L Installation
Manual
M501-E362 UD150L-40E/F XUD150L-40F D150LC-40E/-40F/-45E/-45F K and later
D150LC-45EX/-45FX, GSC-2002S Installation Manual
M501-4017 UD150B-40 Series Parts List J and later
M501-E401 D150BC-40S / D150VC-40S Installation Manual G and later
M501-4030 D150BC-40S / D150VC-40S Parts List C and later
M517-E429 DAR-8000 Remote Maintenance Installation Manual N and later
M517-E235 SONIALVISION G4 (DR300) Remote Maintenance Installation Manual C and later
M517-E483 FPD17-R1 Installation Manual R and later

5. When Replacing the Board on Existing Units


When replacing the UD CONT-2002 board on an existing instrument used at a customer, order the
new UD CONT-2002X board from now on.
Cautions:
 The UD CONT-2002X board cannot be used on the following instruments. Order the UD
CONT-2002 board for them.
 UD150B-40 for Heart Speed-10
 After replacing the board, readjust the tube voltage (adjustment of FVR and FVF).
 Backup data can be transferred to the new board. However, the Flash Volume value is
initialized to 100 (default of the new board) when the system is turned ON for the first time
after board replacement. This occurs due to a difference in the control method. Do not
return the value to the value for the previous board since there is no problem if the Flash
Volume value is at the default of 100. This is the same when a Q stator tube or JG tube is
used.
 Backup data can be used even if the new board is replaced with the previous board for
some reason. In this case, return the Flash Volume value to the value for the previous
board (standard value: approx. 40 to 60) after restoring data. For XTV units, perform flash
calibration as well.
 After replacement between the previous and new boards, TeraTerm serial port settings on
the remote maintenance PC also need to be changed.
 Configure settings referring to the remote maintenance installation manual (for the
DAR-7500, FPD17-R1 installation manual) and confirm that UD logs are correctly
collected.