Professional Documents
Culture Documents
(3/2010)
Service Manual
[05]
Liability Disclaimer
bioMérieux, Inc. makes no express or implied warranty regarding this manual, its
quality, performance, or appropriate use regarding any type of specific procedure.
Furthermore, this manual may be modified by bioMérieux without notice and without
implying any obligation or liability on the part of the company.
Intellectual Property
bioMérieux, the blue logo, BacT/ALERT, BacT/LINK, and FAN are used, pending,
and/or registered trademarks of bioMérieux, SA or one of its subsidiaries.
Intended Audience
Standard Symbols
The following table presents symbols that may appear in the instructions for use or on
the instrument, package inserts, or packaging.
CE-Marking of Conformity
Use by
Manufacturer
Date of manufacture
Keep dry
Biological risks
Radiation warning
i
Standard Symbols
Laser
Temperature limitation
Batch code
Catalog number
Serial Number
Do not reuse
Recyclable
Very toxic
Corrosive
Sodium azide
ii
Standard Symbols
Irritant
Positive control
Negative control
This way up
Do not stack
Humidity limitation
Fuse
Direct current
Alternating current
iii
Standard Symbols
Equipotentiality
ON (supply)
OFF (supply)
iv
Table of Contents
Standard Symbols................................................................................................ i
A Specifications .............................................................................................A-1
A.1 Physical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A.2 Environmental Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A.3 Electrical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Figure 7-1: Scanner Default Settings Barcode Label for Older Barcode Scanner .......... 7-16
Figure 7-2: Scanner Default Settings Barcode Label for Newer Barcode Scanner ........ 7-16
Figure 7-3: Instrument Icon with Fault Codes ................................................................. 7-17
Figure 7-4: Instrument Status Code #710 ....................................................................... 7-17
Figure 7-5: Main Screen ................................................................................................. 7-52
Figure 7-6: Setup Screen with Field Engineer Password Activation ............................... 7-53
Figure 7-7: Diagnostic Test Selection (Setup) Screen .................................................... 7-53
Figure 7-8: Inc Module/Loopback Screen ....................................................................... 7-54
Figure 7-9: Inc Module/Port Stats Screen ....................................................................... 7-55
Figure 7-10: Inc Module/Message Stats Screen ............................................................... 7-56
Figure 7-11: Inc Module/Message Log Screen ................................................................. 7-57
Figure 7-12: Inc Module/Temperature Log Screen ........................................................... 7-58
Figure 7-13: Inc Module/Reading Log Screen .................................................................. 7-59
Figure 7-14: Inc Module/Flag Check Screen .................................................................... 7-60
Figure 7-15: Cell Calibration Screen ................................................................................. 7-61
Figure 7-16: Run Continuously Button .............................................................................. 7-61
Figure 7-17: Inc Module/Drawer/Heater Screen ............................................................... 7-62
Figure 7-18: Inc Module/LED Check ................................................................................. 7-63
Figure 7-19: Inc Module/LED Check (Walking) ................................................................. 7-64
Figure 7-20: Instrument Cntrl/Zip Drive Test Screen ........................................................ 7-65
Figure 7-21: Barcode Scanner Connection ....................................................................... 7-66
Figure 7-22: Instrument Cntrl/Barcode Loopback Screen ................................................. 7-67
Figure 7-23: Instrument Cntrl/Power Event Log Screen ................................................... 7-68
Figure 7-24: FILE : EVENT1.LOG and EVENT0.LOG ...................................................... 7-70
Figure 7-25: Instrument Cntrl/Software Screen ................................................................ 7-76
Figure 7-26: LIS/Loopback Screen ................................................................................... 7-78
Figure 7-27: LIS/Port Stats Screen ................................................................................... 7-79
Figure 7-28: LIS/Message Stats Screen ........................................................................... 7-80
Figure 7-29: LIS/Message Log Screen ............................................................................. 7-81
Figure D-1: CPU PCBA .....................................................................................................D-2
Figure D-2: CPU PCBA (Front View) .................................................................................D-3
Figure D-3: CPU PCBA (Back View) .................................................................................D-4
Figure D-4: ModSig PCBA Connector Locations ...............................................................D-6
Figure D-5: ModSig PCBA Fuse Locations .......................................................................D-8
Figure D-6: Step-5 PCBA Connector Locations ................................................................D-9
Figure D-7: Outlet Thermistor PCBA Connector Location ...............................................D-10
Figure D-8: Inlet Thermistor PCBA Connector Location ..................................................D-10
Figure D-9: Terminal Board 1 Wiring ...............................................................................D-11
This chapter contains information about how this manual is organized and how to use
this manual. It is recommended that you read this chapter first.
IMPORTANT: Read this manual carefully before you attempt to work on or use the
BacT/ALERT® 3D 60 instrument.
Chapter Contents
IMPORTANT: Figures, schematics and their components are not necessarily shown to scale
in this manual. Figures and schematics are included in this manual for
illustrative purposes or as examples.
This manual is for use by bioMérieux Field Service Engineers and factory-trained
Biomedical Engineers to assist in the troubleshooting and repair of the
BacT/ALERT® 3D 60 system. Every effort has been made to ensure that the
information contained in this manual is complete and accurate. All calibration and
adjustment procedures used in this manual are adapted directly from current
engineering documentation.
The information in this manual is organized into seven chapters, arranged so that the
information you need first is in the beginning of the manual. Information required for
maintenance after the system is operating is found in later chapters.
All of the chapters are organized in the same way and include the following:
• About this Chapter — Summary of the chapter’s content and purpose.
• Chapter Contents — A table of contents for the chapter.
• Descriptions and/or Procedures — Descriptions and procedures appropriate for
this chapter’s subject matter.
• Background Information, where applicable and useful.
This manual uses several methods to help you find information and keep your
bearings:
Table of Contents — Located at the front of the manual. It contains the titles of all
chapters/appendices and their sections, and the page number of each title and
section.
List of Figures — Located at the front of the manual. It contains a list of all figures in
the manual and the page number of each figure.
List of Tables — Located at the front of the manual. It contains a list of all tables in
the manual and the page number of each table.
Chapter Contents — Located at the front of each chapter. It lists all sections in the
chapter and their page number.
Page Headers — Located at the top of each page. There are two parts to a header:
the chapter title and the primary section title.
Page Footers — Located at the bottom of each page. There are three parts to a
footer: the manual name, the manual part number, and the page number.
Index — Located at the back of the manual. It contains topical entries and their page
numbers.
1.6.1 Click
This manual uses the word “click” to refer to using a mouse to choose or select a
command, window, button, or option.
1.6.2 Commands
Menu, keyboard and button commands are in Proper Case, bold.
Example: Enterobacteriaceae
• An instruction to use one of these keys begins with the word “Press.”
• Function keys that you must press simultaneously are shown together like this:
Shift + Tab
The term “right-click” refers to pressing the right mouse button one time to perform
these actions. “Double-click” refers to using two clicks in rapid succession to select a
command, window, button, or option.
The names of windows are in Proper Case, but are not bolded.
1.6.7 Press
This manual uses the word “press” to refer to pressing a key on the keyboard in order
to initiate action in the software. “Press” is also used to refer to pressing switches and
buttons on the instrument or other hardware components.
Note: The “press Enter” command is often not stated explicitly in software procedures,
because it is commonly understood that you press this key after you enter data into
the system using a keyboard.
1.6.9 References
References to chapter and section titles in this manual are in Proper Case.
Quick access to references are available if you are viewing this document
electronically in a portable document format (PDF). References may appear with a
blue font, indicating that they are linked to another step, section, procedure, or task in
this manual. Clicking the reference will take you directly to the location of the
reference in this manual. The Table of Contents, List of Figures, List of Tables,
Chapter Contents, and Index are all linked, but do not appear with a blue font.
1.6.10 Select
The word “select” is generally used for selecting menus, menu commands, and
graphical user interface (GUI) navigation.
In the Literal User Input example, you are to type exactly what you see on the page
(April 1 in this example).
This manual uses different types of symbols to alert you to important information.
Symbols and their associated information are labeled in text where they occur and set
off from surrounding paragraphs, as shown in the following examples.
WARNING
Warning is a statement that alerts the user to the possibility of injury,
death, or other serious adverse reactions associated with the use or
misuse of a device.
IMPORTANT: Important relates to content presented in this manual. It is used to reinforce the
importance of your understanding or remembering something.
The BacT/ALERT® 3D 60 has been designed and tested pursuant to the standards
listed below and has been supplied in a safe condition. A CB Certification and
Construction File has been established for the apparatus.
• IEC Publication 61010-1: 2001 Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use — Part 1: General Requirements
• UL 61010A-1: 2002 UL Standard for Safety Equipment for Laboratory Use —
Part 1: General Requirements
• CAN/CSA C22.2 No. 1010.1-92: Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use — Part 1: General Requirements
1. Before applying power to any module, ensure that the Power Entry Module
(PEM) is properly configured for the specific facility power voltage. Also ensure
that fuse(s) of proper rating value are installed.
2. Any interruption of the protective conductor inside or outside the instrument or
disconnection of the protective earth terminal is likely to make the instrument
dangerous to lab personnel and other users.
3. The instrument must be disconnected from all voltage sources before it is
opened for any adjustments, replacement, maintenance, or repair.
4. Any adjustment, maintenance, or repair of the opened instrument while under
voltage should be avoided. However, if this is unavoidable, it shall be carried out
only by a trained and qualified individual who is aware of the hazard(s) involved
with the task. In addition, a second individual must be nearby to render aid if
necessary.
5. Make sure that only fuses with the required rated current and of the specified
type are used for a replacement. The use of makeshift fuses or the short-
circuiting of any fuse holder is extremely dangerous and prohibited.
6. Whenever it is likely that a BacT/ALERT® 3D module has been impaired, it
should be made inoperative by powering it down and disconnecting the power
cord. If there is evidence of moisture within the instrument, turn the facility
power off at the circuit breaker junction box before removing the power cable.
The instrument is likely to be impaired if it:
• Shows visible damage.
• Fails to perform an intended operation.
• Has been subjected to storage or operation under unfavorable conditions.
(i.e.: above 80% humidity, dusty environment, prolonged storage, see
Appendix A, Specifications for tolerances.)
• Has been subject to severe transport stresses.
7. If any of the above conditions occur, a qualified service technician must check
the instrument prior to use.
Before installing the instrument, ensure that a grounded wall receptacle is available
for each module. It must be plugged into a mating grounding type wall receptacle in
accordance with the National Electrical Code and applicable local codes and
ordinances for this type of installation. (Refer to Figure 1-2.)
WARNING
Under no circumstances is the ground prong to be removed. If it, the
plug, or the cord has any physical damage, the cord is to be replaced.
It is extremeley hazardous to operate this (or any) instrument with a
missing ground prong or a faulty power cable in place.
There are no user serviceable fuses in the BacT/ALERT® 3D 60. Only trained and
qualified service technicians are to replace any fuse within the instrument.
With the exception of the internal circuitry in the display, the DC voltages used within
the modules are low voltage and low current. They do not pose any immediate hazard
to technicians; however, facility power of either 115 VAC or 230 VAC is present while
the instrument power switch is in the energized position, and could pose an electrical
hazard if proper precautions are not observed.
WARNING
Do not insert any object, other than a Zip® disk, into the Zip® drive
under any circumstances.
Specimens and inoculated culture bottles are to be treated as though they are
capable of transmitting infectious agents. Handling of any bottles or specimens must
be done with the utmost care. Spills must be cleaned as soon as possible using the
disinfection procedures outlined in the current version of the BacT/ALERT® 3D 60
User Manual and Appendix B of this manual.
The BacT/ALERT® 3D 60 contains moving parts within the left side of the instrument's
chassis. At times, personnel may be exposed to the Agitation Drive Arm, Motor Offset
Round and other internal moving parts. Care must be taken to avoid exposure of
clothing, hands, fingers, arms, etc. Prior to performing maintenance procedures
around moving parts, it is advised that loose clothing (i.e. ties, long sleeves, etc.) be
secured.
The instrument contains a heater to perform the incubation process. The heater may
remain hot for a period of time after the unit is first shut down for maintenance. To
prevent burns, avoid contact with the heater until it is known that it is cool enough to
touch.
The size and measuring standard used for existing hardware (for example, screws,
nuts, and lock washers) may vary on the BacT/ALERT® 3D 60 due to configuration
changes or system setup. To ensure the integrity of the system, it is extremely
important to use care when retaining existing hardware for later use so that the correct
hardware is used at all times.
If a new component is being installed, be sure to use the new hardware or cabling
supplied with the new component and discard the hardware and cabling from the
defective component.
This chapter provides the Field Service Engineer with instructions on how to install the
BacT/ALERT® 3D 60 instrument, set up and configure the instrument, and perform
functional testing.
Chapter Contents
2.2 Requirements
After the BacT/ALERT® 3D 60 instrument is installed, the customer has two options
for handling the bottles in the BacT/ALERT® Classic instrument:
• Allow the BacT/ALERT® Classic instrument to remain operational until the existing
bottles have either turned positive, or have timed out negative.
or
2.3 Preparation
Item Quantity
1/4 in. x 20 x 5 in. Bolt 4 each
1 — Threaded holes are accessible after the four rubber feet are
removed.
1. Determine the desired location of the instrument on the countertop or mounting
surface.
Note: Care should be taken to avoid braces or bulkheads within the counter which could
interfere with hardware installation and mounting. Drilling into these braces may
reduce the integrity of the counter’s strength.
2. Determine the length of the bolts needed for mounting. (Refer to Figure 2-2.)
¼ in. x 20
A + B = C (Calculated Length)
A = 2 ¾ in. (Equipment Interior Dimension)
B = Thickness of your mounting surface
For mounting surfaces up to 2 1/4 in. thick, use the 5 in. bolts supplied
with the instrument. For mounting surfaces or surfaces requiring longer
bolts, use Table 2-2 and Figure 2-2 to determine the required bolt length.
Equivalent Dimensions
1/2 in. = 1.3 cm 4 in. = 10.2 cm
3. Transfer the dimensions from the Mounting Surface (Drilling) Template to the
actual mounting surface. (Refer to Figure 2-3.)
Rear clearance is
5 in. from hole, Drill 3/8 in. holes (four places)
or 4 in. from
instrument.
Top View
13 in.
4. Properly locate and drill four 3/8 in. diameter holes in the mounting surface.
5. Remove the rubber feet from the instrument.
6. Position the instrument. Insert the bolts through 1/4 in. flat washers, through the
mounting surface, and into the threaded inserts of the instrument. (Refer to
Figure 2-3.)
7. Lightly tighten the bolts.
WARNING
The BacT/ALERT® 3D 60 ships without a default voltage setting. The
procedure that follows describes how to insert the fuse(s) for proper
voltage setting. After installing proper fuse(s), discard the remaining
fuse(s).
Note: The Power Entry Module (P/N 805-0017-02) is located in the rear and lower left corner
of the instrument.
Note: When configured for 115 VAC, the PEM has one 6.3 A, 250 V time-delay fuse
(P/N 870-0008-25). In addition, a conversion clip is located across the neutral line in
the fuse holder.
Note: When configured for 230 VAC, each PEM has two 5.0 A, 250 V time-delay fuses
(P/N 870-0008-24).
5. Install the fuse(s) for the proper voltage rating. Determine the proper fuse
installation to use based on the Facility Power Ratings listed in Table 2-3.
a. To configure the BacT/ALERT® 3D 60 for 115 VAC, see Figure 2-5 to install
a 6.3 A, 250 V time-delay fuse.
1 — Conversion Clip
2 — 6.3 A, 250 V Time-Delay Fuse
b. To configure the BacT/ALERT® 3D 60 for 230 VAC, see Figure 2-6 to install
the two 5.0 A, 250 V time-delay fuses. Be sure to remove and dispose of the
conversion clip, if one is present.
Top/Bottom View
Side View
WARNING
The BacT/ALERT® 3D 60 has been designed to minimize risks
associated with MB testing. However, to further reduce the risks of
accidental exposure to infectious agents, take additional precautions.
It is strongly recommended that the instrument be placed in a
laboratory used for the routine culture of M. tuberculosis. For
activities involving the propagation and manipulation of M.
tuberculosis or Mycobacterium species grown in culture, Biosafety
Level 3 Practice, Containment Equipment, and Facilities are required
as recommended by CDC and NIH guidelines.
Note: If countertop or other permanent (bolted) installation is required, see Installing the
Optional Restraint on page 2-4.
1. Connect the UPS battery per the instructions on the UPS. Place the UPS near
the instrument. Connect the UPS serial cable to the 9-pin connector on the rear
of the UPS and the other end to the UPS connector (not Serial UPS) on the
BacT/ALERT® 3D 60 instrument Communications panel. UPS port connection
is shown in Figure 2-8.
Note: The 9-pin cable packed with the UPS should not be used and should be discarded.
Use the UPS Serial Cable (9-Pin) provided in the accessory kit.
Mouse Printer
Barcode
13 Reader 12
Communications Panel
COMM
LIS Mouse Printer Bar Code
6 8 4
9 Surge outlets
Speakers with battery
Keybrd 1
backup.
Modem
5 UPS
10 11 Monitor
Telephone 3
Line Jack
Rear
3D 60 Communications View 2
Panel. See detailed view. Surge outlets
3D 60 without
battery backup
Figure 2-8: Installation and Setup Diagram — Port and Power Connections
IMPORTANT: The Control Module microprocessor (BIOS) clock is set to and must remain at
the Coordinated Universal Time (UTC), formerly Greenwich Mean Time, in a
24 hour format. The BIOS time is set in manufacturing and should not require
readjustment in the field. The current UTC may be obtained from the National
Technical Information Services WWV radio station by calling 1-303-499-7111, or
by accessing the Internet USNO Master Clock at the following web address:
http://tycho.usno.navy.mil/cgi-bin/timer.pl. Contact Instrument Service for
assistance in setting the BIOS time.
Note: The BIOS date and time are the only settings that get adjusted in the BIOS. Also, the
BIOS time must be set to UTC and not local time, unless the local time is the same as
UTC.
7. Indicate that the BIOS time and date are verified on the BacT/ALERT® 3D 60
Installation Checklist on page C-2.
1. From the Main screen, click the Next Screen button. (Refer to Figure 2-10.)
Next Screen
Button
Key Symbol
Button
Note: The password is company confidential. Do not provide the password to the customer.
The Set Maximum Test Time screen (screen ID 2.7) overlays and disables the
Setup screen.
6. From the Set Maximum Test Time screen, verify the maximum test times are
set to the proper values. The site will provide this information.
7. Click the Check button to save the setting.
8. From the Setup screen, click the Set Audible Alarm Options button.
The Set Audible Alarm Options screen (screen ID 2.8) overlays and disables
the Setup screen.
9. From the Set Audible Alarm Options screen, verify that the audible alarms are
enabled (set to 1).
10. Click the Check button to save the settings.
11. Set the audible volume and tone options to suit the requirements of the site
staff. (Refer to Table 3-1 and Table 3-2 in Section 3.3.2.3.)
12. From the Setup screen, click the Set Date/Time button.
The Set Date/Time screen (screen ID 2.1) overlays and disables the Setup
screen.
13. From the Set Date/Time screen, set the date/time to reflect the current date and
time at the site.
Note: Check with the customer to determine the desired format of the date and time.
15. From the Setup screen, click the Calibrate Module Temperature button.
16. From the Calibrate Module Temperature screen, set the optimal temperature to
reflect the requirements of the site.
17. Click the Check button to save the setting.
18. Indicate that instrument installation options are selected on the
BacT/ALERT® 3D 60 Installation Checklist on page C-2.
Note: Allow the Incubation Chamber to stabilize and maintain temperature for a minimum of
4 hours prior to calibrating the temperature and cells.
1. From the Setup screen, click the Setup and Configuration button. (Refer to
Figure 2-11.)
The Setup and Configuration screen overlays and disables the Setup screen.
(Refer to Figure 2-17.)
2. Select the style and type of instrument installation using the Line Scroll Up/
Down buttons.
3. Click the Check button to save settings.
If changes were made to the settings, the software will exit to the C:\ prompt.
4. At the C:\ prompt, press Ctrl + Alt + Delete to restart the software.
Next Screen
Button
Key Symbol
Button
Note: The password is company confidential. Do not provide the password to the customer.
6. From the Setup and Configuration screen, click the Configure Screen button.
(Refer to Figure 2-17.)
Temperature
Format Rack
Selection Configuration
MB
Configuration
7. Click the Checkbox button to select the desired temperature format (for
example, 37.0 or 37,0).
Note: This only changes the temperature format and not the temperature.
IMPORTANT: In this screen, boundary checking is not done on fields that have numerical
entries. Therefore, it is possible to enter out-of-range values.
1. From the Setup and Configuration screen, click the Configure Instrument
Screen button. (Refer to Figure 2-17.)
Positive Reminder Period (Sec) Alarms until the Positive bottle is removed. Set
the value to 0 and the alarm will not sound after it
is shut off for the first time.
Idle Logout Timeout (Sec) Do not set this value to less than 60 seconds.
Note: If the printer is enabled, restart the software for this update to take effect.
1. From the Setup and Configuration screen, click the User Preferences button.
(Refer to Figure 2-17.)
Note: If you select the wrong option, click the Checkbox button again to deselect.
the functionality available, similar to how the instrument operates while not in
21 CFR Part 11 mode.
If 21 CFR Part 11 mode is enabled and in a logged-off state (user is not logged
on), the user has limited access to the instrument’s functions. The following
activities can be performed while logged out of the instrument software:
Note: Minimum and maximum length only apply to accession numbers entered using the
barcode reader. Accession numbers that are entered manually are not limited to these
settings.
• Lifetime (Days) — Lifetime determines the length of time (in days) after an
accession number may be reused. If a zero (0) is entered, then the accession
number is considered unique. If a database management system is
connected, the Lifetime setting is ignored. If a number is entered, both the
Note: The Lifetime field only accepts integer values in days (for example, 4 or 5). It does not
accept fractional values (for example, 4.5).
1. From the Setup and Configuration screen, click the LIS Configuration button.
(Refer to Figure 2-17.)
Note: The values for this screen depend on how the LIS is configured.
2. Select the following options using the Line Scroll Up/Down buttons.
3. Click the Checkbox button next to the following desired options to select.
Note: If you select the wrong option, click the Checkbox button again to deselect.
Note: If the bidirectional setting is not selected, then this setting is ignored and results are
not sent to the LIS.
• Result On Load — If selected, results will be sent to the LIS when the bottle
is loaded. If not selected, this feature is disabled. Auto Transmit Results
must be selected for this feature to work.
• Result On Unload — If selected, results will be sent to the LIS when the
bottle is unloaded. If not selected, this feature is disabled. Auto Transmit
Results must be selected for this feature to work.
• Result On Negative — If selected, results will be sent to the LIS when the
bottle test result changes from non-negative to negative, or vice versa. If not
selected, this feature is disabled. Auto Transmit Results must be selected
for this feature to work.
• Result On Positive — If selected, results will be sent to the LIS when the
bottle test result changes from non-positive to positive, or vice versa. If not
selected, this feature is disabled. Auto Transmit Results must be selected
for this feature to work.
• Test Order Poll Period (Sec) — Enter a number greater than zero to
periodically poll LIS for test orders, where this value defines the period in
seconds. (For example, a value of 3600 would cause the
BacT/ALERT® 3D 60 to poll for test orders every hour.) Enter a 0 if the user
does not want the BacT/ALERT® 3D 60 to periodically poll the LIS for test
orders.
Note: If the bidirectional setting is not selected, then this setting is ignored and results are
not sent to the LIS.
• Test Link Period (Sec) — Enter a value in seconds to enable this feature so
that when there is a lull in events that require communication with the LIS, a
test message is sent. The value specifies the polling period in seconds. Enter
a 0 to disable this feature.
4. Click the Check button to save the settings.
5. Click the Check button to exit the Setup and Configuration screen. (Refer to
Figure 2-17.)
If changes were made to the Baud Rate, Stop Bit, Parity, or Data Bits, the
software will exit to the C:\ prompt.
6. At the C:\ prompt, press Ctrl + Alt + Delete to reboot the software.
IMPORTANT: The bottle barcodes for i FA, i FN, and i PF must be activated. Failure to comply
may result in a “generic” algorithm when these bottle types are loaded and not
recognized.
1. Verify that each component is enabled by viewing the Main screen. Any
disabled component will be displayed by gray “hash marks” denoting the
disabled rack or individual cells within the rack. Use the Enable/Disable Rack,
and Cell screen to enable each component.
2. Indicate that instrument configuration is completed on the BacT/ALERT® 3D 60
Installation Checklist on page C-2.
The software will indicate or “flag” any cells that have cell flag readings that vary more
than ± 1.5% from the initial factory calibration. Recalibrate the flagged cells as
required.
1. From the Main screen, click the Next Screen button. (Refer to Figure 2-29.)
Next Screen
Button
Key Symbol
Button
Note: The password is company confidential. Do not provide the password to the customer.
11. Indicate that the cells are calibrated on the BacT/ALERT® 3D 60 Installation
Checklist on page C-2.
When 21 CFR Part 11 mode is enabled, users are required to enter a user name and
password.
While logged out of the instrument in 21 CFR Part 11 mode, the Main screen is
displayed with the Login button. (Refer to Figure 2-32.)
Login
Button
Figure 2-32: Main Screen While Logged Out — 21 CFR Part 11 Mode
1. From the Main screen, click the Login button. (Refer to Figure 2-33.)
User Name
Password Check
Button
Cancel
Button
Change
Password
Previous Button
Screen
Button
When 21 CFR Part 11 mode is first activated, there are no valid users in the
system. The customer may use the User Name setup and the Password setup.
Once there is at least one user added, the User Name (setup) is no longer
valid.
Note: Field Service Engineers may use the User Name service and the Password
biomerieux3d. The User Name (service) is always valid. Do not provide the service
User Name or Password to customers.
2. Click the User Name field (field turns white) and enter service.
3. Click the Password field (field turns white) and enter biomerieux3d.
4. Click the Check button to log in.
5. At the Main screen, click the Next Screen button. (Refer to Figure 2-35.)
Next Screen
Button
Key Symbol
Button
Configure
Users
Button
Note: The password is company confidential. Do not provide the password to the customer.
Note: The Configure Users button is only active when the instrument is configured for
21 CFR Part 11 mode.
10. Verify the User Configuration screen appears. (Refer to Figure 2-37.)
Add User
Previous
Screen Check
Button Button
User Name
Field
Add User
Button User
List
Previous
Screen Check
Button Button
Add User
Button User
List
Previous
Screen Check
Button Button
1. Connect a phone cable from the wall jack to the TELCO port of the 3D External
Modem. (Refer to Figure 2-40.)
2. Connect a phone cable from the 3D External Modem PHONE port to a
telephone, if available.
3. Connect the serial cable from the 3D External Modem to the
BacT/ALERT® 3D 60 system.
4. Adjust the switch setting on the 3D External Modem so that SWITCHES 3 and 8
are in the down position.
5. Connect AC power to the 3D External Modem.
3. Click the Line scroll up button to view the algorithms listed in Table 2-4.
11 LES/Plastic BacT/ALERT SA
12 LES/Plastic BacT/ALERT SN
13 LES/Plastic BacT/ALERT FA
14 LES/Plastic BacT/ALERT FN
15 LES/Plastic BacT/ALERT PF
17 LES/Plastic BacT/ALERT MP
5. Perform the following Scanning Plastic Bottle Barcodes procedure for any
bottles that are not set to plastic.
6. Indicate that the plastic bottle algorithms were verified on the
BacT/ALERT® 3D 60 Installation Checklist on page C-2.
To ensure that the algorithm for plastic bottles is used by the BacT/ALERT® 3D 60
instrument, each bottle type must be scanned.
Note: If the barcode reader does not scan the printout, then type the bottle ID below the
barcode.
The Bottle ID field is highlighted, and the Accession Number field is no longer
highlighted.
4. Repeat Step 2 and Step 3 for each of the eight remaining barcodes shown in
Figure 2-44.
5. Click the Check button to exit the Load Mode screen.
6. Repeat Verifying Plastic Bottle Algorithms on page 2-34 to verify that all plastic
algorithms are correct.
Plastic SA Bottle
Plastic SN Bottle
Plastic FA Bottle
Plastic FN Bottle
Plastic PF Bottle
Plastic MP Bottle
Note: The password is company confidential. Do not provide the password to the customer.
The Backup Management screen overlays and disables the Setup screen.
5. Click the Previous Screen button to return to the Main screen and verify the
backup is running.
The Backup In Progress icon will appear in the upper left corner of the Backup
Management and Main screens. The icon will disappear every 5 seconds for
half a second. Allow 15–30 minutes for the backup to complete.
6. Remove AC power from the UPS by removing the UPS AC power cord from the
wall outlet. Leave the UPS connected to the instrument and leave the power
switch set to On.
7. While waiting for 1 minute, verify that the UPS emits an audible alarm.
8. Verify that the blue Abnormal Shutdown screen appears on the instrument
monitor screen.
9. Verify that the instrument powers down.
10. After the instrument powers off, restore power to the UPS.
11. Verify that the instrument retains the correct time and has no visible errors
displayed on the monitor screen.
12. Indicate that the UPS Functional Test is complete on the BacT/ALERT® 3D 60
Installation Checklist on page C-2.
CAUTION: Do not plug the printer into the UPS. The current draw for the
heating element in the printer will overload most uninterruptible power
supplies.
4. Enable the Printer in the software. (Refer to Section 2.7.4, Configuring the
Instrument on page 2-19.)
5. Restart the BacT/ALERT® 3D 60 software to activate the printer setting.
6. Run any report to test the printer.
7. Indicate that the printer is installed on the BacT/ALERT® 3D 60 Installation
Checklist on page C-2.
Chapter Contents
Operator Control
Panel (Display)
Barcode
Scanner
Printer
UPS
Keyboard
This section gives a brief description of the assembly's purpose and operation.
• 3D 60 System Control
The Operator Display Monitor and the instrument internal circuitry are the
heart of the system. (Refer to Figure 3-1.) An electronics compartment
houses the internally installed WinSystems® Computer. With the attached
keyboard and barcode scanner, the operator performs all of the controlling,
programming and database management functions from the Operator Display
Monitor.
• Barcode Scanner
The barcode scanner scans bottle labels, identifying the bottles whenever
loading or unloading.
• Keyboard
This is a standard-sized 101-key keyboard. It is used as an alternative input
device to the Barcode Scanner or mouse.
Note: The following information refers to the rear of the BacT/ALERT® 3D 60 and
specifically the Communications Ports. (Refer to Figure 3-2.)
• COMM Port
This port is reserved for future use.
• Printer Port
A printer connects to this port via a parallel cable and produces hard copy
reports.
• Modem Port
Used to connect the 3D 60 to an external modem.
• LIS Port
Used for connection to the facility's LIS.
• Monitor Port
Used for connection to an external VGA or SVGA monitor.
• UPS Port
Used to connect an external UPS to the instrument.
• Mouse Port
Used for connecting the mouse to the instrument.
• Keyboard Port
Used for connecting the keyboard to the instrument.
A Zip® drive has been interfaced, allowing for factory supported software upgrades
and system backups. A 16-bit data bus has been provided for add-on peripherals.
Two different models of the CPU PCBA exist in the field. CPU PCBA (P/N 750-0070-
01) is the older model. CPU PCBA (P/N 750-0070-902) is the newer model. The
newer model functions much as the older model, with the exception of the addition of
a port for the Compact Flash Disk.
The Quad Serial Interface (Com4) board is a daughter board that plugs into the
PC/104 bus on the WinSystems® motherboard. (Refer to Figure 3-6.)
There are four ports on the board. Com 1 is configured as a RS-485 port. Com 2, 3,
and 4 are configured as RS-232 ports. Com 1 and Com 2 are located on connector J6,
while Com 3 and Com 4 are located on connector J3. All communications ports are
software supported and operate at a 9.6K Baud data rate.
The Quad Serial is a four-channel serial (INS8250 compatible) P/C 104 module that is
based on the 16C554 UART. It is ideally suited for applications running DOS
programs or languages that require exact register compatible hardware for program
execution. Each of the channels supports RS-232, with optional RS-485 and RS-422
electrical levels to interface with a wide variety of serial communications devices.
The Quad Serial PCBA is I/O mapped. Each channel requires eight port addresses
and each is independently decoded within the I/O map. The interrupt status register
also requires one I/O port. The addresses are decoded in the PAL and are jumper-
selectable, with eight different I/O maps. The Com4 card is a 16-bit stack-through card
that supports 8-bit data transfers only.
A 4.5-volt reference is on the PCBA for A/D reference and a 2.5-volt reference for
temperature control. Power-fail circuitry detects if the +5VDC is out of tolerance.
LED's on the board show E Fan ON, Power-Enable ON, Blower ON, Heater ON,
Agitation ON, +5V ON for ModSig, +5V ON for Step-5 PCBA, +5V ON for Racks, Door
Switches Closed, six status indicators and RS-485 transmit and receive indicators.
Temperature sensors on the board are used for E-Fan control and to report
electronics over-temperature conditions. External input and output temperature
sensors are used by PID control firmware on the microprocessor for incubation
temperature control and are connected to JP1. Heater, blower and fan ON/OFF time
The front door has two separate interlock switches. These connect to the MODSIG
PCBA through JP11 and JP12. One switch, which the firmware cannot override,
functions as a safety interlock that disables the heater and blower if the door should
open. The second switch is for logic functions and redundancy. Both switches are
monitored by the firmware and by the Step-5 agitation motor control firmware.
Step-5
PCBA
ModSig
PCBA
JP13, JP15, JP17, JP18 and JP19 connect to the Instrument Controller CPU in the
3D 60. JP29 and JP30 are used for transferring serial data signals from the Quad
Serial PCBA. The Serial UPS connector at P7 is for future use. The UPS is controlled
through the connector at P5 and must be connected for proper system operation. The
VGA display monitor is connected at P6. Five-volt power is connected to the board at
JP31 and is supplied to the Zip Drive and the Compact Flash Adapter PCBA through
JP21. Control and data signals for the racks are supplied through JP10.
Jumpers on the ModSig PCBA control the audio frequencies supplied to the remote
speakers. The remote speakers control volume. Volume jumpers on the board are for
future use only. Refer to Table 3-1 and Table 3-2 to identify the different jumper
settings and their effect.
600 Hz In In In In
800 Hz In In In Out
TO Step-5 PCBA
• Door Interlock
Switches
J2 • +5 Power
• Agitate ON
32KB RAM FROM Step-5 PCBA
• Error Signal
68HC11
32KB RAM
• Blower ON
• Heater ON
Heater Inlet/Outlet
Temperature JP1 Drivers JP2 •
•
Agitate ON
E Fan ON
Sensors • Power En ON
Parallel I/O
To/From CPU JP18
P6 VGA Monitor
RS-232,
RS-485 From JP29 +5V
Quad Serial JP21 Compact Flash
Zip® Drive
RS-232 From
Quad Serial PCBA
JP30
J1 Remote Speaker
P9 Modem
The Step-5 PCBA is designed to drive a five-phase stepper motor at the speed of 60
revolutions per minute. The board operates with input voltages of +5V and +12V,
drives the stepper motor directly, and provides starting and stopping functions based
on inputs received from the ModSig PCBA.
The Step-5 PCBA drives a Vextra 5-phase stepper motor at 1 revolution per second
(+10%) to provide rack agitation. The controller board's inputs, outputs, and features
are listed below.
• Outputs:
A Fault Code is sent to the ModSig PCBA if certain conditions occur. The four
conditions that initiate this code are:
Phase 1
Phase 2
Phase 3
Phase 4
Phase 5
If an over current occurs, the motor will be stopped. The fault signal is
generated by the open-collector of Q16, and goes LOW to indicate a fault. By
design, a disconnected power cable will also result in a fault indication.
• Internal Indicators:
• LED 1 Output Codes:
[1 flash]: Motor stopped OK
[2 flashes]: Motor running OK
[3 flashes]: Motor over current
[4 flashes]: Door switch (one switch open, the other closed)
[5 flashes]: Position sensor not changing state
[6 flashes]: Safety timer not functioning (does not go high after door
opened)
[7 flashes]: Motor power bad (12VDC)
Note: LED 1 on the Step-5 PCBA constantly outputs one of the above codes. To interpret
the code, count the number of longer flashes (LED on periods) after the short start
flash.
• Stopped
• Ramping up
• Holding (constant speed)
• Seeking (waiting for position count to reach ramp down point)
• Ramping down
• Fault (over current)
• Motor power bad
Independent
Safety Timer
Interlocks
PLD
Microchip
Position PIC16C74
Sensor Drive Motor
MOSFETS
Stop/Start
Command
(From ModSig High-side
PCBA) Predrivers
DC In Fault
Output
This drive provides an internal resource for loading new operating system software
and performing system data backup. (Refer to Figure 3-1.)
The Compact Flash Disk is a removable mass storage device installed on the
Compact Flash Adapter PCBA or the CPU PCBA. It is a device about the size of a
matchbook and weighs about one-half ounce. Since it is a solid state device utilizing
flash memory, it contains no moving parts and requires no battery for data retention.
The Compact Flash Disk is compatible with IDE/ATAPI standards and serves as the
C drive in the BacT/ALERT® application, where it holds the operating program, bottle
records, system configurations, event log files, and miscellaneous other information
needed for BacT/ALERT® operation. (Refer to Figure 3-12.)
Figure 3-12: Compact Flash Disk (Example Only, Actual Part May Vary)
The Compact Flash Adapter PCBA (with Compact Flash Disk installed) is mounted
near the WinSystems® motherboard. It is connected by a flat ribbon cable to the
motherboard's IDE port.
The Compact Flash Disk stores the operating system, BacT/ALERT® 3D 60 firmware,
instrument configurations, operator settings, layout of the instrument pictorial, along
with bottle record NVRAM data. On 3D 60 Modules that have been upgraded to CPU
PCBA (P/N 750-0070-902), the Compact Flash Adapter PCBA is removed and the
Compact Flash Disk is installed in a port on the underside of the CPU PCBA.
The BacT/ALERT® 3D 60 instrument uses a standard BTA cabinet power supply. The
Power Panel Assembly provides the AC and DC power supply voltages that are used
within the 3D 60. With the use of relays, this will also shut down incubation processes
when instructed to do so by the ModSig PCBA. The power supply provides the four
separate DC voltages (+ 5V and ±12V) used throughout the instrument. (Refer to
Figure 3-14.)
Solid State
Relays
DC Power
Supply
Transformer
3.3.3.2 Transformer
The transformer isolates the instrument from facility power and is installed on the
Power Panel. (Refer to Figure 3-14.)
3.3.3.3 Heater
The heater provides the necessary heat for the incubation process of cultures. (Refer
to Figure 3-15.)
Blower
Heater Coil
Blower Motor
Capacitor
3.3.3.4 Blower
For the incubation process, the blower circulates heat (produced by the heater)
throughout the module. (Refer to Figure 3-15.) The blower stops any time the door is
opened and will start again when the door is closed.
The Step-5 Motor provides agitation to the racks, unless the drawer has been
converted to MB. (Refer to Figure 3-16.) If all three racks are converted to MYCO
(MB), the motor is disconnected to bypass this process. (Refer to Section 6.3.)
Step-5
Motor
A fan circulates air over the electronic components within the Electronics
Compartment to help prevent overheating. (Refer to Figure 3-17.)
Fan Assembly
There are two Thermistors PCBAs in the 3D 60 instrument. The Inlet and Outlet
Thermistors measure input and output air temperature, respectively. The Inlet and
Outlet temperature readings are averaged together to comprise the Incubation
Chamber's temperature used by the temperature control logic.
3.3.4.1 Rack
Each rack contains 20 bottle cells. Unless the instrument is designated as an MB unit,
the racks are continuously rocked (to provide agitation for the specimen bottles). Also,
three fascia plates identify the bottle locations and are not interchangeable between
replacement racks.
3.3.4.2 Cell
There are 60 cells per instrument, with cells identified as 1 through 20 in the first (top)
rack, 21 through 40 in the middle rack and 41 through 60 in the bottom rack.
The cell flag serves as a clip securing the bottles into their cells. The cell flag is also a
key component in bottle detection and cell diagnostics.
A lamp is located next to each cell and provides feedback to the operator. Depending
on the function selected by the operator, illuminated lamps identify the positions of
positive or negative bottles, cells that can be loaded, etc.
The Rack Controller PCBA, using EPROM-stored software, controls the excitation
LEDs. (Refer to Figure 3-20.) The LED light illuminates and reflects on the bottle
sensors, creating emitted measurable values at the photo detectors. The Rack
Controller digitizes each cell's photo-detector output for reading by the ModSig's
Control Module and for bottle presence detection.
Cell indicator LEDs are controlled for operator information. Status and data are
communicated through an RS-485 cable from the Rack Controller to the ModSig
board. The microprocessor software also monitors the bottle rack rocking motion with
a "Hall-Effect" sensor. In addition, the software normalizes illumination readings with
cell and temperature calibration factors. Address lines from the ModSig are used to
confirm that the 3D 60's WinSystems® CPU board is addressing the correct Rack
Controller thus eliminating the need for switches on each Rack Controller. Calibration
values are stored onboard the Rack Controller PCBA in a non-volatile RAM (NVRAM).
The Rack Controller PCBA is located in the rear portion of the rack channel of the
Incubation Chamber. A fully implemented Incubation Chamber will have three Rack
Controller PCBAs.
The microprocessor is the Motorola 68HC11 and the same used on the ModSig
PCBA. This microprocessor has separate 16-bit address and 8-bit data lines defined.
This eliminates the need of having external address and data multiplexing for the
external 32KB of RAM and 32KB of EPROM included on the PCBA.
The clock speed of the 68HC11 is set at 16 MHz by an external crystal. Internally, the
microprocessor contains 512 bytes of EPROM, 1024 bytes of RAM, and no ROM.
External memory for the microprocessor is on the Rack Controller PCBA. A 16-bit free
running counter, driven by a programmable prescaler and other divider chains,
provides for a wide range of internal timing periods for software use. Included is a
watchdog timer.
The Rack Controller PCBA uses 32KB of static RAM in a surface mount package.
The software for most dynamic storage uses this memory. The Rack Controller uses
32KB of EPROM for program storage. Jumpers are included to allow the installation of
larger memory, in the event that the 27C256 microprocessor becomes obsolete.
Memory parts with less than 90ns access are required.
There are two 20-pin connector positions on the board. Power (+5 volts for each
board) is supplied through a cable from the ModSig to the JP2 connector. The other
connector, JP1, is cabled to JP2 of the next rack. No other power is required. A fuse
on the ModSig protects the set of three Rack Controllers in the Incubation Chamber.
The +5 volt input voltage produces two reference voltages. A +4.5V reference is
provided to the A-D converter. A +2.5V reference is provided for the Rack temperature
operational amplifiers as well as an accurate current control of the exciter LED's
current by an operational amplifier.
Each cell has one exciter LED. Each of these LEDs is driven by an output from a
74ACT374 driver.
One indicator LED is provided for each bottle holder or cell. Each of the LEDs is driven
by an output from a 74ACT374 driver. In addition, there are four "status" LEDs for
general purpose and are explained in the firmware.
Agitation 32 KB
Sensor RAM
Cell
68HC11 Microprocessor 32 KB
Sensors OP
w/internal RAM and EPROM RAM
Amp 1-20
Analog
Multiplexer
Drivers Cell
Constant Exciters
Rack 1-20
Temperature Current
Control
Cell
12-Bit Indicators
Reference
A-D 1-20
Voltage
Converter
Drawer ID
RS-485
RCVR/XMTR
Event Request
To/From To/From Next
ModSig or Rack Controller
previous Rack
Controller +5V Out
RS-485
+5V In Rack ID
RS-485 Drawer ID
Rack ID Event Request
Event Request Rack ID Out
Rack ID Out
Each cell has an individual external sensor. The sensor outputs are multiplexed to
cascaded operational amplifiers that drive the analog input of an A-D Converter. The
cell sensor readings are sent, on request, to the instrument. The Rack Controller
firmware detects bottle presence from these photodiode sensors as well.
Circuitry is provided to develop the analog signals to the A-D Converter for the two
temperature sensors on the Rack PCBA. These sensors are part of the calibration
factors for the bottle cells.
The agitation sensor is a "Hall-Effect" sensor. The sensor detects bottle rack motion
for a change in magnetic flux from a rare-earth magnet as the rack is agitated. Each
board has a sensor; however, only one position in each drawer is utilized for
monitoring.
The Event Request signal is a common open drain signal driven by the ModSig PCBA
and all Rack Controller PCBAs. The ModSig, Rack Controller and WinSystems® CPU
board can also read the signal. Its purpose is similar to an interrupt signal; it is to notify
the WinSystems® CPU board that an event such as a bottle removal has occurred.
For the ModSig, this signal is both an output to the WinSystems® CPU board and an
input from the Rack Controller. When the Event Request is pulled low, the Drawer ID
low signal used by the Rack Controllers is disabled. The Rack Controller drives the
Drawer ID lines to indicate which Rack Controller(s) activated the Event Request line.
The Rack Control firmware reads the Event Request line before and after reconfirming
its address, to ensure address lines are not in use.
With only one of four lines low for each drawer, the Rack Controllers can determine
which drawer they are in for each address identification for WinSystems® CPU board
commands. When the Event Request is low, this is disabled; allowing the same lines
to be used by the Rack Controllers to identify which ones activated the Event
Request. The driving circuitry and input circuitry is provided on the Rack Controller.
Firmware on the ModSig aids the WinSystems® CPU board in selecting which Rack
Controllers read status, thus expediting bottle detection state changes.
The ModSig makes only one of the three Rack ID lines low for each rack, and then
polling this line within the Rack PCBAs, the Rack Controllers can determine their
address without having individual address switches. (Refer to Figure 3-22.)
Four outputs from one of the 74ACT374-exciter LED drivers serve as the output
drivers for the Rack ID Out signals, through open-drain inverters.
The communication between the modules is over RS-485 full-duplex circuitry. The
necessary conversion between RS-485 differential signals and single-ended CMOS
compatible serial signals, for use by the software, is contained on the Rack Controller.
The actual RS-485 signals connect to all Rack Controllers through a connector on the
ModSig PCBA.
JP10
ModSig
PCBA
+5V
+5V
16
17
18
Polling line
connects via
JP2 on each
Rack PCBA
17 17 17 17 17 17
18 18 18 18 18 18
A 12-bit A-D Converter is provided to convert the rack temperature sensor's analog
signals to the digital signals used by the Rack Controller software. The same A-D
Converter is used for converting cell sensor analog signals into digital signals.
Current for the exciter LEDs is controlled by using the +2.5 VDC reference voltage,
resistor voltage dividers, an operational amplifier, and a MOSFET. Only one exciter
LED is activated at any one time.
Input power for the 3D 60 is either 115 VAC or 220 VAC. (Refer to Figure 3-23.) The
installation technician sets the proper voltage selection at the Power Entry Module
(PEM). The circuitry is protected by one 6.3A time-delay 250 VAC fuse (120 VAC
version) or two 5A time-delay 250 VAC fuses (240 VAC version). AC voltage is
coupled across the isolation transformer thus insulating the instrument from facility
power for circuit protection including a 120:120VAC or a 240:120VAC (step-down)
voltage ratio.
Facility
Power
UPS
Power
Isolation MOV
Module Transformer
Blower
TB1
Heater
Elec. Comp.
Fan
Common Logic Ground Return 3
ModSig PCBA
Power Panel
+5 VDC GND
Step-5 PCBA
Step-5 Motor
Input power is routed to a metal oxide varistor (MOV) mounted between terminals 2B
and 3B of TB1. The varistor provides inline surge protection for the internal circuitry of
the system. From TB1, power is routed to the fan, DC power supply and all power-
enable relays.
The DC power supply provides the necessary DC voltages on J2 for various sub-units
throughout the 3D 60 instrument. There are three separate voltages: ±12VDC and
+5VDC, with a common DC return providing a logic ground. In addition, the DC power
supply provides a Power Fail signal that is routed to the ModSig PCBA.
The BacT/ALERT 3D 60's power supply is a quad output switcher intended for small
to medium-sized digital systems. These power supplies also contain built-in over-
voltage protection and a power fail detect signal.
Universal input voltage allows these supplies to be powered in any country of the
world without changing any jumpers or individual switch settings. The unit is capable
of supplying the high peak currents necessary to start disk drives and peripheral
devices. An integral filter reduces line-conducted noise below FCC and VDE limit B.
The over-voltage protection circuits are designed to activate at 6.25 VDC (± 0.75
VDC) for 5 VDC output. The power supply has a "Power Fail Detection" circuit that
produces a "power fail detection" (PFD) output. The PFD signal occurs within
milliseconds following an abnormal AC line condition. The delay between the
abnormal AC line condition and the PFD output varies depending on AC line variation
and power supply load.
If this condition is met, the PFD signal will go to a low state within approximately 3ms.
Once the output returns to the +4.75 VDC, the PFD signal will return to a high state in
50ms to 200ms. Table 3-3 and Table 3-4 show the power supply and pin settings.
Input Surge Current Cold Start, 110 VAC/220 VAC 17A maximum/ 34A maximum
P1 AC Ground
J1 P2 AC Neutral
P3 AC Hot
P1 + 5.1 VDC
P2 + 5.1 VDC
P3 + 5.1 VDC
P4 Return
P5 Return
P6 Return
J2 P7 Return
P8 + 12 VDC
P9 + 12 VDC
P10 PFD
P11 - 12 VDC
P12
P13 -5 VDC
The modem interfaces the 3D 60 instrument through a connection to the MODEM port
on the rear of the unit. Digital signals are converted to tones by the modem for
sending to the telephone system through the telephone jack connection and to a
remote modem at the receiving end. Conversely, tones from the remote modem are
converted to digital signals and sent to the instrument through the MODEM port.
Power for the modem is supplied by a wall-mounted power supply that supplies low
voltage required by the modem. The wall-mounted power supply plugs directly into a
standard electrical outlet.
The fax modem serves as an interface to the 3D 60 instrument for remote access. The
remote user is able to download and upload information, including seeing what is on
the display screen that the local operator views. The Technical Support Center is able
to perform remote diagnostics and other troubleshooting of the instrument.
The modem interfaces the instrument through a connection to the MODEM port on the
rear of the unit. Digital signals are converted to tones by the modem for sending to the
telephone system through the telephone jack connection and to a remote modem at
the receiving end. Conversely, tones from the remote modem are converted to digital
signals and sent to the instrument through the MODEM port. Power for the modem is
supplied by a wall-mounted power supply that supplies low voltage required by the
modem. The wall-mounted power supply plugs directly into a standard electrical
outlet.
Computer Overload
Telephone Surge Interface Protector
Protection Port
On/Off
Switch
Test
Switch
Power
On LED
• On/Off Switch — Turns the UPS and BacT/ALERT® 3D 60 unit ON and OFF from
the UPS. A Power On LED is included on the On/Off Switch.
• UPS Comm Port (Computer Interface Port) — The UPS Computer Interface Port
is a standard 9-pin port used to connect a cable that monitors communication
between the UPS and the 3D 60 instrument. The port connection is shown as
Computer Interface Port in Figure 3-24.
• Overload Reset Switch — The Overload Reset switch is a reset switch (breaker)
allowing for UPS reset following a power surge or uncommon UPS shutdown. The
overload is rated 120 VAC, 12 Amp (Max) at 50/60 Hz.
• Test (Check Battery/Alarm Disable) Switch — By pressing the top of Test (toggle)
switch, the UPS battery may be checked. The UPS powers the loads from the
battery. If the UPS emits the battery exhaustion alarm, the battery is weak and
requires extended charge or replacement.
• 120V Backup Receptacles — These 120 VAC receptacles are used when devices
connected to the UPS require full battery backup to allow proper shutdown of the
equipment. A device that may require such battery backup might be the Display
Monitor, since it is required to shut down the 3D 60 instrument in a facility power-
fail situation.
CAUTION: Do not plug the printer into the UPS. The current draw for the
heating element in the printer will overload most uninterruptible power
supplies.
• Accessory Surge Receptacles — These 120V receptacles are used when devices
connected to the UPS do not require battery backup, but may require voltage
surge protection.
• AC Power Cord — The AC Power Cord is a standard three-pronged AC
(grounded) cord.
The International UPS model is very similar to the US domestic model. The voltage
rating, power cord and fuse ratings, however, are for the Interntional power
conversions. (Refer to Figure 3-25 for International UPS components.)
Computer
Interface Overload
Port Protector
On/Off
Switch
Test
Switch
Power
On LED
This screen is primarily used when an instrument is being installed. It allows the Field
Service Engineer to enter information regarding the temperature format selection,
rack configuration, and whether or not the instrument is set to MB. Refer to Figure 3-
26 for an example of the BacT/ALERT® 3D 60 Configuration screen.
Temperature
Format Rack
Selection Configuration
MB
Configuration
1. At the Main screen, click the Next Screen button. This will bring up the Setup
screen. (Refer to Figure 3-27.)
Next Screen
Button
2. From the Setup screen, enter the Field Engineer password 43432121. (Refer to
Figure 3-28.)
Note: The password is company confidential. Do not provide the password to the customer.
3. Click the Key Symbol button. (Refer to Section 3.7 for Screen Diagnostics.)
Keypad
Password
Buttons
Key Symbol
Button
Previous
Screen Configure
Button Screen
Button
Note: To get back to the Main screen, click the Previous Screen button.
This section and its subsections contain illustrations and tables that explain the
screens, buttons, values, and other necessary information that are needed for
diagnostic testing of the BacT/ALERT® 3D 60 instrument. This section and its
subsections should be reviewed prior to performing diagnostic testing procedures.
This section and its subsections should also be used as a reference during diagnostic
testing procedures. (Refer to Chapter 7, Diagnostics and Troubleshooting Information
for instructions on how to perform diagnostic tests.)
Operation Buttons: These buttons invoke operations that are common to all the
diagnostics. Refer to Table 3-6 and associated figures below for details.
Run Continuously
Pause/Run Once
Find Text
Cancel Print
Previous
Screen Button
Button Description
Run When clicked, diagnostic runs continuously and data in the display area is updated
Continuously continuously.
Pause/Run If clicked, while the diagnostic is running continuously, the diagnostic is allowed to run
Once one more time and then it is paused. Once paused, the data in the display area
represents the data generated the last time the diagnostic was run. If clicked while the
diagnostic is already paused, the diagnostic is allowed to run one time and updates the
display data.
Button Description
Find Text This function is used to find a text string in the display data. When clicked, the following
screen appears.
Enter Text to
Search For
Press to Exit
Without Search
Press to Begin
Search
To find the additional instances of the search text, press the F3 function key on your
keyboard. When the last instance is reached, searching automatically wraps back to the
first instance of the text search.
Button Description
Save Output to This function saves all of the diagnostic output to a text file on the Zip® disk. To use this
Zip function:
1. Place a Zip® disk in the instrument's Zip® drive.
2. Click the Save Output to Zip button. In response, the following screen will appear.
A default file name is automatically entered for you. By default the data is saved on the
Zip® disk in a subdirectory named DIAGDUMP. The file name corresponds to the
number of the diagnostic that is currently selected. If you wish to use another filename,
click the destination filename entry box, press the SPACE bar to clear the box and
then type in the desired name.
Note: You cannot change the destination drive. It will always be the Zip® drive. Use
only 250 MB Zip® disks. To initiate the Save, click the Check button. To cancel the
save, click the Cancel button.
Operation will return to the currently selected diagnostic screen. The data is written to
the file in text. This file can be viewed using any text-editing program. For example,
Windows® Notepad. To save, click the Check button.
Print All Output Send all lines of diagnostic output to the printer.
Print Current Send just diagnostic output lines (currently displayed) to the printer.
Page
Cancel Print Button is normally disabled, yet enabled while printing is in progress. Clicking this button
cancels printing.
Exit Terminates the diagnostic, exits this screen and returns to the Diagnostic Test Selection
screen.
Display Output Area Navigation Buttons: The buttons, located just above and
below the diagnostic output display area, allow the user to scroll through the display
data.
Anchor display (Top/ Scroll one line Scroll one page (Up/ Go to Home or End
Bottom) (Up/Down) Down)
Positions display area
Controls whether Moves display area one Moves display area one to the first or last line of
display stays anchored line at a time (up or page (up or down). The the diagnostic output.
on the first or last line of down). The same same function can be The same function can
the diagnostic output. function can be performed by pressing be performed by
performed by pressing the Page Up or Page pressing the Home or
the Up Arrow or Down Down keys on the End keys on the
Arrow keys on the keyboard. keyboard.
keyboard.
Slide Bar: The bar, located to the left of the diagnostic output display area, allows the
user to scroll through the display data.
The raised section of the Slide bar indicates the position of the
displayed data and is relative to the amount of total display data area.
The size of the raised portion of the Slide bar is proportional to the
amount of display data in relationship to the total amount of data
(including scrollable data). For example, if there are 54 lines of total
data and only 27 lines can be displayed at a time, the raised area
would be one-half of the total size of the Slide bar.
Clicking a geographical location (top, middle, bottom, etc.) on the
Slide bar will scroll the to same relative position in the display area.
For example, if there were 2000 lines of data in the display area, and
the slide bar is clicked in the middle, the display area would scroll
approximately to line 1000.
Key Symbol
Button
Main Selection
List
Diagnostic
Name
Module, Drawer,
Rack and Cell
Identifier Code
Diagnostic Output
Display Area
Help Line
(if applicable)
Field Purpose
Date and Time Date and time displayed is the local time. It is
displayed in day, month, year, hour, minute,
second order [DD/MM/YY HH:MM:SS],
regardless of the settings on the Set Date/
Time menu.
Field Purpose
The example screen shown in Figure 3-36 depicts the normal values to be found in
the Loopback screen.
1.1 Should increment — Number of transmit buffer empty interrupts seen by serial
hardware.
1.2 Should increment — Number of receive buffer full interrupts generated by serial
hardware.
2.2 Should remain at 0 — Number of parity errors reported by serial hardware. Parity
error checking is turned off so no errors should be reported.
3.1 This value can be ignored — It is displayed for diagnosis of software activity and
indicates whether the software's transmit data FIFO is empty or not (1 or 0
respectively).
3.2 Should remain at 0/0/0 — Displayed for the purpose of diagnosing software
activity and indicates whether there was fault storing data in the software's transmit
data FIFO and where in the code that fault occurred.
4.1 Should increment — Indicates the number of times data loopback test was run
and passed.
4.2 Should remain at 0 — Indicates the number of times loopback test was run and
failed.
4.2.1 Should remain at 0 — Indicates the number of times loopback data was
transmitted but a time-out occurred before all the data was received.
4.2.2 Should remain at 0 — Indicates the number of times loopback data was
transmitted and received but the received data did not match what was transmitted.
5.1 Should remain at 0 — Indicates the state of the power fail button on the loopback
test cable, which is not used on the 3D 60.
5.2 Should remain at 0 — Indicates the state of the event request button on the
loopback connector, which is not used on the 3D 60.
The example screen shown in Figure 3-37 depicts the normal values to be found in
the Port Stats screen.
1.4 Number of modem status interrupts generated by serial hardware. (Note this
interrupt is disabled as indicated on the screen.)
2.2 Number of parity errors reported by serial hardware. Parity error checking is
turned off so no errors should be reported.
2.3 Number of framing errors reported by serial hardware. A framing error occurs if the
number of serial bits between the start and stop bit are not correct.
3.1 Indicates whether the software's transmit data FIFO is empty or not (1 or 0
respectively).
3.2 Indicates whether there was fault storing data in the software's transmit data FIFO
and where in the code that fault occurred.
The example screen shown in Figure 3-38 depicts the normal values to be found in
the Message Stats screen.
2.1 Number of event requests generated by the module controller. (Should be the
same as 1.2.)
2.3 Number of messages exchanged with module controller without any problem.
2.4 Number of messages sent to module controller where the expected response did
not occur in the required time.
2.5 Number of responses from the Module Controller that had a checksum error.
2.6 Number of message exchanges with the Module Controller that fail for some
reason other than time-outs or corruption.
3.2 Number of messages exchanged between Control Module and rack controller 1.
3.3 Number of messages exchanged with rack controller 1 without any problem.
3.4 Number of messages sent to rack controller 1 where the expected response did
not occur in the required time.
3.5 Number of responses from the rack controller 1 that had a checksum error.
3.6 Number of message exchanges with the rack controller 1 that fail for some reason
other than time-outs or corruption.
The remaining sections duplicate the information in 3.1 through 3.6, but apply to the
racks 2 through 12 respectively.
The Message Log screen displays the recent history of commands and responses
exchanged between the Control Module and the Incubation Module.
The Temperature Log screen displays the history or heater temperature data from the
sensors and sensor on the racks. Bottle testing is not affected by this diagnostic.
The Run Status runs continually and adds new temperature readings to the end of the
display as they occur.
Time of Event in
dd/mm/yy hh:mm:ss format
Event Location
Module Controller
entry shows heater
inlet and outlet sensor
readings along with
average of the readings.
The Drawer open (DRW OPEN) and Close (DRW CLOS) events are also displayed,
but are not shown in the example above and events are logged when they occur.
MOD TEMP events are logged every 5 minutes. RCK TEMP events are logged
approximately once per hour per rack. TEMP SET events are logged when they
occur.
The Reading Log screen displays the most recent readings taken on incubation
module cells. Each line of the display identifies each cell the reading was taken from,
followed by:
The following reading values indicate a fault condition and you will see Fault Code
#51–57 Bottle Rack Hardware Failure reported on the operator screen.
The diagnostic automatically stops running once it has gathered all the required
readings and the RUN button will be in the unpressed state. Pressing the run button
restarts the diagnostic. It will gather new readings and then display the diff % data.
When first entered, or whenever the diagnostic is restarted (RUN button pressed), the
diagnostic gathers readings from the racks in the incubation module. It takes several
seconds to gather all of the data. The rack that is being processed is indicated on the
Flag Check screen. (Refer to Figure 3-42.)
If a rack is not installed or a drawer is not installed, they are skipped and do not
appear in the display. During the gathering phase, keyboard input is ignored.
Once all of the required data is gathered, the diff % data is displayed. (Refer to
Figure 3-43.) This screen groups the cells by the amount (Delta) that their flag reading
differs from its expected reading (i.e. the flag value when the cell was calibrated).
Cells whose
difference is greater
or equal to 1.0%
Cell location
(% diff.)
A given cell may appear in multiple groups. For example, a cell whose difference is
1.5% will appear in the 1.0% and 1.5% groups. Groupings are shown for percentages
of 0.5%, 1.0%, 1.5%, and so on up to 10%
No data is displayed for cells that are loaded, since a flag reading cannot be taken on
those cells. No data is displayed for cells in racks or drawers that are uninstalled.
Note that keyboard input is accepted at this point. Pressing the C key brings up the
Calibration screen. In this manner a cell can be calibrated without leaving the
diagnostic screen. When the Cancel button is pressed on the calibration screen,
operation returns to the Flag Check diagnostic screen.
Pressing 2 causes detailed data to be displayed. This data shows the current flag
reading, the expected flag reading, and the difference and percentage difference for
each cell. If a cell is loaded, the word LOAD is shown on the current reading line.
Pressing 1 returns to the diff % data.
3.7.3.8 Drawer/Heater
This diagnostic displays the state of the drawer switch and heater function in an
incubation module. Bottle testing is not affected by this diagnostic and the Run Status
runs continuously and updates the display continuously.
This diagnostic provides a way to turn all cell LEDs ON and also to blink them in a
pattern to verify that they all work and that no LEDs are shorted together. Bottle
testing is not affected by this diagnostic.
Upon initial entry to diagnostic, all LEDs are turned on and the display appears as
shown in Figure 3-45.
After the Walking LED test is initiated, the display values appear as shown in
Figure 3-46. The displayed data indicates which LED should be on.
The values listed below are the normal values to be found in the Zip Drive Test
screens.
2.1 Number of times test was run and test file could not be opened.
2.2 Number of times test was run and an error occurred while writing.
2.3 Number of times test was run and an error occurred while reading the data or the
data did not match what was written.
2.4 Number of times an error was generated while trying to write the test data.
3.1 Last test data string written to test file on Zip® disk.
3.2 Last data read back from test file on Zip® disk.
The example screen shown in Figure 3-47 depicts the normal values to be found in
the Barcode Loopback screens.
1.1 Should increment — Number of transmit buffer empty interrupts seen by serial
hardware.
1.2 Should increment — Number of receive buffer full interrupts generated by serial
hardware.
2.2 Should remain at 0 — Number of parity errors reported by serial hardware. Parity
error checking is turned off so no errors should be reported.
3.1 This value can be ignored — It is displayed for diagnosis of software activity and
indicates whether the software's transmit data FIFO is empty or not (1 or 0
respectively).
3.2 Should remain at 0/0/0 — Displayed for the purpose of diagnosing software
activity and indicates whether there was fault storing data in the software's transmit
data FIFO and where in the code that fault occurred.
4.1 Should increment — Indicates the number of times data loopback test was run
and passed.
4.2 Should remain at 0 — Indicates the number of times loopback test was run and
failed.
4.2.1 Should remain at 0 — Indicates the number of times loopback data was
transmitted but a time-out occurred before all the data was received.
4.2.2 Should remain at 0 — Indicates the number of times loopback data was
transmitted and received but the received data did not match what was transmitted.
4.2.3 Should remain at — Just before transmitting loopback data, the four possible
combinations or signal states are asserted on the RTS and DTS lines. The state of the
CTS line which is connected to the RTS line is monitored and if found inconsistent
with the assert RTS/DTS signals, this count is incremented.
4.2.4 Should remain at 0 — Just before transmitting loopback data, the four possible
combinations or signal states are asserted on the RTS and DTS lines. The state of the
DSR line which is connected to the DTR line is monitored and if found inconsistent
with the assert RTS/DTS signals, this count is incremented.
The values listed below are the normal values to be found in the Modem Loopback
screens.
1.1 Should increment — Number of transmit buffer empty interrupts seen by serial
hardware.
1.2 Should increment — Number of receive buffer full interrupts generated by serial
hardware.
2.2 Should remain at 0 — Number of parity errors reported by serial hardware. Parity
error checking is turned off so no errors should be reported.
3.1 This value can be ignored — It is displayed for diagnosis of software activity and
indicates whether the software's transmit data FIFO is empty or not (1 or 0
respectively).
3.2 Should remain at 0/0/0 — Displayed for the purpose of diagnosing software
activity and indicates whether there was fault storing data in the software's transmit
data FIFO and where in the code that fault occurred.
4.1 Should increment — Indicates the number of times data loopback test was run
and passed.
4.2 Should remain at 0 — Indicates the number of times loopback test was run and
failed.
4.2.1 Should remain at 0 — Indicates the number of times loopback data was
transmitted but a time-out occurred before all the data was received.
4.2.2 Should remain at 0 — Indicates the number of times loopback data was
transmitted and received but the received data did not match what was transmitted.
4.2.3 Should remain at — Just before transmitting loopback data, the four possible
combinations or signal states are asserted on the RTS and DTS lines. The state of the
CTS line which is connected to the RTS line is monitored and if found inconsistent
with the assert RTS/DTS signals, this count is incremented.
4.2.4 Should remain at 0 — Just before transmitting loopback data, the four possible
combinations or signal states are asserted on the RTS and DTS lines. The state of the
DSR line which is connected to the DTR line is monitored and if found inconsistent
with the assert RTS/DTS signals, this count is incremented.
The example screen shown in Figure 3-48 depicts the normal values to be found in
the Power Event Log screens.
Time of event in
dd/mm/yy hh:mm:ss
format.
Events that appear in the log (but not necessarily shown in the example above) are:
INSTR CNTRL AC (power fault) — this is logged when AC power fails and returns as
reported by the UPS. Instrument will do a controlled shutdown.
INSTR CNTRL UPS (battery low) — this is logged when reported by the UPS.
MODULE POWER FAULT — this is logged when incubation chamber asserts its
power fault signal.
MODULE POWER UP — this is logged when incubation chamber negates its power
fault signal, or whenever the instrument software is started.
RACK POWER UP — this is logged when rack reports a power up event, or whenever
the instrument software is started.
This function allows contents of selected files to be viewed. Bottle testing is not
affected by this diagnostic.
For each file in the c:\btlprocs directory of the flash drive with a .BID extension, the file
name is displayed, followed by a row of dashes, followed by the contents of the file.
For each file in the c:\btlprocs directory of the flash drive with a .BTF extension, the file
name is displayed, followed by a row of dashes, followed by the contents of the file.
For each file in the c:\btlprocs directory of the flash drive with an .APF extension, the
file name is displayed, followed by a row of dashes, followed by the contents of the
file.
For each file in the c:\btlprocs directory of the flash drive with an .LOG extension, the
file name is displayed, followed by a row of dashes, followed by the contents of the
file.
FILE : EVENT1.LOG
------------------------------------------------------------
!! 139719250(06/06/01 08:04:34) 0x38(DTChg) 4294948672
!! 139719253(06/06/01 08:04:37) 0x4(ICFlt) 0
!! 139719253(06/06/01 08:04:37) 0x3(IMFlt) 1 0 0 0 0 0
!! 139719253(06/06/01 08:04:37) 0x3(IMFlt) 2 0 0 0 0 0
!! 139719253(06/06/01 08:04:37) 0x3(IMFlt) 3 0 0 0 0 0
!! 139719253(06/06/01 08:04:37) 0x3(IMFlt) 4 0 0 0 0 0
!! 139719253(06/06/01 08:04:37) 0x3(IMFlt) 5 0 0 0 0 0
!! 139719253(06/06/01 08:04:37) 0x3(IMFlt) 6 0 0 0 0 0
!! 139719258(06/06/01 08:04:42) 0x30(IMPwr) 1
!! 139719258(06/06/01 08:04:42) 0x30(IMPwr) 2
!! 139719535(06/06/01 08:09:19) 0x4(ICFlt) 12
!! 139721277(06/06/01 08:38:21) 0x38(DTChg) 4294948672
!! 139721279(06/06/01 08:38:23) 0x4(ICFlt) 0
!! 139721280(06/06/01 08:38:24) 0x3(IMFlt) 1 0 0 0 0 0
FILE : EVENT0.LOG
------------------------------------------------------------
!! 139809948(07/06/01 09:16:12) 0x38(DTChg) 4294948672
!! 139809951(07/06/01 09:16:15) 0x4(ICFlt) 0
!! 139809951(07/06/01 09:16:15) 0x3(IMFlt) 1 0 0 0 0 0
!! 139809951(07/06/01 09:16:15) 0x3(IMFlt) 2 0 0 0 0 0
!! 139809951(07/06/01 09:16:15) 0x3(IMFlt) 3 0 0 0 0 0
!! 139809951(07/06/01 09:16:15) 0x3(IMFlt) 4 0 0 0 0 0
!! 139809951(07/06/01 09:16:15) 0x3(IMFlt) 5 0 0 0 0 0
!! 139809951(07/06/01 09:16:15) 0x3(IMFlt) 6 0 0 0 0 0
!! 139809956(07/06/01 09:16:20) 0x30(IMPwr) 1
!! 139809956(07/06/01 09:16:20) 0x30(IMPwr) 2
!! 139810161(07/06/01 09:19:45) 0x4(ICFlt) 12
!! 139810363(07/06/01 09:23:07) 0x3(IMFlt) 2 0 0 74 0 0
!! 139811149(07/06/01 09:36:13) 0x4(ICFlt) 0
!! 139812682(07/06/01 10:01:46) 0x38(DTChg) 4294948672
!! 139812694(07/06/01 10:01:58) 0x4(ICFlt) 0
!! 139812695(07/06/01 10:01:59) 0x3(IMFlt) 1 0 0 0 0 0
!! 139812695(07/06/01 10:01:59) 0x3(IMFlt) 2 0 0 0 0 0
!! 139812695(07/06/01 10:01:59) 0x3(IMFlt) 3 0 0 0 0 0
!! 139812695(07/06/01 10:01:59) 0x3(IMFlt) 4 0 0 0 0 0
!! 139812695(07/06/01 10:01:59) 0x3(IMFlt) 5 0 0 0 0 0
!! 139812695(07/06/01 10:01:59) 0x3(IMFlt) 6 0 0 0 0 0
!! 139812700(07/06/01 10:02:04) 0x30(IMPwr) 2
!! 139812700(07/06/01 10:02:04) 0x30(IMPwr) 1
!! 139813009(07/06/01 10:07:13) 0x3(IMFlt) 1 0 20 0 0 0
!! 139813053(07/06/01 10:07:57) 0x3(IMFlt) 1 0 0 0 0 0
!! 139813314(07/06/01 10:12:18) 0x3(IMFlt) 1 0 99 0 0 0
!! 139813764(07/06/01 10:19:48) 0x3(IMFlt) 1 0 51 0 0 0
The event info content varies (depending on the kind of event) and is described in
Table 3-10.
0x1(SWExc) LOG,TEST,MAIN
This event occurs if there is a condition, for example a hardware problem that the software cannot handle, or
there is a design flaw in the software that results in a condition that cannot be handled. Report this information
to the 3D 60 software development team.
0x3(IMFlt) 3 0 41 0 0 0
This information corresponds to information displayed on the Main screen's incubation module pictorial. A fault
code of 0 means there is no fault in the module or corresponding drawer. See BacT/ALERT® 3D 60 User
Manual or Quick Reference card for the meaning of non-zero fault codes.
0x4(ICFlt) 10
This information corresponds to information displayed on the Main screen's Controller Module pictorial. A fault
code of 0 means there is no fault. Refer to the BacT/ALERT® 3D 60 User Manual or Quick Reference card for
the meaning of non-zero fault codes.
0x30(IMPwr) 2
This event indicates that the incubation module has just powered up.
0x38(DTChg) 16432
Internally, the 3D 60 software maintains time in UTC time. When the operator uses the Set Date/Time screen
to enter the local time, the difference between local time and UTC is calculated and this event is logged to
record that difference. This event is an indication that the operator has changed the date or time.
0x39(QCFlt) 3 2 3 12 1
This event is logged any time a cell goes out of calibration or back into calibration (presumably because the
operator has calibrated the cell). The Drawer Number value ranges from 1 to 4 and corresponds with drawers
A through D, respectively. The Rack Number ranges from 1 to 3 and corresponds with the top rack, middle
rack, and bottom rack, respectively. The Cell Number ranges from 1 to 20 and identifies the affected cell. The
cell number that the operator sees depends on the Rack Number. For example, if the Rack Number is 2 and
Cell Number is 5, the operator would see this as cell 25. If the Rack Number is 3 and the Cell Number is 10,
the operator would see this as cell 50. If the Fault present value is 1, the cell is out of calibration. If the Fault
present value is 0, then the cell has been calibrated.
Backup started
0x100(BkupStart)
Backup completed
successfully
0x101(BkupOk)
Backup failed
0x102(BkupFailed)
This event is logged any time that a backup has started but failed to complete successfully.
0x300(CCalSt) 3 2 3 12
This event is logged any time that the operator starts a single cell calibration.
0x301(CcalOK) 3 2 3 12
This event is logged any time that the operator successfully completes single cell calibration.
0x600(ICPwr)
This event is logged any time that the Controller Module is powered up.
Normal Shutdown
0x200(NrmSD)
This event is logged any time that the software is shut down in the normal fashion.
0x201(PwrLossSD)
This event is logged any time that the software is shut down due to a loss of power.
This event is logged any time that the software is shut down for an abnormal reason without saving the
NVRAM. For example, if during power up the software detects that the CPU NVRAM is uninitialized, the
software is abnormally shut down without saving the NVRAM.
0x203(SmxErrSD)
This event is logged any time that the software is shut down because of an SMX error detected at startup.
This event is logged any time that the software is shut down for an abnormal reason with saving the NVRAM.
For example, if the software is started with the appropriate Debug mode for immediate shutdown configuration
set, the software shuts down abnormally, but saves the NVRAM.
0x205(RemUserSD)
This event is logged any time that the software is shut down because of a request passed by the Remote User
software.
0x500(MTSet) 1 37.0
This event is logged any time that the setpoint has been changed.
0x501(IMOfs) 1 -.42
This event is logged any time that the incubation module is powered up.
0x502(IMOfs)
This event is logged any time that the incubation module's temperature is updated.
This event is logged any time that the incubation module's temperature is calibrated.
0x700(RUOn)
This event is logged any time that the software detects that the Remote User software has connected.
0x701(RUOff)
This event is logged any time that the software detects that the Remote User software connection has been
dropped.
3.7.4.10 Software
The Software screen and its subcreens display data used to debug, verify and
validate Controller Module software. Also, displays logs of exception conditions
detected by the software. Bottle testing is not affected by this diagnostic.
The values listed in Figure 3-50 are typical values to be found in the Software
screens. The descriptions of these values are listed below.
3.1 Approximate CMOS clock value the last time the Controller Module software was
terminated
3.2 Approximate local time the Controller Module software was terminated
The Debug screen displays software debug messages intended for development
activity and is intended for the 3D 60 software development team. The format and
content messages put in this log are determined by the software implementor and no
design specifications are imposed.
The Software Exception Log screen displays software exceptions messages intended
for the 3D 60 software development team. Access this screen if a Fault Code #19
Software Exception is being displayed on the Main screen. The format and content
messages put in this log are determined by the software implementor and no design
specifications are imposed.
The Operating System Exception Log screen displays exceptions reported by the 3D
60 operating system (SMX). The format and content messages put in this log are
determined by the operating system vendor (Micro Digital) and no design
specifications are imposed.
The Modem Activity Log screen displays software debug messages intended for
development activity and is intended for the 3D 60 software development team. The
format and content messages put in this log are determined by the software
implementor and no design specifications are imposed.
The Association Tables screen displays current content of various dictionaries and
association tables used to keep track of bottle and pre-load data. An example of the
Association Tables screen content is shown below. Explanatory comments are shown
in italics and do not actually appear on the diagnostic screen.
--- Patient Name Dictionary --- relates a hospital ID (hid) to the patient name
17, AH3HID hash table index, hospital ID, first name, last name
--- HID to Acc Container --- relates hospital ID to accessions that share that hid
--- ACC to HID Dictionary --- relates accession back to its hospital ID
--- ACC to BID Container --- relates accession to bottle IDs that share that accession
--- BarCodeDictionary --- relates bottle ID to a bottle record, key values from that
record are displayed
[a:1003600 n:28 s:28 p:60] [r:2 s:P x:1] [loc:: a: address of bottle record
1A20 ld: 1] n: bottle record index (1-1920) s: bottle record sequence num
bid: AH3BID1 acc: AH3ACC*8895AFA p: hash table index
hid: AH3HID pn: ,
r: result code (1=pos 2=neg 3=ntd 4=near pos)
[a:1003800 n:29 s:29 p:61] [r:3 s:I x:1] [loc:: s: status (P=prelim, I= incomplete, F=final) x: significant test result if
1A40 ld: 1] ld: 0=unloaded 1=loaded
bid: AH4BID1 acc: AH4ACC*1 loc: if "ld: 1", cell where bottle is loaded;
hid: AH4HID pn: AH4LAST, AH4FIRST if "ld: 0", cell where bottle was last loaded
bid: bottle ID
acc: accession number
hid: hospital ID
pn: last, first parts of patient
--- ACC to ACC Container --- relates a base accession to accessions with same base
3, AH3ACC, AH3ACC*8895AFA hash table index, base accession, accession with suffix
4, AH4ACC, AH4ACC*1
--- PRELOAD: ACC to RecNums Container --- relates a base accession to all record numbers in the
preload file that have that accession
14, ACC43, 6
--- PRELOAD: BID to RecNum Dictionary --- relates a bottle ID to the one record number in the preload
file that has that bottle ID
171, AH2BIDA, 12 hash table index, bottle ID, preload file rec num
--- PRELOAD: ACC to RecNum Dictionary --- release an accession to the one record number in the
preload file that has that accession
191, AH4ACC*1, 2 hash table index, accession, preload file rec num
275, AH3ACC*8895AFA, 1
The Bottle Processing Tables screen displays the current content of the tables that
drive the way the 3D processes bottles. See the following example.
ZBLYM, 5, 8, 16, 2
ZPPC, 5, 15, 29, 2
ZBMB, 5, 8, 18, 2
ZBMP, 5, 8, 17, 2
ZBPF, 5, 8, 15, 2
ZBFN, 5, 8, 14, 2
ZBFA, 5, 8, 13, 2
ZBSA, 5, 8, 11, 2
ZMB, 5, 15, 9, 1
ZPN, 5, 15, 7, 1
ZPA, 5, 15, 6, 1
ZFN, 5, 15, 5, 1
ZFA, 5, 15, 4, 1
@O, 8, 8, 29, 2
@A, 8, 8, 28, 2
@C, 8, 8, 27, 2
@D, 8, 8, 26, 2
@E, 8, 8, 25, 2
@S, 8, 8, 24, 2
@I, 8, 8, 22, 2
@Z, 8, 8, 23, 2
@R, 8, 8, 21, 2
@P, 8, 8, 20, 2
!B, 8, 8, 18, 2
!M, 8, 8, 17, 2
!L, 8, 8, 16, 2
!P, 8, 8, 15, 2
!G, 8, 8, 14, 2
!F, 8, 8, 13, 2
!N, 8, 8, 12, 2
!A, 8, 8, 11, 2
SB, 8, 8, 18, 3
SM, 8, 8, 17, 3
SL, 8, 8, 16, 3
SP, 8, 8, 15, 3
SG, 8, 8, 14, 3
SF, 8, 8, 13, 3
SN, 8, 8, 12, 3
SA, 8, 8, 11, 3
MB, 0, 0, 8, 1
ZN, 5, 15, 3, 1
ZA, 5, 8, 2, 1
IA, 8, 8, 50, 2
IN, 8, 8, 51, 2
IL, 8, 8, 52, 2
9, 0, 0, 0, 0
8, 0, 0, 0, 0
7, 0, 0, 0, 0
6, 0, 0, 0, 0
5, 0, 0, 0, 0
4, 0, 0, 0, 0
3, 0, 0, 0, 0
2, 0, 0, 0, 0
1, 0, 0, 0, 0
0, 0, 0, 0, 0
G, 0, 0, 0, 0
U, 0, 0, 1, 0
H, 9, 9, 3, 1
D, 9, 9, 2, 1
0, GENERIC, G, 1, 3, 0, 10, 5, 11
1, UNKNOWN, U, 1, 5, 0, 10, 5, 11
2, AEROBIC, A, 1, 2, 1, 10, 10, 10
7, LYM-BACT, PN, 1, 2, 1, 4, 4, 4
8, MB/BACT, MS, 1, 1, 1, 2, 2, 2
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
MinSkipPts(6)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
SlopProp(0.1E0) SlowSlopProp(0.2E-1)
FastDgdr(0.75E0) SlowDgdr(0.9E0)
AcelSlopThres(20) SlowSlopThres(10)
PosSlopDly(720) DltDgdr(0.995E0)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
SlopProp(0.1E0) SlowSlopProp(0.2E-1)
FastDgdr(0.75E0) SlowDgdr(0.9E0)
AcelSlopThres(20) SlowSlopThres(10)
PosSlopDly(720) DltDgdr(<1.0E-32)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
MinSkipPts(6)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
MinSkipPts(6)
MinSkipPts(6)
SlopProp(0.1E0) SlowSlopProp(0.2E-1)
FastDgdr(0.75E0) SlowDgdr(0.9E0)
AcelSlopThres(20) SlowSlopThres(10)
PosSlopDly(720) DltDgdr(0.995E0)
--- Embedded E2 Parms ---
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
SlopProp(0.1E0) SlowSlopProp(0.2E-1)
FastDgdr(0.75E0) SlowDgdr(0.9E0)
AcelSlopThres(20) SlowSlopThres(10)
PosSlopDly(720) DltDgdr(<1.0E-32)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
MinSkipPts(6)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
MinSkipPts(6)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
MinSkipPts(6)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
MinSkipPts(6)
13 ID(G3) BltFact(0.231E1) Resv(1320)
SlopProp(0.1E0) SlowSlopProp(0.2E-1)
FastDgdr(0.75E0) SlowDgdr(0.9E0)
AcelSlopThres(20) SlowSlopThres(10)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
MinSkipPts(6)
SlopProp(0.1E0) SlowSlopProp(0.2E-1)
FastDgdr(0.75E0) SlowDgdr(0.9E0)
AcelSlopThres(20) SlowSlopThres(10)
PosSlopDly(720) DltDgdr(<1.0E-32)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
MinSkipPts(6)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
MinSkipPts(6)
SlopProp(0.1E0) SlowSlopProp(0.2E-1)
FastDgdr(0.75E0) SlowDgdr(0.9E0)
AcelSlopThres(20) SlowSlopThres(10)
PosSlopDly(720) DltDgdr(0.995E0)
--- Embedded H2 Parms ---
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
MinSkipPts(6)
SlopProp(0.1E0) SlowSlopProp(0.2E-1)
FastDgdr(0.75E0) SlowDgdr(0.9E0)
AcelSlopThres(20) SlowSlopThres(10)
PosSlopDly(720) DltDgdr(<1.0E-32)
SlowDgdr(0.98E0) SlowSlopProp(0.9083E0)
MinSkipPts(6)
18 not used
19 not used
20 not used
21 not used
22 not used
23 not used
24 not used
25 not used
26 not used
27 not used
28 not used
29 not used
30 not used
31 not used
32 not used
33 not used
34 not used
35 not used
36 not used
37 not used
38 not used
39 not used
40 not used
41 not used
42 not used
43 not used
44 not used
45 not used
46 not used
47 not used
48 not used
49 not used
50 not used
51 not used
52 not used
53 not used
54 not used
55 not used
56 not used
57 not used
58 not used
59 not used
60 not used
61 not used
62 not used
63 not used
64 not used
65 not used
66 not used
67 not used
68 not used
69 not used
70 not used
71 not used
72 not used
73 not used
74 not used
75 not used
76 not used
77 not used
78 not used
79 not used
80 not used
81 not used
82 not used
83 not used
84 not used
85 not used
86 not used
87 not used
88 not used
89 not used
90 not used
91 not used
92 not used
93 not used
94 not used
95 not used
96 not used
97 not used
98 not used
99 not used
10 LES/Glass (0x2000)
19 LES/Glass (0x4000)
20 LES/Glass (0x4000)
21 LES/Glass (0x4000)
22 LES/Glass (0x4000)
23 LES/Glass (0x4000)
24 LES/Glass (0x4000)
25 LES/Glass (0x4000)
26 LES/Glass (0x4000)
27 LES/Glass (0x4000)
28 LES/Glass (0x4000)
29 LES/Glass (0x4000)
30 LES/Glass (0x6000)
31 LES/Glass (0x6000)
32 LES/Glass (0x6000)
33 LES/Glass (0x6000)
34 LES/Glass (0x6000)
35 LES/Glass (0x6000)
36 LES/Glass (0x6000)
37 LES/Glass (0x6000)
38 LES/Glass (0x6000)
39 LES/Glass (0x6000)
40 LES/Glass (0x6000)
41 LES/Glass (0x6000)
42 LES/Glass (0x6000)
43 LES/Glass (0x6000)
44 LES/Glass (0x6000)
45 LES/Glass (0x6000)
46 LES/Glass (0x6000)
47 LES/Glass (0x6000)
48 LES/Glass (0x6000)
49 LES/Glass (0x6000)
50 LES/Glass (0x2000)
51 LES/Glass (0x2000)
52 LES/Glass (0x2000)
53 LES/Glass (0x6000)
54 LES/Glass (0x6000)
55 LES/Glass (0x6000)
56 LES/Glass (0x6000)
57 LES/Glass (0x6000)
58 LES/Glass (0x6000)
59 LES/Glass (0x6000)
60 LES/Glass (0x6000)
61 LES/Glass (0x6000)
62 LES/Glass (0x6000)
63 LES/Glass (0x6000)
64 LES/Glass (0x6000)
65 LES/Glass (0x6000)
66 LES/Glass (0x6000)
67 LES/Glass (0x6000)
68 LES/Glass (0x6000)
69 LES/Glass (0x6000)
70 LES/Glass (0x6000)
71 LES/Glass (0x6000)
72 LES/Glass (0x6000)
73 LES/Glass (0x6000)
74 LES/Glass (0x6000)
75 LES/Glass (0x6000)
76 LES/Glass (0x6000)
77 LES/Glass (0x6000)
78 LES/Glass (0x6000)
79 LES/Glass (0x6000)
80 LES/Glass (0x6000)
81 LES/Glass (0x6000)
82 LES/Glass (0x6000)
83 LES/Glass (0x6000)
84 LES/Glass (0x6000)
85 LES/Glass (0x6000)
86 LES/Glass (0x6000)
87 LES/Glass (0x6000)
88 LES/Glass (0x6000)
89 LES/Glass (0x6000)
90 LES/Glass (0x6000)
91 LES/Glass (0x6000)
92 LES/Glass (0x6000)
93 LES/Glass (0x6000)
94 LES/Glass (0x6000)
95 LES/Glass (0x6000)
96 LES/Glass (0x6000)
97 LES/Glass (0x6000)
98 LES/Glass (0x6000)
99 LES/Glass (0x6000)
The Audit Trail screen displays activities and other related information. Each audit trail
entry consists of an event and the time and date the event occurred. Refer to
Table 3-11 for a list of events recorded in the Audit Trail.
Note: Some recorded events may occur while the user is logged out, such as firmware
upgrades, and will not have an associated User ID. In these cases, the user is listed
as Unknown.
Anonymous Bottle First Loaded Time, Date, User ID, Cell Location, Max Test Time,
Sequence Number
Identified Bottle First Loaded Time, Date, User ID, Cell Location, Bottle Type,
Max Test Time, Bottle ID, Accession Number,
Sequence Number
Bottle Identified Time, Date, User ID, Cell Location, Bottle Type,
Max Test Time, Bottle ID, Accession Number,
Sequence Number
Bottle Moved Time, Date, User ID, Original Cell, Destination Cell,
Sequence Number
Bottle Record Changed Time, Date, User ID or LIS, Cell Location, Action
(Bottle Type Changed, Loaded Status Changed,
Max Test Time Per Bottle Changed, Test Result
Manually Changed, Bottle ID Changed, Accession
Number Changed), Sequence Number, Bottle
Type, Loaded Status, Max Test Time, Test Result,
How Determined, Bottle ID, Accession Number
Temperature Calibration Time, Date, User ID, Module ID, New Temperature
Setpoint
Display Time Changed Time, Date, User ID, Amount of Change Relative to
UTC (Universal Coordinated Time)
Module, Drawer, Rack, or Cell Time, Date, User ID, Location, Action (enable or
Enabled or Disabled disable), Device
Maximum Test Time for Bottle Time, Date, User ID, Bottle Type, Max Test Time
Type Changed
Positive Alarm Setting Changed Time, Date, User ID, Action (enable or disable)
Instrument Alarm Setting Time, Date, User ID, Action (enable or disable)
Changed
Operation Error Alarm Setting Time, Date, User ID, Action (enable or disable)
Changed
21 CFR Part 11 Mode Changed Time, Date, User ID, Action (enable or disable)
User Account Changed Time, Date, User ID, Action (user added, user
deleted, password changed or cleared)
Patient Name Changed Time, Date, User ID, Hospital ID, First Name, Last
Name
Cell Gap Detection Faults Time, Date, User ID, Number of Bottles Cleared
Cleared
3.7.5.1 Loopback
The example screen shown in Figure 3-51 depicts the normal values to be found in
the Loopback screens.
1.1 Should increment — Number of transmit buffer empty interrupts seen by serial
hardware.
1.2 Should increment — Number of receive buffer full interrupts generated by serial
hardware.
2.2 Should remain at 0 — Number of parity errors reported by serial hardware. Parity
error checking is turned off so no errors should be reported.
3.1 This value can be ignored — It is displayed for diagnosis of software activity and
indicates whether the software's transmit data FIFO is empty or not (1 or 0
respectively).
3.2 This value can be ignored — It is displayed for diagnosis of software activity and
indicates whether Controller Module is waiting to receive data from the DBMS, or is
transmitting data to the DBMS.
3.0 Should remain at 0 — Displayed for the purpose of diagnosing software activity
and indicates whether there was a fault storing data in the software's receive data
FIFO.
4.1 Should increment — Indicates the number of times data loopback test was run
and passed.
4.2 Should remain at 0 — Indicates the number of times loopback test was run and
failed.
4.2.1 Should remain at 0 — Indicates the number of times loopback data was
transmitted but a time-out occurred before all the data was received.
4.2.2 Should remain at 0 — Indicates the number of times loopback data was
transmitted and received but the received data did not match what was transmitted.
4.2.3 Should remain at 0 — Just before transmitting loopback data, the four possible
combinations or signal states are asserted on the RTS and DTS lines. The state of the
CTS line which is connected to the RTS line is monitored and if found inconsistent
with the assert RTS/DTS signals, this count is incremented.
4.2.4 Should remain at 0 — Just before transmitting loopback data, the four possible
combinations or signal states are asserted on the RTS and DTS lines. The state of the
DSR line which is connected to the DTR line is monitored and if found inconsistent
with the assert RTS/DTS signals, this count is incremented.
The example screen shown in Figure 3-52 depicts the normal values to be found in
the Port Stats screens.
1.4 Number of modem status interrupts generated by serial hardware. (Note this
interrupt is disabled as indicated on the screen.)
2.2 Number of parity errors reported by serial hardware. Parity error checking is
turned off so no errors should be reported.
2.3 Number of framing errors reported by serial hardware. A framing error occurs if the
number of serial bits between the start and stop bit are not correct.
3.0 Indicates whether there was fault storing data in the software's receive data FIFO.
This screen displays recent history of message data exchanged between Controller
Module and LIS. The example screen shown in Figure 3-54 depicts the normal values
to be found in the Message Log screens.
********************************************************
********************************************************
SCANDISK C:
----------------------------
----------------------------
File IO.SYS
File MSDOS.SYS
Directory DOS
File COMMAND.COM
File AUTOEXEC.BAT
File CONFIG.SYS
File PME16.EXE
File BTA3D.INI
Total Entries 7
Deleted Entries 0
Directories 1
Cluster count 1
Traverse subdirectory DOS
Directory .
Directory ..
File ATTRIB.EXE
File CHKDSK.EXE
File COUNTRY.SYS
Total Entries 5
Deleted Entries 0
Directories 2
Cluster count 1
----------------------------
----------------------------
----------------------------
----------------------------
********************************************************
********************************************************
********************************************************
********************************************************
********************************************************
DrvTypeNum = 128
Reserved1 = 0x0
BootSig = 0x29
VolID = 365958411
Vol Label = ZIP250
FileSysType = FAT16
--- Computed data ---
FAT Type = 16
FAT Size = 239
Total Secs = 489440
Bytes/FATEnt = 2
RootDirSector= 479
RootDirSectrs= 32
DataSector = 511
ClusterSector= 511
EOCMark >= 0xFFF8
BADSectorMark= 0xFFF7
Cluster 2 sector = 511
********************************************************
********************************************************
SCANDISK D:
----------------------------
File ZIP250
File 100-1D.RKP
File 100-1E.RKP
File 100-1F.RKP
Directory MTOUCH62
Total Entries 5
Deleted Entries 0
Directories 1
Cluster count 1
Directory .
Directory ..
File MCAL.BAT
File MCAL.OVL
Total Entries 4
Deleted Entries 0
Directories 2
Cluster count 1
----------------------------
----------------------------
----------------------------
----------------------------
********************************************************
********************************************************
Normally, the mouse is used to make selections and enter data. The keyboard is an
alternate method and backup means of inputting data. Items on a screen are selected
by using the Tab key. A box or color change indicates which item is selected.
Pressing Enter will initiate the function of the selected item. For menus that require
entry of a numeric password, the numeric keys can be used instead of clicking the
mouse. The number keys (1–4) are used to enter the password value. The 5 is used
instead of clicking the Key Symbol button. Up/Down Scroll buttons are selected by
using the Tab key. Use > for next, and < for the previous item. Function keys can be
used to invoke the following commands:
• Ctrl + F1 – activates a special Unload mode. This identifies bottles that are
loaded in cells that are disabled or out of calibration. The green drawer LED
illuminates on drawers containing such bottles. When the drawer is opened, the
cell indicators will light for those bottles. Cells with rapidly flashing LEDs are
disabled or unusable. In this mode, bottles can be removed from the indicated
cells one at a time and moved to usable cells whose LED is off.
• Ctrl + F2 – activates a special Unload mode. This identifies cells out of
calibration. All such cells are identified whether they are loaded or not. When the
drawer is opened, the cell indicators will light for cells out of calibration.
• Ctrl + F3 – activates a special Unload mode. This identifies all bottles declared
positive by Fault Code #80 Reading Gap Detected.
• Ctrl + F10 – activates a special Unload mode. This identifies all currently loaded
bottles regardless of their test status. The green drawer indicator is lit on drawers
with such bottles. When the drawer is opened, the cell indicators will light for the
loaded cells.
Note: These special Unload modes function like the regular Unload modes. Bottles can be
removed and identified as they are removed. Bottles can even be moved to a different
cell if so desired.
Note: Anonymous bottles cannot be relocated to another cell until a bottle ID has been
entered for the bottle.
The quick reference card is simply that, a quick reference. (Refer to Figure 3-56 and
Figure 3-57.) The card contains references to errors and fault codes that may appear
on the BacT/ALERT® 3D 60 during operation. The card's purpose is to aid the
operator in determining what may have caused a particular problem. It does not
contain corrective action measures. Refer to Chapter 7, Diagnostics and
Troubleshooting Information of this manual for possible solutions.
3-95
Quick Reference Card
3-96
Quick Reference Card
452501-1EN1
Instrument Components
The intention of this chapter is to provide Field Service Engineers with the basic
operating procedures. These procedures may already be included in the
BacT/ALERT® 3D 60 User Manual.
Chapter Contents
The following procedure is designed to properly bring the system power up and down
gracefully.
Note: If an error occurs, press Esc, then type yes to restart the sequence.
2. Press Enter. A blue screen appears and the following will be displayed:
Normal shutdown in progress
Shutdown complete
3. After the DOS prompt appears, turn the Main Power switch located on the
Power Entry Module (PEM) to Off.
4. Press the On/Off switch to turn the UPS off.
5. If performing work on the interior of the module, disconnect the power cord from
the PEM.
4.1.2 Power Up
1. Reconnect the power cord to the PEM if it had been previously disconnected.
2. Turn Main Power switch to On, providing AC power to the UPS.
3. Push the On/Off switch on the UPS.
Note: The system should restart on its own when power is reapplied to the instrument.
Note: It is recommended that the Zip® disk be formatted. If the backup fails to perform,
replace the Zip® disk with a new Zip® disk.
Note: The contents of the Bottle Status table will be used as a reference after the restore
procedure is completed. It will be used to assure that no bottle status has changed.
13. Record the version of the software shown on the Main screen. (Refer to
Figure 4-1.)
Bottle Status
Table
Software Version
Number
Next Screen
Button
14. Click the Next Screen button to access the Setup screen.
15. Enter the password 43432121.
Note: The password is company confidential. Do not provide the password to the customer.
18. Click a cell with a dark circle around it and an asterisk (*) inside the circle.
(Refer to Figure 4-3.)
Dark Circle
with Asterisk (*)
Previous Screen
Button
19. Verify the Edit Bottle Detail screen appears. (Refer to Figure 4-4.)
Cancel Button
21. Record the bottle data in the blank fields of Figure 4-33 for one bottle in the
Incubation Chamber.
22. Click the Cancel button to return to the Edit Cell Contents screen.
23. Click the Previous Screen button to return to the Main screen.
24. Press Esc, then type yes to exit the software.
IMPORTANT: Ensure that the 3D 60 software backup has completed prior to exiting the
software.
25. At the C:\ > prompt, type xcopy*.Dat D:\backup /v and press Enter.
Note: DOS will ask you if you wish to overwrite the files with the .DAT extension. Press A to
overwrite all files. If asked to create a new file or directory, the backup performed
earlier did not complete properly. Return to Step 10 and begin the procedure again.
26. At the C:\ > prompt, type xcopy*.Bak D:\backup /v and press Enter.
Note: DOS will ask you if you wish to overwrite the files with the .Bak extension. Press A to
overwrite all files. If asked to create a new file or directory, the backup performed
earlier did not complete properly. Return to Step 10 and begin the procedure again.
27. At the C:\ > prompt, type D: and press Enter to verify the backup completed
without errors.
28. At the D:\ > prompt, type DIR/w and press Enter.
29. Verify BACKUP<DIR> appears.
Note: If BACKUP $$$ also appears, the backup failed. Return to Step 10 and begin the
procedure again.
30. At the D:\ > prompt, type CD BACKUP and press Enter.
31. At the D:\Backup> prompt, type DIR/w and press Enter.
Note: The B.25 software will list 20 or more files and folders in the directory. The last file
listed in the backup is BACKUP.OK with 0 bytes. This file is created when the backup
completes without errors and indicates that the backup was successful.
Note: The restore should take 5–10 minutes to perform. If the message RESTORE FAILED
appears (Figure 4-5), remove the new Compact Flash Disk and install the old
Compact Flash Disk. After installation of the old Compact Flash Adapter PCBA is
complete, perform Section 4.2 from the beginning.
7. Remove the Zip® disk when the RESTORE COMPLETE message appears.
(Refer to Figure 4-6.)
8. Verify the C:\ > prompt appears to indicate that the restore is complete.
9. Press Ctrl + Alt + Delete to restart the BacT/ALERT® 3D 60 software.
WARNING
If the Compact Flash Disk Data Backup Restore procedure cannot be
performed, all previous bottle data has been lost. The customer must
be advised to remove all bottles using the Ctrl + F10 function and
subculture the bottles. Failure to comply may result in invalid test
results.
IMPORTANT: If the Compact Flash Disk Data Backup Restore procedure cannot be
performed, the system’s site configuration must be restored to reflect the
current requirements of the site. This includes bottle test time, LIS parameters,
temperature, user preferences, instrument configurations, etc.
IMPORTANT: If the Compact Flash Disk Data Backup Restore procedure cannot be
performed, the bottle barcodes for i FA, i FN, and i PF must be activated. (Refer
to the BacT/ALERT® 3D i FA, BacT/ALERT® 3D i FN, and BacT/ALERT® 3D i PF
Bottle Barcode Activation Instructions.)
10. From the Main screen, verify the numbers shown on the Bottle Status table on
the Main screen are the same as recorded in Step 12 of Section 4.2. (Refer to
Figure 4-7.)
Bottle Status
Table
Software Version
Number
Load Bottles
Button
11. Verify the version of the software shown on the Main screen is the same as
recorded in Step 13 of Section 4.2. (Refer to Figure 4-7.)
12. Verify the bottle information is the same as recorded in Step 21 of Section 4.2.
13. Scan the Plastic SA Bottle barcode. (Refer to Figure 4-8.)
Note: To ensure that the algorithm for plastic bottles are used by the BacT/ALERT® 3D 60
instrument, each bottle type must be scanned. If the barcode reader does not scan the
barcodes shown in Figure 4-8, type the bottle ID below the barcode in the Bottle ID
field.
WARNING
If BPA/BPN or iFA/iFN/iPF barcodes are entered from the keyboard,
the bottle type in the Bottle Type field must be BPA/BPN or iFA/iFN/
iPF, as applicable. If the generic bottle type is shown in the Bottle
Type field after entering BPA/BPN or iFA/iFN/iPF barcodes from the
keyboard, then BPA/BPN or iFA/iFN/iPF bottle types have not been
activated. If this occurs, you must perform the BPA/BPN or iFA/iFN/
iPF activation, as applicable.
Note: After the barcode is scanned, the Accession Number field is highlighted.
Note: Plastic BPA/BPN Bottle or Plastic iFA/iFN/iPF Bottle barcodes will only work after
the BPA/BPN or iFA/iFN/iPF activation, as applicable. If you receive an Instrument
Status Code #923 Invalid Barcode Entry when scanning BPA, BPN, iFA, iFN, or
iPF barcodes, you must activate the BPA/BPN or iFA/iFN/iPF barcodes, as
applicable.
16. Click the Check button to exit the Load Mode screen.
Plastic SA Bottle
Plastic SN Bottle
Plastic FA Bottle
Plastic FN Bottle
Plastic PF Bottle
Plastic MP Bottle
19. Type 6 to access the Software Bottle Tables screen. (Refer to Figure 4-10.)
Line Scroll
Up Button
20. Click the Line Scroll Up button to view the algorithms listed in Table 4-1.
11 LES/Plastic BacT/ALERT SA
12 LES/Plastic BacT/ALERT SN
13 LES/Plastic BacT/ALERT FA
14 LES/Plastic BacT/ALERT FN
15 LES/Plastic BacT/ALERT PF
17 LES/Plastic BacT/ALERT MP
Note: The password is company confidential. Do not provide the password to the customer.
Note: Only 250 Megabyte Zip® disks should be used as the backup media with B.25
software.
6. Click the Previous Screen button until the Main screen appears.
7. Verify the Backup in Process icon is flashing in the upper left corner of the
screen.
Note: When the backup is finished, the Backup in Process icon will stop flashing.
Note: If the backup was successful, BACKUP <DIR> will appear. If the backup failed,
BACKUP $$$ will be displayed.
Note: If the backup fails to perform again, replace the Zip® disk with a new Zip® disk and
perform the procedure again.
Note: The backup can be used to restore the data if the system crashes during the UPS test
of the 3D 60.
1. Insert the Zip® disk that has the backup data copied to it.
2. Press Esc and type yes to exit the software.
3. At the C:\ > prompt, type RESTORE and press Enter.
Note: When the program has completed restoration successfully, a message will appear
indicating that the restore was complete.
IMPORTANT: Bottle readings are not taken when operating from the DOS level.
Note: The DOS Controller Module Backup is only available on systems with the new B.25.23
software version or higher.
Note: It is recommended that the Zip® disk be formatted. If the backup fails to perform,
replace the Zip® disk with a new Zip® disk and repeat Step 2 through Step 5.
Note: Ensure all drawers have been closed for a minimum of 30 minutes before continuing.
1. Ensure the thermometer is switched ON and is set to TEST, and remove the
probe cover.
2. Open the door and tape or lay the Digital thermometer next to the thermometer
in the Incubation Chamber.
3. Place the probe as close to the unit’s thermometer as possible.
4. Close the door and allow the temperature to stabilize for 20 minutes.
5. Open the door and switch thermometer to the HOLD position and remove.
6. Close the door and measure the incubator temperature.
7. Adjust the incubator temperature if the difference between the measured
temperature and the set point temperature of the incubator exceeds ±0.5 °C.
Continue verifying additional incubation modules as required.
8. Inform the customer of the difference between the measured temperature and
the internal thermometer. Note in comments section on the PM Check List who
was informed.
User Name
Password Check
Button
Cancel
Button
Change
Password
Previous Button
Screen
Button
Note: Field Service Engineers may use the User Name service and the Password
biomerieux3d. The User Name service is always valid. Do not provide the service
User Name or Password to customers.
2. Click the User Name field (field turns white) and enter service.
3. Click the Password field (field turns white) and enter biomerieux3d.
4. Click the Check button to log in.
5. At the Main screen, click the Next Screen button. (Refer to Figure 4-14.)
Next Screen
Button
Key Symbol
Button
Configure
Users
Button
Note: The password is company confidential. Do not provide the password to the customer.
Note: The Configure Users button is only active when the instrument is configured for
21 CFR Part 11 mode.
10. Verify the User Configuration screen appears. (Refer to Figure 4-16.)
Add User
Previous
Screen Check
Button Button
User Name
Field
Add User
Button User
List
Previous
Screen Check
Button Button
Add User
Button User
List
Previous
Screen Check
Button Button
User Name
Password Check
Button
Cancel
Button
Change
Password
Previous Button
Screen
Button
Note: Field Service Engineers may use the User Name service and the Password
biomerieux3d. The User Name service is always valid. Do not provide the service
User Name or Password to customers.
2. Click the User Name field (field turns white) and enter service.
3. Click the Password field (field turns white) and enter biomerieux3d.
4. Click the Check button to log in.
5. At the Main screen, click the Next Screen button. (Refer to Figure 4-21.)
Next Screen
Button
Key Symbol
Button
Configure
Users
Button
Note: The password is company confidential. Do not provide the password to the customer.
Note: The Configure Users button is only active when the instrument is configured for
21 CFR Part 11 mode.
10. Verify the User Configuration screen appears. (Refer to Figure 4-23.)
Delete User
Button
Previous
Screen Check
Button Button
Delete User
Button User
List
Previous
Screen Check
Button Button
Note: More than one user name may be selected. If the wrong user name is selected, click
the name again to deselect it.
Delete User
Button Highlighted
Users
Previous
Screen Check
Button Button
17. When all desired users have been deleted, click the Previous Screen button
until the Main screen appears.
User Name
Password Check
Button
Cancel
Button
Change
Password
Previous Button
Screen
Button
Note: Field Service Engineers may use the User Name service and the Password
biomerieux3d. The User Name service is always valid. Do not provide the service
User Name or Password to customers.
2. Click the User Name field (field turns white) and enter service.
3. Click the Password field (field turns white) and enter biomerieux3d.
4. Click the Check button to log in.
5. At the Main screen, click the Next Screen button. (Refer to Figure 4-28.)
Next Screen
Button
Key Symbol
Button
Configure
Users
Button
Note: The password is company confidential. Do not provide the password to the customer.
Note: The Configure Users button is only active when the instrument is configured for
21 CFR Part 11 mode.
10. Verify the User Configuration screen appears. (Refer to Figure 4-30.)
Clear User
Password
Button
Previous
Screen Check
Button Button
Clear User
Password
Button User
List
Previous
Screen Check
Button Button
13. Select the user name(s) associated with the passwords to clear from the user
list.
Note: More than one user name may be selected. If the wrong user name is selected, click
the name again to deselect it.
Note: When you clear a user password, the password is reset to the blank setting. The user
will then have to recreate a password the next time they try to log in.
Clear User
Password
Button Highlighted
Users
Previous
Screen Check
Button Button
Figure 4-32: Clear User Password Screen with User Names Highlighted
This chapter provides the steps necessary to perform the remove and replace actions
of the Lowest Replaceable Unit (LRU). This chapter explains the common field repair
procedures for replacing defective components of the BacT/ALERT® 3D 60.
This chapter also provides procedures that may be utilized to facilitate higher
maintenance. Replacement sections are divided into removal and installation
procedures for that purpose.
IMPORTANT: If you are replacing a component, it is extremely important that you complete all
procedures within the replacment section.
IMPORTANT: Before you perform any of the repair/replacement procedures in this chapter,
use the troubleshooting procedures in Chapter 7, Diagnostics and
Troubleshooting Information to isolate and diagnose instrument problems.
For a list of available spare parts, contact your local bioMérieux subsidiary or refer to
the most recent spare parts database.
Chapter Contents
WARNING
Potential Biohazardous Material! Instrument surfaces and
contaminated bottles are potentially biohazardous and should be
handled according to good laboratory practices. Observe universal
precautions when operating the instrument and when performing
maintenance or troubleshooting.
WARNING
The customer could have a subculture event present if the 3D 60 is
shut down for more than an hour. The B.25 software monitors for a
gap in the readings. If an hour gap is detected, the software will flag
all bottles affected with a Fault Code #80 Reading Gap Detected.
Failure to follow the procedures in this chapter or failure to attend to
fault conditions reported by the instrument within 1 hour may lead to
invalid test results and the need to subculture bottles.
5.2.1 Electronics
• CPU PCBA (P/N 750-0070-01 / 902)
• Quad Serial PCBA
• Compact Flash Adapter PCBA
• ModSig PCBA
• ModSig PCBA Fuses
• Step-5 PCBA
• Step-5 Motor
5.2.4 Agitation
• Fan
• Heater
• Blower Assembly
• Blower Motor Capacitor
• Step-5 Drive Agitation Linkage Assembly
This section is for the replacement of CPU PCBA (P/N 750-0070-902). CPU PCBA
(P/N 750-0070-01) is no longer manufactured. Replacement of CPU PCBA (P/N 750-
0070-01) will require an instrument upgrade to CPU PCBA (P/N 750-0070-902). Refer
to Section E.1 in Appendix E, Instrument Improvements for instructions on this
upgrade.
Set Point
Temperature
Software
Version
b. Record the bottle counts and software version from the Main screen in
Table 5-8.
c. Press Esc and type yes to exit the program. Wait for the C:\ > prompt to
appear.
Note: If the CPU is faulty to the point that access to the C:\ > prompt is impossible, go
directly to Section 5.3.2 and Section 5.3.3 to replace the CPU PCBA.
2. Simultaneously, press Ctrl + Alt + Delete. The screen will go blank and then
text will appear.
3. Press F1 at the bioMérieux splash screen to enter the CMOS Setup screen.
4. Record both the local date and time and CMOS date and time in Table 5-9 that
was printed out earlier.
5. Highlight Z. Save values to EEPROM and Exit and press Enter to reboot the
system. The BacT/ALERT® 3D 60 software will launch to the main screen.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-2.)
Top Panel
Screw Top Panel
Screw
Note: The replacement procedure is simplified if the Quad Serial PCBA cables are not
disconnected.
Screws
6. Tag and disconnect the cable assemblies from the CPU PCBA. (Refer to
Figure 5-4.)
IMPORTANT: Take special care not to damage the plastic bulkhead connectors when
disconnecting the connectors.
J14 J7
J2
J12
J19
J26
7. Remove six screws securing the CPU PCBA to the support standoffs.
1. Install the RAM (P/N 186201-1) into the SODIMM socket located at J103 on the
back of the CPU PCBA. (Refer to Figure 5-5.)
Note: The new CPU PCBA may already have the RAM installed when shipped. If the RAM
was installed in the new CPU PCBA when shipped, disregard Step 1.
2. Install the Compact Flash Disk into the slot on the back of the CPU PCBA.
3. Position the new CPU PCBA on the top panel.
4. Install the screws and standoffs to secure the CPU PCBA to the top panel.
5. Connect the cable assemblies to the CPU PCBA. (Refer to Figure 5-6 and
Table 5-1 for cable and component placement.)
Note: Do not remove the existing jumpers from J31 on the CPU PCBA.
J14 J7
J2
J12
J19
J26
J31
Note: The old connection locations are also given in Table 5-1 to aid in reconnection of the
cables to the new CPU PCBA, as some of the cables may still have the older
connections listed on them.
6. Carefully align the pins for the Quad Serial PCBA and plug into the CPU PCBA.
7. Install two screws to secure the Quad Serial PCBA to the CPU PCBA. (Refer to
Figure 5-3.)
8. Apply power to the instrument and turn on the UPS.
9. Return the top panel to the CLOSED position.
10. Open the door and install two screws to secure the top panel. (Refer to
Figure 5-7.)
Top Panel
Screw Top Panel
Screw
Note: Information was recorded in Table 5-9 during the Initial Checks procedure in
Section 5.3.1.
b. Enter the current Local Time in the Time column next to the Local Time at
Finish in Table 5-10.
c. Subtract the Local Time at Finish from Local Time at Start and enter the
information in the Time column next to Time Spent Replacing CPU PCBA
in Table 5-10.
d. Enter the CMOS Time from Table 5-9 in the Time column next to the
CMOS Time at Start in Table 5-10.
e. Enter the information from the Time column next to Time Spent Replacing
CPU PCBA in Table 5-10 in the Time column next to Time Spent
Replacing CPU PCBA in Table 5-10.
f. Subtract the Time Spent Replacing CPU PCBA from CMOS Time at Start
in Table 5-10.
Note: If the CMOS date and time were not noted in Table 5-9 due to a down system, set the
CMOS date and time to Coordinated Universal Time (UTC).
Note: The current UTC may be obtained from the National Technical Information Services
WWV radio station by calling 1-303-499-7111, or by accessing the Internet USNO
Master Clock at the following web address: http://tycho.usno.navy.mil/cgi-bin/timer.pl.
7. Highlight Z. Save values to EEPROM and Exit and press Enter to reboot the
system.
Note: The password is company confidential. Do not provide the password to the customer.
7. Click the Previous Screen button until the Main screen appears.
8. At the Main screen confirm the following:
a. That the display clock is running.
b. The 3D 60 appears on the pictorial as expected. If not, refer to Step 6b
and reenter configuration data.
9. Turn off the power to the 3D 60.
10. Verify that the 3D 60 continues to run for about 1½ minutes and then powers
down automatically.
11. Turn the power back on to the 3D 60, and verify that the 3D program starts up
and the Main screen appears.
12. Click the Next Screen button to access the Setup screen.
13. Enter the password 43432121.
Note: The password is company confidential. Do not provide the password to the customer.
Note: Only the rack and cells will be able to be enabled or disabled on the 3D 60.
Note: The steps listed here ensures that the 3D 60 functions as expected.
1. Verify the bottle counts are the same as those recorded in the initial checks.
2. Verify the optimal temperature settings.
3. At the Main screen, click the Load Bottles button. Without loading any bottles,
verify the following:
a. The Load Bottles screen appears.
b. Upon opening the door to the Incubation Chamber, verify that the yellow
LED illuminates as well as the cell indicator lights for available cells.
Verify that the yellow LED extinguishes when the door to the Incubation
Chamber is closed.
4. Click the Check button to exit the Load Bottles screen.
5. Refer to Section 5.42 and perform the Post Service Validation procedure.
6. Record the procedure in the Service Log per local procedures.
7. Return the system to operation.
This section is for the replacement of CPU PCBA (LBC-586 PLUS Board, P/N 750-
0070-01). Although CPU PCBA (P/N 750-0070-01) is no longer manufactured,
refurbished boards may be available. If refurbished boards are not available,
replacement of CPU PCBA (P/N 750-0070-01) will require an instrument upgrade to
CPU PCBA (P/N 750-0070-902). Refer to Section E.1 in Appendix E, Instrument
Improvements for instructions on this upgrade.
b. Record the bottle counts and software version from the Main screen in
Table 5-8.
c. Write down the Controller Module ID value displayed. (Refer to
Figure 5-1.)
d. Press Esc and type yes to exit the program. Wait for the C:\ > prompt to
appear.
Note: If the CPU is faulty to the point that access to the C:\ > prompt is impossible, go
directly to Section 5.4.2 and Section 5.4.3 to replace the CPU PCBA.
2. Simultaneously, press Ctrl + Alt + Delete. The screen will go blank and then
text will appear.
3. While the system is rebooting, press the Delete key to enter the CMOS SETUP
menu.
4. Ensure the CMOS SETUP menu appears. (Refer to Figure 5-8.)
7. Record both the local date and time and CMOS date and time in Table 5-9 that
was previously printed out.
8. Press the Esc key to exit the CMOS Setup Utility screen.
9. Press F10 + Y and press Enter to save and exit the BIOS Main screen.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-10.)
Top Panel
Screw Top Panel
Screw
Note: The replacement procedure is simplified if the Quad Serial PCBA cables are not
disconnected.
Screws
Note: Connectors should be removed from the CPU PCBA in the order they are given.
Screws
Screws
J2 J3
D1
J3
J1
D2
C2
J2
C3 C4
J4
C1
+
J6
C6
R2
R3
R1
R4
C5
R5
J4 J7
C7
U6
U4
U7
J9
J8
C8
J11
J10
U10
J18
J5
Y2
U9
R16
U8
J18
J17
U15
U3
U5
U1
U2
Y1
J12
U14
U11
LITHIUM
KEEPER
II
LTC-3PN
J14
J15
J16
J5
J13
J13
J20
U16
T1
J19
J21
J19
U18
J27
U17
J23
J24
J25
J22
J26
J22
J27
J41
J28
J31
J23
J21
J30
U19
U20
LS1
U22
J33
T2
J31
C9
D6 D5 D4 D3
U21
+ +
C11
C10
+
J37
J35
J34 Y3
U25 U26
U24
J38
U23
U28
U29
U27
J39
SIM1
J40
4. Connect the cable assemblies to the CPU PCBA. (Refer to Table 5-2 for cable
and component placement.)
Connector Description
Location
J13, J19, 9-Pin Connector from CPU BIOS Clock Battery (if
J23 installed)
Top Panel
Screw Top Panel
Screw
Note: Information was recorded in Table 5-9 during the Initial Checks procedure in
Section 5.4.1.
b. Enter the current Local Time in the Time column next to the Local Time at
Finish in Table 5-10.
c. Subtract the Local Time at Finish from Local Time at Start and enter the
information in the Time column next to Time Spent Replacing CPU PCBA
in Table 5-11.
d. Enter the CMOS Time from Table 5-9 in the Time column next to the
CMOS Time at Start in Table 5-11.
e. Enter the information from the Time column next to Time Spent Replacing
CPU PCBA in Table 5-10 in the Time column next to Time Spent
Replacing CPU PCBA in Table 5-11.
f. Subtract the Time Spent Replacing CPU PCBA from CMOS Time at Start
in Table 5-10.
Note: If the CMOS date and time were not noted in Table 5-9 due to a down system, set the
CMOS date and time to Coordinated Universal Time (UTC).
Note: The current UTC may be obtained from the National Technical Information Services
WWV radio station by calling 1-303-499-7111, or by accessing the Internet USNO
Master Clock at the following web address: http://tycho.usno.navy.mil/cgi-bin/timer.pl.
5. To change the date or time settings, press Up or Down Arrow key on the
keyboard to highlight the value you want to change.
6. Enter the determined time and/or date from New CMOS Date and Time in
Table 5-11 and press Enter.
7. If no changes were made, press the Esc key, select Y, and press Enter to exit
the BIOS.
8. If changes were made, press F10, select Y, and press Enter to save to CMOS
and exit the BIOS.
9. Verify the system functions normally and that there are not any Fault Codes
present.
10. If Fault Code #71 Invalid Time Stamp or #80 Reading Gap Detected occurs,
verify that the CMOS date and time are correct.
11. Refer to Section 5.3.5 and perform the Re-enter Configuration and Operation
Option Date.
12. Refer to Section 5.3.6 and perform the CPU PCBA Final Checks procedure.
Note: The CPU BIOS batteries should be replaced 10 years after their manufacture date.
The CPU BIOS battery has the part number and date code printed on it with the date
format of MMYYYY.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-15.)
Top Panel
Screw Top Panel
Screw
Screws
Note: The replacement procedure is simplified if the Quad Serial PCBA cables are not
disconnected.
Note: If the CPU BIOS Clock Battery is installed on an older model CPU PCBA, proceed to
Step 8. If the CPU BIOS Clock Battery is installed on a newer model CPU PCBA,
proceed to Step 9.
8. Disconnect the 9-pin connector of the battery assembly from J23, J19, and J13.
(Refer to Figure 5-17.)
9-Pin Connector
9. Disconnect the 3-pin connector from J14 on the CPU PCBA. (Refer to
Figure 5-18.)
J14
J17
CPU BIOS
Clock Battery
1. Position the CPU BIOS Clock Battery on the Velcro® dot on the CPU PCBA.
Note: Only perform Step 2 if the CPU BIOS Clock Battery is being installed on an older
model CPU PCBA. Only perform Step 3 and Step 4 if the CPU BIOS Clock Battery is
being installed on a newer model CPU PCBA. The caution statement below Step 4
only applies to the newer model CPU PCBA.
2. Connect the 9-pin connector of battery assembly to J23, J19, and J13. (Refer to
Figure 5-17.)
3. Connect the 3-pin connector to J14 on the CPU PCBA. (Refer to Figure 5-18.)
4. Ensure the jumper at J17 has been removed.
CAUTION: Failure to remove the jumper at J17 will shorten the battery
life.
8. Install two screws to secure the Quad Serial PCBA to the CPU PCBA. (Refer to
Figure 5-16.)
9. Connect the two cable assemblies to the Quad Serial PCBA.
10. Return the top panel to the CLOSED position.
11. Open the door and install two screws to secure the top panel. (Refer to
Figure 5-15.)
12. Close the door.
13. Refer to Section 4.1.2 and perform the Power Up procedure.
14. Refer to Section 5.42 and perform the Post Service Validation procedure.
15. Record the procedure in the Service Log per local procedures.
16. Return the system to operation.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-20.)
Top Panel
Screw Top Panel
Screw
Screws
2. Carefully align the pins for the Quad Serial PCBA and plug into the CPU PCBA.
3. Install two screws to secure the Quad Serial PCBA to the CPU PCBA. (Refer to
Figure 5-21.)
4. Connect the two cable assemblies to the Quad Serial PCBA.
5. Return the top panel to the CLOSED position.
6. Open the door and install two screws to secure the top panel. (Refer to
Figure 5-20.)
7. Close the door.
8. Refer to Section 4.1.2 and perform the Power Up procedure.
9. Refer to Section 5.42 and perform the Post Service Validation procedure.
10. Record the procedure in the Service Log per local procedures.
11. Return the system to operation.
This procedure is for the replacement of the Compact Flash Disk. The type of CPU
PCBA in use will determine the location of the Compact Flash Disk. In Controller
Modules that still have the older version of the CPU PCBA (P/N 750-0070-01), the
Compact Flash Disk is located in the Compact Flash Adapter PCBA. If the Controller
Module has the newer version of the CPU PCBA (P/N 750-0070-902), the Compact
Flash Adapter PCBA has been removed and the Compact Flash Disk is located on the
CPU PCBA.
Replacing the Compact Flash Disk or loading firmware without performing proper data
backup and restore procedures can, under certain circumstances, result in an
incorrect algorithm being applied to plastic bottles. If problems are encountered after
installation of the Compact Flash Disk, refer to Chapter 7, Diagnostics and
Troubleshooting Information to troubleshoot the problem.
Note: If you are removing the Compact Flash Disk as part of another maintenance
procedure and not as part of the replacement procedure for the Compact Flash Disk
(i.e. to facilitate higher maintenance), it is not required that you perform the Compact
Flash Disk Data Backup procedure; however, it is recommended.
4. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-22.)
Top Panel
Screw Top Panel
Screw
Compact Flash
Disk
Black Plunger
8. If the newer version of the CPU PCBA is in use, remove the Compact Flash
Disk from the CPU PCBA. (Refer to Figure 5-24.)
IMPORTANT: If you are installing the Compact Flash Disk as part of another maintenance
procedure and not as part of the replacement procedure for the Compact Flash
Disk (i.e. to facilitate higher maintenance), and you did not perform the 3D 60
Module Data Backup during removal of the Compact Flash Disk, do not perform
the 3D 60 Module Data Backup Restore procedure.
This replacement procedure is only for 3D 60 Modules that still have the older version
of the CPU PCBA (P/N 750-0070-01). If the 3D 60 Module has the newer version of
the CPU PCBA (P/N 750-0070-902), the Compact Flash Adapter PCBA has been
removed and the Compact Flash Disk is located on the CPU PCBA. Refer to
Section E.1 to ensure that all upgrades associated with the CPU PCBA have been
completed.
3. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-25.)
Top Panel
Screw Top Panel
Screw
6. Remove the Compact Flash Disk from the Compact Flash Adapter PCBA by
pressing the black release plunger and set the Compact Flash Disk aside.
(Refer to Figure 5-26.)
Compact Flash
Disk
Black Plunger
Screws
7. Disconnect the J7 power cable from the Compact Flash Adapter PCBA. (Refer
to Figure 5-27.)
J3
J7
8. Remove four screws securing the Compact Flash Adapter PCBA to the support
standoffs. (Refer to Figure 5-26.)
9. Disconnect the J3 cable from the Compact Flash Adapter PCBA. (Refer to
Figure 5-27.)
10. Remove the Compact Flash Adapter PCBA.
Note: Ensure the J7 power cable is routed around the right side and on top of the Compact
Flash Adapter PCBA before installing the mounting screws.
3. Install four screws to secure the Compact Flash Adapter PCBA to the support
standoffs. (Refer to Figure 5-26.)
4. Connect the J7 power cable to the Compact Flash Adapter PCBA. (Refer to
Figure 5-27.)
5. Install the Compact Flash Disk in the Compact Flash Adapter PCBA. (Refer to
Figure 5-26.)
6. Ensure that the Compact Flash Disk is locked in place.
Note: A soft click will be heard when the Compact Flash Disk locks in place.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-28.)
Top Panel
Screw Top Panel
Screw
Note: JP4 and JP28 connectors do not have any cables connected to them.
Step-5
PCBA
ModSig
PCBA
8. Cut the tie wraps around the bundled cables on the ModSig PCBA.
Note: The cables in Step 9 through Step 19 should be disconnected in the exact order
listed.
Screws
Nuts Nuts
21. Remove eight screws securing the ModSig PCBA to the support standoffs.
(Refer to Figure 5-29.)
22. Disconnect the MCJ4 cable from the J4 connector on the Step-5 PCBA. (Refer
to Figure 5-29.)
23. Remove the ModSig PCBA from the standoffs.
24. Remove two screws securing the ModSig PCBA to the rear bulkhead.
25. Remove two hex nuts from the LIS port on the rear bulkhead. (Refer to
Figure 5-30.)
26. Remove two hex nuts from the MOUSE port on the rear bulkhead.
27. Remove two hex nuts from the BARCODE port on the rear bulkhead.
28. Remove the knurled nut from the EXT SPKR port on the rear bulkhead.
29. Remove the ModSig PCBA from the rear bulkhead and set it aside.
1. Align the rear bulkhead to the mounting holes and peripheral ports on the
ModSig PCBA.
2. Install two screws to secure the ModSig PCBA to the rear bulkhead.
3. Install the knurled nut on the EXT SPKR port on the rear bulkhead. (Refer to
Figure 5-30.)
4. Install two hex nuts on the BARCODE port on the rear bulkhead.
5. Install two hex nuts on the MOUSE port on the rear bulkhead.
6. Install two hex nuts on the LIS port on the rear bulkhead.
7. Position the ModSig PCBA on the support standoffs.
Note: Ensure that there are not any cables under the ModSig PCBA.
8. Loosely install two screws in the rear panel to allow for proper alignment of the
ModSig PCBA.
9. Loosely install eight screws to align the ModSig PCBA on the support standoffs.
10. Ensure ModSig PCBA is aligned correctly.
11. Tighten eight screws to secure the ModSig PCBA to the support standoffs.
12. Tighten two screws in the rear panel.
13. Connect the MCJ4 cable to the J4 connector on the Step-5 PCBA. (Refer to
Figure 5-29.)
Note: The cables in Step 14 through Step 24 should be connected in the exact order listed.
33. Record the procedure in the Service Log per local procedures.
34. Return the system to operation.
The ModSig PCBA contains seven fuses identified as F1, F2, F3, F4, F5, F6, and F7.
F1 and F6 are 3.15-Amp 5x20mm type FST Time Lag fuses
(P/N 870-0008-22). F2, F3, F4, and F7 are 1-Amp 5x20mm type FST Time Lag fuses
(P/N 870-0008-17). Fuse F5 is a 0.5-Amp 5x20mm type FST Time Lag fuse (P/N 870-
0008-14).
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-31.)
Top Panel
Screw Top Panel
Screw
F2
F7
F3
F5
F6
F4 F1
CAUTION: The new fuse must be the same amperage and voltage rating
as the removed fuse. Overfusing could damage the equipment and
create a potential fire hazard.
1. Install the fuse in the retaining clip. (Refer to Table 5-3 for fuse information and
Figure 5-32 for fuse location.)
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-33.)
Top Panel
Screw Top Panel
Screw
Note: Feed the MCJ3 cable underneath the Step-5 PCBA and plug it in.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-35.)
Top Panel
Screw Top Panel
Screw
MCJ3
6. Remove the left side panel by lifting from the top edge of the panel.
7. Locate the setscrew points on the Motor Offset Round. (Refer to Figure 5-37.)
Motor Offset
Round
Drive Agitation
Linkage
Setscrew
Points
9. Rotate the Drive Agitation Linkage 90 degrees clockwise and remove the
second setscrew.
10. Remove four motor mounting screws from the pressed nuts. (Refer to
Figure 5-38.)
Motor
Mounting
Screws
11. Pull the Step-5 Motor out and to the right in the direction shown in Figure 5-39.
Step-5
Motor
12. Remove four nuts and washers securing the motor mounting plate to the Step-5
Motor.
13. Set the Step-5 Motor aside.
Note: Do not tighten the motor mounting screws all the way.
4. Tighten the motor mounting screws in an X pattern until all screws are
completely tightened. Tighten the upper left screw first, the lower right screw
second, the upper right screw third, and the the lower left screw fourth. (Refer to
Figure 5-40 for the tightening sequence.)
First
Third
Fourth Second
5. Position the motor offset round in place with proper orientation to the motor
shaft flats.
Note: Support the opposite end of the Step-5 Motor when pushing the motor offset round on
the Motor Shaft to prevent pushing the isolation grommets from their locations in the
chassis holes.
6. Install the first setscrew in the motor offset round. (Refer to Figure 5-37.)
7. Rotate the drive agitation linkage 90 degrees clockwise and install the second
setscrew.
8. Ensure that the motor offset round and the drive agitation linkage are in the
correct position.
9. Insert paper cardstock over the wide foam at the top of the door.
Note: The paper cardstock will prevent damage to the foam when installing the left side
panel.
12. Connect the MCJ3 cable to the J3 connector. (Refer to Figure 5-36.)
13. Return the top panel to the CLOSED position.
14. Open the door and install two screws to secure the top panel. (Refer to
Figure 5-35.)
15. Close the door.
16. Refer to Section 4.1.2 and perform the Power Up procedure.
17. Refer to Section 5.42 and perform the Post Service Validation procedure.
18. Record the procedure in the Service Log per local procedures.
19. Return the system to operation.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-41.)
Top Panel
Screw Top Panel
Screw
Screws
J3 Connector
7. Remove two screws securing the Outlet Thermistor PCBA to the panel.
8. Remove the Outlet Thermistor PCBA and set it aside.
Note: The Outlet Thermistor PCBA should be positioned so that the writing on the PCBA is
upside down. Refer to Figure 5-42 for the proper orientation.
2. Install two screws to secure the Outlet Thermistor PCBA to the panel. (Refer to
Figure 5-42.)
3. Connect the cable to the J3 connector on the Outlet Thermistor PCBA.
4. Insert paper cardstock over the wide foam at the top of the door.
Note: The paper cardstock will prevent damage to the foam when installing the left side
panel.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-43.)
Top Panel
Screw Top Panel
Screw
Screws
J3 Connector
7. Remove two screws securing the Inlet Thermistor PCBA to the panel.
8. Remove the Inlet Thermistor PCBA and set it aside.
Note: The paper cardstock will prevent damage to the foam when installing the right side
panel.
WARNING
Do not insert any object, other than a Zip® disk, into the Zip® drive
under any circumstances.
CAUTION: Do not forcibly remove the Zip® disk from the instrument.
Forcibly removing the Zip® disk may cause damage to the Zip® disk or
Zip® drive and may cause the system to lock up.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-45.)
Top Panel
Screw Top Panel
Screw
Note: Only perform Step 5 if the Compact Flash Adapter PCBA is installed. If the Compact
Flash Adapter PCBA is not installed, proceed to Step 6.
5. Remove four screws securing the Compact Flash Adapter PCBA to the chassis.
(Refer to Figure 5-46.)
Note: Do not remove the connections to the Compact Flash Adapter PCBA. Allow the
Compact Flash Adapter PCBA to hang down.
Screws
6. Disconnect the IDE cable from the Zip® Drive. (Refer to Figure 5-47.)
IDE Cable
Main Power
Connector
Note: The main power connector is a connector with blue and black wires. Do not pull on the
wires to disconnect the connector.
8. Remove four screws securing the Zip® Drive to the instrument. (Refer to
Figure 5-48.)
IMPORTANT: Hold on to the Zip® Drive as you remove the last screw securing the Zip® Drive
to the chassis. The Zip® Drive can fall and be damaged when the last screw is
removed.
Screws
Top Panel
Cutout
Note: There are typically four to six screws securing the Zip® Drive to the mounting bracket.
The number of screws will depend on where the screws are located. The screws may
be installed from the side, with three on each side of the mounting bracket, or on the
bottom of the mounting bracket. Refer to Figure 5-50.
Screws
Screws
(3 ea. Side)
13. Remove the Zip® Drive from the mounting bracket and set it aside.
Note: There are typically four to six screws securing the Zip® Drive to the mounting bracket.
The number of screws will depend on where the screws are located. The screws may
be installed from the side, with three on each side of the mounting bracket, or on the
bottom of the mounting bracket. Refer to Figure 5-50.
IMPORTANT: Hold onto the Zip® Drive until at least one screw has been installed.
6. Connect the main power connector to the Zip® Drive. (Refer to Figure 5-47.)
7. Disconnect the IDE cable from the Zip® Drive.
Note: Only perform Step 8 and Step 9 if the Compact Flash Adapter PCBA is installed. If the
Compact Flash Adapter PCBA is not installed, proceed to Step 10.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-51.)
Top Panel
Screw Top Panel
Screw
6. Disconnect the HBJ1/JBP1 wiring connector coming from the Heater and
Blower to TB1 and Relays. (Refer to Figure 5-52.)
BHP1
Connector
HBJ1/HBP1
Connector
Screws
9. Lift the Power Panel Assembly up and out of the instrument through the top
panel and set it aside.
Note: The top panel may have to be pushed beyond the normal locked position of the Gas
Spring Assembly in order to clear the Power Panel Assembly from the instrument.
1. Lower the Power Panel Assembly down and into the instrument through the top
panel.
Note: The top panel may have to be pushed beyond the normal locked position of the Gas
Spring Assembly in order to place the Power Panel Assembly in the instrument.
2. Install two screws to secure the Power Panel Assembly to the chassis. (Refer to
Figure 5-53.)
3. Connect the BHP1 wiring connector coming to the electronics compartment to
the power supply. (Refer to Figure 5-52.)
4. Connect the HBJ1/JBP1 wiring connector coming from the Heater and Blower
to TB1 and Relays.
5. Insert paper cardstock over the wide foam at the top of the door.
Note: The paper cardstock will prevent damage to the foam when installing the right side
panel.
Note: Individual power supply subassembly components are not repaired or replaced in
Domestic 3D 60s. Individual components are spared as parts and replaced in
International 3D 60s only.
2. Disconnect the four Transformer wires from the PEM. (Refer to Table 5-4 and
Figure 5-54.)
Color Name
Brown PWRSWA
Orange PWRSWB
Red PWRSWC
Yellow PWRSWD
TB1-4B
TB1-1B
PEM
3. Remove the bolt securing the Transformer to the Power Panel Assembly
chassis. (Refer to Figure 5-55.)
Bolt
4. Remove the screws from Terminal Block TB1-1B and TB1-4B. (Refer to
Figure 5-54.)
5. Disconnect the Green and Brown wires from TB1-4B.
6. Disconnect the Red and Blue wires from TB1-1B.
7. Remove the Transformer and set it aside.
Note: Individual power supply subassembly components are not repaired or replaced in
Domestic 3D 60s. Individual components are spared as parts and replaced in
International 3D 60s only.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-56.)
Top Panel
Screw Top Panel
Screw
Bolt
7. Pull the Transformer out of the instrument far enough to easily access the
Terminal Board (TB1).
8. Remove the screws from TB1-2B and TB1-3B. (Refer to Figure 5-58.)
TB1-3B
TB1-2B
MOV
4. Install the bolt to secure the Transformer to the Power Panel Assembly chassis.
5. Insert paper cardstock over the wide foam at the top of the door.
Note: The paper cardstock will prevent damage to the foam when installing the right side
panel.
Note: Individual power supply subassembly components are not repaired or replaced in
Domestic 3D 60s. Individual components are spared as parts and replaced in
International 3D 60s only.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-59.)
Top Panel
Screw Top Panel
Screw
Screws
Lock Nut
Cable
Connectors
K1 K2 K3 K4
8. Gently pull the solid state relay off of the studs, taking care not to damage the
wires attached to the terminals.
9. Remove four slotted screws securing the cable (lug) connectors to the defective
solid state relay.
10. Tag and disconnect the cable connections from the solid state relay.
11. Remove the solid state relay from the Power Panel assembly and set it aside.
1. Position cable connections on the solid state relay as previously tagged. (Refer
to Figure 5-60.)
2. Install four slotted screws to secure the cable (lug) connectors to the solid state
relay.
3. Position the solid state relay on the Power Panel assembly.
Note: The solid state relays mount inverted with terminals 3 and 4 at the top. (Refer to
Figure 5-60.)
4. Install the lock nut to secure the solid state relay to the Power Panel assembly.
5. Insert paper cardstock over the wide foam at the top of the door.
Note: The paper cardstock will prevent damage to the foam when installing the right side
panel.
Note: Individual power supply subassembly components are not repaired or replaced in
Domestic 3D 60s. Individual components are spared as parts and replaced in
International 3D 60s only.
J2 Connector
J1 Connector
3. Remove four mounting screws from the rear of the Power Panel Assembly.
(Refer to Figure 5-62.)
IMPORTANT: Hold on to the DC Power Supply as you remove the last screw securing the DC
Power Supply to the Power Panel Assembly. The DC Power Supply can fall and
be damaged when the last screw is removed.
Screws
IMPORTANT: Hold on to the DC Power Supply until at least one screw is installed to secure
the DC Power Supply to the Power Panel Assembly. The DC Power Supply can
fall and be damaged until at least one screw is installed.
Note: Individual power supply subassembly components are not repaired or replaced in
Domestic 3D 60s. Individual components are spared as parts and replaced in
International 3D 60s only.
2. Disconnect the ground wire lug from the PEM. (Refer to Figure 5-63.)
PEM
Ground
Wire Lug
3. Disconnect four AC input cable lugs from the PEM. (Refer to Figure 5-64.)
AC Input
Cable Lugs
4. Release the tabs securing the PEM to the System Power Panel Assembly.
5. Remove the PEM and set it aside.
Connector Description
Location
4. Connect the ground wire lug to the PEM. (Refer to Figure 5-63.)
5. Refer to Section 5.16.2 and perform the System Power Panel Assembly
Installation procedure.
6. Installation procedure.
7. Refer to Section 5.42 and perform the Post Service Validation procedure.
8. Record the procedure in the Service Log per local procedures.
9. Return the system to operation.
The Power Entry Module (P/N 805-0017-02) is located in the rear of the Power Panel
Assembly. (Refer to Figure 5-62.) When configured for 115 VAC, the PEM has one
6.3 Amp 250 V slow-blow fuse (P/N 870-0008-25). In addition, a conversion clip is
located across the neutral line in the fuse holder. (Refer to Figure 5-67.) When
configured for 230 VAC, the PEM has two 5.0 Amp 250 V slow-blow fuses (P/N 870-
0008-24). (Refer to Figure 5-68.)
CAUTION: The new fuse MUST be the same size rating as the removed
fuse. Over fusing could damage the equipment and create a fire hazard.
Note: Failure to properly install the spring clip will result in no power to the unit. (Refer to
Figure 5-66.)
Note: Steps for replacing fuses in the 230 VAC version are the same as the steps taken with
the 115 VAC version, with the exception of the fuse size and the 115 V conversion clip
(which is not used). Remove and dispose of the clip if one is present. Refer to
Figure 5-68 for 230 VAC version.
Conversion Clip
6.3A 250V
Slow-blow Fuse
5.0A 250V
Slow-blow Fuse
(2 Required)
6. Refer to Section 5.42 and perform the Post Service Validation procedure.
7. Record the procedure in the Service Log per local procedures.
8. Return the system to operation.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-69.)
Top Panel
Screw Top Panel
Screw
Bolt
7. Pull the Transformer out of the instrument far enough to easily access the
Terminal Board (TB1).
8. Remove the screws from TB1-2A and TB1-4A. (Refer to Figure 5-71.)
TB1-4A
TB1-2A
Note: There are two wire clamps on the two bottom screws that hold wiring in place.
Note: Hold on to the washers as the nuts and screws are removed, as the washers can fall
into the chassis.
Nuts
Wire Clamps
Washers
Screws
14. Remove the fan from the chassis and set it aside.
Note: The paper cardstock will prevent damage to the foam when installing the right side
panel.
Top Panel
Screw Top Panel
Screw
HTR-2
HTR-2 Terminal
Terminal Connector
Mounting
Support
Tab
HTR-1
Terminal
8. Gently pull down on the locking tab and remove the Heater by pulling outward
from the bottom of both ends.
CAUTION: Do not pull down on the locking tab too hard. The locking tab
can be damaged, losing its ability to lock the heater in place.
Note: The paper cardstock will prevent damage to the foam when installing the left side
panel.
Top Panel
Screw Top Panel
Screw
Screw
TC0-2
Thermal
Switch Connector
TC0-1
Connector Screw
Thermal Switch
Mounting Bracket
8. Remove the Thermal Switch from the thermal switch mounting bracket and set
aside.
IMPORTANT: Ensure that the Thermal Switch is not touching the Heater coil.
2. Install two screws and lock washers to secure the Thermal Switch to the thermal
switch mounting bracket. (Refer to Figure 5-76.)
3. Connect the TC0-1 and TC0-2 connectors to the Thermal Switch.
4. Insert paper cardstock over the wide foam at the top of the door.
Note: The paper cardstock will prevent damage to the foam when installing the left side
panel.
Top Panel
Screw Top Panel
Screw
Blower
Mounting
Nuts
BLP1/BLJ1
Note: The paper cardstock will prevent damage to the foam when installing the left side
panel.
2. Remove the boot from the Blower Motor Capacitor. (Refer to Figure 5-79.)
Boot
3. Make a note of the color and location of the cable (lug) connectors on the
Blower Motor Capacitor.
4. Disconnect the cable (lug) connectors from the Blower Motor Capacitor. (Refer
to Figure 5-80.)
Cable (Lug)
Connectors
Lock Nut
5. Remove the lock nut from the Blower Motor Capacitor retaining clamp.
6. Slide the Blower Motor Capacitor out of the retaining clamp and set it aside.
4. Install the boot on the Blower Motor Capacitor. (Refer to Figure 5-79.)
5. Refer to Section 5.16.2 and perform the System Power Panel Assembly
Installation procedure.
6. Refer to Section 5.42 and perform the Post Service Validation procedure.
7. Record the procedure in the Service Log per local procedures.
8. Return the system to operation.
Reference
Thermometer
Reference
Thermometer
3. Using a flat blade screwdriver, rotate the battery cap counterclockwise and
remove the battery cap.
4. Remove the battery from the Reference Thermometer.
2. Remove bottles by using CTRL + F10 function or unload all of the bottles from
the defective rack.
Note: Anonymous bottles cannot be relocated to another cell until a bottle ID has been
entered for the bottle.
4. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-83.)
Top Panel
Screw Top Panel
Screw
Screws
Rack
Cable
Clamp
Screws
10. Disconnect the cables from the Rack Controller PCBA. (Refer to Figure 5-86.)
Connections
11. Loosen two captive screws located at either end of the rack being removed.
(Refer to Figure 5-87.)
Captive Screws
Note: Step 13 through Step 15 should only be performed if removing the Rack Assembly for
replacement. Step 13 through Step 15 should not be performed if removing the Rack
Assembly to facilitate higher maintenance.
13. Remove two screws from the rack cable clamp of the new Rack Assembly.
14. Remove the ground cable from the old rack cable clamp.
Note: Retain the old parts and return them with the defective Rack Assembly.
15. Disconnect the Rack cable connectors (daisy-chained cables) from the Rack
Interface PCBA connectors.
Note: Step 1 through Step 3 should only be performed if installing a new Rack Assembly.
Step 1 through Step 3 should not be performed if installing the Rack Assembly after
removing the Rack Assembly to facilitate higher maintenance.
Note: Retain these parts and return them with the defective Rack Assembly.
Note: Anonymous bottles cannot be relocated to another cell until a bottle ID has been
entered for the bottle.
4. Remove five screws securing the Bottle Holder Cover to the Rack Assembly.
(Refer to Figure 5-88.)
Note: Older configurations of the Rack Assembly may have six screws.
Rack Assembly
Screws
5. Remove the Bottle Holder Cover from the Rack Assembly and set it aside.
Note: Older configurations of the Rack Assembly may have six screws.
Link Assembly
Exterior Bushing
Screw
Interior Bushing
Screw
d. Grasp the interior nut with a wrench and turn the screwdriver
counterclockwise to remove the nut securing the Link Assembly to the link
cover. (Refer to Figure 5-90.)
Exterior Nut
Interior Nut
Link Cover
e. Insert a screwdriver into the head of the exterior bushing screw. (Refer to
Figure 5-89.)
f. Grasp the exterior nut with a wrench and turn the screwdriver
counterclockwise to remove the nut securing the Link Assembly to the
chassis.
g. Remove the Link Assembly from the chassis.
3. To remove the right side Link Assemblies:
a. Remove the right side panel by lifting from the top edge of the panel.
b. Determine which Link Assembly needs to removed.
c. Insert a screwdriver into the head of the bushing screw. (Refer to
Figure 5-91.)
Link Assembly
Bushing Screw
d. Grasp the nut with a wrench and turn the screwdriver counterclockwise to
remove the nut securing the Link Assembly to the chassis. (Refer to
Figure 5-92.)
Nuts
Note: The paper cardstock will prevent damage to the foam when installing the right side
panel.
Note: The paper cardstock will prevent damage to the foam when installing the left side
panel.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 5-93.)
Top Panel
Screw Top Panel
Screw
Linkage Assembly
Connecting Link
Bottom
Bushing Nut
7. From the reverse side of the connecting link, remove the bottom bushing screw
from the Drive Agitation Linkage Assembly.
8. Locate the setscrew points on the Motor Offset Round. (Refer to Figure 5-95.)
Motor Offset
Round
Bushing Screw
Drive Agitation
Linkage
Setscrew
Points
10. Rotate the Drive Agitation Linkage 90 degrees clockwise and remove the
second setscrew.
11. Slide the entire Drive Agitation Linkage Assembly, with the Motor Offset Round
attached, up and out of the instrument.
Note: The bottom part of the Linkage slides through a cutout hole in the support panel.
Note: When removing the entire Drive Agitation Linkage Assembly with the Motor Offset
Round attached, pull the Motor Offset Round slightly forward to slide the Motor Offset
Round off of the Step-5 Motor shaft.
12. Remove the Drive Agitation Linkage Assembly from the Motor Offset Round by
removing the bushing nut on the reverse side of the Motor Offset Round. (Refer
to Figure 5-96.)
Motor Offset
Round
Drive Agitation
Linkage
Bushing Nut
Figure 5-96: Drive Agitation Linkage Assembly with Motor Offset Round
13. Remove the bushing screw from the Drive Agitation Linkage Assembly. (Refer
to Figure 5-95.)
14. Set the Drive Agitation Linkage Assembly aside.
5. Position the motor offset round in place with proper orientation to the motor
shaft flats.
Note: Support the opposite end of the Step-5 Motor when pushing the motor offset round on
the Motor Shaft to prevent pushing the isolation grommets from their locations in the
chassis holes.
6. Insert the Offset Round Spacing tool between the system chassis and the Motor
Offset Round.
7. Install the first setscrew in the motor offset round. (Refer to Figure 5-95.)
8. Rotate the drive agitation linkage 90 degrees clockwise and install the second
setscrew.
9. Remove the Offset Round Spacing tool.
10. Ensure that the motor offset round and the drive agitation linkage are in the
correct position.
11. Align the Drive Agitation Linkage Assembly’s bottom bushing screw hole with
the hole in the connecting link. (Refer to Figure 5-94.)
12. From the reverse side of the connecting link, install the bottom bushing screw
into the Drive Agitation Linkage Assembly.
13. Install the bottom bushing nut. (Refer to Figure 5-94.)
14. Insert paper cardstock over the wide foam at the top of the door.
Note: The paper cardstock will prevent damage to the foam when installing the left side
panel.
Top Panel
Screw Top Panel
Screw
Note: The spring assembly (socket) and the support bracket (ball joint) serve as a ball joint
and socket assembly. It will require force to separate the two from each other.
Protective Cap
Notch for
Screwdriver
7. Remove the bottom protective cap with a small flat blade screwdriver. (Refer to
Figure 5-99.)
8. While holding the support bracket, pull and separate the spring assembly from
the bottom support bracket.
Protective Cap
Notch for
Screwdriver
3. While holding the support bracket, push the spring assembly into the top
support bracket. (Refer to Figure 5-98.)
4. Install the bottom protective cap.
Top Panel
Screw Top Panel
Screw
Note: The door switch LEDs are located on the ModSig PCBA at D55 and D66. (Refer to
Figure 5-102.)
6. Remove two nuts securing the Door Switch to the instrument chassis. (Refer to
Figure 5-101.)
7. Disconnect the Door Switch connector at JP11 or JP12 on the ModSig PCBA,
depending on which Door Switch is defective. (Refer to Figure 5-102.)
Door
Switch
LEDs
JP11 JP12
8. Slide the JP11 or JP12 connector through the circular cut-out hole in the
chassis.
Note: The connector must go through vertically to fit through the hole.
Note: The front mounting screws of the ModSig PCBA may have to be loosened in order to
facilitate sliding the JP11 or JP12 connectors through the hole in the chassis.
Keyboard Port
Barcode Port
Modem Port
2. Disconnect the Fax Modem cable from the fax modem. (Refer to Figure 5-106.)
Note: The international modem has two additional DIP switch settings (9 and 10). Both
additional DIP switch settings should be set to ON.
3. Disconnect the UPS cable connector from the UPS port on the rear panel of the
instrument. (Refer to Figure 5-108.)
UPS Port
4. Disconnect the UPS cable connector from the Computer Interface port on the
rear of the UPS.
5. Set the UPS aside.
Note: Step 1 through Step 9 should only be performed if installing a new UPS. Step 1
through Step 9 should not be performed if installing the UPS after removing the UPS
to facilitate higher maintenance.
Screws
Battery
Compartment
Cover
Note: The red positive lead should already be connected to the battery’s positive terminal.
Black Negative
Lead
UPS Port
Screws
Battery
Compartment
Cover
Red
Positive
Lead
Black
Negative
Lead
9. Disconnect the red positive lead from the battery’s positive terminal.
10. Remove the battery from the UPS and set it aside.
The customer may order new reflective standards that are verified as being within
calibration. Another option is for the Field Engineer to replace the Reflective Standard
Discs. For replacement of the Reflective Standard Discs, use the part numbers listed
in Table 5-7.
CAUTION: Reflective Standard Discs that are not pressed flat or adhere
well to the Reflective Standard may result in errors in calibration testing.
Reflective Standard Discs that are scratched or have smudges may also
result in errors in calibration testing.
Note: The status (empty, occupied, disabled, etc.) for any cell can be determined by viewing
the View Cell Status screen.
a. From the Main screen, click the Incubation Chamber on the Instrument
icon. (Refer to Figure 5-114.)
Incubation
Chamber
Instrument
Icon
Example
Previous Selected
Screen Cell
Button
c. Ensure that the desired cell is empty and free of alarms by viewing the
View Cell Status screen.
Note: The cell pictorial in the View Cell Status Screen must indicate it is empty and free of
alarms by not showing a bold cell outline or red icons.
d. Once the location and cell status are verified, record all of the cell location
information in the ID column of the Cell Information Table.
Note: Cell location information includes the module, drawer, rack, and cell number. (Refer to
Figure 5-116 for an example.)
Note: A blank table is provided at the end of this procedure. (Refer to Table 5-12.)
Cell ID S/N
Module 1 708BS1807
Drawer A N/A
Rack A3 02273324
Cell 49 N/A
Next Screen
Button
Key Symbol
Button
Previous
Screen
Button View Incubation
Module Information
Button
Incubation
Module
Indicator
Module Serial
Number
Rack Serial
Number
Previous
Screen
Button
g. Locate the Module Serial Number on the Module and Rack Information
screen and record the information in the Cell Information Table.
Note: Module and Rack serial numbers must be entered as eight characters, including
leading zeros. Nine character module serial numbers must be entered as eight
characters by omitting the “I” in the serial number.
h. Locate the Rack Serial Number on the Module and Rack Information
screen and record the information in the Cell Information Table.
6. Press Esc and type yes.
7. Insert the BTA3D Field Utilities Zip® disk into the D drive.
8. At the DOS prompt, type D: and press Enter.
9. Type SLGCHK3D and press Enter.
Note: When the program begins, screens will appear in succession to prompt the entry of
the QC Reflectance Standards, Test Reflectance Standards, Module, and Rack serial
numbers of the cell.
10. Enter the QC Reflectance Standard serial number and press Enter.
11. Enter the Test Reflectance Standard serial number and press Enter.
12. Enter the Module serial number and press Enter.
13. Enter the Rack serial number and press Enter.
14. If a mistake is made entering Module or Rack serial numbers, click the Change
Cell button on the Main menu and repeat Step 12 and Step 13.
15. Verify that the Main menu of the QC Reflectance Std Test appears.
16. Use the Up or Down Arrow key to select Run QC Verify and press Enter.
Note: The screen may go blank momentarily before the CAL1 Main menu appears.
17. Use the Up or Down Arrow key to select Current Selection and press Enter.
18. Use the Up or Down Arrow key to select the module containing the selected
test cell and press Enter.
19. Verify the QC Reflectance Standard:
a. Use the Up or Down Arrow key to select CALIBRATE/VERIFY and
press Enter.
b. Use the Up, Down, Left or Right Arrow key to select VER adjacent to
the range containing the verification cell selected in the table and press
Enter.
c. Verify that the Select Single Cell menu appears.
d. Use the Up, Down, Left or Right Arrow key to select the verification cell
selected in the table and press Enter.
e. At the prompt, open the door and insert the 50% value QC Reflectance
Standard.
Note: The cell LED will begin flashing when the screen prompts you to insert the reference
standards.
e. At the prompt, open the door and insert the 50% value Test Reflectance
Standard.
Note: The cell LED will begin flashing when the screen prompts you to insert the reference
standards.
f. At the prompt, remove the 50% value Test Reflectance Standard and
insert the 35% value Test Reflectance Standard.
g. At the prompt, remove the 35% value Test Reflectance Standard and
insert the 20% value Test Reflectance Standard.
h. At the prompt, remove the 20% value Test Reflectance Standard and
insert the 5% value Test Reflectance Standard.
i. At the prompt, remove the 5% value Test Reflectance Standard and close
the door.
j. At the prompt, press any key to continue.
21. When the menu appears, press Esc twice to return to the Main menu.
22. Use the Up, Down, Left or Right Arrow key to select Print Report from the
Main menu and press Enter.
23. Review the screen information.
Note: It is not necessary for the Field Service Engineer to print or retain this report. If,
however, the customer desires, the report may be printed and retained by the
customer.
Note: After the troubleshooting task is complete, return to this procedure and begin at
Step 25.
27. Use the Up, Down, Left or Right Arrow key to select Quit to DOS from the
Main menu and press Enter.
28. At the DOS prompt, type C: and press Enter.
29. Type XBTA3D and press Enter.
30. Verify that the BacT/ALERT® 3D 60 software reboots.
31. Remove the BTA3D Field Utilities Zip® disk from the D drive.
32. If you get a pass test result, the Test Reflectance Standard is verified to be
acceptable for continued use in the field.
33. Place the calibration label near the center of the calibration slug on the Test
Reflectance Standard that received pass test results.
Note: Do not label Reflectance Standards that are unacceptable for certifying.
Note: The calibration label will validate slug calibration for two years from the date it was
tested. Test verification will be performed at approximately 12 months (every annual
PM).
34. If you are unable to get a pass test results after two tries, the Test Reflectance
Standards are unacceptable for certifying and should be replaced with new
Reflectance Standards (P/N 32005-2, article code 06543) that are verified as
being within calibration.
Note: The slug set must be retested and pass test results before applying new calibration
labels.
35. Record the procedure in the Service Log per local procedures.
36. Return the system to operation.
This section provides the procedures for verifying the proper functioning of the
BacT/ALERT® 3D 60 after a service call (repairs, alignments). They are not used after
PMs or installations, as they are covered separately.
• 3D 60 Module
• Field Utilities Zip® Disk
5.42.2 Procedures
1. In the incubation chamber, verify that the internal temperature has stabilized
and then verify that the displayed temperature on the 3D 60 display is within
± 0.5 °C of the calibrated reference thermometer. (Refer to Section 4.7.)
2. Verify that the Barcode Scanner will load a bottle by using the Load Bottles
function:
a. From the Main screen, click the Load Bottles button.
b. Verify the Load Mode screen appears.
c. Verify the Bottle ID field appears white.
d. Scan the bottle.
e. Verify the Bottle ID number from the bottle appears in the Bottle ID field.
f. Click the Check button to return to the Main menu.
Note: If no operator or bottle loading activity has been recorded in a period of 2 minutes, the
instrument will terminate the Load Bottle operation. Operator activity includes pressing
keys on the keyboard, scanning barcodes, clicking the mouse, and loading or
unloading bottles.
3. Disconnect the UPS from the AC wall outlet and verify the audible alarm,
emergence of the red triangle icon and the UPS alarm activate. Immediately
connect the UPS to the AC wall outlet.
Note: Try not to keep AC power off for too long. If the AC is allowed to remain off more than
1 minute, the 3D 60 instrument will begin an orderly shutdown. If this occurs, the
system will require rebooting.
Bottle Counts:
Description Time
Local Date/Time
CMOS Date/Time
Description Time
Description Time
Cell ID S/N
Module
Drawer N/A
Rack
Cell N/A
Chapter Contents
Refer to Section 6.9 and perform the Adjust the 3D 60 Power Supply procedure.
Adjust only if required. The measured voltage should be +5.2 ± 0.5 VDC.
1. From the Main screen, click the Next Screen button. (Refer to Figure 6-1.)
Next Screen
Button
Note: The password is company confidential. Do not provide the password to the customer.
Key Symbol
Button
Calibrate Module
Temperature
Button
Note: The Calibrate Module Temperature screen overlays and disables the Setup screen.
Time Remaining
Until Stable Actual Temp
Indicator Scroll Buttons
Setup Screen
(Disabled)
Previous
Screen Button
6. Enter the temperature from the thermometer in degrees Celsius (°C) on the
Actual Temp scroll buttons. (Refer to Figure 6-3.)
7. Click the Check button to initiate the calibration, or click the Cancel button to
retain the previous calibration.
8. Observe the icon to the left of the Actual Temp scroll buttons.
9. When the icon to the left of the Actual Temp scroll buttons indicates that
temperature is stable, check the temperature to verify the calibration was
successful.
10. Click the Previous Screen button to return to the Setup screen.
6.3 MB Conversion
By removing power from the agitation (Step-5) motor and anchoring the linkage
assembly that rocks the racks, the incubation chamber can be converted to a
stationary (MB) configuration.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 6-4.)
Top Panel
Screw Top Panel
Screw
Note: This removes +12 VDC power to the Step-5 PCBA and Motor.
Step-5 PCBA
J1
J4
Note: This removes communication between the ModSig PCBA and Step-5 PCBA and
+5 VDC Power. This will also remove false (agitation) Stepper Motor errors in the
software.
7. Clip J1 and J4 connectors to one wire of the five Step-5 Motor wires connected
to J3. (Refer to Figure 6-6.)
J1 Connector
J4 Connector
8. Remove the left side panel by lifting from the top edge of the panel.
9. Remove the MB conversion lock screw from Hole #1. (Refer to Figure 6-7.)
Agitation Arm in
the Lowest Position
Hole #2
Hole #1
10. Ensure the agitation arm is at its lowest position. (Refer to Figure 6-7.)
11. Install the MB conversion lock screw into Hole #2. (Refer to Figure 6-7.)
Note: It may be necessary to shift the racks slightly forward or backward to align the screw
with the screw hole in the linkage.
12. Insert paper cardstock over the wide foam at the top of the door.
Note: The paper cardstock will prevent damage to the foam when installing the left side
panel.
6.4 Optics
Bottle Racks are pre-calibrated at the factory, and while they are designed not to
require any re-calibration in the field, an initial calibration is required for validation
purposes.
This procedure is for calibration of the BacT/ALERT® 3D 60 while utilizing the Cal 10
software and calibration rods. This will allow the FSE to calibrate complete racks
instead of calibrating single cells. This procedure will only work on BacT/ALERT® 3D
systems.
IMPORTANT: Before performing this procedure, it should be completely read and fully
understood. Call bioMérieux immediately if you have any questions.
IMPORTANT: Cal 10 software or 10 rod calibration rods are not for customer use.
WARNING
Potential Biohazardous Material! Instrument surfaces and
contaminated bottles are potentially biohazardous and should be
handled according to good laboratory practices. Observe universal
precautions when operating the instrument and when performing
maintenance or troubleshooting.
WARNING
Use of the Cal 10 program stops the bottle reading function. If blood
culture bottle readings have been interrupted for 1 hour or longer, the
bottles must be subcultured. If MP/MB bottle readings have been
interrupted for 4 hours or longer, the bottles must be subcultured. The
B.25 software monitors for a gap in the readings. If an hour gap is
detected, the software will flag all bottles affected with a Fault Code
#80 Reading Gap Detected. Failure to attend to fault conditions
reported by the instrument within 1 hour may lead to invalid test
results and the need to subculture bottles.
WARNING
Bottles loaded into the system during the calibration procedure (or
any procedure requiring the drawer to be open too long) may be
falsely detected as positives due to temperature fluctuations during
and after calibrating the racks. Do not leave drawers open too long.
WARNING
If a Temperature Unstable message is displayed during calibration,
promptly halt the procedure. Start the 3D 60 software and allow
bottles to be monitored. If calibrating multiple incubators, shut the
drawers for the incubator with the temperature error and start the
calibration in a different incubator. Failure to comply may result in a
subculture event.
IMPORTANT: The customer should move any bottles occupying cells targeted for calibration
prior to the FSE performing this procedure. MP/MB bottles should never be
moved.
1. If bottles were not removed from the target cells, press Ctrl + F10 and unload
the bottles from the drawer.
Note: Anonymous bottles cannot be relocated to another cell until a bottle ID has been
entered for the bottle.
Note: The bottles do not have to be scanned when loaded using the special unload function
(Ctrl + F10).
IMPORTANT: If bottles were moved in the incubator, allow the temperature to stabilize for a
minimum of 1 hour before proceeding with this procedure.
12. Verify the Retrieving Flag Readings window appears. (Refer to Figure 6-10.)
The display will continue to report the row in which data is being retrieved until
all data is collected.
13. The Cell Status Report screen will appear when the Retrieving Flag Readings
window disappears. (Refer to Figure 6-11.)
Note: The cells highlighted with red indicate cells that are out of range and should be
calibrated. The cells that contain **** indicate a cell where a bottle is loaded.
18. Using the Up or Down Arrow key, select the drawer and rack to be calibrated
and press Enter.
19. The Calibrating Cells window will appear. (Refer to Figure 6-13.) The bottles
should already be removed from the rack to be calibrated.
WARNING
Any bottles removed at this point will need to be subcultured. Failure
to comply could result in false readings.
40. Examine the Results screen and note if all cells in the status row indicate pass
and not FAIL.
Note: Cells that fail calibration should be calibrated in the 3D 60 software calibration screen.
43. Repeat Step 18 through Step 42 until all desired racks have been calibrated.
44. Press Esc to exit the Reflectance Calibration screen.
45. Using the Up or Down Arrow key, highlight EXIT and press Enter to exit the
Calibration menu.
46. At the C:/> prompt, press Ctrl + Alt + Delete to restart the 3D 60 software.
47. After the system has restarted, verify the instrument is running properly by
observing that the time display is updating approximately every second.
48. Make a note of the time and ensure less than 1 hour has passed since
beginning this procedure.
Note: Proceed to Step 49 only if any cells failed calibration. If all cells passed calibration,
you have completed the procedure.
49. Refer to Section 6.7 and perform the 3D 60 Software Single Cell Calibration
procedure for each cell that failed calibration.
Rack Cells can be individually calibrated using the B.25 software feature found under
the Incubation Module Flag Check diagnostic. (Refer to Section 7.5.8.) This method
of calibration will enable cells that are currently testing to remain operational, while
concurrently calibrating rack cells that are out of calibration.
1. From the Main screen, click the Next Screen button to access the Setup
screen.
2. Verify the Setup screen appears. (Refer to Figure 6-2.)
Key Symbol
Button
Note: The password is company confidential. Do not provide the password to the customer.
4. Click the Key Symbol button to access the Diagnostic Test Selection screen.
(Refer to Figure 6-23.)
Main Selection
List
Sub-selection
Key Symbol
List
Button
Next
Screen
Button
1.7 Flag Check
7. Click the Next Screen button to start the Flag Check diagnostic. Allow enough
time for the diagnostic to gather flag readings. When the diagnostic completes,
the screen in Figure 6-24 appears.
Scroll Up/Down
Button
Print Current
Page Button
8. Observe the cells displayed in the Differences % > 1.5 section on the screen
(use the Scroll Up/Down buttons to find this section, if necessary).
9. Write down the cell numbers or (if printer is available) click the Print Current
Page button to print out the list of cell numbers.
10. Press C to access the Cell Calibration screen. (Refer to Figure 6-25.)
Cell
Selection
Buttons
Calibration
Staircase
Icon
Check Button
Previous
Screen
Button
11. Click the Cell Selection buttons to choose the appropriate cell (1–60).
The first screen displays an X at the base of the Calibration Staircase icon.
This prompts the user to verify the cell is empty and the door is closed. If
instrument Fault Code #20 Drawer Open For Too Long appears on the top of
the Calibration Staircase icon, the door is not closed. Close the door before
going to Step 12 below.
Note: The Cell Selection scroll buttons will be grayed out and inactive until calibration of the
cell is complete.
Note: The cell light of the cell selected for calibration is now illuminated and #1 appears
above the first step of the Calibration Staircase icon.
13. Insert Standard Number One into the selected cell. (A single ring around the
end of the reflectance standard identifies Standard #1.)
14. Click the Check button. #2 will appear above the second step of the
Calibration Staircase icon.
15. Insert Standard Number Two into the selected cell. (Two rings around the end
of the reflectance standard identifies Standard #2.)
16. Click the Check button. #3 will appear above the third step of the Calibration
Staircase icon.
17. Insert Standard Number Three into the selected cell. (Three rings around the
end of the reflectance standard identifies Standard #3.)
18. Click the Check button. #4 will appear above the fourth step of the Calibration
Staircase icon.
19. Insert Standard Number Four into the selected cell. (Four rings around the end
of the reflectance standard identifies Standard #4.)
20. Click the Check button.
21. If a check mark appears at the top of the Calibration Staircase icon, then the
calibration of the cell was successful.
22. Remove Standard Number Four and click the Check button to save the new
calibration values.
23. If Fault Code #60 Cell Failed QC appears at the top of the Calibration
Staircase icon, then the calibration of the cell was unsuccessful. Click the
Cancel button and perform Step 11 through Step 22 to recalibrate the cell.
Note: The calibration process can be canceled, and any new calibration values discarded,
by clicking the Cancel button at any time during the calibration process.
24. Click the Previous Screen button to return to the Flag Check Test screen.
IMPORTANT: Before performing this procedure, it should be completely read and fully
understood. Call bioMérieux immediately if you have any questions.
IMPORTANT: Cal 10 A software and/or calibration rods are not for customer use.
WARNING
Potential Biohazardous Material! Instrument surfaces and
contaminated bottles are potentially biohazardous and should be
handled according to good laboratory practices. Observe universal
precautions when operating the instrument and when performing
maintenance or troubleshooting.
WARNING
Use of the Cal 10 A program stops the bottle reading function. If blood
culture bottle readings have been interrupted for 1 hour or longer, the
bottles must be subcultured. If MP/MB bottle readings have been
interrupted for 4 hours or longer, the bottles must be subcultured. The
B.25 software monitors for a gap in the readings. If an hour gap is
detected, the software will flag all bottles affected with a Fault Code
#80 Reading Gap Detected. Failure to attend to fault conditions
reported by the instrument within 1 hour may lead to invalid test
results and the need to subculture bottles.
WARNING
Bottles loaded into the system during the calibration procedure (or
any procedure requiring the drawer to be open too long) may be
falsely detected as positives due to temperature fluctuations during
and after calibrating the racks. Do not leave drawers open too long.
WARNING
If a Temperature Unstable message is displayed during calibration,
promptly halt the procedure. Start the 3D 60 software and allow
bottles to be monitored. If calibrating multiple incubators, shut the
drawers for the incubator with the temperature error and start the
calibration in a different incubator. Failure to comply may result in a
subculture event.
IMPORTANT: The customer should move any bottles occupying cells targeted for calibration
prior to the FSE performing this procedure. MP/MB bottles should never be
moved.
1. If bottles were not removed from the target cells, press Ctrl + F10 and unload
the bottles from the drawer.
Note: Anonymous bottles cannot be relocated to another cell until a bottle ID has been
entered for the bottle.
Note: The bottles do not have to be scanned when loaded using the special unload function
(Ctrl + F10).
IMPORTANT: If bottles were moved in the incubator, allow the temperature to stabilize for a
minimum of 1 hour before proceeding with this procedure.
12. Verify the Retrieving Flag Readings window appears. (Refer to Figure 6-28.)
The display will continue to report the row in which data is being retrieved until
all data is collected.
13. The Cell Status Report screen will appear when the Retrieving Flag Readings
window disappears. (Refer to Figure 6-29.)
Note: The cells highlighted with red indicate cells that are out of range and should be
calibrated. The cells that contain **** indicate a cell where a bottle is loaded.
18. Using the Up or Down Arrow key, highlight [CAL] and press Enter.
19. Verify the Single Cell Calibration Selection screen appears. (Refer to
Figure 6-31.)
20. Using the Up or Down Arrow key, select the cell that requires calibration and
press Enter.
21. When the 50 Percent window appears, open the applicable drawer and insert
the 50 percent calibration rod into the appropriate rack. (Refer to Figure 6-32.)
39. Press the Space Bar to return to the Reflectance Calibration screen. (Refer to
Figure 6-30.)
40. Repeat Step 18 through Step 39 to calibrate other cells.
41. Using the Up or Down Arrow key, highlight EXIT and press Enter to exit the
Reflectance Calibration screen.
42. Using the Up or Down Arrow key, highlight EXIT and press Enter to exit the
Single Cell Mode Calibration menu.
43. At the C:/> prompt, press Ctrl + Alt + Delete to restart the 3D 60 software.
44. After the system has restarted, verify the instrument is running properly by
observing that the time display is updating approximately every second.
45. Make a note of the time and ensure less than 1 hour has passed since
beginning this procedure.
The DC Power Supply is mounted on the System Power Panel (located on the right-
hand side of the 3D 60's chassis). A metal housing encloses the DC Power Supply.
Since the adjustment POT is only accessible through a hole in the metal housing, the
instrument's top and right side panels must be removed to facilitate maintenance.
IMPORTANT: It is recommended that you perform a backup of the records stored on the
BacT/ALERT® 3D 60 before continuing this procedure. (Refer to Section 4.4.)
1. Press Ctrl + Alt + Delete to exit the software, but do not turn the instrument off.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 6-39.)
Top Panel
Screw Top Panel
Screw
DC Power
Supply
5. On the ModSig PCBA, connect both of the DC voltmeter's POS (RED) lead and
the NEG (BLACK) lead to respective VCC (Orange Wire) and Ground (White
Wire) on JP31. (Refer to Figure 6-41.)
Ground Wire
at JP31
+5 VDC Test
Point at JP31
POT Access
8. On the ModSig PCBA, disconnect both of the DC voltmeter's POS (RED) lead
and the NEG (BLACK) lead from the respective VCC (Orange Wire) and
Ground (White Wire) on JP31. (Refer to Figure 6-41.)
9. Although adjustments are not required, there should be +12 V at J4 on the
Step 5 PCBA. The measured voltage tolerance should be within ±1 VDC. (Refer
to Figure 6-43.)
Ground Wire
at J4
10. At the C:\ > prompt, type XBTA3D and press Enter to restart the software.
11. Insert paper cardstock over the wide foam at the top of the door.
Note: The paper cardstock will prevent damage to the foam when installing the right side
panel.
Top Panel
Screw Top Panel
Screw
1. Press Ctrl + Alt + Delete to exit the software, but do not turn the instrument off.
2. Open the door and remove two screws securing the top panel. (Refer to
Figure 6-45.)
Top Panel
Screw Top Panel
Screw
This chapter contains the most common malfunctions of the BacT/ALERT® 3D 60,
and their corresponding resolutions.
Chapter Contents
Diagnostics and troubleshooting are currently limited to Error Codes and the Cause/
Solution associated with the error. Of course, troubleshooting also depends on the
technician's system knowledge and understanding of the equipment.
Section 7.2 pertains to the symptomatic problems that may appear during normal
operations of the BacT/ALERT® 3D 60. These errors may or may not cause an error
code to appear. Error Code listings and associated troubleshooting procedures can
be found in Section 7.3.
WARNING
Potential Biohazardous Material! Instrument surfaces and
contaminated bottles are potentially biohazardous and should be
handled according to good laboratory practices. Observe universal
precautions when operating the instrument and when performing
maintenance or troubleshooting.
WARNING
The customer could have a subculture event present if the instrument
is shut down for more than an hour. The B.25 software monitors for a
gap in the readings. If an hour gap is detected, the software will flag
all bottles affected with a Fault Code #80 Reading Gap Detected.
Failure to follow the procedures in this chapter or failure to attend to
fault conditions reported by the instrument within 1 hour may lead to
invalid test results and the need to subculture bottles.
Anonymous Bottle The cell is calibrated incorrectly. 1. Calibrate the cell. (Refer to
indicated in an empty Section 6.7 or Section 6.8.)
cell
2. If an anonymous bottle is still
indicated in an empty cell,
proceed to the next malfunction.
The cell contains dirt or debris. 1. Inspect the cell for dirt or
debris.
System locks up after System error. 1. Press Esc, then type yes or
modem connection is press Ctrl + Shift + Alt + 2.
established with
Customer Service 2. At the C:\ > prompt, press Ctrl
+ Alt + Delete to reboot the
system.
Cell Indicator(s) The rack cable connections are 1. Remove the rack cables.
never illuminate loose or not routed correctly. (Refer to Section 5.30.1 and
perform Step 1 through Step 10.)
Note: Investigate all
fault codes reported 2. Install the rack cables. (Refer
for this module before to Section 5.30.2 and perform
making these checks. Step 6 through Step 13.)
The alarm does not The operator has disabled all 1. Navigate to the Set Audible
sound audible alarms. Alarms screen and adjust the
See Note 1 below. alarm options as required.
The cable from the ModSig PCBA 1. Tighten the cable connections
(JP15) to CPU PCBA (J3) is from the ModSig PCBA to the
loose. CPU PCBA.
The cable from the ModSig PCBA 1. Remove and replace the cable.
(JP15) to CPU PCBA (J3) is
defective. 2. If the alarm still does not
sound, proceed to the next
malfunction.
Barcode Scanner The Barcode Scanner needs to 1. Scan the barcode label in
does not beep be reprogrammed to the default Figure 7-1 or Figure 7-2. (The
settings. label scanned will depend on the
model of barcode scanner in use.)
Barcode Scanner The Barcode Scanner trigger is 1. Verify that scanner is pressed
does not auto scan not depressed. down in its cradle so the trigger
switch on the handle is
depressed.
Barcode Scanner The Barcode Scanner needs to 1. Scan the barcode label in
scans, but nothing is be reprogrammed to the default Figure 7-1 or Figure 7-2. (The
read settings. label scanned will depend on the
model of barcode scanner in use.)
Note: Investigate all
fault codes reported 2. If the Barcode Scanner scans,
for this module before but still nothing is read, proceed
making these checks. to the next malfunction.
On the UPS, the The power cables are not 1. Refer to Section 5.39.1 and
Power On LED will connected properly. perform Step 1 through Step 4.
not illuminate and/or
the UPS does not 2. Refer to Section 5.39.2 and
start perform Step 11 through Step 14.
The UPS does not The UPS battery is defective. 1. Replace the UPS battery.
provide expected (Refer to Section 5.40.)
backup time
2. If the UPS still does not provide
expected backup time, proceed to
the next malfunction.
On the UPS, the On The UPS internal temperature is 1. Refer to Section 4.1.1 and
Battery LED is too high. perform the Power Down
illuminated and a procedure.
loud tone is heard
continuously 2. Remove any blockage from the
vents of the UPS.
5. Wait 5 minutes.
Printer will not Printer is configured incorrectly. 1. Turn the printer power off.
function after (Lexmark E352dn only)
following the setup 2. Press and hold both the Back
procedure button and the Check button on
the printer while powering on the
printer.
Printer will not Incompatible printer is installed. 1. Ensure that the printer
function installed is an HP® 1320,
Lexmark E352dn, or Lexmark
E360dn printer.
QC Reflectance Cell used for verification is out of 1. Calibrate the cell used for
Standard fails calibration. verification. (Refer to Section 6.7
verification (alone or or Section 6.8.)
in conjunction with
other failures) on the 2. Refer to Section 5.41 and
first attempt perform Step 4 through Step 24.
Test Reflectance Cell used for verification is out of 1. Calibrate the cell used for
Standard fails calibration. verification. (Refer to Section 6.7
verification (alone or or Section 6.8.)
in conjunction with
other failures) on the 2. Refer to Section 5.41 and
first attempt perform Step 4 through Step 24.
Reflectance Standard CAL1 did not complete a 1. Refer to Section 5.41 and
verification Test verification pass. perform Step 4 through Step 24
Results indicate using an alternate cell.
Failed with a file not
found error on the 2. If the Reflectance Standard
first attempt verification Test Results still
indicate Failed with a file not
found error, proceed to the next
malfunction.
*NOTE1: Investigate any fault codes reported before making any checks. To test the speaker click the Load
Bottles button on the Main screen. The speaker should beep once.
Figure 7-1: Scanner Default Settings Barcode Label for Older Barcode Scanner
Figure 7-2: Scanner Default Settings Barcode Label for Newer Barcode Scanner
WARNING
Failure to follow the procedures in this chapter or failure to attend to
fault conditions reported by the instrument within 1 hour may lead to
invalid test results and the need to subculture bottles.
Fault Codes
Note: Locate specific bottles with fault by clicking the Incubator on the Instrument icon on
the Main screen to display the View Cell Status screen.
Instrument Status Codes appear in a separate window on the Main screen and are
identified by a 98 in the top left corner. A description of each Instrument Status Code
is provided. The following is an example of an Instrument Status Code #710
Incubation Module Initialization.
Status
Code
Fault/Status Number
Malfunction Corrective Action
Displayed
#1 Power Fault In Fuse is blown in Power Entry 1. Replace the PEM fuse. (Refer
Incubation Chamber Module. to Section 5.22.)
Fault/Status Number
Malfunction Corrective Action
Displayed
Power Supply cable is loose or 1. Verify the power cable from the
disconnected. Incubator Power Supply is
connected and secured to
ModSig PCBA.
Fault/Status Number
Malfunction Corrective Action
Displayed
Fault/Status Number
Malfunction Corrective Action
Displayed
Fault/Status Number
Malfunction Corrective Action
Displayed
Fault/Status Number
Malfunction Corrective Action
Displayed
3. Repeat Temperature
Calibration at desired set-point.
(Refer to the Section 6.2.)
Fault/Status Number
Malfunction Corrective Action
Displayed
#4 Incubation Module One or more drawers have been 1. Ensure all drawers are tightly
Temperature is too Low left open or ajar for an extended closed.
period of time.
Note: Bottle Readings 2. If more than ½ hour passes
are not taken while this and the Instrument icon still
error condition exists. displays the fault code, proceed
to the next malfunction.
3. Repeat Temperature
Calibration at desired set-point.
(Refer to the Section 6.2.)
Fault/Status Number
Malfunction Corrective Action
Displayed
Fault/Status Number
Malfunction Corrective Action
Displayed
#6 Agitation Stepper Cable from the Step-5 Motor to 1. Refer to Section 5.12.1 and
Motor Fault the Step-5 PCBA is disconnected, perform Step 1 through Step 4.
loose or defective.
2. Verify cable is securely
connected to the Step-5 PCBA.
#10 Controller Module A/C power cable is missing or 1. Secure both ends of power
Power Fault loose. cable.
Power Supply cable (MS-JP31) is 1. Verify the power cable from the
loose or disconnected between Power Supply is connected and
ModSig PCBA (JP31). secured to the ModSig PCBA.
Fault/Status Number
Malfunction Corrective Action
Displayed
#11 Insufficient Charge UPS cable is loose or 1. Verify the cable from the UPS
in UPS disconnected from the UPS port is connected and secured to the
on the rear panel. UPS port on the rear panel.
Fault/Status Number
Malfunction Corrective Action
Displayed
#19 Software Exception Automatic or manual backup 1. Press Esc, then type yes.
System performance may attempted with a defective Zip®
be compromised while disk. 2. At the C:\ > prompt, press Ctrl
this error condition exists. + Alt + Delete to reboot the
system.
Fault/Status Number
Malfunction Corrective Action
Displayed
#20 Drawer Open For Door has been open too long or 1. Close the door.
Too Long ajar for an extended period.
2. If the Instrument icon does not
clear itself in 30 minutes or the
door was not ajar or open too
long, proceed to the next
malfunction.
#21 Drawer Agitation Mechanical failure in the agitation 1. Replace faulty agitation linkage
Failure linkage. components.
Fault/Status Number
Malfunction Corrective Action
Displayed
Fault/Status Number
Malfunction Corrective Action
Displayed
#39 Rack Temperature Door has been open too long or 1. Close the door.
Fault ajar for an extended period.
2. If the fault code does not clear
itself in 30 minutes or the door
was not ajar or open too long,
proceed to the next malfunction.
Fault/Status Number
Malfunction Corrective Action
Displayed
#41–46 Bottle Rack(s) Cable connecting the Rack 1. Refer to Section 5.9.1 and
Not Responding Controller PCBA to ModSig PCBA perform Step 1 through Step 4.
is loose or defective.
2. Verify cable is securely
connected to the ModSig PCBA
and Rack PCBA.
Fault/Status Number
Malfunction Corrective Action
Displayed
Fault/Status Number
Malfunction Corrective Action
Displayed
#51–57 Bottle Rack One or more cells on bottle rack 1. Perform cell calibration on the
Hardware Failure are out of calibration. entire bottle rack. (Refer to
Section 6.6.)
#60 Cell Failed QC A cell failed the automatic internal 1. Calibrate the cell indicated.
diagnostic check and needs to be (Refer to Section 6.7 or
calibrated. Section 6.8.)
Bottle Rack retaining screws are 1. Remove the bottle rack. (Refer
not tightened completely. to Section 5.30.1.)
Fault/Status Number
Malfunction Corrective Action
Displayed
#62 Erroneous Cell Cell is out of calibration. Empty 1. Disable the affected cell for a
Load Status cells that are in the process of minimum of 10 minutes. Refer to
falling out of calibration can the BacT/ALERT® 3D 60 User
sometimes falsely report that they Manual.
are loaded before the auto QC
logic detects that the cell is out of 2. Enable the affected cell. Refer
calibration. In some cases the cell to the BacT/ALERT® 3D 60 User
transitions from the loaded state Manual.
back to the unloaded state when 3. Calibrate the affected cell.
in fact no bottle has been loaded (Refer to Section 6.7 or
or unloaded from that cell. Section 6.8.)
#71 Invalid Time Stamp BIOS (CMOS) clock and real time 1. Refer to Section 7.6.7.1 and
Bottle Readings are not clock are out of sync with each perform the Scandisk Output Test
taken while this error other, improper time change or Procedure.
condition exists. If not CPU hardware fault involving the
cleared within 1 hour, real time clock. 2. Determine how many bottle
then the affected bottles readings have been missed. If the
must be subcultured. time passed is greater than
1 hour, bottles must be
subcultured.
#72 Invalid Calculation Hardware failure involving the 1. Determine how long bottle
Data non-volatile random access readings have not been taken. If
Bottle Readings are not memory. the time passed is greater than
taken while this error 1 hour, bottles must be
condition exists. Bottles subcultured.
with fault are flagged
positive. 2. Unload affected bottles via the
Unload Positive button on the
Main screen.
Fault/Status Number
Malfunction Corrective Action
Displayed
3. Subculture bottles if
necessary.
#73 Invalid Method Data Information needed for the Each bottle type uses the same
Bottle Readings are not analysis of bottle readings is method data. Eventually all
taken while this error corrupt. bottles of the same type will show
condition exists. the fault.
Bottles with fault are
flagged positive 1 hour 1. Refer to Section 4.3 and
after the fault first perform the Compact Flash
appears. Adapter Data Backup Restore
procedure.
#74 Algorithm Change Incorrect media type has been 1. Unload affected bottles via the
previously assigned to a bottle or Unload Positive button on the
an anonymous bottle has been Main screen.
placed in an inappropriate drawer.
2. Subculture the bottle.
Fault/Status Number
Malfunction Corrective Action
Displayed
#75 Bottle Data Data integrity check has found 1. Unload affected bottles via the
Corrupted corrupt bottle data. Unload Positive button on the
Bottle Readings are not Main screen.
taken while this error
condition exists. Bottles 2. Subculture the bottles.
with fault are flagged 3. Reload bottles via the Load
positive. Bottles function on the Main
screen.
• Original load date and time will
be replaced with reloaded date
and time.
• Maximum test time is set to the
default associated with the
bottle type.
#77 Reading An incorrect bottle type has been 1. Unload affected bottles via the
Polynomial Change previously assigned to a bottle. Unload Positive button on the
Bottle Readings are not Main screen.
taken while this error
condition exists. Bottles 2. Subculture the bottles.
with fault are flagged 3. Reload bottles via the Load
positive. Bottles function on the Main
screen.
• Original load date and time will
be maintained.
Fault/Status Number
Malfunction Corrective Action
Displayed
#78 Incomplete Test This fault occurs when an 1. Unload affected bottles via the
Data identified bottle is reloaded Unload Positive button on the
Bottle Readings are not anonymously and later is Main screen.
taken while this error identified and reloaded in the
condition exists. Bottles instrument. The original bottle 2. Subculture the bottle.
with fault are flagged data cannot be merged with the 3. Reload bottles via the Load
positive. data collected when the bottle Bottles function on the Main
was anonymous and therefore the screen.
data is incomplete.
• Original load date and time will
be maintained.
#80 Reading Gap A gap in bottle readings greater There are two procedures that
Detected than 1 hour has occurred due to can be followed when this fault
power loss to instrument or occurs.
incubator or a negative-to-date
Solution 1: This procedure
bottle was unloaded, then
instructs you to unload and
reloaded more than 1 hour later.
subculture each bottle to clear the
80 fault.
Note: On the Load, Status and
Unload reports, bottles with Fault 1. Print an Unload report before
Code #80 will be marked on the clearing this fault to ensure that
report with a Gap Detection you have a record of all bottles
indicator (!) next to the result. If a with an 80 fault.
negative bottle has a Fault Code
#80 while it is unloaded, the Gap 2. From the Main screen, press
Detection indicator will remain on Ctrl + F3 to enter a special unload
the report. mode.
Fault/Status Number
Malfunction Corrective Action
Displayed
WARNING
These bottles must have a terminal subculture performed if their final status is
negative.
#81 Bottle Loaded in A bottle was loaded in a cell that 1. From the Main screen, press
Cell During QC has not completed the QC cycle. Ctrl + F10 on the keyboard and
verify that the Unload Screen
appears.
Fault/Status Number
Malfunction Corrective Action
Displayed
#99 Bottle Loaded in a Bottle currently loaded in either a 1. Determine how long bottle
Non-Operational Cell disabled cell or a cell that is not readings have not been taken.
Bottle Readings are not taking readings.
taken in cells with this 2. If time is less than 1 hour, it is
error condition. only necessary to relocate the
bottle.
#800 Printer Fault Printer is not connected to the 1. Connect the printer to the
3D 60 is unable to print to 3D 60. 3D 60.
the printer.
2. If the Instrument Status Code
window still displays the status
code, proceed to the next
malfunction.
Fault/Status Number
Malfunction Corrective Action
Displayed
#810 Backup Failed Zip® disk is missing from Zip® 1. Insert a Zip® disk in the drive.
3D 60 was unable to write drive.
backup data to the Zip® 2. Refer to Section 4.4 and
disk. perform the 3D 60 Module Data
Backup procedure.
#820 Zip® Disk Missing Zip® disk is not fully inserted into 1. Remove the Zip® disk and re-
or Zip® Drive Fault the drive. insert into the drive.
Fault/Status Number
Malfunction Corrective Action
Displayed
#901 Corrupted Bottle Corrupted bottle record has been 1. This error will automatically
Record Has Been selected for the Edit Bottle Detail disappear.
Selected screen.
2. Verify there is not a Fault
Code #75 Bottle Data
Corrupted. If there is, follow the
troubleshooting steps for Fault
Code #75 Bottle Data
Corrupted.
#902 Required Data Required data not entered while 1. Remove the bottle from the cell
Missing loading bottles. that was just loaded.
#909 Invalid Cell Load – A bottle was loaded in a cell that 1. Remove the bottle from the cell
Cell QC in Process had not completed the QC cycle. that was just loaded.
Fault/Status Number
Malfunction Corrective Action
Displayed
#910 Invalid Cell Load Indicates a bottle was loaded in a 1. Remove the bottle from the cell
cell belonging to the wrong that was just loaded. The cell ID
drawer type or in a cell that is will be displayed in the error
disabled or non-functional. This window on the 3D 60. The cell will
status code occurs if a bottle is also have a rapidly flashing cell
loaded while an Unload function indicator light. If in the Load
is active. function, load the bottle in an
indicated cell. Removing the
indicated bottle also causes the
Operation error screen to vanish.
#911 Invalid Cell Unload Indicates a bottle was unloaded 1. Replace the bottle from the cell
from a non-indicated cell or that a that was just unloaded. The cell
bottle was unloaded while not in ID will be displayed in the error
an Unload function. window on the 3D 60. The cell will
also have a rapidly flashing cell
indicator light. If in the Load
function, load the bottle in an
indicated cell.
Fault/Status Number
Malfunction Corrective Action
Displayed
#912 Invalid Cell Reload When reloading a bottle, the 1. Remove the bottle from the
bottle was placed in a disabled or non-functional cell whose
non-functional cell. indicator light will be flashing
rapidly. The cell ID will be
displayed in the error window on
the 3D 60. Reload the bottle into
a different cell. Moving the
indicated bottle also causes the
Operation error screen to vanish.
#913 Bottle ID Just Bottle ID is incorrect for the bottle 1. Rescan the bottle and click the
Entered Matches That being loaded or unloaded. Check button after it turns red.
of a Currently Loaded
Bottle 2. If the Instrument Status Code
Bottle ID just entered is window still displays the status
displayed at the top of the code, proceed to the next
Operator Error screen. malfunction.
The cell location of a If the bottle ID is correct for the 1. Record the cell location
bottle loaded with the bottle being loaded or unloaded, displayed in the Error box.
same ID is displayed the bottle ID for the bottle in the
below the bottle ID. cell location indicated was 2. If the error occurred while
incorrect. loading, then do not load the
bottle at this time. Set the bottle
aside and go to Step 4.
Fault/Status Number
Malfunction Corrective Action
Displayed
#921 Barcode Scan Not Barcode was scanned but current No action required to correct this
Allowed in this Mode screen does not accept barcodes. error. It automatically resolves
Frequently occurs if appropriate after 2 seconds.
function is not initiated, such as
load or unload bottles.
#923 Invalid Barcode Scanned barcode cannot be 1. Acknowledge error and rescan
Entry accepted because it is the wrong the bottle.
length, has the wrong starting
character, or has an invalid 2. If the error persists after
character. rescanning the barcode, attempt
1) Unacceptable bottle ID to enter the accession number or
characters are ',",\,*. bottle ID via the keyboard.
2) Unacceptable Accession 3. Ensure barcode is entered into
Number characters are the correct field
',",\,*,@.
4. If error persists, use a generic
bottle ID barcode label.
Fault/Status Number
Malfunction Corrective Action
Displayed
#930 Incorrect or A previously identified bottle is 1. If an error was made during the
Unexpected Barcode unloaded, but the wrong label scanning of the bottle, rescan the
Label Scanned While was scanned while unloading the bottle and click the Check button.
Unloading a Bottle bottle. The instrument determines
that it is not the same barcode
currently associated with the 2. If the Instrument Status Code
bottle and reports this error. window still displays the status
code, proceed to the next
Note: Two bottle ID's are malfunction.
displayed in a box on the
Operation Error screen. At the top
is the bottle ID of the bottle just
unloaded and scanned. The
second bottle ID displayed was
assigned to the bottle when it was
originally loaded or identified.
Fault/Status Number
Malfunction Corrective Action
Displayed
#931 Detection An anonymous or identified bottle 1. If the bottle has multiple labels,
Algorithm Has Been is unloaded and its barcode is it may be that the wrong label was
Changed scanned. The instrument scanned. Try re-scanning the
All affected bottles must determines that the media type bottle ID and if correctly entered,
be subcultured. associated with the barcode does click the Check button.
not match the algorithm used in
analyzing the initial data. 2. If no changes are made and
the Check button is clicked, the
bottle will be flagged positive and
Fault Code #74 Algorithm
Change or #77 Reading
Polynomial Change will appear
in the cell where the bottle was
loaded.
Fault/Status Number
Malfunction Corrective Action
Displayed
#932 Bottle ID Scanned An anonymous bottle is identified 1. If the bottle ID was entered
Matches that of a with a previously unloaded incorrectly, perform the following
Previously Unloaded bottle's ID. procedure:
Bottle
a. Enter the correct bottle ID.
Fault/Status Number
Malfunction Corrective Action
Displayed
#940 Bottle Detail Not Occurs only in the Select Bottle to 1. This Instrument Status Code
Found Edit/Graph screen when: will terminate spontaneously after
1. An empty cell is selected. being briefly displayed. When the
2. The bottle ID entered is not Select bottle to Edit / Graph
recognized as belonging to a screen is displayed again, correct
currently loaded or recently the previous entry or click the
unloaded bottle. Cancel button.
#941 Bottle ID Entered The bottle ID entered on the Edit If the bottle ID is correctly
Matches that of a Bottle Bottle Detail screen is the same entered, perform the following
that is Currently as the bottle ID of a bottle loaded procedure:
Loaded in a Different in a different cell. 1. Click the Cancel button on the
Cell Edit Bottle Detail screen.
Fault/Status Number
Malfunction Corrective Action
Displayed
#942 Bottle Type In the Edit Bottle Detail screen, a 1. If an error was made entering
Changed Requires bottle's media type has been the new media, perform the
Different Algorithm changed to a media type that following procedure:
requires a different data analysis.
a. Click the Cancel button and
return to the Edit Bottle Detail
screen to correct the entry.
Fault/Status Number
Malfunction Corrective Action
Displayed
#943 Identified Bottle The Bottle ID field has been 1. Click the Cancel button to
Changed to cleared for an identified bottle in return to the Edit Bottle Detail
Anonymous the Edit Bottle Detail screen. screen.
#944 Manual Unload of An attempt was made to manually 1. Click the Cancel button to
Anonymous Bottle unload an anonymous bottle from return to the Edit Bottle Detail
the Edit Bottle Detail screen. screen.
#945 Bottle ID Entered In the Edit Bottle Detail screen, a 1. If the bottle ID was entered
Matches That of a bottle has been given the bottle incorrectly, perform the following
Previously Unloaded ID of a previously unloaded bottle. procedure.
Bottle
a. Click the Cancel button on
the error screen.
Fault/Status Number
Malfunction Corrective Action
Displayed
#960 Duplicate Entries Two or more label entry fields on 1. Change one of the duplicate
the Label Entry screen contain labels to a unique entry and click
the same entry when the Check the Check button.
button was clicked.
2. If the Instrument Status Code
window still displays the status
code, contact bioMérieux.
#961 Invalid Blank Field One of the label fields on the 1. Enter a label in the blank field
Label Entry screen was blank and click the Check button.
when the Check button was
clicked. 2. If the Instrument Status Code
window still displays the status
code, contact bioMérieux.
Several diagnostics are provided to test the various communications ports on the
3D 60. The tests require special cables/connectors, as described below.
43013-1 TOOL RS485 This cable is included in the diagnostic kit, but is not
LOOPBACK 3D used on the 3D 60.
43014-1 TOOL CA ASSY RS485 This cable is included in the diagnostic kit, but is not
LOOPBACK 3D used on the 3D 60.
43205-7 CA ASSY BLKHD/MDL This cable is included in the diagnostic kit, but is not
CONT used on the 3D 60.
43005-19 KIT, CALIBRATION/ This kit includes the parts above (P/N 43013-1,
TOOL 3D and COMBO 43014-1, 43205-7 and 43012-1) along with parts
used for calibrating cells.
This section will detail the procedures to troubleshoot problems with various
components of the Incubator.
Note: This procedure will be used to access the Diagnostic Test Selection screen for the
Incubator, Controller Module, and LIS diagnostic procedures.
1. From the Main screen, click the Next Screen button. (Refer to Figure 7-5.)
Next Screen
Button
Note: The password is company confidential. Do not provide the password to the customer.
Key Symbol
Button
5. Verify the Diagnostic Test Selection screen appears. (Refer to Figure 7-7.)
Key
Symbol
Main Selection
Button
List
Sub-Selection
List
Next
Previous Screen
Screen Button
Button
There are also keyboard functions used in the procedure. Pressing the R — Resets
counts on lines 1.1–1.5, 2.1–2.4, 3.2, 4.1, 4.2, 4.2.1, 4.2.2 to 0 and pressing Run
status once started, the diagnostic runs continuously and updates the display data
continuously. (Refer to Section 3.7.3.1 for an example of the Loopback Method
screen content.)
Note: The Instrument Status Code #710 Incubation Module Initialization screen will
appear momentarily. This is normal.
There are also keyboard functions used in the procedure. Pressing the R — Resets
counts on lines 1.1–1.5, 2.1–2.4, and 3.2 to 0 and the Run Status runs continuously
and updates the display data continuously. (Refer to Section 3.7.3.2 for an example of
the Port Stats screen content.)
There are also keyboard functions used in the procedure. Pressing the R — Resets
counts on all lines to 0 and the Run Status runs continuously and updates the display
data continuously. (Refer to Section 3.7.3.3 for an example of the Message Stats
screen content.)
The Run Status runs continually and adds new messages to the end of the display as
they occur. (Refer to Section 3.7.3.4 for an example of the Message Log screen
content.)
6. Let the diagnostic run for 2 minutes and verify that the values on the Message
Log screen are all within the proper range. (Refer to Section 3.7.3.4.)
7. Click the Previous Screen button to exit the Message Log screen.
The Run Status runs continually and adds new temperature readings to the end of the
display as they occur.
The Drawer open (DRW OPEN) and Close (DRW CLOS) events are also displayed,
but are not shown in Figure 7-12 and events are logged when they occur.
MOD TEMP events are logged every 5 minutes. RCK TEMP events are logged
approximately once per hour per rack. TEMP SET events are logged when they
occur. (Refer to Section 3.7.3.5 for an example of the Temperature Log screen
content.)
Time of Event in
dd/mm/yy hh:mm:ss format
Event Location
Cells whose
difference is greater
than or equal to
1.0%
Cell location
(% diff.)
8. Perform a single cell calibration for each of the cells noted in Step 6. (Refer to
Section 6.7 and perform Step 11 through Step 24 for each cell.)
9. Click the Run Continuously button to restart the diagnostic and gather new
readings. (Refer to Figure 7-16.)
10. Click the Previous Screen button to exit the Flag Check screen.
7.5.9 Drawer/Heater
This diagnostic displays the state of the drawer switch and heater function in an
Incubator. Bottle testing is not affected by this diagnostic and the Run Status runs
continuously and updates the display continuously. (Refer to Section 3.7.3.8 for an
example of the Drawer/Heater screen content.)
5. Open the door and verify that all the LEDS on the cells are ON. (Refer to
Figure 7-18.)
Note: Upon initial entry to diagnostic, all LEDs are turned on and the display appears as
shown in Figure 7-18.
This section will focus on the diagnostic testing procedures that relate to the Controller
Module (Instrument Controller).
The Run Status runs continuously and updates the display continuously.
1. Remove the Zip® disk from the Zip® drive and replace it with a known good Zip®
disk.
2. Access the Diagnostic Test Selection screen. (Refer to Section 7.5.1.)
3. Select 2.0 Instrument Cntrl from the Main Selection List.
4. Select Zip Drive Test from the Sub-Selection List.
5. Click the Next Screen button.
6. Let run for at least 2 minutes.
This diagnostic verifies the operation of the serial hardware that makes up the
interface to the barcode scanner. It isolates problems to either the barcode scanner or
3D 60. Bottle testing is not affected by this diagnostic. (Refer to Section 3.7.4.2 for an
example of the Barcode Loopback screen content.)
The only keyboard function is: R — Resets values on lines 1.1–1.5, 2.1–2.4, 3.2, 4.1,
4.2, and 4.2.1–4.2.4 to 0.
The Run Status runs continuously and updates the display continuously.
1. Disconnect the Barcode Scanner cable connector from the BARCODE port on
the 3D 60 rear panel. (Refer to Figure 7-21.)
Barcode
Port
10. Click the Previous Screen button to exit the Barcode Loopback screen.
11. Disconnect the loopback connector P/N 43012-1 from the BARCODE port on
the 3D 60 rear panel.
12. Position the Barcode Scanner in place.
13. Connect the Barcode Scanner cable connector to the BARCODE port on the
3D 60 rear panel. (Refer to Figure 7-21.)
This diagnostic verifies the operation of the serial hardware that makes up the
interface to the external modem. It isolates problems to either the external modem or
3D 60. Bottle testing is not affected by this diagnostic.
The Run Status runs continuously and updates the display continuously. (Refer to
Section 3.7.4.3 for an example of the Modem Loopback screen content.)
1. Disconnect the modem cable connector from the rear panel of the 3D 60.
2. Connect the loopback connector P/N 43012-1 to the MODEM port on the rear
panel of the 3D 60.
3. Access the Diagnostic Test Selection screen. (Refer to Section 7.5.1.)
4. Select 2.0 Instrument Cntrl from the Main Selection List.
The Run Status runs continuously and updates the display continuously. (Refer to
Section 3.7.4.4 for an example of the Power Event Log screen content.)
Time of event in
dd/mm/yy hh:mm:ss
format.
5. Let the diagnostic run for 2 minutes and verify that the values on the Power
Event Log screen are all within the proper range. (Refer to Section 3.7.4.4.)
6. Click the Previous Screen button to exit the Power Event Log screen.
Each line in the event logs for .BID, .BTF, and .APR have the same format. The only
difference between them is that the file name extension will differ depending on the
type of event log accessed.
LOG files are similar to the event logs for .BID, .BTF, and .APR except that log files in
the \LOG directory, specifically EVENT14.LOG, EVENT13.LOG and so on to
EVENT0.LOG are displayed. EVENT0.LOG contains the most recent data and
appears at the bottom of the display. EVENT14.LOG contains the oldest data and
appears at the top of the display. If there is more log data than can be displayed, the
oldest log data is omitted from the display. An example is shown in Figure 7-24. Refer
to Section 3.7.4.9 for details on what the contents of .LOG file mean.
FILE : EVENT1.LOG
------------------------------------------------------------
!! 139719250(06/06/01 08:04:34) 0x38(DTChg) 4294948672
!! 139719253(06/06/01 08:04:37) 0x4(ICFlt) 0
!! 139719253(06/06/01 08:04:37) 0x3(IMFlt) 1 0 0 0 0 0
!! 139719253(06/06/01 08:04:37) 0x3(IMFlt) 2 0 0 0 0 0
!! 139719253(06/06/01 08:04:37) 0x3(IMFlt) 3 0 0 0 0 0
!! 139719253(06/06/01 08:04:37) 0x3(IMFlt) 4 0 0 0 0 0
!! 139719253(06/06/01 08:04:37) 0x3(IMFlt) 5 0 0 0 0 0
!! 139719253(06/06/01 08:04:37) 0x3(IMFlt) 6 0 0 0 0 0
!! 139719258(06/06/01 08:04:42) 0x30(IMPwr) 1
!! 139719258(06/06/01 08:04:42) 0x30(IMPwr) 2
!! 139719535(06/06/01 08:09:19) 0x4(ICFlt) 12
!! 139721277(06/06/01 08:38:21) 0x38(DTChg) 4294948672
!! 139721279(06/06/01 08:38:23) 0x4(ICFlt) 0
!! 139721280(06/06/01 08:38:24) 0x3(IMFlt) 1 0 0 0 0 0
FILE : EVENT0.LOG
------------------------------------------------------------
!! 139809948(07/06/01 09:16:12) 0x38(DTChg) 4294948672
!! 139809951(07/06/01 09:16:15) 0x4(ICFlt) 0
!! 139809951(07/06/01 09:16:15) 0x3(IMFlt) 1 0 0 0 0 0
!! 139809951(07/06/01 09:16:15) 0x3(IMFlt) 2 0 0 0 0 0
!! 139809951(07/06/01 09:16:15) 0x3(IMFlt) 3 0 0 0 0 0
!! 139809951(07/06/01 09:16:15) 0x3(IMFlt) 4 0 0 0 0 0
!! 139809951(07/06/01 09:16:15) 0x3(IMFlt) 5 0 0 0 0 0
!! 139809951(07/06/01 09:16:15) 0x3(IMFlt) 6 0 0 0 0 0
!! 139809956(07/06/01 09:16:20) 0x30(IMPwr) 1
!! 139809956(07/06/01 09:16:20) 0x30(IMPwr) 2
!! 139810161(07/06/01 09:19:45) 0x4(ICFlt) 12
!! 139810363(07/06/01 09:23:07) 0x3(IMFlt) 2 0 0 74 0 0
!! 139811149(07/06/01 09:36:13) 0x4(ICFlt) 0
!! 139812682(07/06/01 10:01:46) 0x38(DTChg) 4294948672
!! 139812694(07/06/01 10:01:58) 0x4(ICFlt) 0
!! 139812695(07/06/01 10:01:59) 0x3(IMFlt) 1 0 0 0 0 0
!! 139812695(07/06/01 10:01:59) 0x3(IMFlt) 2 0 0 0 0 0
!! 139812695(07/06/01 10:01:59) 0x3(IMFlt) 3 0 0 0 0 0
!! 139812695(07/06/01 10:01:59) 0x3(IMFlt) 4 0 0 0 0 0
!! 139812695(07/06/01 10:01:59) 0x3(IMFlt) 5 0 0 0 0 0
!! 139812695(07/06/01 10:01:59) 0x3(IMFlt) 6 0 0 0 0 0
!! 139812700(07/06/01 10:02:04) 0x30(IMPwr) 2
!! 139812700(07/06/01 10:02:04) 0x30(IMPwr) 1
!! 139813009(07/06/01 10:07:13) 0x3(IMFlt) 1 0 20 0 0 0
!! 139813053(07/06/01 10:07:57) 0x3(IMFlt) 1 0 0 0 0 0
!! 139813314(07/06/01 10:12:18) 0x3(IMFlt) 1 0 99 0 0 0
!! 139813764(07/06/01 10:19:48) 0x3(IMFlt) 1 0 51 0 0 0
For each file in the c:\btlprocs directory of the flash drive with a .BID extension, the file
name is displayed, followed by a row of dashes, followed by the contents of the file.
(Refer to Section 3.7.4.6 for an example of the .BID Files screen content.)
Procedure:
For each file in the c:\btlprocs directory of the flash drive with a .BTF extension, the file
name is displayed, followed by a row of dashes, followed by the contents of the file.
(Refer to Section 3.7.4.7 for an example of the .BTF Files screen content.)
Procedure:
For each file in the c:\btlprocs directory of the flash drive with an .APF extension, the
file name is displayed, followed by a row of dashes, followed by the contents of the
file. (Refer to Section 3.7.4.8 for an example of the .APF Files screen content.)
Procedure:
For each file in the c:\btlprocs directory of the flash drive with an .LOG extension, the
file name is displayed, followed by a row of dashes, followed by the contents of the
file. (Refer to Section 3.7.4.9 for an example of the .LOG Files screen content.)
Procedure:
For each file in the c:\btlprocs directory of the flash drive with an PLDSTORE.DAT
extension, the file name is displayed, followed by a row of dashes, followed by the
contents of the file.
Procedure:
7.6.6 Software
The software function displays data used to debug, verify and validate the instrument
software. It also displays logs of exception conditions detected by the software. The
Run Status runs continually and updates display when new information needs to be
displayed. Bottle testing is not affected by this diagnostic. (Refer to Section 3.7.4.10
for an example of the Software screen content.)
7.6.6.1 Debug
The Debug screen displays software debug messages intended for development
activity and is intended for the 3D 60 software development team. The format and
content messages put in this log are determined by the software implementor and no
design specifications are imposed. (Refer to Section 3.7.4.11 for information on the
Debug screen content.)
Procedure:
The Software Exception Log screen displays software exceptions messages intended
for the 3D 60 software development team. Access this screen if a Fault Code #19
Software Exception is being displayed on the Main screen. The format and content
messages put in this log are determined by the software implementor and no design
specifications are imposed. (Refer to Section 3.7.4.12 for information on the Software
Exception Log screen content.)
Procedure:
The Operating System Exception Log screen displays exceptions reported by the 3D
60 operating system (SMX). The format and content messages put in this log are
determined by the operating system vendor (Micro Digital) and no design
specifications are imposed. (Refer to Section 3.7.4.13 for information on the
Operating System Exception Log screen content.)
Procedure:
The Modem Activity Log screen displays software debug messages intended for
development activity and is intended for the 3D 60 software development team. The
format and content messages put in this log are determined by the software
implementor and no design specifications are imposed. (Refer to Section 3.7.4.14 for
information on the Modem Acitivy Log screen content.)
Procedure:
The Association Tables screen displays current content of various dictionaries and
association tables used to keep track of bottle and pre-load data. (Refer to
Section 3.7.4.15 for an example of the Association Tables screen content.)
Procedure:
The Bottle Processing Tables screen displays the current content of the tables that
drive the way the 3D processes bottles. (Refer to Section 3.7.4.16 for an example of
the Bottle Processing Tables screen content.)
Procedure:
The Audit Trail screen displays the audit trail settings that are used to determine the
BacT/ALERT® 3D 60's operation. (Refer to Section 3.7.4.17 for information on the
Audit Trail screen content.)
Procedure:
When replacing a CPU board, it is crucial to transfer the CMOS date/time from the old
board to the new board. This is not possible if the old board is not functioning.
Consequently, when the instrument is powered up, it is possible that Fault Code #71
Invalid Time Stamp is reported. This diagnostic screen provides the best chance of
fixing the CMOS time.
1. If Fault Code #71 Invalid Time Stamp appears, turn off the incubators.
2. Access the Diagnostic Test Selection screen. (Refer to Section 7.5.1.)
3. Select 2.0 Instrument Cntrl from the Main Selection List.
4. Select Software from the Sub-Selection List.
5. Click the Next Screen button.
6. Note the date and time on line 3.1.
7. Determine the amount of time that the instrument was down and add that to
date/time from Step 6.
8. Simultaneously, press Ctrl + Alt + Delete.
Note: The screen will go blank and then text will appear.
9. Immediately begin pressing and releasing the Delete key. Keep doing this until
the CMOS Setup screen appears.
10. Once the CMOS Setup screen appears, use the Arrow keys to select
STANDARD CMOS SETUP (will be highlighted red) and press Enter.
11. Enter the time from Step 7 into the CMOS clock.
12. Press Esc and then F10.
Note: The message Save CMOS and EXIT (Y/N)? will appear.
7.7.1 Loopback
Special equipment needed: P/N 43012-1
This diagnostic verifies that interrupts, transmit data, receive data, clear to send and
data set ready are working on the incubator's communication port. Bottle testing is not
affected by this diagnostic. (Refer to Section 3.7.5.1 for an example of the Loopback
screen content.)
Keyboard functions:
R — Resets counts on lines 1.1–1.5, 2.1–2.4, 3.0, 4.1, 4.2, 4.2.1, 4.2.2, 4.2.3, and
4.2.4 to 0.
Run status: Once started, the diagnostic runs continuously and updates the display
data continuously.
1. Disconnect the LIS cable from the 3D 60 end and replace with the loopback
connector P/N 43012-1.
2. Access the Diagnostic Test Selection screen. (Refer to Section 7.5.1.)
3. Select 3.0 LIS from the Main Selection List.
4. Select Loopback from the Sub-Selection List.
5. Click the Next Screen button.
6. Run the diagnostic for at least 2 minutes and verify that the values on the
Loopback screen are all within the proper range. (Refer to Section 3.7.5.1.)
7. Verify that FAIL count on line 4.2 is 0.
8. Click the Previous Screen button to exit the Loopback screen.
9. Disconnect the loopback connector P/N 43012-1 and connect the LIS cable to
the 3D 60.
Keyboard functions:
Run Status: Runs continuously and updates the display data continuously.
Appendix Contents
WARNING
Handle specimens and inoculated culture bottles as though capable
of transmitting infectious agents.
Appendix Contents
B.1 Spills
Any blood or test specimen spills within the unit need to be removed immediately.
Procedures outlining disinfection for spills are in the BacT/ALERT® 3D 60 User
Manual Safety Precautions and Procedures section.
Required Equipment:
• Gloves
• Cleaner
• Brush
• Paper Towels
• 5.25% bleach
• Mask
• Decontamination Certificate
• 700 PPM free chlorine solution (13 mL of 5.25% bleach to 1 liter of tap water)
Should the instrument require shipment to another location, perform the following
procedure:
1. Remove all tubing and trays and discard into an approved biohazard receptacle.
Remove crusted materials, blood, reagents, etc. using a 700 PPM free chlorine
solution (13 mL of 5.25% bleach to 1 liter of ph 8.0 tap water) and brush.
2. Dried and caked blood should be removed with a mild detergent and/or diluted
bleach solution before disinfection.
3. Using a 10% bleach disinfectant of 700 PPM free chlorine solution (13 mL of
5.25% bleach added to 1 liter of ph 8.0 tap water), wipe down the exterior of the
instrument. Pay particular attention to areas that are frequently touched.
4. Apply the 10% bleach solution or other EPA registered tuberculocidal
disinfectant. Gently cover the spill area with a PAPER TOWEL.
5. Wet down all surfaces with which the spill may have come in contact.
6. Allow all surfaces adequate contact time with the bleach solution (15–30
minutes) before cleanup.
7. After decontamination, wipe with damp (water only) paper towel and thoroughly
dry.
8. All materials used in the clean up should be treated as biohazardous waste.
9. Place instrument in sealed plastic bag before packaging.
Note: Solutions of sodium hypochlorite (household bleach) >5% are the recommended
concentrations and have demonstrated to have broad-spectrum germicidal properties.
Fresh dilutions should be used to prevent the loss of germicidal actions that occur
during storage.
B.3 Assemblies
Using the same equipment as mentioned in Section B.2, perform the following
procedures on assemblies before shipping:
1. Using a 10% bleach disinfectant of 700 PPM free chlorine solution (13 mL of
5.25% bleach to 1 liter of ph 8.0 tap water), wipe down all external surfaces of
the assembly.
2. After the decontamination procedure has been performed on the assembly,
complete the decontamination certificate, noting any areas that were unable to
be decontaminated. Attach the certificate to the assembly.
3. Place the assembly in a sealed plastic bag before packaging.
Contents
o BacT/ALERT® 3D 60 Instrument
o BacT/ALERT® 3D 60 Version B.25 User Manual
o 2 ea. 5 amp fuse
o 1 ea. 6.3 amp fuse
o 2 ea. Zip® Disk
o Decontamination Procedure
o Decontamination Certificate
o Anonymous BTL Labeling Set – WARNING –
o Rack — Bottle Carrier BTA
o Mouse and Mouse Pad
o Barcode Reader and Manual
o Barcode Reader Stand
o Keyboard
o Power Cord
o Printer
o Parallel Printer cable
o 1/4" X 20 X 5" bolt
o 1/4" flat washer
o UPS
o UPS Serial Cable
o CRT Monitor
o Calibration Standards Kit
o Certificate of Accuracy
o Digital Thermometer
o International Install Checklist
o MS-DOS 6.22 Software License
o Quick Reference Card
o Disk, BTA3D Installation and Utilities, Version B.25
o Super Cable Clamp
o Installation Instructions, UPS
o Cell Plug/Bottle Brush Kit
Optional Equipment
o Speaker Set — For 110 V BacT 3D 60
or
o Speaker Set — For 220 V BacT 3D 60
o Fax Modem — Domestic (Optional) — 120 V
or
o Fax Modem — International (Optional)
o Serial Modem Cable (Optional)
o UPS to Printer Power AC Adaptor
o UPS to Printer AC Power Cable
o Flat Panel Monitor
Note: This page may be copied for field use.
Installation
Set AC Power (see page 2-6) ____ _____
Instrument Placement (see page 2-10) ____ _____
Restraint Installation, if required (see page 2-4) ____ _____
Instrument connections made (see page 2-10) ____ _____
Instrument powered up (see page 2-12) ____ _____
Verify BIOS Time and Date (see page 2-13) ____ _____
Set/Verify Optimal Temperature (see page 2-16) ____ ____ _____
Instrument Cell Calibration (see page 2-26) ____ _____
Configure Software (see page 2-14) ____ _____
Set Bottle Test Time (see page 2-14) ____ ____ _____
Enable Audio Alarms (see page 2-14) ____ _____
Select Instrument Installation Options (see page 2-16) ____ ____ ____ _____
Signature Select SelectLink
Configure the Screen (see page 2-17) ____ _____
Configure the Instrument (see page 2-19) ____ _____
Set User Preferences (see page 2-20) ____ _____
Configure the LIS — SelectLink Only (see page 2-23) ____ _____
Check for Disabled Components (see page 2-26) ____ _____
Check for Errors (see page 2-26) ____ _____
Log In — 21 CFR Part 11 (see page 2-28) ____ _____
Functional Testing
Modem Functional Test (see page 2-33) ____ _____
Barcode Reader Functional Test (see page 2-33) ____ _____
Verify Plastic Bottle Algorithms (see page 2-34) ____ _____
UPS Functional Test (see page 2-38) ____ _____
Install Printer (see page 2-39) ____ _____
Record Software Revision and Annotate Logbooks ____ _____
Sheet 1 of 2
The following table shows items that may be connected to the instrument as “shipped”
or “optionally installed” parts:
Comments:
__________________________________________________________________________________________
__________________________________________________________________________________________
__________________________________________________________________________________________
__________________________________________________________________________________________
__________________________________________________________________________________________
Sheet 2 of 2
1. Physical Specifications:
Ɣ Receiving dock and the internal delivery route must be able to accommodate the size and weight of instruments being delivered.
BacT/ALERT® 3D 60 Installation Checklist
3D-60 3D-60
(in) (cm) (lb) (kg)
Height: 24 60.9 Weight
Width: 23.1 59 With Bottles: 98 45
Depth: 19.5 50 Empty: 90 41
2. Electrical Requirements:
Ɣ Properly grounded dedicated A/C power branch circuit capable of providing typical:
20 Amps @ 120 Volts / 60 Hz North American
16 Amps @ 220 Volts / 50 Hz Europe
15 Amps @ 100 Volts / 50 - 60 Hz Japan
13 Amps @ 240 Volts / 50 Hz United Kingdom
Ɣ One branch circuit will provide power to the 3D 60, UPS, Printer, Speakers, Modem, (Total of 2 receptacles for the system are required).
Ɣ NO power strips or extension cords are to be used for power distribution to the BacT/ALERT® 3D 60 System.
Ɣ AC power outlets must be within 8 ft or 244 cm of the 3D 60 installed.
Ɣ Nonconformance to the above Electrical Requirements may result in voiding the equipment warranty or service contract resulting in a charge for service.
Ɣ To ensure proper operation, the supporting electrical circuit must be maintained as a dedicated power source (i.e. not serving large power consuming
devices including but not limited to, centrifuges, air conditioners, water baths, refrigerators, freezers, or any devices that radiate electrical noises)
Ɣ Note: It is recommended that each 3D-60 be connected to Emergency power branch circuits.
The BacT/ALERT® 3D 60 System cannot agitate, heat, or perform bottle readings during a power outage
BacT/ALERT® 3D 60 Standard Laboratory Operating Environment
452501-1EN1
BacT/ALERT® 3D 60 Service Manual
3. Seismic Restraint:
Ɣ Requirements for earthquake protection must be determined and provided by the customer.
4. Environmental Requirements:
452501-1EN1
Ɣ Temperature Range: 50º to 86ºF or 10º to 30ºC, with no wide temperature fluctuations
Ɣ Room Relative Humidity: 10% to 90%
5. Location of the BacT/ALERT® 3D System:
Installation Checklist
Ɣ A single Analog telephone line must be within 25 ft of the 3D 60 instrument for remote access modem capability to be utilized.
Ɣ Nonconformance to the above Modem Requirements may result in voiding the equipment warranty or service contract resulting in a charge for service.
Ɣ The BacT/ALERT® 3D 60 System should not be installed under or near a HVAC register or in direct sunlight or bright overhead lights.
Ɣ The Customer must certify that both the bench and the supporting floor structure are capable of supporting the sum weight of each 3D 60 installed.
Ɣ At least 4 inches or 10 cm of unobstructed clearance must be provided behind the rear of each CM and IM for proper airflow and cooling.
Ɣ To provide adequate service accessibility, a minimum unobstructed clearance of 12 inches or 30.5 cm must be provided above and on both sides of the
System.
Ɣ Customer established or local code restrictions of the size or type of transport vehicle that can deliver to receiving dock.
________________________________________________________________________________________________________________
7. Site Inspection
Ɣ A bioMérieux, Inc. Field Engineer may perform a site inspection to confirm the status of the above requirements prior to scheduling system installation.
Ɣ Routine site inspections by Field Engineers are not performed unless specifically requested or deemed necessary after consultation between Instrument
Service and Sales Management.
Customer Acknowledgement
To the best of my knowledge, I certify that all requirements above have been met or will be implemented prior to system installation.
bioMérieux, Inc., 100 Rodolphe St, Durham, NC 27712, 800-662-2666, Fax: 919-620-2291
C-5
BacT/ALERT® 3D 60 Installation Checklist
BacT/ALERT® 3D 60 Installation Checklist Installation Checklist
Appendix Contents
This section is for the connections for the Controller Module CPU PCBA (P/N 750-
0070-01 / -902). There are two models of CPU PCBA currently in use. Ensure that you
are using the correct subsection to reference connector locations.
J2 J3
D1
J3
J1
D2
C2
J2
C3 C4
J4
C1
+
J6
C6
R2
R3
R1
R4
C5
R5
J4 J7
C7
U6
U4
U7
J9
J8
C8
J11
J10
U10
J18
J5
Y2
U9
R16
U8
J18
J17
U15
U3
U5
U1
U2
Y1
J12
U14
U11
LITHIUM
KEEPER
II
LTC-3PN
J14
J15
J16
J5
J13
J13
U16 J20
T1
J19
J21
J19
U18
J27
U17
J23
J24
J25
J22
J26
J22
J27
J41
J28
J31
J23
J21
J30
U19
U20
LS1
U22
J33
T2
C9
D6 D5 D4 D3
U21
+ +
C11
C10
+
J37
J35
J34 Y3
U25 U26
U24
J38
U23
U28
U29
U27
J39
SIM1
J40
Connector Description
Location
Connector Description
Location
J13, J19, 9-Pin Connector from CPU BIOS Clock Battery (if
J23 installed)
J14 J7
J2
J12
J19
J26
J105
J103
Note: The old connector locations are also given in Table D-2 to aid in reconnection of the
cables to the new CPU PCBA, as some of the cables may still have the older
connections listed on them.
D.2 UPS
Connector Description
Location
Note: The DIP switches on the rear panel of the UPS are used to configure the UPS input
voltage/output voltage range. Output voltage default setting from the factory is 120V.
You should not need to change this setting.
Input Voltage
Output Voltage DIP Switch 1 DIP Switch 2
Range
240V 216V–254V ON ON
Step-5
PCBA
ModSig
PCBA
Connector Description
Location
Connector Description
Location
P2 Mouse Port
P3 Debug Port
P4 Printer Port
P5 UPS Port
P6 Monitor Port
P8 LIS Port
P9 Modem Port
600 Hz In In In In
800 Hz In In In Out
F2
F7
F3
F5
F6
F4 F1
Connector Description
Location
Connector Description
Location
J3
Connector Description
Location
J3
Connector Description
Location
The Power Supply Terminal Strip TB1 is located near the Power Entry Module and
Isolation Transformer Assembly.
TB1-4A
TB1-4B
TB1-3A
TB1-3B
TB1-2A TB1-2B
TB1-1A
TB1-1B
Connector Description
Location
1A K3-1
Connector Description
Location
K1-2 HBP1-3
K2-2 HBP1-6
K3-1 TB1-1A
K4-2 E-FAN
Connector Description
Location
Note: This appendix does not include all historical instrument improvements.
Appendix Contents
A design improvement is available for the BacT/ALERT® Instrument. The CPU PCBA
(P/N 750-0070-01, Model LBC-586 PLUS) is no longer manufactured. If the CPU
PCBA requires replacement, it should be replaced with the CPU PCBA (P/N 750-
0070-902, Model LBC-GX500). This design improvement only applies to the B.25 (or
newer) firmware version. Firmware version B.12 (or prior) is not approved or validated
for use with the new CPU PCBA.
The new CPU PCBA has a compact flash socket on the board. The new CPU PCBA
also incorporates new RAM memory.
The current Zip® Drive will be upgraded to allow the new CPU PCBA to communicate
with the Zip® Drive. The CPU PCBA (P/N 750-0070-01) should be replaced utilizing
the CPU Replacement Kit for 3D 60 Instrument (P/N 450600-901). (Refer to
Table E-1.)
1 ea 395060-1 Box, 12 X 10 X 8
IMPORTANT: Read and understand this procedure before performing it. Contact Technical
Services if you do not understand this procedure.
IMPORTANT: The new style is validated for B.25 (or newer) firmware version only. Do not
install with B.12 firmware version (or older).
IMPORTANT: Installation can be accomplished in under 1 hour. If you are replacing the CPU
PCBA on a functioning system, be aware that a subculture event can occur if
replacement is not completed in 1 hour.
IMPORTANT: Record the BIOS time before replacing the CPU PCBA. If BIOS time is not set
correctly, Fault Code #71 Invalid Time Stamp may result. This could result in a
subculture event for all incubators installed on the system.
IMPORTANT: Current UTC may be obtained by telephone from the National Technical
Information Services WWV by calling 1-303-499-7111 (US) or may be obtained
from the web site http://tycho.usno.navy.mil/cgi-bin/timer.pl.
Note: To minimize the amount of time the system is down, Step 1 through Step 3 are
completed while the system is operational.
Note: Table E-9 and Table E-10 will be used later in this procedure.
2. Record the bottle counts and software version from the Main screen in
Table E-8.
3. Perform a system backup. (Refer to Section 4.4.)
4. Press Esc and type yes to exit the software.
Note: The 1 hour rule begins when the software shuts down.
5. Insert the B.25 Installation and Utilities disk (P/N 514490-3) into the Zip® Drive.
6. At the C:\ > prompt, type D: and press Enter.
7. At the D:\ > prompt, type COPY IOMEGA\GUEST.INI C:\IOMEGA and press
Enter.
8. Select Y and press Enter to overwrite the existing files.
11. Select STANDARD CMOS SETUP and press the Enter key.
12. Verify the CMOS Setup Utility screen appears. (Refer to Figure E-2.)
13. Record both the local date and time and CMOS date and time in Table E-9 that
was printed out earlier.
14. Press the Esc key to exit the CMOS Setup Utility screen.
15. Press F10 + Y and press Enter to save and exit the BIOS Main screen.
18. Open the door and remove two screws securing the top panel. (Refer to
Figure E-3.)
Top Panel
Screw Top Panel
Screw
Cable Clamp
IDE Cable
22. Remove two screws securing the Quad Serial PCBA to the CPU PCBA. (Refer
to Figure E-5.)
Screws
Note: The replacement procedure is simplified if the Quad Serial PCBA cables are not
disconnected.
26. Install the new RAM memory (P/N 186201-1) into the slot on the back of the
new CPU PCBA (P/N 750-0070-02).
Note: The new CPU PCBA may already have the memory installed.
27. Eject the Compact Flash Disk from the Flash Adapter PCBA.
28. Install the Compact Flash Disk into the slot on the back of the new CPU PCBA
(P/N 750-0070-02). (Refer to Figure E-6.)
29. Disconnect the IDE and power cables from the Flash Adapter PCBA.
Note: Do not remove the existing jumpers from J31 on the CPU PCBA.
J2
J7 J12
J19
J26
J31
Note: The old connection locations are also given in Table E-2 to aid in reconnection of the
cables to the new CPU PCBA.
33. Carefully align the pins for the Quad Serial PCBA and plug into the CPU PCBA.
34. Install two screws to secure the Quad Serial PCBA to the CPU PCBA. (Refer to
Figure E-5.)
35. Apply power to the instrument and turn on the UPS.
36. While the system is rebooting, press the F1 key to enter the BIOS screen.
(Refer to Figure E-8.)
37. To determine the time spent on the CPU PCBA replacement and new CMOS
time:
a. Enter the Local Time from Table E-9 in the Time column next to the Local
Time at Start in Table E-10.
b. Enter the current Local Time in the Time column next to the Local Time at
Finish in Table E-10.
c. Subtract the Local Time at Finish from Local Time at Start and enter the
information in the Time column next to Time Spent Replacing CPU PCBA
in Table E-10.
d. Enter the CMOS Time from Table E-9 in the Time column next to the
CMOS Time at Start in Table E-11.
e. Enter the information from the Time column next to Time Spent Replacing
CPU PCBA in Table E-10 in the Time column next to Time Spent
Replacing CPU PCBA in Table E-11.
f. Subtract the Time Spent Replacing CPU PCBA from CMOS Time at Start
in Table E-11.
Note: If the CMOS date and time were not noted in Table E-9 due to a down system, set the
CMOS date and time to Coordinated Universal Time (UTC).
Note: The current UTC may be obtained from the National Technical Information Services
WWV radio station by calling 1-303-499-7111, or by accessing the Internet USNO
Master Clock at the following web address: http://tycho.usno.navy.mil/cgi-bin/timer.pl.
40. Highlight Z. Save values to EEPROM and Exit and press Enter to reboot the
system.
41. If the new Zip® drivers were successfully installed in Step 7, proceed to Step 43.
42. If the system stops with the error message Finding a drive letter for your
Iomega drive:
a. Press Ctrl + Alt + Delete to reboot the system.
Note: This should boot up the system to the Main screen. If the system stops again, repeat
Step 42a. If the system stops again, proceed to Step 42b to manually edit the existing
Zip® driver file. If the system does not stop, proceed to Step 43.
Note: Refer to Figure E-9. The red text indicates information added to the file in Step 42f
and Step 42g.
Note: This may not be applicable if the instrument was inoperable before beginning this
procedure.
51. At the Main screen, click the Next Screen button. (Refer to Figure E-10.)
Next Screen
Button
Key Symbol
Button
Set Maximum
Test Time
Button
Note: The password is company confidential. Do not provide the password to the customer.
The Set Date/Time screen (screen ID 2.1) overlays and disables the Setup
screen.
56. From the Set Date/Time screen, set the date/time to reflect the current date and
time at the site.
Note: Check with the customer to determine the desired format of the date and time.
Note: In some cases, Fault Code #71 Invalid Time Stamp may occur. Verify the CMOS
date and time were configured properly.
Top Panel
Screw Top Panel
Screw
2) None
3) None
Table E-5 lists the functions available from the Main screen of BIOS and lists the sub-
selections available. Functions that do not utilize sub-selection lists will be noted with
N/A in the BIOS Sub-Selection List field of the table to denote that the sub-selection
list is not applicable.
4. Verify the BIOS Time (Coordinated Universal Time) and Date are set correctly.
5. Refer to Table E-5 to determine which item from the main selection list needs to
be selected to access the desired sub-selection settings screen.
Note: Functions that do not utilize sub-selection lists will be noted with N/A in the BIOS
Sub-Selection List field of the table to denote that the sub-selection list is not
applicable.
6. Highlight the desired selection from the main selection list and press Enter.
7. Highlight the desired sub-selection from the sub-selection list and press Enter.
8. Refer to Table E-5 and verify the settings are correct.
A. Time N/A
B. Date N/A
B. I/O Configuration
G. PCI Configuration
The update is a program on Zip® disk that will automatically install the BIOS update
when the program is launched from the D drive. In addition to the new version of
BIOS, the update program will replace the WinSystems® factory settings in CPU
CMOS with the bioMérieux BIOS default settings.
Modification of the BacT/ALERT® 3D 60 BIOS will not impact current bottles under
test. The BIOS update will not affect the BIOS date or time.
BIOS Version
Number and
Build Date
4. If the Main Menu indicates Version Release 070911-01 Built: 10/29/2007, the
CPU PCBA has the latest version of BIOS loaded and no further action is
necessary.
5. If the Main Menu indicates Version: Release 070726-01 Built: 07/26/2007, the
new version of BIOS should be installed.
Note: If the Main Menu indicates Version Release 070911-01 Built: 10/29/2007, do not
continue with the procedure to update the BIOS version. If the Main Menu indicates
Version: Release 070726-01 Built: 07/26/2007, proceed to Step 6. If any version
other than 070911-01 or 070726-01 is shown, complete Step 6 through Step 8 and
contact Global System Support - BacT/ALERT® for assistance. Do not continue with
the procedure to update the BIOS version if any version other than 070911-01 or
070726-01 is installed on the system without guidance from Global System Support -
BacT/ALERT®.
6. Print a copy of Table E-9 and record the local date and time and the CMOS date
and time.
7. Highlight Q. Exit Without Save and press Enter.
8. Select Y when prompted to exit the BIOS screen.
IMPORTANT: Read and understand this procedure before performing it. Contact Technical
Services if you do not understand this procedure.
Note: The BIOS Update Disk is acceptable for use in verifying the functionality of the Zip®
Drive.
b. From the Main screen, click the Next Screen button. (Refer to
Figure E-17.)
Next Screen
Button
Note: The password is company confidential. Do not provide the password to the customer.
Key Symbol
Button
j. Allow 10 passes of the Zip® Drive Test to accumulate. The PASS count
on the Instrument Cntrl/Zip drive test screen will show PASS count: 10.
(Refer to Figure E-19.)
Note: After about 1 minute, the system should start an orderly shutdown.
IMPORTANT: Read and understand this procedure before performing it. Contact Technical
Services if you do not understand this procedure.
IMPORTANT: This update is for the new WinSystems® CPU PCBA (P/N 750-0070-902) only.
The WinSystems® CPU can quickly be identified by its on-board Flash Drive
mounted to the bottom side of the CPU board.
IMPORTANT: The BIOS version can be determined by the Version field on the BIOS Main
Menu. Version release 070726-01 is the original BIOS and should be updated.
1. Press Esc and type yes to exit the Controller Module software.
2. Press Ctrl + Alt + Delete to reboot the software.
3. While the system is rebooting, press F1 to enter the BIOS screen.
4. Highlight Q. Exit Without Save and press Enter.
5. Select Y when prompted to exit the BIOS screen.
6. Once the BacT\ALERT® 3D application has finished loading, press Esc and
type yes to exit the Controller Module software.
7. Insert the BacT\ALERT® 3D BIOS Version Update Zip® Disk into the Zip® Drive.
8. At the DOS prompt, type D: and press Enter.
9. At the D: prompt, type UPDATE and press Enter.
10. At the screen prompt, type Y to proceed with the BIOS update. (Refer to
Figure E-20.)
Note: The update is complete when the system returns the text shown in Figure E-21 and
returns to the D: prompt.
18. Print a copy of Table E-12 and use the checklist to verify that the
BacT/ALERT® 3D 60 operates correctly.
19. Refer to Section 5.42 and perform the Post Service Validation procedure.
20. Record the procedure in the Service Log per local procedures.
21. Return the system to operation.
A new CPU BIOS Clock battery kit (P/N 193810-901) is available for the
BacT/ALERT® 3D 60. The kit can be installed on both the old CPU PCBAs (P/N 750-
0070-01) and the newer CPU PCBAs (P/N 750-0070-02 / -03). The kit provides a new
battery with extended life.
This design improvement was implemented into instrument production in March 2009
beginning with the serial number 711BS1876. For replacement of the CPU BIOS
Clock battery for BacT/ALERT® 3D 60 instruments after this serial number, or
instruments that have already been upgraded to the new CPU BIOS Clock battery,
refer to Section 5.5, CPU BIOS Battery Replacement.
Note: The CPU BIOS batteries should be replaced 10 years after their manufacture date.
The CPU BIOS battery has the part number and date code printed on it with the date
format of MMYYYY.
E.3.1 CPU BIOS Clock Battery Installation Procedure for Older CPU
PCBAs
This procedure is only for systems that have the CPU PCBA with the part number
750-0070-01. Refer to Section E.3.2 to install the CPU BIOS Clock Battery if the CPU
PCBA in the system has the part number 750-0070-02 or higher.
IMPORTANT: Read and understand this procedure before performing it. Contact Technical
Services if you do not understand this procedure.
IMPORTANT: Installation can be accomplished in under 1 hour. If you are installing the CPU
BIOS Clock Battery on a functioning system, be aware that a subculture event
can occur if installation is not completed in 1 hour.
IMPORTANT: Record the BIOS time before installing the CPU BIOS Clock Battery. If BIOS time
is not set correctly, Fault Code #71 Invalid Time Stamp may result. This could
result in a subculture event for all incubators installed on the system.
IMPORTANT: Current UTC may be obtained by telephone from the National Technical
Information Services WWV by calling 1-303-499-7111 (US) or may be obtained
from the website http://tycho.usno.navy.mil/cgi-bin/timer.pl.
Note: To minimize the amount of time the system is down, Step 1 through Step 7 are
completed while the system is operational.
1. Verify the contents of the CPU BIOS Clock battery kit (P/N 193810-901) and
ensure all parts required are included. (Refer to Table E-6.)
2. Open the door and remove two screws securing the top panel. (Refer to
Figure E-22.)
Top Panel
Screw Top Panel
Screw
Note: Table E-9 and Table E-10 will be used later in this procedure.
6. Record the bottle counts and software version from the Main screen in
Table E-8.
7. Perform a system backup. (Refer to Section 4.4.)
8. Press Esc and type yes to exit the Controller Module software.
Note: The 1 hour rule begins when the Controller Module software shuts down.
12. Select STANDARD CMOS SETUP and press the Enter key.
13. Verify the CMOS Setup Utility screen appears. (Refer to Figure E-24.)
14. Record both the local date and time and CMOS date and time in Table E-9 that
was printed out earlier.
15. Press the Esc key to exit the CMOS Setup Utility screen.
16. Press F10 + Y and press Enter to save and exit the BIOS Main screen.
17. Press Esc and type yes to exit the Controller Module software.
Screws
Note: The replacement procedure is simplified if the Quad Serial PCBA cables are not
disconnected.
25. Remove release liner and apply dot to top of U22 (just below beeper). (Refer to
Figure E-27.)
Velcro®
Dot
26. Install 9-pin connector of battery assembly across all 3-jumper locations,
placing key between J23 and J19. (Refer to Figure E-28.)
9-Pin Connector
CPU BIOS
Clock Battery
36. Select STANDARD CMOS SETUP and press the Enter key.
37. Verify the CMOS Setup Utility screen appears. (Refer to Figure E-24.)
38. Verify that the CMOS date and time updated correctly and did not reset. The
date should be the same and the time should update to include the amount of
time the system was down.
39. If the date and time did not update correctly, determine the time spent on the
CPU BIOS Clock Battery installation and new CMOS time:
a. Enter the Local Time from Table E-9 in the Time column next to the Local
Time at Start in Table E-10.
b. Enter the current Local Time in the Time column next to the Local Time at
Finish in Table E-10.
c. Subtract the Local Time at Finish from Local Time at Start and enter the
information in the Time column next to Time Spent Replacing CPU PCBA
in Table E-10.
d. Enter the CMOS Time from Table E-9 in the Time column next to the
CMOS Time at Start in Table E-11.
e. Enter the information from the Time column next to Time Spent Replacing
CPU PCBA in Table E-10 in the Time column next to Time Spent
Replacing CPU PCBA in Table E-11.
f. Subtract the Time Spent Replacing CPU PCBA from CMOS Time at Start
in Table E-11.
Note: If the CMOS date and time were not noted in Table E-9 due to a down system, set the
CMOS date and time to Coordinated Universal Time (UTC).
Note: The current UTC may be obtained from the National Technical Information Services
WWV radio station by calling 1-303-499-7111, or by accessing the Internet USNO
Master Clock at the following web address: http://tycho.usno.navy.mil/cgi-bin/timer.pl.
40. To change the date or time settings, press Up or Down Arrow key on the
keyboard to highlight the value you want to change.
41. Enter the determined time and/or date from CMOS Date and Time in
Table E-11 and press Enter.
42. If no changes were made, press the Esc key, select Y, and press Enter to exit
the BIOS.
43. If changes were made, press F10, select Y, and press Enter to save to CMOS
and exit the BIOS.
44. Verify the system functions normally and that there are not any Fault Codes
present.
45. If Fault Code #71 Invalid Time Stamp or #80 Reading Gap Detected occurs,
verify that the CMOS date and time are correct.
46. Refer to Section 5.42 and perform the Post Service Validation procedure.
47. Record the procedure in the Service Log per local procedures.
48. Return the system to operation.
E.3.2 CPU BIOS Clock Battery Installation Procedure for Newer CPU
PCBAs
This procedure is only for systems that have the CPU PCBA with the part number
750-0070-02 or higher. Refer to Section E.3.1 to install the CPU BIOS Clock Battery if
the CPU PCBA in the system has the part number 750-0070-01.
IMPORTANT: Read and understand this procedure before performing it. Contact Technical
Services if you do not understand this procedure.
IMPORTANT: Installation can be accomplished in under 1 hour. If you are installing the CPU
BIOS Clock Battery on a functioning system, be aware that a subculture event
can occur if installation is not completed in 1 hour.
IMPORTANT: Record the BIOS time before installing the CPU BIOS Clock Battery. If BIOS time
is not set correctly, Fault Code #71 Invalid Time Stamp may result. This could
result in a subculture event for all incubators installed on the system.
IMPORTANT: Current UTC may be obtained by telephone from the National Technical
Information Services WWV by calling 1-303-499-7111 (US) or may be obtained
from the website http://tycho.usno.navy.mil/cgi-bin/timer.pl.
Note: To minimize the amount of time the system is down, Step 1 through Step 7 are
completed while the system is operational.
1. Verify the contents of the CPU BIOS Clock battery kit (P/N 193810-901) and
ensure all parts required are included. (Refer to Table E-6.)
2. Open the door and remove two screws securing the top panel. (Refer to
Figure E-32.)
Top Panel
Screw Top Panel
Screw
Note: Table E-9 and Table E-10 will be used later in this procedure.
6. Record the bottle counts and software version from the Main screen in
Table E-8.
7. Perform a system backup. (Refer to Section 4.4.)
8. Press Esc and type yes to exit the Controller Module software.
Note: The 1 hour rule begins when the Controller Module software shuts down.
11. Record both the local date and time and CMOS date and time in Table E-9 that
was printed out earlier.
12. Highlight Z. Save values to EEPROM and Exit and press Enter.
13. Press Esc and type yes to exit the Controller Module software.
14. Turn off the UPS and remove the power.
15. Remove two screws securing the Quad Serial PCBA to the CPU PCBA. (Refer
to Figure E-34.)
Screws
Note: The replacement procedure is simplified if the Quad Serial PCBA cables are not
disconnected.
17. Remove the Quad Serial PCBA and lay it to the side.
18. Remove the jumper at J17. (Refer to Figure E-35.)
CAUTION: Failure to remove the jumper at J17 will shorten the battery
life.
J14
J17
19. Install the 3-pin connector of the battery assembly into J14.
20. Remove loop side of Velcro® dot from CPU BIOS Clock Battery.
21. Remove release liner and apply dot to unpopulated area below U4, to the left of
the CPU heat sink. (Refer to Figure E-36.)
Velcro® Dot
23. Use the cable tie from the kit to secure the unused portion of the cable near the
battery. (Refer to Figure E-37.)
Cable
Tie
CPU BIOS
Clock Battery
32. Verify that the CMOS date and time updated correctly and did not reset. The
date should be the same and the time should update to include the amount of
time the system was down.
33. If the date and time did not update correctly, determine the time spent on the
CPU BIOS Clock Battery installation and new CMOS time:
a. Enter the Local Time from Table E-9 in the Time column next to the Local
Time at Start in Table E-10.
b. Enter the current Local Time in the Time column next to the Local Time at
Finish in Table E-10.
c. Subtract the Local Time at Finish from Local Time at Start and enter the
information in the Time column next to Time Spent Replacing CPU PCBA
in Table E-10.
d. Enter the CMOS Time from Table E-9 in the Time column next to the
CMOS Time at Start in Table E-11.
e. Enter the information from the Time column next to Time Spent Replacing
CPU PCBA in Table E-10 in the Time column next to Time Spent
Replacing CPU PCBA in Table E-11.
f. Subtract the Time Spent Replacing CPU PCBA from CMOS Time at Start
in Table E-11.
Note: If the CMOS date and time were not noted in Table E-9 due to a down system, set the
CMOS date and time to Coordinated Universal Time (UTC).
Note: The current UTC may be obtained from the National Technical Information Services
WWV radio station by calling 1-303-499-7111, or by accessing the Internet USNO
Master Clock at the following web address: http://tycho.usno.navy.mil/cgi-bin/timer.pl.
43. Refer to Section 5.42 and perform the Post Service Validation procedure.
44. Record the procedure in the Service Log per local procedures.
45. Return the system to operation.
This design improvement was implemented into instrument production in May 2009
beginning with the serial number 905BS2376. For replacement of the Barcode
Scanner for BacT/ALERT® 3D Instruments after this serial number, or instruments
that have already been upgraded to the new Barcode Scanner, refer to Section 5.37.
IMPORTANT: Read and understand this procedure before performing it. Several equipment
upgrades may have been made to this instrument. It is very important that you
know which components are on this system prior to beginning this procedure
to avoid delays. Contact Technical Services if you do not understand this
procedure.
IMPORTANT: Installation can be accomplished in under 1 hour. If you are replacing the
Barcode Scanner on a functioning system, be aware that a subculture event can
occur if replacement is not completed in 1 hour.
Barcode Port
Next Screen
Button
14. Using Table E-13, scan a barcode label in the Sample column and compare the
text displayed on the screen with the text in the Read Data column. (Refer to
Figure E-43.)
Displayed
Data
Cancel
Button
15. Repeat Step 14 until all labels have been scanned and verified.
Note: Labels can be scanned in succession without clicking any buttons on the screen.
There is no need to click the Cancel button until all labels have been entered.
IMPORTANT: The Barcode Scanner should beep for each label scanned, indicating the label
was read. Displayed data must match the Read Data column in Table E-13 for
every label read. If the Barcode Scanner does not work properly, refer to
Chapter 7, Diagnostics and Troubleshooting Information to troubleshoot the
Barcode Scanner. Troubleshooting procedures for the Barcode Scanner
include:
Bottle Counts:
Description Time
Local Date/Time
CMOS Date/Time
Description Time
Description Time
System Function
Printer Function
Modem Function
Keyboard Function
INTERLEAVED 2 OF 5 0123456789
CODABAR $99.95
UPC-E 998875
Appendix Contents
• You are required to wear a grounded wrist strap prior to, and while, handling
ESDS components or assemblies. The wrist strap shall make constant
contact with your bare wrist and not be placed over shirt sleeves or gloves.
• Keep static-sensitive assemblies inside antistatic packages when not in use.
• Prior to carrying an ESDS component or assembly, always place it completely
inside an antistatic bag. (Do not partially wrap the ESDS component or
assembly with an antistatic bag.)
• Discard any antistatic bag with holes or other visible damage.
• If you need to ship an ESDS component or part, package it in appropriate
antistatic packaging. (For example, antistatic tubes, carriers or conductive
bags.) If possible, mark the outside of the package as containing ESDS
components.
• Do not staple antistatic bags containing ESDS components. To seal the bag,
fold the open end of the bag and apply tape to the outside.
• Do not place paper or other foreign items inside the antistatic packages. If
documentation needs to be kept with the ESDS component, tape it on the
outside of the package.
• All tools should be static free, where possible.
• Probing of circuits containing ESDS components with ungrounded
instrumentation (VOM, DVM, etc.) or similar type equipment is prohibited,
except when battery-operated or plastic-cased instruments must be used, or
when required measurement must be obtained with a piece of test equipment
that is isolated from ground.
1. Remove all static generating material (polyethylene film, Styrofoam, paper, etc.)
from the work area.
2. If possible, ground the assembly into which the ESDS components are to be
installed.
3. Verify the proper functioning of wrist strap and cord as defined below.
• Using a multimeter, measure the resistance from the wrist strap to the end of
the cord. Discard and replace the wrist strap if the resistance is less than
800k ohms or greater than 1.2M ohms.
4. Put the grounding wrist strap on, and clip the cord end to a chassis ground point
of the instrument/module you will be repairing.
IMPORTANT: When working on ESDS equipment, you must discharge through a grounded
wrist strap. Do not allow your skin to come into contact with the instrument
chassis until after you connect the wrist strap alligator clip to a known chassis
ground point.
A-D CONVERTER A device that transforms analog electrical signals to a digital format by
using a logarithmic function.
ACCESSION NUMBER A number that uniquely identifies the sample or patient specimen.
AGITATION The rocking motion of the racks within the Incubation Chamber. Required
in order to produce the growth of Microbes.
AGITATION MOTOR The device within the Incubation Chamber that produces agitation
(Step-5).
ANONYMOUS BOTTLE A bottle that has been placed in an Incubation Chamber without going
through the formal bottle loading process. Bottles loaded anonymously
must eventually be identified and assigned a bottle ID. Anonymous bottles
cannot be moved to another cell without being identified and assigned a
bottle ID.
AUDIBLE ALARM The field technician sets the coarse adjustment of volume and frequency
with the use of jumpers. Fine setting of volume specified by the user to flag
positive bottles, instrument failure, or an operator error.
BacT/LINK® The version of BacT/ALERT® 3D 60 software that includes the LIS option.
BARCODE READER Used to scan accession numbers or bottle barcode labels and identifies
bottles when loading or unloading.
BLOWER Device within the Electronics Drawer that circulates warm air in the
Incubation Chamber for the incubation process.
BOTTLE GRAPH A line graph that plots the microbial activity in a blood culture bottle over
time.
BOTTLE TYPE BacT/ALERT® bottle with a specific type of culture medium (i.e.: FAN®
Aerobic).
BUTTON BAR The user can freely move to any screen at any time by selecting (clicking)
the appropriate button from the vertical column of buttons that is always
visible on the right-hand side of the display.
CAUTION A step or action that could easily result in damage to the instrument, or an
incorrect result.
CALIBRATION Test rod used to calibrate cells. There are four separate standards used to
STANDARD calibrate the cells. Each standard is identified by the number of rings
around the end of the standard.
CARET The text entry position indicator that appears in the form of a vertical line
indicating where new text entry will appear. The caret always appears in
the current prompt.
CELL Holds and monitors each culture bottle. Each cell is numbered from one
through 60.
CELL FLAG Secures bottles in cells. Aids in cell diagnostics and with bottle loading and
unloading determinations.
CELL INDICATOR LIGHT Illuminates to show the location of positive, negative, or anonymous
bottles as well as to show where new bottles should be placed/removed.
An indicator light is adjacent to each cell.
CHAMBER The Incubation Chamber contains three racks (60 cells in total) for a total
capacity of 60 culture bottles.
CHECK BUTTON Icon on operator's screen used to accept an input or selection by the
operator.
CLICK Pointing to an item, and then quickly pressing and releasing the mouse
button.
CURSOR A marker for the active site where input would be directed.
D-A CONVERTER A device that transforms digital signals to an analog electrical format by
using a logarithmic function.
DATA SIGNAL Contained within the ModSig PCBA and routes all signals within the
DISTRIBUTION Instrument.
DC POWER SUPPLY Device that produces the required DC voltages used within the Instrument.
DISABLED Describes a user interface item (i.e.: menu item, button, text entry field)
that is not currently selectable by the user. Disabled user interface items
are usually displayed in gray.
EXECUTABLE A software program that can be started from the operating system's
command prompt.
EXTERNAL SPEAKERS External devices typically used to produce alarm sounds remotely and
connected via an external port on the rear of the Instrument.
FIRMWARE In computer terms, firmware is the internal software code that a system
uses to operate.
GENERIC BOTTLE ID A bottle ID that does not indicate one of the pre-defined bottle types.
GRAPHICAL USER A method of user/computer interfacing where the user communicates with
INTERFACE (GUI) the computer by manipulating graphical screen images.
GRAYED See DIMMED. Refers to user interface items that are not selectable.
HEATER Device within the Incubation Module that produces the required heat for
the Incubation process.
INCUBATION CHAMBER The chamber within the Instrument that holds the racks and incubates
BacT/ALERT® 3D 60 bottles. It may also be referred to as the Incubation
Module.
INDICATOR LIGHT OR An identifier for an indicated cell that lights up to indicate location of a
LAMP particular bottle or bottle type. They are located in the racks adjacent to
the cells.
KEYBOARD Provides for an alternate means of input for bottle IDs, accession
numbers, hospital IDs, and names. Serves as a backup input device
should the mouse or Barcode Reader fail.
LCD MONITOR A Liquid Crystal Display device that is used to display video images of
software and firmware functions. Typically called a Flat Panel Display.
METAL OXIDE VARISTOR An electronic device that provides inline surge protection for internal
(MOV) circuitry of the Instrument.
MOMENTARY BUTTON An Instrument display screen button which remains activated (appears to
be pressed) only as long as the corresponding screen area is being
clicked.
MYCO CONVERSION Rack configuration designed to grow and detect Mycobacteria. This
conversion disables the agitation process, which is not required for the
growth of Mycobacteria.
NOTE Any step or actions that may need further clarification or explanation.
NVRAM (Non Volatile Random Access Memory that is capable of retaining data when power is
RAM) removed from the system.
OPTION A member of the set of possible valid entries for the current field.
POWER CONNECTOR A fitting to attach the alternating current (AC) power cord.
POWER ENTRY MODULE Device on the Instrument that contains the fuse(s) and facility voltage
(PEM) setting selection. It also contains the power connector.
POWER PANEL Panel within the Instrument that contains the DC Power Supply, relays,
and isolation transformer.
PRINTER PORT Printer interface port (parallel) on the Instrument for producing hard copy
reports concerning bottle status.
PRINTED CIRCUIT BOARD A fiberboard containing the necessary electronic components that are
(PCBA) interconnected via soldered paths to create a working circuit assembly.
PROMPT A text entry field that appears on a display screen. The primary purpose
for a prompt is for data entry. Read-only prompts are not selectable and
are dimmed. User entries into a prompt are checked for validity according
to the value range internally assigned for that prompt.
QUAD SERIAL PCBA PCBA within the Instrument that provides for the communication between
the Incubation Chamber and the rest of the equipment.
QUICK REFERENCE Outlines instructions for system operation and describes specific fault
CARDS codes. Found in the Operator’s Manual and may be conveniently
relocated by the customer.
RACK Contains 20 cells, each of which can hold and monitor an individual culture
bottle. Racks gently rock to agitate bottles, unless they are set up for MB.
RACK LEVELING Bracket on the rear of the rack assembly providing for a means of the
BRACKET agitation arm to make contact with the rack.
RAM An acronym for Random Access Memory. RAM is volatile memory (unless
it is the special NVRAM).
SCROLL BUTTON A button that allows selection of numerical or symbolic values by clicking
arrows above or below a window that shows the selected character. While
an arrow is being clicked, the center portion shows one character from
among the available list of choices.
SLIDEBAR SWITCH A rectangular icon device which moves between ON and OFF (symbolized
by 1 and 0, respectively) when the corresponding end of the rectangle is
clicked.
TITLE BAR The horizontal bar at the top of a window that contains the title of the
window, screen, or dialog box.
TOGGLE BUTTON A button that becomes active (appears to be clicked) after the
corresponding screen area is clicked, and continues to be active until it is
clicked again.
UP/DOWN SCROLL Icon that is used to select a numerical value of 0 to 9. More than one
BUTTON button is available for larger numbers.
USER INTERFACE The methods by which the computer and user can communicate.
WARNING Identifies actions that can result in injury or death to personnel performing
a procedure (or persons that may be nearby).
WAVEFORM From the lab technician’s perspective, this is another name for a bottle
graph. For the Field Engineer, it could mean either the same as above, or
more commonly the picture presentation on an oscilloscope display.
Zip® DRIVE Disk drive within the Instrument providing for a means of loading software
and backing up data.
Mouse 3-93 Q
Moving Parts 1-9
Quad Serial PCBA 3-7
connector locations D-9
N replacement 5-27
Navigation Buttons Quick Reference Card 3-94
display output area 3-37 description 3-4
O R
Operation Buttons 3-34 Rack
Operator Panel incubator 3-20
description 3-3 Rack Controller PCBA 3-21
Operator Panel Screens 3-30 Reading Log
Optics 6-8 incubator 3-46, 7-58
Organization of the Manual 1-2 Reflective Standard Discs
replacement 5-116
Outlet Thermistor PCBA
connector locations D-10 Requirements
replacement 5-50 system setup 2-2
Restore
P CPU configuration data 5-22
data backup 4-13
Port Stats
Restore CPU Configuration Data 5-13
incubator 3-42, 7-54
LIS 3-85, 7-78
S
Post Service Validation 5-124
Power Entry Module Safety Summary 1-7
connector locations D-13 Scandisk Output 3-88, 7-75
fuses Schematics and Diagrams D-1
replacement 5-74
screen
Power Entry Module Fuse(s) calibrate module temperature 6-3
replacement 5-74
Select 1-6
Power Event Log
Setup System 2-1
controller module 3-54, 7-68
Single Cell Calibration
Power Panel Assembly 3-17
3D 60 software 6-15
Power Supply Cal 10 A software 6-19
adjust 6-25
Single Slug Calibration 6-8
DC power
replacement 5-70 Slide Bar 3-37
metal oxide varistor SLOER Form C-4
replacement 5-65
Software
solid state relay
association tables 3-62, 7-74
replacement 5-67
audit trail 3-81
transformer
replacement 5-62
bottle processing tables 3-63, 7-74
configuration tables 3-83
Power Supply Assembly 3-25
controller module 3-60, 7-72
Power Supply Data 3-27 debug 3-61, 7-72
Printer Port modem activity log 3-62, 7-74
description 3-4 operating system exception log 3-62, 7-73
Purpose of this Manual 1-2 software exception log 3-62, 7-73
T
Temperature
verification 4-14
Temperature Calibration 6-2
Temperature Log
incubator 3-45, 7-57
Terminal Strip TB1
connector locations D-11
Transformer 3-18
replacement 5-62
Troubleshooting
Fault/Status Code 7-16
Symptomatic 7-4
Typographic and Usage Conventions 1-4
U
UPS
connector locations D-5
domestic model
battery replacement 5-114
replacement 5-112
User Input 1-6
V
Verification
incubator temperature 4-14
Note: The Recommended Changes to Service Documentation Forms may only be used for
service documentation for the VITEK® 2, VITEK® 2 Compact, BacT/ALERT 3D®, and
PREVI™ Isola product lines.
SAMPLE
John Doe
John Doe, Field Service Engineer 123-456-7890
03/2010