Professional Documents
Culture Documents
Argene Adenovirus Flyer 4-13
Argene Adenovirus Flyer 4-13
Adenovirus
R-gene® US
respiratory infection. Easy assay set up and rapid turn around time
(< 1 h post extraction).
Reference Method
ADENOVIRUS COMPARISON RESULTS
Swab Specimens + - Total Comments
+ 44 43a 87 sensitivity 91.7% (80.0%-97.7%) 95% CI
Adenovirus R-gene® US
- 4b 1092 1096 specificity 96.2% (94.9%-97.2%) 95% CI
Total 48 1135 1183
a Among the 43 negative samples for rapid culture and positive for Adenovirus, 42 were confirmed as positive for adenovirus using Real Time PCR.
b Among the 4 positive samples for rapid culture and negative for Adenovirus, 1 was confirmed as positive for adenovirus using Real Time PCR.
Reference Method
ADENOVIRUS COMPARISON RESULTS
NP Wash/Aspirate Specimens + - Total Comments
+ 21 21a 42 sensitivity 100% (86.7%-100%) 95% CI
Adenovirus R-gene® US
- 0 351 351 specificity 94.4% (94.9%-97.2%) 95% CI
Total 21 372 393
a Among the 21 negative samples for rapid culture and positive for Adenovirus, 19 were confirmed as positive for adenovirus using Real Time PCR.
FDA cleared for extraction with the Adenovirus R-gene® US will enable the
NucliSENS® easyMAG® detection of all of the following serotypes:
bioMérieux, Inc.
100 Rodolphe Street
Durham, NC 27712
U.S.A.
Tel: (800) 682 2666
www.biomerieux.com/argene
Fax: (800) 968 9494
www.biomerieux-usa.com