You are on page 1of 8

Obstetrics 1

Intrapartum Assessment
Submitted by: Mamauag, Angela Beatriz C.
Trician Villarosa, MD
_________________________________________________________________________________________________________
I. ELECTRONIC FETAL MONITORING o The FHR is detected through the maternal
A. Internal (Direct) Electronic abdominal wall using the ultrasound Doppler
Monitoring principle
o Direct FHR measurement is accomplished by o The instrument contains a transducer that emits
attaching a bipolar spiral electrode directly to the ultrasound and a sensor to detect a shift in
fetus frequency of the reflected sound
o The wire electrode penetrates the fetal scalp, and o The transducer is placed in the maternal abdomen
the send pole is a metal wing in the electrode. at a site where fetal heart action is best detected
o The P wave, QRS complex, and T wave the o Coupling gel must be applied because air conducts
electrical fetal cardiac signal are amplified and fed ultrasound waves poorly
into a cardiotachometer for heart rate calculation. o The device is held in position by an elastic belt
o The peak R-wave voltage is the most reliably o Autocorrelation
detected portion of the fetal electrocardiogram  It is based on the premise that the FHR has
(ECG) regularity, whereas “noise” is random and
o Time (t) in milliseconds between fetal R waves is lacks regularity
fed into a cardiotachometer, and a new FHR is set o Other features of current external monitors
with the arrival of each new R wave include:
o Baseline variability  Ability to monitor twin fetuses
 It is the phenomenon of continuous R-to-R  Concurrently assess maternal heart rate
wave FHR computation formerly known as  Display fetal ECG
beat-to-beat variability (old)  Record maternal pulse oximetry values
o If present, fetal premature atrial contractions o Wireless, abdominal, dermal-path, external array
(PACs) cause the cardiotachometer to rapidly and systems
erratically seek new heart rates and create the  May improve the interpretability of
“spiking” intrapartum FHR tracings in obese women
o Electrical cardiac complexes detected by the fetal  Use electrocardiography to accurately
electrode include those generated by the mother. measure maternal and fetal heart rates yet
o However, the amplitude of the maternal ECG allow patient mobility during labor
signal is diminished when recorded through the
fetal scalp electrode. Thus, maternal cardiac
signals don’t appear in the FHR tracing C. Fetal heart rate patterns
o However, if the fetus is dead, maternal R waves o The interpretation of FHR patterns can be
can be still detected by the scalp electrode as the problematic without definitions and nomenclature
next best signal and are counted by the o Pattern interpretation derives from the FHR that is
cardiotachometer portrayed on the monitor or graph paper
o Thus, the choice of vertical and horizontal scaling
B. External (Indirect) Electronic greatly affects patterns
Monitoring o National Institute Child Health and Human
o Although it avoids membrane rupture, external Development Workshop scaling parameters
monitoring does not provide the precision of recommend that each centimeter on the vertical
internal FHR monitoring axis represents 30 beats per minute (bpm).
o Moreover, in some women, external monitoring  Measurement markers on this axis range from
may be difficult 30 to 240 bpm.
o Along the horizontal axis, the heart render moves
at a speed 3 m/min
o The time between each bold vertical line on this  Maternal administration of atropine
axis is 1 minute (parasympathetic drug) or terbutaline
o FHR patterns during labor are dynamic and (sympathomimetic drug)
interchange rapidly
o Thus, a full description should include an
Baseline variability
evaluation of changes or trends over time
o It is an important index of cardiovascular function
o It reflects a sympathetic and parasympathetic
D. Baseline fetal heart activity
o This refers to FHR baseline characteristics apart “push and pull” mediated by the fetal sinoatrial
node. This produces beat-to-beat fluctuations of
from periodic accelerations or decelerations
the baseline FHR
o Elements include:
o Variability describes these changes during 1
 Rate
minute, which modify the baseline’s waviness
 Variability
o The NICHD workshop defined baseline variability
 Distinct FHR patterns such as sinusoidal or
as irregular fluctuations in the baseline excluding
saltatory.
accelerations and decelerations
Rate o Normal variability shows oscillations that change 6
to 25 bpm
o Beginning at 16 weeks’ gestation, the heart rate
o FHR variability increases with advancing
drops approximately 1 bpm each week gestational age
o This continues postnatally such that the average o Increased Variability
rate is 85 bpm by age 8 years
 Greater variability accompanies fetal
o This normal gradual slowing of the FHR is thought
breathing and body movements
to correspond to maturation of parasympathetic
 In one study of 390 fetuses, greater
(vagal) heart control
intrapartum variability was associated with
o Baseline fetal heart rate
abnormal fetal arterial cord blood gas
 The approximate mean rate rounded to measurements but without increased
increments of 5 bpm during a 10-minute neonatal morbidity rates.
tracing segment o Decreased variability
 In any 10-minute window, the minimum  Absent variability and minimal variability are
interpretable baseline duration must be at defined as no baseline fluctuation or changes
least 2 minutes- otherwise the baseline is ≤5 bpm
considered indeterminate.
 Diminished variability can be an ominous sign
o Bradycardia
that indicates a seriously compromised fetus.
 Third Trimester - the normal mean baseline ○ Diminished variability, when it reflects fetal
FHR ranges between 110 and 160 bpm compromise, likely reflects acidemia rather
 Rate between 100 and 119 bpm, in the than hypoxia
absence of their FHR changes, usually is not  Severe maternal acidemia also can lower fetal
representative of fetal compromise variability
 Low baseline heart rates have also been  Reduced baseline FHR variability is the single
attributed to head compression from input most reliable sign of fetal compromise.
posterior or transverse positions, particularly  Another common cause of diminished
during secondstage labor variability is administration of some analgesic
 Fetal Bradycardia may stem from congenital drugs during labor which include narcotics,
heart block or from fetal hypoxia due to barbiturates, phenothiazines, tranquilizers and
maternal or fetal causes general anesthetics
o Tachycardia  Magnesium sulfate can diminish variability
 Maternal fever from chorioamnionitis and and is widely used in the United States for
other infections fetal neuroprotection, tocolysis, and seizure
 Fetal compromise – associated with prophylaxis in preeclampsia
concomitant HR decelerations
 Cardiac arrhythmia
 In summary, variability is affected by fetal  Major—25 or more bpm to quantify fetal risk
physiology, and its meaning differs depending o Pseudosinusoidal – intrapartum sine wavelike
on clinical setting baseline variation with periods of acceleration.
 A decrease in variability but without o Antepartum sine wave baseline undulation
decelerations is unlikely to reflect fetal portends severe fetal anemia
hypoxia
 A persistently flat FHR baseline—absent
variability—within the normal baseline rate
range and without decelerations may reflect a
prior fetal insult that has resulted in A. Periodic fetal heart rate changes
neurological damage o Periodic FHR – refers to deviations from baseline
that are temporally related to uterine contractions
Cardiac arrhythmia o Acceleration – refers to an increase in FHR above
baseline
o When fetal cardiac arrhythmias are first suspected
o Deceleration - refers to drop below the baseline
using electronic monitoring, findings include
rat
 Baseline bradycardia
 Tachycardia Accelerations
 Abrupt baseline spiking
o Intermittent baseline bradycardia is frequently o Onset of acceleration to a peak in less than 30
due to congenital heart block seconds— in the FHR baseline.
o Conduction defects, most commonly complete o At 32 weeks’ gestation and beyond, the
atrioventricular (AV) block, usually are found in acceleration has a peak of 15 bpm with a duration
association with maternal connective-tissue of 15 seconds or more but less than 2 minutes
diseases o Before 32 weeks, a peak of 10 bpm for 15 seconds
o Scalp electrodes – fetal monitors can be adapted to 2 minutes is considered normal
to output the scalp electrode signals into an o Prolonged acceleration was defined as 2 minutes
electrocardiographic recorder or more but less than 10 minutes
o Fetal heart accelerations during the first or last 30
Sinusoidal heart rate minutes during labor, or both, was a favorable sign
for fetal wellbeing
o True sinusoidal pattern may be observed with
o There is an approximately 50-percent chance of
 Fetal intracranial hemorrhage
acidemia in the fetus who fails to respond to
 With severe fetal asphyxia
stimulation in the presence of an otherwise
 With severe fetal anemia from rh
nonreassuring pattern
alloimmunization
o Most often occur antepartum, in early labor, and
 Fetomaternal hemorrhage
in association with variable decelerations.
 Twin-twin transfusion syndrome
o Proposed mechanisms for intrapartum
 Vasa previa with bleeding
accelerations include:
o Insignificant sinusoidal patterns have been
 Fetal movement
reported following administration
 Stimulation by uterine contractions
 Meperidine
 Umbilical cord occlusion
 Morphine
 Fetal stimulation during pelvic examination
 Alphaprodine
 Fetal scalp blood sampling
 Butorphanol
 Acoustic stimulation
o Sinusoidal pattern also has been described with
 Chorioamnionitis Deceleration
 Fetal distress
o Decrease below the baseline rate.
 Umbilical cord occlusion
o The nomenclature most commonly used in the
o Other investigators have proposed a classification
United States is based on the timing of the
of sinusoidal HR patterns
deceleration in relation to contractions—thus,
 Mild—amplitude 5 to 15 bpm
early, late, or variable in onset related to the
 Intermediate—16 to 24 bpm
corresponding uterine contraction
o The waveform of these decelerations is also o Late decelerations alone do not predict fetal
significant for pattern recognition. acidemia
o Early and late decelerations: the slope of FHR o Late decelerations plus decreased variability,
change is gradual, resulting in a curvilinear and which itself is an indicator of less activity in the
uniform or symmetrical waveform cardioregulatory center of the medulla, are more
o With variable decelerations: the slope of FHR predictive of fetal acidemia and neonatal
change is abrupt and erratic, giving the waveform morbidity
a jagged appearance o The following can induce late deceleration
o Description of decelerations is based on the  hypotension from epidural analgesia (most
pathophysiological events: common)
 Early decelerations are termed head  uterine hyperactivity from oxytocin
compression stimulation (most common)
 Late decelerations are termed uteroplacental  maternal hypotension
insufficiency  excessive uterine activity
 Variable decelerations become cord  placental dysfunction
compression patterns

Early deceleration
Variable deceleration
o Consists of a gradual decrease and return to
o Defined as a drop in the FHR that begins with the
baseline associated with a contraction
contraction’s onset and reaches a nadir in <30
o Physiologically, there is a HR drop with
seconds.
contractions and that this was related to cervical
o The deceleration lasts between 15 secs and 2
dilatation.
minutes, its depth is ≥15 bpm in amplitude
o Generally seen in active labor between 4 and 7 cm
o The onset of deceleration typically varies with
dilatation
successive contractions.
o The degree of deceleration is generally
o Most frequent deceleration pattern encountered
proportional to the contraction strength and
during labor
rarely falls below 100 to 110 bpm or 20 to 30 bpm
o Attributed to umbilical cord occlusion
below baseline
o The slope of FHR change is abrupt and erratic,
o Head compression probably causes vagal nerve
giving the waveform its jagged shape.
activation as a result of dural stimulation, and this
o Often reflects some degree of umbilical cord
mediates the HR deceleration
occlusion
o Represents reflexes that reflect either blood
pressure changes due to interruption of umbilical
Late deceleration
flow or changes
o This deceleration is a smooth, gradual, o Recurrent variable decelerations with minimal to
symmetrical decline in the FHR that begins at or moderate variability are indeterminate
after the contraction peak and returns to baseline o Recurrent variable decelerations with absent
only after the contraction has ended. variability are abnormal
o It reaches its nadir within 30 seconds of its onset o A saltatory FHR baseline is another pattern linked
o Late deceleration’s depth is <10 to 20 bpm below to umbilical cord complications during labor
baseline o The pattern is rapidly recurring couplets of
o These usually are not accompanied by acceleration and deceleration causing relatively
acceleration large oscillations of the FHR baseline
o Often reflect poor uterine perfusion or placental o In the absence of other FHR findings, this pattern
dysfunction. does not signal fetal compromise
o Late decelerations are the 1st FHR consequence of
uteroplacental-induced hypoxia.
Prolonged deceleration
o Fetal acidemia develops after frequent or
prolonged periods of hypoxia.
o This pattern is defined as an isolated deceleration o The only benefits reported for scalp pH testing are
with a depth >15 bpm and length ≥2 minutes but fewer cesarean deliveries for fetal distress
<10 minutes from onset to return to baseline o The scalp pH sampling does not increase delivery
o Difficult to interpret because they are seen in rate for fetal distress. They concluded that scalp
many different clinical situation pH sampling was unnecessary
o Epidural, spinal, or paracervical analgesia may o Fetal scalp blood lactate concentration was used
induce a prolonged deceleration as an adjunct to pH
o Scalp blood pH analysis and scalp blood lactate
analysis are equivalent in predicting fetal acidemia
B. Fetal heart rate patterns during
second stage labor
o Decelerations are virtually ubiquitous during the
second stage labor
B. Scalp stimulation
o Alternative to scalp blood sampling
o Both cord and fetal head compressions are
o HR acceleration in response to pinching of the
implicated as causes decelerations and baseline
scalp with an Allis clamp just before obtaining
bradycardia in this stage
blood was invariably associated with a normal pH
o Failure to provoke acceleration was not uniformly
predictive of fetal acidemia
C. Computerized interpretation
o FHR pattern interpretations are subjective. Thus,
the potential for computer assistance to enhance
the precision identifying abnormal patterns
appeared promising C. Vibroacoustic stimulation
o FHR acceleration in response to vibroacoustic
stimulation has been recommended as a
substitute for scalp sampling
o Response to vibroacoustic stimulation is
II. OTHER INTRAARTUM ASSESSMENT considered normal if FHR acceleration of at least
TECHNIQUES 15 bpm for at least 15 seconds occurs within 15
A. Fetal scalp blood sampling seconds after the stimulation and with prolonged
o Measurements of the pH in capillary scalp blood
fetal movements
may help to identify the fetus in serious distress o Vibroacoustic stimulation is an effective predictor
o It also emphasized that neither normal nor
of fetal acidosis in the setting of variable
abnormal scalp pH results have been shown to be
decelerations
predictive of infant outcome
o Although vibroacoustic stimulation in second-
o The procedure is now used uncommonly and is
stage labor is associated with FHR reactivity, the
not even available at some hospitals quality of the response did not predict neonatal
o The pH of fetal capillary scalp blood is usually outcome or enhance labor management
lower than that of umbilical venous blood and
approaches that of umbilical arterial blood.
o Confirmation of fetal distress
D. Fetal pulse oximetry
 If the pH is > 7.25, labor is observed
o Technology similar to that of adult pulse oximetry.
 If between 7.20 and 7.25, the pH
o Tool measures fetal oxyhemoglobin saturation
measurement is repeated within 30 minutes.
once membranes are ruptured
 If the pH is <7.20, another scalp blood sample
o Unique padlike sensor is inserted through the
is collected immediately, and the mother is
cervix and positioned against the fetal face
taken to an operating room and prepared for
o 30% → lower limit for normal fetal oxygen
surgery
saturation
 Delivery is performed promptly if the low pH
o U.S. Food and Drug Administration (FDA)
is confirmed. Otherwise, labor is allowed to
approved → Nell-cor N-400 Fetal Oxygen
continue, and scalp blood samples are
Monitoring System.
repeated periodically
E. Fetal electrocardiography o Correct maternal hypotension caused by regional
o As fetal hypoxia worsens, the fetal ECG changes. analgesia and discontinuing oxytocin both serve to
o Mature fetus with hypoxemia develops an improve uteroplacental perfusion. Vaginal
elevated ST segment and a progressive rise in the examination excludes a prolapsed cord or
T-wave height that can be expressed as a T:QRS impending delivery
ratio o Three maneuvers that significantly raise fetal
o Increasing T:QRS ratios oxygen saturation levels
 To reflect fetal cardiac ability to adapt to  IV hydration with 500 to 1000 mL lactated
hypoxia Ringer solution given over 20 minutes
 Appear before neurological damage  Lateral versus supine positioning;
o Further worsening of hypoxia then leads to  Administration supplemental oxygen at 10
progressively negative ST-segment deflection that L/min using a nonrebreathing mask
creates a biphasic form
Tocolysis

o Terbutaline sulfate given to relax the uterus can be


III. NONREASSURING FETAL STATUS a temporizing maneuver in the management of
o Nonreassuring designation suggests inability to nonreassuring FHR patterns during labor
remove doubt o A single 250-μg IV or subcutaneous injection is
o Heart rate control stems from interconnected used to inhibit uterine contractions and thereby
mechanisms that depend on blood flow and improve fetal oxygenation
oxygenation o The American College of Obstetricians and
o Normal labor is a process of repeated fetal hypoxic Gynecologists (2019b) cites that evidence is
events and an infrequently lead to significant insufficient to recommend tocolysis for
acidemia nonreassuring FHR patterns. We consider
terbutaline during labor stimulation to resolve
tachysystoleassociated prolonged decelerations.
A. Diagnosis However, the chance of success is balanced
o Category I or II training with a 5-minute Apgar
against terbutaline’s side effects should cesarean
score >7 or with normal arterial blood acid–base
delivery be needed for unresolved bradycardia.
values is not consistent with an acute hypoxic-
o Contraindications
ischemic event
 Maternal cardiopulmonary disease
o Five-tier system had better sensitivity than the
 Poorly controlled hyperthyroidism
three-tier one
 Maternal or fetal tachycardia with loss of
variability

IV. MECONIUM IN THE AMNIOTIC FLUID Amnoinfusion


o Fetuses may pass meconium in response to
o For variable decelerations, other studies also
hypoxia, and meconium therefore signals fetal
support amnioinfusion to reduce variable
compromise
deceleration frequency, improve neonatal
o However, in utero passage meconium may
outcome, and lower cesarean delivery rates.
represent normal gastrointestinal tract maturation
o Amnioinfusion is a reasonable approach to treat
under neural control
repetitive variable decelerations but not late
o Common but transient umbilical cord entrapment
decelerations
leads to vagal stimulation, increased bowel
o Attempts to flush out or dilute thick meconium
peristalsis, and meconium passage
are not recommended
o Despite differing amnioinfusion protocols, most
A. Management options provide a 500- to 800-mL bolus of warmed normal
o Initial management variant FHR patterns aim to saline, which is followed by a continuous infusion
correct any fetal insult, if possible of 3 mL/min
o The woman is moved to lateral position, and
supplemental oxygen is provided by mask.
B. FHR patterns and brain injury
o Complete asphyxia was produced by total o Uterine performance
occlusion umbilical blood flow that led to  Product of contraction intensity in mm Hg
prolonged deceleration multiplied by the number of contractions in a
o Fetal arterial pH did not drop to 7.0 until 10- minute span
approximately 8 minutes after complete cessation  For example, three contractions in 10 minutes,
of oxygenation and umbilical with at least 10 each with 50-mm Hg intensity, would equal
minutes prolonged deceleration were required 150 Montevideo units
before surviving fetuses demonstrated brain o Clinical labor
damage  Reaches values between 80 and 120
o Persistent nonreactive FHR tracing was already Montevideo units
present at the time of admission in 70% of  Three contractions of 40 mm Hg every 10
neurologically impaired fetuses. minutes
 No clear-cut division marks labor onset
o First-stage labor
C. Benefits of EFHR monitoring  25 mm Hg at labor commencement to 50 mm
o Fetal monitor – implies that inanimate technology Hg at its end
in some fashion “monitors.”  Advances from three to five contractions per
o The assumption is made that if a dead or damaged 10 minutes
infant is delivered, the tracing strip must provide  Uterine baseline tone rises from 8 to 12 mm
some clue, because the fetal monitor was Hg
monitoring fetal condition → Led to great o Second stage labor
expectations and the belief that all neonatal  Aided by maternal pushing
deaths or injuries were preventable  Contraction intensity of 80 to 100 mm Hg is
o Electronic fetal monitoring increased the rate of typical
cesarean and operative vaginal deliveries but  Five to six times each 10 minutes
produced no declines in rates of perinatal o Uterine contractions are clinically palpable only
mortality, neonatal seizures, or cerebral palsy  Exceeds 10 mm Hg
o Despite being widely used; electronic fetal  Until the intensity of contractions reaches 40
monitoring has failed as a public health screening mm Hg, the uterine wall can readily be
program. depressed by the finger
o Fetal monitoring increased operative delivery  At greater intensities, the uterine wall then
rates but decreased early neonatal mortality rates becomes so hard that it resists easy
on singleton live births depression
o Temporal increase in fetal monitoring was  Not associated with pain until their strength
associated with a decline in neonatal mortality exceeds 15mm Hg.
rates, especially in preterm gestations o The uterus that performs poorly before delivery is
also prone to atony and puerperal hemorrhage

V. INTRAPARTUM SURVEILLANCE OF
UTERINE CAVITY B. Uterine contraction terminology
o Analysis of electronically measured uterine o Normal uterine activity
activity  Defined as five or fewer contractions in 10
o Permits some generalities concerning the minutes, averaged during a 30-minute span
relationship of certain contraction patterns to o Tachysystole
labor outcome  More than five contractions in 10 minutes,
o Uterine muscle efficiency to effect delivery varies averaged over 30 minutes
greatly  Applied to spontaneous or induced labor

C. Electronic fetal monitoring complications


A. Patterns of uterine activity o Rarely, an intrauterine pressure catheter may
o Contraction intensity- defined as the rise in this lacerate a fetal vessel in the placenta during
pressure above a resting pressure baseline placement
o Also with insertion, placental and possibly uterine
perforation can cause hemorrhage, abruption, and
spurious recordings that have resulted in
inappropriate management
o Severe cord compression has been described from
entanglement with the pressure catheter
o Injury to the fetal scalp or breech by a FHR
electrode is rarely severe. However, face
presentations pose risk for eye or mouth trauma.
o Both the fetus and the mother may be at greater
risk of infection from internal monitoring
o Scalp wounds from the electrode may become
infected, but subsequent cranial osteomyelitis is
rare

You might also like