Philips 50T Fetal Telemetry System - Service Manual
Philips 50T Fetal Telemetry System - Service Manual
M1310-9000B
Printed in Germany February 2002
Edition 1, A.00.01
Notice
Philips makes no warranty of any kind with regard to this material, including,
but not limited to, the implied warranties of merchantability and fitness for a
particular purpose. Philips shall not be liable for errors contained herein or
for incidental or consequential damages in connection with the furnishing,
performance or use of this material.
The information contained in this document is subject to change without
notice.
Philips assumes no responsibility for the use or reliability of its software on
equipment that is not furnished by Philips.
Important
United States federal law restricts this device to sale by or on the order of a
physician.
This device is not intended for home use.
Warning
Failure on the part of the responsible individual hospital or institution
employing the use of this equipment to implement a satisfactory
maintenance schedule may cause undue equipment failure and possible
health hazards.
Printing History
M1310-9000A April 1994
M1310-9000B February 2002
Warning
A warning calls attention to a condition or possible situation that could cause
injury to the user and/or patient.
Caution
A caution calls attention to a condition or possible situation that could damage
or destroy the product or the user’s work.
1.Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Patient Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Protective Earth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Maximum Input/Output Voltages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Telemetry System (Receiver and Transmitter) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Dimensions and Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Channels and Frequencies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Antenna. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Inputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Input Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Output Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Image Rejection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Self-Test Facilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Transducers and Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
External Blue Toco Transducers (M1355A) or (M1355A Opt. C03) . . . . . . . . . . . . . . . . . . . . . . . 8
Blue Ultrasound Transducer (M1356A) or (M1356 Opt. C03) . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
DECG Transducer (M1357A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
DECG/MECG Patient Module (M1364A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
IUP Quartz Transducer (1290C #J05) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
IUP Pressure Transducer (CPJ840J5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Remote Event Marker (15249A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Solving General Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
No LEDs Lit on Receiver (continued overleaf) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Not all Receiver LEDs are Lit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Yellow LED Remains Lit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Transmission INOP Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Contents v
Transmission Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Transducer Mode Not Detected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Cardio Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
TOCO Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
FMP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
FMP Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Event Marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Nurse Call . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
7.Replacing Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Ordering Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Service Tools. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Lists of Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Receiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Dismantling the Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Transmitter Processor Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Transmitter VCXO. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Dismantling the Receiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Power Supply Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Processor Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
RF Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Receiver VCXO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
RF Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Display Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
9.Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
vi Contents
Transmitter Functional Blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
US Gating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
US Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
US Receiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
ECG Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
ECG Receiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Dynamic Compression Circuitry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
FMP Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
Toco Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
Toco Receiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
Toco A/D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
FSK Generator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
Oscillator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
Service Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
EEPROM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
Japan ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
VCXO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
Receiver Functional Blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
Input Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
RF Preamp M1406A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
RF Receiver M1402A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
US Bandpass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
ECG Bandpass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
FSK Bandpass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
FSK Demodulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
Toco D/A Converter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
EEPROM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
Service Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
Oscillator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
Contents vii
France . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Germany . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Italy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Japan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Luxembourg . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Netherlands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
New Zealand. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Norway . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Service USA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Singapore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Spain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Sweden . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Switzerland . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
UK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
USA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
viii Contents
List of Figures
List of Figures ix
x List of Figures
1
Safety
Safety Symbols
Read this information before setting up, using or servicing your Series 50 T
Telemetry System.
FCC WARNING
This equipment generates, uses and radiates radio-frequency energy, and
if it is not installed and used in accordance with this manual, may cause
interference to radio communications.
Caution
Remove the batteries if you do not intend to use the Transmitter for a
long period of time, to avoid damage from leakage.
Chapter 1 - Safety
Patient Safety
Patient Safety
The Telemetry System should only be used by, or under the direct supervision of, a
licensed physician or other health care parctitioner who is trained in the use of fetal
heart rate monitors and in the interpretation of fetal heart rate traces. US federal law
restricts this device to sale by, or on the order of, a physician.
The Telemetry Receiver is a Protection Class 1, Type B instrument. It is
designed to fulfill safety requirements according to IEC 60601-1, UL 544 and CSA-
C22.2 No.601.1-M90.
The telemetry transmitter is a battery operated device, applied parts (patient
connectors) are Type CF .
Environment
Use the system in an environment which is reasonably free from vibration, dust,
corrosive or explosive gases, extremes of temperature and humidity. It operates within
specifications at ambient temperatures between 0 and +45°C. Ambient temperatures
that exceed these limits can affect the accuracy of the system, the transmitter radio
frequency transmission, and can damage the components and circuits.
The system (not including transducers1) can be stored at ambient temperatures
between -40°C and +75°C.
The blue Toco and Ultrasound transducers are water-tight to a depth of 50 cm (19.7
inches).
Warning
Do not use brown transducers to monitor patients under water. Ultrasound and
Toco transducers that are colored blue are watertight and comply with IEC 529
(IP68).
You can immerse the BLUE Ultrasound and Toco tranducers in water ONLY
when connected to the telemetry transmitter. NEVER connect blue transducers
directly to the fetal monitor when they are immersed in water, or likely to come
in contact with water.
1. Transducers can be stored at temperatures between -40°C and +60°C (-40°F to +140°F).
Chapter 1 - Safety
Protective Earth
Protective Earth
To protect hospital personnel and the patient, the cabinet must be grounded.
Accordingly, the receiver is equipped with a 3-wire power cable which grounds it to
the power line ground when plugged into an appropriate 3-wire receptacle. Do not
use a 3-wire to 2-wire adapter with the receiver. Any interruption of the protective
earth grounding will cause a potential shock hazard that could result in serious
personal injury.
Whenever it is likely that the protection has been impaired, the receiver must be made
inoperative and be secured against any unintended operation.
Warning
Check each time before use that the Telemetry System is in perfect working
order and the receiver is properly grounded.
The patient cable must be positioned so that it does not come into contact with any
other electrical equipment.
Before operation, make sure that the receiver is free from condensation. This can
form when equipment is moved from one building to another, and is exposed to
moisture and differences in temperature.
Warning
Possible explosion hazard if used in the presence of flammable anesthetics.
Chapter 1 - Safety
Maximum Input/Output Voltages
Transmitter
1. Service Socket
Maximum voltage of ±12V.
2. Socket to Fetal Monitor
Maximum voltage of ±12V.
3. Power Input Socket
100-120V ~ or 220-240V ~
4. Event Marker/Service Socket
Maximum Voltage of +5V.
5. Toco Socket
Maximum Voltage of +5V.
6. Cardio Socket
Maximum Voltage of +5V.
Chapter 1 - Safety
Technical Specifications
Power Requirements
The Telemetry System is set for the correct voltage at the factory, but before
you connect power, ensure that the voltage switch is in the correct position
for your country.
Operating Voltage: 100V - 120V~ or 220V - 240V (±10%).
Line Frequency: 50 to 60Hz.
Power Consumption: 19VA max.
Battery Type: 3 x 1.5V (AA size).
Environment
The Telemetry System should be used in an environment which is reasonably
free from vibration, dust, corrosive or explosive gases, extremes of
temperature and humidity. It operates within specifications at ambient
temperatures between 0 and 45°C . Ambient temperatures which exceed
these limits can affect the accuracy of the Telemetry System, the Transmitter
radio frequency transmission, and cause damage to the components and
circuits.
Operating Temp: 0 to +45°C (32°F to 113°F).
Storage Temp1: -40 to +75°C (-40°F to 167°F).
Relative Humidity: 5 to 95%.
Antenna Receiver
BNC Female input connector (standard 50 Ohms input). For standalone or
connection to an antenna system.
Transmitter
Built-in Antenna in US and DECG transducer cables.
Inputs Receiver
• Equipotential Grounding Point.
• Service Socket.
Transmitter
• Cardio socket for M1356A ultrasound transducer or the M1357A/
M1364A DECG transducer.
• Toco socket for the M1355A Toco transducer or an IUP Transducer
(CPJ840J5).
• Remote Event Marker and Service Socket.
Outputs Receiver
• 25 pin RS232 Socket for connecting the Fetal Monitor.
There are two types of Toco, Ultrasound and DECG Transducers, the
Standard Series 50 and Telemetry Transducers. The Standard Transducers
have a 2.5m/8ft 2in ventilated cable and the Telemetry Transducers have a
70cm/28in ventilated cable.
Both types of Toco and Ultrasound Transducers are colored blue and are
watertight complying with IEC 529 (IP68)
Warning
Under no circumstances should the transducer be immersed in water
when connected to the Fetal Monitor.
Caution
Blue Toco Transducers have a ventilated cable. As liquid can enter this
cable through capillary action, extra care should be taken not to get the
connector wet.
3. Check the red light (B) situated behind the middle battery. If:
• the red light is lit for a few seconds and then goes out, the self test is
successfully complete.
• the red light blinks, or remains on after three seconds, change the
batteries. If the fault continues, there is a hardware error. See
troubleshooting flowcharts in Chapter 4.
The parameter test checks the signal path to and from the transducer sockets,
but not the transducers themselves.
1. Switch on the monitor, the recorder and the telemetry receiver.
3. Standing within view of the monitor, press the Nurse Call Button on
the transmitter and switch on the transmitter simultaneously. The test
runs while the Nurse Call is pressed.
Error Messages
The following error messages are directly related to telemetry and appear on
the fetal monitor. Refer to the Instructions for Use provided with your monitor
for error messages not related to telemetry monitoring.
Series 50 Family
Err 9 US Toco Invalid telemetry mode. Check the cable from the telemetry
receiver and, if necessary, replace it.
Err 14 US Toco Incorrect transducer Check that the transducer is
connected to transmitter. compatible with Series 50T Fetal
Telemetry System.
Err 16 US Toco Transducers are Disconnect the transducers from
connected to the front the monitor or switch off the
panel of the monitor. telemetry receiver.
8040A*
Err 16 US Toco Transducers are Disconnect the transducers from the monitor or
connected to the front switch off the telemetry receiver.
panel of the monitor.
*. No longer available.
8041A*
*. No longer available
The Telemetry Incorrect interface connection between Follow the instructions in Instructions for Use for
Indicator Lamp on the the monitor and the receiver. details on how to connect the monitor to the
fetal monitor does not receiver.
light when the monitor
and the receiver are Faulty interface cable. Replace interface cable.
switched on.
Receiver Power On Power cable not plugged into the power Plug in and switch on.
Light does not light supply.
when the receiver is
switched on. Fuses need replacing. Replace fuses.
Transmission INOP Receiver and transmitter do not have Check channel number and the serial numbers
light on the receiver is the same channel or serial number. are the same on the receiver and the transmitter.
still lit when the
transmitter is switched Batteries in the transmitter are Change the batteries in the transmitter (refer to
on. exhausted. the Instructions for Use for details.)
Chapter 4 - Troubleshooting
Solving General Problems
Comes On
Comes On
Comes On
Operating
Operating
Operating
3 x 1.5V
Size: AA
Time
Time
Time
Type: LR6
Alkaline (1.8 Ah) 180 min 40 hrs 100 min 16 hrs 80 min 14.5 hrs
NiCd (0.6 Ah) 10 min 12 hrs 6 min 5.5 hrs 5 min 4.5 hrs
Rechargeable
Chapter 4 - Troubleshooting
Solving General Problems
Chapter 4 - Troubleshooting
Solving General Problems
Chapter 4 - Troubleshooting
Solving General Problems
Chapter 4 - Troubleshooting
Solving General Problems
Chapter 4 - Troubleshooting
Solving General Problems
Transmission Range
Chapter 4 - Troubleshooting
Solving General Problems
Chapter 4 - Troubleshooting
Cardio Channel
Cardio Channel
Chapter 4 - Troubleshooting
TOCO Channel
TOCO Channel
Chapter 4 - Troubleshooting
FMP
FMP
FMP Test
To test an ultrasound transducer:
1. Connect the telemetry receiver to the fetal monitor using the interface cable.
2. Switch on the fetal monitor, its recorder and the telemetry receiver.
Chapter 4 - Troubleshooting
FMP
6. Apply a small amount of gel on the transducer surface. Holding the transducer
in one hand, gently tap the transducer surface with your finger.
Chapter 4 - Troubleshooting
Event Marker
Event Marker
Chapter 4 - Troubleshooting
Nurse Call
Nurse Call
Chapter 4 - Troubleshooting
Interfacing to a Fetal Monitor
The Fetal Monitor Interface on the Telemetry Receiver allows digital and
analog ouputs to the Monitor. The following table lists the signals at each
pin.
Prerequisites
Run tweetsrv.exe
You can add /? for HELP how to run
/1 for COM1 (default)
/2 for COM2
/3 for COM3
/4 for COM4
Example :/tweetsrv.exe /2
When the program has been loaded onto your PC, the main menu is
displayed. (If the menu is not displayed an error message is displayed along
the bottom of the screen).
Main Menu
--------------------------------------------------------------------------------------
--M1310A Service Software Rev.A.xx.xx--
MAIN MENU
--------------------------------------------------------------------------------------
Select with >cursor keys < up >, < down >,< enter >
--------------------------------------------------------------------------------------
Program S/N to Programs the Serial Number to the Transmitter (if no Serial Number is
Transmitter present on the Transmitter). This feature is needed when the Transmitter
Printed Circuit Board is exchanged.
Follow the steps as the program requests.
When moving the service connector from one device to the other, a "SIO
RCV error" may be displayed. This is due to the PC’s serial interface circuitry
and can be ignored.
If a Serial Number needs to be RESET, contact the factory Technical
Marketing Engineer to get RESET-CODE needed to RESET the Serial
Number.
Program S/N to Programs the Serial Number to the Receiver (if no Serial Number is present
Receiver on the Receiver). This feature is needed when Receiver-PC-Bd. is exchanged.
Follow the steps as the program requests.
When moving the service connector from one device to the other, a "SIO
RCV error" may be displayed. This is due to the PC’s serial interface circuitry
and can be ignored.
If a Serial Number needs to be RESET, contact factory Technical Marketing
Engineer to get RESET-CODE to RESET the Serial Number.
Power On Selftest The PC acts as a Terminal. The Transmitter or Receiver connected to the PC
must be switched OFF and then ON. The PC displays the Start-up messages
from the connected Transmitter or Receiver.
Show last errors/ Shows the summary of the received Errors and Warnings.
warnings
Read SerNum and When selected from the Main Menu, this sub-menudisplays:
Revisions
• Software Revision
• Serial Protocol Revision
• Serial Number (S/N)
Reset Serial Number This feature is only available for HP Service Personnel. It allows you to reset
the Serial Number on the Transmitter or Receiver Board if the Serial Number
was programmed incorrectly.To use this section a Reset-Code from a factory
Technical Marketing Engineer is needed.
The Technical Marketing Engineer requires the following information:
• Serial Number (S/N)
• Transmitter OR Receiver
• Reason for RESET request
The Technical Marketing Engineer will provide you with:
The RESET-CODE.
Caution
Never erase the Serial Number from both the transmitter and the
receiver, as this status is not recoverable and requires factory assistance.
Warning
Electrical power is dangerous. Before removing any parts from the
Receiver, the power to the Receiver must be disconnected.
Caution
Static electricity will damage sensitive electrical circuitry. Ensure you are
adequately grounded before touching any circuit board or its associated
components.
Ordering Parts
To order a part, contact your local Philips Sales/Service Office, quoting the
part number listed in the tables. To order a part not listed in the tables, give
the following information:
• The model number of the Telemetry System.
• The complete serial number of the Telemetry System.
• A description of the part, including its function and location.
Caution
Occasionally, as well as markings indicating manufacture by Philips,
electronic components will carry standard commercial identification
numbers. These components have been selected to meet specific
operational criteria. Using components purchased through normal
commercial channels may result in degradation of the operation
performance or the reliability of the Receiver or Transmitter.
Service Tools
You should have available all the boards and parts listed in the following
tables, plus the following tools:
• Small crosshead screwdriver.
• Medium slothead screwdriver.
• Large crosshead screwdriver.
• Box screwdriver (HBA).
• Spanner (M6 and M5)
• Digital Volt Meter.
• Small pincers or tweezers (useful, but not essential).
Lists of Parts
Transmitter
Receiver
11 Washer 2190-0676
Transducers
'HVFULSWLRQ 3DUW1XPEHU
4. Using a small flat bladed screwdriver, carefully prise the VCXO off the
board. Do not lean or put excessive pressure upon the board as it
contains sensitive components which can be easily damaged.
Transmitter VCXO
2. Lift top board from bottom board by lifting from the cardio side.
5. Using a suitable flat bladed screwdriver, carefully prise the VCXO out
of the board.This task must be done carefully as excessive pressure
exerted upon the board may damage delicate components.
3. Replace the connector rings onto the cardio and toco sockets.
4. Insert the board into the housing at an angle, Toco side first.
6. Connect the top housing to the bottom housing and snap the two
parts together.
9. Switch on and ensure that the green On LED is lit. Also check the red
LED inside the battery compartment (situated behind the middle
battery) by lifting the battery cover.
2. Remove antenna.
5. Lift off.
Note— Cases with spacers are not interchangable with those earlier
models that do not have them fitted.
To remove the Power Supply Board, first, follwing the instructions for
dismantling the Receiver, then:
1. Remove power switch manipulator (A).
Processor Board
To remove the Processor Board, first follow the instructions for dismantling
the Receiver, then:
RF Module
To remove the RF Module, first follow the instructions for dismantling the
Receiver, then:
1. Disconnect the flat cable (B).
2. Remove the coax cable from the Receiver Board Assembly with M6
spanner (A).
Receiver VCXO
1. Remove the 7 screws on the underside of the Receiver Board Assembly
and remove the cover.
3. When fitting the replacement VCXO, angle the new part and locate its
pins in one side of the socket. Carefully seat the VCXO, applying
gentle pressure until it is securely in position.
RF Amplifier
To remove the RF Amplifier, first follow the instructions for dismantling the
Receiver, then:
1. Remove the 3 screws that hold the RF Amplifier Board.
Display Board
To remove the Display Board, first follow the instructions for dismantling the
Receiver, then:
1. Remove the small flat cable from the Display Board.
Fuses
4. Remove the fuse from the holder and replace it with another of the
correct value.
Introduction
Caution
Static electricity will damage sensitive electrical circuitry. Ensure you are
adequately grounded before touching any circuit board or its associated
components.
The goal of the RF Bench Test is to find out which of the following sections is
defective:
Transmitter VCXO (M1400-6xxxx)
Receiver Preamplifier Board (M1406-60560)
Receiver Assembly (M1402-6x30x)
Receiver VCXO (M1402-61xxx)
If the operating range is decreased or no transmission is possible, the
following tests given in this chapter should be performed.
3. Pull up the housing top and lift out the Transmitter Board Assembly
from the bottom housing.
4. Place the Transmitter Board Assembly into housing top again and load
the batteries. Ensure the power switch manipulator is in the off
position.
3. Unplug the BNC cable between the Preamplifier and the Receiver
assembly at the Preamplifier BNC connector.
8. Remove the BNC connector from the Preamplifier input and connect
this cable to the Spectrum Analyzer.
1. Remove the flat cable and coax cable from the Receiver Assembly.
2. Pull out and turn over the Receiver Assembly and undo all 7 screws.
3. Lift up the top cover of the Receiver Assembly and connect the flat
cable again.
5. If all voltages are present and the green LED is flickering now test
Receiver VCXO.
6. If all voltages are correct and the green LED is not flickering, replace
the Receiver Assembly. Check for processor part number higher than
or equal to 1821-1141 on the Receiver Assembly. If any voltage is
incorrect replace the flat cable. If the problem is still not solved, replace
Receiver Processor Board (503).
You require a Spectrum Analyzer and Active Probe to carry out this test.
1. Check the Receiver VCXO ouput signal at the transistor amplifier.
US Gating
The US Gating circuitry generates the gated 1 MHz bursts for the US driver
and the US receiver necessary in a pulsed doppler system.
The burst repetition rate is 3.2 kHz. The US driver burst has a duration of
96 µsec. The burst for the Receiver’s demodulator, which also has a duration
of 96 µsec, follows the driver bursts with a delay of 32 µsec. This timing gives
a sensitive depth range from 2.4cm up to 16.8cm (with a sound velocity in
human tissue of 1500 m/sec.)
US Driver
The US Driver circuitry drives the ultrasound transducer crystals with the
bursts delivered by the US Gating module. The output amplitude is 4V peak
to peak. The drivers output goes to a high impedance state when the driving
burst is inactive.
US Receiver
The US Receiver consists of three submodules:
• Preamplifier
• Demodulator
• Bandpass filter with amplification
The preamplifier amplifies the small doppler shifted 1MHz signals from the
transducer. These are converted to the low frequency baseband by the
ECG Driver
This circuitry powers an ECG transducer with a 5V peak to peak, 250 kHz
square wave. The driver current is sensed by the ECG Receiver.
ECG Receiver
After converting into a voltage, the signal is amplified and bandpass filtered.
The filter consists of a 15 Hz highpass and a 100 Hz lowpass. The overall
gain from the transducer inputs to the ECG Receiver’s output is 56 dB. In
addition, the ECG Receiver outputs an INOP flag if the 50/60 Hz signal or
an INOP signal (250Hz delivered by the ECG transducer if electrodes have
no contact) of a too high amplitude is detected.
FMP Detector
The FMP Detector consists of two parts:
• The FMP Envelope
• The FMP Detector
The FMP Envelope circuitry builds the envelope signal of the fetal movement
profile doppler signal from the US receiver. It is a full wave rectifier followed
by a 12 Hz lowpass.
The Fetal Movement Profile Detector searches for fetal movement related
signals in the FMP envelope signal. If a movement is detected, a status bit is
Modes
This module checks for connected transducers at the Cardio and Toco
connectors. If a valid transducer is detected, the corresponding operating
modes are set in the Transmitter (ECG/US, TOCO/IUP) and the mode
encoding bits are delivered to the FSK signal generator. If an invalid
transducer is detected, these bits are set to an error code and the Transmitter
frontends are disabled.
Toco Driver
This circuitry drives a connected Toco or IUP transducer with a 1.6 kHz, 5V
peak square wave. The 1.6 kHz frequency is synchronised with the 3.2 kHz
repetition rate of the doppler ultrasound circuitries to avoid interference
between the Toco and ultrasound channels.
Toco Receiver
This module contains a differential intput amplifier, a synchronous
demodulator and a low pass filter, to derive a pressure proportional signal
from the TOCO/IUP transducers bridge.
Toco A/D
The TOCO/IUP pressure signal is A to D converted by a sigma-delta type
converter with a resolution of 12 bits and a 2Hz low pass characteristic. The
repetition rate is 4/sec. These 12 bits are encoded in the serial bitstream by
the FSK Generator module.
To ensure a high accuracy of the Toco pressure signal, correction factors are
stored in a EEPROM during the production test individually for every
Transmitter. Before encoding the Toco pressure values into the serial
bitstream, the pressure values are corrected with these factors. This
mechanism allows the usage of circuitries in the Toco frontends which only
have stability over temperature and aging but not an absolute accuracy.
FSK Generator
The Frequency Shift Key generator (FSK), is a sine wave generator (smoothed
square wave by filters) capable of generating either a frequency of 1.6 kHz
(represents a logic zero) or 2.4 kHz (represents a logic one). It encodes the
following signals in a bitstream:
• Coding pattern identification
• Transmitters Serial Number
• 12 Bit Toco pressure value
Oscillator
This circuitry produces the different clocks needed in the Transmitter. The
oscillator is crystal controlled and starts operation at power supply voltages as
low as 2V. This is necessary because the oscillator also drives the power
supply. To ensure a start up over the complete specified battery voltage range,
the oscillator must start at this low voltage. The clock freqencies are:
• 2 MHz for the mircocontroller
• 1 MHz for the ultrasound drive/receive clocks and the power supply.
• 500 kHz for the Toco sigma-delta A/D converter.
• 250 kHz for the ECG driver and the Toco sigma-delta A/D converter.
Power Supply
The power supply is a switched mode power supply with step up operation.
It works with a 250 kHz clock frequency (delivered by the oscillator) in a
pulse width modulation mode. It is capable of delivering the necessary power
in a fixed 5V output voltage for input voltages between 2.5V and 4.7V. (This
covers the input voltages delivered by three AA type alkaline batteries or three
AA type NiCd/NiHd accumulators).
Other supply voltages are derived from the 5V supply by linear regulators or
by charge pumps runing at 250 kHz.
• 2.5V as virtual ground (linear regulator)
• 8.5V for ultrasound preamplifier and OP-Amps (charge pump as
voltage doubler.)
• -3.5V for OP-Amps (charge pumps as voltage inverter).
Service Interface
The service and production interface uses the marker input as the I/O
connector for settings and internal status readings for service and production
tests.
It is a RS232 similiar (0.5V only levels) serial interface with a specific
protocol. This interface is used for:
• setting the Transmitter's serial number
• reading some internal values for prodution tests
• writing correction factors for the Toco frontend into the EEPROM.
EEPROM
The EEPROM holds the following settings:
• Serial number
• Gain calibration factors
• Country code
• Japanese ID code
Japan ID
This module is only acitve for Japanese options (set by the country code). It
sends the required ID code for Japanese Telemetry Transmitters. If there is
no valid ID code (or country code) in the EEPROM, this module shuts off
the VCXO. RF emission is allowed only if a correct ID code is emitted first
after power up.
VCXO
The Voltage Controlled Xtal (crystal) Oscillator is a "MAGIC" M1400-
6XXXX type VCXO. It generates the 4mW, 2mW or 1mW RF carrier in the
406 to 470 MHz band. This is dependant on county options:
• 4 mW for USA and Australia
• 2 mW for Europe
• 1 mw for Japan
It also performs the FM modulation for this carrier with the composite signal
of the ultasound doppler or the fetal scalp ECG and the FSK subcarrier for
the digital information. The heart rate signals use 60% and the FSK
subcarrier uses 40%of the available RF peak deviation. An additional lowpass
filter at the VCXO's output reducers the RF carrier harmonics.
The outer shield of the heart rate transducer cables is used as an antenna. An
impedence matching network adapts the antenna impedance to the VCXO's
50 Ohms output.
Input Filter
This module builds up a highpass filter with <1dB loss in the useful
frequency range 406 to 470 MHz and >20dB rejection for 100 MHz signals.
It prevents overloading of the RF preamplifier with strong lower frequency
signals.
The filter also provides the preamplifier with a 21V DC power by coupling
this DC voltage into the hot wire of the preamplifiers input cable. A
capacitor decouples the DC power from the antenna input.
RF Preamp M1406A
This MAGIC line amplifier boosts the input signal by 12 dB. This results in
an input sensitivity of about -118 dBm together with the M1402A RF
Receiver. The amplifier is powered from the DC power at the hot wire of the
input coax cable.
RF Receiver M1402A
The MAGIC receiver module M1402A performs the complete baseband
signal recovery. The input signal is prefiltered, mixed down and filtered by a
double superhetrodyne circuitry and demodulated to the baseband signal.
An on board microcontroller does the complete automatic frequency control
tracking of the input signal.
US Bandpass
This module performs a bandfiltering for the ultrasound doppler signal. It
consists of a 60 Hz highpass to remove the low frequency noise of the FM
demodulator, a 450 Hz lowpass to remove the higher frequency noise, and
the FSK subcarrier. The subcarrier attenuation must be so high that it is not
audible in the ultrasound signal (>70 dB at 1.6 kHz).
ECG Bandpass
This module performs a bandfiltering for the fetal scalp ECG signal with a 15
Hz highpass, to remove the low frequency noise (low frequency phase noise of
the RF oscillators) and a 80 Hz lowpass, to remove the high frequency noise
and the FSK subcarrier.
The output of the US or ECG bandpass is selected by the decoded mode
information by the FSK Demodulator and fed to the receiver output as
heartrate output. If the input signal quality is too bad (bit error rate increase
over a certain limit), the heartrate and Toco output is switched off.
FSK Bandpass
This bandpass selects the FSK subcarrier out of the composite signal. It
consists of a 1.6 KHz highpass to remove the low frequency noise and the
heart rate signals, and a 2.4 kHz lowpass to eleminate the high frequency
noise.
FSK Demodulator
The FSK Demodulator consists of:
• FSK Comperator
• FSK Demodulator
The Comperator converts the sine shaped FSK subcarrier signal into a square
wave. The Comperator has a hysteresis to improve the performance of noisy
input signals. The Demodulator demodulates the FSK subcarrier and
recovers the encoded bitstream. The synchronisation is done by searching the
encoded 12 bit serial number. The bitstream is also descrambled and the
CRC is checked. If the test is positive, the complete frame is saved and the
information bits are recovered and appropriate actions are taken:
• Set the Mode outputs
• Display the status values on the front panel.
• Switch On/Off the nurse call beeper
• Switch On all output signals
• Set the fetal movement output
EEPROM
The EEPROM keeps the following settings:
• Serial Number
• Nurse Call volume control
These values all can be set via the Service/Production interface, or in the case
of the nurse call volume control, by pressing the nurse call acknowledge
button if no nurse call is active.
Service Interface
This is a RS232 type serial interface used for servicing and production test
and settings, to read back internal values or to store serial numbers and other
settings in the EEPROM.
Power Supply
This module delivers all the necessary supply voltages. It has an input selector
for 110V or 230V mains input. It outputs the following voltages:
+21V RF Preamplifier
+12V Analog circuitries and the M1402A Receiver Module
-12V Analog circuitries and the M1402A Receiver Module
+5V Digital circuitries and the M1402A Receiver Module
Oscillator
The oscillator produces a CMOS clock signal for the digital circuitries. It is
running at 16.256 MHz. This frequency is divided by 2; the resulting 8.128
MHz frequency is used as a clock signal for the microcontroller.
Preventive Maintenance
Every 12 months, you must carry out a series of preventive maintenance tasks
and performance assurance tests. These ensure that the Telemetry System
continues to perform at its best, and reduces the possibility of failures. The
tasks to be carried out, their sequence, and the estimated time to complete
each one is given in the following table.
1. Mechanical Inspections
2. Carry out Safety Tests
3. Performance Assurance Tests
Mechanical Inspection
To carry out a mechanical inspection of the Receiver:
• Make sure all exposed screws are tight.
• Make sure the boards and connectors are firmly in place.
• Check the external cables for splits, cracks or signs of twisting. Replace
any cables that show serious damage.
Safety Tests
Complete the safety tests in accordance with national requirements.
For care and cleaning instructions, please refer to the Instructions for Use.
The Receiver VCXO operation frequency equals the channel frequency plus
21.4 MHz.
Austria
Australia
Belgium
Canada
Denmark
Finland
France
Germany
Italy
Japan
Luxembourg
Netherlands
New Zealand
Norway
Service USA
Singapore
Spain
Sweden
Switzerland
UK
USA
Index
Receiver functional blocks transmitter, 12 Dynamic Compression
ECG Bandpass, 74 service interface, 76 Circuitry, 70
EEPROM, 74 service program, 33 ECG Bandpass, 75
FSK Bandpass, 74 cable assembly, 33 ECG Driver, 70
FSK Demodulator, 74 check receiver, 36 ECG Receiver, 70
Input Filter, 74 check transmitter, 36 EEPROM transmitter, 73
Oscillator, 74 exit, 37 EEPROM, Receiver, 76
power supply, 74 power on self test, 36 FMP Detector, 70
RF Preamp M1406A, 74 read country information, 37 FSK Bandpass, 75
RF Receiver M1402A, 74 read serial number/ FSK Demodulator, 75
service interface, 74 revisions, 36 FSK Generator, 71
Toco D/A Converter, 74 reset serial number, 37 Input filter, 74
US Bandpass, 74 serial number to receiver, 36 Japan ID, 73
Receiver VCXO Test, 67 serial number to Modes, 71
remote event marker transmitter, 36 Oscillator Receiver functional
cables, 10 show last error/warning, 36 blocks, 76
specifications, 10 using, 34 Oscillator transmitter, 72
replacement and removal, 39 specifications power supply transmitter, 72
display board, 57 antenna, 6 power supply, Receiver, 76
fuses, 58 channels & frequencies, 6 RF Preamp M1406A, 74
power supply board, 53 DECG transducers, 9 RF Receiver M1402A, 75
processor board, 54 environment, 5 service interface, 76
Receiver VCXO, 56 inputs, 6 service interface
RF amplifier, 57 IUP transducer, 10 transmitter, 72
RF Module, 55 power requirements, 5 Toco A/D, 71
Transmitter, 46 remote event marker, 10 Toco D/A Converter, 76
Transmitter processor Telemetry, 5 Toco Driver, 71
board, 47 Toco transducer, 8 Toco Receiver, 71
Transmitter VCXO, 48, 50 storage Transmitter, 69
RF temperatures, 2 US Bandpass, 75
amplifier replacement and US Driver, 69
removal, 57 US Gating, 69
bench repair, 59, 69, 77, 79 T US Receiver, 69
bench test, 59 testing, 7 VCXO, Transmitter, 73
Module replacement and parameter test, 13 Toco
removal, 55 Receiver assembly, 65 A/D, 71
Preamp M1406A, 74 Receiver preamplifier, 64 D/A Converter, 76
Receiver M1402A, 75 receiver self test, 11 Driver, 71
Receiver VCXO, 66 Receiver, 71
RF bench test, 59 transducer, 8
S transducers, 8
self test, 7
safety, 1 transmitter self test, 12 cables, 8
self test, 7 Transmitter VCXO, 60 DECG, 9
receiver, 11 theory of operation, 69 IUP, 10
Toco, 8
Index
ultrasound, 8 Toco Receiver, 69 transmission range, 23
watertight blue, 2 US Driver, 69 yellow LED stays on, 21
Transmitter US Gating, 69
replacement and removal, 50 US Receiver, 69
transmitter VCXO, 69 U
testing, 12 Transmitter processor board ultrasound transducer, 8
Transmitter functional replacement and removal, 47 US Bandpass, 75
blocks Transmitter VCXO US Gating, 69
Dynamic Compression bench test, 60 US Receiver, 69
Circuitry, 69 replacement and removal, 50
ECG Driver, 69 troubleshooting, 17
ECG Receiver, 69 general problems, 17 V
EEPROM, 69 troubleshooting flowchart
FMP Detector, 69 cardio channel, 25 VCXO
FSK Generator, 69 event marker, 29 Receiver bench test, 67
Japan ID, 69 FMP, 27 Transmitter, 73
Modes, 69 no LED’s lit on receiver, 19 Transmitter bench test, 60
Oscillator, 69 not all LED's lit, 21 voltage
power supply, 69 nurse call, 30 fuse values, 58
service interface, 69 Toco channel, 26 output power, 7
Toco A/D, 69 transducer not detected, 24 power requirements, 5
Toco Driver, 69 transmission INOP range, 22 power supply board, 53
Index