Professional Documents
Culture Documents
C/CD/SC/EX Series
Date:
Institution:
Address:
Varian Representative:
Customer Representative:
Machine Configuration Tables (Enter N/A in any boxes that do not apply)
Software Versions
Clinac MLC
Console W/S
Comm Controller
Controller Shaper
Confirm the following energy configurations and purchased options with the Customer
Energies # Energies #
Electron energy # 3
OTHER OPTIONS
TABLE OF CONTENTS
1. RADIATION SURVEY.................................................................................................................... 3
1.1. COLLIMATOR TRANSMISSION ..................................................................................................................... 3
1.2. X-RAY LEAKAGE....................................................................................................................................... 3
2. MECHANICAL TESTS................................................................................................................... 3
2.1. MECHANICAL ISOCENTER VARIATION WITH ROTATION .................................................................................. 3
2.2. FIELD LIGHT & CROSSHAIR ALIGNMENT ...................................................................................................... 3
2.3. JAW POSITION READOUTS ........................................................................................................................ 3
2.4. GANTRY ROTATION READOUT CALIBRATION ............................................................................................... 3
2.5. COLLIMATOR ROTATION READOUT CALIBRATION ......................................................................................... 3
2.6. COUCH MECHANICAL MOTIONS ................................................................................................................. 3
2.7. OPTICAL DISTANCE INDICATOR (ODI) CALIBRATION .................................................................................... 3
3. RADIATION ISOCENTER TEST ................................................................................................... 3
3.1. GANTRY ROTATION SPOKE SHOT .............................................................................................................. 3
3.2. COINCIDENCE OF LIGHT FIELD AND X-RAY FIELD ......................................................................................... 3
4. BEAM PERFORMANCE................................................................................................................ 3
4.1. PHOTON ENERGY DEPTH OF IONIZATION .................................................................................................... 3
4.2. ELECTRON DEPTH OF IONIZATION .............................................................................................................. 3
4.3. PHOTON FIELD FLATNESS AND SYMMETRY ................................................................................................. 3
4.4. ELECTRON FIELD FLATNESS AND SYMMETRY .............................................................................................. 3
4.5. PHOTON AND ELECTRON SYMMETRY INTERLOCKS ....................................................................................... 3
4.6. BEAMSTOPPER ATTENUATION ................................................................................................................... 3
5. DOSIMETRY .................................................................................................................................. 3
5.1. SHORT TERM DOSE REPRODUCIBILITY ....................................................................................................... 3
5.2. DOSE REPRODUCIBILITY WITH MU SETTINGS.............................................................................................. 3
5.3. DOSE REPRODUCIBILITY WITH DOSE RATE (REP RATE) ............................................................................... 3
5.4. DOSE REPRODUCIBILITY WITH GANTRY ANGLE ........................................................................................... 3
6. DYNAMIC THERAPY..................................................................................................................... 3
6.1. ARC DYNAMIC.......................................................................................................................................... 3
6.2. ENHANCED DYNAMIC WEDGE .................................................................................................................... 3
7. LASERGUARD COLLISION PROTECTION SYSTEM................................................................. 3
7.1. PROTECTION ZONE AREA VERIFICATION ..................................................................................................... 3
7.2. PROTECTION ZONE TILT VERIFICATION....................................................................................................... 3
7.3. MOTION STOP FUNCTION VERIFICATION ..................................................................................................... 3
7.4. POWER KEY SWITCH & OVERRIDE FUNCTION VERIFICATION ........................................................................ 3
8. CLINAC OPERATIONAL TRAINING ............................................................................................ 3
REVISION HISTORY
Added EXd model Clinac & IEC 1217 scales to all applicable tables.
O 03/25/04 Removed X-ray contamination test for electron depth of Ionization scans. JT
Various other minor wording changes throughout document
PURPOSE
The main objectives of the Customer Acceptance Procedure (CAP) are to:
• Provide Varian the opportunity to demonstrate that the Clinac was designed,
delivered, and operates according to specifications listed in this document.1
• Familiarize the customer with the performance, daily checkout, and general
maintenance of the Clinac.
SCOPE
The Clinac System Verification documents are developed and approved by Product Support
Engineering (PSE) based on inputs from the Engineering System test procedure and
summary, RAD specifications documents, and Customer Support Installation procedures.
GENERAL PROCEDURE
The CAP consists of a series of specifications and testing procedures. The tests can be
performed either in sequential order, all at once, or in an alternating sequence during logical
periods of the machine installation. The CAP is a subset of the System Verification
Procedure and Summary.
The System Verification Procedure and Summary (SVP & SVS) are accomplished and
certified by the Varian Customer Support Installation Representative during the installation
of the product. Verified data from the SVS may be directly transferred into the CAP at the
customer’s discretion.
The Dosimetry portion of the System Verification Summary should be accomplished during
the CAP. See the Note in the Dosimetry section regarding available worksheet options.
All positions are listed in IEC scale with Varian (VAR) scale in parentheses. When IEC-601
differs from IEC-1217, then they will be called out separately.
All data entry boxes must be filled in. If any boxes do not apply, enter NA.
Many measurements simply require a checkmark to indicate the test has met the required
specification. It is not required to measure and record a numerical value for these tests.
1
Machines having special options and/or specifications may require additional testing.
DEFINITIONS
Inplane: the vertical plane defined by the accelerator gun and target; equivalent to Radial.
Crossplane: the vertical plane that is at right angles to the Inplane surface; equivalent to
Transverse.
Radial: the vertical plane defined by the accelerator gun and target; equivalent to In-Plane.
Transverse: the vertical plane that is at right angles to the Radial surface; equivalent to
Cross-Plane.
REQUIRED EQUIPMENT
Varian Supplied
Hospital Supplied
2
Varian can provide a Wellhofer Buddleship scanning system for USA customers. Customers outside of the USA are
required to provide a scanning system for acceptance testing. However all specifications and procedures are written
for the Wellhofer Scanning System using the Varian protocol. If another scanning system is used and a conflict in the
measurements occurs, the tests should be verified using a Wellhofer Scanning System.
1. RADIATION SURVEY
As soon as the Clinac is able to produce a radiation beam, a qualified radiation expert
provided by the customer shall conduct a survey. In addition to direct beam testing, a
scattering material such as water or solid phantom must be placed in the radiation beam to
perform secondary barrier testing. To obtain worst-case scenario, the following parameters
will be used as minimum conditions:
• The Clinac must be operating at the maximum dose rate
• The Gantry must be tested at selected angular positions with the beam on
• The Collimators on the treatment head must be set for the maximum field size
If dose rates in areas external to the treatment room exceed radiation levels recommended
by the governing agency, machine beam operation is to be immediately suspended until
either the equipment, or the facility is modified or a temporary deviation issued. A
temporary deviation may consist of operating at a limited dose rate, operating with the
Gantry restricted to certain positions, or a combination of both as determined by the
qualified radiation expert. This survey is a preliminary check used to determine temporary
safe installation environment. It is not to be used as the data gathering survey that will be
performed by the customer after installation.
Specification
• Transmission through the moveable Collimators shall not exceed 0.5% of the
central beam intensity. Collimator Transmission tests are performed during
manufacturing testing of the system. Results of this testing are available
from Varian upon request.
Specification
• The average X-ray intensity measured over an area of 100 cm2 at a distance
of 1 meter from the primary beam shall not exceed 0.1% of the intensity at
isocenter. Data is on file at Varian that represents a typical X-ray intensity for
the High Energy Clinac system. This data is available from Varian upon
request.
2. MECHANICAL TESTS
Specification
Test Setup
1. Level the Gantry at 0° (180° VAR). Rotate level end-over-end to
check level accuracy.
2. Install a calibrated 100 cm front pointer.
3. Extend the tip of another front pointer over the front edge of the
Couch and rotate the Gantry between 90° to 270° to accurately set
the distance of the Collimator front pointer to precisely 100 cm TSD.
4. Remove the Couch top front pointer.
5. Tape a piece of 1mm ruled graph paper to the top of the Couch.
6. Position Couch top until the graph paper is just below the front pointer
(but not touching).
7. With the graph paper aligned under the front pointer tip, rotate the
Collimator from 90° to 270°, while observing the pointer run-out.
8. If the walkout is >0.25 mm, shim or adjust the front pointer as
necessary to minimize run-out.
Method of Test
1. With the graph paper aligned under the front pointer tip, rotate the
Collimator from 90° to 270°, while observing the pointer run-out.
Verify that the worst case run-out meets specification.
2. Fill out the applicable box in the Mechanical Isocenter Measurements
table.
Specification
Test Setup
1. The front pointer, Gantry and Couch should be set-up per the
previous step with the collimator at 0° (180° VAR).
2. Tape the short front pointer (or a 2-mm drill bit) to the end of the
Couch so that the 2-mm tip is extended over the end of the Couch
top.
3. With the Couch at 0° (180° VAR), position the Couch vertical axis so
that the tip of the Collimator front pointer is aligned to the center of the
short front pointer tip.
4. Position the Couch longitudinal axis so that the tip of the short front
pointer is approximately 1 mm away from the Collimator front pointer.
5. Position the Couch lateral axis so the tip of the short front pointer is
centered on (but not touching) the tip of the Collimator front pointer.
Method of Test
1. Rotate the Gantry through a full 360° while visually checking the front
pointer run-out every 90°.
2. Verify the front pointer tip run-out is ≤1.0 mm radius throughout the
entire 360° of rotation for counterweight systems (or ≤2.0 mm for
Beamstopper systems)
3. For systems with a retractable Beamstopper, perform this test with the
Beamstopper retracted, and again with it extended.
4. Fill out the applicable boxes in the Mechanical Isocenter
Measurements table. Enter NA in any boxes that do not apply.
Specification
Test Setup
1. The front pointer, Gantry and Couch should be set-up per the
previous section with the collimator at 0° (180° VAR).
2. Rotate the Couch to 90° and align the graph paper under the front
pointer.
Method of Test
1. Slowly rotate the Couch from 90° to 270° while observing the front
pointer run-out every 45°.
2. Verify the worst case run-out is ≤1.0 mm radius for counterweight
systems and ≤2.0 mm for Beamstopper systems.
3. Repeat test with Beamstopper extended for systems with a
Beamstopper.
4. Fill out the applicable boxes in the Mechanical Isocenter
Measurements table. Enter NA in any boxes that do not apply.
Initial Date
Specification
• The edges of the light field at 100 cm TSD shall be parallel and
orthogonal within ±1° (1° equals 5.2 mm of linear error on a 30 cm
field).
• The Field Light source shall intersect mechanical isocenter within ≤1.0
mm at 100 cm TSD.
Test Setup
1. Verify Gantry is at 0° (180° VAR).
2. Open the jaws to 35×35 cm.
Method of Test
Since the jaws are not mechanically linked, the jaw edges
cannot be used to verify Field Light source alignment.
Instead, an independent shadow source must be used that
will not rotate with the Collimator.
Specification
Test Setup
1. Verify Gantry is at 0° (180° VAR) and open the jaws to 35×35 cm.
Method of Test
1. Set the Couch top to 100 cm TSD and tape a piece of graph paper
ruled in millimeters to the Couch top.
2. With the crosshairs aligned on the graph paper, rotate the Collimator
from 90° to 270° and verify crosshair run-out is ≤1.0 mm at isocenter.
3. Verify each crosshair is parallel to the upper and lower jaws. Perform
this test as follows:
Initial Date
Specification
Test Setup
1. Level the Gantry at 0° (180° VAR).
2. Using a calibrated front pointer, set the Couch top to 100 cm TSD.
3. Tape a piece of accurately ruled millimeter graph paper to the Couch
top (or a precision ruler may be used)
4. Turn on the Field Light and align the graph paper (or ruler) to the
Clinac crosshairs.
Method of Test
1. Independently drive each jaw so that the 50% isodensity point of the
projected jaw shadow corresponds to the jaw positions listed in the
following table. Verify the digital readout displays meet specification.
2. Fill out the applicable boxes in the Independent Jaw Digital Readouts
table.
Y1 -8 cm ±2 mm
Y1 20 cm ±2 mm
Y2 20 cm ±2 mm
Y2 -8 cm ±2 mm
X1 -1 cm ±2 mm
X1 20 cm ±2 mm
X2 20cm ±2 mm
X2 -1 cm ±2 mm
Initial Date
Specification
• The projected light field size measured at the 50% isodensity line for
each jaw shall coincide with the symmetric digital position display to
an accuracy of ±2 mm at 100 cm TSD.
Test Setup
1. Use the previous test setup but select the Symmetric jaw mode.
Method of Test
1. Symmetrically drive the X & Y jaws so that the 50% isodensity point of
the projected jaw shadows correspond to the field size positions listed
in the following table. Verify the digital readout displays meet
specification.
2. Fill out the applicable boxes in the Symmetric Jaw Digital Readouts
table.
√ = Pass
Field Size Spec
X Jaws Y Jaws
6×6 cm ±2 mm
30×30 cm ±2 mm
Initial Date
Specification
• The true angular position of the Gantry shall coincide with the Gantry digital
display to an accuracy of ±0.5° and the Gantry mechanical scale to an
accuracy of ±1.0°.
Method of Test
1. Using a precision level placed on a true surface of the Interface Mount
(which should match the Collimator baseplate), level the Gantry at each
position shown in the following table. Verify the Gantry digital and
mechanical displays meet specification.
2. Fill out the applicable boxes in the Gantry Angle Digital & Mechanical
Readouts table.
Initial Date
Specification
• The true angular position of the Collimator shall coincide with the digital
display to an accuracy of ±0.5° and the mechanical scale on the Collimator
assembly to an accuracy of ±1°.
Test Setup
The following steps provide one method of verifying the true angular
position of the Collimator. Other methods can be used as long as
the same end result is achieved.
Method of Test
1. Without moving the Collimator, Verify the Collimator digital and mechanical
readout displays meet specification for the 0° (180° VAR) position and
check off the applicable box in the following table.
2. Close the jaws to 20×20 cm and align the radial (inplane) crosshair line to a
sheet of graph paper placed on the Couch top at 100 cm TSD.
3. Rotate the Collimator to 90° until the crosshair line is again aligned with the
graph paper. Projected crosshair lines may not be perfectly perpendicular
so use the same crosshair line during these tests. Verify the digital and
mechanical readout displays meet specification.
4. Repeat the test with the Collimator rotated to 270º.
5. Fill out the applicable boxes in the Collimator Angle Digital & Mechanical
Readouts table.
6. Re-position the Collimator to the true 0° (180° VAR) position for the next
test.
Initial Date
Specification
Test Setup
1. Use the previous test setup to position the Couch angle to the true 0°
(180° VAR) position.
Method of Test
1. With the Couch angle at 0° (180° VAR), Verify the digital and
mechanical readout displays meet specification in the following table.
2. Rotate the Couch to both 90° and 270° until the crosshair line is again
aligned with the graph paper. Verify the digital and mechanical
readout displays meet specification.
3. Fill out the applicable boxes in the Couch Angle Digital & Mechanical
Readouts table.
Initial Date
Specification
• The Couch longitudinal travel shall coincide with the digital display to
an accuracy of ±1 mm.
Test Setup
1. Replace the wooden service panel with the carbon fiber panel.
2. Install the Varian provided tape measure into the PRO Alignment Tool
so that the 140 cm mark on the tape measure lines up with the sight
window scribe marks.
3. Install the Longitudinal PRO Alignment Tool onto the Couch carbon
fiber top at the 0 position index. The end of the tape measure should
be extended toward the Gantry PRO display.
4. Position the Couch to 0° (180° VAR) and 100 cm TSD.
Method of Test
1. Gently support the end of the tape measure to keep it level and float
the couch top until the crosshair aligns with the 20 cm mark. Verify
the digital display meets specification in the following table. Note: this
will be the 0cm position for scale convention IEC 1217).
2. Float the couch top until the crosshair aligns with the 150 cm mark on
the tape (press down on the tape to keep it flat and bend the end of
the tape down to avoid hitting the PRO display). Verify the digital
display meets specification.
3. Fill out the applicable boxes in the Couch Longitudinal Digital
Readouts table.
Longitudinal Position
Spec √ = Pass
Varian & IEC 601 IEC 1217
20 cm 0 cm ±1 mm
150 cm 130 cm ±1 mm
Initial Date
Specification
• The Couch lateral travel shall coincide with the digital display to an
accuracy of ±1 mm.
Test Setup
1. Using the same PRO Alignment tool setup from the previous test,
center the Couch top laterally by aligning the crosshair to the scribe
mark on the PRO alignment tool.
Method of Test
1. With the couch aligned to the center travel position, verify the digital
display meets specification for the central position in the following
table.
2. Move the tabletop 23 cm to the right by measuring the distance
between the crosshair and the PRO tool scribe mark. Verify the
digital display meets specification.
3. Repeat the test with the tabletop moved 23 cm to the left of center.
4. Fill out the applicable boxes in the Couch Lateral Digital Readouts
table.
Lateral Position
Spec √ = Pass
Varian IEC 601 IEC 1217
77 cm 977 cm -23 cm ±1 mm
100 cm 0 cm 0 cm ±1 mm
123 cm 23 cm +23 cm ±1 mm
Initial Date
Specification
• The Couch vertical travel shall coincide with the digital display to
within ±1 mm.
Test Setup
1. Position the Gantry to 0° (180° VAR).
2. Using a calibrated front pointer, position the tabletop to isocenter with
the carbon fiber panel installed (do not use the wood Service panel).
Remove the front pointer after positioning.
3. Using the provided tape measure, extend and slide the tape measure
between the Couch top and the side support rail (at any location on
either side of the Couch). Tape the end of the tape measure to the
bottom edge of the Couch. End result:
• The tape measure dispenser should be sitting on top of the side
Couch rail
• The tape measure should be extended downward between the
Couch top and the side rail
• The end of the tape measure should be hooked under the
bottom of the Couch and securely taped to the base of the
Couch.
4. Record the tape measurement for the “top edge” of the Couch side
rail. This will be the measurement reference value for 100 cm TSD.
Method of Test
1. With the couch in the test setup position, verify the digital display
meets specification for the 0 cm (100 cm VAR) position in the
following table.
2. Rotate the Gantry 50° in either direction to allow full couch extension.
Vertical Position
Spec √ = Pass
Varian IEC 601 IEC 1217
65 cm 965 cm +35 cm ±1 mm
100 cm 0 cm 0 cm ±1 mm
160 cm 60 cm -60 cm ±1 mm
Initial Date
Specification
• The ODI rangefinder shall indicate the Target to Surface Distance (TSD) to
an accuracy of ±1 mm at 100 cm TSD and to an accuracy of ±5 mm for all
other distances.
Test Setup
1. Use the same test setup from the previous test.
Method of Test
1. Turn on the Field Light and project the crosshairs and ODI rangefinder
display on a piece of white paper. Verify the ODI meets specification for the
100cm position in the following table.
2. Using the same tape measuring technique in the previous test, sequentially
drive the Couch 20 cm above and then 30cm below the 100 cm reference
position. Verify the ODI display meets specification. It will be necessary to
select “Raw Scale” readouts in Service Mode in order to raise the Couch
above isocenter with the Gantry at 0°(180° VAR).
3. Fill out the applicable boxes in the OBI Optical Display table.
80 cm ±5 mm
100 cm ±1 mm
130 cm ±5 mm
Initial Date
Specification
• Central axis X-ray beam variation due to Gantry rotation shall be confined to
a sphere of ≤1.0 mm radius for counterweight systems and ≤2.0 mm radius
for Beamstopper systems.
Test Setup
1. Set the Collimator angle to 0° (180° VAR).
2. Open the upper jaws to 40 cm.
3. Using the Independent jaw mode, close both lower jaws to 0.5 cm so that a
symmetrical 1.0 cm field is projected relative to the crosshair. The lower
jaws must be positioned equidistant to an accurately aligned crosshair for
this test to be valid.
4. Using support blocks (such as solid water slices), position an X-ray film so
that it is standing vertically on the Couch top and perpendicular to the lower
jaws. If using solid water slices, ensure that the slices will not fall!
5. Position the Couch height so that the center of the film is near 100 cm TSD.
Position the Couch lateral so that the approximate center of the film is
aligned to the Clinac crosshair.
6. Rotate the Gantry throughout its full rotational range and verify there will be
no collisions with the couch setup.
Method of Test
1. Rotate Gantry successively from 90°, 0° (180° VAR), 275° (265° VAR), and
185° (355° VAR) while exposing the film at each position. Develop the film.
2. Bisect the 1.0 cm exposures in the center with a sharp line.
3. At the intersection of lines, measure the longest line of the trapezoid. This
represents the diameter error. Refer to Figure 1.
4. For retractable Beamstopper systems, repeat the test with the Beamstopper
extended.
5. Fill out the applicable boxes in the Gantry Radiation Spoke Shot table.
Enter NA in any boxes that do not apply.
Initial Date
VAR/IEC
180°/0°
Area of
VAR/IEC interest
265°/275° VAR/IEC
90°/90°
VAR/IEC
355°/185°
ac_svp1.vsd, 3/15/01
Specification
• The field defining light shall coincide with the 50% isodensity line on an X-ray
film at 100 cm TSD within the specifications listed in the following table.
Test Setup
1. Machine positioning:
o Gantry 0° (180° VAR) and level
o Collimator 0° (180° VAR)
o Field size 30×30 cm (or 20×20 cm per Note above)
o Couch vertical 100 cm TSD
Method of Test
1. Tape a piece of X-ray film to the top of the Couch.
2. Turn on the field light and mark the edges of the field on the film package at
the 50% density region using a small pin or a ballpoint pen. Leave gaps if a
densitometer is going to be used to find the 50% point.
3. Expose the film in Lo-X.
4. Repeat for Hi-X.
5. Develop the films and compare the 50% isodensity lines of the X-ray field
edges to the field light edges.
6. Verify that field light and X-ray field edges are coincident within
specification.
7. Fill out the applicable boxes in the Light Field & X-ray Field Coincidence
table.
Spec
Energy Field Size √ = Pass
C/CD/SC/EXd EX
Initial Date
4. BEAM PERFORMANCE
In preparation for beam scanning, set up the Wellhofer scanning system as shown.
Crossplane
Inplane
Pendant
Motor control cable
Triax Cables
Wellhofer Controller
Plotter (COM2,Serial B
Printer LPT1, PRT/FDD
Laptop Computer
ap_svp2.vsd, 2/25/96
Specification
• The maximum buildup and intensity at 10 cm depth shall meet the
specification shown in the Photon Depth of Ionization Table.
• All Photon Depth of Ionization tests are specified using a water phantom at
100cm TSD and a 10×10 cm field size.
Test Setup
1. Set up and level the water phantom for 100 cm TSD. Refer to the Wellhofer
Operation Manual for proper setup.
2. Setup the Wellhofer for Depth Dose scanning
3. Set the jaws to 10×10 cm.
4. Turn the PFN Servo ON and the Dose Rate Servo OFF.
Method of Test
1. Beam on in Lo-X, Rep Rate 3 and scan the central axis depth of ionization.
2. Repeat the scan for Hi-X.
3. Analyze the scans and verify each energy meets the specifications in the
following table. Record the data in the following table and in the SVS (if not
already completed). Enter NA in any boxes that do not apply.
Initial Date
Specification
• The depth of ionization at 90%, 80%, 50%, and 30% of the maximum buildup
intensity shall be per the specifications shown in the Electron Depth of
Ionization Table for all electron energies.
• All electron Depth of Ionization tests are specified using a water phantom at
100cm TSD and a 15×15 cm field size.
Test Setup
1. Level the Gantry at 0° (180° VAR).
2. Install a 15×15 cm electron applicator and make sure the FFDA insert is
fully seated.
3. Setup the Wellhofer for Depth Dose scanning.3
3
The depths in the Electron Depth of Ionization table are not depth dose values. To obtain percent depth dose
values, appropriate correction factors for probe displacement, etc, which are outlined in ICRU Report No. 35, must be
applied.
4. Turn the PFN Servo ON and the Dose Rate Servo OFF.
Method of Test
1. Using a slow to medium speed, run the central axis depth of ionization
scans for Rep Rate 3 in each electron energy.
2. Analyze each scan and verify the depth intensities meet the specifications
per the following table.
3. Record the data in the table provided and in the SVS (if not already
completed). Enter NA in the boxes that do not apply.
4e- 0.6cm 0.89 ±0.1cm 1.00cm ±0.1cm ±0.07cm 1.26 ±0.1cm ≤2.00cm
6e- 1.0cm 1.71 ±0.1cm 1.90cm ±0.1cm ±0.07cm 2.30 ±0.1cm ≤2.60cm
9e- 1.4cm 2.68 ±0.1cm 2.95cm ±0.1cm ±0.07cm 3.50 ±0.1cm ≤3.90cm
12e- 2.0cm 3.77 ±0.1cm 4.15cm ±0.1cm ±0.07cm 4.89 ±0.1cm ≤5.40cm
15e- 2.6cm 4.68 ±0.1cm 5.20cm ±0.1cm ±0.07cm 6.17 ±0.1cm ≤6.80cm
16e- 2.7cm 4.87 ±0.1cm 5.45cm ±0.1cm ±0.07cm 6.49 ±0.1cm ≤7.30cm
18e- 3.0cm 5.31 ±0.1cm 6.10cm ±0.1cm ±0.07cm 7.41 ±0.1cm ≤8.15cm
20e- 3.3cm 5.52 ±0.1cm 6.55cm ±0.1cm ±0.07cm 8.13 ±0.1cm ≤9.30cm
22e- 3.4cm 5.59 ±0.1cm 6.80cm ±0.1cm ±0.07cm 8.64 ±0.1cm ≤10.00cm
4e-
6e-
9e-
12e-
15e-
16e-
18e-
20e-
22e-
Initial Date
Flatness Specification
• The arithmetic average between minimum and maximum X-ray beam
intensities within the central 80% of the radial and transverse major axes at
100 cm TSD and at a depth of 10 cm, shall not exceed the specifications
listed in the X-ray Beam Symmetry & Flatness tables.
Symmetry Specification
• The maximum variation in integrated dose between any two corresponding
points equidistant from the centerline within the central 80% of the radial and
transverse major axes at 100 cm TSD and a depth of 10 cm, shall not exceed
2.0%.
Test Setup
1. Level the Gantry at 0° (180° VAR).
2. Setup the Wellhofer scanning system for INPLANE scanning. (Refer to
Figure 2)
3. Level the tank in both planes.
4. Using the front pointer set the top surface at 100 cm TSD.
5. Open the jaws to 40×40 cm.
6. Set the probe position to the zero position (using the 0 hot key) and the
depth to 1 cm to allow visual probe alignment to the Clinac crosshairs (the
crosshair should bisect the probe tip).
7. The reference probe should be positioned as far as possible away from the
scanning path so that it is does not cause scatter that could affect the scan
results. (Refer to Figure 2 for a diagram of scanner set up.)
Method of Test
1. All scans should be saved with the following information: Energy, Scanning
Plane, Jaw Size and Date.
2. Scan through the X-ray field central axis at a depth of 10 cm in the radial
plane (Inplane) for each energy at field sizes of 40×40 cm and 10×10 cm.
Radial scans should always be performed prior to transverse (Crossplane)
scans in case any steering adjustments are required.
3. Analyze the scans using the definitions described in Figure 3 & Figure 4,
and record the data in the following table (and in the SVS if not already
completed).
4. Setup the Wellhofer scanning system for CROSSPLANE scanning and
repeat the scans in the transverse plane. For Buddleship tanks, rotate the
couch to 90º or 270º.
5. Analyze the scans and record the data in the following table (and in the SVS
if not already completed). Enter NA in any boxes that do not apply.
Flatness Spec
Actual Symmetry Actual
Energy Field Size C CD SC
EX Flatness4 Spec Symmetry
& EXd
±2.5%
Hi-X 40 x 40 Radial ±3% ≤2%
(±3% 20X)
±2.5%
Hi-X 40 x 40 Trans ±3% ≤2%
(±3% 20X)
Initial Date
The Wellhofer Buddleship and many other scanning systems analyze Flatness as a total value, not in a ± format (i.e.
4
6% = ±3%). Varian specifications are stated in the ± format, so divide the total Flatness value by 2 and enter a ±
value in the table.
50%
Central 80%
Field Width MAX Intensity
MIN Intensity
ap_svp3.vsd 4/24/96
Beam Centerline
Y
X 100% MAX
Intensity
Intensity
Point 1 Point 2
50%
50% Side A Side B Intensity
X/2
0 X
ap-svp3.vsd, 4/24/96
Flatness Specification
• The arithmetic average between minimum and maximum electron beam
intensities within the central 80% of the radial and transverse major axes at
100 cm TSD and at a depth of Dmax, shall not exceed the specifications
listed in the Electron Beam Symmetry & Flatness tables.
Symmetry Specification
• The maximum variation in integrated dose between any two corresponding
points equidistant from the centerline within the central 80% of the radial and
transverse major axes at 100 cm TSD and at a depth of Dmax, shall not
exceed 2.0%.
Energy Depth
4e- 0.6 cm
6e- 1.0 cm
9e- 1.4 cm
12e- 2.0 cm
15e- 2.6 cm
16e- 2.7 cm
18e- 3.0 cm
20e- 3.3 cm
22e- 3.4 cm
Test Setup
1. The test setup described in the Photon Field Flatness and Symmetry Test
should be used but the field sizes for electron scanning will be defined by
the 10×10 cm, 25×25 cm, or 36×36 cm electron applicators.
Method of Test
1. Attach a 25×25 cm applicator and make sure the FFDA insert is fully
seated.
Radial Plane Electron Beam Symmetry and Flatness (within 80% FWHM)
Flatness Spec
Actual Symmetry Actual
Energy Field Size C CD SC
EX Flatness5 Spec Symmetry
& EXd
±5% ±4.5%
E1 25 x 25 Radial ≤2%
(±7% 4e) (±7% 4e)
25 x 25 Radial
E6 ±5% ±4.5% ≤2%
(CL2300/23EX)
10 x 10 Radial
E6 ±5% ±4.5% ≤2%
(CL2300/23EX)
±5% ±5% 6e
E1 10 x 10 Radial ≤2%
(±7% 4e) (±7% 4e)
Initial Date
The Wellhofer Buddleship and many other scanning systems analyze Flatness as a total value, not in a ± format (i.e.
5
6% = ±3%). Varian specifications are stated in the ± format, so divide the total Flatness value by 2 and enter a ±
value in the table
Transverse Plane Electron Beam Symmetry and Flatness (within 80% FWHM)
Flatness Spec
Actual Symmetry Actual
Energy Field Size C CD SC
EX Flatness Spec Symmetry
& EXd
±5% ±4.5%
E1 25 x 25 Trans ≤2%
(±7% 4e) (±7% 4e)
25 x 25 Trans
E6 ±5% ±4.5% ≤2%
(CL2300/23EX)
10 x 10 Trans
E6 ±5% ±4.5% ≤2%
(CL2300/23EX)
±5% ±5% 6e
E1 10 x 10 Trans ≤2%
(±7% 4e) (±7% 4e)
Initial Date
Specification
• The asymmetry interlocks [EXQ1/EXQ2] shall terminate the beam at or
before field asymmetry exceeds 2.0%.
• Photon asymmetry is measured at the 25% and 75% points of the full width,
half maximum value for a 40 cm square field.
• Electron asymmetry is measured at the 10% and 90% points of the full width,
half maximum value for a 25 cm square field.
Test Setup
1. Select the Clinac Service Mode and override DOS1, DOS2, DS12, XDRS,
TIME, and EXQ1 & EXQ2 interlocks. Turn the Dose Rate Servo ON.
2. Open the radial and transverse Angle Steering Servo Loops (on B15 &
B16).
3. Verify the Position Servo Loops are closed.
4. Connect one end of a jumper to B19-TP8 (leave the other end
disconnected)
5. Select Lo-X. Measure and record the ANG R & ANG T programming values
in the following table. Repeat for Hi-e.
6. Open the Factory Red Book (Test Summary or STS section) to the test
titled “X-ray & e-Mode Symm Servo/Phase Margin Loop Gain Calibration.”
Record the Angle Servo values for Lo-X and Hi-e in the following table.
Method of Test
The following steps are written for the Radial symmetry tests. Refer
to the words in parenthesis ( ) when performing the Transverse
symmetry tests.
Always use the Setup – Set All menu selections when changing
energies.
1. Beam on in Low X-ray and adjust pot R54 on B19 so that the voltage at
B19-TP8 measures the same value as the Lo-X ANG R (ANG T)
programming pot voltage.
2. With the beam still on, connect the jumper wire between B19-TP8 and the
ANG R (ANG T) test point on the front of the Program PCB.
3. Adjust B19-R54 until the R SYM (T SYM) meter on the Console monitor
indicates < 1.0. If the EXQ1 (EXQ2) interlock is red, clear the interlock by
depressing the “I” & “D” keys.
4. Slowly adjust R54 in either direction until the EXQ1 (EXQ2) interlock trips.
Immediately note the R SYM (T SYM) meter display.
5. Multiply the trip value by the Lo-X Angle Servo calibration factor in the Red
Book to determine the amount of beam asymmetry when the interlock
occurred. Verify the calculated value meets specification.
For example: If the R SYM meter indicated 1.72 when the EXQ1 interlock
tripped, and the Lo-X Radial Angle Servo calibration value in the Red
Book is 0.98%, the actual beam asymmetry when the interlock occurred
would be: 1.72×0.98 = 1.69% (rounded off)
6. Repeat these tests for the Hi-e energy.
7. Fill out the applicable boxes in the Symmetry Interlocks table.
8. Remove the jumper to restore all original steering values. Clear all
interlocks, and close the Angle Steering Loops.
If time permits, complete these tests for all energies per the
requirements in the SVS, if not already completed.
Symmetry Interlocks
√ = Pass
Energy Field Size Spec
R Sym T Sym
Initial Date
Specification
Method of Test
1. Set the Gantry at 90° or 270°. Position the Beamstopper to the fully
extended position. Set the jaws to 10×10 cm.
2. Using an ionization chamber and buildup cap, measure the central
axis radiation intensity 15 to 20 cm beyond the Beamstopper in the
highest X-ray energy.
3. Without moving the test ionization chamber, repeat measurement with
the Beamstopper retracted. The ratio of measurements shall be
≥1000 to 1.
4. Fill out the applicable boxes in the Beamstopper Attenuation table.
Enter NA if there is no Beamstopper installed.
Beamstopper Attenuation
Initial Date
Method of Test
1. In Physics mode, enter the energies and angles that will cause a
STPR interlock to activate.
2. Save the changes and exit Physics mode.
3. Verify the STPR interlock is activated for the energies/angles requiring
the Beamstopper to be extended.
4. Checkmark the following box if acceptable or enter NA if not
applicable.
Beamstopper Interlock
Initial Date
5. DOSIMETRY
Dosimetry calibration was already performed in the factory. Depending on the protocol
used by the hospital, the absolute dose calibration should be relatively close. Therefore, it
is not necessary to calibrate the absolute dose during these tests. Instead, the dose
integration data shall represent relative data for the sole purpose of verifying the
specifications. The final absolute dosimetry calibration (which is ultimately the responsibility
of the hospital) should be performed by the customer after completion of the acceptance
testing.
There are three choices for recording the test data for beam dosimetry tests.
1. Lotus Notes SVS Worksheet
2. Excel Spreadsheet Worksheet (available on the PSE web site in
Downloads 2, titled HE-CAP-Dosimetry Spreadsheet.xls)
3. SVS Worksheet (paper copy)
Use the worksheet in the Lotus Notes SVS, if available. Integration data is
entered in a provided worksheet and all test results are then automatically
calculated and entered in the required tables. This is the most efficient
method. After completion of the SVS, print a copy for the customer to
provide a permanent record for the Dosimetry section test results.
Use the Excel spreadsheet if the Lotus Notes SVS is not available.
Integration data is entered in a provided worksheet and all test results are
then automatically calculated and entered in the required tables. After
completion, the spreadsheet must be printed out, signed and attached to the
SVS and CAP paper documents. Once attached, it is necessary to write
“See Attached Spreadsheet” diagonally across each of the applicable
tables in the SVS and CAP and initial or sign the box below each table.
Only use the SVS paper worksheet if the other two options above are not
available. This method is very time consuming and permits possible errors
due to potential human calculation errors.
Definitions
RRmax refers to the maximum Rep Rate available on the Clinac iX being accepted
RRmid refers to RR3 for standard Clinacs or RR2 for Clinacs with only three Rep Rates
LDR refers to the Low Dose Rate option
LRRx refers to the Low Dose Rate mode and the x represents the Rep Rate
Test Setup
1. Tests shall be made using a Farmer Secondary Standard, or PTW ion
chamber (or equivalent), located in a water or suitable solid phantom. The
output of the ion chamber should be measured using an electrometer.
2. Accumulated dose and intensity should be measured at central axis at
100 cm TSD using the following depths and field sizes for all tests. Actual
Dmax depths are not required since these dosimetry checks are relative
readings.
3. Ambient temperature variations should not exceed ±5°C during the course
of these tests.
4. The Gantry should be leveled at 0° (180° VAR) for all tests except as
indicated in the Dose Reproducibility with Gantry Angle tests.
5. All Clinac servos must be ON (Dose Rate Servo, Steering Servo Loops and
PFN Servo).
6. Field sizes will be as follows:
o 10×10 cm for all photon energies
o 15×15 cm applicator for all standard electron energies
o 36×36 cm HDTSe applicator for HDTSe energy
Method of Test
1. Refer to the worksheet(s) provided in the Lotus Notes SVS, Excel
spreadsheet or SVS paper document and run all of the required integrations
using the MU settings and Rep Rates shown.
2. Record the integrated charge readings displayed on the electrometer for
each run in the appropriate box in the worksheet(s).
Specification
• Short-term dose reproducibility for each energy shall be ± 1% or 1 monitor
unit, whichever is greater, at a fixed dose rate.
Method of Test
1. Using the worksheet data in the SVS and the equations shown in the
following table, calculate and record the test results (not required if using
the Lotus Notes SVS or Excel spreadsheet).
Short Term Dose Reproducibility [enter NA in any boxes that do not apply]
50 MU 100 MU 300 MU
(500 MU in HDTSe) (1000 MU in HDTSe) (3000 MU in HDTSe )
Hi-X ±1%
Lo-X ±1%
E1 ±1%
E2 ±1%
E3 ±1%
E4 ±1%
E5 ±1%
E6 ±1%
HDTSe- ±1%
Equation 1 (EQ I) = {[READING #1 - READING #2] ÷ READING #2} ×100
Equation 2 (EQ II) = {[READING #2 - READING #3] ÷ READING #2} ×100
Equation 3 (EQ III) = {[READING #1 - READING #3] ÷ READING #3} ×100
Initial Date
Specification
• Dose linearity shall be:
o ±1 % for accumulated doses between 20 and 999 monitor units
o ±2 % for accumulated doses between 10 and 20 monitor units
o ±3 % for accumulated doses between 5 and 10 monitor units
Method of Test
1. Using the applicable AVG column data from the worksheet and the
equations shown in the following table, calculate and record the test results
(not required if using the Lotus Notes SVS or Excel spreadsheet).
Hi- X ±1%
Lo-X ±1%
Hi-e- ±1%
Initial Date
Specification
• Dose reproducibility with variations in dose rate from minimum to maximum
shall be per the specifications listed in the following table.
Method of Test
1. Using the applicable AVG column data from the worksheets and the
equations shown in the following table, calculate and record the test results
(not required if using the Lotus Notes SVS or Excel spreadsheet).
Spec
Energy RR1 RRmax
C/CD/SC/EXd EX
Initial Date
Specification
• Dose reproducibility with variations in Gantry angles shall be per the
specifications listed in the following table.
Method of Test
1. Using the applicable AVG column data from the worksheets and the
equations shown in the following table, calculate and record the test results
(not required if using the Lotus Notes SVS or Excel spreadsheet).
SPEC
ENERGY 90° 270°
C/CD/SC/EXd EX
Initial Date
6. DYNAMIC THERAPY
Specification
• The Dose Difference Standard Deviation (sD) shall be <0.20 MU.
• The Dose Weighted Gantry Position Difference Standard Deviation (sDP) shall
be <0.50°.
Test Setup
1. Position the Couch angle at 0° (180° VAR) and vertical position at isocenter.
2. Enter the treatment room and verify that no obstructions will prevent a
dynamic run.
3. Disable RV interface before performing dynamic tests or place VARiS
Treatment in Standby.
o Enter Physics mode. Select Configuration / RV Interface and disable
Interface Enabling.
o Save the changes before exiting Physics mode.
Method of Test
180° 180°
Hi-X 90MU 0.5
(90° to 270°) (90° to 270°)
45° 45°
Hi-X 720MU 16.0
(270° to 225°) (270° to 315°) SD = <.20 MU
180° 180° SDP = <0.5°
Hi-e 90MU 0.5
(90° to 270°) (90° to 270°)
45° 45°
Hi-e 720MU 16.0
(270° to 225°) (270° to 315°)
180° 180°
Hi-X 90MU 0.5
(90° to 270°) (90° to 270°)
90° 90°
Hi-X 900MU 10.0
(270° to 180°) (270° to 0°) SD = <.20 MU
180° 180° SDP = <0.5°
Hi-e 90MU 0.5
(90° to 270°) (90° to 270°)
90° 90°
Hi-e 900MU 10.0
(270° to 180°) (270° to 0°)
Initial Date
Specification
• The Dose Difference Standard Deviation (sD) shall be <0.20 MU.
• The Dose Weighted Jaw Position Difference Standard Deviation (sDP) shall
be less than 0.15 cm.
Test Setup
1. In Physics Mode, commission the Enhanced Dynamic Wedge and disable
the ORNT I/L.
2. Rotate the gantry to 0º (180º VAR).
3. Disable RV interface before performing dynamic tests or place VARiS
Treatment in Standby.
4. Enter Physics mode. Select Configuration / RV Interface and disable
Interface Enabling.
5. Save the changes before exiting Physics mode.
Method of Test
Initial Date
Specification
• With the Collimator & Gantry at 0° (180° VAR), the protection zone area is
centered and parallel to the Interface Mount and meets the following
requirements:
o The 177.8-mm handle protrusion of the T shaped Gauge (100012420-
01) will clear the protection zone (refer to Figure 5).
o The 36.9-mm Gauge Plug (100011752-02) clears the protection zone.
o The 1-mm Gauge Shim (100011752-04) used with the 36.9-mm Gauge
Plug will penetrate the protection zone on either side of the zone.
• With the Collimator at 90° and the Gantry at 0° (180° VAR), the protection
zone tilt alignment is 3° relative to the Interface Mount (ref Figure 7) and
meets the following requirements:
o The 26-mm Gauge Plug (100011752-01) attached to the rear Interface
Mount pin (gun side) will clear the warning zone.
o The 47.8-mm Gauge Plug (100011752-03) attached to the front
Interface Mount pin (target side) will clear the warning zone.
o If a 1-mm shim is added to either Gauge plug, the warning zone is
penetrated.
Method of Test
1. Position the Gantry and Collimator at 0° (180° VAR).
2. Using the T-shaped alignment gauge device provided with the system,
Verify the protection zone is centered relative to the Interface Mount. Install
the device as shown in Figure 5 and confirm that the red indicator light on
the Gantry PRO display is not illuminated.
Figure 5.
3. Adjust the volume on the Master Control Panel (in the Stand) so that it is
audible to the user in the room.
4. Attach the medium-height gauge plug to the Interface Mount over the
Accessory Mount alignment pin as shown in Figure 6. This gauge plug
should not intrude into the protection zone and the red indicator light should
not illuminate.
5. Place the medium gauge plug onto the opposite alignment pin and Verify
the red indicator light does not illuminate.
6. Reattach the medium height gauge plug at each location with the 1-mm
shim in place. The red light should now illuminate and there should be an
audible alarm.
Figure 6.
Initial Date
Specification
• With the treatment room door closed, the Gantry motion should stop when an
intrusion into the protection zone occurs.
Method of Test:
1. Remove the Beam-On key from the Clinac Dedicated Keyboard. While
remaining inside the treatment room, close the room door to apply –12V
from the Door switch to the LaserGuard control unit.
2. Place the foam collision block provided on the Couch top in a position that
will cause the Collimator to hit the block before it hits any portion of the
Couch (see Figure 8). Make sure the corner of the block is positioned so
that the corner will penetrate the protection zone and not hit in the central
clear zone area (see Figure 5).
3. Using a pendant, carefully rotate the Gantry towards the block. Verify that
Gantry motion stops when the block penetrates the protection zone and that
the red collision indicator on the Gantry PRO is illuminated.
4. Exit the treatment room, close the door and verify the red indicator on the
Console display is illuminated.
Figure 8.
Collision Test
Initial Date
Specification
• The yellow indicator light on the Gantry PRO display and at the Console
display should illuminated when the power key switch is off or either of the
override buttons in the Stand are activated.
Method of Test
1. Turn the key switch to the OFF position. Verify that LaserGuard is
deactivated, and that the yellow indicator on the Gantry PRO and Console
display is illuminated.
2. Re-enable the LaserGuard system by turning the key switch to the ON
position. Press each of the override buttons to ensure the yellow indicators
are illuminated.
Initial Date
Modulator
START button
HVPS circuit breaker
Other circuit breakers
Explain the presence of HV in the Modulator and the Crowbar noise when opening the
d
Explain the doors must be closed to clear the HVCB interlock
Demonstrate Emergency Off switches - must be pulled out to reset
Stand
SF6 Gas system nominal pressure [32 psig] (Morning Checkout item) - demonstrate how
tAir pressure
fill gauge – nominal setting is 45-50 psig
Briefly explain the function of the Klystron
Demonstrate Stand Emergency OFF switches (activated by door button so do not slam
d
Water) Temperature gauge should indicate 40° prior to treatment (Morning Checkout
Closed System Water Pressure gauge (Morning Checkout item)
Water level check (Morning Checkout item)
Explain how to fill water (distilled water only and use funnel)
Briefly explain RF Driver and EMI Power Supplies – point out front panel breakers
Briefly explain Auxiliary Electronics chassis (controls Clinac & Carrousel motions, Bmag
Couch )
Demonstrate Pendant operation (Room Light, Lasers, Field Light & SSD controls)
Explain Motion Enable bars (must depress both MEB’s at the same time)
Demonstrate Motions
Demonstrate symmetric and asymmetric Jaw motions
Explain Pendant storage position (interlock PNDT)
Demonstrate how to change a pendant cable
Demonstrate Side Panel operation (Motion Enable required)
Demonstrate Longitudinal and Lateral brakes (also cause PNDT interlock)
Demonstrate Beam Stopper control (if installed)
Emergency Off buttons (must press again to reset)