Issue Date:
01-03-2022
MRI ERROR L
Effective Date:
02-03-2022
DATE ERROR DIALOG CODE FAULT
20.3.2022 GPA FLOW ERROR 24190 WATER FLOW
21.3.2022 SENSE UNIT 24180 29.5C
22.3.2022 GPA FLOW ERROR 24190 WATER FLOW
24.3.2022 GPA FLOW ERROR 24190 WATER FLOW
25.3.2022 GPA FLOW ERROR 24190 WATER FLOW
SENSE UNIT 24180 COMPRESSOR FLOW ALARM
SENSE UNIT 240067 COMPRESSOR FLOW ALARM
30.3.2022 Humidity exceed 60% 24157 to check air conditioning system is properly set and operating
16.4.2022 HE PA(8776) CHILLER TRIP
18.4.2022 System SHUT DOWN Genarator Fault
8.5.2022 HE pressure Up (3762) HE Compressor Overload
16.5.2022 HE pressure Up(3379) HE Compressor Overload
30.5.2022 HE pressure Up(3297) HE Compressor Overload
12.6.2022 GPA FLOW ERROR 24190 WATER FLOW(SENSE)
26.6.2022 GPA FLOW ERROR 24190 WATER FLOW(SENSE)
30.6.2022 GPA FLOW ERROR 24190 WATER FLOW(SENSE)
3.7.2022 HE PRESSURE INCRESE TO 8668 HE LEVEL DECREASE TO 62%,CHILLER TRIP
4.7.2022 HE PRESSURE INCRESE TO 8669 HE LEVEL DECREASE TO 62%
5.7.2022 HE PRESSURE INCRESE TO 8670 HE LEVEL DECREASE TO 62%
11.7.2022 FLOW TEMPERATURE HIGH CHILLER TRIP*(CIRCUIT 1 ERROR)
15.7.2022 GPA FLOW ERROR 24190 WATER FLOW
SENSE UNIT 24180 COMPRESSOR FLOW ALARM
SENSE UNIT 240067 COMPRESSOR FLOW ALARM
20.8.2022 HE PRESSURE INCRESE TO 8500 HE Compressor Error
14.9.2022 DRF During the scan DRF circuit board
14.9.2022 FLOW TEMPERATURE HIGH chiller trip 2times
15.9.2022 CPU ERROR DRF circuit board ERROR
16.9.2022 CPU ERROR DRF circuit board ERROR
**GPA Flow unstable ,flow temperature increasing and Sense unit for MRI Syst
27.9.2022 SYSTEM ERROR Flow Temperature 30
29.7.2022 SYSTEM ERROR Flow Temperature 30
30.9.2022 SENSE UNIT ERROR GPA FLOW Down,He Level Down
MRI ERROR LOG
STATUS
SYSTEM ERROR
MR NOT READY
SYSTEM ERROR
SYSTEM ERROR
SYSTEM ERROR
SYSTEM ERROR
SYSTEM ERROR
SYSTEM ERROR
SYSTEM ERROR
SYSTEM ERROR
SYSTEM ERROR
SOFTWARE REINSTALL
SYSTEM ERROR
SYSTEM ERROR
SYSTEM ERROR
He comp flow - 1.21 at 12AM / Fixed to 8 / Back to normal range at 2AM
analog to digital convertor error
SYSTEM ERROR /CPU NOT READY
SYSTEM ERROR/CPU NOT READY
ure increasing and Sense unit for MRI System Error still occurs during scan with patient.
SYSTEM ERROR WHILE SACANNING L/SPINE
SYSTEM ERROR WHILE SACANNIG Brain
SYSTEM ERROE ,MR dosen't give Ready,BEFORE SCANNING at 8AM
MDL-MIT-RE-030
01
ACTION REMARK
FIXED
FIXED
FIXED
FIXED
FIXED
FIXED
FIXED
FIXED
FIXED
FIXED
Restart Compressor
Restart Compressor
FIXED
FIXED
FIXED
FIXED
FILLED UP WITH EXTERAL HE GAS CYLINDER
FILLED UP WITH EXTERAL HE GAS CYLINDER
FILLED UP WITH EXTERAL HE GAS CYLINDER
FIXED
FIXED
FIXED
FIXED
FIXED/ flow temp: increase and GPA & He comp: flow sill unstable during scan time Lasted 6days (Fixed at 26.8.2022)
FIXED
g scan with patient.
FIXED Lasted 1 day
FIXED Lasted 1 day
FIXED Lasted 1 day
Issue Date:
01-03-2022
MRI ERROR
Effective Date:
02-03-2022
DATE ERROR DIALOG CODE
11.2.2023 HE Quench _
17.6.2023 CPU not ready _
21.6.2023 CPU not ready _
2.7.2023 Can't do postprocessing(FE) _
7.7.2023 CPU not ready _
8.7.2023 Abnormal sound at MR Gnatry room of cold head area _
MRI ERROR LOG
FAULT STATUS
CIRCUIT BREAKER ERROR SYSTEM ERROR
RFIP UNIT,MAINBOARD ERROR SYSTEM ERROR
MOTHERBOARD ERROR SYSTEM ERROR
INSUFFICIENT DATA STORAGE SYSTEM ERROR
MOTHERBOARD ERROR SYSTEM ERROR
Equipment Error
MDL-MIT-RE-030
01
ACTION REMARK
FIXED/He 76%/Add Portable chiller Lasted 3 months (30.5.2023)
Changed DRF circuit board Lasted 3 days (20.6.2023)
Shut down & Restart the system Lasted 1 day
Changed Power Supply Circuit Lasted 1 day
Reinstall Software Lasted 1 day
Checked equipment at MR Gantry room(MMT) According to OKTR (Machine can use as normal)