BSM-2301K / BSM-2303K

BSM-2351K / BSM-2353K
Bedside Monitor
Technical Data
DESIGN FEATURES
High resolution, 10.4/8.4 inch color display Network capability
The monitor has a high resolution, wide-viewing An optional network card allows networking with
angle TFT color LCD for easy viewing. CNS-9700 series and BSS-9800. Also, by attaching
a ZS-900PK transmitter to the monitor, telemetry
Touch panel and optional touch pen networking is available with a WEP-4200 series and
A touch panel and touch pen provide easy and quick ORG-9700 multi-patient receiver.
operation. You can directly write the patient name
on screen with the touch pen or finger. Any charac- Interbed
ters and figures can be displayed. (Some limitations When LifeScope series bedside monitors are con-
apply when QI-101P is connected.) nected to each other by Ethernet, you can view data
of other bedside monitors.
7 parameters, 6 or 5 waveforms
7 parameters—ECG, Resp (impedance and ther- 3 hour battery operation
mistor method), SpO2, NIBP, Temp, IBP (BSM-2303/ One fully charged new battery gives around three
2353K) and either IBP or CO2 (mainstream)—are hour operation.
available. Up to 6 (BSM-2303/2353K) or 5 (BSM-
2301/2351K) waveforms are displayed. 3 channel thermal array recorder option
A thermal array recorder records waveforms of ECG,
MULTI connector respiration, IBP, pulse and ETCO2.
The MULTI connector gives you high flexibility in the
selection of parameters. When you plug the sensor Multigas measurement (option)
cable into a MULTI connector, the monitor automati- You can measure CO2, N2O, O2, and anesthetic
cally detects the type of parameter and starts agents (Halothane, Isoflurane, Enflurane,
measuring. Sevoflurane, Desflurane) with optional AG-920RA/RK
multigas unit. Gases can be measured quickly with
PWTT triggered NIBP measurement small sampling flow rate.
PWTT is non-invasively measured by continuous
monitoring of ECG and SpO2. If a sudden critical
circulation change occurs between periodic NIBP
measurements, PWTT may detect it and trigger
NIBP measurement to confirm it. PWTT trend
indicates how the patient’s PWTT is changed.
Alarm indicator
An alarm indicator at the top of the monitor can be
seen clearly from a distance. It lights in red or yellow
depending on the severity of the alarm. A green
lamp blinks in synchronization with the QRS.
High quality arrhythmia detection
The multi-template matching method reduces false
alarms and assures high quality arrhythmia detection.
Eight dominant ECG waveform windows let you
confirm accuracy. You can edit the dominant wave-
form in the windows.
OCRG screen
The OCRG (Oxygen Cardiorespirogram) screen with
heart rate trendgraph, respiration waveform is useful
for neonatal monitoring. Also lead selection helps in
difficult neonatal respiration measurement.

2
SYSTEM COMPOSITION
Telemetry
Multigas unit system
AG-920R Transmitter WEP-4200 series
ZS-900PK
CO unit Bedside Multi-patient
2
monitor
LS-NET Central
AG-400R receiver monitor
Interface *1
BSM-2301K
QI-231P
*2
ORG-9700 CNS-9700
BSM-2303K
BSM-2351K series
External BSM-2353K Interface LS-NET Central
monitor *3 QI-101P monitor
CNS-9700 series
Isolation
transformer *4
Bedside
station
BSS-9800 (central mode)
*1: Interface QI-231P is not required if you only connect either CO2 unit, Multigas unit, or external monitor
*2: QI_RGB cable (YS-080P3/YS-080P4)
*3: EIZO Flexscan T962/L680 or equivalent (local purchase)
*4: Medical grade

SPECIFICATIONS
uÿMeasuring Parameters Waveform freeze: Provided
Display waveforms: ECG, respiration, IBP, SpO2 pulse
ECG, respiration (impedance), NIBP, SpO2, tempera- wave, and CO2
ture, pulse, IBP (BSM-2303/2353K), and thermistor Numerical data display: Heart rate, VPC rate, ST
respiration, IBP, or CO2 (MULTI connector) level, respiration rate, IBP (systolic, diastolic, mean),
NIBP (systolic, diastolic, mean), SpO2, pulse rate,
temperature, and ETCO2
uÿDisplay Synchronization mark: Heart rate sync mark, pulse
rate sync mark, respiratory sync mark
Display size:
BSM-2301/2303K: 8.4 inch, TFT color LCD
BSM-2351/2353K: 10.4 inch, TFT color LCD
Waveform display mode: Non-fade moving or non- uÿSound
fade fixed Type of sound: Alarm, synchronization, click
Viewing area: Alarm sound: 3 types
BSM-2301/2303: 170 × 128 mm Synchronization sound: Pitch variable for IBP and
BSM-2351/2353: 211 × 158 mm SpO2
Resolution: 800 × 600 dots
Number of waveforms:
BSM-2301/2351K: 5
BSM-2303/2353K: 6 uÿAlarm
Sweep speed: 1 mm/s, 6 mm/s, 25 mm/s, 50 mm/s Alarm items: Upper/lower limits, apnea, arrhythmia,
(respiration and CO2 low speed: 1.56 mm/s) connector disconnection, electrode off, pulse wave-
Sweep width: form detecting, probe off, cuff/hose check, sensor
BSM-2301/2303K: about 124 mm (at 25 mm/s sweep check, battery weak, operating environment
speed) Alarm levels: Crisis (red, blinking), Warning (yellow,
BSM-2351/2353K: about 154 mm (at 25 mm/s sweep blinking), Advisory (yellow, lit)
speed) Alarm indication: Alarm indicator, highlighted mes-
Waveform display colors: 12 sage, alarm sound
Numeric display colors: 12 Alarm suspend: Provided (for 1 or 2 min)

3
uÿAlarm History Defibrillation-proof: ECG input protected against 400J
discharge (IEC60601-2-27 17.101 compatible)
Number of files: 200 ESU interference filter: Provided
Filters:
ON: Time constant 0.5 s, AC hum filter 0.3 to 23 Hz
uÿTrendgraph (>−3dB), ≤−16 dB (50 Hz or 60 Hz)
OFF: Time constant 3.2 s, 0.05 to 150 Hz (>−3dB)
Trend parameters: Heart rate (or pulse rate), respira- ECG leads:
tion rate, VPC rate, ST level, Event (arrhythmia), 3-electrode: I, II, III
apnea (time), apnea (frequency), SpO2, NIBP (systolic, 6-electrode: I, II, III, aVR, aVL, aVF, 2 from V1 to V6
diastolic, mean), IBP (systolic, diastolic, mean), Display sensitivity: 10 mm/mV ±5% (at ×1 sensitivity)
temperature, ETCO2 Sensitivity control: ×1/4, ×1/2, ×1, ×2, ×4, or Auto
Trend time: 1, 2, 4, 8 or 24 h Pacing spike display: Available
Data storage time: 1 min for 1, 2, 4, 8 hours, 3 min for Heart rate display update interval: Every 3 s or when
24 hours alarm is generated
Alarm limits:
Upper limit: 20 to 300 beats/min in 5 beats/min steps,
uÿVital Signs List Off
Lower limit: Off, 15 to 295 beats/min in 5 beats/min
Parameters: Heart rate (or pulse rate), VPC rate, ST steps
level, NIBP (systolic, diastolic, mean), SpO2, IBP Alarm items: TACHYCARDIA, BRADYCARDIA, ASYS-
(systolic, diastolic, mean), respiration rate, tempera- TOLE
ture, ETCO2 ST alarm limits: ±2.0 mV in 0.01 mV steps, Off
Number of files in list: 120 for periodic list, 120 for
NIBP list
List interval: 1, 5, 15, 30 or 60 min for periodic list, at uÿRespiration (impedance)
NIBP measurement for NIBP list
Measuring lead: R-F or R-L
Measuring impedance available range: 0 to 2 kΩ
uÿRecall Internal noise: ≤ 0.2 Ω, referred to input
Excitor current: 30 ±10µArms at 40 kHz
Item: Arrhythmia Frequency response*: 3.0 Hz ±1 Hz (−3 dB)
Number of recall files: 16 Time constant*: 1.5 s ±0.5 s
Waveform length: 8 s * Hardware specification
Counting range: 0 to 150 breaths/min
Apnea counting range: 5 to 40 s
uÿECG Counting accuracy: ±2 breaths/min
Defibrillation-proof: Respiration input protected
Electrode offset potential tolerance: ±500 mV against 400 J discharge
Input dynamic range: ±5 mV Waveform display sensitivity: 10 mm/Ω ±20% (at ×1
Internal noise: ≤30µVp-p, referred to input sensitivity Zo=480 Ω)
CMRR: ≥95dB Waveform display sensitivity control: ×1/4, ×1/2, ×1,
Input impedance: ≥5 MΩ (at 10 Hz) ×2, ×4
Input bias current: ≤100 nA Measurement On/Off: Available
Heart rate counting method: Moving average/Instan- Display update cycle: Every 3 s or when alarm is
taneous beat-to-beat (Selectable) generated
Heart rate counting range: 0, 12 to 300 bpm (±2 Alarm limits:
beats/min) Upper limit: 2 to 150 breaths/min in 2 breaths/min
Counting accuracy: ±2 bpm steps, Off
Arrhythmia analysis method: Multi-template match- Lower limit: Off, 0 to 148 breaths/min in 2 breaths/min
ing method steps
VPC counting range: 0 to 99 VPCs/min Apnea time: Off, 5 to 40 s in 5 s steps
Arrhythmia message: ASYSTOLE, VT, VF, VPC RUN,
COUPLET, EARLY VPC, BIGEMINY, FREQ VPC,
TACHYCARDIA, BRADYCARDIA
ST measuring range: ±2.5 mV uÿRespiration (thermistor)
Number of ST measurement channels: 1 Counting range: 0 to 150 breaths/min
Pacemaker pulse rejection capability: 0.1 to 2 ms, Apnea counting range: 5 to 40 s
±2 to 700 mV (ANSI/AAMI EC 13-1992 compatible) Counting accuracy: ±2 breaths/min
Pacing pulse detection: On/Off Noise: within 20 Ω, referred to input

4
Frequency response: 3.0 Hz (−3 dB)
Time constant: 1.5 s or more
uÿTemperature
Display sensitivity: 10 mm/100Ω ±20% (at ×1 sensi- Measuring range: 0 to 45°C
tivity) Accuracy: ±0.1°C (25 to 45°C), ±0.2°C (0 to 25°C)
Sensitivity control: ×1/4, ×1/2, ×1, ×2, ×4 Temperature drift: ±0.005°C/°C
Display update cycle: Every 3 s or when alarm is Display update: Every 3 s
generated Alarm limits:
Alarm limits: Upper: 0.1 to 45°C in 0.1°C steps, Off
Upper limit: 2 to 150 breaths/min in 2 breaths/min Lower: Off, 0 to 44.9°C in 0.1°C steps
steps, Off
Lower limit: Off, 0 to 148 breaths/min in 2 breaths/min
steps
Apnea time: 5 to 40 s in 5 s steps, Off
uÿIBP
Measuring range: −50 to 300 mmHg
uÿSpO 2
Accuracy:
−50 to 99 mmHg: ±1 mmHg ±1 digit
Measuring range: 1 to 100% 100 to 300 mmHg: ±1% ±1 digit
Pulse rate counting range: 0, 30 to 300 bpm Auto zero balancing range: ±200 mmHg
Accuracy (amplifier): ±1 digits SpO2 (80 to 100%), Auto zero balancing accuracy: ±1 mmHg
±2 digits SpO2 (50 to 79%), unspecified (<50%) Transducer sensitivity: 50 µV/V/10 mmHg
Pulse rate display update cycle: Every 3 s or when Pulse rate counting range: 0, 12 to 300 beats/min
alarm is generated Pulse rate counting accuracy: ±2 bpm
SpO2 tone modulation: Changes in 20 steps from 81 Noise: Within ±1 mmHg
to 100% Temperature zero drift: ±0.1 mmHg/°C
Waveform sensitivity: ×1/8, ×1/4, ×1/2, ×1, ×2, ×4, ×8 Frequency response: DC to 20 Hz ±3 Hz or DC to
or AUTO 12Hz ±3 Hz
Alarm limits: Blood pressure value display update interval:
Upper limits: 51 to 100% in 1% SpO2 steps, Off Every 3 s or when alarm is generated
Lower limits: Off, 50 to 99% in 1% SpO2 steps BP sync sound: Provided, systolic value 20 to 120
mmHg, changes in 20 steps every 5 mmHg
Alarm limits:
uÿNIBP Upper: 2 to 300 mmHg in 2 mmHg steps, Off
Lower: Off, 0 to 298 mmHg in 2 mmHg steps
Measuring method: Oscillometric
Cuff pressure display range: 0 to 300 mmHg
Accuracy:
0 to 200 mmHg: ±3 mmHg
uÿCO 2

200 to 300 mmHg: ±4 mmHg Measurement method: Mainstream (TG-900/920P:


Cuff inflation time: Adult 7 s, Neonate 5 s semi-quatitative, TG-950P: quantitative)
Initial cuff inflation pressure: Adult 180 mmHg, Measuring range: 0 to 99 mmHg
Neonate 100 mmHg Accuracy:
Cuff inflation maximum pressure limiter: Adult 300 TG-900/920P (at 1 atmospheric pressure, air inspira-
mmHg, Neonate 150 mmHg tion, no condensation):
Safety cuff inflation limiter: Adult 330 mmHg, 0 to 40 mmHg: ±4 mmHg
Neonate 165 mmHg 41 to 76 mmHg: ±10% reading
Cuff inflation time limiter: TG-950P (When no condensation, BTPS (body tem-
Adult ≤180 s, Neonate ≤90 s perature 37°C, ambient pressure, saturated with
Measurement mode: Manual, Periodic (2, 2.5, 5, 10, vapor)
15, 30 min, 1, 2, 4, 8 hrs interval, PWTT), STAT 0 to 40 mmHg: ±2 mmHg
(Continuous) 41 to 70 mmHg: ±5% reading
Measurement completion sound: Generated at every 71 to 99 mmHg: ±7% reading
measurement completion when set in the system setup Warm-up time: 5 s minimum
screen Response time:
Display update cycle: Updated every measurement TG-900/920P: 200 ms (typical) from 10 to 90%
Alarm limits: TG-950P: 60 ms (typical) for steps from 10 to 90%,
Upper: 15 to 260 mmHg in 5 mmHg steps, Off delay time 100 ms ±10 ms
Lower: Off, 10 to 255 mmHg, in 5 mmHg steps Respiration rate counting range:
TG-900P: 3 to 60 breaths/min
TG-920P: 3 to 150 breaths/min
TG-950P: 0 to 150 breaths/min

5
Respiration rate counting accuracy: ±2 breaths/min Paper speed: 25, 50 mm/s
CO2 value display update cycle: Every 3 s or when Parameters: ECG, RR, IBP, pulse, CO2
alarm is generated Recording paper: A723, 50 mm wide
Alarm limits:
Upper: 2 to 99 mmHg in 1 mmHg steps, Off
Lower: Off, 1 to 98 mmHg in 1 mmHg steps
Apnea time: 5 to 40 s in 5 s steps, Off
uÿPower Requirements
Line voltage: 100 to 240 V AC ±10%
Line frequency: 50 or 60 Hz ±2%
uÿInterbed Power consumption:
BSM-2301K/2303K: 86 VA max
Waveforms and data of other Life Scope series bedside BSM-2351K/2353K: 95 VA max
monitor connected by Ethernet can be reviewed.
Display items for personal display:
Waveform: 1 trace (selectable)
Numeric data: ECG, SpO2, Resp, CO2, IBP1, IBP2,
uÿClock Accuracy
IBP3, NIBP, TEMP, O2 At operating temperature 25°C: about ±2 min 40 s/
month
At storage temperature –20 to 60°C: about ±5 min/
uÿCO2 Unit, AG-400RA/K (option) month
Measured paremeters: Sidestream CO2
*For detailed specifications, see the individual techni-
cal data.
uÿBattery Operation
Type of battery: Nickel-metal hydride
Voltage: 10.8 to 15 V DC ±5%
uÿMultigas unit, AG-920RA/K (option) Power consumption: 40W maximum
Operation time: Approx. 3 hours when used under the
Measured gases: CO2 partial pressure, N2O concentra- specified conditions
tion, O2 concentration, anesthetic agent concentration Charging mode: Standard and trickle charging mode,
(Automatic detection of Halothane, Isoflurane, automatic selection
Enflurane, Sevoflurane, Desflurane. Primary and Charging time: approx 16 hours to charge
secondary agents can be measured), respiration rate Battery life: 1 year or up to 200 times charging/dis-
*For detailed specifications, see the individual techni- charging
cal data.

uÿEnvironment
uÿTransmitter, ZS-900PK (option) Operating temperature: 10 to 40°C
Parameters sendable to central monitor: ECG (1ch), Storage temperature: −20 to 60°C
respiration, SpO2, NIBP, IBP (2 ch), temp (2ch), CO2 −15 to 55°C (recording paper)
*For detailed specifications, see the individual techni- Operating humidity: 30 to 90% RH (0 to 40°C non-
cal data. condensing)
Storage humidity: 10 to 90% RH (0 to 40°C non-
condensing)
uÿExternal Output Atmospheric pressure: 70 to 106 kPa
ZS-900PK transmitter
QI-231P interface (option) or external monitor or AG-
400R CO2 unit or AG-920R multigas unit
uÿElectromagnetic Compatibility
IEC60601-1-2 (1993)–Collateral Standard: Electro-
magnetic compatibility–Requirement and tests
uÿRecorder, WS-231P (option) Emissions: CISPR11 Group 1, Class B
Recording method: Thermal array recording
Number of channels: 3 traces (maximum) uÿSafety
Recording width: 46 mm
Resolution: IEC60601-1 (1988)
Amplitude direction of waveforms: 8 dots/mm IEC60601-1 Amendment 1 (1991)
Time direction of waveforms: 40 lines/mm, 8 lines/mm IEC60601-1 Amendment 2 (1995)
(graphic recording)
6
IEC60601-1-1 Amendment 1 (1995)
IEC60601-2-27 (1994) Particular requirements for the
safety of electrocardiographic monitoring
STANDARD ACCESSORIES
IEC60601-2-34 (1994) Particular requirements for the BSM-2301K/2303K /2351K/2353K
safety of direct blood pressure monitoring equipment None
IEC60601-2-30 (1995) Particular requirements for the
safety of automatic cycling in direct blood pressure WS-231P
monitoring equipment Recording paper 1
According to the type of protection against electri- Thermal head cleaner pen 1
cal shock: Class I Equipment (AC Powered), Internally
Powered Equipment (Battery Powered)
According to the degree of protection against
electrical shock:
Defibrillator-proof type CF applied part: ECG, Resp
(impedance/thermistor), IBP, CO2, Temp
Defibrillator-proof type BF applied part: NIBP, SpO2 OPTIONAL ACCESSORIES
According to the degree of protection against
harmful ingress of water: IPX0 (Ordinary Equipment) Accessory kit (An accessory kit must be ordered)
According to the degree of safety of application in YZ-028P2 (100V IEC type)
the presence of a flammable anaesthetic mixture YZ-028P3 (200V IEC type)
with air, or oxygen or nitrous oxide: Equipment not YZ-028P4 (GB IEC type)
suitable for use in the presence of flammable anaes- YZ-028P5 (200V AHA type)
thetic mixture with air or with oxygen or nitrous oxide
According to the mode of operation: CONTINUOUS Consists of:
OPERATION Power cord 1
H for YZ-028P2
N for YZ-028P3/YZ-028P5
uÿDimensions and Weight GB AC for YZ-028P4
Electrode lead, for 3 electrodes 1
Dimensions: K911 for YZ-028P2/YZ-028P3/YZ-028P4
BSM-2301/2303: 253 W × 242 H × 145 D mm BR-903PA for YZ-028P5
BSM-2351/2353: 293 W × 272 H × 149 D mm ECG connection cord, 3/6 electrodes 1
WS-231P: 212 W × 90 H × 140 D mm K922 for YZ-028P2/YZ-028P3/YZ-028P4
Weight: JC-906PA for YZ-028P5
BSM-2301/2303: approx. 4.7 kg, excluding options Disposable electrode set (C-30) 1
BSM-2351/2353: approx. 5 kg, excluding options SpO2 connection cord, K931 1
WS-231P: approx. 1.5 kg Air hose for adult/child, 3.5 m, S902 1
Cuff for adult, YP-963T 1
Cuff for child, YP-962T 1

7
OPTIONS CONSUMABLES
Multigas unit, AG-920RA/K ECG and Respiration (Impedance)
CO2 unit, AG-400RA/K
Connection cable, for AG-400R, YJ-231P Cable/cord
Recorder module, WS-231P K911 Electrode lead, for 3 electrodes, clip type, 0.8
Grounding lead, 544582A m, BR-903P
Fuse T2.0A/250V, 104522 K911A Electrode lead, for 3 electrodes, clip type, 0.8
Battery pack, 10HR-4/3FAUC-NK, X062 m, BR-903PA, AHA
Transmitter, ZS-900PK K910A Electrode lead, for 3 electrodes, snap type,
Adapter for attaching transmitter, DI-230P 0.8 m, BR-913P
Hook, DZ-230P K910B Electrode lead, for 3 electrodes, snap type,
Cart, KC-013P 0.8 m, BR-913PA, AHA
Network card, QI-101P K912 Electrode lead, for 6 electrodes, clip type, 0.8
Network printer card, QI-111P m, BR-906P
Wireless LAN station, QI-210P K912A Electrode lead, for 6 electrodes, clip type, 0.8
Wireless LAN access point, QI-902R/902RK m, BR-906PA, AHA
Touch pen, Y075 K915 Electrode lead, for 6 electrodes, snap type, 0.8
BSM-2300 RGB cable, 10 m, YS-076P2 m, BR-906P
BSM-2300 RGB cable, 2 m, YS-080P2 K915A Electrode lead, for 6 electrodes, snap type,
Interface, QI-231P, output trigger or ECG analog 0.8 m, BR-906PA, AHA
signals to connected instruments, add external display K922 ECG connection cord, 3/6 electrodes, 3 m, JC-
Interface for a display and an external instrument, 906P
QI-236P K922A ECG connection cord, 3/6 electrodes, 3 m,
Adapter for attaching interface, DI-231P JC-906PA, AHA
QI RGB cable, 10 m, YS-080P3 Disposable electrodes
QI RGB cable, 2 m, YS-080P4 G201 for general use, 30 electrodes × 5 sets, 47 × 47
Isolation transformer, SM-800RJ/K mm, cottony tape base, Bs-150
External monitor (local purchase), EIZO Flexscan G204 for general use, 30 electrodes × 5 sets, 47 × 47
T962/ L680 or equivalent mm, cottony tape base, C-150
Minimum specifications of external monitor: G221 for general use, 30 electrodes × 20 sets, 47 ×
1) Resolution: 800 × 600 47 mm, foam pad base, solid gel, G-600
2) Input: Analog RGB 0.7Vp-p G250 for general use, 30 electrodes × 5 sets, 40 mm
3) H-sync: 39KHz dia, form pad base, solid gel, J-150
4) V-sync: 62.5 Hz G210D for infant/neonate to adult, 3 electrodes × 50
5) Sync. signal: Negative sets, 25 × 45 mm, F-150M
6) Safety standards: ICE60601-1 or must use with a G272A for general use, 3 electrodes × 30 sets, 25 ×
medical grade isolation transformer. 45 mm conductive sheet base, DIN type, V-090M3
7) Connection cable (YS-076P2/YS-080P2) must be G272B for general use, 4 electrodes × 10 sets, DIN,
purchased from Nihon Kohden. V-040M4
8) When using LCD display, set waveform display G278A for general use, 3 electrodes × 30 sets, 25 ×
mode to non-fade fixed mode 45 mm, DIN, V-09IO3
G273A for general use, 4 electrodes × 10 sets, 25 ×
CAUTION: 45 mm, DIN, V-04IO4
Some commercially available monitors do not work G272C for general use, 6 electrodes × 10 sets, DIN,
well with BSM-2301K/2303K/2351K/2353K. Before V-060M6
purchasing an external monitor locally, you must G274A for general use, 6 electrodes × 10 sets, 25 ×
confirm whether the monitor works with BSM-2301K/ 45 mm, DIN, V-06IO6
2303K/2351K/2353K. G217 for general use, 3 electrodes × 30 sets, 55 mm
dia, autoclave, D-90
G271A for NICU, 3 electrodes × 40 sets, DIN, 20 × 20
mm, V-120S3
G210C for NICU, 3 electrodes × 50 sets, 18 × 36 mm,
F-150S
G300A for NICU, 3 electrodes × 10 sets, DIN, 20 × 20
mm, N-03IS3

8
Respiration (thermistor method) Sterilized disposable cuffs:
S260 YP-111P: for neonates, 3 cm cuff width
P901 Thermistor respiration pickup for nose, TR-900P S260A YP-112P: for neonates, 4 cm cuff width
P902 Thermister respiration pickup for airway, TR- S260B YP-113P: for neonates, 5 cm cuff width
910P Air hoses:
V911 Airway adapter, YG-001P S903 YN-990P: extension hose, 1.5 m
S901 YN-900P: 1.5 m
SpO2 S902 YN-901P: 3.5 m
S904 YN-920P: for neonate, 1.5 m
Cord S905 YN-921P: for neonate, 3.5 m
K931 SpO2 connection cord, 2.5 m, JL-900P
Probes (BluPRO series)
P225F Finger probe, attached to finger or toe by clip, IBP
TL-201T
P225G Multi-site probe, attached to finger or toe by IBP connection cord
tape, TL-220T K951 for P-23ZL1/P-10EZ1, 3.5m, JP-900P
P259 Cottony tape, 340703, secures probe and cord K952 for TP-400T and Becton Dickinson transducer,
for stable monitoring, stretchable, 20 pcs/set 0.3 m, JP-910P
Disposable probes (BluPRO series) Pressure transducer:
P203A for adult finger or toe, 24 pcs/set, TL-271T P-10EZ1, P-23XL1: product of Becton Dickinson and
P203B for child finger or toe, 24 pcs/set, TL-272T Company, USA
P203C for neonate instep, 24 pcs/set, TL-273T S238 Transducer holder, for TP-400T P-23XL1, ZY-
P203D for infant finger or toe, 24 pcs/set, TL-274T 101U
Multi-site Y probe S239 Adapter2, for P-23XL1, use with S238
P205A Multi-site Y probe (for adults, children, neo- Dome:
nates), 5 pcs/set, TL-260T TA-1011: for TP-400T, P-23XL1, without membrane, 1
P260A Sponge attachment tape S for TL-260T multi- pc
site Y probe, 24 pcs/set TA-1010ND: with membrane, rotatory luer lock, 12 pcs/
P260B Sponge attachment tape L for TL-260T multi- set
site Y probe, 24 pcs/set TA-1019: for P-10EZ1, with membrane, 12 pcs/set
P256 Ear clip adapter for TL-260T, reusable TA-1019M: for P-10EZ1, with membrane, 12 pcs/set,
Disposable probes with wing
P228A for adult/neonate, 80 cm cable, 5 pcs/set, TL- Super dome:
051S TA-1015T: for TP-400T, P-23XL1, with flush device,
P228B for adult/neonate, 160 cm cable, 5 pcs/set, 10 pcs/set
TL-052S TA-1017: for P-10EZ1, with flush device, 10 pcs/set
P229A for child/infant, 80 cm cable, 5 pcs/set, TL- TA-1017M: for P-10EZ1, with flush device, 10 pcs/set,
061S with wing
P229B for child/infant, 160 cm cable, 5 pcs/set, TL- TA-1018: for P-10EZ1, with flush device, 10 pcs/set
062S Criti Flo:
TA-4004, 3 cc/H, 10 pcs/set
NIBP TA-4005, 30 cc/H, 10 pcs/set
Monitoring kit:
Reusable Cuffs (Latex free): SCK-520: for TP-400T, P-23XL1, 5 kits/set,
S943A YP-960T: for infants, 5 cm cuff width SCK-512: for P-10EZ1, arm mount, 5 kits/set,
S943B YP-961T: for children, 7 cm cuff width SCK-560: for P-10EZ1, pole mount, 5 kits/set,
S943C YP-962T: for children, 10 cm cuff width Extension tube:
S944B YP-963T: for adults, 13 cm cuff width PT-06, 0.15 m, 25 pcs/set
S944C YP-964T: for adults, 15 cm cuff width PT-12, 0.3 m, 25 pcs/set
S944D YP-965T: for thigh, 19 cm cuff width PT-24, 0.6 m, 25 pcs/set
Disposable Cuffs: PT-36, 0.9 m, 25 pcs/set
YP-910P: for infants, 6 cm cuff width, 20 pcs PT-48, 1.2 m, 25 pcs/set
YP-912P: for children, 9 cm cuff width, 20 pcs PT-60, 1.5 m, 25 pcs/set
YP-913P: for adults, 12 cm cuff width, 20 pcs
YP-914P: for adults, 14 cm cuff width, 20 pcs DX series (Becton Dickinson)
YP-915P: for adults, 16 cm cuff width, 20 pcs K951 for P-23ZL1/P-10EZ1, 3.5 m, JP-900P
YP-916P: for adults, 20 cm cuff width, 20 pcs Monitoring Kit:
YP-920P: for neonates, 2.5 cm cuff width, 10 pcs DX-100: For cerebral pressure, 10 pcs/set
YP-921P: for neonates, 3 cm cuff width, 10 pcs DX-200: For neonates, 10 pcs/set
YP-922P: for neonates, 4 cm cuff width, 10 pcs DX-300: With flush device, 10 pcs/set
YP-923P: for neonates, 5 cm cuff width, 10 pcs DX-312: For arm mounting, 5 pcs/set
9
DX-360: For pole mounting, 5 pcs/set Quantitative method
DX-360R: For pole mounting, with dumping device, 5 P905 CO2 sensor kit, 4 m, TG-950P (JG-950P + TG-
pcs/set 201T)
DX-360TT: Triple line, 3 pcs/set R802 Airway adapter, Adult, 30 pcs/set, YG-201T
DX-360SD: For pole mounting, 5 pcs/set R803 Airway adapter, Child, 30 pcs/set, YG-202T
SCKD-5005: double line, 5 pcs/set K982 CO2 adapter, 2 m, JG-950P
TA-BPN: Safe needle, for DX-360SD, 50 pcs/set P921 CO2 sensor, TG-201T
QS-90: Arterial blood sampler, for DX-360SD, use with
TA-BPN CO (Sidestream measurement)
Transducer fixing plate 2

TBG: Transducer fixing plate FilterLineTM


TMM: Transducer fixing device V905A OR/EMS, 12 pcs/box, YG-910P
UMM: Transducer fixing device for pole V905B ICU, 12 pcs/box, YG-911P
Nasal FilterLineTM
Edwards Lifesciences: V907A Adult, 25 pcs/box, YG-912P
L901 IBP connection cord, for Edwards Lifesciences V907B Pediatric, 25 pcs/box, YG-913P
transducer, 3.5 m, JP-920P V907C Infant/neonate, 25 pcs/box, YG-914P
Edwards Lifesciences blood pressure transducers are MicrostreamTM airway adapter
available from Edwards Lifesciences Corporation V903E Adult, 24 pcs/box, YG-904P
(www.edwards.com) or their suppliers. V903F Infant/neonate, 24 pcs/box

Gas monitoring
Temperature V915A Dryline A-Adult, water trap, 613712, 10 pcs/
Thermistor Probe: set
P240A For adult rectum/esophagus, YSI-401JG V915B Dryline N-Neonate, water trap, 613721, 10
P241A For child rectum/esophagus, YSI-402JG pcs/set
P242A Disc type, YSI-409JG V916A T-piece S-Straight, 613739, 10 pcs/set
P249 Probe cover, for YSI-401JG, 10 pcs/set V916B T-piece E-Elbow, 613748, 10 pcs/set
P252 Temperature insulation pad, for YSI-409JG, 60 V903F T-piece, for neonate, 618877, 25 pcs/set
pcs/set V917A Sampling line A-Adult, 613766, 25 pcs/set
Sheritemp, Sonatemp, Foley catheter for bladder, V917B Sampling line N-Neonate, 613775, 25 pcs/
contact Kendall Healthcare Products Company directly set
(www.kendallhq.com) Exhaust gas adapter, 6144-008197A
CO2 (Mainstream measurement)
Semi-quantitative method
Recording Paper
P903 CO2 sensor kit, 3 m, TG-900P (JG-900P + TG- A723 FQW50-3-100, 50 mm × 30 m, 10 pcs/set
101T)
R801 Airway adapter, 50 pcs/set, YG-101T
K981 CO2 adapter, JG-900P
P922A CO2 sensor, TG-101T ORDERING INFORMATION
P907 CO2 sensor kit, 3 m, TG-920P (JG-920P + TG-
121T) When ordering, please specify the following information:
K984 CO2 adapter, JG-920P 1) Power line voltage and frequency
P923 CO2 sensor, TG-121T 2) Requirements for consumables
V921 Nasal adapter, YG-120T, 30 pcs/set 3) Requirements for optional accessories
V922 Nasal/oral adapter, YG-121T, 30 pcs/set 4) Transmission frequency of any existing telemetry sys-
V923 Nasal/oral adapter with O2 supply tube adapter, tem—Nihon Kohden’s or any other manufacturers’
YG-122T, 30 pcs/set 5) Specify the part order code in for consumables

This brochure may be revised or replaced by Nihon Kohden at any time without notice.