Electromagnetic Flowmeter

User’s Manual

Jan. 2019

-1-
CONTENTS
Electromagnetic Flowmeter........................................................................................................................ - 1 -
User’s Manual............................................................................................................................................... - 1 -
Jan. 2019....................................................................................................................................................... - 1 -
I. QTLDZ Electromagnetic Flowmeter Converter Operation..................................................................... 4
1. Product Function Introduction............................................................................................................ 4
GB 3836.2-2010 GB 3836.4-2010.................................................................................................... 4
2. Converter Exterior Size and Power Connection Instruction..................................................................5
2.1 Exterior Size..................................................................................................................................... 5
2.2 Power Supply Connection.................................................................................................................5
2.3 Output Signal Terminal Connection.................................................................................................6
3.Operation Converter......................................................................................................................................8
3.1 Key and Display..................................................................................................................................8
3.2 Menu Structure................................................................................................................................... 8
4. Parameter Set and Use.............................................................................................................................10
4.1. Zero Trim.......................................................................................................................................... 10
4.2 Fast Setting....................................................................................................................................... 12
4.3 (BAS) Setting.................................................................................................................................... 13
PV Decimal.............................................................................................................................................. 13
Total Decimal........................................................................................................................................... 13
LCD Rotate.............................................................................................................................................. 13
Noise Limit................................................................................................................................................13
4.4 (SYS) Setting.................................................................................................................................... 14
Signal...................................................................................................................................................... 14
Pulse output............................................................................................................................................. 14
RS485..................................................................................................................................................... 15
Output....................................................................................................................................................... 15
Load Settings...........................................................................................................................................15
4.5 TRIM................................................................................................................................................... 16
3.1Zero Trim.............................................................................................................................................16
3.2Tube Trim............................................................................................................................................16
3.3Loop Trim............................................................................................................................................16
3.4K Character........................................................................................................................................ 17
4.6 CHK.................................................................................................................................................... 17
4.1Loop Test............................................................................................................................................ 17
4.2Pulse Test........................................................................................................................................... 17
4.3Coil Test.............................................................................................................................................. 17
5. Self-diagnosis Information and Fault Handling..................................................................................... 18
5.1 Fault handling................................................................................................................................... 18
6. Detail Setup................................................................................................................................................. 18
Size mm....................................................................................................................................................20
Qmax (m3/h)............................................................................................................................................ 20
Liter/Pulse................................................................................................................................................ 20

-2-
Sensor K...................................................................................................................................................20
Converter K..............................................................................................................................................20
Revise....................................................................................................................................................... 20
Coil Freq................................................................................................................................................... 22
Power Freq...............................................................................................................................................22
Act Zero mV............................................................................................................................................. 22
AZ Range mV.......................................................................................................................................... 22
Save Settings...........................................................................................................................................22
Password Protect....................................................................................................................................22
Clear Total.............................................................................................................................................. 22
FWD Preset (m3).................................................................................................................................. 22
REV Preset (m3)...................................................................................................................................22
7. Electromagnetic Flowmeter Installation..................................................................................................22
Appendix 1 About Password...................................................................................................................... - 1 -
 SYSTEM SETUP...................................................................................................................................- 1 -
New Password....................................................................................................................................- 1 -
Clear total.............................................................................................................................................. - 1 -
 TRIM........................................................................................................................................................ - 1 -
K Character......................................................................................................................................... - 1 -
 DETAIL SETUP......................................................................................................................................- 1 -
New Password......................................................................................................................................- 1 -
New Password......................................................................................................................................- 1 -
Appendix 2 Frequency Output Information..............................................................................................- 2 -

-3-
I. QTLDZ Electromagnetic Flowmeter Converter Operation
1. Product Function Introduction
1.1 Basic Function

 Executive Standard: JB/T9248-1999 GB 3836.1-2010


GB 3836.2-2010 GB 3836.4-2010
 Current speed range:0.03m/s~12m/s
 Repeatability: ±0.1%
 Display
English displaying mode can display the instantaneous flow, the percentage of flow, the total
forward flow and reverse flow
 Output signal:
Analog output:4~20mA Load resistance:<750Ω
Frequency: 0~5kHz 24V
Pulse
RS485 communication
HART(selectable)
 Excitation mode
Three amplitude low frequency excitation
High frequency excitation
 Excitation Frequency
1/16、1/8、1/4、1/2 can be set
 Excitation Electricity
160mA or specified
 Time Constant
0.1~99.9 Second can be set
 Ex-proof Mark
Exd[ia]iaIICT5

1.2 Normal operating conditions


 Ambient Temperature Ranges: -20℃~+55℃
 Relative Humidity:5%~90%
 Power supply: 85~265AC or 18~36VDC Active
 Power consumption:<10W

4
2. Converter Exterior Size and Power Connection Instruction
2.1 Exterior Size

2.2 Power Supply Connection

Note: When connect power wires, please confirm the power supply type first. Please never insert AC
power to DC converter. If connection wrong,it’ll permanent damage the converter!

Power Connection Instruction


NO. Mark Function Remark
L L AC 85~265v Power supply L is for Ac220v Power supply
N N AC 85~265v Power supply N is for Ac220v Power supply
+ + DC 18~36v Power supply+ DC 24v +
- - DC 18~36v Power supply- DC 24v -

5
2.3 Output Signal Terminal Connection
2.3.1 Current Output
Instruction:
a) Analog output signal connection:4~20mA,Max load resistance:500Ω
b) Pulse Output Connection: Frequency: 0~5kHz,V value:24V, Load current: 50mA

Note1: When connect output signal wire, please note the function of insertion cable. (eg.Electricity
Frequency or Communication).Please definitely distinguish the output type (eg. Active output or Passive
output)
Note2: When connection the wire,please never connect the power to the output terminal.Or else it may
permanent damage the converter!

6
Power Connection Terminal Instruction
NO. Mark Function Remark
1 + 4~20mA Output + Load Resistance≤500Ω
4~20mA Output - When use Hart communication,please use
2 -
Exterior Power 24VDC
3 + Frequency or Pulse Output + The Amplitude of Frequency or Pulse output
4 - Frequency or Pulse Output - is +24v,Load current ≤50mA
5 + RS-485 Data +
6 - RS-485 Data -

2.3.2 Sensor Signal Connection:

Note:When connect sensor signal,please don’t mixed the signal wire A B C and the Excitation wire X Y. If
connection wrong,it’ll damage the converter permanent.

Excitation and Input Signal Connection Terminal Instruction


NO. Mark Function Remark
1 X Excitation X
Load Resistance 10~100Ω
2 Y Excitation Y
3 A Input signal A
4 C Input signal common terminal
5 B Input signal B

7
3.Operation Converter

3.1 Key and Display

3.2 Menu Structure

Parameter Set in Measuring Mode


Press “C/CE” key, it’ll show as picture in right
Choose“C/CE”to enter the menu:
(BAS)
- 8 8 8 8 .8
Enter Setup ?
(SYS)
YES:C/CE NO:
(TRIM)
(CHK)
Press“→” to ESC REVISION V8.10

C/CE

8
1. Damping(s)
(0.1~99.1)
BAS 2. PV Decimal
BASIC SETUP (1、2、3)
3. Total Decimal
(1、2、3)
4.LCD Rotate
(0、+90、180、-90)

No.: 0000-9999
Default value=0200
1. New Password can set new password as you
like
Select Item: CHINESE ENGLISH
2. Language Default value= CHINESE

1 Qmax(m3/h)
2 Low Cutoff(%)
3. Signal 3 Direction
4 Max Limit(%)
SYS 5 Limit Time(s)
SYSTEM SETUP 6 Direction
7 Indication

1 Freq Max(Hz)
2 Liter/pulse
4. Pulse Output
3 Pulsewidth(ms)

1 Protocol Communication
5. MODBUS Output mode
2 Baud rate
3 Parity Calibration mode
4 Dev Address Device address
No.: 0000-9999
6. Clear Total
Default value=0020
can set new password as you
like
7.Load Settings 1. Empty Trim: Empty tube Trim
2. Full Trim: Full tube Trim
3. Tube Region %:Tube
sensitivity %

9
1. Zero Trim
4. Empty Trim
2. Tube Trim 5. Full Trim
TRIM
TRANSMITTER TRIM 6. Tube Region %
7. Empty Freq
8. Full Freq

3. Loop Trim 1. 4mA Trim


2. 20mA Trim
3. Loop Mode

4. K Character

CHK
1. Loop Test
OUTPUT CHECK
2. Pulse Test
3. Coil Test

4. Parameter Set and Use


4.1. Zero Trim
When use electromagnetic flowmeter, should do Zero Trim in order to get accurate measure result.
“Zero Trim” means to adjust the working Zero point when set current actual flow as Zero. While doing Zero
Trim, please make sure the medium is static and full of the measuring tube.

4.1.1 Quick Zero Trim


 Condition:the measuring tube is full of medium and medium is static
 Press “↓”and then“→”,the converter come to Zero Trim Program
 LCD display as: -0.868
Zero Trim ?
YES:C/CE NO:

REVISION V8.10

 If Press“→”,Converter will cancel Zero Trim Mode and exit to Measuring Mode.
C/CE
10
 If Press“C/CE”,Converter come to Zero Trip Mode and will display as:

-0.868
Zero Triming
-2.2

REVISION V8.10

C/CE

 After Zero Trim,Converter will automatically come back to Measure Mode.

-8888.8
115.8% m/h
3

82345678.8
TOTAL + m3
OVER LIMIT !
REVISION V8.10

C/CE

4.1.2 Zero Trim in the process of flowmeter calibration


 Condition:the measuring tube is full of medium and medium is static
 First come to ‘Zero Trim”:

 Press“C/CE”,then also press“C/CE”,


 enter set up as Picture 1 in right:

 Press“→”,choose“TRIM”
 Press“↓”to“Zero Trim”as picture 2
 Press“→” as picture 3

 Press“→”or “↓”,choose “yes”


 Press“→”enter to picture 4

 Press“C/CE”,come to“Zero Trim”double confirm as picture 5


 If press“→”,will exit“Zero Trim”.
 “Zero Trim”double confirm interface as picture 5
 Choose“C/CE”,will execute “Zero Trim”
 Choose“→”will exit “Zero Trim”

11
 Execute“Zero Trim” as Picture 6
After adjust zero point,converter
Will come back to measure mode.

4.2 Fast Setting


4.2.1 Change Unit

C/CE


→ →

↓ : change instant ↓ : change total flow ↓: change total flow


flow unit direction unit

Instant Flow Unit: m3/s,m3/m,m3/h,m3/d,gal/s,gal/m,gal/h,gal/d,ft3/s, ft3/m,ft3/h,ft3/d,


Igal/s,Igal/m,Igal/m,Igal/h,Igal/d,bbl/s,bbl/m,bbl/h,bbl/d,Mgal/d,ML/d,ml/s,ml/m,ml/h,ml/d,kg/s,kg/m,kg/h,k
g/d,t/s,t/m,t/h,t/d,STon/m,STon/h,STon/d,LTon/h,LTon/d,lb/s,lb/m,lb/h,lb/d,L/s,L/m,L/h

Total Flow Unit: m3,gal,ft3,Igal,bbl,in3,hL,Mgal,kg,t,STon,LTon,lb,L

4.2.2 Change contrast ratio


Under Measurement mode

Press→ for 5 second

Press C/CE

↓:reduce
→:increase
C/CE:record and ESC

12
4.3 (BAS) Setting
BASIC SETUP (BAS)

Floating number: 99.9 – 0.1


Damping (s) Default = 1.0

Options: 1 2 3
PV Decimal
Default = 1

Options: 1 2 3
Total Decimal
Default = 1

Options: 0 +90 180 -90


LCD Rotate
Default = 0

Options: No Yes
Noise Limit Default = YES
When there’s big flow fluctuation, can set to “YES” to suppression.

13
4.4 (SYS) Setting
SYSTEM SETUP (SYS)

Floating number:max = flow at 15m/s velocity


min = flow at 0.2m/s velocity
Q=(velocity*D*D)/353.67765
For example size=100mm,then:
max = 424.115 m3/h min = 5.655 m3/h
Qmax (m3/h) Default = 100.0
Qmax means when instant flow rate reach Qmax, analog
output is 20mA; Frequency output is the setting value set in
“ Frequency Upper Hz ” .
If change this parameter, analog output and frequency output
will be affected.

Floating number: 9.9 – 0.0 %


Default = 1.0 %
Low cutoff %
When absolute value of instant flow rate is less than Qmax * this
setting value, then instant flow rate will be 0.

Signal Options: Forward, Reverse, bi-direction


Default = Bi-direction
If set as forward, do not calculate and display reverse flow
Direction
If set as reverse, do not calculate and display forward flow
If set as bi-direction, calculate and display both foward and reverse

Options: Forward, Reverse


Default = Forward
Indication
Indication flow meter installation direction.
Set as forward, flow* 1.0
Set as reverse, flow*-1.0

Floating number: 9.999 – 0.1


Density (g/mL) Default = 1.0
When flow unit choose mass flow, density value will be calculated.

Floating number: 5000.0 – 100.0


Default = 2000.0
Freq Max (Hz) Frequency output value when Qmax
Output frequency (Hz) = (current flow rate(m3/h)/ Qmax
(m3/h))×upper frequency limit (Hz)

Floating number: xxx – 0.0


Default = 0.0
Set value=0, frequency output as “ upper frequency limit ”HZ
output
Liter/pulse
Set value >0,
Pulse output
Output (Hz) = [instant flow rate (m3/h)/3.6]/ pulse (L/p)
=instant flow rate (L/s)/ pulse (L/p)
Max frequency output: 5000.0Hz

Floating number: 1000.0 – 0.0


Default = 0.0
Set value=0,output square wave;
Set value>0,pulse width as the set value; If set pulse width > 50%
PulseWidth (ms) of actual pulse output period, automatically reduce pulse width to
the 50% of actual pulse output period.
For example, “pulse width” is 100ms, but at this time actual pulse
frequency=500Hz, then at this time actual pulse period=2ms, then
actual pulse output width=1ms

14
Options: Active L Active H
Default = Active L
Level type of “pulse width” set value in one pulse output period.
Pulse Level For example,pulse level=active L,pulse width=1ms
Pulse frequency=100Hz,
Then at this time pulse period =10ms,include
High level time =9ms,low level time =1ms

Options: INT EXT


Default =INT
Choose power supply when use pulse output.
Pulse Power INT:Power is supplied by itself, output 0-15V pulse directly.
EXT : Inner power cut off, open-collector output, need external
power supply and pull-up resistor, then can have pulse output.

Selection: MODBUS-RTU MODBUS-ASCII


RS485 Protocol
Default = MODBUS-RTU

Options: 1200 2400 4800


Baud Rate 9600 19200 38400
Default = 9600

Selection: 7 8
RS485 Data Bit Default = 8
Under RTU protocol,can not choose 7 bit.
Output
Options: NONE EVEN ODD
Parity
Default = NONE

Options: 1 2
Stop bit
Default = 1

Number: 247 – 1
Dev address
Default = 1

Options: No Yes
Clear Total Default = No
Clear forward and reverse total flow

Options: No Yes
Default = No
Load Settings
If choose YES, then restore factory settings;
All the parameters will be back to the factory settings.

15
4.5 TRIM
TRIM

Options: No Yes
3.1 Default = No
Zero Trim Make sure pipe is full and liquid is in static condition, then use this
Function To make flow meter do zero trim automatically.

Options: No Yes
Default = No
3.2.1
Make sure pipe is empty, then use this
Empty Trim
function,flow meter will record pipe character
value in empty pipe condition.

Options: No Yes
Default = No
3.2.2
Make sure pipe is full, then use this function,
Full Trim
flow meter will record pipe character value in full
pipe condition.

Floating number: 99.9 – 0.0


Default = 40.0
3.2.3 Empty pipe detection sensitivity, suggest to set
Tube Region % 40.0-50.0. If set as 0.0, empty pipe detection is
3.2
forbidden. The value of Tube Region is bigger means
Tube Trim
that empty tube region is more sensitive.

Floating number: 9999.0 –0.0


Default = 1960.0
The value get when do empty pipe frequency trim.
3.2.4 Do not suggest to change this value.
Empty Freq This parameter provide fine adjustment and check, and
the manual recovery of when wrong operation of empty
trim.

Floating number: 9999.0 –0.0


Default = 220.0
The value get when do full pipe frequency trim.
3.2.5
Full Freq Do not suggest to change this value.
This parameter provide fine adjustment and check, and
the manual recovery of when wrong operation of full
trim.

Floating number: 4.5 – 3.5


Default = 4.0

When execute this function, and use accurate ampere


3.3 3.3.1
Loop Trim 4mA Trim meter to measure 4-20mA output. Input the reading
value to the converter, and converter will automatically
achieve trim work.

16
Floating number: 21.5– 19.5
Default = 20.0

When execute this function, and use accurate ampere


3.3.2
20mA Trim meter to measure 4-20mA output. Input the reading
value to the converter, and converter will automatically
achieve trim work.

Options: 4-20mA 4-12-20mA


Default = 4-20mA
4-20mA mode:
4mA(when flow =0.0)
3.3.3
20mA(when flow=Qmax)
Loop Mode
4-12-20mA mode:
4mA(when flow = -Qmax)
12mA(when flow =0.0)
20mA(when flow = +Qmax)

Floating number: 1.03 –0.97


Default = 1.0
3.4 This function is used for flow meter Quadratic periodic test, when test result is very
K Character bad, can use this parameter to do adjustment. Changed parameter value and the
times of adjustment will be recorded.
K Character = 1.0 - test meter relative error

4.6 CHK

Output Check (CHK)

Floating number: 20.0 – 4.0


Default = 12.0
4.1
Loop Test User this function, and use ampere meter at the same time to check 4-20mA output.
Change the set value in allowed range, test the deviation between output value and
the set value.

Floating number: 5000.0 – 1.0


Default = 1000.0
4.2
Pulse Test Execute this function and use accurate cymometer to measure frequency output.
Change present set value in the allowed range to test the deviation between output
value and set value.

Options: OFF X->Y Y->X


Default = OFF
OFF: normal excitation drive
X->Y : Excitation one-way drive X->Y , connect ammeter to excitation circuit in
concatenation way, measure the exciting current;
4.3
Coil Test
Y->X : Excitation one-way drive Y->X , connect ammeter to excitation circuit in
concatenation way,measure the exciting current;

Use this function can check whether the excitation forward/reverse drive current is
consistent.

17
5. Self-diagnosis Information and Fault Handling
Electromagnetic flowmeter converter have self-diagnosis function. It can provide correct alarm information by itself for most
general fault except power supply and hardware circuit fault. Fault Information will display in the bottom right of the LCD
screen.

5.1 Fault handling


5.1.1 No display on the screen
a) Check whether the power supply is switched on
b) Check whether the power fuse is intact
c) Check whether the power supply voltage is conform to the requirements
If above information is normal, please contact with manufacturer.

5.1.2 Excitation Alarm


a) Check whether the Excitation line X and Y are open circuit
b) Check whether the Excitation coil resistance is normal, or else it means that converter may
has fault.

5.1.3 Empty Alarm


a) Check if the liquid medium is full of measuring tube
b) Use wire to make signal input terminal A,B and C short circuit. At this moment if ‘Empty
Alarm’ indicate revocation, it means that converter is normal. This fault may caused by low
conductivity of measuring medium or the setup error of empty tube threshold value and empty tube
measuring range.
c) Check whether signal line connection are correct
d) Check whether sensor electrodes are normal

5.1.4 Inaccuracy
a) Check whether the liquid is full of the measuring tube
b) Check whether the signal line connection are normal
c) Check whether the sensor coefficient,zero point is normal or Calibration setting is normal.

6. Detail Setup
Noted: Detail setup is not show in normal measurement mode. Do not suggest to check and change by
end user without professional instruction.

 Enter Detail Setup Interface

1. Long Press“C/CE”for more than 5 seconds,then enter detail setup interface:

18
2. Press the second key ,then press to choose YES to enter DETAIL SETUP
Interface.

 Return to the normal measure interface

After “Save Settings”,long press for more than 5 seconds to return to the normal measure
Interface as below:

19
Detail Setup

Option: Max 3000mm


Size mm
Default = 100mm

Floating number:
max = flow at 15m/s velocity
min = flow at 0.2m/s velocity
Q=(velocity*D*D)/353.67765
For example size=100mm,then:
Qmax (m3/h)
max = 424.115 m3/h min = 5.655 m3/h
Default = 100.0
Qmax means when instant flow rate reach Qmax, analog output is 20mA;
Frequency output is the setting value set in “ Frequency Upper Hz ” .
If change this parameter, analog output and frequency output will be affected.

Floating number: xxx – 0.0


Default = 0.0
Set value=0, frequency output as “ upper frequency limit ”HZ output
Liter/Pulse Set value >0,
Output (Hz) = [instant flow rate (m3/h)/3.6]/ pulse (L/p)
=instant flow rate (L/s)/ pulse (L/p)
Max frequency output: 5000.0Hz

Floating number: 99.9 – 0.01


Default = 0.16
Sensor K This parameter will get after flow calibration
New Sensor K =
Old Sensor K/(tested meter value/ standard meter value)

Floating number: 99.9 – 0.01


Converter K Default = 8.0
This parameter only have relationship with converter, do not suggest user to
change!

Floating number: +19.0 – -19.0


Default = 0.0
Used to revise error of flow meter lower flow rate
PresetZero=(QvL×Ei)
5.10.1 QvL - voltage value at a certain flow point, unit:mV
PresetZero(mV) Ei - relative error at a certian flow point
Unit :%
For example : when flow velocity=0.15m/s, voltage
=15mV, relative error=2%,Then
PresetZero=15.0×0.02=0.3mV

Floating number: 3.0 – 0.0


Revise 5.10.2
PZ Range Lower Default = 1.0
Lower flow rate under PresetZero Range

Floating number: 3.0 – 0.0


5.10.3
PZ Range Upper Default = 2.0
Upper flow rate under PresetZero Range

5.10.4 Floating number: 15.0 –0.0


Speed_1 m/s Default = 0.1

5.10.5 Floating number: 1.5 –0.5


Revise_1 Default = 1.0

20
Floating number: 15.0 –0.0
5.10.6
Speed_2 m/s Default = 0.2
This set value should > set value in Speed_1m/s

5.10.7 Floating number: 1.5 –0.5


Revise_2 Default = 1.0

Floating number: 15.0 –0.0


5.10.8
Speed_3 m/s Default = 0.3
This set value should > set value in Speed_2m/s

5.10.9 Floating number: 1.5 –0.5


Revise_3 Default = 1.0

Floating number: 15.0 –0.0


5.10.10
Speed_4 m/s Default = 0.4
This set value should > set value in Speed_3m/s

5.10.11 Floating number: 1.5 –0.5


Revise_4 Default = 1.0

Floating number: 15.0 –0.0


5.10.12
Speed_5 m/s Default = 0.5
This set value should > set value in Speed_4m/s

5.10.13 Floating number: 1.5 –0.5


Revise_5 Default = 1.0

Floating number: 15.0 –0.0


5.10.14
Speed_6 m/s Default = 0.6
This set value should > set value in Speed_5m/s

5.10.15 Floating number: 1.5 –0.5


Revise_6 Default = 1.0

Floating number: 15.0 –0.0


5.10.16
Speed_7 m/s Default = 0.7
This set value should > set value in Speed_6m/s

5.10.17 Floating number: 1.5 –0.5


Revise_7 Default = 1.0

Floating number: 15.0 –0.0


5.10.18
Speed_8 m/s Default = 0.8
This set value should > set value in Speed_7m/s

5.10.19 Floating number: 1.5 –0.5


Revise_8 Default = 1.0

Floating number: 15.0 –0.0


5.10.20
Speed_9 m/s Default = 0.9
This set value should > set value in Speed_8m/s

5.10.21 Floating number: 1.5 –0.5


Revise_9 Default = 1.0

Floating number: 15.0 –0.0


5.10.22
Speed_10 m/s Default = 1.0
This set value should > set value in Speed_9m/s

5.10.23 Floating number: 1.5 –0.5


Revise_10 Default = 1.0

21
Floating number: 15.0 –0.0
5.10.24
Speed_11 m/s Default = 1.1
This set value should > set value in Speed_10m/s

5.10.25 Floating number: 1.5 –0.5


Revise_11 Default = 1.0

Options: 3.125Hz 6.25Hz 12.5Hz 25Hz


Coil Freq
Default = 12.5Hz

Options: 50Hz 60Hz


Power Freq
Default = 50Hz

Options: +99.99 – -99.99


Default = 0.0
Zero value get while doing zero trim.
Act Zero mV Do not suggest to change this value.
This parameter provide fine adjustment and check, and the manual recovery of
when wrong operation of zero trim.

Floating number: 99.0 –0.0


AZ Range mV Default = 20.0
Restrict sensor zero mv range,when |zero mv|> set value, then do
not allow to do adjustment

Options: No Yes
Default = No
Save Settings If choose “YES”, then will record factory settings, all the parameters will be
recorded. If the parameter in converter been damaged, then can use this
parameter to record factory settings.

Options: No Yes
Default = No
If choose yes, some menu which is under password protect will need to
Password Protect
Input password, then can enter.
After set “Yes”, Menu which need password to enter, pls check appendix
information.

TOTAL

Options: No Yes
Default = No
Clear Total
Clear forward and reverse total flow

Floating number: 9999999999 – 0


FWD Preset (m3) Default = 0
Can preset forward total flow value by this function.

Floating number: 9999999999 – 0


Default = 0
REV Preset (m3)
Can preset reverse total flow value by this function.

7. Electromagnetic Flowmeter Installation

In order to let you can correct measuring, please be pay attention to the following requirements when
22
choosing the line position.

2.1 The sensor not only installed in the straight pipe, but also can be installation the in level or tilted

pipe,but the required of the center of two electrode attachment on the horizontal state

2.2 the medium in the installation position should be full pipe flow, avoid not full pipe and gas attached to

the electrode.

2.3 For liquid-solid two-phase flow,it better choose the perpendicular installation,it make the liner

material of the sensor attrition uniform,and prolong working life.

2.4 when the flow meter installation position with not full medium, it can raise the end pipeline of the

flowmeter, and make its full tube, forbidden in peak and outlet pipe installation flowmeter

2.5 Modify piping installation method: when the medium velocity out of requires,it should choose a small

caliber flow meter, and should use different caliber taper pipe or modify the sections of pipe, make its same

caliber with sensor, but before and after straight pipe section at least requirement the before straight pipe

section ≥10 DN, after straight pipe section≥5 DN (DN for pipe diameter)

23
2.6 Before and after straight pipe section for the flowmeter ,before pipe section ≥10 DN, after pipe

section≥5 DN

Right Suspending

Dont’ install it at the entrance of pump, install it at the exit of pump

Install exhaust valve at the downstreat of flow meter when drop is more than 5m

Installed at the lowest point when used in open drain pipe

Need 10D of upstreat and 5D of downstreat

Installed at the rising direction

24
Pipe must be full of Medium

Avoid air bladder

Straight Pipe Requirement

25
Appendix 1 About Password
If open “password protect” menu in DETAIL SETUP, Menu which are under password protect and default password
are as below:

 SYSTEM SETUP

(SYS)

Number: 0000 - 9999


New Password
Default = 0200

Number: 0000 - 9999


New password
Default = 0020
Clear total
Options: No Yes
Clear total Default = No
Clear forward and reverse total flow

 TRIM

Number: 0000 – 9999


New Password
Default = 0003

Floating number: 1.03 –0.97


Default = 1.0
K Character This function is used for flow meter Quadratic periodic test,
K Character when test result is very bad, can use this parameter to do
adjustment. Changed parameter value and the times of
adjustment will be recorded.
K Character = 1.0 - test meter relative error

 DETAIL SETUP

Detail Setup

Number: 0000 - 9999


New Password
Default = 1111

TOTAL

Number: 0000 - 9999


New Password
Default = 0020

-1-
Appendix 2 Frequency Output Information
There’re two types of frequency output: active output and passive output. Default is active output.
Active output as below:

Internal External
Imax=2.4mA

24V
10kΩ

F+

F-

Passive ouput as below:

Internal External

Rf
F+
VE

F-

Inner drive max can provide 200mA current.


Thus, require VE / Rf < Imax(200mA)