Professional Documents
Culture Documents
com
National Center for Scientific Research “Demokritos”, GR-15310 Aghia Paraskevi Attikis, Greece
Received 13 November 2007; received in revised form 15 January 2008; accepted 17 January 2008
Available online 26 January 2008
Abstract
The phase variation method is described and applied in order to determine liquid/gas partition coefficients of monomers in aqueous solutions
as well as the effect of the presence of one monomer to the partition coefficient of the other. The method is based on the relationship between
reciprocal peak area and the phase ratio in the vial containing the sample solution. It involves regression analysis of headspace gas chromatography
measurements of a number of sample vials with a variation of phase ratios. The method is applied in the case of the monomers butyl acrylate and
methyl methacrylate. Experimental data for the partition coefficient of the two acrylates are modeled with the Statistical-Associating-Fluid Theory
(SAFT).
© 2008 Elsevier B.V. All rights reserved.
0378-3812/$ – see front matter © 2008 Elsevier B.V. All rights reserved.
doi:10.1016/j.fluid.2008.01.015
22 Z. Kechagia et al. / Fluid Phase Equilibria 266 (2008) 21–30
Table 3
Monomers and internal standard tested
Monomer Internal standard tested
Fig. 1. Schematic of three calibration data sets, where i is the concentration level, and j, k are the experimental and instrumental replicates, respectively.
Fig. 2. Schematic of the different sources of error in chemical calibration using linear least-squares regression.
The symbols are explained at the end of the paper and the
various terms are given elsewhere and are not repeated here
[31–33].
The analytical signals obtained at the different experimental
and instrumental replicates for BuA are presented in Table 4,
and the calibration curve obtained for BuA is depicted in Fig. 3
and for MMA in Fig. 4, respectively.
Table 4
Analytical signals obtained for the center-weighted calibration
Nominal concentration level i (ppm) Experimental replicate j Instrumental replicate k Signal (monomer/IS area)
200 1 1 0.1485
2 1 0.1501
3 1 0.1197
400 1 2 0.3904, 0.3635
2 2 0.3759, 0.3868
600 1 2 0.5910, 0.5953
800 1 2 0.8742, 0.8614
2 2 0.8849, 0.8707
1000 1 1 0.9387
2 1 0.9849
3 1 1.0615
unlike molecules. Mathematically, SAFT is expressed by the v00 is the spherical segment size, u0 /k is the dispersion energy
following equation for the residual Helmholtz free energy per parameter, εhb /k is the association energy and κ is the association
mole: volume. All five parameters are fitted to saturated liquid density
and vapor pressure data from low temperature up to close to the
ares (T, ρ) a(T, ρ) aideal (T, ρ) aref (T, ρ) adisp (T, ρ)
= − = + critical point. In this work, water is modeled as a four hydrogen
RT RT RT RT RT bonding site fluid whereas both BuA and MMA are modeled as
ahs (T, ρ) achain (T, ρ) aassoc (T, ρ) adisp (T, ρ)
= + + + two hydrogen bonding site fluids. In Table 5, the pure component
RT RT RT RT parameters are shown.
(12)
SAFT extension to mixtures is straightforward based on the
where T and ρ are the temperature and density of the system, res- van der Waals one fluid theory [34]. For the cross-association
pectively and the residual Helmholtz free energy is with respect interactions (hydrogen bonding between unlike species), the
to the Helmholtz free energy for the ideal gas at the same T and following combining rules are used:
ρ. Superscripts hs, chain, assoc, and disp refer to hard-sphere
interactions, chain formation, association (hydrogen bonding), εAB εAB
hb εhb
BB κAA + κBB
hb
= κAB = (13)
and dispersion interactions, respectively. If the Helmholtz free k k k 2
energy of the fluid is known, all other thermodynamic proper-
where superscript AB denotes hydrogen bonding between a
ties, such as pressure, chemical potential, etc., can be calculated
molecule of species A and a molecule of species B. Further-
using standard thermodynamic equations. The expressions for
more, in the dispersion term for mixtures a binary interaction
the various terms in Eq. (12) are given elsewhere [33] and are
parameter, kij , is used that is fitted to binary phase equilibrium
not repeated here. SAFT is a five parameter EoS for pure com-
data:
ponents: m is the number of spherical segments per molecules, u u 1/2
uij ii jj
= (1 − kij ) (14)
k k k
The calibration curves obtained for BuA and MMA are depic-
ted in Figs. 3 and 4 respectively. Corner-weighted calibration
was used, as it is known to decrease experimental work without
loss of analytical information. The corresponding values of stan-
dard deviations and sums squares calculated for the different
sources of error are presented in Table 6. The overall error
is 5.3% and 4.7% for the BuA and MMA calibration curves,
respectively, which are acceptable percentages for this type of
measurements.
The partition coefficient measurements were carried out at
three different temperatures. At each temperature the solution
was analyzed from eight different vials, every two correspon-
ding to different phase ratios. A number of solution series were
Fig. 4. Corner-weighted calibration curve of MMA using the full evaporation used in order to verify the results. The sample phase fractions
technique and 2-pentanone as internal standard. as well as the volume of the vial used were determined accor-
26 Z. Kechagia et al. / Fluid Phase Equilibria 266 (2008) 21–30
Table 5
SAFT pure component parameters for the components examined here and percentage average absolute deviation (% AAD) between experimental data [40] and
SAFT correlation for the vapor pressure (Psat ) and saturated liquid density (ρliq ).
Component m v00 (cm3 /mol) u0 /k (K) εHB /k (K) κ T (K) Psat (% AAD) ρliq (% AAD)
ding to investigations by Kolb and Ettre [26], after ensuring chromatograph, due to changes in the external parameters, such
that the values selected in the present work both satisfy the as the pressure in the vial.
easily attainment of equilibrium and a representative sample The FET allows us to reduce the sample size and increase the
aliquot. temperature, thus reducing the effect of the sample matrix. It thus
A very large sample volume increases the time needed for ensures that salts and mineral acids will never enter and conta-
equilibrium, while it has to be enough in order to ensure a repre- minate the analytical instrumentation. Furthermore, the sample
sentative sample aliquot inserted in the column. The sample matrix does not affect the calibration of the analytical procedure,
volume in the vial was selected to range from 1 to 4 ml in order creating the potential for universal standardization. Even though
to obtain a significant concentration of the analyte in the head- in the specific system examined there is no matrix effect, the fact
space. It was found that larger sample volumes greatly increased that the response of each component remains uniform regardless
the time needed to reach equilibrium, without a significant effect of the matrix present, permits that the results be used in order to
on the quantitative analysis. examine the stripping process of BuA and MMA polymers and
Specifically the determination of the partition coefficients copolymers. This work is underway at the Aristotle University
depends on peak area differences, resulting from changing the of Thessaloniki group and will be presented soon.
phase ratio values. For a partition coefficient of K = 10 it was The use of liquid concentration measurements allows us also
calculated, using the phase ratio variation method’s equations, to determine components of low volatility, that will be present
that filling the vial with 4 ml of the sample the concentration in other systems examined. It is generally more convenient to
of the analyte in the headspace vial will be 2.15 times of that dispense microliters of sample into a headspace vial at room
when the sample occupies only 1 ml of the vial volume. This temperature than to provide fast injection through a small ope-
difference permits the calculation of the partition coefficients ning of a hot injection port. Furthermore when measurements
from the peak areas observed at the gas chromatograph. Fur- were carried out using the gas phase the results were of lower
thermore it was calculated that by changing the sample volume accuracy (especially in the case that no internal standard was
by 0.1 ml a variation of the concentration of the analyte in the used), even though care was taken that the carrier gas inlet pres-
headspace vial of 2.6% was observed thus providing a good sure was higher than the pressure reached in the headspace vial
sample to sample reproducibility, since balances and pipettes of and the filling sample loop time was carefully selected. On the
high accuracy were used (±0.1 mg and ±0.5 l, respectively). other hand, when the concentration was measured in the liquid
The size of the sample vial was selected to be 20 ml in order to phase more reproducible results were observed.
have enough headspace gas available for transfer into the column
as well as to obtain sample homogeneity and therefore analyze 5.1. Temperature dependence of the partition coefficient
a representative aliquot.
The measurements were carried out by the FET, using 2- It is known that the relationship between log K and the inverse
pentanone as internal standard, in order to minimize the effects of temperature, 1/T may be considered linear, even though a
of the quantity variations of the sample introduced in the gas slight curvature may be occasionally observed [5,35–37]. Typi-
cal of this slight curvature are the examples of the plots of
chlorobenzene, toluene and n-xylene in water [1,25]. Thus the
Table 6
Values of standard deviations and sums squares for BuA and MMA calculated
existence of such relationship between log K and 1/T is an indica-
for the different sources of error tion of the correctness of the measurements [38]. Determination
of the liquid–gas partition coefficient of an analyte at various
Standard deviations and sum squares BuA MMA
temperatures will permit extrapolation within a reasonable range
Instrumental error to other temperatures. This is especially important for the deter-
SSI 0.00061 0.00506 mination of partition coefficients values at temperatures higher
sI 0.00687 0.01837
than the maximum temperature of the thermostat of the head-
Procedural error space auto sampler (150 ◦ C).
SSP 0.00388 0.00119
The results of the present study are presented in the overall
sP 0.00645 0.00356
tables (Tables 7 and 8). A typical diagram from which the results
Overall error are derived is depicted in Fig. 5. The figure represents the ratio
SS0 0.03425 0.03033
s0 0.05341 0.04655
of the concentration of the initial vial to the equilibrium concen-
tration in the liquid phase versus the volume ratio of the vial
Table 7
Liquid/gas partition coefficients of BuA at various temperatures and various mass fractions in the initial solution
Temperature (K) Concentration ratio of Klg of BuA Klg of BuA Intercept b0 Slope b1 Standard Standard error 95% confidence interval 95% confidence Correlation
BuA:MMA in the initial (expt.) (SAFT) error, sy of estimate, sy,x of b0 interval of b1 factor R2
solution
333.15 0:100 – – – – – – – – –
20:80 6.52 9.09 0.6445 0.0988 0.622 0.080 [0.335, 0.644] [0.072, 0.125] 0.9833
30:70 7.06 9.09 0.6899 0.0977 0.610 0.071 [0.388, 0.992] [0.072, 0.124] 0.9867
50:50 7.33 9.10 0.9953 0.1357 0.969 0.083 [0.691, 1.137] [0.109, 0.162] 0.9927
80:20 9.23 9.10 1.0541 0.1141 0.732 0.081 [0.786, 1.323] [0.093, 0.136] 0.9880
100:0 9.49 9.10 1.0414 0.1097 0.748 0.051 [0.712, 1.371] [0.084, 0.136] 0.9953
343.15 0:100 – – – – – – – – –
20:80 5.12 8.53 0.7635 0.1489 0.996 0.101 [0.371, 1.156] [0.132, 0.165] 0.9897
30:70 5.76 8.53 1.1446 0.1986 1.096 0.124 [0.496, 1.793] [0.171, 0.227] 0.9872
50:50 5.90 8.54 1.0014 0.1696 1.102 0.088 [0.626, 1.376] [0.140, 0.199] 0.9935
80:20 8.34 8.54 1.2850 0.1539 0.894 0.100 [0.841, 1.729] [0.114, 0.193] 0.9874
Table 8
Liquid/gas partition coefficients of MMA at various temperatures and various mass fractions in the initial solution
Temperature (K) Concentration ratio of Klg of MMA Klg of MMA Intercept b0 Slope b1 Standard Standard error of 95% confidence interval 95% confidence Correlation
BuA:MMA in the initial (expt.) (SAFT) error, sy estimate, sy,x of b0 interval of b1 factor R2
solution
333.15 0:100 13.10 13.78 0.7954 0.0607 0.435 0.030 [0.547, 1.044] [0.042, 0.080] 0.9952
20:80 11.95 13.79 0.9369 0.0784 0.493 0.055 [0.692, 1.182] [0.057, 0.099] 0.9876
30:70 11.04 13.79 0.8720 0.0790 0.494 0.037 [0.628, 1.116] [0.058, 0.100] 0.9943
50:50 9.51 13.79 0.7273 0.0765 0.530 0.040 [0.521, 0.933] [0.063, 0.091] 0.9944
80:20 8.30 13.80 0.8810 0.1061 0.681 0.068 [0.631, 1.131] [0.086, 0.126] 0.9899
100:0 – – – – – – – – –
343.15 0:100 11.97 12.48 1.6612 0.1387 1.026 0.152 [1.188, 2.135] [0.102, 0.176] 0.9781
20:80 11.08 12.48 1.4133 0.1275 0.851 0.092 [1.085, 1.755] [0.113, 0.141] 0.9881
30:70 9.34 12.48 1.3354 0.1429 0.789 0.062 [0.869, 1.802] [0.123, 0.163] 0.9939
50:50 7.82 12.49 1.1026 0.1412 0.918 0.117 [0.793, 1.417] [0.117, 0.166] 0.9850
80:20 7.55 12.49 1.2250 0.1621 0.944 0.098 [0.728, 1.721] [0.119, 0.205] 0.9891
100:0 – – – – – – – – –
353.15 0:100 10.86 11.42 1.1465 0.1056 0.613 0.079 [0.817, 1.476] [0.092, 0.119] 0.9834
20:80 10.29 11.42 1.3084 0.1272 0.796 0.130 [0.946, 1.669] [0.112, 0.142] 0.9696
30:70 8.38 11.42 1.2139 0.1449 0.978 0.092 [0.910, 1.518] [0.121, 0.169] 0.9910
50:50 6.65 11.43 1.2433 0.1869 1.157 0.093 [0.877, 1.609] [0.173, 0.200] 0.9935
80:20 6.30 11.43 1.3244 0.2103 1.368 0.131 [0.793, 1.856] [0.169, 0.252] 0.9907
100:0 – – – – – – – – –
27
28 Z. Kechagia et al. / Fluid Phase Equilibria 266 (2008) 21–30
Fig. 5. Concentration ratios of BuA in the liquid phase at the initial solution and
at equilibrium vs. volume ratio of the vial. The concentration ratio of BuA/MMA Fig. 7. Liquid/gas partition coefficients of MMA vs. mass fractions of MMA
in the initial solution is equal to 30/70. (mass of MMA/mass of BuA + MMA) in the initial solution for various tempe-
ratures.
[24] J. Drozd, J. Novak, J. Chromatogr.:Chromatogr. Rev. 165 (1979) 141. [34] S.H. Huang, M. Radosz, Ind. Eng. Chem. Res. 30 (1991) 1994.
[25] L.S. Ettre, C. Welter, B. Kolb, Chromatographia 35 (1993) 73. [35] R. Keymeulen, B. Parewijck, A. Gorna-Binkul, H. Van Langenhove, J.
[26] L.S. Ettre, B. Kolb, Chromatographia 32 (1991) 5. Chromatogr. A 765 (1997) 247.
[27] X.S. Chai, J.Y. Zhu, J. Chromatogr. A 799 (1998) 207. [36] A.-M. Bakierowska, J. Trzeszczynski, Fluid Phase Equilibr. 213 (2003)
[28] M. Markelov, J.P. Guzowski, Anal. Chim. Acta 276 (1993) 235. 139.
[29] J. Schuberth, Anal. Chem. 68 (1996) 1317. [37] C. Bichii, C. Cordero, E. Liberto, P. Rubiolo, B. Sgorbini, P. Sandra, J.
[30] A. Gonzalez-Casado, L. Cuadros Rodriguez, A.M. Garcia Campana, R. Chromatogr. A 1 (2005) 111.
Blanc Garcia, LC–GC Int. (1998) 726. [38] T. Harner, D. Mackay, Environ. Sci. Technol. 29 (1995) 1599.
[31] L.C. Rodriguez, A.M. Garcia Campana, C.J. Linaros, M.R. Ceba, Anal. [39] V.J. Beck, K.J. Arnold, Parameter Estimation in Engineering and Science,
Lett. 26 (1993) 1243. Wiley Series in Probability and Mathematics, 1976.
[32] L.C. Rodriguez, A.M. Garcia Campana, J.M. Bosque Sendra, Anal. Lett. [40] T.E. Daubert, R.P. Danner, Physical and Thermodynamic Properties of Pure
29 (1996) 1231. Compounds: Data Compilation, Hemisphere, New York, 2001.
[33] S.H. Huang, M. Radosz, Ind. Eng. Chem. Res. 29 (1990) 2284.