MANUAL
E-57
Catalog
2
E-57
CHAPTER 1 Summary
The XK3100-B2 Indicator is a new product of our company. It is a kind of general weighing
indicator. It adopted the AD IC of famous semiconductor factory in USA. It has high precision, high
stability, high credibility, zero code adjust and widely of the input signal, upper and lower limit setting
and alarm, peak test (selectable). The indicator can be made many kinds of electronic scale connected
with load cells easily such as platform scale, ground scale, vehicle scale, oil scale and animal scale. It
is a perfect indicator which can be used in business,industry and etc.
Main Function of XK3100-B2:
1. With AC/DC power supply, 4V 4Ah no-maintenance battery and high efficiency recharging circuit
2. Low power cost, can be used for 200 hours after recharging(only with one load cell and not
open the backlight)
3. Auto adjust of zero code (tare automatically), input signal range: - 6 mv ----- +21 mv
4. Calibrate with digital keyboard (locked by hardware and software)
5. 6 digits LCD display, 24 mm high of symbol, backlight can be selected Zero auto track
6. Several divisions and point selecting, calibrate with any point
7. Low voltage check, calibration self-check, overload warning and error information
3
E-57
CHAPTER 3. Installation
1. The front of indicator
Keyboard
Display
Figure 3-1
2. The back of indicator
Power plug
Load cell interface
Figure 3-2
3. Load cell connected with indicator
+V ① ⑤ Shield
-V ② ④ -IN
③ +IN
Figure 3-3
Note: +V, -V (1, 2): Load cell’s excitation
+S, -S (1, 2): Load cell’s feedback signal
+IN, -IN (3,4): Load cell’s output signal
Shield (5): the load cell’s shield cable (connected with GND of AC plug)
4
E-57
CHAPTER 4. Calibration
When you calibrate the indicator, the load cell’s shield cable must connected with GND, otherwise the
indicator reads will have errors. (Figure 3-3)
Calibration should be operated by measure technician is as follows:
1. Weight calibration
STEP OPERATION DISPLAY NOTICE
Power on, press “calibration” and “ Enter” together when self-checking, loosen when hear
1
the buzzer.
【pass】 It will display 【 pass 】 after Self checking and ask to input
2
calibration password
Input calibration 【------1】 Press 【 ↑ 】 to input data , press 【 Enter 】 to confirm;
password “010201” press【Tare】to modify last bit of password or exit calibration.
3
Press【Enter】 【d= xx】 Password right to enter division setting, otherwise exit
calibration
【d= xx】 Division setting
press【 ↑ 】
4 【d= yy】 (1, 2, 5, 10, 20, 50, 100, 200); Press【Enter】 to confirm
press【Enter】
【dot= x】 Confirm the division and enter point setting
【dot= x】 Point setting
5 press【 ↑ 】 【dot= y】 Press【↑】to modify decimal digits, at best 5 digits
press【Enter】 【ZEroxx】 confirm the decimal digits and enter zero range setting
Zero range setting
【ZEroxx】
press【 ↑ 】 Press 【 ↑ 】 to modify the zero range, the most zero tracing
6 【Zeroyy】
press【Enter】 range will be 3d ;Press【Enter】 to confirm
【 Full 】
Confirm the setting and enter capacity setting
【 Full 】 Full capacity setting (DOS prompt and original setting display
Press【 ↑ 】 in turn)
7 Press【Enter】 【 xxxx 】 Press【↑】to modify data (user must keep the scale empty at
【 LoAd】 this time), press【Enter】to confirm present data and enter next
digit, after finish 6 digits will enter weight calibration
【LoAd】 Weight calibration: Set the poise on the platform
Add poise
【LoAd】 Press【↑】and【Enter】to enter poise weight, to make the data
Press【↑】
8 same as the poise weight, then when the Tare signal is on, you
Press【Enter】
can press 【Enter】, the calibration finish. If the tare signal is off,
Press【Enter】
【 xxxx】 press 【Enter】until Tare signal on to finish calibration.
Note: 1. When calibrate, tare flag will be acted as stablization.
2. Press【Tare】If you want to quit without save settings. It keeps the old.
3. Power off the indicator after calibration, then switch on to check if the calibration is correct.
2. Calibration parameter
5
E-57
2. The indicator use currency RS232 UART signal , every 10BIT is a data frames, the format of
frames as follows:
Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit
1 2 3 4 5 6 7 8 9 10
BIT1:start bit
BIT2~BIT9:data bit
BIT10:stop bit
6
E-57
CHAPTER 6. Operation
1. Keyboard instruction:
(1) 【Cali.】 To modify some calibration parameter.
(2) 【Mode】 To select function: Weighing, Counting & percentum.
(3) 【Unit】 To select unit: kg or lb.
(4) 【Set】 To select up and low limits range in weighing model.
To sample the weight in counting and percentum model.
(5) 【Zero/B.L.】 To make weight zero when nothing on the platform but the display is not zero.
Press this key 2-4 seconds, the backlight will be switched on or off.
(6) 【Tare】 To set/delete the tare.
(7) 【↑/ ON/OFF】 To input number.
(8) 【Enter】 To confirm
2. Operation instruction:
When power on, the indicator will be in weighing model. The “Zero” flag and “kg” flag disappear.
You can press【Mode】to select model: weighing, counting, percentum.
Tare:
Press 【 Tare 】 to make the weight window display “Zero” after the weight is stable. The “Tare” flag
disappears. The weight is Net. Remove the poise, the displays is negative. Press 【Tare 】 again, the
displays is zero. The “Tare” flag disappear.
Note: (1). Tare function can set and set continuously only the weight is over zero and stable.
(2). Tare range: 99999 (if the display data is over 99999, not include the decimal, tare function is
ineffective)
Zero:
Press【Zero】to make display zero. (Zero range: <=4%FS, it is invalid under tare state.)
【Weighing mode】
Select Unit: Press【Unit】to select the unit between “kg” and “lb”. The corresponding flag disappears.
【Counting mode】
Press【Set】to sample weight in the counting model. It displays “SAP X” (X is the sample number).
1. Put enough samples on platform and press 【 ↑ 】 to select samples number: 10, 20, 50,100 &
200,500,1000. Press【Enter】confirm.
2. The window displays “LOAD-C” after the sample number is selected. Press【Enter】to finish sample.
3. There are two instance of unit weight lacking:
a. When the weight window displays “-SLAC-”, it means the unit weight of sample is less than 80%
of division. You still can count but the counting may be inaccurate. The signal will disappear
after about 3 seconds.
b. When the weight window displays “-CSL-”, it means the unit weight of sample is not enough at
all. Please press【Set】to re-sample. Press【Enter】to exit and return weighting.
Press【Set】again it will exit from the sample state.
【Percentum mode】
7
E-57
1. Press【Set】in the percentum state to display “LOAD-P”. There are two ways to sample:
a. Put enough samples on the platform and then press【Enter】to finish sample.
b. Press【Unit】 then the scale displays “000000”. Press【↑】and【Enter】to input the weight value.
2. When the sample weighed less than 0.1% FS., the window displays “-CSL-”. It means the sample is
not enough. You should re-sample. Press【Set】to re-sample. Press【Enter】to exit return weighing.
Press【Set】again it will exit from the sample state
Caution: The up and low limit range and alarm ways will be saved, different unit VS different alarm
range and ways.
8