You are on page 1of 6

CT Imaging • Original Research

Downloaded from www.ajronline.org by 2001:448a:50e0:7cca:9cf7:9334:f1b5:711 on 12/05/22 from IP address 2001:448a:50e0:7cca:9cf7:9334:f1b5:711. Copyright ARRS. For personal use only; all rights reserved

Schindera et al.
MDCT Tube Current Modulation

CT Imaging
Original Research

Effect of Patient Size on Radiation


Dose for Abdominal MDCT
with Automatic Tube Current
Modulation: Phantom Study
Sebastian T. Schindera1,2 OBJECTIVE. The purpose of this study was to evaluate in a phantom study the effect of
Rendon C. Nelson2 patient size on radiation dose for abdominal MDCT with automatic tube current modulation.
Thomas L. Toth 3 MATERIALS AND METHODS. One or two 4-cm-thick circumferential layers of fat-
Giao T. Nguyen 4 equivalent material were added to the abdomen of an anthropomorphic phantom to simulate
Greta I. Toncheva4 patients of three sizes: small (cross-sectional dimensions, 18 × 22 cm), average size (26 × 30
cm), and oversize (34 × 38 cm). Imaging was performed with a 64-MDCT scanner with com-
David M. DeLong2
bined z-axis and xy-axis tube current modulation according to two protocols: protocol A had
Terry T. Yoshizumi2,4 a noise index of 12.5 H, and protocol B, 15.0 H. Radiation doses to three abdominal organs
Schindera ST, Nelson RC, Toth TL, et al. and the skin were assessed. Image noise also was measured.
RESULTS. Despite increasing patient size, the image noise measured was similar for pro-
tocol A (range, 11.7–12.2 H) and protocol B (range, 13.9–14.8 H) (p > 0.05). With the two
protocols, in comparison with the dose of the small patient, the abdominal organ doses of the
average-sized patient and the oversized patient increased 161.5–190.6% and 426.9–528.1%, re-
spectively (p < 0.001). The skin dose increased as much as 268.6% for the average-sized patient
and 816.3% for the oversized patient compared with the small patient (p < 0.001).
CONCLUSION. Oversized patients undergoing abdominal MDCT with tube current
modulation receive significantly higher doses than do small patients. The noise index needs
to be adjusted to the body habitus to ensure dose efficiency.

T
he effectiveness of automatic tube consistent quality, the radiation exposure can
Keywords: automatic tube current modulation, CT, current modulation to reduce radi- vary substantially between a thin and a large
image quality, obesity, radiation dose ation dose and improve image patient. The potential risk of very high radia-
quality has been discerned for ab- tion doses to oversized patients when automatic
DOI:10.2214/AJR.07.2891 dominal CT [1–5]. With the technique, tube tube current modulation is used has been noted
Received July 18, 2007; accepted after revision
current is automatically adjusted to the X-ray by several authorities [2, 5, 7]. However, the
September 10, 2007. attenuation of the patient section being scanned exact extent of the radiation dose received by
to keep the radiation exposure as low as possi- oversized patients undergoing high-output CT
1
Department of Diagnostic Radiology, University Hospital ble and to obtain images with a constant speci- with automatic tube current modulation has not
of Bern, Bern, Switzerland.
fied image quality [6]. Various types of auto- been investigated, to our knowledge. Because
2
Department of Radiology, Duke University Medical matic tube current modulation, including z-axis obesity is a growing public health problem in
Center, Box 3808, Erwin Rd., Durham, NC 27710. modulation, xy-axis modulation, and combined many countries and CT plays an important role
Address correspondence to R. C. Nelson z- and xy-axis modulation, have been intro- in the diagnostic evaluation of obese patients
(rendon.nelson@duke.edu).
duced. State-of-the-art MDCT scanners in- with abdominal comorbid conditions, more
3
GE Healthcare Inc., Waukesha, WI. clude xyz-axis tube current modulation, where- data are needed to solve this important clini-
by tube current is modulated to patient-specific cal problem. The purpose of our study was to
4
Division of Radiation Safety, Duke University Medical attenuation in all three planes. evaluate in a phantom study the effect of patient
Center, Durham, NC. In automatic tube current modulation for ab- size on radiation dose in abdominal MDCT
WEB dominal CT, the radiation dose is increased for performed with xyz-axis automatic tube cur-
This is a Web exclusive article. a larger body habitus [1, 3, 4]. With the advent rent modulation.
of high-output X-ray tubes, which are capable
AJR 2008; 190:W100–W105 of producing peak tube currents up to 800 mA, Materials and Methods
0361–803X/08/1902–W100
it is technically feasible to maintain constant Anthropomorphic Phantom
image quality over a wide range of patient An adult-sized whole-body anthropomorphic
© American Roentgen Ray Society sizes. Thus in attempts to obtain CT images of phantom of a woman (model 702, CIRS) was used to

W100 AJR:190, February 2008


MDCT Tube Current Modulation
Downloaded from www.ajronline.org by 2001:448a:50e0:7cca:9cf7:9334:f1b5:711 on 12/05/22 from IP address 2001:448a:50e0:7cca:9cf7:9334:f1b5:711. Copyright ARRS. For personal use only; all rights reserved

Fig. 1—Photograph shows female adult anthropomorphic phantom encased with Fig. 2—Anteroposterior scout CT image of anthropomorphic phantom shows two
two 4-cm-thick fat rings covering upper abdominal portion. Phantom was placed 10-cm-wide fat rings covering four sectional slabs of upper abdomen. Fat rings
on wooden board, which was fixed on CT examination table. were placed at level of upper abdomen (T12–L3).

assess abdominal organ doses and image quality (Fig. GE Healthcare). The modulation technique for protocol. For both scanning protocols, the tube
1). The phantom was fabricated of epoxy resins for the z-axis is an automatic tube current modulation voltage was set at 140 kVp, representing the standard
accurate simulation of the physical density and X-ray feature used to adjust tube current along the z-axis tube voltage for routine abdominal CT in our
interaction of various human tissues. The phantom of the patient despite the wide range of attenuation department. For protocol A, a noise index of 12.5 H
was composed of 38 2.5-cm-thick sectional slabs. caused by changing patient size, anatomic features, was chosen, and for protocol B, 15.0 H. The selected
The cross-sectional diameter at the level of the upper and tissue composition [6]. In contrast, xy-axis noise indexes were recommended by Kalra et al. [1]
abdomen measured 18 cm in the anteroposterior modulation is used to adjust the tube current to for routine abdominopelvic CT examinations. The
direction and 22 cm in the lateral direction. minimize X-rays over angles that are less important minimum and maximum tube currents were 10 and
To simulate patients of different sizes, the in reducing overall image noise. Thus for anatomic 715 mA, respectively. At 140 kVp, the maximum
phantom was custom fitted at the level of the upper features that can be highly asymmetric, such as the tube current allowed by the MDCT scanner was
abdomen (T12–L3) with one or two 4-cm-thick shoulders, X-rays are significantly less attenuated in 715 mA. The gantry rotation time for the simulated
circumferential layers of fat-equivalent material the anteroposterior direction than in the transverse small patient was 0.5 second, and for the simulated
(Fig. 1). Each fat ring had a CT attenuation of direction, so the tube current can be substantially average- and oversized patients, 1.0 second. The
−80 H and measured 10 cm in the longitudinal reduced in this direction without causing a longer gantry rotation time was selected to prevent
direction (Fig. 2). The fat rings covered four substantial increase in overall image noise. use of tube current settings greater than 715 mA
sectional slabs of the upper abdomen. Three A single scout image was used to measure in the two phantom setups with a wider girth. All
types of phantom setups simulated three patients attenuation area and oval ratio, which were used other CT parameters were kept identical for the two
with different body sizes: a small patient (cross- to determine the tube current for the selected CT protocols (Table 1).
sectional diameter of phantom setup without fat
ring, 18 × 22 cm), an average-sized patient (setup
TABLE 1: CT Parameters
with one fat ring, 26 × 30 cm), and an oversized
patient (setup with two fat rings, 34 × 38 cm). The Parameter Protocol A Protocol B
cross-sectional dimensions of the three phantom Detector configuration 64 × 0.625 mm 64 × 0.625 mm
setups were in accordance with published human
Peak kilovoltage (kVp) 140 140
data on patient sizes [1, 8]. For example, the
abdominal anteroposterior by lateral diameter Noise index (H) 12.5 15.0
derived from 153 patients in one study averaged Tube current range (mA) 10–715 10–715
25 × 32 cm [1]. Gantry rotation time (s)
Simulated small patient 0.5 0.5
CT
The anthropomorphic phantom with its three Simulated average-sized and oversized 1.0 1.0
patient
setups was scanned on a 64-MDCT scanner
(VCT, GE Healthcare). Two scanning protocols Beam pitch 1.375 1.375
(A and B) were used with xyz-axis automatic Table feed per gantry rotation (mm) 55 55
tube current modulation, which merges z-axis
Reconstructed slice thickness (mm) 5 5
and xy-axis modulation techniques (Smart mA,

AJR:190, February 2008 W101


Schindera et al.
Downloaded from www.ajronline.org by 2001:448a:50e0:7cca:9cf7:9334:f1b5:711 on 12/05/22 from IP address 2001:448a:50e0:7cca:9cf7:9334:f1b5:711. Copyright ARRS. For personal use only; all rights reserved

Longitudinal scan coverage was 15 cm, which according to the three setups, either on the surface simulated average- and oversized patients within
included the four sectional slabs of the upper of the anthropomorphic phantom or on the surface protocols A and B. We applied a two-way analysis
abdomen encased by the 10-cm-wide fat ring plus of the fat rings. Because the helical beam pattern of of variance with the protocol and the phantom setup
2.5 cm cranial and caudal to the ring (Fig. 2). To MDCT exposes the detectors in a random manner, as the factors. A value of p < 0.05 was considered
avoid errors in radiation dose and image quality we took a conservative approach to determining to indicate a statistically significant difference.
caused by misplacement of the phantom, great effort a representative skin dose by accepting the three All statistical analyses were performed with SAS
was made to use the scanner laser to accurately highest readings of the two CT acquisitions. software system version 9.1.3.
align the phantom with the gantry isocenter. The mean radiation dose and its SD were
One of the authors assessed the mean tube computed for each of the three abdominal organs Image Quality Assessment and
current for the three phantom setups associated with and the skin. For statistical analysis of the radiation Statistical Analysis
protocols A and B and then multiplied by the gantry dose, we compared the two protocols within For image quality assessment, the three phantom
rotation time to acquire the mean tube current–time the same phantom setup and different phantom setups were scanned once with protocols A and
product. The CT dose index was obtained from the setups within the same protocol. For the latter, B without the MOSFET detectors to eliminate
operator console of the scanner. we compared the simulated small patient with the streak artifacts. Image quality was assessed by an

Detector Calibration Method


For measurements of the radiation dose to
the upper abdominal organs and to the skin, we
used a metal oxide semiconductor field effect
transistor (MOSFET) system (model TN-RD-60,
Thomson-Nielsen) with high-sensitivity radiology
dosimeters (TN-1002RD, Thomson-Nielsen). The
MOSFET reader was connected to a notebook
computer (Latitude, Dell), and the data were read
immediately after each CT exposure.
We calibrated the MOSFET detectors as follows.
First, we determined the thickness of copper sheets
to achieve the half-value layer (7.24 mm aluminum
at 120 kVp) of the CT scanner with a conventional
radiographic X-ray tube [9, 10]. Second, we added Fig. 3—Diagrams of axial
0.2-mm copper sheets to the X-ray tube to obtain an CT images show simulated
patients scanned with protocols
equivalent half-value layer of 7.37 mm aluminum A and B. Top, small patient
at 120 kVp. Third, individual MOSFET detectors (phantom without fat ring);
were calibrated at our clinical energy level of 140 middle, average-sized patient
(phantom encased by one fat
kVp. During calibration, detectors were placed
ring); bottom, oversized patient
beside an ion chamber (10x5-6, Radcal). Radiation (phantom encased by two
exposure was read with a radiation monitor (model fat rings). As cross-sectional
9015, Radcal) that had a built-in function for diameter of phantom increases,
image quality is maintained in
automatic temperature and pressure corrections. protocols A and B. Left bottom
Conversion from exposure to absorbed dose was image shows region of interest
computed by multiplication by an f-factor of 0.94 placement for image noise
measurement (circle).
at 140 kVp [9].

Radiation Dose Assessment and TABLE 2: Mean Tube Current–Time Product, Range of the Tube Current
Statistical Analysis Product, and CT Dose Index
To measure radiation dose to the upper
Mean Tube Current–Time
abdominal organs, 13 MOSFET dosimeters were CT Protocol Product (mAs) Tube Current (mA) CT Dose Index (mGy)
placed in drilled holes in the anthropomorphic
Small patient
phantom at the following three abdominal organ
locations: right and left lobes of the liver (five A 102.8 81–198 5.0
MOSFET detectors), stomach (five detectors), B 70.5 56–135 3.5
and spleen (three detectors). To avoid scattered Average-sized patient
radiation from beyond the fat rings, we used only
A 166.7 157–172 14.6
holes that were covered by the fat rings at least 3
cm in the cranial direction and 3 cm in the caudal B 115.2 108–119 9.7
direction. The three phantom setups with the Oversized patient
MOSFET detectors were scanned twice with each
A 576.6 468–631 47.8
of the two protocols. For skin dose measurements,
B 400.1 354–438 33.2
six MOSFET detectors were placed anteriorly,

W102 AJR:190, February 2008


MDCT Tube Current Modulation
Downloaded from www.ajronline.org by 2001:448a:50e0:7cca:9cf7:9334:f1b5:711 on 12/05/22 from IP address 2001:448a:50e0:7cca:9cf7:9334:f1b5:711. Copyright ARRS. For personal use only; all rights reserved

TABLE 3: Radiation Dose and Image Noise


Radiation Dose (mGy)
Protocol Liver Stomach Spleen Skin Image Noise (H)
Small patient
A 7.4 ± 0.8 6.4 ± 3.0 7.4 ± 0.8 8.6 ± 0.4 11.9 ± 0.6
B 5.2 ± 1.3 4.7 ± 1.5 5.0 ± 0.2 6.3 ± 0.5 14.1 ± 0.5
Average-sized patient
A 20.4 ± 1.2 18.6 ± 4.2 19.5 ± 0.8 31.7 ± 0.8 11.7 ± 0.7
B 13.6 ± 8.5 13.0 ± 5.4 13.5 ± 6.4 21.3 ± 2.5 13.9 ± 0.6
Oversized patient
A 41.1 ± 8.3 40.2 ± 13.1 44.5 ± 4.7 78.8 ± 4.5 12.2 ± 0.4
B 27.4 ± 5.4 26.4 ± 9.4 29.4 ± 3.7 54.7 ± 3.4 14.8 ± 0.7
Note—Values are mean ± SD.

investigator on a separate workstation (Advantage sizes, the radiation dose associated with both compared with the small patient (p < 0.001).
Windows 4.2, GE Healthcare) using reconstructed protocols was always higher for the skin As the size of the simulated patient increased,
5-mm-thick transverse images. For the image (range, 6.3–78.8 mGy) than for any of the the radiation dose to the skin yielded a greater
quality assessment, image noise was measured by abdominal organs (range, 4.7–44.5 mGy). percentage increase than did the radiation dose
placement of an approximately 2,000-mm 2 region The radiation doses associated with protocol to the abdominal organs (Fig. 4).
of interest (ROI) in the center of the phantom (Fig. A, which had a noise index of 12.5 H, were
3). Image noise was defined as the SD of the ROI 26.6–34.3% higher than those associated Discussion
value. Three ROI measurements were obtained with protocol B, which had a noise index of With high-output CT scanners with au-
on three axial CT images for each of the three 15.0 H (Table 4). At the same time, for the tomatic tube current modulation technique,
phantom setups scanned with the two protocols. three patient sizes, the image noise increased radiologists can obtain images with constant
Mean and SD were calculated for the image 18.5–21.2% from protocol A to protocol B. image quality despite the size of the patient.
noise values, which were compared among the This increase paralleled the difference be- Because it is clinically desirable to maintain
two protocols and the phantom setups by means tween the 12.5- and 15.0-H noise indexes image quality over a range of patient sizes,
of two-way analysis of variance. A difference of (20% increase). radiologists are tempted to use a constant
p < 0.05 was considered statistically significant. Within both protocols, the lowest radiation operator-selected image quality setting (e.g.,
doses occurred in the small patient and the noise index, reference tube current–time
Results highest doses in the oversized patient (Table 3). product) for patients of all sizes. From a pa-
Within protocols A and B, mean tube Within protocols A and B, the average-sized tient dose perspective, however, this approach
current–time product, range of tube current patient received 161.5–190.6% and the over- is not necessarily recommended. With a pre-
values, and CT dose index all increased in sized patient received 426.9–528.1% higher selected image quality setting, a thin patient
relation to the size of the simulated patient abdominal organ doses than did the small pa- can receive a very low radiation dose and an
(Table 2). Despite the increasing patient size, tient (p < 0.001) (Table 4). The skin doses in- oversized patient, a very high dose. To the
the two protocols had similar image noise creased up to 268.6% for the average-sized pa- best of our knowledge, the relation between
(p > 0.05) (Table 3). For the three patient tient and up to 816.3% for the oversized patient radiation dose and patient size with the use
of automatic tube current modulation in ab-
dominal CT has not been investigated in an
TABLE 4: Percentage Difference in Radiation Doses between Protocols A and anthropometric phantom study.
B and Patient Sizes The CT scanner used in our study deliv-
Patient Size Protocol Liver Stomach Spleen Skin Image Noise
ered a maximum tube current of 715 mA at
a peak tube potential of 140 kVp. Because
Small A vs B –29.4 a –26.6a –32.4 a –26.7a 18.5 of the high tube current capability of the
Average A vs B –33.3 –30.1 –30.8 –32.8 18.8 scanner, the selected noise indexes, and the
Oversized A vs B –33.3 –34.3 –33.9 –30.6 21.2 adjusted gantry rotation times, xyz-axis au-
tomatic tube current modulation maintained
Small vs average A 175.7 190.6 163.5 268.6 –1.7a
image quality for the three simulated patient
B 161.5 176.6 170.0 238.1 –1.4 a sizes. However, the maintenance of constant
Small vs oversized A 455.4 528.1 501.3 816.3 2.5a image quality for larger patients came at the
B 426.9 461.7 488.0 536.0 5.0a cost of significantly increased radiation. For
aNo statistically significant difference (p > 0.05); all other comparisons show a statistically significant
the simulated oversized patient, the abdomi-
difference (p < 0.05). nal organ doses increased up to 528.1% and

AJR:190, February 2008 W103


Schindera et al.
Downloaded from www.ajronline.org by 2001:448a:50e0:7cca:9cf7:9334:f1b5:711 on 12/05/22 from IP address 2001:448a:50e0:7cca:9cf7:9334:f1b5:711. Copyright ARRS. For personal use only; all rights reserved

80 80
Radiation Dose (mGy)

Radiation Dose (mGy)


60 60

40 40

20 20

0 0
20 25 30 35 40 20 25 30 35 40
Lateral Diameter (cm) Lateral Diameter (cm)
A B
Fig. 4—Mean radiation dose for skin (+), spleen (), stomach (), and liver () plotted according to size of simulated patient (lateral diameter).
A and B, Graphs show doses for protocol A (A) and protocol B (B).

the skin doses up to 816.3%, compared with et al. [12] assessed a sized-compensated tube upper abdomen instead of the entire abdo-
the doses in small patients. Thus when auto- current modulation technique using the square- men and pelvis. Total coverage of the abdo-
matic tube current modulation is used with- root projection area of the patient’s scout image men and pelvis was not technically feasible
out adjustment of the noise index to patient to select the noise index for abdominopelvic because of the curvature of the phantom at the
habitus, radiologists and technologists need CT. The square root of the projection area can level of the lower abdomen and pelvis. A com-
to be aware of the resultant very broad range be determined from scout or CT images and plete abdominopelvic CT examination of the
of radiation doses. is the most appropriate patient size metric for three phantom setups with the two protocols
Operator-selected image quality settings patient-dependent protocol planning because it should increase the absolute organ and skin
play a key role in the dose efficiency of au- is a direct measure of patient attenuation and dose numbers but not the percentage change
tomatic tube current modulation. Previous not merely a factor for attenuation, as are pa- between the phantom setups. Second, we used
studies [1, 11] have shown that subjective tient dimensions, weight, and body mass index. a female phantom and did not include a male
image quality increases with patient size on In the study by Udayasankar et al., the prelimi- phantom. The fat deposits of obese men tend
abdominopelvic CT images with constant nary results of size compensation were promis- to be mainly in the visceral region; in obese
image noise. In other words, radiologists ing. Two radiologists graded the image quality women, fat is present primarily in the subcu-
tend to accept greater subjective image qual- of all 100 abdominopelvic CT examinations taneous region [13, 14]. The female phantom,
ity for larger patients. The increased subjec- as acceptable regardless of patient size. The however, which had a smaller cross-sectional
tive image quality for larger patients may be applied noise indexes used for the 100 CT diameter than that of a male phantom and did
explained by the increase in fat deposition examinations ranged between 9.6 and 28.7 H. not have extra visceral fat deposition, simulates
around the abdominal organs that adds to the Another noteworthy finding of our study the worst-case radiation exposure scenario for
inherent tissue contrast between various or- was the greater percentage increase in skin a large patient. Furthermore, because our sim-
gans. This greater acceptance of higher sub- dose compared with abdominal organ dose ulation included an increase in subcutaneous
jective image quality for large patients may as patient size increased. A possible expla- fat only and not an increase in visceral fat, our
allow adjustment of the operator-selected nation for this finding may be the greater abdominal organ radiation dose results have
image quality settings in the use of auto- attenuation of the traversing photon fluence to be considered higher than those for an actu-
matic tube current modulation. The optimal through the wider girth of our simulated pa- al obese patient of corresponding size. Third,
noise indexes adjusted to patient habitus for tient, resulting in lower detection of radiation we evaluated automatic tube current modula-
abdominopelvic CT still have to be evaluated by the abdominal detectors in the center of tion with the xyz-axis modulation technique
in a clinical study [1]. the phantom compared with the skin detec- and did not separately evaluate z-axis and xy-
Selection of image quality for different-sized tors on the body surface. If one assumes that axis modulation techniques. Because clinical
patients undergoing abdominopelvic CT with the female breast can receive doses similar studies [2, 3] have shown extra dose savings
automatic tube current modulation is widely to our reported skin doses, oversized women with the combined technique, we decided to
operator dependent and thus arbitrary. No clear are at risk of receiving high radiation doses investigate only the technique with the greater
reference values for patients of various sizes to the radiosensitive breast with automatic potential for dose reduction.
have been published, to our knowledge. In an tube current modulation CT. Further stud- With abdominal CT performed with xyz-
attempt to reduce the arbitrary nature of param- ies of this technique are needed to assess axis automatic tube current modulation, our
eter selection, manufacturers of CT scanners exposure of the female breast to radiation in phantom study showed a wide range of radia-
should be encouraged to intensify research on different-sized patients. tion doses due to varying patient size. As the
semiautomatic selection of image quality set- There were limitations to our study. First, patient’s body habitus increased, the greatest
tings dependent on patient size. Udayasankar the two fat rings covered only 10 cm of the increase in radiation exposure was measured

W104 AJR:190, February 2008


MDCT Tube Current Modulation
Downloaded from www.ajronline.org by 2001:448a:50e0:7cca:9cf7:9334:f1b5:711 on 12/05/22 from IP address 2001:448a:50e0:7cca:9cf7:9334:f1b5:711. Copyright ARRS. For personal use only; all rights reserved

in the skin. To ensure dose efficiency with modulation. AJR 2007; 188:553–562 multidetector array computed tomography proto-
automatic tube current modulation in pa- 5. Kalra MK, Maher MM, Toth TL, Kamath RS, cols. J Comput Assist Tomogr 2007; 31:544–549
tients of all sizes, we recommend that the Halpern EF, Saini S. Comparison of Z-axis auto- 11. Wilting JE, Zwartkruis A, van Leeuwen MS,
operator-selected image quality settings be matic tube current modulation technique with Timmer J, Kamphuis AG, Feldberg M. A rational
adjusted to the patient’s habitus. fixed tube current CT scanning of abdomen and approach to dose reduction in CT: individualized
pelvis. Radiology 2004; 232:347–353 scan protocols. Eur Radiol 2001; 11:2627–2632
References 6. Kalra MK, Maher MM, Toth TL, et al. Strategies 12. Udayasankar UK, Kalra M, Li J, et al. Multidetector
1. Kalra MK, Maher MM, Kamath RS, et al. Six- for CT radiation dose optimization. Radiology scanning of the abdomen and pelvis: a study for evalu-
teen-detector row CT of abdomen and pelvis: 2004; 230:619–628 ation of size compensated automatic tube current
study for optimization of Z-axis modulation tech- 7. McCollough CH, Bruesewitz MR, Kofler JM Jr. modulation technique in 100 subjects. RSNA 2006.
nique performed in 153 patients. Radiology 2004; CT dose reduction and dose management tools: Oak Brook, IL: Radiological Society of North America,
233:241–249 overview of available options. RadioGraphics 2006. Available at: http://rsna2006.rsna.org/rsna2006/
2. Mulkens TH, Bellinck P, Baeyaert M, et al. Use of an 2006; 26:503–512 V2006/conference/event_display.cfm?em_id=4441276.
automatic exposure control mechanism for dose opti- 8. Ogden K, Huda W, Scalzetti EM, Roskopf ML. Accessed November 14, 2007
mization in multi-detector row CT examinations: Patient size and x-ray transmission in body CT. 13. Kanehisa H, Miyatani M, Azuma K, Kuno S, Fu-
clinical evaluation. Radiology 2005; 237:213–223 Health Phys 2004; 86:397–405 kunaga T. Influences of age and sex on abdominal
3. Rizzo S, Kalra M, Schmidt B, et al. Comparison 9. Hendee WR. Interaction of x and gamma rays in muscle and subcutaneous fat thickness. Eur J
of angular and combined automatic tube current the body. In: Hendee, WR, ed. Medical radiation Appl Physiol 2004; 91:534–537
modulation techniques with constant tube current physics, 2nd ed, Chicago, IL: Year Book, 14. Snehalatha C, Ramachandran A, Satyavani K,
CT of the abdomen and pelvis. AJR 2006; 1979:171–172 Vallabi MY, Viswanathan V. Computed axial to-
186:673–679 10. Hurwitz LM, Yoshizumi TT, Goodman PC, et al. mographic scan measurement of abdominal fat
4. Mulkens TH, Daineffe S, De Wijngaert R, et al. Effective dose determination using an anthropomor- distribution and its correlation with anthropome-
Urinary stone disease: comparison of standard- phic phantom and metal oxide semiconductor field try and insulin secretion in healthy Asian Indians.
dose and low-dose with 4D MDCT tube current effect transistor technology for clinical adult body Metabolism 1997; 46:1220–1224

AJR:190, February 2008 W105

You might also like