Professional Documents
Culture Documents
Pediatric Chest MDCT Using Tube Current Modulation - Effect On Radiation Dose With Breast Shielding.
Pediatric Chest MDCT Using Tube Current Modulation - Effect On Radiation Dose With Breast Shielding.
Coursey et al.
Pediatric Chest MDCT with Tube Current Modulation
Pediatric Imaging
Original Research
A
n estimated 57 million CT exami- and bismuth shielding [13–15]. Automatic
Box 3808, Duke University Medical Center, Durham,
NC 27710. Address correspondence to D. P. Frush
nations were performed in the tube current modulation using both angular
(frush943@mc.duke.edu). United States in 2000 [1]. It is (x,y-axis) and longitudinal (z-axis) tech-
also estimated that approximately niques has been shown to reduce patient
2
Radiation Safety Division, Duke University Medical 7.1 million CT examinations were performed dose [5–12]. Shielding has also been shown
Center, Durham, NC.
in pediatric patients in 2002 [2]. Because of to reduce radiation dose, primarily from at-
3
Department of Radiology, University of Arkansas for the large number of examinations and the rel- tenuation of the primary beam.
Medical Sciences and Arkansas Children’s Hospital, atively high radiation dose delivered by CT, The combined use of automatic tube cur-
Little Rock, AR. techniques should minimize the radiation rent modulation and a shielding device might
dose to the patient while still providing diag- be expected to reduce radiation dose more
CME
This article is available for CME credit. See www.arrs.org
nostic quality—the ALARA (as low as rea- than either technique alone. However, it may
for more information. sonably achievable) principle. Radiation dose also be that automatic tube current modula-
is of special concern for the pediatric patient tion technology that modulates in the z-axis
WEB population because the younger the child at on the basis of the topogram (scout image)
This is a Web exclusive article.
exposure, the greater the potential for cancer may compensate for the presence of the
AJR 2008; 190:W54–W61 development [3, 4]. shield by increasing the tube current and
Two techniques that have independently thus increasing the dose during the diagnos-
0361–803X/08/1901–W54
been shown to lower CT radiation dose are tic scan when the shield is present while the
© American Roentgen Ray Society automatic tube current modulation [5–12] scout image is obtained. To our knowledge,
A B
Fig. 1—Phantom and shield used in this study.
A, Anthropomorphic phantom representing 5-year-old child (Atom Pediatric 5-Year-Old Phantom, model 705-D CIRS) with metal oxide semiconductor field effect transistor
(MOSFET) (Autosense MOSFET system, Best Medical Canada) detectors in place.
B, Bismuth (F & L Medical Products) breast shield (arrow) in place.
dosimetry during MDCT using the combina- TABLE 1: Organ Doses Using MOSFET During Chest MDCT in Phantom
tion of bismuth shielding and automatic tube Representing a 5-Year-Old Child
current modulation has not been reported. Dose (cGy)a
Therefore, the purpose of our study was to
Without With Shield, Scout With Scout Without
determine the effect on radiation dose and Slice, Shield or Without Shield, With Shield, With
to measure image quality (noise) of images Organ Locationb ATCM ATCM ATCM ATCM
obtained using bismuth breast shields with z- Breast (l) 12, 58 0.40 0.31 0.17 0.17
axis automatic tube current modulation dur-
Breast (r) 12, 59 0.35 0.25 0.19 0.19
ing pediatric chest MDCT.
Skin above breast 12, – 0.37 0.28 0.20 0.20
Materials and Methods Skin above sternum 12, – 0.48 0.32 0.29 0.18
An anthropomorphic phantom representing
Esophagus and heart 12, 51 0.36 0.34 0.26 0.19
a 5-year-old child (Atom Pediatric 5-Year-Old
Phantom, model 705-D, CIRS) was used (Fig. Bone marrow, sternum 11, 39 0.31 0.29 0.20 0.12
1). The specific dimensions of the phantom are Bone marrow, ribs (l) 11, 40 0.30 0.24 0.19 0.13
weight, 19 kg; height, 110 cm; thorax dimensions, Bone marrow, ribs (r) 11, 45 0.38 0.32 0.13 0.18
14 × 17 cm. This phantom was scanned using the Bone marrow, thoracic 11, 38 0.46 0.30 0.23 0.21
technique clinically used for the 5-year-old age- and lumbar spine
based chest CT protocol (16-MDCT, LightSpeed, Lung (l) (middle) 12, 53 0.41 0.37 0.27 0.19
GE Healthcare) consisting of 120 kVp, 0.5-second Lung (r) (middle) 12, 56 0.40 0.36 0.26 0.22
rotation time, 1.375 pitch, 16 × 1.25 mm effective
Lung (l) (middle) 13, 68 0.44 0.42 0.38 0.25
collimation, 5-mm slice thickness, 2.5-mm re-
Lung (r) (middle) 13, 71 0.48 0.41 0.40 0.26
construction interval, and a small scanning field
of view, with images reconstructed in a standard Thymus 12, 60 0.43 0.37 0.28 0.23
algorithm. The phantom was first scanned with Thyroid (l)c 9, 24 0.45 0.44 0.24 0.25
a fixed tube current of 65 mA without a breast Thyroid (r)c 9, 25 0.43 0.42 0.26 0.25
shield. The phantom was next scanned using the Liverc 16, 103 0.40 0.42 0.38 0.35
same parameters but with the addition of a 2-ply
Stomachc 16, 109 0.37 0.37 0.36 0.37
bismuth (1.7 g/cm 2) (F & L Medical Products)
Esophagusc 10, 35 0.40 0.36 0.24 0.23
breast shield with a 1-cm-thick foam stand-off
pad. The shield measured 9 cm in the craniocau- Kidney (l)c 16, 116 0.33 0.31 0.31 0.28
dal dimension and was designed to cover the chest Note—MOSFET = metal oxide semiconductor field effect transistor. ATCM = z-axis automatic tube
to the midaxillary line (Fig. 1). The phantom was current modulation. Dash (–) indicates detectors placed on surface of phantom.
a Average of two scans was obtained for each protocol.
then scanned using z-axis automatic tube current bOf CIRS phantom (Atom Pediatric 5-Year-Old Phantom, model 705-D).
modulation (Auto mA, GE Healthcare) with 65 cLocations not underneath breast shield.
mA maximum, 10 mA minimum, and a noise in- diographs obtained with the bismuth breast shield noise level (the noise index) between maximum
dex (measured in SD of Hounsfield units) of 12.0 in place. Auto mA selects tube currents on the ba- (ceiling) and minimum levels, which are also se-
H (SD) (all other parameters remaining the same) sis of patient density, size, and shape, as reflected lected [16]. The noise index approximates noise
with the topogram (scout) frontal and lateral ra- in the localizer radiograph, to maintain the target as measured in the center of a uniform phantom
Downloaded from www.ajronline.org by 54.210.20.124 on 10/28/15 from IP address 54.210.20.124. Copyright ARRS. For personal use only; all rights reserved
A B
C D
Mid scapulae 9.42 (7.65–11.31) 9.62 (7.12–12.4) 0.63 11.87 (9.05–14.38) 11.9 (9.08–14.82) 0.93
Superior mediastinuma 8.96 (6.80–12.24) 9.42 (8.14–10.56) 0.27 11.41 (9.79–13.58) 13.06 (11.20–16.86) 0.01
Heart a 8.50 (7.27–9.93) 9.96 (8.93–11.51) 0.008 10.00 (8.00–10.82) 12.82 (11.55–14.36) 0.001
Upper abdomen 11.4 (10.25–12.95) 11.12 (9.90–12.68) 0.44 11.12 (9.83–11.96) 11.38 (10.06–14.06) 0.50
Note—Numbers in parentheses are minimum–maximum SD of Hounsfield units in ≈ 170 mm 2 region of interest. ATCM = automatic tube current modulation.
aLocation in range of shielding.
With the implementation of tube current the left lung to a 62.5% decrease in the skin surements were 11.4 H for the fixed tube cur-
modulation with the shield present in the scout over the sternum. Of note, breast dose de- rent scan with no shield, 11.1 H for the fixed
image, measured doses decreased at 19 of the creased by 51.6% when automatic tube cur- tube current scan with breast shield, 11.2 H
20 (95%) locations (average dose decrease, rent modulation and the bismuth shield were for automatic tube current modulation and
25.5%) when compared with the fixed tube used with the shield placed after the scout as breast shield with shield present in the scout,
current scan with the breast shield (Table 1). compared with the fixed tube current, age- and 11.4 H for automatic tube current modula-
In the remaining location, the measured dose based chest CT with no shield. tion and breast shield with shield placed after
in the left kidney was unchanged (Table 1). Image noise was measured as described the scout (Table 3). Mean noise measurements
Comparing the examinations with au- earlier (Table 2). For the CT images in range were not significantly different with the addi-
tomatic tube current modulation when the of the shield, average noise measurements tion of the breast shield in the fixed tube cur-
shield was placed before the scout was ob- (SD of Hounsfield units) were 8.7 H for the rent scans, except in the heart, where the noise
tained versus when the shield was placed fixed tube current scan, 9.7 H for the fixed increased from 8.50 to 9.96 H (p = 0.008)
after, measured doses were higher (average tube current scan with breast shield, 10.6 H (Table 2). When the shield was placed after
dose increase, 48%) at 10 of 14 locations for automatic tube current modulation and the scout was obtained with automatic tube
(71.4%) under the shield, unchanged at three breast shield with shield present in the scout, current modulation, the mean noise measure-
locations (21.4%) under the shield, and lower and 12.9 H for automatic tube current modula- ment at the level of the heart was significantly
(average dose decrease, 27.8%) at one loca- tion and breast shield with shield placed after greater (from 10.0 to 12.8 H) than when the
tion (7.1%) under the shield when the shield the scout was obtained (Table 3). Superior to shield was present in the scout (p = 0.001).
was present in the scout image (Table 1). the shield, average noise measurements were Likewise, the mean noise measurement at the
Comparing the overall highest-dose scan 9.3 H for the fixed tube current scan with no level of the superior mediastinum was sig-
(fixed tube current, age-based chest CT proto- shield, 9.4 H for the fixed tube current scan nificantly greater as compared with the scans
col, no shield) with the lowest-dose scan (au- with the breast shield, 11.7 H for automatic when the shield was present in the scout (from
tomatic tube current modulation and bismuth tube current modulation and breast shield 11.4 to 13.1 H) (p = 0.01). Average noise at the
shield with shield placed after scout image with shield present in the scout, and 11.5 H for level of the shield was greater than the target
was obtained), measured doses decreased by automatic tube current modulation and breast noise index (12.0 H [SD]) when the shield was
an average of 51.3% in the 16 locations under shield with shield placed after the scout (Table placed after the scout because automatic tube
the shield, ranging from a 43.2% decrease in 3). Inferior to the shield, average noise mea- current modulation did not compensate for the
presence of the shield and generated a lower
tube current (34–36 mA in the range of the
TABLE 3: Effective Dose and Noise Determination for Chest MDCT in shield) compared with when the shield was
Phantom Representing a 5-Year-Old Child present in the scout (tube current of 40–62
Noiseb mA in the range of the shield). This lower
Effective Above At Shield Below tube current resulted in increased noise.
Scanning Paradigm Dosea (mSv) Shield Level Shield Using the DLP method, the effective dos-
Fixed tube current, age-based chest CT 2.0 9.3 8.7 11.4 es were 2.0 mSv using the fixed tube current
Fixed tube current, age-based chest CT 2.0 9.4 9.7 11.1 technique with and without the breast shield,
plus breast shield as expected because DLP is independent of
ATCM, with shield, shield present in 1.6 11.7 10.6 11.2 the scout, unlike with automatic tube current
scout image modulation, in which the DLP will depend
ATCM, with shield, shield placed after 1.4 11.5 12.9 11.4 on the scan modulation. When automatic
scout image obtained tube current modulation is used, the project-
Note—ATCM = automatic tube current modulation in z-axis using Auto mA (GE Healthcare). ed modulations in tube current are accounted
aBased on dose–length product. for in the DLP. The total effective dose was
bSD of Hounsfield units in ≈ 170 mm2 region of interest.
1.6 mSv using automatic tube current modu-
lation with the bismuth shield present in the The algorithms used for automatic tube cur- used in this investigation (combined z- and x,y-
scout radiograph, yielding a 20% dose reduc- rent modulation differ depending on the man- axis modulation was not available at the time
tion as compared with the fixed tube current ufacturer [28]. The z-axis modulation used in of our study), the increased attenuation present
scans (Table 3). The total effective dose was this investigation (Auto mA) requires the user on the scout image due to the shield (and the
1.4 mSv using automatic tube current modu- to select the noise index (SD of Hounsfield concomitant increase in tube current in this re-
lation with the bismuth shield placed after units) and minimum and maximum tube cur- gion) could negate the benefits of both shield-
Downloaded from www.ajronline.org by 54.210.20.124 on 10/28/15 from IP address 54.210.20.124. Copyright ARRS. For personal use only; all rights reserved
the scout radiograph was obtained, yielding rent values. The range between these values is ing and tube current modulation. We found
a 30% dose reduction as compared with the where the modulation will take place. that Auto mA z-axis modulation compensated
fixed tube current scans (Table 3). Automatic tube current modulation in the for the density of the shield if the shield was
z-axis has been shown to reduce tube current– present in the scout radiograph by increasing
Discussion time product by 18–26% in the setting of the tube current along the z-axis in the range of
Managing the radiation dose during pediat- adult chest CT, depending on the noise index the shield (Fig. 3). As seen in Figure 3, the tube
ric chest CT is important for two reasons. First, [7]. Automatic tube current modulation in current began to increase at the leading edge of
the potential cancer risk has been shown to in- the x,y-axis has been shown to reduce mAs the breast shield. The rate of rise increased dra-
crease as age at radiation exposure decreases by 31–39% in the setting of pediatric chest matically after 15 slices, or approximately 7.5
[3, 4]. Second, radiation-sensitive tissues such CT [6] and to reduce mean effective mAs by cm beyond the leading edge of the shield. The
as the breasts [23–26] are directly irradiated 16.9% in the setting of adult chest CT [5]. reason for this unexpected delay in increasing
during chest CT. Strategies for reducing radia- Note, however, that tube current reductions tube current is not known. Although there is
tion dose for pediatric chest CT include ap- are surrogates for the actual patient dose. The a lag before modulation begins, it is relatively
propriate adjustments in scanning parameters current study provides additional insight into short (80 milliseconds). Further analysis of the
[27], in-plane breast shielding (where the the work of prior studies by also assessing the data with the manufacturer did not provide an
shield is in range of CT) [14–15], and auto- effect on organ doses when automatic tube cur- explanation (Toth T, GE Healthcare, personal
matic tube current modulation [5–12]. rent modulation is used because dosimetry was communication).
Bismuth breast shields have been shown to performed, rather than recording estimated However, if the modulation began sooner
reduce breast radiation dose by 29% in chil- dose changes based on alterations in tube cur- in the upper thorax, we would have expected
dren [14] and by 27% to 52% in adults, de- rent. MOSFET dosimetry offers an opportu- an even greater dose savings with Auto mA
pending on the thickness of the shield [15]. nity for more accurate dose determinations in tube current modulation when the shield was
In line with these investigations, we found a CT [19]. The 37.5% average organ dose reduc- placed after the scout. That tube current did
26% dose reduction to the breast in an anthro- tion in the current anthropomorphic phantom not decrease beyond the shield before reach-
pomorphic phantom representing a 5-year-old study with the use of Auto mA tube current ing the abdomen can be explained by the fact
using a 2-ply bismuth breast shield. The 4% modulation with the shield placed after the that the Auto mA tube current modulation
decrease in dose at four locations not under scout image, then, is slightly higher than previ- software derives the tube current on the ba-
the shield and the 5% increase in dose at one ously reported mAs reductions in the setting of sis of the highest attenuating structure under
location not under the shield when the shield adult chest CT with z-axis automatic tube cur- the detectors at a given time. A portion of
is added is not unexpected in that the uncer- rent modulation. This difference may reflect a the detectors was likely always either over
tainty of the MOSFET system is approximate- combination of different automatic tube cur- the shield or over the upper abdomen at the
ly 10% or less for a typical CT dose of 1 cGy. rent modulation technologies, different noise end of the scanning. However, even though
Three general techniques for automatic indexes, or automatic tube current modulation Auto mA compensated for the shield when
tube current modulation are offered by the when used in a phantom. The 25.5% average the shield was present in the scout, the select-
manufacturer of the scanner used for this in- dose reduction with Auto mA tube current ed tube currents through the positions that
vestigation: z-axis (longitudinal) modulation, modulation with the shield present in the scout included the shield (40–62 mA) still were
x,y-axis (angular) modulation, and a combi- is similar to previous investigations. lower than the fixed tube current of 65 mA
nation of both. With z-axis modulation, the The locations (stomach and left kidney) in used in the fixed-tube-current, age-based
method used in this study, tube current is which the measured dose did not change with scan. Therefore, some but not all of the tube
modulated along the craniocaudal dimension Auto mA were in the upper abdomen, where current benefit was lost when the shield was
of the patient. For example, in the setting of tube current was the same (65 mA) in the fixed present in the scout radiograph.
chest CT with z-axis modulation, a relatively tube current and the automatic tube current Furthermore, we postulated that placing the
lower tube current is delivered through the modulation scans, likely because an equivalent shield after the scout image was obtained would
lungs, where there is less attenuation, while amount of tube current was necessary to main- maximize dose benefits from both shielding
a relatively higher tube current is delivered tain the target noise level. Noise levels in the and Auto mA tube current modulation. In fact,
through the upper abdomen, where there is sections containing the stomach and left kid- placing the shield after the scout was obtained
more soft-tissue attenuation. With x,y-axis ney dosimeters were similar in the automatic resulted in radiation doses (measured using
modulation, a relatively lower tube current is tube current modulation and fixed tube current MOSFET technology) that were lower at 10
delivered through the thinner portions of the scans (data not shown). locations, unchanged at three locations (right
patient (generally the anteroposterior dimen- The use of breast shields in combination breast, left breast, skin over breasts), and in-
sion), and a relatively higher tube current is with automatic tube current modulation has creased at one location in the range of shield-
delivered when there is relatively more at- not been previously addressed. It is conceiv- ing. The effective dose, as expected, decreased
tenuation (generally the lateral dimension). able that with the type of z-axis modulation by 13%, from 1.6 to 1.4 mSv, when the shield
was placed after the scout radiograph with noise increased. The highest-dose scan (fixed nology was investigated. For example, not all
Auto mA tube current modulation. tube current chest CT without shielding) was automatic tube current modulation products
That breast dose and skin dose superficial also the least noisy. The lower noise measured described by McCollough et al. [28] modu-
to the breasts did not change when the shield in the nonmodulated examination was due, we late on the basis of the scout radiograph.
was placed after the scout was obtained while think, to tube current levels that were higher Therefore, the effect of placing the shield
almost all the other organ doses in the range than needed (such as in the lungs). For non- before or after the scout radiograph cannot
Downloaded from www.ajronline.org by 54.210.20.124 on 10/28/15 from IP address 54.210.20.124. Copyright ARRS. For personal use only; all rights reserved
of the shield decreased was somewhat un- modulated examinations, it is likely that pro- be predicted in these settings. The use of an
expected. One explanation is that the breast tocols have evolved to maintain a minimum anthropomorphic phantom representing a
detectors were located near the leading edge noise level for diagnostic quality through single age is also a limitation of this study.
of the shield, where the differences in tube intrinsically higher attenuating areas, mean- However, it is expected that similar dose re-
current were not so great (43–50 mA when ing other regions will have less noise than is ductions would be seen with anthropomor-
shield was present in scout image vs 34–35 needed. For example, in the nonmodulated phic phantoms representing different ages.
mA when shield was placed after scout im- examinations, average noise measurements Also, because there are no respiratory or
age was obtained). Given the spiral nature of cranial to the shield were 9.3 and 9.4 H (SD) cardiac motion artifacts in a phantom study,
the primary beam, it is also possible that the and caudal to the shield were 11.4 and 11.1 H the doses and image noise changes we found
position of the beam (e.g., posteroanterior vs (SD) in the runs with and without the breast may not completely translate into a patient
anteroposterior) relative to the breast detec- shield, respectively (Table 3). study. The use of the SD of Hounsfield units
tors was different in the automatic tube cur- The lowest-dose scan (automatic tube in a region of interest as a proxy for image
rent modulation runs with shield placed be- current modulation and shield, with shield quality is also a limitation of this study. The
fore and after the scout image was obtained. placed after the scout image was obtained) effect of the combined use of bismuth shields
A posteroanterior beam would be more at- was the noisiest scan. The increase in im- and automatic tube current modulation on
tenuated by soft tissues before it reaches the age noise at the level of the heart and supe- diagnostic quality of images needs to be as-
breast detectors, whereas an anteroposte- rior mediastinum (in the range of shielding) sessed. Finally, our conversion factor used to
rior beam would be attenuated by the breast when the shield was placed after the scout compute effective dose was an estimate of a
shield before reaching the detectors. This was significantly (p < 0.01) higher compared chest CT protocol for a 5-year-old because
phenomenon has been reported recently with with when the shield was present in the scout standards are not available for pediatric
MOSFET and breast dosimetry and is a limi- radiograph with automatic tube current mod- chest 16-MDCT. Although the effective dose
tation of this type of surface dosimetry (un- ulation. It is unclear why the heart and not would differ if another conversion factor
like more central dose locations) in helical as the superior mediastinum had a significantly were used, the change in the effective dose
opposed to axial MDCT [29]. higher amount of noise with nonmodulated we determined (13% decrease) when the
It was also unexpected that the average scanning with and without the shield. This shield was placed after the scout compared
dose in the bone marrow of the right ribs may represent a statistical error. with when the shield was present in the scout
increased when the shield was placed after When the shield was placed after the scout is independent of the factor used.
the scout (Table 1). As noted earlier, all other image was obtained, the measured noise in- In conclusion, the use of a bismuth breast
measured doses in the range of the shield de- dex (12.9 H [SD]) in the range of shielding shield in addition to one technique for z-axis
creased or were unchanged when the shield was slightly higher than the selected target automatic tube current modulation further re-
was placed after the scout with Auto mA. noise index of 12.0 H (SD). Although this duces detected radiation dose as determined
The measured doses at this location for each increase in noise is statistically significant, it in a pediatric anthropomorphic phantom.
of the four Auto mA runs (two with the shield does not necessarily equate to a reduction in Dose reduction is greatest when the breast
present in the scout and two with the shield diagnostic quality. For example, a prior inves- shield is placed after the scout radiograph is
placed after the scout) were 0.12, 0.14, 0.13, tigation (Darsie J et al., presented at the 2005 obtained. Although placing the breast shield
and 0.22 cGy. The 0.22-cGy value appears to annual meeting of the Radiological Society of after the scout radiograph is obtained does
be an outlier that may have falsely elevated North America) supports a noise index of 12.9 result in increased image noise in the range
the average. H (SD) to be in the acceptable range for pe- of the shield because the tube current is cal-
Breast doses of 0.1–0.7 Gy have been diatric chest CT. In addition, the target noise culated without factoring in the attenuation
shown to be associated with an increased index was 12.0 H (SD), and the breast shields of the breast shield, the increased image
risk of breast cancer [25, 26]. Although the increased noise above this level by 0.9 H (SD), noise is still close to the noise target level.
breast dose in our study was low (0.4 cGy a difference that is arguably below threshold
without the shield), appropriately minimiz- for visual detection. In the automatic tube References
ing any radiation dose, especially for breast current modulation runs, not unexpectedly, 1. Linton OW, Mettler FA Jr. National conference on
parenchyma in young individuals, is prudent. noise measurements were similar both cra- dose reduction in CT, with an emphasis on pediat-
In keeping with this ALARA principle, we nial to and caudal to the shield. The automatic ric patients. AJR 2003; 181:321–329
found that breast dose was reduced by 26% tube current modulation algorithm modulates 2. Frush DP, Applegate K. Computed tomography
with the breast shield alone and by 52% when tube current to maintain a constant noise level and radiation: understanding the issues. J Am Coll
using the breast shield and automatic tube throughout the scanning. Radiol 2004; 1:113–119
current modulation in an anthropomorphic Limitations of our study include that only 3. Brenner DJ, Elliston CD, Hall EJ, Berdon WE.
phantom representing a 5-year-old child. one type of CT scanner and therefore only Estimated risks of radiation-induced fatal cancer
When measured doses decreased, image one automatic tube current modulation tech- from pediatric CT. AJR 2001; 176:289–296
4. Karlsson P, Holmberg E, Lundell M, Mattsson A, 12. Mastora I, Remy-Jardin M, Suess C, Scherf C, fective dose determination using an anthropomor-
Holm LE, Wallgren A. Intracranial tumors after Guillot JP, Remy J. Dose reduction in spiral CT phic phantom and metal oxide semiconductor field
exposure to ionizing radiation during infancy: a angiography of thoracic outlet syndrome by ana- effect transistor technology for clinical adult body
pooled analysis of two Swedish cohorts of 28,008 tomically adapted tube current modulation. Eur multidetector array computed tomography proto-
infants with skin hemangioma. Radiat Res 1995; Radiol 2001; 11:590–596 cols. J Comput Assist Tomogr 2007 31:544–549
150:357–364 13. Perisinakis K, Raissaki M, Tzedakis A, Theoch- 21. Shrimpton PC, Wall BF. Reference doses for pae-
Downloaded from www.ajronline.org by 54.210.20.124 on 10/28/15 from IP address 54.210.20.124. Copyright ARRS. For personal use only; all rights reserved
5. Tack D, De Maertelaer V, Gevenois PA. Dose re- aropoulous N, Damilakis J, Gourtsoyiannis N. diatric computed tomography. Radiat Prot Dosim
duction in multidetector CT using attenuation- Reduction of eye lens radiation dose by orbital 2000; 90:249–252
based online tube current modulation. AJR 2003; bismuth shielding in pediatric patients undergo- 22. Hollingsworth CL, Frush DP, Yoshizumi TT, et
181:331–334 ing CT of the head: a Monte Carlo study. Med al. Pediatric cardiac-gated CT angiography: as-
6. Greess H, Lutze J, Nömayr A, et al. Dose reduc- Phys 2005; 32:1024–1030 sessment of radiation dose. AJR 2007; 189:12–18
tion in subsecond multislice spiral CT examina- 14. Fricke BL, Donnelly LF, Frush DP, et al. In-plane 23. Travis LB, Hill D, Dores GM, et al. Cumulative
tion of children by online tube current modula- bismuth breast shields for pediatric CT: effects on absolute breast cancer risk for young women
tion. Eur Radiol 2004; 14:995–999 radiation dose and image quality using experi- treated for Hodgkin lymphoma. J Natl Cancer
7. Kalra MK, Rizzo S, Maher M, et al. Chest CT per- mental and clinical data. AJR 2003; 180:407–411 Inst 2005; 97:1428–1437
formed with z-axis modulation: scanning protocol 15. Hopper KD, King SH, Lobell ME, TenHave TR, 24. Land CE, Tokunaga M, Koyama K, et al. Inci-
and radiation dose. Radiology 2005; 237:303–308 Weaver JS. The breast: in-plane x-ray protection dence of female breast cancer among atomic
8. Kalra MK, Maher MM, Toth TL, Kamath RS, during diagnostic thoracic CT—shielding with bomb survivors, Hiroshima and Nagasaki,
Halpern EF, Saini S. Comparison of z-axis auto- bismuth radioprotective garments. Radiology 1950–1990. Radiat Res 2003; 160:707–717
matic tube current modulation technique with 1997; 205:853–858 25. Miller AB, Howe GR, Sherman GJ, et al. Mortali-
fixed tube current CT scanning of abdomen and 16. Kalra MK, Maher MM, Toth TL, et al. Techniques ty from breast cancer after irradiation during fluo-
pelvis. Radiology 2004; 232:347–353 and applications of automatic tube current modu- roscopic examinations in patients being treated for
9. Kalra MK, Maher MM, Kamath RS, et al. Six- lation for CT. Radiology 2004; 233:649–657 tuberculosis. N Engl J Med 1989; 321:1285–1289
teen-detector row CT of abdomen and pelvis: 17. Atom pediatric 5-year-old phantom user manual: 26. Hildreth NG, Shore RE, Dvoretsky PM. The risk
study for optimization of z-axis modulation tech- model 705-D handling instructions. Norfolk, VA: of breast cancer after irradiation of the thymus in
nique performed in 153 patients. Radiology 2004; CIRS, 2002 infancy. N Engl J Med 1989; 321:1281–1284
233:241–249 18. Peet DJ, Pryor MD. Evaluation of a MOSFET ra- 27. Frush DP. Pediatric CT: practical approach to di-
10. Mulkens TH, Bellinck P, Baevaert M, et al. Use of diation sensor for the measurement of entrance minish the radiation dose. Pediatr Radiol 2002;
an automatic exposure control mechanism for surface dose in diagnostic radiology. BJR 1999; 32:714–717
dose optimization in multi-detector row CT ex- 72:562–568 28. McCollough CH, Bruesewitz MR, Kofler JM Jr.
aminations: clinical evaluation. Radiology 2005; 19. Yoshizumi T, Goodman P, Frush DP, et al. Valida- CT dose reduction and dose management tools:
237:213–223 tion of metal oxide semiconductor field effect overview of available options. RadioGraphics
11. Kalendar WA, Wolf H, Suess C, Gies M, Greess H, transistor technology for organ dose assessment 2006; 26:503–512
Bautz WA. Dose reduction in CT by on-line tube during CT: comparison with thermoluminescent 29. Hurwitz LM, Yoshizumi TT, Reiman RE, et al.
current control: principles and validation on phan- dosimetry. AJR 2007; 188:1332–1336 Radiation dose to the female breast from 16-MDCT
toms and cadavers. Eur Radiol 1999; 9:323–328 20. Hurwitz LM, Yoshizumi TT, Goodman P, et al. Ef- body protocols. AJR 2006; 186:1718–1722
1. Joana Santos, Shane Foley, Graciano Paulo, Mark F. McEntee, Louise Rainford. 2015. The impact of pediatric-specific dose
modulation curves on radiation dose and image quality in head computed tomography. Pediatric Radiology 45, 1814-1822.
[CrossRef]
2. J.H. Cho, M.S. Kim, J.D. Rhim. 2015. Comparison of radiation shielding ratios of nano-sized bismuth trioxide and
molybdenum. Radiation Effects and Defects in Solids 170, 651-658. [CrossRef]
Downloaded from www.ajronline.org by 54.210.20.124 on 10/28/15 from IP address 54.210.20.124. Copyright ARRS. For personal use only; all rights reserved
3. Jae-Hwan Cho, Myung-Sam Kim. 2015. Performance Analysis of Low-level Radiation Shielding Sheet with Diamagnetic
Nanoparticles. Journal of Magnetics 20, 103-109. [CrossRef]
4. Marie-Pierre Revel, Isabelle Fitton, Etienne Audureau, Joseph Benzakoun, Mathieu Lederlin, Marie-Laure Chabi, Pascal
Rousset. 2015. Breast Dose Reduction Options During Thoracic CT: Influence of Breast Thickness. American Journal of
Roentgenology 204:4, W421-W428. [Abstract] [Full Text] [PDF] [PDF Plus]
5. Ulla Nikupaavo, Touko Kaasalainen, Vappu Reijonen, Sanna-Mari Ahonen, Mika Kortesniemi. 2015. Lens Dose in
Routine Head CT: Comparison of Different Optimization Methods With Anthropomorphic Phantoms. American Journal of
Roentgenology 204:1, 117-123. [Abstract] [Full Text] [PDF] [PDF Plus]
6. William Moore, Michael J. Bonvento, Dwight Lee, Jared Dunkin, Priya Bhattacharji. 2015. Reduction of Fetal Dose in
Computed Tomography Using Anterior Shields. Journal of Computer Assisted Tomography 39, 298-300. [CrossRef]
7. Stephen Inkoom, Maria Raissaki, Kostas Perisinakis, Thomas G. Maris, John Damilakis. 2015. Location of radiosensitive
organs inside pediatric anthropomorphic phantoms: Data required for dosimetry. Physica Medica . [CrossRef]
8. Ehsan Samei. 2014. Pros and cons of organ shielding for CT imaging. Pediatric Radiology 44, 495-500. [CrossRef]
9. Jack Lambert, John D. MacKenzie, Dianna D. Cody, Robert Gould. 2014. Techniques and Tactics for Optimizing CT Dose in
Adults and Children: State of the Art and Future Advances. Journal of the American College of Radiology 11, 262-266. [CrossRef]
10. P. Akhlaghi, H. Miri Hakimabad, L. Rafat Motavalli. 2014. An overview of exposure parameters, dose measurements and
strategies for dose reduction in pediatric CT examinations. Radioprotection 49, 9-15. [CrossRef]
11. Cynthia H. McCollough. 2013. Standardization Versus Individualization. Health Physics 105, 445-453. [CrossRef]
12. Sabah Servaes, Xiaowei Zhu. 2013. The effects of bismuth breast shields in conjunction with automatic tube current modulation
in CT imaging. Pediatric Radiology 43, 1287-1294. [CrossRef]
13. S. J. Foley, M. F. McEntee, L. A. Rainford. 2013. An evaluation of in-plane shields during thoracic CT. Radiation Protection
Dosimetry 155, 439-450. [CrossRef]
14. Steven Birnbaum. 2013. CT Dose Optimization: The General Radiologist's Perspective. Journal of the American College of
Radiology 10, 637-638. [CrossRef]
15. Claudia Zacharias, Adam M. Alessio, Randolph K. Otto, Ramesh S. Iyer, Grace S. Philips, Jonathan O. Swanson, Mahesh M.
Thapa. 2013. Pediatric CT: Strategies to Lower Radiation Dose. American Journal of Roentgenology 200:5, 950-956. [Abstract]
[Full Text] [PDF] [PDF Plus]
16. Frédéric A. Miéville, Laureline Berteloot, Albane Grandjean, Paul Ayestaran, François Gudinchet, Sabine Schmidt, Francis
Brunelle, François O. Bochud, Francis R. Verdun. 2013. Model-based iterative reconstruction in pediatric chest CT: assessment
of image quality in a prospective study of children with cystic fibrosis. Pediatric Radiology 43, 558-567. [CrossRef]
17. Patrick M. Colletti, Orlando A. Micheli, Kai H. Lee. 2013. To Shield or Not to Shield: Application of Bismuth Breast Shields.
American Journal of Roentgenology 200:3, 503-507. [Abstract] [Full Text] [PDF] [PDF Plus]
18. Rafel Tappouni, Bradley Mathers. 2013. Scan Quality and Entrance Skin Dose in Thoracic CT: A Comparison between
Bismuth Breast Shield and Posteriorly Centered Partial CT Scans. ISRN Radiology 2013, 1-6. [CrossRef]
19. Jie Zhang, M. Elizabeth Oates. 2012. CT Bismuth Breast Shielding: Is It Time to Make Your Own Decision?. Journal of the
American College of Radiology 9, 856-858. [CrossRef]
20. P-L. Khong, D. Frush, H. Ringertz. 2012. Radiological protection in paediatric computed tomography. Annals of the ICRP
41, 170-178. [CrossRef]
21. A. Hojreh, H. Prosch. 2012. Dosisoptimierung bei CT-Untersuchungen von Kindern. Der Radiologe 52, 927-933. [CrossRef]
22. Jenny K. Hoang, Terry T. Yoshizumi, Kingshuk Roy Choudhury, Giao B. Nguyen, Greta Toncheva, Andreia R. Gafton,
James D. Eastwood, Carolyn Lowry, Lynne M. Hurwitz. 2012. Organ-Based Dose Current Modulation and Thyroid Shields:
Techniques of Radiation Dose Reduction for Neck CT. American Journal of Roentgenology 198:5, 1132-1138. [Abstract] [Full
Text] [PDF] [PDF Plus]
23. E. Dougeni, K. Faulkner, G. Panayiotakis. 2012. A review of patient dose and optimisation methods in adult and paediatric CT
scanning. European Journal of Radiology 81, e665-e683. [CrossRef]
24. Andrew J. Einstein, Carl D. Elliston, Daniel W. Groves, Bin Cheng, Steven D. Wolff, Gregory D. N. Pearson, M. Robert Peters,
Lynne L. Johnson, Sabahat Bokhari, Gary W. Johnson, Ketan Bhatia, Theodore Pozniakoff, David J. Brenner. 2012. Effect
of bismuth breast shielding on radiation dose and image quality in coronary CT angiography. Journal of Nuclear Cardiology
19, 100-108. [CrossRef]
25. Cynthia H. McCollough, Jia Wang, Robert G. Gould, Colin G. Orton. 2012. The use of bismuth breast shields for CT should
Downloaded from www.ajronline.org by 54.210.20.124 on 10/28/15 from IP address 54.210.20.124. Copyright ARRS. For personal use only; all rights reserved
46. Sangroh Kim, Donald P. Frush, Terry T. Yoshizumi. 2010. Bismuth shielding in CT: support for use in children. Pediatric
Radiology 40, 1739-1743. [CrossRef]
47. Geoffrey P. Schembri, Anne E. Miller, Richard Smart. 2010. Radiation Dosimetry and Safety Issues in the Investigation of
Pulmonary Embolism. Seminars in Nuclear Medicine 40, 442-454. [CrossRef]
48. Rutger A. J. Nievelstein, Ingrid M. Dam, Aart J. Molen. 2010. Multidetector CT in children: current concepts and dose
reduction strategies. Pediatric Radiology 40, 1324-1344. [CrossRef]
49. Jee-Eun Kim, Beverley Newman. 2010. Evaluation of a Radiation Dose Reduction Strategy for Pediatric Chest CT. American
Journal of Roentgenology 194:5, 1188-1193. [Abstract] [Full Text] [PDF] [PDF Plus]
50. K.-H. Chang, W. Lee, D.-M. Choo, C.-S. Lee, Y. Kim. 2010. Dose reduction in CT using bismuth shielding: measurements
and Monte Carlo simulations. Radiation Protection Dosimetry 138, 382-388. [CrossRef]
51. M. Street, Z. Brady, B. Van Every, K. R. Thomson. 2009. Radiation exposure and the justification of computed tomography
scanning in an Australian hospital emergency department. Internal Medicine Journal 39:10.1111/imj.2009.39.issue-11, 713-719.
[CrossRef]
52. Martin L. Gunn, Kalpana M. Kanal, Orpheus Kolokythas, Yoshimi Anzai. 2009. Radiation Dose to the Thyroid Gland and
Breast From Multidetector Computed Tomography of the Cervical Spine. Journal of Computer Assisted Tomography 33, 987-990.
[CrossRef]
53. Donald P. Frush. 2009. Radiation safety. Pediatric Radiology 39, 385-390. [CrossRef]
54. Jong-Seok Lee, Dae-Cheol Kweon, Beong-Gyu You. 2009. Radiation Dose Reducing Effect during the AEC System in the
Chest and Abdomen of the MDCT Scanning. The Journal of the Korea Contents Association 9, 225-231. [CrossRef]
55. S. Kim, T. T. Yoshizumi, D. P. Frush, C. Anderson-Evans, G. Toncheva. 2009. Dosimetric characterisation of bismuth shields
in CT: measurements and Monte Carlo simulations. Radiation Protection Dosimetry 133, 105-110. [CrossRef]
56. Lynne M. Hurwitz, Terry T. Yoshizumi, Philip C. Goodman, Rendon C. Nelson, Greta Toncheva, Giao B. Nguyen, Carolyn
Lowry, Colin Anderson-Evans. 2009. Radiation Dose Savings for Adult Pulmonary Embolus 64-MDCT Using Bismuth Breast
Shields, Lower Peak Kilovoltage, and Automatic Tube Current Modulation. American Journal of Roentgenology 192:1, 244-253.
[Abstract] [Full Text] [PDF] [PDF Plus]
57. Donald P. Frush. 2008. PEDIATRIC DOSE REDUCTION IN COMPUTED TOMOGRAPHY. Health Physics 95, 518-527.
[CrossRef]
58. Nikhil Bharat Shah, Shari L Platt. 2008. ALARA: is there a cause for alarm? Reducing radiation risks from computed
tomography scanning in children. Current Opinion in Pediatrics 20, 243-247. [CrossRef]
59. Donald P. Frush. 2008. Pediatric abdominal CT angiography. Pediatric Radiology 38, 259-266. [CrossRef]