You are on page 1of 7

M e d i c a l P hy s i c s a n d I n f o r m a t i c s • O r i g i n a l R e s e a r c h

Christianson et al.
Noise Measurement in Clinical CT

Medical Physics and Informatics


Original Research
Downloaded from www.ajronline.org by 200.28.58.4 on 05/09/22 from IP address 200.28.58.4. Copyright ARRS. For personal use only; all rights reserved

Automated Technique to Measure


Noise in Clinical CT Examinations
Olav Christianson1 OBJECTIVE. The purpose of this study was to develop and validate an automated meth-
James Winslow 1 od to measure noise in clinical CT examinations.
Donald P. Frush2 MATERIALS AND METHODS. An automated algorithm was developed to measure
Ehsan Samei1 noise in CT images. To assess its validity, the global noise level was compared with image
noise measured using an image subtraction technique in an anthropomorphic phantom. The
Christianson O, Winslow J, Frush DP, Samei E global noise level was further compared with image noise values from clinical patient CT im-
ages obtained by an observer study. Finally, the clinical utility of the global noise level was
shown by assessing variability of image noise across scanner models for abdominopelvic CT
examinations performed in 2358 patients.
RESULTS. The global noise level agreed well with the phantom-based and clinical image–
based noise measurements, with an average difference of 3.4% and 4.7% from each of these
measures, respectively. No significant difference was detected between the global noise level
and the validation dataset in either case. It further indicated differences across scanners, with
the median global noise level varying significantly between different scanner models (15–35%).
CONCLUSION. The global noise level provides an accurate, robust, and automated
method to measure CT noise in clinical examinations for quality assurance programs. The
significant difference in noise across scanner models indicates the unexploited potential to
efficiently assess and subsequently improve protocol consistency. Combined with other auto-
mated characterization of imaging performance (e.g., dose monitoring), the global noise lev-
el may offer a promising platform for the standardization and optimization of CT protocols.

M
ore than 70 million CT examina- comparing it across institutions and scanner
tions are performed annually in models is becoming routine clinical practice.
the United States, and CT now For example, The Joint Commission will re-
Keywords: CT, dose monitoring, image quality, protocol accounts for approximately 25% quire recording the radiation dose for every
optimization, quality assurance of the annual radiation exposure to the U.S. CT study to maintain accreditation effec-
population [1, 2]. Accordingly, recent studies tive July 1, 2015 [5]. Although radiation dose
DOI:10.2214/AJR.14.13613
have suggested that radiation exposure from tracking is a valuable tool for identifying ra-
Received August 6, 2014; accepted after revision CT examinations may be associated with an diation overdoses and evaluating operation-
December 31, 2014. increased risk of developing cancer [3, 4]. al consistency, this monitoring provides no
Therefore, there is a growing emphasis placed direct information on achieving the desired
1
Ravin Advanced Imaging Laboratories, Duke University, on using the minimum radiation dose neces- image quality [6]. Importantly, because im-
2424 Erwin Rd, Ste 302, Durham, NC 27705. Address
correspondence to O. Christianson
sary. However, using a relatively low radiation age quality is affected by several reconstruc-
(olav.christianson@gmail.com). dose may have a detrimental impact on diag- tion parameters (e.g., kernel, slice thickness,
nostic accuracy. Therefore, research and clini- iterative reconstruction, and display FOV)
2
Department of Radiology, Duke University Medical cal efforts are essential to optimize CT proto- that are not factors in the delivered radiation
Center, Durham, NC.
cols to deliver the minimum radiation dose dose, radiation dose tracking alone is insuffi-
WEB while maintaining diagnostic image quality. cient to determine whether CT protocols are
This is a web exclusive article. In a clinical environment, it is a daunt- consistent or optimized [7].
ing task to ensure that the multitude of ever- An ideal approach to optimize and moni-
AJR 2015; 205:W93–W99 changing CT protocols is consistent across tor CT protocols is to directly evaluate the
0361–803X/15/2051–W93
a fleet of CT scanners to maintain a proper level of image quality in clinical images.
balance of radiation dose and image quali- However, current methods to evaluate image
© American Roentgen Ray Society ty. To this end, tracking radiation dose and quality in patient CT images (e.g., assessing

AJR:205, July 2015 W93


Christianson et al.

600

500

400

Frequency
300
Downloaded from www.ajronline.org by 200.28.58.4 on 05/09/22 from IP address 200.28.58.4. Copyright ARRS. For personal use only; all rights reserved

200

100

0
0 20 40 60 80 100
SD (HU)

A B C
Fig. 1—Noise detection process.
A, Original CT image of patient.
B, Noise map shows soft-tissue regions.
C, Histogram shows noise map for soft-tissue regions. Mode of histogram corresponds to global noise level.

noise, contrast, or the ratio of the two) are cifically, the aims of this work were to develop The Noise Detection Algorithm
highly labor intensive. For example, methods an automated computationally efficient meth- We developed an automated method to calcu-
for noise assessment rely on manual selection od for measuring the global noise level from late noise in patient CT images. Because CT noise
of ROIs in uniform areas to measure image patient CT images, to validate this method varies with location in the image, noise was char-
noise [8], making them impractical for large- in a complex abdominal-thoracic surrogate acterized in terms of the most frequent noise level
scale quality assurance programs. A more re- phantom as well as in clinical patient images, in areas of homogeneous tissue. The metric was
cently reported automated platform is limit- and to show how this new metric can be used termed the “global noise level.”
ed by its reliance on phantom measurements as a tool to evaluate the consistency of image The global noise level was calculated in three
and the complexity associated with image noise across multiple CT scanner models. steps. First, the image was segmented into dif-
quality in nonuniform patient images [9, 10]. ferent tissue types (Fig. 1). The tissue types were
Automated approaches to assess image qual- Materials and Methods roughly separated into four categories: aerated tis-
ity in chest radiography have been notewor- In accordance with our institutional policy, this sue (HU< −800), fat (−300 to < 0 HU), soft-tissue
thy [11]. However, such methods have not quality improvement activity did not constitute (0 < to 100 HU), and bone (> 300 HU). For the
been implemented for CT. human subject research because it pertained to purposes of this study, we limited our analysis to
The objective of this study was to take the a clinical quality control initiative. All data were the soft-tissue components of the image.
first step toward automated assessment of im- anonymized, and the project was in compliance Second, a noise map was generated in which the
age quality in clinical CT images. More spe- with HIPAA guidelines. value at each location corresponded to the SD in an

A B
Fig. 2—Meleagris gallopavo phantom.
A, Representative CT image of phantom shows homogeneous areas as well as regions containing bone, air cavities, and fluid recesses. Boxes indicate ROIs used for
noise measurement in subtracted image.
B, CT image resulting from subtraction of repeated series. Boxes indicate ROIs used for noise measurement in subtracted image.

W94 AJR:205, July 2015


Noise Measurement in Clinical CT

TABLE 1: Image Acquisition Parameters for Cadaver Phantom ROIs did not overlap with any anatomic structure.
The ROIs were made as large as possible without
Parameter Settings
including adjacent anatomic structures and were
Dose levels (% of default clinical setting) 100, 33, and 17 spaced as evenly as possible throughout the pa-
Slice thicknesses (mm) 5 and 1 tient. The noise in the image was determined by
computing the average SD across the ROIs. Fi-
Reconstruction kernels B31f, B35f, B71f, I31fa, I36fa, I75fa
nally, image noise measured by the observers was
Downloaded from www.ajronline.org by 200.28.58.4 on 05/09/22 from IP address 200.28.58.4. Copyright ARRS. For personal use only; all rights reserved

aUses SAFIRE (Sinogram Affirmed Iterative Reconstruction) 3 (Siemens Healthcare).


compared with the global noise level. The Pearson
correlation coefficient was calculated between the
area (i.e., kernel) surrounding that pixel. A design thoracic cavities was needed. Therefore, a cadaver two datasets and used to evaluate the differences
consideration was the kernel size. Kernels that are too of Meleagris gallopavo (turkey) was selected for between the two datasets, with a p value less than
small may not fully sample the low-frequency com- the validation (Fig. 2). This test object combined 0.05 indicating a significant difference.
ponents of the noise, whereas those that are too large homogeneous tissue regions with inhomogeneous
may be overly contaminated with anatomic transi- areas, air cavities, bones, and fluid recesses, mak- Application: Variability in Noise Across
tions. Multiple kernel sizes from 2 to 30 mm were ing it an effective platform on which to validate Scanner Models
investigated across abdominal and thoracic CT pa- the global noise level algorithm. To illustrate the clinical utility of the global
tient images to determine which would give the most The phantom was scanned on a commercial noise level, the global noise level was calculated
robust and accurate results that were relatively in- CT scanner (Somatom Definition Flash, Siemens for 2358 patients imaged on three commercial CT
sensitive to the size. A convolution method was then Healthcare) using two slice thicknesses, three dose scanner models (Discovery CT 750HD [n = 512]
used to calculate the SD for each kernel. The aver- levels, and six reconstruction kernels (Table 1). For and LightSpeed VCT [n = 1644], both GE Health-
age computing time for this processing on a standard each set of scanning parameters, the phantom was care and Somatom Definition Flash [n = 202], Sie-
commercially available laptop computer (ThinkPad scanned twice; the repeated image series were sub- mens Healthcare) using the abdominopelvic pro-
T420, Lenovo) was 0.05 seconds per image. tracted to remove the anatomic structures and iso- tocol (Table 2). To minimize processing time, 10
Third, a histogram was generated for the SDs late the quantum noise in the image. Six ROIs were evenly spaced images were selected from each
corresponding to the targeted tissue type (soft tis- uniformly distributed throughout the soft tissue in study for analysis. The global noise level was aver-
sue). The histogram comprised two principal cat- a representative slice of the phantom (Fig. 2). The aged across those 10 images to provide a represen-
egories: homogeneous areas and transitional areas. noise was averaged across the six ROIs and com- tative noise level for the examination as a whole.
Transitional areas, such as the interface between pared with the global noise level. The Spearman For comparison purposes, the patient effective di-
soft-tissue and bone, constituted a surprisingly small rank correlation coefficient was calculated between ameters and size-specific dose estimates (SSDEs)
portion of the image. Furthermore, because they the two datasets and used to evaluate the differ­ were determined using a previously published
were at the boundary of multiple tissue types, they ences between the two datasets, with a p value less dose monitoring program [6]. ANOVA was used
exhibited higher and more variable ranges of SDs than 0.05 indicating a significant difference. to determine the significance of the difference in
compared with those of homogeneous areas. As a effective diameter, SSDE, and global noise level
result, the histogram had a characteristic high nar- Validation: Patient Images between scanner models (p < 0.05 indicated sig-
row peak corresponding to homogeneous areas and An observer study was performed to further nificance). Because of the large number of patient
a short broad tail corresponding to transitional ar- validate that the global noise level is indicative cases, even very small differences across scanner
eas. The global noise level was determined by iden- of image noise in patient images. Six clinical pa- models may appear significant. Therefore, in ad-
tifying the mode of the histogram peak correspond- tient images, three abdominal and three thoracic, dition to significance testing, the Cohen f statistic
ing to homogeneous tissue. were randomly selected from our clinical dataset. was used to investigate the effect size across scan-
All six images were acquired on a commercial CT ner models for effective diameter, SSDE, and glob-
Validation: Phantom Images scanner (CT 750 HD, GE Healthcare) using clin- al noise level with the following evaluation criteria:
A phantom study was conducted to perform the ical abdominal and chest imaging protocols. All f > 0.10 is a small effect, > 0.25 is a medium effect,
initial validation of the global noise level in repre- images were acquired using the standard clinical and > 0.40 is a large effect [12].
senting image noise. Although several anthropo- protocol; the abdominal images were contrast en-
morphic phantoms were commercially available, hanced. Four physicists took part in the observer Results
none possessed the same level of inhomogeneity study. Previous studies have shown that observ- Stability of the Global Noise Level With
present in patient images. To better evaluate the ers can accurately measure noise in patient im- Kernel Size
global noise level algorithm, a test object more ages [8]. Each observer was instructed to place As noted, first the sensitivity of the global
similar in complexity to human abdominal and five ROIs in areas of soft tissue in a way that these noise level to kernel size was investigated by

TABLE 2: Abdominopelvic Protocols


Reconstruction Slice Noise Index or Quality Gantry Rotation Reconstruction
Scanner Model Name kVp Thickness (mm) Pitch Reference mAs Time (s) Kernel
LightSpeed VCT 120 5 1.375 15 0.5 FBP, standard
Discovery CT 750 HD 120 5 1.375 22 0.5 ASIR 40%, standard
Somatom Definition Flash 120 5 0.8 200 0.5 SAFIRE 3, I31f
Note—FBP = filtered back projection, ASIR = adaptive statistical iterative reconstruction. LightSpeed VCT and Discovery CT 750 HD manufactured by GE Healthcare,
Somatom Definition Flash manufactured by Siemens Healthcare. SAFIRE (Sinogram Affirmed Iterative Reconstruction) 3 is a product of Siemens Healthcare.

AJR:205, July 2015 W95


Christianson et al.

tively. Similarly, changing the slice thickness


70 70
from 5 to 1 mm was expected to increase the
60 60 noise by the square root of five (123% in-
Global Noise Level (HU)

Global Noise Level (HU)


crease in noise). The image noise and global
50 50
noise level, however, only increased by 97%
40 40 and 84%, respectively.
Downloaded from www.ajronline.org by 200.28.58.4 on 05/09/22 from IP address 200.28.58.4. Copyright ARRS. For personal use only; all rights reserved

30 30
Validation: Patient Images
20 20 The global noise level was compared with
image noise determined from an observer
10 10 study in six clinical patient images (Fig. 5).
0 0
Across the six images, the average difference
0 10 20 30 0 10 20 30 between the global noise level and observ-
Kernel Size (mm) Kernel Size (mm)
er noise measurement was 4.7%, which was
A B less than the intraobserver variability of 6.2%.
Fig. 3—Sensitivity of global noise level to ROI size. The agreement between the global noise level
A and B, Graphs show how global noise level varies with ROI size for abdominal image (A) and thoracic image and observer measured noise was slightly bet-
(B). In both cases, there is stable region from around 6–20 mm. ter in the abdominal images than in the tho-
racic images (3.9% and 5.4% for the abdom-
applying the noise detection algorithm to ab- manlike tissues and organs (Fig. 4). Across inal and thoracic images, respectively). The
dominal and thoracic patient CT images using all image acquisition parameters, the average correlation between the two datasets was sig-
a range of ROI sizes. In general, the measured absolute difference between the global noise nificant (ρ = 0.9979 and p = 0.001).
noise level increased with kernel size up to 6 level and image noise was 3.4%. The larg-
mm (Fig. 3). From 6 to 20 mm, the measured est difference between the two measures was Application: Variability in Noise Across
noise level was relatively constant. Beyond 12.0%, which occurred at the highest dose Scanner Models
20 mm, the measured noise level increased level using iterative reconstruction resulting Effective diameter, SSDE and global noise
abruptly as the number of ROIs composed in the image with the lowest overall noise level were determined for 2358 patients im-
completely of homogeneous tissue dropped to magnitude. In this case, the global noise lev- aged on three CT scanner models using the
a critically low level. Beyond this threshold, el differed from the image noise by only 0.7 abdominopelvic protocol (Fig. 6). The median
the measured noise levels were corrupted with HU. The correlation between the two data­ effective diameter differed by 2–8% between
the inability of large kernels to avoid anatom- sets was significant (ρ = 0.9991 and p < 0.01). the scanner models. Although the difference
ic transitions. On the basis of these findings, a Although the global noise level and image in effective diameter across scanner models
6-mm kernel size was determined as ideal for noise agreed well, they did not always match was significant (p < 0.001), the effect size was
the global noise level measurement. the expected noise based on slice thickness small (f = 0.129) indicating minimal practi-
and dose level. For example, decreasing the cal significance in terms of patient diameters
Validation: Phantom Images dose to 50% was expected to increase the across scanner models. The median SSDE
The global noise level was compared with noise by the square root of two (41% increase differed by 9–33% between scanner models.
image noise measured using subtraction in noise). The global noise level and image The difference in SSDE across scanner mod-
techniques in a phantom consisting of hu- noise increased by 36% and 34%, respec- els was significant (p = < 0.001) and the effect

90 20 16
80 18 14
70 16
12
60 14
Noise (HU)

10
Noise (HU)

Noise (HU)

12
50
10 8
40
8
30 6
6
20 4
4
10 2 2
0 0 0
B31f B35f B75f l31f l36f l70f 100%
50% 17% 5 mm 1 mm
Reconstruction Kernel Dose Level Slice Thickness
Global Noise Level Subtraction Method Global Noise Level Subtraction Method Global Noise Level Subtraction Method
A B C
Fig. 4—Phantom validation.
A–C, Graphs show comparison of global noise level with noise determined from subtracted images across reconstruction kernel (A), dose level (B), and slice thickness (C).
(Fig. 4 continues on next page)

W96 AJR:205, July 2015


Noise Measurement in Clinical CT

100
Fig. 4 (continued)— ity control program. Finally, such a quality
Phantom validation. control program facilitates characterization
D, Graph shows global
noise level versus of consistency across multiple platforms, op-
noise measured from erators, shifts, and institutions.
80 subtracted images for all Despite the clinical utility of the global
scanning combinations.
Each point corresponds
noise level and its strong agreement with noise
to noise measured from measured in phantom and clinical patient im-
Global Noise Level (HU)
Downloaded from www.ajronline.org by 200.28.58.4 on 05/09/22 from IP address 200.28.58.4. Copyright ARRS. For personal use only; all rights reserved

60 unique set of scanning ages, there are several limitations of this ap-
parameters and solid proach. First, although large-scale anatomic
line represents identity
line. structures are avoided by the histogram anal-
ysis, small-scale anatomic texture may con-
40
tribute to the global noise level. The strong
agreement between the global noise level and
the noise measured in the phantom using the
20 subtraction technique, which should remove
most anatomic contributions, suggests that
the amount of anatomic structure was small.
On the other hand, the amount of anatom-
0
0 20 40 60 80 100
ic structure likely varies between individu-
Subtraction Method (HU) als, and the phantom used in this study is not
D representative of the spectrum of patient size
and tissue distributions seen clinically. Fur-
size was moderate (f = 0.265), reflective of ac- new paradigm in quality metrology that can thermore, the observer study showed a strong
tual differences between the dosing protocols be used toward a variety of goals. First, it can agreement between the global noise level and
across scanners. The median global noise lev- be used for granular estimation of quality on observer measurements even in the presence
el differed by 15–35% between scanner mod- an image-by-image basis. For example, this of contrast enhancement, which should em-
els. The difference in global noise level across method enables characterization of image phasize the small-scale anatomic structure.
scanner models was significant (p = < 0.001) noise across slices of a CT series. Second, Conversely, anatomic structure contamina-
and the effect size was large (f = 0.463), indi- this approach enables individualized retro- tion at low noise levels could explain the find-
cating a high degree of practical significance spective assessment of the diagnostic opera- ing that the global noise level did not increase
due to differences in both the imaging proto- tion, whereas other techniques treat all pa- as expected with changes in slice thickness
cols and system attributes. tients similarly. Third, patient image quality and dose levels. More work is needed to de-
data facilitate prospective definition of pro- termine the full impact of small-scale ana-
Discussion tocols to achieve the desired diagnostic accu- tomic structures.
We developed an automated computa- racy. Fourth, because image quality and ra- Second, this technique measures noise
tionally efficient method to measure the diation dose are inversely related, measuring magnitude but ignores the spatial correla-
global noise level from patient CT images. both is necessary for a comprehensive qual- tions of the noise. In general, low-frequen-
This method segments the image into tissue
types, generates an SD map, and uses histo-
20
gram analysis to determine the global noise
level in homogeneous regions of soft tissue. 18
We found that the global noise level mea-
sured using this method agreed to within 5% 16

of noise measurements made in a complex 14


Global Noise Level (HU)

anthropomorphic phantom as well as those


from observer assessments in clinical CT ex- 12
aminations. Further, we implemented this
10
algorithm in routine quality control practic- Fig. 5—Observer
es. Sampling 10 images from each series for validation. On graph
of global noise level 8
global noise level measurement kept process- versus noise determined
ing time under 1 second for each acquisition, by human observers, 6
enabling this method to be applied to every each point represents
average across all 4
patient undergoing a CT examination. observers for given
A notable advantage of our findings is that image, and error 2
image quality can be evaluated in terms of bars correspond to
the actual attributes of clinical images rather SD across observer 0
measurements. Solid 0 5 10 15 20
than idealized phantoms. Measuring image line represents identity Observer Noise Measurement (HU)
quality using actual clinical images offers a line.

AJR:205, July 2015 W97


Christianson et al.

45 45
+ +
+
+
+
40 +
+
+
+
40 + +
+
35 +
+
+
+
+ +
+ +
+
Effective Diameter (cm)

30 +
+ +
+
+ +
+
+
Downloaded from www.ajronline.org by 200.28.58.4 on 05/09/22 from IP address 200.28.58.4. Copyright ARRS. For personal use only; all rights reserved

35

SSDE (mGy)
25
+
20 +
30 +
15

10
25

+
+ 5
+ +
20 0
Discovery Lightspeed Somatom Discovery Lightspeed Somatom
CT 750 HD VCT Definition CT 750 HD VCT Definition
Flash Flash

A B
Fig. 6—Variability in surance of resulting clinical images that can
35 + effective diameter, size-
+
+
aid in protocol optimization. The global noise
+
+
specific dose estimate
level also offers a simple retrospective tool to
+
+ (SSDE), and global noise
30 +
+ level across scanner compare image noise in patient images across
+
+ models.
+
+ +
+
+
scanner models. As expected, the SSDE var-
+ +
+ A–C, Box-and-whisker
ied with scanner model and reconstruction al-
25 +
+ + plots show median
+
Global Noise Level (HU)

and 25th and 75th gorithm, thereby making the SSDE difficult
percentiles of effective to use as a primary quality control metric. Be-
20 diameter (A), SSDE (B), cause the noise required for diagnostic image
+
+ and global noise level (C)
for each scanner model. quality depends to a much lesser extent on
15 Outliers (indicated scanner model and reconstruction algorithm,
with + marker) were the global noise level is better suited for estab-
those that exceeded
lishing diagnostic reference levels [20].
10 +
+ 1.5 times interquartile
length. Discovery CT The benefits of the global noise level
+ 750 HD and LightSpeed method are exemplified when comparing
5 VCT manufactured by
GE Healthcare, and
protocols across scanner models. The global
Somatom Definition noise level varied by 15–35% between scan-
0 Flash manufactured by ner models in this study, indicating the possi-
Discovery Lightspeed Somatom Siemens Healthcare. bility of improving the consistency of image
CT 750 HD VCT Definition
Flash quality and reducing patient dose. Assuming
the images from the LightSpeed VCT scan-
C
ner are of diagnostic quality, the noise could
cy noise content, resulting in a coarse lumpy derive some insight into these attributes from be increased on the CT 750 HD and the So-
appearance, is more detrimental to diagnos- patient CT images; however, methods to do matom Definition Flash scanners according-
tic performance than high-frequency noise so have yet to be developed. Currently, these ly. Adjusting the protocols to achieve con-
content, resulting in a fine sandy appearance image attributes must be measured in phan- sistent noise across scanner models would
[13–17]. More work is needed to develop a toms and extrapolated to patient images. decrease the SSDE by 27% and 45% on the
robust technique to quantify the noise tex- They can then be combined with the global CT 750 HD and Somatom Definition Flash
ture present in patient CT images. In the in- noise level through the use of computational scanners, respectively. This newfound po-
terim, the noise texture measured in a uni- observer models to provide a meaningful de- tential for dose reduction shows the potential
form background can be combined with the tectability index that is related to the likeli- clinical impact of an automated image qual-
global noise level to fully characterize the hood of an observer being able to detect the ity monitoring program.
noise present in the image. presence of a lesion [18, 19].
Third, many factors other than noise con- Even with these limitations, the benefits of Conclusion
tribute to image appearance, most notably, using the global noise level are substantial. This study involved the development of an
resolution and contrast. It may be possible to This innovative technique permits quality as- automated computationally-efficient method

W98 AJR:205, July 2015


Noise Measurement in Clinical CT

to measure the global noise level for CT im- mography scans in childhood or adolescence: time has come. Educ Psychol Meas 1996; 56:746–759
ages. The global noise level measured using data linkage study of 11 million Australians. BMJ 13. Boedeker KL, Cooper VN, McNitt-Gray MF. Ap-
this new method agreed well with noise mea- 2013; 346:f2360 plication of the noise power spectrum in modern
surements made using an anthropomorphic 5. Joint Commission. New/revised diagnostic imag- diagnostic MDCT. Part I. Measurement of noise
phantom containing humanlike tissues and ing standards from Joint Commission. J Nucl Med power spectra and noise equivalent quanta. Phys
organs as well as those from patient CT im- 2014; 55:13N Med Biol 2007; 52:4027–4046
ages. Furthermore, we showed how the glob- 6. Christianson O, Li X, Frush D, Samei E. Auto- 14. Chew E, Weiss GH, Brooks RA, Di Chiro G. Ef-
Downloaded from www.ajronline.org by 200.28.58.4 on 05/09/22 from IP address 200.28.58.4. Copyright ARRS. For personal use only; all rights reserved

al noise level can be applied to a large num- mated size-specific CT dose monitoring program: fect of CT noise on detectability of test objects.
ber of patients to analyze the consistency of assessing variability in CT dose. Med Phys 2012; AJR 1978; 131:681–685
image noise between scanner models. The 39:7131–7139 15. Burgess AE. Statistical efficiency of perceptual
automated technique to measure CT noise 7. Goldman LW. Principles of CT: radiation dose decisions. (abstr) SPIE website. proceedings.
in patient images offers a promising new ap- and image quality. J Nucl Med Technol 2007; spiedigitallibrary.org/proceeding.aspx?articleid=
proach for monitoring image noise on a per- 35:213–225 1236924, Published June 15, 1984. Accessed
patient basis, providing a platform to facil- 8. Kalra MK, Maher MM, Kamath RS, et al. Six- March 24, 2015
itate standardization of image quality and teen-detector row CT of abdomen and pelvis: 16. Hanson KM. Detectability in computed tomo-
optimization of imaging protocols. study for optimization of z-axis modulation tech- graphic images. Med Phys 1979; 6:441–451
nique performed in 153 patients. Radiology 2004; 17. Solomon JB Bochud F, Samei E. Design of an-
References 233:241–249 thropomorphic textured phantoms for CT perfor-
1. Brenner DJ, Hall EJ. Computed tomography: an 9. Larson DB, Malarik RJ, Hall SM, Podberesky DJ. mance evaluation. (abstr) spie.org/Publications/
increasing source of radiation exposure. N Engl J System for verifiable CT radiation dose optimiza- Proceedings/Paper/10.1117/12.2043555. Pub-
Med 2007; 357:2277–2284 tion based on image quality. Part II. Process con- lished March 19, 2014. Accessed March 24, 2015
2. Schauer DA, Linton OW. NCRP Report No. 160, trol system. Radiology 2013; 269:177–185 18. Burgess AE, Jacobson FL, Judy PF. Human ob-
ionizing radiation exposure of the population of 10. Solomon JB, Samei E. Are uniform phantoms server detection experiments with mammograms
the United States: medical exposure—are we do- sufficient to characterize the performance of itera- and power-law noise. Med Phys 2001; 28:419–437
ing less with more, and is there a role for health tive reconstruction in CT? (abstr) SPIE website. 19. Gang GJ, Lee J, Stayman JW, et al. Analysis of
physicists? Health Phys 2009; 97:1–5 proceedings.spiedigitallibrary.org/proceeding.aspx? Fourier-domain task-based detectability index in
3. Pearce MS, Salotti JA, Little MP, et al. Radiation articleid=1662019. Published March 13, 2013. tomosynthesis and cone-beam CT in relation to
exposure from CT scans in childhood and subse- Accessed March 24, 2015 human observer performance. Med Phys 2011;
quent risk of leukaemia and brain tumours: a retro- 11. Lin Y, Luo H, Dobbins JT, et al. An image-based 38:1754–1768
spective cohort study. Lancet 2012; 380:499–505 technique to assess the perceptual quality of clinical 20. [No authors listed]. 1990 Recommendations of the
4. Mathews JD, Forsythe AV, Brady Z, et al. Cancer chest radiographs. Med Phys 2012; 39:7019–7031 International Commission on Radiological Protec-
risk in 680,000 people exposed to computed to- 12. Kirk RE. Practical significance: a concept whose tion (Report 60). Ann ICRP 1991; 21:1–2001

AJR:205, July 2015 W99

You might also like