Unsaturated Iron-Binding Capacity
Order information
Analyzer(s) on which cobas c pack(s) can be used
08058776190 Unsaturated Iron‑Binding Capacity (100 tests) System‑ID 2118 001 cobas c 303, cobas c 503
Materials required (but not provided):
12146401216 Fe Standard (1 x 75 mL) Code 20566
05117003190 PreciControl ClinChem Multi 1 (20 x 5 mL) Code 20391
05947626190 PreciControl ClinChem Multi 1 (4 x 5 mL) Code 20391
05117216190 PreciControl ClinChem Multi 2 (20 x 5 mL) Code 20392
05947774190 PreciControl ClinChem Multi 2 (4 x 5 mL) Code 20392
UIBC
Unsaturated Iron-Binding Capacity
The sample types listed were tested with a selection of sample collection Values obtained should fall within the defined limits. Each laboratory should
tubes that were commercially available at the time of testing, i.e. not all establish corrective measures to be taken if values fall outside the defined
available tubes of all manufacturers were tested. Sample collection systems limits.
from various manufacturers may contain differing materials which could Follow the applicable government regulations and local guidelines for
affect the test results in some cases. When processing samples in primary quality control.
tubes (sample collection systems), follow the instructions of the tube
manufacturer. Calculation
Centrifuge samples containing precipitates before performing the assay. cobas c systems automatically calculate the analyte concentration of each
sample in the unit µmol/L (µg/dL, mg/L).
See the limitations and interferences section for details about possible
sample interferences. Conversion factors: µmol/L x 5.59 = µg/dL
Stability:7,8 4 days at 15‑25 °C µmol/L x 0.0559 = mg/L
7 days at 2‑8 °C Limitations - interference
Materials provided Criterion: Recovery within ± 10 % of initial value at an iron concentration of
See “Reagents – working solutions” section for reagents. 60 µmol/L (335 µg/dL).
Icterus:9 No significant interference up to an I index of 60 for conjugated
Materials required (but not provided) and unconjugated bilirubin (approximate conjugated and unconjugated
See “Order information” section bilirubin concentration: 1026 µmol/L or 60 mg/dL).
General laboratory equipment Hemolysis:9 No significant interference up to an H index of 40 (approximate
Assay hemoglobin concentration: 24.8 µmol/L or 40 mg/dL).
For optimum performance of the assay follow the directions given in this Contamination with erythrocytes will elevate results, because the analyte
document for the analyzer concerned. Refer to the appropriate operator’s level in erythrocytes is higher than in normal sera. The level of interference
manual for analyzer‑specific assay instructions. may be variable depending on the content of analyte in the lysed
erythrocytes.
The performance of applications not validated by Roche is not warranted
and must be defined by the user. Lipemia (Intralipid):9 No significant interference up to an L index of 300.
There is poor correlation between the L index (corresponds to turbidity) and
Application for serum and plasma triglycerides concentration.
Test definition Anticoagulants: Complexing anticoagulants such as EDTA, oxalate, and
citrate must not be used.
Reporting time 10 min Drugs: No interference was found at therapeutic concentrations using
Wavelength (sub/main) 700 / 546 nm common drug panels.10, 11
Reagent pipetting Diluent (H2O) Exception: Oxytetracycline causes artificially high UIBC values at the tested
drug level.
R1 41 µL 53 µL Pathologically high levels of albumin (7 g/dL) decrease the apparent UIBC
R3 19 µL 15 µL value significantly.
Sample volumes Sample Sample dilution If the patient’s serum iron exceeds the binding capacity of the transferrin, a
negative UIBC value will result.
Sample Diluent (H2O) In patients treated with iron supplements or metal‑binding drugs, the
Normal 15 µL – – drug‑bound iron may not properly react in the test, resulting in falsely low
values.
Decreased 7.5 µL – –
The physiological function of deferoxamine containing drugs is to bind iron
Increased 15 µL – – to facilitate its elimination from the body. Therefore any deferoxamine
For further information about the assay test definitions refer to the concentration interferes with the UIBC assay.
application parameters setting screen of the corresponding analyzer and In the presence of high ferritin concentrations > 1200 μg/L the assumption
assay. that serum iron is almost completely bound to transferrin is not valid
anymore. Therefore, such iron results should not be used to calculate Total
Calibration Iron Binding Capacity (TIBC) or percent transferrin saturation (% SAT).12
Calibrators S1: H2O In very rare cases, gammopathy, in particular type IgM (Waldenström’s
macroglobulinemia), may cause unreliable results.13
S2: Iron Standard
For diagnostic purposes, the results should always be assessed in
Calibration mode Linear conjunction with the patient’s medical history, clinical examination and other
findings.
Calibration frequency Full calibration
- after reagent lot change ACTION REQUIRED
Special Wash Programming: The use of special wash steps is mandatory
- as required following quality control when certain test combinations are run together on cobas c systems. All
procedures special wash programming necessary for avoiding carry-over is available
Calibration interval may be extended based on acceptable verification of via the cobas link. The latest version of the carry-over evasion list can be
calibration by the laboratory. found with the NaOHD/SMS/SCCS Method Sheet. For further instructions,
refer to the operator’s manual.
Traceability: This method has been standardized against a primary
reference material (weighed in purified material) through iron. Limits and ranges
Quality control Measuring range
For quality control, use control materials as listed in the “Order information” 3-125 µmol/L (16.8‑700 µg/dL, 0.17‑7 mg/L)
section. In addition, other suitable control material can be used. Determine samples having higher concentrations via the rerun function.
The control intervals and limits should be adapted to each laboratory’s Dilution of samples via the rerun function is a 1:2 dilution. Results from
individual requirements. It is recommended to perform quality control samples diluted using the rerun function are automatically multiplied by a
always after lot calibration and subsequently at least every 8 weeks. factor of 2.
Lower limits of measurement
UIBC
Unsaturated Iron-Binding Capacity
Limit of Blank, Limit of Detection and Limit of Quantitation Human serum 1 8.25 0.636 7.7
Limit of Blank = 3 µmol/L (16.8 µg/dL) Human serum 2 24.4 0.674 2.8
Limit of Detection = 3 µmol/L (16.8 µg/dL) Human serum 3 31.3 0.676 2.2
Limit of Quantitation = 3 µmol/L (16.8 µg/dL) Human serum 4 62.2 0.628 1.0
The Limit of Blank, Limit of Detection and Limit of Quantitation were Human serum 5 116 0.742 0.6
determined in accordance with the CLSI (Clinical and Laboratory Standards a) PreciControl ClinChem Multi 1
Institute) EP17‑A2 requirements.
b) PreciControl ClinChem Multi 2
The Limit of Blank is the 95th percentile value from n ≥ 60 measurements of The data obtained on cobas c 503 analyzer(s) are representative for
analyte‑free samples over several independent series. The Limit of Blank cobas c 303 analyzer(s).
corresponds to the concentration below which analyte‑free samples are
found with a probability of 95 %. Method comparison
The Limit of Detection is determined based on the Limit of Blank and the Serum
standard deviation of low concentration samples. UIBC values for human serum and plasma samples obtained on a
The Limit of Detection corresponds to the lowest analyte concentration cobas c 503 analyzer (y) were compared with those determined using the
which can be detected (value above the Limit of Blank with a probability of corresponding reagent on a cobas c 501 analyzer (x).
95 %). Sample size (n) = 65
The Limit of Quantitation is the lowest analyte concentration that can be
reproducibly measured with a total error of 20 %. It has been determined Passing/Bablok15 Linear regression
using low concentration unsaturated iron‑binding capacity samples. y = 1.001x - 0.193 µmol/L y = 1.000x - 0.434 µmol/L
Expected values14
τ = 0.964 r = 0.999
µmol/L
The sample concentrations were between 3.25 and 122 µmol/L.
Females: 24.2‑70.1 µmol/L UIBC values for human serum and plasma samples obtained on a
Males: 22.3‑61.7 µmol/L cobas c 303 analyzer (y) were compared with those determined using the
corresponding reagent on a cobas c 501 analyzer (x).
µg/dL*
Sample size (n) = 61
Females: 135‑392 µg/dL
Passing/Bablok15 Linear regression
Males: 125‑345 µg/dL
y = 1.021x + 0.264 µmol/L y = 1.020x + 0.611 µmol/L
* calculated by unit conversion factor
τ = 0.974 r = 0.999
Serum/plasma iron levels are dependent on diet and subject to circadian
variation. The sample concentrations were between 3.50 and 117 µmol/L.
Each laboratory should investigate the transferability of the expected values References
to its own patient population and if necessary determine its own reference 1 Fairbanks VF, Klee GG. Biochemical aspects of hematology. In: Tietz
ranges. NW, ed. Fundamentals of Clinical Chemistry. 3rd ed. Philadelphia: WB
Specific performance data Saunders 1987:789-824.
Representative performance data on the analyzers are given below. These 2 Bauer JD. Hemoglobin, porphyrin, and iron metabolism. In: Kaplan LA,
data represent the performance of the analytical procedure itself. Pesce AJ, eds. Clinical Chemistry, theory, analysis, and correlation. St.
Results obtained in individual laboratories may differ due to heterogenous Louis: Mosby Company 1984;611-655.
sample materials, aging of analyzer components and mixture of reagents 3 Lauber K, Peheim E, Perritaz R, et al. Latente Eisenbindungskapazität
running on the analyzer. und andere Eisenparameter im Plasma. GIT Labor Medizin
Precision 1991;14:95-96.
Precision was determined using human samples and controls in 4 Stookey LL. FerroZine - a new spectrophotometric reagent for iron.
accordance with the CLSI (Clinical and Laboratory Standards Institute) Anal Chem 1970;42:779-781.
EP05‑A3 requirements with repeatability (n = 84) and intermediate precision 5 Persijn JP, Van der Slik W, Riethorst A. Determination of serum iron
(2 aliquots per run, 2 runs per day, 21 days). Results for repeatability and and latent iron-binding capacity (LIBC). Clin Chim Acta 1971;35:91-98.
intermediate precision were obtained on the cobas c 503 analyzer.
6 Harris WR. Thermodynamics of Anion Binding to Human Serum
Repeatability Mean SD CV Transferrin. Biochemistry 1985;24:7412-7418.
µmol/L µmol/L % 7 Weissman N, Pileggi VJ. Inorganic ions. In: Henry RJ, Cannon DC,
Winkelman W, eds. Clinical Chemistry, Principles and Techniques. 2nd
PCCC1a) 39.2 0.349 0.9 ed. Hagerstown: Harper & Row 1974;4:639-754.
PCCC2b) 50.8 0.299 0.6 8 WHO Publication: Use of anticoagulants in diagnostic laboratory
Human serum 1 8.21 0.305 3.7 investigations, WHO/DIL/LAB/99.1 Rev.2:Jan 2002.
Human serum 2 24.4 0.308 1.3 9 Glick MR, Ryder KW, Jackson SA. Graphical Comparisons of
Interferences in Clinical Chemistry Instrumentation.
Human serum 3 31.5 0.318 1.0 Clin Chem 1986;32:470-475.
Human serum 4 62.5 0.272 0.4 10 Breuer J. Report on the Symposium “Drug effects in Clinical Chemistry
Methods”. Eur J Clin Chem Clin Biochem 1996;34:385-386.
Human serum 5 116 0.408 0.4
11 Sonntag O, Scholer A. Drug interference in clinical chemistry:
recommendation of drugs and their concentrations to be used in drug
interference studies. Ann Clin Biochem 2001;38:376-385.
Intermediate precision Mean SD CV
µmol/L µmol/L % 12 Tietz NW, Rinker AD, Morrison SR. When Is a Serum Iron Really a
Serum Iron? A Follow-up Study on the Status of Iron Measurements in
PCCC1a) 39.0 1.17 3.0 Serum. Clin Chem 1996;42(1):109-111.
PCCC2b) 50.7 1.25 2.5
UIBC
Unsaturated Iron-Binding Capacity