Application

Note: 41260
Analysis of Iron and Steel with desktop
Optical Emission Spectrometer
ARL QuantoDesk CCD-based Metals Analyzer

To these can be added some more specific


Key Words characteristics, including image analysis in discrete
elements (pixels) with exact field registration.
• ARL QuantoDesk
• CCD
• Iron & Steel
• Optical Emission

Introduction
The Thermo Scientific ARL QuantoDesk is a compact
desktop spectrometer for metals analysis based on CCD Each spectrum is stored, and a retrospective analysis
(Charge Coupled Device) technology. A single CCD for “extra” elements is possible without the need to spark
replaces the photomultiplier detectors used in traditional the sample another time.
arc/spark spectrometers and gives the instrument immense
flexibility. The argon flushed sample stand and spark Factory calibration (CARL)
source are comparable to those of larger instruments and Thermo Scientific optical emission spectrometers can be
enable the ARL QuantoDesk to be as capable as a factory calibrated for iron and steel utilizing CARL, a very
conventional laboratory spectrometer. sophisticated multi-variable regression tool that corrects
The ARL QuantoDesk offers the solution to perform a for matrix effects as well as spectral interferences. CARL
rapid check close to the furnace where its performance provides an immediate “turnkey” system which gives the
meets the analytical needs of Quality Control departments. user the highest accuracy possible. The calibrations are
Its simple architecture allows for high accuracy, stability available for the different qualities given below. For each
and reliability, making it the solution for numerous quality, we use certified material as standard samples and
industrial applications in rugged environments. setting-up samples are delivered with the instrument to
maintain the accuracy of the calibration.
Charge Coupled Device (CCD) The calibrations available are for:
In its broadest definition, a charge-coupled device (CCD) • Low alloy steel
image sensor is an analogue integrated circuit that • Cast iron
converts an optical image into an electronic output. As • Cr steel
with many semiconductor components, the evolution of
• Cr-Ni steel
CCD’s from a laboratory concept has been extremely
rapid, boosted by their potential for a wide range of • Mn steel
applications in both consumer products and sophisticated • High speed steel
professional equipment. • Ni resist
Today, our company uses this advanced technology in • High alloy cast iron
its spectrometers. The ARL QuantoDesk’s CCD has the
equivalent of 8,044 detectors of 7 by 7 µm. Because of
this, all the wavelengths (from 170 to 410 nm) necessary
for multi-element, multi-base analysis are available
without the need for additional hardware. Their
integrated nature provides their best-known features,
namely unlimited lifetime and miniaturization.
Sample preparation Table 2: Accuracy example on a low cast iron sample
The sample is usually prepared with a grinding machine.
BS 1C
EL CERTIFIED CONF. INTERVAL MEASURED DIFFERENCE
Sample analysis time VALUE % % VALUE % %

C 2.359 0.019 2.376 0.017


MATERIAL TIME PER RUN
S 0.060 0.0039 0.054 0.0061
Low Alloy Steel 24 s
Mn 1.79 0.011 1.77 0.023
Chrome-Nickel Steel 24 s
Ni 1.99 0.011 2.01 0.016
Cast Iron 31 s
Cr 0.52 0.02 0.51 0.016
Mo 0.433 0.009 0.426 0.007
Accuracy V 0.116 0.005 0.121 0.005
Precision is only a small part of providing accurate analyses. Ti 0.083 0.0032 0.089 0.0057
The most important factor is the accuracy and quality of Co 0.244 0.0086 0.266 0.0225
the calibration standards. Spectral interferences are Al 0.033 0.0023 0.033 0.0001
significantly reduced by the high flexibility of the
ARL QuantoDesk. The example below illustrates the Table 3 : Accuracy example on a stainless steel sample
excellent linearity and the reduced absorption effects
achieved on element carbon in low alloy steel. IARM 18B
EL CERTIFIED CONF. INTERVAL MEASURED DIFFERENCE
VALUE % % VALUE % %
C 0.085 0.0028 0.082 0.0031
Si 3.70 0.04 3.61 0.086
P 0.032 0.0028 0.037 0.0051
Mn 8.58 0.085 8.59 0.014
Ni 8.50 0.06 8.30 0.20
Cr 16.4 0.1 16.5 0.14
Cu 0.362 0.0073 0.360 0.0024
Co 0.124 0.007 0.1327 0.0087

Memory effects
Memory effects are determined by running an alloyed
sample (Ni: 20,7 %, Cr: 25,75 %) and then measuring a
pure sample, until the correct concentrations are
measured. No memory effect is observed: the correct
concentration is obtained at the second run already,
confirming the suitability of the ARL QuantoDesk for
multiple matrices and multiple qualities purposes.
Tables 1 to 3 show that the analysis of CRM’s (Certified
Reference Material) provides results within the uncertainty Performance guarantee
interval (3s) of the certified values in most cases. Our company guarantees the precision shown in Table 4
using homogeneous samples and recommended sample
Table 1: Accuracy example on a low alloy steel sample preparation. The list will be updated as improvements are
CKD187A announced. Please contact your nearest Thermo Fisher
EL CERTIFIED CONF. INTERVAL MEASURED DIFFERENCE Scientific office or consult our website at
VALUE % % VALUE % %
www.thermo.com/oes for the most recent values.
Al 0.019 0.002 0.0239 0.005
C 0.119 0.004 0.126 0.007
The precision is calculated from the formula:
Cr 3.51 0.04 3.526 0.02 where:
Cu 0.036 0.003 0.0281 0.008 i=n
X i the individual readings
Mn 0.525 0.007 0.5334 0.008 (X i - X i )2
X i the arithmetic means of
Mo 0.565 0.008 0.533 0.032 i=1 the individual readings
SD (1s) = +
n-1
Ni 0.085 0.004 0.0904 0.005 n the number of determinations
P 0.035 0.0015 0.0366 0.0016
S 0.0018 0.001 <3*DL
Si 0.567 0.023 0.5447 0.022
V 0.558 0.008 0.551 0.008
W 0.67 0.02 0.6645 0.006
Table 4: ARL QuantoDesk typical detection limits (3 sigma) and precision values (1 sigma) for iron base
ELEMENT Al C Ca Ce Co Cr Cu La Mg Mn Mo

Guaranteed DL [ppm] 30 20 1 50 15 25 15 10 6 20 50
Level % SD SD SD SD SD SD SD SD SD SD SD
0.001 0.00005
0.002 0.0001 0.0002
0.005 0.002 0.00025 0.0007 0.001 0.0005 0.001 0.0003 0.001
0.01 0.001 0.002 0.0005 0.0035 0.001 0.001 0.001 0.0015 0.0005 0.001 0.0045
0.02 0.0015 0.002 0.0035 0.001 0.001 0.001 0.0015 0.001 0.001 0.0045
0.05 0.0015 0.002 0.004 0.0035 0.001 0.0015 0.0015 0.002 0.005
0.1 0.002 0.002 0.005 0.0015 0.002 0.0035 0.002 0.0055
0.2 0.0025 0.003 0.008 0.002 0.0025 0.0025 0.007
0.5 0.003 0.007 0.003 0.005 0.0035 0.01
1 0.015 0.005 0.009 0.0055 0.02
2 0.025 0.01 0.02 0.01 0.03
3 0.03 0.02 0.025 0.02 0.04
4 0.04 0.035 0.035 0.025 0.05
5 0.05 0.035 0.065
10 0.06 0.09
20 0.07

ELEMENT Nb Ni P S Si Sn Ti V W Zn Zr

Guaranteed DL [ppm] 30 25 15 40 30 30 5 10 250 3 20


Level % SD SD SD SD SD SD SD SD SD SD SD
0.001 0.0004
0.002 0.0003 0.00055 0.0004
0.005 0.0015 0.001 0.0003 0.00055 0.0005 0.001
0.01 0.0015 0.0015 0.0025 0.002 0.0025 0.002 0.0003 0.0006 0.0005 0.001
0.02 0.002 0.0015 0.003 0.002 0.0025 0.0025 0.0004 0.0007 0.0005 0.0015
0.05 0.004 0.002 0.003 0.0025 0.003 0.003 0.0005 0.001 0.015 0.0025
0.1 0.0045 0.002 0.005 0.003 0.0035 0.0035 0.001 0.0015 0.015 0.004
0.2 0.005 0.0025 0.005 0.005 0.007 0.002 0.0025 0.02
0.5 0.01 0.004 0.008 0.007 0.003 0.0035 0.025
1 0.015 0.006 0.015 0.01 0.0035 0.0055
2 0.025 0.01 0.02 0.008
3 0.04 0.025 0.02
4 0.05 0.03 0.05
5 0.055
10 0.085
20 0.2
30 0.25

This data applies when homogeneous samples are prepared by recommended sample preparation methods.
The precision given is typical performance. Guaranteed values will be 1.5 times higher.
SD = Standard Deviation. DL = Detection Limit.
Stability Conclusion In addition to these
Stability of the instrument is of the utmost importance The ARL QuantoDesk is a compact, desktop spectrometer offices, Thermo Fisher
when performing routine analysis. Mid-term stability for metals analysis, based on CCD technology, which Scientific maintains
measured in a laboratory environment over 24 hours provides flexibility for the analysis. It can be installed and a network of represen-
shows that the standard deviation achieved is at the level be operational on the same day it is delivered. Simple to tative organizations
of the precision value. handle, the ARL QuantoDesk provides quick, precise and throughout the world.
The example below shows the long-term stability of reliable analysis.
nickel and sulfur in a stainless steel sample recorded over Its accuracy, associated with its excellent stability and
a period of 8 days without any intermediate drift correction. reliability allows to have results you can rely on 24 hours
No standardization was required, as the measured a day, 365 days a year without spending a lot of time in
concentrations stayed almost within the control limits. long lasting and expensive drift corrections.
Africa
In addition, the ARL QuantoDesk brings the +43 1 333 5034 127
following advantages: Australia
• Simple and easy to use +61 2 8844 9500
Austria
• Compact, space saving, desktop unit +43 1 333 50340
• Multibase capability without hardware change Belgium
+32 2 482 30 30
• Menu driven Windows® XP based software Canada
• Touch screen operation as standard +1 800 530 8447
China
• Data storage of more than 10,000 results +86 10 5850 3588
• Extensive grade library Denmark
+45 70 23 62 60
• On line help functions Europe-Other
• Multi-lingual capability +43 1 333 5034 127
France
• High level of performance at an affordable price. +33 1 60 92 48 00
All these features allow you to optimize your Germany
+49 6103 408 1014
productivity and to achieve the shortest payback time:
India
• Your investment costs are reduced with the capability of +91 22 6742 9434
the instrument to cover your potential future needs Italy
without the need for additional hardware +39 02 950 591
Japan
• Your production costs are reduced by the increased +81 45 453 9100
instrument availability thanks to its high stability and Latin America
drift corrections being less frequently required +1 608 276 5659
• Your operating and maintenance costs are reduced Middle East
+43 1 333 5034 127
through simple maintenance Netherlands
With its over 70 years of experience and history of +31 76 587 98 88
innovative technology, our company has become the South Africa
+27 11 570 1840
world leader in OES metals analysis. We work with our Spain
customers to improve the efficiency of their analytical +34 914 845 965
tasks and thereby increase productivity. Sweden / Norway /
Finland
+46 8 556 468 00
Switzerland
+41 21 694 71 11
UK
+44 1442 233555
USA
+1 800 532 4752

www.thermo.com

©2007 Thermo Fisher Scientific Inc. All rights reserved. MS Windows® is a registered trademark of Microsoft corp. All other trademarks are the property of Thermo Fisher Scientific Inc. and its Thermo Electron SA, Ecublens,
Switzerland is ISO certified.
subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

AN41260_E 12/07C