Professional Documents
Culture Documents
638 www.nature.com/jp
Volume Guarantee Ventilation Abubakar and Keszler
Our objective was to compare the effect of VG combined with 7 cmH2O, both remained unchanged throughout the study.
AC and VG combined with SIMV on the stability of VT, oxygen Fraction of inspired oxygen (FiO2) was adjusted to maintain
saturation, respiratory rate (RR), heart rate (HR) and airway oxygen saturation target range of 88 to 92% according to our unit
pressures in preterm infants requiring mechanical ventilation. We protocol. Blood gas measurements were obtained by the clinical
hypothesized that exhaled tidal volume (VTe), oxygen saturation team only as needed. However, no ventilation settings or FiO2
and minute ventilation (MV) will be more stable when VG is needed to be changed in any patient as a result of an undesired
combined with AC, compared to VG combined with SIMV. blood gas value.
RESULTS more variable while infants were on SIMV þ VG. Even though FiO2
A total of 12 infants requiring mechanical ventilation who met the remained unchanged across study modes, oxygen saturation was
study criteria were enrolled. All infants successfully completed the significantly lower and much more variable during SIMV þ VG
study. All infants had received a full course of antenatal than AC þ VG (Table 3).
betamethasone prior to delivery. The baseline characteristics of the
infants at time of study entry are shown in Table 1. Ventilator and DISCUSSION
cardio-respiratory parameters discussed below are shown in Tables
2 and 3. Although the physiologic advantages of synchronized mechanical
PIP and mean airway pressures were significantly higher with ventilation have been well documented, there is no clear consensus
machine-supported breaths of SIMV þ VG compared to AC þ VG on which of the two most commonly used synchronized ventilation
(Table 2). MV (all breaths included) and VTe were similar in the modes is superior. Prior studies have shown that AC facilitates
two groups, but both were significantly more variable during weaning from mechanical ventilation and delivers more consistent
SIMV þ VG compared to AC þ VG. The VTe of machine-supported tidal volume with less fluctuation in blood pressure when
breaths during SIMV þ VG was similar to that of AC þ VG by compared to SIMV.13–16 Adding volume guarantee to either mode
design since the same tidal volume was targeted in both modes. of synchronized conventional ventilation has been shown to
However, the VTe of unsupported breaths during SIMV þ VG was provide more uniform tidal volume delivery.22 The purpose of this
significantly lower and more variable than that of machine study was to compare the effect of adding volume guarantee to
supported breaths during SIMV þ VG or the VTe during AC þ VG
(Table 2).
During SIMV þ VG, the infants had a faster and more variable Table 3 Heart Rate, Respiratory Rate and Oxygenation
RR than while on AC þ VG. Similarly, the HR was higher and
AC+VG SIMV+VG p-value
Exhaled tidal volume (ml) Mean±SD 5.2±0.5 4.9±2.0 5.1±1.8 3.2±1.8 p<0.001w
Variance 0.3 4.4 3.4 3.5 p<0.0001z
assist control ventilation on the stability of tidal volume delivery, pressure), it follows that the ventilator contribution would need to
MV, airway pressures, oxygen saturation and RR compared to SIMV be higher in order to compensate. This appears to be the more
with VG. plausible explanation for the higher working pressure needed
We found that adding volume guarantee to AC ventilation during SIMV to generate identical VTe to AC.
results in more uniform tidal volume delivery with less Our findings are in keeping with earlier studies that have
breath-to-breath fluctuation compared to SIMV þ VG. We demonstrated lower respiratory rates smaller and more consistent
suggest that this is because with AC, every spontaneous breath VT and less fluctuation in blood pressure with AC, compared
is subject to volume regulation, whereas in SIMV only the to SIMV.13–16 An important finding in our study is the lower
machine breaths are supported and any un-supported breaths oxygen saturation and the degree with which it fluctuates
taken by the infant are not subject to volume guarantee. The during SIMV þ VG. This increased fluctuation in blood pressure
algorithm of the VG software limits the magnitude of pressure and oxygen saturation can be deleterious especially in smaller
increase from one breath to the next to approximately 3 cmH2O infants at risk of intraventricular hemorrhage and retinopathy
to avoid potentially dangerous overshoot. Consequently, when of prematurity. We have shown in this study that adding VG
the VTe falls much below the target, several breaths are needed to AC rather than SIMV may add to the advantages of AC
to achieve the target VTe. Since fewer breaths are subject to over SIMV.
volume regulation in SIMV, more time is needed to reach During our data analysis, all breaths were examined manually
target value and larger fluctuations are likely as the infant’s and those that manifested severe distortion, large negative leak
spontaneous respiratory effort fluctuates. (VTe much larger than inspiratory VT) or other obvious artifact,
Small preterm infants with respiratory failure who require such as would result from handling, movement or hiccups were
mechanical ventilation typically cannot generate adequate VT on deleted following pre-set criteria to obtain a truer reflection of
their own. This in part reflects their poor lung compliance, the the ventilator performance. On clinical observation, infants are
high airway resistance of small endotracheal tubes, inadequate occasionally noted to have an interrupted expiratory pattern on the
muscle strength and unfavorable mechanics of the immature chest ventilator. Some of these interrupted expiratory patterns may
wall. Consequently, when ventilated by SIMV, the unsupported appear as a distorted breath during manual examination of the
spontaneous breaths are often quite small, resulting in limited particular breath, resulting in that breath being excluded from
effective alveolar ventilation. In SIMV mode, the limited number of analysis. However, the number of excluded breaths was small and
ventilator breaths must make up for the inefficient spontaneous analysis of both the raw data and data with distorted breaths
breaths by using larger VT. In AC, each breath is supported; removed did not alter the results.
therefore, the ventilator cycles at a faster rate and thus a smaller It should be recognized that the differences between SIMV
VT is required to maintain equal MV, when compared to SIMV.16 and AC observed in this study are likely to be more pronounced
When SIMV is used in pressure-limited mode, the larger ventilator with slower SIMV rates. During the acute phase of RDS when
VT is generated with somewhat higher inspiratory pressure. more rapid SIMV rates are commonly used, there may be little
When the VT is fixed at an equal value by the use of VG in difference between the effectiveness of these modes with or
combination with SIMV, the only mode of compensation available without VG. However, because many infants are exposed to
is for the infant to breathe more rapidly, as was seen in our low SIMV rates for substantially longer periods of time during
study population. weaning, our findings are applicable to the majority of infants
We were surprised to see that in SIMV, compared to AC, higher on SIMV.
PIP was needed to generate the same VTe. Two possible This is a short-term crossover study in a small number of
explanations present themselves. One is that the slower rate of infants to evaluate the immediate effects of adding VG to SIMV or
‘‘normal’’ sized unsupported breaths allowed enough time for AC and was not designed to look at long-term differences between
atelectasis to develop, thereby leading to decreased lung these ventilation modes. However, our findings and those of
compliance. This appears unlikely, as the ventilator rate was set at previous authors suggest that adding VG to assist control
30 in SIMV mode, a time interval not likely to allow for atelectasis. ventilation would provide more gentle ventilation than SIMV þ VG
The observation that the infants were more tachypneic, with a in small infants, especially during weaning. Available evidence does
faster HR and had consistently lower oxygen saturations suggests not support the reluctance of many clinicians to accept AC as a
that the infants had substantially higher work of breathing during weaning mode. In the absence of an appropriately designed
SIMV. In that situation, we speculate that the infants were tiring prospective randomized trial, however, no conclusions can be
and therefore making a smaller contribution to each mechanical drawn about the long-term benefits of AC þ VG over SIMV þ VG
breath. Given that the VT of a synchronized breath is determined by particularly looking at air-leak syndromes, duration of mechanical
the combined positive pressure of the ventilator and the negative ventilation, chronic lung disease and neuro-developmental
inspiratory pressure generated by the infant (the transpulmonary abnormalities.