Professional Documents
Culture Documents
Ser1500 Rev.e
Ser1500 Rev.e
Service Manual
SER1500 Rev. E
02/08
SER1500 E0208 i
Section 5 . . . . . OPERATIONAL VERIFICATION PROCEDURE (OVP)
• Introduction
• Set Up
• Standard Test Settings
• Pressure Gauge Verification
• Alarm/LED Check
• Mixer Alarm
• FIO2
• ƒ (respiratory frequency)
• tI (inspiratory time)
• SPONT V̇1
• SPONT V̇2
• V̇ (flow)
• PEEP/CPAP
• Low Baseline Alarm, RESET and
Alarm Loudness Control
• ▲Paw (High Pressure) Alarm
• ▼Paw (Low Pressure) Alarm
• Ptrig (trigger level) and MODE Select
• Apnea Alarm
• Manual Inflation Button
• NEB (Nebulizer)
• Internal Battery Back-Up
• Control Monitor
• Paw Monitor, (peak, mean, base)
• Proximal Line Purge
• Advanced Trigger Control (ATC)
• Auto Set (AS) Alarms
• Time-Limited Demand Flow
• E100M Test Record
Section 6 . . . . . PREVENTIVE MAINTENANCE PROCEDURE
• Introduction
• Tools Required
• Parts Required
• Maintenance Procedure
Section 7 . . . . . OVERHAUL PROCEDURE
• Introduction
• Tools Required
• Parts Required
• Overhaul Procedure
ii SER1500 E0208
Section 8 . . . . . ELECTRONIC & PNEUMATIC CALIBRATION
• Introduction
• Electronic Verification
• Pressure Gauge Calibration
• Front Panel Calibration
• V̇ (flow)
• tI (inspiratory time)
• ƒ (respiratory frequency)
• Time-Limited Demand Flow
• PEEP/CPAP
• Battery Voltage Adjustment
• Mixer Regulator Calibration
Section 9 . . . . . TROUBLESHOOTING AND ERROR CODES
• Introduction
• Troubleshooting
• Error Codes
Section 10 . . . . ELECTRONIC AND PNEUMATIC COMPONENT
REMOVAL AND REPLACEMENT PROCEDURES
• Introduction
• Tools Required
• Open Front Panel
• Remove Rear Panel
• Mixer (MIX1725A)
• Flow Control Assembly (FCL1500A)
• Flow Block Assembly (BLK1500A)
• Nebulizer and Prox/Machine Solenoids
(SOL1503P and SOL1501P)
• Internal Battery Back-Up (BAT1500A)
• Main Board (PCB1550A)
• Display Board (PCB1511A)
• Electronic Pressure Gauge (MAN1540M)
• Power Supply Board (PCB1530A)
• Lower Display Board (PCB1520A)
• Adapter Board (PCB1570A)
iv SER1500 E0208
1. OPERATOR’S RESPONSIBILITY
SER1500 E0208
SECTION 1
LIMITATION OF LIABILITY
Newport Medical Instruments, Inc. liability, whether arising out of, or
related to manufacture and sale of the goods, their installation,
demonstration, sales representative, use, performance, or otherwise,
including any liability based upon Newport Medical Instruments, Inc.
product warranty, is subject to and limited to the exclusive terms and
conditions as set forth, whether based upon breach of warranty or any
other cause of action whatsoever, regardless of any fault attributable to
Newport Medical Instruments, Inc. and regardless of the form of action
(including, without limitation, breach of warranty, negligence, strict
liability, or otherwise).
Newport Medical Instruments, Inc. shall not be liable for, nor shall the
buyer be entitled to recover, any special incidental or consequential
damages or for any liability incurred by buyer to any third party in any
way arising out of or relating to the goods.
WARRANTY
DEFINITIONS
WARNING A WARNING statement refers to a condition with the
possibility of personal injury if disregarded.
Typing Conventions Within the text of this manual, controls, alarms, buttons, etc. are
designated by the labeling name as they appear on the ventilator, e.g.:
P limit (pressure limit), P trig (pressure trigger), SPONT (spontaneous
mode).
Cautions Use only dry, clean, compressed air and medical grade oxygen. Water in
the air or oxygen supply can cause equipment malfunction and damage.
Mains voltage must correspond to the voltage range selected on the
power module of the ventilator. Always replace a blown fuse with one of
proper rating for corresponding voltage range.
Do not place containers of liquids in the immediate vicinity or on top of
the E100M Ventilator. Liquids that get into the ventilator can cause
equipment malfunction and damage.
Periodically inspect the Air Inlet Water Trap. Drain water from the bowl
when necessary by depressing the pin at the bottom of the bowl.
CONTACT INFORMATION
Address: Newport Medical Instruments, Inc.
1620 Sunflower Avenue
Costa Mesa, CA 92626
Corporate Office Hours: Monday through Friday 8:00 am to 5:00 pm, Pacific Time, USA
Technical Service
Department Hours: Monday through Friday 7:00 am to 4:00 pm, Pacific Time, USA
EU Representative: EU Representative
Braun and Company, Ltd.
11B/11C Harrier Road
Humber Bridge Industrial Estate
Barton on Humber
North Lincolnshire
DN18 5 RP, England
tel: 01652 632273
fax: 01652 633399
Street ______________________________________________________________
Address ____________________________________________________________
__________________________________________________________________
WARRANTY
Newport Medical Instruments, Inc. (NMI) warrants this product to meet the published specifications and
to be free from defects in material and workmanship under normal use for a period of three (3) years or
12,000 hours from date of purchase.THE FOREGOING IS IN LIEU OF ANY OTHER WARRANTY,
EXPRESSED, IMPLIED OR STATUTORY, INCLUDING WITHOUT LIMITATION ANY WARRANTY OR
MERCHANTABILITY, WARRANTY OF FITNESS FOR ANY PARTICULAR PURPOSE, OR WARRANTY
OF ANY KIND AS TO DESIGN. The sole liability of NMI under this warranty is limited to replacing,
repairing or issuing credit at the discretion of NMI for the products, equipment or parts which fail to meet
the published specifications or which become defective during warranty period and which are, upon
examination by NMI, found not to meet the published specifications or to be defective in materials or
workmanship. NMI will not be liable under this warranty unless the following provisions are strictly
complied with: (a) NMI is promptly notified, in writing, upon discovery of the failure of the said product or
equipment to meet the published specifications or of the defects in materials or workmanship, (b) The
defective product, equipment or part thereof is returned to NMI, transportation charges paid by NMI, (c)
The defective part is received by NMI for examination no later than one (1) month following the expiration
of the warranty period and provided, (d) that examination by NMI of said product, equipment or part shall
disclose to NMI's satisfaction that such defect has not been caused by improper usage, accident,
neglect, alteration, abuse, improper installation or unauthorized repair. Products, equipment or parts
replaced under this warranty are warranted only through the terms of the original warranty. NMI neither
assumes nor authorized any other person or entity to assume for it any other warranty, obligation or
liability in connection with its products or equipment whatsoever, and as to the fitness or usefulness of
the equipment manufactured by it for any medical treatment, physical condition or other purpose
whatsoever. In no event shall NMI be liable for personal injury, property damage or any special or
consequential damage to the buyer, user or any other person whomsoever, including, but not limited to,
loss of profits, loss of use of the product or equipment, or for damages of any kind whatsoever based on
a claim for breach of warranty other than a refund of the purchase price of any defective product or
equipment. Any authorization for the repair or alteration by buyer must be in writing from NMI to prevent
the voiding of this warranty.
2. SPECIFICATIONS
Ventilator Overview . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Intended Use. . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Symbols/Labeling Table . . . . . . . . . . . . . . . . . . 2 - 2
Specifications. . . . . . . . . . . . . . . . . . . . 2 - 5 to 2 - 9
SER1500 E0208
SECTION 2
VENTILATOR OVERVIEW
The E100M ventilator provides the following features:
INTENDED USE
Indications For Use: The Newport E100M is intended to provide positive
pressure ventilation for neonatal, infant, pediatric and adult patients, for
use in the hospital environment. The device is intended for use by or at
the direction of a licensed practitioner.
Reset
Apnea (s)
Alarm Silence/Reset Button Apnea Time Delay
▼Battery
▼ ▲ Paw Alarm
Spont
Push to set only ▲ Paw
Adjust High and Low Paw Alarm Spont (Patient Effort) Indicator
—
▲ Paw ▲ Paw
High Airway Set High Airway
Pressure Alarm Indicator Pressure Alarm
▼ Paw ▼ Paw
Low Airway Set Low Airway
Pressure Alarm Indicator Pressure Alarm
AS ATC !
Automatic Control Buttons Refer to Operating Manual
(see Paw Alarm and Ptrig)
NEB
EXH VLV
PROX PATIENT
• •
SPONT V2 Tubing Connection SPONT V2 Flowmeter Connection
TYPE 'B'
Type B
REMOTE
ALARM
OUTPUT
STDBY ON
Main Power “Standby” Main Power “On”
E100M SPECIFICATIONS
Control/Parameter/Indicator Range/Selection
1. MODE A/CMV
Pressure
or SIMV
Volume SPONT + APNEA A/CMV
SPONT
6. PEEP/CPAP 0 to ≥ 25 cmH2O/mbar
15. Spont (Patient Effort) Indicator Blinks once each time airway pressure drops below
the P trig setting.
E100M SPECIFICATIONS
Control/Parameter/Indicator Range/Selection
16. ▼▲ Paw Alarm Set high and low airway pressure alarms
_
▲Paw 5 to 120 cmH2O / 5 to 118 mbar
▼Paw 3 to 118 cmH2O / 3 to 116 mbar
21. Control Monitor Selector Selects parameter (I:E, ƒ tot, VT) which is displayed
in the Control Monitor display.
23. Apnea Selector Selects time delay interval for the Apnea alarm,
Ø (disable), 15, or 30 seconds.
24. Reset (Alarm Silence) Silences an alarm violation for 60 seconds, resets a
“latched” alarm indicator and cancels Power Down
alarm.
25. ALARMS Indicators Indicator for the alarm that is actively violated blinks
red. When the alarm is no longer violated, the
indicator latches (stays lit). Cancel a latched indicator
by pressing the Reset button.
26. tI > tE Indicator A yellow indicator blinks any time the control settings
result in an inverse I:E ratio.
E100M SPECIFICATIONS
Control/Parameter/Indicator Range/Selection
•
28. SPONT V 1 Switch Provides Ø or 8 L/min mixed gas for spontaneous
breathing.
30. External Alarm Silence Input Allows use of the NEWPORT External Alarm Silence
cable which functions the same as the Reset
button. Does not cancel the Power Down alarm.
31. RS-232C Interface Operates at 38,400 baud. Allows for interfacing with
central monitoring and computer systems via 25 pin
D type connector.
32. Remote Alarm Output Provides contact closure when alarming. Operates
with most nurse call systems using a 1/4 in (0.64 cm)
phone type connector.
34. Internal Battery Test Button When pressed during battery power operation, %
battery charge (0 to 99%) is displayed in Control
Monitor window.
35. Power Switch This rocker type switch, (ON) or (STDBY) switches
ventilator on and off. When the ventilator is
connected to A.C. power, the internal battery
charges regardless of switch position.
36. Loss of Gas Alarm Continuous audible alarm. Activated if air or oxygen
supply pressures drop below 28 ± 3 psig.
(No visual alarm indicator.)
37. Power Down Alarm Continuous audible alarm activated when the ventilator
power switch is changed from (ON) to (STDBY).
The alarm is silenced by pressing the Reset button.
(No visual alarm indicator.)
E100M SPECIFICATIONS
39. High Baseline Alarm ▼Paw and ▲Paw alarm indicators blink red and the
(automatic) audible alarm sounds continually if Paw remains
above ▼Paw alarm setting at end exhalation.
40. Low Baseline Alarm ▼Paw alarm indicator blinks red and the audible
(automatic) alarm beeps intermittently if Paw remains below the
Low Baseline alarm setting (see chart pg. 4-14) for >
3 seconds.
Hardware Requirements
Miscellaneous Description
46. Regulatory and Agency Designed to meet the applicable requirements of the
Requirements FDA, CSA, JIS, IEC and TUV.
E100M SPECIFICATIONS
Miscellaneous Description
49. Height 10.5 in (26.7 cm)
SER1500 E0208
SECTION 3
Reset ALARMS
E100M V E N T I L A T O R
▲ Paw
▼ Paw
Apnea (s) ▼ Battery 70 80 90
15 60 100
30 Apnea
0 50 110
40 120
MODE 30 Paw
SPONT 20 (cmH20)
10 0 -10 peak
SPONT +APNEA A/CMV mean
base
SIMV
A/CMV
▼ ▲ Paw Alarm Ptrig Spont
Push to set only ▲ Paw
I:E ƒ tot
AS ATC
V T (L)
•
F IO 2 V (L/min) tI (s) ƒ (b/min)
30 tI > tE
.60 20 40 1.5 30
.50 .70 40
1.0 2.0 20
10 50
.40 .80 5 60
MIN 60 0.5 2.5 10 80
.30 .90 100
70 0.1 1
.21 1.00 3.0 120
MAX
80
100 90 Time-Limited
Manual Demand Flow (s) PEEP/CPAP Plimit
Inflation 0.5 Pbase
0.3 0.7
0.1
0 1.0
NEB PROX
Figure 3-1
E100M FRONT PANEL
MODE Control
A/CMV
(Assist/Control Mandatory Ventilation)
MODE The minimum number of time triggered mandatory breaths is determined
SPONT
SPONT +APNEA A/CMV
by the ƒ control. The total number of mandatory breaths is a result of
SIMV the ƒ setting and patient triggers. Set the P trig as close to baseline
A/CMV pressure as possible so that all patient efforts are recognized as patient
triggers. The indicator for A/CMV is lit green while operating in this
mode.
NOTE: If the P trig setting is not adjusted to a level that allows the
patient’s inspiratory effort to be detected, the A/CMV mode resembles
Control mode.
SIMV
(Synchronized Intermittent Mandatory Ventilation)
MODE The number of time triggered mandatory breaths is determined by the ƒ
SPONT
SPONT +APNEA A/CMV
control. The patient can take spontaneous breaths between the delivery
SIMV of the mandatory breaths. Set the P trig as close to baseline as possible
A/CMV
so that all patient efforts are recognized as patient triggers. This allows
the E100M to synchronize the mandatory breaths and spontaneous
Time-Limited Demand Flow breaths with patient effort. Time-Limited
Demand Flow may be used to support spontaneous efforts with a higher
initial flow which then drops to the lower Spont Flow. The indicator for
SIMV is lit green while operating in this mode.
MODE All breaths are spontaneous. Set the P trig as close to baseline as
SPONT
SPONT +APNEA A/CMV
possible so that all patient efforts are recognized as patient triggers.
SIMV Time-Limited Demand Flow may be used to support spontaneous efforts
A/CMV
with a higher initial flow which then drops to the lower Spont Flow. If the
patient stops breathing and it results in an Apnea alarm violation, A/CMV
breaths are delivered at the face panel control settings until the patient
effort results in four consecutive patient triggered breaths. The indicator
for SPONT + APNEA A/CMV is lit green while operating in this mode.
SPONT
(Spontaneous Ventilation)
MODE All breaths are spontaneous. Set the P trig as close to baseline as
SPONT
SPONT +APNEA A/CMV
possible so that all patient efforts are recognized as patient triggers.
SIMV Time-Limited Demand Flow may be used to support spontaneous efforts
A/CMV
with a higher initial flow which then drops to the lower Spont Flow. The
indicator for SPONT is lit green while operating in this mode.
Ventilation Controls
FIO2
FI O 2 Determines the air/oxygen mixture for all gas delivery to the patient.
.60
.50 .70 Range: .21 to 1.00
.40 .80
.30
NOTE: When the FIO2 is readjusted, the oxygen concentration will
.90
.21 1.00
stabilize in the Reservoir Bag in approximately 30 seconds (depending
on SPONT V̇1 and V̇ 2 settings).
V̇
(mandatory flow)
•
V (L/min) Sets the flow delivered to the patient during mandatory breaths (volume
20
30
40 or pressure), APNEA A/CMV breaths, Time-Limited Demand Flow, and
10 50 Manual Inflation.
5
MIN 60
Volume Ventilation: Adjust V̇ and tI to set VT (tidal volume).
70
MAX
80 Pressure Ventilation: Adjust V̇ to achieve desired inspiratory pressure
100 90
rise/slope.
Normal flow ranges for neonatal, pediatric, and adult patients are
indicated by the three different green lines around the dial.
tI
(inspiratory time-mandatory)
tI (s) Sets the inspiratory time for mandatory breaths (volume or pressure).
tI > tE
1.5
1.0 2.0 Volume Ventilation: Adjust V̇ and tI to set VT (tidal volume).
0.5 2.5
Pressure Ventilation: Adjust tI and ƒ to set appropriate I:E ratio.
0.1 3.0
Range: 0.1 to 3.0 sec
ƒ
(respiratory frequency)
ƒ (b/min) Sets the number of time triggered breaths delivered to the patient in the
30 A/CMV and SIMV modes, and in the event of an Apnea alarm violation in
20 40
60 the SPONT + APNEA A/CMV mode.
10 80
100 Range: 1 to 120 b/min
1 120
P limit
(pressure limit)
Plimit P limit determines the maximum pressure for all breaths. If airway
pressure reaches the P limit setting, flow vents to prevent airway
pressure from going higher. Clockwise rotation increases the maximum
P limit level.
Pressure Ventilation: Use the P limit control to set the peak inspiratory
pressure for all mandatory breaths. The inspiratory pressure rise/slope is
determined by the V̇ setting.
Volume Ventilation:
_ Use the P limit control to set a safe pressure pop-off
above the ▲Paw alarm setting.
WARNING
_ When volume ventilating, P limit must be set higher than
the ▲Paw alarm, otherwise volume may be vented.
PEEP/CPAP
PEEP/CPAP Establishes the pressure in the patient circuit from end exhalation to the
Pbase
start of the next inspiration. Also referred to as base or baseline
pressure.
▲Paw Alarm
(user adjustable)
_
ALARMS If airway pressure reaches the ▲Paw (high pressure) alarm setting a
▲ Paw
▲Paw alarm violation occurs, and any breath in progress cycles to
▼ Paw exhalation. The alarm resets when airway pressure drops below the
_
▼ Battery ▲Paw setting.
Apnea
Audible Alarm: intermittent beep
Visual Alarm: ▲Paw indicator blinks red
▼Paw Alarm
(user adjustable)
ALARMS If airway pressure does not rise above the ▼Paw (low pressure) alarm
▲ Paw
setting during inspiration, then drop below the ▼Paw alarm setting
▼ Paw during exhalation, a ▼Paw alarm violation will occur on mandatory
▼ Battery breaths. The alarm is reset when one mandatory breath is delivered
Apnea without a ▼Paw alarm violation.
NOTE: The Low Baseline alarm default settings are defined in Section 4,
Theory of Operation, Low Baseline Alarm Table, Table 4-4.
Range: 0 to 4 sec
NEB
(nebulizer
NEB gas source)
Activates a 28 psig (1.9 Bar) mixed gas source for powering a low
NEB
volume nebulizer. NEB function is available during all mandatory, Time-
Limited Demand Flow, and Manual Inflation breaths. Push button to
toggle on/off. When in use the indicator is lit green.
NEB On/Off Button
NOTE: The NEB gas source provides an unrestricted flow of 10 L/min.
NEB
When using the Newport Nebulizer Kit (p/n PNK502A) the actual flow to
the patient breathing circuit is 8 L/min. The additional 8 L/min flow is
NEB Connection reflected in the VT calculation displayed in the Control Monitor. You can
compensate for the additional flow by reducing the V̇ control setting by 8
L/min when the NEB indicator is lit.
PROX
(proximal
PROX
mainflow pressure monitoring)
Selects the location for monitoring airway pressure and patient efforts
PROX
(triggering). Monitoring location is at the main flow outlet of the ventilator
or at the proximal monitoring line of the breathing circuit.
PROX On/Off Button PROX Button Indicator unlit: PATIENT main flow outlet monitoring
WARNING Although the E100M allows the user to set an inverse I:E
ratio of up to 4:1, ventilating a patient at a ratio of greater than 1:1.0
is unusual. Care should be taken to ensure that each patient is being
ventilated at an appropriate I:E ratio.
ƒ tot
(total frequency)
I:E ƒ tot
ƒ tot is the total number of time triggered and patient triggered breaths
detected in the last 60 seconds.
18
V T (L)
Range: 0 to 255 b/min
VT (L)
(calculated tidal volume)
I:E ƒ tot Set VT is calculated from the V̇ and tI control settings and is displayed in
.6 5 liters (L). Increasing either control increases tidal volume.
V T (L)
Range: 0 to 5 L
I:E ƒ tot When switching from A.C. power to internal battery power and every 30
minutes of battery power thereafter, the Control Monitor displays the %
75P charge remaining in the battery, and sounds an audible Battery In Use
V T (L)
alert for two seconds.
NOTE: When the ventilator is powered by internal battery power and the
Internal Battery Test button on the rear panel is pressed, the % of
battery charge remaining is displayed in the Control Monitor. The
ventilator must be running on battery power for a minimum of 5 minutes
prior to checking the % charge.
50
40
110
120
NOTE: Automatic selection of monitored values occurs with certain
30
20
10 0 -10
Paw
(cmH20)
peak
control changes or specific conditions. Certain conditions do not allow
mean
base
display of some values. When the display shows “- - ”, the measured
value is out of the monitor’s range.
peak
In A/CMV and SIMV mode, peak pressure is updated at the end of each
Paw
(cmH20)
mandatory breath or manual inflation.
peak
mean
base
In either SPONT mode, peak pressure is not displayed unless APNEA
A/CMV is activated or a Manual Inflation is delivered.
mean
Paw
In all modes, mean pressure is averaged over the last 30 seconds. It is
(cmH20)
base
Paw
In all modes, base (PEEP/CPAP) pressure displays the airway pressure
(cmH20)
peak
at end exhalation of the previous breath. It is updated every 10 seconds
mean
base or at the beginning of inspiration, whichever occurs first.
Ptrig Spont ATC indicator lit green: P trig level managed by the ATC function. P trig
follows baseline changes automatically and ATC re-evaluates the current
P trig setting every 30 minutes if self triggering does not occur.
ATC
▼ Paw
1. Silencing alarms: Press the Reset button to silence all front panel
Apnea (s)
15
▼ Battery alarms for 60 seconds. When the Reset button indicator is lit, all
30
0
Apnea front panel alarms are silenced. Press the button again to cancel the
silence period.
NOTE: The Loss of Gas alarm (pneumatic) and Fault System alarm are
non-silencable.
▼ Battery
ALARMS
▼Battery (low battery) indicator blinks red/no audible alarm: Indicates
▲ Paw
20% charge remaining on internal battery.
▼ Paw
When the ventilator is switched back to A.C. power, the Low Battery
indicator latches (stays lit). Press Reset to cancel latched alarm
indicator.
Pressure Gauge
The pressure gauge displays proximal airway pressure if PROX is
E100M VENTILATOR
activated, or PATIENT mainflow outlet pressure if PROX is not
60
70 80 90
100 activated.
50 110
40 120 _
30
20
10 0 -10
Non-blinking segments: ▲Paw and ▼Paw alarm settings are displayed
as single non-blinking segments. P base is displayed as the appropriate
range of non-blinking segments.
NOTE: If the P trig indicator is set at the same level as P base, the
segment which indicates both blinks.
tI > tE Indicator
(inverse ratio)
tI (s) Any time the ƒ and tI settings cause an inspiratory time that is longer
tI > tE
1.5 than expiratory time (I >E), the tI > tE indicator blinks yellow.
1.0 2.0
0.5 2.5
Range: 4:1 max
0.1 3.0 WARNING Although the E100M allows the user to set an inverse I:E
ratio of up to 4:1, ventilating a patient at a ratio of greater than 1:1.0
is unusual. Care should be taken to ensure that each patient is being
ventilated at an appropriate I:E ratio.
NOTE: The Power Down Alarm cannot be silenced with the External
Alarm Silence Cable.
Range: ≤ 28 psig
•
SPONT V1
8 L/min
0 L/min
!
•
SPONT V2
•
SPONT V2
Figure 3-2
E100M LEFT SIDE PANEL
SPONT V̇ 2
This control provides a user-adjustable amount of mixed gas for patient
spontaneous breathing. It can be used in place of Time-Limited Demand
Flow, or in addition to the SPONT V̇1 gas source.
TYPE 'B'
SERIAL NO.
RS-232C
WARNING:
Condensed water in the air
supply can cause malfunction
ALARM EXTERNAL of this device.
LOUDNESS ALARM
100/120 V ~ .14A SILENCE
220/240 V ~ .07A MEDICAL 02
! AIR OXYGEN
50/60 HZ
REMOTE
ALARM
OUTPUT
FUSE
100/120 V ~ .25A, F
220/240 V ~ .10A, F
ON
LR51420
NRTL/C
0344
STDBY
Figure 3-3
E100M REAR PANEL
I:E ƒ tot NOTE: The E100M must be unplugged from A.C. power and running on
internal battery for a minimum of 5 minutes before testing the % charge.
75P
V T (L)
NOTE: If the ventilator is not being powered by the internal battery and
the Internal Battery Test button is pressed, dashes (---) are displayed
instead of the % charge.
When an alarm violation occurs on the E100M it also triggers the remote
alarm system which informs the remote location that the ventilator is
alarming. The REMOTE ALARM OUTPUT does not indicate which alarm
on the ventilator is being violated.
Alarms that the External Alarm Silence Cable button will not cancel:
Use only the NEWPORT External Alarm Silence Cable with this connector.
It is available as an optional accessory.
RS-232C Interface
RS-232C
The RS-232C interface is a 25 pin D-type connector, operating at 38,400
baud. It allows you to interface with central monitoring and computer
systems.
Figure 3-4
E100M RIGHT SIDE PANEL
SER1500 E0208
SECTION 4
An emergency intake valve allows the patient to pull ambient air into the
breathing circuit in the event of a complete loss of supply gas pressure.
Opening pressure is approximately –2 cmH2O/mbar in this condition.
A/CMV
(Assist/Control Mandatory Ventilation)
In A/CMV, mandatory breaths are time triggered at the ƒ setting. A
patient can trigger mandatory breaths in addition to, or in place of, the
time triggered breaths if the effort they generate causes airway pressure
to meet the Ptrig setting. Every patient effort that causes airway
pressure to meet the Ptrig setting results in a mandatory breath. Tidal
volume is determined by the V̇ and tI settings in volume control, and
determined by V̇, tI and P limit settings in pressure limited ventilation.
SIMV
(Synchronized Intermittent Mandatory Ventilation)
In SIMV, the patient receives a fixed number of time/patient triggered
mandatory breaths and may breathe spontaneously from the ventilator’s
Time-Limited Demand Flow and/or spontaneous flow. See Figure 4-1.
The first breath the patient triggers in any mandatory breath interval is
the patient triggered mandatory breath. The patient has the rest of the
interval to breathe spontaneously. If the patient stops triggering breaths
and one complete mandatory breath interval has elapsed without a
trigger, a time triggered mandatory breath is delivered.
SPONTANEOUS BREATH
PRESSURE
A B C C A A
MANDATORY
100% BREATH
LOCKOUT
INTERVAL
Figure 4-1
100% SIMV WINDOW
WARNING The SPONT V̇1 switch and/or SPONT V̇2 flowmeter must
be turned “on” in SIMV and SPONT modes to provide gas for patient
spontaneous breathing even if Time-Limited Demand Flow is
enabled.
SPONT
(SPONTANEOUS VENTILATION)
In SPONT, the patient may trigger Time-Limited Demand Flow and/or
may breathe exclusively from the spontaneous flow through the
breathing circuit and reservoir bag at the selected FIO2. The patient has
control over each breath. CPAP may be added using the PEEP/CPAP
control. Adjust the P trig control or enable ATC so that the E100M can
detect the patient’s spontaneous efforts. If detected, the Spont indicator
blinks yellow, and the apnea alarm delay interval is reset.
The patient breathes from mandatory flow during mandatory and Time-
Limited Demand Flow breaths, and from SPONT V̇1 and V̇2 flow during
spontaneous breathing. When mandatory flow is delivered to the patient,
SPONT V̇1 and V̇2 are switched off (electronically), and vice-versa. See
Figure 4-2.
DuoFlow System
Flow
Figure 4-2
DuoFlow SYSTEM
Time-Limited Demand Flow can provide a higher level (V̇ setting) of flow
to the patient at the onset of a patient triggered spontaneous breath,
after which the patient completes inspiration from the SPONT V̇1 and V̇2
flow. Providing the patient with a higher level of flow during the time they
have the highest demand can decrease imposed work of breathing.
Use of Time-Limited Demand Flow can allow the caregiver to set a lower
level of SPONT V̇1 and V̇2 since the spontaneous flow no longer needs
to meet the patient’s spontaneous peak flow needs. Setting a lower level
of SPONT V̇1 and V̇2 helps to decrease expiratory resistance for patient
exhalation and decrease gas consumption. The V̇, Time-Limited
Demand Flow, and SPONT V̇1 and V̇2 controls should be adjusted to
maximize the patient's comfort during spontaneous breathing.
F A B B A
L
O
W TIME
P
R
E A B B A
S
S TIME
U Open Closed Open Closed Open
R
E Exhalation Valve
Figure 4-3
DuoFlow Plus System
SIMV with Time-Limited Demand Flow
P limit
(Pressure Limited Ventilation)
Range: 0 to 120 cmH2O / 0 to 118 mbar
P limit
tI (Inspiratory Time)
tI (Inspiratory Time)
Figure 4-4
Pressure Limited Ventilation Open Circuit
Figure 4-5 shows an example of how the V̇ setting affects the inspiratory
pressure slope/rise and pressure plateau. The V̇ setting for breath #1 is
45 L/min and for breath #2 is 30 L/min. The tI and Plimit settings are the
same for both breaths.
Breath #1 has a much faster pressure rise due to the higher flow (50
L/min). As a result, the pressure plateau level (Plimit setting) for breath
#1 is reached rapidly, causing flow entering the airway to begin to
decrease quickly. With breath #2, flow entering the airway does not
begin to decrease until approximately 75% of the elapsed inspiratory
time.
1 #1 #2
.
L/s . . V
V = 50 L/min V = 25 L/min
1
60
cm P
H2O a
w
-20
Figure 4-5
MANDATORY FLOW V̇ SETTING
NOTE: Always verify the Plimit setting after a change is made to the V̇ or
tI controls.
NOTE: The Plimit control is factory preset at its minimum setting (fully
counter-clockwise).
_
NOTE: The purpose of setting the ▲Paw alarm when pressure
ventilating a patient is to have a secondary safety back-up in the event
of a Plimit valve failure.
Ptrig CONTROL
Range: –10 to ≥ 25 cmH20
The Ptrig control knob adjusts the patient trigger sensitivity setting on
the E100M. The trigger sensitivity setting determines the amount of
negative pressure (inspiratory effort) the patient must generate in the
breathing circuit in order for the E100M to detect the effort (trigger). On
“power up” of the ventilator, the trigger is set at –5 cmH2O/mbar.
When you rotate the knob, you will feel it click slightly. Each click
represents a trigger sensitivity level change of ~0.14 cmH2O/mbar.
Rotating it through seven clicks changes the trigger sensitivity by one
cmH2O/mbar, i.e., one pressure gauge segment. (The selected segment
blinks on the electronic pressure gauge.) Rotate the Ptrig control
counter-clockwise to increase the patient effort required to trigger and
clockwise to decrease the effort required.
The Spont indicator blinks when the E100M detects a trigger. The Apnea
time delay interval is reset at the same time.
Adjusting the Ptrig control: Adjust the Ptrig control until you reach a
setting at which it is easy for the patient to trigger the ventilator but not
likely that the ventilator will trigger from non-patient generated circuit
pressure fluctuations. From this point, non-patient generated triggers
will be refered to as "artifact triggers".
When you adjust the PEEP/CPAP control the trigger level automatically
moves in conjunction with the change in baseline pressure. If baseline
pressure changes as a result of adjusting you need to re-adjust the
trigger sensitivity manually.
If the E100M does not artifact trigger at the target trigger setting, as
determined by the definitions listed in the Artifact Trigger Conditions on
page 4-13, the trigger setting remains at the target value.
Trigger is never adjusted closer to baseline than the target trigger level
defined in the ATC Target Trigger Look Up Table and it is never adjusted
to a position further than 2 cmH2O/mbar below the most recently
assessed baseline.
Each time an artifact trigger condition is identified in any mode, ATC will
move the trigger level down (more negative) by 0.14 cmH2O/mbar but it
will never move the trigger more than 2 cmH2O/mbar below baseline.
NOTE: For information on manually adjusting the trigger level, see
previous section, Ptrig Control.
NOTE: When SPONT V̇1 or V̇2 is changed, a shift of the P base may
occur. Press the ATC button to deactivate and press it again to
reactivate the ATC function so that trigger is immediately adjusted to the
new P base.
The Auto-Set
_ (AS ) alarm button activates a function that automatically
selects ▲Paw (high pressure) and ▼Paw (low pressure) alarm settings.
The settings are selected according to the current peak airway pressure
for mandatory breaths at the time the button is pressed._ The AS
button must be pressed again in order to select new ▲Paw and ▼Paw
alarm settings after any change is made to a primary ventilator control.
Primary controls include Mode, Inspiratory Time (tI), Flow (V· ), Frequency
(ƒ) and PEEP/CPAP.
_
The ▲Paw alarm setting will increase a maximum of 10 cmH2O from its
prior setting with each _activation of the AS button. With the exception
of an increase
_ to the ▲Paw alarm setting, the AS function can move the
▼Paw and ▲Paw alarm settings anywhere within their specified ranges.
If the ▲Paw alarm continues to be violated after the 10 cmH2O increase,
the AS function turns off. This helps alert the user to the fact that the
changes made to the ventilator or changes in the patient condition have
resulted in a significant increase in peak pressure. Press the AS button
again to reactivate the AS function after verifying that all primary
controls are set at a safe level. Primary controls include Mode,
Inspiratory Time (tI), Flow (V· ), Frequency (ƒ) and PEEP/CPAP.
▼ ▲ Paw Alarm
Push to set only ▲ Paw
AS alarm indicator unlit: Alarms set manually with the Paw Alarm
control knob or a primary control has been changed since pressing AS
AS
button.
Press the AS button once to cause the high and low pressure alarm
settings to be evaluated and adjusted for the current peak airway
pressure.
Activating the AS alarm function after power up:
After the MODE, ƒ, V̇, tI, PEEP/CPAP and P limit controls are adjusted to
the appropriate settings and the patient is connected to the ventilator,
press the AS alarm button to set the alarms. The indicator on the AS
button will light indicating AS is on, and the ▲Paw and ▼P_ aw alarm
bars on the pressure gauge will blink until the selection of ▲Paw and
▼Paw settings are establised.
WARNING Make certain the patient is being ventilated at an
appropriate peak airway pressure before pressing the AS alarm
button.
2. For three consecutive breaths, the time duration of patient inhalation versus the time duration of
patient exhalation calculates out to an I:E ratio which exceeds the following limits:
• Neonate: I:E > 5:1
• Pediatric: I:E > 4:1
• Adult: I:E > 3:1
3. If the sensitivity trigger level setting is above current baseline and current baseline is within
2 cmH2O/mbar of the baseline that was assessed during the last trigger optimization process.
4. For Pediatric and Adult categories only, baseline pressure isn’t stable (no flat spot) between
triggers.
5. If the rate of pressure drop in the patient circuit that causes the E100M to trigger is less than
0.28 cmH2O/mbar per 100 ms.
A/CMV & SIMV A/CMV & SIMV A/CMV & SIMV A/CMV & SIMV
120 105 120 115 120
110 95 120 105 115
100 85 115 95 105
90 75 105 85 95
80 65 95 75 85
70 55 85 65 75
60 45 75 55 65
50 35 65 45 55
40 30 55 35 45
30 20 45 25 35
25 20 40 20 30
20 15 35 15 25
10 5 25 7 15
9 5 25 6 15
8 5 25 5 15
7 5 25 5 15
6 5 25 5 15
5 AS AS AS AS
off off off off
RS-232C SPECIFICATIONS
Scope
This section defines both the hardware and software protocol for the
RS-232 port on the Newport E100M Ventilator.
Hardware Protocol
Hardware specifications:
Pin Assignment
Pin Number Assignment
2 Transmitted Data
3 Received Data
7 Ground
Software Protocol
The E100M Ventilator has the ability to interface with external devices by
transmitting and receiving messages in variable length data packets. All
packets are preceded by an SYN character, and ended by a checksum.
The purpose of the SYN character is to enable the receiver to detect the
start of a message. Following the SYN character is the identification
character. The identification character consists of a message ID and a
message length. Checksum is the one’s complement of the sum of all
data bytes contained in the package, excluding the SYN character.
Format:
[SYN][ID][Data #1] ........................[Data #n][Checksum]
SYN: 16H
ID: Bit 5-7: identification
Bit 0-4: n number of data bytes in the packets
Checksum: One’s complement of [ID + Data #1 +......+ Data #n]
Every 10 ms, the E100M outputs the real time packet. If a request is
made by an external device for a control or monitor data packet, the
E100M will transmit the requested packet after the real time packet. An
external device can request a packet from the E100M by sending a
message that follows the above format: the identification of the data
packet it is requesting, and no data bytes (i.e. bit 0-4 of ID are 0).
Binary
Data# Bit Parameter Range Range#
Control data
Format: [SYN][ID][Data #1].....[Data #8][Checksum]
ID: 48h
Identification: Bit 5-7: 010 = 2
Length: Bit 0-4: 01000 = 8 data bytes
Binary
Data# Bit Parameter Range Range#
Monitor data
Format: [SYN][ID][Data #1].....[Data #7][Checksum]
ID: 67h
Identification: Bit 5-7: 011 = 3
Length: Bit 0-4: 00111 = 7 data bytes
Binary
Data# Bit Parameter Range Range#
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Set-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Standard Test Settings. . . . . . . . . . . . . . . . . . . . . . . . 5 - 2
Pressure Gauge Verification. . . . . . . . . . . . . . . . . . . . 5 - 3
Alarm/LED Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 3
Mixer Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4
FIO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4
ƒ (respiratory frequency) . . . . . . . . . . . . . . . . . . . . . . 5 - 4
tI (inspiratory time) . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4
SPONT V̇ 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 5
SPONT V̇ 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 5
V̇ (flow) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6
PEEP/CPAP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6
Low Baseline Alarm, Reset and Alarm Loudness. . . 5 - 6
▼P aw Alarm (low pressure) . . . . . . . . . . . . . . . . . . . . 5 - 7
▲P aw Alarm (high pressure) . . . . . . . . . . . . . . . . . . . 5 - 7
P trig (pressure trigger level) and MODE Select . . . . 5 - 8
Apnea Alarm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 8
Manual Inflation Button . . . . . . . . . . . . . . . . . . . . . . . 5 - 8
NEB (nebulizer). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 9
Internal Battery Back-Up . . . . . . . . . . . . . . . . . . . . . 5 - 10
Control Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 10
P aw Monitor (peak, mean, base) . . . . . . . . . . . . . . 5 - 10
Proximal Line Purge . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 11
Advanced Trigger Control (ATC) . . . . . . . . . . . . . . . 5 - 11
Auto-Set Alarms (AS) . . . . . . . . . . . . . . . . . . . . . . . 5 - 11
Time-Limited Demand Flow . . . . . . . . . . . . . . . . . . 5 - 12
E100M Test Record . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 13
SER1500 E0208
SECTION 5
INTRODUCTION
This procedure is intended to assist a qualified hospital technician or
NEWPORT factory authorized service technician establish a routine
verification program to ensure proper operation of the Newport E100M
Ventilator.
An Operational Verification Procedure (OVP) must be performed each
time the E100M is serviced or a minimum of once a year with the
Preventive Maintenance.
SET-UP
1. Connect a reusable adult breathing circuit with proximal line and a
test lung to the Newport E100M to be tested. See Figure 5-1.
2. Connect 50 psig air and 50 psig oxygen supplies to the ventilator
gas inlets.
3. Connect the power cord to an appropriate hospital grade electrical
outlet.
4. Set controls to STANDARD TEST SETTINGS.
5. Install the Reservoir Bag Cap (CAP100A) on the Reservoir Bag
Connection .
6. Make a copy of the Test Record at the end of this section and check
off each test as it is performed.
Reset ALARMS
E100 M V ENTILATOR
Paw
P aw
Apnea (s) Battery
15
30 Apnea
0/
MODE Paw
SPONT
Calibration
(cmH 2O)
peak
SPONT +APNEA A/CMV mean
base
SIMV
Analyzer
A/CMV
Paw Alarm P trig Spont
Push to set only P aw
I:E tot
AS ATC
VT (L)
0.1
0/ 1.0
NEB PROX
Figure 5-1
NOTE: To set Plimit, press and hold the Manual Inflation button. Turn Plimit
knob clockwise to increase pressure and counter-clockwise to decrease
pressure.
Reset ALARMS
E100 M V ENTILATOR
Paw
P aw
Apnea (s) Battery
15
30 Apnea
0/
MODE Paw
SPONT
Calibration
(cmH 2O)
peak
SPONT +APNEA A/CMV mean
base
SIMV
Analyzer
A/CMV
Paw Alarm P trig Spont
Push to set only P aw
I:E tot
AS ATC
VT (L)
0.1
0/ 1.0
NEB PROX
Figure 5-2
MIXER ALARM
1. Reduce O2 supply pressure to 28 psig ± 3 psig, verify that the mixer
audible alarm sounds.
2. Increase O2 supply pressure to 44 psig, verify that the mixer audible
alarm ceases.
3. Return O2 supply pressure to 50 psig.
4. Repeat steps 1 thru 3 for air supply.
5. Remove adult breathing circuit from the E100M.
FIO2
1. Set SPONT V̇ 2 to 15 L/min.
2. Connect standard oxygen analyzer to PATIENT outlet.
3. Set FIO2 control to .21, verify O2% per tolerance on Test Record.
4. Repeat step 3 for settings of .40, .60, .80 and 1.0.
5. Set SPONT V̇ 2 to minimum.
6. Set FIO2 to .60.
ƒ (respiratory frequency)
1. Set RT-200 to Function 45 (Rate bpm), and press Enter key, then
press Zero key.
2. Remove a section of large bore tubing from the breathing circuit and
connect one end to the PATIENT outlet of the ventilator and the
other end to the High Flow Range of the RT-200. See Figure 5-3.
3. Set t I to 0.2 sec.
4. Set MODE to SIMV.
5. Remove reservoir bag cap from reservoir outlet.
6. Set ƒ to 1, verify RT-200 reading per tolerance on Test Record.
7. Repeat step 6 for settings of 10, 20, 30, 60, 90 and 120.
8. Set ƒ to 10.
t I (inspiratory time)
1. Set RT-200 to Function 46 (Inspiratory) then press Enter key.
2. Set t I to 0.1, verify RT-200 reading per tolerance on Test Record.
3. Repeat step 2 for setting of 1.0, 2.0 and 3.0.
4. Press and hold the Manual Inflation button and verify inspiration
ends per tolerance on Test Record.
5. Release Manual Inflation button.
6. Set MODE to SPONT.
Reset ALARMS
E100 M V ENTILATOR
P aw
P aw
MODE Paw
SPONT
Calibration
(cmH 2O)
peak
SPONT +APNEA A/CMV mean
base
SIMV
Analyzer
A/CMV
Paw Alarm P trig Spont
Push to set only P aw
I:E tot
AS ATC
VT (L)
0.1
0/ 1.0
NEB PROX
Figure 5-3
SPONT V̇ 1
1. Set RT-200 to Function 35 (180 lpm O2 (%)/N2), press Enter, press
F1, then 0-6-0, then press Zero.
2. Install reservoir bag cap on reservoir outlet.
3. Flip up the SPONT V̇1 switch (8 L/min) on the left side of the E100M.
4. Verify that the RT-200 reads 7 to 9 L/min.
5. Flip down the SPONT V̇1 switch (Ø L/min).
6. Verify that the RT-200 reads less than 1 L/min.
SPONT V̇ 2
1. Set SPONT V̇2 to maximum.
2. Verify that the RT-200 reads a minimum of 36 L/min.
3. Set SPONT V̇2 to minimum.
4. Verify that the RT-200 reads less than 1 L/min.
5. Remove reservoir bag cap from reservoir outlet.
V̇ (flow)
1. Set MODE to SIMV.
2. Set V̇ to 5 L/min. and verify RT-200 reading per tolerance on Test
Record.
3. Repeat step 2 for settings of 10, 20, 40, 70 and 100 L/min.
4. Set V̇ to 40 L/min.
5. Set SPONT V̇2 to 10 L/min.
6. During mechanical inspiration, verify RT-200 reading (40 L/min) still
meets tolerance on Test Record.
7. Disconnect tube from the RT-200 and the PATIENT outlet.
8. Reassemble the breathing circuit, then reconnect the circuit and test
lung. See Figure 5-1.
PEEP/CPAP
1. Set MODE to SPONT.
2. Install the reservoir bag cap on the reservoir outlet.
3. Set PEEP/CPAP to maximum.
4. Select base by pressing the Paw selector button. Confirm that the
base LED is lit.
5. Confirm that the Electronic Pressure Gauge display and digital
pressure readout exceed 25 cmH2O/mbar.
6. Slowly reduce PEEP/CPAP to minimum and verify PEEP/CPAP can
be set to any value between 25 and 1 cmH2O/mbar.
NOTE: The Apnea alarm LED indicator will also activate after 30
seconds.
7. Briefly press and release the Manual Inflation button to reset Apnea
alarm. The Apnea alarm LED changes from flashing to latched
(steadily lit).
8. Re-attacj test lung, verify that the audible alarm is silenced and
▼Paw LED changes from flashing to latched (steadily lit) after
pressure rises above 3 cmH2O/mbar.
9. Press Reset and verify that the Apnea and ▼Paw LED indicators
extinguish.
10. Set PEEP/CPAP to minimum.
11. Set SPONT V̇2 to minimum.
APNEA ALARM
1. Set MODE to SPONT.
2. Gently squeeze and slowly release test lung to cause pressure to
meet Ptrig, simulating one spontaneous effort. Make sure Spont LED
blinks. Verify Apnea alarm LED and audible alarm are activated 30
seconds after the Spont indicator LED blinks.
3. Create another spontaneous effort simulation and verify Apnea alarm
LED latches (steadily lit) and audible alarm is silenced.
4. Set Apnea to 15 sec selection and repeat step 2 and 3.
MANUAL INFLATION BUTTON
1. Press and release the Manual Inflation button.
2. Verify a manual inflation occurs each time the Manual Inflation button
is pressed.
3. Set PEEP to 0 cmH2O/mbar.
4. Set SPONT V̇2 to minimum.
NEB (Nebulizer)
1. Set MODE to SIMV.
2. Set V̇ (flow) to 30.
3. Set t I to 3.0.
4. Set Plimit to 40 cmH2O.
5. Remove test lung from circuit.
6. Set the RT-200 to Function 36.
7. Connect the NEB outlet to the RT-200 High Flow Range inlet. See
Figure 5-4.
8. Press the NEB button, verify that the NEB LED indicator activates.
9. Press and hold Manual Inflation button until inspiration is terminated.
10. Verify flow from NEB outlet is 10 –2/+1 L/min during each manual
inspiration.
11. Press NEB button and verify NEB LED indicator extinguishes.
12. Repeat step 8 and verify that no gas is delivered from NEB outlet.
(NEB indicator should be off (unlit).)
Reset ALARMS
E100 M V ENTILATOR
Paw
P aw
Apnea (s) Battery
15
30 Apnea
0/
MODE Paw
SPONT
Calibration
(cmH 2O)
peak
SPONT +APNEA A/CMV mean
base
SIMV
Analyzer
A/CMV
Paw Alarm P trig Spont
Push to set only P aw
I:E tot
AS ATC
VT (L)
0.1
0/ 1.0
NEB PROX
Figure 5-4
CONTROL MONITOR
1. Set tI to 1.0, ƒ to 10 and set V̇ to 30.
2. Set MODE to A/CMV.
3. Select VT on the Control Monitor and verify LED activates.
4. Verify that the Control Monitor displays 0.45 to 0.55.
5. Set tI to 2.0 and set V̇ to 60.
6. Verify that the Control Monitor displays 1.8 to 2.20.
7. Select I:E and verify that the LED activates.
8. Verify that the Control Monitor displays 1:1.9 to 1:2.1.
9. Set ƒ to 60.
10. Verify that the Control Monitor displays 4.0:1 and the tI > tE LED
activates.
11. Set ƒ to 10, set tI to 1.0.
12. Verify that the Control Monitor displays 1:4.5 to 1:5.5.
13. Select ƒ tot and verify LED activates.
14. Verify that the Control Monitor displays 9 to 11. (Allow up to 60
seconds for reading to stabilize.)
FIO2
OVP Tests Pass
Set Tolerance Actual Value
Pressure Gauge Verification .21 20.9 to 24.0
Alarm/LED check .40 37.0 to 43.0
.60 57.0 to 63.0
Mixer Alarm
.80 77.0 to 83.0
FIO2
1.00 97.0 to 100
ƒ (respiratory frequency)
tI (inspiratory time) ƒ (respiratory frequency)
SPONT V̇1 Set Tolerance Actual Value
SPONT V̇2 1 1 to 2
V̇ (flow) 10 9 to 11
PEEP/CPAP 20 19 to 21
Low Baseline Alarm 30 28 to 32
Reset Button 60 57 to 63
90 85 to 95
Alarm Loudness Control
120 114 to 126
Low Pressure Alarm ( ▼Paw)
High Pressure Alarm (▲Paw) tI (inspiratory time)
Ptrig (trigger level) and MODE Select Set Tolerance Actual Value
Apnea Alarm 0.1 0.10 to 0.20
Manual Inflation Button 1.0 0.90 to 1.10
NEB (nebulizer) 2.0 1.90 to 2.10
Internal Battery Back-Up 3.0 2.90 to 3.10
Control Monitor Manual 3.90 to 4.10
Paw Monitor (peak, mean, base)
Proximal Line Purge V̇ (flow)
Set Tolerance Actual Value
Advanced Trigger Control (ATC)
5 2 to 8 L/min
Auto-Set (AS) Alarm
10 7 to 13 L/min
Time-Limited Demand Flow
20 17 to 23 L/min
40 36 to 44 L/min
70 63 to 77 L/min
100 90 to 110 L/min
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Parts Required . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Maintenance Procedure . . . . . . . . . . . . . . . . . . . 6 - 1
SER1500 E0208
SECTION 6
INTRODUCTION
Preventive Maintenance should be performed every 3000 hours or one
time per year, whichever comes first.
TOOLS REQUIRED
• 3/4 inch open end wrench
• Diagonal pliers
• Diagonal cutters
• T25 torx driver
• T15 torx driver
PARTS REQUIRED
PMK1500A (Preventive Maintenance Kit)
Kit Includes:
Qty. Part # Description
2 FTR1700P Mixer Inlet Filter
1 JFK100P Water Trap Filter Kit
1 DIA200P Exhalation Valve Diaphragm
1 FRK1545A Filter Restrictor Kit
MAINTENANCE PROCEDURE
Open Front Panel
Refer to Section 12, Final Assembly drawing, #E100MB-X, item #6 for
location of SCT0408P and item #32 for location of SCT0408M.
1. Remove the two (2) screws (SCT0408P) from Reservoir Outlet on
bottom of ventilator and remove the Outlet.
2. Remove the two (2) screws (SCT0408M) from each side of the front
panel.
3. Open front panel by pulling from the top. The retaining strap
(SST1500P) will limit how far the panel will open.
FTR1700P Replacement
Refer to Section 12, Box Assembly drawing, #BOX1550A, item #33 for
location of MIX1725A.
1. Remove the air inlet fitting from MIX1725A using a 3/4 inch wrench.
2. Remove the filter.
3. Insert the new filter.
4. Reinstall the air inlet fitting.
5. Repeat steps 1-4 for removal of the O2 side of mixer.
NOTE: The O2 side of the mixer is left handed thread.
JFK100P Replacement
Refer to Section 12, Final Assembly drawing, #E100MB-X, item #25, for
location of water trap (WTR305A).
1. Unscrew the bowl from the water trap (WTR305A).
2. Unscrew the filter holder.
3. Remove and replace the filter.
4. Remove and replace O-ring.
5. Reinstall the filter holder.
6. Reinstall the bowl.
DIA200P Replacement
1. Remove the exhalation valve located under MIX1725A from the
ventilator.
2. Unscrew the exhalation valve cap assembly.
3. Remove the knurled plastic nut from the diaphragm stem.
4. Remove and replace the diaphragm.
5. Install and tighten the knurled plastic nut.
6. Reinstall the cap assembly on the exhalation valve.
FRK1545A Replacement
Refer to Section 11, E100M System Pneumatic Diagram, #SPD1550A,
for tube locations.
1. Remove tube # 24 from 3-way toggle desig #10.
2. From the 3 way toggle, remove the desig #11 restrictor and
tube # 42 .
3. Remove tube # 13 from TEE1551P desig #6.
4. Remove tube # 17 from mixer outlet port desig #4.
5. Remove tube # 43 from exhaust port desig #6.
6. Remove tube # 50 from prox fitting desig #3.
7. Remove tube # 49 from prox solenoid desig #3.
8. To intall new FRK1545A, reverse steps 1-7.
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Parts Required . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Overhaul Procedure. . . . . . . . . . . . . . . . . . . . . . . 7 - 2
SER1500 E0208
SECTION 7
INTRODUCTION
An overhaul should be performed every 12,000 hours or one time every
5 years, whichever occurs first.
TOOLS REQUIRED
Diagonal pliers
Diagonal cutters
T25 torx driver
T15 torx driver
#2 Philips screwdriver
1/4 inch open end wrench
5/16 inch open end wrench
3/8 inch open end wrench
7/16 inch open end wrench
1/2 inch open end wrench
9/16 inch open end wrench
PARTS REQUIRED
OVL1500A & MIX1725E
Overhaul Kit and Mixer for E100M (must order both part #s)
Kit Includes:
Qty. Part # Description
1 MIX1725E Rebuilt Exchange Mixer
1 JFK100P Water Trap Filter Kit
1 DIA200P Exh. Valve Diaphragm
1 FRK1545A Filter Restrictor Kit
1 VLV150P Spont Flow Valve
1 VLV900M Umbrella Valve
1 SOL200P Master Solenoid Valve
1 BAT1500A Internal Battery Back-Up
1 CM1270 One Way Valve
1 ORG1500P O-ring
1 ORG1003P O-ring
2 ORG400P O-ring
1 GKT1500M Gasket
OVERHAUL PROCEDURE
Open Front Panel
Refer to Section 12, Final Assembly drawing, #E100MB-X, item #6 for
location of SCT0408P and item #32 for location of SCT0408M.
1. Remove two (2) screws (SCT0408P) from Reservoir Outlet on bottom
of ventilator and remove the Outlet.
2. Remove two (2) screws (SCT0408M) from each side of the front
panel.
3. Open front panel by pulling from the top of panel. The retaining strap
(SST1500P) will limit how far the panel will open.
MIX1725A Replacement
1. Disconnect the three (3) tubes from mixer.
2. Remove the four (4) nuts from left side of box and remove mixer.
3. To install new mixer, reverse steps 1 and 2.
JFK100P Replacement
Refer to Section 12, Final Assembly drawing, #E100MB-X, item #25 for
location of the water trap (WTR305A).
1. Unscrew bowl from water trap (WTR305A).
2. Unscrew filter holder.
3. Remove and replace filter.
4. Remove and replace O-ring.
5. Reinstall filter holder.
6. Reinstall bowl.
DIA200P Replacement
1. Remove exhalation valve, located under MIX1725A, from ventilator.
2. Unscrew exhalation valve cap assembly.
3. Remove knurled plastic nut from diaphragm stem.
4. Remove and replace diaphragm.
5. Install and tighten knurled plastic nut.
6. Reinstall cap assembly on exhalation valve.
FRK1545A Replacement
Refer to Section 11, E100M System Pneumatic Diagram, #SPD1550A,
for tube locations.
1. Remove tube # 24 from 3-way toggle desig #10.
2. From the 3 way toggle, remove the desig #11 restrictor and
tube # 42 .
3. Remove tube # from TEE1551P desig #6.
4. Remove tube # 17 from mixer14outlet port desig #4.
5. Remove tube # 43 from
14 exhaust
15 port desig #6.
6. Remove tube # 50 from
15 prox 38
fitting desig #3.
7. Remove tube # 49 from
38
prox solenoid desig #3.
18
8. To intall new FRK1545A, reverse steps 1-7.
18 42
VLV150P Replacement 42 47
Refer to Section 12, Box Assembly drawing, #BOX1550A, for part
47 2
location and assembly details. Be sure to apply ADH222P and
PPS1000P per details # 2 and 5 .
1. Disconnect the four (4)
5 tubes1from valve.
2. Remove the two (2) nuts holding mounting plate to the side of
1
ventilator box.
3. Remove the two (2) screws holding valve to mounting plate.
4. Remove the four (4) brass fittings from valve.
5. To install new valve, reverse steps 1-4.
15
42
Refer to Section 12, Flow Block Assembly drawing, #BLK1545A, for part
location and 47assembly details. Be sure to apply GRS129P to all O-rings
per detail # 2 .
1. Remove 5tube from side of Flow Block (BLK1500M).
2. Remove 1Carrier (CRV1002M) from top of Flow Block.
3. Remove VLV900M and ORG1003P from Carrier.
4. To install new VLV900M and ORG1003P, reverse steps 1-4.
5. Remove Block Plug (PLG1500M) from side of Flow Block.
6. Remove and replace ORG1500P.
7. Reinstall Block Plug to Flow Block.
14
8. Remove the two (2) screws (SCR6150P) from back of Flow Block.
15
9. Remove the Flow Block from Outlet Block (BLK1501M).
38
10. Remove and replace CM1270 and GKT1500M.
11. Reinstall Flow Block 18
to Outlet Block by reversing steps 8 and 9.
42
SOL200P Replacement 47
2
Refer to Section 12, Master Solenoid Assembly drawing, #SOL1500A,
5
for part location and assembly details. Be sure to apply PPS1000P to all
pipe thread per detail # 1 .
1. Disconnect the six (6) tubes from Master Solenoid Assembly
(SOL1500A).
2. Disconnect the two (2) wires from VLV1501P.
3. Loosen or remove the four (4) screws from the Master Solenoid Base
(BMS820P).
4. Remove SOL200P from Base.
5. Remove Manifold (MFD1500M) from SOL200P (do not remove any
parts from Manifold).
6. Remove the three (3) brass fittings from SOL200P.
7. To install new SOL200P, reverse steps 1-6.
BAT1500A Replacement
Refer to Section 12, Back Panel E100M drawing, #PNL1550A, for part
location and assembly details.
1. Disconnect the positive and negative battery cables.
2. Remove the battery by pulling it from the top of the rear panel.
3. To reinstall the battery, reverse steps 1 and 2.
ORG400P Replacement
Refer to Section 12, Front Panel E100M drawing, #BZL1550A, for part
location and assembly details.
1. Remove caps from tI and ƒ knobs, loosen nuts on knobs and
remove knobs.
2. Remove and replace O-rings (ORG400P) from pointers.
3. Reinstall knobs, tighten nuts and reinstall caps.
NOTE: Do not replace any tubing or tees that are a part of FRK1545A
filter restrictor kit.
FINAL PROCEDURE
Verify Electronic and Pneumatic Calibration by following the procedures
outlined in Section 8 (Electronic and Pneumatic Calibration). Once
calibration has been verified and meets all requirements as described in
Section 6, reinstall rear panel, close front bezel and perform the
Operational Verification Procedure (OVP) in Section 5. If the ventilator
successfully passes the OVP, the ventilator can be returned to patient use.
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Electronic Verification . . . . . . . . . . . . . . . . . . . . 8 - 1
Pressure Gauge Calibration . . . . . . . . . . . . . . . 8 - 3
Front Panel Calibration . . . . . . . . . . . . . . . . . . . 8 - 3
V̇ (flow) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 3
tI (inspiratory time) . . . . . . . . . . . . . . . . . . . . . . 8 - 4
ƒ (respiratory frequency) . . . . . . . . . . . . . . . . . . 8 - 4
Time-Limited Demand Flow . . . . . . . . . . . . . . . 8 - 4
PEEP/CPAP. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 5
Battery Voltage Adjustment . . . . . . . . . . . . . . . 8 - 5
Mixer Regulator Calibration . . . . . . . . . . . . . . . 8 - 5
SER1500 E0208
SECTION 8
INTRODUCTION
This procedure is intended to assist a qualified NEWPORT factory
authorized service technician in troubleshooting electronic failures and
calibration errors.
Equipment Required:
• Air supply (50 psig)
• Oxygen supply (50 psig)
• 680 ohm resistor
• Digital multimeter (DMM) Fluke 87 or equivalent
• T25 torx driver
• T15 torx driver
• 5/64 inch hex key
• RT-200 Calibration Analyzer or 0-60 psig gauge or equivalent
• 50 mL syringe or equivalent
• Container of water (for bubble test)
Reset ALARMS
E100M V E N T I L A T O R
▲ Paw
▼ Paw
Apnea (s) ▼ Battery 70 80 90
15 60 100
30 Apnea
0 50 110
40 120
MODE 30 Paw
SPONT 20 (cmH20)
10 0 -10 peak
SPONT +APNEA A/CMV mean
base
SIMV
A/CMV
▼ ▲ Paw Alarm Ptrig Spont
Push to set only ▲ Paw
I:E ƒ tot
AS ATC
V T (L)
•
F IO 2 V (L/min) tI (s) ƒ (b/min)
30 tI > tE
.60 20 40 1.5 30
.50 .70 1.0 2.0 40
20
10 50
.40 .80 5 60
MIN 60 0.5 2.5 10 80
.30 .90 100
70 0.1 1
.21 1.00 3.0 120
MAX
80
100 90 Time-Limited
Manual Demand Flow (s) PEEP/CPAP Plimit
Inflation 0.5 Pbase
0.3 0.7
0.1
0 1.0
NEB PROX
Figure 8-1
tI (inspiratory time)
The potentiometer scaled reading for tI will be displayed in the Control
Monitor when I:E ratio has been selected and the indicator is lit.
1. Turn tI knob until the Control Monitor reads 1.00.
2. Verify that the tI knob reads 1.0 on the front panel scale.
3. To adjust the tI knob to 1.0 on the front panel scale, remove the cap
from the tI knob and loosen the locking nut.
4. Adjust the tI knob so that the Control Monitor reads 1.00 and the tI
knob reads 1.0.
5. Retighten locking nut and reinstall the cap.
ƒ (respiratory frequency)
The potentiometer scaled readings for ƒ will be displayed in the Control
Monitor when ƒ tot has been selected and the indicator is lit.
1. Turn ƒ knob until the Control Monitor reads 30.
2. Verify that the ƒ knob reads 30 on the front panel scale.
3. To adjust the ƒ knob to 30 on the front panel scale, remove the cap
from the ƒ knob and loosen the locking nut.
4. Adjust the ƒ knob so that the Control Monitor and the ƒ knob read 30.
5. Retighten the locking nut and reinstall cap.
PEEP/CPAP
The potentiometer scaled readings for PEEP/CPAP will be displayed in the
Paw display when base has been selected and the base indicator is lit.
Verify PEEP/CPAP:
1. Turn the PEEP/CPAP knob to minimum.
2. Verify that the Paw display reads 0.
3. Turn PEEP/CPAP to maximum.
4. Verify that the Paw display reads a minimum of 30.
5. Return PEEP/CPAP knob to minimum.
If display at minimum setting is not 0, or display at maximum setting is not
> 30 then you must calibrate the PEEP/CPAP control according to the
following directions.
Calibration of PEEP/CPAP:
1. Remove the two screws securing the reservoir outlet on the bottom of
the ventilator and remove the outlet.
2. Remove the two screws from each side of the front panel and open
the front panel.
3. Remove the cap from the PEEP control knob, loosen the nut on the
knob and remove the knob.
4. Remove the two screws securing the BKT1504M with the PEEP
needle valve and potentiometer to the front bezel.
5. Carefully pull the needle valve assembly out of the ventilator so you
can loosen the gears.
NOTE: Do not unsolder the wires on the potentiometer from the PEEP
needle valve assembly.
6. Disconnect the tube from the outlet of the needle valve assembly. (The
outlet is the stem that comes out at a 90° angle to the needle valve.)
7. Connect a tube long enough to go from the outlet of the needle valve
to a cup of water on the workbench. (Water is used to check for
bubbles.)
8. Loosen the set screw securing the gear to the shaft of the
potentiometer. Rotate the shaft of the potentiometer fully clockwise.
9. Connect the 50 psig gas source to the ventilator, verify that flow
comes out of the needle valve outlet by the presence of bubbles in the
water.
10. Slowly rotate the shaft of the PEEP needle valve until no bubbles are
present in the cup of water. Rotate the shaft an additional 1/3 turn.
11. Secure the gear to the shaft of the potentiometer with the set screw.
12. Rotate the shaft of the needle valve fully clockwise and fully counter
clockwise. Verify the gears on the PEEP needle valve and the
potentiometer align properly throughout the rotation and the
potentiometer creates the minimum and maximum stops for the needle
valve. (Repeat steps 8 through 11 if these conditions are not met.)
13. Remove the tube from the outlet of the PEEP needle valve and
reconnect the tube from the ventilator to the PEEP needle valve.
14. Secure the bracket BKT1504M with the PEEP needle valve and
potentiometer to the ventilator with two screws.
15. Close and secure the front panel with two screws.
16. Reconnect the reservoir outlet to the bottom of the vent and secure
with two screws.
17. Perform steps 1 through 5 in Verify PEEP/CPAP above.
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 4
SER1500 E0208
SECTION 9
INTRODUCTION
The following information is intended to assist a qualified hospital
technician or NEWPORT factory authorized service technician to
determine the cause of functional failures. NEWPORT does not guarantee
that the suggested corrective action will rectify the problem.
TROUBLESHOOTING
Problem Potential Cause Suggested Corrective Action
Continuous low gas pressure/ Gas supplies did not Ensure both gas supplies exceed
gas failure alarm from mixer. exceed 40 psig upon 40 psig at mixer inlet upon
initial application of initial application of pressure.
pressure.
Mixer does not alarm when No gas to either side of Ensure gas pressure is
gas pressure drops below mixer gas inputs. applied to one side of the
25 psig. mixer.
Can't achieve or maintain Leak in patient/test circuit. Check circuit for leaks.
25 cmH2O/mbar of Replace circuit if necessary.
PEEP/CPAP.
SPONT V̇1 or V̇2 flow does not Tube #11 between spont Remove occlusion or replace
shut off during inspiration of flow valve (VLV150P) and tube if necessary.
a mandatory breath. master solenoid assembly
(SOL1500A) occluded.
No proximal purge from Prox Internal proximal circuit Check all restrictors, hoses
outlet when Prox is selected. occluded. and filter in prox circuit for
occlusion. Replace any
necessary parts.
ERROR CODES
The ventilator will not cycle, ERROR “E01” Read only memory error.
a constant audible alarm is Replace PCB1550A.
activated and an error code is
displayed in the Control ERROR “E02” Random access memory
Monitor window. error. Replace PCB1550A.
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Open Front Panel. . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Remove Rear Panel. . . . . . . . . . . . . . . . . . . . . . 10 - 1
Mixer (MIX1725A) . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Flow Control Assembly (FCL1500A) . . . . . . . . . 10 - 2
Flow Block Assembly (BLK1545A) . . . . . . . . . . 10 - 2
Nebulizer and Prox/Machine Solenoids
(SOL1503P and SOL1501P). . . . . . . . . . . . . 10 - 2
Internal Battery Back-up (BAT1500A) . . . . . . . . 10 - 2
Main Board (PCB1550A) . . . . . . . . . . . . . . . . . . 10 - 3
Display Board (PCB1511A) . . . . . . . . . . . . . . . . 10 - 3
Electronic Pressure Gauge (MAN1540M) . . . . . 10 - 3
Power Supply Board (PCB1530A) . . . . . . . . . . 10 - 4
Lower Display Board (PCB1520A) . . . . . . . . . . 10 - 4
Adapter Board (PCB1570A) . . . . . . . . . . . . . . . 10 - 4
SER1500 E0208
SECTION 10
INTRODUCTION
This section will assist you in the removal and replacement of electronic
and pneumatic components.
TOOLS REQUIRED
Small #1 phillips screwdriver
Medium #2 phillips screwdriver
TOL001P (special NEWPORT tool to loosen nuts on knobs)
7/16 inch open end wrench
1/4 inch nut driver
T25 torx driver
T15 torx driver
MIXER (MIX1725A)
1. Disconnect the three (3) tubes from the mixer.
2. Remove the four (4) nuts from the left side of the box and remove
the mixer.
3. To reinstall the mixer, reverse steps 1 and 2.
MAIN BOARD
(PCB1550A)
1. Remove the six (6) cables from PCB1550A.
2. Remove the caps from Paw Alarm and Ptrig knobs, loosen the nuts
on the knobs, then remove knobs.
3. Remove nuts from the Paw Alarm and Ptrig potentiometer shafts.
4. Remove the tube from the transducer.
5. Remove the six (6) screws from PCB1550A, then remove board.
6. To reinstall the board, reverse steps 1-5.
DISPLAY BOARD
(PCB1511A)
NOTE: PCB1511A and MAN1540M are connected by ribbon cables.
Use caution when removing them.
1. Remove PCB1550A.
2. Disconnect cable from J2 on PCB1511A.
3. Remove the cap from the MODE knob, then loosen the nut on the
knob, remove knob.
4 Remove the nut and washer from MODE control shaft.
5. Remove the two (2) standoffs (STO1500P) and the four (4) standoffs
(STO1505P).
6. For ease of removal, gently push on the manometer display while
pulling on the PCB1511A board.
7. To reinstall the PCB1511A board, reverse steps 1-6.
2. Remove PCB1511A.
NOTE: PCB1511A and MAN1540A are connected by ribbon cables.
Use caution when removing.
2.1 Disconnect cable from J2 on PCB1511A board.
2.2 Remove cap from MODE control knob, loosen nut on knob and
remove knob.
2.3 Remove nut and washer from MODE control shaft.
2.4 Remove the two (2) standoffs (STO1500) and the four (4)
standoffs (STO1505P).
3. For ease of removal, gently push on the pressure gauge display
while pulling on the PCB1511A board.
4. Remove the four (4) display ribbon cables from the connectors on
PCB1511A.
5. To install new MAN1540M, reverse steps 1-4.
ADAPTER BOARD
(PCB1570A)
1. Remove ribbon cable from Front Panel Membrane Switch to
PCB1570A.
2. Remove PCB1570A from PCB1550A.
3. To reinstall PCB1570A board, reverse steps 1-2.
SER1500 E0208
ELECTRONIC AND PNEUMATIC DIAGRAMS WITH PARTS LIST
SER1500 E0208
SECTION 11
PCB1511 SCHEMATIC
SCH1511
11-1
SECTION 11
PCB1520 SCHEMATIC
SCH1520
SER1500 E0208
SECTION 11
PCB1530 SCHEMATIC
SCH1530
11-3
ELECTRONIC AND PNEUMATIC DIAGRAMS WITH PARTS LIST
C25
15pF
Y1
24MHz
35
C24 34
15pF 15
RST
GAIN = 100K / 523 = 191.2
FLOW ASST POT 61
VCC
C17 .15uF PEEP POT 60
ASUP IT POT 59
ASUP
RATE POT 58
10 HZ LOPASS C19 C20 C21
C18
C15 R8 3
523 ASUP .1uF .1uF .1uF .1uF
.1uF C16
.1uF 1
1%
ASUP XD1 68
TP3 0cm H2O = 0.433V 67
7
8
1
5 1 +OUT 3 U3 PRESSURE1 120cm H2O = 3.94V 66
VS +OUT U4A
4
+ 6 3 65
+ R33 BATT SENSE
2 3 -OUT 2 1 64
GND -OUT - 2 ANASPARE1 63
AMP04 - C71 ANASPARE2 62
4.7K
LMC660 .1uF
SDX01G2 5%
4
5
C22 C23 4
11
.1uF .1uF 5
ASUP cw 7
FLOW 8
GAIN 9
R9 10
RV2 100K
47.5K ASUP C70 11
1% 12
13
R11 .01uF \ALARMSW 14
cw U4D 33.2K R13
RV1 1% 47.5K 6
U4B
1%
4
4
0.433V 12 0.433V 5 S1
+ 14 + 7
5K 13 R12 6
- 3.92K - ZERO
LMC660 1%
R10 LMC660
RN1 VCC
11
11
3.92K
1% \ALRM 1 14
R1 COM
\TRIG 2 13
R2 R13
\PRESSURE 3 12
R3 R12
4 11
R4 R11
5 10
R5 R10
SCL 6 9
R6 R9
SDA 7 8
R7 R8
10K
JMP1
JMP2
\SILENCE
\APNEATIME \PRESSURE
\TRIG
\ALRM
\SELECT
\PROX
\NEB
MEMBRAN
CONNECTO
TP1
VDD VCC VCC ASUP
ANALOG DIGITAL
GND PLANE GND PLANE
U4C
4
10
+ 8
9
-
LMC660
11
U21 74HC14 14 7
U8 62256 28 14
U7 27C256 32 16
U6 74HC573 20 10
U5 80C552 2 36,37
REF PART NO. VCC GND
POWER CONNECTION
SER1500 E0208
SECTION 11
U21D
8 9
VCC
74HC14
U21C
U5
16 U21B TRIGGERA
XTAL1 P1.0/CT0I
17 TRIGGERB
XTAL2 P1.1/CT1I
18 4 3 ALARMA
RST P1.2/CT2I
19 ALARMB
P1.3/CT3I
20
AVDD P1.4/T2 TRIGGER
21 74HC14
AVSS P1.5/RT2 INOP1
22
AVREF+ P1.6/SCL SCL
23
AVREF- P1.7/SDA SDA
U21A
24 DATA
STADC P3.0/RXD
25 CLOCK 2 1
P3.1/TXD
26 LATCH
P5.0/ADC0 P3.2/INT0
27 POL
P5.1/ADC1 P3.3/INT1
28 \HI VOLT ON 74HC14
P5.2/ADC2 P3.4/T0
29
P5.3/ADC3 P3.5/T1 HIPRESS
30 \WR AD[0..7]
P5.4/ADC4 P3.6/WR
31 \RD
P5.5/ADC5 P3.7/RD
P5.6/ADC6 U6 U7 U22
57 AD0 AD0 2 19 A0 A0 11 13 AD0 A0 10 11 AD0
P5.7/ADC7 P0.0/AD0 D1 Q1 A0 O0 A0 D0
56 AD1 AD1 3 18 A1 A1 10 14 AD1 A1 9 12 AD1
P0.1/AD1 D2 Q2 A1 O1 A1 D1
55 AD2 AD2 4 17 A2 A2 9 15 AD2 A2 8 13 AD2
PWM0 P0.2/AD2 D3 Q3 A2 O2 A2 D2
54 AD3 AD3 5 16 A3 A3 8 18 AD3 A3 7 15 AD3
PWM1 P0.3/AD3 D4 Q4 A3 O3 A3 D3
53 AD4 AD4 6 15 A4 A4 7 19 AD4 A4 6 16 AD4
P0.4/AD4 D5 Q5 A4 O4 A4 D4
52 AD5 AD5 7 14 A5 A5 6 20 AD5 A5 5 17 AD5
P4.0/CSMR0 P0.5/AD5 D6 Q6 A5 O5 A5 D5
51 AD6 AD6 8 13 A6 A6 5 21 AD6 A6 4 18 AD6
P4.1/CSMR1 P0.6/AD6 D7 Q7 A6 O6 A6 D6
50 AD7 AD7 9 12 A7 A7 4 22 AD7 A7 3 19 AD7
P4.2/CSMR2 P0.7/AD7 D8 Q8 A7 O7 A7 D7
A8 29 A8 25
P4.3/CSMR3 A8 A8
39 A8 11 A9 28 12 A9 24
P4.4/CSMR4 P2.0/A08 C A9 NC A9
40 A9 1 A10 24 26 A10 21
P4.5/CSMR5 P2.1/A09 OC A10 NC A10
41 A10 A11 27 1 A11 23
P4.6/CMT0 P2.2/A10 A11 DU A11
42 A11 74HC573 A12 3 17 A12 2
P4.7/CMT1 P2.3/A11 A12 DU A12
43 A12 A13 30 A13 26
P2.4/A12 A13 A13
44 A13 A14 31 A14 1
EW P2.5/A13 A14 A14
45 A14
P2.6/A14
46 A15 23 20
P2.7/A15 CE CS
25 \RD 22
OE OE
47 2 \WR 27
PSEN VCC VPP WE
48
ALE
49 27C256 62256
EA
A15 13 12
74HC14
MONITOR MICROPROCESSOR
JP2
DIN 1 2
JP3 3 4 S2
CLK
1 LOAD 5 6 1
POL
2 7 8 2
LATCH \ALARMSW
3 9 10 3
CLOCK ALARMB
4 11 12 4
DATA ALARMA
5 13 14 5
\HI VOLT ON
6 15 16 VCC 6
7 17 18
8 19 20 ENCODER
9 \SPONT 21 22 \A/C
\SPONTBKUP 23 24 \SIMV
S3
VCC
1
2
NE SWITCH
3
OR TRIGGERB
4
DISPLAY PCB TRIGGERA
5
CONNECTOR
VCC 6
ENCODER
JP1
VCC
1
2 JP4
3 1
4 2
5 NEB SOLN 3
6 PROX SOLN 4
7 EXH SOLN 5
|LINK
8 EXH SOLN SOURCE 6
|PCB15502.SCH
9
|PCB15503.SCH
10
SOLENOID CONNECTOR
Sheet 1 of 3
SCH1550
PCB1550 SCHEMATIC
SCH1550 (1 of 3)
11-4
ELECTRONIC AND PNEUMATIC DIAGRAMS WITH PARTS LIST
ASUP
cw
SPARE FOR FUTURE USE
FLOW TRIM RV6
C45 .15uF
10 Hz LOPASS
ASUP
ASUP
C43 R14 C44 0cm H2O = 0.433
.1uF 523 .1uF 120cm H2O = 3.94
1% FLOW POT
ASUP XD2 TP4
U16A PRESSURE2
IT POT
7
8
1
5 1 +OUT 3 U15
VS +OUT RATE POT
4
+ 6 3 R34
2 3 -OUT 2 + 1
GND -OUT - 2
- C32 C33 C34 C59
ASUP AMP04 4.7K
SDX01G2 LMC660 5% C66
5
.1uF .1uF .1uF .1uF .1uF
11
R15
47.5K
1% cw
\SPONT
\SIMV
cw ASUP
\A/C
U16D GAIN
RV4 \SPONTBKUP
RV3 EXHSOL
4
0.433V 12 R17 100K NEB SOL
+ 33.2K C65
5K 14 R19
13 1% U16B 47.5K INOP2 REMALARM
1%
4
- 0.433V 5 .01uF
LMC660 + 7
R16
R18 6
11
3.92K
3.92K -
1%
1%
LMC660
1
11
R1
COM
14
VCC
\NEB
\PROX
\SIL
\BATTEST
\SELECT
\APNEATIME
VCC
TP6
ASUP ASUP
U23 4.1V
8 1
VIN VOUT
7 5
FDBK ERR
3
SHTDN
R36 + C55
C68 6 110K C69
TAP 1uF
1uF 2 1% .01uF
SENS
16V 16V
LP2951
R35
47.5K
1%
ANALOG SUPPLY
TP5
VCAP
D7 R20
VCAP
10
MBRS130LT3
5% C56
1F
SUPERCAP
VCAP POWER
D1
1N4001
5
6
7
8
Q
SI
4
\EXHSOL
JP7
BATT SENSE 1
VDD VCC
2
3
2
3
VCC 4
C67 C46 C47 C48 C49 C50 C51 C52 C53
SPARE
U16C
4
10
+ 8
9
-
LMC660
11
POWER CONNECTION
REF # PART # VCC GND
U9 74HC573 20 10
U8 S80C552 2 36,37
U14 MAX232A 16 15
U13 MAX693A 3 4
U19 X2402 8 4
U11 62256 28 14
U22 27C256 32 16
U12 SCC2692 44 22
SER1500 D0105
SECTION 11
C35
15pF
Y2 JP6
24MHz SCL
C36 1
SDA
15pF 2
R2I
3
T2O
4
VCC
5
U8 U19 OPTION CONNECTOR
VCC 6
35 16 PROX SOL 1
XTAL1 P1.0/CT0I A0
34 17 2 6
XTAL2 P1.1/CT1I FLOW A1 SCL
RST 15 18 REMSIL 3 5
RST P1.2/CT2I A2 SDA
19 7
P1.3/CT3I TRIGGER WP
61 20
VCC AVDD P1.4/T2 ALARM
60 21 X2402
AVSS P1.5/RT2 LOAD
59 22
ASUP AVREF+ P1.6/SCL SCL
58 23
AVREF- P1.7/SDA SDA
3 24
STADC P3.0/RXD DIN
25
P3.1/TXD CLK
1 26
P5.0/ADC0 P3.2/INT0 HIPRESS
68 27 UART INT
P5.1/ADC1 P3.3/INT1
67 28 WDI
P5.2/ADC2 P3.4/T0
66 29
P5.3/ADC3 P3.5/T1 SOLENOID
BATT SENSE 65 30 \WR AD[0..7]
P5.4/ADC4 P3.6/WR
64 31 \RD
P5.5/ADC5 P3.7/RD
ANASPARE3 63 U9 U10 U11
P5.6/ADC6
ANASPARE4 62 57 AD0 AD0 2 19 A0 A0 11 13 AD0 A0 10 11 AD0
P5.7/ADC7 P0.0/AD0 D1 Q1 A0 O0 A0 D0
56 AD1 AD1 3 18 A1 A1 10 14 AD1 A1 9 12 AD1
C62 P0.1/AD1 D2 Q2 A1 O1 A1 D1
C63 4 55 AD2 AD2 4 17 A2 A2 9 15 AD2 A2 8 13 AD2
.1uF PWM0 P0.2/AD2 D3 Q3 A2 O2 A2 D2
.1uF 5 54 AD3 AD3 5 16 A3 A3 8 18 AD3 A3 7 15 AD3
PWM1 P0.3/AD3 D4 Q4 A3 O3 A3 D3
53 AD4 AD4 6 15 A4 A4 7 19 AD4 A4 6 16 AD4
P0.4/AD4 D5 Q5 A4 O4 A4 D4
7 52 AD5 AD5 7 14 A5 A5 6 20 AD5 A5 5 17 AD5
P4.0/CSMR0 P0.5/AD5 D6 Q6 A5 O5 A5 D5
8 51 AD6 AD6 8 13 A6 A6 5 21 AD6 A6 4 18 AD6
P4.1/CSMR1 P0.6/AD6 D7 Q7 A6 O6 A6 D6
9 50 AD7 AD7 9 12 A7 A7 4 22 AD7 A7 3 19 AD7
P4.2/CSMR2 P0.7/AD7 D8 Q8 A7 O7 A7 D7
10 A8 29 A8 25
P4.3/CSMR3 A8 A8
11 39 A8 11 A9 28 12 A9 24
P4.4/CSMR4 P2.0/A08 C A9 NC A9
12 40 A9 1 A10 24 26 A10 21
P4.5/CSMR5 P2.1/A09 OC A10 NC A10
13 41 A10 A11 27 1 A11 23
P4.6/CMT0 P2.2/A10 A11 DU A11
14 42 A11 74HC573 A12 3 17 A12 2
P4.7/CMT1 P2.3/A11 A12 DU A12
43 A12 A13 30 A13 26
P2.4/A12 A13 A13
6 44 A13 A14 31 U21E A14 1
EW P2.5/A13 A14 A14
45 A14
P2.6/A14
46 A15 23 A15 11 10 20
P2.7/A15 CE CS
25 \RD 22
OE OE
2
3
4
5
6
7
RN2 47 2 \WR 27
PSEN VCC VPP WE
10K 48 74HC14
R2
R3
R4
R5
R6
R7
ALE
49 27C256 TI SN74HC14D 62256
EA
RAM ADDR 8000-FFFF
80C552 (PRM1501A)
R13
R12
R10
R11
R9
R8
SO14 U12
A0 2 28 AD0
A0 D0
A1 4 18 AD1
13
12
10
11
9
8
A1 D1
A2 6 27 AD2
A2 D2
A3 7 19 AD3
A3 D3
CONTROLLER MICORPROCESSOR A15 39 26 AD4
CEN D4
A[0..15] \WR 9 20 AD5
WRN D5
\RD 10 25 AD6
RDN D6
UART INT 24 21 AD7
INT D7
RST 38 U14
R21 RST
35 12 13 R1I
VCC RXDA R1O R1I
32 33 11 14 T1O
4.7K OP0 TXDA T1I T1O
5% 14 11 9 8 R2I
OP1 RXDB R2O R2I
31 13 10 7 T2O
OP2 TXDB T2I T2O
15
OP3 C37 .1uF
30 8 1
22pF OP4 IP0 \PROX C1+
16 5 3 C39
OP5 IP1 \NEB C1-
29 40 C38 4 2
OP6 IP2 \SIL C2+ V+ VCC
C41 17 3 5 6
OP7 IP3 C2- V-
43 \BATTEST .1uF
IP4
Y3 36 42 MAX232A C40
X1 IP5 \SELECT .1uF
VCC 3.686MHz 37 41 .1uF
X2 IP6 \APNEATIME
C42
SCC2692
D12 22pF
1N4001 UART ADDR 0
PROX SOLN
Q7
SI9410
5
6
7
8
PROX SOL 4
2
3
VCC
D13
1N4001
VCC
NEB SOLN
Q8 R22
3.3K TP7
SI9410 5% RST
5
6
7
8
D9
RST
NEB SOL 4
R23
MBRS130LT3
VCC 20K VCC
5%
2
3
U13 R24
1 16 10K
VBATT RST
2
3
4 Q2 2 15 5%
INOP VOUT RST \PWROFF
SI9400 14
WDO \WDO
5 13
BAT ON CEI
12
EXH SOLN SOURCE CEO
6 11 WDI
LOWLN WDI
7 10
5
6
7
8
OSC IN PFO
8 9
OSCSEL PFI
MAX691A
D10 JP5
1N4001
1
REAR PANEL
2
CONNECTOR
3
VCAP 4
EXH SOLN 5
6
\BATTEST
7
5
6
7
8
13
SDA I2C 13,14
14
\SOLENOID
R26
Q5 47
BSS138 5%
ALRMON
VCC
REMALARM K1
Q6 N.C.
BSS84
D14
SG051B00
1N4001
PCB1550 SCHEMATIC
SCH1550 (2 of 3)
11-5
SER1500 E0208
INOP
U17A
2 U20A U20C
INOP1 1 1 9
U20B U17B U17C
3 3 4 8 5 8
INOP2 2 6 10 4 10 U18
74HC02 5 6 9 6 7
1C0 1Y ALRMON
74HC00 74HC00 VCAP 5
1C1
74HC00 74HC02 74HC02 4
1C2
\SILLOCK 3
\WDO ALARM R25 1C3
10K
10 9
5% 2C0 2Y
11
2C1
12
\SILENCE 2C2
13
2C3
14
\PWROFF A
2
B
1
1G
D8 15
2G
MBRS130LT3
VCAP BACKUP ALARM/VENT INOP CIRCUIT 74HC153
\SILLOCK
\SIL
U20D
12
SOLENOID 11
\SOLENOID
13
74HC00
U17D
11
13
EXHSOL \EXHSOL
12
74HC02
VCC
SCH1550 (3 of 3)
VCAP
PCB1550 SCHEMATIC
C57 C60 C58 SCH1550 Sheet 3 of 3
.1uF .1uF .1uF
11-6
SECTION 11
SECTION 11
PCB1570A
MUFFLER*
73 MICRON
.030 .025
RESTRICTOR
FILTER
73 MICRON
8L
SPONT V
QUICK CONNECT FILTER
.012
5 MICRON
FILTER
MIXER
.004 .005
.004
RESTRICTORS
.004
LINE
AIR
MIXED
GAS
PRESSURE
TRANSDUCER
MIXED
PROX PURGE
GAS
FLOW
METER PROXIMAL /
MACHINE
SOLENOID
AIR 02
P
REGUL
PROXIMAL
CONNECTOR
SER1500 E0208
SECTION 11
MUFFLER*
8LPM
SPONT V
EXHAUST
.0228
LPM RESTRICTOR
EMERGENCY
VALVE RESERVOIR
IN
CYL BAG
EXH
PILOT
VALVE
FLAPPER
VALVE
MAIN
FLOW
VALVE CYL
FLOW
CONTROL
MUFFLER
IN
RESTRICTOR
.025
.0189 RESTRICTOR
NEB SOLENOID
.26mm
(.102)
RESTRICTOR
EXHALATION
PEEP SOLENOID
LATOR
NEBULIZER PATIENT
CONNECTOR EXHALATION
VALVE
OUTLET
MUFFLER*
TO POP-OFF
PATIENT VALVE
CIRCUIT
PNEUMATIC DIAGRAM
PNU1550A
11-8
ELECTRONIC AND PNEUMATIC DIAGRAMS WITH PARTS LIST
3
3-WAY
TOGGLE
2
2
24 14
3
10 1
46
12 11 18
13
2
4 3
42 19
6
3
4 5 4
13 12
SPONT 9
VALVE
11
14 31
15
3
2 1
2 1
16
30
5
6
2
XD1
1
2
MASTER SOLENOID
ASSY.
CYL
3
IN
40
32
1
6
FLOW 18
CONTROLLER
ASSY.
34
17 16
39
2 3
36 1
EXH
2
49
1
2
1 2
(IN) (OUT)
7
20
1
(IN)
27
2 6
7
33
1 3
24 23
50
22
FLOW
3
BLOCK
ASSY. 3
4
5 48
PROX N
SER1500 E0208
SECTION 11
19 1 18
7
4 4 4
17
2 1 2 1 2 1 2
1 1 2 47 52 53 48
41 15 16
21
5 6 1 2 20
51 45
4 19
5
3
6 18 6
3
2
44
3
6
4
7 9
7
7 2 8
26
1 2
43
EXHAUST
9 6 PORTS
8 5
6 17
5
MIXER
11 OUTLET
PORTS
3
37 28
10 12
3 2
29
.
SPONT V2
21 FLOW
1
2 OUT
IN
21
6
55
5
20 2 OUT IN 1
25 38
4 6
FLOW
3 20
NEB SOLENOID
PROX
1
2
2
54
2 2 1
(OUT) (OUT) (IN)
FLOW
10 PEEP REGULATOR
35
2 1
E100M SYSTEM
NEB. EXH-VALVE PNEUMATIC DIAGRAM
SPD1550A
11-9
ELECTRONIC AND PNEUMATIC DIAGRAMS WITH PARTS LIST
JP1 (PCB1550A)
6
1 2 3 4 5 6 7 8 9 10
N\C
PEEP
JP3
CPAP
2
1
1
A B
3
2
MANUAL
T.L. INFLATION (PCB1550A)
DEMAND
FLOW
JP6
3
1
2
FLOW
CONTRL
RATE
2
JP2
3 1 1
2 3
INSP
TIME
3 1
JP7 JP5 JP1 JP4
2
JP2 IN
D E T A I L "A"
RATE
3 3 3
1 1
2
TIME
INSP
2 2
1
FLOW
S E E DETA IL A
2 CONTROL
OUT
1
CAP
3
PEEP P
NE
SOL.
PROX
SOL.
EXH
SO
3 1
1 1 1
2 2 2 2
T.L.
DEMAND
FLOW 1
JP1
A
PCB1520A
2
MA
INFL
G3
S E E DETA IL A
SER1500 E0208
SECTION 11
REF
FUSE
+
HOLDER
J3
1
2
BATTERY
PUSH
BUTTON
3 1
1 2 2
-
EXTERNA L
BATTERY
SPEAKER
17
REF
_
4 + R
ME
OR
REF SF
AN
TR
1 2 3 7
2
REF
ALARM
1 2 3
RS232 A C E
2
1 4
B D F
REMOTE 2 3
SILENCE
1 G H
NURSE POWER
CALL 2 ENTRY
INP MODULE
5 3
G1
EB 2 3
OL. HOUR
METER
MASTER 1 2
1 SOLENOID 3
B
2 G2
ANUAL 3
LATION
MAIN 2
SWITCH 1
3
EXH
EXH
CYL
IN
13
J1
J2 G4
11-10
12. ASSEMBLY DRAWINGS
WITH PARTS LIST
SER1500 E0208
ASSEMBLY DRAWINGS WITH PARTS LIST
2
32 1 1
2 PLCS
SHOWN 47
OPPOSITE SIDE 46
10
28
4
35
36 8
33
7 2
4 PLCS
SER1500 E0208
SECTION 12
1 6
9 PLCS
14 37
TYPE 'B'
16
11 SERIAL # :
RS- 232C 34
18
ALARM EXTERNAL
LOUDNESS ALARM
SILENCE
10 0 / 120 V .14A
220/240 V .07A !
50 / 60 HZ
REMOTE
ALARM
OUTPUT
6
100/ 120 V .25A, F
220/ 240 V .10A, F
Cer tifie d t o the r equiremen ts o f the C ana dian Ele ctrical Co de. C SA
has no t i nvestig ated o ther ph ysiolo gical e ff ects. D ANGER-Explo sion
Haz ard. D o no t us e i n pr esenc e o f F lam mable ana esthe tics.
Total s ystem cha ssisrisk cu rrent shou ld no t e xc eed 100uA.
Cer tifie s elon le s e xig enc es d u c ode C ana dian de l'ele ctricit e.
L'A CNOR n'a p as e tud ie le s a utr es e ff ects ph ysiolo gique s p ossi ble s.
DANGER-RI SQUE D'EXP LOSION. Nep as em ployer en pr esenc e
d'ane sthe siq ue s i nflamable s. L e c ou rant de ris que t otal
admissi ble a u cha ssis ne noit p as dep assar 100uA.
ON LR51420 R
NRT L/C
0050
STDBY
44 25
12 13
19
11
1 49
2 PLCS
8 30
17 20
E100M Final Assembly
E100MB-X
12-1
SECTION 12
29
31 3 PL
39
33
2 18
4 PLCS
19
N
EW
N MO R T
P
13 5 I
50 5
51
30
F
1O
.50 2
.60
.40
.70
.80
1.0
0
SEE DE
TAIL A
17
30
2 PLCS
35
25
26
52
37 2 9
2 PLCS 4P
LC
S
34
3 16
36
2 PLCS
2 PLCS
39
15
2
2 PLCS
14
12 13 10
5
11 13 5
42
2 PLCS
SER1500 E0208
SECTION 12
21 43 6
LCS
39 20
24 4
4 PLCS
28
23
40
46
24 7
28
3
8
26
REF 46
22 44 7
7 PLCS
47 44
2 PLCS
41 1
10
G
38 39
W
B
R
4 PLCS
4
1
1 REF
32
6
13 5
1
5
7 12
2 PLCS
12 4
24 4
3 PLCS
49
DETAIL A
8
SCALE = NON
E E100M Box Assembly
2 PLCS
BOX1550A
12-3
SECTION 12
SER1500 E0208
SECTION 12
12-6
SECTION 12
9 6
12∞±5∞
A
8 10
7 TOP VIEW
SCALE: NONE
3
6
45∞±5∞ REF
60∞±5∞
VIEW A-A
SCALE : NONE
6
5
10
EXH
EXH
CYL
IN
4
10
10
Master Solenoid
PART OF ITEM 1 Assembly
4 PLCS
FRONT VIEW
SOL1500A
SCALE: NONE
27
5
4 2 PLCS
D CK
RE BA NE
L
(+) PA
CK
BLA
(-)
25
CA
U
RE TIO
S N
TO TRIC : U
S T .S
OF ALE S TH .A. F
A P BY IS D ED
HY OR EV ERA
SIC ON IC LL
IAN TH E AW
. EO
RD
ER
10 26
TY
PE
"B
"
28
REF
17
16
15
REF
19
4
29
21
11
13
32
12
23
23
REF
SER1500 E0208
SECTION 12
5
6 7
4
4 PLCS 3
2 3
2 PLCS
9 2
4 PLCS
30
1
26
REF 2
28
15
INT
E
BA RNA
TT L
TE ERY
ST
19 REF
2 PLCS
18
2 21
REF
33
22
29
REF
14 4
1
E100M Back Panel
PNL1550A
12-10
SECTION 12
24
26 3
31 2 PLCS
6 PLCS
7 PLCS 45
IN
IN IN
39
25
53
1 52
13 66
49
PIN 1
41
41
REF
63
4 PLCS
33
10 21
23
30 MO
DE
54 SP
ON
T
SP
ON
T
+A
10 21
SIM PN
EA
V A/C
A/C
64
23 13 8 3 PLCS
REF
30
V (l
/m
in)
20 30
10
40
5
MIN 50
13 T
60 I (se
REF
MA c)
X 1.0 1.5
70
11
10 80
0 0.5
90 2.0
12 0.1 (b/m
in)
2.5 20
14
30
TIM
MA E-L
NU IM 3.0 10 40
AL DE
MA ITED
ND 60
LFO
0.3 W
0.5 80
1
8
0.1 0.7 10
0
12
0
9 0
10
1.0
30
12
3 PLCS
58
17
18
28
9
23
27
40
42
46
58 29
23
36
37 4
2 PLCS
SER1500 E0208
SECTION 12
2
J3 ON COMPONENT SID
E
16 3 9
3 15 6 PLCS
4 PLCS
1 11 3 56
10 35
2 PLCS
65
12
62
2 PLCS
4
IN
13
80 90
70 10
0
11
50 0
40 12
0
30
OUT
10
0
-10
30 6
31
68
FLOW
67 6
21
69
30
3
16 3 9
2 PLCS
26 3
3 43
2 PLCS
17
REF
44
55
32
34
33
2 PLCS
21
30 3 PLCS
31
12-12
SECTION 12
SEE DE
TAIL A
1 B 2
IN
4 4
1 3
2
2 3
OUT 4
1 4
VIEW B-B
6 8
2
7
NUT 3 9
5
5 6
7 3
BRACKET MOUNTING
SCREWS (2 P
LCS) 2
DETAIL "A"
77 81
57
56
(15)
54
81
17 2X
30 2X
21 2X
20 83 2X
38 2X
22 2X
ALARM PISTON
ASSEMBLY
PROPORTIONAL
1 2X
VALVE ASSEMBLY
3 2X
44
43 2X
67
2X 6
2X 45 82
5
2X 83 25 83 2X
30 2X
46
31
48 2X
73 2X
47 25
DETAIL 90 2X
POPORTIONAL 32 99 2X
BLOCK
ASSEMBLY 4X 33
SER1500 E0208
SECTION 12
DETAILED 72
REGULATOR (ITEM 79, 1.0"LG)
ASSEMBLY 78
(ITEM 79, 3" LG) 79
(4 1/4" LG)
59
15 2 X
58
100
(58)
(58)
1.20 1
AFFIX & CENTER
LABEL
85
69
55 82
19
82 WARNING:
68 Condensed water in the air
supply can cause malfunction
of this device.
MEDICAL OXYGEN
AIR O2
VIEW A-A
LABEL
LOCATION
SCALE: NONE
MIXER ASSEMBLY
MIX1725A
12-17
SECTION 12
1 2
3
A 4 B
A B
1
1
5
PART OF
2 ITEM 1
2 REF
REF
REF
1 REF
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 - 1
RGA (Return Goods Authorization) . . . . . . . . . . . 13 - 1
Packaging: Complete Unit . . . . . . . . . . . . . . . . . 13 - 1
Packaging: Parts or Accessories . . . . . . . . . . . . 13 - 1
SER1500 E0208
SECTION 13
INTRODUCTION
This section will help ensure that returns are properly packaged, the risk
of shipping damage is reduced and that the return is delivered to the
correct Department within Newport Medical Instruments. Please follow
the instructions below. This will help you avoid any inconveniences or
delays when returning merchandise.