Contents

Chapter 1 Brief Introduction...........................................................................................................1


1.1 Product Structure and Composition......................................................................................1
1.2 Scope of Application............................................................................................................1
1.3 System Structure...................................................................................................................1
Chapter 2 Safety Precaution...........................................................................................................2
Chapter 3 External Characteristics.................................................................................................4
3.1 Panel Instruction...................................................................................................................4
3.2 Packing.................................................................................................................................5
Chapter 4 Operation Instruction.....................................................................................................7
4.1 Operation Flow Chart...........................................................................................................7
4.2 Basic Operation Steps...........................................................................................................8
4.3 Bolus Function....................................................................................................................13
4.4 Fat Emulsion Infusion.........................................................................................................14
4.5 Change the Supplier of Infusion Pipe.................................................................................14
4.6 Load the Infusion Pipe Parameters.....................................................................................16
4.7 Battery Charge....................................................................................................................18
Chapter 5 Alarms and Solutions for Alarms.................................................................................20
Chapter 6 Product Maintenance....................................................................................................22
6.1 Cleaning and Sterilization...................................................................................................22
6.2 Preventative Testing............................................................................................................22
6.3 Environmental Production..................................................................................................23
Chapter 7 Electromagnetic Compatibility and Interference..........................................................24
Chapter 8 Product Specification...................................................................................................25
Chapter 9 Infusion Accuracy Chart..............................................................................................26

I
Chapter 1 Brief Introduction
1.1 Product Structure and Composition
Infusion Pump is composed of host computer and rechargeable battery.

1.2 Scope of Application


It is used in hospitals where patients need to be given steady and continuous intravenous
infusion or precise medication.

1.3 System Structure


The Infusion Pump is composed of several parts:

(1) The microcomputer system: The “brain” of the whole system, giving intellectualized
control and management over the whole system and processing the detected signals. It uses
two single chip Micyoco (SCM) systems for mutual backup copying and supervision. When
something is wrong with an SCM, the other one can give timely warning and cut the power of
the host computer, which will then stop running completely, so that the patient's safety can be
ensured.
(2) The pump device: The “heart” of the whole system and the main driving force of the
infusion. It drives the peristaltic pump with stepper motor.
(3) The inspection device: Infrared sensor (detecting air bubble and flux), pressure sensor
(detecting occlusion), Hall transducer (detecting motor control)etc.. They can give
corresponding signals, and after magnifying them, send them over to computer for
processing, and then operate correspondingly with the control instruction educed.
(4) The alarm device: After the signal induced by the sensor is processed by the
microcomputer, an alarm signal will be educed, which will be responded by the alarm device
to arouse people’s attention for right treatment. There are mainly two kinds of alarms,
photoelectric alarm (LBD) and sound alarm (loudspeaker and buzzer).
(5) The input and display device: Key in part of the various parameters in charge of infusion
enactment, such as infusion volume and infusion rate. It will display various parameters and
the current operation progress on the LCD displayer.
(6) Interior rechargeable battery part: This part can supply electricity for continual use of
Infusion Pump while AC powder is not available.

1
Chapter 2 Safety Precaution
Warnings

 The device must be operated by professional medical staff, such as doctors, nurses and
medical electric experts, etc.Z
 During infusion process, keep watching the amount of the left liquid in the infusion bag
or infusion bottle. Do not reply on the alarm function of the Infusion Pump only.
 Avoid using this pump in combination with MRI (Nuclear Magnetic Resonance Image)
or medical equipments analogous to this instrument, otherwise the electromagnetic
interference will cause system failure or system collapse.
 When using this pump, do not use equipments which produce electromagnetic field such
as mobile phone near this pump, otherwise the electromagnetic interference will cause
system failure or system collapse .
 Avoid using this Infusion Pump in the flammable atmosphere.
 In the high pressure atmosphere, the occlusion inspection may not correctly work.
 Before infusion, making sure the blood vessels have been protected.
 In the infusion line, the occlusion caused by pipe knot and filter coagulation or intubation
would cause the rise of inner pressure of the infusion pipe. At this moment, excluding the
cause of the occlusion can cause too much liquid to be injected to the patient’s body.
Proper measures should be taken to prevent too much liquid to be injected into the
patient’s body. For example, before eliminating the occlusion cause, squeeze the infusion
pipe.
 Use the Infusion Pump 120CM above or below the patient’s heart.
 Avoid using the Infusion Pump when there are alarms.
 If another set of infusion system or accessories is connected to the Infusion Pump to
realize the balance, the operation may be different from the specifications of the Infusion
Pump.
 Avoiding repeatedly using disposable infusion components or repeatedly sterilizing the
used components. After finishing using the components; dispose them in according to the
proper instructions.
 Only the components, connectors and disposable products which comply with the
National Standards can be used in the infusion of the system. Please consult the

2
distributor who sells the product to you about the related information.
Attention:
 When infusion lasts over 24 hours, use the new infusion components to replace the used
ones.
 Please use your fingers to press the button of the panel. Sharp tool would damage the
surface of the panel.
 Avoid the direct sun shine, abnormal high temperature or humidity.
 Use a piece of wet soft cloth with warm water to wipe the surface of the Infusion Pump
when any liquid drops on it.
 Avoid conduct autoclaving to the Infusion Pump or expose it in chemical substance.
 Before using the inner battery, check the inner battery to make sure it has sufficient
power. When necessary, recharge the battery.
 When using AC power, use the power cord included in the packing carton.
 If the Infusion Pump does not work according to the instruction manual, and the cause is
not clear, stop the Infusion Pump and report the error (including the information of the
infusion components, infusion amount, infusion rate, machine No., the type of the
infused liquid, etc.) to the distributor or manufacturer who sells the product to you.
 If inner error is inspected, the self-diagnosing function of the system will stop the work
of the equipment and give alarm.
 If the infusion set in use has not been tested, the accuracy of it can not be maintained.
The biggest error would reach 40% or above.

3
Chapter 3 External Characteristics
3.1 Panel Instruction

16

15
3 14

4 13
5 12
6 11
10
7 9

Serial
Description Function
No.
1 Indicator Light It will be twinkling under working status.
2 Tubing groove It is used for installing IV set.
3 ↑100 key 100 Used to add 100 to the initial value in the setting.
4 ↓100 key 100 Used to reduce 100 from the initial value in the setting.
5 ↓10 key Used to reduce 10 from the initial value in the setting.
6 "STOP" key Stop infusion or stop the alarm
7 "CLEAR" key Make the accumulated value to "0" in "stop" status.
Used to select the parameters for setting in the "setting"
8 "SELECT" key
interface.
1. Press the key for more than 2 seconds to turn on /off the
9 "POWER" key pump.
2. Open or Close the light version in the pump of open status.
Enter into setting of flow rate, volume limit, system
10 "SET" key
parameters.

4
1. START: start to infuse.
START/ 2. BOLUS: Press “BOLUS” key in the infusion status without
11
BOLUS key release, the pump starts to fast infusion at the maximum
rate(Rate 600ml/h).
Pull out the door lock when install the IV set and put it down
12 Door lock
after install the IV set.
13 ↓1 key Used to reduce 1 from the initial value in the setting.
14 ↑1 key Used to add 1 to the initial value in the setting.
15 ↑10 key Used to add 10 to the initial value in the setting.
Display the setting parameters, working status and so on after
16 LCD
turn on the pump.

3.2 Packing
1) Product label pattern (pasted on the shell of the pump)

2) Label Marks and Signification

Mark Description

Attention! Check the Instruction Manual!

Class II equipment

5
Type BF applied part

IP21 Keep equipment away from dropping water

Pollution-free Treatment

Date of Production

Symbol for "SERIAL NUMBER"

Caution against rain during transportation

Fragile product, handle with care

Keep upright during transportation

6
Stacking 5 layers at most for the same packaging

3) List of accessories

① Infusion Pump one unit


② AC power cord one set
③ Instruction manual one set
④ Certificate of approval one set
⑤ Maintenance card one set

Note: If any accessory is missing in the box, please contact us or our agents in your local
area.

7
Chapter 4 Operation Instruction
4.1 Operation Flow Chart

Fix the machine Fix the infusion pump on the stable IV stand.

Connect to the power source Press for 3~5 seconds.

Install IV set Install IV set

Set infusion parameters Press SET key to set parameters

Eliminate air bubbles from IV set Press "START" key to infuse and eliminate the
air bubbles from IV set, then press stop key.

Clear accumulated volume Press "CLEAR" key.

Infusion accessories insert into


patients Connect Infusion accessories with patients.

Start infusion Press "START" key.

Infusion completion Press "STOP" key.

Turn off the power supply Press for 3~5 seconds.

8
4.2 Basic Operation Steps
Step 1: Fix Infusion Pump

1) Contra rotate clamp knob, spare enough space for holding IV stand, refer to Diagram 1.
2) Fix the pump on the IV stand as Diagram 2, clockwise rotate clamp knob and fix the
pump on the IV stand.

Contra rotate clamp


knob

Diagram 1

Clockwise rotate
clamp knob

Diagram 2

Attention:

 Put the machine in the horizontal position.


 Please check the stability of IV stand before fixing the Infusion Pump.

Step 2: Connect to the power

Plug the power cord into the outlet of the Infusion Pump as Diagram 3 shows.

9
Diagram 3

Notice:

 Applicable power supply scope is AC100V-240V, 50Hz/60Hz.


 The AC power cord should be plugged into the outlet securely.

Step 3: Turn on the power supply

Press key for 3-5 seconds as Diagram 4 shows. The pump starts self-diagnosis and the
LCD will display the version No. and serial No. of this pump. After that it display as Diagram
5 shows.

Notice:

 If LCD displays AC power supply symbol, it means the pump is supplied by AC power,
and the battery is charging automatically.

 During night or in dim light, press key (just on touch) to turn on the night vision.

AC power supply

Diagram 4 Diagram 5

Step 4: Install the Infusion Pump

1) Pull out the door lock as Diagram 6 shows.

10
2) Draw the infusion pipe straight and then put the infusion pipe into the groove making
sure it reaches the bottom as Diagram 7 shows.

Pull out the


door lock

Diagram 6

doo ththe j

Diagram 7
3) Put down the door lock gently by holding the top of it until it reaches back in place as
Diagram 8 shows.
When the green indicator light is on, it means proper installation. Otherwise reinstall it until
the light is on.

11
The indicator
light is on

Pull down the handle

Diagram 8

Notice:

 If the pipe is loose or pulled too tightly, the infusion volume would get wrong.
 The infusion pipe must be discarded after 24 hours of continuous using.

Step 5: Set Basic Parameters

1) Press “set” key as Diagram 9 shows. The LCD shows 3 values as Diagram 10 shows, the
upper one is flow rate, the middle one is volume limit, the bottom one is bed No.
2) Press “select” key as Diagram 11 shows to choose the parameter which needs to be set.
When the value is flashing, which means it’s selected and can be adjusted by pressing
number keys.
3) After setting all the parameters, press “set” key to save the parameter and exit.

Notice: When the Infusion Pump is used together with our infusion supervision system, you
could set the bed No.. Otherwise, there is no need to set bed No..

Diagram 9 Diagram 10 Diagram 11

Step 6: Eliminate the Accumulated Volume

12
Press “clear” key for once as Diagram 12 shows, then the accumulated volume becomes “0”
as Diagram 13 shows.
Notice: During the infusion process, the accumulated volume can only be observed. If you
need to eliminate it,You should stop the infusion and eliminate it.

Diagram 12 Diagram 13

Step 7: Start infusion

After setting all the parameters and properly installing infusion pipe, it’s time to infuse.

Press "Start" key as Diagram 14 shows, and the indicator light is flashing as Diagram 15. The
motor begins to run, and the pump starts infusing.

Warning: Please check and make sure there is no air bubble in the pipe before infusing.
Otherwise press “start” key to discharge the air bubble from the pipe then press “stop ”key
after that, at last press “Clear” key to eliminate the accumulated volume.(When discharging
the air bubble, make sure the IV set needle didn’t insert into patient’s body.)

Notice: The pump will stop automatically once the alarm for serious low battery capacity
during the infusion process.

13
It is infusing when
the indicator light is
flashing.

Diagram 14 Diagram 15

Step 8: Infusion Finish

When the accumulated volume meets the volume limit as Diagram 16 shows, the LCD
displays "OVER" and sends an audible and visible alarm to remind the user that the infusion
is finished. Press "stop" key to stop infusion as Diagram 17 shows.

Diagram 16 Diagram 17

Step 9: Turn off AC Power

Press power key for 3-5 seconds as Diagram 18 shows, and then release it, the power
is turned off.

14
Diagram 18

4.3 Bolus Function


In the process of infusion, if the on-going rate is lower than 600ml/h and yet you need a
moment's fast-rate infusion, you may press "BOLUS" key by keeping finger on it for a while.
The pump shall infuse at 600ml/h as long as you keep your finger on "BOLUS" key as
Diagram 19 shows. After removing your finger from the key, it will run at the original
infusion rate.

Notice:

 Bolus function should be run under the condition of normal infusion.


 Bolus infusion won’t affect any alarm functions.

Diagram 19

4.4 Fat Emulsion Infusion


If you need to infuse fat emulsion for patient after you install the infusion pipe filled with fat
emulsion and press "start" key, the indicator light on top left will not be on as Diagram 20
shows, and the pump will not start operation (The indicator light is usually on when installing
the infusion pipe with simple medicine).You need to press "down 10" and "down 1" keys
together as Diagram 21 shows. Then the pump shall start operation.

15
Notice: Only when the concentration of the fat emulsion is more than or equal to 0.833% can
it be infused.

The indicator light


is not on after
installing the pipe.

Diagram 20 Diagram 21

4.5 Change the Supplier of Infusion Pipe


We test and set the parameters under the brand of "Double Dove”. If you use other brands of
infusion pipe, please reset the parameters as following steps.

Step 1:Set the Occlusion Value for Alarm

1) Install a new brand of infusion pipe filled with liquid. Press "up 100" and "down 100"
keys together as Diagram 22 shows. The LCD displays 3 values as Diagram 23 shows.
Record the upper one pressure value.
2) Press "set" and "stop" keys together as Diagram 24 shows. The LCD will display
another 3 values as Diagram 25 shows, and the upper one is the occlusion value for
occlusion alarm. The higher the value, the less sensitive the pressure sensor shall be.
Adjust a suitable occlusion value by pressing number keys as Diagram 26 shows. At last
press “set” key to save the setting.
Notice: Occlusion value>recorded pressure value+50.

Pressure value

Diagram 22 Diagram 23 Diagram 24

16
Occlusion alarm value

Accuracy

Diagram 25 Diagram 26
Step 2: Adjust Accuracy
1) Set the infusion rate at 250ml/h,and the volume limit is 10ml. Start infusion after
installing the new brand of infusion pipe properly. Use the measuring cup to measure the
liquid volume flowed from the infusion pipe.
2) Press SET and STOP keys together as Diagram 24 shows, the LCD displays 3 values as
Diagram 25 shows, the middle one is accuracy, refer to point 1, use the measuring cup to
measure the actual liquid volume flowed from the new infusion pipe. If the liquid
flowed into the measuring cup is 1ml more than 10ml, add 2 to the accuracy value. If the
actual liquid flowed into the measuring cup is 1ml less than 10ml, reduce 2 from the
accuracy value. If the actual liquid flowed into the measuring cup is the same as the
preset volume 10ml, no need to adjust the accuracy. Press SET key to select the
accuracy as Diagram 27 shows. Adjust a suitable accuracy value by pressing the number
keys as Diagram 26 shows, then press SET key to save it.
Example1: The actual liquid flowed into the measuring cup is 11ml, and the initial accuracy
value is 50, we should adjust the accuracy to 52.
Example2: The actual liquid flowed into the measuring cup is 8ml, and the initial accuracy
value is 50, we should adjust the accuracy to 46.
3) Repeat step 1 and step 2 until the accuracy value is accurate. (That means the actual
liquid volume flowed into the measuring cup is the same as the volume limit.)
Attention:
 Infusion accuracy: ±5%
 In order to reduce infusion test error, the infusion pipe should be filled with liquid,
without any air bubbles inside before operating the Infusion Pump.

17
Diagram 27

4.6 Load the Infusion Pipe Parameters


1) The Infusion Pump can save three kinds of infusion pipe parameters. It displays A, B or
C on the top right corner of LCD as Diagram 28 shows, and the A, B or C stands for the
type of infusion pipe in use.
2) Press STOP and SET keys together as Diagram 29 shows, the LCD shows the infusion
pipe’s parameters in use as Diagram 30 shows.
a. If you want to load the parameter of Type A infusion pipe, press CLEAR and
DOWN 100 keys at the same time as Diagram 31 shows, and the LCD will display
A on the top right corner.
b. If you want to load the parameter of Type B infusion pipe, press CLEAR and
DOWN 10 keys at the same time as Diagram 32 shows, and the LCD will display
B on the top right corner.
c. If you want to load the parameter of Type C infusion pipe, press CLEAR and
DOWN 1 keys at the same time as Diagram 33 shows, and the LCD will display C
on the top right corner.
3) Load the type of infusion pipe you need, and then press SET key to save as Diagram 34
shows.

Type A Type B Type C

Diagram 28

18
Occlusion
alarm value

Accuracy

Diagram 29 Diagram 30

Diagram 31 Diagram 32

Diagram 33 Diagram 34

4.7 Battery Charge


1) After connecting to the power and turning on the power supply, the Infusion Pump will
display AC power supply symbol and battery symbol as Diagram 36 shows, and begins
to charge automatically. At this time, press UP 100 and DOWN 100 keys together as
Diagram 36 shows, the LCD displays 3 values as Diagram 37 shows, the middle one is
battery capacity value. When the battery is fully charged, the value is 640. If the value is
lower than 640, it will charge automatically and the grids are in alteration state. When
the value is more than 640, it will automatically stops charging and the battery displays

19
in full grids without alteration.
2) When pulling out the wire cord as Diagram 38 shows, it will display the battery supply
symbol, which means the pump is supplied by battery and the grids mean the capacity of
battery. When the capacity is blank, the pump will send out alarm to remind the user that
the battery capacity is low and needs to recharge. The pump will give serious low
battery alarm with battery indicator flicks and send out an urgent alarm if the AC power
is not connected after a period of battery shortage. If it is under infusion state, the pump
will stop infusion automatically and keep alarm status before AC power is connected. At
the same time, the pump can not infuse.
Attention: It will take about 8~14 hours to fully recharge the battery after it is used
up.

AC power
supply symbol

Diagram 35 Diagram 36

Low battery
Battery supply symbol

Diagram 37 Diagram 38

20
Chapter 5 Alarms and Solutions for Alarms
Description Display Reason Solution

Press "stop" key to stop the


Alarm for Preset infusion infusion and the alarm. Then
infusion volume press "clear" key to clear the
completion completion. accumulated volume. After that,
press "start" key to start infusion.

Solution for No.1:


Press "stop" key to stop infusion
1.Air bubble in
and silence the alarm, remove the
the pipe
air bubble in the pipe, and then
2.Infusion pipe
press "start" key to restart
Air bubble installation
infusion.
alarm error
Solution for No.2:
3.Malfunction
Reinstall infusion pipe properly.
with the
Solution for No.3:
infrared sensor
Contact manufacturer or agent to
maintain.
Solution for No.1:
Press "stop" key to stop infusion
and silence the alarm. Press
1. Infusion "start" key to restart infusion after
loop occlusion removing the occlusion in the
2.The loop.
occlusion Solution for No.2:
Occlusion
value is much Refer to this instruction
alarm
too sensitive. manual『2.4 Method to change
3. Something different brand of Infusion Pipe:
wrong with the First step: Set the Occlusion value
sensor. for Alarm』
Solution for No.3:
Contact manufacturer or agent to
maintain.

21
1. Too low
battery. Solution for No.1:
2.Battery Connect to AC power supply to
Alarm for aging or charge the battery.
low battery something Solution for No.2:
wrong with the Contact manufacturer or agent to
battery charge maintain.
circuit.
1. No AC
power supply Solution for No.1:
for the Check if the power cord is not
Alarm for
infusion pump. plugged in or not properly
no AC
2. Something plugged in.
power
wrong with the Solution for No.2:
supply
power supply Contact manufacturer or agent to
circuit of repair.
infusion pump.
Solution for No.1:
1.Air bubble in Press "stop" key to stop the
the infusion infusion and silence the alarm.
pipe Then restart infusion after
2.Infusion pipe removing the air bubble in the
Alarm for
installation pipe.
installation
error Solution for No.2:
error
3. Something Reinstall the infusion pipe
wrong with the properly.
sensor of Solution for No.3:
infusion pump. Contact manufacturer or our
agent to repair.
Press "stop" key to stop infusion
and eliminate the alarm sound.
Alarm for Something
Then press "start" key to restart
abnormal wrong with the
infusion. If such alarm sounds
control motor.
again, please contact
manufacturer or agent to repair.

22
Chapter 6 Product Maintenance
6.1 Cleaning and Sterilization
1) Use cloth to clean the pump at once when there is liquid dropped on it.
2) Pay serious attention to cleaning the pump. Use a tampon moistened with 75% alcohol
to wipe the pump shell, after disinfection, keep the room ventilation 2 hours.
3) Make sure to turn the pump off and disconnect it from the AC power supply before
wiping it.
4) Don’t use xylene, acetone or anything analogous to clean the pump, as such chemicals
may cause damage to the plastic module.

6.2 Preventative Testing


1) Inspection of injection rate
Use the measuring bucket for testing injection volume every six months.
2) Battery Inspection
 Batteries belong to expendables. Replacing of battery needed when
they are used up. If replacing batteries are required, please contact its distributor or
the manufacturer. Model: Lithium polymer battery (7.4V,1600mAh).
 Using the battery every month till its depletion and the power of the
pump turns off automatically for insuring battery performance and prolonging its
life. After battery depletion, recharge about 8—14 hours for next use.
 The following inspection need to be taken every six months.
a. Connect to AC power about 8 hours for recharging.
b. Turn on the pump and install the infusion set.
c. Set up the infusion rate: 25ml/h and start infusion.
d. Continuous operation of the enteral feeding pump till it turns off based on low
battery.
 If the eteral feeding pump takes 90 minutes or even long time from start
infusion to shutdown, the battery is in good condition
 If the enteral feeding pump takes 45—90 minutes from start infusion to
shutdown, the battery life close to its end.
 If the enteral feeding pump takes less than 45 minutes from start infusion
to shutdown, the battery is up to its end and recharge of battery is
required.
e. After battery inspection, recharge the battery again for usage next time.

23
3) Routine Maintenance
Interval Routine Maintenance Procedure
Thoroughly clean external surfaces of the pump before and after
As per Hospital Policy prolonged period of storage.
1. Inspect AC Power supply plug and cable for damage.
2. Perform functional tests as outlined in the Technical Service
At least once per
Manual.
year(Refer to TSM for
3. Operate the pump on battery power until the low battery alarm
identification of parts)
sounds. Then charge the battery to confirm operation and
charging normally.

6.3 Environmental Production


The service life of this product is 3 years. Meanwhile over its service life should be
discarded, please contact the manufacturer or distributor for more relevant information.
(1) The enteral feeding pump that are no longer in use could be delivered to its distributors or
manufacturers for proper recycling.
(2) Used lithium polymer batteries could be delivered to its distributor or manufacturers for
handling, or handing it as per applicable laws and regulations.

24
Chapter 7 Electromagnetic Compatibility and Interference
This pump has a function to prevent external interference, including high-intensity radio
frequency radiation, magnetic field and electrostatic. Users should avoid the use of the
mobile phone within 0.5 meters from the machines.
This pump is quite low in electromagnetic frequency, which will not interference the
surrounding electronic equipments. But this pump produces certain amount of
electromagnetic radiation, which comply with IEC/EN 60601-1-2 and IEC/EN60601-2-24
standard. If interference caused when this pump is used with other equipments, measures
must be taken to reduce this interaction, such as separating one from the other machine.
Avoid using this pump in combination with nuclear magnetic resonance imaging (MRI) or
something analogous to this equipment, otherwise the electromagnetic interference will cause
mechanical failure or machine collapse.
When using this pump, do not use equipments which produce electromagnetic field such as
mobile phone near this pump, otherwise the electromagnetic interference will cause
mechanical failure or machine collapse.

25
Chapter 8 Product Specification
Bolus rate 600ml/h
Flow rate range 1ml/h~600ml/h
Preset volume range 1ml~9999 ml
Infusion accuracy ±5%
KVO rate 1ml/h(keep vein open, KVO infusion rate)
Power supply AC100V~240V, 50Hz/60Hz
Battery Rechargeable lithium polymer battery, 7.4V, 1600mAh
Maximum power consumption 25VA, running more than 4 hours after being fully recharged.
When Infusion Pump is connected to AC power, the battery will
Battery charge
automatically recharge (about 8-14 hours to fully recharge).
Fuses T 2AL 250V~
flow rate, preset volume, accumulated infusion volume, battery
Displayed information
capacity, bed No., air bubbles, occlusion, empty, infusion completion
Over, occlusion, air bubbles, low pressure, control abnormal, No AC
Alarm information
power supply, installation error of infusion pipe, serious low battery
Maximum size of the outer shell 265 mm×230 mm×175 mm (length × width × height)
Maximum weight < 3.0kg
Classification Class Ⅱ, Type BF
Work mode Continuous operation
Outer Shell material ABS Plastic
Environment temperature 5℃~+40℃, air pressure (86~106)kPa,
Operating conditions
relative humidity ≤80%
Environment temperature -20℃~+50℃, air pressure (50~106)kPa,
Storage conditions
relative humidity ≤95%
Applicable infusion pipes National standard infusion pipes

IEC60601-1:1988+A1:1991+A2:1995, IEC60601-2-24:1998,
Applied standard
IEC60601-1-2:2007, ISO14971:2007, EN980:2008

26
Chapter 9 Infusion Accuracy Chart
The following is infusion accuracy curve; it indicates the infusion movement in a certain
period after starting infusing and the performance achieve a normal infusion volume.
Note:
Infusion accuracy does not show the clinical standard, such as age, weight and other drugs
the patient use.
Infusion accuracy may be affected by environmental conditions (pressure, temperature,
humidity,the size of IV set and injectors, etc.)

27
Applied standard: EN60601-2-24

28
Chapter 10 Recommended Infusion Component
We test the Infusion Pump and set the parameters under the brand of “Dragon heart”. If you
use other brands of infusion pipes which are complied with the national standard, please reset
the parameters and adjust the accuracy as per this instruction manual.
The connection will be slacked if use the pipe-line equipment which is equipped with sliding
end connector.
Infusion accuracy may be influenced by solution consistency and proportion.

29