Chapter 1 Brief Introduction
1.1 Product Structure and Composition
Infusion Pump is composed of host computer and rechargeable battery.
(1) The microcomputer system: The “brain” of the whole system, giving intellectualized
control and management over the whole system and processing the detected signals. It uses
two single chip Micyoco (SCM) systems for mutual backup copying and supervision. When
something is wrong with an SCM, the other one can give timely warning and cut the power of
the host computer, which will then stop running completely, so that the patient's safety can be
ensured.
(2) The pump device: The “heart” of the whole system and the main driving force of the
infusion. It drives the peristaltic pump with stepper motor.
(3) The inspection device: Infrared sensor (detecting air bubble and flux), pressure sensor
(detecting occlusion), Hall transducer (detecting motor control)etc.. They can give
corresponding signals, and after magnifying them, send them over to computer for
processing, and then operate correspondingly with the control instruction educed.
(4) The alarm device: After the signal induced by the sensor is processed by the
microcomputer, an alarm signal will be educed, which will be responded by the alarm device
to arouse people’s attention for right treatment. There are mainly two kinds of alarms,
photoelectric alarm (LBD) and sound alarm (loudspeaker and buzzer).
(5) The input and display device: Key in part of the various parameters in charge of infusion
enactment, such as infusion volume and infusion rate. It will display various parameters and
the current operation progress on the LCD displayer.
(6) Interior rechargeable battery part: This part can supply electricity for continual use of
Infusion Pump while AC powder is not available.
1
Chapter 2 Safety Precaution
Warnings
The device must be operated by professional medical staff, such as doctors, nurses and
medical electric experts, etc.Z
During infusion process, keep watching the amount of the left liquid in the infusion bag
or infusion bottle. Do not reply on the alarm function of the Infusion Pump only.
Avoid using this pump in combination with MRI (Nuclear Magnetic Resonance Image)
or medical equipments analogous to this instrument, otherwise the electromagnetic
interference will cause system failure or system collapse.
When using this pump, do not use equipments which produce electromagnetic field such
as mobile phone near this pump, otherwise the electromagnetic interference will cause
system failure or system collapse .
Avoid using this Infusion Pump in the flammable atmosphere.
In the high pressure atmosphere, the occlusion inspection may not correctly work.
Before infusion, making sure the blood vessels have been protected.
In the infusion line, the occlusion caused by pipe knot and filter coagulation or intubation
would cause the rise of inner pressure of the infusion pipe. At this moment, excluding the
cause of the occlusion can cause too much liquid to be injected to the patient’s body.
Proper measures should be taken to prevent too much liquid to be injected into the
patient’s body. For example, before eliminating the occlusion cause, squeeze the infusion
pipe.
Use the Infusion Pump 120CM above or below the patient’s heart.
Avoid using the Infusion Pump when there are alarms.
If another set of infusion system or accessories is connected to the Infusion Pump to
realize the balance, the operation may be different from the specifications of the Infusion
Pump.
Avoiding repeatedly using disposable infusion components or repeatedly sterilizing the
used components. After finishing using the components; dispose them in according to the
proper instructions.
Only the components, connectors and disposable products which comply with the
National Standards can be used in the infusion of the system. Please consult the
2
distributor who sells the product to you about the related information.
Attention:
When infusion lasts over 24 hours, use the new infusion components to replace the used
ones.
Please use your fingers to press the button of the panel. Sharp tool would damage the
surface of the panel.
Avoid the direct sun shine, abnormal high temperature or humidity.
Use a piece of wet soft cloth with warm water to wipe the surface of the Infusion Pump
when any liquid drops on it.
Avoid conduct autoclaving to the Infusion Pump or expose it in chemical substance.
Before using the inner battery, check the inner battery to make sure it has sufficient
power. When necessary, recharge the battery.
When using AC power, use the power cord included in the packing carton.
If the Infusion Pump does not work according to the instruction manual, and the cause is
not clear, stop the Infusion Pump and report the error (including the information of the
infusion components, infusion amount, infusion rate, machine No., the type of the
infused liquid, etc.) to the distributor or manufacturer who sells the product to you.
If inner error is inspected, the self-diagnosing function of the system will stop the work
of the equipment and give alarm.
If the infusion set in use has not been tested, the accuracy of it can not be maintained.
The biggest error would reach 40% or above.
3
Chapter 3 External Characteristics
3.1 Panel Instruction
16
15
3 14
4 13
5 12
6 11
10
7 9
Serial
Description Function
No.
1 Indicator Light It will be twinkling under working status.
2 Tubing groove It is used for installing IV set.
3 ↑100 key 100 Used to add 100 to the initial value in the setting.
4 ↓100 key 100 Used to reduce 100 from the initial value in the setting.
5 ↓10 key Used to reduce 10 from the initial value in the setting.
6 "STOP" key Stop infusion or stop the alarm
7 "CLEAR" key Make the accumulated value to "0" in "stop" status.
Used to select the parameters for setting in the "setting"
8 "SELECT" key
interface.
1. Press the key for more than 2 seconds to turn on /off the
9 "POWER" key pump.
2. Open or Close the light version in the pump of open status.
Enter into setting of flow rate, volume limit, system
10 "SET" key
parameters.
4
1. START: start to infuse.
START/ 2. BOLUS: Press “BOLUS” key in the infusion status without
11
BOLUS key release, the pump starts to fast infusion at the maximum
rate(Rate 600ml/h).
Pull out the door lock when install the IV set and put it down
12 Door lock
after install the IV set.
13 ↓1 key Used to reduce 1 from the initial value in the setting.
14 ↑1 key Used to add 1 to the initial value in the setting.
15 ↑10 key Used to add 10 to the initial value in the setting.
Display the setting parameters, working status and so on after
16 LCD
turn on the pump.
3.2 Packing
1) Product label pattern (pasted on the shell of the pump)
Mark Description
Class II equipment
5
Type BF applied part
Pollution-free Treatment
Date of Production
6
Stacking 5 layers at most for the same packaging
3) List of accessories
Note: If any accessory is missing in the box, please contact us or our agents in your local
area.
7
Chapter 4 Operation Instruction
4.1 Operation Flow Chart
Fix the machine Fix the infusion pump on the stable IV stand.
Eliminate air bubbles from IV set Press "START" key to infuse and eliminate the
air bubbles from IV set, then press stop key.
8
4.2 Basic Operation Steps
Step 1: Fix Infusion Pump
1) Contra rotate clamp knob, spare enough space for holding IV stand, refer to Diagram 1.
2) Fix the pump on the IV stand as Diagram 2, clockwise rotate clamp knob and fix the
pump on the IV stand.
Diagram 1
Clockwise rotate
clamp knob
Diagram 2
Attention:
Plug the power cord into the outlet of the Infusion Pump as Diagram 3 shows.
9
Diagram 3
Notice:
Press key for 3-5 seconds as Diagram 4 shows. The pump starts self-diagnosis and the
LCD will display the version No. and serial No. of this pump. After that it display as Diagram
5 shows.
Notice:
If LCD displays AC power supply symbol, it means the pump is supplied by AC power,
and the battery is charging automatically.
During night or in dim light, press key (just on touch) to turn on the night vision.
AC power supply
Diagram 4 Diagram 5
10
2) Draw the infusion pipe straight and then put the infusion pipe into the groove making
sure it reaches the bottom as Diagram 7 shows.
Diagram 6
doo ththe j
Diagram 7
3) Put down the door lock gently by holding the top of it until it reaches back in place as
Diagram 8 shows.
When the green indicator light is on, it means proper installation. Otherwise reinstall it until
the light is on.
11
The indicator
light is on
Diagram 8
Notice:
If the pipe is loose or pulled too tightly, the infusion volume would get wrong.
The infusion pipe must be discarded after 24 hours of continuous using.
1) Press “set” key as Diagram 9 shows. The LCD shows 3 values as Diagram 10 shows, the
upper one is flow rate, the middle one is volume limit, the bottom one is bed No.
2) Press “select” key as Diagram 11 shows to choose the parameter which needs to be set.
When the value is flashing, which means it’s selected and can be adjusted by pressing
number keys.
3) After setting all the parameters, press “set” key to save the parameter and exit.
Notice: When the Infusion Pump is used together with our infusion supervision system, you
could set the bed No.. Otherwise, there is no need to set bed No..
12
Press “clear” key for once as Diagram 12 shows, then the accumulated volume becomes “0”
as Diagram 13 shows.
Notice: During the infusion process, the accumulated volume can only be observed. If you
need to eliminate it,You should stop the infusion and eliminate it.
Diagram 12 Diagram 13
After setting all the parameters and properly installing infusion pipe, it’s time to infuse.
Press "Start" key as Diagram 14 shows, and the indicator light is flashing as Diagram 15. The
motor begins to run, and the pump starts infusing.
Warning: Please check and make sure there is no air bubble in the pipe before infusing.
Otherwise press “start” key to discharge the air bubble from the pipe then press “stop ”key
after that, at last press “Clear” key to eliminate the accumulated volume.(When discharging
the air bubble, make sure the IV set needle didn’t insert into patient’s body.)
Notice: The pump will stop automatically once the alarm for serious low battery capacity
during the infusion process.
13
It is infusing when
the indicator light is
flashing.
Diagram 14 Diagram 15
When the accumulated volume meets the volume limit as Diagram 16 shows, the LCD
displays "OVER" and sends an audible and visible alarm to remind the user that the infusion
is finished. Press "stop" key to stop infusion as Diagram 17 shows.
Diagram 16 Diagram 17
Press power key for 3-5 seconds as Diagram 18 shows, and then release it, the power
is turned off.
14
Diagram 18
Notice:
Diagram 19
15
Notice: Only when the concentration of the fat emulsion is more than or equal to 0.833% can
it be infused.
Diagram 20 Diagram 21
1) Install a new brand of infusion pipe filled with liquid. Press "up 100" and "down 100"
keys together as Diagram 22 shows. The LCD displays 3 values as Diagram 23 shows.
Record the upper one pressure value.
2) Press "set" and "stop" keys together as Diagram 24 shows. The LCD will display
another 3 values as Diagram 25 shows, and the upper one is the occlusion value for
occlusion alarm. The higher the value, the less sensitive the pressure sensor shall be.
Adjust a suitable occlusion value by pressing number keys as Diagram 26 shows. At last
press “set” key to save the setting.
Notice: Occlusion value>recorded pressure value+50.
Pressure value
16
Occlusion alarm value
Accuracy
Diagram 25 Diagram 26
Step 2: Adjust Accuracy
1) Set the infusion rate at 250ml/h,and the volume limit is 10ml. Start infusion after
installing the new brand of infusion pipe properly. Use the measuring cup to measure the
liquid volume flowed from the infusion pipe.
2) Press SET and STOP keys together as Diagram 24 shows, the LCD displays 3 values as
Diagram 25 shows, the middle one is accuracy, refer to point 1, use the measuring cup to
measure the actual liquid volume flowed from the new infusion pipe. If the liquid
flowed into the measuring cup is 1ml more than 10ml, add 2 to the accuracy value. If the
actual liquid flowed into the measuring cup is 1ml less than 10ml, reduce 2 from the
accuracy value. If the actual liquid flowed into the measuring cup is the same as the
preset volume 10ml, no need to adjust the accuracy. Press SET key to select the
accuracy as Diagram 27 shows. Adjust a suitable accuracy value by pressing the number
keys as Diagram 26 shows, then press SET key to save it.
Example1: The actual liquid flowed into the measuring cup is 11ml, and the initial accuracy
value is 50, we should adjust the accuracy to 52.
Example2: The actual liquid flowed into the measuring cup is 8ml, and the initial accuracy
value is 50, we should adjust the accuracy to 46.
3) Repeat step 1 and step 2 until the accuracy value is accurate. (That means the actual
liquid volume flowed into the measuring cup is the same as the volume limit.)
Attention:
Infusion accuracy: ±5%
In order to reduce infusion test error, the infusion pipe should be filled with liquid,
without any air bubbles inside before operating the Infusion Pump.
17
Diagram 27
Diagram 28
18
Occlusion
alarm value
Accuracy
Diagram 29 Diagram 30
Diagram 31 Diagram 32
Diagram 33 Diagram 34
19
in full grids without alteration.
2) When pulling out the wire cord as Diagram 38 shows, it will display the battery supply
symbol, which means the pump is supplied by battery and the grids mean the capacity of
battery. When the capacity is blank, the pump will send out alarm to remind the user that
the battery capacity is low and needs to recharge. The pump will give serious low
battery alarm with battery indicator flicks and send out an urgent alarm if the AC power
is not connected after a period of battery shortage. If it is under infusion state, the pump
will stop infusion automatically and keep alarm status before AC power is connected. At
the same time, the pump can not infuse.
Attention: It will take about 8~14 hours to fully recharge the battery after it is used
up.
AC power
supply symbol
Diagram 35 Diagram 36
Low battery
Battery supply symbol
Diagram 37 Diagram 38
20
Chapter 5 Alarms and Solutions for Alarms
Description Display Reason Solution
21
1. Too low
battery. Solution for No.1:
2.Battery Connect to AC power supply to
Alarm for aging or charge the battery.
low battery something Solution for No.2:
wrong with the Contact manufacturer or agent to
battery charge maintain.
circuit.
1. No AC
power supply Solution for No.1:
for the Check if the power cord is not
Alarm for
infusion pump. plugged in or not properly
no AC
2. Something plugged in.
power
wrong with the Solution for No.2:
supply
power supply Contact manufacturer or agent to
circuit of repair.
infusion pump.
Solution for No.1:
1.Air bubble in Press "stop" key to stop the
the infusion infusion and silence the alarm.
pipe Then restart infusion after
2.Infusion pipe removing the air bubble in the
Alarm for
installation pipe.
installation
error Solution for No.2:
error
3. Something Reinstall the infusion pipe
wrong with the properly.
sensor of Solution for No.3:
infusion pump. Contact manufacturer or our
agent to repair.
Press "stop" key to stop infusion
and eliminate the alarm sound.
Alarm for Something
Then press "start" key to restart
abnormal wrong with the
infusion. If such alarm sounds
control motor.
again, please contact
manufacturer or agent to repair.
22
Chapter 6 Product Maintenance
6.1 Cleaning and Sterilization
1) Use cloth to clean the pump at once when there is liquid dropped on it.
2) Pay serious attention to cleaning the pump. Use a tampon moistened with 75% alcohol
to wipe the pump shell, after disinfection, keep the room ventilation 2 hours.
3) Make sure to turn the pump off and disconnect it from the AC power supply before
wiping it.
4) Don’t use xylene, acetone or anything analogous to clean the pump, as such chemicals
may cause damage to the plastic module.
23
3) Routine Maintenance
Interval Routine Maintenance Procedure
Thoroughly clean external surfaces of the pump before and after
As per Hospital Policy prolonged period of storage.
1. Inspect AC Power supply plug and cable for damage.
2. Perform functional tests as outlined in the Technical Service
At least once per
Manual.
year(Refer to TSM for
3. Operate the pump on battery power until the low battery alarm
identification of parts)
sounds. Then charge the battery to confirm operation and
charging normally.
24
Chapter 7 Electromagnetic Compatibility and Interference
This pump has a function to prevent external interference, including high-intensity radio
frequency radiation, magnetic field and electrostatic. Users should avoid the use of the
mobile phone within 0.5 meters from the machines.
This pump is quite low in electromagnetic frequency, which will not interference the
surrounding electronic equipments. But this pump produces certain amount of
electromagnetic radiation, which comply with IEC/EN 60601-1-2 and IEC/EN60601-2-24
standard. If interference caused when this pump is used with other equipments, measures
must be taken to reduce this interaction, such as separating one from the other machine.
Avoid using this pump in combination with nuclear magnetic resonance imaging (MRI) or
something analogous to this equipment, otherwise the electromagnetic interference will cause
mechanical failure or machine collapse.
When using this pump, do not use equipments which produce electromagnetic field such as
mobile phone near this pump, otherwise the electromagnetic interference will cause
mechanical failure or machine collapse.
25
Chapter 8 Product Specification
Bolus rate 600ml/h
Flow rate range 1ml/h~600ml/h
Preset volume range 1ml~9999 ml
Infusion accuracy ±5%
KVO rate 1ml/h(keep vein open, KVO infusion rate)
Power supply AC100V~240V, 50Hz/60Hz
Battery Rechargeable lithium polymer battery, 7.4V, 1600mAh
Maximum power consumption 25VA, running more than 4 hours after being fully recharged.
When Infusion Pump is connected to AC power, the battery will
Battery charge
automatically recharge (about 8-14 hours to fully recharge).
Fuses T 2AL 250V~
flow rate, preset volume, accumulated infusion volume, battery
Displayed information
capacity, bed No., air bubbles, occlusion, empty, infusion completion
Over, occlusion, air bubbles, low pressure, control abnormal, No AC
Alarm information
power supply, installation error of infusion pipe, serious low battery
Maximum size of the outer shell 265 mm×230 mm×175 mm (length × width × height)
Maximum weight < 3.0kg
Classification Class Ⅱ, Type BF
Work mode Continuous operation
Outer Shell material ABS Plastic
Environment temperature 5℃~+40℃, air pressure (86~106)kPa,
Operating conditions
relative humidity ≤80%
Environment temperature -20℃~+50℃, air pressure (50~106)kPa,
Storage conditions
relative humidity ≤95%
Applicable infusion pipes National standard infusion pipes
IEC60601-1:1988+A1:1991+A2:1995, IEC60601-2-24:1998,
Applied standard
IEC60601-1-2:2007, ISO14971:2007, EN980:2008
26
Chapter 9 Infusion Accuracy Chart
The following is infusion accuracy curve; it indicates the infusion movement in a certain
period after starting infusing and the performance achieve a normal infusion volume.
Note:
Infusion accuracy does not show the clinical standard, such as age, weight and other drugs
the patient use.
Infusion accuracy may be affected by environmental conditions (pressure, temperature,
humidity,the size of IV set and injectors, etc.)
27
Applied standard: EN60601-2-24
28
Chapter 10 Recommended Infusion Component
We test the Infusion Pump and set the parameters under the brand of “Dragon heart”. If you
use other brands of infusion pipes which are complied with the national standard, please reset
the parameters and adjust the accuracy as per this instruction manual.
The connection will be slacked if use the pipe-line equipment which is equipped with sliding
end connector.
Infusion accuracy may be influenced by solution consistency and proportion.
29