Professional Documents
Culture Documents
Model: GE300/GE320
Service Manual
For In Vitro Diagnostic Use
Intellectual Property Rights
Genrui owns this manual and all intellectual property rights (including
copyright) in this manual. Nobody is allowed to use, disclose or let
others obtain part or whole information of this manual in any manner
whatsoever without the written permission of Genrui. Nobody is
allowed to take the following actions to part or all contents of this
manual, including (but not limited to) shoot, reproduce, copy, or
translate it into other languages without the written permission of
Genrui.
Genrui reserves the right to modify the content of this manual without
prior notice.
I
Statement
Regarding this manual, Genrui will not take any warranty responsibility,
expressed or implied, including (but not limited to) warranties of
merchantability and fitness for any particular purpose.
Genrui will not take the responsibility for the products’ safety, reliability
and performance based on the following situations:
II
Warranty Service
Services to be charged:
Man-made damage
III
Customer Service Department
Manufacturer: Genrui Biotech Inc.
Address: 4-10F, Building 3, Geya Technology Park, Guangming
District, 518106, Shenzhen, China
Website: www.genrui-bio.com
E-mail Address: service@genrui-bio.com
Tel: +86 755 26835560
Fax: +86 755 26678789
IV
This analyzer can only be operated by test professionals, doctors
or laboratory technicians who have been trained by Genrui or its
distributors.
It is important for the hospital or organization that employs this
equipment to carry out a reasonable service/maintenance plan.
Neglect of this may result in machine breakdown or injury of
human health.
Be sure to operate the analyzer under the situation specified in
this manual; otherwise, the analyzer will not work normally and the
analysis results will be unreliable, which would damage the
analyzer components and cause personal injury.
V
Table of Contents
Chapter 1 Instrument Overview ............................................................ 1
1.1 Instrument Structure ................................................................ 1
1.2 Block Diagram of the Instrument ............................................. 4
1.3 Each Module’s Function and Components.............................. 5
1.3.1 Sampling Module .......................................................... 5
1.3.2 Distribution Valve Module .............................................. 6
1.3.3 Peristaltic Pump Module ............................................... 7
1.3.4 Electrode Module .......................................................... 8
1.3.5 CO2 Module ................................................................... 9
Chapter 2 Software Upgrade .............................................................. 10
Chapter 3 Software Debugging .......................................................... 11
3.1 Functional Test ...................................................................... 12
3.1.1 COM ............................................................................ 12
3.1.2 Print Test ..................................................................... 13
3.1.3 CO2 Test ...................................................................... 13
3.1.4 Card ............................................................................. 14
3.1.5 Probe Up ..................................................................... 14
3.1.6 Probe Down................................................................. 14
3.1.7 Valve Sensor ............................................................... 14
3.1.8 Location Sensor .......................................................... 14
3.1.9 Pump ........................................................................... 15
3.1.10 Distribution Valve....................................................... 15
3.1.11 Tube Cleaning ........................................................... 15
3.1.12 Aging Test .................................................................. 15
3.2 Restore Factory Settings ....................................................... 16
3.3 Modify Standard A, B Concentration Value ........................... 17
3.4 Modify Slope Factor .............................................................. 18
VI
Chapter 4 Maintenance and Service .................................................. 19
4.1 Maintenance .......................................................................... 19
4.1.1 Daily Maintenance ....................................................... 19
4.1.2 Weekly Maintenance ................................................... 19
4.1.3 Liquid System Inspection ............................................ 20
4.1.4 Deprotein Solution Instruction ..................................... 22
4.2 Service ................................................................................... 23
4.2.1 Common System Errors .............................................. 23
4.2.2 Ion Calibration and Measurement Abnormal............... 24
4.2.3 CO2 Calibration and Measurement Abnormal ............. 27
4.2.4 Common Liquid System Errors ................................... 29
4.3 Common Mechanical Parts’ Assembly and Dismantlement .. 31
4.3.1 Dismantlement and Assembly of Distribution valve .... 31
4.3.2 Dismantlement and Assembly of Peristaltic Pump...... 35
4.3.3 Dismantlement and Assembly of Sampling Module .... 37
4.3.4 Dismantlement and Assembly of Aspirating Cup ........ 42
4.3.5 Disassembly and Installation of Electrodes ................ 46
4.3.6 Dismantlement and Assembly of Liquid Location Sensor
.............................................................................................. 48
4.3.7 Dismantlement of CO2 Component ............................. 49
Appendix 1 Liquid System Diagram ................................................... 52
VII
Chapter 1 Instrument Overview
7---Front panel
1
Chapter 1 Instrument Overview
2
Chapter 1 Instrument Overview
3
Chapter 1 Instrument Overview
Probe
Display
Distribution
Memory Valve
Main MCU
Peristaltic
COM
Pump
R Card Digital
Potentiometer
Auxiliary MCU
Positioner
Barcode Thermal Printer AD Board
4
Chapter 1 Instrument Overview
Fig. 1-3-1
5
Chapter 1 Instrument Overview
Fig. 1-3-2
6
Chapter 1 Instrument Overview
The peristaltic pump module consists of pump head, pump tube, tube
support, mounting plate and motor, as shown in Fig. 1-3-3. The motor
of the peristaltic pump module drives the pump head to rotate and
realize the liquid pumping by squeezing and releasing the pump tube
alternatively. This module is the power source of the whole liquid
system.
Fig. 1-3-3
7
Chapter 1 Instrument Overview
Fig. 1-3-4
8
Chapter 1 Instrument Overview
Fig. 1-3-5
9
Chapter 2 Software Upgrade
JP5
JP6
10
Chapter 3 Software Debugging
Fig. 3-1
11
Chapter 3 Software Debugging
Click “Basic Test”, then the interface as shown in Fig. 3-2 displays.
Fig. 3-2
3.1.1 COM
12
Chapter 3 Software Debugging
This function could test the inner thermal printer works normally or not.
Make sure there is printing paper inside, then click “Print Test” in Fig.
3-2, if the printer is normal, it will display the instrument model and
software version on the printing paper.
Acid Pump: Click “Acid Pump”, the acid pump starts to work.
Waste Pump: Click “Waste P.”, the waste pump starts to work.
Pressure Value: Click “Pa” and press the “Aspirate” key to aspirate
sample, the current testing pressure is displayed.
13
Chapter 3 Software Debugging
3.1.4 Card
This function could test whether the reagent card could be detected or
not. Click “Card” in Fig. 3-2, if a valid card could be detected, it will
display “√” on the right side, otherwise display “×”.
3.1.5 Probe Up
This function could test whether the upper location sensor of the probe
works normally or not. Click “Probe Up” in Fig. 3-2, if the upper location
sensor works normally, it will display “√” on the right side, otherwise
display “×”.
This function could test whether the lower location sensor of the probe
works normally or not. Click “Probe Down” in Fig. 3-2, if the lower
location sensor works normally, it will display “√” on the right side,
otherwise display “×”.
This function could test whether the valve sensor works normally or not.
Click “Val Sensor” in Fig. 3-2, if the valve sensor works normally, it will
display “√” on the right side, otherwise display “×”.
This function could test whether the liquid system location sensor
works normally or not. Click “Loc Sensor” in Fig. 3-2, if the location
sensor works normally, it will display “√” on the right side, otherwise
display “×”.
14
Chapter 3 Software Debugging
3.1.9 Pump
This function could test whether the peristaltic pump motor works
normally or not. Click “Pump” in Fig. 3-2 and observe whether the
peristaltic pump motor rotates normally or not.
This function could test whether the distribution valve motor works
normally or not. Click “Allot Val” in Fig. 3-2 and observe whether the
valve motor rotates normally or not. The distribution valve motor
rotation is unobvious, please observe or listen if there is rotating sound.
This function is used to clean the liquid path and electrodes before
packaging the instrument after the production test is complete.
This function is for production aging test, it’s safe to ignore them.
15
Chapter 3 Software Debugging
This function could restore factory settings. Click “Initialize” under the
service interface to enter the initialize interface. Click “Yes” to enter into
the interface as shown in Fig. 3-3. After the initialization, all the data
will be deleted, so do not execute this function without the owner’s
agreement. Use with caution.
Fig. 3-3
16
Chapter 3 Software Debugging
Under the service interface, click “Set STD” to enter into the interface
as shown in Fig. 3-4. Click the corresponding ion’s A, B value to modify
the concentration value of standard A and B, so as to make the
concentrations be consistent with the tested values and get more
accurate results.
Fig. 3-4
17
Chapter 3 Software Debugging
Under the service interface, click “Set Slope” to enter into the interface
as shown in Fig. 3-5. Click the corresponding ion button to modify the
corresponding slope coefficient, set the calibration slope within a
certain range. Modifying the slope factor only affects the slope of the
calibration and does not affect the measured value. When the
calibration slope does not pass and is urgently needed, the calibration
slope can be modified temporarily and you can perform quality control.
Fig. 3-5
18
Chapter 4 Maintenance and Service
3). Drain the waste bottle every day and clean the waste bottle.
2). Check the inventory of the electrode filling solution, if less than 2/3
of the volume, replace it timely.
3). Check whether there is salt deposit around the electrode pole, if any,
use a wet cotton ball to clean it and use a dry cotton ball to dry it, then
reinstall it.
19
Chapter 4 Maintenance and Service
6). Observe the electrodes’ potential millivolt value under the potential
tracking interface to evaluate whether need to replace the reference
electrode and reference filling solution or not.
1). Perform the ion calibration program, observe the liquid flowing
situation inside the electrode channel and the tube, the normal
situation while switching standard A and standard B is big segments of
standard solution separated by big segments of air. While executing
CO2 calibration program, the whole tube of standard C is full of lactic
acid without bubbles. Observe whether the total liquid volume of the
CO2 reaction chamber is normal or not (the reaction pool total volume
is decided by the adjustment of the sample aspirating amount and
standard C amount), and whether there is no standard C or the total
reaction liquid volume exceeds the reaction pool’s total volume.
2). Check the tubes segment by segment. Pull out the waste tube,
probe tube and distribution valve outlet tube, then use a syringe to
push through the tube and solve the problems like poor connection,
tube blockage, tube conglutination, cracking etc..
20
Chapter 4 Maintenance and Service
3). If there’s big amount of bubbles at the connector and contact pin’s
position, it is air leaking; If abnormal sound occurred while the
peristaltic pump is rotating, maybe the pump tube is conglutinated or
clogged, check each tube and the pump tube segment by segment.
5). If the tube has good air impermeability, check if the tube is clogged
or the junction is disconnected.
21
Chapter 4 Maintenance and Service
7). Timely drain the tube connecting the CO 2 pool cover with the
pressure sensor. Clean the inner side of CO2 reaction chamber, take
down the reaction pool cover, remove the protein residual and unclog
the waste outlet, then re-assemble them and connect the tubes well. If
the reaction pool’s air-leaking, report for repairement timely.
8). If the pump tube is severely deformed (as shown in the following
figure), it will lead to insufficient aspirating amount, which can not
satisfy the instrument’s testing range, the tube should be replaced in
time.
22
Chapter 4 Maintenance and Service
4.2 Service
4.2.1 Common System Errors
Problems Possible reasons Solutions
Reconnect the
The printer interface cable
printer interface
loosened
cable
23
Chapter 4 Maintenance and Service
24
Chapter 4 Maintenance and Service
25
Chapter 4 Maintenance and Service
26
Chapter 4 Maintenance and Service
27
Chapter 4 Maintenance and Service
28
Chapter 4 Maintenance and Service
29
Chapter 4 Maintenance and Service
30
Chapter 4 Maintenance and Service
Mechanical Parts
4.3.1 Disassembly and Installation of Distribution Valve
1). Turn off the power.
2). Pull out the tubes connected with the distribution valve, avoiding
liquids flowing into the instrument. (As shown in the figure below)
31
Chapter 4 Maintenance and Service
3). Use the screwdriver to remove the four flat screws and take down
the distribution valve assembly. (As shown in the figure below)
4). Use the screwdriver to remove the pan head screws that fix the
sensor, the hex wrench to take apart the distribution valve fixing plate,
then respectively take down the distribution valve’s fixing plate and
sensor. (As shown in the figure below)
32
Chapter 4 Maintenance and Service
5). Use the hex wrench to loosen the two stop screws and take down
the distribution valve shell and axle. (As shown in the figure below)
33
Chapter 4 Maintenance and Service
6). Pull out the distribution valve’s axle, check if there’s any problem on
the O-type rubber ring, if yes, replace it. (As shown in the figure below)
7). Use the screwdriver to remove the three flat screws, then separate
the shell, cover and pressure plate, check if the O-type rubber ring’s
abraded, if yes, replace it. (As shown in the figure below)
34
Chapter 4 Maintenance and Service
8). If without above problems, use the multimeter to check whether the
sensor has problem or not.
2). Take down the tube twining on the peristaltic pump, then check if it
is severely deformed. If it’s out of the instrument’s adjustable range,
should be replaced. (As shown in the figure below)
35
Chapter 4 Maintenance and Service
3). Use the screwdriver to remove the four flat screws, take down the
peristaltic pump assembly. (As shown in the figure below)
4). Use the hex wrench to remove the two M3 pump head’s tightening
screws (During this process, check if the pump head’s tightly fixed on
the electrode axle), then take down the pump head. (As shown in the
figure below)
5). Check whether each rotating point of the pump head running
smoothly or not.
36
Chapter 4 Maintenance and Service
1). Lift the probe to the sampling position, turn off the power and take
down the probe decoration. (As shown in the figure below)
37
Chapter 4 Maintenance and Service
2). Take down the probe connected tube, remove the probe and check
if there’s dirt inside the probe, if yes, use the through-hole needle to
clean the dirt, then use the distilled water to clean and dry it. (As shown
in the figure below)
38
Chapter 4 Maintenance and Service
1). Turn off the power, remove the sampling probe cover. (As shown in
the figure below)
2). Remove the probe connected tube, use the screwdriver to loosen
the six screws fixing the double probes, take down the probes and their
clamp, check if there’s dirt inside the probe, if yes, use the
through-hole needle to clean the dirt, then use the distilled water to
clean and dry it. (As shown in the figure below)
39
Chapter 4 Maintenance and Service
1). Use the screwdriver to remove the 10 screws fixing the back cover
and take it down. (As shown in the figure below)
2). Use the screwdriver to remove the 4 hex socket screws fixing the
motion component, then take out the motion component. (As shown in
the figure below)
40
Chapter 4 Maintenance and Service
4). Check the tight degree of the synchronous belt, if too loose, adjust it
appropriately.
5). Check if there’s enough lubricating oil at the screw nut and the
bearing.
41
Chapter 4 Maintenance and Service
1). Take down the aspirating cup’s decoration. (As shown in the figure
below)
2). Use the screwdriver to take down the two M3 screws fixing the
aspirating cup, remove the tubes connecting with the aspirating cup,
then take down the aspirating cup. (As shown in the figure below)
42
Chapter 4 Maintenance and Service
3). Press on the two button placement of the aspirating cup shell to
disassemble it, check if there’s any problem, then use the distilled
water to wash away the salt deposit, etc.. (As shown in the figure
below)
43
Chapter 4 Maintenance and Service
1). Take down the aspirating cup’s decoration. (As shown in the figure
below)
2). Pull out the tube connected with the aspirating cup, remove the two
M3 inner hexagon screws fixing the lower part of the aspirating cup,
then take it out. (As shown in the figure below)
44
Chapter 4 Maintenance and Service
3). Use the hex wrench to remove the two hexagon screws fixing the
Aspirating cup’s silica gel cover. Take part the aspirating component,
check if there’s any problem, then use the distilled water to wash away
the residual proteins or salt deposit, etc.. (As shown in the figure
below)
4). After the cleaning, dry it, then install it reversely as the
disassembling steps(While doing the installation, take care to adjust
the position of the Aspirating cup, make sure the double probes
aligning at the Aspirating cup’s guide holes).
45
Chapter 4 Maintenance and Service
2). Loose the electrodes’ fixing nut. (As shown in the figure below)
46
Chapter 4 Maintenance and Service
3). Pull out each electrode from the two sides to disassemble the
electrode module and the guide shaft. (As shown in the figure below)
6). Assemble the electrodes onto the guide shaft according to the
original sequence, take care to install the O type rubber ring well
between each electrode. (As shown in the figure below)
8). Connect the tubes well, respectively insert each electrode’s plug
into the corresponding socket.
47
Chapter 4 Maintenance and Service
2). Use the screwdriver to remove the two flat screws fixing the sensor
base, then take down the sensor base. (As shown in the figure below)
3). Use the screwdriver to remove the two sensor fixing screws and
take down the sensor. (As shown in the figure below)
4). Use the multimeter to check if there’s any problem, if yes, replace it;
otherwise install it reversely as the disassembling steps.
48
Chapter 4 Maintenance and Service
2). Loosen the M3 screws by hand directly, open the CO 2 door. (As
shown in the figure below)
3). Remove the pump tube of standard C pump and the tube
connecting the CO2 reaction chamber with the aspirating cup
component, use the screwdriver to take down the 4 screws fixing the
CO2 component on the CO2 shell fixing plate. Then pull out the CO 2
component slightly, take down the flat cable from the backside of CO 2
component. (As shown in the figure below)
49
Chapter 4 Maintenance and Service
5). Refer to 4.3.2 for the peristaltic pump’s disassembling steps, check
if the peristaltic pump head rotates smoothly, the pump tube is
adhesive or deformed.
50
Chapter 4 Maintenance and Service
9). Check the whole air impermeability of the CO2 reaction system,
make sure there is no blocking, air leaking and pollution, then install it
reversely according to the disassembling steps.
51
Appendix 1 Liquid System Diagram
52
Add:4-10F, Building 3, Geya Technology Park, Guangming District,
518106, Shenzhen, China
Web:www.genrui-bio.com
P0 1 . 9 1 . 7 0 0 0 4 3- 0 1