Professional Documents
Culture Documents
Objective: To e valuate the relationships between di- tern were regressed with directly measured val ues for
rectly m easured work of breathing (WOB) and va ri - WOB.
ables of the breathing pattern commonly used at the Results: All breathing pattern vari ables poorly pre-
b edside to infer WOB for intubated, sp ontane ous ly dicted WOB as evidenced b y the low val ue s for thc
breathing patients treated with pressure supp o rt ven- coefficients ofdetermination (~). Breathing frequency
tilation (PSV). correlated positively with WOB (r=0.47, p<O.OOI) and
Design: In vivo measurements of the WOB were predicted or explained only 22 % (~=.22) of thc vari-
obtained on a consecutive series of adults. Breathing ance in WOB. VT correlatcd ncgatively and flVT and
frequency (f), tidal volume (VT), the index of rapid, TIITToT cach correlated positively with WOB. How-
shallow breathing (fIV T), the duration of r e spiratory ever, these variables predicted only 20 to 27 % of the
muscle contraction expressed as the ratio of inspira- variance in WOB. The breathing pattern score corre-
tory time over total respiratory cycle time (TIITToT)~ latcd negatively with WOB and predicted only 43% of
and a breathing pattern sco re (applied to approxi- the variance in WOB. A prediction model taking aU
mately 50% of the patients) which ranks f, VT, ste rno- va ri ab les into consideration using multiple r egression
cleidomastoid muscle activity, substernal retraction, analysis p redicted only 50% of the variance in WOB;
and abdominal paradox on a scale were variables ofthe thus, it too was a poor to moderate predictor ofWOB.
breathing pattern were also measured. The greater Conclusion: Our data reveal that WOB should be
the breathing pattern score, the lower the WOB and measured directly because variables of the breathing
vice versa. pattern commonly used at the bedside appear to be
Setting: Surgical ICUs in two university teaching hos- inaccurate and misleading inferences of the WOB. The
pitals. clinical implication of the se findings involve s the tra-
Patients: Sixty-seven adults (42 men and 25 women, ditional and empirical practice of titrating PSV based
aged 20 to 78 yea rs ) who had acute respiratory failure on the breathing pattern. Wc do not imply that the
from various etiologies were studied. All patients were patient's breathing pattern should b e ignorcd, nor
breathing spontaneously receiving continuous positive undcnnine its importance, for it provides useful diag-
airway pressure and PSV. nostic information. It appears, however, that rclying
Interventions: Intraesophageal pressure (indirect mea- primarily on the breathing pattern alone does not
surement of intrapleural pressure) was measured with provide enough infonnation to accurately assess the
an esophageal balloon integrated into a nasogastric respiratory muscle workload. Using the breathing
tube. VT was obtained by positioning a flow sen sor pattern as the primary guideline for sele cting a level
between the "Y" pieee of breathing circuit and the of PSV may result in inappropriate respiratory muscle
e nd otr ach e al tube. Data from these measurements workloads. A more comprehensive strategy is to em-
were directed to a bedside respiratory monitor (Bi- ploy WOB m easurements and the breathing pattern in
core; Allied Healthcare Products; Riverside, Calif) that a complementary manner when titrating PSV in criti-
calculates WOB u sing the CampbeU diagram. Patients cally ill patients. (CHEST 1995; 108:1338-44)
received PSV at levels d eemed reasonable to unload
the respiratory muscles. All measurements were ob-
tained after 15 to 20 min at each le vel ofPSV, averaged CPAP=continuous positive ainvay pressure ; febreathing
over I min, and then variable s of the breathing pat- frequency; flVT=ratio of brcathing frequency over tidal
volume; PSV=pressurc support ventilation; TIITTOT=ratio
of inspiratory time over total respiratory cycle time;
'From the De pa rtments of AnestheSiology (Drs. Bann er and Kirby) Vretidal volumc; WOB=work of breathing
an d Physiology (D r. Bann er), University of F lorida College of
Medicine , ana Dep art ment of Respiratory Care (Mr. Blanch),
Shan ds Hospi tal, Gainesvi lle, F la, ana the Dep art men t of Surg ery,
D ivision of Surgical Inte nsive Ca re, (D r. Kirton and Mr. De- Ke y words: breathing freq uency; breathing pattern; chest
Haven), Universi ty of Miami Schoo l of Medicine.
Th is paper was presented in part at the ann ual meeting of the wall compliance; esophageal pr essure; mechanical ventila-
Amencan Society of Anesth eslologists, October 12, 199:f, Wash - tion; pressure support venti lation; respiratory failure; respi-
ington, DC .
Man usc ript received November 3, 1994; revision accepted May 5, ratory monitor; respiratory muscle loading; work of breath-
1995. ing
~ STOMACH
VENTILATOR
.- - . . J ) :~~~~5 ~
C)
' - - - 7 "."
'---------./P"'U~TAC? 1
(MEASUREMENT Of VOLUME
ANO flOW RATE)
y
1
ESOPHAGUS I NTRAESOPHAGEAL
BALLOON CATHETER
fOR MEASURI NG
ESOPHAGEAL PRESSURE
~J C)iJi'W
MONITOR TOTAL WORK Of BREATHING
<r-;
IDIIIIIl RESISTIVE WORK
o + +
ESOPHAGEAL PRESSURE
~
I
ELASTIC WORK
IWORK Of BREATHING
1,5 Joule / L
FIGURE 1. Clinical meth od of measurin g the WOB performed by the patient (including physiologic plus
impo sed work). WOB is computed using the Campb ell diagram, which shows the change in volume plot-
ted over the change in esophageal pressure (inference of intrapleural pressure) durin g spontaneous inha-
lation (I) and exhalation (E). The chan ge in volume is measured at the connection between the "Y" piece
of the breathing circuit and the endotracheal tub e with a miniature pneu motachograph (flow sensor ).
Esophageal pressure is measured with an intraesophageal baUoonpositioned in the middle to lower third
of the esophagus. These data are directed to a computerized, port able monitor (Bicore, model CP-lOO).
Th e esophageal pressure-volume loop moves in a clockwise direction; the slope of the loop is lung com-
pliance (CL). Chest wall compliance (Ccw) is obtained previously by mechanically ventilating a relaxed
pat ient. Under these conditions, the esophageal pressure volume loop moves in a counterclockwise
direction (not shown); the slope of the loop is Ccw. (This compliance value is stored in the monitor's
computer memory and is used to construct the Campbell diagram.) Inspiratory resistive WOB includes
the physiologic resistive work on the airways and the resistive work on the breathing apparatus (vertical
lines). Elastic WOB is the tria ngular-shaped area subtended by the lung and Ccw curves (diagonal lines).
Total measured WOB , the sum of resistive and elastic work, is 1.5 JIL in this example.
It is inferred that the greater the breathing patt ern score, the lower attached to the pneumotachograph ; this was used to measure CPAP
the WOB and vice versa. The selected variables and scoring range and PSV. Data from these measurements and from the measure-
are based on our collective experience at qu antifYingchanges in the ment of chest wall compliance were processed with a portabl e,
breathin g pattern . Th e variables used in the breathing pattern score comm ercially available respiratory monitor (Bicore, model CP- 100)
are commonly used by many clinicians to assess breat hing. Only 30 that calculates WOB by the method described by Campbell' ! and
patients at one of the institutions (University of Florid a) wer e eval- Agostoni et al12 (Fig 1). This is a well-established method that has
uated with the breathing patt ern score. been used by many investigators to assess WOB during ventilatory
WOB performed by the patient was also directly measured at all support.13·16 The nasogastric tub e with the esophageal balloon and
levels of PSV. Total measured WOB , which includ es physiologic the pneum otachograph are components of the monitor , which also
(elastic and resistive ) work and the imposed resistive work of the provides real-time measurements of f, fi'VT, and TIffTOT . Th e de-
breathing appa ratus (endotracheal tub e , breath ing circuit tub ing, gree of ste rnocleidomastoid muscle activity, substernal retraction,
and ventilat or's demand-flow system ) during spont aneous ventila- and abdominal paradox were ranked based on clinical presentation,
tion was calculated by integrating the change in intraesophageal ie, observati on and palpation by a board-certified anesthesiologist
pressure (indirect measurement of intr apleu ral pressure ) and VT.9 who was an attendin g physician in critical care medicine.
Intraesophageal pressure was measured (relative to atmospheric Accuracy in measuring chest wall compliance requires a relaxed
pressur e) by the nasal insertion of an esophageal balloon that is in- and mech anically ventilated patient. Patients were initially given 1
tegrated into a nasogastric tub e (Smart Cath, Nasogastric Balloon to 2 mg of midazolam for relaxation and then the mechani cal ven-
Cathete r; Bicore: Allied Health care Products; Riversid e, Calif). tilator rate was increased transiently to about 12 breaths/min. Un-
Correct position was confirm ed with an occlusion test as described der these condition s of mechanical inflation with a preselected VT
by Baydur et al:1O following occlusion of the airway opening, the esophageal pressure increases. The monitor integrates the changes
changes in pressure at the airway and in the esophagus are close to in esophageal pressure and volume to produce a pressure-volume
identical during spontaneous inspiratory efforts. VTwas obtained by loop moving in a counterclockwise direction, the slope of which was
integrating the flow signal of a pneum otachograph (flow sensor ) taken as the chest wall compliance. This compliance value was
positioned betwe en the "Y" piece of the breathing circuit and the stored in the monitor's compute r memo ry. Measured chest wall
endotrache al tub e . Airway pressure was measured from a cathe ter compliance values were 109± 37 mUcm H20. Next, 0.2 mg of flu-
CO 1.0
formed by the ventilator. Q) Q)
•
•
.... 0>
it
In a validation study of the accuracy of the monitor, 18 WOB was CO c
6
measured by using the monitor and simultaneously using conven- '0
~ E 0. 3
tional equipment to construct a Ca mpbell diagram. The relationship .... 0
between the two sets of measurem ents of WO B was nearly perfect o z
(r'!=0.99, p <O.oo I) . Bias was minimal (- 0.05 JIL ) and precision was ~ o 5 10 15 20 25 30 35 40
excellent (+0.03 JIL ). '------..y----'
All measu rement s (over 200 detenninations) were obtained af- Clinically acceptable range
ter IS to 20 min at each level of PSV, 5 to 10 pe r test, averaged over Spo ntaneo us Breat hing Frequency (per m in )
1 min, and then reco rded. Measurements were record ed only if they
eould be consistently rep rod uced during each set of tests; thus, in- FIG URF: 2. Helationship be twee n \ VO B and spontaneous f is de-
tern al test eonsistency (reliability of measurem ent) was ensure d. picted. Although a positive and significant eorre lation was found
When appropriate, arte rial blood gas and hemodynamic data were (r=0.47, p <O.OOl ), spo ntaneous f pred icted or explained only 22%
of the variance in WOB , ie, a poor predictor. With in a clinically
co llected at various levels of PSV. All variables or inferences of
accep table f range of 15 to 25 breaths/min, some patient s performed
WOB were regressed with directly measured values for WOB at all no work (predisposing to disuse respira tory muscle atrophy), oth-
levels of PSV. Data were analyzed by linear, polynomial (seeond ers performed work in a norm al range, and some perform ed work
order), and multiple regre ssion analyses. Alpha was set at 0.05 for in a fatiguing range (predisposing to resp iratory muscle fatigue ).
statistical significance.
R ESULTS
sideration . Only three variables were significant pre-
Measured values for WOB ranged from 0 to 2.2 JIL. dictors ofWOB: breathing pattern score, TIffTOT, and
The level of PSV applied ranged from 5 to 50 em H 20 . VT. However, because it predicted only 50% of the
All breathing pattern variables poorly predicted WOB variance in WOB, the model was a poor to moderate
as evidenced by the low values for the coefficients of predictor of WOB (Table 3).
determination ( ~). For each variable, ~ predicts or For most ventilatory conditions, all patients were in
explains the amoun t of variance in WOB. A variable hemodynamically stable condition and blood gas ex-
with an ~ value between 0.64 and 0.81 is considered change was within acceptable ranges.
high and thus, a fairly good predictor. 19 The f corre-
DISC USSIO;o.J
lated positively with WOB and predicted only 22% of
the variance in WOB (Table 2 and Fig 2). VT correlated The main finding of this study is that, for adults with
negatively and for fNT and TIffTOT each correlated abnormal pulmonary mechanics and loaded respira-
positively with WOB. However, these variables pre- tory muscles who are in respiratory failure and being
dicted 20 to 27% of the variance in WOB (Table 2). treated with PSV, the f and the breathing pattern are
The breathing patt ern score correlated negatively with poor inferences of the WOE. Our study reveals that
WOB and it predicted only 43% of the variance in WOB should be measured directly because f, for ex-
WOB (Table 2, Fig 3). ample, appears to be an inaccurate and misleading
By using multiple regression analysis, a prediction variable from which to infer the respiratory muscle
model was formulated that took all variables into con- workload. The clinical implication of these findings
-l • •
(r2=O.50, p<O.OOl )
• •
...... WOB=2.14+0.0002 (D- 0.44 (VT)*- O.OO3 (fN T)
2 +2.13 (TuTToT)*- 0.21 (breathing pattern score)"
Q)
•• ••••
::J
0 *VT, the duration of respiratory muscle contraction expressed as
:3 TuTTOT, and the breathing pattern score were the only significant
1 .5
(p<0.05) predictors (regressors) in the model, where f is breathing
• •
C)
c
•
frequen cy (breaths/min) and fNT is the index of rapid, shallow
x:
......
til • breathing (breaths/rnin-L). The factor "Institution" (University of
Florida and University of Miami) is not shown because it also was not
•
Q)
L-
eo a Significant regressor. The model predicts or explains Dilly 50% of the
......
0
.5
• • variance in WOB, ie, a relatively poor to moderate prediction model.
sc
L-
0
3 dex of rapid shallow br eathin g does not app ear specific
0 enough to infer WOB when applying PSV. Moreover ,
0 2 4 6 8 10 regarded to be an accurate predictor of weaning fail-
BreathIng Patte rn Scor e ur e, the index has also been criticized as poorly
predicting extubation outcorne.P Th e ind ex lacked
FIGURE 3. Relationship between WOB and th e breathin g pattern
score (Ta ble 1) is shown . WOB correlated negatively and signifi-
sufficient sen sitivity and speci ficity to differen tiate
cantly with the breathin g pattern score (r=- 0.65, p <O.OO l) . H ow- between patients who were successfully extubated and
ever, the breathing patte rn score was a poor pred ictor explaining tho se who we re not.
only 43% of the variance in WOB. ideally, a breath ing pattern score
of 10 is an inference of a normal wor kload (0.3 to 0.6 JIL). How- Because it predicted only 50% of the variance in
eve r, with this score, some patients pe rforme d no work, others WOB , the prediction model is not considered a clin-
pe rforme d work in a norm al range, and some perform ed work in ically acceptable method of inferring WOB . Other
a potentially fatiguing range.
factors explaining th e variance in WOB not included in
the mod el may be carbon dioxide minute production ,
oxygen consumption, lung and ches t wall compliances,
brings into qu estion the traditional practice of titrat ing airway resistance, physiologic dead space, and peak
PSV based on breathing pattern, as this may resu lt in inspiratory flow rate demand. A mor e comprehensive
inappropriate respiratory muscle workloads. This point prediction mod el, including the aforem entioned vari-
is illustrated in Figure 2. Whil e receiving PSV and ables, may increase the value of r2 and , thus , its pre-
breathing at a clinically acceptable range between 15 dictive ability. However , it is easier to measur e WOB
and 25 breaths/min, the WOB for some pati ents was dir ectly than to measure and regress a number of
eithe r zero, in a normal range (0.3 to 0.6 J/L 4,20,21), or physiologic variables. Predicting WOB based on a host
at an abnormally high rang e. Similar findin gs were of clinical variables, even assuming that all measure-
observed with th e breathin g pattern score (Fig 3). ments are obtained correctly, appears dicey.
Disuse respiratory muscle atrophy may result when the We do not dispute the fact that as pu lmonary me -
muscles are totally un loaded and performing no work chanics det eriorate, th e respiratory muscles are loaded,
for an exte nde d period. 22 Respiratory muscle fatigue is WOB increases , and the br eathing pattern changes.
likely to resu lt when the muscles are forced to perform Specifically, these changes are vagally mediated by af-
workloads at an abnormally high rang e for too long a ferent or sen sory fibers (load sensors) in the lungs and
period. With eithe r severe respiratory muscle atrophy respiratory tract. Three types of affere nt fibers that
or fatigue, pati ents may becom e dependent and not modulate the breathing pattern have been describ ed :
wean from ventilatory support. (1) slowly adapting receptors; (2) rapidly adapting re-
Our findin gs are in accordance with oth er studies . cep tors (also termed "deflation," "cough," or "irritant"
Silas et al23 reported that the index of rapid shallow receptors ), both of which are pulmonary stretch or
breathing (fNT) as described by Tobin et al24 was a mechanoreceptors; and (3) chemos ensitive or C-fibe r
poor predictor ofWOB because it predicted only 16% endings.26 Slowly adapting receptors are found in the
of the varianc e in WOB in patients before extu bation. bronchial smooth muscle fibers, rapidly adap ting re-
The ind ex of rapid shallow br eathing predicted only cep tors in the supe rficial layers of th e respiratory tract
20% of th e variance in WOB in our patients. Most of mucosa, and C-fibers in the airway epithelium .27 These
our patients were not ready for weaning from ventila- recep tors det ect changes in pulmonary mechanics and
tory support but were treated during the initial-to-late thoracic gas volume (functional residual capacity).27,28
ph ases of respiratory failure . Althou gh regard ed as an Fo llowing a decrease in lung compliance (increase
inference of the respiratory muscle workload, the in- in respiratory muscle load), an increase in discharge