G ()  mod.

liquiUV
Enzymatic UV test for the
quantitative determination of GOT (ASAT)
MINDRAY BS-300
NOTE:
RECOMMENDATION TO APPLY THE REAGENTS ON THE ANALYZER.
THE LABORATORY HAS TO VALIDATE THE SETTINGS IN ITS OWN RESPONSIBILITY.
HUMAN WILL ACCEPT NO RESPONSIBILITY FOR THE CORRECTNESS OF THIS APPLICATION AS WE
HAVE NO CONTROL ON SOFT- AND HARDWARE MODIFICATIONS OF THE RESPECTIVE ANALYZER.

Package Size
[REF] 12211 16 x 5 ml GOT (ASAT) IFCC mod. liquicolor Reagent kit
12011 10 x 10 ml GOT (ASAT) IFCC mod. liquicolor Reagent kit
12021 8 x 50 ml GOT (ASAT) IFCC mod. liquicolor Reagent kit
12031 4 x 250 ml GOT (ASAT) IFCC mod. liquicolor Reagent kit
13160 4 x 5 ml AUTOCAL
Recommended control serum
13511 6 x 5 ml HumaTrol N
13512 6 x 5 ml HumaTrol P
13951 6 x 5 ml SERODOS
13151 6 x 5 ml SERODOSplus
[IVD]
R1 = [BUF], ready for use
R2 = [SUB], ready for use
¼ For detailed information kindly refer to the most recent package insert.
Calibration
Calibration frequency
User defined; re-calibration is recommended
• after reagent lot change
• as required following quality control procedures.
Control frequency
It is recommended that 2 levels of control be assayed daily.

Parameters
No # Test GOT Method Kinetic Direction Decrease
Unit U/l Decimals 1 Prim. Wave 340 Sec. Wave 0
Sample vol 25 R1 Vol 200 R2 Vol 50 Line Limit 20

Antigen Response Full Name GOT


Incubation 25 Check Antigen Lower 0 Print Name #
Reaction 5 - 15 Substrate 0 Upper 0

R1 Blank Mix R Blank Linearity


Lower 0 Lower 0 Lower 0 Factor 0
Upper 0 Upper 0 Upper 400 Speed 0

Calibration
Request
Rule Two point linear
Replicates 2 Difference 0
Interval 0 Black Response 0/0
Sensitivity 0 Coeff Difference 0
Correlation 0 Non-Linear SD 0

# Data entry by the user